Metrohm 805 User Manual

805 Dosimat
Manual
8.805.8001EN
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
805 Dosimat
Manual
8.805.8001EN 05.2010 dm /ek
Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on http://prod-
ucts.metrohm.com under Literature/Technical documentation.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 Introduction 1
1.1 Instrument description ......................................................... 1
1.2 About the documentation ................................................... 1
1.2.1 Symbols and conventions ........................................................ 1
1.3 Safety instructions ................................................................ 2
1.3.1 General notes on safety ........................................................... 2
1.3.2 Electrical safety ........................................................................ 2
1.3.3 Working with liquids ................................................................ 3
1.3.4 Flammable solvents and chemicals ........................................... 3
1.3.5 Recycling and disposal ............................................................. 4
2 Overview of the instrument 5
3 Installation 7
Table of contents
3.1 Setting up the instrument .................................................... 7
3.1.1 Packaging ................................................................................ 7
3.1.2 Checks .................................................................................... 7
3.1.3 Location .................................................................................. 7
3.2 Connecting the 805 Dosimat ............................................... 7
4 Working with the exchange unit 9
5 Handling and maintenance 11
5.1 General information ........................................................... 11
5.1.1 Care ...................................................................................... 11
5.1.2 Maintenance by Metrohm Service .......................................... 12
5.2 Quality Management and validation with Metrohm ....... 12
5.3 GLP - Validation .................................................................. 13
6 Troubleshooting 14
6.1 Problems ............................................................................. 14
7 Appendix 17
7.1 Buret data ........................................................................... 17
805 Dosimat
8 Technical specifications 18
8.1 Dosing drive ........................................................................ 18
8.2 Power supply ...................................................................... 18
8.3 Safety specifications ........................................................... 18
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
8.4 Electromagnetic compatibility (EMC) ................................ 18
8.5 Ambient temperature ......................................................... 19
8.6 Dimensions .......................................................................... 19
9 Conformity and warranty 20
9.1 Declaration of Conformity ................................................. 20
9.2 Quality Management Principles ........................................ 21
9.3 Warranty (guarantee) ......................................................... 22
10 Accessories 23
10.1 Scope of delivery ................................................................ 23
10.1.1 805 Dosimat 2.805.0010 ....................................................... 23
10.2 Optional accessories ........................................................... 24
10.2.1 805 Dosimat 2.805.0010 ....................................................... 24
Index 26
■■■■■■■■
IV
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 Front 805 Dosimat ............................................................................ 5
Figure 2 Rear 805 Dosimat ............................................................................. 6
Figure 3 Attaching the exchange unit ............................................................. 9
Figure 4 Contact pins for the exchange unit .................................................. 11
Table of figures
805 Dosimat
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
The 805 Dosimat is a versatile dosing drive which can be used with a number of different Metrohm devices (e.g. Titrando or sample changer). The 805 Dosimat and the corresponding 806 Exchange Unit are suitable as a buret not only for simply dosing auxiliary solutions but also for titrations.

1.1 Instrument description

Thanks to various 806 Exchange Units with 1, 5, 10, 20 or 50 mL dosing cylinders, the 805 Dosimat is suitable for flexible use as a buret (dosing unit and drive). Older exchange units (without data chip) can also be used.
The exchange units can be set up on the Dosimat and also removed again in one easy manual step.

1 Introduction

1.2 About the documentation

Caution
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe operation of the instrument.

1.2.1 Symbols and conventions

The following symbols and styles are used in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
Instruction step
Carry out these steps in the sequence shown.
Warning
805 Dosimat
This symbol draws attention to a possible life hazard or risk of injury.
Warning
This symbol draws attention to a possible hazard due to electrical current.
■■■■■■■■
1

1.3 Safety instructions

1.3 Safety instructions

1.3.1 General notes on safety

■■■■■■■■■■■■■■■■■■■■■■
Warning
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
Warning
This symbol draws attention to a possible biological hazard.
Caution
This symbol draws attention to a possible damage of instruments or instrument parts.
Note
This symbol marks additional information and tips.
This instrument may only be operated in accordance with the specifica­tions in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

1.3.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
Warning
Warning
■■■■■■■■
2
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
Warning
Never open the housing of the instrument. The instrument could be dam­aged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Always pull the mains cable out of the mains connection socket before connecting or disconnecting electrical appliances on the rear panel of the instrument.

1.3.3 Working with liquids

Caution
Periodically check all system connections for leaks. Observe the relevant regulations in respect to working with flammable and/or toxic fluids and their disposal.

1.3.4 Flammable solvents and chemicals

Warning
All relevant safety measures are to be observed when working with flam­mable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. laboratory
flue).
Keep all sources of flame far from the workplace. Clean up spilled fluids and solids immediately. Follow the safety instructions of the chemical manufacturer.
805 Dosimat
■■■■■■■■
3
1.3 Safety instructions

1.3.5 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
■■■■■■■■■■■■■■■■■■■■■■
■■■■■■■■
4
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
805 Dosimat
On
Status
1
2 3
4
6
5
Metrohm
2 Overview of the instrument

2 Overview of the instrument

Figure 1 Front 805 Dosimat
Guide openings
1
For centering the exchange unit
Contact pins
3
For the data chip
LED "On"
5
Lights up when the Titrando and a controller (Touch Control or computer) are connected to the mains and switched on.
Piston rod
2
Moves the piston of the exchange unit up and down.
Coupling
4
For switching the flat stopcock
LED "Status"
6
Indicates the current status of the internal dosing drive
805 Dosimat
■■■■■■■■
5
■■■■■■■■■■■■■■■■■■■■■■
Nr
Made by Metrohm Herisau Switzerland
1
2
Figure 2 Rear 805 Dosimat
Type plate
1
Contains specifications concerning mains voltage, instrument type and serial number.
Connection cable
2
For connecting to a control device.
■■■■■■■■
6
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
3 Installation

3.1 Setting up the instrument

3.1.1 Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

3.1.2 Checks

Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.

3.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.

3 Installation

Place the instrument in a location of the laboratory suitable for operation and free of vibrations, if possible protected from corrosive atmospheres and contamination by chemicals.
The instrument should be protected against excessive temperature fluctua­tions and direct sunlight.

3.2 Connecting the 805 Dosimat

Note
In the case of a Titrando with exchange unit, the MSB Port 1 is used by the internal drive. The MSB 1 is therefore not available for an 805 Dosi­mat .
The 805 Dosimat is controlled by devices which are equipped with a 'Me­trohm Serial Bus' (MSB). These are e.g. Titrandos or sample changers. Connect the 805 Dosimat on the rear of the corresponding device to one of the connectors (MSB 1/MSB 2 to MSB 4 ). The 805 Dosimat can how- ever not be connected to devices with an MSB connector which do not support tandem dosing.
805 Dosimat
■■■■■■■■
7
3.2 Connecting the 805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
Warning
Connect the 805 Dosimat only to control devices which have been switched off beforehand. The control device will only be able to recognize the Dosimat during the switch-on sequence.
Observe the alignment of the connection socket. Never use excessive force to plug in the connection cable! Doing so could damage the device electronics.
■■■■■■■■
8
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
4 Working with the exchange unit
The 806 exchange unit has an integrated data chip which allows to save data for the exchange unit and for the reagent. The data is being edited in the Touch Control or in the PC software. The assembly of the exchange unit is described in the manual for the 806 exchange unit.

4 Working with the exchange unit

805 Dosimat
Figure 3 Attaching the exchange unit
■■■■■■■■
9
■■■■■■■■■■■■■■■■■■■■■■
To attach the exchange unit, proceed as follows:
Slide the exchange unit from the front onto the 805 Dosimat and push
1
all the way to the rear, so that it snaps in and the LED Status flashes slowly.
It must snap in audibly.
If the exchange unit is attached correctly, its initialization is being activated by a micro switch which is triggered by the guide bolts of the exchange unit. The exchange unit is being recognized and the data is being read automat­ically from the data chip. After this, the LED Status lights up constantly.
The following table provides an overview showing all operating states of the internal dosing device which are indicated with the LED Status:
LED Status
Operating status of the dosing device
off No exchange unit is attached.
flashes constantly The Titrando is ready to dose or
titrate. The exchange unit has been attached correctly and recognized and is now in the exchange posi­tion, i.e. the exchange unit can be removed.
flashes slowly The Titrando is dosing or filling or
the exchange unit is not in the exchange position, respectively.
An 806 intelligent exchange unit has been attached and the data on the integrated data chip is being read or written.
flashes fast Error at the dosing drive (see Chap.
5)
■■■■■■■■
10
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
1
5 Handling and maintenance

5.1 General information

5.1.1 Care

The 805 Dosimat requires no special care. Take care to ensure however that it is not exposed to any excessive contamination nor to any corrosive influ­ences. These could under certain circumstances result in functional disrup­tions and a reduction in the service life of the inherently sturdy mechanics and electronics.
Monthly inspections are called for in the event that alkali, corrosive or high­concentration reagents are used. If non-problematic reagents are used, then the inspection intervals can be extended to between six and twelve months.
Spilled chemicals and solvents should be removed immediately. Above all, the plug connection on the rear of the instrument should be protected from contamination.

5 Handling and maintenance

Caution
Although this is extensively prevented by design measures, the mains plug of the control device should be unplugged immediately if aggressive media has penetrated the inside of the instrument, so as to avoid serious damage to the instrument electronics. In such cases, the Metrohm Service must be informed.
Figure 4 Contact pins for the exchange unit
Contact pins
1
For exchanging data with the exchange unit.
The contact pins for data exchange with the data chip are made of titanium and are exceptionally resistant to both chemicals and abrasion.
805 Dosimat
■■■■■■■■
11

5.2 Quality Management and validation with Metrohm

Caution
Take care to ensure that the contact pins do not become contaminated. Wipe off any contaminations at once.
In the event of more serious contamination, you can use a moist cloth (possibly with a small amount of dishwashing detergent or ethanol).
■■■■■■■■■■■■■■■■■■■■■■

5.1.2 Maintenance by Metrohm Service

Maintenance of the 805 Dosimat is best carried out as part of an annual service, which is performed by specialist personnel of the Metrohm com­pany. If working frequently with caustic and corrosive chemicals, a shorter maintenance interval could be necessary.
The Metrohm service department offers every form of technical advice for maintenance and service of all Metrohm instruments.
5.2 Quality Management and validation with Metrohm
Quality Management
Metrohm offers you comprehensive support in implementing quality man­agement measures for instruments and software. Further information on this can be found in the brochure «Quality Management with Metrohm» available from your local Metrohm agent.
Validation
Please contact your local Metrohm agent for support in validating instru­ments and software. Here you can also obtain validation documentation to provide help for carrying out the Installation Qualification (IQ) and the Operational Qualification (OQ). IQ and OQ are also offered as a service by the Metrohm agents. In addition, various application bulletins are also available on the subject, which also contain Standard Operating Proce- dures (SOP) for testing analytical measuring instruments for reproducibility and correctness.
Maintenance
Electronic and mechanical functional groups in Metrohm instruments can and should be checked as part of regular maintenance by specialist person­nel from Metrohm. Please ask your local Metrohm agent regarding the precise terms and conditions involved in concluding a corresponding main­tenance agreement.
■■■■■■■■
12
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
You can find information on the subjects of quality management, valida­tion and maintenance as well as an overview of the documents currently available at www.metrohm.com/com/ under Support.

5.3 GLP - Validation

Every Dosimat and every exchange unit manufactured by the Metrohm company is subjected to rigorous quality controls prior to shipment. Every exchange unit is issued a quality certificate attesting conformance with the quality criteria of Metrohm. GLP (Good Laboratory Practice) requires, among other things, periodic inspection of analytical measuring devices with respect to precision and correctness on the basis of standard operating pro­cedures Standard Operating Procedure, SOP). This may also include an inspection of dosing accuracy.
5 Handling and maintenance
Note
Recommended literature
Metrohm brochure "Quality management with Metrohm", detailed infor-
mation concerning the principles and procedural methods of Good Laboratory Practice
Metrohm Applications Bulletin 283/1 Validation of Metrohm burets
The validation of burets is carried out by the Metrohm service with special software.
The Metrohm agents worldwide offer the possibility of on-site inspections and certifications of Dosimats and exchange units with respect to accuracy. It is recommended that an accuracy inspection be performed when the dos­ing cylinders and dosing pistons of an exchange unit are replaced .
805 Dosimat
■■■■■■■■
13

6.1 Problems

6 Troubleshooting

6.1 Problems
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
Data of the exchange unit cannot be read.
Dosimat cannot be actuated by the con­trol device.
Exchange unit recog­nized either not at all or incorrectly.
LED "Status" is flash­ing fast.
Data chip mechanically damaged or impaired by chemicals.
Connection between Dosi­mat and control device is interrupted or an error has occurred on the Dosimat.
The exchange unit has not correctly been attached.
The dosing drive is overloa­ded because the flat stop­cock is blocked.
Remove the exchange unit and set it up
again.
Have the data chip replaced by the Metrohm
Service Dept.
Check the cable connections. Switch the instrument off and on again. Check the dosing and filling rate. Contact the Metrohm Service Dept.
Remove the exchange unit and set it up
again.
Check the correct placing of the exchange
unit.
Check the piston and flat stopcock position. Switch the instrument off and on again. If necessary contact the Metrohm Service
Dept.
Switch off the Touch Control or exit PC Control. Check wether the exchange unit can be removed. If it cannot be removed, check wether the flat stopcock still can be turned. Switch it manually to the exchange position turning it to the right (see manual for the exchange unit). Remove the exchange unit and proceed as described in the manual for the exchange unit.
■■■■■■■■
14
The dosing drive is overloa­ded because the piston is blocked.This error is dis­played by the software (Touch Control or PC Con­trol / tiamo).
Switch the control device off and then on again. The dosing device is being initialized during switching on. Remove the exchange unit and clean it as described in the manual for the exchange unit in chapter "Care and mainte­nance". Contact the Metrohm service if remov­ing the exchange unit is not possible.
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
Problem Cause Remedy
6 Troubleshooting
LED "Status" is not illuminated although an exchange unit is attached.
No dosing takes place at all.
The data of the exchange unit cannot be read because the data chip has been damaged mechani­cally or by chemicals.
The exchange unit has not correctly been attached.
Tubing connections are blocked or the exchange unit is not assembled cor­rectly.
Have the data chip being replaced by the Metrohm service.Until the data chip is being replaced you can remove the data chip yourself in order to be able to still use the exchange unit. The cylinder volume is automatically recognized nevertheless, but no data can be read from the exchange unit or be saved on it anymore.
Remove the exchange unit and attach it again until it snaps in. The LED flashes during data is read from an intelligent exchange unit (806) and lights up constantly if the exchange unit has cor­rectly been recognized.
Check whether the tubings are correctly con-
nected.
Check whether the dosing tip is blocked. Check whether the filling tubing is blocked. Check whether all of the openings of the
bottle attachment have accidentally been closed (vacuum in the supply bottle!). One of the openings must be open as a pressure compensation or has to be equipped with an (open) adsorber tube.
Remove the exchange unit and check
whether the dosing piston is connected to the drive.
The entire system is blocked.
The exchange unit cannot be attached to the Dosimat.
The exchange unit cannot be attached.
805 Dosimat
An exceptional error has occurred on the Dosimat or the control device.
The dosing piston is mis­aligned.
The piston rod of the exchange unit is misaligned.
The flat stopcock of the exchange unit is not in the exchange position.
The piston rod in the exchange unit is not in the right position.
Check the cable connections. Switch the instrument off and on again.
Carefully pull out or push in the piston with the
6.2739.010 key until the recess of the piston rod is tight on the buret base.
Switch the instrument off and on again.
Switch the flat stopcock manually to the exchange position (lever switch directed to the right).
Switch the piston rod to the right position (see manual for the exchange unit).
■■■■■■■■
15
6.1 Problems
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
The exchange unit cannot be removed and the LED "Status" is flashing slowly.
The exchange unit cannot be removed from the Dosimat.
The flat stopcock is blocked.
It is being dosed or filled and/or the Dosimat is not in the exchange position.
Piston and/or flat stopcock are not in filling position.
The flat stopcock is conta­minated, corroded or worn out.
Stop the run or carry out a "filling".
Carry out the [Filling] on the control device.
Check the cable connections to the control instrument.
Switch the instrument off and on again.
Carefully loosen the flat stopcock out of its
holder and switch the control device off and on again.
Put the flat stopcock into water with a small
amount of dishwashing detergent or into ethanol and eventually clean it with ultra­sonic. Rinse it thoroughly and place it again into the flat stopcock holder.
Replace the defectous flat stopcock.
■■■■■■■■
16
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
7 Appendix

7.1 Buret data

The 806 Exchange Unit is equipped with a data chip which contains the specifications for the dosing unit, the tubing connections and the reagent used.
Indications on dosing unit / tubing connections
Order number of the dosing unit Serial number of the dosing unit Serial number of the dosing cylinder Length and diameter of the tubings on the dosing ports Validation date
Indications on the reagent
Name of the reagent Titer of the reagent Concentration of the reagent Production and expiry date of the reagent

7 Appendix

The 805 Dosimat makes it possible to read and overwrite data with the aid of a suitable device (e.g. Titrando). Consult the respective manual in order to determine whether the Metrohm device which you are using is capable of accomplishing this.
805 Dosimat
■■■■■■■■
17

8.1 Dosing drive

8 Technical specifications

8.1 Dosing drive
■■■■■■■■■■■■■■■■■■■■■■
Cylinder volume of
1 ml, 5 ml, 10 ml, 20 ml or 50 ml
the dosing unit
Resolution
Dosing / filling time
20'000 steps per cylinder volume
18 seconds each for the cylinder volume

8.2 Power supply

from control device
Dosing device con­nector
±12 V, 5 V , 6 W
Mini DIN plug, 8-pin, (MSB)

8.3 Safety specifications

Design and testing
EN/IEC 61010-1:2001 UL 61010-1: 2004 CSA-C22.2 No. 61010-1: 2004 Degree of protection IP40 Protection class III
Safety instructions
The documentation contains safety instructions which have to be fol­lowed by the user in order to ensure safe operation of the instrument.

8.4 Electromagnetic compatibility (EMC)

Emission
Immunity
■■■■■■■■
18
Standards fulfilled:
EN/IEC 61326-1:2006 EN/IEC 61000-6-3: 2006 EN 55011 / CISPR 11: 2007
Standards fulfilled:
EN/IEC 61326-1:2006 EN/IEC 61000-6-2: 2005 EN/IEC 61000-4-2: 2001 EN/IEC 61000-4-3: 2006 EN/IEC 61000-4-4: 2004 EN/IEC 61000-4-5: 2001 EN/IEC 61000-4-6: 2001
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
EN/IEC 61000-4-11: 2004 EN/IEC 61000-4-14: 2004 NAMUR: 2004

8.5 Ambient temperature

8 Technical specifications
Nominal function range
Storage
Transport
+5…+45 °C (at a maximum of 85% humidity)
–20…+60 °C
–40…+60 °C

8.6 Dimensions

Width
Height
Depth
Weight
Material (Housing)
142 mm
164 mm approx. 450 mm with exchange unit
231 mm
2550 g (without exchange unit)
PBT (polybutylene terephthalate)
805 Dosimat
■■■■■■■■
19

9.1 Declaration of Conformity

9 Conformity and warranty

9.1 Declaration of Conformity
This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
■■■■■■■■■■■■■■■■■■■■■■
Name of commodity
Electromagnetic com­patibility
Safety specifications
805 Dosimat
Microprocessor-controlled dosing device for liquid-handling purposes in laboratories.
This instrument has been built and has undergone final type testing accord­ing to the standards:
Emission: EN/IEC 61326-1: 2006, EN/IEC 61000-6-3: 2006,
EN 55011 / CISPR 11: 2007
Immunity: EN/IEC 61326-1: 2006, EN/IEC 61000-6-2: 2005,
EN/IEC 61000-4-2: 2001, EN/IEC 61000-4-3: 2006, EN/IEC 61000-4-4: 2004, EN/IEC 61000-4-5: 2001, EN/IEC 61000-4-6: 2001, EN/IEC 61000-4-11: 2004, EN/IEC 61000-4-14: 2004, NAMUR: 2004
EN/IEC 61010-1: 2001, UL 61010-1: 2004, CSA-C22.2 No. 61010-1: 2004, protection class I
Manufacturer
■■■■■■■■
20
This instrument meets the requirements of the CE mark as contained in the EU directives 2006/95/EC (LVD), 2004/108/EC (EMC). It fulfils the following specifications:
EN 61326-1 Electrical equipment for measurement, control and
laboratory use – EMC requirements
EN 61010-1 Safety requirements for electrical equipment for
measurement, control and laboratory use
Metrohm Ltd., CH-9101 Herisau/Switzerland
Metrohm Ltd. is holder of the SQS certificate ISO 9001:2000 Quality man­agement system for development, production and sales of instruments and accessories for ion analysis.
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
9 Conformity and warranty
Herisau, 6 January, 2009
D. Strohm
Vice President, Head of R&D

9.2 Quality Management Principles

Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number 10872-02, issued by SQS (Swiss Association for Quality and Management Systems). Internal and external audits are carried out periodically to assure that the standards defined by Metrohm’s QM Manual are maintained.
The steps involved in the design, manufacture and servicing of instruments are fully documented and the resulting reports are archived for ten years. The development of software for PCs and instruments is also duly docu­mented and the documents and source codes are archived. Both remain the possession of Metrohm. A non-disclosure agreement may be asked to be provided by those requiring access to them.
The implementation of the ISO 9001:2000 quality management system is described in Metrohm’s QM Manual, which comprises detailed instructions on the following fields of activity:
Instrument development
The organization of the instrument design, its planning and the intermediate controls are fully documented and traceable. Laboratory testing accompa­nies all phases of instrument development.
A. Dellenbach
Head of Quality Management
805 Dosimat
Software development
Software development occurs in terms of the software life cycle. Tests are performed to detect programming errors and to assess the program’s func­tionality in a laboratory environment.
Components
All components used in the Metrohm instruments have to satisfy the quality standards that are defined and implemented for our products. Suppliers of components are audited by Metrohm as the need arises.
Manufacture
The measures put into practice in the production of our instruments guar­antee a constant quality standard. Production planning and manufacturing procedures, maintenance of production means and testing of components, intermediate and finished products are prescribed.
■■■■■■■■
21

9.3 Warranty (guarantee)

Customer support and service
Customer support involves all phases of instrument acquisition and use by the customer, i.e. consulting to define the adequate equipment for the ana­lytical problem at hand, delivery of the equipment, user manuals, training, after-sales service and processing of customer complaints. The Metrohm service organization is equipped to support customers in implementing standards such as GLP, GMP, ISO 900X, in performing Operational Qualifi­cation and Performance Verification of the system components or in carrying out the System Validation for the quantitative determination of a substance in a given matrix.
9.3 Warranty (guarantee)
Metrohm guarantees that the deliveries and services it provides are free from material, design or manufacturing errors. The warranty period is 36 months from the day of delivery; for day and night operation it is 18 months. The warranty remains valid on condition that the service is provided by an authorized Metrohm service organization.
■■■■■■■■■■■■■■■■■■■■■■
Glass breakage is excluded from the warranty for electrodes and other glassware. The warranty for the accuracy corresponds to the technical spec­ifications given in this manual. For components from third parties that make up a considerable part of our instrument, the manufacturer's warranty pro­visions apply. Warranty claims cannot be pursued if the Customer has not complied with the obligations to make payment on time.
During the warranty period Metrohm undertakes, at its own choice, to either repair at its own premises, free of charge, any instruments that can be shown to be faulty or to replace them. Transport costs are to the Customer's account.
Faults arising from circumstances that are not the responsibility of Metrohm, such as improper storage or improper use, etc. are expressly excluded from the warranty.
■■■■■■■■
22
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■

10 Accessories

10.1 Scope of delivery

Note
Subject to change without notice.

10.1.1 805 Dosimat 2.805.0010

Qty. Order no. Description
1 1.805.0010 805 Dosimat
10 Accessories
1 6.2739.010 Wrench
For exchange units.
Length (mm): 190
1 8.805.8001EN Manual 805 Dosimat
805 Dosimat
■■■■■■■■
23

10.2 Optional accessories

10.2 Optional accessories

10.2.1 805 Dosimat 2.805.0010

Order no. Description
6.2151.010 Cable MSB plug / socket
Extension cable for MSB connections and for Touch Control - Titrando. Mini DIN socket - Mini DIN plug
Length (m): 2
6.3026.110 Exchange unit 1 mL
Exchange unit with integrated data chip with 1 mL glass cylinder and light pro­tection. PCTFE/PTFE flat stopcock, FEP tubing connection, antidiffusion buret tip and standard amber glass reagent bottle.
Volume (mL): 1
■■■■■■■■■■■■■■■■■■■■■■
6.3026.150 Exchange unit 5 mL
Exchange unit with integrated data chip with 5 mL glass cylinder and light pro­tection. PCTFE/PTFE flat stopcock, FEP tubing connection, antidiffusion buret tip and standard amber glass reagent bottle.
Volume (mL): 5
■■■■■■■■
24
805 Dosimat
■■■■■■■■■■■■■■■■■■■■■■
Order no. Description
6.3026.210 Exchange unit 10 mL
Exchange unit with integrated data chip with 10 mL glass cylinder and light pro­tection. PCTFE/PTFE flat stopcock, FEP tubing connection, antidiffusion buret tip and standard amber glass reagent bottle.
Volume (mL): 10
6.3026.220 Exchange unit 20 mL
Exchange unit with integrated data chip with 20 mL glass cylinder and light pro­tection. PCTFE/PTFE flat stopcock, FEP tubing connection, antidiffusion buret tip and standard amber glass reagent bottle.
Volume (mL): 20
10 Accessories
6.3026.250 Exchange unit 50 mL
Exchange unit with integrated data chip with 50 mL glass cylinder and light pro­tection. PCTFE/PTFE flat stopcock, FEP tubing connection, antidiffusion buret tip and standard amber glass reagent bottle.
Volume (mL): 50
805 Dosimat
■■■■■■■■
25

Index

Index
■■■■■■■■■■■■■■■■■■■■■■
A
Accuracy .................................. 13
B
Buret data ................................ 17
C
Certification ............................. 13
Connection cable ....................... 6
Contact pins ......................... 5, 11
Control ....................................... 7
Coupling .................................... 5
D
Data chip ........................... 15, 17
Data exchange ......................... 17
Device type ................................ 6
Dosing accuracy ....................... 13
E
Electromagnetic compatibility ... 18
Electrostatic charge .................... 3
Emission ................................... 18
Exchange unit .......... 9, 15, 16, 15
F
Flat stopcock blocked ............... 14
G
GLP .................................... 12, 13
Good Laboratory Practice ......... 13
Guarantee ................................ 22
Guide openings .......................... 5
I
Immunity .................................. 18
L
LED
On ....................................... 5
Status ........... 5, 10, 15, 16, 14
M
Maintenance ............................ 11
O
Order number .......................... 17
P
Piston blocked .......................... 14
Piston rod ................................... 5
Power supply ............................ 18
Production date ........................ 17
Q
Quality certificate ..................... 13
Quality control ......................... 13
Quality Management ................ 12
R
Reagent
Concentration .................... 17
Expiry date ......................... 17
Name ................................. 17
Production date .................. 17
Titer ................................... 17
S
Safety instructions ...................... 2
Safety specifications ................. 18
Serial number ....................... 6, 17
Service ....................................... 2
Service Agreement ................... 12
SOP .......................................... 13
T
Tubing diameter ....................... 17
Tubing length ........................... 17
V
Validation ........................... 12, 13
Validation date ......................... 17
W
Warranty .................................. 22
■■■■■■■■
26
805 Dosimat
Loading...