Medical Switching Power Supply |
|||
|
|
|
|
|
|
|
XPiQ inc. |
|
|
|
110 Watts |
|
|
|
|
PM110 Series |
|
|
Intelligent Design Quality Product |
|
|
|
|
|
|
|
|
|
|
|
|
|
• |
4000 VAC Isolation |
|
|
|
|
• |
||
|
|
|
|
|
|
|
|
|
|
|
100% Burn-in |
|
|
|
|
|
• |
|
|
|
|
|
Low Safety Leakage Current |
|
|
|
|
|
• |
|
|
|
|
|
Wide Input Voltage 85 to 264 VAC |
|
|
|
|
|
• |
|
|
|
|
|
Meets EMI Requirements of |
|
|
|
|
|
Medical Equipment |
|
|
|
|
|
• |
|
|
|
|
|
Both Line and Neutral Fused for |
|
|
|
|
|
European Requirements |
|
|
|
|
|
|
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage |
• |
85 to 264 |
VAC |
Input Frequency |
• |
47 to 440 |
Hz |
Input Current |
• |
3.20A (rms) for 115 VAC |
|
|
|
1.8A (rms) for 230 VAC |
|
Safety Ground |
• |
50µA max. at 110 VAC 60 Hz |
|
Leakage Current |
• |
100µA max. at 240 VAC 50 Hz |
Output
Total output power: |
• |
0 to 110 watts |
Ripple and noise: |
• |
1% peak to peak maximum |
Overvoltage |
• |
Provided on output #1 only; |
Protection: |
|
set at 112-132% of its nominal |
|
|
output voltage |
Overcurrent |
• |
All outputs protected to short |
Protection: |
|
circuit conditions |
Temperature |
• |
All outputs ±0.04%/°C |
Coefficient: |
|
maximum |
Transient Response: |
• |
Maximum excursion of 4% or |
|
|
better on all models, recovering |
|
|
to 1% of final value 500us after |
|
|
a 25% step load change |
Line Regulation |
• |
±0.5% max. at full load |
General
Efficiency |
• |
65% minimum at 110 watt output |
|
Hold-up time |
• |
12 msec minimum at nominal input |
|
Inrush Current |
• |
15 amps at 115 VAC or 30 amps at |
|
|
|
230 VAC at 25°C cold start |
|
Withstand Voltage |
• |
4000 |
VAC from input to output |
|
|
1500 |
VAC from input to ground |
|
|
500 VAC from output to ground |
|
Insulation Resistance |
• |
10 Mohm minimum from |
|
|
|
output to ground |
Environmental
Operating |
• |
0°C to +70°C* |
Temperature |
|
|
Storage |
• |
-40°C to +85°C |
Temperature |
|
|
Relative Humidity |
• 5% to 95% non-condensing |
|
MTBF |
• 100,000 hours minimum at full load |
|
|
|
at 25°C ambient |
EMI Requirement |
• Meets the conducted limits of |
|
|
|
EN 55011 (CISPR11) Level B |
Safety Requirements |
• Approved to: (Not for patient contact) |
a)UL2601
b)CSA C22.2 No.601.1-M90 per CUL
c)IEC 601-1/EN60601-1 (per TUV)
*Derate linearly from 100% load at 50°C to 50% load at 70°C
• |
• |
www.XP-iQ.com
PH: 508 429.9883
FAX: 800 226.2100 Email: sales@xp-iq.com Holliston, MA 01746 USA