Dräger Medizintechnik |
|
|
|
Error list
General
The error numbers shown on the EL display refer to a faulty assembly.
The abbreviations shown on the EL display have the following meaning:
Abbreviation on EL |
Faulty assembly |
Location |
display |
|
|
|
|
|
|
|
|
p_sens |
WT2 Sensor PCB |
Sensor unit |
|
|
|
p_cont |
WT2 Controller PCB |
Display housing |
|
|
|
p_actu |
WT2 Actuator PCB |
E-box |
|
|
|
p_matr |
WT2 Mattress PCB |
E-box |
|
|
|
soft |
Software error |
|
|
|
|
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01
6150.000 Caleo |
10/2000 |
Repair Instructions |
Page 1 of 14 |
Dräger Medizintechnik |
|
|
|
Caleo error list
• Click on an error code to jump to relevant description of the error code.
Micro-controller errors cause a direct program abort. Other errors cause a program abort only after they have occurred several times. All errors and their occurrence are written in the error log.
Error code numbers shown on the EL display
1 |
2 |
3 |
|
5 |
6 |
7 |
8 |
9 |
|
11 |
12 |
13 |
14 |
15 |
16 |
|
|
|
|
|
22 |
23 |
24 |
25 |
|
27 |
28 |
29 |
30 |
31 |
32 |
|
|
35 |
36 |
37 |
38 |
39 |
40 |
41 |
42 |
43 |
|
|
46 |
47 |
48 |
49 |
50 |
51 |
52 |
53 |
54 |
55 |
56 |
57 |
58 |
59 |
60 |
61 |
62 |
|
64 |
65 |
66 |
67 |
68 |
69 |
70 |
71 |
72 |
|
|
|
|
|
|
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01
6150.000 Caleo |
10/2000 |
Repair Instructions |
Page 2 of 14 |
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01
Dräger Medizintechnik |
|
|
|
|
|
|
|
No. |
Plain text |
Cause |
Remedy |
|
|
|
|
|
|
|
|
|
|
|
|
1 |
Sensor drift too |
The measuring |
• Check the WT2 Sensor PCB using the |
|
|
|
|||||
|
|
|
|
|
high |
paths (air and skin) |
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
have drifted out of |
"Sensor PCB". |
|
|
|
|
|
|
||
|
|
|
|
|
|
the tolerance range. |
• If necessary, replace the |
|
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
WT2 Sensor PCB, see Repair |
|
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
Instructions "Replacing the WT2 |
|
|
|
|
|
|
|
10-minute test and |
|
|
|
|
|
|
|
Sensor PCB". |
|
|
|
|
|
|
|
leads to INOP.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
2 |
Comm. error on |
Communication |
• Check cable connection between WT2 |
|
|
|
|
|
CAN bus |
between WT2 |
Actuator PCB and WT2 Controller PCB; |
|
|
|
|
|
|
Controller PCB, |
replace cable connection, if necessary. |
|
|
|
|
|
|
WT2 Actuator PCB, |
• Check cable connection between WT2 |
|
|
|
|
|
|
and WT2 Sensor |
|
|
|
|
|
|
|
Actuator PCB and WT2 sensor PCB; |
|
|
|
|
|
|
|
PCB is faulty. (Note: |
|
|
|
|
|
|
|
replace cable connection, if necessary. |
|
|
|
|
|
|
|
The error is |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
detected during |
• Replace WT2 Controller PCB, if |
|
|
|
|
|
|
operation.) |
necessary. |
|
|
|
|
|
|
|
• Replace WT2 Actuator PCB, if |
|
|
|
|
|
|
|
necessary. |
|
|
|
|
|
|
|
• Replace WT2 Sensor PCB, if necessary. |
|
|
|
|
|
|||
|
|
|
|
3 |
36 degree ref. |
The "36 degree |
• Check the WT2 Sensor PCB using the |
|
|
|
|
|
faulty |
test" with the |
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
reference |
"Sensor PCB". |
|
|
|
|
|
|
resistance failed. |
• If necessary, replace the |
|
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
WT2 Sensor PCB, see Repair |
|
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
Instructions "Replacing the WT2 |
|
|
|
|
|
|
|
10-minute test and |
|
|
|
|
|
|
|
Sensor PCB". |
|
|
|
|
|
|
|
leads to INOP.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
5 |
Air excess temp. |
The excess |
• Check the WT2 Sensor PCB using the |
|
|
|
|
|
test failed |
temperature circuit |
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
test with a reference |
"Sensor PCB". |
|
|
|
|
|
|
resistance failed. |
• If necessary, replace the |
|
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
WT2 Sensor PCB, see Repair |
|
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
Instructions "Replacing the WT2 |
|
|
|
|
|
|
|
10-minute test.) |
|
|
|
|
|
|
|
Sensor PCB". |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
6 |
Invalid |
The configuration is |
• Switch device off and on and re- |
|
|
|
|
|
configuration |
invalid. (Note: The |
configure device. |
|
|
|
|
|
|
error is detected |
• Replace the WT2 Controller PCB, see |
|
|
|
|
|
|
during the power-on |
|
|
|
|
|
|
|
Repair Instructions "Replacing the |
|
|
|
|
|
|
|
test and leads to |
|
|
|
|
|
|
|
WT2 Controller PCB". |
|
|
|
|
|
|
|
INOP.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
6150.000 Caleo |
10/2000 |
Repair Instructions |
Page 3 of 14 |
Dräger Medizintechnik |
|
|
|
|
|
No. |
Plain text |
|
Cause |
|
Remedy |
|
|
7 |
40 degree ref. |
|
The "40 degree |
|
• Check the WT2 Sensor PCB using the |
|
|
|
|
||||
|
|
|
faulty |
|
test" with the |
|
DrägerService Mode, see diagnosis |
|
|
|
|
|
reference |
|
"Sensor PCB". |
|
|
|
|
|
resistance failed. |
|
• If necessary, replace the |
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
|
WT2 Sensor PCB, see Repair |
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
|
Instructions "Replacing the WT2 |
|
|
|
|
|
|
10-minute test and |
|
|
|
|
|
|
|
|
Sensor PCB". |
|
|
|
|
|
|
leads to INOP.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
8 |
AirSafetyRelay |
|
The air heater safety |
|
• Check the WT2 Actuator PCB using the |
|
|
|
blocked |
|
relay or the control |
|
DrägerService Mode, see diagnosis |
|
|
|
|
|
of the air heater |
|
"Digital signals". |
|
|
|
|
|
safety relay is faulty. |
|
• If necessary, replace the |
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
10-minute test and |
|
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|
leads to INOP.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
9 |
AirSafetyRelay |
|
The control of the |
|
• Check the WT2 Actuator PCB using the |
|
|
|
Feedb. error |
|
air heater safety |
|
DrägerService Mode, see diagnosis |
|
|
|
|
|
relay sends no |
|
"Digital signals". |
|
|
|
|
|
feedback signal. |
|
• If necessary, replace the |
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
10-minute test and |
|
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|
leads to INOP.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
11 |
Air heater off |
|
The air heater is off, |
|
• Check fuse on WT2 Actuator PCB and |
|
|
|
|
|
but should be on. |
|
replace fuse, if necessary. |
|
|
|
|
|
(Note: The error is |
|
• Check air heater thermostat and replace |
|
|
|
|
|
detected during |
|
|
|
|
|
|
|
|
thermostat, if necessary. |
|
|
|
|
|
|
operation and leads |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
to INOP.) |
|
• Check air heater and replace air heater, |
|
|
|
|
|
|
|
if necessary. |
|
|
|
|
|
|
|
• Check the WT2 Actuator PCB using the |
|
|
|
|
|
|
|
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
|
"Actuator PCB". |
|
|
|
|
|
|
|
• If necessary, replace the |
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|
|
|
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01
6150.000 Caleo |
10/2000 |
Repair Instructions |
Page 4 of 14 |
For internal use only. Copyright reserved.
R6150000FL1.fm 08.02.01
Dräger Medizintechnik |
|
|
|
|
|
|
|
No. |
Plain text |
Cause |
Remedy |
|
|
|
|
|
|||
|
|
|
|
12 |
Air heater on |
The air heater is on, |
• Check the WT2 Actuator PCB using the |
|
|
|
|
|
|
but should be off. |
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
||
|
|
|
|
|
|
||
|
|
|
|
|
|
(Note: The error is |
"Actuator PCB". |
|
|
|
|
|
|
detected during |
• If necessary, replace the |
|
|
|
|
|
|
||
|
|
|
|
|
|
operation and leads |
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
|
to INOP.) |
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|||
|
|
|
|
13 |
HumiditySafety |
The humidifier |
• Check the WT2 Actuator PCB using the |
|
|
|
|
|
Relay blocked |
safety relay or the |
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
control of the |
"Digital signals". |
|
|
|
|
|
|
humidifier safety |
• If necessary, replace the |
|
|
|
|
|
|
relay is faulty. (Note: |
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
|
The error is |
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|
|
10-minute test.) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
14 |
HumiditySafety |
The control of the |
• Check the WT2 Actuator PCB using the |
|
|
|
|
|
Relay Feedb. |
humidifier safety |
DrägerService Mode, see diagnosis |
|
|
|
|
|
error |
relay sends no |
"Digital signals". |
|
|
|
|
|
|
feedback signal. |
• If necessary, replace the |
|
|
|
|
|
|
(Note: The error is |
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
|
detected during the |
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
|
10-minute test.) |
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
|
|
|
15 |
Boiler off |
The humidifier |
• Check fuse on WT2 Actuator PCB and |
|
|
|
|
|
|
heater is off, but |
replace fuse, if necessary. |
|
|
|
|
|
|
should be on. (Note: |
• Check humidifier thermostat and replace |
|
|
|
|
|
|
The error is |
|
|
|
|
|
|
|
thermostat, if necessary. |
|
|
|
|
|
|
|
detected during |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
operation.) |
• Check humidifier and replace humidifier, |
|
|
|
|
|
|
|
if necessary. |
|
|
|
|
|
|
|
• Check the WT2 Actuator PCB using the |
|
|
|
|
|
|
|
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
|
"Actuator PCB". |
|
|
|
|
|
|
|
• If necessary, replace the |
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|||
|
|
|
|
16 |
Boiler on |
The humidifier |
• Check the WT2 Actuator PCB using the |
|
|
|
|
|
|
heater is on, but |
DrägerService Mode, see diagnosis |
|
|
|
|
|
|
should be off. (Note: |
"Actuator PCB". |
|
|
|
|
|
|
The error is |
• If necessary, replace the |
|
|
|
|
|
|
detected during |
|
|
|
|
|
|
|
WT2 Actuator PCB, see Repair |
|
|
|
|
|
|
|
operation.) |
|
|
|
|
|
|
|
Instructions "Replacing the |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
WT2 Actuator PCB". |
|
|
|
|
|
|
|
|
6150.000 Caleo |
10/2000 |
Repair Instructions |
Page 5 of 14 |