Mobile Temperature System for testing components, hybrids, parts, modules, subassemblies and printed circuit boards at precise temperature

TEMPERATURE PERFORMANCE AND AIRFLOW CAPACITY

Temperature Range:

 

+125° to -55°C: approximately 10 seconds or less

 

1,3

-80° to +225°C (60 Hz System)

 

 

-75° to +225°C (50 Hz System)

Typical Temperature Transition Rate (air)1,2

-55° to +125°C: approximately 10 seconds or less

 

 

 

System air flow output

 

2.4 to 9 liters/second (5 to 18 scfm) CONTINUOUS

 

 

 

Temperature accuracy

 

1.0°C (when calibrated against NIST transfer

 

 

Temperature set, display and resolution

± 0.1°C

 

 

Temperature Control:

Control to within ± 0.1°C; SELF-TUNING available in DUT Control

DUT Sensor Ports

Type T and Type K Thermocouple ports standard

DUT Control

 

 

Remote interface ports

 

IEEE-488, RS232C Serial, and Start Test/End of Test/Stop

 

 

on First Fail (ST/ET/SFF)

 

 

 

1Note: Ultimate low temperature and system performance may vary under operating conditions less than or greater than nominal. 2Note: Transition performed under nominal operating conditions.

3Note: Low temperature extremes can be achieved at reduced airflow rates.

 

 

 

 

 

 

 

USER FEATURES

 

test programs may be created

 

Modes of Operation

 

Two: Manual and Program Modes

 

Test set-up configurations

 

12 Manual Mode test programs and 12 Program Mode

 

 

 

 

 

 

 

 

Test set-up storage

 

11 Manual Mode test programs, 11 Program Mode

 

 

 

 

 

test programs and 11 affiliated “Define Parameters”

 

 

 

 

 

test set-ups can be stored for later recall.

 

 

 

 

 

 

 

 

 

 

Ramp/soak/cycle configurations

 

In Program Mode, up to 12 sequences per test set-up;

 

 

 

 

 

table is displayed on-screen

 

 

 

 

 

 

 

 

 

 

DUT temperature control

 

Patented4 Dual Loop Temperature Control

 

 

 

 

 

 

 

 

 

 

Status indicators

 

Real-time status displayed on-screen and via remote

 

 

 

 

 

interface

 

 

 

 

 

 

 

 

 

 

Purge flow tester interface

 

Dry air purge protects tester electronics

 

 

 

protection

 

from condensation

 

 

 

 

 

 

 

 

 

 

Purge flow capacity

 

0.25 to 1.5 liters per second (0.5 to 3 scfm) airflow

 

 

 

 

 

 

 

 

 

 

Calibration

 

Automated, simplified and accurate for all airflows

 

 

 

 

 

and DUT types

 

 

 

 

 

 

 

 

 

 

Thermal head raising and lowering

 

Pneumatic control on thermal head,

 

 

 

 

 

operated manually or via remote interface

 

 

 

 

 

 

 

 

 

 

Head positioner movement

 

Manual locking (4 locks), 360° head rotation; head

 

 

 

 

 

can be manually pivoted, turned, tilted and vertically

 

 

 

 

 

swung for ease of interface at tester site.

 

 

 

 

 

 

 

 

 

 

Manipulator (arm) movement

 

Motorized raising and lowering of arm;

 

 

 

 

 

330° positioning range around the base unit

 

 

 

 

 

 

 

 

 

 

 

 

4Note: US Patent no. 4,734,872

 

 

 

 

 

 

 

Part No. SL10290 R/C

Specifications System ®ThermoStream TP04310A Model