Modular IC System for cation elimination........................................................18
Fig. 9
: Assembling the actuator..................................................................................34
Fig. 10
: Adjusting the braking range ............................................................................37
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
II
1.1 Instrument description
1 Introduction
1.1 Instrument description
The 793 IC Sample Prep Module consists of a reactor block (SP
Module A) with its corresponding control unit; this contains an addi-
tional 2-channel peristaltic pump.
Inline sample pretreatment like neutralization or cation exchange for ion
chromatographic analyses can be carried out with the cation exchanger
built into the SP Module A. The SP Module A contains three cation ex-
change units which are in turn used for cation exchange, regenerated
with acid and rinsed with water. In order to perform each analysis under
comparable conditions a freshly regenerated cation exchange unit is
normally used. Switching is either carried out automatically by the IC
system or manually. Due to possible interference of sulfate with the IC
determination, perchloric acid is normally used for cation exchanger regeneration of the SP Module A instead of sulphuric acid.
At the control unit the SP Module A can either be switched to the next
position manually or by remote control via the remote interface. The
built-in peristaltic pump, which delivers the solutions required for operating the SP Module A, can also be switched on by remote control or
manually.
The 793 IC Sample Prep Module is particularly suitable for retrofitting a
modular anion IC system for inline sample pretreatment in a simple
manner. Two typical applications are described in detail in these instructions for use: Neutralization (e.g. exchange of Na
Cation Separation (e.g. exchange of heavy metals by H
the 793 IC Sample Prep Module can also be combined with all other
HPLC components currently on the market.
Although in principle the techniques of cation exchange by the 793 IC
Sample Prep Module and suppression by the 753 Suppressor Module
are comparable, the 793 IC Sample Prep Module is not suitable for
chemical suppression after ion chromatographic separation of anions.
+
by H+) and
+
). Of course
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
1
1 Introduction
1.2 Parts and controls
11114
IC Sample Prep Module793
POWER
REMOTEPUMP
ONON
22223
OFFOFF
36
33
45
44
5
55
REMOTE
ACTUATOR
ON
OFF
67
66
Met r ohm
78
77
STEP
8
88
1210
1212
Fig. 1
: Front of the 793 IC Sample Prep Module
1111
Power on lamp
Lights up when instrument is switched
on
2222 Pump Remote ON/OFF
On/Off switch for remote control of the
pump
3333 Pump ON/OFF
On/Off switch for the pump (manual
operation mode)
4444 Tubing cartridge
For 6.1826.010 Pump tubing
5555 Contact pressure lever
For adjusting the contact pressure
6666 SP Module A Remote ON/OFF
On/Off switch for remote control of the
actuator
11
1112
1111
7777 SP Module A position display
Is lit up when the actuator is "in position"
8888 Actuator step
Switches the actuator to the next
position (manual operation mode)
9999 Holding clamp
For locking the tubing cartridge into
place
10
10 Snap-action lever
1010
For releasing the tubing cartridge
11
11 Pump drive
1111
Roller head with contact rollers
12
12 Mounting pin
1212
For attaching the tubing cartridge
10
1010
9999
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
2
1.2 Parts and controls
14
13
1313
Type
1413
1414
WARNING - Fire Hazard -
For continued protection replace only
with the same type and rating of fuse
20
2016
2020
Fig. 2
13
13 Mains switch
1313
For switching the instrument on/off:
I = ON 0 = OFF
110V-120V
220V-240V
50 - 60 Hz 50 VA
0,5A(T)
0,25A(T)
Remote
PumpActuator
A
1718
1717
ActuatorBreakpoint
CB
Made by Metrohm Herisau Switzerland
1617
1616
''''t
S
19
1915
1919
18
1818
: Rear of the 793 IC Sample Prep Module
17
17 Connection for SP Module A
1717
15
1515
14
14 Serial number 18
1414
15
15 Remote interface for pump
1515
18 Adjustment screw for braking
1818
19
19 Mains connection plug
1919
Remote I/O lines for connection of
external devices (e.g. 732 IC Detector)
16
16 Remote interface for actuator
1616
20
20 Fuse holder
2020
Remote I/O lines for connection of
external devices (e.g. 732 IC Detector)
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
range
Mains connection, see section 2.2
Changing the fuses, see section 2.2
3
1 Introduction
21
21
Reactor block
2121
SP Module A
Ion an alys is
22
22 Actuator connection
2222
with permanently fixed inlet and
outlet capillaries
23
23 Connection cable
2323
Connection cable to control unit
21
2122
2121
2223
2222
1
2
3
23
2323
Fig. 3
: 1.793.0110 SP Module A
1.3 Information about the Instructions for Use
Please read through these Instructions for Use carefully before you put
the 793 IC Sample Prep Module into operation. The Instructions for
Use contain information and warnings to which the user must pay
attention in order to assure safe operation of the instruments.
1.3.1 Organization
These Instructions for Use8.793.1003 for the 793 IC Sample Prep
Module provide a comprehensive overview of the startup procedure,
operation, fault rectification and technical specifications of these instruments. The Instructions for Use are organized as follows:
Section 1 Introduction
Section 2 Installation
Section 3 Operation
Section 4 Faults – Maintenance
Section 5 Appendix
General description of instruments, parts and controls
and safety notes
Setting up, mains connection, attachment of accessories, connection to IC system
Manual operation and operation via 732 IC Detector
Fault rectification, maintenance
Technical data, standard equipment, options, warranty,
declarations of conformity, index
To find the required information on the instruments please use either
the Table of contents or the Index at the back.
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
4
1.3 Information about the Instructions for Use
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these Instructions for Use:
<PUMP> Switch or key
15
15 Part or control of 793
1515
89
89 Part or control of 732/733
8989
"
Range
"
Parameter or entry value
at 732 IC Detector
>PARAM/detector
range: 1.00 mS/cm
Display
Text in display 1111 of the 732 IC
Detector
Hazard
This symbol draws attention to a
possible danger to life or of injury if
the associated directions are not
followed correctly.
Warning
This symbol draws attention to
possible damage to instruments or
instrument parts if the associated
directions are not followed correctly.
Caution
This symbol marks important
information. First read the associated directions before you continue.
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
Comment
This symbol marks additional
information and tips.
5
1 Introduction
1.4 Safety notes
1.4.1 Electrical safety
While electrical safety in the handling of the 793 IC Sample Prep Module is assured in the context of the specifications IEC 1010-1 (protection
class I, degree of protection IP20), the following points should be noted:
x Mains connection
Setting the mains voltage, checking the mains fuse and the mains
connection must be effected in accordance with the instructions in
section 2.2.
x Opening the 793 IC Sample Prep Module
When the 793 IC Sample Prep Module is connected to the power
supply the instrument must neither be opened nor should parts be
removed from it, otherwise there is a danger of coming into contact
with components which are live. Always disconnect the instrument
from all voltage sources before you open it and ensure that the mains
cable is disconnected from mains connection 19
19 !
1919
x Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 793 IC Sample Prep Module, you should earth yourself and
any tools you are using by touching an earthed object (e.g. housing of
the instrument or a radiator) to eliminate any static charges which
exist.
1.4.2 General precautionary rules
x Handling of solvents
Check all pump tubings and all input and output lines periodically for
possible leaks. Follow the relevant instructions regarding the handling
of flammable and/or toxic solvents and their disposal.
x Regular exchange of pump tubings
Pump tubings are consumable material and must be replaced from
time to time (see Section 4.2.4). Suitable measures must be taken so
that any leak which might occur in the pump tubing or connections
during unattended operation will cause no damage (placing the
instrument at the bottom, collection device for any liquid which may
leak out).
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
6
2.1 Setting up the instrument
2 Installation
2.1 Setting up the instrument
2.1.1 Packaging
The 793 IC Sample Prep Module is supplied together with the separately packed accessories in special packagings containing shockabsorbing foam linings designed to provide excellent protection. The
actual instrument is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as
only they can assure damage-free transport of the instrument.
2.1.2 Check
After receipt, immediately check whether the shipment is complete and
has arrived without damage (compare with delivery note and list of
accessories in section 5.2). In the case of transport damage, see
instructions in section 5.4.1 "Warranty".
2.1.3 Location
Position the instruments in the laboratory at a location convenient for
operation, free from vibrations and protected against a corrosive
atmosphere and contamination by chemicals.
2.1.4 Arrangement of the instruments
The best position for the 793 IC Sample Prep Module is beside the Modular IC System (e.g. 733 IC Separation Center).
The 793 IC Sample Prep Module should always be placed at the
bottom of the instrument stack so that any leaks which may occur in
the pump tubing or connections cannot cause damage to the other
instruments by leakage of corrosive liquids (e.g.acid).
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
7
2 Installation
2.2 Mains connection
Follow the instructions below for connecting to the power supply. If
the instrument is operated with the mains voltage set wrongly and/or
wrong mains fuse there is a danger of fire!
2.2.1 Setting the mains voltage
Before switching on the 793 IC Sample Preparation Module for the first
time, check that the mains voltage set on the instrument (see Fig. 2)
matches the local mains voltage. If this is not
the mains voltage on the instrument as follows:
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 19
793 IC Sample Preparation Module.
the case, you must reset
19 of the
1919
2 Remove fuse holder
Using a screwdriver, loosen fuse holder 20
connection 19
3 Check fuse
Carefully take the fuse installed for the desired mains voltage
out of fuse holder 20
of the fuse in the fuse holder is marked by the white arrow
imprinted next to the mains voltage range):
100}}}}120 V 0.5 A (slow-blow) Metrohm No. U.600.0013
220}}}}240 V 0.25 A (slow-blow) Metrohm No. U.600.0010
4 Insert fuse
Change fuse if necessary and reinsert in fuse holder 20
5 Install fuse holder
Depending on the desired mains voltage, insert fuse holder in
the 793 IC Sample Preparation Module so that the corresponding mains voltage range can be read normally and the adjacent
white arrow points to the white bar imprinted below the fuse
holder (see Fig. 4).
19 and take out completely.
1919
20 and check its specifications (the position
2020
20 below the mains
2020
20.
2020
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
8
2.2 Mains connection
2.2.2 Fuses
13
13
1313
19
19
1919
20
20
2020
220 – 240 V
100 ---- 120 V
220 ---- 240 V
100 – 120 V
220 ---- 240 V
100 ---- 120 V
13
13 Mains switch
1313
19
19 Mains connec-
1919
tion plug
22220000 Fuse holder
Fig. 4
: Setting the mains voltage
One of the two fuses 0.5 A/slow-blow for 100}120 V or 0.25 A/slowblow for 220}240 V is installed in fuse holder 20
20 of the 793 IC Sample
2020
Prep Module as standard.
Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is danger of fire!
For checking or changing fuses, proceed as described in section 2.2.1.
2.2.3 Mains cable and mains connection
Mains cable
The instrument is supplied with one of three mains cables:
x 6.2122.020 with plug SEV 12 (Switzerland, })
x 6.2122.040 with plug CEE(7), VII (Germany, })
x 6.2133.070 with plug NEMA 5-15 (USA, })
which are three-cored and fitted with a plug with an earthing pin. If a different plug has to be fitted, the yellow/green lead (IEC standard) must
be connected to protective earth (protection class I).
Any break in the earthing inside or outside the instrument can make it
a hazard!
Mains connection
Plug the mains cable into mains connection plug 19
Sample Prep Module (see Fig. 4).
19 of the 793 IC
1919
2.2.4 Switching the instrument on/off
The 793 IC Sample Prep Module is switched on and off using mains
switch 13
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
13. When the instrument is switched on lamp 1111 lights up.
1313
9
2 Installation
2.3 Connection to a Modular IC System for
Neutralization
An important application for the 793 IC Sample Prep Module is neutralization of an alkaline sample for anion determination by IC with chemical
suppression. Strong bases (e.g. 30 % NaOH) should not be led unlimited to the SP Module A. To provide a sufficient capacity for exchanging
+
by H+ Ions, only a small fraction (e.g. 20 µL) is directed to the SP
Na
Module A using a sample loop. The analyte anions are preconcentrated
subsequently on a preconcentration column and finally eluted to the
separation column in counter flow by the eluent.
Therefore, such an IC system corresponds to the Modular IC System 6 (MIC 6: Anion system with chemical suppression, preconcentration
and matrix elimination) enhanced by the 793 IC Sample Prep Module.
If the sample doesn’t require matrix elimination and preconcentration, it
can be treated directly by the SP Module A. This could be an option for
samples containing only low amounts of the cations which have to be
eliminated. For this purpose, a system as described in section 2.4
should by used.
The following pages describe the electronic connections of this system
in combination with the 766 IC Sample Processor on the base of full
control by «IC Net 2.1».
2.3.1 Electrical connection
The 793 IC Sample Prep Module is connected to an IC system consisting of 732 IC Detector, 733 IC Separation Center (two injectors), 709 IC
Pump, 752 IC Pump Unit, 753 IC Suppressor Module and 762 IC Interface as shown in Fig. 5.
6.2128.13 0
793
6.2128.180
753
PC
6.2134.100
6.2143.21 0
6.2125.120
6.2143.200
732
6.2125.090
733
B (LOOP) A (PCC)
6.2128 .180
6.2115.070
762
6.2134.09 0
Fig. 5
752
6.2128.180
709
6.2141.11 0
6.2134.08 0
: Connection 793 IC Sample Prep Module -
Modular IC System for Neutralization
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
10
2.3 Connection to a Modular IC System for Neutralization
A
Please note, that the suppressor of the 753 IC Suppressor Module is
not switched by the 766 Sample Processor program. The remote connection 16
16 (suppressor/actuator) of the 753 IC Suppressor Module is
16 16
connected to the event line “fill” of the 733 valve B instead, using the
6.2128.180 cable. Thus the suppressor is switched when the fill position
is accessed by valve B.
2.3.2 Connection of the SP Module A
The following description implies the usage of the 2.733.0120 IC Separation Center (two injectors).
The 1.793.0110 SP Module A must first be inserted in the 733 IC Separation Center and connected to the 793 IC Sample Prep Module. Additionally, the suppressor block of the 753 Suppressor Module is inserted
and connected, which is described in detail in the corresponding
8.753.1001 Instructions for Use.
The connection of all parts should be done as described by the following scheme:
) Sample loop at valve B
is filled.
B) Sample is neutralized
by SP Module A;
Anions are preconcentrated via valve A..
sample
sample
FillInject
SP
B
Module
A
A
preconcentration
752793709
H2ORegenerantRegenerantEluentH2OH
column
InjectFill
SP
B
Module
A
A
752793709
ORegenerantRegenerantEluentH2OH
H
2
preconcentration
column
InjectInject
separation
column
separation
column
MSM
753
MSM
753
O
2
O
2
C) Preconcentrated
anions are eluted to
the separation column.
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
sample
SP
B
Module
A
752793709
ORegenerantRegenerantEluentH2OH
H
2
preconcentration
column
separation
A
column
MSM
753
O
2
11
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.