Medical Switching Power Supply
|
XPiQ inc. |
55 Watts |
||
|
PMP55 Series |
|||
|
Intelligent Design Quality Product |
|||
|
|
|
|
|
|
|
• |
CE Marked - LVD & EMC |
|
|
|
|
|
• |
|
|
|
|
Universal Input |
|
|
|
|
• |
|
|
|
|
100% Burn-In |
|
|
|
|
• |
|
|
|
|
Low Leakage Current |
|
|
|
|
• |
|
|
|
|
4000 VAC Isolation |
|
|
|
|
• |
|
|
|
|
Input Surge Protection |
|
|
|
|
• |
|
|
|
|
Line & Neutral Fused |
|
|
|
|
|
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage |
• |
85-264 VAC |
Input Frequency |
• |
47 to 440 Hz |
Input Current |
• |
1.50A (rms) for 115 VAC |
|
|
0.90A (rms) for 230 VAC |
Safety Ground |
• |
50µA max. at 110 VAC 60 Hz |
Leakage Current |
• |
100µA max. at 240 VAC 50 Hz |
Output |
|
|
Total Output Power |
• |
0-55 watts |
Ripple and Noise |
• |
1% peak to peak max. |
Overvoltage |
• |
Provided on output #1 only |
Protection |
|
set at 112-132% of its nominal |
|
|
output voltage |
Overcurrent |
• |
All outputs protected to short |
Protection: |
|
circuit conditions |
Temperature |
• |
All outputs ±0.04%/°C |
Coefficient: |
|
maximum |
Transient Response |
• |
Maximum excursion of 4% or |
|
|
better on all models. Recovers |
|
|
to 1% of final value within |
|
|
500µs after a 25% step load |
|
|
change |
General
Efficiency |
• |
70% minimum, single outputs |
|
|
65% minimum, all other models |
Hold-up time |
• |
10 msec minimum at 115 VAC |
Line Regulation |
• |
±0.5% maximum at full load |
Inrush Current |
• |
15 amps at 115 VAC or |
|
|
30 amps at 230 VAC at 25°C |
|
|
cold start |
Withstand Voltage |
• |
4000 VAC from input to output |
|
|
1500 VAC from input to ground |
|
|
500 VAC from output to ground |
Environmental
Operating |
• |
0°C to +70°C* |
Temperature |
|
|
Storage |
• |
-40°C to +85°C |
Temperature |
|
|
Relative Humidity |
• 5% to 95% non-condensing |
|
MTBF |
• 100,000 hours minimum at full load |
|
|
|
at 25°C ambient |
EMI Requirement |
• Meets conducted and radiated |
|
|
|
limits of EN 55011 (CISPR11) Class B |
Safety Requirements |
• Approved to: (not for patient contact) |
a)UL 2601
b)CSA C22.2 No. 601.1-M90 per CUL
c)IEC 601-1/EN60601-1
*Derate linearly from 100% load at 50°C to 50% load at 70°C
• |
• |