Fresenius 2008T User's troubleshooting manual

2008T BlueStarTM Hemodialysis Machine
Operator’s Troubleshooting Guide
TABLE OF CONTENTS
Installation
Auto Prime
Miscellaneous
Indications for Use
NOTE: This troubleshooting guide is intended to be a supplement to the information in
the 2008T Operator’s Manual, P/N 490122, including Warnings and Cautions. Users must read and understand the information in the 2008T Operator’s Manual before using the 2008T BlueStarTM hemodialysis machine.
1
2
5
8
14
INSTALLATION
The 2008T BlueStar Standard Upgrade Kit (P/N 191137) installs a different functional software version than expected (e.g., installs version 2.64 instead of version 2.72).
1. Read and follow the installation instructions included with the upgrade kit.
2. The 2008T BlueStar software installation requires a two-step installation process as noted in the instructions. The updater will install a temporary software version which will not match
the Flash Drive label during the first installation. The latest functional software version will be installed during the second installation.
2008T BlueStarTM Machine Operator’s Troubleshooting Guide 1
AUTO START
Auto Start not available.
1. 2008T BlueStar Premium Upgrade Kit (P/N 191138) must be installed and the Premium features enabled.
2. Once installed, Auto Start is accessible in Service Mode / Maintenance / Scheduler / Auto Start.
Auto Start automatic self-tests do not complete.
1. A conductivity time-out occurs.
2. Select Both Tests to run self-tests manually.
NOTE: The Auto Start Dialysis button will be green and the status box message “Connect Concentrates” will be displayed. Once the acid connector is removed from its port, the machine will wait up to seven minutes for dialysate conductivity and temperature to stabilize. If they do not stabilize, automatic testing will be cancelled, and the self-tests must be run manually.
To reduce the chances of this situation occurring, it is suggested that both the acid and bicarbonate concentrate supplies be connected at the same time to achieve conductivity. Follow facility policies and procedures.