Service Manual
Hemodialysis System
5008
Edition: 1/08.04 |
|
Part number: M35 179 1 |
0123 |
Fresenius Medical Care
Page
Table of Contents
1Index
2Important Information
2.1 |
Organization of the Technical Document................................................................................ |
2-1 |
2.2 |
How to Use the Technical Document ...................................................................................... |
2-1 |
2.3 |
Precautions for Working on the System ................................................................................. |
2-2 |
2.4 |
Addresses .................................................................................................................................. |
2-2 |
3Specifications
4Installation
|
4.1 |
Preface........................................................................................................................................ |
4-1 |
|
4.2 |
Important Information on Initial Start-Up ................................................................................ |
4-1 |
|
4.3 |
Initial Start-Up Report ............................................................................................................... |
4-2 |
|
4.4 |
Explanations on the Initial Start-Up Report ............................................................................ |
4-9 |
5 |
Setup |
|
|
|
5.1 |
Operator Setup .......................................................................................................................... |
5-1 |
|
5.2 |
Technician's SETUP ................................................................................................................ |
5-13 |
|
5.3 |
Information Regarding the Setting of Concentrates in the Technician’s Setup................ |
5-21 |
6 TSC / TMC / Maintenance
6.1 |
Important Information ............................................................................................................... |
6-1 |
6.2Test Report – Technical Safety Checks, Technical Measurements Checks and Maintenance Procedures6-3
6.3Explanations on Technical Safety Checks, Technical Measurement Checks and Maintenance Procedures6-11
6.4 |
TSC / TMC Report .................................................................................................................... |
6-27 |
Fresenius Medical Care 5008 SM 1/08.04 |
i |
Kapitel :
7Error Messages
8Tools (Service Equipment)
9Calibration / Adjustment
10Repair
11Functional Description
11.1 |
Overall System......................................................................................................................... |
11-1 |
11.2 |
Overview of P.C.B.s................................................................................................................. |
11-3 |
11.3 |
Monitor...................................................................................................................................... |
11-4 |
11.4 |
EBM (Extracorporeal Blood Module) ..................................................................................... |
11-6 |
11.5 |
Hydraulics Unit ........................................................................................................................ |
11-8 |
11.6 |
Power Supply Unit ................................................................................................................. |
11-10 |
11.7 |
Pneumatic Unit....................................................................................................................... |
11-12 |
11.8 |
Hydraulics Unit ...................................................................................................................... |
11-14 |
12 Service Program (Option)
ii |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 1: Index
How to use the index: E.g., index entry 1-3 is to be interpreted as: Chapter 1, page 3
A |
R |
|
Addresses 2-2 |
Reinfusion 5-2 |
|
Alarm processing 5-4 |
T |
|
AutoFlow 3-6 |
||
B |
Temperature 3-6 |
|
U |
||
Blood pump 5-1 |
||
BPM 5-10 |
Ultrafiltration 5-3 |
|
BTM 5-11 |
||
User interface 5-5 |
||
|
||
C |
|
|
Concentrates 3-6 |
|
|
D |
|
|
DIASAFE®plus 3-8 |
|
|
E |
|
|
EcoFlow 3-6 |
|
|
Emergency 5-8 |
|
|
External connections 3-3 |
|
|
F |
|
|
Flow diagram 11-14 |
|
|
H |
|
|
Heparin pump 3-8 |
|
|
O |
|
|
OCM 3-9, 5-9 |
|
|
ONLINEplus™ 5-9 |
|
|
Operating programs 3-4 |
|
|
Operator Setup 5-1 |
|
|
Optical detector 3-7 |
|
|
Override conditions 3-4 |
|
|
P |
|
|
Patient card 5-8 |
|
Fresenius Medical Care 5008 SM 1/08.04 |
1-1 |
Kapitel 1: Index
1-2 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 2: Important Information
Page identification |
Page number 1-3 is to be interpreted as: Chapter 1, page 3. |
Document changes |
Document changes will be released as new editions or supplements. In |
|
general: this manual is subject to change without notice. |
Editorial information |
The current edition of this technical document is: |
|
1/08.04 = 1st edition, August 2004 |
Intended use |
This technical document is intended for service technicians and is to be |
|
used for first studies (to acquire a basic knowledge) and for reference |
|
purposes (for TSC, maintenance and repair). The study of this |
|
document, however, does not replace the training courses offered by |
|
the manufacturer. |
Requirements |
Knowledge of the current Operating Instructions of the respective |
|
system. |
|
Background experience in mechanics, electrical and medical |
|
engineering. |
Note and Caution symbols |
|
|
Explanation of the Note and Caution symbols used: |
|
|
|
Note |
|
Informs the operator that in case of a failure to follow the steps as |
|
described, a specific function will be executed incorrectly or will not be |
|
executed at all, or will not produce the desired effect. |
|
|
|
|
|
Caution |
|
Advises the operator against certain procedures or actions that could |
|
cause damage to the equipment or may have adverse effects on |
|
operators and patients. |
|
|
Fresenius Medical Care 5008 SM 1/08.04 |
2-1 |
Chapter 2: Important Information
Authorized persons |
Assembly, extensions, adjustments, modifications or repairs may only |
|
be carried out by the manufacturer or persons authorized by him. |
Measuring equipment and |
The activities described in this technical document require the |
accessories |
availability of the necessary technical measuring equipment and |
|
accessories. |
Precautions |
Before turning power on, repair any visible damage. |
|
Prior to opening the system and when working on the open system, the |
|
following precautions have to be taken: |
|
– Protect the components against ingress of liquids. |
|
– Do not touch live parts. |
|
– All plugs, connections and components may only be disconnected or |
|
connected if de-energized. |
ESD precautions |
When repairing the system and replacing spare parts, observe |
|
applicable ESC precautions (e.g. EN 100 015-1). |
Monitor support arm |
If the 5008 hemodialysis system is to be placed in a horizontal position |
|
for servicing, the monitor support arm must be protected with the |
|
transport protection to prevent it from flipping over. |
To be observed after |
A disinfection and a T1 test must be performed after working on the |
working on the system |
system. |
To be observed after |
After a disinfection program has been aborted or if the system is to be |
aborting a disinfection |
preserved, the hemodialysis system must be disconnected from the |
program |
water supply after a maximum of 3 days. When the system is returned |
|
to use, check that the pressure of the water supply meets the prescribed |
|
minimum pressure. |
Please address any inquires to:
Fresenius Medical Care AG 61346 Bad Homburg Germany
Phone: + 49 6172 609-0 www.fmc-ag.com
2-2 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 2: Important Information
Service |
Fresenius Medical Care |
Central Europe |
Deutschland GmbH |
|
Geschäftsbereich Zentraleuropa |
|
Kundendienst / Servicecenter |
|
Steinmühlstraße 24 I |
|
61352 Bad Homburg |
|
Germany |
|
Phone: +49 6172 609-7100 |
|
Fax: +49 6172 609-7102 |
|
E-mail: ServicecenterD@fmc-ag.com |
Service |
Fresenius Medical Care |
International |
Deutschland GmbH |
|
Service Support International |
|
Hafenstraße 9 |
|
97424 Schweinfurt |
|
Germany |
|
Phone: +49 9721 678-333 (hotline) |
|
Fax: +49 9721 678-130 |
Local Service |
|
|
|
Fresenius Medical Care 5008 SM 1/08.04 |
2-3 |
Chapter 2: Important Information
2-4 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 3: Specifications
Dimensions, weight and housing material
Dimensions |
Height: approx. 162 cm (approx. 210 cm incl. IV pole) |
|
Width: approx. 48 cm (on base incl. brake) |
|
Depth: approx. 72 cm (approx. 86 cm with extended concentrate rack) |
Weight |
Approx. 135 kg (without options) |
Housing material |
PU vacuum cast resin |
Type label |
|
1
2
3 6
4 |
5 |
|
1Type identification, serial number
2Power requirements
3Protection against ingress of liquids: drip-proof
4Caution, consult accompanying documents
5Degree of protection against electric shock: Type B
6CE mark
Electrical safety (classification according to EN 60601-1, IEC 601-1)
Type of protection against |
Safety class I |
electric shock |
|
Degree of protection |
Type B, symbol: |
against electric shock |
|
Fresenius Medical Care 5008 SM 1/08.04 |
3-1 |
Chapter 3: Specifications
Applicable only to the BPM blood pressure cuff: Degree of protection against electric shock
Degree of protection against ingress of liquids
Leakage currents
EMC specifications according to EN 60601-1-2 (IEC 601-1-2)
Electric supply
Line voltage
Connection to power supply
Operating current dialysis
Power supply (internal)
Battery
Fuses
Main power switch
Operating conditions
Water inlet pressure
Water inlet temperature
Water inlet rate
Type CF, symbol:
Drip-proof, symbol: IPX1
According to EN 60601-1
RFI emissions:
Limit class A according to EN 55011, Group 1
Immunity:
–Electrostatic discharge, atmospheric discharge: 8 kV
–Electromagnetic fields: 27 MHz – 1000 MHz: 3 V/m
–Bursts: Power line (alternating current): 1 kV
–Surge voltages (alternating current): 2 kV
100 to 230 V AC, ±10 %, 47 to 63 Hz
(The decisive criterion is the line voltage and the operating current specified on the type label of the system)
16 A at 230 V, regulation according to VDE 0100 part 0107
Approx. 6 A, (at 230 V)
at a water inlet temperature of 17 °C Dialysate temperature 37 °C Dialysate flow: 500 ml/min
+24 V ± 3 %, 20 A short-circuit proof +18 V ± 3 %, 14 A short-circuit proof 480 W total power output
Lead-acid battery (maintenance-free) 24 V, 7 Ah
2 x G 16 A (miniature circuit-breaker) rear of power supply unit
1.5 to 6.0 bar
5 °C to 30 °C
with "Integrated hot rinse": 85 °C to 95 °C
1.5 l/min; at an inlet pressure of 1.5 bar
3-2 |
Fresenius Medical Care 5008 SM 1/08.04 |
|
Chapter 3: Specifications |
|
|
Water drain |
0 to 100 cm above the floor, minimum 5 cm free fall. The water drain |
|
must be located at a lower level than the dialyzer position. |
Concentrate supply |
0 to -100 mbar; maximum suction height 1 m |
|
with Central Delivery System (option): 0.05 to 2.0 bar |
Heat dissipation |
Dialysis: |
|
approx. 400 Watt (at an ambient temperature of 20 °C) |
Range of operating |
15 °C to 35 °C |
temperature |
|
Atmospheric pressure |
700700 hPa to 1060 hPa |
Relative humidity |
30 % to 75 %, temporarily 95 % |
Stability |
5° |
IV pole load capacity |
Maximum: 5 kg |
|
Maximum load capacity of one hook: 5 kg |
External connection options |
|
|
|
|
Caution |
|
Any additional equipment connected to the analog and digital interfaces |
|
of the machine must comply with the applicable EN specifications (e.g. |
|
EN 60950 for data processing equipment and EN 60601 (IEC 601) for |
|
electro-medical equipment). |
|
Apart from this requirement, all configurations must comply with the |
|
system standard EN 60601-1-1 (IEC 601-1-1), or their applicability with |
|
regard to safety has to be proven by a certificate issued by a testing |
|
agency authorized to test the ready-for-use machine. |
|
The connection of additional equipment to the signal input or output |
|
component affects the system configuration and anyone connecting |
|
additional equipment is therefore responsible for compliance with the |
|
system standard EN 60601-1-1 (IEC 601-1-1). |
|
|
|
|
|
Caution |
|
The external alarm indicators do not relieve the operator of the |
|
obligation to observe the local alarms of the system. |
|
|
LAN |
Interface for the exchange of data. |
|
Electrically isolated by transformer. |
|
Port: RJ 45 |
RS232 |
Interface for the exchange of data. |
|
Electrically isolated by optocoupler. |
|
Port: DSUB 9-pin |
Fresenius Medical Care 5008 SM 1/08.04 |
3-3 |
Chapter 3: Specifications
Service/diagnostics |
(Protected by cover!) |
|
For inhouse computer diagnostics. |
|
Port: DSUB 15-pin |
24 V |
(Protected by cover!) |
|
24 V connection (2 A fuse) |
|
Port: Flanged socket, 4-pin |
Alarm output |
For the connection of an external alarm indicator (nurse call). (Potential- |
|
free alarm output. Alternating contact maximum 24 V/24 W). |
|
Port: 5-pin diode plug via a shielded line; shield grounded on either side. |
Override conditions |
|
|
When overriding a safety system the responsibility for the patient’s |
|
safety rests with the operator of the machine. |
Audible alarm suppression |
Mute alarm time: maximum 2 minutes |
|
(adjustable in the SETUP) |
Alarm override |
After confirmation of the error message and start of the blood systems: |
|
Arterial and venous pressure alarm for approx. 10 seconds (window |
|
inactive) |
|
Air detector alarm for approx. 2 seconds |
Blood leak override |
Override time: maximum 2 minutes |
Override air-bubble |
Override time: after starting removal of air: approx. 4 seconds |
detector |
|
Operating programs |
|
T1 test |
Automatic test for verification of the operating and safety systems. |
|
The T1 test is mandatory, |
|
– after power on (not following a power failure) |
|
– after a cleaning program |
Preparation |
Defined by the optical detector located below the venous bubble |
|
catcher. |
|
Preparation is terminated as soon as the optical detector senses |
|
opaque fluid in the blood lines. |
Priming and rinsing the |
Minimum rinse volume 500 ml; automatic switching to rinsing, if level in |
blood lines |
bubble catcher detected. Automatic raising of the fluid level during the |
|
rinse phase. |
Reinfusion |
Reinfusion volume adjustable in the SETUP. |
|
Return to dialysis still possible. |
Dialysis |
Bicarbonate dialysis |
ISO-UF |
Ultrafiltration without dialysate flow (Bergström method) |
3-4 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 3: Specifications
Cleaning programs |
Rinse clear/rinse/mandatory rinse: |
||
|
Time adjustable in the TECHNICIAN's SETUP, |
||
|
Temperature: approx. 37 °C, |
||
|
Flow: 600, 800 ml/min (adjustable in the SETUP) |
||
|
Cold disinfection/degreasing, cold disinfection: |
||
|
Time adjustable in the TECHNICIAN's SETUP, |
||
|
Temperature: approx. 37 °C, |
||
|
Flow: max. 900 ml/min |
||
|
Heat disinfection: |
||
|
Time adjustable in the TECHNICIAN's SETUP, |
||
|
Flow: max. 900 ml/min |
||
|
In all programs: |
||
|
Blood pump stops, arterial and venous line occlusion clamp closed. |
||
|
Progress of the program (time-counting) is interrupted in the event of a |
||
|
flow alarm. |
|
|
|
The cleaning programs can be aborted. |
||
|
The chemical disinfection program is followed by a mandatory rinse. |
||
Flush |
Rinsing of the water supply area |
||
Dialysate circuit and safety systems |
|
|
|
Blood leak detector |
Threshold of response ≤ 0.5 ml blood loss per minute into the dialysate |
||
|
at a hematocrit of 0.25. |
||
|
(flow rate 100 ml/min to 1000 ml/min) |
||
Transmembrane pressure |
Display range: –100 to 400 mmHg |
||
|
Resolution: 5 mmHg |
||
|
Definition: |
|
|
|
TMP |
= |
Pbo – (Pdi + Pdo) / 2 + Offset |
|
TMP |
= |
Transmembrane pressure |
|
Pbo |
= |
Blood pressure on the outlet side of the dialyzer |
|
Pdi |
= |
Dialysate pressure on the inlet side of the dialyzer |
|
Pdo |
= |
Dialysate pressure on the outlet side of the dialyzer |
|
Offset |
= |
Flow-dependent pressure fluctuations |
Ultrafiltration |
Selectable UF rate: 0 ml/h to 4000 ml/h (in 10 ml increments) |
||
|
Maximum rate internally adjustable to 1, 2, 3, or 4 l/h. |
||
|
Pump volume accuracy: ±1 % (at Pdi > –500 mbar) |
||
|
The UF rate/effective blood flow ratio is being monitored during the |
||
|
treatment. If an incongruity occurs a warning will be displayed after |
||
|
approx. 10 seconds. |
||
Pressure holding test |
Event-controlled |
Fresenius Medical Care 5008 SM 1/08.04 |
3-5 |
Chapter 3: Specifications
Balancing |
Accuracy: ±0.1 % related to the total dialysate volume |
||
Maximum balancing error |
F |
= |
FUF + FBil |
|
|||
|
F |
= |
Maximum balancing error |
|
FUF |
= |
Ultrafiltration error |
|
Fbil |
= |
Balancing error |
|
Example: |
|
|
|
Ultrafiltration error: with 1000 ml in 1 hour: ±1 % = ±10 ml/h |
||
|
Balancing error: with 30 l fluid flow in 1 hour at a dialysate flow of |
||
|
500 ml/min: ±0.1 %= ± 30 ml/h |
||
|
Maximum balancing error: |
||
|
F = FUF + FBil = (±10 ml/h) + (±30 ml/h) = ±40 ml/h |
||
Degassing |
Method: Negative pressure |
||
Dialysate concentration |
Display range: 12.8 to 15.7 mS/cm |
||
(conductivity) |
Resolution: 0.1 mS/cm |
||
|
Accuracy: 0.1 mS/cm |
||
|
Method: |
|
|
|
Temperature-compensated electronic conductivity meter with |
||
|
adjustable alarm limits. |
||
Concentrates |
Entering concentration types |
||
|
Adjustment range: 125 to 151 mmol/l, depending on the concentrate |
||
|
used ±10 % of the base value. |
||
|
Bicarbonate readjustment range: corresponds to ±8 mmol/l |
||
bibag® |
Bicarbonate concentrate preparation from the bibag® |
||
|
Temperature range: 15 to 35 °C |
||
Dialysate temperature |
Adjustment range: (prescribed temperature) 34.0 °C to 39.0 °C |
||
|
Resolution: 0.5 °C |
||
|
Measuring accuracy: ±0.2 °C |
||
Dialysate flow |
Display range: 100 to 1000 ml/min |
||
|
Resolution: 100 ml/min |
||
|
Desired values: 100 to 1000 ml/min |
||
|
Measurement by means of time pulse monitoring and balancing |
||
|
chamber volume |
||
|
Auto flow: dialysate flow controlled in relation to the blood flow, |
||
|
determined by the dialyzer. |
||
|
EcoFlow: dialysate flow automatically reduced to 100 ml/min in |
||
|
Preparation |
||
Rinse and chemical |
Desired temperature: 37 °C |
||
disinfection temperature |
Resolution: 0.5 °C |
||
|
Measuring accuracy: ±0.2 °C |
||
Rinse and chemical |
Desired value: 600 ml/min |
||
disinfection flow |
|
|
|
3-6 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 3: Specifications
Hot rinse and heat |
Desired temperature: 85 °C |
disinfection temperature |
Resolution: 0.5 °C |
|
Measuring accuracy: ±2.0 °C |
Hot rinse and heat |
Desired value: 600 ml/min |
disinfection flow |
|
Concentration of |
Dilution: Disinfectant is diluted with purified water in the dialysis system |
disinfectant |
at a ratio of 1+24. |
Flow alarm |
Dependent on the programmed flow |
Extracorporeal blood circuit and safety systems |
|
Arterial pressure |
Display range: –300 to +300 mmHg |
measurement |
Resolution: 5 mmHg |
|
Accuracy: 7 mmHg (typical) |
|
OD senses non-opaque presence: |
|
Alarm window width: –300 to +300 mmHg |
|
OD senses opaque presence: |
|
Alarm window width: +40 to +200 mmHg |
|
Default value adjustable in the SETUP, factory setting 120 mmHg |
Blood pump |
Delivery rate: 30 to 600 ml/min |
|
Resolution: 10 ml/min (with a line diameter of 8 mm) |
|
Accuracy: < 5 % (without lines) |
|
Line diameter: 4.4 mm, 6.4 mm, 8.0 mm |
|
Blood pump stop alarm: 60 seconds |
|
Spring-loaded rollers, fully occluding, pressure-limited to 2 bar with 8 x |
|
2.1 pump line segment (when using the prescribed tubing systems). |
|
(The blood pump design allows manual operation, hand crank in the |
|
rotor, in clockwise direction only.) |
Venous pressure |
Display range: –100 to +500 mmHg |
measurement |
Resolution: 5 mmHg |
|
Accuracy: 7 mmHg (typical) |
|
OD senses non-opaque presence: |
|
Alarm window width: –100 to +500 mmHg |
|
OD senses opaque presence: |
|
Alarm window width: 40 to 200 mmHg |
|
Default value adjustable in the SETUP, |
|
Factory setting 120 mmHg |
|
adjustable over a range of 20 to 500 mmHg |
|
(adjustable from -100 to 500 mmHg via SETUP.) |
Fill level detector |
Method: |
|
Capacitive measurement |
|
Switching point 13 mm, ±4 mm from upper edge |
Optical detector |
Method: |
|
Infrared transmission |
Fresenius Medical Care 5008 SM 1/08.04 |
3-7 |
Chapter 3: Specifications
|
Distinguishes between |
|
|
OD light (saline or air in the tubing system) |
|
|
OD dark (blood in the tubing system). |
|
Air bubble detector |
Method: |
|
|
Ultrasonic transmission measurement on the line |
|
|
Sensitivity: |
|
|
– Air bubbles: |
|
|
Bubble volume ≥ 20 µl |
|
|
– Blood foam (air-blood mixture) |
|
|
Air alarm: |
|
|
– BP rate < 100 ml/min: |
|
|
Air bubble: Volume ≥ 20 µl |
|
|
Blood foam |
|
|
– BP rate ≥ 100 ml/min: |
|
|
10 air bubbles with an air bubble volume of |
< 50 µl each |
|
or 1 air bubble with an air bubble volume of |
≥ 50 µl, |
|
Blood foam |
|
|
The specified data refer to the most unfavorable case with a BP rate of |
|
|
0 to 600 ml/min when using the blood lines specified in chapter |
|
|
Consumables. |
|
Heparin pump |
Delivery rate: 0.5 to 10 ml/h |
|
|
Resolution: 0.1 ml/h |
|
|
Accuracy: ±5 % for delivery rates of 0.5 to 10 ml/h and a measuring time |
|
|
of 2 hours up to 1.2 bar counter-pressure |
|
|
(calibrated for 30 ml Fresenius heparin syringes) |
|
|
With delivery rates of <1.0 ml/h the tolerance may exceed the specified |
|
|
±5 %. |
|
|
Stop time: 0 minutes up to 2 hours. |
|
|
Resolution: 1 minute |
|
|
Bolus injection: 1.0 up to 20.0 ml |
|
|
Resolution: 0.1 ml |
|
|
30 ml Fresenius heparin syringe |
|
Audible alarm |
Setting range of the loudness of the audible alarm: |
|
|
Factory setting ≥ 65 db (adjustable) |
|
|
Minimum setting: ≥ 65 db |
|
DIASAFE®plus (option)
Filter life: maximum 12 weeks.
Monitored by the dialysis system and a warning (Filter change) is displayed.
3-8 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 3: Specifications
When using ONLINEplus™ (option): Filter life: maximum 100 treatments.
Monitored by the dialysis system and a warning (Filter change) is displayed. If the warning is ignored, ONLINEplus™ will be disabled after the respective number has been exceeded.
After 90 treatments the number of the remaining treatments will be displayed in the cleaning programs.
OCM (option)
Measuring accuracy of the clearance: ± 6 % standard deviation
Shortest measuring interval: 25 min
Time scale of the display: 10 s
ONLINEplus™ (option) |
|
|
|
Delivery rate: 25 to 600 ml/min (inside line diameter: 8.0 mm) |
|
|
Resolution: 1 ml/min |
|
|
Exchange volume: substituate goal 500 l adjustable in relation to |
|
|
treatment parameters |
|
|
Accuracy: < 5 % (without lines) |
|
|
(This specification only applies to the range from 30 to 350 ml/min. With |
|
|
delivery rates of < 30 ml/min the deviation may be greater.) |
|
|
Volume counter display: 0.1 to 210 liters |
|
|
Resolution: 0.1 liter |
|
|
Spring-loaded rollers, fully occluding, pressure-limited to < 1.3 bar. |
|
|
(The blood pump design allows manual operation, hand crank in the |
|
|
rotor, in clockwise direction only.) |
|
|
Auto sub: The sub rate is determined as a function of: |
|
|
– |
UF rate |
|
– |
Blood flow |
|
– |
Hematocrit (HCT) |
|
– |
Total protein (TP) |
|
– |
Filter performance |
Single Needle (option) |
|
|
Blood pump |
During Single Needle operation 180 seconds. |
|
stop alarm |
|
|
Single Needle pump |
|
|
Stroke volume |
10 to 50 ml in increments of 5 ml |
|
External compliance |
50 ml or 60 ml stroke volume |
|
chamber |
|
|
Fresenius Medical Care 5008 SM 1/08.04 |
3-9 |
Chapter 3: Specifications
Auto SN |
+20 % (programmable in the Operator setup.) |
Delivery rate of the |
|
Single Needle pump |
|
BPM (option) |
|
Blood pressure |
Display Area |
|
– Systole: 30 mmHg to 280 mmHg |
|
– Diastole: 10 mmHg to 240 mmHg |
|
– MAP: 20 mmHg to 255 mmHg |
|
Resolution: 1 mmHg |
|
Accuracy of measured value ±3 mmHg |
Pulse |
Display range: 20 to 245 1/min |
|
Resolution: 1/min |
BTM (option) |
|
Required blood flow for |
≥ 120 ml/min |
accurate BTM function |
(The measuring and control functions of the BTM are deactivated if the |
|
blood flow is < 100 ml/min.) |
Temperature |
Accuracy of the fistula temperatures (if correct ambient temperature is |
measurement |
indicated): ± 0.5 °C |
|
Error in fistula temperatures per °C error of the set ambient temperature |
|
0.08 °C (at a blood flow of 100 ml/min) |
|
0.03 °C (at a blood flow of 300 ml/min) |
|
Body temperature change accuracy: ± 0.2 °C |
Recirculation |
Accuracy of recirculation measurement |
measurement |
(for 2.5 °C venous bolus amplitude): ± 2 % |
|
Maximum bolus amplitude: – 3 °C or + 3 °C |
|
Maximum duration of the bolus: up to 10 min |
|
Maximum dialysate temperature range used by the BTM: |
|
33.5 °C to 39.5 °C |
Body temperature control |
Allowed range of desired values for body temperature change rate: |
|
– 0.5 °C/h to + 0.5 °C/h |
|
Maximum dialysate temperature range used by the BTM: |
|
33.5 °C to 39.5 °C |
3-10 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 3: Specifications
Network (option)
Caution
The responsible organization of the network is responsible for protecting the machine from excessive network load (e.g. by accumulation of broadcast messages or port scans). If necessary, the connection to the network must be established via a router or a firewall, for example.
The system configurator is responsible for the further secure data processing, e.g. in PC software applications.
The responsible organization of the network is responsible for the protection of the not encrypted, transferred data.
The data transfer of alarm states via the network must not be used as an external alarm alert (nurse call).
Fresenius Medical Care 5008 SM 1/08.04 |
3-11 |
Chapter 3: Specifications
3-12 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 4: Installation
Instructions for all technicians who are authorized to commission our hemodialysis systems.
We, as manufacturers, permanently aim at delivering systems of highest quality.
To reach this aim, we need your support.
Please commission our hemodialysis systems by uniformly using the enclosed "initial start-up report" and enter the values determined in the columns provided.
The following is applicable:
Corrections are necessary only if the measured values are outside of the tolerances specified!
We will then evaluate the initial start-up reports, which will enable us to monitor the quality of our systems on their delivery.
After initial start-up, please asap send – by mail or by fax – the completed form (Initial Start-Up Report) back to the following address:
Fresenius Medical Care Deutschland GmbH Werk Schweinfurt
Herrn Alfred Laus, Abt. BM Hafenstraße 9
97424 Schweinfurt
Fax: 09721/ 678450
Thank you very much for your help!
4.2Important Information on Initial Start-Up
This technical document is intended for initial start-up only. It is not intended for restarting hemodialysis systems that have been shut down or have been put out of service temporarily.
The initial start-up must be performed by the Technical Service of Fresenius Medical Care or a person authorized by them!
Any information on initial start-up and the specifications in the Operating Instructions must be observed.
When bringing the hemodialysis system from a cooler to a warmer room, allow approx. 2 hours for the system to adjust to the ambient temperature before turning the unit on.
Fresenius Medical Care 5008 SM 1/08.04 |
4-1 |
Chapter 4: Installation
4.3Initial Start-Up Report
5008 |
Initial Start-Up Report |
Fresenius Medical Care
Technician's name: |
|
Service report number: |
|
|
|
Customer/Customer no.: |
|
|
|
|
|
Inventory no.: |
Device no. |
Operating hours: |
|
|
|
Device type including option(s): |
|
|
|
|
|
4-2 |
Fresenius Medical Care 5008 SM 1/08.04 |
Chapter 4: Installation
No. |
Description |
Measure- |
|
|
|
ment |
|
|
|
value |
|
|
|
|
|
1 |
Preparation |
|
|
|
|
|
|
1.1 |
Hemodialysis system without visible shipping damage. |
|
|
|
|
|
|
1.2 |
Remove the transport protection for the monitor support arm. |
|
|
|
Install the IV pole. |
|
|
|
|
|
|
1.3 |
Connect the water supply tubing. |
|
|
|
Connect the drain and the flush tubings. |
|
|
|
Protect the tubings from slipping out. |
|
|
|
Standard: Length 3 m, internal diameter 6 mm |
|
|
|
Tubing dimensions, adjusted: Length _____ m, internal diameter_____ mm |
|
|
|
|
|
|
1.4 |
Connect the CDS tubings and protect them from slipping out. |
|
|
|
|
|
|
|
Apply a shrink tube marking for the CDS tubings. |
|
|
|
|
|
|
|
CDS 1 |
|
|
|
|
|
|
|
CDS 2 |
|
|
|
|
|
|
|
BIC |
|
|
|
|
|
|
1.5 |
Remove the shipping plugs from the overflows. |
|
|
|
|
|
|
1.6 |
When turning the hemodialysis system on, perform an audible check of the watchdog alarm. |
|
|
|
|
|
|
1.7 |
Rinse out the anti-freeze. |
|
|
|
|
|
|
1.8 |
Select the Filter change program. |
|
|
|
Connect the filter. DIASAFE®plus/ONLINEplus™ |
|
|
|
In the service mode, delete mandatory disinfection. |
|
|
|
Then completely run the rinse program. |
|
|
|
|
|
|
2 |
SETUP settings (Technician's SETUP/Operator SETUP) |
|
|
|
|
|
|
2.1 |
Check the SETUP on the hemodialysis system. |
|
|
|
|
|
|
3 |
Check – water inlet flow / adjustment – degassing |
|
|
|
|
|
|
3.1 |
Check MaxWaterFlow. |
________ |
Corr.: |
|
Desired value: 1300 ml/min to 1550 ml/min |
|
Yes |
|
|
|
No |
|
|
|
|
3.2 |
Perform the degassing adjustment. |
|
|
|
|
|
|
4 |
Check – dialysate flow |
|
|
|
|
|
|
4.1 |
Check flow at 800 ml/min. |
________ |
|
|
Desired value: 770 ml/min to 830 ml/min |
|
|
|
|
|
|
Fresenius Medical Care 5008 SM 1/08.04 |
4-3 |
Chapter 4: Installation
No. |
Description |
Measure- |
|
|
|
|
ment |
|
|
|
|
value |
|
|
|
|
|
|
|
5 |
Check – temperature |
|
|
|
|
|
|
|
|
5.1 |
Check PT07 (temperature) at 37 °C. (flow 500 ml/min) |
________ |
Corr.: |
|
|
Desired value: 36.8 °C to 37.2 °C (display on hemodialysis system) |
|
Yes |
|
|
Measure the reference temperature with an external measuring instrument. |
________ |
No |
|
|
|
|||
|
Difference = Reference temperature minus PT07 |
|
|
|
|
Desired value – difference: –0.5 °C to +0.2 °C |
|
|
|
|
|
|
|
|
6 |
Check – conductivity |
|
|
|
|
|
|
|
|
6.1 |
Check CD7 (conductivity). |
________ |
Corr.: |
|
|
Desired value: approx. 13.5 mS/cm to approx. 14.5 mS/cm |
|
Yes |
|
|
Measure the reference conductivity with an external measuring instrument. |
________ |
No |
|
|
|
|||
|
Difference = Reference conductivity minus CD7 |
|
|
|
|
Desired value – difference: ±0.2 mS/cm |
|
|
|
|
|
|
|
|
7 |
Check – blood leak detector |
|
|
|
|
|
|
|
|
7.1 |
Check the blood leak: |
________ |
Corr.: |
|
|
Desired value: 4.8 V to 5.2 V |
|
Yes |
|
|
|
|
No |
|
7.2 |
Check the dimness: |
________ |
||
|
||||
|
Desired value: 4.8 V to 5.2 V |
|
|
|
|
|
|
|
4-4 |
Fresenius Medical Care 5008 SM 1/08.04 |
|
|
Chapter 4: |
Installation |
|
|
|
|
|
|
|
|
|
|
|
No. |
Description |
Measure- |
|
|
|
|
ment |
|
|
|
|
value |
|
|
|
|
|
|
|
8 |
Check – dialysate pressure |
|
|
|
|
|
|
|
|
8.1 |
Zero point S03/S07 |
|
|
Corr.: |
|
|
|
|
Yes |
|
Reference measuring instrument: 0 mbar |
________ |
|
|
|
|
No |
||
|
|
|
|
|
|
Check S03. |
________ |
|
|
|
Desired value: +16 mbar to +76 mbar |
|
|
|
|
|
|
|
|
|
Check S07. |
________ |
|
|
|
Desired value: +16 mbar to +76 mbar |
|
|
|
|
|
|
|
|
8.2 |
Slope S03/S07 (+) |
|
|
|
|
|
|
|
|
|
Reference measuring instrument: +533 mbar (± 26 mbar) |
________ |
|
|
|
|
|
|
|
|
Check S03. |
________ |
|
|
|
Desired value: S03 = Display of reference measuring instrument + (+16 mbar |
|
|
|
|
to +76 mbar) |
|
|
|
|
|
|
|
|
|
Check S07. |
________ |
|
|
|
Desired value: S07 = Display of reference measuring instrument + (+16 mbar |
|
|
|
|
to +76 mbar) |
|
|
|
|
|
|
|
|
8.3 |
Slope S03/S07 (–) |
|
|
|
|
|
|
|
|
|
Reference measuring instrument: –533 mbar (± 26 mbar) |
________ |
|
|
|
|
|
|
|
|
Check S03. |
________ |
|
|
|
Desired value: S03 = Display of reference measuring instrument + (+16 mbar |
|
|
|
|
to +76 mbar) |
|
|
|
|
|
|
|
|
|
Check S07. |
________ |
|
|
|
Desired value: S07 = Display of reference measuring instrument + (+16 mbar |
|
|
|
|
to +76 mbar) |
|
|
|
|
|
|
|
|
Fresenius Medical Care 5008 SM 1/08.04 |
4-5 |
Chapter 4: Installation
No. |
Description |
|
Measure- |
|
|
|
|
ment |
|
|
|
|
value |
|
|
|
|
|
|
9 |
Check – electrical safety |
|
|
|
|
In Germany according to DIN VDE 0751-1, edition 10/2001. |
|
|
|
|
In other countries, observe the local regulations! |
|
|
|
|
|
|
|
|
9.1 |
Visual inspection performed. |
|
|
|
|
|
|
|
|
9.2 |
Protective earth resistance maximum 0.3 ohms (with power cord) |
______ Ω |
|
|
|
|
|
|
|
9.3 |
Leakage current measurement (device leakage current) |
|
|
|
|
Differential current measurement according to figure C.6 |
|
|
|
|
or |
|
|
|
|
Direct measurement according to figure C.5 |
|
|
|
|
Nominal voltage of power supply: |
__________ V |
|
|
|
Device leakage current mains polarity 1 |
__________ µA |
|
|
|
with line voltage |
__________ V |
|
|
|
scaled to nominal voltage (maximum 500 µA, see Additional requirements) |
_____ µA |
|
|
|
Device leakage current mains polarity 2 |
__________ µA |
|
|
|
with line voltage |
__________ V |
|
|
|
scaled to nominal voltage (maximum 500 µA, see Additional requirements) |
_____ µA |
|
|
|
Test equipment used: ________________________ |
|
|
|
|
|
|
|
|
10 |
Check – zero point arterial/venous pressure display and venous clamp |
|
|
|
|
|
|
|
|
10.1 |
Zero point arterial pressure display |
|
|
Corr.: |
|
Check the arterial pressure display (standby operation). |
|
________ |
Yes |
|
|
No |
||
|
Desired value: –5 mmHg to +5 mmHg |
|
|
|
|
|
|
|
|
|
|
|
|
|
10.2 |
Zero point venous pressure display |
|
|
Corr.: |
|
Check the venous pressure display (standby operation). |
________ |
Yes |
|
|
No |
|||
|
Desired value: –5 mmHg to +5 mmHg |
|
|
|
|
|
|
|
|
|
|
|
|
|
10.3 |
Check – venous clamp: |
|
|
|
|
A pressure change within 3 minutes must not exceed the following values: |
|
|
|
|
Arterial pressure display |
|
|
|
|
Maximum pressure change: ±5 mmHg |
|
|
|
|
Reference measuring instrument for pressure display |
|
|
|
|
Maximum pressure drop: –0.1 bar |
|
|
|
|
|
|
|
|
4-6 |
Fresenius Medical Care 5008 SM 1/08.04 |
|
|
Chapter 4: |
Installation |
|
|
|
|
|
|
|
|
|
|
|
No. |
Description |
Measure- |
|
|
|
|
ment |
|
|
|
|
value |
|
|
|
|
|
|
|
11 |
Final check |
|
|
|
|
|
|
|
|
11.1 |
Check the error memory. |
|
|
|
|
|
|
|
|
11.2 |
Save calibration data and SETUP settings on a data disk. |
|
|
|
|
|
|
|
|
11.3 |
Perform the T1 test. |
|
|
|
|
|
|
|
|
11.4 |
Run the disinfection program |
|
|
|
|
(with Puristeril 340 or Puristeril plus or Diasteril or Citrosteril). |
|
|
|
|
|
|
|
|
11.5 |
Check the alarm function during the disinfection program. |
|
|
|
|
Open the shunt interlock. |
|
|
|
|
Audible alarm and traffic light |
|
|
|
|
|
|
|
|
11.6 |
Check absence of disinfectant by means of test strips (not with Citrosteril). |
|
|
|
|
|
|
|
|
11.7 |
Record entries in the medical device register and on the machine card. |
|
|
|
|
|
|
|
|
11.8 |
Operating Instructions and accessories package complete and appropriate for the system. |
|
|
|
|
|
|
|
|
Fresenius Medical Care 5008 SM 1/08.04 |
4-7 |
Chapter 4: |
Installation |
|
|
|
|
|
|
|
|
|
|
|
|
|
Date: |
|
Signature: |
Stamp: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
The system has been released for the intended use. |
Yes |
No |
||
|
|
|
|
|
|
|
|
|
|
Test equipment used: |
________________________ |
|||
Temperature, conductivity, pressure |
||||
(type, serial number): |
________________________ |
|||
Protective earth resistance, leakage current |
________________________ |
|||
(type, serial number): |
________________________ |
|||
|
|
|
|
|
|
|
|
|
|
Comments: |
|
|
||
|
|
|
|
|
|
|
|
|
|
Date: |
|
Signature: |
Stamp: |
|
|
|
|
|
|
4-8 |
Fresenius Medical Care 5008 SM 1/08.04 |