XPiQ inc PMP110-23, PMP110-31, PMP110-32, PMP110-40, PMP110-41 Datasheet

...
0 (0)
XPiQ inc PMP110-23, PMP110-31, PMP110-32, PMP110-40, PMP110-41 Datasheet

AC/DC Medical Desktop

 

 

 

 

110 Watts

XPiQ inc.

PMP110 Series

Intelligent Design Quality Product

 

 

 

 

 

 

CE Marked - LVD & EMC

 

 

 

 

 

 

Universal Input

 

 

 

 

 

 

100% Burn-In

 

 

 

 

 

 

Low Leakage Current

 

 

 

 

 

 

4000 VAC Isolation

 

 

 

 

 

 

Input Surge Protection

 

 

 

 

 

 

Line & Neutral Fused

 

 

 

 

 

Specification

All specifications typical at nominal line, full load and 25°C

Input

Input Voltage

85-264 VAC

Input Frequency

47 to 440 Hz

Input Current

3.20 A (rms) for 115 VAC

 

 

0.90 A (rms) for 230 VAC

Safety Ground

50 µA max at 110 VAC 60 Hz

Leakage Current

100 µA ma. at 240 VAC 50 Hz

Output

 

 

Total Output Power

0-110 watts

Ripple and Noise

1% pk-pk max

Overvoltage

Provided on output #1 only

Protection

 

set at 112-132% of its nominal

 

 

output voltage

Overcurrent

All outputs protected to short

Protection:

 

circuit conditions

Temperature

±0.04%/°C

Coefficient:

 

 

Transient Response

Maximum excursion of 4% or

 

 

better on all models. Recovers

 

 

to 1% of final value within

 

 

500 µs after a 25% step load

 

 

change

General

Efficiency

65% minimum

Hold-up time

12 msec minimum at 115 VAC

Line Regulation

±0.5% maximum at full load

Inrush Current

15 A at 115

VAC or

 

 

30 A at 230

VAC at +25°C

 

 

cold start

 

Withstand Voltage

4000

VAC from input to output

 

 

1500

VAC from input to ground

 

 

500 VAC from output to ground

Environmental

Operating

0 °C to +70 °C, derate linearly from

Temperature

 

100% load at +50 °C to 50% load

 

 

at +70 °C

Storage

-40 °C to +85 °C

Temperature

 

 

Relative Humidity

5% to 95% non-condensing

MTBF

100,000 hours minimum at full load

 

 

at +25 °C ambient

EMC & Safety

EMI Requirement

Meets conducted and radiated

 

 

limits of EN 55011 (CISPR11) Class B

Safety Requirements

Approved to: (not for patient contact)

 

 

a) UL 2601

 

 

b) CSA C22.2 No. 601.1-M90 per CUL

 

 

c) IEC 601-1/EN60601-1

 

PH: 508

429.9883

 

 

 

 

 

 

FAX: 800

226.2100

 

 

www.XPiQ.com

 

 

Email: sales@xpiq.com

 

 

 

 

Holliston, MA 01746 USA

 

 

 

 

 