Type 8045
Electromagnetic flowmeter
Quickstart
English
1. |
About the Quickstart......................................................................... |
3 |
|
2. |
Intended use................................................................................................ |
4 |
|
3. |
Basic safety information................................................................ |
4 |
|
4. |
General information............................................................................ |
6 |
|
5. |
Description of the name plate................................................... |
6 |
|
6. |
Technical data............................................................................................ |
7 |
|
7. |
Installation and wiring.................................................................. |
11 |
|
8. Adjustment, commissioning.......................................................................... |
23 |
||
9. |
Maintenance and troubleshooting................................... |
30 |
|
10. |
Packaging, Transport................................................................... |
30 |
|
11. |
Storage....................................................................................................... |
30 |
|
12. |
Disposal of the product........................................................... |
31 |
We reserve the right to make technical changes without notice.
Technische Änderungen vorbehalten.
Sous réserve de modifications techniques.
© Bürkert SAS, 2012 - 2013
Operating Instructions 1308/1_EU-ML 00563855 ORIGINAL_FR
Type 8045
About the Quickstart
1.About the Quickstart
The quickstart describes the life cycle of the device. Please keep this manual in a safe place, accessible to all users and any new owners.
Important safety information.
Respect the safety instructions. Study in particular the chapters entitled Intended Use and Basic Safety Instructions.
• The quickstart must be read and understood.
The quickstart explains how to install, set, and start-up the device.
A detailed description of the device can be found in the related operating instructions on the elivered CD.
1.1.Symbols used
danger
Warns against an imminent danger.
•Failure to observe this warning can result in death or in serious injury.
Warning
Warns against a potentially dangerous situation.
•Failure to observe this warning can result in serious injury or even death.
attention
Warns against a possible risk.
•Failure to observe this warning can result in substantial or minor injuries.
note
Warns against material damage.
•Failure to observe this warning may result in damage to the device or system.
Indicates additional information, advice or important recommendations.
Refers to information contained in this manual or in other documents.
→→Indicates a procedure to be carried out.
1.2.Definition of the word "device"
The word "device" used within this manual refers to the flowmeter type 8045.
|
|
|
|
English |
3 |
|
|
|
Type 8045
Intended use
2.Intended use
Use of the device that does not comply with the instructions could present risks to people, nearby installations and the environment.
•The 8045 flowmeter is intended exclusively to measure flow rate in liquids.
•This device must be protected against electromagnetic interference, ultraviolet rays and, when installed outdoors, the effects of climatic conditions.
•This device must be used in compliance with the characteristics and commissioning and use conditions specified in the contractual documents and in the operating instructions.
•Requirements for the safe and proper operation of the device are proper transport, storage and installation, as well as careful operation and maintenance.
•Only use the device as intended.
→→Observe any existing restraints when the device is exported.
3.Basic safety information
This safety information does not take into account:
•any contingencies or occurences that may arise during installation, use and maintenance of the devices.
•the local safety regulations for which the operating company is responsible including the staff in charge of installation and maintenance.
Danger due to high pressure in the installation.
•Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Danger due to electrical voltage.
•Shut down the electrical power source of all the conductors and isolate it before carrying out work on the system.
•Observe all applicable accident protection and safety regulations for electrical equipment.
Danger due to high temperatures of the fluid.
•Use safety gloves to handle the device.
•Stop the circulation of fluid and drain the pipe before loosening the process connections.
Danger due to the nature of the fluid.
•Respect the prevailing regulations on accident prevention and safety relating to the use of aggressive fluids.
|
|
4 |
English |
|
|
Type 8045
Basic safety information
Various dangerous situations
To avoid injury take care:
•not to use the device for the measurement of gas flow rates.
•not to use the device in explosive atmospheres.
•not to use the device in an environment incompatible with the materials it is made of.
•not to subject the device to mechanical loads (e.g. by placing objects on top of it or by using it as a step).
•not to make any external or internal modifications to the device.
•to prevent any unintentional power supply switch-on.
•to ensure that installation and maintenance work are
carried out by qualified, authorised personnel in possession of the appropriate tools.
•to guarantee a defined or controlled restarting of the process, after a power supply interruption.
•to use the device only if in perfect working order and in compliance with the instructions provided in the operating instructions.
•to observe the general technical rules when installing and using the device.
note
The device may be damaged by the fluid in contact with.
•Systematically check the chemical compatibility of the component materials of the device and the fluids likely to come into contact with it (for example: alcohols, strong or concentrated acids, aldehydes, alkaline compounds, esters, aliphatic compounds, ketones, halogenated aromatics or hydrocarbons, oxidants and chlorinated agents).
note
Elements / Components sensitive to electrostatic discharges
•This device contains electronic components sensitive to electrostatic discharges. They may be damaged if they are touched by an electrostatically charged person or object. In the worst case scenario, these components are instantly destroyed or go out of order as soon as they are activated.
•To minimise or even avoid all damage due to an electrostatic discharge, take all the precautions described in the EN 61340- 5-1 and 5-2 norms.
•Also ensure that you do not touch any of the live electrical components.
|
|
|
|
English |
5 |
|
|
|
Type 8045
General information
4.General information
4.1.Manufacturer's address and international contacts
To contact the manufacturer of the device, use following address: Bürkert SAS
Rue du Giessen BP 21
F-67220 TRIEMBACH-AU-VAL
You may also contact your local Bürkert sales office.
The addresses of our international sales offices are available on the internet at: www.burkert.com
4.2.Warranty conditions
5.Description of the name plate
|
1 |
2 |
3 |
4 |
|
5 |
|
|
in France |
FLOW 8045 SST LONG |
|
|
SUPPLY: 18-36V |
300 mA |
|||
DO1: 5-36V= 100mA |
|
|
|
DO2/3:250V 3A Max |
||||
DI1: 18-36V= 10mA |
IP65 |
|
|
Fluid: PN16:-15/110°C |
||||
Made |
S/N 11 162 |
|
|
|
|
|
|
|
|
00449673 |
|
|
|
|
|
W43M |
|
|
13 |
12 11 |
|
10 |
9 |
8 |
7 |
6 |
|
|
FLOW |
|
|
|
|
|
|
|
|
|
|
14 |
|
|
|
|
The condition governing the legal warranty is the conforming use of the 8045 in observance of the operating conditions specified in this manual.
4.3.Information on the Internet
You can find the user manuals and technical data sheets regarding the type 8045 at: www.burkert.com
|
|
6 |
English |
|
|
Type 8045
Technical data
1.Measured value and type of the device
2.Specification of the flow sensor
3.Specification of the DO1 digital output
4.Specifications of the relay outputs DO2 and DO3
5.Electrical power supply and current consumption
6.Manufacturing code
7.Conformity logo
8.Warning: Before using the device, take into account the technical specifications described in these operating instructions.
9.Fluid temperature range and fluid nominal pressure
10.Protection class of the device
11.Specification of the DI1 digital input
12.Serial number
13.Order code
14.Shows the flow direction
Fig. 1: Name plate of the 8045 flowmeter
6.Technical daTA
6.1.Conditions of use
Ambient temperature |
-10 to +60 °C |
|
|
Air humidity |
< 85%, non condensated |
|
|
Height above see level |
max. 2000 m |
|
|
Protection class acc. |
IP65, device wired and cable glands |
to EN 60529 |
tightened and cover lid screwed tight. |
|
|
6.2.Conformity to standards and directives
The device conforms to the EC directives through the following standards:
•EMC: EN 61000-6-2, EN 61000-6-3
•LVD: EN 61 010-1
•Vibration: EN 60068-2-6
•Shock: EN 60068-2-27
Pressure: article 3§3 of the Pressure Directive 97/23/CE. Acc. to the Pressure Directive 97/23/CE: the device can only be used in the following cases (depending on the max. pressure, the DN of the pipe and the fluid)
Type of fluid |
Conditions |
Fluid group 1, par. 1.3.a |
Forbidden |
|
|
|
|
English |
7 |
|
|
|
Type 8045
Technical data
Type of fluid |
Conditions |
|
Fluid group 2 par. 1.3.a |
DN ≤ 32 |
|
or DN > 32 and PNxDN ≤ 1000 |
||
|
||
Fluid group 1 par. 1.3.b |
PNxDN ≤ 2000 |
|
|
DN ≤ 200 |
|
Fluid group 2 par. 1.3.b |
or PN ≤ 10 |
|
|
or PNxDN ≤ 5000 |
6.3.Mechanical data
|
Material (8045 with |
Material (8045 |
|
Part |
with flow sensor in |
||
flow sensor in PVDF) |
|||
|
|
stainless steel) |
|
Housing / seal |
PC / NBR |
Black PPA / NBR |
|
|
|
|
|
Cover with lid / |
PC / silicone |
PSU / silicone |
|
seal |
|
|
|
Front foil |
Polyester |
||
|
|
|
|
M20x1,5 cable |
PA / neoprene |
||
glands / seal |
|
|
|
Screws |
Stainless steel |
||
|
|
|
|
Nut |
PC |
PPA |
|
|
|
|
|
Flow sensor |
PVDF |
Stainless steel 316L |
|
(exposed to the |
|
(DIN 1.4404) |
|
fluid) |
|
|
|
Material (8045 with |
Material (8045 |
|
Part |
with flow sensor in |
||
flow sensor in PVDF) |
|||
|
stainless steel) |
||
Seal |
FKM |
8045 with a G2" nut: |
|
|
|
FKM |
|
Earth ring of the |
Stainless steel 316L |
- |
|
flow sensor |
(DIN 1.4404) or Alloy |
|
|
|
C22 |
|
|
Electrodes holder |
- |
PEEK |
|
|
|
|
|
Electrodes |
Stainless steel 316L (DIN 1.4404) or Alloy |
||
|
C22 |
6.4.General data
Min. fluid conductivity |
20 µS/cm |
|
|
Fluid temperature |
The fluid temperature may be |
|
restricted by the fluid pressure, the |
|
material the flow sensor is made of |
|
and the material the S020 fitting used |
|
is made of. |
• 8045 with flow sensor |
• 0 to +80 °C |
in PVDF |
|
• 8045 with flow sensor in • -15 to +110 °C stainless steel
|
|
8 |
English |
|
|
Type 8045
Technical data
Fluid pressure |
The fluid pressure may be restricted |
|
by the fluid temperature, the material |
|
the flow sensor is made of, the |
|
material the S020 fitting used is made |
|
of and the DN of the fitting used. |
• 8045 with flow sensor |
• PN10 |
in PVDF |
|
•8045 with flow sensor in • PN16 stainless steel
P (bar) |
|
|
A |
|
|
|
10 |
|
|
|
|
|
|
9 |
|
|
|
|
PVDF (PN10) / Metal |
|
8 |
|
|
|
|
||
|
|
|
|
|
|
|
7 |
PVC + PP |
|
|
|
|
|
6 |
|
|
|
|
||
5 |
|
|
|
|
|
|
4 |
|
|
|
|
|
PP (PN10) |
3 |
|
PVC (PN10) |
|
|
||
|
|
|
|
|||
2 |
|
|
|
|
||
|
|
|
|
|
|
|
1 |
|
|
|
|
|
|
0 |
|
|
|
|
|
T (°C) |
|
0 |
+20 |
+40 |
+60 |
+80 |
|
|
|
A: range of use
Fig. 2: Fluid pressure /temperature dependency for a 8045 with PVDF flow sensor and a fitting S020 in metal, PVC, PVDF or PP
P (bar) |
|
|
|
A |
|
|
|
|
|
|
|
|
|
16 |
|
|
|
|
|
|
15 |
|
|
Metal |
|
|
|
14 |
|
|
|
|
|
|
|
|
(PN16) 1) |
|
|
|
|
13 |
|
|
|
|
DN100 for measuring |
|
12 |
PVDF |
|
|
|
devices with a clamp con- |
|
11 |
|
|
|
|
nection (PN10) |
|
10 |
|
|
|
|
|
|
9 |
|
|
|
|
|
|
8 |
PVC + PP |
|
PVDF (PN10) |
|||
7 |
|
|
|
|
||
|
|
|
|
|
|
|
6 |
|
|
|
|
|
|
5 |
|
|
|
|
|
|
4 |
|
|
PVC (PN10) |
|
|
|
3 |
|
|
|
|
||
2 |
|
|
|
PP (PN10) |
|
|
1 |
|
|
|
|
||
0 |
-20 |
0 |
+20 |
+40 |
+60 |
+80 +100 +120 |
|
||||||
|
|
|
|
|
|
T (°C) |
1) Except DN100 for measuring devices with a clamp connection
A: range of use
Fig. 3: Fluid pressure /temperature dependency for a 8045 with stainless steel flow sensor and a fitting S020 in metal, PVC, PVDF or PP
|
|
|
|
English |
9 |
|
|
|
Type 8045
Technical data
6.5.Electrical data
Power supply |
• 18-36 V DC, filtered and |
||
|
|
|
regulated |
|
|
• |
oscillation rate: ±5 % |
Current consumption |
300 mA max. (at 18 V DC) |
||
|
|
|
|
Transistor output DO1 |
|
|
|
• |
type |
• NPN / PNP (wiring dependent), |
|
|
|
|
open collector |
• |
function |
• pulse output (by default), user |
|
|
|
|
configurable |
• |
frequency |
• |
0-250 Hz |
• |
Electrical data |
• 5-36 V DC, 100 mA max. |
|
• duty cycle if f > 2 Hz |
• |
1/2 |
|
• |
min. pulse duration if |
• |
250 ms |
|
f < 2 Hz |
|
|
• |
protections |
• |
galvanically isolated, and |
|
|
|
protected against over- |
|
|
|
voltages, polarity reversals and |
|
|
|
short-circuits |
Relay outputs (DO2 and |
|
|
|
DO3) |
|
|
|
• |
operating |
• hysteresis (by default), configu- |
|
|
|
|
rable, normally open |
• |
electrical data of the load |
• |
250 V AC / 3 A or 30 V DC / |
|
|
|
3 A (resistive load) |
• |
max. breaking capacity |
• 750 VA (resistive load) |
|
• |
life span |
• |
min. 100000 cycles |
Current output AO1 |
|
|
|
• |
specification |
• 4-20 mA, sink or source (wiring |
|
|
|
|
dependent), 22 mA to indicate |
|
|
|
a fault |
• |
max. loop impedance |
• 1300 W at 36 V DC, 1000 W at |
|
|
|
|
30 V DC, 700 W at 24 V DC, |
|
|
|
450 W at 18 V DC |
Digital input (DI1) |
|
|
|
• |
supply voltage |
• |
18-36 V DC |
• |
input impedance |
• |
15 kW |
• |
min. pulse duration |
• |
200 ms |
• |
protections |
• galvanically isolated, and pro- |
|
|
|
|
tected against polarity reversals |
|
|
|
and voltage spikes |
|
|
10 |
English |
|
|
Type 8045
Installation and wiring
6.6.Electrical connection
Type of connection |
Through two M20x1,5 cable |
||
|
|
glands |
|
Cable specifications |
|
|
|
• |
cable type |
• |
shielded |
• |
Cross section |
• |
0,5 to 1,5 mm2 |
• |
Diameter of each cable: |
|
|
|
-- if only one cable is used |
|
-- 6...12 mm |
|
per cable gland |
|
|
|
-- if two cables are used per |
|
-- 4 mm, with the supplied |
|
cable gland |
|
multi-way seal |
7.Installation and wiring
7.1.Safety instructions
danger
Risk of injury due to high pressure in the installation. Risk of injury due to electrical voltage.
Risk of injury due to high fluid temperatures. Risk of injury due to the nature of the fluid.
Warning
Risk of injury due to non-conforming installation.
•The electrical and fluid installation can only be carried out by qualified and skilled staff with the appropriate tools.
•Install appropriate safety systems (correctly rated fuse and/or circuit-breaker);
•Respect standard NF C 15-100 / IEC 60364.
•Respect the assembly instructions for the fitting used.
Risk of injury due to unintentional switch on of power supply or uncontrolled restarting of the installation.
•Take appropriate measures to avoid unintentional activation of the installation.
•Guarantee a set or controlled restarting of the process subsequent to any intervention on the device.
|
|
|
|
English |
11 |
|
|
|
Type 8045
Installation and wiring
Protect this device against electromagnetic interference, ultraviolet rays and, when installed outdoors, the effects of the climatic conditions.
7.2.Recommandations for installing the 8045 on the pipe
→→Choose an S020 fitting appropriate to the velocity of the fluid inside the pipe.
→→Install the device on the pipe in such a way that the upstream and downstream distances are respected according to the design of the pipes, refer to standard EN ISO 5167-1 and Fig. 4 :
flow direction
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
50 x DN |
5 |
|
x DN |
|
40 x DN 5 x DN |
||||||||||||||||||||||
With control valve |
|
Pipe with 2 elbows at 90° in 3 |
|||||||||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
dimensions |
|||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
25 x DN 5 x DN |
|
|
20 x DN 5 x DN |
||||||||||||||||||||||||
Pipe with 2 elbows at 90° |
Pipe with 1 elbow at 90° or 1 |
||||||||||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
T-piece |
||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
18 x DN 5 x DN |
15 x DN 5 x DN |
With pipe expansion |
With pipe reduction |
Fig. 4: Upstream and downstream distances depending on the design of the pipes.
|
|
12 |
English |
|
|
Type 8045
Installation and wiring
→→Respect the following additional mounting conditions to ensure that the measuring device operates correctly:
-- Preferably install the device at a 45° angle to the horizontal centre of the pipe to avoid having deposits on the electrodes and false measurements due to air bubbles (see Fig. 5);
45°
45°
Fig. 5: Mounting angle on the pipe
-- Ensure that the pipe is always filled in the section around the device (see Fig. 6).
-- When mounting vertically ensure that the flow direction is in an upward direction (see Fig. 6).
Horizontal mounting |
|
Correct |
Incorrect |
Vertical mounting
|
flow direction |
Correct |
Incorrect |
Fig. 6: Filling of the pipe
-- Prevent the formation of air bubbles in the pipe in the section around the device (see Fig. 7).
-- Always mount the device upstream a possible injection point in the pipe of a high-conductivity fluid (for example: acid, base, saline,...).
|
|
|
|
English |
13 |
|
|
|
Type 8045
Installation and wiring
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Correct |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Incorrect |
|
flow direction |
|||||||||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Correct Incorrect
Fig. 7: Air bubbles within the pipe
→→If necessary, use a flow conditioner to improve measurement precision.
7.3.Installation into the pipe of a 8045 with a G2'' nut
In order to ensure a high accuracy of the measurements and good stability of the „flow zero“ point, install the device into the processed medium at least 24 H before calibration.
→→Install the S020 fitting into the pipe taking into account the recommendations in chap. 7.2.
→→Check that there is a seal on the fitting and that it is not damaged. Replace the seal if necessary.
→→Check that there is a seal (see mark 6, Fig. 8) on the flow sensor. →→Insert the nut (see mark 3, Fig. 8) on the fitting.
→→Insert the snap ring (mark 2 Fig. 8) into the groove (mark 5 Fig. 8).
→→Position the arrow on the side of the device in the direction of the flow: the totalizers will increment.
→→Insert the device (mark 1 Fig. 8) into the fitting. →→Tighten the nut (mark 3, Fig. 8) by hand on the device.
|
|
14 |
English |
|
|
Type 8045
Installation and wiring
FLOW |
1 |
6 |
flow direction |
Electrodes |
2 |
3 |
5 |
4 |
Fig. 8: Installation into the pipe of a 8045 with a G2'' nut
7.4.Installation into the pipe of a 8045 with a clamp connection
In order to ensure a high accuracy of the measurements and good stability of the „flow zero“ point, install the device into the processed medium at least 24 H before calibration.
→→Install the S020 fitting into the pipe taking into account the recommendations in chap. 7.2.
→→Install the seal (mark 3 Fig. 9) on the S020 fitting.
→→Make sure that the polarizing pin (mark 4 Fig. 9) is on the fitting.
→→Insert the device (mark 1 Fig. 9) into the fitting. Position the arrow on the side of the device in the direction of the flow: the totalizers will increment.
→→Tighten by hand the clamp collar (mark 2 Fig. 9). →→Charge the pipe to make sure the installation is tight.
1
2
4 |
3 |
flow direction
Fig. 9: Installation into the pipe of a 8045 with a clamp connection
|
|
|
|
English |
15 |
|
|
|
Type 8045
Installation and wiring
7.5.Wiring
danger
Risk of injury due to electrical voltage.
•Shut down the electrical power source of all the conductors and isolate it before carrying out work on the system.
•Observe all applicable accident protection and safety regulations for electrical equipment.
•Use a high quality electrical power supply (filtered and regulated).
•Make sure the installation is equipotential. See chap. 7.5.1.
•Use shielded cables with a temperature limit of 80 °C minimum.
•Do not install the cables near high voltage or high frequency cables; If this cannot be avoided, observe a min. distance of 30 cm.
•Protect the power supply by means of a 300 mA fuse and a switch.
•Protect the relays by means of a max. 3 A fuse and a circuit breaker (depending on the process).
•Do not apply both a dangerous voltage and a safety extralow voltage to the relays.
Insert the supplied stopper gasket into the unused cable gland to ensure the tightness of the device.
If two cables are used in the same cable gland, first insert the supplied multi-way seal.
To wire the device:
→→Loosen the screw from the lid. →→Flip the lid.
→→Loosen the 4 screws from the cover of the housing. →→Remove the cover.
→→Loosen the nuts of the cable glands.
→→Insert the cable through the nut then through the cable gland.
→→Make sure the earth cable coming from the housing and, on a version with stainless steel sensor, the cable coming from the flow sensor, are connected as shown in Fig. 15, chap. 7.5.2.
→→Wire acc. to chap. 7.5.1 to 7.5.3.
7.5.1.Equipotentiality of the installation
To ensure the equipotentiality of the installation (power supply - device - fluid):
→→Connect together the various earth spots in the installation to eliminate the potential differences that may occur between different earthes.
→→Observe faultless earthing of the shield of the power supply cable, at both ends.
|
|
16 |
English |
|
|
Type 8045
Installation and wiring
→→Connect the negative power supply terminal to the earth to suppress the effects of common mode currents. If this connection cannot be made directly, a 100 nF/50 V capacitor can be fitted between the negative power supply terminal and the earth.
→→Special attention has to be paid if the device is installed on plastic pipes because there is no direct earthing possible. Proper earthing is performed by earthing together the metallic instruments such as pumps or valves, that are as close as possible to the device. If no such instrument is near the device, insert metallic earth rings inside the plastic pipes upstream and downstream the device and connect these parts to the same earth. The earth rings must be in contact with the fluid.
Power cable shield
Power supply
18-36VDC
+
-
1)
Pipes in plastic
Valve, pump,... (or earthing rings, not provided, inserted into the pipe)
1) If a direct earth connection is not possible, fit a 100 nF / 50 V capacitor between the negative power supply terminal and the earth.
Fig. 10: Equipotentiality skeleton diagram with pipes in plastic
|
|
|
|
English |
17 |
|
|
|
Type 8045
Installation and wiring
Power cable shield
Power supply
18-36VDC
+
-
1)
Metal pipe
1) If a direct earth connection is not possible, fit a 100 nF / 50 V capacitor between the negative power supply terminal and the earth.
Fig. 11: Equipotentiality skeleton diagram with pipes in metal
7.5.2.Terminal assignment and use of the selectors
|
|
|
1 |
|
|
|
A |
|
|
|
|
|
|
|
|
|
2 |
|
CURRENT |
Iout L+ L- PE |
P+ P- |
|
|
|
SINK SOURCE |
AO1 |
Supply |
PULSE |
|
|
|
|
18...36 Vdc |
DO1 |
PE |
|
|
|
PE DI1 |
|
|
|
|
|
|
|
|
4 pin ribbon ca- |
|
|
- |
|
3 |
|
|
|
|
||
ble coming from B |
OFF ON |
|
+ |
|
|
|
|
|
|||
the flow sensor |
DO2 |
|
|
|
|
4
DO3
5
Selector A : see Fig. 13
Selector B : see Fig. 14
|
|
18 |
English |
|
|
Type 8045
Installation and wiring
Terminal block 1
Iout: 4-20 mA output (AO1) L+: V+ (positive voltage)
L-: 0V (power supply ground)
PE: functional earth, wired in the factory (see Fig. 15) P+: positive transistor output (DO1)
P-: negative transistor output (DO1)
Terminal block 2
PE: shieldings of both the power supply cable and the AO1 and DO1 output cables
Terminal block 3
PE: functional earth of the DI1 digital input -: negative signal of the DI1 input
+: positive signal of the DI1 input
Terminal block 4: wiring the DO2 relay output Terminal block 5: wiring the DO3 relay output
Fig. 12: Terminal assignment
Use switch A |
to configure the wiring of the 4-20 mA cur- |
|||||||||||||||||
rent output in sinking or sourcing mode. |
||||||||||||||||||
|
|
|
|
|
|
|
|
|
A |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
CURRENT |
|
CURRENT |
|||||||||||||||
SINK |
SOURCE |
SINK |
|
SOURCE |
||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|||||||
Wire the current output |
Wire the current output in |
|||||||||||||||||
in sourcing mode. |
sinking mode. |
|||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Fig. 13: Sink/source switch |
|
|
|
|
|
|
|
|
|
|
||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||
|
|
B to lock/unlock the |
ENTER |
|||||||||||||||
Use the switch |
|
|
|
|
|
key to prevent |
||||||||||||
|
|
|
|
|
||||||||||||||
unauthorized access to the configuration of the device. |
B
|
OFF ON |
|
ENTER |
The |
key is unlocked (de- |
|
fault position). |
OFF ON
ENTER |
|
The |
key is |
locked. |
|
Fig. 14: Switch to lock/unlock the ENTER key
|
|
|
|
English |
19 |
|
|
|
1 2 3 4 5 6
Earth cable coming from the housing.
On a version with stainless steel flow sensor, a second cable is coming from the sensor.
Fig. 15: Connection of the earth wire coming from the housing (made in the factory)
7.5.3.Wiring the DO2 and DO3 relay outputs
|
m |
DO2 |
|
|
|
230 VAC |
3 A |
|
|
|
|
|
3 A |
|
230 VAC |
|
|
|
|
DO3 |
Fig. 16: Wiring of the DO2 and DO3 relay outputs |
Type 8045
Installation and wiring
7.5.4.Wiring the AO1 current output
For safety reasons, secure the cables using a non-con- ducting cable clamp.
The 4-20 mA output can be wired in either sourcing or sinking mode.
Position the A switch to the right.
+ |
300 mA |
18-36 V DC |
- |
|
+ |
|
- |
|
|
|
|
4-20mA input at external |
|
(*) |
device |
I |
Power supply |
|
|
|
A |
|
|
CURRENT |
Iout L+ L- |
PE P+ P- |
|
|
SINK SOURCE |
AO1 |
Supply |
PULSE |
|
|
|
18...36 Vdc |
DO1 |
PE |
|
|
|
|
Fig. 17: Wiring of the 4-20 mA output (AO1) in sourcing mode
*) If a direct earth connection is not possible, fit a 100 nF/50 V capacitor between the negative power supply terminal and the earth
|
|
20 |
English |
|
|
Type 8045
Installation and wiring
Position the A switch to the left.
+
-
4-20mA input at external device
A
300 mA |
18-36 V DC |
|
+ |
|
- |
|
(*) |
I |
Power supply |
|
CURRENT |
Iout L+ L- |
PE P+ P- |
|
|
SINK SOURCE |
AO1 |
Supply |
PULSE |
|
|
|
18...36 Vdc |
DO1 |
PE |
|
|
|
|
Fig. 18: Wiring of the 4-20 mA output (AO1) in sinking mode
*) If a direct earth connection is not possible, fit a 100 nF/50 V capacitor between the negative power supply terminal and the earth
7.5.5.Wiring the DO1 transistor output
18-36 V DC |
300 mA |
+ |
5-36 VDC |
+ |
|
- |
|
- |
|
|
|
Power supply |
(*) |
+ |
|
|
|
- |
|
|
|
|
PLC |
Iout L+ L- PE |
P+ P- |
|
|
AO1 |
Supply |
PULSE |
|
18...36 Vdc |
DO1 |
PE |
|
|
|
|
Fig. 19: NPN wiring of the DO1 transistor output
*) If a direct earth connection is not possible, fit a 100 nF/50 V capacitor between the negative power supply terminal and the earth
|
|
|
|
English |
21 |
|
|
|
|
|
|
|
|
Type 8045 |
||
|
|
|
|
|
Installation and wiring |
||
18-36 V DC |
300 mA |
|
+ |
7.5.6. Wiring the DI1 digital input |
|||
+ |
|
5-36 VDC |
|
|
|
||
- |
|
|
- |
|
|
|
18-36 V DC |
|
|
|
|
|
|
||
|
|
|
|
|
|
|
+ |
Power supply |
(*) |
|
+ |
|
|
|
- |
|
|
|
- |
|
|
|
(*) |
|
|
|
|
PLC |
|
|
Power supply |
|
|
|
|
|
|
|
|
|
|
|
|
CURRENT |
Iout L+ L- |
PE P+ P- |
|
|
|
|
|
SINK SOURCE AO118...36 Vdc |
DO1 |
|
|
|
|
|
|
|
Supply |
PULSE |
|
|
|
|
|
|
PE DI1 |
|
PE |
|
|
|
|
|
|
|
|
|
Iout L+ L- |
PE P+ P- |
|
|
- |
|
|
|
Supply |
PULSE |
|
|
+ |
|
Switch |
|
AO118...36 Vdc |
DO1 |
PE |
|
|
||
|
|
|
DO2 |
|
|
|
|
|
|
|
|
|
|
|
|
Fig. 20: PNP wiring of the DO1 transistor output |
|
|
|
||||
*) If a direct earth connection is not possible, fit a 100 nF/50 V |
|
|
|
||||
capacitor between the negative power supply terminal and the earth |
|
|
|
||||
|
|
|
|
DO3 |
|
|
|
|
|
22 |
English |
|
|
Type 8045
Adjustment, commissioning
|
|
|
|
|
Power supply |
|
|
|
|
|
18-36 V DC |
|
|
|
|
|
+ |
|
|
|
|
|
- |
|
|
|
|
|
(*) |
CURRENT |
Iout L+ L- |
PE |
P+ P- |
|
|
SINK SOURCE |
AO1 |
Supply |
|
PULSE |
|
18...36 Vdc |
|
DO1 |
PE |
||
|
|
PE DI1 |
|
|
|
|
|
|
|
|
|
|
|
- |
|
|
Switch |
|
|
+ |
|
|
|
|
|
|
|
|
|
DO2 |
|
|
|
|
|
DO3 |
|
|
|
|
|
Fig. 21: Possible wirings of the DI1 digital input |
*) If a direct earth connection is not possible, fit a 100 nF/50 V capacitor between the negative power supply terminal and the earth
8.Adjustment, commissioning
8.1.Safety instructions
Warning
Risk of injury due to non-conforming operating.
Non-conforming operating could lead to injuries and damage the device and its surroundings.
•The operators in charge of operating must have read and understood the contents of these operating instructions.
•In particular, observe the safety recommendations and intended use.
•The device/installation must only be operated by suitably trained staff.
Warning
Danger due to non-conforming commissioning.
Nonconforming commissioning could lead to injuries and damage the device and its surroundings.
•Before commissioning, make sure that the staff in charge have read and fully understood the contents of the operating instructions.
•In particular, observe the safety recommendations and intended use.
•The device / the installation must only be commissioned by suitably trained staff.
•Before commissioning the device, enter the K factor of the fitting used. See chap. 8.6.
|
|
|
|
English |
23 |
|
|
|
Type 8045
Adjustment, commissioning
8.2.Operating levels of the device
The device has two operating levels: the Process level and the Configuration level.
Table 1: Default settings of the device
Function |
Default value |
|
Function |
Default value |
|
LANGUAGE |
English |
|
OUTPUT |
Hysteresis |
|
|
|
|
DO3 |
|
|
UNIT of the |
l/min. |
|
3-= |
0.000 |
|
|
|
||||
flow rate |
|
|
|
|
|
UNIT of the |
litre |
|
|
3+= |
0.000 |
totalizers |
|
|
|
|
|
K-FACTOR |
1.000 |
|
|
Not inverted |
|
|
|
|
|
|
|
OUTPUT |
4mA= 0.000 |
|
|
time delay = 0 |
|
AO1 |
|
|
|
|
|
20mA= 0.000 |
|
INPUT DI1 |
disable |
||
|
|
||||
|
|
|
|
|
|
OUTPUT |
pulse |
|
FILTER |
5, slow |
|
DO1 |
|
|
|
|
|
PU= 0.00 litre |
|
FREQUENC. |
50Hz |
||
|
|
||||
|
|
|
|
|
|
OUTPUT |
Hysteresis |
|
CUT-OFF |
0.000 |
|
DO2 |
|
|
|
|
|
2-= 0.000 |
|
BACKLIT |
level 9, acti- |
||
|
|
||||
|
|
|
|
vated for 30s |
|
|
2+= 0.000 |
|
K-SENSOR |
Kw= 1.000 |
|
|
|
|
|
|
|
|
Not inverted |
|
FLOW-W. |
W-= 0.000 |
|
|
|
|
|
|
|
|
time delay = 0 |
|
|
W+= 0.000 |
|
|
|
|
|
|
|
|
|
24 |
English |
|
|
Type 8045
Adjustment, commissioning
Configuration level
Process level
12.6 L/S
> 5 s |
> 2 s |
|
ENTER |
0...... |
9 |
16.45 MA |
|
Parameters menu |
|
0......9 |
ENTER |
|
LANGUAGE |
|
|
||
|
|
|
|
87654 L |
> 5 s |
|
UNIT |
|
|
||
|
|
|
|
|
|
|
K-FACTOR |
|
|
|
OUTPUT |
231 L. |
|
0......9 |
INPUT |
|
|
||
|
|
|
FILTER |
0......9 + |
|
|
TOTAL |
|
|
FREQUENC. |
|
> 2 s |
|
|
|
To reset the daily totalizer (identified by |
|
CUT-OFF |
|
a dot after the volume units). |
|
|
BACKLIT
Test Menu
CAL AO1
CALIB 0
0......9 FLOW
K-SENSOR
FLOW-W.
END
ENDSAVE N/Y
1)Accessible when the device status LED is orange or red.
2)If the ENTER key is unlocked.
Information menu 1)
MEAS. OVF
CAL. FAIL
0......9 NEG. FLOW
END
ENTER
or
2)
Process level
|
|
|
|
English |
25 |
|
|
|
Type 8045
Adjustment, commissioning
8.3.Description of the navigation keys and the state LEDs
•Scrolling up the parameters
•Incrementing the figure selected
Device state LED: see following table.
•Selecting the displayed parameter
•Confirming the settings
Status LED of relay DO3 (LED ON = contact closed)
Status LED of relay DO2 (LED ON = contact closed)
•Reading the messages
•Scrolling through the parameters
•Selecting the figure on the left
|
|
26 |
English |
|
|
Type 8045
Adjustment, commissioning
8.4.Using the navigation keys
You want to... |
Press... |
|
|
|||
move between parameters within a level or a menu. |
• |
|
|
to go to the next parameter. |
||
|
|
|
||||
|
• |
0......9 |
to go to the previous parameter. |
|||
access the Parameters menu. |
ENTER |
|
|
|
|
|
|
|
+ |
simultaneously for 5 s, in the Process level |
|||
|
|
|
||||
access the Test menu. |
|
|
|
|
|
ENTER |
|
0......9 |
+ |
|
+ |
simultaneously for 5 s, in the Process level |
|
access the Information menu. |
|
|
|
|
|
|
|
|
|
for 2 s, in the Process level, when the device status LED is orange or red. |
|||
reset the daily totalizer. |
|
|
|
|
|
|
|
0......9 |
+ |
|
simultaneously for 2 s, when the daily totalizer is displayed in the |
||
|
Process level |
|
||||
select the displayed parameter. |
ENTER |
|
|
|
|
|
|
|
|
|
|
|
|
confirm the displayed value. |
ENTER |
|
|
|
|
|
|
|
|
|
|
|
|
modify a numerical value. |
• |
|
|
to increase the blinking digit. |
||
|
0......9 |
|||||
|
• |
|
|
to select the digit at the left of the blinking digit. |
||
|
• |
0......9 |
+ |
|
to move the decimal point. |
|
|
|
|
English |
27 |
|
|
|
Type 8045
Adjustment, commissioning
8.5.Choosing the display language
When the device is energized for the first time, the display language is English.
LANGUAGE ENGLISH
DEUTSCH
0...... |
9 |
FRANÇAIS |
|
|
ITALIANO
ESPANOL
UNIT
→→Confirm the displayed language: The selected language is immediately active.
Fig. 22: Diagram of the "LANGUAGE" parameter of the Parameters menu
→→If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press ENTER to save the settings or not and go back to the Process level.
8.6.Entering the K factor of the fitting used:
The device determines the flow rate in the pipe using the fitting K factor.
The K factor of the fitting used can be entered here.
The device will use the new K factor as soon as "SAVE YES" is confirmed when leaving the Parameters menu.
|
|
28 |
English |
|
|
Type 8045
Adjustment, commissioning
K-FACTOR K=10.000
The display shows the K factor of the fitting, may it have been entered or determined by a teach-in procedure; this K factor is currently used by the device
→→Edit the parameter.
K=2.8500
→→Enter the K factor (value between 0,0001 and 9999,9) of the fitting used.
→→Confirm the displayed value.
OUTPUT |
|
|
RETURN |
|
|
|
VALID N/Y |
|
|
|
|||||
|
|
|
|
|
|
|
|
Is only displayed if the K factor has been changed. →→Confirm the entered K factor or not.
Fig. 23: Entering the K factor of the fitting used
→→ ENTER
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press to save the settings or not and go back to the Process level.
|
|
|
|
English |
29 |
|
|
|
Type 8045
Maintenance and troubleshooting
9.Maintenance and troubleshooting
9.1.Safety instructions
danger
Risk of injury due to high pressure in the installation. Risk of injury due to electrical voltage.
Risk of injury due to high fluid temperatures. Risk of injury due to the nature of the fluid.
Warning
Risk of injury due to non-conforming maintenance.
•Maintenance must only be carried out by qualified and skilled staff with the appropriate tools.
•Ensure that the restart of the installation is controlled after any interventions.
→→If a problem occurs, refer to the complete operating instructions available on the CD delivered with the device.
10.Packaging, Transport
note
Damage due to transport
Transport may damage an insufficiently protected device.
•Transport the device in shock-resistant packaging and away from humidity and dirt.
•Do not expose the device to temperatures that may exceed the admissible storage temperature range.
•Protect the electrical interfaces using protective plugs.
11. Storage
note
Poor storage can damage the device.
•Store the device in a dry place away from dust.
•Storage temperature of the device: -20 to +60 °C.
|
|
30 |
English |
|
|