The pneumatic grip controller provides open/close function, air pressure
regulation, and flow control for pneumatic grips.
For grip operation, the unit comes with a foot switch. The foot switch sequence is
user-selectable for adaptable and efficient specimen loading. The controller
comes with temperature-resistant Teflon tubing for grip connections.
This section provides information about safety issues that pertain to pneumatic
grip controllers in general and a definition for the graphical hazard labeling that
is affixed to your product.
When you prepare to operate the pneumatic grip controller, consider the
following:
•Failure to use this equipment in accordance with the manufacturer’s
•Ensure the controller is off before connecting the pneumatic tubing.
•Always wear protective devices such as eye protectioin.
•Ensure air supply is properly filtered (see “Specifications” on page 10).
recommendations can result in unpredicatbale grip operation, which can
cause equipment damage or personal injury.
Hazard Placards
•Follow all precautions and safety information associated with you
pneumatic grips.
Hazard placards contain specific safety information and are affixed directly to the
system so they are plainly visible.
The following hazard action icons are used on MTS products. These icons
indicate recommended actions that will minimize the chance of personal injury or
equipment damage. They are typically placed on or near the area of concern.
They are intended to alert the user to actions that should be taken.
I
CONDESCRIPTION
General waring. Possible hazard(s) exists in
the vicinity where this icon is located. See
the manual for additional information.
Do not start, operation, or service pneumatic
grip controller until you read and understand
operator’s manual.
1. Disconnect power to the pneumatic grip controller.
NoteTo release the tubing, press in the locking collar (see 3 on the previous
page) and pull the tubing out of the connection.
2. Using the supplied tubing and cutter, cut the tubing to the appropriate
length.
3. Push the tubing into the four grip connections firmly.
4. Connect the foot switch to the Switch Connection.
5. Firmly push the tubing into the Compressed Air Input.
6. Connect E-stop Input.
7. Connect E-stop Output. This connector allows you to pass the E-stop signal
to another device. If the pneumatic grip controller is the last device in the Estop chain, this connector must be terminated (jumpered).