Although all the information given in this documentation has been
checked with great care, errors cannot be entirely excluded. Should you
notice any mistakes please send us your comments using the address
given above.
Documentation in additional languages can be found on
http://products.metrohm.com under Literature/Technical documenta-
6.2.1804 Ti Stand 2.804.0010, 2.804.0040 ................................... 19
■■■■■■■■
IV
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
Table of figures
Figure 1Front 804 Ti Stand ............................................................................. 5
Figure 2Rear 804 Ti Stand .............................................................................. 5
Figure 3Connecting the 804 Ti Stand ........................................................... 11
Figure 4Connecting the Dosino to the 804 Ti Stand ..................................... 11
Table of figures
804 Ti Stand
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
1.1 Instrument description
The 804 Ti Stand is a titration stand and controller for the propeller stirrer
802 Stirrer.
The titration stand is directly connected to an analysis device via an MSB
connection cable.
1.2 About the documentation
Caution
Please read through this documentation carefully before putting the
instrument into operation. The documentation contains information
and warnings which the user must follow in order to ensure safe operation of the instrument.
1 Introduction
1.2.1 Symbols and conventions
The following symbols and styles are used in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the
second to the instrument part in the figure.
Instruction step
Carry out these steps in the sequence shown.
Warning
This symbol draws attention to a possible life hazard
or risk of injury.
Warning
This symbol draws attention to a possible hazard due
to electrical current.
Warning
This symbol draws attention to a possible hazard due
to heat or hot instrument parts.
804 Ti Stand
■■■■■■■■
1
1.3 Safety instructions
1.3 Safety instructions
1.3.1 General notes on safety
Warning
This instrument may only be operated in accordance with the specifications in this documentation.
■■■■■■■■■■■■■■■■■■■■■■
Warning
This symbol draws attention to a possible biological
hazard.
Caution
This symbol draws attention to a possible damage of
instruments or instrument parts.
Note
This symbol marks additional information and tips.
This instrument has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
instrument, the following instructions must be observed carefully.
1.3.2 Electrical safety
The electrical safety when working with the instrument is ensured as part
of the international standard IEC 61010.
Only personnel qualified by Metrohm are authorized to carry out service
work on electronic components.
Never open the housing of the instrument. The instrument could be
damaged by this. There is also a risk of serious injury if live components
are touched.
There are no parts inside the housing which can be serviced or replaced
by the user.
Warning
Warning
■■■■■■■■
2
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see
rear panel of the instrument).
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be
destroyed by discharges.
Always pull the mains cable out of the mains connection socket before
connecting or disconnecting electrical appliances on the rear panel of
the instrument.
1.3.3 Working with liquids
Caution
Periodically check all system connections for leaks. Observe the relevant
regulations in respect to working with flammable and/or toxic fluids
and their disposal.
1.3.4 Flammable solvents and chemicals
Warning
All relevant safety measures are to be observed when working with
flammable solvents and chemicals.
■Set up the instrument in a well-ventilated location (e.g. laboratory
flue).
■Keep all sources of flame far from the workplace.
■Clean up spilled fluids and solids immediately.
■Follow the safety instructions of the chemical manufacturer.
804 Ti Stand
■■■■■■■■
3
1.3 Safety instructions
1.3.5 Recycling and disposal
This product is covered by European Directive 2002/96/EC, WEEE – Waste
from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative
effects on the environment and public health.
More details about the disposal of your old equipment can be obtained
from your local authorities, from waste disposal companies or from your
local dealer.
■■■■■■■■■■■■■■■■■■■■■■
■■■■■■■■
4
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
3.804.300x
MSB
1
4
2 Device overview
Figure 1Front 804 Ti Stand
2 Device overview
LED function display
1
Lights up, when the stirrer is switched on.
Figure 2Rear 804 Ti Stand
MSB connector
1
Metrohm Serial Bus.
For connecting additional devices
Clamping screw
2
For fixing the positioning height or the
swing position of the magnetic stirrer.
MSB connection cable
3
For connecting to a control device.
804 Ti Stand
Propeller stirrer connector
4
For connecting the 802 Stirrer
■■■■■■■■
5
3.1 Setting up the instrument
3 Installation
3.1 Setting up the instrument
3.1.1 Packaging
The instrument is supplied in highly protective special packaging together
with the separately packed accessories. Keep this packaging, as only this
ensures safe transportation of the instrument.
3.1.2 Checks
Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.
3.1.3 Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
■■■■■■■■■■■■■■■■■■■■■■
Place the instrument in a location of the laboratory which is suitable for
operation, free of vibrations, protected from corrosive atmosphere, and
contamination by chemicals.
The instrument should be protected against excessive temperature fluctuations and direct sunlight.
■■■■■■■■
6
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
6.2001.060
6.2016.070
1
2
3.2 Mounting the 804 Ti Stand
Mounting the stand
3 Installation
Mount the stand as follows:
Screw the 6.2001.060 stand plate with the accompanying four
1
screws tightly to the bottom of the instrument.
Screw the 6.2016.070 stand plate with the hexagon screw (hexagon
2
key supplied) firmly to the stand plate.
804 Ti Stand
■■■■■■■■
7
3.2 Mounting the 804 Ti Stand
2
1
■■■■■■■■■■■■■■■■■■■■■■
Fastening the titration stand
Fasten the titration stand as follows:
Attach the titration stand on the support rod and push it down-
1
wards.
With the clamping screw, fix the positioning height or the swing
2
position of the titration stand.
■■■■■■■■
8
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
6.2021.020
6.2013.010
Attaching the clamping ring and the electrode holder
Fasten the clamping ring and the electrode holder as follows:
Push the 6.2013.010 clamping ring onto the support rod.
1
Attach the 6.2021.020 electrode holder on the support rod.
2
Keep the red button pressed down and push the electrode holder
3
downwards to the required height.
Fix the position of the electrode holder with the aid of the clamping
4
ring.
3 Installation
804 Ti Stand
■■■■■■■■
9
3.3 Connecting the 804 Ti Stand
1
2
■■■■■■■■■■■■■■■■■■■■■■
Mounting the propeller stirrer
Mounting the propeller stirrer as follows:
Insert the propeller stirrer 802 Stirrer without the stirring propeller
1
from above into the center opening of the electrode holder.
Plug the stirring propeller from below to the propeller stirrer.
2
3.3 Connecting the 804 Ti Stand
Connect the titration stand as follows:
Exit the control software.
1
Connect the connection cable of the titration stand to one of the
2
sockets marked with MSB on the rear of the control device.
■■■■■■■■
10
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
MSB
3.804.300X
3 Installation
Figure 3Connecting the 804 Ti Stand
Connecting an additional device (e.g. Dosimat or Dosino) to the MSB
3
connector of the titration stand.
Figure 4Connecting the Dosino to the 804 Ti Stand
Start the control software.
4
804 Ti Stand
■■■■■■■■
11
4.1 Rotational speed 804 Ti Stand
4 Technical specifications
4.1 Rotational speed 804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
Maximum rota-
±1700...1900 U/min
tional speed
Setting of the
±15 steps
rotational speed
Increase of rota-
±115...125 U/min
tional speed per
step
4.2 Power supply
Voltage
Power consumption
Fuse
+12 V, -12 V, +5 V
4 W
Electronic overload protection
4.3 Safety specifications
Design and testing
■EN/IEC 61010-1:2001
■UL 61010-1: 2004
■CSA-C22.2 No. 61010-1: 2004
■Degree of protection IP40
■Protection class III
This is to certify the conformity to the standard specifications for electrical
appliances and accessories, as well as to the standard specifications for
security and to system validation issued by the manufacturing company.
■■■■■■■■■■■■■■■■■■■■■■
Name of commodity
Electromagnetic
compatibility
Safety specifications
804 Ti Stand
Titration stand and controller for 802 Stirrer.
This instrument has been built and has undergone final type testing
according to the standards:
EN/IEC 61010-1: 2001, UL 61010-1: 2004,
CSA-C22.2 No. 61010-1: 2004, degree of protection IP40, protection
class III
Manufacturer
■■■■■■■■
14
This instrument meets the requirements of the CE mark as contained in
the EU directives 2006/95/EC (LVD), 2004/108/EC (EMC). It fulfils the following specifications:
EN 61326-1Electrical equipment for measurement, control
and laboratory use – EMC requirements
EN 61010-1Safety requirements for electrical equipment for
measurement, control and laboratory use
Metrohm Ltd., CH-9101 Herisau/Switzerland
Metrohm Ltd. is holder of the SQS certificate ISO 9001:2000 Quality management system for development, production and sales of instruments
and accessories for ion analysis.
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
5 Conformity and warranty
Herisau, 6 January, 2009
D. Strohm
Vice President, Head of R&D
5.2 Quality Management Principles
Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number
10872-02, issued by SQS (Swiss Association for Quality and Management
Systems). Internal and external audits are carried out periodically to assure
that the standards defined by Metrohm’s QM Manual are maintained.
The steps involved in the design, manufacture and servicing of instruments
are fully documented and the resulting reports are archived for ten years.
The development of software for PCs and instruments is also duly documented and the documents and source codes are archived. Both remain
the possession of Metrohm. A non-disclosure agreement may be asked to
be provided by those requiring access to them.
The implementation of the ISO 9001:2000 quality management system is
described in Metrohm’s QM Manual, which comprises detailed instructions on the following fields of activity:
Instrument development
The organization of the instrument design, its planning and the intermediate controls are fully documented and traceable. Laboratory testing
accompanies all phases of instrument development.
A. Dellenbach
Head of Quality Management
804 Ti Stand
Software development
Software development occurs in terms of the software life cycle. Tests are
performed to detect programming errors and to assess the program’s
functionality in a laboratory environment.
Components
All components used in the Metrohm instruments have to satisfy the quality standards that are defined and implemented for our products. Suppliers of components are audited by Metrohm as the need arises.
Manufacture
The measures put into practice in the production of our instruments guarantee a constant quality standard. Production planning and manufacturing
procedures, maintenance of production means and testing of components, intermediate and finished products are prescribed.
■■■■■■■■
15
5.3 Warranty (guarantee)
Customer support and service
Customer support involves all phases of instrument acquisition and use by
the customer, i.e. consulting to define the adequate equipment for the
analytical problem at hand, delivery of the equipment, user manuals, training, after-sales service and processing of customer complaints. The
Metrohm service organization is equipped to support customers in implementing standards such as GLP, GMP, ISO 900X, in performing Operational Qualification and Performance Verification of the system components or in carrying out the System Validation for the quantitative determination of a substance in a given matrix.
5.3 Warranty (guarantee)
Metrohm guarantees that the deliveries and services it provides are free
from material, design or manufacturing errors. The warranty period is 36
months from the day of delivery; for day and night operation it is 18
months. The warranty remains valid on condition that the service is provided by an authorized Metrohm service organization.
■■■■■■■■■■■■■■■■■■■■■■
Glass breakage is excluded from the warranty for electrodes and other
glassware. The warranty for the accuracy corresponds to the technical
specifications given in this manual. For components from third parties that
make up a considerable part of our instrument, the manufacturer's warranty provisions apply. Warranty claims cannot be pursued if the Customer
has not complied with the obligations to make payment on time.
During the warranty period Metrohm undertakes, at its own choice, to
either repair at its own premises, free of charge, any instruments that can
be shown to be faulty or to replace them. Transport costs are to the Customer's account.
Faults arising from circumstances that are not the responsibility of
Metrohm, such as improper storage or improper use, etc. are expressly
excluded from the warranty.
■■■■■■■■
16
804 Ti Stand
■■■■■■■■■■■■■■■■■■■■■■
6 Accessories
6.1 Scope of delivery
Note
Subject to change without notice.
6.1.1 804 Ti Stand without stand 2.804.0010
Qty.Order no.Description
11.804.0010804 Ti Stand
Titration stand with permanently attached cable for MSB (Metrohm
Serial Bus).
6 Accessories
18.804.8001ML Manual 804 Ti Stand
6.1.2 804 Ti Stand with stand 2.801.0040
Qty.Order no.Description
11.804.0010804 Ti Stand
Titration stand with permanently attached cable for MSB (Metrohm
Serial Bus).
16.2001.060Stand plate without support rod
For mounting the 801, 804, 803 stirrers on a Titrando, Titrino plus,
Dosimat plus and on the 856 and 867 measuring modules.