145-166 1/5/02 11:12 am Page 13
|
|
Medical Switching Power Supply |
|||
|
|
|
|
|
|
|
XPiQ inc. |
|
|
|
200 Watts |
|
|
|
|
PM200 Series |
|
|
Intelligent Design Quality Product |
|
|
|
|
|
|
|
|
|
|
|
|
|
• |
4000 VAC Isolation |
|
|
|
|
• |
||
|
|
|
|
|
|
|
|
|
|
|
100% Burn-in |
|
|
|
|
|
• |
|
|
|
|
|
Low Safety Leakage Current |
|
|
|
|
|
• |
|
|
|
|
|
Short Circuit Protection |
|
|
|
|
|
• |
|
|
|
|
|
Meets EMI Requirements of |
|
|
|
|
|
Medical Equipment |
|
|
|
|
|
• |
|
|
|
|
|
Both Line and Neutral Fused for |
|
|
|
|
|
European Requirements |
|
|
|
|
|
|
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage |
• 90 to 130 VAC or 180 to 264 VAC |
|
autoranging |
Input Frequency |
• 47 to 440 Hz |
Input Current |
• 4.5A (rms) for 115 VAC |
|
2.6A (rms) for 230 VAC |
Safety Ground |
• 50µA max. at 110 VAC 60 Hz |
Leakage Current |
• 100µA max. at 240 VAC 50 Hz |
Output
Total output power: |
• 15 to 200 watts |
|
Ripple and noise: |
• 2% peak to peak maximum |
|
Overvoltage |
• Provided on output #1 only; |
|
Protection: |
|
set at 112-132% of its nominal |
|
|
output voltage |
Overcurrent |
• All outputs protected to short |
|
Protection: |
|
circuit conditions |
Temperature |
• |
All outputs ±0.04%/°C |
Coefficient: |
|
maximum |
Transient Response: • |
Maximum excursion of 4% or |
|
|
|
better on all models, recovering |
|
|
to 1% of final value 500us after |
|
|
a 25% step load change |
Line Regulation |
• ±0.5% max. at full load |
General
Efficiency |
• |
70% minimum at 200 watt output |
|
Hold-up time |
• |
20 msec minimum at nominal input |
|
Inrush Current |
• |
60A max at 25°C cold start |
|
Withstand Voltage |
• |
4000 |
VAC from input to output |
|
|
1500 |
VAC from input to ground |
|
|
500 VAC from output to ground |
Environmental
Operating |
• |
0°C to +70°C* |
Temperature |
|
|
Storage |
• |
-40°C to +85°C |
Temperature |
|
|
Relative Humidity |
• |
5% to 95% non-condensing |
MTBF |
• |
85,000 hours minimum at full load |
|
|
at 25°C ambient |
EMI Requirement |
• |
Meets the conducted limits of |
|
|
EN 55011 (CISPR11) Level B |
Safety Requirements • Approved to: (Not for patient contact)
a)UL2601
b)CSA C22.2 No.601.1-M90 per CUL
c)IEC 601-1/EN60601-1 (per TUV)
*Derate linearly from 100% load at 50°C to 50% load at 70°C
• |
PH: 508 |
429.9883 |
• |
|
|
||
|
FAX: 800 |
226.2100 |
|
157 www.XP-iQ.com |
Email: sales@xp-iq.com |
||
Holliston, MA 01746 USA |