Fresenius 4008 Technical manual

4.5 (6)

4008 E / 4008 B / 4008 H / 4008 S

Hemodialysis system

Technical Manual

Edition: 5/03.09 Part no.: M40 618 1

Software 4.5/5.3 and higher

Fresenius Medical Care

Important information on the Technical Manual

How to use the Technical Manual

Identification

The document can be identified by the following information on the title page and

 

on the labels, if any:

 

– Edition of the technical document

 

– Part number of the technical document

Page identification

The page identification 1-3, for example, refers to: chapter 1, page 3.

Editorial information

The editorial information 1/01.05, for example, refers to the 1st edition, January

 

2005.

Changes

Changes to the Technical Manual will be released as new editions or supple-

 

ments. In general: This manual is subject to change without notice.

Significance of the

Explanation of the Caution and Note symbols used:

safety precautions

Caution

Advises the operator against certain procedures or actions that could cause damage to the equipment or may have adverse effects on operators and patients.

Note

Informs the operator that if the steps are not followed as described, a specific function will be executed incorrectly or will not be executed at all, or will not produce the desired effect.

Fresenius Medical Care 4008 5/03.09 (TM)

0-1

Important information on the system

Technician’s qualification

Purpose

This Technical Manual is intended for service technicians and is to be used for

 

first studies (to acquire a basic knowledge) and for reference purposes (for TSC,

 

Maintenance and repair). The Technical Manual, however, does not replace the

 

training courses offered by the manufacturer.

Requirements

Knowledge of the current Operating Instructions for the respective system.

 

Background experience in mechanics, electrical and medical engineering.

Precautions for working on the system

Authorized persons

Assembly, extensions, adjustments, modifications or repairs may only be carried

 

out by the manufacturer or persons authorized by him.

Test equipment and

The activities described in this technical document require the availability of the

accessories

necessary technical test equipment and accessories.

Specifications

For the specifications of the respective system, refer to the current Operating

 

Instructions. Observe the information on the specifications.

Precautions

Before turning power on, repair any visible damage.

 

Prior to opening the system and when working on the open system, the following

 

precautions have to be observed:

 

– Protect the components against ingress of fluids.

 

– Do not touch live parts (e.g. connectors of the power cable or heater).

 

– Disconnect and connect all jacks, connectors and components only when the

 

system is turned off.

ESD precautions

When repairing the system and replacing spare parts, observe the applicable

 

ESD precautions.

Hygienic measures

The system and the consumables are generally considered to be contaminated

 

and must therefore be sufficiently disinfected by the responsible organization as

 

specified by the manufacturer.

0-2 Fresenius Medical Care 4008 5/03.09 (TM)

Addresses

Please address any inquiries to:

Manufacturer

Fresenius Medical Care AG & Co. KGaA

 

D-61346 Bad Homburg

 

+49 (0)6172-609-0

 

www.fmc-ag.com

Service

Fresenius Medical Care

Central Europe

Deutschland GmbH

 

Geschäftsbereich Zentraleuropa

 

Kundendienst / Servicecenter

 

Steinmühlstraße 24

 

61352 Bad Homburg

 

Germany

 

Phone: +49 (0)6172-609-7100

 

Fax: +49 (0)6172-609-7102

 

E-mail: ServicecenterD@fmc-ag.com

International

Fresenius Medical Care

service

Deutschland GmbH

 

Service Support International

 

Hafenstraße 9

 

D-97424 Schweinfurt

 

Germany

 

Phone: +49 (0)9721-678-333 (hotline)

 

Fax: +49 (0)9721-678-130

Local service

 

 

 

 

 

 

 

Fresenius Medical Care 4008 5/03.09 (TM)

0-3

0-4 Fresenius Medical Care 4008 5/03.09 (TM)

Table of contents

Section

 

Page

1

Description of system functions and malfunctions .................................................

1-

1.1

Description of the T1 test ...............................................................................................

1-3

1.2

Functional description of the modules............................................................................

1-78

1.3

Functional description of the hydraulic unit ....................................................................

1-83

2

Technical Safety Checks / Maintenance ....................................................................

2-

2.1Technical Safety Checks and Maintenance

 

for 4008 hemodialysis systems and options ..................................................................

2-3

2.2

TSC report ......................................................................................................................

2-25

3

Adjustment instructions ..............................................................................................

3-

3.1

Overview of the DIP switches in the 4008......................................................................

3-9

3.2

Calibration mode ............................................................................................................

3-13

3.3

Hydraulics .......................................................................................................................

3-15

3.4

Air detector .....................................................................................................................

3-31

4

Calibration program .....................................................................................................

4-

5

Diagnostics program....................................................................................................

5-

5.1

General notes .................................................................................................................

5-3

5.2

Menu structure ................................................................................................................

5-5

5.3

Reading the analog inputs of CPU I ...............................................................................

5-7

5.4

Reading the analog inputs of CPU II ..............................................................................

5-9

5.5

Reading the digital inputs of CPU I ................................................................................

5-10

5.6

Reading the digital inputs of CPU II ...............................................................................

5-15

5.7

Writing the analog outputs of CPU I ...............................................................................

5-19

5.8

Writing the analog outputs of CPU II ..............................................................................

5-20

5.9

Writing the digital outputs of CPU I ................................................................................

5-21

5.10

Writing the digital outputs of CPU II ...............................................................................

5-27

5.11

Writing/Reading the digital outputs of CPU I ..................................................................

5-30

5.12

ONLINEplus™ module .....................................................................................................

5-31

5.13

HPU ................................................................................................................................

5-32

6

Setup menu ...................................................................................................................

6-

6.1

Overview Setup menu settings.......................................................................................

6-3

6.2

Overview .........................................................................................................................

6-7

6.3

Main menu 4008 E/B ......................................................................................................

6-9

6.4

Main menu 4008 H/S ......................................................................................................

6-37

7

Miscellaneous ...............................................................................................................

7-

Fresenius Medical Care 4008 5/03.09 (TM)

0-5

Section

 

Page

8

Circuit diagrams and circuit descriptions .................................................................

8-

8.1

Block diagram 4008 ........................................................................................................

8-3

8.2

Block diagram of voltage supply ....................................................................................

8-4

8.3

Block diagram of screen 4008 H/S .................................................................................

8-5

8.4

Connection layout diagram .............................................................................................

8-6

8.5

P.C.B. LP 450-2 Level detector control (LD) ..................................................................

8-9

8.6

P.C.B. LP 493 Blood leak detector .................................................................................

8-11

8.7

P.C.B. LP 624 Control board (BP) ..................................................................................

8-12

8.8

P.C.B. LP 630 Motherboard ...........................................................................................

8-13

8.9

P.C.B. LP 631 CPU 1 .....................................................................................................

8-14

8.10

P.C.B. LP 632 CPU 2 .....................................................................................................

8-16

8.11

P.C.B. LP 633 Input board .............................................................................................

8-18

8.12

P.C.B. LP 634 Output board ...........................................................................................

8-20

8.13

P.C.B. LP 635 Display board ..........................................................................................

8-22

8.14

P.C.B. LP 636 External connectors ................................................................................

8-24

8.15

P.C.B. LP 950 Control board (HEP) ...............................................................................

8-25

8.16

P.C.B. LP 644-4 Display board (HEP)............................................................................

8-26

8.17

P.C.B. LP 645 Position sensor membrane pump...........................................................

8-27

8.18

P.C.B. LP 649-2 Display board (4008 B/S) ....................................................................

8-28

8.19

P.C.B. LP 742 Interference filter.....................................................................................

8-30

8.20

P.C.B. LP 748 Display board (BP) .................................................................................

8-31

8.21

P.C.B. LP 763 Multi interface board (COMMCO III) ......................................................

8-32

8.22

P.C.B. LP 922 Display board (4008 S) ...........................................................................

8-33

8.23

P.C.B. LP 923 Traffic light (4008 H/S) ...........................................................................

8-34

8.24

P.C.B. LP 924 Display board (4008 H) ..........................................................................

8-35

8.25

P.C.B. LP 941 Hydraulics processor ..............................................................................

8-36

8.26

Heater board (power supply unit 4008) ..........................................................................

8-37

8.27

Power board (power supply unit 4008) ..........................................................................

8-38

0-6 Fresenius Medical Care 4008 5/03.09 (TM)

Table of contents

1 Description of system functions and malfunctions

Section

 

Page

1.1

Description of the T1 test ............................................................................................

1-3

1.1.1

T1 test flow diagram, serial program steps ....................................................................

1-3

1.1.2

T1 test flow diagram, parallel program steps .................................................................

1-5

1.1.3

Description of the T1 test incl. error messages ..............................................................

1-7

1.1.4

Description of system errors during the cleaning programs ..........................................

1-53

1.1.5

Error messages after turning power on ..........................................................................

1-70

1.1.6

Error messages during dialysis ......................................................................................

1-71

1.2

Functional description of the modules ......................................................................

1-78

1.2.1

Blood pump (arterial) ......................................................................................................

1-78

1.2.2

Blood pump (Single-Needle), optional ...........................................................................

1-79

1.2.3

Heparin pump .................................................................................................................

1-80

1.2.4

Air detector .....................................................................................................................

1-82

1.3

Functional description of the hydraulic unit .............................................................

1-83

 

Fig.: Flow diagram ..........................................................................................................

1-83

1.3.1

Description of the hydraulic unit .....................................................................................

1-85

1.3.2

Theory of operation of the balancing chamber ..............................................................

1-87

1.3.3

Central delivery system option .......................................................................................

1-91

1.3.4

Program sequences during the cleaning programs .......................................................

1-92

 

Fig.: Flow chart of cleaning programs – overview ..........................................................

1-92

Fresenius Medical Care 4008 5/03.09 (TM)

1-1

1-2 Fresenius Medical Care 4008 5/03.09 (TM)

1.1Description of the T1 test

1.1.1T1 test flow diagram, serial program steps

MODULE

 

 

TEST

 

 

T1

 

 

ACCUMULATOR

 

 

START T1 TEST

 

 

TEST OK

no

STORAGE

 

 

 

 

 

 

?

 

ERROR NUMBER

 

 

 

yes

 

 

TEST

 

 

TEST

 

 

BYPASS

 

 

BLOOD LEAK DETECTOR

 

 

TEST OK

no

STORAGE

TEST OK

no

STORAGE

?

 

ERROR NUMBER

?

 

ERROR NUMBER

yes

 

 

yes

 

 

TEST

 

 

TEST

 

 

OPT. DETECTOR

 

 

TEMPERATURE

 

 

TEST OK

no

STORAGE

TEST OK

no

STORAGE

?

 

ERROR NUMBER

?

 

ERROR NUMBER

yes

 

 

yes

 

 

TEST

 

 

NEG. PRESSURE

 

 

BLOOD SYSTEM

 

 

HOLDING TEST

 

 

TEST OK

no

STORAGE

TEST OK

no

STORAGE

?

 

ERROR NUMBER

?

 

ERROR NUMBER

yes

 

 

yes

 

 

TEST

 

 

POS. PRESSURE

 

 

VENOUS PRESSURE

 

 

HOLDING TEST

 

 

TEST OK

no

STORAGE

TEST OK

no

STORAGE

?

 

ERROR NUMBER

?

 

ERROR NUMBER

yes

 

 

yes

 

 

TEST

 

 

TEST

 

 

AIR DETECTOR

 

 

UF-FUNCTION

 

 

TEST OK

no

STORAGE

TEST OK

no

STORAGE

?

 

ERROR NUMBER

?

 

ERROR NUMBER

yes

 

 

yes

 

 

TEST

 

 

TEST

 

 

DISPLAY

 

 

CONDUCTIVITY

 

 

TEST OK

no

STORAGE

TEST OK

no

STORAGE

?

 

ERROR NUMBER

?

 

ERROR NUMBER

yes

 

 

yes

 

 

TEST

 

 

Basic hydraulics

 

Advanced hydraulics

ARTERIAL PRESSURE

 

 

 

 

 

TEST

 

TEST DIASAFE PLUS /

 

 

 

 

 

 

 

DIASAFE/HDF FILTER

 

ONLINE PLUS /

 

 

 

 

 

HPU TEST

TEST OK

no

STORAGE

 

 

 

?

 

ERROR NUMBER

 

 

 

yes

 

 

TEST OK

no

STORAGE

 

 

?

 

ERROR NUMBER

 

 

 

 

 

 

 

yes

 

 

 

 

 

1

 

 

Fresenius Medical Care 4008 5/03.09 (TM)

1-3

1

 

T1 TEST

yes

UNSUCCESSFUL

DIALYSIS START KEY

 

no

INCORRECT

TEST STEP

RETURN

TEST OK

no

ERROR

?

 

DISPLAY

 

yes

 

 

 

 

 

FURTHER

yes

NEXT INCORRECT

INCORRECT

 

TEST STEP

TEST STEPS

 

 

 

 

no

 

 

 

 

 

RETURN

 

 

1-4 Fresenius Medical Care 4008 5/03.09 (TM)

Fresenius 4008 Technical manual

1.1.2T1 test flow diagram, parallel program steps

MODULE

T1

START T1 TEST

TEST

 

 

TEST

 

TEST

 

 

TEST

 

 

TEST

 

 

TEST

 

 

OPT. DETECTOR

 

 

BLOOD SYSTEM

 

BYPASS

 

 

DISPLAY

 

 

ACCUMULATOR

 

 

ARTERIAL PRESSURE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

yes

 

 

yes

 

 

 

yes

 

 

yes

 

 

yes

 

 

yes

TEST OK

TEST OK

TEST OK

TEST OK

TEST OK

TEST OK

?

1

?

 

?

 

 

 

?

1

?

1

?

1

 

no

 

no

 

 

no

 

 

 

no

 

no

 

no

 

 

 

 

STORAGE

 

 

STORAGE

 

STORAGE

 

 

STORAGE

 

 

STORAGE

 

 

STORAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

ERROR NUMBER

 

 

ERROR NUMBER

 

ERROR NUMBER

 

 

ERROR NUMBER

 

 

ERROR NUMBER

 

 

ERROR NUMBER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST

 

TEST

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

VENOUS PRESSURE

 

TEMPERATURE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

yes

 

yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST OK

TEST OK

 

 

 

 

 

 

 

 

 

 

 

 

??

 

 

 

 

no

 

 

 

 

 

no

 

 

 

 

 

 

 

 

 

 

 

STORAGE

 

 

 

 

STORAGE

 

 

 

 

 

 

 

 

 

 

 

ERROR NUMBER

ERROR NUMBER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST

 

 

 

 

NEG. PRESSURE

 

 

 

TEST

 

 

 

 

 

 

AIR DETECTOR

 

 

 

 

HOLDING TEST

 

 

 

BLOOD LEAK DETECTOR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

yes

 

 

 

 

yes

 

 

 

yes

 

 

 

TEST OK

 

TEST OK

 

TEST OK

 

 

 

 

 

?

 

 

 

 

 

 

?

 

 

 

 

?

 

 

 

 

 

 

 

 

 

no

1

 

 

no

 

 

 

 

no

1

 

 

 

 

 

 

 

 

 

 

 

 

 

STORAGE

 

 

 

 

STORAGE

 

 

 

STORAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERROR NUMBER

 

 

ERROR NUMBER

 

 

 

ERROR NUMBER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

POS. PRESSURE

 

 

 

 

 

 

no

 

 

 

 

yes

 

 

TEST

 

 

 

 

 

 

 

 

Conductivity?

 

 

 

 

 

HOLDING TEST

 

 

 

 

 

 

 

 

 

 

 

 

 

CONDUCTIVITY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST OK

 

 

yes

 

 

 

 

 

 

 

 

 

 

 

 

TEST OK

yes

?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

?

no

 

 

 

no

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STORAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STORAGE

 

 

ERROR NUMBER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERROR NUMBER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST

 

 

 

 

 

 

TEST

 

 

 

 

 

 

 

POS. PRESSURE

 

 

CONDUCTIVITY

 

 

 

 

 

UF-FUNCTION

 

 

 

 

 

HOLDING TEST

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEST OK

yes

TEST OK

yes

TEST OK

yes

?

 

?

 

?

 

no

 

no

 

no

 

STORAGE

 

STORAGE

 

STORAGE

 

ERROR NUMBER

 

ERROR NUMBER

 

ERROR NUMBER

 

 

 

 

 

 

 

 

Basic hydraulics

 

 

Advanced hydraulics

TEST

 

TEST

 

 

TEST

 

 

TEST DIASAFE PLUS /

DIASAFE/HDF FILTER

 

 

 

 

 

 

UF-FUNCTION

 

 

DIASAFE/HDF FILTER

 

 

ONLINE PLUS /

 

 

 

 

 

 

 

 

 

yes

 

 

 

 

 

 

 

HPU-TEST

TEST OK

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

?

 

TEST OK

yes

TEST OK

yes

 

 

 

 

no

?

 

 

?

 

 

 

 

 

 

 

1

no

1

 

no

1

 

 

 

STORAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERROR NUMBER

 

STORAGE

 

 

STORAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERROR NUMBER

 

 

ERROR NUMBER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fresenius Medical Care 4008 5/03.09 (TM)

1-5

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T1 TEST

yes

 

 

UNSUCCESSFUL

DIALYSIS START KEY

 

 

 

no

INCORRECT

 

 

 

 

 

RETURN

TEST STEP

 

 

 

 

 

 

 

 

TEST OK

no

ERROR

 

 

?

 

DISPLAY

 

 

 

yes

 

 

 

 

 

 

 

 

 

FURTHER

no

NEXT INCORRECT

 

 

INCORRECT

 

 

 

TEST STEP

 

 

TEST STEPS

 

 

 

 

 

 

 

 

yes

 

 

 

 

 

 

 

 

 

RETURN

 

 

1-6 Fresenius Medical Care 4008 5/03.09 (TM)

1.1.3Description of the T1 test incl. error messages

Prerequisites for starting and running the test

Error message

Description

Power failure

Power failure while the test is in progress

Dialines not conn

The dialysate lines are not in the interlock shunt.

Shunt Cover open

The interlock shunt is open.

Connect Conc.Line

 

Wrong conc. supply

The concentrate connector is in the rinse chamber, or concen-

 

trate is not connected at all. The error message depends on the

 

central delivery system preselected in the setup menu.

Blood Sensed by OD

The optical detector senses blood in the system.

Flow alarm

Line to or from the dialyzer kinked, malfunctions in the hydraul-

 

ics.

Water alarm

Water supply interrupted.

XXX not calibrated

A valid calibration value is missing in the NOVRAM.

Overview of the individual test steps

 

Bypass test .....................................................................................................................

1-8

Optical detector test .......................................................................................................

1-10

Blood system test ...........................................................................................................

1-12

Venous pressure system test .........................................................................................

1-16

Air detector test ..............................................................................................................

1-18

Display test .....................................................................................................................

1-22

Arterial pressure system test ..........................................................................................

1-24

Battery test .....................................................................................................................

1-26

Blood leak test ................................................................................................................

1-28

Temperature test ............................................................................................................

1-30

Negative pressure holding test .......................................................................................

1-32

Positive pressure holding test ........................................................................................

1-34

UF function test ..............................................................................................................

1-39

Conductivity test .............................................................................................................

1-42

Diasafe/HDF filter test ....................................................................................................

1-44

Online plus / Diasafe plus filter / HPU test .....................................................................

1-48

Fresenius Medical Care 4008 5/03.09 (TM)

1-7

Bypass test Test description:

Check of the following functions:

Heater relay

Bypass (electric)

Check of the temperature range changeover

Illustration:

X632/A26 HOT_RINSE

 

 

 

 

 

 

 

 

 

 

 

 

LP 639

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(4008E/H)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power

 

X639/

 

 

 

 

 

 

 

 

 

 

 

 

 

A20

 

 

 

 

 

 

 

 

 

 

 

 

Logic

 

 

 

 

 

X632/A10

 

 

H_REL_W

X639/A12

 

LP 647

 

 

 

 

 

X632/A9

 

 

EM_H_OFF

X639/A17

 

 

 

 

 

 

X632/C25

 

 

V24_EN

 

 

(4008S/B)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

X632/C26

 

 

V24B_EN

 

 

 

 

 

 

 

 

 

 

X632/B22

 

 

CI

 

 

 

 

 

 

 

 

LP 632

 

 

 

 

 

 

 

 

 

 

X632/A6

V26

 

 

 

 

 

 

 

 

CPU 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

X632/A6

 

 

V26

 

 

 

 

 

 

 

 

 

 

 

X634R/

 

 

 

 

 

 

X632/A5

 

 

V24B

 

X634R/

 

 

 

X632/A4

V24

 

A18

C22

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LP 634

 

X632/B27

X632/B28

 

 

 

 

 

 

 

 

 

 

 

 

 

LP 633

 

 

Output

X634R/

 

 

 

 

 

 

 

 

 

Input

X634R/

 

 

 

 

 

 

 

 

 

 

A23

 

board

C24

board

X631/A20

X631/A21

X634L/A25

X634L/C12

X634L/C25

LP 631

 

 

 

 

 

CPU 1

LP 630

DATA BUS

 

 

 

 

 

 

 

Mother

 

 

 

 

board

 

 

 

LP 635

LP 649

(4008

E)

(4008

B)

 

Display

 

 

board

 

LP 924

LP 922

(4008

H)

(4008

S)

 

 

 

 

Testgenerierung/Generation of Test

Rueckmeldung/Acknowledgement

V24

V24

V24B

V24B

V26

V26

1-8 Fresenius Medical Care 4008 5/03.09 (TM)

Error description:

 

Error message

Description

F 01 Bypass

The heater relay is switched off.

 

– Acknowledgement (H_REL_W, X639/A12) → X632/A10, 0 V are

 

missing.

F 02 Bypass

The heater relay cannot be switched off by CPU2.

 

– Acknowledgement (H_REL_W, X639/Y12) → X632/A10, 12 V are

 

missing.

 

– Control line (EM_H_OFF, X632/A9) → X639/A17, 12 V are missing.

F 03 Bypass

The temperature measurement range is set to hot rinse.

 

– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are miss-

 

ing.

 

– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are

 

missing.

F 04 Bypass

The extended bypass cannot be correctly switched by CPU2 (V24 = off,

 

V26 = on, V24B = off).

 

– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.

 

– Acknowledgement (V26, X637/C2) → X632/A6, 0 V are missing.

 

– Acknowledgement (V24B, X637/C23) → X632/A5, 24 V are miss-

 

ing.

F 05 Bypass

The extended bypass cannot be correctly switched off by CPU2 (V24 =

 

on, V26 = off, V24B = on).

 

– Acknowledgement (V24, X637/C1) → X632/A4, 0 V are missing.

 

– Acknowledgement (V26, X637/C2) → X632/A6, 24 V are missing.

 

– Acknowledgement (V24B, X637/C23) → X632/A5, 0 V are missing.

F06 Bypass

CPU1 fails to set the temperature control to hot rinse.

 

– Control line (HOTRINSE, X634R/C24) → X639/A20, 12 V are miss-

 

ing.

 

– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 12 V

 

are missing.

F 07 Bypass

The extended bypass cannot be correctly switched by CPU1 (V24 = off,

 

V26 = on, V24B = off).

 

– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.

 

– Acknowledgement (V26, X637/C2) → X632/A6, 0 V are missing.

 

– Acknowledgement (V24B, X637/C23) → X632/A5, 24 V are miss-

 

ing.

F08 Bypass

CPU1 fails to reset the temperature control to dialysis.

 

– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are miss-

 

ing.

 

– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are

 

missing.

F09 Bypass

The extended bypass cannot be correctly switched off by CPU1 (V24 =

 

on, V26 = off, V24B = on).

 

– Acknowledgement (V24, X637/C1) → X632/A4, 0 V are missing.

 

– Acknowledgement (V26, X637/C2) → X632/A6, 24 V are missing.

 

– Acknowledgement (V24B, X637/C23) → X632/A5, 0 V are missing.

F95 Bypass

System error

Fresenius Medical Care 4008 5/03.09 (TM)

1-9

Optical detector test Test description:

Attenuation of the optical detector.

Check of the acknowledgement of the optical detector.

Illustration:

 

 

LP 632

X632/C16

 

 

 

LDSA

 

 

 

CPU 2

X632/C15

 

 

 

ODSA

 

 

 

 

 

X632/A30

 

 

OD_OUT

 

 

 

 

 

 

 

 

 

 

 

X633L/C7

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LP 633

 

LP 634

/B27

X632/B28

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Input

 

Output

 

 

 

 

 

 

 

 

 

board

 

board

 

 

 

 

 

 

 

 

 

 

 

 

 

 

X633L/C8

/A20

X631/A21

IN

LP 631

OD

 

CPU 1

LP 630

DATA BUS

 

 

 

Mother

 

 

board

 

 

 

 

 

X351/5

X351/7

LP 635

LP 649

 

Pven

 

 

(4008

E)

(4008

B)

 

 

X351/10

 

 

 

 

 

 

 

Display

 

 

 

 

 

 

 

 

 

board

 

 

 

 

 

 

 

 

LP 924

LP 922

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(4008

H)

(4008

S)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Testgenerierung/Generation of Test

Rueckmeldung/Acknowledgement

1-10

Fresenius Medical Care 4008 5/03.09 (TM)

Error description:

 

 

Error message

Description

F01 opt. Detector

CPU1 interprets the optical detector in a different way than does CPU2.

 

Acknowledgement (OD_OUT, X633L/C7) → X632/A30 and the

 

 

digital input of P.C.B. LP 633 measure different levels.

F02 opt. Detector

CPU2 fails to recognize blood in the system.

 

Acknowledgement (OD_OUT, X633L/C7) → X632/A30, 0 V are

missing.

– Detuning (ODSA, X632/C15) → X351/7 not 12V.

F03 opt. Detector CPU1 fails to recognize blood in the system.

Acknowledgement (OD_OUT, X633L/C7) → digital input on P.C.B. LP 633.

Detuning (ODSA, X632/C15) → X351/7 not 12V.

F04 opt. Detector CPU2 recognizes that the optical detector senses opaque fluid (required because of the test in the cleaning program).

Acknowledgement X632/A30 not 12V.

AD28 defective.

F96 opt. Detector

System error.

Fresenius Medical Care 4008 5/03.09 (TM)

1-11

Blood system test Test description:

Check of the following functions:

Blood alarm acknowledgement

Blood pump switch-off

Illustration:

 

X632/C10

 

CLP_CTL

 

 

 

 

 

 

X632/C14

 

LDA2

 

X632/C21

 

BL_AL

 

 

 

LP 632

 

BPSB_VEN

 

 

X632/B11

 

CPU 2

 

 

 

 

 

X632/A11

 

BPSB_ART

 

 

 

 

 

 

 

 

 

 

X632/A15

SN_ART

 

 

 

 

 

 

 

 

X634L/

 

 

 

 

 

 

 

 

 

 

 

 

C15

 

X632/B27

X632/B28

LP 633

 

 

 

 

X634L/

LP 634

 

 

 

 

 

 

A14

 

(4008E/H)

 

 

 

 

 

 

 

 

 

 

 

Input

 

 

 

 

A15

 

Output

 

 

board

 

 

 

 

 

 

 

board

 

 

 

 

 

 

 

 

ART

 

 

LP 647

 

 

 

 

 

 

 

 

_

 

 

 

 

 

 

 

 

 

 

 

 

BPST

 

 

(4008S/B)

 

 

X633L/

A13

X633L/

 

 

 

X634L/ B15

X634L/

 

 

 

 

 

 

 

 

C13

 

 

 

 

 

 

 

B14

 

C14

 

 

 

 

 

 

 

 

 

 

 

X631/A20

X631/A21

 

 

 

 

 

 

 

 

 

 

 

LP 631

 

 

 

 

 

 

 

 

 

 

 

 

LP 630

 

 

 

 

 

 

BPSST_A

 

 

CPU 1

DATA BUS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Mother

 

 

 

 

 

 

 

 

V

 

 

 

board

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BPSST

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

X348a/2

X348a/6

X348a/3

X348a/1

X348/V6

X348/V3

X348/V1

X351/8

X351/6

LP 635

LP 649

 

 

 

 

 

 

 

 

 

Pven

 

 

 

 

 

 

 

 

 

 

 

(4008 E)

(4008 B)

 

 

 

 

 

 

 

 

 

 

 

Display

board

LP 924

LP 922

(4008 H)

(4008 S)

Testgenerierung/Generation of Test

Rückmeldung/Acknowledgement

1-12

Fresenius Medical Care 4008 5/03.09 (TM)

Error description:

 

 

Error message

Description

F09 Bloodsystem

Acknowledgement that CPU2 recognizes that the arterial blood pump

 

is inactive (BP not running).

 

– Acknowledgement (BPSB_ART, X348a/6) → X632/A11, 12 V miss-

 

 

ing.

 

– Control line (BPSST_ART, X634L/B14) → X348a/1, 12 V missing or

 

 

(BPST_ART, X634L/A14) → X348a/3, 12 V are missing.

F10 Bloodsystem

Acknowledgement that CPU1 recognizes that the arterial blood pump

 

is inactive (BP not running).

 

– Acknowledgement (BPSB_ART, X348a/6) → X633L/A11, 12 V are

 

 

missing.

 

– Control line (BPSTT_ART, X634L/B14) → X348a/1, 12 V missing or

 

 

(BPST_ART, X634L/A14) → X348a/3, 12 V missing.

 

– Level is raised during the T1 test.

F11 Bloodsystem

The arterial blood pump cannot be stopped by CPU1.

 

CPU2 recognizes that the arterial blood pump remains active.

 

– Control line (BPSST_ART, X634L/B14) → X348a/1, 0 V missing,

 

 

as well as (BPST_ART, X634L/A14) → X348a/3, 0 V missing.

 

Acknowledgement (BPSB_ART, X348a/6) → X632/A11, 0 V are

 

 

missing.

 

– The level is raised during the T1 test, or the up/down key on the air

 

 

detector is blocked and the level is constantly raised.

F12 Bloodsystem

The arterial blood pump cannot be stopped by CPU1.

 

CPU1 recognizes that the arterial blood pump remains active.

 

– Control line (BPSST_A, X634L/B14) → X348a/1, 0 V missing,

 

 

as well as (BPST_ART, X634L/A14) → X348a/3, 0 V missing.

 

– Acknowledgement (BPSB_ART, X348a/6) → X633L/A11, 0 V are

 

 

missing.

F13 Bloodsystem

Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or

 

4008 HDF pump is connected (= ADKS active)

 

Acknowledgement that CPU2 detects that the pump is inactive (pump

 

is not running).

 

Acknowledgement (BPSB_VEN, X348V/6) → X632/ B11, 12V

 

 

missing

 

– Control line (BPSST_VEN, X634L/B15) → X348V/1, 12V missing or

 

 

(BPST_VEN, X634L/A15) → X348V/3, 12V missing

 

– Transistor T9 on P.C.B. LP 754 defective

 

– IC5 on P.C.B. LP 632 defective

 

In 4008 HDF an HDF treatment was performed, followed by a

 

 

cleaning program with the substituate pump running, then the T1

 

 

test has been re-started.

 

 

The substituate pump must be switched off because otherwise the

 

 

test step will fail to be passed (problem was corrected with SW 3.20

 

 

in 4008 H/S systems: the substituate pump will be switched off

 

 

automatically on starting a cleaning program).

Fresenius Medical Care 4008 5/03.09 (TM)

1-13

F14 Bloodsystem

Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or

 

4008 HDF pump is connected (= ADKS active)

 

Acknowledgement that CPU1 detects that the pump is inactive (pump

 

is not running).

 

– Acknowledgement (BPSB_VEN,X348V/6) → X633L/A13, 12V

 

missing

 

– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 12V or

 

(BPST_VEN, X634L/A15) → X348V/3 not 12V

 

– IC16 on P.C.B. LP 633 defective

 

– P.C.B. LP 633 recognizes Single-Needle pump although it is not

 

connected.

F15 Bloodsystem

Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or

 

4008 HDF pump is connected (= ADKS active)

 

CPU1 fails to stop the corresponding blood pump.

 

CPU2 detects that the pump remains active.

 

– Control line (BPSST_VEN, X634L/B15) → X348V/1, 0V missing as

 

well as (BPST_VEN, X634L/A15) → X348V/3 not 0V

 

– Acknowledgement (BPSB_VEN, X348V/6) → X632/B11, 0V miss-

 

ing

 

– Transistor T9 on P.C.B. LP 754 defective

 

– IC5 on P.C.B. LP 632 defective

 

– During the test the lines are inserted on the corresponding pump

 

using the Start/Stop key.

 

– P.C.B. LP 633 recognizes Single-Needle pump although it is not

 

connected.

F16 Bloodsystem

Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or

 

4008 HDF pump is connected (= ADKS active)

 

CPU1 fails to stop the corresponding blood pump.

 

CPU1 detects that the pump remains active.

 

– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 0V as well

 

as (BPST_VEN, X634L/A15) → X348V/3 not 0V

 

– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 0V

 

– IC16 on P.C.B. LP 633 defective

 

– P.C.B. LP 633 recognizes Single-Needle pump although it is not

 

connected.

F17 Bloodsystem

Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or

 

4008 HDF pump is connected (= ADKS active)

 

Although the recognition of the venous blood pump (ADKS) is not

 

acknowledged, the 24-V supply voltage of the pump can be switched

 

off.

 

– Acknowledgement line (ADKS, X348V/7) → X633L/A10 not 12V

 

– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 12V

 

– Acknowledgement (BPSB_VEN, X348V/6) → X632/B11 not 12V

 

– Online-HDF has already been switched on during the T1 test.

 

– IC16 on P.C.B. LP 633 defective.

1-14

Fresenius Medical Care 4008 5/03.09 (TM)

F18 Bloodsystem Applicable for SW 5.00/4.10 and higher, check of the BPUS signal (CPU, P.C.B. LP 632)

At the beginning of the test step a maximum of 40s may pass until rotation has stopped.If the blood pump is being activated, the rotation stop alarm must have been cleared.

Acknowledgement line (BPUS, X348A/8) → X632/A13 not 0V

Acknowledgement line (BPUS, X348A/8) → X632/A13 not 12V

Blood pump speed is set to “0”: preset speed during the T1 test.

F19 Bloodsystem Applicable for SW 5.00/4.10 and higher, check of the BPUS signal (CPU, P.C.B. LP 631 via LP 633)

At the beginning of the test step a maximum of 40s may pass until rotation has stopped. If the blood pump is being activated, the rotation stop alarm must have been cleared.

1.Acknowledgement line (BPUS, X348A/8) → X633L/A12 not 0V

2.Acknowledgement line (BPUS, X348A/8) → X633L/A12 not 12V

F20 Bloodsystem Check of the actual arterial BP rate.

The actual rate of the arterial BP is not zero. The actual rate of the arterial BP does not increase.

If SN is installed: The actual rate of the venous BP is not zero. The actual rate of the venous BP does not increase.

Acknowledgement line (BPR_ART, X348A/10) → X633L/B3 not 0V or acknowledgement line (BPR_ART, X348A/10) → X632/A14 not 0V

Acknowledgement line (BPR_ART, X348A/10) → X633L/B3 no increase or acknowledgement line (BPR_ART, X348A/10) → X632/ A14 no increase

If SN is installed:

Acknowledgement line (BPR_VHDF, X348V/10) → X633L/B4 not 0V

Acknowledgement line (BPR_VHDF, X348V/10) → X633L/B4 no increase

F95 Bloodsystem

System error.

Fresenius Medical Care 4008 5/03.09 (TM)

1-15

Venous pressure system test Test description:

Verification of the lower limit by checking the venous zero point.

The upper limit is tested by detuning the venous pressure unit in positive direction. (The venous line clamp is closed during the test.)

Illustration:

LP 632

X632/C16

 

LDSA

 

 

CPU 2

 

 

 

 

 

X632/C18

 

PV_DET

 

 

 

X632/C17

 

P_VEN

 

 

 

 

 

 

X632/B27 X632/B28

LP 633

LP 634

 

 

Input

Output

 

 

board

board

 

 

X633L/B5

X634R/C18

 

 

 

V

X631/A20

X631/A21

 

VENT

 

 

LP 631

 

 

CPU 1

LP 630

DATA BUS

 

 

 

 

 

Mother

 

 

 

board

 

 

 

X351/1

X351/2

LP 635

LP 649

PVEN

(4008 E)

(4008 B)

 

Display

 

 

board

X351/4

X351/10

LP 924

LP 922

 

 

(4008 H)

(4008 S)

 

 

Testgenerierung/Generation of Test

Rueckmeldung/Acknowledgement

1-16

Fresenius Medical Care 4008 5/03.09 (TM)

Error description

 

Error message

Description

F01 Venous

CPU1 (input board) shows a venous zero point deviation of more than

 

±12 mmHg (60 s).

 

– Control (VENT_VALVE, X634R/C18) → X351/1 of the vent valve in

 

the LD is defective.

 

– Acknowledgement (P_VEN, X351/4) → X633L/B5 that the voltage

 

value is outside the zero point tolerance.

 

– P-venous has not been calibrated.

F02 Venous

CPU2 shows a venous zero point deviation of more than ±12 mmHg

 

(60 s).

 

– Control (VENT_VALVE, X634R/C18) → X351/1 of the vent valve in

 

the LD is defective.

 

– Acknowledgement (P_VEN, X351/4) → X632/C17, the voltage

 

value is outside the zero point tolerance.

 

– P-venous has not been calibrated.

F03 Venous

With detuning in positive direction, the achieved change in the venous

 

display is less than 100 mmHg (7 s).

 

– The test detuning is defective (PV_DET, X632/C18) → X351/2.

 

– Acknowledgement (P_VEN, X351/4) → X633L/B5, the change in

 

voltage is too low.

 

– P-venous has not been calibrated.

F04 Venous

The deviation in the measured value between CPU1 and CPU2 is

 

higher than ±12 mmHg (if Pven > 100 mmHg).

 

– Acknowledgement (P_VEN, X351/4) → X633L/B5 and X632/C17

 

measure different voltage values.

 

– P-venous has not been calibrated.

F95 Venous

System error.

Fresenius Medical Care 4008 5/03.09 (TM)

1-17

Air detector test Test description:

Test of the air detector by checking the alarm state.

Switch-off of the venous line clamp in the air detector module.

Illustration:

1-18

Fresenius Medical Care 4008 5/03.09 (TM)

Error description:

 

Error message

Description

F01 Airdetector

CPU1 interprets the air detector signal in a different way than does

 

CPU2.

 

– Acknowledgements (LDA1, X351/14) → X632/C13 and X633L/C10

 

recognize different signal levels.

F02 Airdetector

The air detector alarm is not recognized by CPU2.

 

– Acknowledgement (LDA1, X351/14) → X632/C13, 0 V are missing.

 

– Transmission weakening (LDSA, X632/C16) → X351/10, 12 V are

 

missing.

F03 Airdetector

Air detector clamps acknowledgement (CPU2) activated (clamp

 

closed).

 

– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.

 

– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.

 

– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.

F04 Airdetector

Air detector clamps acknowledgement (CPU1) activated (clamp

 

closed).

 

– Acknowledgement (LDA2, X351/6) → X633L/C13, 24 V are miss-

 

ing.

 

– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.

 

– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.

F05 Airdetector

The blood alarm signal has not been cleared (indicates an alarm).

 

– Acknowledgement (BL_AL, X634L/C15) → X632/C21, 12 V are

 

missing.

 

If the HDF option is used, this signal is not tested (special function).

F06 Airdetector

Closing of the air detector clamp via the CPU2 control line was not

 

possible.

 

– Clamp control (CLP_CTL, X632/C10) → X351/8, 0 V are missing.

 

– Acknowledgement (LDA2, X351/6) → X632/C14, 0 V are missing.

F07 Airdetector

Opening of the air detector clamp via the CPU2 control line was not

 

possible.

 

– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.

 

– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.

F08 Airdetector

Closing of the air detector clamp via the CPU1 control line was not

 

possible, or CPU2 acknowledgement is incorrect.

 

– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V are missing.

 

– Acknowledgement (LDA2, X351/6) → X632/C14, 0 V are missing.

Fresenius Medical Care 4008 5/03.09 (TM)

1-19

F09 Airdetector

Closing of the air detector clamp via the CPU1 control line was not

 

possible, or CPU1 acknowledgement is incorrect.

 

– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V are missing.

 

– Acknowledgement (LDA2, X351/6) → X633L/C13, 0 V are missing.

F10 Airdetector

The blood alarm message is missing.

 

Acknowledgement (BL_AL, X634R/C15) → X632/C21, 0 V are

 

 

missing.

 

 

If the HDF option is used, this signal is not tested (special function).

F11 Airdetector

Air detector clamps acknowledgement (CPU2) activated (clamp

 

closed).

 

– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.

 

– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.

 

– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.

F12 Airdetector

Air detector clamps acknowledgement (CPU1) activated (clamp

 

closed).

 

– Acknowledgement (LDA2, X351/6) → X633L/C13, 24 V are miss-

 

 

ing.

 

– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.

 

– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.

F13 Airdetector

The blood alarm signal has not been cleared (indicates alarm).

 

Acknowledgement (BL_AL, X634L/C15) → X632/C21, 12 V are

 

 

missing.

 

 

If the HDF option is used, this signal is not tested (special function).

F14 Airdetector

Raise level key on the air detector is constantly active.

 

– Acknowledgement (LEVEL_UP, X351/3) → X632/C11 not 0V.

F15 Airdetector

Acknowledgement of the supply voltage for the ultrasonic output stage

 

not between 6.5 and 13.5 V after 3 seconds.

 

– Adapter board AD28 not connected

 

Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7)

 

 

not 12V.

 

– Relay on AD28 failed to drop.

F16 Airdetector

Acknowledgement of the supply voltage for the ultrasound output stage

 

not >14.5V after 3 seconds.

 

– Adapter board AD28 not connected.

 

Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7)

not 16V/24V.

– Relay on AD28 is not controlled.

– No 10-Hz signal at ALARM_REST (X351/12)

1-20

Fresenius Medical Care 4008 5/03.09 (TM)

F17 Airdetector

Acknowledgement of the supply voltage for the ultrasound output stage

 

not between 6.5 and 13.5 V after 3 seconds.

 

– Adapter board AD28 not connected

 

– Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7)

 

not 12V

 

– Relay on AD28 failed to drop

F95 Airdetector

System error.

Fresenius Medical Care 4008 5/03.09 (TM)

1-21

Display test Test description:

Check of all displays and indicators on the monitor front

Display test

Status LED

Alarm LED

Seven-segment display, all dark

Seven-segment display, all 8888

Bar graph

CPU1/CPU2 alarm tone

This display test must be monitored by the user!

Illustration:

 

LP 632

X632/C29

CPU2_AL

 

 

 

 

CPU 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

X634R/A16

 

 

 

 

 

LP 633

 

LP 634

 

 

 

 

 

Input

 

Output

X632/ B27

X632/ B28

 

 

 

 

 

 

board

 

board

 

 

 

 

 

 

 

 

 

+LS X634L/A13 -LS X634L/B13

X631/ A20

X631/ A21

 

LP 631

 

CPU 1

LP 630

DATA BUS

 

 

 

Mother

 

 

board

 

LP 635

LP 649

 

 

 

 

(4008

E)

(4008

B)

 

 

 

 

 

Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

board

 

 

 

 

 

 

 

 

LP 924

LP 922

 

 

 

 

(4008

H)

(4008

S)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test Display

8888

8888

8888

8888

Testgenerierung/Generation of Test

Rueckmeldung/Acknowledgement

1-22

Fresenius Medical Care 4008 5/03.09 (TM)

Loading...
+ 320 hidden pages