Metrohm 867pH User Manual

867 pH Module
Manual
8.867.8003EN
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
867 pH Module
8.867.8003EN
Manual
07.2013 ek
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://documents.metrohm.com.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 Introduction 1
1.1 pH Module system ................................................................ 1
1.2 Instrument description ......................................................... 2
1.3 Measuring modes and commands ...................................... 2
1.4 About the documentation ................................................... 3
1.4.1 Symbols and conventions ........................................................ 3
1.5 Safety instructions ................................................................ 4
1.5.1 General notes on safety ........................................................... 4
1.5.2 Electrical safety ........................................................................ 5
1.5.3 Working with liquids ................................................................ 6
1.5.4 Recycling and disposal ............................................................. 6
2 Overview of the instrument 7
Table of contents
3 Installation 9
3.1 Setting up the instrument .................................................... 9
3.1.1 Packaging ................................................................................ 9
3.1.2 Checks .................................................................................... 9
3.1.3 Location .................................................................................. 9
3.2 Connecting a controller ........................................................ 9
3.2.1 Operation ................................................................................ 9
3.3 Connecting MSB devices .................................................... 13
3.3.1 Connecting a dosing device ................................................... 14
3.3.2 Connecting a stirrer or titration stand .................................... 15
3.3.3 Connecting a Remote Box ..................................................... 16
3.4 Connecting USB devices ..................................................... 17
3.4.1 General ................................................................................. 17
3.4.2 Connecting a USB hub ........................................................... 17
3.4.3 Connecting a printer .............................................................. 18
3.4.4 Connecting a balance ............................................................ 18
3.4.5 Connecting a PC keyboard (only for operation with Touch
Control) ................................................................................. 20
3.4.6 Connecting a barcode reader ................................................. 20
3.5 Connecting sensors ............................................................ 22
3.5.1 Connecting a pH, metal or ion-selective electrode .................. 22
3.5.2 Connecting a reference electrode .......................................... 22
3.5.3 Connecting a polarizable electrode ........................................ 23
3.5.4 Connecting a temperature sensor or an electrode with inte-
grated temperature sensor .................................................... 24
3.5.5 Connecting an iConnect ........................................................ 24
867 pH Module
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
4 Operation and maintenance 26
4.1 General notes ...................................................................... 26
4.1.1 Care ...................................................................................... 26
4.1.2 Maintenance by Metrohm Service .......................................... 26
4.2 Quality Management and qualification with
Metrohm ............................................................................. 27
5 Troubleshooting 28
5.1 General ................................................................................ 28
6 Appendix 30
6.1 Remote interface ................................................................ 30
6.1.1 Pin assignment of the remote interface .................................. 30
7 Technical specifications 34
7.1 Measuring interface ........................................................... 34
7.1.1 Potentiometry ........................................................................ 34
7.1.2 Temperature .......................................................................... 35
7.1.3 Polarizer ................................................................................ 35
7.2 Power connection ............................................................... 36
7.3 Safety specifications ........................................................... 36
7.4 Electromagnetic compatibility (EMC) ................................ 36
7.5 Ambient temperature ......................................................... 37
7.6 Reference conditions .......................................................... 37
7.7 Dimensions .......................................................................... 37
7.8 Interfaces ............................................................................. 37
8 Warranty (guarantee) 39
9 Accessories 41
9.1 Scope of delivery ................................................................ 41
9.1.1 2.867.0010 ........................................................................... 41
9.1.2 2.867.0110 ........................................................................... 42
9.1.3 2.867.0210 ........................................................................... 46
9.2 Optional accessories ........................................................... 50
Index 54
■■■■■■■■
IV
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 867 pH Module system ..................................................................... 1
Figure 2 Front 867 pH Module ........................................................................ 7
Figure 3 Rear of 867 pH Module ..................................................................... 8
Figure 4 Connecting the Touch Control ......................................................... 10
Figure 5 Connecting the computer ................................................................ 12
Figure 6 MSB connections ............................................................................ 13
Figure 7 Connecting a dosing device ............................................................. 15
Figure 8 Connecting an MSB stirrer ............................................................... 15
Figure 9 Connecting the propeller stirrer to the titration stand ...................... 16
Figure 10 Connecting the Remote Box ............................................................ 16
Figure 11 Connecting a printer ....................................................................... 18
Figure 12 Connecting a pH, metal or ion-selective electrode ........................... 22
Figure 13 Connecting a reference electrode .................................................... 23
Figure 14 Connecting a polarizable electrode .................................................. 23
Figure 15 Connecting a temperature sensor or an electrode with integrated tem-
perature sensor ............................................................................... 24
Figure 16 Connecting the iConnect ................................................................. 25
Figure 17 Connecting an electrode to the iConnect ........................................ 25
Figure 18 Connectors of the Remote Box ........................................................ 30
Figure 19 Pin assignment of remote socket and remote plug .......................... 30
Table of figures
867 pH Module
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
MSB
Controller
PC Keyboard
Barcode
Reader
USB/RS-232 Converter
Touch Control
USB Sample Processor
Robotic Titrosampler
USB Hub
Relay Box
Remote Box
Balance
Dosing Interface
Dosino
Stirrer / Ti Stand
Dosimat
On
Status
Titrando
USB
Cond. cell
Sensors
Printer
Tower-PC
867 pH Module
1 Introduction

1.1 pH Module system

1 Introduction

867 pH Module
Figure 1 867 pH Module system
■■■■■■■■
1

1.2 Instrument description

1.2 Instrument description
The 867 pH Module has the following characteristics:
Operation
A touch-screen Touch Control or a high-performance PC software is used for operation.
Dosing
Dosinos and Dosimats can be used for dosing. Dosing is required for standard additions and when auxiliary solutions are to be dosed.
MSB connectors
Four MSB connectors (Metrohm Serial Bus) for connecting dosing devi­ces (Dosimat with exchange unit or Dosino with dosing unit), stirrers and Remote Boxes.
USB connectors
Two USB connectors, with which devices such as printers, PC key­boards, barcode readers or additional control instruments (USB Sample Processor, Titrando, Dosing Interface, etc.) can be connected.
Measuring interface
One measuring input each for:
– a potentiometric electrode (pH, metal or ion-selective electrode) – a separate reference electrode – a temperature sensor (Pt1000 or NTC) – a polarizable electrode – an iConnect (measuring interface for electrodes with integrated
data chip, so-called iTrodes)
■■■■■■■■■■■■■■■■■■■■■■

1.3 Measuring modes and commands

The following measuring modes and commands are supported:
MEAS
The following measuring modes can be selected for measurements:
pH (pH measurement) – U (potentiometric voltage measurement) – T (temperature measurement) – Ipol (voltametric measurement with selectable polarization cur-
rent)
Upol (amperometric measurement with selectable polarization
voltage)
Conc (concentration measurement)
■■■■■■■■
2
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
STDADD
The measuring modes for standard addition are listed separately only in
tiamo™. In Touch Control they are integrated in the measuring mode
MEAS Conc.
The following measuring modes can be selected for measurements:
auto (automatic addition of the standard addition solution by
specifying a potential difference)
dos (automatic addition of the standard addition solution by
specifying the individual volume increments)
man (manual addition of the standard addition solution)
CAL
Electrode calibration.
Measuring mode:
pH (calibration of pH electrodes) – Conc (calibration of ion-selective electrodes)
ELT
Electrode test for pH electrodes.
This mode is listed separately only in tiamo™. In Touch Control, the
electrode test is part of the CAL calibration mode.
Dosing commands
The following commands for dosing can be selected:
PREP (rinsing the cylinder and tubings of an exchange unit or
dosing unit) – EMPTY (emptying the cylinder and tubings of a dosing unit) – ADD (dosing a specified volume) – LQH (carrying out complex dosing tasks with a Dosino)

1.4 About the documentation

CAUTION
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe opera­tion of the instrument.

1.4.1 Symbols and conventions

The following symbols and formatting may appear in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
867 pH Module
■■■■■■■■
3

1.5 Safety instructions

■■■■■■■■■■■■■■■■■■■■■■
Instruction step
Carry out these steps in the sequence shown.
Method Dialog text, parameter in the software
File New Menu or menu item
[Next] Button or key
WARNING
This symbol draws attention to a possible life-threat­ening hazard or risk of injury.
WARNING
This symbol draws attention to a possible hazard due to electrical current.
WARNING
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
1.5 Safety instructions

1.5.1 General notes on safety

WARNING
This instrument may only be operated in accordance with the specifica­tions in this documentation.
WARNING
This symbol draws attention to a possible biological hazard.
CAUTION
This symbol draws attention to possible damage to instruments or instrument parts.
NOTE
This symbol highlights additional information and tips.
■■■■■■■■
4
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■

1.5.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
1 Introduction
WARNING
WARNING
Mains voltage
WARNING
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
Protection against electrostatic charges
WARNING
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.
867 pH Module
■■■■■■■■
5
1.5 Safety instructions

1.5.3 Working with liquids

CAUTION
Periodically check all system connections for leaks. Observe the relevant regulations in respect to working with flammable and/or toxic fluids and their disposal.

1.5.4 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
■■■■■■■■■■■■■■■■■■■■■■
■■■■■■■■
6
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
2 Overview of the instrument
Figure 2 Front 867 pH Module

2 Overview of the instrument

"On" LED
1
Lights up when the pH Module is ready for operation.
867 pH Module
■■■■■■■■
7
Figure 3 Rear of 867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Electrode connector (Ind.)
1
For connecting pH, metal or ion-selective electrodes with integrated or separate refer­ence electrode. Socket F.
Temperature sensor connector (Temp.)
3
For connecting temperature sensors (Pt1000 or NTC). Two B sockets, 2 mm.
Electrode connector (iConnect)
5
For connecting electrodes with integrated data chip (iTrodes).
USB connector (USB 1 and USB 2)
7
USB ports (type A) for connecting printer, keyboard, barcode reader, USB Sample Pro­cessor, etc.
Power socket
9
Electrode connector (Ref.)
2
For connecting reference electrodes, e.g. Ag/AgCl reference electrode. Socket B, 4 mm.
Electrode connector (Pol.)
4
For connecting polarizable electrodes, e.g. double Pt wire electrodes. Socket F.
Connector (Controller)
6
For connecting a Touch Control or a PC with installed PC software. Mini DIN, 9-pin.
Type plate
8
Contains specifications concerning supply voltage, instrument type and serial number.
MSB connector (MSB 1 to MSB 4)
10
Metrohm Serial Bus. For connecting external dosing devices, stirrers or Remote Boxes. Mini DIN, 9-pin.
■■■■■■■■
8
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
3 Installation

3.1 Setting up the instrument

3.1.1 Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

3.1.2 Checks

Immediately after receipt, check whether the shipment has arrived com­plete and without damage by comparing it with the delivery note.

3.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.

3 Installation

Place the instrument in a location of the laboratory which is suitable for operation, free of vibrations, protected from corrosive atmosphere, and contamination by chemicals.
The instrument should be protected against excessive temperature fluctu­ations and direct sunlight.

3.2 Connecting a controller

3.2.1 Operation

Two different versions are available for operating the 867 pH Module:
A Touch Control with touch-sensitive screen. It forms a "stand-alone
instrument" together with the 867 pH Module.
A computer enables operation of the 867 pH Module with the help of
a PC software, e.g. tiamo.
CAUTION
Take care to ensure that the power supply cable is pulled out of the power socket before either setting up or disconnecting connections between the instruments.
867 pH Module
■■■■■■■■
9
3.2 Connecting a controller
3.2.1.1 Connecting a Touch Control
NOTE
The plug is protected against accidental disconnection of the cable by means of a pull-out protection feature. If you wish to pull out the plug, you will first need to pull back the outer plug sleeve marked with arrows.
Connect the Touch Control as follows:
Insert the plug of the Touch Control connection cable into the
1
Controller socket.
■■■■■■■■■■■■■■■■■■■■■■
Figure 4 Connecting the Touch Control
Connect the MSB devices (see Chapter 3.3, page 13).
2
Connect the USB devices (see Chapter 3.4, page 17).
Connect the pH Module to the power supply.
3
Switch on the Touch Control.
4
The Touch Control power supply is supplied through the pH Module. Automatic system tests are performed on both instruments at the time of activation. The On LED on the front of the pH Module lights up when the system test has been completed and the instrument is ready for operation.
■■■■■■■■
10
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
CAUTION
The Touch Control must be shut down properly by deactivation with the power switch on the rear of the instrument before the power sup­ply is interrupted. If this is not done, then there is a danger of data loss. Because of the fact that the power supply for the Touch Control is pro­vided through the pH Module, you must never disconnect the pH Mod­ule from the power supply (e.g. by deactivating with a connector strip) before you have deactivated the Touch Control.
If you would prefer not to position the Touch Control directly next to the pH Module, then you can lengthen the connection with the 6.2151.010 cable. The maximum connection length permitted is 5 m.
3.2.1.2
Connecting a computer
The 867 pH Module requires a USB connection to a computer in order to be able to be controlled by a PC software. Using a 6.2151.000 controller cable, the instrument can be connected directly, either to a USB socket on a computer, to a connected USB hub or to a different Metrohm control device.
You need administrator rights for the installation of driver software and control software on your computer.
Cable connection and driver installation
A driver installation is required in order to ensure that the 867 pH Module is recognized by the PC software. To accomplish this, you must comply with the procedures specified. The following steps are necessary:
1
Installing the software
Insert the PC software installation CD and carry out the installa-
tion program directions.
Exit the program if you have started it after the installation.
2
Establishing the cable connections
Connect all peripheral devices to the instrument, see Chapter 3.3,
page 13 and see Chapter 3.4, page 17.
Connect the instrument to the power supply if you have not
already done this. The "On" LED on the 867 pH Module is not yet illuminated!
Connect the instrument to a USB connector (Type A) of your com-
puter (see manual of your computer). The 6.2151.000 cable is used for this purpose.
867 pH Module
■■■■■■■■
11
3.2 Connecting a controller
■■■■■■■■■■■■■■■■■■■■■■
Figure 5 Connecting the computer
The instrument is recognized. Depending on the version of the Win­dows operating system used, the driver installation proceeds differ­ently afterwards. Either the necessary driver software is installed automatically or an installation wizard is started.
Follow the instructions of the installation wizard.
3
The "On" LED on the 867 pH Module lights up when the driver installation has been completed and the instrument is ready for oper­ation.
If problems should occur during installation, contact your company's IT support team.
NOTE
The plug on the instrument end of the 6.2151.000 controller cable is protected against accidental disconnection by means of a pull-out pro­tection feature. If you wish to pull out the plug, you will first need to pull back the outer plug sleeve marked with arrows.
Registering and configuring the instrument in the PC soft­ware
The instrument must be registered in the configuration of your PC soft­ware. Once that has been done, you can then configure it according to your requirements. Proceed as follows:
■■■■■■■■
12
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
MSB
Stirrer / Ti Stand
Dosimat / Dosino
Dosimat
Remote Box
Dosino / Dosimat
Dosino
Relay Box
Ti Stand / Stirrer
1
Setting up the instrument
Start the PC software.
The instrument is automatically recognized. The configuration dia­log for the instrument is displayed.
Make configuration settings for the instrument and its connec-
tors.
More detailed information concerning the configuration of the instrument can be found in the documentation for the respective PC software.

3.3 Connecting MSB devices

In order to connect MSB devices, e.g. stirrers or dosing devices, Metrohm instruments are equipped with up to a maximum of four connectors on what is referred to as the Metrohm Serial Bus (MSB). Various kinds of peripheral devices can be connected in sequence (in series, as a "daisy chain") at a single MSB connector (8-pin Mini DIN socket) and con­trolled simultaneously by the respective control instrument. In addition to the connection cable, stirrers and the Remote Box are each equipped with their own MSB socket for this purpose.
3 Installation
The following figure provides an overview of the instruments that can be connected to an MSB socket, along with a number of different cabling variations.
Figure 6
MSB connections
The control instrument determines which peripheral devices are suppor­ted.
867 pH Module
■■■■■■■■
13
3.3 Connecting MSB devices
■■■■■■■■■■■■■■■■■■■■■■
NOTE
When connecting MSB devices together, the following must be observed:
Only one device of the same type can be used at a single MSB con-
nector at one time.
Type 700 Dosino and 685 Dosimat dosing devices cannot be con-
nected together with other MSB instruments on a shared connector. These dosing devices must be connected separately.
CAUTION
Exit the control software before you plug in MSB instruments. When it is switched on, the control instrument automatically recognizes which device is connected to which MSB connector. The operating unit or the control software enters the connected MSB devices into the system configuration (device manager).
MSB connections can be extended with the 6.2151.010 cable. The maxi­mum connection length permitted is 15 m.

3.3.1 Connecting a dosing device

Four dosing devices can be connected to the instrument (MSB 1 to MSB
4).
The types of dosing devices that are supported are:
800 Dosino 700 Dosino 805 Dosimat 685 Dosimat
Proceed as follows:
1
Connecting a dosing device
Exit the control software. Connect the connection cable of the dosing device to one of the
sockets marked with MSB on the rear of the control instrument.
Start the control software.
■■■■■■■■
14
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Figure 7 Connecting a dosing device

3.3.2 Connecting a stirrer or titration stand

You can use the following instruments:
With built-in magnetic stirrer (stirring "from below"):
– 801 Stirrer – 803 Ti Stand
Without built-in magnetic stirrer (stirring "from above"):
– 804 Ti Stand with propeller stirrer 802 Stirrer
3 Installation
Connect a stirrer or a titration stand as follows:
1
Connecting the stirrer or titration stand
Exit the control software. Connect the connection cable of the magnetic stirrer or of the
titration stand to one of the sockets marked with MSB on the rear of the control instrument.
804 Ti Stand only: Connect the propeller stirrer to the stirrer con-
nector (socket with stirrer symbol) of the titration stand.
Start the control software.
Figure 8 Connecting an MSB stirrer
867 pH Module
■■■■■■■■
15
3.3 Connecting MSB devices
Figure 9 Connecting the propeller stirrer to the titration stand

3.3.3 Connecting a Remote Box

Instruments that are controlled via remote lines and/or that send control signals via remote lines can be connected via the 6.2148.010 Remote Box. In addition to Metrohm, other instrument manufacturers also use similar connectors that make it possible to connect different instruments together. These interfaces are also frequently given the designations "TTL Logic", "I/O Control" or "Relay Control" and generally have a signal level of 5 volts.
■■■■■■■■■■■■■■■■■■■■■■
Control signals are understood to be electrical line statuses or electrical pulses (> 200 ms) which display the operating status of an instrument or which trigger or report an event. Sequences on a variety of instruments can thus be coordinated in a single complex automation system. No exchange of data is possible, however.
Proceed as follows:
1
Connecting the Remote Box
Exit the control software. Connect the Remote Box connection cable to one of the sockets
marked with MSB on the rear of the control instrument.
Start the control software.
Figure 10 Connecting the Remote Box
■■■■■■■■
16
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
You can, for example, connect an 849 Level Control (fill level monitoring in a canister) or a 731 Relay Box (switch box for 230/110 volt alternating current sockets and low-voltage direct current outlets). The Remote Box also has an MSB socket at which a further MSB device, e.g. a dosing device or a stirrer, can be connected.
You will find precise information concerning the pin assignment of the interface on the Remote Box in the appendix.

3.4 Connecting USB devices

3.4.1 General

The 867 pH Module has two USB connectors (type A sockets) for periph­eral devices with USB interfaces. The 867 pH Module functions as a USB hub (distributor) no matter how it is operated. If you wish to connect more than two devices to the USB, you can also use an additional, com­mercially available USB hub.
CAUTION
3 Installation
CAUTION
If you operate the 867 pH Module with the aid of the Touch Control, take care to ensure that the Touch Control is switched off when you set up or disconnect connections between the various instruments. If you use a PC software to control the 867 pH Module, you should exit the program before you set up or disconnect the USB connections.

3.4.2 Connecting a USB hub

If you wish to connect more than two devices to the USB connector of the 867 pH Module, you can also use an additional commercially available USB hub (distributor). If you operate the 867 pH Module with the help of the Touch Control, then you should use a USB hub with its own power supply.
Connect the USB hub as follows:
Switch off the Touch Control and/or exit the PC software.
1
With the aid of the 6.2151.020 cable, connect the USB connector of
2
the 867 pH Module (type A) with the USB connector of the hub (type B, see manual for the hub).
867 pH Module
■■■■■■■■
17
3.4 Connecting USB devices
USB 1
USB 2
MSB 1
MSB 2
MSB 3
MSB 4
Made by Metrohm Herisau Switzerland
Controller
100 -200 V f = 50 - 60 Hz P = 45 W
Metrohm
Switzerland
Nr.
Typ
Switch on the Touch Control.
3
The USB hub is recognized automatically.

3.4.3 Connecting a printer

Printers that are connected to the 867 pH Module with Touch Control must meet the following requirements:
Printer languages: HP-PCL (PCL 3 to 5, PCL 3GUI), Canon BJL Com-
mands or Epson ESC P/2
Printer resolution: 300 dots/inch or 360 dots/inch (Epson) Paper size: A4 or Letter, single-sheet feed.
Connect the printer as follows:
Switch off the Touch Control.
1
With the aid of the 6.2151.020 cable, connect the USB connector of
2
the 867 pH Module (type A) with the USB connector of the printer (type B, see manual for the printer).
Switch on the printer first, then the Touch Control.
3
Configure the printer in the device manager of the Touch Control
4
(see Touch Control manual).
■■■■■■■■■■■■■■■■■■■■■■
Figure 11 Connecting a printer

3.4.4 Connecting a balance

Operation with a PC software:
Operation with Touch Control:
■■■■■■■■
18
– Connect the balance directly to the serial connector (COM) of
the computer. This is usually 9-pin and marked with the symbol IOIOI.
– You will need the 6.2148.050 USB/RS-232 adapter to connect a
balance.
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
The following table offers an overview of the balances that you can use together with the 867 pH Module and of which cable you will need for connection to the RS-232 interface:
Balance Cable
AND ER, FR, FX with RS-232 inter-
6.2125.020 + 6.2125.010
face (OP-03)
Mettler AB, AG, PR (LC-RS9) In the scope of delivery for the
balance
Mettler AM, PM, PE with interface option 016
or
Mettler AJ, PJ with interface option 018
6.2146.020 + 6.2125.010
Also from Mettler: ME 47473 adapter and either ME 42500 hand switch or ME 46278 foot switch
Mettler AT 6.2146.020 + 6.2125.010
Also from Mettler: ME 42500 hand switch or ME 46278 foot switch
Mettler AX, MX, UMX, PG, AB-S,
6.2134.120
PB-S, XP, XS
Mettler AE with interface option 011 or 012
6.2125.020 + 6.2125.010
Also from Mettler: ME 42500 hand switch or ME 46278 foot switch
867 pH Module
Ohaus Voyager, Explorer, Analyti-
Cable AS017-09 from Ohaus
cal Plus
Precisa balances with RS-232-C
6.2125.080 + 6.2125.010
interface
Sartorius MP8, MC, LA, Genius,
6.2134.060
Cubis
Shimadzu BX, BW 6.2125.080 + 6.2125.010
Operation with Touch Control
Connect the balance as follows:
Plug in the USB plug of the USB/RS-232 adapter at the USB connec-
1
tor of the 867 pH Module.
■■■■■■■■
19
3.4 Connecting USB devices
Connect the RS-232 interface of the USB/RS-232 adapter with the
2
■■■■■■■■■■■■■■■■■■■■■■
RS-232 interface of the balance (see table for cable).
Switch on the Touch Control.
3
Switch on the balance.
4
Activate the RS-232 interface of the balance if necessary.
5
Configure the RS-232 interface of the USB/RS-232 adapter in the
6
device manager of the Touch Control (see Touch Control manual).
3.4.5 Connecting a PC keyboard (only for operation with Touch Con­trol)
The PC keyboard is used as an aid for text and numerical input.
Connect the PC keyboard as follows:
Insert the USB plug of the keyboard into one of the USB sockets of
1
the 867 pH Module.
Switch on the Touch Control.
2
The keyboard is recognized automatically and entered in the device manager.
Configure the keyboard in the device manager of the Touch Control
3
(see Touch Control manual).

3.4.6 Connecting a barcode reader

The barcode reader is used as an aid for text and numerical input. You can connect a barcode reader with USB interface.
Operation with Touch Control
Connect the barcode reader as follows:
Insert the USB plug of the barcode reader into one of the USB sock-
1
ets of the 867 pH Module.
Switch on the Touch Control.
2
■■■■■■■■
20
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
The barcode reader is recognized automatically and entered in the device manager.
Configure the barcode reader in the device manager of the Touch
3
Control (see Touch Control manual).
Settings on the barcode reader:
Program the barcode reader as follows (see also the manual for the bar­code reader):
Switch the barcode reader to programming mode.
1
Specify the desired layout for the keyboard (USA, Germany, France,
2
Spain, German-speaking Switzerland).
This setting must match the setting in the device manager (see the Touch Control manual).
Make sure that the barcode reader is set in such a way that Ctrl char-
3
acters (ASCII 00 to 31) can be sent.
Program the barcode reader in such a way that the ASCII character
4
02 (STX or Ctrl B) is sent as the first character. This first character is normally referred to as the "Preamble" or "Prefix Code".
Program the barcode reader in such a way that the ASCII character
5
04 (EOT or Ctrl D) is sent as the last character. This last character is normally referred to as the "Postamble", "Record Suffix" or "Postfix Code".
Exit the programming mode.
6
867 pH Module
■■■■■■■■
21

3.5 Connecting sensors

■■■■■■■■■■■■■■■■■■■■■■
3.5 Connecting sensors
The measuring interface contains the following measuring inputs:
Ind. for a potentiometric electrode (pH, metal or ion-selective elec-
trode)
Ref. for a separate reference electrode Temp. for a temperature sensor (Pt1000 or NTC) Pol. for a polarizable electrode iConnect for an iConnect (measuring interface for electrodes with
integrated data chip, so-called iTrodes)

3.5.1 Connecting a pH, metal or ion-selective electrode

Connect the pH, metal or ion-selective electrode as follows:
Plug the electrode plug into the Ind. socket of the 867 pH Module.
1
Figure 12 Connecting a pH, metal or ion-selective electrode
NOTE
The electrode cable is protected against accidental disconnection of the cable by means of a pull-out protection. If you wish to pull out the plug again, you will first need to pull back the outer plug sleeve.

3.5.2 Connecting a reference electrode

Connect the reference electrode as follows:
Plug the electrode plug into the Ref. socket of the 867 pH Module.
1
■■■■■■■■
22
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Figure 13 Connecting a reference electrode

3.5.3 Connecting a polarizable electrode

Connect the polarizable electrode as follows:
Plug the electrode plug into the Pol. socket of the 867 pH Module.
1
3 Installation
Figure 14 Connecting a polarizable electrode
NOTE
The electrode cable is protected against accidental disconnection of the cable by means of a pull-out protection. If you wish to pull out the plug again, you will first need to pull back the outer plug sleeve.
867 pH Module
■■■■■■■■
23
3.5 Connecting sensors
■■■■■■■■■■■■■■■■■■■■■■

3.5.4 Connecting a temperature sensor or an electrode with integrated temperature sensor

A temperature sensor of the Pt1000 or NTC type can be connected to the Temp. connector.
Connect the temperature sensor or the electrode with integrated temper­ature sensor as follows:
Insert the plugs of the temperature sensor into the Temp. sockets of
1
the pH Module.
Figure 15 Connecting a temperature sensor or an electrode with
Always insert the red plug into the red socket. This is the only way that shielding against electrical interference can be ensured.

3.5.5 Connecting an iConnect

An external 854 iConnect measuring interface can be connected to the 867 pH Module.
Connect the iConnect as follows:
Plug the iConnect plug into the iConnect socket of the 867 pH
1
Module.
Take care to ensure that the marking on the plug matches the mark­ing on the pH Module as shown in the figure.
integrated temperature sensor
NOTE
■■■■■■■■
24
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
Figure 16 Connecting the iConnect
The iConnect is detected automatically and entered as measuring input into the device properties of the pH Module. If an electrode is connected to the iConnect that is not yet included in the list of sen­sors for the control software, then a corresponding message will be displayed.
The iConnect can be plugged in and unplugged while the pH Module is switched on.
An electrode with integrated data chip, referred to as iTrode, is connected to the iConnect.
Connect the electrode as follows:
Remove the protective caps on the iConnect and the electrode.
1
Plug in the electrode on the iConnect as illustrated below.
2
Figure 17 Connecting an electrode to the iConnect
867 pH Module
The guide pin guarantees correct connection in such a way that the contact pins cannot be damaged.
Screw the iConnect tightly.
3
■■■■■■■■
25

4.1 General notes

4 Operation and maintenance

4.1 General notes

4.1.1 Care

The 867 pH Module requires appropriate care. Excess contamination of the instrument may result in functional disruptions and a reduction in the lifetime of the otherwise sturdy mechanics and electronics.
Spilled chemicals and solvents should be removed immediately. Above all, the plug connections on the rear of the instrument (in particular the power socket) should be protected from contamination.
CAUTION
Although this is largely prevented by design measures, the power plug should be unplugged immediately if aggressive media have found their way into the interior of the instrument to prevent serious damage to the instrument electronics. In such cases, Metrohm Service must be informed.
■■■■■■■■■■■■■■■■■■■■■■

4.1.2 Maintenance by Metrohm Service

Maintenance of the 867 pH Module is best carried out as part of annual service, which is performed by specialist personnel from Metrohm. A shorter maintenance interval may be necessary if you frequently work with caustic and corrosive chemicals.
Metrohm Service provides professional technical consultation at all times for the maintenance and servicing of all Metrohm instruments.
■■■■■■■■
26
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
4 Operation and maintenance

4.2 Quality Management and qualification with Metrohm

Quality management
Metrohm offers you comprehensive support in implementing quality man­agement measures for instruments and software. Further information on this can be found in the brochure "Metrohm Quality Management" available from your local Metrohm representative.
Qualification
Please contact your local Metrohm representative for support in qualifica­tion of instruments and software. The Installation Qualification (IQ) and Operational Qualification (OQ) are offered by Metrohm represen­tatives as a service. They are carried out by trained employees using stan­dardized qualification documents and in accordance with the currently applicable requirements of the regulated industry. Further information on this can be found in the brochure "Analytical Instrument Qualifica- tion – Confidence in quality with IQ/OQ".
Maintenance
The electronic and mechanical functional groups of Metrohm instruments can and should be checked by specialist personnel from Metrohm as part of a regular preventive maintenance schedule. Please ask your local Metrohm representative regarding the precise terms and conditions involved in concluding a corresponding maintenance agreement. Further information on this can be found in the brochure "Metrohm Care Con- tracts – Protect your investment the smart way" available from your local Metrohm representative.
867 pH Module
■■■■■■■■
27

5.1 General

5 Troubleshooting

5.1 General
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
The measured value is evidently wrong.
No or heavily fluctu­ating measuring sig­nal.
The calibration is wrong. Check the calibration and repeat, if neces-
sary.
Check the buffer or standard solutions and
repeat the calibration, if necessary.
The manual temperature input is wrong.
The pH or ISE membrane or the diaphragm is contami­nated.
The electrolyte or the mea­suring chain is overaged.
The measuring chain is defective.
The measuring chain is not connected.
Air is trapped in or in front of the diaphragm.
The measuring chain is defective.
Enter the correct measuring temperature.
Clean the membrane or the diaphragm.
Replace the electrolyte or the measuring chain.
Replace the measuring chain.
Connect the measuring chain.
Remove the air.
Replace the measuring chain.
The measured value remains unstable and does not fulfil the drift criterion.
■■■■■■■■
28
The pH or ISE membrane or the diaphragm is contami­nated.
The pH value or the tem­perature of the measuring solution is not stable.
Unsuitable measuring chain. The conductivity is too low or an organic solu­tion is present.
The measuring chain is not connected.
Clean the membrane or the diaphragm.
Measure under exclusion of air. Control the temperature of the measuring
solution.
Use a suitable measuring chain.
Connect the measuring chain.
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Problem Cause Remedy
5 Troubleshooting
Sluggish measured value setting.
Insufficient slope during calibration.
The Delta U poten­tial step set is not achieved with 'auto' standard addition.
The measuring chain is
Replace the measuring chain.
defective.
The pH or ISE membrane or
Clean the membrane or the diaphragm.
the diaphragm is contami­nated.
The pH or ISE membrane or
Clean the membrane or the diaphragm.
the diaphragm is contami­nated.
No hydrated layer is pres­ent on the glass membrane after measurements in
Between measurements, place separate elec­trodes in distilled water and combined elec­trodes in reference electrolyte.
water-free solutions.
The buffer or standard sol-
Use fresh buffer or standard solutions.
utions are old.
The electrode is worn out. Replace the electrode.
The streaming potential of the indicator electrode or reference electrode results in a potential difference between stirred addition
Reduce the dosing rate. Increase the stirring rate. Increase Delta U. Touch Control:
Select Addition = auto dos.
and non-stirred measure­ment.
867 pH Module
■■■■■■■■
29

6.1 Remote interface

1
2
3
13
1
14
25
1
13
14
25

6 Appendix

6.1 Remote interface
The 6.2148.010 Remote Box allows devices to be controlled which cannot be connected directly to the MSB interface of the pH Module.
■■■■■■■■■■■■■■■■■■■■■■
Figure 18 Connectors of the Remote Box
Cable
1
For connecting to the pH Module.
Remote connector
3
For connecting instruments with a remote interface.
MSB connector
2
Metrohm Serial Bus. For connecting external dosing devices or stirrers.

6.1.1 Pin assignment of the remote interface

Figure 19
The above figure of the pin assignment applies for all Metrohm instru­ments with 25-pin D-Sub remote connector.
Pin assignment of remote socket and remote plug
■■■■■■■■
30
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
+5 V
t
p
t
p
6 Appendix
Inputs
approx. 50 kΩ Pull-up
tp > 20 ms
active = low, inactive = high
Outputs
Open Collector
tp > 200 ms
active = low, inactive = high
IC = 20 mA, V
CEO
= 40 V
+5 V: maximum load = 20 mA
The following tables offer information concerning the assignment of the individual pins and their function:
Table 1
Assignment Pin No. Function
Inputs and outputs of the remote interface
*
Input 0 21 Start
Input 1 9 Stop
Input 2 22
Input 3 10 Quit
Input 4 23
Input 5 11
Input 6 24
Input 7 12
Output 0 5 Ready
Output 1 18 Conditioning OK
Output 2 4 Determination
Output 3 17 EOD
Output 4 3
Output 5 16 Error
Output 6 1
Output 7 2 Warning
867 pH Module
■■■■■■■■
31
6.1 Remote interface
■■■■■■■■■■■■■■■■■■■■■■
Assignment Pin No. Function
*
Output 8 6
Output 9 7
Output 10 8
Output 11 13
Output 12 19
Output 13 20
0 volts / GND 14
+5 volts 15
0 volts / GND 25
* Signal activated only for operation with Touch Control.
Table 2
Explanation of the individual functions
Function Explanation
Start The current method is started at the time of acti-
vation.
t
> 100 ms
pulse
Stop The current method is canceled (Stop) at the time
of activation.
t
> 100 ms
pulse
Quit The current command in the determination run
will be canceled at the time of activation.
t
> 100 ms
pulse
Ready The instrument is ready to receive a start signal.
Conditioning OK The line is set when Conditioning with SET titra-
tion and KFT titration is at OK. The line remains set until the determination is started with
[START].
Determination The instrument performs a data-generating
determination.
EOD End of Determination.
Pulse (t
= 200 ms) after a determination or
pulse
after a buffer/standard solution during calibration using a Sample Processor.
■■■■■■■■
32
Error The line is set for error message display.
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
6 Appendix
Function Explanation
Warning The line is set for warning message display.
867 pH Module
■■■■■■■■
33

7.1 Measuring interface

7 Technical specifications

7.1 Measuring interface
The 867 pH Module has one galvanically isolated measuring interface.
The measuring cycle is 100 ms for all measuring modes.

7.1.1 Potentiometry

One high-ohm measuring input (Ind.) for pH, metal or ion-selective elec­trodes and one measuring input (Ref.) for separate reference electrodes.
■■■■■■■■■■■■■■■■■■■■■■
Input resistance
> 1 · 1012 Ω
Offset current < 1 · 10
Measuring mode pH
Measuring
–13 - +20 pH
range
Resolution 0.001 pH
Measuring accuracy
±0.003 pH (±1 digit, without sensor error, under reference conditions)
Measuring mode U
Measuring
–1,200 - +1,200 mV
range
Resolution 0.1 mV
Measuring accuracy
±0.2 mV (±1 digit, without sensor error, under reference conditions)
Measuring mode Conc
Measuring
10
–21
- 10
range
Resolution
Measuring
(±1 digit, without sensor error, under reference conditions)
accuracy
–12
A (under reference conditions)
20
■■■■■■■■
34
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■

7.1.2 Temperature

A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC type with automatic temperature compensation.
R (25 °C) and B value can be configured for NTC sensors.
Measuring range
Pt1000 –150 - +250 °C
NTC –5 - +250 °C
(R (25 °C) = 30,000 Ω and B (25/50) = 4,100 K)
Resolution
Pt1000 0.1 °C
NTC 0.1 °C
Measuring accu­racy
Pt1000 ±0.2 °C
(Applies for measuring range –20 - +150 °C; ±1 digit; without sensor error, under reference conditions)
NTC ±0.6 °C
(Applies for measuring range +10 - +40 °C; ±1 digit; without sensor error, under reference conditions)
7 Technical specifications

7.1.3 Polarizer

One measuring input (Pol.) for polarizable electrodes.
Measuring mode Ipol
Polarization current
Measuring range
Resolution 0.1 mV
Measuring accuracy
Measuring mode Upol
Polarization voltage
Measuring range
Determination with adjustable polarization current
–122.5 - +122.5 µA (increment: 0.5 µA) –125.0 - +125.0 µA: non-guaranteed values, dependent on reference voltage +2.5 mV
–1,200 - +1,200 mV
±0.2 µA (±1 digit; without sensor error, under reference conditions)
Determination with adjustable polarization voltage
–1,225 - +1,225 mV (increment: 25 mV) –1,250 - +1,250 mV: non-guaranteed values, dependent on reference voltage +2.5 mV
–120 - +120 µA
867 pH Module
■■■■■■■■
35

7.2 Power connection

Resolution 0.01 µA
Measuring
accuracy
7.2 Power connection
Supply voltage 100 - 240 V
Frequency 50 - 60 Hz
■■■■■■■■■■■■■■■■■■■■■■
Power consump-
45 W
tion
Fuse 2 x 1.6 ATH

7.3 Safety specifications

Design and testing
Safety instructions The documentation contains safety instructions which have to be fol-
EN/IEC 61010-1 UL 61010-1 CSA-C22.2 No. 61010-1 Protection class I
lowed by the user in order to ensure safe operation of the instrument.

7.4 Electromagnetic compatibility (EMC)

Emission
Standards fulfilled:
EN/IEC 61326-1 EN/IEC 61000-6-3 EN/IEC 61000-6-4 EN 55022 / CISPR 22
Immunity Standards fulfilled:
EN/IEC 61326-1 EN/IEC 61000-6-2 EN/IEC 61000-4-2 EN/IEC 61000-4-3 EN/IEC 61000-4-4 EN/IEC 61000-4-5 EN/IEC 61000-4-6 EN/IEC 61000-4-11 EN/IEC 61000-4-14 NAMUR
■■■■■■■■
36
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■

7.5 Ambient temperature

7 Technical specifications
Nominal function range
Storage –20 - +60 °C
Transport –40 - +60 °C
+5 - +45 °C (at a maximum of 85% humidity)

7.6 Reference conditions

Ambient tempera­ture
Relative humidity 60%
Instrument status Instrument in operation at least 30 min
Validity of the data
+25 °C (± 3 °C)
After adjustment

7.7 Dimensions

Width
Height 108 mm
142 mm
Depth 230 mm
Weight 2,660 g
Material Steel sheet

7.8 Interfaces

USB connectors
USB ports
Controller port
Touch Control With integrated Touch Control cable.
Computer With 6.2151.000 cable.
2 USB downstream ports (type A sockets), 500 mA each, for connect­ing peripheral devices such as printers, keyboards, barcode readers or RS-232/USB boxes (Metrohm order no. 6.2148.020).
Connector "Controller"
USB upstream port with auxiliary power supply (Mini DIN socket) for connecting Touch Control or computer for controlling the 867 pH Module.
867 pH Module
■■■■■■■■
37
7.8 Interfaces
■■■■■■■■■■■■■■■■■■■■■■
MSB connectors (Metrohm Serial Bus)
Dosing device Connector for a maximum of four external dosing devices, models Dos-
imat or Dosino (MSB 1 to MSB 4).
Stirrer Connector for a maximum of four stirrers.
Stirrer control: switching on/off manually or coordinated with the titra­tion sequence. Speed in 15 steps and shift direction can be selected.
Remote Box Connector for a maximum of four Remote Boxes. Remote Boxes can
be used to actuate and monitor external devices.
■■■■■■■■
38
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
8 Warranty (guarantee)
Metrohm guarantees that the deliveries and services it provides are free of defects in materials, design or manufacturing.
The general warranty period is 36 months (exclusions below) from the date of delivery or 18 months in the event of continuous operation. The warranty remains valid on the condition that the servicing is provided by a service organization authorized by Metrohm at defined intervals and with a defined scope.
The warranty period for anion suppressors of the type "MSM" is 120 months from the date of delivery or 60 months in the case of continuous operation.
The warranty period for IC separation columns is 90 days after start-up.
For third-party components that are recognizable as such, the manufac­turer's warranty regulations apply.

8 Warranty (guarantee)

Consumables and materials with limited storage life and glass breakage in the case of electrodes or other glass parts are excluded from the warranty.
Warranty claims cannot be asserted if the ordering party has failed to meet its payment obligations according to schedule.
During the warranty period, Metrohm undertakes either to replace free of charge or to credit the purchaser for any modules or components that can be shown to be faulty. Any transport or customs fees that may apply are the ordering party’s responsibility.
The precondition for this is that the ordering party has to specify the arti­cle number, the article designation, an adequate error description, the delivery date and (if applicable) the serial number or chip data in the Sup­port Tracker. Metrohm then decides whether a replacement or a credit note is to be issued or whether the faulty part has to be returned using the Return Material Authorization (RMA). If a replacement or credit note is issued, the ordering party undertakes to store the faulty part for at least 24 months in accordance with the current storage directives (in compli­ance with ESD guidelines) and to hold it in readiness for onsite inspection or for return shipment to Metrohm. Metrohm reserves the right to invoice the ordering party for these articles, including retroactively, in the event of noncompliance with these preconditions.
867 pH Module
The same warranty periods that are specified for a corresponding new part apply to parts that are replaced or repaired within the above-men­tioned warranty periods. However, replacement or repair of a part does not extend the warranty period of the entire system.
■■■■■■■■
39
■■■■■■■■■■■■■■■■■■■■■■
Deficiencies arising from circumstances that are not the responsibility of Metrohm, such as improper storage or improper use, etc., are expressly excluded from the warranty.
Metrohm also offers a 120 month spare parts availability guarantee and a 60 month PC software support warranty, calculated from the date on which the product is withdrawn from the market. The content of this war­ranty is the ability of the customer to obtain functioning spare parts or appropriate software support at market prices during the time of the war­ranty period.
If Metrohm AG is unable to meet this obligation due to circumstances beyond the control of Metrohm AG, then the ordering party shall be offered alternative solutions at preferential conditions.
■■■■■■■■
40
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
9 Accessories
NOTE
Subject to change without notice.

9.1 Scope of delivery

9.1.1 2.867.0010

9 Accessories

Qty.
Order no. Description
1 1.867.0010 pH Module
Module for pH/ion measurement as supplement to a Titrando or "stand-alone" in combination with a 900 Touch Control.
1 6.2103.130 Adapter red 2 mm plug / 4 mm socket
For connecting plug B (4 mm) to 2 mm socket.
1 6.2103.140 Adapter black 2 mm plug / B socket 4 mm
For connecting plug B (4 mm) to 2 mm socket.
867 pH Module
■■■■■■■■
41
9.1 Scope of delivery
Qty. Order no. Description
1 6.2122.0x0 Power supply cable with IEC 60320 line
socket, type C13
Cable plug according to customer requirements.
Switzerland: SEV 1011, Type 12
6.2122.020
Germany, …: CEE 7, Type VII
6.2122.040
USA, …: NEMA 5-15, Type 498
6.2122.070
Length: 1.5 m
1 8.867.8003EN 867 pH Module Manual

9.1.2 2.867.0110

■■■■■■■■■■■■■■■■■■■■■■
Qty.
Order no. Description
1 1.854.0010 iConnect
1 1.867.0010 pH Module
1 1.900.0010 Touch Control
1 6.0278.300 iUnitrode with Pt 1000
Combined pH glass electrode for measurements in difficult media, with memory chip for sensor data such as article and serial number, calibration data and calibration history.
Shaft material: Glass Measuring range: 0 - 14 Measuring unit: pH Temperature range, long-
term (°C): Temperature range, short­term (°C): Shaft diameter top (mm): 12
Shaft diameter bottom (mm): 12 Min. immersion depth (mm): 30 Temperature sensor: Pt1000 Indicator electrode shape: Cylinder Membrane glass type: U Membrane resistance: 150 - 500 Electrode zero point (mV): 0 - +/-15 Electrode slope: > 0.97
0 - 80
0 - 100
■■■■■■■■
42
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Qty. Order no. Description
Internal reference electrode type: Diaphragm: Fixed ground-joint diaphragm
Reference electrolyte type: c(KCl) = 3 mol/L Bridge electrolyte type: c(KCl) = 3 mol/L Electrolyte flow rate (μl/h): 5 - 30
1 6.2001.060 Support stand without support rod
To attach an 801, 804, 803 Stirrer to a Titrando, Titrino plus, Dosi­mat plus or 856 and 867.
1 6.2013.010 Clamping ring
LL system
9 Accessories
For support rods with a diameter of 10 mm.
Material: Metal Width (mm): 20 Height (mm): 16
1 6.2016.070 Support rod / 400 mm
Material: Stainless steel 18/9 Outer diameter (mm): 10 Length (mm): 400
867 pH Module
■■■■■■■■
43
9.1 Scope of delivery
Qty. Order no. Description
1 6.2021.020 Electrode holder
Electrode holder for 4 electrodes and 2 buret tips
Material: Plastic
1 6.2103.130 Adapter red 2 mm plug / 4 mm socket
For connecting plug B (4 mm) to 2 mm socket.
■■■■■■■■■■■■■■■■■■■■■■
1 6.2103.140 Adapter black 2 mm plug / B socket 4 mm
For connecting plug B (4 mm) to 2 mm socket.
1 6.2307.230 Buffer solutions pH 4, 7 and 9
Mixed buffer solutions pH 4.00/7.00/9.00 (25 °C) in single use sachets, colourless, box of 3 x 10 x 30 mL
1 6.2325.000 pHit kit
Maintenance kit for pH electrodes The kit contains: 50 mL cleaning solution50 mL 3M KCl solution50 mL storage solution2 Storage ves­selsInstructions for use
■■■■■■■■
44
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Qty. Order no. Description
1 6.2621.070 Hexagon key 5 mm
5 mm.
Length (mm): 80
1 6.2621.130 Hexagon key 2 mm
2 mm.
9 Accessories
1 6.6064.010 USB Memory Stick for 900 Touch Control
1 6.2122.0x0 Power supply cable with IEC 60320 line
socket, type C13
Cable plug according to customer requirements.
Switzerland: SEV 1011, Type 12
6.2122.020
Germany, …: CEE 7, Type VII
6.2122.040
USA, …: NEMA 5-15, Type 498
6.2122.070
Length: 1.5 m
867 pH Module
■■■■■■■■
45
9.1 Scope of delivery
Qty. Order no. Description
1 8.867.8003EN 867 pH Module Manual

9.1.3 2.867.0210

Qty. Order no. Description
1 1.854.0010 iConnect
1 1.867.0010 pH Module
1 6.0278.300 iUnitrode with Pt 1000
Combined pH glass electrode for measurements in difficult media, with memory chip for sensor data such as article and serial number, calibration data and calibration history.
Shaft material: Glass Measuring range: 0 - 14 Measuring unit: pH Temperature range, long-
term (°C): Temperature range, short­term (°C): Shaft diameter top (mm): 12
Shaft diameter bottom (mm): 12 Min. immersion depth (mm): 30 Temperature sensor: Pt1000 Indicator electrode shape: Cylinder Membrane glass type: U Membrane resistance: 150 - 500 Electrode zero point (mV): 0 - +/-15 Electrode slope: > 0.97 Internal reference electrode
type: Diaphragm: Fixed ground-joint diaphragm
Reference electrolyte type: c(KCl) = 3 mol/L Bridge electrolyte type: c(KCl) = 3 mol/L Electrolyte flow rate (μl/h): 5 - 30
■■■■■■■■■■■■■■■■■■■■■■
0 - 80
0 - 100
LL system
■■■■■■■■
46
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Qty. Order no. Description
1 6.2001.060 Support stand without support rod
To attach an 801, 804, 803 Stirrer to a Titrando, Titrino plus, Dosi­mat plus or 856 and 867.
1 6.2013.010 Clamping ring
For support rods with a diameter of 10 mm.
Material: Metal Width (mm): 20 Height (mm): 16
9 Accessories
1 6.2016.070 Support rod / 400 mm
Material: Stainless steel 18/9 Outer diameter (mm): 10 Length (mm): 400
1 6.2021.020 Electrode holder
Electrode holder for 4 electrodes and 2 buret tips
Material: Plastic
867 pH Module
■■■■■■■■
47
9.1 Scope of delivery
Qty. Order no. Description
1 6.2103.130 Adapter red 2 mm plug / 4 mm socket
For connecting plug B (4 mm) to 2 mm socket.
1 6.2103.140 Adapter black 2 mm plug / B socket 4 mm
For connecting plug B (4 mm) to 2 mm socket.
■■■■■■■■■■■■■■■■■■■■■■
1 6.2151.000 Cable USB A – mini-DIN 8-pin
Controller cable
Length (m): 1.8
1 6.2307.230 Buffer solutions pH 4, 7 and 9
Mixed buffer solutions pH 4.00/7.00/9.00 (25 °C) in single use sachets, colourless, box of 3 x 10 x 30 mL
1 6.2325.000 pHit kit
Maintenance kit for pH electrodes The kit contains: 50 mL cleaning solution50 mL 3M KCl solution50 mL storage solution2 Storage ves­selsInstructions for use
■■■■■■■■
48
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Qty. Order no. Description
1 6.2621.070 Hexagon key 5 mm
5 mm.
Length (mm): 80
1 6.2621.130 Hexagon key 2 mm
2 mm.
9 Accessories
1 6.6056.231 tiamo ™ 2.3 Light CD: 1 license
tiamo ™ 2.3 light computer program for controlling a titration sys­tem. Up to two instruments can be connected Graphical method edi­tor with numerous templates Layout Manager for customizing the user interface Professional database with reevaluation High-perform­ance report generator No parallel titration, no data export 1 license Dialog languages: German, English
1 6.2122.0x0 Power supply cable with IEC 60320 line
socket, type C13
Cable plug according to customer requirements.
Switzerland: SEV 1011, Type 12
6.2122.020
Germany, …: CEE 7, Type VII
6.2122.040
USA, …: NEMA 5-15, Type 498
6.2122.070
Length: 1.5 m
1 8.867.8003EN 867 pH Module Manual
867 pH Module
■■■■■■■■
49

9.2 Optional accessories

9.2 Optional accessories
Order no. Description
2.801.0010 801 Stirrer
Magnetic stirrer without stand for supplementing the Titrino plus, Dosimat plus, Titrandos, Sample Processors, 805 Dosimats and 780/781 pH Meters. With per­manently attached cable for MSB (Metrohm Serial Bus).
2.802.0040 802 Stirrer (propeller stirrer) for 804 Ti Stand
Rod stirrer with 6.1909.010 Propeller stirrer.
■■■■■■■■■■■■■■■■■■■■■■
■■■■■■■■
50
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Order no. Description
2.804.0010 804 Ti Stand without stand rod
Titration stand and controller for 802 Stirrer. The 804 Ti Stand together with the optional 802 Stirrer provides an alternative to the magnetic stirrer. Without stand.
2.814.0010 814 USB Sample Processor (1T/1P)
USB Sample Processor with one work station and one membrane pump for automatic processing of small to medium numbers of routine samples. The device can be controlled in «stand alone» operation using the Touch Control unit or by means of a PC with either the titration software tiamo™ or the chro­matography software MagIC Net™.
9 Accessories
867 pH Module
■■■■■■■■
51
9.2 Optional accessories
Order no. Description
2.814.0020 814 USB Sample Processor (1T/2P)
USB Sample Processor with one work station and two membrane pumps for automatic processing of small to medium numbers of routine samples. The device can be controlled in «stand alone» operation using the Touch Control unit or by means of a PC with either the titration software tiamo™ or the chro­matography software MagIC Net™.
2.814.0030 814 USB Sample Processor (1T/0P)
USB Sample Processor with one work station and two pump connections for automatic processing of small to medium numbers of routine samples. The device can be controlled in «stand alone» operation using the Touch Control unit or by means of a PC with either the titration software tiamo™ or the chro­matography software MagIC Net™.
■■■■■■■■■■■■■■■■■■■■■■
6.2065.000 Stacking frame for 846 Dosing Interface, 856 Conduc­tivity Module,867 pH Module
For fixing the Reagent Organizer on top of the Dosing Interface
■■■■■■■■
52
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Order no. Description
6.2151.000 Cable USB A – mini-DIN 8-pin
Controller cable
Length (m): 1.8
9 Accessories
867 pH Module
■■■■■■■■
53

Index

Index
■■■■■■■■■■■■■■■■■■■■■■
Numbers/Symbols
6.2151.000 controller cable ..... 11
685 Dosimat ............................ 14
700 Dosino .............................. 14
800 Dosino .............................. 14
801 Stirrer ................................ 15
803 Ti Stand ............................. 15
804 Ti Stand ............................. 15
805 Dosimat ............................ 14
A
ADD ........................................... 3
B
Balance .................................... 18
Barcode reader
Connect ............................. 20
C
Calibration mode
CAL ...................................... 3
Computer
Connect ............................. 11
Connect
Balance .............................. 18
Barcode reader ................... 20
Computer ........................... 11
Dosing device ..................... 14
iConnect ............................ 24
MSB devices ....................... 13
PC keyboard ....................... 20
Printer ................................ 18
Remote Box ........................ 16
Stirrer ................................. 15
Titration stand .................... 15
Touch Control .................... 10
USB hub ............................. 17
Connector
MSB ..................................... 2
USB ...................................... 2
Controller
Connector ............................ 8
D
Dosing command
ADD ..................................... 3
EMPTY ................................. 3
LQH ..................................... 3
PREP ..................................... 3
Dosing device
Connect ............................. 14
Driver software
Install ................................. 11
E
Electrode
Connect ............................. 22
Electrode test
ELT ....................................... 3
Electrostatic charge .................... 5
EMPTY ....................................... 3
G
GLP .......................................... 27
Guarantee ................................ 39
I
iConnect
Connect ............................. 24
Installation
Driver software ................... 11
Instrument description ................ 2
Instrument type .......................... 8
K
Keyboard
Connect ............................. 20
L
LQH ........................................... 3
M
Mains voltage ............................. 5
Maintenance ............................ 26
Maintenance Agreement .......... 27
MEAS ......................................... 2
Measured value ........................ 28
Unstable ............................. 28
Measured value setting ............. 29
Measuring interface .................... 2
Measuring mode
MEAS ................................... 2
STDADD ............................... 3
Measuring signal
Fluctuating ......................... 28
Metrohm Serial Bus MSB, see also
"MSB" ...................................... 13
MSB
Connect devices ................. 13
Connector
MSB connector ........................... 2
O
Overview of the instrument ........ 7
P
PC keyboard
Connect ............................. 20
Pin assignment ......................... 30
Power connection ...................... 8
PREP ........................................... 3
Printer ...................................... 18
Q
Quality Management ................ 27
R
Remote
Interface ............................. 30
Pin assignment ................... 30
Remote Box
Connect ............................. 16
S
Safety instructions ...................... 4
Sensor
Connect ............................. 22
Serial number ............................. 8
Service ....................................... 5
STDADD ..................................... 3
Stirrer
Connect ............................. 15
System test ............................... 10
T
Temperature sensor
Connect ............................. 22
Titration stand
Connect ............................. 15
Touch Control
Connect ............................. 10
U
USB
Connector ............................ 8
USB connector ........................... 2
USB hub
Connect ............................. 17
............................ 8
■■■■■■■■
54
867 pH Module
■■■■■■■■■■■■■■■■■■■■■■
Index
V
Validation ................................. 27
W
Warranty .................................. 39
867 pH Module
■■■■■■■■
55
Loading...