01.10.98 |
only. Copyright reserved. |
GBU6141223T01.fm |
For internal use |
Conversion Instructions for Mounting the New Water Supply Level Controller 2M 22130
The new water supply level controller (order no. 2M 22130) is to be mounted on the Incubator 8000, Incubator 8000 SC/IC/NC in the following manner:
•Drill two holes for the water container hook using the drilling template (fig. 1)
•Screw on the hook for the water container (fig. 1)
•Drill the two holes for the hose clip (fig. 2)
•Screw on the hose clip (fig. 2)
•Fit the hose, which connects the water container to the water supply level controller, on the hose clip as shown in the Instructions for Use (90 28 905 - us/e).
•Connect the hose and adapter from the conversion kit to the water supply level controller (fig. 3)
•Connect the water supply level controller to the water cooker of the Incubator 8000, Incubator 8000 SC/IC/NC (see connection hole on fig. 2).
•Connect water container hose to the water supply level controller.
water container hook
Fig. 1: Drilling template for water container hook
6141.223 Incubator 8000 NC |
09.97 |
Conversion Instructions |
Page 1 |