Metrohm 794 User Manual

CH-9101 Herisau/Switzerland E-Mail info@metrohm.com Internet www.metrohm.com
794 Basic Titrino
Program version 5.794.0010
Instructions for Use
Table of contens

Table of contents

1 Introduction...........................................................................................................1
1.1 Instrument description........................................................................................... 1
1.2 Controls and parts................................................................................................. 2
2
Manual operation..................................................................................................6
2.1 Keypad ..................................................................................................................6
2.2 Principle of data input............................................................................................ 7
2.3 Text input ...............................................................................................................8
2.4 Tutorial ................................................................................................................... 9
2.4.1 Entering data, setting the dialog language....................................................9
2.4.2 Development of a method, titration of an acid.............................................11
2.4.3 Storage and loading of methods ................................................................. 16
2.4.4 pH calibration.............................................................................................. 17
2.4.5 Statistics, acid capacity of drinking water.................................................... 18
2.5 Configuration, key <CONFIG>...........................................................................22
2.6 Selection of the mode, key <MODE> ................................................................ 24
2.7 Parameters, key <PARAM> ............................................................................... 26
2.7.1 Parameters for DET and MET ...................................................................... 26
2.7.2 Parameters for SET ......................................................................................36
2.7.3 Parameters for MEAS................................................................................... 45
2.7.4 Parameters for CAL......................................................................................47
2.7.5 Parameters for TIP .......................................................................................50
2.8 Result calculations............................................................................................... 52
2.9 Statistics calculation ............................................................................................55
2.10 Common variables ..............................................................................................57
2.11 Data output.......................................................................................................... 58
2.12 TIP, Titration procedure ....................................................................................... 62
2.13 Method memory, keys <USER METH> ............................................................. 66
2.14 Calibration data, key <CAL.DATA> ...................................................................68
2.15 Current sample data, key <SMPL DATA> .........................................................69
2.16 Silo memory for sample data ..............................................................................70
2.17 Storing determination results and silo calculations............................................. 73
2.17.1 Storing determination results .......................................................................73
2.17.2 Silo calculations ........................................................................................... 74
3
Operation via RS232 Interface...........................................................................77
3.1 General rules .......................................................................................................77
3.1.1 Call up of objects .........................................................................................78
3.1.2 Triggers ........................................................................................................79
3.1.3 Status messages .........................................................................................80
3.1.4 Error messages............................................................................................82
3.2 Remote control commands................................................................................. 85
3.2.1 Overview.......................................................................................................85
3.2.2 Description of the remote control commands ........................................... 102
3.3 Properties of the RS 232 Interface.....................................................................125
3.3.1 Handshake................................................................................................. 125
3.3.2 Pin Assignment .......................................................................................... 129
3.3.3 What can you do if the data transfer does not work? ................................ 131
794 Basic Titrino
Table of contents
4
Error messages, Troubleshooting...................................................................133
4.1 Error and special messages ............................................................................. 133
4.2 Diagnosis .......................................................................................................... 137
4.2.1 General ...................................................................................................... 137
4.2.2 Procedure .................................................................................................. 137
4.2.3 Equipment required: .................................................................................. 138
4.2.4 Diagnosis steps ......................................................................................... 138
4.3 Initialise and test RAM....................................................................................... 147
4.4 Releasing a locked spindle ............................................................................... 148
5
Preparations .....................................................................................................149
5.1 Setting up and connecting the instruments ...................................................... 149
5.1.1 Titrino with Stirrer or Titration Stand........................................................... 149
5.1.2 Connection of a printer .............................................................................. 150
5.1.3 Connection of a balance............................................................................ 151
5.1.4 Connection of a Sample Changer ............................................................. 152
5.1.5 Connection of a recorder........................................................................... 153
5.1.6 Connection of a computer......................................................................... 154
5.2 Connection of electrodes, preparing titration vessel ........................................ 155
6
Appendix ...........................................................................................................158
6.1 Technical specifications.................................................................................... 158
6.2 Pin assignment of the "Remote" socket ............................................................ 161
6.2.1 Lines of the "Remote" socket during the titration....................................... 163
6.2.2 Possible configurations of the activate pulse in SET and CAL.................. 164
6.3 User methods.................................................................................................... 165
6.3.1 General ...................................................................................................... 165
6.3.2 "Titer_pH" ................................................................................................... 166
6.3.3 "Blank"........................................................................................................ 167
6.3.4 "Chloride" ................................................................................................... 168
6.3.5 "TAN-TBN".................................................................................................. 169
6.3.6 "Diazo"........................................................................................................ 170
6.3.7 "Br-Index" ................................................................................................... 171
6.3.8 "Sapon.No"................................................................................................. 172
6.3.9 "Ca-Mg"...................................................................................................... 173
6.3.10 "EDTA-NTA"................................................................................................ 174
6.3.11 "Metals" ...................................................................................................... 175
6.3.12 "Perox.No" .................................................................................................. 176
6.3.13 "FormolNo""................................................................................................ 177
6.3.14 "P2O5Fert""................................................................................................. 180
6.4 Validation / GLP................................................................................................. 183
6.5 Warranty and conformity ................................................................................... 184
6.5.1 Warranty..................................................................................................... 184
6.5.2 EU Declaration of conformity..................................................................... 185
6.5.3 Certificate of Conformity and System Validation ....................................... 186
6.6 Scope of delivery and ordering designations ................................................... 187
Index.........................................................................................................................190
794 Basic Titrino
1.1 Instrument description

1 Introduction

1.1 Instrument description

The 794 Basic Titrino is an all-purpose titrator. Titration modes of the 794 Basic Titrino are constant or – depending on the titration curve – variable dosing of the titration reagent and Endpoint-Titration. All operating modes of the Titrino can be combined to perform extensive analytical se­quences. Ready-to-start methods for the most common applications are stored in the internal method memory. The operator is free to modify and overwrite this methods or to create and store his own titration sequences. Data exchange with a PC is possible with the Metrodata VESUV Software and with Metro­data TiNet Software complete remote control, data acquisition and evaluation via PC is enabled.
794 Basic Titrino
1
1.2 Controls and parts

1.2 Controls and parts

Front view of instrument:
794 Basic Titrino
2
1.2 Controls and parts
1 Exchange Unit
2 Display
3 Setting of display contrast
4 Controls the dosing rate during manual dosing with <DOS> and subse-
quent filling
5 Control keys and indicator lamps on the Titrino
Key <DOS> Dosing key. Dispensing is performed as long as <DOS> is being
pressed. Used e.g. to prepare the Exchange Unit. The dispensing rate can be set with potentiometer (4).
Key <STOP/FILL> - Stops procedures, e.g. titrations, conditioning.
- Filling after manual dosing with <DOS>.
Key <START> Starts procedures, e.g. titrations, conditioning.
Identical with key <START> of the separate keypad. Indicator lamps: "Statistics" Lamp is on when the "statistics" function (calculation of mean and
standard deviation) is on. "Silo" Lamp is on when silo memory (for sample data) is on.
794 Basic Titrino
3
1.2 Controls and parts
Rear view of instrument:
794 Basic Titrino
4
1.2 Controls and parts
6 RS232 interface
for the connection of printer, balance or a computer
7 Remote lines
(input/output) for the connection of the Remote Box, Sample Changers, robots etc.
8 Connection of electrodes and temperature sensor
2 high-impedance measuring inputs for pH and U measurements (Ind I/ Ind II).
They can either be used separately or for differential potentiometry, see page
155. Important: If you work with both measuring inputs in the same vessel, the same reference electrode must be used.
1 measuring input for polarized electrodes (Pol).
1 measuring input for PT100 or Pt1000 temperature sensor.
9 Connection for separate keypad
10 Analog output for the connection of a recorder
11 Connection for power cable
With power supplies where the voltage is subject to severe HF disturbances, the Ti­trino should be operated via an additional power filter, e.g. Metrohm 615 model.
12 Mains switch
13 Earthing socket
14 Connection for stirrer
728 Magnetic Stirrer, 802 Rod Stirrer, 703 or 727 Ti Stand Supply voltage: 9 VDC (I 200 mA)
15 Display of the set mains voltage
Before switching on for the first time, check that the set mains voltage matches the voltage of your power supply. If this is not the case, disconnect mains cable and change voltage.
16 Rating plate
with fabrication, series and instrument number
794 Basic Titrino
5
2.1. Keypad

2 Manual operation

2.1 Keypad

CONFIG Configuration. PARAM Parameters. SMPL DATA Sample data. STATISTICS ON/OFF switching of statistics calcu-
lations of consecutive determination, see page 55.
MEAS/HOLD ON/OFF switching of measurements
between titrations and hold during titrations.
SILO ON/OFF switching of silo memory for
sample data, see page 69. CAL.DATA Calibration data, see page 68. C-FMLA Calculation values, see page 54. DEF Formulas, data output, sequence for
TIP, see page 52ff. USER METH Management of internal method
memory, see page 66. PRINT Printing of reports, see page 60. REPORTS Result output. MODE Mode selection, see page 24. <,> Keys for text input. SELECT Selection of special values (dialog
marked with ":") CLEAR Clears values, set special values. ENTER Stores values. STOP Stops methods. QUIT Quits inquiries, waiting times, printing. START Starts methods.
The third functions (inscriptions in the triangle) on the keys of the keypad are used for formula entry, see page 52.
794 Basic Titrino
6
2.2 Principle of data input

2.2 Principle of data input

If you press a key you will find the corresponding
menu in the display. Example key <CONFIG>:
configuration >peripheral units
>peripheral units send to: IBM
configuration >auxiliaries
DET pH ********
In the first line you see where you are: You pressed key <CONFIG> and you are now in the menu "con­figuration".
In our example you are in the menu "configuration"
on the title ">peripheral units". By pressing <CONFIG> you can move to the other titles of this menu.
If a dialog text is marked with ">", it contains a
group of inquiries. You branch into this group by pressing <ENTER>. Example inquiries of "peripheral units": The first line indicates again where you are. If a dialog text of an inquiry is marked with ":", you can select a value with the key <SELECT>.
A value is stored with <ENTER> and the cursor
moves to the next inquiry.
Repeated pressing of <ENTER> moves you
through the inquiries of the group ">peripheral units“, after the last inquiry of this group you leave this group and return to the level above. The next group of the menu "configuration“ appears: ">auxiliaries“
With key <QUIT> you leave an inquiry or a group of
inquiries, it always
In this example you leave the menu "configuration"
by pressing <QUIT> and return to the display of ti­tration mode and the chosen method.
moves you one level up.
794 Basic Titrino
7
2.3. Text input

2.3 Text input

<USER METH>
user methods >recall method
<USER METH>
<ENTER>
>store method method name:
<CLEAR>
<<>
>store method method name: ABCDEFG
<QUIT>
<ENTER>
>store method method name: text
Example storing a method:
Press key <USER METH>, the group ">recall
method" appears.
Choose ">store method" by pressing <USER
METH> and press <ENTER>. The name of the method which is currently in the working memory is displayed.
Delete this name with <CLEAR>.
Open the "text writing mode" with key <<>.
Select the character marked by the blinking cursor
with the Keys <<> and <>>, confirm it with <ENTER>. Select the next character...
When you confirmed the last character, i.e. your name is complete, you quit the text writing mode with <QUIT>. Confirm now the name with <ENTER>.
If your text fills the whole text field, just press <ENTER> to store the text.
During text input you can correct typing errors with
<CLEAR>: <CLEAR> deletes the characters one by one.
If you wish to modify an existing name (e.g. if you
have names like Text 1, Text 2, Text 3), do not delete the existing name before you start the text input mode. Proceed then as follows:
1. Press <USER METH>, place the cursor to
">store method" and press <ENTER>.
2. Open the text writing mode directly: Press key
<<>.
3. <CLEAR> now deletes the characters one by
one or you can add additional characters.
794 Basic Titrino
8
2.4 Tutorial

2.4 Tutorial

This short operating course teaches you to work quick and efficient with the 794 Basic Titrino, by means of the most important applications.
Set up your Titrino and connect the peripheral devices needed, see chapter 5.

2.4.1 Entering data, setting the dialog language

We can thus make a start and first take a look at the fundamentals of the entry of data. We change the dialog language.
DET pH ********
<CONFIG>
configuration >peripheral units
<CONFIG>
configuration >auxilliaries
<ENTER>
>auxilliaries dialog: english
<SELECT>
>auxilliaries dialog: español
<ENTER>
>ajustes varios fecha 2002-01-02
Switch on the Titrino. It is now in the standby state, it
shows you the active mode and method name.
Press the key <CONFIG>, the display shows:
This is the title of the group "peripheral units". This group contains various inquiries about peripheral units.
Press again <CONFIG>. You see the title of a new
group of inquiries.
This "auxiliaries" group contains the inquiry for the dialog language.
Pressing the <ENTER> key takes you to the inquir-
ies of the group "auxiliaries". Note the ">" sign. All ti­tles of inquiry groups are prefixed by this sign.
This is the first inquiry of the "auxiliaries" group: the selection of the dialog language.
You select the various dialog languages with the
<SELECT> key. Press <SELECT> repeatedly until "español" appears in the display. Note the sign ":". It appears if the values can be se­lected with the key <SELECT>.
Accept the new "value" with <ENTER>.
The next inquiry "fecha" (date) of the group "ajustes varios" (auxiliaries) is shown.
You can open this inquiry by pressing <ENTER> too and go through all the inquiries of this group this way.
794 Basic Titrino
9
2.4. Tutorial
<QUIT>
configuration >ajustes varios
<QUIT>
DET pH ********
Because this inquiry follows no colon ":" the value can't be selected by <SELECT>, the date "fecha" has to be entered with the numeric keys.
Exit the inquiries with <QUIT>. You are one level
higher in the "configuration" menu showing the title "ajustes varios" (auxiliaries).
Press <QUIT> once again to exit the "configuration"
menu and return to the standby state.
All the dialog texts will now be displayed in Spanish. If you prefer English as the displayed dialog lan­guage, proceed as before and select "English".
794 Basic Titrino
10
2.4 Tutorial

2.4.2 Development of a method, titration of an acid

Selection of the mode
<MODE>
mode mode DET
<ENTER>
DET pH ******** DET: pH
4 × <SELECT>
DET pH ********
For the titration put a Exchange Unit with c(NaOH) = 0.1 mol/L on the Titrino and rinse the tubing and the buret Tipp with <DOS>. Fill the buret again with <STOP/FILL>. Plug a combined pH glass electrode into measuring input 1 (Ind I). Pipette 2 mL c(HCl) = 0.1 mol/L into your titration vessel, dilute with ca. 20 mL dist. water. Put a magnetic stirrer in the titration vessel and place buret tip and electrode in the probe, see page 157 for their arrangement. Start the stirrer and press <START>
During the titration the first line of the display shows the current measured value and the volume already dispensed:
Press <MODE> repeatedly until "DET" appears in
the display. For a description of the DET mode see page 25.
Confirm "DET" with <ENTER>
Now select the measured quantity: Press
<SELECT> repeatedly until "pH" appears again in the display. Confirm the measured quantity "pH" with <ENTER>.
You are now ready to titrate.
pH 3.02 0.351 ml
As soon as the instrument has found an equivalence point, this is shown on the second line:
pH 7.64 2.083 ml EP1
Let the titration continue for a short while, e.g. until ca. pH 11.50. Now stop it with <STOP>.
DET pH ******** EP1 2.083 ml pH 7.64
The first line now shows the mode "DET pH" and the name of the chosen method (here "********" because this method has not been saved under a own name until now). Of interest is the second line, which shows the equivalence point found.
794 Basic Titrino
11
2.4. Tutorial
If more than one equivalence point has been found, the others can be viewed with <ENTER>.
Calculation of the result: formula entry
<DEF>
def >formula
<ENTER>
RS?
<1>
RS1=
RS1=EP1*C01*C02/C00
<ENTER>
>formula RS1 text RS1
>formula RS1 decimal places 2
>formula RS1 unit %
<SELECT>
<ENTER>
RS?
2 x <QUIT>
With this equivalence point a result can be calculated.
Press the <DEF> key.
Press <ENTER> to move on to the formula entry.
The display now shows "RS?".
Press "1", i.e. the first formula.
You now can enter a formula. Note here the top inscrip­tion on the keys of the keypad and the numbers. The following symbols can be used: EP# EP's with 1-digit number, e.g. EP1. RS# Previously calculated results, e.g. RS1 in the sec-
ond formula.
C## Calculation constant with 2-digit number, e.g. C01.
C00 is reserved for the sample size. For the mean­ing of the calculation constants see page 53.
Calculate the content of your hydrochloric acid in g/L: RS1=EP1*C01*C02/C00 End point*conc.(titrant)*molecular weight/sample size To correct a formula, delete it with <CLEAR>.
Confirm the formula with <ENTER>.
You may enter a text for the result output, see page
8.
Enter the desired number of decimal places for the
result.
Select the desired unit g/L with <SELECT>, or type
a text as unit, see page 8.
The Titrino prompts for the calculation of the next result.
Quit the formula entry by pressing <QUIT> twice.
In order to be able to calculate the result, enter the calculation constants used in the for­mula.
794 Basic Titrino
12
2.4 Tutorial
Entry of the calculation constants
<C-FMLA>
C-fmla >C01 0.0
Press <C-FMLA>.
The constants which have been used in the formula are requested: input with digit keys, confirm with
<ENTER>. C01: Concentration of your titrant= 0.1 mol/L. Enter
0.1.
C02: Molar mass of HCl = 36.47 g/mol
Entry of the sample size
4 x <SMPL DATA>
smpl data smpl size 1.0 g
<2>
<ENTER>
smpl data smpl unit: g
<SELECT>
<ENTER>
Press <SMPL DATA> repeatedly until "sample size"
appears in the display.
Enter 2.
Use <SELECT> to select the unit "mL" and confirm
the new value with <ENTER>.
The result is now calculated and can be displayed in place of the equivalence point. If your method already includes a formula at the end of the titration, the calculated result is displayed directly after the titration. As we have entered the formula later, we now have to select the result display:
Display of the result
<SELECT>
DET pH ******** >display results
<ENTER>
DET pH ******** RS1 3.80 g/l
Press <SELECT> repeatedly until ">display results"
appears in the display.
Press <ENTER> to move to the result display.
794 Basic Titrino
13
2.4. Tutorial
If you have a printer connected, you probably wish to have the curve and a result report printed out automatically at the end of a titration. To install a printer, see page 150.
Print reports
4 x <DEF>
def >report
<ENTER>
report:
6 x <SELECT>
<;>
9 x <SELECT>
report:curve;full
<ENTER>
<QUIT>
Press <DEF> repeatedly until the display shows:
Press <ENTER> to move to the definition of reports.
With <SELECT> you select the individual repot
blocks. Use a ";" as delimiter between the report blocks to print more than one block. If you wish to print out a curve and a full result report, enter "curve;full".
Confirm the entry with <ENTER> and quit the inquir-
ies with <QUIT>.
Now print your reports by pressing <PRINT><REPORTS><ENTER>. Your printout will look like the following:
'cu 794 Titrino 01102 794.0010 date 2002-01-03 time 17:34 1 start V 0.000 ml DET pH ********
1.0 ml/div dpH=2.0/div
−−−−−−−−−−−−−−−−
'fr 794 Titrino 01102 794.0010 date 2002-01-03 time 17:34 1 pH(init) 2.88 DET pH ******** smpl size 2 ml EP1 2.083 ml 7.64 RS1 3.80 g/l manual stop
−−−−−−−−−−−−−−−−
Identification of the report type (cu=curve)
Start volume, mode and method name Scaling of curve axis
Identification of the report type (fr=full report) Device type with identification and program version
Initial pH, mode and method name
Volume and pH value of EP1
Calculated result
794 Basic Titrino
14
2.4 Tutorial
So that you do not have to stop your titration manually each time, let us add a stop crite­rion for the titration.
Stop criterion
2 x <PARAM>
parameters >stop conditions
<ENTER>
>stop conditions stop V: abs.
<ENTER>
>stop conditions stop V: 99.99 ml
<ENTER>
>stop conditions stop pH OFF
<ENTER>
2 x <QUIT>
Press <PARAM> twice to display the title ">stop
conditions".
Open this group of inquiries with <ENTER>.
Absolute stop volume. This can be used as a safety measure to prevent overflow of the titration vessel.
The magnitude of the absolute stop volume. Set a value that appears suitable for your titration vessel.
Set the pH value, e.g. pH=11.5 as the expected stop criterion.
If several stop criteria have been set, the one that is reached first applies.
Confirm your entries with <ENTER> and quit the inquiry with <QUIT>.
The development of your method is now complete. Before we store it in the method mem­ory, you should check it again. Prepare a fresh sample and restart the titration with <START>.
If everything appears to be all right, you can now store the method in the method mem­ory.
794 Basic Titrino
15
2.4. Tutorial
Y

2.4.3 Storage and loading of methods

Now store the method you have just developed in the method memory.
Storage of a method
2 × <USER METH>
user methods >store method
<ENTER>
>store method method name: ********
<ENTER>
DET pH Acid
If you have a printer connected, you can print out the contents of your method memory. Key sequence: <PRINT><USER METH><ENTER>
Stored methods can be loaded at any time.
Loading a method from the method memory
<USER METH>
user methods >recall method
<ENTER>
>recall method method name: ********
<SELECT>
or
direct entry
<ENTER>
DET pH Acid
Press <USER METH> repeatedly until the title
">store method" appears in the display.
Open the inquiry with <ENTER>.
Enter an identifier, e.g. 1 or Acid.
For text input see page 8.
The method now runs under the name "Acid". It is ready to titrate.
Press <USER METH>. The display shows the title
">recall method".
Open the inquiry <ENTER>
ou can select the desired method with <SELECT> (it shows all methods of the method memory one af­ter the other) or by directly enter the method name.
Load the method with <ENTER>
The method is ready to titrate.
794 Basic Titrino
16
2.4 Tutorial

2.4.4 pH calibration

pH calibrations are not a requirement for pH titrations, when the equivalence point is de­termined from a curve. For end-point titrations, however, where titration is performed to a fixed, specified pH value, a calibration should be performed.
Selection of the calibration mode, CAL
<MODE>
mode mode CAL
<ENTER>
CAL ******** pH(as) 7.00 slope 1.000
Calibration procedure
<START>
CAL cal.temp. 25.0°C
<ENTER>
25.0°C buffer 1 pH 7.00
<ENTER>
25.0°C buffer 2 pH 4.00
<ENTER>
oder
<STOP>
CAL ******** pH(as) 6.89 slope 0.985
The calibration data can be viewed at any time under the <CAL.DATA> key. Our calibra­tion data are stored under ">input 1".
The calibration report can be printed out at any time with the key sequence:
<PRINT><CAL.DATA><ENTER>
Press <MODE> repeatedly until "CAL" appears in
the display and confirm the mode with <ENTER>.
The instrument is ready for a 2-point calibration. The second display line shows the current calibration data for measuring input 1.
Immerse your electrode in the first buffer and start
the calibration procedure.
Inquiry of calibration temperature.
Enter the current temperature. If you have a tem-
perature sensor connected, this inquiry is skipped.
pH value of the first buffer.
Enter the pH value of the buffer at your calibration
temperature.
The voltage of the first buffer is measured. When the measured value has met the set drift criterion, the measurement is stopped and the pH value of the sec­ond buffer is requested.
Now enter the pH value of the second buffer. If you
require a 1-point calibration, you can also terminate the calibration with <STOP>.
At the end of the calibration, the calibration data ob­tained are displayed: asymmetry pH and slope.
794 Basic Titrino
17
2.4. Tutorial

2.4.5 Statistics, acid capacity of drinking water

Let us now determine the acid capacity of drinking water. For this, the SET (set endpoint titration) mode is used to titrate to pH = 4.3.
First select the mode SET pH (keys <MODE> and <SELECT>).
Now set the end point.
Entry of the end point and the control parameters
<PARAM>
parameters >SET1
<ENTER>
>SET1 EP at pH OFF
>SET1 EP at pH 4.3
<ENTER>
>SET1 dynamics OFF
>SET1 dynamics 3
<ENTER>
>SET1 max.rate 10.0 ml/min
<ENTER>
>SET1 min.rate 25.0µl/min
2 × <QUIT>
Mount an Exchange Unit with c(HCl) = 0.1 mol/L as titrant. Add 25 mL drinking water as sample and start the titration with <START>. During the titration the first line of the dis­play shows the measured value and dispensed volume. The "c" in pHc shows that the electrode assembly has been calibrated. The second line shows a "control bar", which indicates the control deviation of the current measured value from the set end point.
Press <PARAM>. The display shows:
Control parameters for EP1
Set the end point to pH = 4.3.
Dynamics means the control range in which the
adding of the titrant is controlled in dependence on the measured value (OFF: largest control range, i.e. slow titration). With buffered systems such as drinking water, this value can be set lower, e.g. to 3. I.e. 3 pH units be­fore the set EP controlled dosing starts.
Maximum possible titration rate inside the control range.
Minimum titration rate inside the control range.
Quit the inquiry with <QUIT>.
pHc 6.34 0.426 ml #======
794 Basic Titrino
18
2.4 Tutorial
If the titration runs too slow or too fast, you can change the control parameters during the titration. If you wish to titrate faster, change the following parameters:
dynamics: lower value, the control range becomes smaller (Attention: possibility of
titrating over the set EP)
max.rate: bigger value
min.rate: bigger value
You will find further details of the control parameters on page 42.
After the titration, end point volume and pH value at the end point are displayed.
For the evaluation enter:
the formula to calculate the m value (key <DEF>) RS1=EP1*C01*C02
with an accuracy of RS1 decimal places 2 the unit RS1 unit: mmol/L
and the calculation constants C01 1 (concentration of titrant ×10) (key <C-FMLA>) C02 4 (factor for the sample size 100 mL/25 mL)
If you have a printer connected, select the automatic titration report (key <DEF>): "re­port:full". You can print out the results of the previous titration with keys <PRINT><REPORTS><ENTER>.
If the previous titration has run to your satisfaction, you can start thinking about perform­ing statistics calculations. Have you already added a new sample to the titration vessel? If you are no longer certain, you can find out immediately with <MEAS/HOLD>.
Rapid measurement between titrations
Press <MEAS/HOLD>. The pH value of your sample is displayed. You can stop the measurement with a second <MEAS/HOLD>.
Statistics calculations
Now switch on the statistics calculations. Press <STATISTICS>. The "Statistics" LED is on. Duplicate determinations are now performed.
Perform 2 titrations.
794 Basic Titrino
19
2.4. Tutorial
After the second titration you receive a printout with statistics calculations:
'fr 794 Titrino 01102 794.0010 date 2002-01-03 time 17:34 3 pHc(init) 6.29 SET pH ******** EP1 0.0628 ml 4.26 m value 5.02 mmol/l mean(2) +/s s/% m value 5.04 0.028 mmol/l 0.56 ==============
If you have no printer connected, you can view the mean value and standard deviation:
Press <SELECT> until "display mean" appears in the display.
With <ENTER> you obtain the mean value.
A second <ENTER> displays the number of single values which have been used for
the mean calculation.
If you press <SELECT> again, the display "display std.deviation" appears. Once
again, you can view this value, by pressing <ENTER>.
Perhaps you have noticed that the two values differ too greatly? In any case, we shall per­form a third determination with the same sample. The result of this determination will be another value which is incorporated in the statistics calculations.
Addition of more determinations for the statistics calculations
5 × <PARAM>
parameters >statistics
<ENTER>
>statistics status: ON
<ENTER>
>statistics mean: n= 2
<3>
<ENTER>
2 × <QUIT>
Press <PARAM> until the display shows:
Statistics calculations can be switched on either with
the <STATISTICS> key or in the inquiry "status:". We shall leave them switched on and proceed to the next inquiry.
Mean value of 2 single determinations.
Enter "3" to include an additional determination.
Quit the inquiry.
Perform another titration.
794 Basic Titrino
20
2.4 Tutorial
You can now decide which of the results is an "outlier". You can delete it from the statistics calculation.
Deleting a result from the statistics calculation
5 × <PARAM>
parameters >statistics
3 × <ENTER>
>statistics res.tab: original
2 × <SELECT>
>statistics res.tab: delete n
<ENTER>
>statistics delete n= 1
<2>
<ENTER>
2 × <QUIT>
Press <PARAM> until the display shows:
Select with <ENTER> the inquiry of the result table
"res.tab:"
To delete a single result with index n from the result
table, press <SELECT> so that "delete n" appears.
Enter index n of the result you wish to delete; in our
example the second:
Quit the inquiry by pressing <QUIT>.
Mean value and standard deviation are recalculated and can be viewed in the display.
With <PRINT><REPORTS><ENTER> you can print a fresh report.
With <PRINT> <STATISTICS><ENTER> you can print a statistics overview, in this report the deleted result is marked with a "" in the report.
794 Basic Titrino
21
2.5. Configuration, key <CONFIG>

2.5 Configuration, key <CONFIG>

CONFIG
configuration >peripheral units
>peripheral units
send to: IBM
balance: Sartorius
record: U
Key <CONFIG> serves to enter device specific data. The set values apply to all modes.
peripheral units: Selection of printer, balance and the curve at the ana­log output. auxiliaries: e.g. setting of dialog language, date, time, etc. RS232 settings: RS parameters for the COM interface. common variables: Values of common variables.
The display texts of the Titrino are shown to the left. The values are the default values.
Settings for peripheral units
Selection of printer (Epson, Seiko, Citizen, HP, IBM) at the Titrino Rs232 interface.
"Epson", for Epson "Seiko", e.g. for DPU-414 "Citizen", e.g. for iDP 562 RS, Custom DP40-S4N "HP" e.g. for Desk Jet types. Place curves always at the beginning of a page as you cannot have them over 2 pages. "IBM" for all printers with IBM character set Table 437 and IBM graphics, as well as for the data transmission to a computer with Metrodata software TiNET or VESUV.
Selection of balance (Sartorius, Mettler, Mettler AT, AND, Precisa)
Sartorius: Models MP8, MC1 Mettler: Models AM, PM, AX, MX, UMX and
balances with 011, 012, or 016 interface Mettler AT: Model AT AND: Models ER-60, 120, 180, 182, FR-200, 300
and FX-200, 300, 320 Precisa: Models with RS232C interface
Selection of the curve for the output at the analog output (U, dU/dt, V, dV/dt, U(rel), T
U:. Voltage dU/dt: Measured value drift V: Volume dV/dt Volume drift U(rel): Control deviation with SET T: Temperature with MEAS T
794 Basic Titrino
22
2.5 Configuration, key <CONFIG>
>auxiliaries
dialog: english
date 2001-01-03
time 08:13
run number 0
auto start OFF
start delay 0 s
dev.label.
program 794.0010
>RS232 settings
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
RS control: ON
General settings
Selection of dialog language (english, deutsch, francais, español, italiano, portugese, svenska)
Current date (YYYY-MM-DD)
Format: Year-month-day, entry with leading zeros.
Current time (HH-MM) Format: Hours-minutes, entry with leading zeros.
Current run number for result output (0...9999) The sample number is set to 0 when the instrument is switched on and incremented on every determination.
Automatic starts of titrations. (1...9999, OFF) Number of automatic starts ("number of samples"). Used for instrument interconnections in which the external instrument does not initiate a start. Not advisable in connections with Sample Changers.
Start delay (0...999999 s) Delay time after start of methods. Abort start delay time with <QUIT>.
Individual identification of devices (up to 8 ASCII characters)
Will be printed in the result report, see page 59.
Display of program version
Settings of RS232 interface see also page 125ff.
Baud rate (300, 600, 1200, 2400, 4800, 9600)
Data bit (7, 8)
Stop bit (1, 2)
Parity (even, odd, none)
Handshake (HWs, HWf, SWline, SWchar, none)
see page 125ff.
Control via RS232 interface (ON, OFF) "OFF" means that the receipt of commands via the RS232 interface is blocked. Data output
is possible.
794 Basic Titrino
23
2.6. Selection of the mode, key <MODE>
>common variables
C30 0.0 etc.
Values of the common variables
Common variables C30...C39 (0..
The values of all common variables are displayed. For creating of common variables see page 57.

2.6 Selection of the mode, key <MODE>

MODE
>
;
Press key <MODE> until the desired mode is displayed and confirm with <ENTER>. Select the measured quantity pH, U, Ipol, Upol, (T) with <SELECT> and confirm it also with <ENTER>.
The following modes can be selected:
DET: Dynamic Equivalence-point Titration
MET: Monotonic Equivalence-point Titration
SET: Set Endpoint Titration.
CAL: pH Calibration.
MEAS: Measuring.
TIP: Titration Procedure. Linking of various
commands and methods to a titration procedure.
These standard modes are equipped with a set of standard parameters. They only need few settings in order to be ready to work. TIP is an empty "shell". The TIP sequence has to be defined with <DEF>, see page 62.
±
999 999)
794 Basic Titrino
24
2.6 Selection of the mode, key <MODE>
Overview of the titration modes
DET
Dynamic Equivalence
point Titration
Titration
Evaluation
Applications
Reagent feeding: Variable volume increments, depending on the slope of the curve.
U/mV
Acquisition of measured values: Drift controlled ("equilibrium titration")
and/or after a fixed equilibration time.
The evaluation of EP' s is based on the zero crossing of the second derivative with a Metrohm correction for the distortion of the curve from superimposed jumps. Can be combined with selectable recognition criteria.
Recognition criteria:
as for MET
Suitable titration mode for most problems. Specially recommended if jumps lie very close together and for very flat jumps.
Note: The reagent feeding algorithm is based on meas­ured data. The curve should therefore not deviate markedly from S-shape.
V/mL
MET
Monotonic Equivalence
point Titration
Reagent feeding: Constant volume incre­ments, independent of the slope of the curve.
U/mV
Acquisition of measured values: Drift controlled ("equilibrium titration")
and/or after a fixed equilibration time.
The evaluation of EP's is based on the Fortuin inter­polation.
Recognition criteria:
all EP's only the last EP only the greatest EP EP windows
For
slow titration reactions
(diazotations, coupling reactions)
sluggish electrode
response.
V/mL
SET, KFT
Endpoint Titration
Titration to preset end-point.
U/mV
Control range
Acquisition of measured values: Continuously
Volume that has been dis­pensed up to the endpoint (EPX in mL).
For rapid, quantitative
determinations in ana­lytical chemistry. Re­quirement: EP of the ti­tration reaction is known and does not change during a de­termination series.
If an excess of titrant
must be avoided.
EP
V/mL
794 Basic Titrino
25
2.7. Parameters, key <PARAM>

2.7 Parameters, key <PARAM>

The key <PARAM> is used for the entry of values that
PARAM
determine the modes. Values marked with "cond." are accessible during the conditioning in the SET mode. "**titr." means that these values can be changed dur­ing the titration. They influence the ongoing determina­tion. Other values can only be changed in the inactive state. The display texts of the Titrino are shown to the left. The values are the default values.

2.7.1 Parameters for DET and MET

parameters >titration parameters
>titration parameters
meas.pt.density 4 DET
min.incr. 10.0 µl DET
V step 0.10 ml MET
titration parameters determine the course of the titration and measured value acquisition. stop conditions: Parameters for the automatic termination of the titra­tion. statistics: Calculation of mean values and standard deviation, see also page 55. evaluation: Parameters for the evaluation of EP's, fix EP's, and pK/HNP. preselections: ON/OFF of various auxiliary functions such as auto­matic requests after the start and activate pulse.
General titration parameters
Measuring point density (0...9) 0 means highest density, 9 lowest. Selection of the meas.pt.density, see page 32.
Minimum increment (0...999.9 µL) The increment is dispensed at the beginning of the titration and in the region of the equivalence point.
Size of volume increment (0...9.999 mL) Dosing step. Small volume increments are needed to determine blank values or to assure accuracy with highly unsymmetrical curves. Selection of size of the increment, see page 33.
794 Basic Titrino
26
Loading...
+ 168 hidden pages