Teachware
Metrohm AG
CH-9101 Herisau
teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked
with great care, errors cannot be entirely excluded. Should you notice any
mistakes please send us your comments using the address given above.
Figure 11Connecting a remote cable ............................................................. 14
Figure 12Attaching the exchange unit ........................................................... 14
Figure 13Keypad 870 KF Titrino plus .............................................................. 17
Figure 14Directory structure on the USB stick ................................................. 45
Figure 15Rotational speed depending on stirring rate .................................... 70
Figure 16Pin assignment of remote socket and plug ...................................... 74
Figure 17Remote status diagram .................................................................... 75
Figure 18Connecting the RS-232/USB Box to the PC ...................................... 76
Table of figures
870 KF Titrino plus
■■■■■■■■
VII
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
1.1 Instrument description
The 870 KF Titrino plus is an instrument used for volumetric water determination according to Karl Fischer. There are method templates available
which are already configured except for a few parameters. If parameters are
changed, then these will be saved automatically. The methods can be exported on a connected USB stick. This function makes it possible for you to copy
methods quickly and easily from one instrument to another.
1.1.1 Titration and measuring modes
The following titration and measuring modes are supported:
■KFT
Volumetric water determination according to Karl Fischer.
Measuring modes:
– Ipol (voltametric measurement with adjustable polarization cur-
rent)
– Upol (amperometric measurement with adjustable polarization
voltage)
1 Introduction
1.1.2 Connectors
The instrument is equipped with the following connectors:
■MSB connector (Metrohm Serial Bus)
■USB (OTG) connector
■Electrode connector
■Remote connector
1.1.3 Intended use
The 870 KF Titrino plus is designed for usage as a titrator in analytical laboratories. Its main application field is volumetric titration.
The present instrument is suitable for processing chemicals and flammable
samples. The usage of the 870 KF Titrino plus therefore requires that the
user has basic knowledge and experience in the handling of toxic and caustic
substances. Knowledge with respect to the application of the fire prevention
measures prescribed for laboratories is also mandatory.
For connecting a stirrer.
The 6.2151.100 adapter can be used to connect, for example, a printer,
a USB Stick or a USB keyboard.
For connecting polarizable electrodes.
For connecting instruments with a remote interface.
870 KF Titrino plus
■■■■■■■■
1
1.2 About the documentation
1.2 About the documentation
Caution
Please read through this documentation carefully before putting the
instrument into operation. The documentation contains information and
warnings which have to be followed by the user in order to ensure safe
operation of the instrument.
1.2.1 Symbols and conventions
The following symbols and styles are used in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the second to the instrument part in the figure.
■■■■■■■■■■■■■■■■■■■■■■
Instruction step
Carry out these steps in the sequence shown.
MethodDialog text, parameter in the software
File ▶ New
[Next]Button or key
Menu or menu item
Warning
This symbol draws attention to a possible life hazard
or risk of injury.
Warning
This symbol draws attention to a possible hazard due
to electrical current.
Warning
This symbol draws attention to a possible hazard due
to heat or hot instrument parts.
Warning
This symbol draws attention to a possible biological
hazard.
■■■■■■■■
2
Caution
This symbol draws attention to a possible damage of
instruments or instrument parts.
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
1.3 Safety instructions
1.3.1 General notes on safety
Warning
This instrument may only be operated in accordance with the specifications in this documentation.
This instrument has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
instrument, the following instructions must be observed carefully.
1.3.2 Electrical safety
The electrical safety when working with the instrument is ensured as part of
the international standard IEC 61010.
1 Introduction
Note
This symbol marks additional information and tips.
Warning
Only personnel qualified by Metrohm are authorized to carry out service
work on electronic components.
Warning
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are
touched.
There are no parts inside the housing which can be serviced or replaced
by the user.
Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
870 KF Titrino plus
Only operate this instrument with a mains voltage specified for it (see rear
panel of the instrument).
■■■■■■■■
3
1.3 Safety instructions
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be
destroyed by discharges.
Always pull the mains cable out of the mains connection socket before
connecting or disconnecting electrical appliances on the rear panel of the
instrument.
1.3.3 Tubing and capillary connections
Caution
Leaks in tubing and capillary connections are a safety risk. Tighten all
connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can
be used to loosen connections.
■■■■■■■■■■■■■■■■■■■■■■
Check the connections regularly for leakage. If the instrument is used
mainly in unattended operation, then weekly inspections are mandatory.
1.3.4 Flammable solvents and chemicals
Warning
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
■Set up the instrument in a well-ventilated location (e.g. laboratory
flue).
■Keep all sources of flame far from the workplace.
■Clean up spilled fluids and solids immediately.
■Follow the safety instructions of the chemical manufacturer.
1.3.5 Recycling and disposal
This product is covered by European Directive 2002/96/EC, WEEE – Waste
from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative
effects on the environment and public health.
■■■■■■■■
4
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
More details about the disposal of your old equipment can be obtained from
your local authorities, from waste disposal companies or from your local
dealer.
870 KF Titrino plus
■■■■■■■■
5
2 Overview of the instrument
2
5
4
3
1
■■■■■■■■■■■■■■■■■■■■■■
Figure 1Front 870 KF Titrino plus
Piston rod
1
Of the dosing drive.
Coupling
3
For switching the flat stopcock.
Keypad
5
Contact pins
2
For the data chip.
Display
4
■■■■■■■■
6
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
1
2
5
4
3
6
2 Overview of the instrument
Figure 2Rear 870 KF Titrino plus
Type plate
1
Contains specifications concerning mains
voltage and serial number.
USB (OTG) connector
3
For connecting printers, USB sticks, USB hubs,
etc.
Remote connector
5
For connecting instruments with a remote
interface. D-Sub, 9-pin.
Mains connection socket
2
MSB connector
4
Metrohm Serial Bus.
For connecting a stirrer. Mini DIN, 9-pin.
Electrode connector (Pol.)
6
For connecting polarizable electrodes, e.g.
double Pt electrodes. Socket F.
870 KF Titrino plus
■■■■■■■■
7
■■■■■■■■■■■■■■■■■■■■■■
EP
KFT Ipol
KFT Upol
EP
V/mL
Continuous
dosing
Initial
dosing
V/mL
V/mL
U/mV
t/s
I/µA
Control range
Control range
3 Water determination according to Karl Fischer
(KFT)
Karl Fischer Titration is a method for volumetric water determination. Conditioning is carried out automatically both before and after the actual
titration. The reagent dosing is controlled in such a way that a predefined
endpoint is reached as quickly and as accurately as possible. The volume
steps and the rate of reagent dosing are regulated by the difference between
the current measured value and the predefined endpoint. This means that
titration is performed more slowly in the control range and that smaller volumes are added. The titration is stopped at the endpoint either driftcontrolled or after a waiting time. The volume dosed until the endpoint is
used to calculate the water content of the sample.
■■■■■■■■
8
Figure 3Reagent dosing for KFT
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
4.1 Setting up the instrument
4.1.1 Packaging
The instrument is supplied in highly protective special packaging together
with the separately packed accessories. Keep this packaging, as only this
ensures safe transportation of the instrument.
4.1.2 Checks
Immediately after receipt, check whether the shipment has arrived complete
and without damage by comparing it with the delivery note.
4.1.3 Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
4 Installation
Place the instrument in a location of the laboratory suitable for operation
and free of vibrations, if possible protected from corrosive atmospheres and
contamination by chemicals.
The instrument should be protected against excessive temperature fluctuations and direct sunlight.
4.2 Connecting a sensor
Connecting a polarizable electrode
Figure 4Connecting a polarizable electrode
870 KF Titrino plus
■■■■■■■■
9
4.3 Connecting a stirrer
Note
The electrode cable is protected against accidental disconnection of the
cable by means of a pull-out protection feature. If you wish to remove
the plug, then you must first retract the outer plug sleeve.
4.3 Connecting a stirrer
You can connect the following stirrers:
■801 Stirrer
■803 Ti Stand
■804 Ti Stand (requires rod stirrer)
■■■■■■■■■■■■■■■■■■■■■■
Figure 5Connecting a stirrer
Caution
Make sure that the flat side of the plug matches the marking on the
socket.
4.4 Connecting a balance
Balances are equipped with a serial RS-232 interface as a rule. To connect a
balance, you require a 6.2148.030 RS-232/USB Box.
■■■■■■■■
10
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
6.2151.100
4 Installation
Figure 6Connecting a balance
When a 6.2151.020 USB cable is used, then the 6.2148.030 RS-232/USB
Box can be connected to the 870 KF Titrino plus by means of a USB hub or
a 6.2151.100 adapter (see Chapter 4.5, page 11).
Connect the 9-pin plug of the respective balance connecting cable to the
RS 232/1 connector. Consult the user manual of the balance in order to
select the correct connecting cable.
The parameters for the RS-232 interface on the instrument must match those
on the balance (see "Editing the COM1 settings", page 47). Additionally
consult the user manual of the balance.
4.5 Connecting keyboard, printer and other USB devices
The 870 KF Titrino plus has a USB (OTG) connector. Use the provided
6.2151.100 adapter USB MINI (OTG) - USB A for connecting USB devices as
e.g. printers, keyboards or USB sticks, see the following figure.
Figure 7Connecting USB devices
870 KF Titrino plus
■■■■■■■■
11
4.5 Connecting keyboard, printer and other USB devices
Caution
Switch the instrument off before connecting or disconnecting a USB
device or a USB stick.
The 870 KF Titrino plus can recognize the device immediately after switching on.
The following devices can directly be operated on the USB connector withthe 6.2151.100 adapter:
■USB sticks (for the backup or storing of methods)
■6.2147.000 numerical USB keypad
■6.2148.030 RS-232/USB Box (for connecting balances or for the RS-232
remote control)
■USB hub (with or without an own power supply)
The 6.2147.000 numerical USB keypad serves for comfortable numerical input and for navigating in the dialog. In addition, it provides two USB
connectors. Connect additional USB devices to the keypad.
■■■■■■■■■■■■■■■■■■■■■■
Note
Most of the USB devices need a so-called hub in order to work correctly.
A USB hub is a distributor to which several USB devices can be connected.
USB hubs are available in specialty stores in a number of different models.
The USB (OTG) connector of the 870 KF Titrino plus has no such hub. The
6.2147.000 numerical USB keypad has a USB hub and two USB connectors.
The following devices can only be connected to a 6.2147.000 numer-ical keypad or to a USB hub:
■Printer (with USB connector, use the 6.2151.020 connecting cable)
■Barcode reader (with USB cable)
■Mouse (PC mouse with USB cable, for navigating in the dialog)
The following devices can only be connected to a USB hub:
■PC keyboard (with USB cable, for the comfortable input of letters and
numbers)
■Keypad with numerical keypad (with USB cable)
■■■■■■■■
12
If you wish to connect a variety of instruments without own power
supply, then you must possibly use a USB hub with own power supply (self
powered). The USB (OTG) connector of the 870 KF Titrino plus is not
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
USB MINI (OTG)-USB
USB stick
6.2151.100
USB MINI (OTG)-USB
USB stick
Keypad
6.2147.000
6.2151.100
Printer
USB MINI (OTG)-USB
USB stick
USB-Hub
RS-232/USB Box
6.2148.030
6.2151.100
Printer
4 Installation
designed for supplying power to several devices with elevated electricity
requirements.
Also observe the instructions in chapter 10.4, page 71.
Examples:
Figure 8Connecting the USB stick
Figure 9Connecting the 6.2147.000 USB keyboard with USB stick and
printer
Figure 10Connecting the USB hub with USB stick, printer and the
4.6 Connecting instruments to the remote connector
■■■■■■■■■■■■■■■■■■■■■■
4.6 Connecting instruments to the remote connector
The 870 KF Titrino plus can be integrated in an automation system with the
aid of a remote cable.
Figure 11Connecting a remote cable
A variety of different connecting cables are available for connecting
Metrohm instruments (e.g. sample changers) (see Optional Accessorieschapter).
4.7 Attaching the exchange unit
Figure 12Attaching the exchange unit
To attach the exchange unit, proceed as follows:
Slide the exchange unit from the front onto the 870 KF Titrino plus and
1
push all the way to the rear.
■■■■■■■■
14
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
It must snap in audibly.
870 KF Titrino plus
■■■■■■■■
15
5.1 Switching the instrument on and off
5 Operation
5.1 Switching the instrument on and off
Switching on the instrument
Proceed as follows:
■Press the red [STOP] key.
1
The instrument is initialized and a system test performed. This process takes some time.
■If a buret unit has been attached, then a request appears to carry
out the PREP function:
■■■■■■■■■■■■■■■■■■■■■■
All tubings and the cylinder are rinsed with the PREP (Preparing)
function. The preparing of the buret unit is described in chapter
"Preparing the buret unit (PREP)", page 34.
■Confirm the message with [OK].
The display of this message can be switched off in the system settings (see "PREP warning", page 41).
The main dialog is displayed:
Switching off the instrument
The instrument is switched off with the [STOP] key. The fact that the key
needs to be pressed down for an extended time prevents accidental switch
off.
■■■■■■■■
16
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
Proceed as follows:
■Keep the red [STOP] key pressed down for at least 3 s.
1
A progress bar is displayed. If the key is released during this time, then
the instrument will not be switched off.
5.2 Fundamentals of operation
5.2.1 The keypad
Figure 13Keypad 870 KF Titrino plus
5 Operation
BACKApply the input and exit the dialog.
⇧⇩Move the selection bar either up or down by one
line at a time. Select the character to be entered
in the text editor.
⇦⇨Select the character to be entered in the text and
number editor. Select the individual functions in
the function bar.
OKConfirm the selection.
STOPStop an ongoing method run or a manual func-
tion. Switch the instrument on/off.
STARTStart a method run or a manual function.
5.2.2 Structure of the dialog windows
870 KF Titrino plus
The current dialog title is displayed on the left-hand side of the title line. The
current status of the system is displayed in the upper right-hand corner.
readyThe instrument is in normal state.
■■■■■■■■
17
5.2 Fundamentals of operation
cond.busyThe working medium is being conditioned.
cond.okThe working medium is conditioned.
busyA method has been started.
Some dialogs have a so-called function bar on the bottom line. The functions
contained therein can be selected with the arrow keys [⇦] or [⇨] and executed with [OK].
5.2.3 Navigating in the dialog
The selection bar is displayed in inverted style. Use the arrow keys [⇧] and
[⇩] to move the selection bar upward or downward one line at a time. If a
dialog text is marked with ">", then additional settings are available in a
subordinate dialog. Use [OK] to access this dialog.
■■■■■■■■■■■■■■■■■■■■■■
Example: System settings
Use the [BACK] key to return to the next higher level.
5.2.4 Entering text and numbers
In the editing dialog for text or numerical input you can select the individual
characters with the arrow keys. Use [OK] to apply the character in the input
field. The following functions are available:
■■■■■■■■
18
870 KF Titrino plus
■■■■■■■■■■■■■■■■■■■■■■
5 Operation
Editing functionDescription
Accept
Cancel
Delete
[BACK]
The modification is applied and the editing dialog
is exited.
The editing dialog is exited without applying the
modification.
The contents of the input field is deleted completely.
The character left of the cursor is deleted (backspace).
Text editor only
The cursor within the input field is shifted to the
left by one character each time that [OK] is
pressed.
Text editor only
The cursor within the input field is shifted to the
right by one character each time that [OK] is
pressed.
The modification is applied and the editing dialog
is exited.
The [BACK] key has the same function as Accept.
A commercially available USB keyboard can be connected to make it easier
to enter text and numbers. The assignment of the keys on the PC keyboard
is described in chapter 10.4.2, page 71.
5.2.5 Selecting from a selection list
In a selection list, select the individual entries with the arrow keys [⇧] and
[⇩]. Accept the selection with [OK] or [BACK].
870 KF Titrino plus
■■■■■■■■
19
5.3 Methods
5.3 Methods
5.3.1 Method templates
The 870 KF Titrino plus contains method templates which are already configured except for a few parameters.
The following method templates can be selected:
KFT IpolMethod with adjustable polarization current.
Titer IpolTiter determination with adjustable polarization
Blank IpolBlank value determination with adjustable polari-
KFT Ipol–BlankMethod with adjustable polarization current.
■■■■■■■■■■■■■■■■■■■■■■
The blank value is not taken into account for the
calculation.
current.
zation current.
KFT UpolMethod with adjustable polarization voltage.
Titer UpolTiter determination with adjustable polarization
Blank UpolBlank value determination with adjustable polari-
KFT Upol–BlankMethod with adjustable polarization voltage.
5.3.2 Loading a method
To load a method, proceed as follows:
1
Open the method table
■In the main dialog, select Method and press [OK].
The blank value is taken into account for the calculation.
The blank value is not taken into account for the
calculation.
voltage.
zation voltage.
The blank value is taken into account for the calculation.
■■■■■■■■
20
The method table with the stored templates opens:
870 KF Titrino plus
Loading...
+ 74 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.