AC/DC Medical Desktop
|
|
|
|
130 Watts |
XPiQ inc. |
||
PMP130 Series |
Intelligent Design Quality Product |
||
|
|
||
|
|
|
|
• |
CE Marked - LVD & EMC |
|
|
|
|
• |
|
|
|
Universal Input |
|
|
|
• |
|
|
|
100% Burn-In |
|
|
|
• |
|
|
|
Low Leakage Current |
|
|
|
• |
|
|
|
4000 VAC Isolation |
|
|
|
• |
|
|
|
Input Surge Protection |
|
|
|
• |
|
|
|
Line & Neutral Fused |
|
|
|
|
|
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage |
• |
85-130 VAC or 180 to 264 VAC |
|
|
auto ranging |
Input Frequency |
• |
47 to 63 Hz |
Input Current |
• |
3.0 A (rms) for 115 VAC |
|
|
1.7 A (rms) for 230 VAC |
Safety Ground |
• |
50 µA max at 110 VAC 60 Hz |
Leakage Current |
• |
100 µA max at 240 VAC 50 Hz |
Output |
|
|
Total Output Power |
• |
See Tables |
Ripple and Noise |
• |
1% pk-pk max |
Overvoltage |
• |
Set at 112-132% of its nominal |
Protection |
|
output voltage |
Overcurrent |
• |
Output protected to short |
Protection: |
|
circuit conditions |
Temperature |
• |
±0.04%/°C |
Coefficient: |
|
|
Transient Response |
• |
Maximum excursion of 4% or |
|
|
better on all models. Recovers |
|
|
to 1% of final value within |
|
|
500 µs after a 25% step load |
|
|
change |
Remote Sense |
• |
Optional, please consult factory |
General
Efficiency |
• |
70% minimum |
|
Hold-up time |
• |
10 msec minimum at 110 VAC |
|
Line Regulation |
• |
±0.5% maximum at full load |
|
Inrush Current |
• |
15 |
A at 115 VAC or |
|
|
30 |
A at 230 VAC at +25°C cold start |
Withstand Voltage |
• |
4000 VAC from input to output |
|
|
|
1500 VAC from input to ground |
|
|
|
500 VAC from output to ground |
|
Insulation Resistance |
• |
10 |
Mohm minimum from |
|
|
output to ground |
|
Switching Frequency |
• |
60 kHz to 75 kHz |
Environmental
Operating |
• |
0 °C to +70 °C, derate linearly from |
Temperature |
|
100% load at +50 °C to 50% load |
|
|
at +70 °C |
Storage |
• |
-40 °C to +85 °C |
Temperature |
|
|
Relative Humidity |
• |
5% to 95% non-condensing |
MTBF |
• |
100,000 hours minimum at full load |
|
|
at +25 °C ambient |
EMC & Safety
EMI Requirements |
• |
In compliance with radiated and |
|
|
conducted limits of CISPR 11 |
|
|
(EN 55011) Class B |
Safety Requirements |
• |
Approved to: (not for patient contact) |
|
|
a) UL 2601 |
|
|
b) CSA C22.2 No. 601.1 |
|
|
c) IEC 601-1/EN60601-1 |
• |
PH: 508 |
429.9883 |
• |
|
|
|
FAX: 800 |
226.2100 |
|
|
www.XPiQ.com |
Email: sales@xpiq.com |
|
|
|
Holliston, MA 01746 USA |
|