Metrohm 791VA User Manual

Page 1
791 VA Detector
Manual
8.791.1013
Page 2
Page 3
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
791 VA Detector
Manual
8.791.1013 01.2006 chs
Page 4
Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://products.metrohm.com under Literature/Technical documenta-
tion.
Page 5
Table of contents

Table of contents

1 Introduction ........................................................................................... 1
1.1 Instrument description .............................................................................1
1.2 Parts and controls.....................................................................................2
1.3 Information on the Instructions for Use ................................................4
1.3.1 Organization ..................................................................................4
1.3.2 Notation and pictograms ..............................................................5
1.4 Safety notes................................................................................................6
1.4.1 Electrical safety..............................................................................6
1.4.2 General precautionary rules..........................................................6
2 Installation.............................................................................................. 7
2.1 Setting up the instrument.........................................................................7
2.1.1 Packaging......................................................................................7
2.1.2 Check ............................................................................................7
2.1.3 Location.........................................................................................7
2.1.4 Arrangement of the instruments....................................................7
2.2 Mains connection ......................................................................................8
2.2.1 Setting the mains voltage.............................................................. 8
2.2.2 Fuses ............................................................................................. 9
2.2.3 Mains cable and mains connection .............................................. 9
2.2.4 Switching the instrument on/off.....................................................9
2.3 656 Electrochemical detector................................................................10
2.3.1 Installation and startup ................................................................10
2.3.2 Connection to 791 VA Detector...................................................10
2.4 6.5303.030 ELCD cell for IC...................................................................11
2.4.1 Assembling the ELCD cell...........................................................11
2.4.2 Connection to 791 VA Detector...................................................13
2.5 Analog output connections ...................................................................14
791 VA Detector
3 Operation...............................................................................................15
3.1 Electrochemical detection .....................................................................15
3.1.1 Classes of substances which can be determined...................... 15
3.1.2 Influence on detector signal........................................................16
3.1.3 Selection of polarization voltage .................................................17
3.1.4 Procedure for unknown substances ...........................................17
3.1.5 Practical information about ELCD...............................................18
3.2 Electrodes.................................................................................................20
3.2.1 Working electrodes .....................................................................20
3.2.2 Reference electrode ....................................................................22
3.2.3 Auxiliary electrode .......................................................................22
3.3 Operating element functions.................................................................23
3.4 Startup.......................................................................................................26
3.4.1 Preparing the ELCD cell..............................................................26
3.4.2 Zero balancing and equilibration phase .....................................27
3.4.3 Measuring procedure..................................................................29
I
Page 6
Table of contents
4 Maintenance – Malfunctions.......................................... 31
4.1 Maintenance and servicing ...................................................................31
4.2 Shutdown..................................................................................................31
4.3 Malfunctions and their rectification .....................................................32
4.4 Instrument test with the dummy cell.................................................... 34
4.4.1 Check of "meas" and "damp" commutators and
of current amplifier offset ............................................................34
4.4.2 Check output voltage.................................................................. 35
4.4.3 Check of current compensation .................................................36
4.4.4 Check of current and voltage overload ...................................... 38
4.4.5 Check sensitivity knob
4.4.6 Check of noise from current amplifier......................................... 40
8 ............................................................. 39
5 Appendix................................................................................................. 41
5.1 Technical data..........................................................................................41
5.2 Standard equipment ...............................................................................43
5.2.1 2.791.0010 VA Detector for HPLC .............................................. 43
5.2.2 2.791.0020 VA Detector for IC ....................................................44
5.3 Optional accessories.............................................................................. 45
5.4 Warranty and conformity .......................................................................46
5.4.1 Warranty ......................................................................................46
5.4.2 EU Declaration of conformity ...................................................... 47
5.4.3 Certificate of conformity and system validation.......................... 48
5.5 Index..........................................................................................................49
List of figures
Fig. 1: Front of the 791 VA Detector ............................................................................. 2
: Rear of the 791 VA Detector.............................................................................. 3
Fig. 2
Fig. 3
: Setting the mains voltage.................................................................................. 9
Fig. 4
: Connection 656 – 791 .....................................................................................10
Fig. 5
: ELCD cell assembly ........................................................................................12
Fig. 6
: Connection of ELCD cell at detector block..................................................... 13
Fig. 7
: Connection of ELCD cell at 791 VA Detector.................................................. 13
Fig. 8
: Selection of polarization voltage .....................................................................17
Fig. 9
: Circuit of 6.2813.020 Dummy Cell................................................................... 34
791 VA Detector
II
Page 7
1.1 Instrument description

1 Introduction

1.1 Instrument description

The 791 VA Detector allows the use of electrochemical/amperometric detection in IC and HPLC; various working electrodes are available de­pending on the application. It can be used in series with another detec­tor (e.g. 732 Conductivity detector) or simply as a «stand alone» detec­tor.
The self-ventilating measuring cell functions according to the time­proved wall-jet principle with three-electrode technology. A silver/silver chloride system is used as the reference electrode and a solid gold pin as the auxiliary electrode. The carbon paste electrode and the silver electrode have proved to be the best working electrodes; but other electrode materials such as glassy carbon, gold, platinum and impreg­nated graphite are available for special applications.
The 791 VA Detector is available in the two following versions:
2.791.0010 VA Detector for HPLC
The measuring cell is built into the 656 Electro­ chemical detector.
2.791.0020 VA Detector for IC
The measuring cell belongs to the accessories and is built into the 733 IC Separation Center.
791 VA Detector
1
Page 8
1 Introduction

1.2 Parts and controls

13
2 4567
VA Detector791
I
comp
0.02 nA
+
~
+
overload meas
015
U
pol
10 mV
I
0.01
comp
x5damp
.
I
nAx50
50
10
0,5
0,1
5
1
Met rohm
I
(U = 1 V)
out
µ
0,1
0,5
1
5
100
10
A
4
11 89
Fig. 1
Mains pilot lamp
1
Lit up when instrument switched
: Front of the 791 VA Detector
10
7 Commutator
Changing the polarity of I
on
2 Overload display 8 Adjusting knob
Setting the current measuring sensi­tivity
3 Commutator
Changeover Stand-by/Measure
4 Commutator
Switching on/off damping
5 Commutator
Switching on/off multiplication of
by a factor of 5
I
comp
9 Potentiometer
Setting the absolute compensation current I
comp
10 Potentiometer
Setting the relative compensation current I
comp
11 Digital switch
Setting the polarization voltage
comp
6 Commutator
Switching on/off multiplication of
by a factor of 50
I
comp
791 VA Detector
2
Page 9
1.2 Parts and controls
12
151413
Type 1.791.0010
f = 50-60 Hz Made by Metrohm Herisau Switzerland S = 5 VA
Fuse
100-240 V: 1A(T)
WARNING - Fire Hazard -
For continued protection replace only with the same type and rating of fuse
pilot voltage
Nr.
16 17 18 19
AE RE WE shield
output 0...1 V
com 0...10 mV
Fuse holder
12
changing the fuses, see section 2.2
26 20
25
Fig. 2
: Rear of the 791 VA Detector
20 Selector
Selection of polarity for analog output
13 Mains voltage indicator 21 Selector
Selection of full-scale deflection for analog output
14 Earthing socket 22 Analog output (live)
15 Connection AE
23 Analog output (common)
Connection of auxiliary electrode
16 Serial number 24 Pilot voltage
Input for external voltage input for potentiostat control
17 Connection RE
Connection of reference electrode
25 Mains switch
To switch instrument on/off: I = ON 0 = OFF
21222324
18 Connection WE
26 Mains connection plug
Connection of working electrode
19 Connection for protective screen
791 VA Detector
Mains connection, see section 2.2
3
Page 10
1 Introduction

1.3 Information on the Instructions for Use

Please read through these Instructions for Use carefully before you put the 791 VA Detector into operation. The Instructions for Use contain information and warnings to which the user must pay attention in order to assure safe operation of the instrument.

1.3.1 Organization

These 8.791.1013 Instructions for Use for the 791 VA Detector pro­vide a comprehensive overview of the installation, startup procedure, operation, fault rectification and technical specifications of this instru­ment. The Instructions for Use are organized as follows:
Section 1 Introduction
General description of instrument, parts and controls and safety notes
Section 2 Installation
Installation of accessories, connection to IC system
Section 3 Operation
Electrodes, operating element functions, startup
Section 4 Maintenance – Faults
Maintenance, fault rectification
Section 6 Appendix
Technical data, standard equipment, options, warranty, declarations of conformity, index
To find the required information on the instruments you will find it an advantage to use either the Table of contents or the Index at the back.
791 VA Detector
4
Page 11
1.3 Information on the Instructions for Use

1.3.2 Notation and pictograms

The following notations and pictograms (symbols) are used in these In­structions for Use:
15 Part or control of 791
Hazard
This symbol draws attention to a possible danger to life or of injury if the associated directions are not followed correctly.
Warning This symbol draws attention to possible damage to instruments or instrument parts if the associated directions are not followed cor­rectly.
Caution This symbol marks important information. First read the associ­ated directions before you con­tinue.
Comment This symbol marks additional information and tips.
791 VA Detector
5
Page 12
1 Introduction

1.4 Safety notes

1.4.1 Electrical safety

While electrical safety in the handling of the 791 VA Detector is assured in the context of the specifications IEC 1010-1 (protection class 1, de­gree of protection IP40), the following points should be noted:
Mains connection
Setting of the mains voltage, checking the mains fuse and the mains connection must be effected in accordance with the instruc-
tions in section 2.2.
Opening the 791 VA Detector
If the 791 VA Detector is connected to the power supply, the instru­ment must not be opened nor must parts be removed from it, other­wise there is a danger of coming into contact with components which are live. Hence, always disconnect the instrument from all voltage sources before you open it and ensure that the mains cable is disconnected from mains connection 26 !
Protection against static charges
Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the 791 VA Detector, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist.

1.4.2 General precautionary rules

Handling of solvents
Check all lines periodically for possible leaks. Follow the relevant instructions regarding the handling of flammable and/or toxic solvents and their disposal.
791 VA Detector
6
Page 13

2.1 Setting up the instrument

2 Installation

2.1 Setting up the instrument

2.1.1 Packaging

The 791 VA Detector is supplied together with the separately packed accessories in special packagings containing shock-absorbing foam linings designed to provide excellent protection. The instrument itself is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they assure transport of the instrument free from damage.

2.1.2 Check

After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of accessories in section 5.2). In the case of transport damage, see instructions in section 5.4.1 "Warranty".

2.1.3 Location

Position the instrument in the laboratory at a location convenient for operation, free from vibrations and protected against a corrosive atmosphere and contamination by chemicals.

2.1.4 Arrangement of the instruments

The 791 VA Detector can be placed on the 732 IC Detector or the 733 IC Separation Center.
791 VA Detector
7
Page 14
2 Installation

2.2 Mains connection

Follow the instructions below for connecting to the power supply. If the instrument is operated with a mains voltage set wrongly and/or wrong mains fuse, there is a danger of fire!

2.2.1 Setting the mains voltage

Before switching on the 791 VA Detector for the first time, check that the mains voltage set on the instrument (see Fig. 3) matches the local mains voltage. If this is not on the instrument as follows:
the case, you must reset the mains voltage
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 26 of the 791 VA Detector.
2 Remove fuse holder
Using a screwdriver, loosen fuse holder 12 and take out com­pletely.
3 Change mains voltage
Completely remove voltage selection insert 13 by hand, rotate it through 180° and reinsert it. The required mains voltage (115 or 230 V) must now be visible from the front.
4 Check fuses
Carefully take both fuses out of fuse holder 12 and check their specifications:
100240 V 1 A (slow-blow) Metrohm-No. U.600.0016
5 Insert fuses
Change both fuses if necessary and reinsert in fuse holder 12.
6 Install fuse holder
Push fuse holder 12 back into the opening of 791 VA Detector by hand until it clicks into place properly.
791 VA Detector
8
Page 15
2.2 Mains connection
100 – 120 V220 – 240 V
13
12
230
115
12 Fuse holder
13 Voltage selection
insert with display of voltage
25
26
Fig. 3
: Setting the mains voltage

2.2.2 Fuses

Two fuses 1 A/slow-blow for 100240 V are installed in the fuse holder 12 of the 791 VA Detector as standard.
Ensure that the instrument is never put into operation with fuses of another type, otherwise there is danger of fire!
For checking or changing fuses, proceed as described in section 2.2.1.

2.2.3 Mains cable and mains connection

Mains cable
The instrument is supplied with one of three mains cables
25 Mains switch
26 Mains connection
plug
6.2122.020 with plug SEV 12 (Switzerland, …)
6.2122.040 with plug CEE(7), VII (Germany, …)
6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a dif­ferent plug has to be fitted, the yellow/green lead (IEC standard) must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it a hazard!
Mains connection
Plug the mains cable into mains connection plug 26 of the 791 VA De­tector (see Fig. 3).

2.2.4 Switching the instrument on/off

The 791 VA Detector is switched on and off using mains switch 25. When the instrument is switched on lamp 1 lights up.
791 VA Detector
9
Page 16
2 Installation

2.3 656 Electrochemical detector

2.3.1 Installation and startup

Inserting the detector cell in shield housing, fitting the electrodes and startup of the 656 Electrochemical detector (order number 2.656.0020) are described in detail in the 656 Instructions for use.

2.3.2 Connection to 791 VA Detector

The electrodes used in the detector cell are each connected to the cor­responding connection of the 791 VA Detector with a 6.2120.020 Elec­trode cable as shown in Fig. 4. In addition the two earthing sockets are connected with a 6.2106.020 Cable.
6.2120.0 20 6.2106.0 20
656
AE WE RE
Fig. 4
AE RE WE
: Connection 656 – 791
791
791 VA Detector
10
Page 17

2.4 6.5303.030 ELCD cell for IC

2.4 6.5303.030 ELCD cell for IC

2.4.1 Assembling the ELCD cell

The reference electrode and working electrode must be inserted in the
6.5303.030 ELCD Cell supplied with the 2.791.0020 VA Detector. The cell is then screwed into the detector block built into the 733 IC Separa­tion Center. Proceed as follows (see Fig. 5 and Fig. 6):
1 Insert reference electrode
Reference electrode 39 (6.0727.000, details see section 3.2.2) included in the supply is already filled with c(KCl) = 3 mol/L and protected with a cap. It is assembled as follows (see Fig. 5):
Screw off the cap of electrolyte storage vessel 41 of reference
electrode 39. Normally PTFE gasket 31 remains in the cap and can be left there.
Insert one of the supplied PTFE gaskets 31 into the corre-
sponding opening of cell body 32 and screw reference elec­trode 39 into this opening.
2 Insert working electrode
Working electrode 28 is not supplied with the 791 VA Detector and must be ordered separately. Details about the working electrodes and information about pretreatment (e.g. polishing) can be found in section 3.2.1. A working electrode is assembled as follows (see Fig. 5):
Insert working electrode 28 in screw nipple 29.
Push one of the supplied PTFE gaskets 31 sufficiently far
along the working electrode from below.
Carefully insert working electrode 28 into the corresponding
opening of cell body 32 from above and press it in until it clearly clicks into position.
Push screw nipple 29 downwards and screw it in.
3 Connect ELCD cell to the detector block
The ELCD cell is screwed onto the 1.733.0X10 Detector block mounted inside the 733 Separation Center. Proceed as follows (see Fig. 5 and Fig. 6):
Screw cell body 32 onto detector block 42 with the help of
the two red knurled screws 43.
Push one of the supplied PTFE gaskets 31 over screw nipple
33 and screw this into the corresponding opening of cell body 32.
Place a screw nipple 36 and a PTFE gasket 35 on the outlet
end of the outlet capillary 37 permanently connected to de­tector block 42.
Insert outlet capillary 37 with screw nipple 36 in place into
screw nipple 33 and press in outlet capillary 37 until it reaches the stop.
791 VA Detector
11
Page 18
2 Installation
Screw nipple 36 onto screw nipple 33 to fix outlet capillary 37
is position.
4 Discharge
Pull outlet capillary 38 which is already mounted on cell body
32 through one of the openings in the rear panel of the 733 IC
Separation Center so that it emerges outside the instrument.
Insert outlet capillary 38 in a sufficiently large waste container
and fix it in position.
4139 3140
2731
28
29
30
31
35
36
38
: ELCD cell assembly
Fig. 5
27
ELCD cell (6.5303.030) 35 PTFE gasket (6.2704.020)
28 Working electrode 36 Screw nipple (4.420.2580)
29 Screw nipple (6.2730.000) 37 PEEK outlet capillary from
1.733.0X10 Detector block
32
31
33
34
35
36
37
30 Sealing ring (E.301.0028) 38 PTFE outlet capillary (6.1803.000)
31 PTFE gasket (6.2704.010) 39 Reference electrode (6.0727.000)
32 Cell body 40 Ag/AgCl Reference system
33 Screw nipple (6.2731.030) 41 Electrolyte vessel
34 Sealing ring (E.301.0004)
791 VA Detector
12
Page 19
2.4 6.5303.030 ELCD cell for IC
42
37
43
28
32
39
43
38
: Connection of ELCD cell at detector block
Fig. 6
Working electrode 39 Reference electrode
28
32 Cell body 42 Detector block (1.733.0X10)
37 PEEK outlet capillary from
1.733.0X10 Detector block
38 PTFE outlet capillary
(6.1803.000)

2.4.2 Connection to 791 VA Detector

The ELCD cell electrodes are each connected to the corresponding connection of the 791 VA Detector with a 6.2120.020 Electrode cable as shown in Fig. 7. The 791 VA Detector must be switched off and in the "stand-by" position when this is taking place.
(6.0727.000)
43 Knurled screw (V.900.4006)
791 VA Detector
6.2120.020
RE
Fig. 7
6.2120.020
6.2120.020
WE
AE
AE
AE RE WE
: Connection of ELCD cell at 791 VA Detector
791
13
Page 20
2 Installation

2.5 Analog output connections

The 791 VA Detector has two analog output sockets: 22 (live) and 23 (common) for connecting a chart recorder of a different data recording instrument. The full-scale deflection range for the analog output (1 V or 10 mV) can be set with switch 21; the polarity can be changed with switch 20 (see section 3.3).
The 6.2115.010 Cable can be used for connecting recorders with ba­nana plug connections; the 6.2115.060 Cable can be used for connect­ing data recording instruments with terminals.
791 VA Detector
14
Page 21
3.1 Electrochemical detection

3 Operation

3.1 Electrochemical detection

3.1.1 Classes of substances which can be determined

A precondition for the use of electrochemical detection is that the sub­stances to be determined are electrochemically active on the particular working electrode used, i.e. they can easily be oxidized or reduced. The electrical current produced by this reaction is proportional to the con­centration of the substance throughout a wide range. It is measured, amplified and recorded as a function of time by the 791 VA Detector in the form of a chromatogram.
The following table provides an overview of the classes of substances and ions which can be detected by oxidation or reduction. The ap­proximate polarization voltage obtained when an Ag/AgCl/c(KCl) = 3 mol/L reference electrode is used is given; this depends to a large ex­tent on the working electrode and eluents used (see section 3.2). The limits of detection which can be achieved are in the lower ppb range.
Structural formula Classes/substances Polarization voltage
R
OH
R
NH
2
5
3
N
H
Aromatic hydroxy compounds
Antioxidants Catecholes Flavones Halogenated phenols Hydroxybiphenyls Hydroxycumarines Methoxyphenols Oestrogens Phenols Tocopherols
Aromatic amines
Anilines Benzidines Sulfonamides
Indols
Indolyl-3-compounds 5-Hydroxy-indols
+800 +1200 mV
+ 800 mV + 1000 mV + 1200 mV + 800 mV + 1000 mV + 800 mV + 1000 mV + 1200 mV + 800 mV
+ 1000 mV + 600 mV + 1200 mV
+ 1000 mV + 800 mV
791 VA Detector
S
N
H
Phenothiazines + 1000 mV
15
Page 22
3 Operation
Structural formula Classes/substances Polarization voltage
RSH
O
O
R NO
R
R
Mercaptans + 800 mV
Miscellaneous
Ascorbic acid Carotenes Purine derivatives Vitamin A
Quinones
Vitamin K
1
+ 800 mV + 800 mV + 1000 mV + 1000 mV
– 700 mV
2
Nitro compounds
Nitrophenol Nitroglycerin Isosorbidedinitrate
Inorganic anions
Nitrite Sulfide Thiosulfate Chlorite Bromide Iodide Cyanide Thiosulfate Thiocyanate
– 700 mV – 800 mV – 800 mV
+ 1100 mV + 0 mV + 1100 mV + 1100 mV + 150 mV + 150 mV + 0 mV + 1100 mV + 1200 mV

3.1.2 Influence on detector signal

The sensitivity of detection, i.e. the height of the detector signal and the background current in the electrochemical detection depend on the fol­lowing parameters:
polarization voltage
temperature (approx. 1.5 %/°C)
flow rate of eluent
surface of electrodes and interior of cell
Eluent composition
(pH value, type and concentration of conductive salt, type and concentration of organic components, etc.)
Each change of one of these parameters will change the background current and the detector signal. Therefore, every parameter change needs a stabilization time to elapse; this is normally a few minutes for
791 VA Detector
16
Page 23
3.1 Electrochemical detection
current ranges > 5 nA. Please pay particular consideration to this on changing the eluent or the sample solution: the "same" eluent may have a different background current because eluents are never exactly the same (ionic strength, pH value, etc.!).

3.1.3 Selection of polarization voltage

The hydrodynamic voltammograms show the voltage limit from the point where a substance is oxidized or reduced. This voltage limit is generally higher than the voltage predicted by thermodynamics be­cause of inhibitions of the electrode reaction. Increasing voltage values give rise to increasing currents up to a maximum value, the diffusion­controlled limiting current.
In Fig. 8 for example, analytes a, b, and c are detected at voltage U whereas at voltage U
, component a is detected selectively. Therefore
1
the polarization voltage to be set has to be optimized for sensitivity on one hand and selectivity on the other.
The polarization voltage to be set also depends on the type and condi­tion of the working electrode (see section 3.5) as well as on the eluent. The latter could be illustrated by Fig. 8, saying that the curves are from one single analyte taken in 3 eluents with different pH values.
I
Diffusion­controlled limiting current
abc
U
1
Fig. 8
: Selection of polarization voltage
U
2
Background current
U
pol
;
2

3.1.4 Procedure for unknown substances

The hydrodynamic current/voltage curve of an analyte with unknown electrochemical characteristics is recorded as follows:
Short circuit the system, i.e. connect the injector directly (without
separating column) to the electrochemical detector.
Set current range of 791 VA Detector to 100 500 nA.
Set polarization voltage of 791 VA Detector to +1200 mV.
Inject analyte (20 100 ng).
791 VA Detector
17
Page 24
3 Operation
If no signal is observed, change the pH value of the eluent (a proto­nated or deprotonated substance may be oxidized more easily). Gen­erally, a higher polarization voltage has to be applied with lower pH val­ues (approx. 50 100 mV per pH unit).
st
1
experiment with eluent of pH 6
nd
2
experiment with eluent of pH 3
If there is still no signal, the analyte is "inactive", i.e. not suitable for elec­trochemical detection.
As soon as a signal is observed, decrease the polarization voltage in steps of 100 mV: a hydrodynamic current/voltage curve results.
For detection, select the lowest polarization voltage U good signal is still observed.

3.1.5 Practical information about ELCD

In principle the electrochemical detector can be connected to all HPLC and IC systems. In order to carry out work successfully, particularly in trace-level quantities, all components such as pump, dampening sys­tem, injector, separating column, capillaries, couplings, etc. must be in perfect condition. In electrochemical detection it is possible that inter­ferences may occur which, for example, are not observed with UV or conductivity detection. These can often be traced back to contaminants from the water supply system (e.g. Fe(II) ions), contaminants from the separating column (e.g. organic residues) or variations in pressure caused by leaking connections.
Particular attention must be given to the following points:

Cleanliness

The general demands placed on trace analysis work also apply here. Each possible source of corrosion of steel components must be avoided as Fe(II) ions may interfere. The eluent should also not contain any complexing agents.
at which a
pol

Pulsation

The noise of the detector signal depends on the pressure variation in­side the cell and on the back-pressure (max. 1 bar), which is deter­mined by the length of outlet capillary 38 (i.d. 0.3 mm). The use of a pulsation dampener (e.g. Metrohm 6.2620.150 Pulsation dampener) for dampening such pulsations has proved itself in practice and is also adequate if only single-piston pumps are used.

Eluent

Most of the solvents normally used in HPLC and IC can be used to pre­pare eluents. The conductivity which is necessary for detection is achieved by the addition of a conductive salt. Concentrations of 1…10 g/L are normally adequate for this purpose, this results in a conductivity of approx. 110 mS/cm. Sulfates, nitrates, phosphates, acetic acid,
791 VA Detector
18
Page 25
3.1 Electrochemical detection
sulfuric acid, perchloric acid, lithium perchlorate, etc. can be used for this. Chlorides and hydroxycarboxylic acids should not be used; some reagents which form ion pairs can be used.
Solvent gradients can be used in less sensitive ranges (current range > 100 nA). The eluent is degassed as usual by vacuum or passing a stream of helium through it.

Stationary phase

In principle all reversed-phase and ion exchanger materials can be used. New columns must be purged well and for a sufficiently long time for any contaminants which may be present to be washed out. This "equlibration phase" of a column may take more than 24 hours in the most sensitive areas.

Sample solution

Whenever possible, the eluent used should also be used as the solvent for the sample in order to minimize interference owing to changing sur­roundings (as small a front peak as possible). Very small amounts of the active substance should be used, in general < 100 ng.
791 VA Detector
19
Page 26
3 Operation

3.2 Electrodes

3.2.1 Working electrodes

The quality of a working electrode WE can be assessed in terms of the following criteria:
Background current (after adequate running-in time):
this should be small and constant
Interference level ("noise"):
this should be low
Response characteristics:
after switching on, the low background current should be quickly reached
It should nevertheless be borne in mind that increases in both back­ground current and noise levels are often due to other causes than a poor-quality working electrode (for example, contamination of the HPLC system, malfunction of the pulsation dampener, etc.). Generally speak­ing, both background current and noise levels tend to increase with higher polarization voltages, and are usually greater in eluents based on mixed solvents than in those composed of purely aqueous solutions.

Glassy carbon electrode GCE

Ordering designation 6.0805.010
Sensor material Glassy carbon, GC
Chemical resistance good resistance to acetonitrile, methanol, wa-
ter, etc.
Voltage range (–800) 0 +1200 mV (pH < 7)
Regeneration
Sprinkle a little aluminium oxide powder (Al on the polishing cloth (ordering number of both these accesso­ries: 6.2802.000) and moisten with distilled water. Hold the elec­trode vertically and polish with a circular motion for about 2060 s, exerting only light pressure. Rinse with distilled water and clean with dry cloth. A glassy carbon electrode should always be polished for a short while immediately before being fitted to the detector cell.
, grain size 0.3 µm)
2O3

Ultra-Trace graphite electrode

Ordering designations 6.1204.100 Ultra-Trace graphite tip
+ 6.2103.110 Contact pin
Sensor material Graphite (turnover is three times larger than in
GC-electrode; does not need tamping)
Chemical resistance good resistance to acetonitrile, methanol, wa-
ter, etc.
Voltage range –800 0 +1200 mV (pH < 7)
Regeneration
With 6.2827.000 Trimming tool (see electrode data sheet).
791 VA Detector
20
Page 27
3.2 Electrodes

Carbon paste electrode CPE

Ordering designation 6.0807.000
Sensor material paste composed of spectroscopic grade car-
bon powder and paraffin
Chemical resistance good resistance to aqueous solutions; limited
resistance only to water/acetonitrile and wa­ter/methanol mixtures
Voltage range (–200) 0 +1200 mV (pH < 7)
Regeneration
The sensor surface of the CPE is more vulnerable to both me­chanical damage and chemical attack than that of the GCE, it is more easily regenerated: often it suffices merely to rub the tip of the electrode backwards and forwards a few times against the surface of a sheet of ordinary, smooth paper, or alternatively the upper layer of carbon paste can be renewed (see below) and then polished in this manner.
Filling
Carefully remove the old carbon paste from the approximately
2.5 mm deep cavity with a suitable sharp instrument. Take care not to damage the platinum contact wire at the bottom of the cav­ity. Using a small spatula, fill bit by bit with the fresh 6.2801.020 carbon paste and tamp well down using the 6.2826.000 stopper. The cavity should be filled over the brim with paste. Now rub off the excess paste by continuous circular motion against a sheet of paper with a hard, smooth, non-porous surface; the circular mo­tion plus the light pressure will polish the surface of the paste, which must be smooth and shiny when the operation is finished. There should be no cracks in the peripheral zone where the paste is up against the plastic material of the electrode shaft.

Metal electrodes

Ordering designations 6.1204.120 Pt tip
6.1204.130 Ag tip
6.1204.140 Au tip + 6.2103.110 Contact pin
Chemical resistance good resistance to acetonitrile, methanol, wa-
ter, etc.
Voltage range Pt: (–1200) 0 +200 mV (pH < 7)
Ag: (–800) 0 +400 mV (pH > 7) Au: (–400) 0 +1200 mV (pH < 7)
Regeneration
Sprinkle a little aluminium oxide powder (Al on the polishing cloth (ordering number of both these accesso­ries: 6.2802.000) and moisten with distilled water. Hold the elec­trode vertically and polish with a circular motion for about 2060 s, exerting only light pressure. Rinse with distilled water and clean with dry cloth. A metal electrode should always be pol­ished for a short while immediately before being fitted to the de­tector cell.
, grain size 0.3 µm)
2O3
791 VA Detector
21
Page 28
3 Operation

Amalgamated electrode

"Home-made" amalgamated electrodes are prepared as follows:
Base material 6.1204.140 Au tip
Preparation Put a few drops of Hg on the front side of the
electrode and leave it for a few minutes. Knock off excess Hg and rinse electrode with distilled water.
Chemical resistance good resistance to acetonitrile, methanol, water Voltage range –1200 0 mV
Regeneration polish electrode with aluminium oxide and re-
amalgamate

3.2.2 Reference electrode

The reference system of the 6.0727.000 Reference electrode (RE) is Ag/AgCl/c(KCl) = 3 mol /L (see Fig. 5). The electrolyte vessel 41 can be refilled. The KCl solution has to be renewed every 12 months.
Ordering designations for KCl solutions: 6.2308.020 (250 mL)

3.2.3 Auxiliary electrode

A gold electrode is built-in in the detector cell as an auxiliary electrode.
6.2313.000 (1 L)
791 VA Detector
22
Page 29

3.3 Operating element functions

3.3 Operating element functions

Mains switch

The 791 VA Detector is switched on and off using mains switch 25 on the rear of the instrument (see Fig. 2):
I Instrument switched on 0 Instrument switched off
~
After the instrument has been switched on, the mains pilot lamp 1 lights up to show that the instrument is ready for use.

Relative compensation current

rel. is set using the
comp
I
comp
The relative compensation current I potentiometer 10 (see Fig. 1). The value set is related to the setting of the adjusting knob 8 (current I).
.
I0.01
Setting range –500 (red) +500 (white) × 0.01 x I
Resolution 0.01 x I per scale division

Absolute compensation current

The absolute compensation current I potentiometer 9 (see Fig. 1). The value set is not related to the setting of the adjusting knob 8 (current I).
abs. is set using the
comp
I
comp
0.02 nA
Setting range
direct 0 … 1000 × 0.02 nA = 20 nA with commutator 5 5 × 20 nA = 100 nA with commutator 6 50 × 20 nA = 1000 nA
Resolution
direct 0.02 nA per scale division with commutator 5 0.10 nA per scale division with commutator 6 1.00 nA per scale division
The polarity of the absolute compensation current is set sepa­rately with commutator 7.

Changeover Stand-by/Measure

meas
For this changeover commutator 3 (see Fig. 1) is used.
Button depressed
Measuring position, i.e. measuring cell connected.
Button not depressed
Stand-by position, i.e. measuring cell disconnected.
791 VA Detector
23
Page 30
3 Operation

Damping

damp
The damping is switched on and off using commutator 4 (see Fig. 1).
Button depressed
Signal damped for all ranges set on adjusting knob 8 in or­der to suppress interference (spurious signals).
Button not depressed
No signal damping (exception: the most sensitive range,
0.1 nA, is always damped to some extent).
x50
x5
Multiplication of I
For this changeover commutator 5 (see Fig. 1) is used.
Button depressed
abs. by a factor of 5
comp
Multiplication of the absolute compensation current by a factor of 5.
Button not depressed
No multiplication of the absolute compensation current.
Multiplication of I
abs. by a factor of 50
comp
For this changeover commutator 6 (see Fig. 1) is used.
Button depressed
Multiplication of the absolute compensation current by a factor of 50.
Button not depressed
No multiplication of the absolute compensation current.
+

Polarity of I

For this changeover commutator 7 (see Fig. 1) is used.
Button depressed
comp
abs.
All current values set on the potentiometer 9 have a positive sign.
Button not depressed
All current values set on the potentiometer 9 have a nega­tive sign.
791 VA Detector
24
Page 31
3.3 Operating element functions
overload

Overload display

If the electronic current measuring circuit is overloaded, the red light 2 is lit up (see Fig. 1).
Remedies
Select a less sensitive current range with adjusting knob 8 or compensate current with potentiometer 9 and/or 10 (for further measures see section 4.3).

Polarization voltage

is set using the digital switch 11 (see
pol
+
015
U
10 mV
The detection voltage U Fig. 1).
Setting range –1990 0 +1990 mV
pol
Resolution 10 mV

Current measuring sensitivity

nA
50
10
5
1
0,5 100
0,1
(U
out
I
= 1 V)
µ
A
The sensitivity of current measurement is set using adjusting
0,1
knob 8 (see Fig. 1). The set range applies to the full-scale
0,5
deflection of the analog signal at sockets 22 and 23 (1 V or
1
5
10 mV depending on the setting of switch 21).
4
10
Setting ranges 0.1, 0.5, 1, 5, 10, 50 nA
0.1, 0.5, 1, 5, 100, 10'000 µA

Polarity of analog output

The polarity (+ or –) of the analog output signal is set using switch 20 (see Fig. 2).

Full-scale deflection of analog output

0...1 V
0...10 mV
The full-scale deflection (1 V or 10 mV) of the analog output signal is set using switch 21 (see Fig. 2).
791 VA Detector
25
Page 32
3 Operation

3.4 Startup

3.4.1 Preparing the ELCD cell

The ELCD cell must be correctly connected to the HPLC or IC system before it is filled (see section 2.3 and section 2.4). Then proceed as fol­lows:
1 Fill reference electrode (only if necessary)
Reference electrode 39 (6.0727.000) is supplied in a filled condition. However, the KCl electrolyte solution must be re­placed every 1 to 2 months. Proceed as follows:
Screw Ag/AgCl reference system 40 out of electrolyte vessel
41 (see Fig. 5).
Fill electrolyte vessel 41 to the brim with c(KCl) = 3 mol/L.
Screw Ag/AgCl reference system 40 back onto electrolyte
vessel 41. Take care that no air bubbles are found in the elec­trolyte vessel (to check: hold reference electrode against the light).
2 Connect reference electrode
Insert reference electrode 39 into cell body 32 (see Fig. 5).
Connect reference electrode 39 to connection 17 of the 791
VA Detector with 6.2120.020 Cable (see Fig. 7).
3 Connect auxiliary electrode
Connect the auxiliary electrode in cell body 32 to connection
15 of the 791 VA Detector 791 with 6.2120.020 Cable (see
Fig. 7).
4 Prepare working electrode
Prepare and polish working electrode 28 according to section
3.2.1.
Insert working electrode 28 into cell body 32 (see Fig. 5).
Do not
at this point.
5 Start pump
Switch on delivery drive of HPLC or IC pump.
connect working electrode 28 to the 791 VA Detector
791 VA Detector
26
Page 33
3.4 Startup

3.4.2 Zero balancing and equilibration phase

Before the first measurement is made the working electrode must be al­lowed to equilibrate and the chart recorder or data recording instrument must be zero-balanced. Proceed as follows:
1 Chart recorder settings
If a chart recorder is used to record the analog signals then proceed as follows (the working electrode must not nected to the 791 VA Detector):
Switch on 791 VA Detector.
Set commutator 3 on the 791 VA Detector to "Stand-by".
Set the required polarity (+ or ) with switch 20; use switch
21 to set the required voltage range for the analog output on the 791 VA Detector (1 V or 10 mV).
Set potentiometer 9 on the 791 VA Detector to 0.
Use knob 8 on the 791 VA Detector to set the current sensitiv- ity to 0.5 µA.
Switch on recorder.
Set the voltage measuring range on the recorder to the same
setting as on the 791 VA Detector (1 V or 10 mV).
Use potentiometer 10 on the 791 VA Detector to adjust the zero point of the recorder.
be con-
2 Settings with a data recording program
If a PC recording and evaluation program is used for recording the analog signals then proceed as follows (the working elec­trode must not
be connected to the 791 VA Detector):
Switch on 791 VA Detector.
Set commutator 3 on the 791 VA Detector to "Stand-by".
Set the required polarity (+ or ) with switch 20; use switch
21 to set the required voltage range for the analog output on the 791 VA Detector (1 V or 10 mV).
Set potentiometers 9 and 10 on the 791 VA Detector to 0.
Use knob 8 on the 791 VA Detector to set the current sensitiv- ity to 0.5 µA.
Switch on the data recording program on the PC and start
recording data.
If required, use potentiometer 10 on the 791 VA Detector to adjust the zero point (so that the mV display on the PC shows
0).
791 VA Detector
27
Page 34
3 Operation
3 Overload test
For this test the working electrode must not be connected to the 791 VA Detector:
Set the polarization voltage on the 791 VA Detector to the
required value for the following determinations with digital switch 11.
Set commutator 3 on the 791 VA Detector to "meas". Over-
load display 2 must light up briefly and then go out again.
If overload display 2 remains lit up for a longer period of time
this indicates that there is a faulty electrical contact between the reference electrode and the auxiliary electrode. Eliminate the cause (e.g. air in the system or in the reference electrode, blocked diaphragm, faulty cable connection) and repeat the test.
If the overload test is OK, set commutator 3 on the 791 VA
Detector back to "Stand-by".
4 Equilibrating the working electrode
Connect working electrode 28 to connection 18 of the 791 VA
Detector with the 6.2120.020 Cable.
Set commutator 3 on the 791 VA Detector to "meas".
Observe equilibration curve: the curve should flatten out
noticeably within a relative short time (1020 min), otherwise the working electrode must be repolished or refilled.
In current measuring ranges > 5 nA a constant base current
is usually achieved with one hour. At higher sensitivities the working electrode should be equilibrated overnight.
5 Zero balancing the base current
(only for recorders)
Set the current measuring sensitivity on the 791 VA Detector
with knob 8 to the required value.
Use potentiometer 9 on the 791 VA Detector to adjust the
base current to the zero point of the recorder. If the setting range of potentiometer 9 is not sufficient then it can be ex­tended by using commutator 5 (× 5) or 6 (× 50).
791 VA Detector
28
Page 35
3.4 Startup

3.4.3 Measuring procedure

In order to be able to carry out measurements it is essential that the working electrode has achieved a stable base current (see section
3.4.2). If the value of the base current regains its original value after the determination then as many determinations as required can be carried out. Attention should be given to the following points:
1 Carrying out the determination
Inject the sample solution into the HPLC or IC system.
Record and evaluate the current curve.
2 Allow ELCD system to run overnight
If the ELCD system is to be used again on the following day then the 791 VA Detector should not be switched off over­night. The working electrode is constantly under the polariza­tion current and is quickly ready for new measurements.
The HPLC or IC pump is set to a minimal flow rate and runs overnight. This means that the base current is stabilized after only a few minutes.
3 Switching off the ELCD system
Set commutator 3 on the 791 VA Detector to "Stand-by".
Unplug working electrode 28.
Switch off the 791 VA Detector.
791 VA Detector
29
Page 36
3 Operation
791 VA Detector
30
Page 37

4.1 Maintenance and servicing

4 Maintenance – Malfunctions

4.1 Maintenance and servicing

Care

The 791 VA Detector requires proper care and attention. Excessive con­tamination of the instrument could possibly lead to malfunctions and a shorter service life of the inherently rugged mechanical and electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is especially important to protect the plug connections at the rear of the instrument (particular the mains plug) against contamination.
Although constructional measures have been designed to virtually eliminate such a situation, should corrosive media penetrate the interior of the instrument the mains plug of the 791 VA Detector must be immediately disconnected to prevent extensive damage to the instrument electronics. Inform Metrohm service if your instrument have been damaged in such a way.
The instrument must not be opened by untrained personnel. Please comply with the safety notes in section 1.4.1.

Maintenance by Metrohm service

Maintenance of the 791 VA Detector is best done as part of an annual service performed by specialists from the Metrohm company. If work is frequently performed with caustic and corrosive chemicals, it may be necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice on the maintenance and servicing of all Metrohm instruments.

4.2 Shutdown

If the ELCD cell is shut down for a considerable length of time, the en­tire HPLC or IC system (without rinsed free from salt with methanol/water (1:4) to avoid crystallization of eluent salts with the corresponding subsequent damage.
column and suppressor) must be
791 VA Detector
31
Page 38
4 Maintenance – Malfunctions

4.3 Malfunctions and their rectification

Malfunction Cause Rectification
No signal • VA Detector is in "stand-by"
mode.
VA Detector switched off or
disconnected from mains voltage.
Recorder switched off or
disconnected from mains voltage.
Recorder disconnected from
VA Detector.
Electrodes disconnected.
Broken interconnecting
cables.
VA Detector shows "overload"
Current range set exceeded
by too high a concentration of substance.
The VA Detector is either
incorrectly balanced or not balanced at all.
Reference electrode discon-
nected.
Auxiliary electrode discon-
nected.
Polarization voltage badly
adjusted.
Press "meas" button on VA
Detector.
Connect VA Detector and
switch on.
Connect recorder and switch
on.
Connect recorder to the VA
Detector.
Connect electrodes.
Use new interconnecting
cables.
Wait or set a higher current
range at the VA Detector.
Balance the recording instru-
ment and VA Detector correctly.
Connect reference electrode.
Connect auxiliary electrode.
Adjust polarization voltage
correctly.
Excessively high background current
Baseline unstable
Polarization voltage too high.
Eluent contaminated.
Eluent contaminated by Fe(II)
ions from rusty steel parts (traces only suffice!)
Highly retarded or enriched
components from the separat­ing column.
Pressure fluctuations due to
poor seals, blocked filters or capillaries, defective pump, etc.
Eluent contaminated by Fe(II)
ions from rusty steel parts (traces only suffice!).
Defective reference electrode.
Polarization voltage should
generally not be set higher than 1200 mV.
Prepare fresh eluent using only
the purest (AR grad) chemicals.
Check entire HPLC or IC system
for rust; check resistance of all materials to the reagents in use.
Rinse the column well with
eluent or regenerate it, use new column if necessary.
Check HPLC or IC system for
these faults.
Check entire HPLC or IC system
for rust; check resistance of all materials to the reagents in use.
Replace reference electrode.
791 VA Detector
32
Page 39
4.3 Malfunctions and their rectification
Malfunction Cause Rectification
Noise level too high
Spikes on baseline
Inadequate pulsation
damping.
Electrostatic screening
defective.
Background current too high.
Eluent outlet tubing under
electrostatic influence.
Defective working electrode.
Defective 791 VA Detector.
Air in detector cell.
Air entering detector cell.
If a CPE working electrode is
in use, the carbon paste is not in good condition.
Use pulsation dampener.
Check screening, if necessary
connect steel column and/or its outlet to socket 14 of the VA Detector by means of a cable.
Check screening, if necessary
connect steel column and/or its outlet to socket 14 of the VA Detector by means of a cable.
Immerse eluent outlet tubing
completely in spent solution in waste recipient or renew PTFE tubing if necessary.
Replace working electrode.
Check 791 VA Detector using
the dummy cell (see section
4.1.4)
Unscrew working electrode and
deaerate detector cell.
Degas eluent or subject it to
appropriate treatment.
Refill CPE and ensure paste is
well tamped down.
Isolated spikes on baseline
Substance turnover too low or non­existent
Substance turnover falls off
Front peak too high Environment altered by
Electrostatic discharges from
operating personnel on the apparatus.
Mains interference.
Electromagnetic interference
from environment.
Polarization voltage too low.
Active surface of working
electrode blocked by sub­stances.
Active surface of working
electrode blocked by sub­stances.
Quantity of substance too
great.
sample solution.
Eluent contaminated by Fe(II)
ions from rusty steel parts (traces only suffice!).
Such discharges cannot be
completely avoided, but are not usually too troublesome at normal (i.e. not too dry) atmos­pheric humidity.
Connect all instruments to the
power supply via a suitable filter (e.g. Metrohm 615 Mains Dis­tributor).
Change "environment".
Increase polarization voltage.
Polish working electrode or
renew paste in CPE.
Polish working electrode or
renew paste in CPE.
The sample quantity should be
less than 100 ng per injection to avoid blocking of the electrode surface.
Whenever possible, use eluent
as solvent for the sample solu­tion.
Check entire HPLC or IC system
for rust; check resistance of all materials to the reagents in use.
791 VA Detector
33
Page 40
4 Maintenance – Malfunctions

4.4 Instrument test with the dummy cell

The correct functioning of the 791 VA Detector can be checked with the included 6.2813.020 Dummy cell. Several simple checks can be carried out with this extremely simplified electrical simulation of a detector cell.
Fig. 9 shows the construction of the dummy cell. The checks described below can be carried out with the recording instrument which is con­nected (chart recorder, integrator, PC data recording program) or, for precise measurements, with a voltmeter which is connected to analog outputs 22 and 23. The dummy cell is connected to the corresponding cable instead of the electrode. Each test can be carried out separately. The settings of any operating elements which are not mentioned are ir­relevant.
AE
10 M
1%
10
1 µF
RE
Fig. 9
: Circuit of 6.2813.020 Dummy Cell
WE

4.4.1 Check of "meas" and "damp" commutators and of current amplifier offset

1 Set digital switch 11 U
to +1000 mV.
pol
+
100
2 Set adjusting knob 8 to 0.1 µA.
3 Set potentiometer 9 I
791 VA Detector
34
abs. to 0.
comp
50
10
5
1
0,5 100
0,1
0.02 nA
I
comp
0,1
0,5
1
5
4
10
Page 41
4.4 Instrument test with the dummy cell
4 Set all commutators 3 7 to "off" position
(non-depressed position).
5 Set selector 20 on the rear panel to +.
6 Set selector 21 on the rear panel to 010 mV.
7 Bring recorder of voltmeter to 0 using potenti-
ometer 10 I
rel.. The value shown on the
comp
knob scale must not vary from 0 by more than 120 scale divisions.
8 Depress commutator 3 "meas": The recorder or
voltmeter should read –10 mV.
9 Depress commutator 4 "damp" and wait until
recorder of voltmeter have attained a stable position.
0...1 V
0...10 mV
I
comp
meas
damp
.
I0.01
10 Set commutator 3 "meas" to "off" position: The
time constant τ to be measured, i.e. the time which has to elapse until 63.2 % of full-scale deflection is reached, should have a value of app. 3 s (consider τ of recorder or voltmeter).

4.4.2 Check output voltage

1 Set digital switch 11 U
2 Set adjusting knob 8 to 0.1 µA.
to 0 mV.
pol
meas
+
000
50
10
5
1
0,5 100
0,1
0,1
0,5
1
5
4
10
791 VA Detector
3 Set commutators 4 7 to "off" position (non-
depressed position).
35
Page 42
4 Maintenance – Malfunctions
4 Set selector 20 on the rear panel to –.
5 Set selector 21 on the rear panel to 01 V.
6 Set recorder sensitivity to 1000 mV.
7 Set commutator 3 "meas" to "on" position
(depressed position).
8 Set potentiometer 9 I
abs. to 0.
comp
9 Bring recorder of voltmeter to 0 using potenti-
ometer 10 I
10 Alter the settings of digital switch 11 U
comp
rel..
pol
in
steps of 100 mV and 10 mV: The settings shown on the digital switch should be repro­duced on the recorder or voltmeter within a tolerance of <12.5 mV and <1.25 mV respec­tively.
0...1 V
0...10 mV
meas
I
comp
0.02 nA
+
I
comp
.
I0.01
140
11 Set digital switch 11 U
to – and alter the
pol
settings in steps of 100 mV and 10 mV: The settings shown on the digital switch should be reproduced on the recorder or voltmeter within a tolerance of <12.5 mV and <1.25 mV respec­tively.

4.4.3 Check of current compensation

1 Set digital switch 11 U
2 Set adjusting knob 8 to 0.1 µA.
to 0 mV.
pol
250
+
000
50
10
5
1
0,5 100
0,1
0,1
0,5
1
5
4
10
791 VA Detector
36
Page 43
4.4 Instrument test with the dummy cell
3 Set selector 20 on the rear panel to –.
4 Set selector 21 on the rear panel to 01 V.
5 Set commutator 3 "meas" to "on" position
(depressed position).
6 Set potentiometer 9 I
abs. to 0.
comp
7 Bring recorder of voltmeter to 0 using potenti-
ometer 10 I
8 If potentiometer 10 I
comp
rel..
rel. is now moved
comp
counterclockwise by 100 divisions (= 1 turn) then this should correspond to 1000 mV on the recording instrument or voltmeter.
0...10 mV
0.02 nA
0...1 V
meas
I
comp
I
comp
I
comp
.
I0.01
.
I0.01
9 If potentiometer 9 I
abs. is now moved
comp
clockwise by 100 divisions (= 1 turn) then this should correspond to the following values on the recording instrument or voltmeter:
direct: +20 mV
×5 button 5 depressed: +100 mV ×50 button 6 depressed: +1000 mV
10 Set commutator 7 (+) to "on" position
(depressed position).
11 Similarly check the functioning of potentiometer
10 I
rel. according to items 79.
comp
I
comp
0.02 nA
I
comp
+
.
I0.01
791 VA Detector
37
Page 44
4 Maintenance – Malfunctions

4.4.4 Check of current and voltage overload

1 Set digital switch 11 U
to +10 mV or
pol
–10 mV.
2 Set adjusting knob 8 to 50 nA.
3 Set selector 20 on the rear panel to –.
4 Set selector 21 on the rear panel to 01 V.
5 Set potentiometer 9 I
abs. to 0.
comp
+
001
50
10
5
1
0,5 100
0,1
0...1 V
0...10 mV
I
comp
0.02 nA
0,1
0,5
1
5
4
10
6 Set commutators 3 7 to "off" position (non-
depressed position).
7 Bring recorder of voltmeter to 0 using potenti-
ometer 10 I
comp
rel..
8 Set commutator 6 "×50" to "on" position
(depressed position).
9 Turn potentiometer 9 I
abs. clockwise until
comp
overload display 2 lights up: The voltage at the recorder output should be app. 10.512 V. Without digital voltmeter, read the value on the scale of potentiometer 9: The value should be between 500 and 620 scale divisions.
10 Reset potentiometer 9 I
abs. to 0: The
comp
overload display 2 extinguishes.
I
comp
.
I0.01
x50
I
comp
0.02 nA
0.02 nA
791 VA Detector
38
I
comp
Page 45
4.4 Instrument test with the dummy cell
11 Set commutator 3 "meas" to "on" position
(depressed position).
12 Unplug dummy cell at "AE": the overload
display 2 lights up.

4.4.5 Check sensitivity knob 8

1 Unplug dummy cell.
2 Set adjusting knob 8 to 50 nA.
3 Set commutators 3 7 to "off" position (non-
depressed position).
meas
50
10
5
1
0,5 100
0,1
0,1
0,5
1
5
4
10
4 Set potentiometer 9 I
abs. to 0.
comp
5 Set selector 21 on the rear panel to 01 V.
6 Bring recorder of voltmeter to 0 using potenti-
ometer 10 I
rel. (set line recorder first to mid-
comp
scale).
7 Turn adjusting knob 8 through all positions and
check the deviation from zero after a waiting time of app. 5 s. Deviation should be < ±1 V.
I
comp
0.02 nA
0...1 V
0...10 mV
I
comp
.
50
10
5
1
0,5 100
0,1
I0.01
0,1
0,5
1
5
4
10
791 VA Detector
39
Page 46
4 Maintenance – Malfunctions

4.4.6 Check of noise from current amplifier

1 Connect dummy cell.
2 Set digital switch 11 U
to 0 V.
pol
3 Set adjusting knob 8 to 1 nA.
4 Set commutator 3 "meas" to "on" position
(depressed position).
5 Set commutators 4 7 to "odd" position
(non-depressed position).
6 Set selector 21 on the rear panel to 01 V.
+
000
50
10
5
1
0,5 100
0,1
meas
0...1 V
0...10 mV
0,1
0,5
1
5
4
10
7 Allow the recording instrument to record for a
longer period of time (first adjust chart recorder to middle of the scale). Typical deflections caused by noise are about 10 mVpp or less (the 656 housing must be earthed).
791 VA Detector
40
Page 47
5.1 Technical data

5 Appendix

5.1 Technical data

Measuring instrument
Measuring technique Direct current amperometry
Polarization voltage U
Current measurement Measurement: direct, not integrated
Damping Switchable
Compensation current I
absolute
Range: –1990 … 0 … +1990 mV
pol
Resolution: 10 mV Deviation: ± 1 % ± 1 mV
Sensitivity: 0.1, 0.5, 1, 5, 10, 50 nA
0.1, 0.5, 1, 5, 100, 10'000 μA Deviation: ± 15 % for 0.1 nA 5 nA ± 1.5 % for 10 nA 0.1 μA ± 1.0 % for 0.5 100 μA ± 4.0 % for 10'000 μA Time constants: 5.0 s for < 0.1 nA 0.6 s for 0.1 ... 0.5 nA 0.3 s for 0.5 ... 1 nA 0.1 s for 1 ... 5 nA
Time constant: 1.7 s
comp
Basic range: –20 0 +20 nA Resolution: 0.02 nA Multiplication: 5 ×, 50 ×
relative
Potentiostat Input resistance: 1012 Ω (typ. value)
Analog output
Signal U = f ( I )
Polarity Switchable
Max. load < 5 mA
Input resistance at connected instrument
Range: –5.00 … 0 … +5.00· I Resolution: 0.01· I
Input current: < 5 pA (typ. value) Zero-point drift: 20 μV/K (typ. value) Output voltage: ± 10 V Output current: 10 mA
1 V or 10 mV DC for the full current range I
> 2 kΩ
791 VA Detector
41
Page 48
5 Appendix
Mains connection
Voltage 115 V: 100...120 V ± 10 %
230 V: 220...240 V ± 10 % Switchable with voltage selector insert in the fuse holder (see section 2.2.1)
Frequency 50...60 Hz
Power consumption 5 VA
Fuse 5 mm dia., 20 mm length
100240 V: 1 A (slow-blow)
Safety specifications
Construction/testing According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions The Instructions for Use include information and
warnings which must be heeded by the user to assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference Standards met:
EN55011 (class B), EN55022 (class B), EN 50081-1/2
Immunity to interference Standards met:
IEC801-2/IEC1000-4-2 (class 3), IEC801-4/ IEC1000-4-4 (class 3), IEC1000-3/EN61000-3, IEC1000-4-11/EN61000-4-11
Ambient temperature
Nominal operating range +5+40°C
(at 2080 % atmospheric humidity)
Storage, transport –40+70°C
Housing
Material of cover Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, CFC-free
Material of base Steel, enameled
Dimensions
Width 260 mm
Height 129 mm
Depth 366 mm
Weight 3.6 kg (with accessories)
791 VA Detector
42
Page 49
5.2 Standard equipment
59

5.2 Standard equipment

Subject to changes ! All dimensions are given in mm.

5.2.1 2.791.0010 VA Detector for HPLC

The 2.791.0010 VA Detector includes the following parts:
Quant. Order No. Description
1 6.2105.030 Cable
Connection cable 791 VA Detector (analog output) – recorder
1 6.2122.0X0 Mains cable
to customer's specifications:
Cable socket Cable connector
Type IEC 320/C 13 Type SEV 12 (CH…)............................... 6.2122.020
Type IEC 320/C 13 Type CEE (7), VII (D…) .......................... 6.2122.040
Type CEE (22), V Type NEMA 5-15 (USA…)...................... 6.2122.070
1 6.2813.020 Dummy Cell
Electrical simulation of the ELCD cell for testing and checking functions
1 8.791.1013 Instruction for Use (English)
for 791 VA Detector
1.2 m
791 VA Detector
43
Page 50
5 Appendix
67
59

5.2.2 2.791.0020 VA Detector for IC

The 2.791.0020 VA Detector includes the following parts:
Quant. Order No. Description
1 6.0727.000 Reference electrode
Mini reference electrode for ELCD cell with Ag/AgCl/c(KCl) = 3 mol/L reference system; including cap
3 6.2120.020 Electrode cable
Electrode cable with mini socket for connection pins 2 mm and mini plug type F
1 6.2122.0X0 Mains cable
to customer's specifications:
Cable socket Cable connector
Type IEC 320/C 13 Type SEV 12 (CH…) ...............................6.2122.020
Type IEC 320/C 13 Type CEE (7), VII (D…) ...........................6.2122.040
Type CEE (22), V Type NEMA 5-15 (USA…).......................6.2122.070
1 6.2617.000 Tool for PTFE gasket
To remove 6.2704.010 and 6.2704.020 PTFE gaskets
∅ 18
83
2 m
4
67
4 6.2704.010 PTFE gasket
Gasket for reference electrode, working electrode and eluent inlet
3 6.2704.020 PTFE gasket
Gasket for eluent inlet
1 6.2813.020 Dummy Cell
Electrical simulation of the ELCD cell for testing and checking functions
1 6.5303.030 ELCD cell
1 8.791.1013 Instruction for Use (English)
for 791 VA Detector
4.3
12
6
4.5
791 VA Detector
44
Page 51
5.3 Optional accessories
M3

5.3 Optional accessories

Order No. Description
6.0805.010 Mini Glassy Carbon electrode Diameter of active zone: 2.8 mm.
6.0807.000 Mini Carbon paste electrode Diameter of active zone: 3 mm.
6.1204.XXX Electrode tips for working electrode
Together with 6.2103.110 Contact pin forms the working electrode. The following electrode tips are available:
Order No. Disk material Shaft material
6.1204.100 Ultra Trace Graphite PVC
6.1204.120 Pt PEEK
6.1204.130 Ag PEEK
6.1204.140 Au PEEK
Disk diameter: 2.0 +0 / –0.05 mm
6.2103.110 Contact pin
Together with 6.1204.1X0 Electrode tip forms the working electrode.
7
45
7
45
52.5
21.3
7
6.2801.020 Carbon paste
Carbon paste for 6.0807.000 Mini carbon paste electrode
6.2826.000 Filling tool
Filling tool for 6.0807.000 Mini carbon paste electrode
20
83
791 VA Detector
45
Page 52
5 Appendix

5.4 Warranty and conformity

5.4.1 Warranty

The warranty on our products is limited to defects that are traceable to material, construction or manufacturing error which occur within 12 months from the day of delivery. In this case, the defects will be recti­fied in our workshops free of charge. Transport costs are to be paid by the customer.
For day and night operation, the warranty is limited to 6 months.
Glass breakage in the case of electrodes or other parts is not covered by the warranty. Checks which are not a result of material or manufac­turing faults are also charged during the warranty period. For parts of outside manufacture insofar as these constitute an appreciable part of our instrument, the warranty stipulations of the manufacturer in question apply.
With the regard to the guarantee of accuracy, the technical specifica­tions in the instruction manual are authoritative.
Concerning defects in material, construction or design as well as the absence of guaranteed features, the orderer has no rights or claims ex­cept those mentioned above.
If damage of the packaging is evident on receipt of a consignment or if the goods show signs of transport damage after unpacking, the carrier must be informed immediately and a written damage report demanded. lack of an official damage report releases Metrohm from any liability to pay compensation.
If any instruments and parts have to be returned, the original packaging should be used if at all possible. This applies above all to instruments, electrodes, burette cylinders and PTFE pistons. Before embedment in wood shavings or similar material, the parts must be packed in a dust­proof package (for instruments, use of a plastic bag is imperative). If open assemblies are enclosed in the scope of delivery that are sensi­tive to electromagnetic voltages (e.g. data interfaces etc.) these must be returned in the associated original protective packaging (e.g. con­ductive protective bag). (Exception: assemblies with built-in voltage source belong in a non-conductive protective packaging). For damage which arises as a result of non-compliance with these instructions, no warranty responsibility whatsoever will be accepted by Metrohm.
791 VA Detector
46
Page 53
5.4 Warranty and conformity

5.4.2 EU Declaration of conformity

EU Declaration of Conformity

The METROHM AG company, Herisau, Switzerland hereby certifies, that the in­strument:
791 VA Detector
meets the requirements of EC Directives 89/336/EEC and 73/23/EEC.
Source of the specifications:
EN 50081-1/2 Electromagnetic compatibility, basic specification;
Emitted Interference
EN 61010 Safety requirements for electrical laboratory measurement
and control equipment
Description of the instrument:
High-sensitivity current measuring instrument for electrochemical detection
in HPLC and IC.
Herisau, September 11, 1998
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
791 VA Detector
47
Page 54
5 Appendix

5.4.3 Certificate of conformity and system validation

Certificate of Conformity and System Validation
This is to certify the conformity to the standard specifications for electrical appli­ances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
Name of commodity: 791 VA Detector
Name of manufacturer: Metrohm Ltd., Herisau, Switzerland
Principal technical information: Voltages: 100…120, 220…240 V Frequency: 5060 Hz
This Metrohm instrument has been built and has undergone final type testing according to the standards:
IEC801-2/IEC1000-4-2 (class 3), IEC801-4/IEC1000-4-4 (class 3), IEC1000-3/EN 61000-3, IEC1000-4-11/EN 61000-4-11, EN55011 (class B), EN55022 (class B), EN50081-1/2 Electromagnetic compatibility
IEC1010, EN61010, UL3101-1 — Security specifications
The technical specifications are documented in the instruction manual.
Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for quality assurance in design/development, production, installation and servicing.
Herisau, September 11, 1998
Dr. J. Frank Ch. Buchmann
Development Manager Production and Quality Assurance Manager
791 VA Detector
48
Page 55
5.5 Index

5.5 Index

A
Absolute compensation current........... 23
Adjusting knob 8
Figure...................................................2
Operation.......................................... 25
Test ................................................... 39
Ag electrode ......................................... 21
Amalgamated electrode....................... 22
Ambient temperature............................ 42
Analog output 22
Connection ....................................... 14
Figure...................................................3
Technical data .................................. 41
Analog output 23
Connection ....................................... 14
Figure...................................................3
Technical data .................................. 41
Appendix............................................... 41
Arrangement of the instruments..............7
Au electrode .................................... 21,22
Auxiliary electrode
Connect to 791................................. 26
Notes ................................................ 22
B
Back pressure ...................................... 18
Background current ........................ 20,32
Base current ......................................... 29
Baseline unstable ................................. 32
C
Cable (6.2105.030)............................... 43
Cable (6.2106.020)............................... 10
Cable (6.2115.010)............................... 14
Cable (6.2115.060)............................... 14
Cable (6.2120.020)..................... 10,26,44
Carbon paste........................................ 45
Carbon paste electrode .................. 21,45
Care ...................................................... 31
Caution ....................................................5
Cell body 32
Assembly .......................................... 11
Figure........................................... 12,13
Certificate of conformity
and system validation ...................... 48
Changeover Stand-by/Measure........... 23
Chart recorder ................................. 14,27
Check ......................................................7
Classes of substances ......................... 15
Classes of substances
which can be determined................. 15
Cleanliness ........................................... 18
Comment.................................................5
Commutator 3
Figure...................................................2
Operation.......................................... 23
Test ................................................... 35
Commutator 4
Figure...................................................2
Operation.......................................... 24
Test ................................................... 35
Commutator 5
Figure ..................................................2
Operation ..........................................24
Commutator 6
Figure ..................................................2
Operation ..........................................24
Commutator 7
Figure ..................................................2
Operation ..........................................24
Compensation current ..........................41
Conformity.............................................47
Connection 15 "AE"
Connecting the
auxiliary electrode .............................13
Figure ..................................................3
Connection 17 "RE"
Connecting the
reference electrode ......................13,26
Figure ..................................................3
Connection 18 "WE"
Connecting the working electrode.... 13
Connection of working electrode......28
Figure ..................................................3
Connection 19
Figure ..................................................3
Connection 656 – 791...........................10
Contact pin (6.2103.110) ......................45
Current amplifier offset.......................... 34
Current compensation ..........................36
Current measurement ........................... 41
Current measuring sensitivity................25
D
Damping...........................................24,41
Data recording instrument ....................27
Declaration of conformity ......................47
Degree of protection ..........................6,42
Detector block 42
ELCD cell assembly ..........................11
Figure ................................................13
Detector signal ...................................... 16
Diffusion-controlled limiting current ...... 17
Digital switch 11
Figure ..................................................2
Dimensions ...........................................42
Disposal...................................................6
Dummy cell .................................34,43,44
E
Earthing ................................................6,9
Earthing socket 14
Figure ..................................................3
ELCD cell 27
Assembly...........................................11
Connection to VA Detector 791 ........13
Figure ................................................12
Ordering designation ........................44
Startup...............................................26
Electrical safety .......................................6
Electrochemical detection.....................15
Electrochemical Detector 656
Connection to VA Detector 791 ........10
Installation ......................................... 10
Electrode cable (6.2120.020) ...............10
Electrodes ............................................. 20
Electrolyte vessel 41
Assembly...........................................11
Figure ................................................ 12
Filling reference electrode ................26
Electromagnetic compatibility ..............42
Eluent ....................................................18
EMC.......................................................42
Emitted interference..............................42
Equilibration phase ............................... 27
EU Declaration of conformity................ 47
F
Filling tool .............................................. 45
Front ........................................................ 2
Front peak ............................................. 33
Full-scale deflection .............................. 25
Fuse holder 12
Change fuses...................................... 8
Figure ............................................... 3,9
Fuses............................................... 8,9,42
G
General precautionary rules ...................6
Glassy Carbon electrode .................20,45
Gold electrode ......................................45
H
Handling of solvents ...............................6
Hazard..................................................... 5
Housing.................................................42
I
Immunity to interference .......................42
Information on the
Instructions for Use............................. 4
Installation ............................................... 7
Instructions for Use (8.791.1013) ...........4
Instrument description ............................ 1
Instrument test ......................................34
Interference level................................... 20
Introduction ............................................. 1
K
Knurled screw 43
ELCD cell assembly.......................... 11
Figure ................................................ 13
L
Leaks.......................................................6
Location .................................................. 7
791 VA Detector
49
Page 56
5 Appendix
M
Mains cable................................... 9,43,44
Mains connection.................................. 42
Procedure ........................................... 8
Safety notes ........................................6
Mains connection plug 26
Figure ............................................... 3,9
Mains connection................................ 9
Mains frequency ................................... 42
Mains pilot lamp 1
Figure .................................................. 2
Ready for use display ....................9,23
Mains switch 25
Figure ............................................... 3,9
Operation ..........................................23
Switching the instrument on/off .......... 9
Mains voltage........................................42
Setting ................................................. 8
Mains voltage indicator 13
Figure .................................................. 3
Maintenance .........................................31
Malfunctions.......................................... 32
Measuring procedure ...........................29
Measuring technique ............................ 41
Metal electrodes ..............................21,45
Metrohm service ...................................31
Multiplication ......................................... 24
N
No signal ............................................... 32
Noise ........................................... 18,20,40
Noise level............................................. 33
Notation................................................... 5
O
Opening the 791 VA Detector................. 6
Operating element functions ................23
Operation ..............................................15
Optional accessories ............................ 45
Organization............................................ 4
Outlet capillary 37
Assembly ..........................................11
Figure ........................................... 12,13
Outlet capillary 38
Assembly ..........................................12
Figure ........................................... 12,13
Output voltage ......................................35
Overload........................................... 32,38
Overload display 2
Figure .................................................. 2
Operation ..........................................25
Overload test......................................... 28
P
Packaging ............................................... 7
Parts and controls...................................2
Pictograms.............................................. 5
Pilot voltage 24
Figure .................................................. 3
Platinum electrode ................................ 45
Polarity.........................................24,25,41
Polarization voltage................ 15,17,25,41
Potentiometer 9
Figure.................................................. 2
Operation .......................................... 23
Potentiometer 10
Figure.................................................. 2
Operation .......................................... 23
Potentiostat ........................................... 41
Power consumption.............................. 42
Practical information about ELCD ........ 18
Precautionary rules ................................. 6
Protection class ..............................6,9,42
Protective earth....................................... 9
Pt electrode........................................... 21
PTFE gasket 31
Assembly ..........................................11
Figure................................................ 12
Ordering designation........................ 44
PTFE gasket 35
Assembly ..........................................11
Figure................................................ 12
Ordering designation........................ 44
Pulsation ............................................... 18
Pulsation dampener ............................. 18
R
Rear......................................................... 3
Reference electrode 39
Assembly ..........................................11
Connect to 791 ................................. 26
Figure........................................... 12,13
Filling................................................. 26
Notes................................................. 22
Ordering designation........................ 44
Reference system 40
Figure................................................ 12
Filling reference electrode ................ 26
Relative compensation current............. 23
Response characteristics ..................... 20
Rinsing the IC system........................... 31
S
Safety directions ...................................42
Safety notes ............................................ 6
Safety specifications............................. 42
Sample solution .................................... 19
Screw nipple 29
Assembly ..........................................11
Figure................................................ 12
Screw nipple 33
Assembly ..........................................11
Figure................................................ 12
Screw nipple 36
Assembly ..........................................11
Figure................................................ 12
Sealing ring 30
Figure................................................ 12
Sealing ring 34
Figure................................................ 12
Selection of polarization voltage .......... 17
Selector 20
Figure.................................................. 3
Selector 21
Figure.................................................. 3
Serial number 16
Figure.................................................. 3
Service .................................................. 31
Servicing ............................................... 31
Setting the mains voltage....................... 9
Setting up the instrument ....................... 7
Settings
Mains voltage ..................................... 8
Shutdown.............................................. 31
Silver electrode ..................................... 45
Spikes ................................................... 33
Spilled chemicals ................................. 31
Standard equipment............................. 43
Startup .................................................. 26
Static charges......................................... 6
Stationary phase................................... 19
Storage ................................................. 42
Substance turnover .............................. 33
Switch 20
Operation.......................................... 25
Switchover ........................................ 14
Switch 21
Operation.......................................... 25
Switchover ........................................ 14
Switching on/off the instrument......... 9,23
T
Technical data ...................................... 41
Tool (6.2617.000) ................................. 44
Transport ................................................ 7
Transport damage................................ 46
U
Ultra Trace Graphite electrode........ 20,45
V
Voltage selection insert 13
Change fuses ..................................... 8
Figure.................................................. 9
W
Warning................................................... 5
Warranty................................................ 46
Waste container.................................... 12
Weight................................................... 42
Working electrode 28
Assembly .......................................... 11
Connect to 791 ................................. 26
Equilibrate......................................... 28
Figure........................................... 12,13
Overview ........................................... 20
Z
Zero balancing...................................... 27
791 VA Detector
50
Loading...