Teledyne 100E User Manual 2

4.5 (4)

Operation Manual

Model 100E

UV Fluorescence SO2 Analyzer

Also supports operation of:

when used in conjunction with:

M100EU Analyzer

M100EU Addendum, PN05831

M100EH Analyzer

M100EH Addendum, PN 04621

M108E Analyzer

M108E Addendum, PN 03940

M108EU Analyzer

M108E Addendum, PN03949 and

 

M100EU Addendum, PN 04621

© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION

9480 CARROLL PARK DRIVE

SAN DIEGO, CA 92121-5201

USA

Toll-free Phone: 800-324-5190

Phone: 858-657-9800

Fax: 858-657-9816

Email: api-sales@teledyne.com

Website: http://www.teledyne-api.com/

Copyright 2008 - 2011

04515F DCN6048

Teledyne Advanced Pollution Instrumentation

06 April 2011

ABOUT THIS MANUAL

This M100E manual, PN 04515, is comprised of the following documents:

Part No.

Rev

Name/Description

04515

F

M100E Operation Manual (main content)

05036

E

Appendix A, Menu Trees and related software documentation, Rev G.4.

05360

Q

Spare Parts List (in Appendix B of this manual)

04526

S

Recommended Spare Parts Stocking Levels (in Appendix B of this manual)

04728

A

AKIT, Spares (in Appendix B of this manual)

04357

A

AKIT, Expendables (in Appendix B of this manual)

01475

A

AKIT, Expendables for IZS option (in Appendix B of this manual)

04796

E

Appendix C, Test Funct & Repair Qstnr

0401101

A

Interconnect List, M100E (in Appendix D of this manual)

04011

A

Interconnect Diagram, M100E (in Appendix D of this manual)

04354

D

PCA, 04003, Pressure Flow Sensor Board (in Appendix D of this manual)

04524

D

Schematic, Relay PCB (in Appendix D of this manual)

05703

A

PCA, 05702, Motherboard, E-Series Gen 4 (in Appendix D of this manual)

04181

H

PCA, 04180, PMT Preamp (in Appendix D of this manual)

04259

A

PCA, 04258, Keyboard Display Interface (in Appendix D of this manual)

05064

C

PCA, 05063, Dual UV Detector (in Appendix D of this manual)

04693

E

PCA, 04692, UV Lamp Driver (in Appendix D of this manual)

 

 

 

04932

C

PCA, Thermo-Electric Cooler Board (in Appendix D of this manual)

04468

B

PCA, 04467, Analog Output Series Res (in Appendix D of this manual)

NOTE

Please read this manual in its entirety before operating the instrument.

04515F DCN6048

i

Table of Contents Teledyne API M100E Analyzer Operation Manual

REVISION HISTORY

M100E Operation Manual, PN04515

Document

PN

Rev

DCN

Change Summary

2011 April 06, Rev F, DCN6048

 

 

 

M100E Operation

04515

F

6048

Administrative corrections in Revision History

Manual

 

 

 

section: typos.

2011 March 11, Rev E, DCN6019

 

 

M100E Operation

04515

E

6019

Combine with M100E “Text” (PN05138) for single

Manual

 

 

 

document/PN; obsolete 05138.

 

 

 

 

Change Disk-on-Module replacement procedure

 

 

 

 

from A-Crosser to ICOP CPU (Mntnc section)

 

 

 

 

Add MODBUS quick setup to Remote Op section

 

 

 

 

Add desiccant bag change to maintenance sched.

2009 Dec 15, M100E Manual, PN045150100 Rev D, DCN5637

Top Assy Manual

045150100

D

5637

Initial development of this manual’s top assy

 

 

 

 

construct, which is the fully assembled manual

 

 

 

 

comprised of the PDF conversions of all documents

 

 

 

 

listed in the About This Manual section. Revision

 

 

 

 

starts at “D” to be in step with Rev History of original

 

 

 

 

manual.

Title/front matter

04515

D

5637

Show current Rev letter of all documents comprising

 

 

 

 

the top assy of this manual (in About This Manual).

 

 

 

 

Initiate Revision History

Text

05138

C

5637

Insert new CPU information in place of old, Sections

 

 

 

 

11.4.1 11.4.1.1 and 11.4.1.2 and new Fig 11-11.

 

 

 

 

Replace all occurrences of disk-on-chip (DOC) with

 

 

 

 

disk-on-module (DOM).

 

 

 

 

Replace illustrations Fig 5-2, 5-6, 5-7, 5-8, 7-5. 7-12,

 

 

 

 

for greater detail and/or clarity.

 

 

 

 

Revise RS-485 configuration instructions, Sec. 7.1.4,

 

 

 

 

to reflect new CPU.

 

 

 

 

Revise Multidrop RS-232 Set-up Instructions, Sec

 

 

 

 

7.1.7, to reflect E-series parameters.

 

 

 

 

Add desiccant bag replacement to maintenance

 

 

 

 

requirements (per DCR 6559)

For the purpose of capturing this manual’s construct with its initial Revision History implementation, the following list documents the current Rev of each part comprising this manual at Rev D; any future changes to these documents will be recorded in the Revision History section and their Rev letters updated in the About This Manual section:

Title

04515

D

 

 

 

 

 

 

 

Text

05138

C

 

 

Relay PCA Addend.

05118

B3

 

 

S/W (Rev G.3) (AppxA)

05036

D

 

 

ii

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual Table of Contents

Document

PN

Rev

DCN

Change Summary

SPL

05360

Q

 

 

RSSL

04526

S

 

 

 

 

 

 

 

AKIT, Spares

04728

A

 

 

 

 

 

 

 

AKIT, Expendables

04357

A

 

 

AKIT, Expendables, IZS

01475

A

 

 

Repair Request (AppxC)

04796

D

 

 

Interconnect List

0401101

A

 

 

 

 

 

 

 

Interconnect Diagram

04011

A

 

 

 

 

 

 

 

PCA04003PressFlwSnsr

04354

D

 

 

PCB,Relay

04524

D

 

 

PCA05702MoBrd

05703

A

 

 

PCA04180PMT Preamp

04181

H

 

 

 

 

 

 

 

PCA04258Kybrd Int

04259

A

 

 

 

 

 

 

 

PCA05063 UVDetect

05064

C

 

 

PCA04692 LmpDrvr

04693

E

 

 

PCA ThermCooler

04932

C

 

 

PCA Analog Outpt

04468

B

 

 

 

 

 

 

 

04515F DCN6048

iii

Table of Contents

Teledyne API M100E Analyzer Operation Manual

This page intentionally left blank.

iv

04515F DCN6048

TABLE OF CONTENTS

 

SECTION I GENERAL INFORMATION ..................................................................................

13

1. INTRODUCTION ................................................................................................................

15

1.1. Safety Messages...........................................................................................................................................

15

1.2. M100E Overview...........................................................................................................................................

16

1.3. M100E Documentation..................................................................................................................................

16

1.3.1. Using This Manual .................................................................................................................................

17

2. SPECIFICATIONS, APPROVALS AND WARRANTY .......................................................

19

2.1. Specifications ................................................................................................................................................

19

2.2. EPA Equivalency Designation ......................................................................................................................

20

2.3. CE Mark Compliance ....................................................................................................................................

21

2.3.1. Emissions Compliance...........................................................................................................................

21

2.3.2. Safety Compliance.................................................................................................................................

21

2.4. Warranty........................................................................................................................................................

21

3. GETTING STARTED ..........................................................................................................

23

3.1. M100E Analyzer Layout ................................................................................................................................

23

3.2. Unpacking the M100E Analyzer....................................................................................................................

27

3.3. Electrical Connections...................................................................................................................................

28

3.3.1. Power Connection..................................................................................................................................

28

3.3.2. Analog Output Connections..................................................................................................................

29

3.3.3. Connecting the Status Outputs..............................................................................................................

29

3.3.4. Connecting the Control Inputs ...............................................................................................................

31

3.3.5. Connecting the Serial Ports ...................................................................................................................

32

3.3.6. Connecting to a LAN or the Internet ......................................................................................................

32

3.3.7. Connecting to a Multidrop Network........................................................................................................

32

3.4. Pneumatic Connections ................................................................................................................................

32

3.4.1. Calibration Gases ..................................................................................................................................

32

3.5. Initial Operation .............................................................................................................................................

36

3.6. Initial Calibration of the M100E .....................................................................................................................

42

3.6.1. Interferents for Measurements...............................................................................................................

42

3.6.2. Initial Calibration Procedure for M100E Analyzers without Options ......................................................

43

3.6.3. O2 Sensor Calibration Procedure...........................................................................................................

47

3.6.4. CO2 Sensor Calibration Procedure ........................................................................................................

47

4. FREQUENTLY ASKED QUESTIONS & GLOSSARY .......................................................

49

4.1. FAQ’s ............................................................................................................................................................

49

4.2. Glossary ........................................................................................................................................................

51

5. OPTIONAL HARDWARE AND SOFTWARE .....................................................................

53

5.1. Optional Pumps (OPT 10 to OPT 13) ...........................................................................................................

53

5.2. Rack Mount Kits (OPT 20A, OPT 20B, OPT 21, OPT 23)............................................................................

53

5.3. Carrying Strap/Handle (OPT 29)...................................................................................................................

54

5.4. Current Loop Analog Outputs (Option 41) ....................................................................................................

54

5.5. Optional Kits (OPT 42A, OPT 43, OPT 45)...................................................................................................

56

5.6. NO Optical Filter (OPT 47B) .........................................................................................................................

56

5.7. Calibration Valves Options............................................................................................................................

56

5.8. Communication Options................................................................................................................................

60

5.8.1. RS-232 Modem Cable (OPT 60A, OPT 60B, OPT 60C) .......................................................................

60

5.8.2. Concentration Alarm Relay (Option 61).................................................................................................

60

5.8.3. RS-232 Multidrop (Option 62)................................................................................................................

61

5.8.4. Ethernet (Option 63A) ............................................................................................................................

62

5.8.5. Ethernet + Multidrop (OPT 63C) ............................................................................................................

63

5.9. Second Gas Sensors ....................................................................................................................................

63

5.9.1. Oxygen Sensor (Option 65A).................................................................................................................

63

5.9.2. Carbon Dioxide Sensor (Option 67A) ....................................................................................................

65

5.10. Additional Manuals......................................................................................................................................

68

5.11. Extended Warranty (OPT 92B and OPT 93C)............................................................................................

68

04515F DCN6048

v

Table of Contents

Teledyne API M100E Analyzer Operation Manual

5.12. Special Features .........................................................................................................................................

 

69

SECTION II OPERATING INSTRUCTIONS............................................................................

 

71

6. OPERATING INSTRUCTIONS...........................................................................................

 

73

6.1. Overview of Operating Modes ......................................................................................................................

 

73

6.2. Sample Mode ................................................................................................................................................

 

74

6.3. Calibration Mode ...........................................................................................................................................

 

78

6.4. Setup Mode...................................................................................................................................................

 

80

6.5. SETUP – CFG: Viewing the Analyzer’s Configuration Information ..............................................................

82

6.6. SETUP – CLK: Setting the Internal Time-of-Day Clock................................................................................

 

83

6.7. SETUP – RNGE: Analog Output Reporting Range Configuration................................................................

85

6.8. SETUP – VARS: Using the Internal Variables..............................................................................................

 

93

6.9. SETUP – DIAG: Using the Diagnostics Functions........................................................................................

 

95

7. REMOTE OPERATION AND ADVANCED FEATURES ..................................................

119

7.1. SETUP – COMM: Setting Up the Analyzer’s Communication Ports..........................................................

119

7.1.3. RS-232 COM Port Cable Connections ...............................................................................................

 

121

7.1.4. RS-485 Configuration of COM2..........................................................................................................

 

122

7.1.6. Ethernet Card Configuration ...............................................................................................................

 

124

7.1.7. MODBUS Setup..................................................................................................................................

 

131

7.1.8. Multidrop RS-232 Set Up ....................................................................................................................

 

133

7.2. Using the Internal Data Acquisition System (iDAS) ...................................................................................

 

139

7.3. Remote Operation......................................................................................................................................

 

158

8. CALIBRATION PROCEDURES .......................................................................................

 

175

8.1. Calibration Preparations.............................................................................................................................

 

175

8.2. Manual Calibration .....................................................................................................................................

 

177

8.3. Manual Calibration Checks ........................................................................................................................

 

179

8.4. Manual Calibration with Zero/Span Valves ................................................................................................

 

180

8.5. Manual Calibration with IZS Option ...........................................................................................................

 

182

8.6. Manual Calibration Checks with IZS or Zero/Span Valves ........................................................................

 

183

8.7. Manual Calibration in DUAL or AUTO Reporting Range Modes ...............................................................

185

8.8. Automatic Calibration (AutoCal).................................................................................................................

 

186

8.9. Calibration Quality......................................................................................................................................

 

189

8.10. Calibration of Optional Sensors ...............................................................................................................

 

189

8.10.1. O2 Sensor Calibration Procedure......................................................................................................

 

189

8.10.2. CO2 Sensor Calibration Procedure...................................................................................................

 

193

9. EPA PROTOCOL CALIBRATION....................................................................................

 

197

9.1. Calibration Requirements...........................................................................................................................

 

197

9.2. Level 1 Calibrations versus Level 2 Checks ..............................................................................................

 

201

9.3. ZERO and SPAN Checks ..........................................................................................................................

 

202

9.4. Precisions Calibration Procedures and Checks.........................................................................................

 

202

9.5. Dynamic Multipoint Span Calibration .........................................................................................................

 

203

9.6. Special Calibration Requirements for Dual Range or Auto Range............................................................

204

9.7. References.................................................................................................................................................

 

204

SECTION III TECHNICAL INFORMATION...........................................................................

 

205

10. INSTRUMENT MAINTENANCE ....................................................................................

 

207

10.1. Maintenance Schedule.............................................................................................................................

 

209

10.2. Predictive Diagnostics..............................................................................................................................

 

211

10.3. Maintenance Procedures .........................................................................................................................

 

212

10.3.7. Performing a Sample Flow Check ....................................................................................................

 

216

10.3.8. Hydrocarbon Scrubber (Kicker) ........................................................................................................

 

216

11. THEORY OF OPERATION ............................................................................................

 

219

11.1. Measurement Principle ............................................................................................................................

 

219

11.1.1. SO2 Ultraviolet Fluorescence............................................................................................................

 

219

11.2. The UV Light Path....................................................................................................................................

 

222

11.2.1. UV Source Lamp...............................................................................................................................

 

223

11.2.2. The Reference Detector....................................................................................................................

 

223

11.2.3. The PMT ...........................................................................................................................................

 

223

vi

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual

Table of Contents

11.2.4. UV Lamp Shutter & PMT Offset........................................................................................................

223

11.2.5. Optical Filters ....................................................................................................................................

224

11.3. Pneumatic Operation ...............................................................................................................................

228

11.3.4. Hydrocarbon Scrubber (Kicker) ........................................................................................................

231

11.4. Electronic Operation.................................................................................................................................

233

11.5. Power Supply and Circuit Breaker ...........................................................................................................

247

11.6. Communications Interface .......................................................................................................................

248

11.7. Software Operation ..................................................................................................................................

252

12. TROUBLESHOOTING & REPAIR.................................................................................

255

12.1. General Troubleshooting..........................................................................................................................

255

12.1.2. Fault Diagnosis with Test Functions .................................................................................................

258

12.2. Status LEDs .............................................................................................................................................

261

12.2.1. Motherboard Status Indicator (Watchdog)........................................................................................

261

12.2.2. CPU Status Indicators.......................................................................................................................

261

12.2.3. Relay Board Status LEDs .................................................................................................................

262

12.3. Gas Flow Problems..................................................................................................................................

263

12.4. Calibration Problems................................................................................................................................

264

12.5. Other Performance Problems ..................................................................................................................

267

12.6. Subsystem Checkout ...............................................................................................................................

269

12.6.8. RS-232 Communication....................................................................................................................

274

12.7. Repair Procedures ...................................................................................................................................

279

12.7.1. Disk-on-Module (DOM) Replacement...............................................................................................

279

12.8. Technical Assistance ...............................................................................................................................

293

13. A PRIMER ON ELECTRO-STATIC DISCHARGE.........................................................

295

13.1. How Static Charges are Created .............................................................................................................

295

13.2. How Electro-Static Charges Cause Damage...........................................................................................

296

13.3. Common Myths About ESD Damage ......................................................................................................

297

13.4. Basic Principles of Static Control .............................................................................................................

298

13.4.1. General Rules ...................................................................................................................................

298

13.5. Basic anti-ESD Procedures for Analyzer Repair and Maintenance.........................................................

299

13.5.1. Working at the Instrument Rack .......................................................................................................

299

13.5.2. Working at an Anti-ESD Work Bench ...............................................................................................

299

13.5.3. Transferring Components from Rack to Bench and Back ................................................................

300

13.5.4. Opening Shipments from Teledyne API’s Customer Service...........................................................

300

13.5.5. Packing Components for Return to Teledyne API’s Customer Service ...........................................

301

INDEX ................................................................................................................................

303

APPENDIX A – MENU TREES and VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX B - SPARE PARTS, M100E

APPENDIX C - REPAIR QUESTIONNAIRE, M100E

APPENDIX D - ELECTRONIC SCHEMATICS, M100E

LIST OF FIGURES

Figure 3-1:

Front Panel Layout.......................................................................................................................

23

Figure 3-2:

Rear Panel Layout .......................................................................................................................

24

Figure 3-3:

M100E Layout (Basic Unit – No Valve Options)..........................................................................

25

Figure 3-4:

M100E Internal Gas Flow (Basic Configuration) .........................................................................

26

Figure 3-5:

Analog Output Connector ............................................................................................................

29

Figure 3-6:

Status Output Connector .............................................................................................................

30

Figure 3-7:

Control Input Connector...............................................................................................................

31

Figure 3-8:

Pneumatic Connections–Basic Configuration–Using Bottled Span Gas.....................................

34

Figure 3-9:

Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator.............................

35

Figure 3-10:

Startup – Front Panel Display......................................................................................................

37

Figure 3-11:

Warning Messages ......................................................................................................................

38

04515F DCN6048

 

vii

Table of Contents

Teledyne API M100E Analyzer Operation Manual

Figure 3-12:

Functional Check .........................................................................................................................

41

Figure 3-13:

Reporting Range Verification.......................................................................................................

43

Figure 3-14:

Dilution Ratio Setup .....................................................................................................................

44

Figure 3-15:

SO2 Span Gas Setting .................................................................................................................

45

Figure 3-16:

Zero/Span Calibration Procedure ................................................................................................

46

Figure 5-1:

Carrying Strap Handle and Rack Mount Brackets.......................................................................

54

Figure 5-2:

Current Loop Option Installed on the Motherboard .....................................................................

55

Figure 5-3:

Pneumatic Diagram of the M100E with Z/S Option Installed.......................................................

57

Figure 5-4:

Pneumatic Diagram of the M100E with IZS Options Installed.....................................................

58

Figure 5-5:

Concentration Alarm Relay..........................................................................................................

60

Figure 5-6:

Multidrop Card Seated on CPU above Disk on Module ..............................................................

62

Figure 5-7:

Ethernet Card...............................................................................................................................

62

Figure 5-8:

Rear Panel with Ethernet Installed ..............................................................................................

63

Figure 5-9:

Oxygen Sensor - Theory of Operation.........................................................................................

64

Figure 5-10:

Internal Pneumatics with O2 Sensor ............................................................................................

65

Figure 5-11:

CO2 Sensor Theory of Operation.................................................................................................

66

Figure 5-12:

Internal Pneumatics with CO2 ......................................................................................................

67

Figure 5-13:

CO2 Sensor Option PCA Layout and Electronic Connections.....................................................

68

Figure 6-1:

Front Panel Display......................................................................................................................

73

Figure 6-2:

Viewing M100E TEST Functions .................................................................................................

76

Figure 6-3:

Viewing and Clearing M100E WARNING Messages ..................................................................

77

Figure 6-4:

SETUP – Enable Calibration Password Security ........................................................................

79

Figure 6-5:

SETUP – Enter Calibration Mode Using Password.....................................................................

80

Figure 6-6:

SETUP – Configuration Information ............................................................................................

82

Figure 6-7:

SETUP – Clock ............................................................................................................................

83

Figure 6-8:

SETUP – Clock Speed Variable ..................................................................................................

84

Figure 6-9:

SETUP – Analog Output Connector Key .....................................................................................

85

Figure 6-10:

SETUP RNGE – Reporting Range Mode ....................................................................................

87

Figure 6-11:

SETUP RNGE – Single Range Mode..........................................................................................

88

Figure 6-12:

SETUP RNGE – Dual Range Mode ............................................................................................

89

Figure 6-13:

SETUP RNGE – Auto Range Mode ............................................................................................

90

Figure 6-14:

SETUP RNGE – Concentration Units Selection..........................................................................

91

Figure 6-15:

SETUP RNGE – Dilution Ratio ....................................................................................................

92

Figure 6-16:

SETUP – VARS Menu .................................................................................................................

94

Figure 6-17:

DIAG Menu ..................................................................................................................................

96

Figure 6-18:

DIAG – Signal I/O Menu ..............................................................................................................

97

Figure 6-19:

DIAG – Analog Output Menu .......................................................................................................

98

Figure 6-20:

DIAG – Analog I/O Configuration Menu....................................................................................

101

Figure 6-21:

DIAG – Analog Output Calibration Mode..................................................................................

102

Figure 6-22:

DIAG – Analog Output Calibration Mode – Single Analog Channel.........................................

103

Figure 6-23:

DIAG – Analog Output – Auto Cal or Manual Cal Selection for Channels ...............................

104

Figure 6-24:

Setup for Calibrating Analog Outputs .......................................................................................

105

Figure 6-25:

Analog Output – Voltage Adjustment........................................................................................

106

Figure 6-26:

Analog Output – Offset Adjustment ..........................................................................................

107

Figure 6-27:

Setup for Calibrating Current Outputs ......................................................................................

108

Figure 6-28:

Analog Output – Zero and Span Value Adjustment for Current Outputs..................................

109

Figure 6-29:

DIAG – Analog Output – AIN Calibration..................................................................................

110

Figure 6-30:

DIAG – Optic Test.....................................................................................................................

111

Figure 6-31:

DIAG – Electrical Test...............................................................................................................

112

Figure 6-32:

DIAG – Lamp Calibration..........................................................................................................

113

Figure 6-33:

DIAG – Pressure Calibration ....................................................................................................

114

Figure 6-34:

DIAG – Flow Calibration ...........................................................................................................

115

Figure 6-35:

DIAG – Test Channel Output....................................................................................................

116

Figure 7-1:

SETUP – COMM Menu.............................................................................................................

119

Figure 7-2:

COMM – Analyzer ID ...............................................................................................................

120

Figure 7-3:

Back Panel Connector Pin-Outs for RS-232 and COM2 in RS-232 Mode...............................

121

Figure 7-4:

CPU Connector Pin-Outs for RS-232 and COM2 in RS-232 Mode..........................................

122

Figure 7-5:

CPU RS-485 Setup...................................................................................................................

123

viii

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual

Table of Contents

Figure 7-6:

Rear Panel Connector Pin-Outs for COM2 in RS-485 Mode ...................................................

123

Figure 7-7:

CPU Connector Pin-Outs for COM2 in RS-485 Mode..............................................................

124

Figure 7-8:

COMM – LAN / Internet Configuration Properties ....................................................................

127

Figure 7-9:

COMM – Manually Configuring the Network IP Addresses (Step 1)........................................

128

Figure 7-10:

COMM – Manually Configuring the Network IP Addresses (Step 2)........................................

129

Figure 7-11:

COMM – Change Host Name..................................................................................................

130

Figure 7-12:

Location of JP2 on RS232-Multidrop PCA (OPT 62)................................................................

133

Figure 7-13:

RS232-Multidrop PCA Host/Analyzer Interconnect Diagram ...................................................

134

Figure 7-14:

COMM – Enable Hessen Protocol (Example) ..........................................................................

136

Figure 7-15:

COMM – COMM Port Baud Rate .............................................................................................

137

Figure 7-16:

COMM – COMM Port Baud Rate .............................................................................................

138

Figure 7-17:

Default iDAS Channels Setup...................................................................................................

143

Figure 7-18:

iDAS – Data Acquisition Menu.................................................................................................

144

Figure 7-19:

iDAS – Editing iDAS Data Channels........................................................................................

145

Figure 7-20:

iDAS – Editing Data Channel Name ........................................................................................

146

Figure 7-21:

iDAS

– Trigger Events .............................................................................................................

147

Figure 7-22:

iDAS – Editing iDAS Parameters.............................................................................................

148

Figure 7-23:

iDAS

– Configuring Parameters for a Specific Data Parameter ..............................................

149

Figure 7-24:

iDAS

– Define the Report Period .............................................................................................

151

Figure 7-25:

iDAS

– Define the Report Period .............................................................................................

152

Figure 7-26:

iDAS – RS-232 Report Function..............................................................................................

153

Figure 7-27:

iDAS

– Disabling / Enabling Data Channels............................................................................

154

Figure 7-28:

iDAS

– Holdoff Feature............................................................................................................

155

Figure 7-29:

Sample APICOM User Interface for Configuring the iDAS.......................................................

156

Figure 7-30:

iDAS Configuration Through a Terminal Emulation Program...................................................

157

Figure 7-31:

Status Output Connector ..........................................................................................................

158

Figure 7-32:

Control Inputs with Local 5 V Power Supply.............................................................................

160

Figure 7-33:

Control Inputs with External 5 V Power Supply ........................................................................

160

Figure 7-34:

COMM – Remote Access by Modem .......................................................................................

164

Figure 7-35:

COMM – Initialize the Modem ..................................................................................................

165

Figure 7-36:

APICOM Remote Control Program Interface............................................................................

167

Figure 7-37:

COMM – Activating Hessen Protocol .......................................................................................

169

Figure 7-38:

COMM – Select Hessen Protocol Type ....................................................................................

170

Figure 7-39:

COMM – Select Hessen Protocol Response Mode..................................................................

171

Figure 7-40:

COMM – Status Flag Bit Assignment .......................................................................................

173

Figure 8-1:

Setup for Manual Calibration without Z/S Valve or IZS Option (Step 1)...................................

177

Figure 8-2:

Setup for Manual Calibration without Z/S Valve or IZS Option (Step 2)...................................

177

Figure 8-3:

Setup for Manual Calibration without Z/S Valve or IZS Option (Step 3)...................................

178

Figure 8-4:

Setup for Manual Calibration Checks .......................................................................................

179

Figure 8-5:

Setup for Manual Calibration with Z/S Valve Option Installed (Step 1) ....................................

180

Figure 8-6:

Setup for Manual Calibration with Z/S Valve Option Installed (Step 2) ....................................

180

Figure 8-7:

Setup for Manual Calibration with Z/S Valve Option Installed (Step 3) ....................................

181

Figure 8-8:

Manual Calibration with IZS Option ..........................................................................................

182

Figure 8-9:

Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 1) ............................

183

Figure 8-10:

Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 2) ............................

184

Figure 8-11:

Manual Calibration in Dual/Auto Reporting Range Modes.......................................................

185

Figure 8-12:

AUTO CAL – User Defined Sequence......................................................................................

188

Figure 8-13:

O2 Sensor Calibration Set Up ...................................................................................................

189

Figure 8-14:

O2 Span Gas Concentration Set Up .........................................................................................

190

Figure 8-15:

Activate O2 Sensor Stability Function .......................................................................................

191

Figure 8-16:

O2 Zero/Span Calibration..........................................................................................................

192

Figure 8-17:

CO2

Sensor Calibration Set Up.................................................................................................

193

Figure 8-18:

CO2

Span Gas Concentration Setup ........................................................................................

193

Figure 8-19:

Activate CO2 Sensor Stability Function ....................................................................................

194

Figure 8-20:

CO2 Zero/Span Calibration .......................................................................................................

195

Figure 9-1:

Dynamic Multipoint Span Calibration........................................................................................

204

Figure 10-1:

Sample Particulate Filter Assembly ..........................................................................................

212

Figure 10-2:

Critical Flow Orifice Assembly ..................................................................................................

214

04515F DCN6048

ix

Table of Contents

Teledyne API M100E Analyzer Operation Manual

Figure 10-3:

Simple Leak Check Fixture.......................................................................................................

217

Figure 10-4:

Hydrocarbon Scrubber Leak Check Setup ...............................................................................

217

Figure 11-1:

UV Absorption...........................................................................................................................

220

Figure 11-2:

UV Light Path............................................................................................................................

222

Figure 11-3:

Source UV Lamp Construction .................................................................................................

223

Figure 11-4:

Excitation Lamp UV Spectrum Before/After Filtration...............................................................

224

Figure 11-5:

PMT Optical Filter Bandwidth ...................................................................................................

225

Figure 11-6:

Effects of Focusing Source UV in Sample Chamber................................................................

226

Figure 11-7:

Gas Flow and Location of Critical Flow Orifice.........................................................................

229

Figure 11-8:

Flow Control Assembly & Critical Flow Orifice..........................................................................

230

Figure 11-9:

M100E Hydrocarbon Scrubber (Kicker)....................................................................................

231

Figure 11-10:

M100E Electronic Block Diagram .............................................................................................

233

Figure 11-11:

CPU Board Annotated ..............................................................................................................

235

Figure 11-12:

M100E Sensor Module .............................................................................................................

236

Figure 11-13:

M100E Sample Chamber..........................................................................................................

237

Figure 11-14:

PMT Housing Assembly............................................................................................................

238

Figure 11-15:

Basic PMT Design ....................................................................................................................

239

Figure 11-16:

PMT Cooling System ................................................................................................................

240

Figure 11-17:

PMT Preamp Block Diagram ....................................................................................................

241

Figure 11-18:

Relay Board Status LED Locations ..........................................................................................

243

Figure 11-19:

Power Distribution Block Diagram ............................................................................................

247

Figure 11-20:

Interface Block Diagram............................................................................................................

248

Figure 11-21:

M100E Front Panel Layout .......................................................................................................

249

Figure 11-22:

Keyboard and Display Interface Block Diagram .......................................................................

250

Figure 11-23:

Basic Software Operation .........................................................................................................

252

Figure 11-24:

Calibration Slope and Offset.....................................................................................................

253

Figure 12-1:

Viewing and Clearing Warning Messages ................................................................................

256

Figure 12-2:

Example of Signal I/O Function ................................................................................................

260

Figure 12-3:

CPU Status Indicator ................................................................................................................

261

Figure 12-4:

Location of Relay Board Power Configuration Jumper.............................................................

270

Figure 12-5:

Manual Activation of the UV Light Shutter ................................................................................

275

Figure 12-6:

Sensor Module Wiring and Pneumatic Fittings.........................................................................

280

Figure 12-7:

Sensor Module Mounting Screws .............................................................................................

281

Figure 12-8:

Sample Chamber Mounting Bracket.........................................................................................

282

Figure 12-9:

Hex Screw Between Lens Housing and Sample chamber .......................................................

283

Figure 12-10:

UV Lens Housing / Filter Housing.............................................................................................

284

Figure 12-11:

PMT UV Filter Housing Disassembled .....................................................................................

284

Figure 12-12:

Disassembling the Shutter Assembly .......................................................................................

286

Figure 12-13:

Shutter Assembly......................................................................................................................

287

Figure 12-14:

Location of UV Reference Detector Potentiometer ..................................................................

288

Figure 12-15:

PMT Assembly - Exploded View...............................................................................................

290

Figure 12-16:

Pre-Amplifier Board Layout.......................................................................................................

292

Figure 13-1:

Triboelectric Charging...............................................................................................................

295

Figure 13-2:

Basic Anti-ESD Work Station....................................................................................................

298

LIST OF TABLES

 

Table 3-1:

Front Panel Description ...............................................................................................................

23

Table 3-2:

Inlet / Outlet Connector Description.............................................................................................

24

Table 3-3:

Ventilation Clearance...................................................................................................................

28

Table 3-4:

Analog Output Pin-Outs...............................................................................................................

29

Table 3-5:

Status Output Signals ..................................................................................................................

30

Table 3-6:

Control Input Signals....................................................................................................................

31

Table 3-7:

NIST-SRM's Available for Traceability of SO2 Calibration Gases ...............................................

33

Table 3-8:

Front Panel Display during System Warm-Up.............................................................................

38

Table 3-9:

Possible Startup Warning Messages – M100E Analyzers w/o Options ......................................

39

Table 3-10:

Possible Startup Warning Messages – M100E Analyzers with Options .....................................

40

Table 5-1:

Optional Pumps ...........................................................................................................................

53

x

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual

Table of Contents

Table 5-2:

Optional Pumps ...........................................................................................................................

53

Table 5-3:

Optional Kits.................................................................................................................................

56

Table 5-4:

Zero/Span Valve Operating States ..............................................................................................

57

Table 5-5:

IZS Valve Operating States .........................................................................................................

58

Table 5-6:

IZS Permeation Tubes for SO2 ....................................................................................................

60

Table 5-7:

Modem Cable Options .................................................................................................................

60

Table 5-8:

CO2 Sensor Ranges ....................................................................................................................

65

Table 5-9:

CO2 Sensor Specifications...........................................................................................................

66

Table 5-10:

Extended Warranty Options.........................................................................................................

69

Table 6-1:

Analyzer Operating Modes ..........................................................................................................

74

Table 6-2:

Test Functions Defined................................................................................................................

75

Table 6-3:

List of Warning Messages............................................................................................................

77

Table 6-4:

Primary Setup Mode Features and Functions .............................................................................

80

Table 6-5:

Secondary Setup Mode Features and Functions ........................................................................

81

Table 6-6:

Password Levels..........................................................................................................................

81

Table 6-7:

Variable Names (VARS) ..............................................................................................................

93

Table 6-8:

M100E Diagnostic (DIAG) Functions...........................................................................................

95

Table 6-9:

DIAG - Analog I/O Functions .......................................................................................................

99

Table 6-10:

Analog Output Voltage Ranges ...................................................................................................

99

Table 6-11:

Analog Output Current Loop Range .........................................................................................

100

Table 6-12:

Analog Output Pin Assignments ...............................................................................................

100

Table 6-13:

Voltage Tolerances for Analog Output Calibration ...................................................................

105

Table 6-14:

Current Loop Output Calibration with Resistor .........................................................................

109

Table 6-15:

Test Parameters Available for Analog Output A4.....................................................................

117

Table 7-1:

Ethernet Status Indicators.........................................................................................................

124

Table 7-2:

LAN/Internet Configuration Properties......................................................................................

126

Table 7-3:

Internet Configuration Keypad Functions .................................................................................

130

Table 7-4:

COMM Port Communication modes .........................................................................................

135

Table 7-5:

Front Panel LED Status Indicators for iDAS.............................................................................

139

Table 7-6:

iDAS Data Channel Properties .................................................................................................

140

Table 7-7:

iDAS Data Parameter Functions...............................................................................................

141

Table 7-8:

Status Output Pin Assignments ................................................................................................

159

Table 7-9:

Control Input Pin Assignments .................................................................................................

159

Table 7-10:

Terminal Mode Software Commands .......................................................................................

161

Table 7-11:

Command Types.......................................................................................................................

162

Table 7-12:

Serial Interface Documents.......................................................................................................

167

Table 7-13:

RS-232 Communication Parameters for Hessen Protocol .......................................................

168

Table 7-14:

M100E Hessen Protocol Response Modes..............................................................................

170

Table 7-15:

Default Hessen Status Bit Assignments ...................................................................................

172

Table 8-1:

NIST-SRM's Available for Traceability of SO2 Calibration Gases ............................................

176

Table 8-2:

AutoCal Modes .........................................................................................................................

186

Table 8-3:

AutoCal Attribute Setup Parameters.........................................................................................

186

Table 8-4:

Example Auto-Cal Sequence....................................................................................................

187

Table 8-5:

Calibration Data Quality Evaluation ..........................................................................................

189

Table 9-1:

Activity Matrix for Calibration Equipment & Supplies................................................................

198

Table 9-2:

Activity Matrix for Calibration Procedure...................................................................................

198

Table 9-3:

Activity Matrix for Quality Assurance Checks ...........................................................................

200

Table 9-4:

Definition of Level 1 and Level 2 Zero and Span Checks.........................................................

201

Table 10-1:

M100E Preventive Maintenance Schedule...............................................................................

209

Table 10-2:

Predictive Uses for Test Functions ...........................................................................................

211

Table 11-1:

Relay Board Status LED’s ........................................................................................................

243

Table 11-2:

Front Panel Status LED’s..........................................................................................................

249

Table 12-1:

Warning Messages - Indicated Failures ...................................................................................

257

Table 12-2:

Test Functions - Possible Causes for Out-Of-Range Values ...................................................

259

Table 12-3:

Relay Board Status LEDs .........................................................................................................

262

Table 12-4:

DC Power Test Point and Wiring Color Code...........................................................................

270

Table 12-5:

DC Power Supply Acceptable Levels .......................................................................................

271

Table 12-6:

Relay Board Control Devices....................................................................................................

272

04515F DCN6048

xi

Table of Contents

Teledyne API M100E Analyzer Operation Manual

Table 12-7:

Analog Output Test Function - Nominal Values .......................................................................

273

Table 12-8:

Status Outputs Check Pin Out..................................................................................................

273

Table 12-9:

Example of HVPS Power Supply Outputs ................................................................................

277

Table 12-10:

Example of HVPS Power Supply Outputs ................................................................................

288

Table 13-1:

Static Generation Voltages for Typical Activities ......................................................................

296

Table 13-2:

Sensitivity of Electronic Devices to Damage by ESD ...............................................................

296

xii

04515F DCN6048

SECTION I

GENERAL INFORMATION

04515F DCN6048

13

This page intentionally left blank.

14

04515F DCN6048

1. INTRODUCTION

1.1. SAFETY MESSAGES

Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these messages carefully.

A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:

General Safety Hazard: Refer to the instructions for details on the specific hazard.

CAUTION: Hot Surface Hazard.

CAUTION: Electrical Shock Hazard.

TECHNICIAN SYMBOL: All operations marked with this symbol are to be performed by qualified maintenance personnel only.

CAUTION

General Safety Hazard

The M100E Analyzer should only be used for the purpose and in the manner described in this manual. If you use the M100E in a manner other than that for which it was intended, unpredictable behavior could ensue with possible hazardous consequences.

Technical Assistance regarding the use and maintenance of the M100E or any other Teledyne API product

can be obtained by contacting Teledyne API’s Customer Service Department by: Phone at 800-324-5190

Sending an email to api-customerservice@teledyne.com or

Accessing various service options on our website at http://www.teledyne-api.com/.

04515F DCN6048

15

Introduction

Teledyne API M100E Analyzer Operation Manual

1.2. M100E OVERVIEW

The Model 100E (also referred to as M100E) UV Fluorescence SO2 Analyzer is a microprocessor controlled analyzer that determines the concentration of sulfur dioxide (SO2), in a sample gas drawn through the instrument’s sample chamber where it is exposed to ultraviolet light causing any SO2 present to fluoresce. The instrument measures the amount of fluorescence to determine the amount of SO2 present in the sample gas.

The M100E’s exceptional stability is achieved with the use of an optical shutter to compensate for sensor drift and a reference detector to correct for changes in UV lamp intensity. Additionally, an advanced optical design combined with a special scrubber, called a "kicker" that removes hydrocarbons (which fluoresces similarly to SO2) prevents inaccuracies due to interferents.

Calibration of the instrument is performed in software that stores SO2 concentration measurements made when specific, known concentrations of SO2 are supplied to the analyzer. The microprocessor uses these calibration values along with other performance parameters such as the sensor offset, UV lamp intensity, the amount of stray light present, and measurements of the temperature and pressure of the sample gas to compute the final SO2 concentration.

Built-in data acquisition capability, using the analyzer's internal memory, allows the logging of multiple parameters including averaged or instantaneous concentration values, calibration data, and operating parameters such as pressure and flow rate. Stored data are easily retrieved through the serial port or optional Ethernet port via our APICOM software or from the front panel, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends. Multiple averaging periods of one minute to 365 days are available for over a period of one year.

Some of the features of your M100E UV Fluorescence Sulfur Dioxide Analyzer are:

Standard two year warranty

Ranges, 0-50 ppb to 0-20,000 ppb, user selectable

Dual ranges and auto ranging

Microprocessor controlled for versatility

Multi-tasking software to allow viewing test variables while operating

Continuous self checking with alarms

Dual bi-directional RS-232 ports for remote operation (optional RS-485 or Ethernet)

Digital status outputs indicate instrument operating condition

Adaptive signal filtering tp optimize response time

Temperature & Pressure compensation

Internal Zero & Span check (optional)

Internal data logging with 1 min to 365 day multiple averages

Critical flow orifices to provide flow stability

1.3. M100E DOCUMENTATION

The documentation (part number 04515) for this instrument is available in several different formats:

Printed format, or;

Electronic format on a CD-ROM.

The electronic manual is in Adobe® Systems Inc. “Portable Document Format”. The Adobe® Acrobat Reader® software, which is necessary to view these files, can be downloaded for free from the internet at http://www.adobe.com/.

16

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual

Introduction

 

 

Additional documentation for the M100E SO2 Analyzer is available from Teledyne API’s website at http://www.teledyne-api.com/manuals/.

APICOM software manual, part number 03945.

Multidrop manual, part number 01842.

iDAS Manual, part number 02837.

1.3.1. USING THIS MANUAL

NOTE

Throughout this manual, words printed in capital, bold letters, such as SETUP or ENTR represent messages as they appear on the analyzer’s display.

This manual has the following structure:

TABLE OF CONTENTS

Outlines the contents of the manual in the order the information are presented. This is a good overview of the topics covered in the manual. There is also a list of appendices, figures and tables.

SECTION I – GENERAL INFORMATION

INTRODUCTION

A brief description of the M100E analyzer architecture as well as a description of the layout of the manual and what information is located in its various sections and chapters.

SPECIFICATIONS AND WARRANTY

A list of the analyzer’s performance specifications and if applicable a description of the conditions and configuration under which EPA equivalency was approved as well as the Teledyne API’s warranty statement.

GETTING STARTED

Instructions for setting up, installing and running your analyzer for the first time.

GLOSSARY

Answers to the most frequently asked questions about operating the analyzer and a glossary of acronyms and technical terms.

OPTIONAL HARDWARE & SOFTWARE

A description of optional equipment to add functionality to your analyzer.

04515F DCN6048

17

Introduction

Teledyne API M100E Analyzer Operation Manual

SECTION II – OPERATING INSTRUCTIONS

BASIC OPERATION OF THE M100E ANALYZER

Step-by-Step instructions for using the display/keyboard to set up and operate the M100E analyzer.

ADVANCED FEATURES OF THE M100E ANALYZER

Step-by-Step instructions for using the M100E analyzer’s more advanced features such as the iDAS system, the DIAG and VARS menus and the and the TEST channel analog output.

REMOTE OPERATION OF THE M100E Analyzer

Information and instructions for interacting with the M100E analyzer via its several remote interface options (e.g. via RS-232, Ethernet, its built in digital control inputs/outputs, etc.)

M100E VALIDATION AND VERIFICATION

Methods and procedures for verifying the correct operation of your M100E Analyzer as well as step by step instructions for calibrating it

EPA PROTOCOL CALIBRATION

Specific information regarding calibration requirements for analyzers used in EPA monitoring.

SECTION III – TECHNICAL INFORMATION

THEORY OF OPERATION

An in-depth look at the various principals by which your analyzer operates as well as a description of how the various electronic, mechanical and pneumatic components of the analyzer work and interact with each other. A close reading of this section is invaluable for understanding the analyzer’s operation.

MAINTENANCE SCHEDULE AND PROCEDURES

Description of preventative maintenance procedures that should be regularly performed on you analyzer to assure good operating condition.

GENERAL TROUBLESHOOTING & REPAIR OF THE M100E ANALYZER

This section includes pointers and instructions for diagnosing problems with the analyzer in general and the Terminus as well as instructions on performing repairs of on the Terminus.

A PRIMER ON ELECTRO-STATIC DISCHARGE

This section describes how static electricity occurs; why it is a significant concern and; how to avoid it and avoid allowing ESD to affect the reliable and accurate operation of your analyzer.

APPENDICES

For easier access and better updating, some information has been separated out of the manual and placed in a series of appendices at the end of this manual. These include version-specific software menu trees, warning messages, definitions Modbus registers and serial I/O variables as well as spare part listings, repair questionnaires, interconnect drawing, detailed pneumatic and electronic schematics.

NOTE

The flowcharts in this manual contain typical representations of the analyzer’s display during the various operations being described. These representations are not intended to be exact and may differ slightly from the actual display of your instrument.

18

04515F DCN6048

2. SPECIFICATIONS, APPROVALS AND WARRANTY

2.1. SPECIFICATIONS

 

Table 2-1: M100E Basic Unit Specifications

 

 

 

Min/Max Range

 

In 1ppb increments from 50 ppb to 20,000 ppb of Full Scale, dual ranges or auto ranging

(Physical Analog Output)

 

 

Measurement Units

 

ppb, ppm, µg/m3, mg/m3 (user selectable)

Zero Noise1

 

0.2 ppb RMS

Span Noise1

 

0.5% of reading, above 50 ppb

Lower Detectable Limit2

 

0.4 ppb

Zero Drift (24 hours)

 

0.5 ppb

Zero Drift (7 days)

 

1 ppb

Span Drift (24 hours)

 

0.5% of Full Scale

Span Drift (7 days)

 

1% of Full Scale

Linearity

 

1% of Full Scale

Precision

 

0.5% of reading1 above 50 ppb

Lag Time1

 

20 seconds

Rise/Fall Time1

 

95% in <100 sec

Sample Flow Rate

 

650 cm3/min. ±10%

Temperature Range

 

5-40 oC

Humidity Range

 

0 - 95% RH, non-condensing

Dimensions H x W x D

 

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Weight, Analyzer

 

41.45 lbs (18.80 kg) with internal pump

(Basic Configuration)

 

36.45 lbs (16.53 kg) without internal pump

AC Power Rating

 

100 V, 50/60 Hz, 2.50 A

 

 

115 V, 50/60 Hz, 2.08 A

 

 

220 V, 50/60 Hz, 1.14 A

 

 

240 V, 50/60 Hz, 1.04 A

Environmental

 

Installation category (over-voltage category) II; Pollution degree 2

Recorder Offset

 

± 10 %

Analog Outputs (Optional

 

Select up to 3 channels

4-20 mA)

 

 

Analog Output Ranges

 

10 V, 5 V, 1 V, 0.1 V selectable

Analog Output Resolution

 

1 part in 4096 of selected full-scale voltage

Status Outputs

 

8 Status outputs from opto-isolators

 

 

 

Control Inputs

 

6 Control Inputs, 3 defined, 3 spare

Serial I/O

 

One (1) RS-232; One (1) RS-485 (2 connecters in parallel)

 

 

Baud Rate : 300 – 115200: Optional Ethernet Interface

Certifications

 

EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A,

 

 

ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.

 

 

IEC 61010-1:90 + A1:92 + A2:95,

Approvals

 

USEPA EQSA-0495-100

 

 

MCERTS certified SIRA MC050067/00

 

 

EN14212, Approved, CE and Others.

1 As defined by the USEPA.

2 Defined as twice the zero noise level by the USEPA.

04515F DCN6048

19

Specifications, Approvals and Warranty

Teledyne API M100E Analyzer Operation Manual

2.2. EPA EQUIVALENCY DESIGNATION

The M100E Analyzer is designated as Reference Method Number EQSA-0495-100 as per 40 CFR Part 53 when operated under the following conditions:

Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).

Ambient temperature range of 5 oC to 40 oC.

Line voltage range of 100-120 VAC or 220-240 VAC, at 50 or 60 Hz.

Sample filter: Equipped with PTFE filter element in the internal filter assembly.

Sample flow of 650 +/- 65 cm3/min.

Vacuum pump (internal) capable of 14"Hg Absolute pressure @ 1 slpm or better.

Software settings:

Dynamic span

OFF

 

 

Dynamic zero

OFF

 

 

Dilution factor

OFF

AutoCal

ON or OFF

Dual range

ON or OFF

Auto-range

ON or OFF

 

 

Temp/Pressure compensation

ON

 

 

Under the designation, the analyzer may be operated with or without the following optional equipment:

Rack mount with chassis slides.

Rack mount without slides, ears only.

Zero/span valve options.

Internal zero/span (IZS) option with either:

SO2 permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.

SO2 permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified. Under the designation, the IZS option cannot be used as the source of calibration.

4-20mA isolated analog outputs.

Status outputs.

Control inputs.

RS-232 output.

Ethernet output.

Zero air scrubber.

4-20mA, isolated output.

20

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual

Specifications, Approvals and Warranty

 

 

2.3.CE MARK COMPLIANCE

2.3.1.EMISSIONS COMPLIANCE

The Teledyne-Advanced Pollution Instrumentation UV Fluorescence SO2 Analyzer M100E was tested and found to be fully compliant with:

EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.

Tested on 21 February 2003 - 08 March 2003 at CKC Laboratories, Inc., Report Number CE03-021A.

2.3.2. SAFETY COMPLIANCE

The Teledyne-Advanced Pollution Instrumentation UV Fluorescence SO2 Analyzer M100E was tested and found to be fully compliant with:

IEC 61010-1:90 + A1:92 + A2:95,

Issued by CKC Laboratories on 4 April 2003, Report Number WO-80146.

2.4. WARRANTY

WARRANTY POLICY (02024D)

Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should equipment failure occur, Teledyne API assures its customers that prompt service and support will be available.

COVERAGE

After the warranty period and throughout the equipment lifetime, Teledyne API stands ready to provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first level of field troubleshooting are to be performed by the customer.

NON-API MANUFACTURED EQUIPMENT

Equipment provided but not manufactured by Teledyne API is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturer’s warranty.

GENERAL

During the warranty period, Teledyne API warrants each Product manufactured by Teledyne API to be free from defects in material and workmanship under normal use and service. Expendable parts are excluded.

If a Product fails to conform to its specifications within the warranty period, Teledyne API shall correct such defect by, at Teledyne API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product.

The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other than in accordance with the instruction provided by Teledyne API, or (iii) not properly maintained.

04515F DCN6048

21

Specifications, Approvals and Warranty

Teledyne API M100E Analyzer Operation Manual

THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED HEREIN. API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF Teledyne API'S PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE

TERMS AND CONDITIONS

All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.

22

04515F DCN6048

3. GETTING STARTED

3.1. M100E ANALYZER LAYOUT

Figure 3-1: Front Panel Layout

Table 3-1: Front Panel Description

Field

Mode

Message

Concentration

Keypad

Definition

STATUS LED’s

Significance

Displays the name of the analyzer’s current operating mode

Displays a variety of informational messages such as warning messages, operational data, test function values and response messages during interactive tasks.

Displays the actual concentration of the sample gas currently being measured by the analyzer in the currently selected units of measure

Displays dynamic, context sensitive definitions for the row of keys just below the display.

 

Name

Color

 

State

 

Definition

 

 

 

 

Off

 

Unit is not operating in sample mode, iDAS is disabled.

 

SAMPLE

Green

 

On

 

Sample Mode active; Front Panel Display being updated; iDAS data being stored.

 

 

Blinking

 

Unit is operating in sample mode, front panel display being updated, iDAS hold-off mode

 

 

 

 

 

 

 

 

 

 

 

is ON, iDAS disabled

 

 

 

 

Off

 

Auto Cal disabled

 

CAL

Yellow

 

On

 

Auto Cal enabled

 

 

 

 

Blinking

 

Unit is in calibration mode

 

FAULT

Red

 

Off

 

No warnings exist

 

 

Blinking

 

Warnings exist

 

 

 

 

 

 

 

 

 

 

 

 

04515F DCN6048

23

Getting Started

Teledyne API M100E Analyzer Operation Manual

Figure 3-2: Rear Panel Layout

 

Table 3-2: Inlet / Outlet Connector Description

 

 

rear panel Label

Function

 

 

SAMPLE

Connect a gas line from the source of sample gas here.

Calibration gasses are also inlet here on units without zero/span/shutoff valve options

 

installed.

EXHAUST

Connect an exhaust gas line of not more than 10 meters long here that leads outside

the shelter or immediate area surrounding the instrument.

SPAN

On units with zero/span/shutoff valve options installed, connect a gas line to the source

of calibrated span gas here.

ZERO AIR

Internal Zero Air: On units with zero/span/shutoff valve options installed but no internal

zero air scrubber attach a gas line to the source of zero air here.

24

04515F DCN6048

Teledyne 100E User Manual 2

Teledyne API M100E Analyzer Operation Manual

Getting Started

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-3: M100E Layout (Basic Unit – No Valve Options)

04515F DCN6048

25

Getting Started

Teledyne API M100E Analyzer Operation Manual

(Kicker)

Figure 3-4: M100E Internal Gas Flow (Basic Configuration)

NOTE

For pneumatic diagrams of M100E SO2 Analyzer’s with various calibration valve options refer to Section 5.7.

26

04515F DCN6048

Teledyne API M100E Analyzer Operation Manual

Getting Started

 

 

3.2. UNPACKING THE M100E ANALYZER

CAUTION

GENERAL SAFETY HAZARD

To avoid personal injury, always use two persons to lift and carry the M100E.

1.Verify that there is no apparent external shipping damage. If damage has occurred, please advise the shipper first, then Teledyne API.

2.Included with your analyzer is a printed record of the final performance characterization performed on your instrument at the factory. It is titled Final Test and Validation Data Sheet (P/N 04551). This record is an important quality assurance and calibration record for this instrument. It should be placed in the quality records file for this instrument.

3.Carefully remove the top cover of the analyzer and check for internal shipping damage by:

Removing the locking screw located in the top, center of the Front panel;

Removing the two flat head, Phillips screws on the sides of the instrument;

Sliding the cover backwards until it clears the analyzer’s front bezel, and;

Lifting the cover straight up.

4.Inspect the interior of the instrument to ensure that all circuit boards and other components are in good shape and properly seated.

5.Check the connectors of the various internal wiring harnesses and pneumatic hoses to ensure that they are firmly and properly seated.

6.Verify that all of the optional hardware ordered with the unit has been installed. These are listed on the paperwork accompanying the analyzer.

NOTE

Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges too small to be felt by the human nervous system. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty.

Refer to Chapter 13 for more information on preventing ESD damage.

CAUTION

ELECTRICAL SHOCK HAZARD

Never disconnect PCAs, wiring harnesses or electronic subassemblies while under power.

04515F DCN6048

27

Getting Started

Teledyne API M100E Analyzer Operation Manual

3.2.1. VENTILATION CLEARANCE

Whether the analyzer is set up on a bench or installed into an instrument rack, be sure to leave sufficient ventilation clearance.

Table 3-3: Ventilation Clearance

AREA

MINIMUM REQUIRED

CLEARANCE

 

Back of the instrument

4 in.

Sides of the instrument

1 in.

Above and below the

1 in.

instrument

 

Various rack mount kits are available for this analyzer. Refer to Section 5.2 of this manual for more information.

3.3.ELECTRICAL CONNECTIONS

3.3.1.POWER CONNECTION

Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 Amps of current at your AC voltage and that it is equipped with a functioning earth ground.

CAUTION

ELECTRICAL SHOCK HAZARD

High Voltages are present inside the analyzers case. Power connection must have functioning ground connection.

Do not defeat the ground wire on power plug.

Turn off analyzer power before disconnecting or connecting electrical subassemblies.

Do not operate instrument with the cover off.

CAUTION

GENERAL SAFETY HAZARD

The M100E analyzer can be configured for both 100-120 V and 220-240 V at either 50 or 60 Hz.

To avoid damage to your analyzer, ensure that the AC power voltage matches the voltage indicated on the Analyzer’s serial number label tag (refer to Figure 3-2) before plugging the M100E into line power.

28

04515F DCN6048

Loading...
+ 358 hidden pages