Fresenius 5008 Service manual

4.5 (19)

Service Manual

Hemodialysis System

5008

Edition: 1/08.04

 

Part number: M35 179 1

0123

Fresenius Medical Care

Page

Table of Contents

1Index

2Important Information

2.1

Organization of the Technical Document................................................................................

2-1

2.2

How to Use the Technical Document ......................................................................................

2-1

2.3

Precautions for Working on the System .................................................................................

2-2

2.4

Addresses ..................................................................................................................................

2-2

3Specifications

4Installation

 

4.1

Preface........................................................................................................................................

4-1

 

4.2

Important Information on Initial Start-Up ................................................................................

4-1

 

4.3

Initial Start-Up Report ...............................................................................................................

4-2

 

4.4

Explanations on the Initial Start-Up Report ............................................................................

4-9

5

Setup

 

 

 

5.1

Operator Setup ..........................................................................................................................

5-1

 

5.2

Technician's SETUP ................................................................................................................

5-13

 

5.3

Information Regarding the Setting of Concentrates in the Technician’s Setup................

5-21

6 TSC / TMC / Maintenance

6.1

Important Information ...............................................................................................................

6-1

6.2Test Report – Technical Safety Checks, Technical Measurements Checks and Maintenance Procedures6-3

6.3Explanations on Technical Safety Checks, Technical Measurement Checks and Maintenance Procedures6-11

6.4

TSC / TMC Report ....................................................................................................................

6-27

Fresenius Medical Care 5008 SM 1/08.04

i

Kapitel :

7Error Messages

8Tools (Service Equipment)

9Calibration / Adjustment

10Repair

11Functional Description

11.1

Overall System.........................................................................................................................

11-1

11.2

Overview of P.C.B.s.................................................................................................................

11-3

11.3

Monitor......................................................................................................................................

11-4

11.4

EBM (Extracorporeal Blood Module) .....................................................................................

11-6

11.5

Hydraulics Unit ........................................................................................................................

11-8

11.6

Power Supply Unit .................................................................................................................

11-10

11.7

Pneumatic Unit.......................................................................................................................

11-12

11.8

Hydraulics Unit ......................................................................................................................

11-14

12 Service Program (Option)

ii

Fresenius Medical Care 5008 SM 1/08.04

Chapter 1: Index

1 Index

How to use the index: E.g., index entry 1-3 is to be interpreted as: Chapter 1, page 3

A

R

Addresses 2-2

Reinfusion 5-2

Alarm processing 5-4

T

AutoFlow 3-6

B

Temperature 3-6

U

Blood pump 5-1

BPM 5-10

Ultrafiltration 5-3

BTM 5-11

User interface 5-5

 

C

 

Concentrates 3-6

 

D

 

DIASAFE®plus 3-8

 

E

 

EcoFlow 3-6

 

Emergency 5-8

 

External connections 3-3

 

F

 

Flow diagram 11-14

 

H

 

Heparin pump 3-8

 

O

 

OCM 3-9, 5-9

 

ONLINEplus™ 5-9

 

Operating programs 3-4

 

Operator Setup 5-1

 

Optical detector 3-7

 

Override conditions 3-4

 

P

 

Patient card 5-8

 

Fresenius Medical Care 5008 SM 1/08.04

1-1

Kapitel 1: Index

1-2

Fresenius Medical Care 5008 SM 1/08.04

Chapter 2: Important Information

2 Important Information

2.1Organization of the Technical Document

Page identification

Page number 1-3 is to be interpreted as: Chapter 1, page 3.

Document changes

Document changes will be released as new editions or supplements. In

 

general: this manual is subject to change without notice.

Editorial information

The current edition of this technical document is:

 

1/08.04 = 1st edition, August 2004

2.2How to Use the Technical Document

Intended use

This technical document is intended for service technicians and is to be

 

used for first studies (to acquire a basic knowledge) and for reference

 

purposes (for TSC, maintenance and repair). The study of this

 

document, however, does not replace the training courses offered by

 

the manufacturer.

Requirements

Knowledge of the current Operating Instructions of the respective

 

system.

 

Background experience in mechanics, electrical and medical

 

engineering.

Note and Caution symbols

 

 

Explanation of the Note and Caution symbols used:

 

 

 

Note

 

Informs the operator that in case of a failure to follow the steps as

 

described, a specific function will be executed incorrectly or will not be

 

executed at all, or will not produce the desired effect.

 

 

 

 

 

Caution

 

Advises the operator against certain procedures or actions that could

 

cause damage to the equipment or may have adverse effects on

 

operators and patients.

 

 

Fresenius Medical Care 5008 SM 1/08.04

2-1

Chapter 2: Important Information

2.3Precautions for Working on the System

Authorized persons

Assembly, extensions, adjustments, modifications or repairs may only

 

be carried out by the manufacturer or persons authorized by him.

Measuring equipment and

The activities described in this technical document require the

accessories

availability of the necessary technical measuring equipment and

 

accessories.

Precautions

Before turning power on, repair any visible damage.

 

Prior to opening the system and when working on the open system, the

 

following precautions have to be taken:

 

– Protect the components against ingress of liquids.

 

– Do not touch live parts.

 

– All plugs, connections and components may only be disconnected or

 

connected if de-energized.

ESD precautions

When repairing the system and replacing spare parts, observe

 

applicable ESC precautions (e.g. EN 100 015-1).

Monitor support arm

If the 5008 hemodialysis system is to be placed in a horizontal position

 

for servicing, the monitor support arm must be protected with the

 

transport protection to prevent it from flipping over.

To be observed after

A disinfection and a T1 test must be performed after working on the

working on the system

system.

To be observed after

After a disinfection program has been aborted or if the system is to be

aborting a disinfection

preserved, the hemodialysis system must be disconnected from the

program

water supply after a maximum of 3 days. When the system is returned

 

to use, check that the pressure of the water supply meets the prescribed

 

minimum pressure.

2.4 Addresses

Please address any inquires to:

Fresenius Medical Care AG 61346 Bad Homburg Germany

Phone: + 49 6172 609-0 www.fmc-ag.com

2-2

Fresenius Medical Care 5008 SM 1/08.04

Chapter 2: Important Information

Service

Fresenius Medical Care

Central Europe

Deutschland GmbH

 

Geschäftsbereich Zentraleuropa

 

Kundendienst / Servicecenter

 

Steinmühlstraße 24 I

 

61352 Bad Homburg

 

Germany

 

Phone: +49 6172 609-7100

 

Fax: +49 6172 609-7102

 

E-mail: ServicecenterD@fmc-ag.com

Service

Fresenius Medical Care

International

Deutschland GmbH

 

Service Support International

 

Hafenstraße 9

 

97424 Schweinfurt

 

Germany

 

Phone: +49 9721 678-333 (hotline)

 

Fax: +49 9721 678-130

Local Service

 

 

 

Fresenius Medical Care 5008 SM 1/08.04

2-3

Chapter 2: Important Information

2-4

Fresenius Medical Care 5008 SM 1/08.04

Fresenius 5008 Service manual

Chapter 3: Specifications

3 Specifications

Dimensions, weight and housing material

Dimensions

Height: approx. 162 cm (approx. 210 cm incl. IV pole)

 

Width: approx. 48 cm (on base incl. brake)

 

Depth: approx. 72 cm (approx. 86 cm with extended concentrate rack)

Weight

Approx. 135 kg (without options)

Housing material

PU vacuum cast resin

Type label

 

1

2

3 6

4

5

 

1Type identification, serial number

2Power requirements

3Protection against ingress of liquids: drip-proof

4Caution, consult accompanying documents

5Degree of protection against electric shock: Type B

6CE mark

Electrical safety (classification according to EN 60601-1, IEC 601-1)

Type of protection against

Safety class I

electric shock

 

Degree of protection

Type B, symbol:

against electric shock

 

Fresenius Medical Care 5008 SM 1/08.04

3-1

Chapter 3: Specifications

Applicable only to the BPM blood pressure cuff: Degree of protection against electric shock

Degree of protection against ingress of liquids

Leakage currents

EMC specifications according to EN 60601-1-2 (IEC 601-1-2)

Electric supply

Line voltage

Connection to power supply

Operating current dialysis

Power supply (internal)

Battery

Fuses

Main power switch

Operating conditions

Water inlet pressure

Water inlet temperature

Water inlet rate

Type CF, symbol:

Drip-proof, symbol: IPX1

According to EN 60601-1

RFI emissions:

Limit class A according to EN 55011, Group 1

Immunity:

Electrostatic discharge, atmospheric discharge: 8 kV

Electromagnetic fields: 27 MHz – 1000 MHz: 3 V/m

Bursts: Power line (alternating current): 1 kV

Surge voltages (alternating current): 2 kV

100 to 230 V AC, ±10 %, 47 to 63 Hz

(The decisive criterion is the line voltage and the operating current specified on the type label of the system)

16 A at 230 V, regulation according to VDE 0100 part 0107

Approx. 6 A, (at 230 V)

at a water inlet temperature of 17 °C Dialysate temperature 37 °C Dialysate flow: 500 ml/min

+24 V ± 3 %, 20 A short-circuit proof +18 V ± 3 %, 14 A short-circuit proof 480 W total power output

Lead-acid battery (maintenance-free) 24 V, 7 Ah

2 x G 16 A (miniature circuit-breaker) rear of power supply unit

1.5 to 6.0 bar

5 °C to 30 °C

with "Integrated hot rinse": 85 °C to 95 °C

1.5 l/min; at an inlet pressure of 1.5 bar

3-2

Fresenius Medical Care 5008 SM 1/08.04

 

Chapter 3: Specifications

 

 

Water drain

0 to 100 cm above the floor, minimum 5 cm free fall. The water drain

 

must be located at a lower level than the dialyzer position.

Concentrate supply

0 to -100 mbar; maximum suction height 1 m

 

with Central Delivery System (option): 0.05 to 2.0 bar

Heat dissipation

Dialysis:

 

approx. 400 Watt (at an ambient temperature of 20 °C)

Range of operating

15 °C to 35 °C

temperature

 

Atmospheric pressure

700700 hPa to 1060 hPa

Relative humidity

30 % to 75 %, temporarily 95 %

Stability

IV pole load capacity

Maximum: 5 kg

 

Maximum load capacity of one hook: 5 kg

External connection options

 

 

 

 

Caution

 

Any additional equipment connected to the analog and digital interfaces

 

of the machine must comply with the applicable EN specifications (e.g.

 

EN 60950 for data processing equipment and EN 60601 (IEC 601) for

 

electro-medical equipment).

 

Apart from this requirement, all configurations must comply with the

 

system standard EN 60601-1-1 (IEC 601-1-1), or their applicability with

 

regard to safety has to be proven by a certificate issued by a testing

 

agency authorized to test the ready-for-use machine.

 

The connection of additional equipment to the signal input or output

 

component affects the system configuration and anyone connecting

 

additional equipment is therefore responsible for compliance with the

 

system standard EN 60601-1-1 (IEC 601-1-1).

 

 

 

 

 

Caution

 

The external alarm indicators do not relieve the operator of the

 

obligation to observe the local alarms of the system.

 

 

LAN

Interface for the exchange of data.

 

Electrically isolated by transformer.

 

Port: RJ 45

RS232

Interface for the exchange of data.

 

Electrically isolated by optocoupler.

 

Port: DSUB 9-pin

Fresenius Medical Care 5008 SM 1/08.04

3-3

Chapter 3: Specifications

Service/diagnostics

(Protected by cover!)

 

For inhouse computer diagnostics.

 

Port: DSUB 15-pin

24 V

(Protected by cover!)

 

24 V connection (2 A fuse)

 

Port: Flanged socket, 4-pin

Alarm output

For the connection of an external alarm indicator (nurse call). (Potential-

 

free alarm output. Alternating contact maximum 24 V/24 W).

 

Port: 5-pin diode plug via a shielded line; shield grounded on either side.

Override conditions

 

 

When overriding a safety system the responsibility for the patient’s

 

safety rests with the operator of the machine.

Audible alarm suppression

Mute alarm time: maximum 2 minutes

 

(adjustable in the SETUP)

Alarm override

After confirmation of the error message and start of the blood systems:

 

Arterial and venous pressure alarm for approx. 10 seconds (window

 

inactive)

 

Air detector alarm for approx. 2 seconds

Blood leak override

Override time: maximum 2 minutes

Override air-bubble

Override time: after starting removal of air: approx. 4 seconds

detector

 

Operating programs

 

T1 test

Automatic test for verification of the operating and safety systems.

 

The T1 test is mandatory,

 

– after power on (not following a power failure)

 

– after a cleaning program

Preparation

Defined by the optical detector located below the venous bubble

 

catcher.

 

Preparation is terminated as soon as the optical detector senses

 

opaque fluid in the blood lines.

Priming and rinsing the

Minimum rinse volume 500 ml; automatic switching to rinsing, if level in

blood lines

bubble catcher detected. Automatic raising of the fluid level during the

 

rinse phase.

Reinfusion

Reinfusion volume adjustable in the SETUP.

 

Return to dialysis still possible.

Dialysis

Bicarbonate dialysis

ISO-UF

Ultrafiltration without dialysate flow (Bergström method)

3-4

Fresenius Medical Care 5008 SM 1/08.04

Chapter 3: Specifications

Cleaning programs

Rinse clear/rinse/mandatory rinse:

 

Time adjustable in the TECHNICIAN's SETUP,

 

Temperature: approx. 37 °C,

 

Flow: 600, 800 ml/min (adjustable in the SETUP)

 

Cold disinfection/degreasing, cold disinfection:

 

Time adjustable in the TECHNICIAN's SETUP,

 

Temperature: approx. 37 °C,

 

Flow: max. 900 ml/min

 

Heat disinfection:

 

Time adjustable in the TECHNICIAN's SETUP,

 

Flow: max. 900 ml/min

 

In all programs:

 

Blood pump stops, arterial and venous line occlusion clamp closed.

 

Progress of the program (time-counting) is interrupted in the event of a

 

flow alarm.

 

 

The cleaning programs can be aborted.

 

The chemical disinfection program is followed by a mandatory rinse.

Flush

Rinsing of the water supply area

Dialysate circuit and safety systems

 

 

Blood leak detector

Threshold of response ≤ 0.5 ml blood loss per minute into the dialysate

 

at a hematocrit of 0.25.

 

(flow rate 100 ml/min to 1000 ml/min)

Transmembrane pressure

Display range: –100 to 400 mmHg

 

Resolution: 5 mmHg

 

Definition:

 

 

TMP

=

Pbo – (Pdi + Pdo) / 2 + Offset

 

TMP

=

Transmembrane pressure

 

Pbo

=

Blood pressure on the outlet side of the dialyzer

 

Pdi

=

Dialysate pressure on the inlet side of the dialyzer

 

Pdo

=

Dialysate pressure on the outlet side of the dialyzer

 

Offset

=

Flow-dependent pressure fluctuations

Ultrafiltration

Selectable UF rate: 0 ml/h to 4000 ml/h (in 10 ml increments)

 

Maximum rate internally adjustable to 1, 2, 3, or 4 l/h.

 

Pump volume accuracy: ±1 % (at Pdi > –500 mbar)

 

The UF rate/effective blood flow ratio is being monitored during the

 

treatment. If an incongruity occurs a warning will be displayed after

 

approx. 10 seconds.

Pressure holding test

Event-controlled

Fresenius Medical Care 5008 SM 1/08.04

3-5

Chapter 3: Specifications

Balancing

Accuracy: ±0.1 % related to the total dialysate volume

Maximum balancing error

F

=

FUF + FBil

 

 

F

=

Maximum balancing error

 

FUF

=

Ultrafiltration error

 

Fbil

=

Balancing error

 

Example:

 

 

Ultrafiltration error: with 1000 ml in 1 hour: ±1 % = ±10 ml/h

 

Balancing error: with 30 l fluid flow in 1 hour at a dialysate flow of

 

500 ml/min: ±0.1 %= ± 30 ml/h

 

Maximum balancing error:

 

F = FUF + FBil = (±10 ml/h) + (±30 ml/h) = ±40 ml/h

Degassing

Method: Negative pressure

Dialysate concentration

Display range: 12.8 to 15.7 mS/cm

(conductivity)

Resolution: 0.1 mS/cm

 

Accuracy: 0.1 mS/cm

 

Method:

 

 

Temperature-compensated electronic conductivity meter with

 

adjustable alarm limits.

Concentrates

Entering concentration types

 

Adjustment range: 125 to 151 mmol/l, depending on the concentrate

 

used ±10 % of the base value.

 

Bicarbonate readjustment range: corresponds to ±8 mmol/l

bibag®

Bicarbonate concentrate preparation from the bibag®

 

Temperature range: 15 to 35 °C

Dialysate temperature

Adjustment range: (prescribed temperature) 34.0 °C to 39.0 °C

 

Resolution: 0.5 °C

 

Measuring accuracy: ±0.2 °C

Dialysate flow

Display range: 100 to 1000 ml/min

 

Resolution: 100 ml/min

 

Desired values: 100 to 1000 ml/min

 

Measurement by means of time pulse monitoring and balancing

 

chamber volume

 

Auto flow: dialysate flow controlled in relation to the blood flow,

 

determined by the dialyzer.

 

EcoFlow: dialysate flow automatically reduced to 100 ml/min in

 

Preparation

Rinse and chemical

Desired temperature: 37 °C

disinfection temperature

Resolution: 0.5 °C

 

Measuring accuracy: ±0.2 °C

Rinse and chemical

Desired value: 600 ml/min

disinfection flow

 

 

 

3-6

Fresenius Medical Care 5008 SM 1/08.04

Chapter 3: Specifications

Hot rinse and heat

Desired temperature: 85 °C

disinfection temperature

Resolution: 0.5 °C

 

Measuring accuracy: ±2.0 °C

Hot rinse and heat

Desired value: 600 ml/min

disinfection flow

 

Concentration of

Dilution: Disinfectant is diluted with purified water in the dialysis system

disinfectant

at a ratio of 1+24.

Flow alarm

Dependent on the programmed flow

Extracorporeal blood circuit and safety systems

Arterial pressure

Display range: –300 to +300 mmHg

measurement

Resolution: 5 mmHg

 

Accuracy: 7 mmHg (typical)

 

OD senses non-opaque presence:

 

Alarm window width: –300 to +300 mmHg

 

OD senses opaque presence:

 

Alarm window width: +40 to +200 mmHg

 

Default value adjustable in the SETUP, factory setting 120 mmHg

Blood pump

Delivery rate: 30 to 600 ml/min

 

Resolution: 10 ml/min (with a line diameter of 8 mm)

 

Accuracy: < 5 % (without lines)

 

Line diameter: 4.4 mm, 6.4 mm, 8.0 mm

 

Blood pump stop alarm: 60 seconds

 

Spring-loaded rollers, fully occluding, pressure-limited to 2 bar with 8 x

 

2.1 pump line segment (when using the prescribed tubing systems).

 

(The blood pump design allows manual operation, hand crank in the

 

rotor, in clockwise direction only.)

Venous pressure

Display range: –100 to +500 mmHg

measurement

Resolution: 5 mmHg

 

Accuracy: 7 mmHg (typical)

 

OD senses non-opaque presence:

 

Alarm window width: –100 to +500 mmHg

 

OD senses opaque presence:

 

Alarm window width: 40 to 200 mmHg

 

Default value adjustable in the SETUP,

 

Factory setting 120 mmHg

 

adjustable over a range of 20 to 500 mmHg

 

(adjustable from -100 to 500 mmHg via SETUP.)

Fill level detector

Method:

 

Capacitive measurement

 

Switching point 13 mm, ±4 mm from upper edge

Optical detector

Method:

 

Infrared transmission

Fresenius Medical Care 5008 SM 1/08.04

3-7

Chapter 3: Specifications

 

Distinguishes between

 

 

OD light (saline or air in the tubing system)

 

 

OD dark (blood in the tubing system).

 

Air bubble detector

Method:

 

 

Ultrasonic transmission measurement on the line

 

Sensitivity:

 

 

– Air bubbles:

 

 

Bubble volume ≥ 20 µl

 

 

– Blood foam (air-blood mixture)

 

 

Air alarm:

 

 

– BP rate < 100 ml/min:

 

 

Air bubble: Volume ≥ 20 µl

 

 

Blood foam

 

 

– BP rate ≥ 100 ml/min:

 

 

10 air bubbles with an air bubble volume of

< 50 µl each

 

or 1 air bubble with an air bubble volume of

≥ 50 µl,

 

Blood foam

 

 

The specified data refer to the most unfavorable case with a BP rate of

 

0 to 600 ml/min when using the blood lines specified in chapter

 

Consumables.

 

Heparin pump

Delivery rate: 0.5 to 10 ml/h

 

 

Resolution: 0.1 ml/h

 

 

Accuracy: ±5 % for delivery rates of 0.5 to 10 ml/h and a measuring time

 

of 2 hours up to 1.2 bar counter-pressure

 

 

(calibrated for 30 ml Fresenius heparin syringes)

 

With delivery rates of <1.0 ml/h the tolerance may exceed the specified

 

±5 %.

 

 

Stop time: 0 minutes up to 2 hours.

 

 

Resolution: 1 minute

 

 

Bolus injection: 1.0 up to 20.0 ml

 

 

Resolution: 0.1 ml

 

 

30 ml Fresenius heparin syringe

 

Audible alarm

Setting range of the loudness of the audible alarm:

 

Factory setting ≥ 65 db (adjustable)

 

 

Minimum setting: ≥ 65 db

 

DIASAFE®plus (option)

Filter life: maximum 12 weeks.

Monitored by the dialysis system and a warning (Filter change) is displayed.

3-8

Fresenius Medical Care 5008 SM 1/08.04

Chapter 3: Specifications

When using ONLINEplus™ (option): Filter life: maximum 100 treatments.

Monitored by the dialysis system and a warning (Filter change) is displayed. If the warning is ignored, ONLINEplus™ will be disabled after the respective number has been exceeded.

After 90 treatments the number of the remaining treatments will be displayed in the cleaning programs.

OCM (option)

Measuring accuracy of the clearance: ± 6 % standard deviation

Shortest measuring interval: 25 min

Time scale of the display: 10 s

ONLINEplus™ (option)

 

 

 

Delivery rate: 25 to 600 ml/min (inside line diameter: 8.0 mm)

 

Resolution: 1 ml/min

 

Exchange volume: substituate goal 500 l adjustable in relation to

 

treatment parameters

 

Accuracy: < 5 % (without lines)

 

(This specification only applies to the range from 30 to 350 ml/min. With

 

delivery rates of < 30 ml/min the deviation may be greater.)

 

Volume counter display: 0.1 to 210 liters

 

Resolution: 0.1 liter

 

Spring-loaded rollers, fully occluding, pressure-limited to < 1.3 bar.

 

(The blood pump design allows manual operation, hand crank in the

 

rotor, in clockwise direction only.)

 

Auto sub: The sub rate is determined as a function of:

 

UF rate

 

Blood flow

 

Hematocrit (HCT)

 

Total protein (TP)

 

Filter performance

Single Needle (option)

 

 

Blood pump

During Single Needle operation 180 seconds.

stop alarm

 

 

Single Needle pump

 

 

Stroke volume

10 to 50 ml in increments of 5 ml

External compliance

50 ml or 60 ml stroke volume

chamber

 

 

Fresenius Medical Care 5008 SM 1/08.04

3-9

Chapter 3: Specifications

Auto SN

+20 % (programmable in the Operator setup.)

Delivery rate of the

 

Single Needle pump

 

BPM (option)

 

Blood pressure

Display Area

 

– Systole: 30 mmHg to 280 mmHg

 

– Diastole: 10 mmHg to 240 mmHg

 

– MAP: 20 mmHg to 255 mmHg

 

Resolution: 1 mmHg

 

Accuracy of measured value ±3 mmHg

Pulse

Display range: 20 to 245 1/min

 

Resolution: 1/min

BTM (option)

 

Required blood flow for

≥ 120 ml/min

accurate BTM function

(The measuring and control functions of the BTM are deactivated if the

 

blood flow is < 100 ml/min.)

Temperature

Accuracy of the fistula temperatures (if correct ambient temperature is

measurement

indicated): ± 0.5 °C

 

Error in fistula temperatures per °C error of the set ambient temperature

 

0.08 °C (at a blood flow of 100 ml/min)

 

0.03 °C (at a blood flow of 300 ml/min)

 

Body temperature change accuracy: ± 0.2 °C

Recirculation

Accuracy of recirculation measurement

measurement

(for 2.5 °C venous bolus amplitude): ± 2 %

 

Maximum bolus amplitude: – 3 °C or + 3 °C

 

Maximum duration of the bolus: up to 10 min

 

Maximum dialysate temperature range used by the BTM:

 

33.5 °C to 39.5 °C

Body temperature control

Allowed range of desired values for body temperature change rate:

 

– 0.5 °C/h to + 0.5 °C/h

 

Maximum dialysate temperature range used by the BTM:

 

33.5 °C to 39.5 °C

3-10

Fresenius Medical Care 5008 SM 1/08.04

Chapter 3: Specifications

Network (option)

Caution

The responsible organization of the network is responsible for protecting the machine from excessive network load (e.g. by accumulation of broadcast messages or port scans). If necessary, the connection to the network must be established via a router or a firewall, for example.

The system configurator is responsible for the further secure data processing, e.g. in PC software applications.

The responsible organization of the network is responsible for the protection of the not encrypted, transferred data.

The data transfer of alarm states via the network must not be used as an external alarm alert (nurse call).

Fresenius Medical Care 5008 SM 1/08.04

3-11

Chapter 3: Specifications

3-12

Fresenius Medical Care 5008 SM 1/08.04

Chapter 4: Installation

4 Installation

4.1Preface

Instructions for all technicians who are authorized to commission our hemodialysis systems.

We, as manufacturers, permanently aim at delivering systems of highest quality.

To reach this aim, we need your support.

Please commission our hemodialysis systems by uniformly using the enclosed "initial start-up report" and enter the values determined in the columns provided.

The following is applicable:

Corrections are necessary only if the measured values are outside of the tolerances specified!

We will then evaluate the initial start-up reports, which will enable us to monitor the quality of our systems on their delivery.

After initial start-up, please asap send – by mail or by fax – the completed form (Initial Start-Up Report) back to the following address:

Fresenius Medical Care Deutschland GmbH Werk Schweinfurt

Herrn Alfred Laus, Abt. BM Hafenstraße 9

97424 Schweinfurt

Fax: 09721/ 678450

Thank you very much for your help!

4.2Important Information on Initial Start-Up

This technical document is intended for initial start-up only. It is not intended for restarting hemodialysis systems that have been shut down or have been put out of service temporarily.

The initial start-up must be performed by the Technical Service of Fresenius Medical Care or a person authorized by them!

Any information on initial start-up and the specifications in the Operating Instructions must be observed.

When bringing the hemodialysis system from a cooler to a warmer room, allow approx. 2 hours for the system to adjust to the ambient temperature before turning the unit on.

Fresenius Medical Care 5008 SM 1/08.04

4-1

Chapter 4: Installation

4.3Initial Start-Up Report

5008

Initial Start-Up Report

Fresenius Medical Care

Technician's name:

 

Service report number:

 

 

 

Customer/Customer no.:

 

 

 

 

 

Inventory no.:

Device no.

Operating hours:

 

 

 

Device type including option(s):

 

 

 

 

 

4-2

Fresenius Medical Care 5008 SM 1/08.04

Chapter 4: Installation

No.

Description

Measure-

 

 

 

ment

 

 

 

value

 

 

 

 

 

1

Preparation

 

 

 

 

 

 

1.1

Hemodialysis system without visible shipping damage.

 

 

 

 

 

 

1.2

Remove the transport protection for the monitor support arm.

 

 

 

Install the IV pole.

 

 

 

 

 

 

1.3

Connect the water supply tubing.

 

 

 

Connect the drain and the flush tubings.

 

 

 

Protect the tubings from slipping out.

 

 

 

Standard: Length 3 m, internal diameter 6 mm

 

 

 

Tubing dimensions, adjusted: Length _____ m, internal diameter_____ mm

 

 

 

 

 

 

1.4

Connect the CDS tubings and protect them from slipping out.

 

 

 

 

 

 

 

Apply a shrink tube marking for the CDS tubings.

 

 

 

 

 

 

 

CDS 1

 

 

 

 

 

 

 

CDS 2

 

 

 

 

 

 

 

BIC

 

 

 

 

 

 

1.5

Remove the shipping plugs from the overflows.

 

 

 

 

 

1.6

When turning the hemodialysis system on, perform an audible check of the watchdog alarm.

 

 

 

 

 

1.7

Rinse out the anti-freeze.

 

 

 

 

 

 

1.8

Select the Filter change program.

 

 

 

Connect the filter. DIASAFE®plus/ONLINEplus

 

 

 

In the service mode, delete mandatory disinfection.

 

 

 

Then completely run the rinse program.

 

 

 

 

 

 

2

SETUP settings (Technician's SETUP/Operator SETUP)

 

 

 

 

 

 

2.1

Check the SETUP on the hemodialysis system.

 

 

 

 

 

 

3

Check – water inlet flow / adjustment – degassing

 

 

 

 

 

 

3.1

Check MaxWaterFlow.

________

Corr.:

 

Desired value: 1300 ml/min to 1550 ml/min

 

Yes

 

 

 

No

 

 

 

 

3.2

Perform the degassing adjustment.

 

 

 

 

 

 

4

Check – dialysate flow

 

 

 

 

 

 

4.1

Check flow at 800 ml/min.

________

 

 

Desired value: 770 ml/min to 830 ml/min

 

 

 

 

 

 

Fresenius Medical Care 5008 SM 1/08.04

4-3

Chapter 4: Installation

No.

Description

Measure-

 

 

 

ment

 

 

 

value

 

 

 

 

 

5

Check – temperature

 

 

 

 

 

 

5.1

Check PT07 (temperature) at 37 °C. (flow 500 ml/min)

________

Corr.:

 

Desired value: 36.8 °C to 37.2 °C (display on hemodialysis system)

 

Yes

 

Measure the reference temperature with an external measuring instrument.

________

No

 

 

 

Difference = Reference temperature minus PT07

 

 

 

Desired value – difference: –0.5 °C to +0.2 °C

 

 

 

 

 

 

6

Check – conductivity

 

 

 

 

 

 

6.1

Check CD7 (conductivity).

________

Corr.:

 

Desired value: approx. 13.5 mS/cm to approx. 14.5 mS/cm

 

Yes

 

Measure the reference conductivity with an external measuring instrument.

________

No

 

 

 

Difference = Reference conductivity minus CD7

 

 

 

Desired value – difference: ±0.2 mS/cm

 

 

 

 

 

 

7

Check – blood leak detector

 

 

 

 

 

 

7.1

Check the blood leak:

________

Corr.:

 

Desired value: 4.8 V to 5.2 V

 

Yes

 

 

 

No

7.2

Check the dimness:

________

 

 

Desired value: 4.8 V to 5.2 V

 

 

 

 

 

 

4-4

Fresenius Medical Care 5008 SM 1/08.04

 

 

Chapter 4:

Installation

 

 

 

 

 

 

 

 

 

 

No.

Description

Measure-

 

 

 

 

ment

 

 

 

 

value

 

 

 

 

 

 

 

8

Check – dialysate pressure

 

 

 

 

 

 

 

 

8.1

Zero point S03/S07

 

 

Corr.:

 

 

 

 

Yes

 

Reference measuring instrument: 0 mbar

________

 

 

 

No

 

 

 

 

 

Check S03.

________

 

 

 

Desired value: +16 mbar to +76 mbar

 

 

 

 

 

 

 

 

 

Check S07.

________

 

 

 

Desired value: +16 mbar to +76 mbar

 

 

 

 

 

 

 

 

8.2

Slope S03/S07 (+)

 

 

 

 

 

 

 

 

 

Reference measuring instrument: +533 mbar (± 26 mbar)

________

 

 

 

 

 

 

 

 

Check S03.

________

 

 

 

Desired value: S03 = Display of reference measuring instrument + (+16 mbar

 

 

 

 

to +76 mbar)

 

 

 

 

 

 

 

 

 

Check S07.

________

 

 

 

Desired value: S07 = Display of reference measuring instrument + (+16 mbar

 

 

 

 

to +76 mbar)

 

 

 

 

 

 

 

 

8.3

Slope S03/S07 (–)

 

 

 

 

 

 

 

 

 

Reference measuring instrument: –533 mbar (± 26 mbar)

________

 

 

 

 

 

 

 

 

Check S03.

________

 

 

 

Desired value: S03 = Display of reference measuring instrument + (+16 mbar

 

 

 

 

to +76 mbar)

 

 

 

 

 

 

 

 

 

Check S07.

________

 

 

 

Desired value: S07 = Display of reference measuring instrument + (+16 mbar

 

 

 

 

to +76 mbar)

 

 

 

 

 

 

 

 

Fresenius Medical Care 5008 SM 1/08.04

4-5

Chapter 4: Installation

No.

Description

 

Measure-

 

 

 

 

ment

 

 

 

 

value

 

 

 

 

 

 

9

Check – electrical safety

 

 

 

 

In Germany according to DIN VDE 0751-1, edition 10/2001.

 

 

 

In other countries, observe the local regulations!

 

 

 

 

 

 

 

 

9.1

Visual inspection performed.

 

 

 

 

 

 

 

9.2

Protective earth resistance maximum 0.3 ohms (with power cord)

______ Ω

 

 

 

 

 

 

9.3

Leakage current measurement (device leakage current)

 

 

 

 

Differential current measurement according to figure C.6

 

 

 

or

 

 

 

 

Direct measurement according to figure C.5

 

 

 

 

Nominal voltage of power supply:

__________ V

 

 

 

Device leakage current mains polarity 1

__________ µA

 

 

 

with line voltage

__________ V

 

 

 

scaled to nominal voltage (maximum 500 µA, see Additional requirements)

_____ µA

 

 

Device leakage current mains polarity 2

__________ µA

 

 

 

with line voltage

__________ V

 

 

 

scaled to nominal voltage (maximum 500 µA, see Additional requirements)

_____ µA

 

 

Test equipment used: ________________________

 

 

 

 

 

 

 

10

Check – zero point arterial/venous pressure display and venous clamp

 

 

 

 

 

 

 

10.1

Zero point arterial pressure display

 

 

Corr.:

 

Check the arterial pressure display (standby operation).

 

________

Yes

 

 

No

 

Desired value: –5 mmHg to +5 mmHg

 

 

 

 

 

 

 

 

 

 

 

10.2

Zero point venous pressure display

 

 

Corr.:

 

Check the venous pressure display (standby operation).

________

Yes

 

No

 

Desired value: –5 mmHg to +5 mmHg

 

 

 

 

 

 

 

 

 

 

 

10.3

Check – venous clamp:

 

 

 

 

A pressure change within 3 minutes must not exceed the following values:

 

 

 

Arterial pressure display

 

 

 

 

Maximum pressure change: ±5 mmHg

 

 

 

 

Reference measuring instrument for pressure display

 

 

 

 

Maximum pressure drop: –0.1 bar

 

 

 

 

 

 

 

 

4-6

Fresenius Medical Care 5008 SM 1/08.04

 

 

Chapter 4:

Installation

 

 

 

 

 

 

 

 

 

 

No.

Description

Measure-

 

 

 

 

ment

 

 

 

 

value

 

 

 

 

 

 

 

11

Final check

 

 

 

 

 

 

 

 

11.1

Check the error memory.

 

 

 

 

 

 

 

 

11.2

Save calibration data and SETUP settings on a data disk.

 

 

 

 

 

 

 

 

11.3

Perform the T1 test.

 

 

 

 

 

 

 

 

11.4

Run the disinfection program

 

 

 

 

(with Puristeril 340 or Puristeril plus or Diasteril or Citrosteril).

 

 

 

 

 

 

 

 

11.5

Check the alarm function during the disinfection program.

 

 

 

 

Open the shunt interlock.

 

 

 

 

Audible alarm and traffic light

 

 

 

 

 

 

 

 

11.6

Check absence of disinfectant by means of test strips (not with Citrosteril).

 

 

 

 

 

 

 

 

11.7

Record entries in the medical device register and on the machine card.

 

 

 

 

 

 

 

11.8

Operating Instructions and accessories package complete and appropriate for the system.

 

 

 

 

 

 

 

Fresenius Medical Care 5008 SM 1/08.04

4-7

Chapter 4:

Installation

 

 

 

 

 

 

 

 

 

 

 

 

Date:

 

Signature:

Stamp:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The system has been released for the intended use.

Yes

No

 

 

 

 

 

 

 

 

Test equipment used:

________________________

Temperature, conductivity, pressure

(type, serial number):

________________________

Protective earth resistance, leakage current

________________________

(type, serial number):

________________________

 

 

 

 

 

 

 

 

Comments:

 

 

 

 

 

 

 

 

 

 

 

 

Date:

 

Signature:

Stamp:

 

 

 

 

 

 

4-8

Fresenius Medical Care 5008 SM 1/08.04

Loading...
+ 138 hidden pages