SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the
intellectual property rights to this product and this manual. This manual may refer to information
protected by copyrights or patents and does not convey any license under the patent rights of
Mindray, nor the rights of others. Mindray does not assume any liability arising out of any
infringements of patents or other rights of third parties.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaption and
translation of this manual in any manner whatsoever without the written permission of Mindray
is strictly forbidden.
Mindray in China and other countries. All other trademarks that appear in this manual are used
only for editorial purposes without the intention of improperly using them. They are the property
of their respective owners.
Contents of this manual are subject to changes without prior notice.
For this manual, the issued Date is January 2015 (Version 2.0).
, , and are the registered trademarks or trademarks owned by
WARNING
Federal Law (USA) restricts this device to sale by or on the order of a physician.
NOTE
This manual describes all features and options. The equipment may not have all of
them. Contact Mindray service department for any questions.
I
Manufacturer’s Responsibility
Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the condition
that:
All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel; and
The electrical installation of the relevant room complies with the applicable national
and local requirements; and
This product is operated under strict observance of this manual.
Warranty
Mindray warrants that components within its products will be free from defects in
workmanship and materials for a period of three years from the date of purchase except that
disposable or one-time use products are warranted to be free from defects in workmanship
and materials up to a date one year from the date of purchase or the date of first use,
whichever is sooner.
This warranty does not cover consumable items such as, but not limited to, batteries,
external cables, and sensors.
Mindray shall not be liable for any incidental, special, or consequential loss, damage, or
expense directly or indirectly arising from the use of its products. Liability under this
warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or
replacing the affected products, at Mindray’s option, at the factory or at an authorized
distributor, for any product which shall under normal use and service appear to Mindray to
have been defective in material or workmanship. Recommended preventative maintenance,
as prescribed in the service manual, is the responsibility of the user and is not covered by
this warranty.
No agent, employee, or representative of Mindray has any authority to bind Mindray to any
affirmation, representation, or warranty concerning its products, and any affirmation,
representation or warranty made by any agent, employee, or representative shall not be
enforceable by buyer or user.
THIS WARRANTY IS EXPRESSLY IN LIEU OF, AND MINDRAY EXPRESSLY DISCLAIMS, ANY
OTHER EXPRESS OR IMPLIED WARRANTIES, INCLUDING ANY IMPLIED WARRANTY OF
MERCHANTABILITY, NON-INFRINGEMENT, OR FITNESS FOR A PARTICULAR PURPOSE, AND OF
ANY OTHER OBLIGATION ON THE PART OF MINDRAY.
Damage to any product or parts through misuse, neglect, accident, or by affixing any
II
non-standard accessory attachments, or by any customer modification voids this warranty.
Mindray makes no warranty whatsoever in regard to trade accessories, such being subject to
the warranty of their respective manufacturers.
A condition of this warranty is that the equipment or accessories which are claimed to be
defective be returned when authorized, freight prepaid to Mindray DS USA, Inc., Mahwah,
New Jersey 07430 or its authorized representative. Mindray shall not have any responsibility
in the event of loss or damage in transit.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting
from the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty does not extend to:
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure events, such as (i) flood, fire and
earthquake or other similar elements of nature or acts of God; (ii) riots, war, civil
disorders, rebellions, or revolutions in any country; or (iii) any other cause beyond the
reasonable control of Mindray.
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible.
Others not caused by instrument or part itself.
III
Return Policy
In the event that it becomes necessary to return a unit to Mindray, follow the instructions below.
1. Obtain a return authorization.
Contact the Mindray Service Department and obtain a Mindray Customer Service
Authorization Number. The Mindray Customer Service Authorization Number must
appear on the outside of the shipping container. Return shipments will not be accepted
if the Mindray Customer Service Authorization Number is not clearly visible. Please
provide the model number, serial number, and a brief description of the reason for
return.
2. Freight policy
The customer is responsible for freight charges when this product is shipped to
Mindray for service (including any relevant customs fees or other freight related
charges).
3. Return address
Please send the part(s) or equipment to the address offered by Customer Service
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan,
Shenzhen 518057 P.R. China
+86 755 81888998
+86 755 26582680
www.mindray.com
Mindray DS USA, Inc.
800 MacArthur Boulevard Mahwah, New Jersey 07430 USA
1.800.288.2121, 1.201.995.8000
www.mindray.com
IV
Preface
Manual Purpose
This manual provides detailed information about the assembling, dissembling, testing and
troubleshooting of the equipment to support effective troubleshooting and repair. It is not
intended to be a comprehensive, in-depth explanation of the product architecture or technical
implementation.
Observance of the manual is a prerequisite for proper equipment maintenance and prevents
equipment damage and personnel injury.
Intended Audience
This manual is for biomedical engineers, authorized technicians or service representatives
responsible for troubleshooting, repairing and maintaining the equipment.
Passwords
A password may be required to access different modes of the following equipment. The
passwords are listed below:
User maintenance: 888888
Configuration mode: 315666
It is recommended that the user should change the passwords for user maintenance and
configuration mode once they take ownership of the equipment.
1.1 Manual Information .................................................................................................................................... 1-1
1.2 Safety Information ....................................................................................................................................... 1-1
2.2 Connectors for Peripheral Devices ........................................................................................................ 2-2
2.3 Main Unit ........................................................................................................................................................ 2-2
2.4 Front Housing Assembly ........................................................................................................................... 2-4
2.5 Main Board ..................................................................................................................................................... 2-4
4.2.3 SpO2 test ............................................................................................................................................ 4-2
4.2.4 NIBP Test ............................................................................................................................................ 4-4
4.2.5 Temperature Test ............................................................................................................................ 4-6
4.2.6 Nurse Call Reply Performance Test .......................................................................................... 4-7
4.3 Electrical Safety and Other Tests ............................................................................................................ 4-8
4.3.1 Electrical Safety and Other Test Frequencies ....................................................................... 4-8
4.3.2 Electric Safety Tests ........................................................................................................................ 4- 8
4.3.3 Power-on Test .................................................................................................................................. 4-8
4.4.3 Software Version ........................................................................................................................... 4-11
4.4.4 Monitor Information ................................................................................................................... 4-12
5.2 Part Replacement ........................................................................................................................................ 5-1
5.3 Checking Equipment Status .................................................................................................................... 5-1
5.4 Checking Software Version ...................................................................................................................... 5-1
7.2 Main Unit ........................................................................................................................................................ 7-1
7.2.2 Parts List ............................................................................................................................................. 7-1
7.3 Front Housing Subassembly (Touchscreen) ...................................................................................... 7-2
7.3.2 Parts List ............................................................................................................................................. 7-3
7.4.2 Parts List ............................................................................................................................................. 7-4
7.5 Main Bracket Assembly ............................................................................................................................. 7-6
7.5.2 Parts List ............................................................................................................................................. 7-6
7.6 Power Management Board Assembly ................................................................................................. 7-7
7.6.2 Parts List ............................................................................................................................................. 7-7
7.7.2 Parts List ............................................................................................................................................. 7-8
7.8 Predictive Temperature Assembly ......................................................................................................... 7-9
7.8.2 Parts List ............................................................................................................................................. 7-9
8 Hardware and Software Upgrade ......................................................................................... 8-1
2.0 Modify temperature module test method, update parts list
1.2 Safety Information
WAR NIN G
Indicates a potential hazard or unsafe practice that, if not avoided, will result in
death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
minor personal injury or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the
most from your product.
1-1
1.2.1 Warnings
WAR NIN G
All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel.
There is high voltage inside the equipment. Never disassemble the equipment
before it is disconnected from the AC power source or the battery.
When you disassemble/reassemble a parameter module, a patient leakage current
test must be performed before it is used again for monitoring.
The equipment must be connected to a properly installed power outlet with
protective earth contacts only. If the installation does not provide for a protective
earth conductor, disconnect it from the power line and operate it on battery power,
if possible.
Disposal of the packaging material should observe the applicable waste control
regulations and keeping it out of children’s reach.
1.2.2 Cautions
CAUTION
Make sure that no electromagnetic radiation interferes with the performance of the
equipment when preparing to carry out performance tests. Mobile phone, X-ray
equipment or MRI devices are a possible source of interference as they may emit
higher levels of electromagnetic radiation.
Before connecting the receiver to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the unit’s
label or in this manual.
Protect the equipment from damage caused by drop, impact, strong vibration or
other mechanical force during servicing.
1.2.3 Notes
NOTE
Refer to operator’s manual for detailed operation and other information.
1-2
1.3 Equipment Symbols
Some symbols may not appear on your equipment.
Caution
Alternating current (AC)
Alarm Reset
Alarm Paused
Admit a new patient
DEFIBRILLATION –PROOF
TYPE CF APPLIED PART
Network connector
USB connectors
Serial number
DATE OF MANUAFACTURE
Power ON/OFF
Battery indicator
NIBP Start/Stop key
Record
Insertion Direction
Input/Output
Equipotentiality
Manufacturer
Protection against fluid ingress
CATALOGUE NUMBER
Temperature limit
Atmospheric pressure
limitation
AUTHORISED
REPRESENTATIVE IN THE
EUROPEAN COMMUNITY
The product bears CE mark indicating its conformity with the provisions of the
Council Directive 93/42/EEC concerning medical devices and fulfills the
essential requirements of Annex I of this directive.
1-3
Humidity limitation
Refer to instruction manual/
booklet
Interference may occur in the
vicinity of equipment marked
with this symbol
The following definition of the WEEE label applies to EU member states only.
This symbol indicates that this product should not be treated as household
waste. By ensuring that this product is disposed of correctly, you will help
prevent bringing potential negative consequences to the environment and
human health. For more detailed information with regard to returning and
recycling this product, please consult the distributor from whom you purchased
it.
* For system products, this label may be attached to the main unit only.
The presence of this label indicates the machine was certified by ETL with the
statement:
Conforms to AAMI Std ES 60601-1, IEC Std 60601-1-6, IEC Std 60601-1-8, IEC Std
60601-2-49, IEC Std 80601-2-30, ISO Std 80601-2-56, ISO Std 80601-2-61.
The Monitor is intended for monitoring physiologic parameters, including SpO2, Pulse Rate, NIBP
and Temperature, on adult, pediatric, and neonatal patients in healthcare facilities by clinical
physicians or appropriate medical staff under the direction of physicians.
The equipment also:
Provides audible and visual alarm indications in case of patient or equipment problems;
Provides display, review, storage and printing of monitored information;
Incorporates multiple input devices such as buttons, knob, and touchscreen; and,
Enables program upgrade over the network.
Monito
Nurse Call
Reserved port
Other device
Mindray Proprietary LAN
i
F
-
i
W
EMR Server
AP
CMS/Gatewa
The above figure shows a system consisting of the vital signs monitor and its peripheral devices.
The vital signs monitor:
Can be used for monitoring the physiological parameters, giving alarms and reviewing
patient data, etc.
Supports recorder.
Supports nurse call function.
Supports Wi-Fi module, wired network, remote view, and communication with the
Hypervisor Central Monitoring System (hereinafter called CMS).
Supports external AC power source and an internal battery.
Supports clinical data acquisition: by SD card or by USB drive.
2-1
2.2 Connectors for Peripheral Devices
1
2
3
4
5
6
1. Connector for Temperature probe
2. Network connector: It is a standard RJ45 connector used to communicate with external
devices, such as central monitoring system, e-Gateway, or used to export data or
upgrade the system.
3. Multi-function connector: connects to the hospital's nurse call system, or connects
external devices through DIAP protocol.
4. USB connector: connects to barcode scanner or USB disk.
5. AC Power Input
6. Equipotential grounding terminal: When the equipment and other devices are to be
used together, their equipotential grounding terminals should be connected together
to eliminate the potential difference between them.
2.3 Main Unit
The main unit of the vital signs monitor consists of three parts:
Front housing assembly: main board, Wi-Fi module, keypad board assembly (knob),
display, touchscreen, and alarm lamp board.
Rear housing assembly: power module (AC/DC), power management and interface
board (including SpO
board.
SpO
2
External module:Temperature module.
isolation power), recorder, speaker, battery, NIBP module, and
2
2-2
The following figure shows the main unit architecture of the vital signs monitor.
Front Housing Assembly
Alarm lamp
board
Knob
Power
On/Off
keypad
(Indicator)
SD card
Main board
Wi-Fi module
I2C
LVD
S
Backlight
Touchscree
n control
board
LCD display
Touch
screen
Rear Housing Assembly
Speaker
Recorder
2600mAh
battery
15V
AC-DC power
module
AC-IN
Power
management
and interface
board
SpO
2
isolation power
External interface
RJ45/USB/Multifunctional
(nurse call, extended ports)
Pump
NIBP module
SpO2 board
(Standard
Mindray,OEM
optional)
NIBP socket
SpO2socket
External Module
Predictive
Temp
module
Temp
probe
2-3
2.4 Front Housing Assembly
Alarm lamp board
Touchscreen
control board
Wi-Fi
antenna
Main board
Screws securing the front and rear housing
Wi-Fi
module
Keypad
2.5 Main Board
The main board is the control center of the equipment. It provides communication, display, and
data storage functions, including:
Communication with SpO
connector;
Communication with power management board, keypad board, and recorder through
connectors;
Providing drive for display and backlight;
Providing 2 USB connectors, a network connector, and a multifunctional connector;
Communication with touchscreen control board through I2C;
Providing drive for SD card; and,
Providing drive for Wi-Fi module.
board, NIBP module, and Temperature module through
2
2-4
Keypad
The keypad scans and detects the input of keys and encoder, integrates the power on/off key,
and connects AC and battery indicators.
Alarm Lamp Board
The alarm lamp board is located at the top of front housing. It has two-color indicators, red and
yellow. The alarm lamp board directly connects the main board through a cable. It is controlled
directly by the main board.
Touchscreen and Touchscreen Control Board
The touchscreen control board drives the touchscreen and implements communication with the
vital signs monitor.
Wi-Fi Module (Optional)
The Wi-Fi Module enables the equipment to connect to an 802.11 b/g/n Mindray proprietary
network.
2.6 Rear Housing Assembly
Speaker
Power
NIBP
module
Recorder
Rear housing assembly consists of power module (AC/DC), power management and interface
board (including SpO
isolation power), recorder, speaker, battery, NIBP module, and SpO2 board.
2
2-5
management
and interface
board
AC/DC
power
module
SpO2board
AC/DC Power Module
The AC/DC power board transforms the input AC into DC power, which is the power source for all
voltages in the equipment.
Power Management and Interface Board
The power management and interface board provides the following functions:
Charge and discharge of battery and charge detection;
DC/DC conversion: outputs 12V and 5V DC power;
Control over power on/off key and AC, BAT indicator;
Communication transmission among parameter modules;
Providing isolation power for SpO
Providing external connectors, filter and protection for these connectors.
module; and,
2
Recorder
The recorder receives data from the main board and sends them to the thermal printhead for
printing.
NIBP Module
The NIBP module consists of blood pressure measurement board and pump and valve assembly,
providing measurement acquisition of blood pressure data. The main functions of the NIBP
module are:
NIBP measurement; and,
Data exchange with the main board.
SpO2 Board
The SpO
to the main board. The power management interface board provides isolation power for it.
An external Temp module can be attached to the monitor.
The independently developed Mindray Temperature module consists of an isolation power board,
Temperature measurement board, and probes. The Temperature measurement board collects
Temperature signals, processes algorithm and sends measurement results to the main board.
2-6
3 Equipment Installation
3.1 Unpacking the Equipment
Open the package and take out the packing list. Check that all the articles included in the
packing list are available and the quantity and specification are correct. Make sure that:
All the optional parts purchased by the customer have been received.
Notify Mindray North America if your order is incorrect or is incomplete.
In case of damage during transportation, keep the packing material and notify Mindray
North America immediately.
Keep the packing material until all equipment is checked and accepted.
3.2 Preparation for Installation
3.2.1 Preparation for Installation Site
1. Ensure that the site meets all safety, environmental and power requirements.
2. Ensure that a network connector is available if the equipment needs to be connected to
network.
3.2.2 Environmental Requirements
To avoid explosion hazard, do not use the equipment in the presence of flammable anesthetics,
vapors or liquids. The environment where the equipment will be used should be reasonably free
from vibration, dust and corrosive substances. If these conditions are not met, the system may
not function normally.
The environmental specification is as follows:
Main Unit
Item
Operating
environment
Storage
environment
Temperature (℃)
0 to 40 (without Temperature
module)
5 to 40 (with Temperature
module)
-30 to 70 10% to 95% 120.0 to 805.5
NOTE
The environmental specifications of unspecified accessories are the same as those of
the main unit.
Relative humidity
(noncondensing)
15% to 95% 427.5 to 805.5
Altitude (mmHg)
3-1
3.2.3 Electrical Requirements
Check cables and power cords. Make sure that:
1. System cables, power cords, and power plugs are not damaged, and that the pins are
not loose. If damage is noted, discontinue use and replace.
2. Patient cables and connections are secure.
WARNING
Use only properly grounded power outlets.
Use the supplied power cord only!
Vol tage
Current
Frequency
100 to 240V AC
0.9 to 0.5A
50/60 Hz
3.3 Equipment Installation
Follow the procedure below to install the equipment:
1. Ensure the unit and accessories are not damaged.
2. Install the battery (optional). For detailed operations, please refer to the operator’s
manual of the vital signs monitor.
3. Connect AC power.
4. Connect the accessories.
Installation Support
The vital signs monitor can be mounted on a wall bracket or on a trolley support. The wall
bracket or trolley support can be ordered optionally. Each type of mounting bracket is delivered
with a complete set of mounting hardware and instructions. For detailed installation information,
please refer to Wall-mount Bracket Instructions for Use (PN: 0010-20-42933) and Rollstand Instructions for Use (PN: 0010-20-42934).
CAUTION
Use only Mindray supplied or approved mounting solutions.
The mounting bracket should be installed by qualified service personnel.
3-2
3.3.1 Preparation for Power on
1. Before using the equipment, check for any mechanical damage and make sure that all
external cables, plug-ins and accessories are properly connected.
2. Plug the power cord into the AC power source. If you run the equipment on battery
power, ensure that the battery is sufficiently charged.
3. Press the Power ON/OFF button (
) on the front panel to turn on the equipment.
3-3
FOR YOUR NOTES
3-4
4 Testing and Maintenance
4.1 Introduction
The expected service life of the equipment is five years. To ensure the equipment always
functions normally, qualified service personnel should perform regular inspection, maintenance
and test. This chapter provides a checklist of the testing procedures for the equipment with
recommended test equipment and frequency. The service personnel should perform the testing
and maintenance procedures as required and use appropriate test equipment.
The testing procedures provided in this chapter are intended to verify that the equipment meets
the performance specifications. If the equipment or a module fails to perform as specified in any
test, repairs or replacements must be done to correct the problem. If you have any question,
contact our Technical Support Department.
CAUTION
All tests should be performed by qualified service personnel only.
Care should be taken to avoid changing the settings in [User Settings >>] and
[Factory Maintenance >>] menus to avoid loss of data.
Service personnel should acquaint themselves with the test tools and make sure
that test tools and cables are available.
4.2 Performance Tests
Performance test are designed to ensure that measurement results are accurate. The following
sections provide a list of performance and accuracy tests and their recommended frequencies.
4.2.1 Performance Test Frequencies
Check/Maintenance Item Frequency
Visual inspection 1. When first installed or reinstalled.
SpO2test
Leakage test
NIBP test
TEMP test
Nurse call function test
Accuracy test
NIBP cuff
overpressure test
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3. Every two years.
Note: Per year is recommended for NIBP leakage
and accuracy tests.
If the user suspects that the nurse call function does
not work well.
4-1
4.2.2 Visual Inspection
Perform an overall inspection on the appearance of the equipment. The test is passed if the
equipment has no obvious signs of damage. Follow these guidelines when inspecting the
equipment:
Carefully inspect the case, display screen, buttons, and knob for obvious signs of
damage.
Inspect all external connections for loose connectors, bent pins or frayed cables.
Inspect all connectors on the equipment for loose connectors or bent pins.
Make sure that safety labels and data plates on the equipment are clearly legible.
4.2.3 SpO
Test Meth od 1
Too l requir ed:
None
Test procedure:
1. Connect SpO
2. Place the SpO
3. Verify the Pleth Wave and Pulse Rate are displayed on the screen.
4. Remove the SpO
Measurement validation
The SpO
measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, and
only about two-thirds of the measurements can be expected to fall within the specified accuracy
compared to CO-oximeter measurements.
NOTE
The SpO
test
2
sensor for adult to the SpO2 connector of the monitor. Set [Patient
2
Category] to [Adult].
sensor on your finger.
2
sensor from your finger and make sure that the SpO2 Sensor Off alarm
2
is triggered.
accuracy has been validated in human studies against arterial blood sample reference
2
simulator can only be used to verify that the pulse oximeter operates
2
properly. It cannot be used to verify the accuracy of the pulse oximeter or the SpO
sensor. To verify the accuracy, clinical tests are required.
2
4-2
Test Meth od 2
Too l requir ed:
SpO
simulator, Index-2 recommended
2
Test procedure:
1. Connect the SpO
2. Selected the model and manufacturer of the SpO
and set the simulator as follows: SpO
sensor to the SpO2 simulator.
2
to 96% and Pulse Rate to 80 bmp.
2
module to be tested on the simulator,
2
3. Set the patient type to [Adult], [Pediatric], and [Neonatal] respectively. Observe the
monitor and make sure the displayed SpO
and Pulse Rate value fall in the following
2
range.
ManufacturerSpO
Nellcor
Masimo
Pulse Rate
2
96% ± 2% (Adult, pediatric)
96% ± 3% (Neonate)
96% ± 2% (Adult, pediatric)
96% ± 3% (Neonate)
80 ± 3 bpm
80 ± 3 bpm
4-3
4.2.4 NIBP Test
Leakage Test
NOTE
Perform NIBP leakage test before any other NIBP test and calibration.
Tools required:
Adult NIBP Cuff
NIBP Hose
Cylinder
Follow this procedure to perform the leakage test:
1. In the [Patient Demographics] menu, set [Patient Category] to [Adult].
2. Connect the NIBP cuff to the NIBP connector on the monitor.
3. Apply the cuff to the cylinder as shown below.
Monitor
Connector for NIBP cuff
NIBP Hose
Cylinder
Cuff
4. Select [Main]→[Maintenance >>]→[NIBP Leakage Test]. The message [Leakage
Tes t in g… ] is displayed in the NIBP parameter area.
After about 20 seconds, the monitor will automatically deflate. This means the test is complete. If
no message is displayed in the NIBP parameter area, it indicates that the system has no leakage. If
the message [NIBP Pneumatic Leak] is displayed, it indicates that the system may have a leak.
Check the tubing and connections for leakages. Ensure that the tubing and connections are all
correct, perform a leakage test again. If the problem persists, contact our Technical Support
Department.
You may also perform a manual leakage test:
1. Perform steps 1 to 4 in the NIBP Accuracy Test section.
2. Raise the pressure in the rigid vessel to 250 mmHg with the manometer bulb. Then,
wait for 5 seconds to allow the pressure to stabilize.
3. Record the current pressure value, wait 60 seconds and record the pressure again.
4. Compare the two pressure values and verify the difference is 6mmHg or less.
4-4
NIBP Accuracy Test
Too l requir ed:
T-shape connector
Appropriate tubing
Manometer bulb
Rigid Vessel with volume 500 ± 25 ml
Reference manometer (calibrated with accuracy equal to or better than 0.75 mmHg)
Follow this procedure to perform the accuracy test:
1. Connect the equipment as shown below.
Monitor
Connector for NIBP cuff
Manometer bulb
Appropriate tubing
Standard
sphygmomanometer
Rigid Vessel
2. Before inflation, check that the reading of the manometer is 0. If not, turn off the
manometer bulb to let the whole airway open to the atmosphere. Turn on the
manometer bulb after the reading is 0.
4. Check the manometer values and the monitor values. Both should be 0 mmHg.
5. Raise the pressure in the rigid vessel to 50 mmHg with the manometer bulb. Then, wait
for 10 seconds to let the measured values become stable.
6. Compare the manometer values with the monitor values. The difference between the
manometer and displayed values should be ± 3 mmHg. If it is greater than ± 3 mmHg,
contact Mindray Technical Support.
7. Raise the pressure in the rigid vessel to 200 mmHg with the manometer bulb. Then,
wait for 10 seconds to let the measured values become stable. Repeat step 6.
NOTE
You can use an NIBP simulator to replace the manometer bulb and the reference
manometer to perform the test.
4-5
NIBP Cuff Overpressure Test
Tools required:
T-shape connector
Appropriate tubing
Manometer bulb
Metal Vessel with volume 500 ± 25 ml
Reference manometer (calibrated, with accuracy equal to or better than 0.75 mmHg)
Follow this procedure to perform the calibration:
1. Perform steps 1 to 4 in the NIBP Accuracy Test section.
3. In [Overpressure Protection Circuit] menu, set [Patient Category] to [Adu/Ped].
Raise the pressure to 330 mmHg. After the pressure value is stabilized, select [Calibrate]
to start calibration.
4. In the [Overpressure Protection Circuit] menu, set [Patient Category] to [Neonatal].
Raise the pressure to 165 mmHg. When the pressure value is stabilized, select
[Calibrate] to start calibration.
5. All calibration results are displayed in the [Calibrate NIBP] menu. If the calibration fails,
check the test system for leakage and perform another calibration.
4.2.5 Temperature Test
Required Tool:
Thermostatic oil tank (HART 7102 recommended) or Water Bath and Precision
Thermometer for reference
Test procedure:
1. Set the temperature of the oil tank or water bath to 37°C and conduct the test after the
temperature stabilizes.
2. Set temperature unit to °C.
3. Select temperature parameter area to access [Te mp Se tup ] menu, and then set
temperature type to [Monitor].
4. In Monitor Mode, remove the Temperature probe from the probe sheath, insert a probe
cover, and place the probe into the oil tank or water bath.
5. Wait till the Temp value displayed on the monitor stabilizes. Verify that the displayed
value is ±0.1°C of the oil bath thermostat setting or reference thermometer.
Contact our Technical Support Department if the Temperature test fails.
4-6
NOTE
Due to the different application environment and the test object in vivo and vitro
conditions, there are deviations in the measurement result. The maximum deviation
of 2.5℃ may exist in predictive Temperature measurement by liquid bath
techniques.
4.2.6 Nurse Call Reply Performance Test
Too l requir ed:
Oscilloscope
Test procedure:
1. Connect the nurse call cable to the Multi-function connector on the rear panel.
3. In the [Nurse Call Setup] menu, select the [Alarm Level] and [Alarm Category], and set
[Conta ct Type] to [Normally Open, and [Signal Type] to [Continuous].
4. Click
5. Select [Main]→[Maintenance >>]→[Demo >>]→enter the required password→[Ok]
to enter demo mode.
6. Make the monitor to generate an alarm and check that the oscilloscope outputs
continuous high level when there is an alarm.
7. Quit the demo mode, and repeat step 2.
8. In the [Nurse Call Setup] menu, select the [Alarm Level] and [Alarm Category], and
set [Conta ct Type] to [Normally Open, and [Signal Type] to [Pulse].
9. Repeat steps 4 and 5.
10. Create an alarm condition and verify that the oscilloscope displays positive pulses of 1
second width when there is an alarm.
to save the changes and quit the menu.
4-7
4.3 Electrical Safety and Other Tests
4.3.1 Electrical Safety and Other Test Frequencies
Check/Maintenance Item Frequency
1. Following any repair or replacement of any
module
Electric Safety Tests
Power- on test
Touchscreen calibration
Recorder check Following any repair or replacement of the recorder.
Bar code scanner test
Functional test
Battery check
Performance test
4.3.2 Electric Safety Tests
See Appendix A Electrical Safety Inspection for electrical safety tests.
4.3.3 Power-on Test
This test is to verify that the equipment can power up correctly. This test is passed if the
equipment starts up by following this procedure:
1. Insert the battery in the battery compartment, and connect the equipment to the AC
mains. The AC mains indicator and battery indicator light up.
2. If monitor is physically damaged.
3. Every two years.
For details, refer to A Electrical Safety Inspection.
1. When first installed or reinstalled.
2. Following any repairs or replacement of any main
unit parts.
1. When the touchscreen appears abnormal.
2. After the touchscreen is replaced.
Following any repair or replacement of the bar code
scanner.
1. When first installed.
2. Whenever a battery is replaced.
Once per year or if the battery run time reduced
significantly.
2. Press the Power ON/OFF button (
The work status indicator lights up inside the Power ON/OFF button.
3. The equipment gives a beep, which indicates that the selftest for alarm audio is passed,
and the alarm lamp turns yellow, then red, and then off, which indicates that the
selftest for alarm lamp has passed.
4. The welcome screen clears and the normal operation screen is displayed.
4. After the calibration is completed, the message [Screen Calibration Completed!].
Select [Ok] to confirm the completion of calibration.
symbol will appear at different positions of the screen.
symbol.
4.3.5 Recorder check
Too l requir ed:
None
Test procedure:
1. Print SpO
be clear.
2. Remove the paper from the recorder then try to print and verify that an error message
(recorder out of paper) appears. Reinstall the paper and verify the recorder functions
properly.
3. Set the recorder to print trend data. Verify the recorder prints trends correctly.
Contact our Technical Support Department if the recorder test fails.
Pleth waveform. The recorder should print correctly and the printout should
2
4.3.6 Bar Code Scanner Test
Too l requir ed:
None
Test procedure:
1. Aim the bar code scanner at the target bar code. Adjust the field of view to capture the
bar code.
2. Hold the trigger until the bar code scanner gives a beep, indicating the bar code is
successfully decoded. Meanwhile, the indicator turns green and scanned characters are
displayed on the monitor.
Please contact our service personnel if the bar code scanner fails to work normally.
4-9
4.3.7 Battery Check
Too l requir ed:
None
Func tional Test
1. Remove the battery (if equipped), then connect the equipment to AC power.
2. Verify that the equipment works correctly when powered form an AC source.
3. Disconnect AC power and reinstall the battery.
4. Verify the equipment functions properly on battery power.
Performance Test
Run the equipment on battery power and verify it performs in accordance to the specifications
stated in the Battery chapter in the operator’s manual (P/N: 046-004882-00).
4.4 Factory Maintenance
4.4.1 Accessing Factory Maintenance Menu
Select [Main]→[Maintenance >>]→[Factory Maintenance >>]→enter the required password→
[Ok] to access the Factory Maintenance menu.
4-10
4.4.2 Drawing Waves
There are two methods for drawing waves: [Color] and [Mono].
Color: selecting Color will have smoother waveforms.
Mono: selecting Mono will have a wider viewing angle.
4.4.3 Software Version
Selecting the [Software Version >>] will show software version information, as shown below:
1
2
1. System software version
2. Power software version
3. Module software version
4-11
3
4.4.4 Monitor Information
Selecting [Monitor Information >>] will show the status of the monitor as shown below:
NOTE
If the main board is replaced, the serial number of the monitor must be entered on
the new main board in the Electronic SN Setup menu.
4-12
5 Troubleshooting
5.1 Overview
In this chapter, equipment problems are listed along with possible causes and
recommended corrective actions. Refer to the tables to check the monitor, identify and
correct problems.
The tables provided are meant as a general guide. For more detailed troubleshooting
assistance, contact our Technical Support Department.
5.2 Part Replacement
Printed circuit boards (PCBs), major parts and components in the monitor are replaceable.
Once you isolate a PCB you suspect defective, follow the instructions in chapter 6 Disassembly and Repair to replace the PCB with a known good one and check that the
trouble disappears or the equipment passes all performance tests.
To obtain information on replacement parts or order them, refer to chapter 7 Parts.
5.3 Checking Equipment Status
Some troubleshooting tasks may require you to identify the hardware version and status of
your monitor. To check equipment status:
1. Select [Main]→[Maintenance >>]→[Monitor Information >>]. Then you can
view the information on system start time, selfcheck, etc.
2. Select [Main]→[Maintenance >>]→[Factory Maintenance >>]→enter the
required password→[Ok]→[Monitor Information >>]. You can also view the
information on the monitor’s current status.
5.4 Checking Software Version
Some troubleshooting may involve software compatibility. Thus it requires you to know
your monitor configuration and software version. For detailed information on version
compatibility, please contact our Customer Service Department. To view information on
the system configuration and system software version,
Select [Main]→[Maintenance >>]→[Factory Maintenance >>]→Enter the required
password→[Ok]→[Software Version >>]. You can also view the information on system
software version and module software version.
5-1
5.5 Checking Technical Alarms
Before troubleshooting the monitor, check for technical alarm messages. Correct all
Technical Alarms before continuing.
For detailed information on technical alarm message, possible cause and corrective action,
refer to the monitor’s operation manual.
5.6 Troubleshooting Guide
5.6.1 Power On/Off Failure
Symptoms Possible Cause Troubleshooting
AC mains not
connected or battery
too low
The monitor
fails to start.
Cable defective
Power board defectiveReplace the power board.
Power management
board defective
The main board failed.Replace the main board.
5.6.2 Display Failures
Check that AC mains is properly connected or
battery capacity is sufficient.
1. Check that the cable between the keypad
board and main board is correctly connected.
2. Check that the cable between the power
board and power management board is
correctly connected.
3. Check that the cable between the main
board and power management board is
correctly connected.
Replace the power management board.
Symptoms Possible Cause Troubleshooting
1. Check if the cable between the display and
main board and the backlight cable are
The display is
blank or
black.
Images
overlapped or
Cable defective
Main board defective Replace the main board.
Display defective Replace the display.
Main board error
correctly connected.
2. Check that the cables and connectors are
not damaged.
Replace the main board, or upgrade the main
board with the upgrade software.
5-2
distorted
Touchscreen
does not
respond.
Tou ch
position
invalid
Cable defective
Touchscreen disabled
Cable defective
Touchscreen control
board defective
Touchscreen defective Replace the touchscreen.
The main board failed. Replace the main board.
Touchscreen not
calibrated
Check if the cable between the display and
main board and the backlight cable are
correctly connected.
Check if there is a
above the [Main] quickkey. If yes, press and
hold the [Main Menu] quickkey for more than
3 seconds to enable the touchscreen.
1. Check that the cable between the
touchscreen and touchscreen control board is
correctly connected.
2. Check that the cable between the
touchscreen control board and main board is
correctly connected.
Replace the touchscreen control board.
Calibrate the touchscreen.
5.6.3 Alarm Lamp Failures
Symptoms Possible Cause Troubleshooting
1. Check that the cable between the alarm
lamp board and main board is correctly
The Alarm
Lamp does not
illuminate
properly or not
at all.
Cable defective
Alarm lamp board
defective
The main board
failed.
connected.
2. Check that the cables and connectors are
not damaged.
Replace the alarm lamp board
Replace the main board.
5.6.4 Button and Knob Failures
symbol displayed
Symptoms Possible Cause Troubleshooting
Buttons do not
work
Cable defective
Keypad board failure Replace the keypad board.
Check that the cable between the keypad
board and main board is correctly connected.
5-3
1. Check that the cable between the knob and
Cable defective
Knob does not
work
Knob failure Replace the knob encoder.
Keypad board failure Replace the keypad board.
keypad board is correctly connected.
2. Check that the cable between the keypad
board and main board is correctly connected.
5.6.5 Sound Failures
Symptoms Possible Cause Troubleshooting
No hardkey or
knob sound,
or hardkey or
knob sound
abnormal
No alarm
sound or
alarm sound
abnormal
The key volume is set
to 0.
Cable defective
Speaker defective Replace the speaker.
The main board failed. Replace the main board.
Power management
board defective
The alarm volume is
set to 0.
Cable defective
Speaker defective Replace the speaker.
The main board failed. Replace the main board.
Power management
board defective
Select [Main]→[General Setup >>] to adjust
the key volume.
Check that the cable between the speaker and
interface board is properly connected.
Replace the power management board.
Select [Main]→[Maintenance >>]→]User Settings >>]→enter the required password→
[Ok]→[Alarm Setup >>] and set the
[Minimum Alarm Volume] to a proper level in
the prompt menu. Select [Main]→[General Setup >>] to adjust the key volume.
1. Check that the cable between the speaker
and interface board is properly connected.
Replace the power management board.
5-4
5.6.6 Battery Failures
Symptoms Possible Cause Troubleshooting
Battery defective Replace the battery.
Check that the cable between the battery
interface board and power management board
is correctly connected.
Replace the power management board.
Replace the battery interface board.
Battery
cannot be
charged
Cable defective
Power management
board defective
Battery interface
board defective
NOTE
If the Battery Module fails, it may affect other modules or boards in the main
unit.
Components of the main unit are powered by the power module. In the event
that a component malfunctions, check if the operating voltage is correct.
5.6.7 Recorder Failures
Symptoms Possible Cause Troubleshooting
1. Check if the recorder status indicator lights.
2. If yes, select [Main]→[Maintenance >>]→
[Factory Maintenance >>]→enter the required
password→[Ok]→[Device Config. >>] and
check the recorder box in the prompt menu to
enable the recorder. Otherwise, check for other
possible causes.
Check that the cable between the recorder and
main board is correctly connected.
Stop the recorder and re-install the paper roll.
1. Check the thermal print head and the paper
roller for foreign matter.
2. Clean the thermal print head with an
appropriate cleaning solution such as contact
cleaner or isopropal alcohol.
5-5
No printout
Poor print
quality or
paper not
feeding
properly
Recorder module
disabled
Paper reversed Re-install the paper roll.
Cable defective
Recorder defective Replace the recorder.
Paper roll not
properly installed
Print head dirty
Recorder defective Replace the recorder.
5.6.8 Output Interface Failure
Symptoms Possible Cause Troubleshooting
1. Check that the cable between the power
management board and main board is
correctly connected.
2. Check that the cable between power
management board and interface board is
correctly connected.
Replace the connector board.
Replace the power management board.
Select [Main]→ [Transfer Data to USB]
Unable to use
the USB devices;
USB drive data
transfer failure
Cable defective
The connector board
failed.
Power management
board defective
The main board failed. Replace the main board.
Improper setup
5.6.9 Data Storage Failure
Symptoms Possible Cause Troubleshooting
Fails to review
archived patient
data
SD card failure
Abnormal patient
admitting
SD card full; unavailable
for more patient data
The main board failed. Replace the main board.
SD card not formatted Format the SD card.
SD card failure Replace the SD card.
SD card is locked Unlock the SD card.
Main board defective Replace the main board.
Admit the patient properly.
Delete uneeded patient data, clear the
related alarm, and readmit the patient.
5-6
5.6.10 Wired Network Related Problems
Symptoms Possible Cause Troubleshooting
Check LAN cable connection. LAN
cable shall not be longer than 50
meters.
Check for IP address conflict. If yes,
reconfigure the IP address.
1. Check that the cable between the
power management board and main
board is correctly connected.
2. Check that the cable between power
management board and interface
board is correctly connected.
Replace the connector board.
Replace the power management
board.
Check LAN cable connection. LAN
cable shall not be longer than 50
meters.
Check LAN cable connection. LAN
cable shall not be longer than 50
meters.
A monitor can only be viewed by four
other monitors at the same time under
the View Others mode. The excessive
view requests system will be ignored.
Check for IP address conflict. If yes,
reconfigure the IP address.
Unable to connect the
wired network
The monitor is
frequently off line or
disconnects from the
network.
The monitor is
connected to a LAN
but cannot view other
patients under the
View Others mode
Incorrect LAN cable
connection
Incorrect IP address
configuration
Cable defective
The connector board
failed.
Power management
board defective
The main board failed. Replace the main board.
Incorrect LAN cable
connection
Incorrect LAN cable
connection
Excessive requests for
viewing the monitor at
the same time
Incorrect IP address
configuration
5.6.11 Wi-Fi Related Problems
Symptoms Possible Cause Troubleshooting
The monitor is
frequently off line or
disconnects from the
Wi-Fi network.
The Wi-Fi signal is unstable in
the operating area.
The monitor's Wi-Fi antenna is
detached or not properly
connects the Wi-Fi module.
Antenna damaged Replace the Wi-Fi antenna.
5-7
Check the signal quality of the
Mindray Wi-Fi network.
Disassemble the monitor and
fix the Wi-Fi antenna.
Symptoms Possible Cause Troubleshooting
Wi-Fi module defective Replace the Wi-Fi module.
Check for IP address conflict. If
yes, reconfigure the IP
address.
Check the signal quality of the
Mindray Wi-Fi network.
Fix the Wi-Fi antenna.
Unable to connect the
Wi-Fi network.
Incorrect IP address
configuration
The Wi-Fi signal is unstable in
the operating area.
The monitor's Wi-Fi antenna is
detached or not connects the
Wi-Fi module.
Antenna damaged Replace the Wi-Fi antenna.
Wi-Fi module defective Replace the Wi-Fi module.
Main board defective Replace the main board.
5.6.12 Module defective
Symptoms Possible Cause Troubleshooting
1. Check that the cable between the
external converter board inside the
Module defective
Failed to connect the
external parameter
modules
Main unit defective
Cable defective
Module can be
loaded, but "XX
communication
stopped" is reported
or some parameters
cannot be used
"XX" indicates the configured modules, such as NIBP, SpO2 and Temperature.
inside the module
Parameter module
defective
Converter board
defective inside the
module
module and the converter board is
correctly connected.
2. Replace the converter board.
1. Check that the cable between the
main board and power management
board is correctly connected.
2. Replace the power management
board.
3. Replace the main board.
Check the cables connecting the
converter board and corresponding
parameter module.
Replace the corresponding module.
Replace corresponding converter board.
5.6.13 Technical Alarm Messages
Please refer to the operator’s manual.
5-8
6 Disassembly and Repair
6.1 Tools Required
To disassemble and replace the parts and components, the following tools may be required:
Before disassembling the equipment, finish the following preparations:
Stop monitoring, turn off the equipment, and disconnect all the accessories and
peripheral devices.
Disconnect the AC power source and remove the battery.
WARNING
Before disassembling the equipment, be sure to eliminate the static charges first.
When disassembling the parts labeled with static-sensitive symbols, make sure you
are wearing electrostatic discharge protection such as antistatic wristband or gloves
to avoid damaging the equipment.
Properly connect and route the cables and wires when reassembling the equipment
to avoid short circuit.
Follow correct sequence to disassembly the equipment. Otherwise, the equipment
may be damaged permanently.
Be sure to disconnect all the cables before disassembling any parts. Be sure not to
damage any cables or connectors.
Place the screws and parts from the same module together to facilitate
reassembling.
Use care during reassembly to ensure cables are properly routed and all gaskets are
intact and correctly positioned.
6-1
T
6.3 Disassembling the Main Unit
NOTE
The recorder can be disassembled separately.
To disassemble the equipment, place the equipment on a work surface free from
foreign material, avoiding damaging the antiglare screen, touchscreen and the
knob. Be careful not to break the two cotters on the front ends of rear housing.
All repairs should be performed in an anti-static environment by qualified service
personnel.
6.3.1 Disassembling the Temperature Module (Optional)
1. Lay the monitor on a table as shown below. Unscrew the three M3×6 screws, pull the
Temperature module up, and disconnect the Temperature cable.
emperature
cable
6-2
6.3.2 Removing the Recorder
Unscrew the two M3×6 screws and pull the clamps as indicated to remove the recorder. Then
disconnect the cable.
Clamps
NOTE
The recorder can be disassembled separately.
6.3.3 Separating the Front and Rear Half of the Monitor
1. Lay the monitor on a table as shown below. Unscrew the four M3 screws.
6-3
2. Remove the recorder or recorder cover, and pull the rear housing out as indicated
below to separate the front and rear housing.
3. Stand the monitor and separate the front housing assembly and rear housing assembly
with caution. Disconnect the cable between the main board and power management
board and then take off the front panel.
Cable
NOTE
When reassembling the equipment, be sure to check if the front housing waterproof
strip is correctly placed.
6-4
6.3.4 Removing the Parameter Connector Panel Assembly
Lay the rear housing assembly of the equipment on a table. Disconnect the cable for AC
receptacle and SpO
Parameter connector panel assembly
cable, and then the silicon tube.
2
Silicon tube
Cable for AC
input
receptacle
SpO2signal
cable
6-5
6.3.5 Disassembling the Main Bracket Assembly
1. For monitors without a multi-function connector, use a flat blade screwdriver to gently
lift the cover release on the inside of the unit, and then remove the cover.
Cover
6-6
2. Disconnect the speaker cable, recorder cable, power cord and battery interface board
cable.
3. Unscrew the five self-tapping screws and one grounding screw. Then remove the main
bracket assembly.
Battery interface
board cable
Speaker cable
Recorder
cable
Grounding screw
Power cord
6-7
5 self-tapping
screws
6.3.6 Removing the Parameter Board (SpO
Optional) and Power
2
Management Board
1. Unscrew the three screws and remove the power management board assembly from
the main bracket assembly.
2. Unscrew the two M3 screws and remove the parameter board from the power
management board assembly.
Parameter Board
6-8
2 M3 screws
3. Unscrew the two M3 screws and remove the power management board assembly.
2 M3 screws
Power
management
board
6.3.7 Disassembling Pumps and Valves
1. Cut the two cable ties and remove the pump.
Cable tie
2. Unscrew the three screws as indicated and remove the NIBP board.
6-9
6.3.8 Disassembling AC/DC Power Board and Battery Converter Board
1. Unscrew the four screws as indicated and remove the AC/DC power board.
2. Unscrew the two M3 nuts to remove the battery interface board.
6.4 Disassembling the Front Housing Assembly
NOTE
To disassemble the equipment, place the equipment on a work surface free from
foreign material, avoiding damaging the antiglare screen, touchscreen and the
knob.
Remember to install the screen support pad properly during reassembly.
Operations relating to optional parts may not apply to your equipment.
Position the touchscreen properly with the flexible cable facing down.
6-10
T
6.4.1 Removing the Touchscreen Control Board
Remove the two M3 screws as shown below. Disconnect the cable between main board
and touchscreen board, and the touchscreen control board cable, and then remove the
touchscreen control board.
ouchscreen
Control Board
cable
Cable between
the main board
and touchscreen
board
6.4.2 Removing the Wi-Fi Module (Optional)
1. Remove the two antennas on the front panel from the slots as shown below:
Antenna
2. Remove the antennas from the Wi-Fi module PCBA.
2 M3 screws
Antenna
6-11
3. Push the clamps aside to remove the Wi-Fi module.
Clamps
6.4.3 Removing the Main Control Board
Disconnect the alarm lamp cable, cable for display backlight, display cable, and the
cable between the main board and keypad board respectively. Unscrew the four M3×8
screws and remove the main board, as shown below:
Cable
Cable
6-12
Cable
6.4.4 Removing SD Card
Remove the main board for Accutorr 7, and push the SD card as indicated below to
remove the SD card.
SD card
6.4.5 Removing the Touchscreen
Unscrew the seven PT3×8 screws as indicated below. Remove the touchscreen
assembly and then the touchscreen.
Tou ch sc re en
6-13
6.4.6 Disassembling the Display
Unscrew the four M3×6 screws indicated below to remove the screen.
6.4.7 Removing the Keypad
Unplug the encoder cable and unscrew the three PT3×8 screws indicated below.
Remove the keypad.
6.4.8 Removing the Encoder
Pull the knob off the encoder shaft. Remove the retaining nut. Then remove the encoder.
Knob
6-14
Encoder
shaft
T
T
T
6.4.9 Removing the Alarm Lamp
Unscrew the four PT2×6 screws indicated below and remove the alarm lamp board and
alarm lamp.
6.5 Disassembling the Temperature Module (Optional)
6.5.1 Removing the Temperature Module PCBA and Module Power Board
PCBA
Unscrew the two M3×6 screws and the two M3 screws. Remove the cover board and metal sheet.
Disconnect the Temperature board cable and the cable between Temperature isolation power
board and Temperature board. Unscrew the four M3×6 screws, you can remove the Temperature
module PCBA. Then unscrew the three M3×6 screws, you can remove the Temperature module
power board PCBA.
4 M3x6 screws
Screw for cover board
Cover board
emperature
module power
board PCBA
3 M3X6 screws
Screw for cover board
Metal sheet
Screw for metal sheet
6-15
emperature
board cable
Cable between
the isolation
power board and
Tem p board
emperature
module PCBA
T
6.5.2 Disassembling the Temperature On-Position Detection Board PCBA
1. Unscrew the four M3×6 screws as indicated and remove the temperature module
housing.
2. Unscrew the four M2 screws as indicated below and remove the Temperature
on-position detection board PCBA.
emperature on-position detection board PCBA
NOTE
Remember to assemble the silicon button for the Temperature on-position detection
switch during reassembly.
6-16
7 Parts
7.1 Introduction
This section contains the exploded views and parts lists of the main unit to identify the parts
during disassembling the monitor and replacing the parts.
This manual is based on fully loaded configuration. Your equipment may not contain all options
and therefore may not be appear the same as the unit in this manual.
7.2 Main Unit
7.2.1 Exploded View
7.2.2 Parts List
Item
No.
1 Accutorr 7 Front Housing Assembly 115-022922-00 Touch screen
11 Screw, pan head Phillips, M3×6 M04-002505--- N/A
14 Touchscreen control board PCBA 051-000881-00 N/A
15 Main board PCBA 115-023203-00 N/A
16 Screw, Pan Head W/Washer Phillips M3X6 M04-004012--- N/A
17 ST3.3X8 screw 030-000338-00 N/A
18 Keypad board adjusting sleeve
19 Keypad board cushion
115-022921-00
M04-051003--- N/A
115-018259-00 N/A
115-020467-00 N/A
front
housing
service kit
20 Keypad PCBA 051-001359-00 N/A
21 Encoder 801-0010-00010-00 N/A
22 Accutorr 7 silicon buttons 049-000607-00 N/A
23 Knob 043-003372-01 N/A
7-3
7.4 Rear Housing Assembly
7.4.1 Exploded View
7.4.2 Parts List
Item
No.
1 Rear Housing Subassembly 115-017698-00 N/A
2 Battery door assembly 043-003359-01 N/A
3 AC socket assembly 115-017699-00 N/A
4 Parameter connector panel assembly 115-017711-00 without SpO2
5 Parameter connector board support 043-003168-00 N/A
6 Power management board 115-018262-00 With multi-IO
7 Screw, Pan head with washer, Phillips M3X6 M04-004012--- N/A
8 Valve assembly 115-017679-00 N/A
9 Pump
10 Shock absorption cushion for pump
/ Fixing strip
Description FRU part number Remarks
801-9261-00040-00
7-4
NIBP pump
service kit
Item
No.
Description FRU part number Remarks
11 ST3.3X8 screw 030-000338-00 N/A
12 Speaker
13 Speaker pad
/
/
/
Cable between the power management
board and power board
Cable between the interface board and
main board
Cable between NIBP module and power
management board
801-9261-00010-00N/A
009-003237-00 N/A
9211-20-87225 N/A
009-003238-00 N/A
/ Recorder cable 009-001969-00 N/A
/ multifunctional connector cap 043-003311-01 N/A
7-5
7.5 Main Bracket Assembly
7.5.1 Exploded View
7.5.2 Parts List
Item No. Description FRU part number
1 Screw, Pan head w/washer M04-004012---
2 Power board insulator 047-010575-00
3 Power board 022-000125-00
4 Knob, Battery latch 0380-00-0593
5 Nut with washer
6 Battery spring
7 6301 battery interface PCBA
7-6
115-018254-00
7.6 Power Management Board Assembly
7.6.1 Exploded View
7.6.2 Parts List
Item
Description FRU part number Remarks
No.
Power management
1
board service kit
2 Plastic hexagon nut M90-000156--- N/A
3 SpO2 shield 047-010576-00 N/A
4 Plastic hexagon bolt 099-000129-00 N/A
5 Screw, pan head M04-051001-01 N/A
6 SpO2 board
Screw, Pan head
7
w/washer
115-018262-00
100-000106-00 Nellcor SpO
040-000109-01 SpO2 board, Masimo MS-2013
M04-004012--- N/A
7-7
with multifunctional
connector
board (MDU)
2
7.7 Parameter Connector Panel Assembly
7.7.1 Exploded View
7.7.2 Parts List
Item
Description FRU part number Remarks
No.
1 Parameter panel
2 NIBP fitting
3 SpO2 signal cable
/ Decorative NIBP socket 043-003304-01 N/A
/ Decorative IBP socket 043-001893-01 N/A
115-017711-00 Panel assembly (without SpO
009-003243-00 Masimo SpO
009-003244-00 Nellcor SpO2 module
7-8
module
2
)
2
7.8 Predictive Temperature Assembly
7.8.1 Exploded View
7.8.2 Parts List
Item
No.
1 Screw, Pan head w/washer M04-004012---
2 Predictive Temperature module power board PCBA 801-6006-00043-00
3 Predictive Temperature housing 043-003326-01
4 Temperature on-position detection board PCBA 051-001419-00
5 External compartment for Temperature module 043-003312-01
6 Temperature cover 049-000547-01
7 Silicon buttons M09A-20-62064
8 Screw, Pan head Self-Tapping PT2X6 M04-051003---
Description FRU part number
7-9
Item
No.
Description FRU part number
9 Screw, Flat Head Phillips, M3X6 M04-005005---
10 Predictive Temperature module PCBA 051-001435-00
/ Predictive Temperature board cable 009-003368-00
/
/
Cable between the Temperature module and power
management board
Cable between the isolation power board and Temperature
board
009-003239-00
009-003240-00
7-10
8 Hardware and Software Upgrade
8.1 Hardware Upgrade
The monitor supports upgrade of the following functions:
SpO
Temperature measurement
8.1.1 Upgrade Package
measurement
2
Upgrade package Description of upgrade package PN of upgrade
package
SpO2 Masimo SpO2 upgrade kit 115-023204-00
Nellcor SpO2 upgrade kit 115-018270-00
Temperature Temperature module 115-017716-00
Note: measurement accessories are not included in the above upgrade packages.
8.1.2 Upgrading Parameter Modules
8.1.2.1 Upgrading Nellcor SpO
Contents of upgrade package:
A Nellcor SpO
An SpO
A connector panel assembly for Nellcor SpO
Two M3×6 screws
1. Remove the power management board and connector panel assembly as described in
section 6.3 Disassembling the Main Unit.
2. Assemble the SpO
described in section 6.3.6 Removing the Parameter Board (SpO2 Optional) and Power Management Board.
3. Install the power management board assembly with the SpO
panel assembly in the service kit into the main unit as described in section 6.3 Disassembling the Main Unit.
board;
2
insulator;
2
2
board; and,
2
board and insulator onto the power management board as
2
board and the connector
2
8-1
8.1.2.2 Upgrading Masimo SpO
2
Contents of upgrade package:
A power management board assembly with Masimo SpO
A Masimo SpO
connector panel assembly.
2
; and,
2
1. Remove the power management board, connector panel assembly and the stopple of
multifunctional connector (if there is one) as described in section 6.3 Disassembling the Main Unit.
2. Install the power management board assembly with the SpO
board and the connector
2
panel assembly in the service kit into the main unit as described in section 6.3 Disassembling the Main Unit.
8.1.3 Upgrading Temperature
Contents of upgrade package:
A Temperature module with cables; and,
Two M3×6 screws.
Remove the decorative cover from the Temperature module connector. Install the Temperature
module onto the main unit as described in section 6.5 Disassembling the Temperature Module (Optional).
2. In the prompt menu, check the upgraded functions.
3. Click
to save the changes and quit the menu.
4. Restart the monitor and the software for upgraded parameters are enabled.
8.2 Software Upgrade
Software upgrades must be performed by Mindray, NA authorized service personnel. Call Service
Dispatch 1 800 288-2121 ext: 7875.
8-2
A Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive
maintenance program. They are a proven means of detecting abnormalities that, if undetected,
could prove dangerous to either the patient or the operator. Additional tests may be required
according to local regulations.
All tests can be performed using commercially available safety analyzer test equipment. Follow
the instructions of the analyzer manufacturer.
The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is
emphasized as a mandatory step if an approved agency status is to be maintained. The safety
analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of line
voltage and grounding, as well as total current loads.
A.1 Power Cord Plug
A.1.1 The Power Plug
Tes t Ite m Acceptance Criteria
The power plug pins No broken or bent pin. No discolored pins.
The plug body No physical damage to the plug body.
The power plug
The power cord
The strain relief
The power plug No loose connections.
No physical damage to the strain relief. No plug
warmth for device in use.
No physical damage to the cord. No deterioration
to the cord.
For devices with detachable power cords, inspect
the connection at the device.
For devices with non-detachable power cords,
inspect the strain relief at the device.
A-1
A.2 Device Enclosure and Accessories
A.2.1 Visual Inspection
Tes t Ite m Acceptance Criteria
No physical damage to the enclosure and accessories.
The enclosure and
accessories
No physical damage to meters, switches, connectors, etc.
No indication of exposure to fluid spills (e.g., water, coffee,
chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
A.2.2 Physical Inspection
Tes t Ite m Acceptance Criteria
No unusual noises (e.g., a rattle inside the case).
The enclosure and accessories
No unusual smells (e.g., burning or smoky smells,
particularly from ventilation holes).
No taped notes that may suggest device deficiencies or
operator concerns.
A.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and legible.