Service Manual
5-Lead, 5L, LT, XG
®
5-Lead, 5L, LT, XG
Service Manual
®
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 i
Table of Contents
Foreword ......................................................................................................................................v
Warnings, Precautions and Notes....................................................................................................v
Operation (5-Lead, 5L, LT and XG) ....................................................................................1 - 1
Introduction (5-Lead, 5L, LT) .......................................................................................................... 1 - 1
Controls, Indicators and Connectors (5-Lead, 5L, LT) ........................................................................ 1 - 3
Front Panel........................................................................................................................... 1 - 4
Left Side Panel...................................................................................................................... 1 - 9
Right Side Panel ................................................................................................................... 1 - 11
Remote Color Display............................................................................................................1-12
Operation (5-Lead, 5L, LT)............................................................................................................ 1 - 13
Setting-up / Turning Power On............................................................................................... 1 - 14
Factory - Default Control Settings ............................................................................................ 1 -15
Display................................................................................................................................ 1 - 16
Use of Menus....................................................................................................................... 1 - 23
Initiation of NIBP Measurements ............................................................................................. 1 - 25
ECG Acquisition................................................................................................................... 1 - 29
Invasive Pressure Acquisition.................................................................................................. 1-34
Sequence for Establishing SpO
2
............................................................................................. 1 - 34
Sequence for Establishing SpO
2
with Nellcor® Pulse Oximetry ................................................. 1 - 39
Respiration Monitoring ..........................................................................................................1-42
Gas Monitoring.................................................................................................................... 1 - 43
Alarms ................................................................................................................................ 1 - 49
How to Set the Clock............................................................................................................. 1 - 53
User Configuration Mode ...................................................................................................... 1 - 53
Recorder (optional) ............................................................................................................... 1 - 54
Status Messages................................................................................................................... 1 - 57
Monitor Problem Solving........................................................................................................ 1-62
Connection to External Devices............................................................................................... 1 - 64
Menus................................................................................................................................. 1 - 64
Introduction (XG)......................................................................................................................... 1 - 77
Controls, Indicators and Connectors (XG)....................................................................................... 1-77
Front Panel........................................................................................................................... 1 - 79
Left Side Panel...................................................................................................................... 1 - 87
Right Side Panel ................................................................................................................... 1 - 90
Rear Panel ........................................................................................................................... 1 - 91
Remote Color Display (optional) ............................................................................................. 1 - 92
Gas Module (Optional)- Front Panel ........................................................................................ 1 - 93
Gas Module (Optional)- Rear Panel......................................................................................... 1 - 95
Operation (XG)........................................................................................................................... 1 - 97
Setting-up / Turning Power On............................................................................................... 1 - 99
Factory - Default Control Settings ............................................................................................ 1 -99
Display................................................................................................................................ 1 - 102
Use of Menus....................................................................................................................... 1 - 109
Initiation of NIBP Measurements ............................................................................................. 1 - 110
ECG Acquisition................................................................................................................... 1 - 115
Invasive Pressure Acquisition.................................................................................................. 1-118
Sequence for Establishing SpO
2
............................................................................................. 1 - 119
Sequence for Establishing SpO
2
with Masimo® Pulse Oximetry ................................................ 1 - 124
Sequence for Establishing SpO
2
with Nellcor® Pulse Oximetry .................................................. 1 - 127
Respiration Monitoring ..........................................................................................................1-132
Mainstream CO
2
Gas Monitoring........................................................................................... 1 - 134
Sidestream CO
2
Gas Monitoring............................................................................................ 1 - 139
Gas Module Option.............................................................................................................. 1-146
Alarms ................................................................................................................................ 1 - 149
How to Set the Clock............................................................................................................. 1 - 153
User Configuration Mode ...................................................................................................... 1 - 153
Ta ble of C onte nts
ii 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
Recorder (optional) ............................................................................................................... 1 - 155
Status Messages................................................................................................................... 1 - 158
Monitor Problem Solving........................................................................................................ 1-168
Connection to External Devices (optional) ................................................................................ 1 - 172
Using the Passport XG with the Visa Central Station (optional).................................................... 1 - 172
Menus................................................................................................................................. 1 - 175
Theory of Operation.........................................................................................................2 - 1
Block Diagrams........................................................................................................................... 2 - 1
Detailed Circuit Descriptions......................................................................................................... 2 - 7
Interconnect Board................................................................................................................ 2 - 8
Frontend Board/Daughter Board ............................................................................................ 2 - 10
SpO
2
Board......................................................................................................................... 2-26
CPU Board...................................................................... .............. ............... ............... ......... 2 - 40
Line Battery Board................................................................................................................. 2 - 49
+12 / -23.5 Volt Converter Board .......................................................................................... 2 - 54
Planar EL Panel Board ...........................................................................................................2-55
CO2 Interface Module .......................................................................................................... 2 -57
NIBP Module: NIBP Pneumatic Board...................................................................................... 2 - 59
Panel Board / Display Interface.............................................................................................. 2 - 63
Termination Board ................................................................................................................ 2 - 66
Specifications....................................................................................................................3 - 1
Performance Specifications........................................................................................................... 3 - 1
Power Pack Electrical Ratings........................................................................................................ 3 - 8
Environmental Conditions........................................................ .............. ....................................... 3 - 8
Physical Characteristics................................................................................................................ 3 - 8
Agency Compliance.................................................................................................................... 3 - 8
Repair Information ...........................................................................................................4 - 1
Introduction ................................................................................................................................ 4 - 1
Safety Precautions....................................................................................................................... 4 - 1
Troubleshooting Guidelines .......................................................................................................... 4 - 2
Equipment and Special Tools Required........................................................................................... 4-3
Disassembly Instructions ............................................................................................................... 4 - 3
Mounting Instructions.... .............. ............... .............. ............... .............. ............... ........................ 4 - 11
Rolling Stand Assembly Instructions......................................................................................... 4 - 11
Bedrail Hook Mount..............................................................................................................4-11
Bed Mount........................................................................................................................... 4 - 11
Passport Mounting - Rolling Stand/Wall Mount ........................................................................ 4 - 11
Power Supply Mounting......................................................................................................... 4 -11
How to Use the Remote Color Display Mounting Kit......................................................................... 4 - 15
Cable Assembly Details................................................................................................................ 4 - 16
Assembly and Schematic Diagrams...................................................................................5 - 1
Replacement Parts............................................................................................................ 6 - 1
Introduction ................................................................................................................................ 6 - 1
Available Replacement Parts and Sub-Assemblies............................................................................ 6 - 1
Product Variations and Options..................................................................................................... 6 - 1
Exchange Program...................................................................................................................... 6 - 2
Replacement Parts Pricing Information............................................................................................ 6 - 2
Ordering Information................................................................................................................... 6 - 2
Abbreviations ............................................................................................................................. 6 - 4
Calibration .......................................................................................................................7 - 1
Introduction ................................................................................................................................ 7 - 4
Warnings and Guidelines ............................................................................................................ 7 - 4
Test Equipment and Special Tools Required .................................................................................... 7 - 4
Inputs......................................................................................................................................... 7 - 4
Diagnostics and Calibration Procedure .......................................................................................... 7-5
NIBP Diagnostics.................................................................................................................. 7 - 5
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 iii
Table of Contents
Error Journal ........................................................................................................................ 7 - 11
Exception Stack Frame .......................................................................................................... 7-11
Communication Test..............................................................................................................7-11
Recorder.............................................................................................................................. 7 - 14
Voltage Test......................................................................................................................... 7 - 14
Display Test ......................................................................................................................... 7 - 14
Keyboard Test...................................................................................................................... 7 - 17
SpO
2
Diagnostics (Datascope SpO
2
Installed).......................................................................... 7 - 18
SpO2 Diagnostics (Nellcor SpO
2
Installed).............................................................................. 7 - 20
SpO2 Diagnostics (Masimo SpO
2
Installed)............................................................................. 7 - 20
Power-Up Verification .................................................................................................................. 7 - 20
Initial Set-Up............................................................................................................................... 7 - 21
ECG Tests .................................................................................................................................. 7 - 21
IBP1 and IBP2 Verification............................................................................................................ 7 - 23
Temperature Verification .............................................................................................................. 7 - 23
SpO
2
Verification........................................................................................................................ 7 - 24
NIBP Verification......................................................................................................................... 7 - 24
Mainstream CO
2
Adapter Calibration ........................................................................................... 7 - 24
Mainstream CO
2
Sensor Calibration Verification...................................................................... 7 - 25
Mainstream CO
2
Sensor Calibration....................................................................................... 7 - 25
Sidestream Pump Calibration........................................................................................................ 7 - 26
Sidestream CO
2
Pump Verification / Calibration Procedure....................................................... 7 - 27
Sidestream Adapter Calibration.............................................................................................. 7 - 28
Sidestream CO
2
Sensor Calibration Verification....................................................................... 7 - 29
Sidestream CO
2
Sensor Calibration........................................................................................ 7 - 29
Battery Operation Verification....................................................................................................... 7 - 30
Leakage Current Test ................................................................................................................... 7 - 30
Cosmetic Checks......................................................................................................................... 7 - 32
Special Options Menu ................................................................................................................. 7 - 32
Preventive Maintena nce....................................................................... .............................8 - 1
Preventive Maintenance Schedule.................................................................................................. 8 - 1
Mechanical / Physical / Visual Inspection...................................................................................... 8-1
Preventive Maintenance Kit - Install Kit Annually .............................................................................. 8-1
Perform Verification and NIBP Calibration - Annually....................................................................... 8 - 2
Perform Verification and CO
2
Pump Calibration - Annually .......................................................... 8 - 2
User Preventive Maintenance Introduction....................................................................................... 8-2
Care and Cleaning of Monitor ............................................................................................... 8 - 2
Care and Cleaning of Datascope Flexisensors and Datasensors ................................................. 8 - 2
Cleaning CO
2
Sensors and Adapters...................................................................................... 8 - 3
Cleaning External CO
2
Sampling Components......................................................................... 8 - 3
Cleaning Internal CO
2
Sampling Components.......................................................................... 8 - 3
Sterilization and Cleaning of Reusable Cuffs............................................................................ 8 - 4
Battery Replacement and Maintenance.................................................................................... 8 - 5
Recorder Paper Replacement.................................................................................................. 8 - 5
Care and Storage of Thermal Chart Paper............................................................................... 8 - 5
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 iv
Table of Contents
This page intentionally left blank.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 v
Foreword Introduction
Foreword
This Service Manual is intended as a guide for technically qualified personnel during repair
and calibration procedures for the Passport 5-Lead, 5L, LT and XG monitors. The information
has been divided into the eight main sections as listed in the table of contents. A detailed
table of contents of each section is also provided on the first page of each section.
This publication may have been updated to reflect product design changes and/or manual
improvements. Any such changes to this manual would be accomplished by supply
replacement pages and instructions for inserting them into the manual.
For Passport external communication protocols (Datasette revision T or greater) see Service
Manual Supplements: P/N 0070-00-0302, Bedside-to-VISA Communication Protocol; P/N
0070-00-0304, Accutorr Communication Protocol; P/N 0070-00-0306, Passport
Communication Protocol (EzEm) and P/N 00700-00-0307, Datascope Improved ASCII
Protocol (DIAP).
NOTE: In order to ensure the proper performance of your
monitoring equipment and to prevent the voiding of the
warranty , it is recommended that only parts and accessories
provided by Datascope be u s e d with you monitor.
Warnings, Precautions and Notes
A WARNING is provided to alert the user to potential serious outcomes (death, injury, or
serious adverse events) to the patient or the user.
A CAUTION is provided to alert the user to use special care necessary for the safe and
effective use of the device. They may include actions to be taken to avoid effects on patients
or users that may not be potentially life threatening or result in serious injury, but ab out which
the user should be aware. Cautions are also provided to alert the user to adverse effects on
this device of use or misuse and the care necessary to avoid such effects.
A NOTE is provided when additional general information is applicable.
NOTE: Unauthorized servicing may void the remainder of the
warranty. Check with the factory or with a local authorized
Datascope representative to determine the warranty status
of a particular instrument.
Warnings
WARNING: The use of remote displays will enla rge all waveforms and
may alter aspec t ratios dep endin g on the remote de vice and
its set-up.
WARNING: Ensure that the conductive parts of ECG electrodes do not
contact other conductive parts including earth ground.
WARNING: Thoracic respiration measurement may inte rfer e wi th some
pacemakers. Refer to the pacemaker’s manual.
Introduction W arnings, Precau tions and Notes
vi 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
WARNING: Do not use the following 3-Lead ECG cables with Passport 5L
(P/N, 0998-00-0126-xx) and 5L-CE (P/N, 0998-00-0131-
xx): 0012-00-0620-05, -06, -07, -08; 0012-00-0722-05, -
06, -07, -08; 0012-00-0723-05, -06, -07, -08; 0012-00-
0724-05, -06, -07, -08. The above 3-Lead cables contain a
jumper wire which connects RL (right leg) to LL (left leg). As a
result, when you view aVR, aVL and aVF you will see the
appropriate ECG waveforms (even though you are using a
3-lead cable). However, there may be excessive noise on
these waveforms which can corrupt the heart rate
calculation. See Section 5 in the Operating Instructions for a
list of the proper ECG cables.
WARNING: When equipped with Nellcor ® SpO
2
, use only Nellcor®
oxygen transducers including Nellcor® OXISENSOR™
patient dedicated adhesive sensors. Use of other oxygen
transducers may cause improper oximete r pe rform ance.
WARNING: Tissue damage or inaccurate measurements may be caused
by incorrect sensor application or use, such as wr apping it
too tightly, app lying supplemental tape, failing to inspect
the sensor site periodically, or failing to position it
appropriately. Carefully read the sensor directions for use,
the Passport operating instructions, and all precautionary
information before use .
WARNING: Excessive ambient light may cause inaccurate
measurements. Cover the sensor site with opaque materi al.
WARNING: Inaccurate measurements may be caused by incorrect
sensor application or use; significant levels of dysfunctio nal
hemoglobins, (e.g., carboxyhemoglobin or methemoglobin);
or intra-v ascul ar dy es such as i ndocy anine green methyl ene
blue; exposure to excessive illumination, such as surgical
lamps (especially ones with a x enon light source), bilirubin
lamps, fluorescent lights, infrared heating lamp s, o r dire ct
sunlight; excessive patient movement; venous pulsations;
electro-surgi cal interference; and placement of a sensor o n
an extremity that has a blood pressure cuff, arterial
catheter, or intra-vascular line.
WARNING: In certain situations in which perfusion and signal strength
are low, such as in patients with thick or pigmented skin,
inaccurately low SpO
2
readings will result. Verification of
oxygenation should be made, especially in preterm infants
and patients with chronic lun g disease, before instituting
any therapy or intervention.
WARNING: Insure that the conductive parts of ECG electrodes do not
contact other conductive parts including earth ground.
WARNING: Many patients suffer from poor peripheral perfusion due to
hypothermia, hypovolem ia, se vere vasoconstriction,
reduced cardiac output, etc. These symptoms may cause a
loss in vital sign readings.
Introduction W arnings, Precau tions and Notes
vii 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
WARNING: The site should be checked at least every eight (8) hours
(every four (4) hours with the Adult re-usable finger sensor).
Ensure proper adhesion, skin integrity, and proper
alignment. Nail polish and fungus may effect readings.
Exercise ex t r eme caution with poorly perfused pa t i ents.
Skin erosion and pressure necrosis can be caused when
sensors are not frequently monitored. Assess the site ev ery
two (2) hours with poorly perfused patients.
WARNING: Do not use the following 3-Lead ECG cables with Passport
XG (P/N’s 0998-00-0133-xx, 0998-00-0134-xx and 0998-
00-0137-xx): 0012-00-0620-05, -06, -07, -08; 0012-00-
0722-05, -06, -07, 08; 0012-00-0723-05, -06, -07, -08;
0012-00-0724-05, -06, -07, -08. The above 3-Lead cables
contain a jumper wire which connects RL (right leg) to LL (left
leg). As a result, when you view aVR, aVL and aVF you will
see the appropriate ECG waveforms (even though you are
using a 3-lead cable). However, there may be excessive
noise on these waveforms which can corrupt the heart rate
calculation. See Section 5 in the Datascope P/N’s 0070-00-
397, 0070-00-0440, 0070-00-0503 of the Passport
Operating Instructions for a list of the proper ECG cables.
WARNING: If the sensor or patient cable is damaged in any way,
discontinue use immediately. To prevent damage do not
soak or immerse the sensor in any liquid solution. Do not
attempt to sterilize.
WA RNING: The maxim um sa mplin g ra te at the nasa l ca nnula i s 2 00 ml /
min. This device should not be used on patients whose
breathing could be impaired by this vac u um flow rate.
WARNING: Connection of the Passport’s exhaust port to the hospital’s
waste gas scavenge system is recommended to prevent
exposure of hospital personnel to the patient’s respiratory
sample.
WARNING: In an intubated configuration, ensure all tubing connections
are tight. If tubing disconnects on the monitor side of the
airway adapter, ventilator pressures cause inaccurate CO
2
waveforms and readouts to be displayed.
WA RNING: The maxim um sa mplin g ra te at the nasa l ca nnula i s 2 00 ml /
min. This device should not be used on patients whose
breathing could be impaired by this vac u um flow rate.
WARNING: Connection of the Gas Module exhaust port (50) to the
hospital’s waste gas scavenge system is stro ngly
recommended to prevent exposure of hospital personnel to
the patient’s respiratory sample. Vacuum (negative
pressure) should not exceed 1 mmHg at the Gas Module
Pump Exhaust fitting (50). Excessive scavenge vacuum may
result in damage to the Gas Module’s internal pum p.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 viii
Warnings, Precautions and Notes Introduction
Cautions
CAUTION: Connection of non-isolated devices to the optional RS232
Connector on this unit may cause chassis leakage to exceed
the specification standards.
CAUTION: Only use the Abbreviated Operating Check List if you are
already familiar with this product. If not, please continue
with the remainder of this chapter, Detailed Operating
Instructions.
CAUTION: Alway s at t a c h al l desired peripher al e qui pment prior to
powering on the Passport. If required to attach more or
change what is connected, turn the Passport off, attach
what is needed, and then turn the Passport on again.
CAUTION: Observe extreme cauti on on all patien ts (n eon a t es,
pediatrics, and adults) when NIBP is set to the Continuous
Mode. A 5 minute limit is placed on continuous
measurements. After 5 minutes the unit will take a
measurement once every 5 minutes. Reports have been
made of nerve injury occurring during use of automatically
cycled blood pressure cuffs. See the Appendix, “Precautions
when Using Automatically Cycled Blood Pressure
Cuffs”.Observe extreme caution on all patients (neonates,
pediatrics, and adults) when NIBP is set to the Continuous
Mode. A 5 minute limit is placed on continuous
measurements. After 5 minutes the unit will take a
measurement once every 5 minutes.
Reports have been made of nerve injury occurring during
use of automatically cycled blood pressure cuffs. See the
Appendix, “Precautions when Using Automatic ally Cycl ed
Blood Pressure Cuffs”.
CAUTION: It is the users responsibility to confir m co rrect operation of
the Passport with external devices. It is essential that a
biomedical engineer verify correct operation of the Passport
with an external device. Testing should include verification
that the maximum delay for cardioversion synchronization
does not exceed 60 ms (mon itor an d defibrillator). Cable
should be labeled with the devices they have been te sted
for use with.
CAUTION: Do not use this special “Y” shap ed power cable for any
devices other than the Passport and Gas Module II.
CAUTION: Vacuum (negative pressure) should not exceed 1 mmHg at
the Passport Pump Exhaust fitting (39). Excessive scavenge
vacuum may result in an “OCCLUSION” message or
damage to the Passport’s internal pump. The scavenge
system if used must be on during calibrati on.
CAUTION: It is the users responsibility to confir m co rrect operation of
the Passport XG with external devices. Before using this
cable it is essential that a biomedical engineer verify correct
operation of the Passport XG with an external device using
this cable. Testing should include verification that the
maximum delay for cardioversion synchronization does not
exceed 60 ms (monitor and defibrillator). Cable should be
labeled with the devices they have been tested for use with.
Introduction W arnings, Precau tions and Notes
ix 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
CAUTION: When cleaning sensors do not use excessive amounts of
liquid. Wi pe the sensor surface with a soft cloth, dampened
with the cleaning solution.
CAUTION: Using dark colored soaks may stain the cuffs. Test a single
cuff to ensure that no damage will occur. ETO sterilization
may also be used.
CAUTION: When ironing o r pre ssin g the cuffs, be aware that the
Velcro® fasteners can melt at temperatures above 325°F,
162°C.
Introduction W arnings, Precau tions and Notes
x 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
This page intentionally left blank.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 1
1.0
Operation (5-Lead, 5L, LT and XG)
Contents of this Chapter ..................................................Page
1.1 Introduction (5-Lead, 5L, LT) ..................................... 1-1
1.2 Controls, Indicators and Connectors (5-Lead, 5L, LT) . 1-3
1.3 Operation (5-Lead, 5L, LT) ........................................ 1-13
1.4 Introduction (XG) ..................................................... 1-77
1.5 Controls, Indicators and Connectors (XG) ................. 1-77
1.6 Operation (XG) ........................................................ 1-97
1.1 In troduction (5-Lead, 5L, LT )
This section of the Service Manual provides general information about the Datascope
Passport Monitors 5-Lead, 5L and LT. All functions, standard and optional, are described in
this section. The pictures of the unit and the displays should be used as samples and may not
be exactly as your unit appears. Table 1-1 on page 1-2 lists the functions available on the
Passport 5-Lead/5L and the Passport LT.
Introduction (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 2 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
NOTE: Sections 1.2 and 1.3 are included as a review of instrument
functions and operation, although the reader is encouraged
to refer to the Operating Instructions, P/N 0070-00-0324,
for more complete details.
TABLE 1-1
PASSPORT 5L PASSPORT LT
FUNCTIONS* STANDARD OPTIONAL STANDARD OPTIONAL
NIBP
••
IBP
••
ECG
••
SpO
2
- Datascope
••
SpO
2
- Nellcor
•
Respiration
••
CO
2
•
Recorder
••
Temperature
••
* For detailed instructions of each function see the Table of Contents for sections and page numbers.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 3
Operation (5-Lead, 5L, LT and XG ) Controls, Indicators and Connectors (5-Lead, 5L, LT)
1.2 Controls, Indicators and Connectors (5-Lead, 5L, LT)
The keys on the front panel of the Passport Monitor are classified as single action, repeat
action, or delayed action keys.
A single action key provides one action each time it is pressed, regardless of how long it is
held.
A repeat action key provides the an action when pressed, then waits half a second before
repeating the action until the key is released.
A delayed action key provides an action, but only after the key has been held pressed for a
(key specific) period of time.
1. p and q (Set-up) 8. ZERO 1 (IBP) (Opt ional15. MUTE (Alarms)
2. SELECT (Set-up) 9. ZERO 2 (IBP) (Optional)16. TREND/RETURN (Trends)
3. END (Set-up) 10. START (NIBP) 17. TIME UP p and DOWN
q (Trends)
4. CENTRAL SILENCE
(Optional)
11. INTERVA L (NIBP) 18. BEEP VOLUME
5. LEAD (ECG) 12. DEFLATE (NIBP) 19. RECORD (Optional)
6. SIZE (ECG) 13. VOL UM E (A larms) 20. DC INPUT
7. FREEZE (ECG) 14. CHECK/RETURN
(Alarms)
21. DISPLAY
PARAGRAPH
NUMBER DESCRIPTION
CONTROL/DISPLAY
NUMBER
PAGE
NUMBER
1.2.1 Front Panel 1 - 21 1-4
1.2.2 Left Side Panel 22 - 30 1-9
1.2.3 Right Side Panel 31 - 34 1-11
1.2.4 Remote Color Display 1-12 12 - 14
Controls, Indicators and Connectors (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
1.2.1 Front Panel
FIGURE 1-1 Front Panel Controls
Introduction
The keys on the front panel of the Passport Monitor are classified as single action, repeat
action, or delayed action keys.
A single action key provides one action each time it is pressed, regardless of how long it is
held.
A repeat action key provides the an action when pressed, then waits half a second before
repeating the action until the key is released.
A delayed action key provides an action, but only after the key has been held pressed for
a (key specific) period of time.
NOTE: Only one key function will be recognized a t any time. The
Passport will ignore multiple key se lections.
All key actions are acknowledged by a key click, except for BEEP VOL and ALARM VOL. If a
key is not available a double key click will sound.
1. p and q (UP and DOWN) (Set-Up)
Two repeat action keys used to move the highlighted screen cursor or change the setting/
value of a menu item. As the cursor is moved, the highlighted menu window is displayed.
2. SELECT (Set-Up)
A repeat action key used to select the function or value indicated by the highlighted cursor.
Datascope
Passport
R
TM
ALARMS
VOLUME
SET-UP
SEL ECT
END
CENTRAL
SILENCE
LEAD
SIZ E
FREEZE
ZERO 1
ZERO 2
START
INTE RVAL
DEFLAT E
ECG IBP NI BP
CHECK
RETURN
MUTE
TREN D
RETURN
TIM E
BEEP
VO LUM E
RECORD
TRENDS
TIME
!
1
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 2019
21
D.C. INPUT
MODEL
EL
5
1
14
2 3
4 5
6
7
11 20
18
17
9
8 10
12
13
15
16
19
21
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 5
Operation (5-Lead, 5L, LT and XG ) Controls, Indicators and Connectors (5-Lead, 5L, LT)
3. END (Set-Up)
A single action key which causes the display to return to the main screen display (as
specified in the DISPLAY section). This key is always available.
4. CENTRAL SILENCE (Optional)
A single-action k ey which m utes the a larms of the c orrespon ding cha nnel on the VISA Centra l
Station. The amount of time the alarms are silenced, is determined by the VISA.
5. LEAD (ECG)
A single action key which selects the ECG lead to be displayed. Each depression of the key
selects and displays the next ECG lead from the list. The list wraps around after the last entry
is selected. Choices are: I, II, III, aVR, aVL, aVF or V.
6. SIZE (ECG)
A single action key which selects the ECG size to be displayed. Each depression of the key
selects and displays the next ECG size from the list. The list wraps around after the last entry
is selected.
7. FREEZE (Screen)
A single action key which enables or releases the screen freeze function. The freeze key
stops or starts the ECG waveform (waveform 1), except when waveform 2 is used for
cascaded ECG. When this is the case, pressing the freeze key the f irst time causes the
currently displayed ECG waveform data to be transferred to waveform 2 and frozen.
Waveforms 1 and 3 continue to move. Pressing FREEZE again causes waveform 2 to return
to cascaded ECG.
8. ZERO 1 (IBP) (Optional)
A delayed action key which zeros the current pressure in the BP1 channel. If the transducer
zero process is unsuccessful, “UNABLE TO ZERO” is displayed in the window containing BP
value. This key is only available when a pressure transducer is present.
9. ZERO 2 (IBP) (Optional)
A delayed action key which zeros the current pressure in the BP2 channel. If the transducer
zero process is unsuccessful, “UNABLE TO ZERO” is displayed in a temporary window
within the parameter window. This key is only available when a pressure transducer is
present.
10. START (NIBP)
A single action key which initiates an NIBP measurement. This function is not available if a
measurement is in progress.
11. INTERVAL (NIBP)
A repeat action key which selects the interval setting for NIBP measurements. Values wrap
around at the lowest/highest choices. Choices are: Off, Continuous, 1, 2.5, 5, 10, 15, 20,
30, 60, and 120 minutes. Five seconds after the selection is made and the key is released,
the new interval takes effect.
Controls, Indicators and Connectors (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
12. DEFLATE (NIBP)
A single action key which stops any NIBP measurement in progress, including any timed
measurement sequence, and deflates the cuff. When a timed measurement is stopped an
“NIBP: DEFLATE” message followed by an “NIBP: IDLE” message displays in the NIBP
message window until the timer mode is restarted by either pressing START or changing the
interval.
13. VOLUME (Alarms and Alarm LED)
A repeat action key which adjusts the alarm volume in 5 steps. There is no off position for this
tone. The tone wraps to the minimum volume once the maximum is reached. The red alarm
bell LED illuminates when an alarm condition exists. Refer to “Alarms” on page 1-49.
14. CHECK/R ET UR N (Alarms)
A single action key which provides access to the alarm menu to view and select alarm
values. Press SELECT to change the alarm values. Pressing CHECK/RETURN a second time
or the END key, returns the display to the main screen and exits the alarm menu.
15. MUTE (Ala rms)
A single action and delayed action key which silences all currently alarming parameters or
suspends all alarms. A single key press will silence the current alarm tone for 2 minutes. Any
new alarms that occur while the alarm tone is muted will disable the mute and sound the
alarm tone. An alarm mute symbol is displayed next to each muted parameter. The word
Mute is displayed above the menu selections.
When the key is pressed and held for 3 seconds, all set alarms will be suspended for 2
minutes. This is indicated by the Alarm Mute Symbol and the words All Mute displayed
above the menu selections.
When “Aud alm standby” is set to ON (in the User Configuration Menu), and the MUTE key
is pressed and held for 4 seconds, all alarms are indefinitely suspended. This is indicated by
the Alarm Mute Symbol in all parameter windows and the words Aud Alm Sby flashing
above the menu selections.
The All Mute and Aud Alm Sby modes can be exited by pressing the mute key.
NOTE: When “Aud alm standby” is set to ON and the unit is power
off and then on again, the unit will power up in the Aud alm
standby mode.
16. TRENDS/RETURN (Trends)
A single action and delayed action key which displays and clears trended data. The most
recent page of data is displayed when the key is first pressed. Press this key a second time to
view additional trended data. Pressing and holding the key for 3 seconds clears all trended
data.
17. TIME UP p and q TIME DOWN (Trends)
Two repeat action keys which are used to scroll through the trend data.
Alarm Mute
Symbol
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 7
Operation (5-Lead, 5L, LT and XG ) Controls, Indicators and Connectors (5-Lead, 5L, LT)
18. BEEP VOLUME
A repeat ac ti o n k ey w h ich ad j us ts th e bee p v ol u me in 5 s te ps plu s OFF. Wh en t h e m axi mu m
volume is reached, the volume will wrap around to OFF , then to the minimum volume
19. RECORD (Optional)
A combination single action and delayed action key which initiates or stops a recording.
Pressing the key once in itiates a printout. Holding the key pressed for three seconds, when
waveforms are displayed, initiates a continuous printing of waveforms. Pressing the key
while the printing is in progress, stops the recorder. The recorder prints either waveforms or
trend data, depending on what is displayed (i.e. trend data is printed when trend list is
displayed.). The Recorder Set-up will determine which waveform or waveforms are printed.
The Recorder Set-up will also determine the Record Destination - Local (recording from th e
Passport), Remote (recording from the VISA, no recording from the Passport), or Both
(recording from the Passport and VISA).
NOTE: When a remote (VISA) recordin g is printe d, the information
that is printed is determined by the set-up in the VISA.
20. DC INPUT
A green LED used to indicate that the POWER Switch (33) is in the ON position.
FIGURE 1-2 Display
MENU
AREA
PARAMETER
AREA
ECG AREA
MULT I-FUNCTION AREA
MESSAGE AREA
Controls, Indicators and Connectors (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
21. DISPLAY
The Display is used to present information which is divided into 5 graphic display areas.
They are:
A. MENU Display Area
B. ECG Display Area
C. PARAMETER Display Area
D. MULTI-FUNCTION Display Area
E. MESSAGE Display Area
NOTE: Only one key function will be recognized a t any time. The
Passport will ignore multiple key se lections.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 9
Operation (5-Lead, 5L, LT and XG ) Controls, Indicators and Connectors (5-Lead, 5L, LT)
1.2.2 Left Side Panel
FIGURE 1-3 Left Side Panel with Datascope SpO
2
FIGURE 1-4 Left Side Panel with Nellcor® SpO
2
SpO
2
CUFF
T
ECG /EK G
E.S.I.S.
All Patient Connections
Electrica lly Isolat e d
Press to Open
Replace only with:
IBP 1
IBP 2
CO
2
BATT.
PI 2
PI 1
!
!
o
22
23
24
25
26
27
28
29
30
10cm
22
23
24
29
30
28
25
26
27
SpO
2
CUFF
T
ECG /EKG
E.S.I.S.
All Patie nt Co nnections
Electrica lly Isolate d
Press to Op en
Replace only with:
IBP 1
IBP 2
CO
2
BATT.
PI 2
PI 1
!
!
10cm
o
22
23
24 25
26
27
28
29
30
22
23
26
27
28
29
30
25
24
Controls, Indicators and Connectors (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
22. RECORDER (Optional)
A two trace thermal strip chart recorder with integral paper spool.
23. BATTERY PACK
Two internal, user replaceable, rechargeable, sealed lead acid batteries that provide power
to operate the monitor when not connected to the power pack. The battery packs may be
removed and replaced independently while the unit is operating.
24. CUFF
A connector used to attach the NIBP cuff assembly to the monitor.
25. SpO
2
An 8-pin DIN type or 9-pin sub miniature D (Nellcor) type female connector used to attach
the Sp
O
2
sensor assembly to the monitor.
26. T (Temperature)
A standard three wire phone jack used to mate with either the YSI series 400 or 700*
temperature probes. The Passport automatically identifies which probe is connected.
27. ECG / EKG
A six-pin AAMI (ECG-D10/75) standard connector (AMP P/N 864900-1; Datascope P/N
0131-00-0079) used for patient cable connections.
*Feature applicable only if available installed in your unit.
28. IBP1 (Optional)
A six-pin male connector (Datascope P/N 0131-00-0094) used for Datascope specified
pressure transducers listed in Chapter 5, Accessories.
29. IBP2 (Optional)
A six-pin male connector (Datascope P/N 0131-00-0094) used for Datascope specified
pressure transducers listed in Chapter 5, Accessories.
30. CO
2
(Optional)
A 20 pin connector used to attach the CAPNOSTAT C
O
2
sensor to the monitor.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 11
Operation (5-Lead, 5L, LT and XG ) Controls, Indicators and Connectors (5-Lead, 5L, LT)
1.2.3 Right Side Panel
FIGURE 1-5 Right Side Panel
31. DATASETTE
A user replaceable software cartridge used for installing updated software revisions.
32. DC POWER INPUT (CONNECTOR)
A four terminal connector to supply low voltage DC power to the unit and charge the
batteries. The power pack provides power to the unit independent of battery installation.
33. J1
A communication interface connector used to conn ect the Passport to a VISA Central Station
Monitor, Remote Color Display, Nurse Call, DPD Defibrillator or other peripheral devices.
34. POWER
A recessed rocker switch which interrupts power to the main unit but does not prevent
charging of the batteries.
31
32
33
34
Controls, Indicators and Connectors (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 12 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
1.2.4 Remote Color Display
FIGURE 1-6 Remote Color Display
For instructions on mounting the remote display to a wall, refer to Operating Instructions.
WARNING: The use of remote displays will enla rge all waveforms and
may alter aspec t ratios dep endin g on the remote de vice and
its set-up.
2. Connect PASSPORT M onitor
Video Interface cable,
P/N 0012-00-0983. to the Remote
Color Display Interface cable,
P/N 0012-00-0994.
1. Connect Interface cable,
P/N 0012-00-0994, to the Analog
Input connector on the Remote
Color Display.
3. Connect the PASSPORT
Monitor Video Interface Cable to
J1 on the PASSPORT Monitor
REMOTE COLOR DISPLAY, REAR
PASSPORT MONITOR , F RONT VIEW
Remote Color Display Interface Cable
P/N 0012-00-0994
P
assport Video Interface Cable
P/N 0012-00-0983
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 13
Operation (5-Lead, 5L, LT and XG ) Operation (5-Lead, 5L, LT)
1.3 Operation (5-Lead, 5L, LT)
Abbreviated Operating Check List
CAUTION: Only use the Abbreviated Operating Check List if you are
already familiar with this product. If not, please continue
with the remainder of this chapter, Detailed Operating
Instructions.
A. Setting-Up
1. Set POWER switch to OFF.
2. Connect, if desired, peripheral equipment (i.e., Remote Color Display, P.C., etc.).
3. Attach power pack and/or install charged batteries as needed.
4. Set POWER switch to ON.
5. Using the menus and keyboard, set (when appropriate) the following:
--
B. Initiating an NIBP Measurement
1. Select cuff.
2. Attach cuff hose to NIBP connector and place cuff on patient.
3. Select timer interval, if desired.
4. Select cuff pressure, if necessary.
5. Press START to begin an NIBP measurement.
6. Press DEFLATE to suspend measurement.
C. Establishing SpO
2
1. Select the appropriate sensor.
2. Attach sensor to Sp
O
2
connector and apply to the patient.
3. Set either waveform 2 or 3 in the set-up menu to display the Sp
O
2
waveform, if desired.
• Patient Size and Trend Clear • HR Information
• Set Up Information (speed, waveform • Sp
O
2
Information 2 and 3, etc.)
• ECG Information • Recorder Information (when available)
• Resp. Information • Alarm Limits
• Alarm Volume • NIBP Information
•Beep Volume •IBP Information
Operation (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 14 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
D. Establishing CO
2
/ CAPNOSTAT*
1. Plug sensor into the side of the monitor.
2. If a new CAPNOSTAT is used, calibrate the system.
3. Apply sensor to the breathing circuit.
4. Set Waveform 2 or 3 to CO
2
.
5. Set the Resp. Source to CO
2
.
* CAPNOSTAT is a tra demark of NOVAMETRIX, Inc.
E. Recording Information
1. Select wave to be recorded through Record Menu or if desired to record tabular trend,
press TREND/RETURN.
2. Press Record to start recording function.
3. Press Record again to stop the recording function.
F. Temperature Measurement
1. Attach the desired temperature probe to the Temperature connector on the side of the
Passport. The Passport automatically detects which probe is connected, 400 or 700
series.
Detailed Operating Instructions
This section of the Service Manual provides guidelines and step-by-step instructions for
proper operation of the monitor. Numbers in parentheses ( ) relate to the displays and
controls described in “Controls, Indicators and Connectors (5-Lead, 5L, LT)” on page 1-3
1.3.1 Setting-up / Turning Power On
1. Set the POWER switch (34) to OFF.
2. Attach any peripheral equipment, i.e., Visa Central Station, Remote Color Display.
CAUTION: Always attach all desired peripheral equipment pr ior to
powering on the Passport. If required to attach more or
change what is connected, turn the Passport off, attach
what is needed, and then turn the Passport on again.
3. Attach the power pack to the DC POWER INPUT connector (32). If battery operation is
required, ensure that two fully charged batteries are installed.
NOTE: If power pack is connected ensure that the screw is
tightened so that the cord does not detach.
4. Set the POWER switch (34) to ON.
Internal self tests will run and the display will come on. A “DIAGNOSTIC IN
PROGRESS” message will display and once the self tests have been completed a “SELF
TEST COMPLETE” message will display. After this the main screen is displayed.
If any failures occur, see “Status Messages” on page 1-57 for further instructions.
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 15
Operation (5-Lead, 5L, LT and XG ) Operation (5-Lead, 5L, LT)
1.3.2 Factory - Default Control Settings
The following are the factory initial (default) control settings. This is the state that the unit will
power up in unless a “current configuration” has been saved from the set up menu.
TAB LE 1- 2 Default Settings
MENU FUNCTION DEFAULT SETTINGS
Patient Size Adult
Clear Trend Memory No
Pacer Enhanc ement Off
Set-up Speed 25 mm/sec.
Resp Speed 12.5 mm/sec.
Waveform 2 Cascade ECG
Waveform 3 Pleth
Resp Source ECG
Powerup Settings No change
Record Wave Selection ECG
Record on Alarm No
Recorder Destination Local
Alarms Low High
Heart Rate Off Off
SpO
2
85 Off
IBP1 Sys Off Off
IBP1 Dia Off Off
NIBP Sys Off Off
NIBP Dia Off Off
IBP2 Mean Off Off
Resp Rate Off Off
Apnea Delay Off Adult, Off Ped., 10 se c
Neonate
ETCO
2
Off Off
NIBP Start Pressure 180 mmHg Adult, 140 mmHg
Ped., 120 mmHg Neonate
IBP IBP1 Scale Range 150 mmHg
IBP2 Scale Range 37.5 mmHg
Respiration Scale 3 cm/ohm
HR Source Auto
SpO
2
Pleth Scale* 3
CO
2
Scale Range 40 Torr
Trend Trigger NIBP
Interval Timer 5 min
ECG Lead II
Size 1 cm/mV
NIBP Interval Off
Volume Beep Off
Alarm 1
Operation (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 16 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
Changes can be made and saved to these default settings. The default settings can also be
restored. Refer to “Setting-up / Turning Power On” on page 1-14.
* If your unit is equipped with Nellcor® SpO
2
, there is no pleth scale adjustment. This is replaced with operat-
ing mode options 1, 2, and 3.
** Farhenheit is the default setting for English units only. All other language units default to Centigrade.
1.3.3 Display
FIGURE 1-7 Window Layout
The Display is divided into five graphic areas as shown above.
MENU Display Area - displays the main menu selections available with the cursor and
select keys, the battery symbol and the mute categories.
ECG Display Area - displays the ECG trace, ECG information and pacer enhancement
status.
P ARAM ET ER Display Area - displays the current values of patient parameters.
Temperature Scale ºF**
HrSpO
2
Size Hr SpO
2
Size HR small, SpO
2
large
Audio Alarm de lay Audio Alarm delay Off
Aud alm stand by Aud alm standby Off
Serial Output T ype Serial Output Type Visa
TAB LE 1-2 D ef au lt Settings (Continued)
MENU FUNCTION DEFAULT SETTINGS
MENU
AREA
PARAMETER
AREA
ECG AREA
MUL TI-FUNCTION AREA
MESSAGE AREA
Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 17
Operation (5-Lead, 5L, LT and XG ) Operation (5-Lead, 5L, LT)
MULTI-FUNCT I ON Display Area - displays additional waveforms and temporary boxes
for menu functions.
MESSAGE Display Area - displays messages relating to NIBP, SpO
2
, CO
2
and recorder
operation.
1.3.3.1 Menu Display Area
FIGURE 1-8 Window Layout
The menu area displays the main menu selections available. These are accessed by using the
UP p and DOWN q arrow keys (1) and SELECT (2) set-up keys. One of the menu item is
always highlighted by the cursor. This window also contains the TIME, DATE, PATIENT SIZE
and MUTE CATEGORY (when active) information.
For each patient size available there are different choices within the menu selections.
T able 1-3 on page 1-18 indicates the choices for each menu set-up and also where there are
different selections for each patient size.
A list of all menus and the selections available is provided in T able 1-2 on page 1-15. For a
graphic representation of each of these menus, see “Use of Menus” on page 23.
NOTE: The “IBP”, “Recorder” and “CO
2
” menu items only appear
only on models equipped with these options.
NOTE: The display sample shown above may not be an exact
representation of your unit.
Operation (5-Lead, 5L, LT) Operation (5-Lead, 5L, LT and XG)
1 - 18 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual
TABLE 1-3 Menu List
MENU
LIST MENU ITEM CHOICES
Patient Size Adult, Pediatric, Neonate
Clear Trend Memory No, Yes
Pacer Enhancement On, Off
Setup Speed 12.5, 25, 50 mm/sec.
Resp Source Off, ECG, CO
2
Resp Speed 3.125, 6.25, 12.5, 25 mm/sec.
Waveform 2 BP1, ECG casc., Pleth., Resp., CO
2
Waveform 3 BP1, BP2, Pleth., Resp., CO
2
Power up Settin gs No Change, Save Cur rent, Restore Factory
Record Record on Alarm Yes, No
Wave Selectio n ECG, Pleth., Re sp., ECG and BP1, ECG and BP2, BP1
and BP2, ECG and Resp., CO
2
Record Destination Local, Remote Both
Alarm HR Low (Off, 30-100), High (Off, 100-250) bpm
SpO
2
Low (50-99), High (Off, 80-100) %
IBP1 Sys Adult: Low (Off, 5 - 130), High (Off, 70 - 240) mmHg
Ped.: Low (Off, 5-130), High (Off, 40 - 180) mmHg
Neo.: Low (Off, 5-130), High (Off, 40 - 180) mmHg
IBP1 Dia Adult: Low (Off, 5 - 90), High (Off, 40 - 130) mmHg
Ped.: Low (Off, 5 - 50), High (Off, 50 - 100) mmHg
Neo.: Low (Off, 5 - 50), High (Off, 50 - 100) mmHg
NIBP Sys Adult: Low (Off, 50 - 150), High (Off, 70 - 240) mmHg
Ped.: Low (Off, 15 - 130), High (Off, 40-180) mmHg
Neo.: Low (Off, 15 - 130), High (Off, 40-180) mmHg
NIBP Dia Adult: Low (Off, 30 - 120), High (Off, 40 - 130) mmHg
Ped.: Low (Off, 10 - 50), High (Off, 50 - 100) mmHg
Neo.: Low (Off, 10 - 50), High (Off, 50 - 100) mmHg
IBP2 (Mean) Adult: Low (Off, 2 - 100), High (Off, 5 - 150) mmHg
Ped.: Low (Off, 2 - 50), High (Off, 5 - 100) mmHg
Neo.: Low (Off, 2 - 50), High (Off, 5 - 100) mmHg
Respiration Low (Off, 5 - 50), High (Off, 30 - 200) bpm
Apnea Delay Adult Off, 10 -40 secs.
Ped. Off, 10 - 30 secs.
Neonate 10 - 20 secs.
ETCO
2
Torr: Low (Off, 5 - 60), High (Off, 20 - 100) %: Low
(Off, 1 - 8), High (Off, 2 - 11) kPa: Low (Off, 1.0 -
8.0), High (Off, 2.0 - 11.0)
NIBP Start Pressure Adult: 100-260 mmHg
Ped.: 60-180 mmHg
Neonate: 40-120 mmHg
IBP BP1 Scale Range 37.5, 75, 150, 300 mmHg
BP2 Scale Range 37.5, 75, 150, 300 mmHg
Resp Scale 1, 2, 3, 4, 5
HR H R S ou r ce Auto, ECG, BP1, SpO
2
SpO
2
(Standard)
Pleth Size 1, 2, 3, 4, 5