AC/DC Medical Desktop
|
|
|
|
110 Watts |
XPiQ inc. |
||
PMP110 Series |
Intelligent Design Quality Product |
||
|
|
||
|
|
|
|
• |
CE Marked - LVD & EMC |
|
|
|
|
• |
|
|
|
Universal Input |
|
|
|
• |
|
|
|
100% Burn-In |
|
|
|
• |
|
|
|
Low Leakage Current |
|
|
|
• |
|
|
|
4000 VAC Isolation |
|
|
|
• |
|
|
|
Input Surge Protection |
|
|
|
• |
|
|
|
Line & Neutral Fused |
|
|
|
|
|
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage |
• |
85-264 VAC |
Input Frequency |
• |
47 to 440 Hz |
Input Current |
• |
3.20 A (rms) for 115 VAC |
|
|
0.90 A (rms) for 230 VAC |
Safety Ground |
• |
50 µA max at 110 VAC 60 Hz |
Leakage Current |
• |
100 µA ma. at 240 VAC 50 Hz |
Output |
|
|
Total Output Power |
• |
0-110 watts |
Ripple and Noise |
• |
1% pk-pk max |
Overvoltage |
• |
Provided on output #1 only |
Protection |
|
set at 112-132% of its nominal |
|
|
output voltage |
Overcurrent |
• |
All outputs protected to short |
Protection: |
|
circuit conditions |
Temperature |
• |
±0.04%/°C |
Coefficient: |
|
|
Transient Response |
• |
Maximum excursion of 4% or |
|
|
better on all models. Recovers |
|
|
to 1% of final value within |
|
|
500 µs after a 25% step load |
|
|
change |
General
Efficiency |
• |
65% minimum |
||
Hold-up time |
• |
12 msec minimum at 115 VAC |
||
Line Regulation |
• |
±0.5% maximum at full load |
||
Inrush Current |
• |
15 A at 115 |
VAC or |
|
|
|
30 A at 230 |
VAC at +25°C |
|
|
|
cold start |
|
|
Withstand Voltage |
• |
4000 |
VAC from input to output |
|
|
|
1500 |
VAC from input to ground |
|
|
|
500 VAC from output to ground |
Environmental
Operating |
• |
0 °C to +70 °C, derate linearly from |
Temperature |
|
100% load at +50 °C to 50% load |
|
|
at +70 °C |
Storage |
• |
-40 °C to +85 °C |
Temperature |
|
|
Relative Humidity |
• |
5% to 95% non-condensing |
MTBF |
• |
100,000 hours minimum at full load |
|
|
at +25 °C ambient |
EMC & Safety
EMI Requirement |
• |
Meets conducted and radiated |
|
|
limits of EN 55011 (CISPR11) Class B |
Safety Requirements |
• |
Approved to: (not for patient contact) |
|
|
a) UL 2601 |
|
|
b) CSA C22.2 No. 601.1-M90 per CUL |
|
|
c) IEC 601-1/EN60601-1 |
• |
|
PH: 508 |
429.9883 |
• |
||
|
||||||
|
||||||
|
|
|
|
FAX: 800 |
226.2100 |
|
|
www.XPiQ.com |
|
|
Email: sales@xpiq.com |
|
|
|
|
|
Holliston, MA 01746 USA |
|
||
|
|
|
|