Mindray Datascope Passport V User manual

Copyright
Intellectual Property Statement
Mindray DS USA, Inc. (hereinafter called Mindray DS) owns the intellectual property
rights to this product and this manual. This manual may refer to information protected by
copyrights or patents and does not convey any license under the copyright and the patent
rights of Mindray DS, nor the rights of others.
Disclosure of the information in this manual in any manner whatsoever without the written
permission of Mindray DS is strictly forbidden. Release, amendment, reproduction,
distribution, rental, adaption and translation of this manual in any manner whatsoever
without the written permission of Mindray DS is strictly forbidden.
is a trademark or a registered trademark of Shenzhen Mindray
Bio-Medical Electronics Co., Ltd. All third-party trademarks that appear in this manual are
used solely for editorial purposes and are the property of their respective owners.
Contents of this manual are subject to changes without prior notice.
© 2008-2010 Mindray DS USA, Inc. All rights reserved.
WARNING
z Federal Law (USA) restricts this device to sale by or on the order of a physician.
Pulse oximeter Service Manual I
Manufacturer’s Responsibility
All information contained in this manual is believed to be correct. Mindray DS shall not be
liable for errors contained herein nor for incidental or consequential damages in
connection with the furnishing or use of this manual.
Mindray will not be liable for the effects on safety, reliability and performance of this
product if:
any installation operations, expansions, changes, modifications and repairs of this
product are not conducted by Mindray DS authorized personnel; and
the electrical installation of the relevant room does not comply with the applicable
national and local requirements; and
the product is not used in accordance with the instructions for use.
Warranty
This warranty is exclusive and is in lieu of all other warranties, expressed or implied,
including warranties of merchantability or fitness for any particular purpose.
Exemptions
Mindray DS's obligation or liability under this warranty does not include any transportation
or other charges or liability for direct, indirect or consequential damages or delay resulting
from the improper use or application of the product or the use of parts or accessories not
approved by Mindray DS or repairs by people other than Mindray DS authorized
personnel.
This warranty shall not extend to
Any Mindray DS product which has been subjected to misuse, negligence or
accident; or
Any Mindray DS product from which Mindray DS's original serial number tag or
product identification markings have been altered or removed; or
Any product of any other manufacturer.
II Pulse oximeter Service Manual
Return Policy
In the event that it becomes necessary to return a unit to Mindray DS, follow the
instructions below.
1. Return authorization.
Contact the Customer Service Department and obtain a Customer Service Authorization
number. This number must appear on the outside of the shipping container. Returned
shipments will not be accepted if the number is not clearly visible. Please provide the
model number, serial number, and a brief description of the reason for return.
2. Freight policy
The customer is responsible for freight charges when this product is shipped to Mindray
DS for service (this includes customs charges).
3. Return address
Please send the part(s) or equipment to the address offered by the Customer Service
Department.
Contact Information
Manufacturer: Mindray DS USA, Inc.
Address: 800 MacArthur Blvd.Mahwah, New Jersey 07430 USA
Tel: 1.800.288.2121, 1.201.995.8000
Website: www.mindray.com
Pulse oximeter Service Manual III
Safety Precautions
1 . Meaning of Signal Words
In this manual, the signal words WARNING and CAUTION are used regarding safety
and other important instructions. The signal words and their meanings are defined as follows.
Please understand their meanings clearly before reading this manual.
Signal word Meaning
WARNING
CAUTION
CAUTION Indicates a potentially hazardous situation which, if not avoided, may
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
result in property damage.
2 . Meaning of Safety Symbols
Symbol Description
Type-BF applied part
"Attention" (Refer to the operation manual.)
3 . Safety Precautions
Please observe the following precautions to ensure the safety of service engineers as well as
operators when using this system.
IV Pulse oximeter Service Manual
WARNING: Do not connect this system to outlets with the same circuit
breakers an d fuses that control current to devices such as life-support systems. If this system malfunctions and generates an overcurrent, or when there is an instantaneous current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped. Do not use flammable gasses such as anesthetics, or flammable liquids such as ethanol, near this product, because there is danger of explosion.
CAUTION: 1. Malfunctions due to radio waves
(1) Use of radio-wave-emitting devices in the proximity of this
kind of medica l electroni c system may interfere with its operation. Do not bring or use devices which generate radio waves, such as cellular telephones, transceivers, and radio controlled toys, in the room where the system is installed.
(2) If a user brings a device which generates radio waves near
the system, they must be instructed to immediately turn OFF the device. This is necessary to ensure the proper operation of the system.
2. Do not allow fluids such as water to contact the system or peripheral devices. Electric shock may result.
Pulse oximeter Service Manual V
VI Pulse oximeter Service Manual
Content
Content
Chapter 1 Introduction...................................................................................................1-1
1.1 General ......................................................................................................................1-1
1.2 Functions....................................................................................................................1-1
1.3 Parameters Measurement..........................................................................................1-1
Chapter 2 Circuit Principle.................. ..........................................................................2-1
2.1 Overview ....................................................................................................................2-1
2.2 Reliability Design........................................................................................................ 2-7
2.3 EMC Design...............................................................................................................2-8
Chapter 3 Specifications...............................................................................................3-1
3.1 General ......................................................................................................................3-1
3.2 Parameter Specifications ........................................................................................... 3-1
3.3 Display and Control.................................................................................................... 3-1
3.4 Input/Output Communications....................................................................................3-2
3.5 Electrical specifications ..............................................................................................3-2
3.6 Printing.......................................................................................................................3-2
3.7 Physical Characteristics.............................................................................................3-3
3.8 Environment and Safety............................................................................................. 3-3
3.9 EMC ........................................................................................................................... 3-4
Chapter 4 Structure........................................................................................................4-1
4.1 Explosive Diagram ..................................................................................................... 4-1
4.2 Batteries Installation and Maintenance....................................................................... 4-2
Chapter 5 Test and Prompt List....................................................................................5-1
5.1 Test Procedure ........................................................................................................... 5-1
5.2 Prompt List.................................................................................................................5-2
Chapter 6 Maintenance and Cleaning ..........................................................................6-1
6.1 Maintenance...............................................................................................................6-1
6.2 Cleaning.....................................................................................................................6-1
Pulse oximeter Service Manual I
Content
6.3 Disinfection & Sterilization..........................................................................................6-2
6.4 Technical Support.......................................................................................................6-2
II Pulse oximeter Service Manual
Introduction
Chapter 1 Introduction
1.1 General
The DPM1 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate
monitor. It operates only on battery power using existing DPM1 disposable and reusable finger
and ear sensors labeled for patients ranging from neonates to adults.
Parameters monitored by the DPM1 pulse oximeter include: arterial oxygen saturation (SpO
pulse rate (PR) and pulse strength. The DPM1 pulse oximeter employs a finger SpO
to measure patient’s SpO
screen.
The DPM1is operated and controlled by three buttons, which are Power Button, Backlight
Button and ID Confirm Button. The DPM1 oximter is also capable of data manage and
exporting the patient’s trend data to a PC for printing through the SpO2/communication
multiplex port.
, PR and pulse strength, and all of these are displayed on the LCD
2
2
),
2
sensor
1.2 Functions
DPM1 has the functions shown as below
1. Monitoring-----SpO
2. Printing-----patient ID, trend data, measurement time.
3. Power Saving------automatic standby, automatic shutdown.
4. Warning-----memory full, ID full, low battery, standby, technical error.
Printing is available only when the device is equipped with a communication cable and works
with a PC with a printer.
, PR and pulse strength.
2
1.3 Parameters Measurement
Parameters monitored by DPM1 pulse oximeter includes: SpO2, PR and pulse strength.
DPM1measures SpO
to determine oxygen saturation of hemoglobin.
DPM1 also can determine pulse rate and pulse strength, which are indicated on the LCD
screen after processing.
by pulsating oximetry, which is a continuous and non-invasive method
2
Pulse oximeter Service Manual 1-1
Loading...
+ 23 hidden pages