User manual
Useful
information
General guidelines
1
The purpose of these instructions is to explain to the user how to use the GS-57
sealer safely and correctly.
Anyone whose work involves using the sealer must be familiar with these instructions
and the safety regulations.
These instructions must always be available to personnel, to prevent incorrect use of
the sealer and to ensure trouble-free operation.
Copyright
These instructions may only be given to third parties with the written consent of
GETINGE. All documents are protected by copyright.
The passing-on and copying of documents, including extracts and the communication
or use for gain of their contents is not permitted, except by express consent.
Infringements are punishable and will lead to claims for damages. We reserve all
rights to exercise the provisions of commercial copyright law.
Warranty and liability
We accept liability within the terms of the stated warranty obligations for errors and
omissions, to the exclusion of further claims.
We reserve the right to make technical changes to the sealer covered by this
instruction manual, for the purpose of further development.
We accept no liability for damage and operational disturbances arising from incorrect
operation, failure to observe these instructions, or improper repairs. Only original
spare parts and accessories approved by GETINGE may be used.
GETINGE accepts no liability for damage resulting from the use of non-approved
spare parts and accessories.
The drawings and graphics are not to scale.
Explanations
Important instructions given in the technical documentation are indicated as follows:
Warning
UM-GS57-V1.03-GB 3
Page 4
User manual
Safety regulations
2
These instructions must be read and strictly complied with.
2.1 Before use
Users must familiarize themselves with the instructions in this manual before using
the GS-57 sealer.
2.2 Field of application
• The GS-57 sealer is to be used only for sealing sterilization packages. No
liability will be accepted for use of the sealer for any other purpose.
• The sealer may only be used for its intended purpose, in accordance with the
generally recognized rules of technology and the industrial health and safety
regulations. It must not be used if it has defects that might be dangerous to
employees or third parties.
• The sealer only may be used under supervision.
2.3
• Before the case is opened, the sealer must be disconnected from the electric
power supply. Open the housing by loosening the 2 housing cover screws (not
marked) on the backside of the unit.
• Only an authorized technician or a specialized dealer may undertake servicing
work on the sealer.
Danger of electric shock
2.4 Damage to the sealer
• If the packaging material becomes trapped, it must not be pulled out of the
sealer. This would damage the sealer.
• The sealer is not suitable for sealing Polyethylene (PE) bags. The sealer may
be damaged if an unsuitable material is used.
• The reverse button must not used to feed packaging material into the sealer
from the left (see section 5.13, page 20).
2.5
• If the pressure of the pressure rollers is incorrectly set, the seam will not be
bacteria-tight (see section 5.11, page 16).
•
The user must set the correct sealing temperature for each packaging material
and sealer. This is done by trial sealing operations.
The usual sealing temperatures are between 120 °C and 200 °C. Our
Contamination of sterile goods
“Danger of patient infection”
UM-GS57-V1.03-GB 4
Page 5
User manual
recommendation is to set the sealing temperature to 120 °C and to raise it
gradually until a bacteria-tight seam is obtained.
• If the temperature is too low, the seam will be incomplete or weak.
• If the temperature is too high, the package will be difficult to peel open. Paper
fibers may be released and contaminate the sterile goods.
•
Released paper fibers may result in patient infection.
2.6
• When the printer has been running for some time, the print head will be hot.
Avoid touching the print head when changing the ink ribbon. (see Section 5.12.2
page 19).
Hot surface
UM-GS57-V1.03-GB 5
Page 6
User manual
Installation
3
Set up the GS-57 sealer on a clean, smooth surface as follows:
•
Position the sealer as required.
• Leave at least 15 cm free space behind the sealer to ensure good cooling and
trouble-free operation.
• Remove the printer blockade by lifting the sealers cover and cut the ty-rap
connected on the printer.
• Plug the mains cable into the socket on the back of the sealer.
•
Check that the voltage and frequency of the supply are 230Vac 60Hz. If the
voltage and frequency of the supply are 115Vac 50Hz replace the two 4A slow
fuses inside the power switch by two 8A slow fuses. These fuses are delivered
with the sealer.
•
Plug the mains lead into the power socket.
• The GS-57 sealer is now installed.
Note that the printers can only print if the print seam distance is large enough.
Paper guide accessories
GETINGE recommend the use of a paper guide for convenient feeding and troublefree transport of the material.
• Install the paper guide on the sealers left side.
•
To set the seam width, slacken the knurled nut from the paper guide. Move
the paper guide to the required position and tighten the knurled nut.
• With the feed stop fully forward, the seam is about 15 mm away from the cutoff edge of the material; with the feed plate fully back it is about 35 mm away
from it.
Feeder plate accessory (Optional)
GETINGE recommend the use of a feeder plate for convenient feeding and troublefree transport of the material. The feeder plate is an optional accessory for the sealer.
The feeder plate is easily to mount by hanging it above the sealers front plate.
UM-GS57-V1.03-GB 6
Page 7
User manual
Overview of controls
4
1 Date of manufacture
2 Force measurement
3 Printer setting
4 Display
5 Personal ID code
6 Counter reading
7 Delete key in:
- input mode
Select key for:
- menu language
- Expiry date, week, month, year
8 Return key
9 Text selection and input up to 100 texts
10 Batch code
11 Menu
12 Print width setting
13 Alphanumeric keypad 0-9 or A-Z
14 Reserve button for sealing material
15 Temperature setting
- Select temperature
16 Expiry date
UM-GS57-V1.03-GB 7
Page 8
User manual
Operating instructions
5
The GS-57 sealer seals all sterilization packages that are normally commercially
available. The sealer is not suitable for sealing PE bags.
Damage to the sealer
See section2.4, page 4.
Important
►
►After every complete entry in the chosen mode (text entry, text selection,
The return key must be pressed after every function key or data entry, as
confirmation. In the instructions below, the return key is represented by
this symbol: ↵↵↵↵.
temperature entry by user or process data), the sealer is ready to seal
although the chosen menu is still displayed. “Ready to seal” only
appears on the display when the material to be seal is fed in.
5.1
Set the green switch on the back of the sealer to (I).
The following information appears on the display:
Heating up
Temp: 20°C
When the sealing temperature has been reached, the following information appears
on the display:
Ready to seal
Temp: 180°C Press: 65 N
5.2
Switching on the sealer
Delete Key
In input mode
•
In select mode
Press: 65 N
the delete key:
Is used to delete characters one by one.
the delete key is used:
• The sealer heats up to the
programmed temperature in about
30 seconds.
• The sealer is ready to seal.
•
Subsequent settings can now be done.
•
• In “Expiry date” mode to change week, month and year.
• In
“Menu”
In
“Temperature”
mode to change the language.
mode to select the temperature.
UM-GS57-V1.03-GB 8
Page 9
User manual
5.3
Return key
• Every entry must be confirmed with the return key.
5.4 Enter date of manufacture
The date of manufacture is entered only once, when the sealer is used for the first
time.
•
Press the button.
•
Time: 14:00
Date: 07-18-06
•
Press
• Enter the password (the factory default password is 1234).
Time and date can be set.
• The set values appear blinking on the display in sequence.
The sealer is in input mode.
• The alphanumeric keypad is used for data input.
• Hour ↵↵↵↵ (confirm), minutes ↵↵↵↵ (confirm). Day ↵↵↵↵ (confirm), month ↵↵↵↵ (confirm), year ↵↵↵↵ (confirm).
• Input mode is ended and the sealer is ready to seal.
↵↵↵↵
to confirm.
The display looks like this.
►
Important
If you make a mistake, it’s possible to delete single characters with the
delete key . This can only be done if the input has not been
confirmed. If the input has been confirmed, the procedure must be
repeated.
UM-GS57-V1.03-GB 9
Page 10
User manual
Month
Day
Year
5.5
You can enter any numerical value (period) from 0 to 99. Only when days are set it is
possible to enter any numerical value (period) from 0 to 999. The display shows the
previously set expiry date, eg days, weeks, month or year.
Enter expiry date
• Press the key.
Period: 10 Days
Expire: 03-20-07
• Press ↵↵↵↵ to confirm.
•
The display value (period) that can be set, blinks.
The sealer is in input mode.
•
The alphanumeric keypad is used for data input.
• Enter a new period and press ↵↵↵↵ to confirm.
• Press the delete key and enter the calculation period, eg days, weeks,
month or year, the press
• Input mode is ended and the sealer is ready to seal.
Example calculation:
Date: 03.20.06 numerical value of 10 (period: day, weeks, etc.).
Days 03-30-06
Weeks 05-29-06
Month exact 01-20-07 Exact date
1st of month 01-01-07 Rounded to the first of the month
16th of month 01-16-07 Rounded to the sixteenth of the month
Year exact 03-20-16 Exact date
Year M/1 03-01-16 Rounded to the first of the month
Year M/16 03-16-16 Rounded to the sixteenth of the month
• The display looks like this.
↵
↵
to confirm.
↵↵
UM-GS57-V1.03-GB 10
Page 11
User manual
5.6
You can enter up to 100 texts each of 24 characters.
To enter text:
Text entry
Text-choice: 0
• Press ↵ to confirm.
• The text number blinks on the display.
• The sealer is in input mode.
•
Using the alphanumeric keypad, choose the desired text number from 0 to 99,
then press ↵↵↵↵ to confirm.
• The text stored under that number (if any) is displayed.
•
To enter a new text, or to overtype the existing text, press the
• If you have previously entered a password or chosen text 0, you can enter the
new text with the keypad.
• For example, if you want to enter a lower-case
using the “<>” key you can select uppercase and lowercase letters, uppercase
and lowercase special characters and the relevant numeral. Choose the
desired character from the active group of characters.
The following sequence of characters appears:
•
When entry is complete, press
• Input mode is ended and the sealer can continue sealing.
• Press the key.
Uppercase letters:
Lowercase letters:
Uppercase special characters:
Lowercase special characters:
Numeral:
2
ABC
abc
↵
↵
to confirm.
↵↵
•
The display looks like this.
Ä ÅÆÇ
деажз
↵
↵
key again.
↵↵
“c”
, press the
“2”
key. By
Important
►
► If you want to enter two letters from the same key, eg “A” and “B”
► If you make a mistake, use the delete key ←←←← to delete the incorrect
►
UM-GS57-V1.03-GB 11
If you press the key too many times, keep pressing it until the desired
character appears.
immediately after each other, wait for a second, so you do not overtype
the first letter entered.
character.
To enter a space, press the 1 (SPACE) key once.
Page 12
User manual
►Text no. 0 can be changed without a password. Texts 1-99 can be
password-protected. If this is done, the texts can be selected but not
edited.
5.6.1 Text selection for the print menu
• Press the key.
Text-choice: 0
• Press ↵↵↵↵ to confirm.
• The text number that can be set blinks on the display.
The sealer is in input mode.
• Use the alphanumeric keypad to enter the desired text number (0 to 99) and
↵
press
• The text is displayed.
To print the selected text, see section 5.12, page 16.
• When text selection is complete, press ↵↵↵↵ to confirm.
•
Input mode is ended and the sealer is ready to seal.
5.6.2
To enter texts with the computer, connect the sealer PC socket to the serial port of
your computer with the supplied cable or connect the sealer to the computer with the
USB cable. Switch on the sealer and the computer. Install the sealer communication
software from the supplied CD and follow the attached instructions.
↵
to confirm.
↵↵
Text entry with a computer
•
The display looks like this.
UM-GS57-V1.03-GB 12
Page 13
User manual
5.7
You can enter up to 14 characters on the alphanumeric keypad. The input procedure
is as described in Text entry (see section 5.6, page 11.)
Pers. nr.:
5.8
You can enter up to 14 characters on the alphanumeric keypad. The input procedure
is as described in Text entry (see section 5.6, page 11.)
Enter personal ID code
• Press the key.
• The display looks like this.
• Press ↵↵↵↵ to confirm.
•
The display value that can be set blinks.
The sealer is in input mode.
• Enter personal ID code and press ↵↵↵↵ to confirm.
• Input mode is ended and the sealer is ready to seal.
Enter production code
• Press the key.
Lot nr: 2785SP
•
Press ↵ to confirm.
•
The display value that can be set blinks.
The sealer is in input mode.
• Enter a new production code and press ↵↵↵↵ to confirm.
• Input mode is ended and the sealer is ready to seal.
• The display looks like this.
UM-GS57-V1.03-GB 13
Page 14
User manual
5.9
If you want to do a certain number of sealing operations, the sealer can be set up as
follows.
Counts
Enter counter reading
• Press the key.
:
0
x
Maximum: 0
• Press ↵ to confirm.
•
The current setting disappears from the display.
The sealer is in input mode.
• The counter can be set to any value (eg 0). Press ↵↵↵↵ to confirm.
• The desired maximum reading (eg 100) can be set. Press ↵↵↵↵ to confirm.
•
The setting is saved and the sealer is ready to seal.
Important
►Taking the above setting as an example, when you have sealed 100 bags,
the sealer reports “Counter maximum” and accepts no more material
when the conveyor belt stops.
►
►
► The maximum value that can be set is 1000.
5.10
If a sealing operation fails, you can reverse the counter reading by 1.
If you do not want a maximum function, set the maximum value to 0. This
disables the function.
Set the sealing temperature
• The display looks like this.
You can enter the sealing temperature in two ways.
1. Select temperature freely
This is the procedure for setting the target temperature:
• Press the key.
Seal temp: 150°C
Set temp: 150°C ± 5°C
• Press ↵↵↵↵ to confirm
UM-GS57-V1.03-GB 14
• The display looks like this.
Page 15
User manual
• If necessary, enter the password (the factory default password is 1234).
•
The current setting disappears from the display.
The sealer is in input mode.
• The alphanumeric keypad is used to enter the target temperature.
• Now enter the desired target temperature for the material you are using and press ↵↵↵↵ to confirm.
•
Input mode is ended and the sealer is ready to seal.
• The sealer now cools down or heats up.
Important
►
2. Selecting a preset temperature
The maximum temperature is 200°C, the minimum temperature is 50°C.
• Press the key.
Seal temp: 150°C
Set temp: 150°C ± 5°C
• You can choose between four stored temperatures by pressing the delete key
←←←←
. Then press
• Input mode is ended and the sealer is ready to seal.
• The sealer now cools down or heats up.
Important
↵
↵
to confirm.
↵↵
• The display looks like this.
►
►
To store the temperatures
If the set temperature does not correspond to the target temperature,
transport does not start. If this happens, wait until the correct
temperature is reached.
Contamination of the sterilized goods
“Danger of patient infection”
See section 2.5, page 4.
(see section 6.4, page 23.)
UM-GS57-V1.03-GB 15
Page 16
User manual
Pressure:
65 N
5.11
Pressure measurement is optional. If pressure measurement is not used, this key
does nothing.
Set press.: 65 N ÷
Pressure measurement is set at the factory to 65 N. There is usually no need to
change this setting. The pressure settings may only be changed by service
personnel.
5.12
The sealer is equipped with one printer (single-line) or two printers (two-line).
The sealer can print different data at the same time during the sealing process:
This is how to set the print information:
Print:
Select >
Pressing the individual symbols activates the following data:
Pressure measurement
• Press the key.
÷ 9N
÷ ÷
Printer setting
• Press the key.
•
Press
•
Press
• The sealer is in input mode.
• To delete a symbol, press it again.
↵
↵
to confirm.
↵↵
↵
↵
to confirm.
↵↵
• Date of manufacture (= Fab)
• Date of expiry (= Exp)
• Personal ID number (= Pnr)
•
Batch number (= Lot)
•
Text selection (= Txt)
• The display looks like this.
•
The display looks like this.
UM-GS57-V1.03-GB 16
Page 17
User manual
• With one printer, all the data are printed on one line.
With two printers, the batch number and the text are printed on the second
line. This reduces the line length.
• Press ↵↵↵↵ to confirm.
•
After confirmation, the character width and line length appear automatically on
the display.
Symbol width: 1.6mm
Print-length: 68mm
Three printing widths are programmed at the factory.
The correct printing width for the packaging material in use can be selected by
pressing the <> symbol on the alphanumeric keypad.
Important
►
► Behind every symbol selected for printing there are data that were
►
►
►
►
The printing width must be matched to the size of the packaging
material. The packaging material must be slightly wider than the printing
width. The line width is only indicated when at least one printed is
switched on.
previously entered and confirmed.
Only the information shown on the display is printed.
When the packaging material is fed in, the paper side must be down.
It’s also possible to print the Barcode on the material. The barcode can
be switched on by connecting the sealer to a computer on the serial port.
(see programmers manual for further instructions)
If the printer is not used for a long time, we recommend removing the
print cassette (see section 5.12.2, page 19) to prevent the ink drying out.
•
The display looks like this
UM-GS57-V1.03-GB 17
Page 18
User manual
5.12.1
You can use the GS 46 sealer to print out the process data for documentation
purposes.
This is how to print the process data:
Print Process-report:
03-18-07 / 180°C 65N
Printing process data
• Insert a piece of packaging material.
•
The sealer prints the following process data:
This information is printed on the bag. This can be detached for
documentation purposes. When the process data have been printed, sealing
can continue.
• Press the key twice.
Date of manufacture
Temperature
(Pressure measurement)
Speed
Total counter reading
• The display looks like this.
UM-GS57-V1.03-GB 18
Page 19
User manual
5.12.2
When the print quality deteriorates, the printing cassette must be replaced. A shortterm improvement in the print quality can be achieved as follows:
Changing the printing cassette
Hot surface
See section 2.6, page 5.
1. Short-term improvement in print quality
• Switch off the sealer.
•
Open the cover (1).
• Move the blue lever (4) to the midway position (4), or if necessary to the upper
position (6).
The print quality will be improved for a short time.
2. Changing the printing cassette
• Switch off the sealer.
• Open the cover (1).
•
Move the blue lever (4) to the lowest position.
• Tilts back the printer (2).
• Pull out the printing cassette (3) in the direction of the arrow.
•
Insert a new printing cassette (7) at an angle into the carrier on the ink ribbon
holder and move the carrier up until it latches into place.
•
Leave the blue lever (4) in the lowest position.
• Close cover (1). The GS-57 sealer is ready for use.
UM-GS57-V1.03-GB 19
Page 20
User manual
5.13
This is what to do if material has become trapped:
5.14
With the label printer, which is available as a accessory, you can create barcode
labels. For this, the label printer is connected to the socket marked LABEL on the
back of the sealer, on the right, using the cable supplied. Now install the label printer
and load the labels and the transfer band, as described in the printer instructions.
The printer can print a label for every bag that passes through, or only on request.
To print a label for each bag, switch on the label printer, as described in section,
switching components on and off (see section 7.1, page 24.). If you have more than one
sealer with label printers, you can select the label printers with numerals between 1
and 9. The barcode begins with the numeral with which the label printer was
switched on.
In order to generate a single label, press the printer symbol three times and then
press Enter to confirm that you want an extra label. This function works with the
printer switched on or off. With the printer switched off, the barcode begins with the
numeral last used.
Reverse button
• Press the button.
•
The transport mechanism runs in the opposite direction and the
material can easily be removed.
“Damage to the sealer”
See section 2.4, page 3.
Label printing
UM-GS57-V1.03-GB 20
Page 21
Setting the default values
6
These are the default settings of the GS-57 sealer:
•
Date and time to Central European time
• Date sequence day - month – year
• User password (1 – 2 – 3 – 4)
• Standby setting at 50 °C and 5 minutes
This is how to change the default values:
M
Menu
•
1. Password 3. Heating
2. Language 4. Settings
• Press the key on the alphanumeric keypad.
Enter the password (the factory default password is 1234).
•
The display looks like this.
User manual
6.1
Menu
Password:
Changing the password
The functions for making changes are protected by a
password. If text 0 is selected the text can be freely edited.
M
• Press the “1” key on the alphanumeric keypad.
• The display value that can be set blinks. The sealer is in input mode.
• The alphanumeric keypad is used for data input.
You can enter any combination of four digits.
• Enter a new password. If you enter 0, the password option is disable. Press ↵↵↵↵
to confirm.
• Input mode is ended and the sealer is ready to seal.
• Press the key.
•
The display looks like this.
• Press ↵ to confirm.
UM-GS57-V1.03-GB 21
Page 22
User manual
Menu
Menu
6.2
Changing the language
M
•
• Press the key.
Press the “2” key on the alphanumeric keypad.
Language: English
Date: mm – dd - yy
• To select one of the languages stored, press the delete key, then
press ↵↵↵↵ to confirm. The date code also will change, it is possible to choose
three different date codes with every language.
• Input mode is ended and the sealer is ready to seal.
6.3
The sealer has a standby mode. After a set time of five minutes, for instance, the
sealing temperature of the sealer drops automatically. The sealer is in standby mode.
Pressing any key returns the temperature to the set sealing temperature. When the
sealer is switched on the conveyor belt stats every 10 minutes for a few seconds.
Changing heating standby parameters
M
• Press the “3” key on the alphanumeric keypad.
• Press the key.
Standby temp: 50°C
Standby timer: 5 min
•
Press
• The set values appear blinking on the display in sequence.
The sealer is in input mode.
• The alphanumeric keypad is used for data input.
•
Enter new standby temperature, press
confirm.
•
Input mode is ended and the sealer is ready to seal.
↵
↵
to confirm.
↵↵
• The display looks like this.
•
The display looks like this.
↵
↵
to confirm. Enter time, press
↵↵
↵
↵
↵↵
to
Important
►
►
UM-GS57-V1.03-GB 22
To disable the standby function set the time to 0.
The standby timer can be set between 1 and 20 minutes.
Page 23
User manual
6.4
To make it easier to set the sealing temperature, four different temperatures can be
stored.
This is how to change a preset temperature:
Menu
Changing a preset temperature
M
• Press the “4“ key on the alphanumeric keypad.
• Press the key. If necessary, enter a password.
Set 1: 120°C Set 2: 150°C
Set 3: 180°C Set 4: 200°C
• Press ↵↵↵↵ to confirm.
•
The first temperature that can be set blinks.
The sealer is in input mode.
• The alphanumeric keypad is used for data input.
• The temperatures can be changed. Set 1 and press ↵↵↵↵ to confirm. Set 2 and press ↵↵↵↵ to confirm. Set 3 ↵↵↵↵ and press
↵
↵ to confirm.
↵↵
Set 4 ↵↵↵↵ and press ↵↵↵↵ to confirm.
• Input mode is ended and the sealer is ready to seal.
• The display looks like this.
Important
►
The maximum temperature is 200°C, the minimum temperature is 50°C.
The stored temperatures can be selected. (see section 5.10, page 14.)
UM-GS57-V1.03-GB 23
Page 24
User manual
Service functions
7
A service technician can use the service functions of the sealer as described in this
paragraph. The service password should be known by the technician.
7.1
This is how to switch a component on or off:
Menu
Printer 1: 1 Press 1
Printer 2: 0 Label 0
7.2
Switching components on and off
M
•
• Press the “5” key on the alphanumeric keypad.
• Press ↵↵↵↵ to confirm.
• The first temperature that can be set blinks.
•
• Input mode is ended and the sealer is ready to seal.
• Press the key on the keypad.
Enter the service password.
The sealer is in input mode.
Switch components on or off one after another (1 = ON / 0= OFF).
Showing the total counter reading
• The display looks like this.
M
Menu
• Enter the service password.
•
Press the “6” key on the alphanumeric keypad.
Total: 00000565
• The reset function is barred.
• For example, 565 packages have been sealed.
• Press the key on the keypad.
• The display looks like this.
UM-GS57-V1.03-GB 24
Page 25
User manual
Menu
Temp :
180°C
Sens1:
0
7.3
Sens2: 0
The transport sensor (sensor 1) is on the left-hand side of the sealer. The counter
sensor (sensor 2) is on the right-hand side of the sealer. If the sensors no longer
work properly you can order spare parts for the sealer or send the sealer to
GETINGE for repair.
Testing sensors
M
•
Enter the service password.
• Press the “7” key on the alphanumeric keypad.
• Press the key on the keypad.
• The display looks like this.
l
UM-GS57-V1.03-GB 25
Page 26
User manual
Troubleshooting guide
8
State of device Fault Corrective action
Display doesn’t light up
Pressure error on display
”pressure to low”
”Pressure to high”
Temperature error on display Various problems Contact an authorized service
Bad seal result
Crumpled paper Paper crumpled up in device Press reverse and remove
Error warning on display Every possible warning Needs service by an
No connection between
sealer and computer
label printer doesn’t print
In all other cases the device needs service or repair by an authorized technician or a
specialized dealer.
Device not connected Connect device
Power switch not pushed Push power switch
Defect fuse Replace fuse
Defect power cable Replace power cable
Various problems Contact an authorized service
technician
technician
Seal-seam not closed Temperature to cold; set-
temperature according to
specification from the sealbag manufacturer
Seal-seam rippled or burned Temperature to warm; set-
temperature according to
specification from the sealbag manufacturer
Seal-bag fed in upside down Feed in seal-bag on a proper
way
Paper rests in device Press reverse and remove
paper rests
paper
authorized technician or a
specialized dealer
Wrong printer height Adjust printer height Bad printing result
Defect ink ribbon Change printing cassette
RS232 cable or USB cable
not connected
Defect RS232 cable or USB
cable
RS232 cable connected to
the wrong serial port
Wrong settings Change settings
RS232 cable not connected Connect RS232 cable
Defect RS 232 cable Replace RS232 cable
RS232 cable connected to
the wrong serial port
Connect RS232 cable or USB
cable
Replace RS232 cable or USB
cable
Connect RS232 cable in the
right serial port
Connect RS232 cable in the
right serial port
UM-GS57-V1.03-GB 26
Page 27
User manual
Technical data
9
Sealer GS-57
Temperature settings 50 - 200°C
Dimensions w x d x h 665 x 280 x 215 mm
Dimensions with feeder plate w x d x h 755 x 505 x 215 mm
Weight 21 kg approx
Transport speed 10 m/min approx
Seal seam dimensions 4 x 2.5 mm approx
Seal-edge settings 15-35 mm
Max. power consumption 850W
Power supply 100-240V / 50-60Hz
Mains fuse 4A (115V : 8A)
Printing features Date of manufacture, expiry date
alphanumeric personal code and batch code
Text
Symbols to DIN 980
Sealing parameters
Bar code with external printer
Options Second printer
Force measurement
Every GS-57 sealer is set up to produce a seal seam that complies with
DIN 58953-7. Every sealer has a certificate of initial validation.
Always process and dispose faulty devices or parts in an environment-friendly way
or send them back to the manufacturer.