Metrohm 854 User Manual

CH-9101 Herisau info@metrohm.com www.metrohm.com
854 iConnect

Introduction

The 854 iConnect is the measuring input for elec­trodes with an integrated memory chip (iTrodes). The analog-digital converter in the 854 iConnect transforms the analog measuring signal of the electrode into digital pulses directly at the sensor. As a result of the digital data transfer, the meas­uring signal is no longer prone to electrostatic influences. A transmission free from interferences can be guaranteed no matter how long the elec­trode cable is.

Connecting the iConnect

The 854 iConnect can be connected to the fol­lowing devices:
Titrandos with an iConnect connector
867 pH Module

Connecting to a Titrando

Either one or two 854 iConnect can be connec­ted to a Titrando depending on the Titrando's model version.
Plug the iConnect plug of the 854 iConnect
into the socket "iConnect" of the Titrando. Make sure that the marking on the plug matches the marking on the Titrando as shown in figure 1.
Fig. 1: Connecting the 854 iConnect
The 854 iConnect is detected automatically
and entered as measuring input into the device properties of the Titrando. If an elec­trode is connected to the 854 iConnect that is not yet listed in the sensor list of the con­trol software, a corresponding message is dis­played.
The 854 iConnect can be plugged in and discon­nected without problems while the instrument is switched on.

Connecting an electrode

Remove the protective cap and connect the
electrode as shown in figure 2. The guide pin guarantees correct connection in such a way that the contact pins cannot be damaged.
Tighten the screw coupling with two fingers.
Fig. 2: Connecting an electrode
Note
Screw on the protective cap when the 854 iConnect is not in use as a protection against chemicals and mechanical impact.

Technical specifications

Measuring interface

Potentiometry:
Input resistance
Offset current < 1 · 10
Polarizer:
Polarization current
1)
Ipol
Polarization voltage
2)
Upol
1)
–125.0 / +125.0 µA: non-guaranteed values,
dependent on reference voltage +2.5 V
2)
–1250 / +1250 mV: non-guaranteed values,
dependent on reference voltage +2.5 V

Measuring ranges

pH values:
Range –13…+20 pH
Resolution 0.001 pH
Measuring accuracy 1)±0.003 pH
Voltage:
2)
Range –1200 mV…
Resolution 0.1 mV
Measuring accuracy 1)±0.2 mV
Current:
3)
Range –120 µA…+120 µA
Resolution 0.01 µA
> 1 · 1012 Ω
-12
A under
reference conditions
–122.5…+122.5 µA in steps of 2.5 µA
–1225…+1225 mV in steps of 25 mV
+1200 mV
Temperature:
Range Pt1000 NTC
–150 °C…+250 °C –5 °C…+250 °C
4)
Resolution Pt1000 NTC
0.1 °C
0.1 °C
Measuring accuracy 1)
Pt1000 ±0.2 °C (–20 °C…
+150 °C)
NTC ±0.6 °C (+10 °C…
+40 °C)
1)
± 1 digit, without sensor error, under reference
conditions
2)
potentiometric and voltametric
3)
amperometric
4)
for an NTC sensor with R (25 °C) = 30'000 Ω
and B (25/50) = 4100 K.

Power supply

Power supply via data cable, e.g.
from Titrando
Operating voltage 5 V
Power consumption approx. 10 mA dur-
ing measurement

Safety specifications

Design and testing According to
EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1 Protection class III

Electromagnetic compatibility

Emission Standards fulfilled:
EN/IEC 61326-1 EN/IEC 61000-6-3 EN 55022 / CISPR 22
8.854.8001EN 1/2
Immunity Standards fulfilled:
EN/IEC 61326-1 EN/IEC 61000-6-2 EN/IEC 61000-4-2 EN/IEC 61000-4-3

Ambient temperature

Nominal function range
Storage –20 °C…+60 °C
Transport –40 °C…+60 °C

Reference conditions

Ambient temperature +25°C (±3°C)
Relative humidity ≤ 60%
Operating tempera­ture status
Validity of the data After adjustment
+5 °C…+45 °C (at a maximum of 85 % humidity)
Instrument in opera­tion at least 30 min
obtained from your local authori­ties, from waste disposal compa­nies or from your local dealer.

Scope of delivery

2.854.0010 iConnect
854 iConnect 1.854.0010
Manual 854 iConnect, DE, EN
Optional accessories
iAquatrode Plus with Pt1000
iUnitrode with Pt1000
iSolvotrode 6.0279.300
iAg Titrode 6.0470.300
iPt Titrode 6.0471.300
8.854.8001DE,
8.854.8001EN
6.0277.300
6.0278.300
ranty provisions apply. Warranty claims cannot be pursued if the Customer has not complied with the obligations to make payment on time.
During the warranty period Metrohm undertakes, at its own choice, to either repair at its own premises, free of charge, any instruments that can be shown to be faulty or to replace them. Transport costs are to the Customer's account.
Faults arising from circumstances that are not the responsibility of Metrohm, such as improper stor­age or improper use, etc. are expressly excluded from the warranty.

Further documents

Documentation in additional languages can be found on http://products.metrohm.com under
Literature/Technical documentation.

Dimensions

Material of housing Polypropylene (PP)
with 5 % steel fibers
Diameter 23 mm
Length of housing 71 mm
Length of cable 1500 mm
Weight (without elec­trode)

Recycling and disposal

This product is covered by Euro­pean Directive 2002/96/EC, WEEE – Waste from Electrical and Elec­tronic Equipment. The correct disposal of your old equipment will help to prevent negative effects on the environ­ment and public health. More details about the disposal of your old equipment can be
8.854.8001EN 2/2
51 g
iEcotrode plus 6.0280.300
iAg ring electrode, comb.
iPt ring electrode, comb.
6.0450.300
6.0451.300

Warranty (guarantee)

Metrohm guarantees that the deliveries and serv­ices it provides are free from material, design or manufacturing errors. The warranty period is 36 months from the day of delivery; for day and night operation it is 18 months. The warranty remains valid on condition that the service is pro­vided by an authorized Metrohm service organi­zation.
Glass breakage is excluded from the warranty for electrodes and other glassware. The warranty for the accuracy corresponds to the technical specifi­cations given in this manual. For components from third parties that make up a considerable part of our instrument, the manufacturer's war-