XPiQ inc PCM90PD00, PCM90PD01, PCM90PD02, PCM90PD03, PCM90PD04 Datasheet

...
0 (0)
XPiQ inc PCM90PD00, PCM90PD01, PCM90PD02, PCM90PD03, PCM90PD04 Datasheet

PCM90.qxd 6/6/03 3:54 pm Page 1

AC/DC Medical Desktop

 

 

 

 

 

 

90 Watts

 

 

 

XPiQ inc.

PCM90 Series

 

 

 

Intelligent Design Quality Product

 

 

 

 

 

 

 

 

 

 

Active PFC, Meets EN61000-3-2, -3

 

 

 

 

 

 

 

 

 

 

 

 

Low Leakage Current

 

 

 

 

PRELIMINARY

 

 

 

 

 

 

 

 

 

Single, Dual and Triple Outputs

 

 

 

 

 

 

 

 

 

 

2003

 

 

 

 

Both Line and Neutral Fused

 

 

 

 

for European Applications

 

 

 

 

Q4

 

 

 

 

 

 

 

 

 

4000 VAC Isolation

 

 

 

 

 

 

 

 

 

Medical Safety Agency Approvals

 

 

 

 

 

 

 

Specification

Input

Input Voltage

90-264 VAC

Input Frequency

47 to 63 Hz

Power Factor

0.99 at 100 VAC or 0.94 at

 

 

240 VAC

Input Current

1.3 A maximum

Inrush Current

15

A max at 115 VAC or

 

 

30

A max at 230 VAC cold start

Safety Ground

0.3 mA max at 115 VAC, 60 Hz

Leakage Current

0.3 mA max at 230 VAC, 50 Hz

Output

Output Voltage

See Tables

Output Power

90 Watts

Minimum Load

0 minimum load required for single

 

 

output models

Line Regulation

±1% maximum at full load

Load Regulation

±5% maximum, except 3.3 V model

 

 

is ±7%

Ripple and Noise

1% pk-pk maximum

Transient Response

Maximum excursion of 4% or

 

 

better on all models. Recovers

 

 

to 1% of final value within

 

 

500 µs after a 25% step load change

Temperature

All outputs ±0.04%/°C

coefficient:

 

maximum

Hold-up time

16 msec minimum

Overvoltage

Provided on output #1 only

Protection

 

112 to 132% recycle input to reset

Overcurrent

All outputs 110-150%

Protection:

 

with auto recovery

General

Efficiency

83% typical

MTBF

100,000 hours minimum at full load

 

 

at +25 °C ambient

Isolation Voltage

4000 VAC from input to output

 

 

1500 VAC from input to ground

 

 

500 VAC from output to ground

Insulation Resistance

100 Mohm minimum from

 

 

output to ground

Size

6.63” x 3.07” x 1.97”

Weight

800 grams approx

Environmental

Operating

0 °C to +70 °C

Temperature

 

Derate linearly from 100% load at

 

 

+40 °C to 50% load at +70 °C

Humidity

5 to 95% RH, non-condensing

Storage

-40 °C to +85 °C

Temperature

 

 

EMC & Safety

EMI Requirement

EN55011 (CISPR11) Class B

 

 

conducted and radiated

Safety Approvals

UL2601-1, CSA C22.2 No 601.1

 

 

per cUL EN60601-1 CE Mark,

 

 

IEC 60601-1

* Preliminary datasheet - see website for current specs

 

PH: 508

429.9883

 

 

 

 

 

 

 

 

 

 

 

FAX: 800

226.2100

 

 

www.XPiQ.com

 

 

Email: sales@xpiq.com

 

 

 

 

Holliston, MA 01746 USA

 

 

 

 

 