Error list

General

The error numbers shown on the EL display refer to a faulty assembly.

The abbreviations shown on the EL display have the following meaning:

Abbreviation on EL

Faulty assembly

Location

display

 

 

 

 

 

 

 

 

p_sens

WT2 Sensor PCB

Sensor unit

 

 

 

p_cont

WT2 Controller PCB

Display housing

 

 

 

p_actu

WT2 Actuator PCB

E-box

 

 

 

p_matr

WT2 Mattress PCB

E-box

 

 

 

soft

Software error

 

 

 

 

For internal use only. Copyright reserved.

R6150000FL1.fm 08.02.01

6150.000 Caleo

10/2000

Repair Instructions

Page 1 of 14

Drager Caleo Incubator Service instructions

Dräger Medizintechnik

 

 

 

Caleo error list

• Click on an error code to jump to relevant description of the error code.

Micro-controller errors cause a direct program abort. Other errors cause a program abort only after they have occurred several times. All errors and their occurrence are written in the error log.

Error code numbers shown on the EL display

1

2

3

 

5

6

7

8

9

 

11

12

13

14

15

16

 

 

 

 

 

22

23

24

25

 

27

28

29

30

31

32

 

 

35

36

37

38

39

40

41

42

43

 

 

46

47

48

49

50

51

52

53

54

55

56

57

58

59

60

61

62

 

64

65

66

67

68

69

70

71

72

 

 

 

 

 

 

For internal use only. Copyright reserved.

R6150000FL1.fm 08.02.01

6150.000 Caleo

10/2000

Repair Instructions

Page 2 of 14

For internal use only. Copyright reserved.

R6150000FL1.fm 08.02.01

Dräger Medizintechnik

 

 

 

 

 

 

 

No.

Plain text

Cause

Remedy

 

 

 

 

 

 

 

 

 

 

 

 

1

Sensor drift too

The measuring

• Check the WT2 Sensor PCB using the

 

 

 

 

 

 

 

 

high

paths (air and skin)

DrägerService Mode, see diagnosis

 

 

 

 

 

 

have drifted out of

"Sensor PCB".

 

 

 

 

 

 

 

 

 

 

 

 

the tolerance range.

• If necessary, replace the

 

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

WT2 Sensor PCB, see Repair

 

 

 

 

 

 

detected during the

 

 

 

 

 

 

Instructions "Replacing the WT2

 

 

 

 

 

 

10-minute test and

 

 

 

 

 

 

Sensor PCB".

 

 

 

 

 

 

leads to INOP.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Comm. error on

Communication

• Check cable connection between WT2

 

 

 

 

 

CAN bus

between WT2

Actuator PCB and WT2 Controller PCB;

 

 

 

 

 

 

Controller PCB,

replace cable connection, if necessary.

 

 

 

 

 

 

WT2 Actuator PCB,

• Check cable connection between WT2

 

 

 

 

 

 

and WT2 Sensor

 

 

 

 

 

 

Actuator PCB and WT2 sensor PCB;

 

 

 

 

 

 

PCB is faulty. (Note:

 

 

 

 

 

 

replace cable connection, if necessary.

 

 

 

 

 

 

The error is

 

 

 

 

 

 

 

 

 

 

 

 

 

detected during

• Replace WT2 Controller PCB, if

 

 

 

 

 

 

operation.)

necessary.

 

 

 

 

 

 

 

• Replace WT2 Actuator PCB, if

 

 

 

 

 

 

 

necessary.

 

 

 

 

 

 

 

• Replace WT2 Sensor PCB, if necessary.

 

 

 

 

 

 

 

 

 

3

36 degree ref.

The "36 degree

• Check the WT2 Sensor PCB using the

 

 

 

 

 

faulty

test" with the

DrägerService Mode, see diagnosis

 

 

 

 

 

 

reference

"Sensor PCB".

 

 

 

 

 

 

resistance failed.

• If necessary, replace the

 

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

WT2 Sensor PCB, see Repair

 

 

 

 

 

 

detected during the

 

 

 

 

 

 

Instructions "Replacing the WT2

 

 

 

 

 

 

10-minute test and

 

 

 

 

 

 

Sensor PCB".

 

 

 

 

 

 

leads to INOP.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

Air excess temp.

The excess

• Check the WT2 Sensor PCB using the

 

 

 

 

 

test failed

temperature circuit

DrägerService Mode, see diagnosis

 

 

 

 

 

 

test with a reference

"Sensor PCB".

 

 

 

 

 

 

resistance failed.

• If necessary, replace the

 

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

WT2 Sensor PCB, see Repair

 

 

 

 

 

 

detected during the

 

 

 

 

 

 

Instructions "Replacing the WT2

 

 

 

 

 

 

10-minute test.)

 

 

 

 

 

 

Sensor PCB".

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

Invalid

The configuration is

• Switch device off and on and re-

 

 

 

 

 

configuration

invalid. (Note: The

configure device.

 

 

 

 

 

 

error is detected

• Replace the WT2 Controller PCB, see

 

 

 

 

 

 

during the power-on

 

 

 

 

 

 

Repair Instructions "Replacing the

 

 

 

 

 

 

test and leads to

 

 

 

 

 

 

WT2 Controller PCB".

 

 

 

 

 

 

INOP.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6150.000 Caleo

10/2000

Repair Instructions

Page 3 of 14

Dräger Medizintechnik

 

 

 

 

 

No.

Plain text

 

Cause

 

Remedy

 

 

7

40 degree ref.

 

The "40 degree

 

• Check the WT2 Sensor PCB using the

 

 

 

 

 

 

 

faulty

 

test" with the

 

DrägerService Mode, see diagnosis

 

 

 

 

 

reference

 

"Sensor PCB".

 

 

 

 

 

resistance failed.

 

• If necessary, replace the

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

 

WT2 Sensor PCB, see Repair

 

 

 

 

 

detected during the

 

 

 

 

 

 

 

Instructions "Replacing the WT2

 

 

 

 

 

10-minute test and

 

 

 

 

 

 

 

Sensor PCB".

 

 

 

 

 

leads to INOP.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8

AirSafetyRelay

 

The air heater safety

 

• Check the WT2 Actuator PCB using the

 

 

 

blocked

 

relay or the control

 

DrägerService Mode, see diagnosis

 

 

 

 

 

of the air heater

 

"Digital signals".

 

 

 

 

 

safety relay is faulty.

 

• If necessary, replace the

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

detected during the

 

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

10-minute test and

 

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

leads to INOP.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

9

AirSafetyRelay

 

The control of the

 

• Check the WT2 Actuator PCB using the

 

 

 

Feedb. error

 

air heater safety

 

DrägerService Mode, see diagnosis

 

 

 

 

 

relay sends no

 

"Digital signals".

 

 

 

 

 

feedback signal.

 

• If necessary, replace the

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

detected during the

 

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

10-minute test and

 

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

leads to INOP.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

Air heater off

 

The air heater is off,

 

• Check fuse on WT2 Actuator PCB and

 

 

 

 

 

but should be on.

 

replace fuse, if necessary.

 

 

 

 

 

(Note: The error is

 

• Check air heater thermostat and replace

 

 

 

 

 

detected during

 

 

 

 

 

 

 

thermostat, if necessary.

 

 

 

 

 

operation and leads

 

 

 

 

 

 

 

 

 

 

 

 

 

to INOP.)

 

• Check air heater and replace air heater,

 

 

 

 

 

 

 

if necessary.

 

 

 

 

 

 

 

• Check the WT2 Actuator PCB using the

 

 

 

 

 

 

 

DrägerService Mode, see diagnosis

 

 

 

 

 

 

 

"Actuator PCB".

 

 

 

 

 

 

 

• If necessary, replace the

 

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

 

 

 

For internal use only. Copyright reserved.

R6150000FL1.fm 08.02.01

6150.000 Caleo

10/2000

Repair Instructions

Page 4 of 14

For internal use only. Copyright reserved.

R6150000FL1.fm 08.02.01

Dräger Medizintechnik

 

 

 

 

 

 

 

No.

Plain text

Cause

Remedy

 

 

 

 

 

 

 

 

 

12

Air heater on

The air heater is on,

• Check the WT2 Actuator PCB using the

 

 

 

 

 

 

but should be off.

DrägerService Mode, see diagnosis

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(Note: The error is

"Actuator PCB".

 

 

 

 

 

 

detected during

• If necessary, replace the

 

 

 

 

 

 

 

 

 

 

 

 

operation and leads

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

 

to INOP.)

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

 

 

 

 

13

HumiditySafety

The humidifier

• Check the WT2 Actuator PCB using the

 

 

 

 

 

Relay blocked

safety relay or the

DrägerService Mode, see diagnosis

 

 

 

 

 

 

control of the

"Digital signals".

 

 

 

 

 

 

humidifier safety

• If necessary, replace the

 

 

 

 

 

 

relay is faulty. (Note:

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

 

The error is

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

 

detected during the

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

 

10-minute test.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

14

HumiditySafety

The control of the

• Check the WT2 Actuator PCB using the

 

 

 

 

 

Relay Feedb.

humidifier safety

DrägerService Mode, see diagnosis

 

 

 

 

 

error

relay sends no

"Digital signals".

 

 

 

 

 

 

feedback signal.

• If necessary, replace the

 

 

 

 

 

 

(Note: The error is

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

 

detected during the

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

 

10-minute test.)

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15

Boiler off

The humidifier

• Check fuse on WT2 Actuator PCB and

 

 

 

 

 

 

heater is off, but

replace fuse, if necessary.

 

 

 

 

 

 

should be on. (Note:

• Check humidifier thermostat and replace

 

 

 

 

 

 

The error is

 

 

 

 

 

 

thermostat, if necessary.

 

 

 

 

 

 

detected during

 

 

 

 

 

 

 

 

 

 

 

 

 

operation.)

• Check humidifier and replace humidifier,

 

 

 

 

 

 

 

if necessary.

 

 

 

 

 

 

 

• Check the WT2 Actuator PCB using the

 

 

 

 

 

 

 

DrägerService Mode, see diagnosis

 

 

 

 

 

 

 

"Actuator PCB".

 

 

 

 

 

 

 

• If necessary, replace the

 

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

 

 

 

 

16

Boiler on

The humidifier

• Check the WT2 Actuator PCB using the

 

 

 

 

 

 

heater is on, but

DrägerService Mode, see diagnosis

 

 

 

 

 

 

should be off. (Note:

"Actuator PCB".

 

 

 

 

 

 

The error is

• If necessary, replace the

 

 

 

 

 

 

detected during

 

 

 

 

 

 

WT2 Actuator PCB, see Repair

 

 

 

 

 

 

operation.)

 

 

 

 

 

 

Instructions "Replacing the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WT2 Actuator PCB".

 

 

 

 

 

 

 

 

6150.000 Caleo

10/2000

Repair Instructions

Page 5 of 14