PCM90.qxd 6/6/03 3:54 pm Page 1
AC/DC Medical Desktop
|
|
|
|
|
|
90 Watts |
|
|
|
XPiQ inc. |
|
PCM90 Series |
|
|
|
Intelligent Design Quality Product |
|
|
|
|
|
||
|
|
|
|
|
|
|
• |
Active PFC, Meets EN61000-3-2, -3 |
|
||
|
|
• |
|
||
|
|
|
|
|
|
|
|
|
Low Leakage Current |
|
|
|
|
|
PRELIMINARY |
|
|
|
|
|
|
• |
|
|
|
|
Single, Dual and Triple Outputs |
|
|
|
|
|
|
• |
|
|
|
|
|
2003 |
|
|
|
|
Both Line and Neutral Fused |
|
|
|
|
|
for European Applications |
|
|
|
|
|
Q4 |
• |
|
|
|
|
|
|
|
|
|
|
4000 VAC Isolation |
|
|
|
|
|
|
• |
|
|
|
|
Medical Safety Agency Approvals |
|
|
|
|
|
|
|
|
Specification
Input
Input Voltage |
• |
90-264 VAC |
|
Input Frequency |
• |
47 to 63 Hz |
|
Power Factor |
• |
0.99 at 100 VAC or 0.94 at |
|
|
|
240 VAC |
|
Input Current |
• |
1.3 A maximum |
|
Inrush Current |
• |
15 |
A max at 115 VAC or |
|
|
30 |
A max at 230 VAC cold start |
Safety Ground |
• |
0.3 mA max at 115 VAC, 60 Hz |
|
Leakage Current |
• |
0.3 mA max at 230 VAC, 50 Hz |
Output
Output Voltage |
• |
See Tables |
Output Power |
• |
90 Watts |
Minimum Load |
• |
0 minimum load required for single |
|
|
output models |
Line Regulation |
• |
±1% maximum at full load |
Load Regulation |
• |
±5% maximum, except 3.3 V model |
|
|
is ±7% |
Ripple and Noise |
• |
1% pk-pk maximum |
Transient Response |
• |
Maximum excursion of 4% or |
|
|
better on all models. Recovers |
|
|
to 1% of final value within |
|
|
500 µs after a 25% step load change |
Temperature |
• |
All outputs ±0.04%/°C |
coefficient: |
|
maximum |
Hold-up time |
• |
16 msec minimum |
Overvoltage |
• |
Provided on output #1 only |
Protection |
|
112 to 132% recycle input to reset |
Overcurrent |
• |
All outputs 110-150% |
Protection: |
|
with auto recovery |
General
Efficiency |
• |
83% typical |
MTBF |
• |
100,000 hours minimum at full load |
|
|
at +25 °C ambient |
Isolation Voltage |
• |
4000 VAC from input to output |
|
|
1500 VAC from input to ground |
|
|
500 VAC from output to ground |
Insulation Resistance |
• |
100 Mohm minimum from |
|
|
output to ground |
Size |
• |
6.63” x 3.07” x 1.97” |
Weight |
• |
800 grams approx |
Environmental
Operating |
• |
0 °C to +70 °C |
Temperature |
|
Derate linearly from 100% load at |
|
|
+40 °C to 50% load at +70 °C |
Humidity |
• |
5 to 95% RH, non-condensing |
Storage |
• |
-40 °C to +85 °C |
Temperature |
|
|
EMC & Safety
EMI Requirement |
• |
EN55011 (CISPR11) Class B |
|
|
conducted and radiated |
Safety Approvals |
• |
UL2601-1, CSA C22.2 No 601.1 |
|
|
per cUL EN60601-1 CE Mark, |
|
|
IEC 60601-1 |
* Preliminary datasheet - see website for current specs
• |
|
PH: 508 |
429.9883 |
• |
||
|
||||||
|
||||||
|
|
|
|
|
||
|
|
|
|
FAX: 800 |
226.2100 |
|
|
www.XPiQ.com |
|
|
Email: sales@xpiq.com |
|
|
|
|
|