Sirona Orthophos XG Service manual

mäìë
T
R
Prog.
S
kV
mA
P1
loqelmelp=ud
=ap=L=`ÉéÜ loqelmelp=ud=R=ap=L=`ÉéÜ loqelmelp=ud=P=ap
pЙкобЕЙ=j~ем~д
bеЦдблЬ
Easypad
ORTHOPHOS XG 5 / 3 / 3 PPE DS
Multipad
sЙклбзе=NPKM
Plus
DS
About the Service Manual:
This Service Manual is valid for the following unit versions:
Plus
– ORTHOPHOS XG – ORTHOPHOS XG 5 DS / Ceph – ORTHOPHOS XG 3 DS*
i
NOTE
These two system versions differ regarding the equipment offered by their user interfaces and their possible configuration levels (see page 1-5).
i
NOTE
Board DX41 is no longer installed in units with hardware version BA or higher (as of November 2006).
In addition, you also require:
Spare parts list: Order No. 59 38 423
– ORTHOPHOS XG
ORTHOPHOS XG 3 DS
DS / Ceph
Plus
DS / Ceph, ORTHOPHOS XG 5 DS / Ceph,
Wiring diagrams: Order No. 59 38 332
Plus
– ORTHOPHOS XG
ORTHOPHOS XG 3 DS
Installation Instructions
– ORTHOPHOS XG – ORTHOPHOS XG 5 DS / Ceph: Order No. 60 04 902 – ORTHOPHOS XG 3 DS: Order No. 60 51 416
Tools
– Screwdriver, medium sized – Torx offset screwdrivers TX10, TX20, TX25 – Open-end wrench, 13 mm A/F – Socket wrench, 13 mm A/F – Side cutters – Spirit level
Auxiliary devices
– Digital multimeter, Accuracy Class 1 – Soldering tool for repairing cables – Cable ties – Teflon tape – Loctite
DS / Ceph, ORTHOPHOS XG 5 DS / Ceph,
Plus
DS / Ceph: Order No. 59 87 651
General information/Software update 1
Messages
Troubleshooting
Adjustment
2
3
4
Service routines
Repair
Maintenance
5
6
7
Contents
1 General information ..................................................................... 1-5
1.1 Safety .................................................................................. 1-5
1.2 XG
1.3 Programs and functions of the
1.4 Programs and functions of the XG
1.5 Operation notes ................................................................. 1-11
1.6 Demo mode – Operation without radiation release ........... 1-13
1.7 Demo mode – Repacking and transport ........................... 1-15
1.8 List of software versions .................................................... 1-16
1.9 Software update ................................................................ 1-22
1.10 Selecting More details ....................................................... 1-32
1.11 The most important modules and components ................. 1-34
1.12 Cabling overview ............................................................... 1-38
1.13 Illustrations of boards ........................................................ 1-42
1.14 Removing the covers ........................................................ 1-50
Plus
and XG 5 / 3 / 3 PPE system classes ...................... 1-6
XG 5 / 3 / 3 PPE system class ............................................ 1-8
Plus
system class ........... 1-9
2 Messages .................................................................................... 2-3
2.1 Help messages ................................................................... 2-4
2.2 System messages ............................................................... 2-6
2.3 Status displays .................................................................... 2-6
2.4 Error messages ................................................................... 2-6
2.5 List of error messages ......................................................... 2-9
2.6 List of available service routines ....................................... 2-62
3 Troubleshooting .......................................................................... 3-3
3.1 Error logging memory .......................................................... 3-4
3.2 Check the CAN bus ............................................................. 3-6
3.3 Checking the boards ......................................................... 3-12
3.4 Checking the motors ......................................................... 3-14
3.5 Checking the light barriers ................................................ 3-15
3.6 Device leakage current too high ........................................ 3-16
3.7 Checking the cables .......................................................... 3-17
3.8 Error analysis of X-RAY control signal path: up to unit hardware revision AG
(with board DX41) ............................................................. 3-18
3.9 Error analysis of X-RAY control signal path: unit HW version BA or higher
(without board DX41) ........................................................ 3-22
3.10 Check data paths/Generate test images ........................... 3-25
59 38 399 D3352 D3352.076.01.13.02 07.2008
V
Contents
4 Adjustment ................................................................................... 4-3
4.1 Important information concerning adjustment ...................... 4-3
4.2 Diaphragm/system adjustment menu .................................. 4-5
4.3 Adjusting the panoramic X-ray unit ...................................... 4-9
4.4 Adjusting the cephalometer
4.5 TSA sensor adjustment
4.6 Resetting the adjustment
(XG
(XG
(XG
Plus
/ XG 5, if ceph is installed) ..................................... 4-56
Plus
full version only) .................................................. 4-107
Plus
/ 5 / 3 / 3 PPE) ..................................................... 4-115
5 Service routines ........................................................................... 5-9
5.1 Selecting the Service menu ............................................... 5-10
5.2 Selecting a service routine ................................................. 5-15
5.3 Service routines with SIDEXIS .......................................... 5-18
5.4 Service routine S001 ......................................................... 5-21
5.5 Service routine S002 ......................................................... 5-23
5.6 Service routine S005 ......................................................... 5-28
5.7 Service routine S007 ......................................................... 5-40
5.8 Service routine S008 ......................................................... 5-48
5.9 Service routine S009 ......................................................... 5-56
5.10 Service routine S012 ......................................................... 5-59
5.11 Service routine S014 ......................................................... 5-67
5.12 Service routine S015 ......................................................... 5-74
5.13 Service routine S017 ......................................................... 5-77
5.14 Service routine S018 ....................................................... 5-109
Plus
5.15 Service routine S020 (XG
5.16 Service routine S021 (not for XG 3 / 3 PPE) ................... 5-121
5.17 Service routine S032 ....................................................... 5-126
5.18 Service routine S033 (not for XG 3 / 3 PPE) ................... 5-130
5.19 Service routine S034 ....................................................... 5-134
5.20 Service routine S037 ....................................................... 5-151
only) .................................. 5-118
VI D3352.076.01.13.02 07.2008
59 38 399 D3352
Contents
6 Repair .......................................................................................... 6-5
6.1 Replacing the height adjustment motor (M1_4)/spindle ...... 6-6
6.2 Replacing the ring motor (M1_3) ....................................... 6-14
6.3 Replacing the PAN actuators (M1_1/2) ............................. 6-17
6.4 Replacing the headrest ..................................................... 6-19
6.5 Replacing the Easypad (XG
or the Multipad (XG 5 / 3 / 3 PPE) ..................................... 6-21
6.6 Replacing the control panel ............................................... 6-24
6.7 Replacing/adjusting the FH light localizer (PAN) .............. 6-26
6.8 Replacing/adjusting the MS light localizer
laser module (PAN) ........................................................... 6-28
6.9 Replacing/adjusting the FH light localizer (Ceph) ............. 6-30
6.10 Removing the light localizers from the ring (for the XG
Plus
and XG 5, if present) ................................. 6-32
6.11 Replacing the support piece (bite block holder) ................ 6-33
6.12 Replacing the motor-driven diaphragm (for XG
Plus
/ XG 5) ............................................................. 6-34
6.13 Replacing the fixed diaphragm
(only XG 3 / 3 PPE) ........................................................... 6-37
6.14 Replacing the X-ray tube assembly .................................. 6-38
6.15 Replacing the fan (tube assembly) .................................... 6-42
6.16 Replacing the PAN (TSA) sensor holder ........................... 6-43
6.17 Replacing the ceph sensor holder ..................................... 6-47
6.18 Replacing the sensor ........................................................ 6-48
6.19 Replacing the light barriers ............................................... 6-49
6.20 Replacing circuit boards .................................................... 6-59
6.21 Replacing cables ............................................................... 6-91
Plus
)
59 38 399 D3352 D3352.076.01.13.02 07.2008
VII
Contents
7 Maintenance ................................................................................ 7-3
7.1 Checking the height adjustment .......................................... 7-4
7.2 Checking the forehead and temple supports ....................... 7-6
7.3 Checking the sensor holder (pan and ceph) ........................ 7-7
7.4 Checking the support piece (bite block holder) .................... 7-9
7.5 Checking the light localizers .............................................. 7-10
7.6 Checking the X-ray images ................................................ 7-12
7.7 Checking the tube data ...................................................... 7-13
7.8 Checking the diaphragm .................................................... 7-17
7.9 Checking the cables for damage ....................................... 7-35
7.10 Checking the idling rollers .................................................. 7-36
7.11 Checking the grounding straps .......................................... 7-37
7.12 Checking the cable shields ................................................ 7-38
7.13 Checking the protective ground wires ................................ 7-39
7.14 Checking the device leakage current ................................. 7-43
VIII D3352.076.01.13.02 07.2008
59 38 399 D3352
ORTHOPHOS XG
1General information
Tab 1
Contents
1.1 Safety ......................................................................1 – 5
1.2 XG
Plus
and XG 5 / 3 / 3 PPE system classes..........1 – 6
1.2.1 Overview of the ORTHOPHOS XG system
classes and versions ............................................ 1 – 7
1.3 Programs and functions of the
XG 5 / 3 / 3 PPE system class ................................1 – 8
1.3.1 XG 5 system version ............................................ 1 – 8
1.3.2 XG 3 system version ............................................ 1 – 8
1.3.3 XG 3 PPE (pay per exposure) system version..... 1 – 8
1.4 Programs and functions of the XG
1.4.1 Programs and functions of the basic version ....... 1 – 9
1.4.2 Programs and functions of the full version ......... 1 – 10
Plus
system class1 – 9
1.5 Operation notes.....................................................1 – 11
1.6 Demo mode – Operation without radiation release1 – 13
1.6.1 Switching the demo mode ON............................ 1 – 14
1.6.2 Switching the demo mode OFF.......................... 1 – 14
1.7 Demo mode – Repacking and transport ...............1 – 15
1.8 List of software versions........................................1 – 16
1.8.1 ORTHOPHOS XG 3 PPE .................................. 1 – 16
1.8.2 ORTHOPHOS XG 3 .......................................... 1 – 17
1.8.3 ORTHOPHOS XG 5 / Ceph ............................... 1 – 18
1.8.4 ORTHOPHOS XG
Plus
/ Ceph............................. 1 – 20
1.9 Software update ....................................................1 – 22
1.9.1 Important information on the software update.... 1 – 22
1.9.2 Permissible software updates ............................ 1 – 23
1.9.3 Performing a software update ............................ 1 – 24
1.9.4 Software update for ORTHOPHOS XG 5
to V02.30 or higher ............................................. 1 – 29
1.9.5 Software Update for ORTHOPHOS XG
from V02.20 to V02.30 or higher ........................ 1 – 30
1.9.6 Software update for ORTHOPHOS XG
from V02.22 to V02.30 or higher ........................ 1 – 31
Plus
Plus
1.10 Selecting More details...........................................1 – 32
1 – 2 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1
1.11 The most important modules and components .....1 – 34
1.11.1 Slide ................................................................... 1 – 35
1.11.2 Stand .................................................................. 1 – 36
1.11.3 Cephalometer
(arm mounted on left side or on right side)......... 1 – 37
1.11.4 Remote control ................................................... 1 – 37
1.12 Cabling overview...................................................1 – 38
1.13 Illustrations of boards ............................................1 – 42
1.13.1 Boards in the slide.............................................. 1 – 42
1.13.2 Boards in the stand ............................................ 1 – 46
1.13.3 Boards in the cephalometer
(arm mounted on right side or on left side)......... 1 – 47
1.13.4 Boards in the remote control .............................. 1 – 48
1.14 Removing the covers.............................................1 – 50
59 38 399 D3352 D3352.076.01.13.02 07.2008
1 – 3
Tab 1
1 – 4 D3352.076.01.13.02 07.2008
59 38 399 D3352

Tab 1 1.1 Safety

1 General information

1.1 Safety
It is essential that you comply with the warning and safety information con­tained in this Service Manual.
All such information is highlighted by one of three signal words, i.e. CAUTION, WARNING or DANGER.
CAUTION
Nonobservance may result in minor physical injuries or material damage and malfunctions.
WARNING
Nonobservance may lead to serious physical injury or death.
DANGER
Immediate danger to life and limb. Threat of serious physical injury or death.
1.1
1.1
bеЦдблЬ
59 38 399 D3352 D3352.076.01.13.02 07.2008
1 – 5
1.2 XG
T
R
Prog.
S
kV
mA
Plus
and XG 5 / 3 / 3 PPE system classes Tab 1
Plus
1.2
XG
and XG 5 / 3 / 3 PPE system classes
Comfort class with Easypad (XG
Basic class with Multipad (XG 5 / 3 / 3 PPE)
Plus
The XG
)
Plus
and XG 5 / 3 / 3 PPE system versions differ with regard to their
user interface. While the ORTHOPHOS XG (XG with a color touchscreen (Easypad), the (XG 5 / 3 / 3 PPE) basic class is equipped with a simpler control panel with a single-line display (Multipad).
Due to their different control panels, the operating procedures for these two system classes also vary.
All ORTHOPHOS XG units are digital systems (DS).
The ORTHOPHOS XG tion with a cephalometer (ceph). The cephalometer for the XG sion is also available with the ceph arm mounted either on the left or the right side. Only the cephalometer version with the arm mounted on the left side is offered for the XG 5 system version. Operation of the XG 3/3 PPE system ver­sion is not possible with a cephalometer.
i
NOTE
The version with the ceph arm mounted on the left side is basically described and illustrated in this Service Manual. Unless described separately, all dis­mantling and assembly steps must be performed laterally reversed for the ver­sion with the ceph arm mounted on the right side.
Remote control operation is possible in all system versions.
The (XG however, the basic system class (XG 5 / 3 / 3 PPE) is available only in a basic version.
The XG
This Service Manual describes both system classes in the highest possible configuration.
Plus
) system class comes in a basic and a full version (see page 1-9),
Plus
system class optionally includes TSA operation.
Plus
and XG 5 system versions are suitable for opera-
Plus
) features a control panel
Plus
system ver-
For more information on the different installation versions, please refer to the Installation Instructions.
i
NOTE
The empty digits in the single-line display of the Multipad are marked with underscores in this Manual. They have been added to enhance clarity and are not present on the real Multipad.
S005_______1____
n
(
123
1 – 6 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1 1.2 XG
T
R
Prog.
S
kV
mA
T
R
Prog.
S
kV
mA
T
R
Prog.
S
kV
mA
T
R
Prog.
S
kV
mA
1.2.1 Overview of the ORTHOPHOS XG system
Comfort class (Easypad, basic or full version, TSA option, remote control option)
Plus
and XG 5 / 3 / 3 PPE system classes
classes and versions
1.2
1.2
bеЦдблЬ
ORTHOPHOS XG
Plus
with ceph arm on right sidewith ceph arm on left side
Basic class (Multipad, basic equipment, remote release optional)
ORTHOPHOS XG 5with ceph arm on left side
ORTHOPHOS XG 3
ORTHOPHOS XG 3 PPE
59 38 399 D3352 D3352.076.01.13.02 07.2008
1 – 7

1.3 Programs and functions of the XG 5 / 3 / 3 PPE system class Tab 1

1.3
Programs and functions of the XG 5 / 3 / 3 PPE system class
1.3.1 XG 5 system version
8 panoramic programs z P 1, P 1C, P 10
z P 1L, P 1R
z TM 1.1/TM 1.2
z S 1
z MS 1
5 Ceph programs z C 1, C 2, C 3, C 3F, C 4
Functions z Half-views of right/left side possible only with P1
(not upper/lower jaw).
z Constant magnification factor of 1.25 possible only with program P 1.
z No Quickshot function.
z No "jaw shape" program presetting.
z No artifact-free display mode.
z No orthodontic image series P 1 - C 3 - C 4 without cooling periods.
z No TSA configuration level.
z No TSA operation possible.
z No welcome screen.
1.3.2 XG 3 system version
5 panoramic programs z P 1, P 1C
z P 1L, P 1R
z TM 1.1/TM 1.2
Functions z Half-views of right/left side possible only with P1
(not upper/lower jaw).
z Constant magnification factor of 1.25 possible only with program P 1.
z No Quickshot function.
z No "jaw shape" program presetting.
z No artifact-free display mode.
z No orthodontic image series P 1 - C 3 - C 4 without cooling periods.
z No TSA configuration level.
z No TSA operation possible.
z No welcome screen.
z
1.3.3 XG 3 PPE (pay per exposure) system version
ORTHOPHOS XG 3 with PPE concept (for USA only).
Functionality and programs identical to those of system version XG 3.
1 – 8 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1 1.4 Programs and functions of the XG
Plus
system class
1.4 Programs and functions of the XG
class
The XG
z Basic version
z Full version
This Service Manual describes the full version of the system.
1.4.1 Programs and functions of the basic version
10 panoramic programs z P 1, P 1C, P 2, P 10
z P 1L, P 1R
z P 12
z TM 1.1/TM 1.2
z S 1
z MS 1
5 Ceph programs z C 1, C 2, C 3, C 3F, C 4
Functions z Half-views of right/left side possible only with P1
z Preselection of individual quadrants not possible.
z Constant magnification factor of 1.25 possible only with program P 1.
z No Quickshot function.
z No "jaw shape" program presetting.
z No artifact-free display mode.
z No orthodontic image series P 1 - C 3 - C 4 without cooling periods.
z No TSA configuration level (TSA operation retrofittable).
z No welcome screen with patient data from SIDEXIS.
z Ceph: No shadowing in upper head region with programs C3 and C3F.
z Ceph: No shadowing in thyroid region with programs C1 and C2.
Plus
system class is available in two configuration levels:
(not upper/lower jaw).
Plus
system
1.4
1.4
bеЦдблЬ
59 38 399 D3352 D3352.076.01.13.02 07.2008
1 – 9
1.4 Programs and functions of the XG
Plus
system class Tab 1
1.4.2 Programs and functions of the full version
The full version offers the following programs and functions in addition to those featured in the basic version:
Additional Pan programs z Constant magnification factor of 1.25 P 2C, P 10C
z Artifact-free display P 1A, P 2A, P 10A
Additional functions z Selectable right/left partial view and individual quadrants
P 1A, P 2/A/C, P 10/A/C, P12.
z Selectable upper/lower jaw partial view and individual quadrants
P 1/A/C, P 2/A/C, P 10/A/C.
z Temporomandibular joint views TM 2, TM 3, TM 4, TM 5, TM 6
z Sinus views S2, S3, S4.
z Selectable "jaw shape" program setting.
z Quickshot function possible (for Pan and Ceph).
z Orthodontic image series P 1 - C 3 - C 4 without cooling periods possible.
z TSA configuration level
z TSA operation possible (optional)
z Welcome screen with patient data from SIDEXIS
(SIDEXIS V01.50 and higher).
z Ceph: Shadowing in upper head region with programs C3 and C3F.
z Ceph: Shadowing in thyroid region with programs C1 and C2.
1 – 10 D3352.076.01.13.02 07.2008
59 38 399 D3352

Tab 1 1.5 Operation notes

1.5 Operation notes
Rated line voltage The ORTHOPHOS XG X-ray unit can be operated in the following
rated line voltage ranges:
z 200 V - 240 V
z 50/60 Hz
The permissible line voltage fluctuations are as follows:
z 200 - 230 V: ± 10 %
The internal line impedance must not exceed max. 0.8 Ω.
i
NOTE
Only permanent electrical connection of the system is allowed in Germany.
Remote control The system can be equipped with...
z a 1 - 3 m coiled cable with release button inside the treatment room or ...
z a remote control with or without coiled cable located outside the X-ray
room (see installation instructions).
1.5
1.5
bеЦдблЬ
Warm-up time After it is switched ON, the system requires a warm-up time of approx. 1 min.
Self-adjustment routine At the same time, a mechanical and electronic self-adjustment routine is exe-
cuted. If a button is pressed during the self-adjustment routine, an error mes­sage is displayed on the Multipad (XG 5 / 3 / 3 PPE) or Easypad (XG
Cooling period The cooling period between two exposures is maintained by an automatic
exposure blocking function according to the pulse/pause ratio. A countdown of the waiting time is displayed on the Multipad (XG 5 / 3 / 3 PPE) or Easypad (XG
Turn-off time The turn-off time must amount to at least 60s.
Demo units If the X-ray unit is to be presented as a demo unit at trade fairs or exhibitions,
it must be ensured that radiation release is blocked (see “Demo mode – Operation without radiation release” on page 1-13).
Software version The overall system software version is determined by the software statuses of
the EEPROMs on the boards (see “List of software versions” on page 1-16).
Wireless phone interference with medical electrical equipment
Disposal The X-ray tube assembly contains a tube with potential implosion hazard,
To ensure safe operation of medical electrical equipment, the use of mobile wireless phones in practice or hospital environments is prohibited.
a small amount of beryllium, a lead lining as well as mineral oil.
Plus
).
Plus
).
Error messages Error messages are displayed on the single-line display of the Multipad
(XG 5 / 3 / 3 PPE) or on the touchscreen of the Easypad (XG
59 38 399 D3352 D3352.076.01.13.02 07.2008
Plus
).
1 – 11
1.5 Operation notes Tab 1
Help messages in case exposure readiness cannot be attained
If you have to remove covers from the unit
Help messages are displayed on the single-line display of the Multipad (XG 5 / 3 / 3 PPE) or on the touchscreen of the Easypad (XG
Proceed according to section "1.14 Removing the covers".
When removing covers, always remember that direct sunlight or bright room lighting can cause system malfunctions due to activated light barriers.
Therefore: avoid direct sunlight and bright room lighting above the unit!
When attaching the covers: be sure to screw the sheet metal cover back on.
Plus
).
CAUTION
For reasons of electromagnetic compatibility, be sure to fasten all screws.
Reattach the covers.
Secondary diaphragm Do not manually move or otherwise exert force on the secondary diaphragm
(e.g. when removing it from its packaging).
Measurements Always switch the unit OFF before connecting a measuring instrument.
Select the correct current/voltage type and adjust the measuring range to match the expected readings.
Perform continuity tests only on units which are switched off.
If several exposures with radiation must be taken to check a measurement, make sure that the prescribed cool-down intervals are observed. They are maintained by an automatic exposure blocking function (see operating instructions).
The pulse/pause ratio is 1: 10, i.e. a 10-second pause is maintained for each second of radiation emission. The pulse/pause ratio is automatically main­tained (automatic exposure blocking).
A pulse/pause ratio of 1:20 is better for the X-ray tube.
WARNING
It is essential that you observe the radiation protection regulations ap­plicable in your country prior to radiation release.
The test rotations triggered by pressing the T key on the Easypad and then the release button are executed without radiation.
When replacing parts Switch the unit OFF before replacing parts.
The unit must be disconnected from the junction box of the building installa­tion before replacing any parts near the power supply, power switch, board DX 32 or the X-ray tube assembly!
Please always wear an ESD wrist band to protect sensitive components on printed circuit boards (ESD).
Always check the system and adjust it as required after replacing a board or the X-ray tube assembly.
The article numbers for ordering spare parts can be found in the spare parts list, Order No. 59 38 423. The diagrams contained in the spare parts list pro­vide a useful guide when replacing parts.
1 – 12 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1 1.6 Demo mode – Operation without radiation release
1.6 Demo mode Operation without radiation
release
1.6
1.
2.
1.6
bеЦдблЬ
A
B
4.3.
Board DX6
59 38 399 D3352 D3352.076.01.13.02 07.2008
S2
J6
i
NOTE
Starting with system SW version V02.28, the complete possible functionality of the system class is simulated, regardless of the current configuration.
1 – 13
1.6 Demo mode – Operation without radiation release Tab 1
1.6.1 Switching the demo mode ON
5.
Info menu for XG 5 / 3 / 3 PPE
P1______14.1__64_8
n
(
When operated in demo mode, the unit must not release any radiation. For this reason, you must take the following safety measures:
z Switch the unit OFF.
1. Remove the cover of the tube assembly.
2. Loosen screws A and remove cover plate B.
3. Set dip switch S2 (DX6) to position 2.
4. Pull cable L5 (X-RAY) off of connector J6 (DX6).
Radiation release is now no longer possible.
123
(
SYSTEM SOFTWARE____
n
(
123
Scroll through the list until the demo mode is displayed
Info screen for XG
Plus
(
5. Switch the unit ON and check the mode with the info menu (XG 5 / 3 / 3 PPE) or the info screen (XG
Demo mode: ON means that: The demo mode is switched ON (Radiation release is not possible)
Demo mode: OFF means: The demo mode is switched OFF (Radiography, X-ray radiation are possible!)
z Switch the unit OFF again and reattach cover plate B and the tube assem-
bly covers by following the dismantling procedure in reverse order.
Plus
).
1.6.2 Switching the demo mode OFF
z Switch the unit OFF.
1. Remove the cover of the tube assembly.
2. Loosen screws A and remove cover plate B.
3. Set dip switch S2 (DX6) to position 1.
4. Pull cable L5 (X-RAY) off of connector J6 (DX6).
Radiation release is now once again possible.
5. Switch the unit ON and check the mode with the info menu (XG 5 / 3 / 3 PPE) or the info screen (XG
Demo mode: ON means that: The demo mode is switched ON (Radiation release is not possible)
Demo mode: OFF means: The demo mode is switched OFF (Radiography, X-ray radiation are possible!)
z Switch the unit OFF again and reattach cover plate B and the tube assem-
bly covers by following the dismantling procedure in reverse order.
Plus
).
Orthophos XG configuration
Mac adress : 010001200000 IP adress : 192.168.15.19 Subnet mask : 155.155.255.0 Default Gateway : 192.168.15.1 Net API version : 1 Net API revision : 14
Demomode : ON
Tube Exposures : 2217
2/2
)
65
59 38 399 D3352
1 – 14 D3352.076.01.13.02 07.2008
Tab 1 1.7 Demo mode – Repacking and transport
1.7 Demo mode Repacking and transport
1.7
1.
A
i
NOTE
If the unit is installed with a floor stand, its height in­creases by 30 mm. The bite block height value dis­played on the Multipad (XG 5 / 3 / 3 PPE) or Easypad (XG
Plus
) remains the same, however!
1.7
bеЦдблЬ
Panoramic X-ray unit z Switch the panoramic X-ray unit ON and move it to its packing height by
pressing the Easypad (XG
– Bite block height = 965 mm (displayed as height on
Multipad (XG 5 / 3 / 3 PPE) or Easypad (XG
– Bottom edge of slide cover = 702 mm
i
NOTE
The bottom edge of the slide cover must be at the same height as the mark­ings A in the column.
Cephalometer z Start service routine S034, test step 6.
z Move the cephalometer to the packing position
(see page 5-143).
z For information on packing the units and on the packing condition, see the
relevant installation instructions.
59 38 399 D3352 D3352.076.01.13.02 07.2008
UP/DOWN keys on the Multipad (XG 5 / 3 / 3 PPE) or
Plus
).
Plus
)
1 – 15

1.8 List of software versions Tab 1

1.8
i
List of software versions
NOTE
Any software combinations other than those listed here are not allowed. If the software version of any particular module does not match the overall software version, the overall software version will be marked with an asterisk on the info screen (e.g. 02.20*).
1.8.1 ORTHOPHOS XG 3 PPE
Panoramic unit
Board DX6 DX71 DX11 DX81
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
02.60 02.33 02.40 02.33
02.60 02.33 02.40.01 02.33
02.60 02.34 02.42 02.33
02.60 02.34 02.42 02.33
02.60 02.34 02.43 02.33
Remarks
compatible with SIDEXIS versions V01.61 and higher
compatible with SIDEXIS versions V01.61 and higher
compatible with SIDEXIS versions V01.61 and higher
compatible with SIDEXIS versions V01.61 and higher
compatible with SIDEXIS versions V01.61 and higher
Remote control
Board DX42
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
Sidexis XG
V 01.53
V 01.54
V 01.55
V 01.61
V 02.00
V 2.2
V 2.3
Remarks Database
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
requires ORTHOPHOS XG 3 PPE Overall Software Version V02.28.2 or higher
02.30
02.30
02.31
02.31
02.31
Remarks
old Database
new Database
1 – 16 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1 1.8 List of software versions
1.8.2 ORTHOPHOS XG 3
Panoramic unit
Board DX6 DX71 DX11 DX41 DX81
Overall software V 02.25
Overall software V 02.27
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
02.57 02.29 02.37 02.22 02.30
02.60 02.31 02.39 02.23 02.32
02.60 02.33 02.40 (02.23*) 02.33
02.60 02.33 02.40.01 (02.23*) 02.33
02.60 02.34 02.42 (02.23*) 02.33
02.60 02.34 02.42 (02.23*) 02.33
02.60 02.34 02.43 (02.23*) 02.33
Remarks
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
1.8
1.8
bеЦдблЬ
Remote control
Board DX42
Overall software V 02.25
Overall software V 02.27
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
Sidexis XG
V 01.51
V 01.53
V 01.54
V 01.55
V 01.61
V 02.00
V 2.2
V 2.3
Remarks Database
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
requires ORTHOPHOS XG 3 Overall Soft­ware Version V02.25 or higher
02.27
02.28
02.30
02.30
02.31
02.31
02.31
Remarks
old Database
new Database
59 38 399 D3352 D3352.076.01.13.02 07.2008
* As of hardware version BA (from November 2006) new systems will be delivered without
board DX41 (see also page 1-44).
1 – 17
1.8 List of software versions Tab 1
1.8.3 ORTHOPHOS XG 5 / Ceph
Panoramic unit Cephalometer
Board DX6 DX71 DX11 DX41 DX61 DX81 DX91
Overall software V 02.18.1
Overall software V 02.20
Overall software V 02.22
Overall software V 02.25
Overall software V 02.27
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
02.45 02.23 02.31 02.19 02.20 02.25 02.32
02.56 02.28 02.33 02.22 02.26 02.29 02.34
02.56 02.28 02.35 02.22 02.30 02.29 02.38
02.57 02.29 02.37 02.22 02.30 02.30 02.40
02.60 02.31 02.39 02.23 02.33 02.32 02.42
02.60 02.33 02.40 (02.23*) 02.33 02.33 02.42
02.60 02.33 02.40.01 (02.23*) 02.34 02.33 02.43
02.60 02.34 02.42 (02.23*) 02.34 02.33 02.43
02.60 02.34 02.42 (02.23*) 02.34 02.33 02.44
02.60 02.34 02.43 (02.23*) 02.35 02.33 02.44
Remote control
Board DX42
Overall software V 02.18.1
Overall software V 02.20
Overall software V 02.22
Overall software V 02.25
Overall software V 02.27
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
02.22
02.27
02.27
02.27
02.28
02.30
02.30
02.31
02.31
02.31
Remarks
compatible with SIDEXIS versions up to and incl. V01.44
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
Remarks
1 – 18 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1 1.8 List of software versions
Sidexis XG
V 01.33 - V01.44
V 01.45
V 01.50
V 01.51
V 01.53
V 01.54
V 01.55
V 01.61
V 02.00
V 2.2
V 2.3
* As of hardware version BA (from November 2006) new systems will be delivered without
board DX41 (see also page 1-44).
Remarks Database
requires ORTHOPHOS XG 5 Overall Software Version < V02.20
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
requires ORTHOPHOS XG 5 Overall Software Version V02.20 or higher
old Database
new Database
1.8
1.8
bеЦдблЬ
59 38 399 D3352 D3352.076.01.13.02 07.2008
1 – 19
1.8 List of software versions Tab 1
1.8.4 ORTHOPHOS XG
Panoramic unit Cephalometer
Board DX6 DX7 DX11 DX41 DX61 DX81 DX91
Overall software V 02.17
Overall software V 02.19
Overall software V 02.20
Overall software V 02.22
Overall software V 02.25
Overall software V 02.27
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
02.41 02.25 02.30 02.16 02.17 02.23 02.31
02.52 02.31 02.32 02.20 02.25 02.26 02.34
02.56 02.35 02.33 02.22 02.26 02.29 02.34
02.56 02.38 02.35 02.22 02.30 02.29 02.38
02.57 02.40 02.37 02.22 02.32 02.30 02.40
02.60 02.42 02.39 02.23 02.33 02.32 02.42
02.60 02.44 02.40 (02.23*) 02.33 02.33 02.42
02.60 02.44 02.40.01 (02.23*) 02.34 02.33 02.43
02.60 02.44 02.42 (02.23*) 02.34 02.33 02.43
02.60 02.44 02.42 (02.23*) 02.34 02.33 02.44
02.60 02.44 02.43 (02.23*) 02.35 02.33 02.44
Plus
/ Ceph
Remarks
compatible with SIDEXIS versions up to and incl. V01.44
compatible with SIDEXIS versions up to and incl. V01.44
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.45 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
compatible with SIDEXIS versions V01.53 and higher
Remote control
Board DX42
Overall software V 02.17
Overall software V 02.19
Overall software V 02.20
Overall software V 02.22
Overall software V 02.25
Overall software V 02.27
Overall software V 02.28
Overall software V 02.28.2
Overall software V 02.30
Overall software V 02.31
Overall software V 02.32
02.17
02.24
02.27
02.27
02.27
02.28
02.30
02.30
02.31
02.31
02.31
Remarks
1 – 20 D3352.076.01.13.02 07.2008
59 38 399 D3352
Tab 1 1.8 List of software versions
Sidexis XG
V 01.33 - V01.44
V 01.45
V 01.50
V 01.51
V 01.53
V 01.54
V 01.55
V 01.61
V 02.00
V 2.2
V 2.3
* As of hardware version BA (from November 2006) new systems will be delivered without
board DX41 (see also page 1-44).
Remarks Database
requires ORTHOPHOS XG ware Version < V02.20
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
requires ORTHOPHOS XG Overall Software Version V02.20 or higher
Plus
Overall Soft-
Plus
Plus
Plus
Plus
Plus
Plus
Plus
Plus
Plus
Plus
old Database
new Database
1.8
1.8
bеЦдблЬ
59 38 399 D3352 D3352.076.01.13.02 07.2008
1 – 21

1.9 Software update Tab 1

1.9
Software update
1.9.1 Important information on the software update
i
NOTE
Read the information provided on the ORTHOPHOS XG Software CD and on the SIRONA dealer page on the internet very carefully. It always con­tains the latest information on the software update.
i
NOTE
Before performing a software update, it is essential to read the program-ver­sion-specific information in sections 1.9.4 to 1.9.6.
CAUTION
Important for TSA operation:
For downgrading from a software version V2.20 to one < V2.20 with the TSA mode activated, the TSA sensor holder must be equipped with the transport safety device and set to the PAN position!
CAUTION
Important for XG 21007 (without Ceph) or 41001 (with Ceph):
A downgrade to a software version V02.25 must not be performed on these units in any case. Such a downgrade would result in permanent and irreparable failure of the user interface!
Plus
units with serial numbers
CAUTION
Important for units with the following serial numbers or higher:
ohne Ceph mit Ceph
XG 3 PPE 104000
XG 3 104000
XG 5 63500 83500
Plus
XG
A downgrade to a software version < V02.32 cannot be performed on these units.
23500 43500
1 – 22 D3352.076.01.13.02 07.2008
59 38 399 D3352
Loading...
+ 548 hidden pages