Fresenius 4008 HDF User manual

4.7 (7)

4008 HDF

Technical Manual

Fresenius Medical Care

4008 HDF

Technical Manual

4008 HDF Technical Manual

The Technical Manual contains all information necessary for performing maintenance and repair work.

The 4008 HDF option reflects the latest state of technology and complies with the requirements of EN 60601-1

Assembly, extension, adjustment, modification or repair may only be carried out by the manufacturer or persons authorized by him.

Any inquiries should be addressed to:

 

Head office:

Manufacturer:

Fresenius Medical Care

Fresenius Medical Care

Deutschland GmbH

Deutschland GmbH

Borkenberg 14

Plant Schweinfurt

D-61440 Oberursel/Ts., Germany

Hafenstraße 9

Tel.: 06171-60-0

D-97424 Schweinfurt, Germany

Telex: 410805 fres d

Tel.: 09721-678-0

Fax: 06171-251-58

Fax: 09721-678-200

Local Service:

 

 

 

 

 

Part No. 674 398 1

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

0-1

0-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

How to use the Technical Manual

Search and find

What?

Where?

 

Tables Contents

Pages 0-5 and at the beginning of each chapter

Purpose

This manual is intended for:

 

– first studies (to acquire basic knowledge)

 

– reference purposes (for start-up, maintenance and repair)

Organization

The manual is divided into 6 chapters:

 

0

General Notes

 

 

1

Description of machine functions

 

2

Technical safety checks

 

3

Calibration instructions

 

4

Circuit descriptions and circuit diagrams

 

5

Spare parts

 

Numbering system

Page number 1-3 is to be interpreted as: Chapter 1, Page 3

Qualification

This manual is intended for service technicians

 

– who are familiar with the current Operating Instructions (Operating Instruc-

 

 

tionseratining to this Technical Manual are available under part no. 674 406 1)

 

– who have the necessary background experience in mechanics, electrical and

 

 

medical engineering

 

 

– who have been authorized by the manufacturer to perform maintenance and

 

 

repair work

 

 

– who have access to the necessary auxiliary and measuring equipment

Restrictions

The study of this manual does not represent an alternative to the training courses

 

offered by the manufacturer.

Manual changes

Manual changes will be released as new editions, supplement sheets or product

 

information.

 

Note:

Modifications relating to circuit diagrams and component layouts (SP/BP) do not necessarily involve a change of the footer (edition).

Refer to the index field of the respective circuit diagram or component layout for the respective state of these diagrams. The identification on the P.C.B. permits the user/technician to verify if the circuit diagram / component layout matches the P.C.B. actually installed.

In general, this manual is subject to modification.

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

0-3

Representation

New circuit symbols are used in the circuit diagrams. Potential data given in the

 

circuit diagrams and setting instructions refer to the respective earth.

Component marking

For example:

 

 

24 means ground for 24 V voltage.

 

 

Example:

75

 

 

in circuit diagrams

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1R5

This refers to a resistor with a position number 75 with a resistance of 1.5 Ohm.

The decimal point is replaced by a unit symbol (to reduce the possibility of errors).

 

Resistors:

Capacitors:

 

 

R1:

0.1 Ω

µ 1:

0.1 µ

F

 

1R5:

1.5 Ω

1µ 5:

1.5 µ

F

 

1K5:

1.5 kΩ

1000µ :

1000 µ F

Note:

When repairing or exchanging replacement parts make sure to take the applica-

 

ble ESD precautions (e.g. EN 100 015-1).

 

During repair/troubleshooting in the hydraulic unit, protect the components from

 

dialysate.

 

 

 

Technical data: The technical data for the 4008 HDF option is to be found in Chapter 1 of the Operating Instructions.

0-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Table of contents

Section

 

Page

1

Description of machine functions ...............................................................................

1-

1.1

Description ......................................................................................................................

1-3

1.2

Description of extended T1 test ......................................................................................

1-8

1.3

Error messages ...............................................................................................................

1-10

2

Technical safety checks ...............................................................................................

2-

3

Calibration instructions ...............................................................................................

3-

3.0

General information on the calibration instructions .......................................................

3-3

3.1

Calibrating the UF2 pump ...............................................................................................

3-5

3.2

VDE inspections..............................................................................................................

3-7

3.3

Calibrating the 4008 HDF scale ......................................................................................

3-8

3.4

Calibrating the substituate sensor ..................................................................................

3-13

3.5

Calibrating the HDF blood pump ....................................................................................

3-15

3.6

Repair instructions ..........................................................................................................

3-17

4

Circuit descriptions and circuit diagrams ..................................................................

4-

4.1

LP 625 display board ......................................................................................................

4-3

4.2

LP 754 control board HDF ..............................................................................................

4-9

4.3

LP 760 HDF motor control ..............................................................................................

4-19

4.4

LP 761 HDF drive keyboard ...........................................................................................

4-27

4.5

LP 762 scale comm ........................................................................................................

4-33

4.6

LP 763 SSE serial interface extender .............................................................................

4-39

5

Spare parts ....................................................................................................................

5-

5.0

How to use the spare parts catalog ................................................................................

5-3

5.1

P.C.B.s ............................................................................................................................

5-4

5.2

Scales .............................................................................................................................

5-6

5.3

Substituate lift .................................................................................................................

5-8

5.4

UF2 pump / blood pump (HDF) / valve V 126 .................................................................

5-10

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

0-5

0-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Table of contents

1 Description of machine functions

Section

 

Page

1.1

Description ....................................................................................................................

1-3

1.1.1

Components ....................................................................................................................

1-3

1.1.2

Description ......................................................................................................................

1-3

1.1.3

Component tests .............................................................................................................

1-3

 

Fig.: Block diagram .........................................................................................................

1-4

 

Fig.: Flow diagram ..........................................................................................................

1-5

 

Fig.: Flow diagram ..........................................................................................................

1-6

1.2

Description of extended T1 test / error messages in T1 test ...................................

1-8

1.2.1

Test UF-Function ............................................................................................................

1-8

1.3

Error messages during treatment and the HDF test ................................................

1-10

1.4

Error messages, substituate pump .............................................................................

1-13

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-1

1-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1.1Description

1.1.1Components

The components of the 4008 HDF option are permanently connected to the hemodialysis machine. The 4008 HDF option comprises the following components:

Substituate lift

Scales (weighing range < 20 kg) with data interface and taring facility

Substituate pump with bidirectional data interface

UF2 pump

1.1.2Description

The scales determine the actual weight of the substituate reserve and signals this to the hemodialysis machine. This calculates the substituate rate depending on the set treatment time.

A substituate sensor mounted on the substituate lift recognises whether:

Line set is fitted without substituate or

Line set is fitted with substituate.

The substituate pump delivers the substituate solution to the venous bubble catcher. The filtrate from the dialysis fluid circuit is removed by the UF2 pump.

1.1.3Component tests

The scales are checked by the operator by means of a plausibility test before treatment. The substituate sensor is tested while the 4008 HDF is filling.

The 4008 HDF test tests:

The substituate pump

The UF2 pump (electrically and hydraulically)

(If the 4008 HDF test is not carried out then the connections are to be tested but once). The UF2 pump is tested cyclically during treatment.

The electric control system of the UF2 pump is checked continuously.

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-3

Fig.: Block diagram

Scales

Serial communication

W

CPU 1

Operating system

Safety system

Serial communication

CPU 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Serial communication

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

UF2 pump

 

 

Substituate

 

 

 

 

 

TMP

 

 

 

pump

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Fig.: Flow diagram

XHDF2–4

Scales

Substituate bag

Substituate sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

X348v/4–5

andtaringline

 

X634L/

A–C23

 

 

X634L/

A–C24

X634R/

C11

 

 

 

 

 

 

 

 

serialinterface

serialinterface

LP634

Output

board

 

 

X634R/

A24

 

 

 

 

 

 

 

 

 

X633L/

C14

X633L/

C23

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LP633

Input

board

interface

extender

X633R/

C28–30

pressuredialy.Acknowledgem.,

pumpUFControl,

pumpUF2Control,

UF2Acknowledgement,pump

UFAcknowledgement,pump

pumpUF2Control,

pumpUFControl,

dialysateAcknowledgement,pressure

 

 

763LP

SSE

Serial

 

 

 

X633L/

B6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

serial LP631

interface

X632/

A29

 

 

X632/

C27

 

 

 

 

 

 

 

Balancingchamber

 

 

 

CPU1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LP632

CPU2

 

 

 

X632/

B24

 

 

 

 

 

 

 

 

 

X632/

C7

 

 

X632/

A7

 

 

 

 

 

 

 

 

 

 

 

Substituate pump

Venousbubblecatcher

Airdetector

Venousline clamp

Patient

Arterial bloodpump

 

 

 

Dialyzer

 

 

V24

V26

V24b

Dialysate

pressure

transducer

V126

Flow pump

UF pump

UF2 pump

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-5

Fresenius 4008 HDF User manual

Fig.: Flow diagram

1-6 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Legend

 

 

2

Temperature sensor

77

Heat exchanger

3

Temperature sensor

78

Relief valve

5

Float switch

84

Disinfection valve

6

Level sensor

85

Disinfectant conncector

7

Conductivity measuring cell

86

Recirculation valve

8

Blood leak detector

87

Discharge valve

9

Pressure transducer

88

Multifunction block

10

Reed contact for concentrate

88a

Degassing chamber

12

Reed contact for bicarbonate

88b

Secondary air separator

20

UF2 pump

88c

Primary air separator

21

Flow pump

89

Degassing orifice

22

UF pump

90a

Rinse chamber concentrate

23

Concentrate pump

90b

Rinse chamber bicarbonate

24

Dialyzer valve 1

91

Rinse valve

24b

Dialyzer valve 2

92

Vent valve

25

Bicarbonate pump

94

Concentrate suction tube

26

Bypass valve

95

Bicarbonate suction tube

29

Degassing pump

97

Ventilation pump

30

Drain valve

99

Rinse valve

31

Balancing chamber valve 1

100

Rinse valve

32

Balancing chamber valve 2

102

Concentrate valve in central delivery system

33

Balancing chamber valve 3

104

Bicarbonate valve in central delivery system

34

Balancing chamber valve 4

109

Temperature sensor

35

Balancing chamber valve 5

111

Hydrophobic filter

36

Balancing chamber valve 6

112

Vent valve

37

Balancing chamber valve 7

114

Dialysate filter

38

Balancing chamber valve 8

115

Disinfection valve sensor

41

Water inlet valve

116

Sampleing valve

43

Fill valve

117

Check valve (concentrate)

54

Heater rod

118

Check valve (bicarbonate)

61

Pressure reducing valve

119

Filter (concentrate)

63

Filter/water inlet

120

Filter (bicarbonate)

65

Loading pressure valve

121

Concentrate connector in central delivery system

66

Heater block

122

Bicarbonate connector in central delivery system

66a

Water inflow chamber

123

Pressure switch for V 102

66b

Heater rod chamber

124

Pressure switch for V 104

66c

Float chamber

125

Adapter plate

68

Balancing chamber

126

4008 HDF vent valve

71

Filter/concentrate

148

Filter/rinse valve

72

Filter/bicarbonate

149

Filter/rinse valve

73

Filter/external dialysate

150

Filter

74

Filter/UF

151

Orifice

75

External flow indicator

152

4008 HDF hydrophobic filter

76

Filter/fill valve

 

 

Hydraulic measurement points

AReduced water inlet pressure

BBalancing chamber loading pressure

CFlow pump pressure

DDegassing pump pressure

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-7

1.2Description of extended T1 test / error messages in the T1 test

Possible error messages displayed in the T1 test are listed in the Technical Manual for the hemodialysis machine. If the HDF option is installed and activated the following error codes can also appear:

1.2.1Test UF-Function Test description:

CPU 1 starts up the UF2 pump at a defined rate.

CPU 2 controls the hydraulic and electrical function of the UF2 pump. CPU 2 blocks the control line of the UF2 pump and checks for standstill. UF2 counter check.

Figure:

1-8 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Error description:

 

 

 

 

Error message

Description

 

 

F11 UF-Function

The interval between the strokes of the UF2 pump was less than

 

220 ms. Correct delivery of the required volume is not safeguard-

 

ed because the stroke return travel is too short.

 

– A pump rate which was too high was signalled by the CPU 1.

F12 UF-Function

The pulse time for the UF2 pump is less than 180 ms. Correct

 

delivery of the required volume is not safeguarded because the

 

discharge time is too short.

 

 

 

– Monoflop on LP 634 defect (IC 42/R65/C45).

F 13 UF-Function

The pulse for the UF2 pump is longer than 500 ms. A maximum

 

rate of 5000 ml/h is not possible.

 

 

 

– Monoflop on LP 634 defect (IC 42/R65/C45).

F14 UF-Function

UF2 pump not active during the test (4 s).

 

 

 

Acknowledgement (UF_P2, X637/B26) →

X632/C7 no LOW

 

 

pulse

 

 

 

Control line (UF_P2, X634L/A–C24) →

X637/B26 no LOW

 

 

pulse

 

 

F15 UF-Function

The UF2 pump cannot be stopped by the CPU2.

 

Control line (UF_P_EN, X632/C28) →

X634R/A22 no 5 V.

 

– Reset input to IC 42/pin 13 on LP 634 defect.

F16 UF-Function

The UF pump acknowledgement from CPU1 is faulty.

 

Acknowledgement (UF_P2, X637/B26) →

X633/C23 no LOW

 

 

pulse

 

 

F17 UF-Function

The change in pressure after a UF2 pump stroke less than

 

20 mmHg.

 

 

 

– UF2 pump mechanically defect.

 

 

 

Control line (UF_P2_CRTL, X632/B24) → X634R/C11 no

 

 

HIGH pulse.

 

 

F20 UF-Function

Difference in pressure between UF pump stroke and UF2 pump

 

stroke more than 20 %.

 

 

 

– Stroke rate for UF pump or UF2 pump not set correctly.

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-9

1.3Error messages during treatment and the HDF test

Message on the

Cause

Possible error elimination

display

 

 

 

 

 

F321 HDF-failure

The UF pump or UF2 pump

Check level sensor (6) (de-

 

test will not be started as long

fect).

 

as the level sensor (6) detects

 

 

air. The level sensor (6) will

 

 

detect air one minute after the

 

 

4008 HDF test has been se-

 

 

lected

 

 

 

 

F322 HDF-failure

The UF pump or UF2 pump

Check V126 (does not open,

 

has carried out more than 50

venous bubble catcher col-

 

strokes and still no air has

lapses), check V43 (whether

 

been detected.

leaking or open).

 

 

Check UF pump or UF2 pump,

 

 

respectively (mech. defect).

 

 

 

F323 HDF-failure

The substituate pump runs at a

Check substituate tubing

 

rate of 10 ml/min. to raise the

(clamped off, not connected to

 

fluid level by a defined amount

the venous bubble catcher).

 

in the secondary air separator

 

 

(the fluid will just be detected).

 

 

Air will still be detected after 2

 

 

minutes.

 

 

 

 

F324 HDF-failure

Since fluid has been recog-

Check substituate tubing

 

nized in the secondary air sep-

(clamped off).

 

arator the substituate pump

 

 

starts at a rate of 10 ml/min.

 

 

until a change in weight follows

 

 

on the scales (i.e. a jump in

 

 

grammes). There was no dif-

 

 

ference in weight established

 

 

after one minute.

 

 

 

 

F325 HDF-failure

If the expected change in

Check bag (fluctuations in air

 

weight (i.e. a jump in

draft)

 

grammes) is higher or lower

Check scales (for drifting).

 

than –1 g, the respective UF

 

 

pump test will be repeated.

 

 

The UF pump test was repeat-

 

 

ed more than 5 times.

 

 

 

 

1-10 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Message on the

Cause

Possible error elimination

display

 

 

 

 

 

F326 UF-failure

The fluid level in the second-

Check substituate hose

 

ary air separator is raised

(clamped off, not connected

 

again by the substituate pump

with the venous bubble

 

after 100 strokes of the UF

catcher).

 

pump or UF2 pump, respec-

 

 

tively, until fluid is detected.

 

 

Plus run-on time at a rate of 10

 

 

ml/min. for the gramme jump.

 

 

Air will still be detected after a

 

 

specific period of time (de-

 

 

pending on the delivery rate of

 

 

the substituate pump).

 

 

 

 

UF1 volume-Error

One UF pump did not pass the

Check UF pump or UF2 pump,

or

test. The filling volume for the

respectively (not calibrated,

UF2 volume-Error

secondary air separator is not

mechanical defect).

 

within the given tolerance of

 

 

100 ml; ± 5 ml. Should the test

 

 

give a reading of over 105 ml

 

 

the cause could also be air

 

 

taken in the flow from a badly

 

 

vented dialysator.

 

 

 

 

F327 UF-failure

Interval between two strokes

Check CPU 1 (defect).

 

of the UF pump less than

 

 

220 ms.

 

 

 

 

F328 UF-failure

Pulse time of a stroke of UF

Check LP 634 (regulating

 

pump less than 180 ml.

monoflop defect).

 

 

 

F329 UF-failure

Pulse time of a stroke of UF

Check LP 634 (regulating

 

pump more than 500 ms.

monoflop defect).

 

 

 

F330 UF-failure

Starting time for the UF pump

Check LP 634 (regulating end

 

more than 10 seconds.

stage defect).

 

 

 

F331 UF-failure

Difference between desired or

Check CPU 1 / CPU 2

 

actual delivery rate of the UF

(communication problems).

 

pump greater than + 10%.

 

 

 

 

F332 UF-failure

UF pump stops longer than the

Check LP 634 (regulating end

 

maximum period time.

stage defect).

 

 

Check UF pump

 

 

(interruption, control).

 

 

Check CPU 1 / CPU 2

 

 

(communication problems)

 

 

 

F333 UF-failure

Despite switched off ultrafiltra-

Check CPU 1 / CPU 2

 

tion the change in delivery rate

(communication problems)

 

of the UF pump is greater than

 

 

10 ml.

 

 

 

 

F334 UF-failure

Interval between two strokes

Check CPU 1 (defect).

 

of the UF2 pump less than

 

 

220ms.

 

 

 

 

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-11

Message on the

Cause

Possible error elimination

display

 

 

 

 

 

 

F335

UF-failure

Pulse time of a stroke of the

Check LP 634 (regulating

 

 

UF2 pump less than 180 ms.

monoflop defect).

 

 

 

 

F336

UF-failure

Pulse time of a stroke of the

Check LP 634 (regulating

 

 

UF2 pump more than 500 ms.

monoflop defect).

 

 

 

 

F337

UF-failure

Reaction time of the UF2

Check LP 634 (regulating end

 

 

pump more than 10 seconds.

stage defect).

 

 

 

 

F338

UF-failure

Difference in desired or actual

Check CPU 1 / CPU 2

 

 

delivery rate of UF2 pump

(communication problems).

 

 

greater than ± 10 %.

 

 

 

 

 

F339

UF-failure

UF2 pump stops longer than

Check LP 634 (regulating end

 

 

the maximum period time.

stage defect).

 

 

 

Check UF pump (interruption,

 

 

 

control).

 

 

 

Check CPU 1 / CPU 2

 

 

 

(communication problems).

 

 

 

 

F340

UF-failure

Despite switched off 4008HDF

Check CPU 1 / CPU 2

 

 

change in delivery rate of the

(communication problems).

 

 

UF2 pump greater than 10 ml.

 

 

 

 

 

F341

UF-failure

Failure of UF pump.

Check UF pump

 

 

 

(spring, screen)

 

 

 

 

F342

UF-failure

Failure of UF2 pump.

Check UF2 pump

 

 

 

(spring, screen)

 

 

 

 

F343

UF-failure

Volume difference between UF

Check UF pump and UF2

 

 

pump and UF2 pump.

pump (delivery volume)

 

 

 

 

F344

HDF-failure

Balance failure recognized by

Check monitor / scales

 

 

CPU2 greater than ± 500 ml.

(communication problems).

 

 

 

 

F345

HDF-failure

Bolus exceeded by more than

Check monitor / scales

 

 

+ 20 ml, detected by CPU2.

(communication problems).

 

 

 

 

F346

HDF-failure

CPU2 failed to perform the cy-

Check V24B and V43 feed-

 

 

clic UF pump test within 5 min-

back lines.

 

 

utes.

Check sense of rotation of as-

 

 

 

piration pump.

 

 

 

Check hydraulics unit for

 

 

 

leaks.

 

 

 

 

F348

HDF-failure

Weight change on the scales

Check scales

 

 

during the weighing test great-

Check substituate bag

 

 

er than 2g.

 

 

 

 

 

F349

HDF-failure

The delivery rate correction

Check substituate bag

 

 

factor could still not be deter-

(weight fluctuations)

 

 

mined after 20 minutes treat-

Check HDF control system

 

 

ment time.

(greater fluctuations)

 

 

 

 

1-12 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1.4Error messages, substituate pump

Code

Description of failure

 

 

E.01

Line diameter outside the permitted range

 

 

E.02

Non-defined hexaswitch position

 

 

E.03

Venous pressure transducer not balanced

 

 

E.04

Failure, running time monitoring system, SN operation

 

 

E.05

SN stroke volume outside the permitted range

 

 

E.06

The SN pressure thresholds outside the value range of the AD converter

 

 

E.07

Not defined

 

 

E.08

Failure in the AD conversion

 

 

E.09

Not defined

 

 

E.10

Not defined

 

 

E.11

Not defined

 

 

E.12

Failure, speed monitoring (hall sensor)

 

 

E.13

Failure, monitoring system, current sensing resistors

 

 

E.14

Failure, monitoring system, current sensing resistors

 

 

E.15

Failure, speed monitoring system

 

 

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

1-13

1-14 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

2 Technical safety checks

General Notes

This chapter includes all necessary technical safety checks (TSC).

These inspections must be carried out every 12 months.

The technical safety inspections stipulated for the hemodialysis machine must be carried out in addition to these technical saftey checks.

The technical safety checks are to be recorded in the equipment log.

A technical safety checks report is to be found at the page 2-3.

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

2-1

2-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

 

 

 

Technical safety checks report

Seite 1/1

 

 

 

 

 

 

 

Manufacturer: .............................

Date: .............................

Machine: .............................

Technician: .............................

Operating hours: .............................

 

Extent of checks and time limits for technical safety checks.

TIME LIMIT: every 12 months

The following checks and inspections have to be carried out on the machine at least every 12 months by persons who are capable of carrying out such technical safety checks efficiently as a result of their training, knowledge and experience gained in practice and are not subject to instructions as to their checking activity.

(See also BMA Bulletin dated 02.04.1987)

TSC

WA

No.

Description

Desired value/function

OK

 

 

 

 

 

 

 

 

1

UF2 pump

 

 

 

 

 

 

 

TSC

 

1.1

Delivery volume (1 ml/stroke)

60 strokes = 60 ml ± 1 %

 

 

 

 

 

 

 

 

 

2

VDE inspections (values according to EN 60601-1)

 

 

 

 

 

 

TSC

 

2.1

Protective ground resistance

max. 0.3 Ω

 

 

 

 

 

 

 

TSC

 

2.2

Summarized leakage current

Must fulfill both conditions:

 

 

 

 

 

1. No more than 1.5 times the “summari-

 

 

 

 

 

zed value first measured”

 

 

 

 

 

(“Summarized value first measured”:

 

 

 

 

 

Refer to the machine card enclosed with

 

 

 

 

 

the machine.)

 

 

 

 

 

2. No more than 1 mA

 

 

 

 

 

 

 

The technical safety checks are to be recorded in the equipment log and the results of the checks documented.

If the machines are not safe in function and/or operation they are to be repaired or the operator is to be informed as to the dangers involved when using the machine in its given state.

The correct completion of the listed work and the correctness of statements made are hereby certified.

Signature, technician:

Signature, customer:

........................... , date .............................

................................ , date .............................

....................................................................

.........................................................................

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

2-3

2-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Table of contents

3 Calibration instructions

Section

 

Page

3.0

General information on the calibration instructions ................................................

3-3

3.1

Calibrating the UF2 pump ............................................................................................

3-5

3.2

VDE inspections ............................................................................................................

3-7

3.3

Calibrating the 4008 HDF scales .................................................................................

3-8

3.4

Calibrating the substituate sensor ..............................................................................

3-13

3.5

Calibrating the HDF blood pump .................................................................................

3-15

3.5.1

Pressure transducer .......................................................................................................

3-15

3.5.2

Checking SN switchover points ......................................................................................

3-16

3.6

Repair instructions .......................................................................................................

3-17

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

3-1

3-2 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

3.0General information on the calibration instructions

Measuring instruments:

The same measuring instruments as used for the hemodialysis machines are to be employed. Required in addition is a calibration weight PT6 – 5000 g.

Fig.: P.C.B. overview 4008 HDF

LP 762

LP 763 LP 761

Fresenius Medical Care

Fresenius Medical Care 4008

 

 

 

 

 

 

 

 

Best.

Auswahl

 

 

 

I/O

 

Arterieller

 

Venöser

 

 

 

 

 

 

 

 

Luftdetektor

 

 

 

TMP

Ultrafiltration

Leitfähigkeit

 

Fluß

 

Druck

 

Druck

 

 

 

 

 

 

 

 

 

UF Menge

ml

 

 

 

ml/min

Spülen

 

 

 

 

 

 

 

Reset

 

 

 

900

 

 

 

 

 

 

 

 

 

 

mS/cm

 

700

 

kPa

mmHg

kPa

mmHg

kPa

mmHg

 

 

 

(25° C)

 

500

Heiß reinigen

 

280

 

500

 

500

 

 

 

15.5

 

300

Blutleck

30

200

60

 

60

 

UF Rate

ml/h

 

 

 

 

 

 

400

 

400

 

 

 

15

 

 

Desinfektion

20

100

50

50

 

 

 

 

 

 

10

40

300

40

300

 

 

 

14.5

I/O

Set

 

0

0

 

 

 

 

 

Single

30

200

30

200

 

 

 

14

 

 

–10

–100

 

 

 

 

 

 

 

Needle

 

20

100

20

100

UF Ziel

ml

 

13.5

 

 

Ü berbrücken

–20

–200

10

10

 

 

 

 

 

Test

–30

 

0

0

0

0

 

 

 

13

Temperatur

 

–40

–300

 

 

 

 

 

 

 

 

 

° C

Vorbereiten

 

 

 

 

 

 

UF Restzeit h:min

Konz.

Bic.

 

39

 

 

 

 

 

 

 

 

 

 

37

Dialyse Start

 

 

 

 

 

 

 

 

 

 

 

35

 

 

 

 

 

 

 

UF

 

Variation

 

 

 

Alarm

 

 

 

 

 

 

I/O

Prog.

I/O

Prog.

 

Set

Ton Aus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pven.

 

 

 

 

 

 

Bolus

 

 

 

 

 

 

 

 

 

Start

 

 

Rate

 

 

 

 

 

 

 

 

 

Stop

 

 

 

 

 

 

 

 

 

 

 

Start

Stop

LP 625

Start

Stop

LP 760

LP 754

Fresenius Medical Care 4008 HDF 3/11.97 (TM)

3-3

Fig.: DIP switch 4008 HDF

SW1

ON

1 2 3 4 5 6 7 8

OFF

LP 632 CPU 2

SW 1, DIP switch 8:

ON: 4008 HDF test selectable (required/not required) OFF: 4008 HDF test carried out automatically

3-4 Fresenius Medical Care 4008 HDF 3/11.97 (TM)

Loading...
+ 74 hidden pages