olympus AU 400, AU 600, AU 640 User Manual

5 (1)
olympus AU 400, AU 600, AU 640 User Manual

Olympus AU 400 - 600 - 640 instructions

rev 422.0.2 - 2010-08-01

Chema Diagnostica

Via Campania 2/4 I-60030 Monsano | An Italy

T +39 0731 605064 | F +39 0731 605672 www.chema.com | mail@chema.com

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Olympus AU 400 - 600 - 640 instructions

rev 422.0.2 - 2010-08-01

Warning!

These programs should be used only as guideline. To check the reagent correct setting on the instrument, use only good quality control sera and work accordingly to with good laboratory practice. Instrumental factor must be checked. Chema Diagnostica will not be responsible for instrumental bad programming.

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 3 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

ACID PHOSPHATASE - Codes AC 0120 TC

Preparation:

POWDER SINGLE REAGENT. FOLLOW PREPARATION PROGRAM INDICATED IN INSERT

 

SHEET. INSTALL AS R1

Storage:

REFRIGERATE AT 2-8°C

TEST NAME:

ACP

 

 

 

 

 

Sample:

Volume

20 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

200 μl

Dilution

0 μl

 

 

 

R2 volume

0 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 800

Min OD

 

Max OD

 

Method:

RATE

 

L

- 0.1

H

2.5

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

- 0.1

First H

0.3

 

 

 

Last L

- 0.1

First H

0.3

Measuring point 1: First 12

Last 27

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

20%

 

Dynamic range:

 

 

No Lag Time:

NO

 

L 1

H 60

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 30

Unit: U/L

Factor: 853

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 4 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

ALBUMIN - Codes BC 0100 / 0500 / 1000 / 1500 CH

Preparation:

R1 - LIQUID READY TO USE SINGLE REAGENT

Storage:

REFRIGERATE (2-8°C)

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

ALB

 

 

 

 

 

Sample:

Volume

2 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

295 μl

Dilution

0 μl

 

 

 

R2 volume

0 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 600

Sec. 800

Min OD

 

Max OD

 

Method:

END

 

L

 

H

 

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

- 0.1

First H

0.5

 

 

 

Last L

- 0.1

First H

0.5

Measuring point 1: First 0

Last 4

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

 

 

Dynamic range:

 

 

No Lag Time:

YES

 

L 0.1

H 6.0

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 90

Unit: g/dl

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 5 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

ALKALINE PHOSPHATASE FL (IFCC) - Codes AF F080 / F245 / F400 / F600 CH

Preparation:

R1 - INSTALL LIQUID READY TO USE REAGENT A

 

R2 - INSTALL LIQUID READY TO USE REAGENT B

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

ALP

 

 

 

 

 

Sample:

Volume

5 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

200 μl

Dilution

0 μl

 

 

 

R2 volume

50 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 480

Min OD

 

Max OD

 

Method:

RATE

 

L

-0.1

H

2.5

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

-0.1

First H

1.5

 

 

 

Last L

-0.1

First H

1.5

Measuring point 1: First 13

Last 19

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

20%

 

Dynamic range:

 

 

No Lag Time:

NO

 

L 5

H 3000

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 60

Unit: U/L

Factor: 2757

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 6 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

ALKALINE PHOSPHATASE (DGKC) FL - Codes AL F080 / F245 / F400 / F600 CH

Preparation:

R1 - INSTALL LIQUID READY TO USE REAGENT A

 

R2 - INSTALL LIQUID READY TO USE REAGENT B

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

ALP

 

 

 

 

 

Sample:

Volume

5 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

200 μl

Dilution

0 μl

 

 

 

R2 volume

50 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 480

Min OD

 

Max OD

 

Method:

RATE

 

L

- 0.1

H

2.0

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

- 0.1

First H

1.2

 

 

 

Last L

- 0.1

First H

1.2

Measuring point 1: First 20

Last 24

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

20%

 

Dynamic range:

 

 

No Lag Time:

NO

 

L 7

H 3000

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 30

Unit: U/L

Factor: 2757

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 7 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

AMYLASE FL - Codes AM F060 / F120 / F245 CH

Preparation:

R1 - LIQUID READY TO USE SINGLE REAGENT

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

AMY

 

 

 

 

 

Sample:

Volume

6 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

240 μl

Dilution

0 μl

 

 

 

R2 volume

0 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 480

Min OD

 

Max OD

 

Method:

RATE

 

L

- 0.1

H

2.0

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

- 0.1

First H

0.3

 

 

 

Last L

- 0.1

First H

0.3

Measuring point 1: First 5

Last 10

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

20%

 

Dynamic range:

 

 

No Lag Time:

NO

 

L 10

H 3000

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 30

Unit: U/L

Factor: 3178

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 8 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

AMYLASE EPS FL - Codes EA F080 / F245 CH

Preparation:

R1 - INSTALL LIQUID READY TO USE REAGENT A

 

R2 - INSTALL LIQUID READY TO USE REAGENT B

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

AMY

 

 

 

 

 

Sample:

Volume

8 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

192 μl

Dilution

0 μl

 

 

 

R2 volume

48 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 480

Min OD

 

Max OD

 

Method:

RATE

 

L

- 0.1

H

2.0

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

- 0.1

First H

0.3

 

 

 

Last L

- 0.1

First H

0.3

Measuring point 1: First 20

Last 27

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

20%

 

Dynamic range:

 

 

No Lag Time:

NO

 

L 6

H 1500

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 30

Unit: U/L

Factor: 3480

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 9 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

PANCREATIC ISOAMYLASE FL - Code PA F080 / F245 CH

Preparation:

R1 - INSTALL LIQUID READY TO USE REAGENT A

 

R2 - INSTALL LIQUID READY TO USE REAGENT B

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

AMYP

 

 

 

 

 

Sample:

Volume

5 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

200 μl

Dilution

0 μl

 

 

 

R2 volume

50 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 480

Min OD

 

Max OD

 

Method:

RATE

 

L

- 0.1

H

2.0

Reaction slope: +

 

Reagent OD limit:

 

 

 

 

 

First L

- 0.1

First H

0.3

 

 

 

Last L

- 0.1

First H

0.3

Measuring point 1: First 20

Last 27

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

20%

 

Dynamic range:

 

 

No Lag Time:

NO

 

L 2

H 2500

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 30

Unit: U/L

Factor: 3480

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 10 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

BICARBONATE FL - Code BR F060 / F245 / F400 CH

Preparation:

R1 - LIQUID READY TO USE SINGLE REAGENT

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

CO2

 

 

 

 

 

Sample:

Volume

2.5 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

250 μl

Dilution

0 μl

 

 

 

R2 volume

0 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 410

Sec. 480

Min OD

 

Max OD

 

Method:

FIXED

 

L

0.1

H

2.5

Reaction slope: -

 

Reagent OD limit:

 

 

 

 

 

First L

- 2.0

First H

2.5

 

 

 

Last L

- 2.0

First H

2.5

Measuring point 1: First 1

Last 9

 

 

 

 

Measuring point 2: First

Last

 

 

 

 

Linearity:

 

 

Dynamic range:

 

 

No Lag Time:

 

 

L 1

H 50

 

 

 

 

 

Correlation factor:

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 30

Unit: mmol/L

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 11 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

BILIRUBIN TOTAL FL - Code DT F125 / F500 CH

Preparation:

R1 - INSTALL LIQUID READY TO USE REAGENT A

 

R2 - INSTALL LIQUID READY TO USE REAGENT B

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

BILT

 

 

 

 

 

 

Sample:

Volume

10 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

200 μl

Dilution

0 μl

 

 

 

R2 volume

50 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 520

Sec. 800

Min OD

 

Max OD

 

Method:

END

 

 

L

 

H

 

Reaction slope: +

 

 

Reagent OD limit:

 

 

 

 

 

 

First L

- 0.1

First H

0.5

 

 

 

 

Last L

- 0.1

First H

0.5

Measuring point 1: First

Last

27

 

 

 

 

Measuring point 2: First

Last

10

 

 

 

 

Linearity:

 

 

 

Dynamic range:

 

 

No Lag Time:

 

 

 

L 0.05

H 20

 

 

 

 

 

 

Correlation factor:

 

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 60

Unit: mg/dl

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 12 of 41

Chema Diagnostica - Guidelines for automatic analyzers - Olympus AU 400 - 600 - 640

Analyzer:

Olympus AU 400 - 600 - 640

Application:

BILIRUBIN DIRECT FL - Code DD F125 / F500 CH

Preparation:

R1 - INSTALL LIQUID READY TO USE REAGENT A

 

R2 - INSTALL LIQUID READY TO USE REAGENT B

Storage:

REFRIGERATE AT 2-8°C

Stability:

AS INDICATED IN THE LABEL

TEST NAME:

BILD

 

 

 

 

 

 

Sample:

Volume

10 μl

Dilution

0 μl

Pre-dilution Rate: 1

Reagents:

R1 volume

200 μl

Dilution

0 μl

 

 

 

R2 volume

50 μl

Dilution

0 μl

 

 

Wavelenght:

Pri. 540

Sec. 660

Min OD

 

Max OD

 

Method:

END

 

 

L

 

H

 

Reaction slope: +

 

 

Reagent OD limit:

 

 

 

 

 

 

First L

- 0.1

First H

0.5

 

 

 

 

Last L

- 0.1

First H

0.5

Measuring point 1: First

Last

27

 

 

 

 

Measuring point 2: First

Last

10

 

 

 

 

Linearity:

 

 

 

Dynamic range:

 

 

No Lag Time:

 

 

 

L 0.05

H 20

 

 

 

 

 

 

Correlation factor:

 

 

 

 

 

 

A 1

B 0

 

 

On-board stability period: 60

Unit: mg/dl

IUS.422.0.2

Directions for use

Edition 2010-08-01

page 13 of 41