Fluke 87 IV, 89 IV User Manual

0 (0)
Fluke 87 IV, 89 IV User Manual

Page 1 of 3, Document #2139

87/89 Series IV Extended Specifications

©1999 Fluke Corporation

Rev. B-4/99

Fluke 87/89 Series IV

True-rms Digital Multimeter

Extended Specifications

Nominal Specifications

Function

Ranges/Description

DC Voltage

0 to 1000V

DC Current

0 to 10A (20A for 30 seconds)

AC Voltage, True-rms

15 mV to 1000V - 100 kHz bandwidth

AC Current, True-rms

25 μA to 10A (20A for 30 seconds)

Resistance

0 to 30 Mohms

Conductance

0 to 50 Nanosiemens

Capacitance

0.01 nF to 50 mF

Diode Test

3.1V

Temperature

-200°C to 1350°C (-328°F to 2462°F)

Frequency

0.5 Hz to 1000 kHz

Accuracy (Basic DC V)

0.025%

(Basic AC V)

0.4%

Detailed Specifications

Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with relative humidity to 90%. Accuracy specifications are given as ± ([% of reading] + [number of least significant digits])

 

 

 

 

 

Accuracy

 

 

 

 

 

45 Hz-

20 Hz-

1 kHz-

10 kHz-

20 kHz-

Function

Range

Resolution

1 kHz

45 Hz

10 kHz

20 kHz

100 kHz

AC mV 1,2

500.00 mV

0.01 mV

0.4% + 40

2% + 80

5% + 40

5% + 40

6% + 40

 

3000.0 mV

0.1 mV

0.4% + 40

2% + 80

0.4% + 40

1.5% + 40

6% + 40

AC V 1,2

5.0000V

0.0001V

0.4% + 40

2% + 80

0.4% + 40

1.5% + 40

6% + 40

 

50.000V

0.001V

0.4% + 40

2% + 80

0.4% + 40

1.5% + 40

6% + 40

 

500.00V

0.01V

0.4% + 40

2% + 80

0.4% + 40

Not specified

Not specified

 

1000.0V

0.1V

0.4% + 40

2% + 80

0.4% + 40

Not specified

Not specified

dBV

-56 to -6

0.01 dB

0.1 dB

0.2 dB

0.5 dB

0.5 dB

0.5 dB

 

-6 to +34

0.01 dB

0.1 dB

0.2 dB

0.1 dB

0.2 dB

0.5 dB

 

+34 to +60

0.01 dB

0.1 dB

0.2 dB

0.1 dB

Not specified

Not specified

Features

Feature

Description

Dual Displays

50,000 count primary display

 

5,000 count secondary display

Backlight with 2

Bright white backlight for clear readings

brightness selections

in poorly lighted areas

Fast Autorange

Meter automatically selects best range -

 

instantly

AC+DC True RMS, ac

Choices for AC only, AC and DC dual

rms specified to 100 kHz

display, or AC+DC readings

dBm, dBV

User selectable impedance references

 

for dBm

AutoHOLD

Uses Touch Hold® feature to capture readings

Continuity/Open Test

Beeper sounds for Ohms readings below

 

threshold or for momentary

 

open circuit indication

Fast Bar Graph

51 Segments for peaking and nulling

Duty Cycle/Pulse Width

Measure time a signal is on or off in %

 

or milliseconds

MIN MAX/Fast MIN MAX

Record Maximum, Minimum, and Average

with elapsed and

values. Real Time for MAX or MIN, elapsed

Real Time Stamp

time for AVG.

 

Fast MIN MAX captures peaks to 250 μsec.

Closed Case Calibration

No internal adjustments needed

Battery/Fuse

Battery or fuse replacement without

Access Door

voiding calibration

Hi-Impact Overmold Case

Integrated Protective Holster provides

 

superior impact protection for your meter

 

 

 

 

 

Accuracy

 

 

 

 

 

 

 

45 Hz-

20 Hz-

 

 

1 kHz-

 

20 kHz-

Function

Range

Resolution

1 kHz

45 Hz

 

 

20 kHz

 

100 kHz

AC μA

500.00 μA

0.01 μA

0.75% + 20

1% + 20

 

0.75% + 20

 

6% + 4

 

5.000 μA

0.1 μA

0.75% + 5

1% + 5

 

 

0.75% + 10

 

2% + 40

AC mA

50.000 mA

0.001 mA

0.75% + 20

1% + 20

 

0.75% + 20

 

9% + 40

 

400.00 mA3

0.01 mA

0.75% + 5

1% + 5

 

 

1.5% + 10

 

4% + 4

AC A

5.0000A

0.0001A

1.5% + 20

1.5% + 20

 

6% + 40

 

Not specified

 

10.000A4

0.001A

1.5% + 5

1.5% + 5

 

5% + 10

 

Not specified

 

 

 

 

 

 

 

 

 

 

 

Accuracy

 

Accuracy Dual Display AC or AC+DC8

 

 

 

 

 

20 Hz-

45 Hz-

 

1kHz-

Function

Range

Resolution

DC

 

45 Hz

1 kHz

 

20 kHz

DC mV

500.00 mV

0.01 mV

0.03% + 26

 

2%

+ 80

0.6% + 40

 

 

 

3000.0 mV

0.1 mV

0.025% + 5 (89-IV)

2% + 80

0.6% + 40

 

 

 

 

 

0.025% + 10 (87-IV)

 

 

 

 

 

 

6% + 405

DC V

5.0000V

0.0001V

0.025% + 10

 

2%

+ 80

0.5% + 40

 

 

 

50.000V

0.001V

0.03% + 3

 

2%

+ 80

0.5% + 40

 

 

 

500.00V

0.01V

0.1% + 2

 

2%

+ 80

0.5% + 40

 

Not specified

 

1000.0V

0.1V

0.1% + 2

 

2%

+ 80

0.5% + 40

 

Not specified

DC μA

500.00 μA

0.01 μA

0.25% + 207

 

7%

+ 10

7% + 10

 

 

9% + 40

 

5.000 μA

0.1 μA

0.25% + 2

 

1%

+ 10

0.75% + 10

2% + 40

DC mA

50.000 mA

0.001 mA

0.15% + 10

 

1%

+ 10

0.75% + 10

2% + 40

 

400.00 mA3

0.01 mA

0.15% + 2

 

1.5% + 10

1.5% + 10

 

3% + 40

DC A

5.0000A

0.0001A

0.5% + 10

 

7%

+ 20

7% + 20

 

 

12% + 40

 

10.000A4

0.001A

0.5% + 2

 

1.5% + 10

1.5% + 10

 

3% + 40

1For the 5,000 count mode, divide the number of least significant digits (counts) by 10.

2A residual reading of 8 to 80 digits with leads shorted, will not affect stated accuracy above 5% of range.

3500.00 mA for 30 seconds maximum.

44.20A overload for 30 seconds maximum.

5DC 5V range, -3dB typical @ 10 kHz.

6In RF field of 3V/m, add 100 counts from 100 to 120 MHz, 60 counts from 270 to 300 MHz, 40 counts from 320 to 335 MHz.

7In RF field of 3V/m, add 200 counts from 80 to 95 MHz.

8See AC conversion notes for AC mV and V.

Above specifications are subject to change without notice.

Loading...
+ 2 hidden pages