Read this manual before use.
Keep this manual in a safe place for future reference.
Page 2
Introduction
This manual describes the basic operations and hardware functions of the LV-N10 Series.
Read this manual carefully to ensure safe performance and function of the LV-N10 Series.
Keep this manual in a safe place for future reference.
Ensure that the end user of this product receives this manual.
Symbols
The following symbols are used in this manual to alert you to matters concerning the
prevention of injury and product damage.
Always read these sections.
DANGER
WARNING
CAUTION
NOTICE
Important
Point
Reference
It indicates a hazardous situation which, if not avoided, will result in
death or serious injury.
It indicates a hazardous situation which, if not avoided, could result
in death or serious injury.
Failure to follow the instructions may lead to minor or moderate
injury.
Failure to follow the instructions may lead to product damage as
well as property damage.
Indicates cautions and limitations that must be followed during
operation.
Indicates additional information on proper operation.
Indicates tips for better understanding or useful information.
Indicates reference pages.
Page 3
Safety Precautions
General Precautions
• This product is only intended to detect objects. Do not use this
product for the purpose of protecting a human body or a part of
the human body.
WARNING
NOTICE
• This product is not intended for use as an explosion-proof product. Do not use this product in a hazardous location and/or
potentially explosive atmosphere.
• This product uses DC power. Do not apply AC power. The product may explode or burn if an AC voltage is applied.
• Do not wire the amplifier line along with power lines or high-tension lines, as the sensor may malfunction or be damaged due to
noise.
• When using a commercially available switching regulator,
ground the frame ground terminal and ground terminal.
• Do not use the LV-N10 Series outdoors, or in a place where
extraneous light can enter the light-receiving element directly.
• Due to individual dispersion characteristics and the difference
in sensor head models, the maximum sensing distance or displayed value may not be the same on all units.
Safety Precautions on Laser Product
• This product uses a semiconductor laser for the light source.
• Use of controls or adjustments or performance or procedures
other than those specified herein may result in hazardous radiation exposure.
WARNING
FDA (CDRH) Part 1040.10
• Follow the instructions mentioned in this manual. Otherwise,
injury to the human body (eyes and skin) may result.
Precautions on Class 1 laser products
• Do not stare into the beam.
• Do not disassemble this product. Laser emission from this
product is not automatically stopped when it is disassembled.
Model
Wavelength660 nm655 nm660 nm
Output310 øW290
IEC 60825-1Class 1 laser product
LV-NH32/NH35/NH37/
NH42/NH62/NH100/
NH110/NH300
*
LV-S31/S41/S41L/S61/
S71/S72
Class 1 laser product
96129E
LV-S62/S63
øW
1
Page 4
Safety Precautions
* The laser classification for FDA (CDRH) is implemented based on IEC60825-1 in accordance with the
requirements of Laser Notice No.50.
Reference
Laser Transmission OFF input
When the laser transmission OFF input is set for external input, the laser
transmission can be stopped by turning the external input ON (2 ms or
longer). The transmission is stopped while the external input is ON. Once
the external input is turned OFF, the laser transmission will resume within
20 ms.
Precautions on Regulations and Standards
UL Certification
This product is a UL/C-UL Listed product. (Contact KEYENCE for information on
heads which support UL-certification.)
• UL File No. E301717
• Category NRKH, NRKH7
• Enclosure Type 1 (Based on UL50)
Be sure to consider the following requirements when using this product as a UL/C-UL
Listed Product.
• Use a power supply with Class 2 output defined in NFPA70 (NEC: National Electrical Code).
• Power supply voltage is 10 - 30 VDC.
• Power supply/External input/ Control output shall be connected to a single Class 2
source only.
• Use with an over current protection device which is rated 30V or more and not
more than 1A.
• Use this product under pollution degree 2.
CE Marking
Keyence Corporation has confirmed that this product complies with the essential
requirements of the applicable EC Directive, based on the following specifications.
Be sure to consider the following specifications when using this product in a member
state of the European Union.
● EMC Directive (2004/108/EC)
• Applicable standard EMI : 60947-5-2, Class A
EMS : 60947-5-2
When connecting with the NU-CL1, always install in a conductive enclosure (control
panel, etc.), and wrap a ferrite core (E04SR401938 manufactured by Seiwa Electric
Mfg. Co., Ltd.) one turn around the sensor head cable.
These specifications do not give any guarantee that the end-product with this product
incorporated complies with the essential requirements of the EMC Directive. The
manufacturer of the end-product is solely responsible for the compliance of the
end-product itself according to the EMC Directive.
2
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 5
Manual Organization
1
2
3
4
5
6
Before Using
Installation and
Connection
Basic Operation
Settings for
Advanced
Functions
Specifications
Appendix
Outlines the package contents and identifies part
names and functions.
1
Provides procedures for installing sensor amplifiers
and cables, as well as operating precautions.
2
Explains basic instructions for operating and setting
the sensor amplifiers.
3
Describes settings for advanced functions of the
LV-N10 Series.
4
Provides the specifications, circuit diagrams and
dimensions of the LV-N10 Series.
5
Provides the troubleshooting instructions and initial
settings (default values).
6
<Points for Using This Manual>
•
When you
cedures"
•
When you
•
When you
•
When you
•
When you
"Forgot the operation methods"
"Want to try out the LV-N10"
"Want to fully utilize the various functions"
"Want to know the meanings of terms used"
"Want to troubleshoot the LV-N10"
- Digital Laser Sensor LV-N10 Series User's Manual -
- Digital Laser Sensor LV-N10 Series User's Manual -
1-1
Page 10
1
*
*
*
*
*
*
1-1
Checking the Package Contents
Before using the LV-N10 Series, make sure that the following equipment and
accessories are included in the package.
We have thoroughly inspected the package contents before shipment. However, in the
Before Using
event of missing, defective or broken items, contact your nearest KEYENCE office.
Sensor Amplifier
LV-N10 Series
Sensor amplifier × 1 Instruction manual × 1
Sensor Head
1-2
Transmitter
Focus ring
Transmitter
Receiver
Transmitter/
Receiver
LV-NH 32
Focus lock
Receiver
Plastic
screw
driver
L
O
C
K
LV-NH 37
LV-NH 62
ReflectorMounting bracket set
(R-6) × 1 (R-7) × 1
- Digital Laser Sensor LV-N10 Series User's Manual -
* There is a bar LED monitor on the sensor head LV-NH32/NH35/NH37/NH42/
NH62/NH100/NH110/NH300.
Bar LED Monitor
(Sequenced with sensor amplifier)
Stable light
reception
Setting
value
Stable
blocking
+10%
and higher
-10%
and higher
If stable light reception or stable blocking is disabled
because of the surrounding environment or changes
in the workpiece, readjust the sensitivity.
Turns ON according to
the received light intensity
with respect to the set
value.
The stability level of the
current detection is
displayed.
Bar LED monitor
excess gain display
With the 2-output type sensor amplifier, if the channel switch is set to 1, the output 1
excess gain is shown. If set to 2, the output 2 excess gain is shown.
- Digital Laser Sensor LV-N10 Series User's Manual -
1-3
Page 12
1
1-2
Part Names
Sensor Amplifier
Before Using
Display/control unit
Expansion connector
*1
Hook
Hold lock lever
Expansion connector
Dust cover
*1*2
Connection cable
(LV-N11غ/N12غ/N11MN)
M8 connector
(LV-N11Cغ/N12Cغ)
w/o cable
(LV-N10)
*1 When shipped from the factory, the expansion cover is installed.
*2 This is not installed on the main unit (LV-N11N/N11P/N11CN/N11CP/N11MN).
1-4
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 13
Display/control unit
1-2 Part Names
(1)-2
(2)
(1)-1
(10)
(3)
(9)
(4)
(5)
(8)
(6)-1
(7)
*
* Not available for the LV-N10.
ItemDescription
Operation indicator
(1)-1
(2-output type)
Operation indicator
(1)-2
(1-output type)
(2) [SET] button
Setting value
(3)
(Displayed in green)
Current value
(4)
(Displayed in red)
Indicates the current output (detection) status of channels 1 and 2 separately.
Indicates the current output (detection) status.
Used when setting sensitivity, etc.
"Adjusting the Sensitivity" (page 3-5)
Displays a setting value or advanced setting item in this
area of 7-segment green indicators.
Displays the current value (received light intensity), or a
selection from advanced settings, in this area of 7-segment red indicators.
(5) Manual buttonUsed to adjust the setting value or select an option.
Channel select
(6)-1
switch
(2-output type)
Toggles between channels 1 and 2 for configuring the
received light intensity display or sensitivity setting.
"2-output Settings ()" (page 4-33)
Changes power modes.
Power select switch
(6)-2
(1-output type)
SEL: Allows you to set a power mode using the "Chang-
ing Power Modes" function of basic setup.
M:Fixes the power mode to "MEGA mode".
"Locking in MEGA Mode" (page 3-22)
(7) [PRESET] button
(8) [MODE] button
Used for presetting or setting values or parameters.
"Adjusting the Sensitivity" (page 3-5)
Used for toggling L-on/D-on, proceeding to advanced
settings, or confirming selections.
Lights when a DATUM mode is in effect.
(9) DTM indicator
"DATUM1 mode" (page 4-9)
"DATUM2 mode" (page 4-11)
(10) PST indicator
Lights when preset value is set.
"Adjusting the Sensitivity" (page 3-5)
(6)-2
1
Before Using
*
- Digital Laser Sensor LV-N10 Series User's Manual -
1-5
Page 14
1
1-2 Part Names
MEMO
Before Using
1-6
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 15
Installation and Connection
This chapter provides procedures for installing sensor amplifiers and
cables, as well as operating precautions.
Connecting the Sensor Amplifier to the Sensor Head
2-3Mounting and Adjusting the Sensor Head..........2-6
..... 2-5
2
- Digital Laser Sensor LV-N10 Series User's Manual -
2-1
Page 16
2
Installation and Connection
2-1
Installing Sensor Amplifiers
Mounting the Sensor Amplifier
Mounting on a DIN rail
Align the claw at the bottom of the main body
1
with the DIN rail, as shown on the right.
While pushing the main body in the direction
of the arrow (1), push down in the direction of
arrow (2).
To release the amplifier, raise the amplifier
2
body in the direction of arrow (3) while pushing in the direction of arrow (1).
Installation on a wall
(3)
(2)
(1)
Reference
1
2
Connecting multiple amplifiers
Up to 16 expansion units can be connected to 1 main unit. Note, however, that the 2output type is considered as 2 expansion units.
This method applies only when using the main unit independently. If the
main unit is connected with an expansion unit(s), use the method of mounting on a DIN rail.
Mount the amplifier on the amplifier mounting bracket (OP-73880, sold separately),
using the same manner as "Mounting on a
DIN rail".
Secure the unit with two M3 screws as
shown in the illustration.
Mount on DIN rail and install on metal surface when connecting
WARNING
multiple amplifiers or mounting main units together.
OP-73880
2-2
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 17
Point
• Contact your nearest KEYENCE office when connecting a unit
other than the N-bus (KEYENCE’s wire-saving system) compatible
sensor amplifier, including LV-N10 Series, or the NU Series communication module.
• Turn the power off before connecting multiple expansion units.
• Do not touch the expansion connector with your bare hands.
• When using the LV-N10 Series as a main unit, use the products
within the expansion unit’s power voltage range if the power voltage range of the expansion unit is narrower than the LV-N10
Series.
Remove the protection covers from the main unit
1
and expansion unit(s).
Mount the main unit and expansion unit(s) on the
2
DIN rail.
Slide the main unit and expansion unit(s) together.
3
Engage the 2 claws of the expansion unit with the
recesses on the main unit side until you hear/ feel
a click.
Attach the separately sold end units (OP-26751: a
4
set of 2 units) to the DIN rail in the same manner
as step (2).
2-1 Installing Sensor Amplifiers
2
Installation and Connection
(2)
(1)
Secure the amplifiers between the end units.
5
Tighten the screws at the top (2 screws × 2 units)
with a Phillips screwdriver to fix the end units.
- Digital Laser Sensor LV-N10 Series User's Manual -
OP-26751 (a set of 2 units)
2-3
Page 18
2
2-1 Installing Sensor Amplifiers
Wiring Diagrams for Sensor Amplifiers
Point
• Be sure to turn off the power before wiring.
• Insulate each input or output cable that will not be used.
Installation and Connection
Cable type (LV-N11/N12)
Brown*
Blue*
Black
White
Pink
* LV-N11N/N11P only
10 to 30 VDC
Output 1
Output 2
External input
M8 connector type (LV-N11C/N12C)
(1)*
10 to 30 VDC
(3)*
(2)
External input
(4)
Output
2
4
3
1
Pin and wire color table
Connected
pin No.
1Brown
2
3
4
OP-73864
(Cable length: 2 m)
OP-73865
(Cable length: 10 m)
* LV-N11CN/N11CP only
Wire color
White
Blue
Black
Monitor output type (LV-N11MN)
Brown
Blue
Black
Orange
Pink
*Connect to a device having an input impedance of 10kΩ or more.
Reference
Refer to "Circuit Diagrams" (page 5-5) for the input/output circuit
*
10 to 30 VDC
Output
Monitor output (1 to 5 V)
External input
diagrams.
2-4
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 19
2-2
Connecting the Sensor Amplifier to the Connector
1
2
3
Point
Connecting the Sensor Amplifier to the Sensor Head
Open the dust cover, and move the hold
lock lever down.
Lift the hook up, and insert the connector completely.
Lower the hook to the position
shown in the drawing, and pull
up the hold lock lever.
• When connecting to a different sensor head model, the settings for
the previously connected model will be saved.
Either reconfirm or initialize the settings.
"Initialization of Settings (Reset to Initial Values)" (page 3-21)
• When shortening the sensor head cable, follow the instructions
given in the "Sensor Head Connector Assembly Procedures"
included with the sensor head.
2
Installation and Connection
- Digital Laser Sensor LV-N10 Series User's Manual -
2-5
Page 20
2
2-3
Mounting and Adjusting the Sensor Head
LV-NH32/NH35/NH37/NH42/NH62/S63
Use the included mounting bracket.
The tightening torques are shown below.
Insert the plate nut* between the
sensor section and mounting bracket.
* When using LV-NH100/NH110, the long plate nut is used for the transmitter, and the
short plate nut for the receiver.
When using LV-NH300, the long plate nut is used for the receiver, and the short
plate nut for the transmitter.
2-6
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 21
2-3 Mounting and Adjusting the Sensor Head
Adjust the beam axis.
2
(1)
Adjust the beam axis upward by tightening the
(a) screw, and downward by loosening the
screw. Fix the angle setting with the (b) nut.
(2) Adjust the horizontal angle by loosening the
fixing section (c).
LV-S31
Mount using the Included mounting
1
bracket.
Place a workpiece at the position to be
2
set as the detection center.
Turn the trimmer with the Included adjustment
3
screwdriver, and adjust so that the (2) JUS indicator lights in green.
(a)
(b)
2
(c)
F
F
J
N
N
F
F
J
N
N
Installation and Connection
F
(1)
(2)
(3)
Trimmer
F
F
J
N
If the (1) FAR indicator is ON, turn the trimmer
N
clockwise
until (2) turns ON.
If the (3) NEAR indicator is ON, turn the trimmer
counterclockwise
until (2) turns ON.
F
J
N
N
When the LV-S31 is connected, the workpiece position is numerically displayed on the
main screen (red display) using the center position as [5000]. If adjust the workpiece
position precisely, set the display of sensor amplifier approximate [5000]. The display
changes as follows according to the workpiece position.
DisplayMeaning
Near
----Out of detection range
The detector is nearer to the sensor side than numerical dis-
nEAr
play range
Centering on 5000, the value increases as the workpiece dis-
5000
tance increases from the sensor
The workpiece is farther from the sensor than the numerical
FAr
Far
display range
----Out of detection range
- Digital Laser Sensor LV-N10 Series User's Manual -
2-7
Page 22
2
2-3 Mounting and Adjusting the Sensor Head
LV-S41/S41L
Always use the included mounting bracket.
Attach the mounting bracket as shown on
1
the right.
Match section A of the fitting onto section
Installation and Connection
B of the sensor head
Attach onto a flat surface with M3 screws as
2
shown on the right.
(M3 screws are not included.)
LV-S62
Mount with the optional L-mounting bracket (OP-84350), back surface mounting
bracket (OP-84349) or horizontal mounting bracket (OP-84351).
When mounting the sensor head in the direction opposite of that
shown above, set the spot selection switch before mounting.
Adjust the beam axis.
2
The beam axis can be adjusted in the
upward direction by tightening the
screw as shown with the arrow. It can
be adjusted in the downward direction
by loosening the same screw.
- Digital Laser Sensor LV-N10 Series User's Manual -
To move the beam axis
upward
To move the beam axis
downward
2-9
Page 24
2-3 Mounting and Adjusting the Sensor Head
LV-S71/S72
2
Installation and Connection
Attach so that the side on which
1
the T (transmitter) and R (receiver)
are printed faces upward. The
operation indicator lights are on
the printed side.
Adjust the beam axis.
2
The beam axis can be adjusted in the downward
direction by tightening the screw as shown with
the arrow. It can be adjusted in the upward direction by loosening the same screw.
Adjust so that the beam spot is centered on
the receiver.
When adjusting, attach the beam axis alignment plate, included with the sensor head, onto
the end of the receiver
Remove the beam axis alignment plate cap
when finished adjusting.
to aid in alignment.
Fixing nut
Axis beam
adjustment screw
Beam axis
alignment
plate
The tightening torque is
1.2 N·m or less.
To move the beam axis downward
To move the beam axis upward
2-10
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 25
Basic Operation
This chapter explains basic instructions for operating and setting the
sensor amplifier.
- Digital Laser Sensor LV-N10 Series User's Manual -
3-1
Page 26
3-1
Quick Reference
The main setting operations are explained according to purpose. Refer to Chapter 4
for information on advanced function settings and explanations not given below.
3
Basic Operation
Operation
indicator
(1-output type)
PurposeDescriptionOperation procedures
Changing the
output
Adjusting the
sensitivity and
integrating the
display to "100.0"
and ".0"
Adjusting the
sensitivity
Shifting the
received light
intensity to ."0"
Operation indicator
(2-output type)
[SET] button
1 Change the output. (L-on/D-on)
Set the current received light intensity
2
to "100.0". (Preset)
When preset is valid, register the
3
received light intensity ".0".
(Work-preset)
Set the received light intensity slightly
higher than when the setting was
4
made, to "100.0". (Maximum sensitivity preset)
Automatically register "100.0" and ".0"
5
when workpiece passes by. (Full Auto
preset)
6 Cancel the various preset functions.Press and hold the [PRESET] button.3-6
Set the display for preset execution to
"1000" (LV-NH100/NH110) or "3000"
7
(LV-NH300). (Preset area display
mode)
Set the setting value at the midpoint
between the received light intensity
8
values when a workpiece is present
and absent. (2-point calibration)
Set the setting value slightly higher
than the received light intensity value
9
at which the setting was made. (Maximum sensitivity calibration)
Set the setting value automatically
10
when a workpiece is passing through.
(Full auto calibration)
S
et the setting value to the base point
11
where the workpiece is positioned.
(Positioning calibration)
12 Finely adjust the setting value directly.
Set the current display to ."0". (Zero
13
shift)
14 Cancel the zero shift function.Press and hold the [PRESET] button.3-17
PST
indicator
DTM
indicator
Manual button
1. Press the [MODE] button.
2. Switch with the ( ) button.
Reflective model: Press the [PRESET] button
when a workpiece is present. Thrubeam/Retroreflective model: Press the [PRESET] button
when no workpiece is present.
After step 2, press the [PRESET] button +
button in the state to be set as ".0".
While the PST indicator is OFF, press and hold
the [PRESET] button. Reflective model: When
no workpiece is present. Thrubeam/Retroreflective model: When a workpiece is present.
Press and hold the [PRESET] button while the
PST indicator is OFF.
1. Press and hold the [MODE] button, button and [SET] button simultaneously.
2. Press the [MODE] button.
3. Select "Pr-d ArEA" with ( ) button, and
press and hold the [MODE] button.
1. Press the [SET] button once when a workpiece is present.
2. Press the [SET] button once when no workpiece is present.
Reflective model: Press and hold the [SET] button when no workpiece is present. Thrubeam/
Retro-reflective model: Press and hold the
[SET] button when a workpiece is present.
Press and hold the [SET] button while the workpiece passes through.
1. Press the [SET] button once when no workpiece is present.
2. Press and hold the [SET] button at the positioning point.
Press the ( ) button.
Press the [PRESET] button + button when
the PST indicator is OFF.
Channel select
switch
(2-output type)
[MODE]
button
[PRESET]
button
Power select
switch
(1-output type)
Reference
page
3-4
3-6
3-7
3-8
3-9
3-11
3-12
3-13
14
3-
3-15
1-5
3-17
3-2
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 27
PurposeDescriptionOperation procedures
1. Press and hold the [SET] button and [PRESET] button.
Loading the
recommended set-
tings
Initializing the set-
tings
Displaying the out-
put 2 display
screen with the
2-output type
Switching to the
maximum received
light intensity
power mode
Preventing
incorrect
operations
Others
(Advanced
function settings,
etc.)
Saving and
loading the
settings
Load the recommended settings.
15
(Recipe function)
Initializing (Restore to factory default
16
settings)
17 Switch to the output 2 display screen.
Adjust the power mode to the MEGA
18
mode.
19 Activating the key lock
20 Deactivating the key lock
Activating the password-protected key
21
lock
Deactivating the password-protected
22
key lock
23 Setting the advanced functionsPress and hold the [MODE] button.4-1
Setting to rescale at each preset exe-
24
cution so that analog output is "5 V"
output in respect to "100.0".
Switching the display to extended dis-
25
play or received light intensity hold
display, etc. (sub-display)
Resetting the following values
* Received light intensity hold value
* Excess gain hold value
26
* Output when output 2 is in limit setting detection mode
* Count value when output 2 is in
counter output mode
27 Saving the settings (custom save)
28 Loading the settings (user reset)
2. Display the LoAd screen with the ( )
button, and press the [MODE] button.
3. Select the recipe such as r-1 FALL with the
( ) button.
4. Press the [MODE] button to execute.
1. Press and hold the [SET] button and [PRE-
SET] button.
2. Press the [MODE] button while on the rSt
screen.
3. Select init with the ( ) button.
4. Press the [MODE] button to execute.
Set the channel switch to .
* The output 2 sensitivity and advanced settings can be modified in this state.
Set the power select switch to .
Press and hold the [MODE] button and the
( ) button simultaneously.
Press and hold the [MODE] button and the
( ) button simultaneously.
1. Press the ( ) button10 times while
holding down the [MODE] button.
2. Input the password with the ( ) button.
3. Press the [MODE] button to execute.
1. Press the ( ) button 10 times while
holding down the [MODE] button.
2. Input the password with the ( ) button.
3. Press the [MODE] button to deactivate the
key lock.
1. Press and hold the [MODE] button, but-
ton and [SET] button simultaneously.
2. Press the [MODE] button once or twice.
3. Select "Pr-A PrST" with ( ) button, and
press the [MODE] button.
After setting the sub-display with the advanced
function settings, press the [MODE] button
twice.
Press and hold the [MODE] button and [SET]
button.
1. Press and hold the [SET] button and [PRE-
SET] button.
2. Display the SAvE screen with the ( )
button, and press the [MODE] button.
3. Select yES with the ( ) button.
4. Press the [MODE] button to execute.
1. Press and hold the [SET] button and [PRE-
SET] button.
2. Press the [MODE] button on the rSt screen.
3. Select USEr with the ( ) button.
4. Press the [MODE] button to execute.
3-1 Quick Reference
Reference
page
3-19
3-21
4-33
3-22
3-23
3-23
3-24
3-24
4-21
2
4-2
4-24
4-34
4-35
4-38
4-39
3
Basic Operation
- Digital Laser Sensor LV-N10 Series User's Manual -
3-3
Page 28
3-2
Switching Output
Output Switch (L-on/D-on)
This function configures when the output turns ON.
When the current received light inten-
1
sity is displayed, press the [MODE] button.
3
Basic Operation
The current output condition (L-on or D-
on) is displayed.
Press the button to switch the out-
2
put condition, and then press the
[MODE] button.
Select "D-on" if you want to output the ON
signal when the beam is blocked (a workpiece is present.) for a thrubeam or retro-reflective model.
Select "L-on" if you want to output the ON signal when the beam is received (a
workpiece is present.) for the reflective model.
The output condition is switched, and the current received light intensity is dis-
*2
played.
*1 If you do nothing for 3 seconds or more or press the [MODE] button, the received
light intensity display is automatically restored.
*2 When using the sub-display, the screen will switch between the current received
light intensity J L-on/D-on screen J sub-display J current received light intensity
and so forth each time the [MODE] button is pressed.
"Sub Display" (page 4-22)
Point
• When in area detection mode, rising edge detection mode, falling
edge detection mode, counter output mode, or when the LV-S31 is
connected, this function works as a normally-open/normallyclosed switch.
"Area detection mode" (page 4-14)
"Rising edge detection mode" (page 4-16)
"Falling edge detection mode" (page 4-16)
"Counter output mode" (page 4-35)
• With the 2-output type, channels 1 and 2 can be configured separately.
*1
3-4
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 29
3-3
Adjusting the Sensitivity
In this manual, the value at which the sensor amplifier's ON/OFF output switches is
expressed as the "setting value". The process of adjusting the setting value is expressed
as "adjusting the sensitivity". This section explains the method of adjusting the sensitivity.
List of Sensitivity Adjustment Methods
The methods of adjusting the LV-N10 series sensitivity is largely classified into 2 types.
(1) Preset
At the same time as adjusting the sensitivity, the received light intensity can be
calibrated to "100.0" or ".0" using simple operations.
This method can decrease the variation of the received light intensity depending on
the contents of the detection and the workpiece, and is useful for predictive
maintenance.
However, this is not suitable for when difference in received light intensity varies little
with or without a workpiece, such as when detecting a transparent workpiece.
(2) Calibration
The sensitivity can be adjusted with simple operations. The received light intensity is
not compensated.
This method is used to adjust without calibrating the received light intensity, or when
highly accurate detection is required.
Calibration can be performed in the preset state.
Basic method of selecting sensitivity adjustment method
3
Basic Operation
Sensitivity
adjustment
method
Preset
Calibration
Basic
At times
like thi s
Basic
At times
like thi s
UsageFunctionDetailsReference
Using a thrubeam/retro-reflective model
Using a reflective model
"100.0" and ".0" cannot be set
regardless of whether a workpiece is present or absent.
The moving workpiece moves
quickly
Using a thrubeam/retro-reflective / reflective model
The moving workpiece moves
quickly
Using in an easily contaminated environment
Using with positioning
Using with highly accurate
detection
Preset
Maximum sensitivity preset
Work- prese t
Full auto preset
Two-point
calibration
Full auto
calibration
Maximum sensitivity calibration
Positioning calibration
Percentage calibration
The sensitivity is adjusted just by
pressing the [PRESET] button when
a workpiece is absent.
The sensitivity is adjusted just by
pressing and holding the [PRESET]
button when a workpiece is absent.
The states at which to display
"100.0" and ".0" can be set randomly.
The sensitivity can be calibrated
using a workpiece which moves at
high speed.
The setting can be established just
by pressing the [SET] button on ce
when a workpiece is present and
absent.
The sensitivity can be calibrated
using a workpiece which moves at
high speed.
This setting prevents malfunctions
even when using in an easily contaminated environment.
A setting suitable for positioning can
be made.
This is useful for calibrating from an
external device, such as a PLC.
- Digital Laser Sensor LV-N10 Series User's Manual -
3-6
3-8
3-7
3-9
3-12
3-9
3-8
3-15
3-16
3-5
Page 30
3
3-3 Adjusting the Sensitivity
Preset Function
Enabling the preset function
When the PST indicator is not lit, press the [PRESET] button
The PST indicator lights in
green. The current value is
set to "100.0" and the
setting value is set to
"50.0".
Basic Operation
Green PST lights up
Workpiece
Setting value
is
"
.
"
Current value is
"
."
Disabling the preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the preset function has been disabled.
• Once the preset function is disabled, the setting value is recalculated while
retaining the ratio of the setting value and received light intensity.
3-6
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 31
3-3 Adjusting the Sensitivity
Work-Preset Function
This function calibrates the current value to ".0". After the preset function has been
executed with "100.0" displayed, and then executing this function with ".0" displayed,
2 random set points can be calibrated to "100.0" and ".0".
Enabling the work-preset function
Impotant
The work-preset function can be used with the preset function (when
preset is enabled).
While the preset function is enabled, press the [PRESET] button and button
simultaneously.
3
Basic Operation
The received light intensity at
that point is set to ".0".
100.0
The value set to "100.0" using
the preset function does not
change.
Reference
Even if the received light intensity is low during preset and is high during
3.0
.0
[PRESET]+
Green PST
lights up
Current value is ".
work-preset, the value is set to "100.0" during preset and ".0" during workpreset. When the actual received light intensity increases, the display will
approach ".0". (The preset saturation level is decreased with respect to
"100.0".)
Disabling the work-preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the work-preset function has been
disabled.
"
- Digital Laser Sensor LV-N10 Series User's Manual -
3-7
Page 32
3
3-3 Adjusting the Sensitivity
Maximum Sensitivity Preset Function
This function calibrates the reference state to ".0" and the state at which the received
light intensity is slightly higher as "100.0".
This function is useful with the reflective model to detect while using the background
as a reference.
Enabling the maximum sensitivity preset function
When no workpiece is present for the
1
reflective model, or when a workpiece is
present for thrubeam model/retro-
Basic Operation
reflective model, press and hold the
[PRESET] button for 3 seconds or more
while the PST indicator is OFF.
Release the button when "
Calibration is complete after the setting
value flashes momentarily, and then stops
(lights up).
The PST indicator lights in green, and the
setting value is set to "
Auto
" flashes.
50.0
".
Thrubeam Model/Retro-reflective Model
Received light intensityReceived light intensity
Reflective Model
With workpiece*
Saturation point
Setting value
No workpiece*
1
"100.0"
"50.0"
".0"
No workpiece
Saturation point
Setting value
With workpiece
1
* Where detection occurs on a target having a background, the maximum sensitiv-
ity setting ignores the background. Maximum sensitivity setting is not available if
the background is more reflective than the workpiece.
Point
• The maximum sensitivity preset function cannot be executed when
the received light intensity is saturated (higher than the value
listed in the Extension display on page 4-30). ("---- ----" will
appear during step 1.)
• The maximum sensitivity preset function cannot be used when the
LV-NH100, NH110 or NH300 is connected, but by holding down the
[PRESET] button in the light ON state, the amount of light blocked
at light intensity .0 and light block 100.0 is displayed.
Disabling the maximum sensitivity preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the maximum sensitivity preset function has been disabled.
3-8
- Digital Laser Sensor LV-N10 Series User's Manual -
"100.0"
"50.0"
".0"
Page 33
3-3 Adjusting the Sensitivity
Full Auto Preset Function
This function automatically judges 2 states (workpiece presence/absence, etc.), and
calibrates the current value to "100.0" and ".0".
This is useful when the detector is moving at high speed, etc.
Enabling the full auto preset function
When the PST indicator is OFF, con-
1
tinue pressing the [PRESET] button
until "Auto" flashes while the workpiece passes through.
After the workpiece has completely
2
passed through, release the [PRESET]
button.
Calibration is complete after the setting
value flashes momentarily, and then
stops (lights up).
The PST indicator lights in green, and the
setting value is set to "50.0".
3
Basic Operation
2
MAX
Setting
value
Time
The area near the maximum value
of the received light intensity while
the [PRESET] button is pressed is
set as "", and the area near
the minimum value is set as "".
"100.0"
Point
light intensity
Received
"50.0"
".0"
The full auto preset function cannot be executed when the received
light intensity is saturated (higher than the value listed in Extension
display on page 4-30). ("---- ----" will appear during step 2.)
MIN
1
- Digital Laser Sensor LV-N10 Series User's Manual -
3-9
Page 34
3
3-3 Adjusting the Sensitivity
Disabling the full auto preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the full auto preset function has been
disabled.
Point
Basic Operation
Each preset function cannot be used when the following functions
are set. Disable the function or change the setting before executing
the preset function.
• Zero shift function -> "Zero Shift Function" (page 3-17)
• Zero shift calibration -> "Adjusting the Sensitivity" (page 3-5)
• Zero shift input -> "External Input" (page 4-17)
• DATUM 1/2 mode -> "Restrictions for Sensitivity Settings in Each
Detection Mode" (page 6-8)
• Rising/falling edge detection mode" -> "Restrictions for Sensitivity
Settings in Each Detection Mode" (page 6-8)
• The preset function is not suited for transparent workpieces such
as thrubeam/retro-reflective models and other cases of detection
with low light intensity differences.
• After changing any of the following settings, disable each preset
function once and then execute again.
"Power Modes" (page 4-4)
"Preset Saturation Function" (page 4-27)
"Display Gain" (page 4-30)
"Sensitivity" (page 4-32)
• If the received light intensity raw value is 50 or less (200 or less
when the light intensity is set to FuLL), the display will be "100.0" or
less when the preset function is executed.
3-10
Reference
• If the [PRESET] button is pressed when the preset function is enabled
(PST indicator is lit), the current received light intensity changes to
"100.0" and the setting value does not change.
• Expansion units can be preset with operations from the main unit.
"Common Key-Operations Function" (page 4-32)
• Periodic presetting is possible with signals input from an external source.
"External Input" (page 4-17)
• With the preset function, a process is carried out to ignore minute
received light intensity changes which do not affect the detection. The
change amount to be ignored can be adjusted randomly.
"Preset Saturation Function" (page 4-27)
If the sub-display is changed to "Extension", the original received light
intensity can be confirmed even when using the preset function.
Extension ->"Sub Display" (page 4-22)
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 35
3-3 Adjusting the Sensitivity
Preset display mode (only when LV-N100/NH110/NH300 is connected)
The "100.0" (default value) displayed when preset is executed can be changed to
"1000" or "3000" * according to the area width.
The preset display mode "ArEA" is convenient for displaying the amount of light blocked
in the area, etc.
* "1000" is displayed when LV-N100/NH110 is connected, and "3000" when
LV-NH300 is connected.
When received light intensity is dis-
1
played, press and hold the [MODE] button, [SET] button and button
simultaneously for 3 seconds or more.
ProG YES or ProG no is displayed.
Press the [MODE] button.
2
Using the button, select Pr-d PrSt
3
for the normal display or Pr-d ArEA for
the area display, and then press and
hold the [MODE] button.
Setting example
(1) If the preset function is executed when the power is ON, the display at that point
When LV-NH300 is connected and preset display mode "ArEA" is selected
will b e scaled to "3000".
Received light intensity
Received light intensity after preset
3
Basic Operation
Press the
[PRESET]
When light is ONWhen light is blockedWhen light is ONWhen light is blocked
3000
0
(2) If the [PRESET] button is press and hold in the light ON state, the amount of
blocked light at light intensity "0" and light block "3000" is displayed.
Received light intensity
Press and hold
[PRESET]
When light is ONWhen light is blockedWhen light is ONWhen light is blocked
Point
If the preset display mode is set to
is fixed to
oFF
, and the value exceeding
Received light intensity after preset
3000
0
ArEA
, the preset saturation function
1000
or
3000
will also be dis-
played. The preset saturation function setting screen will not appear.
"Preset Saturation Function" (page 4-27)
- Digital Laser Sensor LV-N10 Series User's Manual -
3-11
Page 36
3
g
3-3 Adjusting the Sensitivity
Two-point Calibration
Two-point calibration is the most basic method of sensitivity setting.
The setting value can be established by simply pressing the [SET] button once each
when a workpiece is present. and when it is absent.
Press the [SET] button once
1
when no workpiece is
present.
Basic Operation
Press the [SET] button once
2
when a workpiece is present.
Calibration is complete after the
setting value flashes momentarily, and then stops (lights up).
Press the button to adjust
the setting value.
Reference
3-12
Step 1 or step 2 may be performed first.
When performing two-point calibration on channel 2 of the 2-output type,
set the channel switch to "2".
li
Received
ht intensity
* If the difference between the 2 conditions is too small, "- - - -" flashes
1
Setting
value
2
after calibration is complete.
However, the setting value will still established.
- Digital Laser Sensor LV-N10 Series User's Manual -
The setting value is set to the middle
of the values at which the [SET] button
was pressed first and pressed second.
Time
Page 37
3-3 Adjusting the Sensitivity
Maximum Sensitivity Calibration
This sensitivity setting method is useful if the received light intensity is reduced by
dust or dirt.
The setting value is calibrated to be slightly than the light intensity received when it is
determined.
When a workpiece is absent for the
1
reflective model, or when a workpiece
is present for the thrubeam model/
retro-reflective model, press and hold
the [SET] button for 3 seconds or
more.
Release the button when "SEt" flashes.
Calibration is complete after the setting
value flashes momentarily, and then stops
(lights up).
Press the button to adjust the setting
value.
3
Basic Operation
Reference
With workpieceNo workpiece*
* Where detection occurs on a target having a background, the maximum sensi-
When performing two-point calibration on channel 2 of the 2-output type,
set the channel switch to "2".
Thrubeam Model/Retro-reflective Model
Received light intensityReceived light intensity
No workpiece
Setting valueSetting value
1
tivity setting ignores the background. Maximum sensitivity setting is not available if the background is more reflective than the workpiece.
- Digital Laser Sensor LV-N10 Series User's Manual -
With workpiece*
Reflective Model
1
3-13
Page 38
3
g
3-3 Adjusting the Sensitivity
Full Auto Calibration
This method automatically sets sensitivity using a moving workpiece.
Sensitivity can be easily set by passing a workpiece without shutting down operating
equipment.
While the workpiece passes through,
1
press and hold the [SET] button until
"SEt" flashes.
Basic Operation
After the workpiece has completely
2
passed through, release the [SET] button.
Calibration is complete after the setting
value flashes momentarily, and then
stops (lights up).
Reference
3-14
• If detection is not stable after the setting operation, for example due to
vibration, press the button to adjust the setting value.
• When performing two-point calibration on channel 2 of the 2-output type,
set the channel switch to "2".
li
Received
ht intensity
MIN
1
- Digital Laser Sensor LV-N10 Series User's Manual -
2
MAX
Setting
value
Time
The setting value is determined
as the middle value between the
maximum and minimum light
intensity received values
obtained while holding down the
[SET] button.
Page 39
Positioning Calibration
g
This method is used when you want to position a workpiece.
Press the [SET] button
1
once when no workpiece is
present.
Position a workpiece such
2
that its edge aligns with the
center of the projecting
beam. Then, press and hold
the [SET] button for 3 seconds or more. Release the
button when "SEt" flashes.
Calibration is complete after
the setting value flashes
momentarily, and then stops
(lights up).
Press the button to
adjust the setting value.
Workpiece
3-3 Adjusting the Sensitivity
3
Basic Operation
Reference
The setting value is determined as the value of light intensity received when
the workpiece comes into position.
When performing two-point calibration on channel 2 of the 2-output type,
set the channel switch to "2".
li
Received
ht intensity
Setting
value
2
1
- Digital Laser Sensor LV-N10 Series User's Manual -
Time
3-15
Page 40
3
3-3 Adjusting the Sensitivity
Other Calibration Methods
Percentage calibration
A setting value is established with a percentage in respect to the current received light
intensity. When used together with external calibration input, percentage calibration
can be performed from an external device, such as a PLC, allowing highly accurate
detection of transparent workpieces and small workpieces, etc.
Refer to "Sensitivity Setting" (page 4-4) for details.
Basic Operation
Zero shift calibration
This sensitivity setting performs the zero shift function and basic calibration (two-point
calibration / maximum sensitivity calibration / full auto calibration) simultaneously.
The lower of the received light intensity values specified at the time of sensitivity
setting will automatically be set to 0.
Refer to "Sensitivity Setting" (page 4-4) for details.
Point
Percentage calibration and zero shift calibration cannot be set with
channel 2 of the 2-output type models.
3-16
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 41
3-4
Setting the Current Received Light Intensity to 0 (Zero Shift)
Zero Shift Function
This function adjusts the current received light intensity display to "0". It is primarily
used with reflective sensor heads. If the received light intensity cannot be set to "0"
when a reflective model is installed, this function can be used to set the received light
intensity to "0" with no workpiece present. This makes the difference in received light
intensity easier to distinguish.
Enabling the zero shift function
When the PST indicator is off, press
1
the [PRESET] button and button
simultaneously.
The PST indicator lights in green, and the
received light intensity changes to "0".
Green PST lights up
Current value is "
Disabling the zero shift function
When the PST indicator is on, press and hold the [PRESET] button.
1
The PST indicator turns off, indicating that the zero shift function has been disabled.
Operating Principle of the Zero Shift Function
< For reflective models >
Received light intensity
650
450
400
300
200
0
In applications for distinguishing colors or
for detecting objects on a background,
the received light intensity will not be "0"
even when no workpiece is present.
Without workpiece
(Background is detected)
Received light intensity
450
250
200
100
0
Applying zero shift input to the lower
level of received light intensity (with
no workpiece present) enhances
the detection display visibility.
3
Basic Operation
"
- Digital Laser Sensor LV-N10 Series User's Manual -
3-17
Page 42
3-4 Setting the Current Received Light Intensity to 0 (Zero Shift)
Point
The zero shift function cannot be used in combination with the preset
function. To use the zero shift function, make sure that the preset
function is disabled (the PST indicator is not lit).
3
Basic Operation
Reference
Zero shift of the expansion unit can be set from the main unit.
"Common Key-Operations Function" (page 4-32)
3-18
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 43
3-5
Recommended settings (recipes) for each sensor head and application are preregistered in the LV-N10 Series.
Selecting Recipe
Point
Loading the Recommended Settings (Recipe Function)
When a recipe is loaded, all settings other than those registered in the
recipe are returned to the default values. Complete all other settings
after loading the recipe.
3
Press and hold the [SET] and [PRE-
1
SET] buttons simultaneously for 3 seconds or more.
The "rSt" display flashes.
Press the button to display "LoAd",
2
and press the [MODE] button.
Press the button to select a
3
desired recipe, and press the [MODE]
button.
After the settings are loaded, the screen displays "LoAd oK", which is then
replaced with the current received light intensity.
Basic Operation
- Digital Laser Sensor LV-N10 Series User's Manual -
3-19
Page 44
3-5 Loading the Recommended Settings (Recipe Function)
List of Recipes and Recommended Sensor Heads
3
Basic Operation
Recipe
Fall
r-1 FALL)
(
Percentage
calibration
(
r-2 SEtP)
Zero-Shift
(
r-3 0SEt)
MEGA
(
r-4 MEGA)
AREA
r-5 ArEA)
(
0 DATUM
(
r-6 0dtM)
Recommended
head
Thrubeam model
Retro-reflective
model
Thrubeam model
Retro-reflective
model
Reflective model
Reflective model
Area model
Distance setting
model
Retro-reflective
model
Description
To detect a falling workpiece.
The falling edge of the received light intensity is
detected, and output as one-shot.
The received light intensity hold value can be confirmed by pressing the [MODE] button twice on the
output 1 screen.
To use the external input and periodically compensate for the reduction in received light intensity due
to contamination on sensor head face.
A decrease in the light-ON mode's received light
intensity can be detected and output to an external
source by using the 2-output type's output 2.
To use the external input and "0" the light intensity
received without a workpiece present.
The background received light intensity is set as 0
even when calibrating with the buttons on the display.
(Zero Shift calibration)
To 0 the lower of level received light intensity values
in order to display higher values of received light
intensity.
The display can be extended up to 5 digits by pressing the [MODE] button twice on the output 1 screen.
This is recommended when outputting within a specific received light intensity range. When the [SET]
button is pressed once at the reference received light
intensity, the upper/lower setting value limits will be
set at μ10% from the reference received light intensity.
This is recommended when detecting a transparent
workpiece. The received light intensity at the state
with no workpiece is set as ".0", and the displayed
value is increased as the amount of light blocked by
the workpiece increases.
When the [SET] button is pressed or when calibrated
with external inputs, the received light intensity -5%
at that point is set as the setting value.
3-20
Reference
Refer to "List of Recipe Function Settings" (page 6-6) for the details on the
items set with recipe loading.
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 45
3-6
Initialization of Settings (Reset to Initial Values)
The sensor amplifier can be reset to the factory default settings.
1
Initialization
Press and hold the [SET] and [PRESET] buttons simultaneously for 3 seconds or more.
The "rSt" display flashes.
3
Press the [MODE] button.
2
Press the button to display "init".
3
Press the [MODE] button.
4
After the settings are initialized, the screen
displays "oK", which is then replaced with
the current received light intensity.
Reference
Refer to "Factory Default Setting (Default Value) List" on page 6-5.
Basic Operation
- Digital Laser Sensor LV-N10 Series User's Manual -
3-21
Page 46
3
Basic Operation
3-7
Locking in MEGA Mode
MEGA Mode Lock (1-Output Type Only)
This function is available for the 1-output type models (LV-N11C / N12C/N11MN)
only.
The sensor amplifier can be locked in MEGA mode, such that it always operates in
MEGA mode regardless of the power mode selected in the basic setup.
"Power Modes" (page 4-4)
Slide the power select switch to the
1
"M" side.
Sliding the power select switch back to the
"SEL" side restores the power mode that
was set before sliding the power select
switch to MEGA mode.
Reference
• When the amplifier is locked in MEGA mode, the power mode may not be
changed in the basic setup, as indicated by the flashing of "Loc".
"Power Modes" (page 4-4)
• Likewise, when the amplifier is in the key locked state, the power mode
may not be changed, as indicated by the flashing of "Loc".
"Key Lock" (page 3-23)
3-22
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 47
3-8
Disabling the Key Operation
Key Lock
The key lock function disables all key operation to prevent unauthorized use.
Activating key lock
Press and hold the [MODE] button and
1
(or ) simultaneously for 3
seconds or more.
3
The screen displays "Loc", disabling key
operation and displaying the current
received light intensity.
Deactivating key lock
Press and hold the [MODE] button and
1
(or ) simultaneously for 3
seconds or more.
The screen displays "unL", enabling key
operation.
Reference
• The key operations on the expansion unit can be locked from the main
unit.
"Common Key-Operations Function" (page 4-32)
• By using the network unit NU series, key locks which could only be disabled via the network can be set. This function allows you to change only
the necessary settings with a PLC and touch panel. Refer to the NU
Series User's Manual for details.
Basic Operation
- Digital Laser Sensor LV-N10 Series User's Manual -
3-23
Page 48
3
3-8 Disabling the Key Operation
Key Lock with PIN Number
A PIN number can be set when activating key lock. Only users knowing the PIN
number can operate the unit.
Activating key lock with a PIN number
Press (or ) 10 times while holding
1
down the [MODE] button.
Basic Operation
The screen displays "Loc 0".
Press the button to set a desired
2
number (up to 4 digits).
Press the [MODE] button.
3
The screen displays "Loc", disabling key
operation and displaying the current
received light intensity.
Deactivating key lock with a PIN number
Hold down the [MODE] button while
1
pressing (or ) 10 times.
The screen displays "Loc 0".
Press the button to specify the
2
PIN number, and then press the
[MODE] button.
The screen displays "unL", enabling key
operation.
3-24
Point
If the PIN number is lost, contact your nearest KEYENCE office.
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 49
Settings for Advanced Functions
This chapter describes settings for advanced functions of the LV-N10
Series.
4-1List of Settings .................................................... 4-2
- Digital Laser Sensor LV-N10 Series User's Manual -
4-1
Page 50
4-1
List of Settings
4
Basic Settings…(page 4-4)
5'6
5W24
7.64
/')#
J52
(KPG
5VF
5'6
5'6
'0&
5V)
5V)
5V)
MODE
5'V2
15'V
MODE
(W0E
&K52
5;5
TURBO mode
SUPER mode
ULTRA mode
MEGA mode
HIGH SPEED mode*
FINE mode
Normal sensitivity setting method
Percentage Calibration*
Zero-shift calibration
Settings complete
Go to detection setup mode
Go to display setup mode
Go to system setup mode
Press and hold for 3 seconds or more
6WT$
MODE
MODE
MODE
Settings for Advanced Functions
Return to normal display
Detection Settings (Func)…(page 4-7)
Timer OFF
Off-delay timer *
On-delay timer *
One-shot timer *
Normal (light intensity)
detection mode
DATUM1 mode *
DATUM2 mode *4 *
Area detection mode
Rising Edge Detection Mode *
Falling Edge Detection Mode *
8
*
9
*
Return to normal display
Alternate
display
External input off
External calibration input
Preset input
Zero shift input
Reset input
Light transmission OFF input
Pause function *
Sleep function
Analog scaling
MODE
Settings complete
Go to display setup mode
Go to system setup mode
Return to detection
setup mode
Display Settings (diSP)…(page 4-22)
…(page 4-4)
…(page 4-4)
…(page 4-4)
…(page 4-4)
…(page 4-4)
1
…(page 4-4)
…(page 4-4)
2
…(page 4-5)
…(page 4-6)
Return to normal display
Normal display method
Reverse display
Sub-display off
Extension display
Bar display
Excess gain (%) display
Light intensity hold display *
Excess gain hold (%) display *
L-on / D-on display
Enable the saturation of the
Preset function *
Disable the saturation of the
Preset function
Settings complete
Go to system setup mode
Go to detection setup mode
Return to display setup mode
…(page 4-22)
…(page 4-22)
…(page 4-22)
…(page 4-23)
…(page 4-23)
…(page 4-24)
10
…(page 4-24)
10
…(page 4-26)
…(page 4-26)
…(page 4-27)
11
…(page 4-27)
System Settings (SYS)…(page 4-29)
3
3
3
6
*7
…(page 4-7)
…(page 4-7)
…(page 4-7)
…(page 4-7)
…(page 4-8)
4 *5
…(page 4-9)
5
…(page 4-11)
…(page 4-14)
5
…(page 4-16)
5
…(page 4-16)
…(page 4-17)
…(page 4-17)
…(page 4-17)
…(page 4-17)
…(page 4-17)
…(page 4-17)
…(page 4-18)
…(page 4-19)
…(page 4-20)
*12
*13
Return to normal display
Disable Eco Function
Enable Eco Function
Reduce power consumption
(response time 4 times slower)
Standard current value display
Maximum current value display
(4 times hysteresis)
Normal operation
Twice the number of
interference-prevention units as
(response time 2 times slower)
Disable common key operations
Enable common key operations
Standard sensitivity
High resolution sensitivity
Settings complete
Go to detection setup mode
Go to display setup mode
Return to system setup mode
…(page 4-29)
…(page 4-29)
…(page 4-29)
…(page 4-30)
…(page 4-30)
…(page 4-31)
56&
…(page 4-31)
…(page 4-32)
…(page 4-32)
4-2
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 51
4-1 List of Settings
2-Output ()
Return to normal display
*1 hsp cannot be selected when LV-S31/S62/S63 is connected.
*2 You can press the button to set between the range of -99P and 99P.
*3 Press the button to set between the range of 1 and 9999 (ms).
*4 Press the button to set the adj ustment sensitivity to a range of between LEu1 and LEu3 and set the war ning output level to a range of
between 0P and 100P.
*5 Cannot be selected when LV-S31 is connected.
*6 Press the button to switch between oFF /on/KEEP.
*7 The analog output upper limit can be set between 50 and 65535.
*8 Not available for zero line types.
*9 Monitor output type only (LV-N11MN).
*10 Press the button to toggle between Std/P~P_/b~b_/P_b~/P~b_.
*11 Press the button to set between the range of 100P and 200P.
*12 Not displayed when LV-S31 is connected.
*13 Main unit only.
*14 Only 2-output types (LV-N11 / N12).
*15 Press the button to select the User (user reset) or Auto (auto reset) resetting methods.
*16 Select the count method from oUr1, oUr2 or Auto. When Auto is selected, the one shot time can be set between 1 and 99 99 (ms).
Reference
*14
…(page 4-33)
Light intensity detection mode
Limit setting output mode *
Warning output mode
Counter output mode *
Timer OFF
Off-delay timer *
On-delay timer *
One-shot timer *
Settings complete
MODE
MODE
MODE
MODE
MODE
MODE
MODE
• Press the button and the button simultaneously to return to the
…(page 4-37)
15
…(page 4-34)
…(page 4-35)
16
…(page 4-35)
…(page 4-37)
3
…(page 4-37)
3
…(page 4-37)
…(page 4-37)
3
MODE
previous setting option.
• When the button is held down, the settings menu will end and the
MODE
received light intensity setting screen will appear.
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-3
Page 52
4
4-2
Basic Settings
Power Modes
Refer to page 4-2 "Basic Settings" for setting methods.
The detection stability will increase when the response time is delayed.
You can select from the following 6 power modes.
The response time slows down in the order of hSP, FinE, turb, Supr, ULtr, MEGA.
DisplayPower mode nameResponse time
*
hSP
FinEFINE mode250 øs
turbTURBO mode (Default)500 øs
SuPrSUPER mode1 ms
Settings for Advanced Functions
* Cannot be selected when the LV-S31/S62/S63 sensor head is connected.
ULtrULTRA mode4 ms
MEGAMEGA mode16 ms
HIGH SPEED mode80 øs
Sensitivity Setting
Refer to page 4-2 "Basic Settings" for setting methods.
You can select from the following 3 sensitivity setting methods.
- Digital Laser Sensor LV-N10 Series User's Manual -
Reference
page
Page 53
Percentage calibration
4-2 Basic Settings
The setting value can be specified as a
percentage in respect to the current received
light intensity.
The percentage calibration target value can be
set in the range of "-99P" (-99%) to 99P (99%)
Default: -10P (-10%)
Sensitivity setting
In the desired reference state, press
1
the [SET] button.
Calibration is complete after the setting
value flashes momentarily, and then
stops (lights up).
Reference
• If the external input setting is set to "SEt" (external calibration input), periodic percentage calibration is possible from external devices, enabling
stable detection even on workpieces having small sensitivity differences.
"External Input" (page 4-17)
• If the detection mode is set to "ArEA (area detection mode)", Hi and Lo
are set simultaneously according to the received light intensity.
(Example)
If the percentage calibration is set
to 10P (10%) when the reference
received light intensity is "100",
the Hi and Lo setting values will be
set as follows.
Hi setting value: 110
Lo setting value: 90
When the percentage calibration
Setting example
target value is set to "-10%".
Received light intensity
1000
900
Hi110
Received
light intensity
Reference
4
Settings for Advanced Functions
100
90Lo
- Digital Laser Sensor LV-N10 Series User's Manual -
4-5
Page 54
4-2 Basic Settings
Zero shift calibration
This sensitivity setting performs the zero shift function and basic calibration (two-point
calibration / maximum sensitivity calibration / full auto calibration) simultaneously.
The lower of the received light intensity values specified at the time of sensitivity
setting will automatically be set to "0".
4
Reference
The light intensity received with a workpiece present will be applied with the
same amount of correction as for that received with no workpiece.
The following basic calibrations can be set while in the zero shift calibration.
• Two-point calibration
• Maximum sensitivity calibration
Settings for Advanced Functions
• Full auto calibration
Setting example
If two-point calibration is performed with a reflective model when the light
intensity received with a workpiece is "2500" and without a workpiece is
"500":
After two point calibration
(
Before zero shift calibration
Received light intensityReceived light intensity
500 2500500 250002000
)
Setting
value
1500
The light intensity received
with no workpiece is
subtracted from the whole.
Zero shift calibration is executed
The state with no
workpiece (the lowest of
received light intensity) is
set to "0".
Setting
value
1000
4-6
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 55
4-3
Detection Settings (Func)
Output Timer
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
There are 3 types of timers available.
DisplayFunction
toFFUses no output timer. (Default)
Off-delay timer
Turns off the output at a specified
oFFd
on-d
Shot
* Example of the light-ON (L-on) mode for the reflective model and the dark-ON (d-
on) mode for the thrubeam/retro-reflective
time after the detection signal
goes off.
Setting range: 1 to 9999 ms
Default: 10 ms
On-delay timer
Turns on the output at a specified
time after the detection signal
goes on. Useful if the ON duration
is limited because the workpiece
is shaky such as by vibration.
Setting range: 1 to 9999 ms
Default: 10 ms
One-shot timer
Turns on the output and keeps it
on for a specified period after the
detection signal goes on.
Setting range: 1 to 9999 ms
Default: 10 ms
Control output
Control output
Control output
Control output
Timing chart
ON
OFF
ON
OFF
ON
OFF
Timer-set period
ON
OFF
Timer-set period
*
With workpiece Without workpieceWithout workpiece
With workpiece Without workpieceWithout workpiece
Timer-set period
With workpiece Without workpieceWithout workpiece
With workpiece Without workpieceWithout workpiece
4
Settings for Advanced Functions
Reference
• When using the 2-output type, different timer functions and timer times
can be set separately for channels 1 and 2.
"Output Timer for Output2" (page 4-37)
• The timer can only control on/off switching of the sensor output.
- Digital Laser Sensor LV-N10 Series User's Manual -
4-7
Page 56
4-3 Detection Settings (Func)
Detection Mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
The table below lists the detection modes that can be selected.
4
DisplayDetection modeFunction
Normal detectionNormal detection mode (Default)-
Std
The received light intensity when
there is no workpiece is always cali-
dtM1
DATUM1 mode
*1 *2
Settings for Advanced Functions
dtM2
ArEA
_|~d
~L_d
DATUM2 mode*1
Area detection mode
Rising edge detection
*2
mode
Falling edge detection
*2
mode
brated to "
value is also calibrated so that the setting value and received light intensity
ratio is always constant.
The received light intensity when
there is no workpiece is always calibrated to "
*2
also calibrated so that the setting
value and received light intensity ratio
is always constant.
Detection is made only when the
received light intensity goes out of a
given range.
Detection is made only when the
received light intensity increases.
Detection is made only when the
received light intensity decreases.
100.0", and the setting
.0", and the setting value is
Reference
page
4-9
4-11
4-14
4-16
4-16
*1 In the DATUM mode, correction intervals can be set in the range of LEv1 to LEv3.
The default is LEv1.
Also, the warning output level can be set in the range of 0P (0%) to 100P (100%).
The default is 50P.
"Adjusting the correction interval" (page 4-12)
"Changing the warning output level" (page 4-13)
*2 This cannot be selected when LV-S31 is connected.
4-8
Point
• Refer to "Detection Mode for Output2" (page 4-33) for details on
the 2-output type channel 2 detection mode.
• Refer to "Restrictions on Each Detection Mode" (page 6-8) for
restrictions on the sensitivity setting methods.
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 57
4-3 Detection Settings (Func)
p
)
DATUM1 mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In DATUM1 mode, the light intensity received
without a workpiece is always corrected to
"100.0". The setting value is also corrected
according to the correction amount, so that the
ratio of setting value and received light intensity
is kept constant, with the result of stable
detection. The display of the setting value does
not change.
The DATUM mode is effective in environments where received light intensity varies
gradually, for example where the sensor head is prone to contamination or large
temperature changes.
Sensitivity setting in DATUM1 mode
The sensitivity setting procedure
below is an example of two-point
calibration (where the received
light intensity is "100.0" when a
workpiece is absent and "20.0"
when a workpiece is present)
using the thrubeam/retro-reflective
model.
Press the [SET] button with no work-
1
piece present.
100.0
Setting
value
60.0
20.0
0
Received light intensity
Calibration point (Step 1)
1
2
Calibration
Fixed proportion
oint (Step 2
The DTM indicator
lights up.
"100.0"
"5.0"
(Initial value)
4
Settings for Advanced Functions
100.0
60.0
Press the [SET] button with a work-
2
piece present.
The light intensity received in full-light
receiving state is displayed as "100.0".
- Digital Laser Sensor LV-N10 Series User's Manual -
Display changes
.".
to "
4-9
Page 58
4-3 Detection Settings (Func)
Point
• When the received light intensity stays lower than a given warning
level, the received light intensity correction stops and the DTM
indicator flashes.
"Changing the warning output level" (page 4-13)
• The light intensity correction stops if the raw received light intensity drops below 50 (200 when the light intensity expansion setting
is set to FuLL). The raw received light intensity can be confirmed by
pressing the [MODE] button for longer than 3 seconds on the
Basic screen.
"Display Gain" (page 4-30)
• This cannot be selected when LV-S31 is connected.
4
Settings for Advanced Functions
Reference
• By pressing the [PRESET] button,
the current value (received light
intensity) can be forcibly calibrated
to "100.0". Calibration is completed
when the current value stops flashing. (At this time, the setting value is internally calibrated so that the
received light intensity and setting value ratio stay constant.)
• You can also set the interval for executing correction, and/or the level of
received light intensity change for issuing a warning.
"Adjusting the correction interval" (page 4-12)
"Changing the warning output level" (page 4-13)
• The saturation level is fixed to 101P (101%). To display a value higher
than 100.0, turn the preset saturation function OFF.
"Preset Saturation Function" (page 4-27)
4-10
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 59
4-3 Detection Settings (Func)
DATUM2 mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In DATUM2 mode, the light intensity received
without a workpiece is always corrected to
".0". The setting value is also corrected
according to the correction amount, so that
the ratio of setting value and received light
intensity is kept constant, with the result of
stable detection. The display of the setting
value does not change.
The DATUM mode is effective in environments where received light intensity varies
gradually, for example where the sensor head is prone to contamination or large
temperature changes.
Sensitivity setting in DATUM2 mode
The sensitivity setting procedure
below is an example of two-point
calibration (where the received light
intensity is ".0" when a workpiece is
absent and "80.0" when a
workpiece is present) using the
thrubeam/retro-reflective model.
Press the [SET] button with no work-
1
piece present.
Setting
value
40.0
80.0
100.0
Received light intensity
Fixed proportion
Calibration point (Step 1)
1
.0
2
Calibration point (Step 2)
The DTM indicator
lights up.
".0"
"5.0"
(Initial value)
4
Settings for Advanced Functions
.0
40.0
Press the [SET] button with a work-
2
piece present.
The light intensity received in the full-light
receiving state is displayed as ".0".
- Digital Laser Sensor LV-N10 Series User's Manual -
Display changes
.".
to "
4-11
Page 60
4-3 Detection Settings (Func)
Point
• When the received light intensity stays higher than a given warning level, the received light intensity correction stops and the DTM
indicator flashes.
"Changing the warning output level" (page 4-13)
• The light intensity correction stops if the raw received light intensity drops below 50 (200 when the light intensity expansion setting
is set to FuLL). The raw received light intensity can be confirmed by
pressing the [MODE] button for longer than 3 seconds on the
Basic screen.
"Display Gain" (page 4-30)
• This cannot be selected when LV-S31 is connected.
4
Settings for Advanced Functions
Reference
Adjusting the correction interval
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
A desired correction interval can be selected from the 3 levels:
HIGH SPEED4.2 s520 ms33 ms
FINE/TURBO/SUPER6.6 s820 ms51 ms
ULTRA/MEGA13 s1.6 s100 ms
Reference
• By pressing the [PRESET] button,
the current value (received light
intensity) can be forcibly calibrated
to "0". Calibration is completed
when the current value stops flashing. (At this time, the setting value is internally calibrated so that the
received light intensity and setting value ratio stay constant.)
• You can also set the interval for executing correction, and/or the level of
received light intensity change for issuing a warning.
"Adjusting the correction interval" (page 4-12)
"Changing the warning output level" (page 4-13)
Power mode
LEv1 (LEVEL1) is recommended.
LEv1
(LEVEL1)
LEv2
(LEVEL2)
LEv3
(LEVEL3)
4-12
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 61
4-3 Detection Settings (Func)
Changing the warning output level
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
A warning is output
when the amplifier
detects variation of the
received light intensity
in the range where the
output does not turn
on.
When the received light
intensity stays between
the warning level and
Display example
When the warning output level is set to 50%, dtm1, D-on:
Received light intensity
50%
100%
100.0
75.0
50.0
A warning is output if the received light
intensity, when no workpiece exists, does
not increase to the warning level because of
beam axis deviation, etc.
Warning output level
Setting value
set value, the received
light intensity calibration stops, the DTM
indicator flashes, and a
warning is output.
A level for outputting a
warning can be set in
Warning output
Control output
0
ON
OFF
ON
OFF
the range of 0P (0%) to
100P (100%).
(Default: 50P (50%)).
• 100P (100%) : A warning is not output even if the received light intensity
changes.
• 50P (50%): A warning is output if the received light intensity drops to the middle between the received light intensity when no workpiece is
present and the setting value.
• 0P (0%): A warning is output if the received light intensity drops even
slightly.
4
Settings for Advanced Functions
Reference
The output type (LV-N11 / N12) can output the warning state to an external device with the 2-output setting ().
For details, see "Warning output mode" (page 4-35).
- Digital Laser Sensor LV-N10 Series User's Manual -
4-13
Page 62
4
4-3 Detection Settings (Func)
Area detection mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In this mode, the output signal is
delivered (in N.O. condition) if the
received light intensity is in the area
defined by Hi (upper limit) and Lo
(lower limit).
Settings for Advanced Functions
Hi (Upper limit
setting value)
Lo (Lower limit
setting value)
Control output
in N.O. condition
Control output
in N.C. condition
0
ON
OFF
ON
OFF
4-14
Reference
Point
• Output-1 can be configured to be ON (N.O.) or OFF (N.C.) when the
received light intensity is above the Lo and under the Hi settings.
"Output Switch (L-on/D-on)" (page 3-4)
• Area detection can be set just for output-1 scale. Output-2 scale operates
according to the "Output 2 detection mode".
"Detection Mode for Output2" (page 4-33)
Always ensure that "upper limit setting value > lower limit setting
value". If "upper limit setting value ≤ lower limit setting value", the
received light intensity will always be output outside the Hi and Lo
range.
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 63
4-3 Detection Settings (Func)
Methods for setting upper and lower limits
The upper and lower limits can be set using one of the 2 methods:
● Setting sensitivity using the percentage calibration mode
See "Percentage calibration" (page 4-5).
●Sensitivity setting using other calibration modes
Sensitivity is set separately for Hi and Lo.
Press to display "Hi" or "Lo".
1
The "Hi" or "Lo" value, and the current setting values are displayed alternately.
When the [MODE] button is pressed while
the display alternately flashes, the "Hi" or
"Lo" display changes.
If you do nothing for 3 seconds or more,
the received light intensity display is automatically restored.
Press the [SET] button to perform sen-
2
sitivity setting (two-point calibration,
maximum sensitivity setting, or other).
4
Settings for Advanced Functions
Calibration is complete after the setting
value flashes momentarily, and then stops
(lights up).
- Digital Laser Sensor LV-N10 Series User's Manual -
4-15
Page 64
4
4-3 Detection Settings (Func)
Edge detection mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In the edge detection mode, the sensor amplifier detects changes of received light
intensity over a fixed period.
This mode is used to turn on the output only when a workpiece is moving into or out of
the detection range of the sensor head.
• To turn the output ON when the received light intensity drops:
Select "~L_d" (Falling edge detection mode)
• To turn the output ON when the received light intensity increases:
Select "_|~d" (Rising edge detection mode)
Timing chart
Operation example
Settings for Advanced Functions
(1), (3):
(2):
(4):
When the falling edge detection mode is selected with a thrubeam model:
The output is OFF because the
received light intensity is not
changing.
The output is ON because the
received light intensity is decreasing.
The output remains OFF because
the received light intensity is
changing but is increasing.
Point
• Combining the edge detection mode and any of the functions below
• If the interference prevention is "dobL" (double):
• If the eco mode is "FuLL" (Full):
• This cannot be selected when LV-S31 is connected.
Received light intensity
(1)(2)(3)(4)
ON
Control output
OFF
makes it more difficult to detect gradual changes in light intensity.
"Interference Prevention" (page 4-31)
"Power Save" (page 4-29)
Sensitivity setting
Press the [SET] button with no workpiece present.
1
The amplifier is now configured to ignore
changes in received light intensity for the duration that the [SET] button is pressed.
Calibration is complete after the setting value
flashes momentarily, and then stops (lights up).
Reference
• When detecting workpieces running on a conveyor, the [SET] button can be pressed for a longer period
to ignore the fluctuations of received light intensity due to conveyor vibration.
• If the setting value is so low that objects other than the workpiece of interest
are also detected, fine adjustments can be made with the button.
4-16
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 65
4-3 Detection Settings (Func)
External Input
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
On sensor amplifiers with an external input option (LV-N11/N12/N11C/ N12C/
N11MN), the external input line can be controlled to utilize the following functions:
DisplayFunction
oFFDisables use of the external input function. (Default)
SEt
PrSt
ShFt
rSt
LoFF
PAUS
SLEP
External calibration input
Uses the external input to perform calibration.
Preset input
*1
Executes a preset at the rising edge of the external input.
"Preset Function" (page 3-6)
Zero shift input
Executes a zero shift at the rising edge of the external input.
*2
"Setting the Current Received Light Intensity to 0 (Zero Shift)"
(page 3-17)
Reset input
Resets the display at the rising edge of the external input when using
the hold function, limit setting detection, or counter output mode.
"Light intensity hold display" (page 4-24), "Excess gain hold (%)
display" (page 4-26)
"Limit setting output mode" (page 4-34)
"Counter output mode" (page 4-35)
Transmission OFF input
Keeps the laser transmission off during signal input.
Pause mode transition input
Locks the output condition during external input.
"Pause mode transition input" (page 4-18)
Sleep mode transition input
Uses the power save mode during external input.
"Sleep mode transition input" (page 4-19)
4
Settings for Advanced Functions
When the preset input is selected, the zero shift function is disabled and cannot be set.
*1
*2 When the zero shift input is selected, each preset function is disabled and cannot
be set.
Reference
• For external input cabling, see "Wiring Diagrams for Sensor Amplifiers" (page 2-4).
• Provide a short-circuit time of 2 ms or more, and 25 ms or more when
external calibration input is selected.
- Digital Laser Sensor LV-N10 Series User's Manual -
4-17
Page 66
4-3 Detection Settings (Func)
Pause mode transition input
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
4
During signal input, the output can
be locked in a desired state
regardless of received light input.
There are 3 selections of output
Reflective model
Received light intensity
L-on
Signal input
Signal input
status that can be locked with the
pause function.
Settings for Advanced Functions
Pause OFF
Paus e ON
Pause KEEP
Output
Output
Output
ON
OFF
ON
OFF
ON
OFF
DisplayFunction
oFFThe output is locked in OFF state during signal input. (Default)
onThe output is locked in ON state during signal input.
KEEP
The output is locked in its existing state when a signal is received
from the external input.
4-18
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 67
4-3 Detection Settings (Func)
Sleep mode transition input
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
This function holds the sensor amplifier in the power save state (sleep mode) during
input of the external signal.
Once the sleep mode is entered, the following events occur.
• The display of received light intensity and setting value goes out.
• Light transmission stops.
• The output is turned off (for both L-on and Don)
• A single segment on the digital display pulses across the display.
The normal display is restored after any key is pressed.
If no operation is made for the next 4 seconds, the display unit turns back off.
Parameter Save
If "SEt" (external calibration input), "PrSt" (preset input) or "ShFt" (zero shift input) is
selected as an external input and the "Parameter Save by Ext-Input" option is set to
"no", the settings changed with external input will not be written to the EEPROM.
This prevents frequent writing to the EEPROM when inputting external signals and
thus prevents the EEPROM from reaching the end of its service life (approx. 1 million
writes).
4
Settings for Advanced Functions
When received light intensity is dis-
1
played, press and hold the [MODE],
[SET], and buttons simultaneously
for 3 seconds or more.
To enable the restriction of writing settings,
use the button to select "no".
Press and hold the [MODE] button.
2
The current received light intensity is displayed.
Reference
• When "no" is set, the values changed with external inputs will not be
changed even after the power is turned off and on.
• When "yes" (default) is set, the settings are written to the EEPROM
approx. 3 seconds after the settings are changed with external inputs.
- Digital Laser Sensor LV-N10 Series User's Manual -
4-19
Page 68
4-3 Detection Settings (Func)
y
Analog scaling (LV-N11MN only)
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
The LV-N11MN is equipped with an analog voltage output.
In the default state, an analog output can be made in the range of 1 to 5 V in respect
to the current received light intensity 0 to 4000.
The received light intensity corresponding to the 5 V analog voltage can be changed
within the following range.
4
Received light intensity setting range corresponding to 5 V: 50 to
Received light intensity setting range corresponding to 1 V: Fixed to
Settings for Advanced Functions
Output voltage
5V
1V
50
4000 (default value)
Point
• The received light intensity calibrated with each preset function or
each DATUM mode is not applied on the analog output.
• The analog scaling mode can be set so that the received light
intensity calibrated with each preset function is applied onto the
analog output.
"Analog scaling mode (LV-N11MN only)" (page 4-21)
• If the [PRESET] button is pressed on the analog scaling setting
screen when each preset function is enabled, the analog scaling
value is automatically set so that 5V is output in respect to the current value "100.0" at that point.
• The received light intensity calibrated with the zero shift function
is applied on the analog output.
65535
65535
(defa ult value:
0
Received light intensit
4000
)
4-20
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 69
4-3 Detection Settings (Func)
Analog scaling mode (LV-N11MN only)
The analog output can be automatically scaled so that the 5 V is output in respect to
the received light intensity "100.0". This function is used simultaneously with the
various preset function.
This function is useful when frequently using the preset function as the analog output
can be constantly sequenced with the ".0 to 100.0" display, etc.
When received light intensity is dis-
1
played, press and hold the [MODE] button, [SET] and button
simultaneously for 3 seconds or more.
ProG YES or ProG no is displayed.
Press the [MODE] button once or twice.
2
Using the button, select Pr-AStd
3
to not change the analog upper limit
value when the preset function is executed, and select Pr-A PrSt to change
the analog upper limit value sequentially when the preset function is executed. Then, hold down the [MODE] button.
Reference
When "Std" (default) is set, the analog output upper limit will not change
even if the various preset function is used.
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-21
Page 70
4-4
Display Settings (diSP)
Display Reverse
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
4
The current value and setting value
displays can be inverted.
DisplayFunction
oFF(OFF)
on(ON)Reversed display
Normal display (Default)
Normal display
Reversed display
Sub Display
Settings for Advanced Functions
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
In addition to the received light intensity and setting value, the following information
can be displayed on the sub display.
Operation
DisplayInformation
oFF(OFF)
StdE
(Extension)
bAr(Bar)
PEr(Percent)
HLd(Hold)
HLdP
(Percent Hold)
Ldon(L-on/D-on)
Yes: Operation is possible. No: Operation is not possible.
*1
For external calibration, the [SET] button can be operated to set values up to "9999".
*2
The displayed value is that before each preset function or zero shift function is executed.
*3 If the detection mode is set to DATUM1 mode or DATUM2 mode, the value before
DATUM correction is displayed.
*4 If the detection mode is set to DATUM1 mode or DATUM2 mode, the excess gain
hold display cannot be selected.
*5
If the detection mode is set to the area detection mode, rising edge detection mode,
falling edge detection mode, or when the LV-S31 is connected,
*6 When set to the edge detection mode, bar display, excess gain display, and excess
gain hold display cannot be selected.
*2
*6
*3
4-22
No sub display (Setting value /
received light intensity) (Default)
Extension display (Displaying up to
5 digits)
Bar display (Bar excess gain / received
light intensity)
Excess gain(%) display (Bar excess
*6
gain / excess gain)
Received-light intensity hold display
(5 types selectable)
Excess gain hold(%) display (5 types
*4*6
selectable)
L-on/D-on display (no/nC display)
- Digital Laser Sensor LV-N10 Series User's Manual -
*5
buttons
[SET]
Ye sYe s-
*1
No
YesYes4-23
YesYes4-24
Ye sN o4 - 2 4
Ye sN o4 - 2 6
Ye sN o4 - 2 6
no/nC
Reference
page
No4-23
is displayed.
Page 71
Displaying a sub display
After selecting the
desired sub-screen display, press the [MODE]
button twice in the normal
display state (setting
value / received light
intensity).
Press the [MODE] button
once again to return to
the normal display.
Setting value/
Received light intensity
4-4 Display Settings (diSP)
MODE
No operation for 3
sec or more
MODEMODE
Sub display
Select L-on/D-on
Extension display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The received light intensity usually displayed with 4 digits can be extended to a 5-digit
display.
The maximum received light intensity value which can be displayed differs according
to the power mode and display gain. Refer to "Display Gain" (page 4-30) for more
details.
Reference
• The upper limit of the setting value remains a 4-digit value (9999). A
larger value cannot be set.
• Each preset function, zero shift function, and each calibration function
cannot be used during the extension display.
• The received light intensity calibrated with each preset function, zero shift
function, each DATUM mode or edge detection mode is not applied on
the extended display.
Bar display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
Received light intensity can be displayed as a
bar graph.
Bars are displayed in real time in the range of
85% to 115% at intervals of 5%, where 100%
is the setting value.
Up to 8 bars are displayed.
If the setting value is 1000:
Display example
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-23
Page 72
4-4 Display Settings (diSP)
Excess gain (%) display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
4
Received light intensity is displayed as an excess
gain* relative to the setting value. On the screen,
the received light intensity is displayed as a bar
graph and a percentage number.
* Excess gain = (Received light intensity/Setting
value) x 100
Light intensity hold display
Settings for Advanced Functions
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The maximum peak values, and the minimum bottom values, of received light intensity
can be displayed continuously. You can select from 5 possible display combinations:
DisplayDisplayed values
Std (Standard)
P~P_ (Peak Max/Peak Min)
b~b_ (Bottom Max/Bottom Min)
P_b~ (Peak Min/Bottom Max)
P~b_ (Peak Max/Bottom Min)
How to reset the held values
To reset the peak and/or bottom values that are currently held, use one of the following
procedures:
If the setting value is 1000
Display example
and the received light
intensity is 2000 (200%):
Updates the peak or bottom value each time the current received light intensity falls below or rises above,
respectively, the setting value. (Default)
Displays the maximum and minimum of peak values
since the power is turned on. (Cumulative)
Displays the maximum and minimum of bottom values since the power is turned on. (Cumulative)
Displays the minimum of peak values and the maximum of bottom values since the power is turned on.
(Cumulative)
Displays the maximum of peak values and the minimum of bottom values since the power is turned on.
(Cumulative)
• Press and hold the [MODE] and [SET] buttons simultaneously.
• Set the external input (in) to "reset input (rSt)" and short-circuit the pink wire (pin
(2)). "External Input" (page 4-17)
• Turn off the power.
4-24
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 73
Timing chart for standard setting (Std)
[Peak value]
• Samples while the current received light intensity > setting value ((1), (3), etc.).
• If the current received light intensity is lower
than the setting value, the previous peak value
is updated.
[Bottom value]
• Samples while the current received light intensity < setting value ((2), (4), etc.).
• If the current received light intensity is higher
than the setting value, the previous bottom
value is updated.
Timing chart for peak value
[Cumulative maximum peak value]
• Samples while the current received light intensity > setting value ((1) to (4)).
• If the current received light intensity is lower than
the setting value, the previous peak value and
current peak value are compared. If the current
peak value is higher, the value is updated.
[Cumulative minimum peak value]
• Samples while the current received light intensity > setting value ((1) to (4)).
• If the current received light intensity is lower than
the setting value, the previous peak value and
current peak value are compared. If the current
peak value is lower, the value is updated.
4-4 Display Settings (diSP)
Peak value
Received light
intensity
Setting value
Received light intensity
Received light intensity
Bottom value
(1)(3)
(2)
Cumulative maximum
of peak
Received light
intensity
Setting value
Cumulative minimum
of peak
(2)(1)(3)(4)
(4)
4
Settings for Advanced Functions
Timing chart for bottom value
[Cumulative maximum bottom value]
• Samples while the current received light intensity < setting value ((1) to (4)).
• If the current received light intensity is higher
than the setting value, the previous bottom
value and current bottom value are compared.
If the current bottom value is higher, the value
is updated.
[Cumulative minimum bottom value]
• Samples while the current received light intensity < setting value ((1) to (4)).
• If the current received light intensity is higher
than the setting value, the previous bottom
value and current bottom value are compared.
If the current bottom value is lower, the value
is updated.
- Digital Laser Sensor LV-N10 Series User's Manual -
Received
light intensity
(1)(3)
Cumulative maximum
of bottom
Received light intensity
Setting value
Cumulative minimum
of bottom
(2)
(4)
4-25
Page 74
4-4 Display Settings (diSP)
Excess gain hold (%) display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
This function is the same as the "received-light intensity hold display", except that the
value being held is displayed as excess gain instead of received light intensity.
"Light intensity hold display" (page 4-24)
L-on / D-on display
4
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
Settings for Advanced Functions
The current output condition (L-on or D-on) can
be displayed when the setting value is displayed.
The received light intensity is displayed in terms
of received light intensity as usual.
If the detection mode is set to the area detection mode, rising edge detection mode,
falling edge detection mode, or when the LV-S31 is connected, no/nC is displayed.
4-26
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 75
4-4 Display Settings (diSP)
Preset Saturation Function
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The saturation level can be set in the range of 100P (100%) to 200P (200%) when the
preset saturation function is ON.
For example, if this function is set to 150P, the received light intensity will be corrected
to "150.0" when the preset function is enabled, and the digital display will display
"100.0".
The saturation level can be set to a higher value for stable detection even in
environments which are prone to some contamination, such as dust.
Turn off this function if you want to display values above "100.0" in applications where
the preset function or the DATUM1 mode is used.
DisplayFunction
Fixes the display to "100.0" (displays no value higher than "100.0")
on (ON)
oFF (OFF)
Point
when the actual received light intensity is above 100.0. (Default:
110P)
Displays the actual received light intensity as is, even when it is
above "100.0".
Setting too high of a saturation level (specifying too large a value)
decreases the detectability for small differences in light intensity or
transparent workpieces.
4
Settings for Advanced Functions
*1 When the detection mode is set to DATUM1, the preset saturation function ON/
OFF state can be selected, but the saturation level will be fixed to 101P (101%),
and the setting screen will not appear.
*2 When the detection mode is set to DATUM2, the preset saturation function set-
ting screen will not appear.
*3 If the preset display mode is set to ArEA when the LV-NH100/NH110/NH300 is
connected, the saturation level is fixed to oFF, and the preset saturation function
setting screen will not appear.
Reference
• Refer to "Adjusting the Sensitivity" (page 3-5) for execution of preset.
• Refer to the next page for the detailed operating principle.
- Digital Laser Sensor LV-N10 Series User's Manual -
4-27
Page 76
4
4-4 Display Settings (diSP)
Operating principles of the preset function
(Current received light intensity is 3300 in this example.)
Preset input is used to register the
light intensity as "100.0".
The actual received light intensity (3300)
is set to "110%" and all actual received
light intensities exceeding "100%" are
displayed as "100.0".
When light intensity falls below 100%, a
value lower than "100.0" is displayed.
(The percentage value 110% (i.e., the
saturation level) at the time of registration
can be any value between 100 and
Settings for Advanced Functions
200%.)
"Preset Saturation Function" (page 4-
27)
Received light intensity
4095
3300
110.0
100.0
3000
2700
0
(After preset input)
All values in this
range are 100.0.
90.0
If the displayed light intensity falls below
"100.0", the preset input can be performed again to restore stable detection.
If light intensity is not displayed as "100.0"
after preset input, the sensor amplifier is not
in stable detection condition.
Reference
Even when the maximum sensitivity preset function or full auto preset function is used, the preset saturation level setting function operates based on
the "100.0" point.
Received light intensity
4095
3300
2970
110.0
100.0
2700
0
(After another preset input for correction)
All values in this
range are 100.0.
4-28
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 77
4-5
System Settings (SYS)
Power Save
Refer to page 4-2 "System Settings (SYS)" for setting methods.
Power consumption can be reduced by turning off the digital display and output
indicators.
DisplayFunction
oFF (OFF)Disables the power save function. (Default)
on (ON)
FuLL (FULL)
ALL (ALL)
Immediately after the ON or FULL option is selected, the option is enabled, and the
power save display results.
The normal display is restored after any key is operated (except for eco ALL).
If no operation is made for the next 30 seconds, the power save display returns.
Enables the power save function. (A single segment
pulses across the digital display.)
Reduces power consumption. (The power save function is
ON, and response time is 4 times longer.)
Eco ALL (The digital display and output indicators are all
off).
This item becomes selectable by pressing and holding
while FuLL is lit.
The power consumption in Eco ALL is the same as that in
Eco ON.
4
Settings for Advanced Functions
Point
When FuLL is selected, response time is 4 times as long as usual (i.e.,
as compared to when oFF or on is selected).
When ALL is selected, response time is the same as usual.
Deactivating eco ALL
Press and hold the [MODE] and [PRE-
1
SET] buttons simultaneously.
The screen returns to the display of current received light intensity.
If no operation is made for the next 30 seconds, the eco ALL state is restored.
- Digital Laser Sensor LV-N10 Series User's Manual -
4-29
Page 78
4
4-5 System Settings (SYS)
Display Gain
Refer to page 4-2 "System Settings (SYS)" for setting methods.
Received light intensity can be increased 4 times without sacrificing response time.
FuLL (Full)Full light intensity display (Hysteresis increases 4 times)
The maximum received light intensity value which can be displayed differs according
to the power mode setting and display screen.
The maximum received light intensity values which can be displayed are shown below.
• This function is not supported in the hSP (HIGH SPEED) mode and
the light intensity expansion setting screen is not displayed.
• When the preset function is used in DATUM1 or DATUM2 mode,
correctible values are limited to 200.
• This cannot be selected when LV-S31 is connected.
4-30
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 79
4-5 System Settings (SYS)
Interference Prevention
Refer to page 4-2 "System Settings (SYS)" for setting methods.
Malfunction may result from "interference", a phenomenon in which light transmission
occurs simultaneously with any other sensor amplifier.
However, when the number of sensor amplifiers increases, the light transmission
timing automatically shifts and prevents this interference.
Setting this function to "dobL" (Double) on all of the connected amplifier units doubles
the number units that do not mutually interfere.
DisplayFunction
Std (Standard)Normal operation (Default)
dobL (Double)
Point
Reference
• Selecting "dobL" also doubles response time over the Std state.
• When selecting "dobL", the main unit and its connected expansion
units must all be set to "dobL".
• This function does not work if the eco function is set to
"FuLL"(Full).
• If the number of interference prevention units is set to Std, the number of
units that do not interfere when the LV-N10 Series main unit and expansion units connected will be as follows. If set to dobL, the number of units
that do not interfere will be doubled.
When unit other than LV-S31 is connected: 0 units for HIGH SPEED, 2
units for FINE/TURBO/SUPER, and 4 units for ULTRA/MEGA.
When LV-S31 is connected: 2 units for FINE, and 4 units for TURBO/
SUPER/ULTRA/MEGA.
• Consult with KEYENCE for the interference prevention when using the
LV-N10 Series together with other series.
The interference prevention doubles over the Standard
state.
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-31
Page 80
4-5 System Settings (SYS)
Common Key-Operations Function
Refer to page 4-2 "System Settings (SYS)" for setting methods.
Some operations on the main unit are applied equally and simultaneously to all
expansion units connected to it.
DisplayFunction
oFF (OFF)Disables common key operations. (Default)
on (ON)Enables common key operations.
4
The operations that can be applied to all expansion units are as follows:
Settings for Advanced Functions
• Enabling/disabling the preset function
"Preset Function" (page 3-6)
• Enabling/disabling the zero shift function
"Zero Shift Function" (page 3-17)
• Enabling/Disabling the key lock function
"Key Lock" (page 3-23)
Reference
This function is effective only with the main unit to which expansion units
are connected. This function cannot be used with an independent main unit
or with the expansion units.
Sensitivity
Refer to page 4-2 "System Settings (SYS)" for setting methods.
When set to rESo, the hysteresis during detection will decrease. This is useful for
detecting workpieces with little received light intensity difference such as when
detecting a transparent workpiece with a retro-reflective model.
DisplayFunction
Std (Standard)Standard (Default)
rESo (Reso)High resolution
4-32
Point
Note that when the sensitivity is set to rESo and the hysteresis drops,
any cause of received light intensity variation such as vibration could
result in malfunction.
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 81
4-6
* This function is only available for the 2-output type models (LV-N11 / N12).
2-output Settings ()
Detection Mode for Output2
Refer to page 4-3 "2-Output Settings ()" for setting methods.
*1 If output 2 is set to Std, the DATUM1, DATUM2, area detection, and edge detection
modes are disabled in output 2.
*2 You can select the USEr (user reset) or Auto (auto reset) resetting methods after
outputting the limit setting under the following restriction:
- If output 1 is set to the DATUM1 mode or DATUM2 mode when LiMt is selected,
the detection will be performed with the value prior to DATUM calibration.
- If output 1 is set to the area detection mode or edge detection mode, LiMt cannot be selected with output 2.
*3 You can select the count method from oUr1, oUr2 or Auto.
When Auto is selected, the one-shot time can be set within the range of 1 to 9999
(ms).
*1
(Default)
*2
*3
Limit setting output mode (Output 2 turns on if received
light intensity for output 1 decreases).
"Limit setting output mode" (page 4-34)
Warning output mode (Output 2 turns on when an error
occurs).
"Warning output mode" (page 4-35)
Counter output mode (Counts the number of output-1 outputs.)
"Counter output mode" (page 4-35)
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-33
Page 82
4
4-6 2-output Settings ()
Limit setting output mode
Refer to page 4-3 "2-Output Settings ()" for setting methods.
Setting output 2 to "LiMt" (limit setting output mode) enables detection of reductions in
received light intensity for output 1.
Specifically, if the cumulative minimum of peak value for output 1 falls below the limit
setting value, it is recognized as a reduction in received light intensity and therefore
output 2 is turned on.
"Light intensity hold display" (page 4-24)
Timing chart
For period (1), the peak value
is not below the limit setting
Settings for Advanced Functions
value, so output 2 remains
OFF.
For period (2), the peak value
is be low the lim it setting value,
so output 2 turns ON.
For period (3), the peak value
again is above the limit setting
value, so output 2 turns OFF if
Auto (Auto) is selected. If USEr
(User) is selected, the
cumulative minimum of peak
values is the peak value
during period (2), which is below the limit setting value, so output 2 remains ON.
"How to reset after the output is activated" (page 4-34)
Received light intensity
ON
OFF
ON
OFF
ON
OFF
(2)(1)(3)
Peak hold value
Limit setting for output 2
Setting value of received
light intensity for output 1
Output1
Output2
(When 75'4 is selected)
Output2
(When #W6Q is selected)
How to reset after the output is activated
DisplayResetting method
User reset
USEr (User)
- Press and hold the [MODE] and [SET] buttons simultaneously.
- Execute reset using external input.
- Cycle the power.
Auto reset
- Press and hold the [MODE] and [SET] buttons simulta-
Auto (Auto)
neously.
- Execute reset using external input.
- Cycle the power.
- When "peak value > setting value + hysteresis"
4-34
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 83
4-6 2-output Settings ()
Warning output mode
Refer to page 4-3 "2-Output Settings ()" for setting methods.
If output 2 is set to "ALrt" (warning output mode), output 2 turns on (or off if the output
condition is set to N.C.) when either of the following errors occur:
• Correction error in DATUM1 or DATUM 2 mode
• ErH
Reference
For error recovery procedures, see "Troubleshooting" (page 6-2).
Counter output mode
Refer to page 4-3 "2-Output Settings ()" for setting methods.
When output 2 is set to [Cnt (counter output mode)], detection which counts the
output-1 output (OFF J ON edge) up to 9999 and makes a comparative judgment
with the setting value is possible.
The following 3 types of counter functions are available.
DisplayFunction in brief
oUr1
(Over1)
oUr2
(Over2)
Auto
(Auto)
* When the counter value reaches 9999 (maximum value), 9999 is held thereafter.
Over counter 1
Turns ON at a value higher than the setting. The count con-
tinues until reset.
Over counter 2
The output turns ON only when the value is the same as the
setting value. The count continues until reset.
Auto reset
Outputs one-shot when the value is the same as the setting
value. The count value is automatically reset. The one-shot
time (Shot) can be set within the range of 1 to 9999 (ms).
Default: 10 ms
*
*
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-35
Page 84
4
4-6 2-output Settings ()
Normal channel 2 display during counter output mode
The screens shown to the right are alternately displayed.
Press to adjust the setting value.
Alternate
display
Setting value Count value
How to reset the count value
• Press and hold the [MODE] and [SET] buttons simultaneously.
• Execute reset using external input.
• Cycle the power.
Settings for Advanced Functions
4-36
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 85
4-6 2-output Settings ()
Output Timer for Output2
Refer to page 4-3 "2-Output Settings ()" for setting methods.
There are 3 types of timers available.
DisplayFunction
toFFUses no output timer. (Default)
Off-delay timer
Turns off the output at a specified
oFFd
time after the detection signal
goes off.
Setting range: 1 to 9999 ms
Default: 10 ms
On-delay timer
Turns on the output at a specified
time after the detection signal
on-d
goes on. Useful if the ON duration
is limited because the workpiece
is shaky such as by vibration.
Setting range: 1 to 9999 ms
Default: 10 ms
One-shot timer
Turns on the output and keeps it
Shot
on for a specified period after the
detection signal goes on.
Setting range: 1 to 9999 ms
Default: 10 ms
Control output
Control output
Control output
Control output
ON
OFF
ON
OFF
OFF
ON
OFF
Timing chart
With workpiece Without workpieceWithout workpiece
With workpiece Without workpieceWithout workpiece
With workpiece Without workpieceWithout workpiece
ON
Timer-set period
With workpiece Without workpieceWithout workpiece
Timer-set period
*
4
Settings for Advanced Functions
Timer-set period
* The above charts show operation examples in the light-ON (L-on) mode for the
reflective model and in the dark-ON (d-on) mode for the retro-reflective or thrubeam model.
Reference
• Different output timers and different timer times can be set separately for
channels 1 and 2.
"Output Timer" (page 4-7)
• The timer can only control on/off switching of the sensor output.
- Digital Laser Sensor LV-N10 Series User's Manual -
4-37
Page 86
4
4-7
Custom Save (Settings Save)
The current setting value and advanced function settings can be saved in the sensor
amplifier.
"Settings for Advanced Functions" (page 4-1)
Point
1
Settings for Advanced Functions
2
Settings Save/Recall
Executing a custom save will override the previously saved settings
with the current settings.
Press and hold the [SET] and [PRESET] buttons simultaneously for 3
seconds or more.
The "rSt" display flashes.
Press the button to display
"SAvE", and then press the [MODE]
button
4-38
Press the button to display "YES",
3
and then press the [MODE] button.
After the settings are saved, the screen
displays "oK", which is then replaced with
the current received light intensity.
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 87
4-7 Settings Save/Recall
User Reset (Settings Recall)
The settings saved with the custom save function can be recalled.
Point
Executing a user reset will override all of the previous settings with
the settings that were saved using the custom save function.
Reference
1
2
3
The factory settings will be recalled if a user reset is executed without previously executing a custom save.
Press and hold the [SET] and [PRESET] buttons simultaneously for 3
seconds or more.
The "rSt" display flashes.
Press the [MODE] button.
Press the button to display
"USEr", and then press the [MODE]
button.
After a user reset is executed, the
screen displays "oK", which is then
replaced with the current received light
intensity.
4
Settings for Advanced Functions
- Digital Laser Sensor LV-N10 Series User's Manual -
4-39
Page 88
4
4-7 Settings Save/Recall
MEMO
Settings for Advanced Functions
4-40
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 89
Specifications
This chapter provides the specifications, circuit diagrams and
dimensions of the LV-N10 Series.
* 80 øs cannot be selected when LV-S31 / S62 / S63 is connected.
Timer variable (1 ms to 9999 ms) Maximum error to setting value is μ 10% or less
(Stand-alone) 1 output max: 100 mA or less, 2 output total: 100 mA or less
PNP outputPNP op en collector 30 V, Residual voltage 1.2 V or less (Output current: 10 mA or less) /
(Stand-alone) 1 output max: 100 mA or less, 2 output total: 100 mA or less
Response time: 1ms (HIGH SPEED / FINE / TURBO),1.2 ms (SUPER), 1.8 ms (ULTRA), 4.2 ms (MEGA)
Connect with
other than LV-S31
Connect with
LV-S 31
*4
Power vo ltage
Power
NPN
consump-
*5
tion
PNP
Ambient
temperature
Relative humidity35 to 85% RH (no condensation)
Vibration10 to 55 Hz, compound amplitude 1.5 mm, 2 hours for each of X, Y, Z axes
Shock resistance
CablePVC
*1 The input time is 25 ms (ON) / 25 ms (OFF) only when external calibration input is selected.
*2 Increases 30 mW (1 mA) for HIGH SPEED Mode
*3 When expanding, always install on a DIN rail (installed on the metal plate), and keep the output current to 20 mA / unit or
less. When expanding units, the working ambient temperature will change according to the following conditions:
1 or 2 units connected: - 20 to + 55 °C; 3 to 10 units connected: - 20 to + 50 °C; 11 to 16 units connected: - 20 to + 45 °C
When using the 2 output type, one unit is counted as 2 units.
*4 When connecting 9 or more expansion units, ensure that the power voltage is 20 V or more.
*5
It increases by 15% when connected to the LV-NH100/NH110/NH300. It dose not include the power consumption of the load.
Power consumption when expansion units are connected is the total power consumption of each amplifier unit.
Example: When one main unit (LV-N11N) is connected to 2 expansion units (LV-N12N) and they are used with LV-NH100 heads in
HIGH SPEED mode.
(1.15
×
*6 Use with the over current protection device which is rated 30 V or more and not more than 1 A.
860 mW × 1) + (1.15
FINE: 2, TURBO / SUPER / ULTRA / MEGA: 4 (These numbers double when "Double" is selected.)
24 VDC (operating voltage 10 to 30 VDC (including ripple)), ripp le (P-P) 10% or less, Class 2 or LPS
Normal: 830 mW or less (at 30 V. 30 mA at 24 V, 56 mA or less at 12 V)
Eco on: 710 mW or less (at 30 V. 26 mA at 24 V, 48 mA or less at 12 V)
Normal: 950 mW or less (at 30 V. 33 mA at 24 V, 60 mA or less at 12 V)
Eco on: 815 mW or less (at 30 V. 29 mA at 24 V, 52 mA or less at 12 V)
Eco full: 650 mW or less (at 30 V. 24 mA at 24 V, 40 mA or less at 12 V)
×
860 mW × 2) = max. 2967 mW
2 V or less (Output current: 10 to 100 mA)
(Multiple connections) 1 output max: 20 mA or less
2.2 V or less (Output current: 10 to 10 0 mA)
(Multiple connections) 1 output max: 20 mA or less
Voltage output: 1 to 5 V, load resistance: 10 k
repeat accuracy: μ 0.5% of F.S.
Input time 2 ms (ON) / 20 ms (OFF) or more
HIGH SPEED: 0, FINE / TURBO / SUPER: 2, ULTRA / MEGA: 4
(These numbers double when "Double" is selected.)
Eco full: 550 mW or less (at 30 V. 21 mA at 24 V, 40 mA or less at 12 V)
- 20 °C to + 55 °C (no freezing )
2
, 3 times for each X, Y, Z axis direction
500 m/s
Ω or more,
*1
*2
*2
*3
- Digital Laser Sensor LV-N10 Series User's Manual -
LV- N1 0
LV-N 11 M N
*6
*2
*2
–
Page 93
5-2
Circuit Diagrams
Cable Type Input/output Circuit Diagram
LV-N11N / N12NLV-N11P / N12P
Load
Load
PLC, etc.
DC10-30V
Short-circuit current
1 mA or less
0V
Overcurrent protection circuit
Sensor main circuit
Overcurrent protection circuit
DC3.3V
Sensor main circuit
*1
Brown
Black
(Control output 1)
White
(Control output 2)
Pink
*1
Blue
*1 LV-N11N only*1 LV-N11P only
M8 Connector Type Input/output Circuit Diagram
LV-N11CN / N12CNLV-N11CP / N12CP
Load
PLC, etc.
4
123
DC10-30V
Short-circuit current
1 mA or less
0V
Overcurrent protection circuit
Sensor main circuit
M8 connector pin layout
*1
(1)
(4)
Overcurrent protection circuit
DC3.3V
Sensor main circuit
(Control output)
(2)
*1
(3)
*1 LV-N11CN only*1 LV-N11CP only
M8 connector pin layout
*1
Brown
Black
(
Control output 1)
White
(Control
output 2)
Load
Pink
*1
Blue
*1
(1)
(4)
(Control output)
(2)
*1
(3)
DC10-30V
PLC, etc.
Short-circuit current
2 mA or less
Load
0V
DC10-30V
PLC, etc.
Short-circuit current
2 mA or less
Load
0V
4
123
5
Specifications
Monitor Output Type Input/output Circuit Diagram
LV-N11MN
10 to 30 VDC
Load
Control output 1
Device
Impedance:10 kΩ or more
)
PLC, etc.
Short-circuit current
1 mA or less
0 V
Overcurrent protection circuit
Protection circuit
3.3 VDC
Sensor main circuit
- Digital Laser Sensor LV-N10 Series User's Manual -
6-2Factory Default Setting (Default Value) List ........ 6-5
6-3List of Recipe Function Settings......................... 6-6
6-4Restrictions on Each Detection Mode ................ 6-8
6
- Digital Laser Sensor LV-N10 Series User's Manual -
6-1
Page 96
6
6-1
Troubleshooting
Frequently Asked Questions
CategoryTroubleCauseAction
• Review the installation environment.
• The received light intensity can be stably dis-
The received light intensity is inconsistent. How can it be stabilized?
Received light
intensity display
The received light intensity is too
low. How can it be increased?
The received light intensity is "---",
"nEAr" or "Far" when LV-S31 is connected.
Nothing is displayed.
Appendix
Screen
display
"-" scrolls from left to right on the
display
Output
operation
The received light intensity at which
the output turns ON and turn s OFF
is slightly different.
It may be affected by vibration, thermal characteristics caused by thermal changes, or the state of the
workpiece surface, etc.
This may be the result of a long
detection distance. When using the
reflective model, the intensity of the
light reflected from the workpiece
may be low, or there may be contamination. When using the thrubeam model or retro-reflective
model, the beam axis may be deviated, etc.
The workpiece is either too close or
too far from the sensor.
Eco (ALL) mode is selected.
The power is cut off or the power
cable is disconnected.
Eco (FuLL) mode is selected.
The sleep mode is enabled by the
external input.
The output timer is set.
Hysteresis
played at 100.0 by using the preset function.
"Preset Function" (page 3-6)
• When using the DATUM1/2 mode, the set value
and received light intensity are both calibrated to
maintain a set ratio.
"Detection Mode" (page 4-8)
• Review the installation environment and detection distance.
• The received light intensity value can be
increased by raising (delaying) the power mode.
"Power Modes" (page 4-4)
• The apparent received light intensity can be
increased 4-fold by using the light intensity
expansion setting.
"Display Gain" (page 4-30)
• Check that the workpiece is installed within the
adjustment range (50 to 200 mm).
• Place the workpiece at the position to be set as
the detection center, and adjust it so that the
JUST indicator lights in green (or the received
light intensity value is near 5000). (Refer to page
2-7)
• Disable the Eco function.
"Power Save" (page 4-29)
• Disable the Eco function.
• Disable the Eco function.
"Power Save" (page 4-29)
• Turn the external input OFF, or select a mode
other than sleep mode transition input with the
external input function.
"Sleep mode transition input" (page 4-19)
• Check the output timer setting.
"Output Timer" (page 4-7)
• A slight hysteresis is set to prevent chattering in
the output. If this level of received light intensity
is notable during detection, the detection tolerance may be low. Please review the detection
details.
6-2
- Digital Laser Sensor LV-N10 Series User's Manual -
Page 97
6-1 Troubleshooting
CategoryTroubleCauseAction
• Press and hold the [MODE] button and check
the raw received light intensity level. Try adjusting the beam axis, removing contamination, or
reviewing the installation environment. Take
actions to raise the original received light intensity.
• Disable the zero shift function.
"Zero Shift Function" (page 3-17)
• Select the sensitivity setting other than the zero
shift calibration.
"Sensitivity Setting" (page 4-4)
• Select the detection method other than the rising
/ falling edge detection mode.
"Detection Mode" (page 4-8)
• Select a mode other than zero shift input for the
external input function.
"External Input" (page 4-17)
• Provide a sufficient difference in received light
intensity when using the preset function and
work-preset function.
• Make sure that the received light intensity does
not exceed the values listed in the Extension display on page 4-30.
• Cancel each preset function.
"Preset Function" (page 3-6)
• Select a mode other than preset input with th e
external input function.
"External Input" (page 4-17)
• Set a detection mode other than ris ing edge
detection/falling edge detection mode with the
detection mode settings.
"Detection Mode" (page 4-8)
• Set a detection mode other than DATUM1/
DATUM2 mode with the detection mode settings.
"Detection Mode" (page 4-8)
• Press and hold the [MODE] button and check
the raw received light intensity level. Try adjusting the beam axis, removing contamination, or
reviewing the installation environment. Take
actions to raise the original received light intensity.
• Check the installation environment and confirm
that the beam axis has not deviated and that the
workpiece surface is not contaminated.
• To continue calibration, increase the warning output
level value. Calibration will not stop as frequently.
• Change the detection mode or sensitivity setting
referring to "Restrictions for Sensitivity Setti ngs
in Each Detection Mode" (page 6-8).
• Initialize the setting.
"Initialization of Settings (Reset to Initial
Values)" (page 3-21)
• Initialize the setting.
"Initialization of Settings (Reset to Initial
Values)" (page 3-21)
Preset related
Zero shift related
DATUM relat ed
Sensitivity setting
(calibration)
related
Others
A value less than "100.0" is displayed even when the [PRESET]
button is pressed.
When each preset function is executed, "---- ----" appears, and no
change is seen.
"---- ----" appears even when
using the zero shift function, and
each preset function with no effect
is being used.
Nothing happens even when the
zero shift function is used.
The received light intensity is not
calibrated to 100.0/.0 in the
DATUM1/2 modes.
Pressing the [SET] button has no
effect.
Initializing all settings-
Cannot remember the key lock PIN
number.
A setting error occurred when connecting to the communication series
NU Series.
The raw received light intensity is 50
or less. * (200 when using the
received light intensity expansion
setting.)
The zero shift function is enabled.
The zero shift calibration (0SEt) is
selected as the sensitivity setting.
The rising / falling edge detection
mode is selected.
Zero shift input is selected with the
external input function.
The difference in received light
intensity during preset function execution and workpiece preset function execution is small.
The state was saturated when using
the maximum sensitivity preset
function and full auto preset function. The state was saturated when
using either the preset function or
workpiece preset function.
The preset function is enabled.
Preset input is selected with the
external input function.
The rising edge detection/falling
edge detection mode is enabled.
The DATUM1/DATUM2 mode is
enabled.
The raw received light intensity is 50
or less. * (200 when using the
received light intensity expansion
setting FuLL)
The raw received light intensity is
less than the warning output level.
The sensitivity setting (calibration)
is restricted in specific detectio n
modes.
-• Contact your nearest KEYENCE office.
The setting is incorrect.
6
Appendix
- Digital Laser Sensor LV-N10 Series User's Manual -
6-3
Page 98
6-1 Troubleshooting
Error Displays and Corrective Actions
6
Error
display
ErC
ErH
ErE
Loc
Overcurrent in the control output.
Head cable is broken, or sensor head
is not connected.
Failed to write/load the internal data.
Key lock function is valid, or if using
M8 connector type, the power mode
was changed in the MEGA mode fixed
state.
Appendix
The DTM
indicator
flashes.
Correction error in DATUM1 or
DATUM 2 mode.
CauseSolution
• Check the load and return the current within the
rated range.
• Check that the output wire is not contacting any
other wire or frame.
• Check that sensor head is connected.
• Check whether head cable is not broken.
• Check the head cable connection to the connector.
• After checking, turn the power OFF and ON.
• Turn the power OFF and ON. If the data is not
recovered, initialize the settings.*
"Initialization of Settings (Reset to Initial Val-
ues)" (page 3-21)
• Disable the key lock.
If the key lock cannot be disabled, it may be
locked with a password, or locked by the NU
Series communication unit.
"Disabling the Key Operation" (page 3-23)
• Check that the sensor amplifier is locked in
MEGA mode.
"Locking in MEGA Mode" (page 3-22)
• Check that the received light intensity has not
dropped.
• Adjust the DATUM warning output level.
• Ensure that the received light intensity value is
50 or more.
*"ErE" display may remain if the LV-N10 Series is damaged.
For errors other than those listed above, contact your nearest KEYENCE office.
6-4
- Digital Laser Sensor LV-N10 Series User's Manual -
*1 Not available for the 0-line type.
*2 Only monitor output type.
*3 Main unit only.
*4 Only 2-output type
*5 Only when LV-NH100/NH110/NH300 is connected.
6
Appendix
ItemDefault value
Setting value (Detection mode: Std selected)50
Reference
page
4-4
Setting value (Detection mode: dtM1 selected)5.0 4-9
Setting value (Detection mode: dtM2 selected)5.0 4-11
Setting value (Detection mode: ArEA selected) 504-14
Setting value (Detection mode: _|~d selected)504-16
Setting value (Detection mode: ~|_d selected)504-16
Percentage calibration target value-10P4-5
Time r valu eAll 104-7
2-output timer valueAll 104-37
Reference
Refer to "Initialization of Settings (Reset to Initial Values)" (page 3-21)
for the initializing method.
- Digital Laser Sensor LV-N10 Series User's Manual -
6-5
Page 100
6-3
List of Recipe Function Settings
6
Reference
Refer to "Selecting Recipe" (page 3-19) for details on selecting recipes.
The following details are set using the recipe function.
Application itemSetting itemSetting
N.O. (1-ch)3-4
Output condition
Fall detection
(r-1 FALL)
Percentage calibration
(r-2 SEtP)
Appendix
Reflective background
cancellation
(r-3 0SEt)
Full light intensity*2
(r-4 MEGA)
Area detection
(r-5 ArEA)
0-datum
(r-6 0dtM)
*1
*2
*3
*4
Output timer
Detection mode
Sub display
Setting value
Output condition
Sensitivity setting
Target percentage value
Preset saturation level
External input
Output2 detection mode
Limit setting value
Sensitivity setting
External input
Sensitivity setting
External input
Power mode
Display gain
Sub display
Detection mode
Sensitivity setting
Percentage target value
Output condition
Detection mode
Sensitivity setting
Percentage target value
External input
Sub display
Sensitivity
Output2 detection mode
Limit setting value
no
D-on (2-ch)3-4
d-on
One shot (1-ch)4-7
Shot
ON delay (2-ch)4-37
on-d
Falling edge detection mode4-16
~L_d
Hold display4-24
HLd
20 (1-ch)-
20
300 (2-ch)-
300
D-on3-4
d-on
Percentage calibration4-5
SEtP
-5%4-5
-5P
101%4-27
101P
External calibration input4-17
SEt
Limit setting detection mode4-33
LiMt
1004-34
100
Zero shift calibration4-6
0SEt
Zero shift input4-17
ShFt
Zero shift calibration4-6
0SEt
Zero shift input4-17
ShFt
MEGA mode4-4
MEGA
Full4-30
FuLL
Extension display4-22
StdE
Area detection mode4-14
ArEA
Percentage calibration4-5
SEtP
-10%4-5
-10P
D-on3-4
d-on
DATUM2 mode4-11
dtM2
Percentage calibration4-5
SEtP
-5%4-5
-5P
External calibration input4-17
SEt
Extension display4-22
StdE
High resolution4-32
RESo
Limit setting output4-33
LiMt
1004-34
100
Reference
page
6-6
- Digital Laser Sensor LV-N10 Series User's Manual -
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.