Mindray BS-200 User manual

4.6 (14)

BS-200

Chemistry Analyzer

Operation Manual

© 2006 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.

For this Operation Manual, the issued Date is 2006-03 (Version: 1.2).

Intellectual Property Statement

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties.

Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rent, adaption and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

, , , , are the registered trademarks or trademarks owned by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners.

Responsibility on the Manufacturer Party

Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this manual.

Mindray is responsible for safety, reliability and performance of this product only in the condition that:

all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel;

the electrical installation of the relevant room complies with the applicable national and local requirements;

the product is used in accordance with the instructions for use.

WARNING:

It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.

i

NOTE:

This equipment is to be operated only by medical professionals trained and authorized by Mindray or Mindray-authorized distributors.

Warranty

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.

Exemptions

Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel.

This warranty shall not extend to:

any Mindray product which has been subjected to misuse, negligence or accident;

any Mindray product from which Mindray's original serial number tag or product identification markings have been altered or removed;

any product of any other manufacturer.

Return Policy

Return Procedure

In the event that it becomes necessary to return this product or part of this product to Mindray, the following procedure should be followed:

1Obtain return authorization: Contact the Mindray Service Department and obtain a Customer Service Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping container. Returned shipments will not be accepted if the Mindray number is not clearly visible. Please provide the model number, serial number, and a brief description of the reason for return.

2Freight policy: The customer is responsible for freight charges when this product is shipped to Mindray for service (this includes customs charges).

3Return address: Please send the part(s) or equipment to the address offered by Customer Service department

Company Contact

Manufacture: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, P.R.China, 518057

Phone: +86 755 26582479 26582888

Fax: +86 755 26582500 26582501

ii

Preface

Before using the BS-200 Chemistry Analyzer, please read this operation manual thoroughly for relevant operation instructions.

Who Should Read This Manual

This manual is written for clinical laboratory professionals to

perform daily operating tasks;

perform system maintenance and troubleshooting; learn about the BS-200 hardware and software.

WARNING:

The BS-200 Chemistry Analyzer is to be operated only by medical professionals trained and authorized by Mindray or Mindray-authorized distributors.

What Can You Find in This Manual

This operation manual covers principles, operations, daily maintenance and troubleshooting of the system. Please operate and service the system strictly as instructed by this manual.

Conventions Used in This Manual

This manual uses certain typographical conventions to clarify meanings in the text.

Bold font indicates a chapter title, such as 4 Maintenance

Bold and Italic font indicates text displayed on the screen, such as Sample Request.

Safety Symbols

This chart explains the symbols used in this manual.

When you see …

Then …

WARNING:

Read the statement following the symbol. The

 

statement is alerting you to an operating hazard

 

that can cause personal injury.

BIOHAZARD:

Read the statement following the symbol. The

statement is alerting you to a potentially

 

 

biohazardous condition.

 

 

 

1

Preface

When you see …

Then …

CAUTION:

Read the statement following the symbol. The

 

statement is alerting you to a possibility of

 

system damage or unreliable results.

NOTE:

Read the statement following the symbol. The

statement is alerting you to information that

 

 

requires your attention.

 

 

Labels Used on the System

The labels attached to the panels of the system use symbols with the text to clarify the meaning of the text. The chart below explains the symbols on the labels.

Serial Number

Date of Manufacture

Manufacturer

The device is fully in conformance with the Council Directive Concerning In Vitro Diagnostic Medical Devices 98/79/EC.

The following definition of the WEEE label applies to EU member states only: The use of this symbol indicates that this product should not be treated as household waste. By ensuring that this product is disposed of correctly, you will help prevent bringing potential negative consequences to the environment and human health. For more detailed information with regard to returning and recycling this product, please consult the distributor from whom you purchased the product.

In Vitro Diagnostic equipment

Biohazard Warning: risk of potentially biohazardous infection

Warning: risk of personal injury or equipment damage

Warning: risk of electric shock

Warning: risk of burn

Alternating current (AC)

ON (MAIN POWER)

OFF (MAIN POWER)

2

Preface

ON (Power)

OFF (Power)

Graphics

All graphics, including screens and printout, are for illustration purpose only and must not be used for any other purposes.

EC Representative

Name:

Shanghai International Holding Corp. GmbH (Europe)

Address:

Eiffestrasse 80 D-20537 Hamburg Germany

Phone:

+49

40 2513174

Fax:

+49

40 255726

3

Preface

Safety Precautions

Observe the following safety precautions when using the BS-200 Chemistry Analyzer. Ignoring any of these safety precautions may lead to personal injury or equipment damage.

WARNING:

If the system is used in a manner not specified by Mindray, the protection provided by the system may be impaired.

Preventing Electric Shock

Please observe the following instructions to prevent electric shock.

WARNING:

When the MAIN POWER is on, users must not open the rear cover or side cover.

Liquid ingression may lead to electric shock or equipment damage. In case of liquid ingression, shut off the power supply and contact Mindray Service Department or your local distributor.

Preventing Personal Injury Caused by Moving Parts

Please observe the following instructions to prevent personal injury caused by moving parts.

WARNING:

Do not touch such moving parts as probe and mixing bar, when the system is in operation.

Do not put your finger or hand into any open part when the system is in operation.

Preventing Personal Injury Caused by Photometer Lamp

Please observe the following instructions to prevent personal injury caused by photometer lamp.

WARNING:

Light sent by the photometer lamp may hurt your eyes. Do not stare into the lamp when the system is in operation.

If you want to replace the photometer lamp, first switch off the MAIN POWER and then wait at least 30 minutes for the lamp to cool down before touching it. Do not touch the lamp before it cools down, or you may get burned.

4

Preface

Preventing Infection

Please observe the following instructions to protect against the biohazardous infection.

BIOHAZARD:

Inappropriately handling samples may lead to biohazardous infection. Do not touch the sample, mixture or waste with your hands. Wear gloves and lab coat and, if necessary, goggles.

In case your skin contacts the sample, follow standard laboratory safety procedure and consult a doctor.

Certain reagents are strongly acid or alkaline. Exercise caution when using the reagents. In case your skin or clothes contact the reagents, wash them off with soap and clean water. In case the reagents spill into your eyes, rinse them with much water and consult an oculist.

Treating Waste

Please observe the following instructions to prevent environmental pollution and personal injury caused by waste.

BIOHAZARD:

Dispose of the waste in accordance with your local or national guidelines for biohazard waste disposal and consult the manufacturer or distributor of the reagents for details.

Preventing Fire or Explosion

Please observe the following instructions to prevent fire and explosion.

WARNING:

Do not use flammable substance around the system.

5

Preface

Precautions on Use

To use the BS-200 Chemistry Analyzer safely and efficiently, please pay attention to the following operation notes.

Intended Use

WARNING:

The BS-200 Chemistry Analyzer is a chemistry system that is designed for the in vitro quantitative determination of clinical chemistries in serum, plasma, urine or CSF samples. Please consult Mindray first if you want to use the system for other purposes.

To draw a clinical conclusion, please also refer to the patient’s clinical symptoms and other test results.

Operator

WARNING:

The BS-200 Chemistry Analyzer is to be operated only by medical professionals trained and authorized by Mindray or Mindray-authorized distributors.

Environment

CAUTION:

Please install and operate the system in an environment specified by this manual. Installing and operating the system in other environment may lead to unreliable results and even equipment damage.

To relocate the system, please contact Mindray Service Department or your local distributor.

Preventing Interference by Electromagnetic Noise

CAUTION:

Electromagnetic noise may interfere with operations of the system. Do not install devices generating excessive electromagnetic noise around the system. Do not use such devices as mobile phones or radio transmitters in the room housing the system. Do not use other CRT displays around the system.

Do not use other medical instruments around the system that may generate electromagnetic noise to interfere with their operations.

6

Preface

Operating the System

CAUTION:

Operate the system strictly as instructed by this manual. Inappropriate use of the system may lead to unreliable test results or even equipment damage or personal injury.

Before using the system for the first time, run the calibration program and QC program to make sure the analyzer is in proper state.

Be sure to run the QC program every time you use the system, otherwise the result may be unreliable.

Do not open the cover of the sample/reagent disk when the system is in operation.

The RS-232 port on the analyzing unit is to be used for connection with the operation unit only. Do not use it for other connections. Only use the supplied cable for the connection.

The operation unit is a personal computer with the control software installed. Installing other software or hardware on this computer may interfere with the system operation. Do not run other software when the system is working.

Do not use this computer for other purposes. Inappropriate use of it will probably introduce computer virus, which may spread into the system through floppy disks, software or network.

Do not touch the display, mouse or keyboard with wet hands or hands with chemicals.

Don’t place the MAIN POWER to ON again within 10 seconds since placing it to OFF; otherwise the system may enter the protection status. If it does so, place the MAIN POWER to OFF and place it to ON again.

Maintaining the System

CAUTION:

Maintain the system strictly as instructed by this manual. Inappropriate maintenance may lead to unreliable results or equipment damage or personal injury.

To wipe off dust from the system surface, use a soft, clean and wet (not too wet) cloth, soaked with soap water if necessary, to clean the surface. Do not use such organic solvents as ethanol for cleaning. After cleaning, wipe the surface dry with dry cloth.

Switch off all the powers and disconnect the power plug before cleaning. Take necessary measures to prevent water ingression into the system, otherwise it may lead to equipment damage or personal injury.

Replacement of such major parts as photometer, probe, mixing bar and syringe plunger assembly must be followed by a calibration.

7

Preface

Samples

CAUTION:

Use serum samples that are completely separated from blood clots or urine samples that are free from suspended matter. If fibrin exists in the serum samples or suspended matter exists in the urine samples, the probe may be blocked.

Medicines, anticoagulants or preservative in the samples may lead to unreliable results.

Hemolysis, jaundice or chylomicron in the samples may lead to unreliable test results, so sample blanks are recommended.

Store the samples properly. Improper storage may change the compositions of the samples and lead to unreliable results.

Sample volatilization may lead to unreliable results. Do not leave the sample open for a long period.

Not all the tests the reagents claim capable of analyzing can be analyzed on the system. Consult the reagent suppliers for details.

Certain samples need to be processed before being analyzed by the system. Consult the reagent suppliers for details.

The system has a specific requirement on the sample volume. Refer to this manual for proper sample volume.

Load the sample to proper tube position on the sample disk before the analysis begins; otherwise you will not obtain correct results.

Reagents, Calibrators and Controls

CAUTION:

Use proper reagents, calibrators and controls in the system.

Select appropriate reagents according to performance characteristics of the system. Consult the reagent suppliers, Mindray or Mindray-authorized distributor for details, if you are not sure about your reagent choice.

Store and use the reagents, calibrators and controls strictly as instructed by the suppliers. Otherwise, you may not obtain reliable results or best performance of the system.

Perform calibration after changing the reagents. Otherwise, you may not obtain reliable results.

Contamination caused by carryover among reagents may lead to unreliable test results. Consult the reagent suppliers for details.

8

Preface

Setting up the System

CAUTION:

To define such parameters as sample volume, reagent volume and wavelength, follow the instructions in this manual and the instructions of reagents.

Backing up Data

NOTE:

The system automatically stores the data to the built-in hard disk. However, data loss is still possible due to mis-deletion or physical damage of the hard disk. Mindray recommends you to regularly back up the data to such medium as CDs.

Computer and Printer

NOTE:

Refer to their operation manuals for details.

External Equipment

WARNING:

External equipment connected to the system, such as PC and printer, shall be consistent with IEC 60950 or EN 60950.

9

Contents

Preface..........................................................................................................................................

1

Who Should Read This Manual .............................................................................................

1

What Can You Find in This Manual........................................................................................

1

Conventions Used in This Manual .........................................................................................

1

Safety Precautions .................................................................................................................

4

Precautions on Use................................................................................................................

6

1 System Description ..........................................................................................................

1-1

1.1

Hardware Introduction .........................................................................................

1-1

 

1.1.1

Analyzing Unit ......................................................................................

1-1

 

1.1.2

Operation Unit ......................................................................................

1-6

 

1.1.3

Output Unit ...........................................................................................

1-6

1.2

Software Introduction...........................................................................................

1-6

 

1.2.1

Software Interface ................................................................................

1-6

 

1.2.2

Main Interface Components .................................................................

1-8

2 Installation .........................................................................................................................

 

2-1

2.1

Unpacking............................................................................................................

2-1

2.2

Installation Requirements ....................................................................................

2-1

 

2.2.1

Installation Environment Requirements................................................

2-1

 

2.2.2

Power Requirements............................................................................

2-2

 

2.2.3

Temperature and Humidity Requirements............................................

2-2

 

2.2.4

Water Supply and Drain Requirements................................................

2-3

 

2.2.5

Space and Accessibility Requirements.................................................

2-3

2.3

Connecting Deionized Water Tank ......................................................................

2-4

2.4

Connecting Waste Tank.......................................................................................

2-5

2.5

Installing/Removing Sample/Reagent Disk .........................................................

2-5

2.6

Installing/Removing Sample Tubes .....................................................................

2-6

2.7

Installing/Removing Reagent Bottles...................................................................

2-7

2.8

Installing/Removing Cuvettes ..............................................................................

2-7

3 Basic Operations

..............................................................................................................

3-1

3.1

Daily Procedure ...................................................................................................

3-1

3.2

Preparing for Analysis..........................................................................................

3-2

 

3.2.1

Checking before Startup.......................................................................

3-2

 

3.2.2

Power-on ..............................................................................................

3-3

 

3.2.3

Starting the Control Software ...............................................................

3-3

 

3.2.4

Setting Up the Analyzer........................................................................

3-4

 

3.2.5

Preparing Reagents..............................................................................

3-5

3.3

Starting Analysis ..................................................................................................

3-5

 

3.3.1

Reagent Blank......................................................................................

3-5

 

3.3.2

Calibration ............................................................................................

3-5

 

3.3.3

QC ........................................................................................................

3-6

 

3.3.4

Samples................................................................................................

3-6

3.4

Processing Results..............................................................................................

3-7

 

3.4.1

Editing Results of Samples ..................................................................

3-7

 

3.4.2

Printing Results of Samples .................................................................

3-7

3.5

Finishing Analysis ................................................................................................

3-7

 

3.5.1

Exiting the Control Software.................................................................

3-7

 

3.5.2

Shutdown..............................................................................................

3-7

 

3.5.3

Operations after Shutdown...................................................................

3-8

I

Contents

4

Advanced Operations.......................................................................................................

4-1

 

4.1

Sample Request ..................................................................................................

4-1

 

 

4.1.1

Sample Information ..............................................................................

4-3

 

 

4.1.2

Delete a Sample...................................................................................

4-5

 

 

4.1.3

Change Position ...................................................................................

4-6

 

 

4.1.4

Requesting Samples or Modifying Information ....................................

4-7

 

4.2

QC Request.........................................................................................................

4-7

 

4.3

Start ...................................................................................................................

 

4-10

 

4.4

Probe Stop.........................................................................................................

4-11

 

4.5

Stop....................................................................................................................

 

4-12

 

4.6

Current Results..................................................................................................

4-13

 

4.7

Replace..............................................................................................................

 

4-15

 

4.8

Re-log ................................................................................................................

 

4-17

 

4.9

Exit.....................................................................................................................

 

4-17

 

4.10

History................................................................................................................

 

4-18

 

 

4.10.1

Conditions...........................................................................................

4-20

 

 

4.10.2

Add Off-system Test Results ..............................................................

4-22

 

 

4.10.3

Compensate Results ..........................................................................

4-23

 

 

4.10.4

Edit Results ........................................................................................

4-25

 

 

4.10.5

Reaction Curve...................................................................................

4-27

 

 

4.10.6

Delete Results ....................................................................................

4-28

 

 

4.10.7

Print Results .......................................................................................

4-29

 

 

4.10.8

Result Trend Curve ............................................................................

4-30

 

4.11

Reagent .............................................................................................................

 

4-32

 

4.12

Calibration..........................................................................................................

4-33

 

 

4.12.1

Calibration Request............................................................................

4-33

 

 

4.12.2

Results................................................................................................

4-37

 

 

4.12.3

Calibrator............................................................................................

4-45

 

4.13

QC .....................................................................................................................

 

4-47

 

 

4.13.1

Real-time QC......................................................................................

4-48

 

 

4.13.2

Daily QC .............................................................................................

4-49

 

 

4.13.3

Day to Day QC ...................................................................................

4-52

 

 

4.13.4

Control ................................................................................................

4-55

 

4.14

Status.................................................................................................................

 

4-57

 

 

4.14.1

Sample Disk .......................................................................................

4-58

 

 

4.14.2

Reagent Disk......................................................................................

4-61

 

 

4.14.3

Reaction Disk .....................................................................................

4-65

 

4.15

Statistics.............................................................................................................

 

4-65

 

 

4.15.1

Worklist...............................................................................................

4-66

 

 

4.15.2

Results................................................................................................

4-67

 

 

4.15.3

Workload ............................................................................................

4-70

 

 

4.15.4

Charges..............................................................................................

4-72

 

4.16

Setup .................................................................................................................

 

4-74

 

 

4.16.1

Test .....................................................................................................

4-75

 

 

4.16.2

Profile .................................................................................................

4-85

 

 

4.16.3

Calculation..........................................................................................

4-87

 

 

4.16.4

Off-system Test...................................................................................

4-89

 

 

4.16.5

Carryover............................................................................................

4-90

 

 

4.16.6

System................................................................................................

4-91

 

 

4.16.7

Hospital...............................................................................................

4-96

 

 

4.16.8

User ....................................................................................................

4-98

 

 

4.16.9

Print ..................................................................................................

4-101

 

4.17

Maintenance ....................................................................................................

4-105

 

 

4.17.1

Daily Maintenance............................................................................

4-105

 

 

4.17.2

Log....................................................................................................

4-107

 

 

4.17.3

Import/Export....................................................................................

4-109

 

 

4.17.4

Commission......................................................................................

4-111

5

Maintenance ......................................................................................................................

 

5-1

II

 

 

 

 

Contents

 

5.1

Preparation ..........................................................................................................

5-1

 

5.2

Daily Maintenance ...............................................................................................

5-2

 

 

5.2.1

Checking Remaining Deionized Water.................................................

5-2

 

 

5.2.2

Emptying Waste Tank...........................................................................

5-3

 

 

5.2.3

Checking Connection of Deionized Water............................................

5-3

 

 

5.2.4

Checking Connection of Wastewater ...................................................

5-4

 

 

5.2.5

Checking Syringe .................................................................................

5-5

 

 

5.2.6

Checking Probe....................................................................................

5-6

 

 

5.2.7

Checking Mixing Bar ............................................................................

5-7

 

5.3

Weekly Maintenance ...........................................................................................

5-7

 

 

5.3.1

Cleaning Probe.....................................................................................

5-7

 

 

5.3.2

Cleaning Mixing Bar .............................................................................

5-8

 

 

5.3.3

Washing Deionized Water Tank ...........................................................

5-9

 

 

5.3.4

Washing Waste Tank..........................................................................

5-10

 

 

5.3.5

Cleaning Sample/Reagent Compartment ..........................................

5-11

 

 

5.3.6

Cleaning Panel of Analyzing Unit .......................................................

5-11

 

5.4

Monthly Maintenance ........................................................................................

5-12

 

 

5.4.1

Cleaning Wash Well of Probe.............................................................

5-12

 

 

5.4.2

Cleaning Wash Well of Mixing Bar .....................................................

5-12

 

5.5

Maintenance Every Six Months.........................................................................

5-12

 

5.6

Irregular Maintenance........................................................................................

5-13

 

 

5.6.1

Unclogging Probe...............................................................................

5-13

 

 

5.6.2

Replacing Probe.................................................................................

5-16

 

 

5.6.3

Replacing Mixing Bar .........................................................................

5-17

 

 

5.6.4

Replacing Plunger Assembly of Syringe ............................................

5-18

 

 

5.6.5

Replacing Lamp .................................................................................

5-20

 

5.7

Maintenance Log ...............................................................................................

5-22

6

Troubleshooting................................................................................................................

6-1

7

Calculation Methods.........................................................................................................

7-1

 

7.1

Analytical Methods ..............................................................................................

7-1

 

 

7.1.1

Endpoint ...............................................................................................

7-1

 

 

7.1.2

Fixed-Time............................................................................................

7-2

 

 

7.1.3

Kinetic...................................................................................................

7-3

 

7.2

Calculation Process.............................................................................................

7-5

 

 

7.2.1

Absorbance ..........................................................................................

7-5

 

 

7.2.2

Response .............................................................................................

7-6

 

 

7.2.3

Calibration Parameters.........................................................................

7-9

 

 

7.2.4

Concentration .....................................................................................

7-12

 

 

7.2.5

QC rules .............................................................................................

7-14

Appendix A

 

Specifications............................................................................................

A-1

 

A.1

Technical Specifications.......................................................................................

A-1

 

A.2

Power Requirements ...........................................................................................

A-1

 

A.3

Environmental Requirements ..............................................................................

A-2

 

 

A.3.1

Storage Requirements..........................................................................

A-2

 

 

A.3.2

Operation Requirements ......................................................................

A-2

 

A.4

Dimension and Weight ........................................................................................

A-2

 

A.5

Other Specifications.............................................................................................

A-2

 

A.6

Input/Output Devices ...........................................................................................

A-2

 

A.7

Interface...............................................................................................................

A-3

Appendix B

 

Supplies .....................................................................................................

B-1

Appendix C

 

Index...........................................................................................................

C-1

III

1 System Description

This chapter includes the following two sections:

Hardware Introduction

Software Introduction

The BS-200 Chemistry Analyzer is a chemistry system that is designed for the in vitro quantitative determination of clinical chemistries in serum, plasma, urine or CSF samples.

NOTE:

Not all the tests the reagents claim capable of analyzing can be analyzed on the system. Consult the reagent suppliers for details.

1.1Hardware Introduction

The system consists of the following three units – the analyzing unit, operation unit and output unit.

1.1.1 Analyzing Unit

The analyzing unit consists of the following major parts:

Sample/Reagent Disk

Dispenser

Mixer

Reaction Disk

Photometric System

1-1

Mindray BS-200 User manual

System Description

Figure 1-1 Analyzing unit

 

Cover

 

Syringe

 

Dispenser

 

Mixer

MAIN POWER

Reaction Disk

Sample/Reagent Disk

Power

 

Upper Cabinet

 

Lower Cabinet (optional)

1.1.1.1 Sample/Reagent Disk

The sample/reagent disk holds sample tubes and reagent bottles.

Figure 1-2 Sample/Reagent disk

Sample/reagent Disk

The disk is composed of two circles – sample disk on the outer circle and reagent disk on the inner circle.

1-2

System Description

The sample disk provides 40 sample tube positions and the reagent disk provides 40 reagent bottle positions. On the reagent disk, No. 39 is for detergent and No. 40 is for distilled water.

NOTE

Mindray recommends you to use the following detergents:

Acid: 0.1mol/l hydrochloric acid; Alkaline: 0.5% (V/V) javel water

The sample disk can hold the following sample tubes:

Micro sample cup and centrifugal tube

Collection tube: Ф12×68.5, Ф12×99, Ф12.7×75 and Ф12.7×100

Plastic tube

The reagent disk can only hold Mindray bottles, which are available in two types, 40ml and 20ml.

The sample/reagent disk is located in the sample/reagent compartment, which has a refrigerator to keep the temperature at 4-15 ˚C.

WARNING:

Make sure the disk cover is closed; otherwise it may degrade the refrigeration and damage the probe.

Before running the analyzing unit, make sure that the disk cover is closed and the round red mark on the cover is aligned with its counterpart on the panel. Otherwise the probe may be damaged.

NOTE

The refrigerator will be put into service once the MAIN POWER is turned on.

Do not use sample tubes and reagent bottles other than the specified ones.

1.1.1.2 Dispenser

The dispenser is composed of a probe, arm and rotor.

1-3

System Description

Figure 1-3 Dispenser

Arm

 

Rotor

Probe

Wash Well

 

The probe aspirates certain amount of sample from the designated sample tube, or reagent from the designated reagent bottle, and then dispenses them into the designated cuvette on the reaction disk.

After dispensing the sample or the reagent, the probe moves to the wash well for cleaning.

Sample volume: 3µl-45µl; precision: 0.5µl.

Reagent volume: 30µl -450µl; precision: 1µl.

The dispenser is capable of preheating the reagent, detecting the sample/reagent level, capable of tracking level and protecting against collision in the vertical direction.

WARNING:

When the analyzing unit is in operation, do not place any part of your body or any obstacle in the route the arm moves. Otherwise, it may lead to personnel injury or equipment damage.

1.1.1.3 Mixer

The mixer is composed of a mixing bar, arm and rotor.

1-4

System Description

Figure 1-4 Mixer

Arm

Mixing Bar

Rotor

Wash Well

The mixing bar thoroughly stirs the reaction mixture (reagent and sample) in the cuvette. After stirring, it moves to the wash well for cleaning.

For the single-reagent test, the mixer starts to work after the sample is dispensed into the cuvette.

For the double-reagent test, the mixer starts to work after the sample or the second reagent is dispensed into the cuvette.

1.1.1.4 Reaction Disk

The reaction disk holds the cuvettes, in which the sample reacts with the reagent(s) and colorimetric readings are taken.

Figure 1-5 Reaction disk

Dispensing Position

Mixing Position

Reaction Disk

The reaction disk can hold 8 cuvette segments (80 cuvettes).

During the analyzing process, the reaction disk rotates to dispensing position or mixing position as needed. The colorimetric readings are taken when the specified cuvette passes through the optical axis.

The cuvettes adopted are

1-5

System Description

Disposable;

5mm×6mm×30mm (5mm optical path);

900µl (capable of holding 180-500µl of the reaction mixture).

The reaction disk is placed in the temperature-controlled chamber, which keeps a constant temperature at 37±0.3˚С.

BIOHAZARD:

Be sure to dispose of the used cuvettes in compliance with the local regulations.

CAUTION:

The cuvettes should not be re-used. Otherwise, the system performance may be degraded.

1.1.1.5 Photometric System

The photometric system, which locates in the analyzing unit, measures the absorbance of the reaction mixture in the cuvette.

The photometric system provides 8 wavelengths: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm, 700nm (optional).

1.1.2 Operation Unit

The operation unit is a computer with the control software of BS-200 Chemistry Analyzer installed. It manages running of the analyzing unit, as well as operation and data processing.

1.1.3 Output Unit

The output unit is a printer that prints out the test results and other data.

1.2Software Introduction

NOTE:

In this manual, “click” refers to moving the pointer of the mouse to the desired item and click the left button of the mouse.

1.2.1 Software Interface

The interface of the control software is shown in Figure 1-6.

1-6

System Description

Figure 1-6 Interface of the control software

System Status Area

Small Buttons Area

37.0

 

 

Group Buttons

 

Area

Shortcut Buttons

Area Working Page

Area

Comment Area

Operator Area

Warning Messages Area

Clear Button

System status area

This area displays the system status, current temperature of the reaction disk and current time.

Small buttons area

Click the small button , and the BS-200 Operation Manual will be displayed. Click the small button , and the version of software will be displayed.

Group buttons area

Displays group buttons which include History, Reagent, Calibration, QC, Status,

Statistics, Setup and Maintenance.

Click a group button, and a relevant working page will be displayed.

Shortcut buttons area

Displays shortcut buttons which include Sample Request, QC Request, Start,

Probe Stop, Stop, Current Results, Replace, Re-log and Exit.

Click a shortcut button, and corresponding operation will be performed.

Working page area

Displays values and graphs for parameters, procedures and results.

When the pointer of the mouse points to an element in current working page, the comment area on the lower part of the working page will display the explanation of the element.

Operator area

Displays the name of the current operator.

Warning messages area

Displays the warning and error messages.

Click the button to view the previous message, and click the button to view the next one.

1-7

System Description

Clear button

Click the button to clear contents displayed in the warning messages area.

1.2.2 Main Interface Components

Dialog box

The dialog box is one of the most common components. See the following example:

Dialog Box

Tab

See the figure below for an example. Click a tab and you can access the working page that it indexes.

Tabs

Drop-down list box

Click and a list will display, as the figure below shows. Click the desired item to select it.

Drop-down List Box

Button

Click a button and you can access the function it indexes, as the figure below shows.

1-8

System Description

Button

Radio button

Click a radio button to select the option it represents.

Note that for a given group of radio buttons, you can only select one of them. See the figure below.

Radio Buttons

Check box

Click a check box to select the option it represents and click it again to deselect it.

Note that for a given group of check boxes, you can choose more than one of them at one time. See the figure below.

Check Boxes

Edit box

You can enter characters in the edit box from keyboard. See the figure below.

There’re two types of edit boxes, one can only accept characters input from the keyboard, while the other can accept characters not only input from the keyboard but also selected by clicking or .

Edit Box

Edit Box

Scroll bar

When the contents to be displayed are too many for one screen, the scroll bar will appear to help you see the hidden contents.

Move the pointer to the scroll bar, press left button of the mouse and hold it, then you can drag the scroll bar left/right or up/down to see the hidden contents.

1-9

System Description

Scroll Bar

Scroll Bar

List

The list can list the names of tests, profiles or others, as the figure below shows. Click a test to select it, and click it again to deselect.

List tree

List tree can list the affiliation among options, as the figure below shows. Click the “-” to hide the subordinate options, and the “-” becomes “+”.

Click the “+” to expand the subordinate options and display their affiliation, then “+” turns to be “-”.

If an option has no subordinate options, there is no “+” or “-” to the left of it.

1-10

System Description

List Tree

Drag bar

Drag bar is used to select a level continuously, as the figure below shows. Click the drag bar and hold it, then you can drag it to the position needed.

Drag Bar

1-11

Loading...
+ 223 hidden pages