Mindray DS USA, Inc., 2008. All rights reserved. Contents of this publication may not be reproduced in any
orm without permission of Mindray DS USA, Inc.
0070-10-0604-01Duo Service Manual
Table of Contents
Foreword....................................................................................................................................................... iii
Warnings, Precautions And Notes ....................................................................................................................iii
Theory of Operation ......................................................................................................... 1 - 1
Warnings and Guidelines................................................................................................................................ 2 - 2
Test Equipment and Special Tools Required........................................................................................................ 2 - 2
Calibration and System Checks ........................................................................................................................ 2 - 3
Device Appearance and Installation Checks................................................................................................ 2 - 3
Unit of Measure Mode ............................................................................................................................. 2 - 3
Software Version Mode............................................................................................................................ 2 - 4
Safety Test ..................................................................................................................................................... 2 - 5
Test Equipment ........................................................................................................................................ 2 - 5
Case Leakage ......................................................................................................................................... 2 - 5
Test Equipment ........................................................................................................................................ 2 - 6
Test Procedure......................................................................................................................................... 2 - 6
Test Equipment ........................................................................................................................................ 2 - 8
Test Procedure......................................................................................................................................... 2 - 8
Parts ................................................................................................................................ 3 - 1
Power Supply Removal............................................................................................................................. 4 - 7
Duo™ Service Manual0070-10-0604-01i
Table of Contents
This page intentionally left blank.
ii0070-10-0604-01Duo™ Service Manual
ForewordIntroduction
Foreword
This service manual gives a detailed description of the Duo Portable Patient Monitor,
including circuit descriptions, test and calibration procedures, and spare parts listings. This
manual is intended as a guide for technically qualified personnel during repair, testing, or
calibration procedures.
Warnings, Precautions And Notes
Please read and adhere to all warnings, precautions, and notes listed here and in the
appropriate areas throughout this manual.
A WARNING is provided to alert the user to potential serious outcomes (death, injury, or
serious adverse events) to the patient or the user.
A CAUTION is provided to alert the user to use special care necessary for the safe and
effective use of the device. They may include actions to be taken to avoid effects on patients
or users that may not be potentially life threatening or result in serious injury, but about which
the user should be aware. Cautions are also provided to alert the user to adverse effects on
this device of use or misuse and the care necessary to avoid such effects.
A NOTE is provided when additional general information is applicable.
Duo™ Service Manual0070-10-0604-01iii
IntroductionWarnings, Precautions And Notes
This page intentionally left blank.
iv0070-10-0604-01Duo™ Service Manual
1.0
Theory of Operation
1.1Introduction
The Duo is a compact, lightweight, portable patient monitor intended for monitoring the
following vital signs: blood pressure, SpO
patients. The Duo monitor can be powered by either the internal rechargeable Lithium-Ion
battery or external 100~240 volt 50/60 Hz AC.
(optional), and pulse rate on adult and pediatric
2
Duo™ Service Manual0070-10-0604-011 - 1
Hardware OverviewTheory of Operation
1.2Hardware Overview
Patient Connectors
Fan
Driver
AC Input
Power Supply
Lithium-Ion
Battery
FIGURE 1-1 Interconnection Block diagram
*The Isolated SpO2 Power Supply is present with optional Masimo and Nellcor SpO2 ONLY.
**The RS232 - TTL Inverter and Serial Interface Connector is used for software updates by Service
1.2.1Power Supply Board
Overview
The AC/DC converter transforms the AC input voltage (90 - 264 vac 50/60 Hz) to a DC
voltage used to charge the internal Lithium-Ion battery and supply power to the +12 vdc and
+3.3 vdc DC/DC converters. The battery charging circuit will actively charge the battery
while the Duo is connected to an AC source. Battery charging takes place whether the
monitor is on, off, or in use. The power supply will automatically switch to the internal battery
if an AC source is not present.
External
Interface
NIBP Module
Keypad Switches
CPU / Display pcb
RS232 - TTL
Inverter**
Module
SpO
2
Isolated SpO
Power Supply*
2
Serial
Interface**
NOTE:The power supply board MUST be connected to a resistive
load to operate properly and to avoid damage due to an
over current condition.
Power ON / OFF
Control Circuit
AC Input
AC / DC
Converter
Battery and Charging
Management Circuit
1 - 20070-10-0604-01Duo™ Service Manual
3.3v DC/DC
Converter
12v DC/DC
Converter
Theory of OperationHardware Overview
FIGURE 1-2 Power Supply Block Diagram
Power Supply Voltage Test Points:
LocationFunction
Measure across C5Primary Rectified Voltage (DC). Range: 105 - 374 volts.
C5 Negative LeadPrimary Ground.
Measure across Q1 pin1 and
C5 Negative Lead
Measure across C12+10.5 vdc input for Fan Driver board.
C12 Negative LeadSecondary Ground.
C47 Positive Lead+5 vdc supply for Power On/Off Control.
C50 Positive Lead+3.3 vdc output.
C68 Positive Lead+12 vdc output.
110k Hz Drive frequency.
Duo™ Service Manual0070-10-0604-011 - 3
Hardware OverviewTheory of Operation
1.2.2Fan Driver Board Overview
The Fan Drive Board is active during the battery charging cycle. The Temperature Detector
senses the temperature of the heat sink of the Secondary Rectifier diode and turns on the fan
when the heat sink reaches a certain temperature.
Temperature
Detector
DC
Input
FIGURE 1-3 Fan Driver Block Diagram
Fan Driver
Circuit
Fan Driver Board Test Points
LocationFunction
Measure across C202+5 vdc Fan power.
C202 Negative LeadGround.
Measure between Q202 pin1 and GroundDrive frequency when the Fan is activated.
1.2.3CPU/Display Board Overview
The CPU/Display board controls the SpO2 Module and NIBP Module through
communications via UART devices. The CPU Board receives user commands from the
Keypad. The power supply board provides +3.3 vdc and +12 vdc to the CPU board. These
voltages are monitored by an A/D converter located on the CPU board. The CPU also
controls an integral LED display array and indicator LEDs. The main processor has a built-in
serial port that is used to load software. The processor also uses a FPGA to communicate
with the NIBP Module, the optional SpO
on JP1) and to drive the LED arrays and indicators.
Fan
module (SpO2 presence is detected via a jumper
2
1 - 40070-10-0604-01Duo™ Service Manual
Theory of OperationHardware Overview
A/D Converter
CPU
To
External
Interface
To
SpO2
Module
Serial Interface
To
NIBP
Module
FPGA
FIGURE 1-4 CPU/Display Board Block Diagram
CPU Board Test Points
LocationFunction
VPP in+12 vdc power supply
VDD in+3.3 vdc power supply
BVBattery voltage
5B in+5 vdc power supply
RTC/ E
2
PROM
User Interface
LED Array
Display
Keypad
Indicator
LEDs
ADV+10.5 vdc ADV out
RSTCPU reset line
XT211.0592 MHz clock
1.2.4NIBP Module Overview
The Duo monitor determines non-invasive blood pressure using the oscillometric method. The
blood pressure cuff is inflated until the pressure in the cuff is sufficient to block blood flow in
the brachial artery. As the cuff begins to deflate, blood beginning to flow through the artery
will cause the artery to pulsate. These pulsations are transmitted through the blood pressure
cuff and connecting hose to the pressure transducer in the NIBP module. The output of the
pressure transducer is an analog pulsating signal. This signal is filtered by a high-pass filter
and then amplified. The amplified analog signal is then converted to a digital signal. The
digital signal is then processed to determine the systolic, diastolic, and mean pressures as
well as heart rate.
Duo™ Service Manual0070-10-0604-011 - 5
Hardware OverviewTheory of Operation
1.2.5SpO2 Overview
Pulse oximetry (SpO2) measurement is used to determine the oxygen saturation level of the
patient's blood. The SpO2 numeric display indicates the amount of hemoglobin that has
bonded with oxygen molecules to form oxyhemoglobin. By analyzing the pulse in the
fingertip using specified algorithm and consulting the clinical data table, we can obtain the
SpO
value. The SpO2 sensor consists of two LEDs (one red and one infrared) and a
2
photodetector. The two LEDs are alternately lighted at a precise frequency. When the
capillary vessels of the fingertip are filled, a certain amount of light from the LEDs is
absorbed by blood in the capillaries. The remaining red and infrared light is then picked up
by the photodetector. The photodetector detects the varying light intensity due to pulsations
and transmits the changing light intensity in the form of changing electronic signals. The
amount of light absorption is then compared to the known fixed LED output by the SpO
board. The pulse rate is counted and the SpO
value is determined by using an algorithm
2
contained in the software on the SpO2 board.
2
1 - 60070-10-0604-01Duo™ Service Manual
2.0
Calibration and Performance
Verification
2.1Introduction
The following procedures are provided to verify the proper operation of the Duo monitor. A
menu driven interface is used to execute all verification tests. Performance tests should be
performed at least once per year and after any preventive maintenance or repair has been
performed.
Duo™ Service Manual0070-10-0604-012 - 1
Warnings and GuidelinesCalibration and Performance Verification
2.2Warnings and Guidelines
In the event that the instrument cover is removed, observe the following warnings and
guidelines:
1. Do not short component leads together.
2. Perform all steps in the exact order they are given.
3. Use extreme care when reaching inside the opened instrument. Do not contact exposed
metal parts that may become ‘‘live’’.
4. Read through each step in the procedure so it is understood prior to performing the step.
2.3Test Equipment and Special Tools Required
• 0-300 mmHg Digital or Mercury manometer with bulb and valve
• 500 cc Test Chamber/Dummy Cuff. P/N 0138-00-0001-03
•DVM
•SpO
• NIBP simulator
• Safety Analyzer (Dempsey model 431 or equivalent)
• Oscilloscope
• Laptop or PC (software upgrade)
simulator
2
2 - 20070-10-0604-01Duo™ Service Manual
Calibration and Performance VerificationCalibration and System Checks
2.4Calibration and System Checks
2.4.1Device Appearance and Installation Checks
Inspect the Duo monitor to ensure that:
• The outer housing is clean and has no scratches or cracks
• When the device is gently shaken, there are no loose components
• All keys are smooth and free for operation
• Labels are complete, clean, and accurate
• All connectors/accessory modules are installed securely
Ensure monitor is securely fastened to its rolling stand (if used)
2.4.2Maintenance Functions/Non-Monitoring Modes
• When entering the maintenance functions/non-monitoring mode, the monitor will perform
a self-test, however the verification of functional LEDs will not be displayed
• In the maintenance mode, the standby mode will not be active
• In the maintenance mode, the auto-shutoff will activate if no key is pressed for a period of
15 minutes.
2.4.3Unit of Measure Mode
The unit of measure mode is used to change between mmHg and kPa. To access the Unit of
Measure mode:
1. Turn the monitor off.
2. Simultaneously press and hold the POWER and CLEAR buttons.
3. Press the CLEAR button to cycle through the unit of measure choices.
4. Once the desired unit of measure is displayed, turn the Duo off to save that setting.
Duo™ Service Manual0070-10-0604-012 - 3
Calibration and System ChecksCalibration and Performance Verification
2.4.4Software Version Mode
Use the following procedure to view the software version.
1. Ensure that the Duo is powered OFF.
2. Press and hold the Patient Size key.
3. While continuing to hold the Patient Size key, press and hold the
Power ON/OFF key for two (2) seconds until the Duo beeps.
4. Release both keys.
5. After an additional 2-second delay, Duo will display “100” in the Pulse Rate tile and a
number in the Information Codes tile.
6. Pressing the Clear key will cause the number displayed in the Pulse Rate tile to cycle
through a sequence of four numbers indicating which software version is being
displayed in the Information Codes tile as shown in the following table.
PULSE RATE TILEINFORMATION CODES TILE SHOWS
100Host Software Revision Level
200NIBP Software Revision Level
300SpO2 Software Revision Level
7. To return to normal operation, press the Power ON/OFF key for two (2) seconds to
turn the Duo off.
8. You may then turn the Duo back on to resume normal operation.
NOTE:The Duo cannot be placed directly back into normal
monitoring mode from Software Version Mode. It must first
be powered OFF.
2 - 40070-10-0604-01Duo™ Service Manual
Calibration and Performance VerificationSafety Test
2.5Safety Test
2.5.1Test Equipment
• Safety Analyzer (Dempsey model 431 or equivalent)
2.5.2Case Leakage
1. Plug the line cord of the unit into the safety analyzer.
2. Connect the case ground lead of the analyzer to the equipotential lug of the Duo
monitor.
3. Perform the leakage tests under the following conditions:
a. Case grounded:
Normal polarity
Normal polarity with open neutral
Reverse polarity
b. Case ungrounded:
Normal polarity
Normal polarity with open neutral
Reverse polarity
4. Verify the current reading is <100 uA under normal operating conditions; <300 uA
under single fault conditions for 120 VAC and <500 uA under single fault conditions for
230 VAC.
Duo™ Service Manual0070-10-0604-012 - 5
NIBP CalibrationCalibration and Performance Verification
2.6NIBP Calibration
2.6.1Test Equipment
• NIBP simulator
• NIBP test chamber/dummy cuff
• Manometer with bulb
2.6.2Test Procedure
2.6.2.1Transducer Accuracy
1. Connect the 500 cc Test Chamber and calibrated manometer via a ‘‘T’’ fitting to the
NIBP fitting on the Duo monitor under test.
2. Ensure the Duo is not turned on. Simultaneously, press and hold the POWER and NIBP
START/STOP buttons.
When the monitor enters the NIBP Calibration Mode, message code 525 will be
displayed in the Information Codes window. Release the buttons simultaneously.
3. Momentarily press the NIBP START/STOP button to start the NIBP calibration. Vent the
Test Chamber and verify the Duo and the manometer read zero. Using the bulb,
pressurize the test chamber to 50 mmHg and verify the Duo reading agrees with the
manometer +/- 3 mmHg. Using the bulb, increase the pressure to 200 mmHg and verify
the Duo reading agrees with the manometer +/- 3 mmHg.
2.6.2.2Pneumatic Leak Test
1. Connect the 500 cc test chamber to the NIBP fitting on the Duo monitor under test.
2. From the NIBP Calibration Mode (code 525) momentarily press the CLEAR button on
the Duo keypad. The Duo will then switch to the Pneumatic Test Mode and will display
message code 550 in the Information Codes window.
3. Momentarily press the NIBP START/STOP button to start the leak test. The Duo under
test will automatically pressurize the test chamber to approximately 180 mmHg.
4. After approximately 20 seconds, the Duo under test will vent the pressure in the test
chamber and display a message code E06 (Pass) or E07 (Fail) in the systolic window.
2.6.2.3Dynamic Repeatability Test
1. Restart unit and allow it to enter normal operating mode.
2. Use polyurethane tubing to connect the Duo monitor to a calibrated NIBP simulator and
the 500 cc test chamber/dummy cuff via a ‘‘T’’ fitting.
3. Select Adult patient size for both the Duo under test and the NIBP simulator.
4. Select a target simulated blood pressure within the ‘‘normal’’ range on the simulator.
5. Take 10 successive NIBP readings and compare the systolic, diastolic, mean and heart
rate readings for consistency. Readings should not deviate more than +/- 5 mmHg for
the NIBP readings and +/- 2 bpm or 2%, whichever is greater for heart rate.
2 - 60070-10-0604-01Duo™ Service Manual
Calibration and Performance VerificationNIBP Calibration
NOTE:The actual measured values displayed on the Duo monitor
may not compare with the selected target pressure on the
simulator. This test is intended to confirm the REPEATABILITY,
not accuracy, of dynamic NIBP readings. Accuracy can only
be confirmed by performing the NIBP Calibration outlined in
section 2.6 of this manual.
Duo™ Service Manual0070-10-0604-012 - 7
SpO2 Ver i f i c a t i o nCalibration and Performance Verification
2.7SpO2 Verification
2.7.1Test Equipment
•SpO2 simulator
2.7.2Test Procedure
1. Connect the appropriate SpO2 probe connector to the Duo monitor.
2. Connect the SpO
3. Set the simulator target values to:
= 98%
SpO
2
Pulse Rate = 70
4. Verify that the displayed SpO2 and pulse rate values on the Duo monitor are +/- 2% of
the simulator target values.
5. Change the simulator values.
6.
Verify the displayed values on the
probe to the SpO2 simulator.
2
Duo
monitor are equal to the simulator values +/- 2%.
2 - 80070-10-0604-01Duo™ Service Manual
3.0
Parts
3.1Introduction
This section contains exploded views of the Duo monitor, internal modules, and parts list.
Duo Service Manual0070-10-0604-013 - 1
IntroductionParts
9
SpO2 Module
See SpO
4
Detail
2
6
5
8A
8
7
5
1
2
FIGURE 3-1 Duo Exploded View
3
3 - 20070-10-0604-01Duo Service Manual
PartsIntroduction
18
19
10
10
11
11
12
12
13
13
12
12
FIGURE 3-2 Masimo SpO2 Detail
14
14
Duo Service Manual0070-10-0604-013 - 3
IntroductionParts
18
16a
12
16
15
10
FIGURE 3-3 Nellcor SpO2 Detail
17
17a
3 - 40070-10-0604-01Duo Service Manual
PartsIntroduction
19
5
21
20
3
4
5
11
5
22
Battery Latch Detail
FIGURE 3-4 Main Frame
Duo Service Manual0070-10-0604-013 - 5
IntroductionParts
30
27
28
33
30
31
31
32
29
1
1
34
41
45
35
35
25
28
26
29
24
27
23
26
FIGURE 3-5 Rear Case Assembly
33
36
3 - 60070-10-0604-01Duo Service Manual
PartsIntroduction
9
9
34
37
35
38
39
36
40
37
38
41
40
43
FIGURE 3-6 Front Case Assembly
39
42
Duo Service Manual0070-10-0604-013 - 7
IntroductionParts
FIGURE 3-7 Battery Connector Assembly Detail
3 - 80070-10-0604-01Duo Service Manual
PartsIntroduction
g
Screws,
mounting plate*
Washers, locking*
Utility basket
Mounting screw
for basket*
Casters, Non locking
Duo monitor
mounting plate*
Mounting screw
for basket*
Casters, lockin
FIGURE 3-8 Duo Rolling Stand
Replacement Parts, Duo Rolling Stand
DESCRIPTIONPART NUMBER
Duo rolling stand, valueDUOROLLSTD
Duo monitor mounting kit0406-00-0857-01
Casters, Non locking0401-00-0045
Casters, Locking0401-00-0046
Utility basket0202-00-0166
*Included in Duo monitor mounting kit
Washers, locking*
Mounting bolts, pole*
Duo Service Manual0070-10-0604-013 - 9
Parts ListingParts
3.2Parts Listing
REF. NUMBERPART NUMBERDESCRIPTION
10211-00-0146Housing Screw (metric panhead)
20380-00-0475Rear Housing
30441-00-0107Chassis
40671-00-0045Power Supply Board
50211-00-0145Metric Panhead Screw
60671-00-0044CPU/Display Board
70213-00-0032Self Tapping Screw
80380-00-0472Nellcor Connector Shroud
8A0380-00-0473Masimo Connector Shroud
90380-00-0476Front Housing
100012-00-1595SpO
110671-00-0246Masimo Isolated Power Board
120211-00-0143Screw
130671-00-0243Masimo SpO
140012-00-1474Masimo Flex Cable
150671-00-0247Nellcor Isolated Power Board
160671-00-0242Nellcor SpO
16a0671-00-0066Nell-3 SpO
170012-00-1457Nellcor Flex Cable
17a0012-00-1661Nellcor Flex Cable
180386-00-0308SpO
190104-00-0037NIBP Module
200671-00-0063Fan Driver Board
210671-00-0043Battery Connector Board
220380-00-0481Battery Latch
230380-00-0474Battery Door
240346-00-0052Battery Door Tether
250012-00-1592Fan with cable
260348-00-0216Fan Gasket
270213-00-4014Screw
280386-00-0310Fan Mounting Plate
290380-00-0471Filler Panel
300367-00-0084Handle
310211-00-0147Handle Screw (metric panhead)
320334-00-1603-03Rear Label, Lower, S/N
330348-00-0202Foot
340380-00-0480-01Power Switch Plunger
350380-00-0480-02Clear Switch Plunger
360380-00-0480-03Patient Size Switch Plunger
370380-00-0480-04Start NIBP Switch Plunger
N/S - Not Shown
Power Cable
2
Board
2
Board
2
Board
2
Mounting Plate
2
3 - 100070-10-0604-01Duo Service Manual
PartsParts Listing
REF. NUMBERPART NUMBERDESCRIPTION
380330-00-0052Keypad Overlay
390103-00-0411Pneumatic Fitting
400386-00-0309Mounting Plate
41See table belowRear Label, Upper
N/SSee table belowDisplay Overlay
N/SSee table belowConnector Label
N/S - Not Shown
Display Overlay
LANGUAGEOPTIONPART NUMBER
EnglishNIBP Only0330-00-0053-01
EnglishNIBP/SpO
2
0330-00-0053-11
Upper Rear Label
LANGUAGEPART NUMBER
English0334-00-1641-01
Connector Label
DESCRIPTIONPART NUMBER
No SpO
Masimo SpO
Nellcor SpO
2
2
2
0334-00-1602-01
0334-00-1602-02
0334-00-1602-04
Duo Service Manual0070-10-0604-013 - 11
Parts ListingParts
This page intentionally left blank.
3 - 120070-10-0604-01Duo Service Manual
4.0
Repair Information
4.1Introduction
This chapter of the Duo Service Manual provides the necessary technical information
needed to perform repairs on the instrument. The most important prerequisites for effective
troubleshooting are a thorough understanding of the instrument functions as well as an
understanding of the theory of operation.
To P1 on Main
Control Board
To P1 on Main
Control Board
External
Interface
Serial
Interface
BatteryP1
Fan
RS232<>TTL
Inverter
Power
Supply
Board
Fan
P2P1
Drive
Board
P2
P3
FIGURE 4-1 Module Interconnection
Isolated
Power
P1P2
Board
Isolated
Power
P1P2
Board
P5
Main
P3
Control
Board
P4
P1
P2
Masimo
SpO2
J3J1
Module
Nellcor
SpO2
J2J1
Module
P1
NIBP
Module
SpO2
Sensor
SpO2
Sensor
Cuff
Duo™ Service Manual0070-10-0604-014 - 1
Troubleshooting GuideRepair Information
4.2Troubleshooting Guide
Error Codes and Solutions
MESSAGE/
PROBLEM REASON SOLUTION
E01NIBP Self Test ErrorNIBP Module hardware failure
E02NIBP Communications
Error
E03Loose CuffCuff is not properly wrapped or no cuff is present
E04Air LeakCuff, hose or connector is damaged, internal leak
E05Air Pressure FailureStable pressure value is not available (e.g. hoses
E06Successful Pneumatic
Test
E07Pneumatic Test Failed/
Pneumatic Leak
E08Weak SignalCuff is too loose or patient pulse is too weak
E09Range ExceededNIBP value exceeds the upper measurement limit
E10Excessive Motion
Signal Saturated
E11Over PressurePressure has exceeded the specified upper safety
E12NIBP System FailureOperation of blood pressure pump system has
E13NIBP Time OutMeasuring time has exceeded 120 seconds
E14NIBP Illegally ResetUnexpected NIBP reset
E15NIBP Reset FailedNIBP reset failed
E16NIBP Communications
CRC Error
E17NIBP Patient Size
Change Error
E20Masimo SpO
Interference
E21Masimo SpO
Perfusion
E22Masimo SpO
Much Light
E23Masimo SpO
Unrecognized Sensor
E24Masimo SpO
Communication Error
E25Masimo SpO
Fault
E26Masimo SpO
Fault
E28Masimo SpO
E29Masimo SpO
Signal IQ
2
Low
2
Too
2
2
2
Board
2
Sensor
2
TimeoutSpO2 data has been determined continuously for
2
Low
2
Communications with NIBP Module have failed
are pinched or occluded)
Indicates NIBP pneumatic test was successful
Leak detected during the pneumatic test
Monitor is detecting too much motion and/or
noise to obtain a reading
limit
failed
NIBP Serial Communication CRC failure
Attempt to change Patient Size failed
Noise detected on the pulse signal prevents pulse
discrimination
Patient perfusion is low
There is too much ambient room light for the
sensor to function properly
The monitor does not recognize the sensor
The monitor and the SpO2 module are not
communicating
The Masimo SET board has failed to operate
properly
Defective sensor
more than 2 minutes, so SpO
from the display
data has timed out
2
The SpO2 signal quality is poor
4 - 20070-10-0604-01Duo™ Service Manual
Repair InformationTroubleshooting Guide
Error Codes and Solutions (Continued)
MESSAGE/
PROBLEM REASON SOLUTION
E34Masimo SpO2 Pulse
Rate Exceeded
E34Nellcor SpO
Exceeded
E40Nellcor SpO
Interference
E41Nellcor SpO
PR
2
2
Check
2
Sensor
E42Nellcor SpO
Communication Error
E43Nellcor SpO
Pulse
E44Nellcor SpO
Signal
E45Nellcor SpO
2
Weak
2
Weak
2
Board
2
Fault
E46Nellcor SpO
E47Nellcor SpO
MotionMotion is detected. The message is removed
2
TimeoutSpO2 data has been determined continuously for
2
E501Unit Battery Voltage
Low
E504Unit Keyboard Error 1Error with front panel keypad board
E505Monitor Shut Off
Failure
E506SpO
Module Not
2
Recognized
Pulse Rate value exceeds the measurement range
Pulse Rate value exceeds the measurement range
Noise is detected on the pulse signal, preventing
pulse discrimination from the noise. The
interference may be due to motion, excess
infrared light or electrical/optical interference.
The message is removed when the noise is
removed
The Nellcor module senses an unstable or illegal
sensor. This may be due to a poor connection or
a bad sensor. The user is required to reconnect
the same sensor or connect a new sensor. The
message will be removed once the Nellcor
module clears the error
The front end module is having problems
communicating (i.e.: framing errors or bad
checksums) with the Nellcor module
A pulse rate can not be determined and all other
measurement conditions are normal. The message
is removed when a pulse is detected
Noise is detected but a pulse rate can not be
discriminated. The message is removed when a
pulse is detected
The SpO2 board has malfunctioned
when No Pulse status is detected or when motion
ceases
more than 2 minutes, so SpO
from the display
data has timed out
2
Battery voltage is low
Monitor cannot be turned off normally
Monitor cannot communicate with SpO2 module
during self-test
Duo™ Service Manual0070-10-0604-014 - 3
Troubleshooting GuideRepair Information
Monitor Failures
MESSAGE/
PROBLEM REASON SOLUTION
No display after
power-on, power
indicator does not
light.
NIBP or SpO
not function.
2
will
Bad line fuseReplace fuse
Bad power supplyReplace power supply
Bad CPU/Display
board
CPU/Display board or
module failure.
Replace CPU/Display board
Isolate and replace defective board/module
4 - 40070-10-0604-01Duo™ Service Manual
Repair InformationTroubleshooting Guide
Module Failures
MESSAGE/
PROBLEM REASON SOLUTION
NIBP cuff cannot
be inflated.
Intermittently won't
take an NIBP
reading.
NIBP readings
inappropriately
high or low for
patient condition.
No SpO
SpO
inaccurate
readingSpO2 sensor or cable
2
value is
2
Pinched or leaking hose
or cuff
Loose cuff or patient
movement
Incorrect cuff size.
Incorrectly applied cuff
NIBP module is out of
calibration
damaged or
disconnected
Sensor not on patient
SpO2 sensor or cable
damaged or
disconnected
Sensor not on patient
Coloring agent (dye)
has been injected into
patient
Patient movementKeep patient quiet
Patient is coldWarm patient and retry
Patient is wearing nail
polish
Check hose and cuff. Replace as needed
Keep the patient quiet. Reapply cuff
Use appropriate size cuff. Ensure correct cuff
application
Calibrate/replace NIBP module
Check sensor placement and connection. Replace
if damaged
Check sensor placement and connection. Replace
if damaged
Retry after the coloring agent has dissipated
Remove nail polish
Duo™ Service Manual0070-10-0604-014 - 5
Disassembly InstructionsRepair Information
4.3Disassembly Instructions
Before disassembling the unit, perform the following:
• Turn off the unit and remove the line cord
• Remove all cables and hoses
• Remove the battery
• Perform all maintenance on a properly grounded work station.
4.3.1Tools Needed
• Phillips Screwdriver
• 5 mm nutdriver
4.3.2Front Housing Removal
1. Remove four (4) 3 x 30 mm Phillips panhead machine screws from the corners of the
Rear Housing.
2. Carefully separate the front and rear housings and disconnect the cables from the CPU/
Display PCB connectors P02, P03 and P07. Disconnect the NIBP tubing from the front
housing pressure fitting.
4.3.3SpO2 Interface Board Removal
1. Disconnect the cable from P01 on the CPU/Display pcb.
2. Remove three (3) 5mm hexnuts from their stand-offs on the SpO2 pcb.
3. Lift the SpO
Interface pcb from the stand-offs.
2
4.3.4SpO2 Board Removal
1. Remove three (3) 5mm hex stand-offs from the SpO2 Board.
2. Remove one (1) 3 x 6 mm Phillips panhead machine screw from the SpO2 Board.
3. Lift the SpO
Board from the mounting bracket.
2
4.3.5CPU/Display Board Removal
1. Remove four (4) 3 x 6 mm Phillips panhead machine screws from the corners of the
CPU/Display Board.
2. Lift the CPU/Display Board from the Front Housing.
4.3.6NIBP Module Removal
1. Remove four (4) 3 x 6 mm Phillips panhead machine screws from the corners of the NIBP
module frame.
2. Lift the NIBP module from the battery housing frame.
4 - 60070-10-0604-01Duo™ Service Manual
Repair InformationDisassembly Instructions
4.3.7Power Supply Removal
1. Remove the battery cable from P1 on the power supply PCB.
2. Remove four (4) 3 x 6 mm Phillips panhead machine screws from the corners of the
power supply PCB.
3. Remove one (1) 3 x 6 mm Phillips panhead machine screw from the line power entry
connector.
4. Slide the Power Supply pcb towards the bottom (open end) of the battery housing frame
until it can go no farther.
5. Lift the power supply PCB from the battery housing frame. Use care to avoid breaking
the positioning tab from the line power entry connector.
Mindray Medical Netherlands B.V.• P.O. Box 26 • 3870 CA Hoevelaken • The Netherlands •
Tel: +31 33 25 44 911 • Fax: +31 33 25 37 621
Mindray (UK) Limited • 3 Percy Road • St. John’s Park • Huntingdon • Cambridgeshire PE29 6SZ •
United Kingdom • Tel: 01480 416840 • Fax: 01480 436588
Mindray Medical France SARL • Europarc Créteil •123, Chemin des Bassins •
94035 Créteil Cedex • France • Tel: (0)1.45.13.91.50 • Fax: (0)1.45.13.91.51
Mindray Medical German GmbH • Zwischen den Bächen 4 • 64625 Bensheim • Germany •
Tel: +49.6251.17524-0 • Fax: +49.6251.17524-20
Mindray Medical International Ltd. • 2813 Office Tower, Convention Plaza • No 1 Harbour Road •
Wanchai • Hong Kong • Tel: +852 2793 5596 • Fax: +852 2344 8824
0002-08-8918
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.