GETINGE GS-57 User Manual

Page 1
Sealer GS-57
User manual
V1.03
Page 2
Contents
1 General guidelines .............................................................................................. 3
2 Safety regulations................................................................................................ 4
2.1 Before use.................................................................................................... 4
2.2 Field of application ....................................................................................... 4
2.3 Danger of electric shock............................................................................... 4
2.4 Damage to the sealer................................................................................... 4
2.5 Contamination of sterile goods..................................................................... 4
2.6 Hot surface................................................................................................... 5
3 Installation ........................................................................................................... 6
4 Overview of controls............................................................................................ 7
5 Operating instructions.......................................................................................... 8
5.1 Switching on the sealer ................................................................................ 8
5.2 Delete Key.................................................................................................... 8
5.3 Return key.................................................................................................... 9
5.4 Enter date of manufacture............................................................................ 9
5.5 Enter expiry date .........................................................................................10
5.6 Text entry ....................................................................................................11
5.6.1 Text selection for the print menu ......................................................12
5.6.2 Text entry with a computer ...............................................................12
5.7 Enter personal ID code................................................................................13
5.8 Enter production code .................................................................................13
5.9 Enter counter reading..................................................................................14
5.10 Set the sealing temperature ........................................................................14
5.11 Pressure measurement ...............................................................................16
5.12 Printer setting ..............................................................................................16
5.12.1 Printing process data........................................................................18
5.12.2 Changing the printing cassette .........................................................19
5.13 Reverse button............................................................................................20
5.14 Label printing...............................................................................................20
6 Setting the default values ...................................................................................21
6.1 Changing the password ..............................................................................21
6.2 Changing the language ...............................................................................22
6.3 Changing heating standby parameters........................................................22
6.4 Changing a preset temperature...................................................................23
7 Service functions ................................................................................................24
7.1 Switching components on and off ...............................................................24
7.2 Showing the total counter reading...............................................................24
7.3 Testing sensors...........................................................................................25
8 Troubleshooting guide........................................................................................26
9 Technical data ....................................................................................................27
10 Wear and replacement parts .............................................................................28
UM-GS57-V1.03-GB 2
Page 3
Useful
information
General guidelines
1
The purpose of these instructions is to explain to the user how to use the GS-57 sealer safely and correctly.
Anyone whose work involves using the sealer must be familiar with these instructions and the safety regulations.
These instructions must always be available to personnel, to prevent incorrect use of the sealer and to ensure trouble-free operation.
Copyright
These instructions may only be given to third parties with the written consent of GETINGE. All documents are protected by copyright.
The passing-on and copying of documents, including extracts and the communication or use for gain of their contents is not permitted, except by express consent. Infringements are punishable and will lead to claims for damages. We reserve all rights to exercise the provisions of commercial copyright law.
Warranty and liability
We accept liability within the terms of the stated warranty obligations for errors and omissions, to the exclusion of further claims.
We reserve the right to make technical changes to the sealer covered by this instruction manual, for the purpose of further development.
We accept no liability for damage and operational disturbances arising from incorrect operation, failure to observe these instructions, or improper repairs. Only original spare parts and accessories approved by GETINGE may be used.
GETINGE accepts no liability for damage resulting from the use of non-approved spare parts and accessories.
The drawings and graphics are not to scale.
Explanations
Important instructions given in the technical documentation are indicated as follows:
Warning
UM-GS57-V1.03-GB 3
Page 4
Safety regulations
2
These instructions must be read and strictly complied with.
2.1 Before use
Users must familiarize themselves with the instructions in this manual before using
the GS-57 sealer.
2.2 Field of application
The GS-57 sealer is to be used only for sealing sterilization packages. No liability will be accepted for use of the sealer for any other purpose.
The sealer may only be used for its intended purpose, in accordance with the generally recognized rules of technology and the industrial health and safety regulations. It must not be used if it has defects that might be dangerous to employees or third parties.
The sealer only may be used under supervision.
2.3
Before the case is opened, the sealer must be disconnected from the electric power supply. Open the housing by loosening the 2 housing cover screws (not marked) on the backside of the unit.
Only an authorized technician or a specialized dealer may undertake servicing work on the sealer.
Danger of electric shock
2.4 Damage to the sealer
If the packaging material becomes trapped, it must not be pulled out of the sealer. This would damage the sealer.
The sealer is not suitable for sealing Polyethylene (PE) bags. The sealer may be damaged if an unsuitable material is used.
The reverse button must not used to feed packaging material into the sealer from the left (see section 5.13, page 20).
2.5
If the pressure of the pressure rollers is incorrectly set, the seam will not be bacteria-tight (see section 5.11, page 16).
The user must set the correct sealing temperature for each packaging material and sealer. This is done by trial sealing operations. The usual sealing temperatures are between 120 °C and 200 °C. Our
Contamination of sterile goods
“Danger of patient infection”
UM-GS57-V1.03-GB 4
Page 5
recommendation is to set the sealing temperature to 120 °C and to raise it gradually until a bacteria-tight seam is obtained.
If the temperature is too low, the seam will be incomplete or weak.
If the temperature is too high, the package will be difficult to peel open. Paper
fibers may be released and contaminate the sterile goods.
Released paper fibers may result in patient infection.
2.6
When the printer has been running for some time, the print head will be hot. Avoid touching the print head when changing the ink ribbon. (see Section 5.12.2
page 19).
Hot surface
UM-GS57-V1.03-GB 5
Page 6
Installation
3
Set up the GS-57 sealer on a clean, smooth surface as follows:
Position the sealer as required.
Leave at least 15 cm free space behind the sealer to ensure good cooling and trouble-free operation.
Remove the printer blockade by lifting the sealers cover and cut the ty-rap connected on the printer.
Plug the mains cable into the socket on the back of the sealer.
Check that the voltage and frequency of the supply are 230Vac 60Hz. If the voltage and frequency of the supply are 115Vac 50Hz replace the two 4A slow fuses inside the power switch by two 8A slow fuses. These fuses are delivered with the sealer.
Plug the mains lead into the power socket.
The GS-57 sealer is now installed.
Note that the printers can only print if the print seam distance is large enough.
Paper guide accessories
GETINGE recommend the use of a paper guide for convenient feeding and trouble­free transport of the material.
Install the paper guide on the sealers left side.
To set the seam width, slacken the knurled nut from the paper guide. Move the paper guide to the required position and tighten the knurled nut.
With the feed stop fully forward, the seam is about 15 mm away from the cut­off edge of the material; with the feed plate fully back it is about 35 mm away from it.
Feeder plate accessory (Optional)
GETINGE recommend the use of a feeder plate for convenient feeding and trouble­free transport of the material. The feeder plate is an optional accessory for the sealer. The feeder plate is easily to mount by hanging it above the sealers front plate.
UM-GS57-V1.03-GB 6
Page 7
Overview of controls
4
1 Date of manufacture
2 Force measurement
3 Printer setting
4 Display
5 Personal ID code
6 Counter reading
7 Delete key in:
- input mode Select key for:
- menu language
- Expiry date, week, month, year
8 Return key
9 Text selection and input up to 100 texts
10 Batch code
11 Menu
12 Print width setting
13 Alphanumeric keypad 0-9 or A-Z
14 Reserve button for sealing material
15 Temperature setting
- Select temperature
16 Expiry date
UM-GS57-V1.03-GB 7
Page 8
Operating instructions
5
The GS-57 sealer seals all sterilization packages that are normally commercially available. The sealer is not suitable for sealing PE bags.
Damage to the sealer
See section2.4, page 4.
Important
After every complete entry in the chosen mode (text entry, text selection,
The return key must be pressed after every function key or data entry, as confirmation. In the instructions below, the return key is represented by this symbol: ↵↵↵.
temperature entry by user or process data), the sealer is ready to seal although the chosen menu is still displayed. “Ready to seal” only appears on the display when the material to be seal is fed in.
5.1
Set the green switch on the back of the sealer to (I).
The following information appears on the display:
Heating up Temp: 20°C
When the sealing temperature has been reached, the following information appears on the display:
Ready to seal Temp: 180°C Press: 65 N
5.2
Switching on the sealer
Delete Key
In input mode
In select mode
Press: 65 N
the delete key:
Is used to delete characters one by one.
the delete key is used:
The sealer heats up to the programmed temperature in about 30 seconds.
The sealer is ready to seal.
Subsequent settings can now be done.
In “Expiry date” mode to change week, month and year.
In
“Menu”
In
“Temperature”
mode to change the language.
mode to select the temperature.
UM-GS57-V1.03-GB 8
Page 9
5.3
Return key
Every entry must be confirmed with the return key.
5.4 Enter date of manufacture
The date of manufacture is entered only once, when the sealer is used for the first time.
Press the button.
Time: 14:00 Date: 07-18-06
Press
Enter the password (the factory default password is 1234). Time and date can be set.
The set values appear blinking on the display in sequence. The sealer is in input mode.
The alphanumeric keypad is used for data input.
Hour ↵↵↵↵ (confirm), minutes ↵↵↵↵ (confirm). Day ↵↵↵ (confirm), month ↵↵↵ (confirm), year ↵↵↵ (confirm).
Input mode is ended and the sealer is ready to seal.
↵↵↵↵
to confirm.
The display looks like this.
Important
If you make a mistake, it’s possible to delete single characters with the delete key . This can only be done if the input has not been confirmed. If the input has been confirmed, the procedure must be repeated.
UM-GS57-V1.03-GB 9
Page 10
Month
Day
Year
5.5
You can enter any numerical value (period) from 0 to 99. Only when days are set it is possible to enter any numerical value (period) from 0 to 999. The display shows the previously set expiry date, eg days, weeks, month or year.
Enter expiry date
Press the key.
Period: 10 Days Expire: 03-20-07
Press ↵↵↵↵ to confirm.
The display value (period) that can be set, blinks. The sealer is in input mode.
The alphanumeric keypad is used for data input.
Enter a new period and press ↵↵↵↵ to confirm.
Press the delete key and enter the calculation period, eg days, weeks,
month or year, the press
Input mode is ended and the sealer is ready to seal.
Example calculation: Date: 03.20.06 numerical value of 10 (period: day, weeks, etc.).
Days 03-30-06 Weeks 05-29-06 Month exact 01-20-07 Exact date 1st of month 01-01-07 Rounded to the first of the month 16th of month 01-16-07 Rounded to the sixteenth of the month Year exact 03-20-16 Exact date Year M/1 03-01-16 Rounded to the first of the month Year M/16 03-16-16 Rounded to the sixteenth of the month
The display looks like this.
to confirm.
↵↵
UM-GS57-V1.03-GB 10
Page 11
5.6
You can enter up to 100 texts each of 24 characters. To enter text:
Text entry
Text-choice: 0
Press to confirm.
The text number blinks on the display.
The sealer is in input mode.
Using the alphanumeric keypad, choose the desired text number from 0 to 99, then press ↵↵↵ to confirm.
The text stored under that number (if any) is displayed.
To enter a new text, or to overtype the existing text, press the
If you have previously entered a password or chosen text 0, you can enter the new text with the keypad.
For example, if you want to enter a lower-case using the “<>” key you can select uppercase and lowercase letters, uppercase and lowercase special characters and the relevant numeral. Choose the desired character from the active group of characters. The following sequence of characters appears:
When entry is complete, press
Input mode is ended and the sealer can continue sealing.
Press the key.
Uppercase letters: Lowercase letters: Uppercase special characters: Lowercase special characters: Numeral:
2
ABC abc
to confirm.
↵↵
The display looks like this.
Ä ÅÆÇ деажз
key again.
↵↵
“c”
, press the
“2”
key. By
Important
If you want to enter two letters from the same key, eg “A” and “B”
If you make a mistake, use the delete key ←← to delete the incorrect
UM-GS57-V1.03-GB 11
If you press the key too many times, keep pressing it until the desired character appears.
immediately after each other, wait for a second, so you do not overtype the first letter entered.
character.
To enter a space, press the 1 (SPACE) key once.
Page 12
Text no. 0 can be changed without a password. Texts 1-99 can be
password-protected. If this is done, the texts can be selected but not edited.
5.6.1 Text selection for the print menu
Press the key.
Text-choice: 0
Press ↵↵↵↵ to confirm.
The text number that can be set blinks on the display.
The sealer is in input mode.
Use the alphanumeric keypad to enter the desired text number (0 to 99) and
press
The text is displayed. To print the selected text, see section 5.12, page 16.
When text selection is complete, press ↵↵↵↵ to confirm.
Input mode is ended and the sealer is ready to seal.
5.6.2
To enter texts with the computer, connect the sealer PC socket to the serial port of your computer with the supplied cable or connect the sealer to the computer with the USB cable. Switch on the sealer and the computer. Install the sealer communication software from the supplied CD and follow the attached instructions.
to confirm.
↵↵
Text entry with a computer
The display looks like this.
UM-GS57-V1.03-GB 12
Page 13
5.7
You can enter up to 14 characters on the alphanumeric keypad. The input procedure is as described in Text entry (see section 5.6, page 11.)
Pers. nr.:
5.8
You can enter up to 14 characters on the alphanumeric keypad. The input procedure is as described in Text entry (see section 5.6, page 11.)
Enter personal ID code
Press the key.
The display looks like this.
Press ↵↵↵↵ to confirm.
The display value that can be set blinks. The sealer is in input mode.
Enter personal ID code and press ↵↵↵↵ to confirm.
Input mode is ended and the sealer is ready to seal.
Enter production code
Press the key.
Lot nr: 2785SP
Press ↵ to confirm.
The display value that can be set blinks. The sealer is in input mode.
Enter a new production code and press ↵↵↵↵ to confirm.
Input mode is ended and the sealer is ready to seal.
The display looks like this.
UM-GS57-V1.03-GB 13
Page 14
5.9
If you want to do a certain number of sealing operations, the sealer can be set up as follows.
Counts
Enter counter reading
Press the key.
:
0
x
Maximum: 0
Press ↵ to confirm.
The current setting disappears from the display. The sealer is in input mode.
The counter can be set to any value (eg 0). Press ↵↵↵↵ to confirm.
The desired maximum reading (eg 100) can be set. Press ↵↵↵↵ to confirm.
The setting is saved and the sealer is ready to seal.
Important
Taking the above setting as an example, when you have sealed 100 bags,
the sealer reports “Counter maximum” and accepts no more material when the conveyor belt stops.
The maximum value that can be set is 1000.
5.10
If a sealing operation fails, you can reverse the counter reading by 1.
If you do not want a maximum function, set the maximum value to 0. This disables the function.
Set the sealing temperature
The display looks like this.
You can enter the sealing temperature in two ways.
1. Select temperature freely
This is the procedure for setting the target temperature:
Press the key.
Seal temp: 150°C Set temp: 150°C ± 5°C
Press ↵↵↵↵ to confirm
UM-GS57-V1.03-GB 14
The display looks like this.
Page 15
If necessary, enter the password (the factory default password is 1234).
The current setting disappears from the display. The sealer is in input mode.
The alphanumeric keypad is used to enter the target temperature.
Now enter the desired target temperature for the material you are using and press ↵↵↵ to confirm.
Input mode is ended and the sealer is ready to seal.
The sealer now cools down or heats up.
Important
2. Selecting a preset temperature
The maximum temperature is 200°C, the minimum temperature is 50°C.
Press the key.
Seal temp: 150°C Set temp: 150°C ± 5°C
You can choose between four stored temperatures by pressing the delete key ←←←←
. Then press
Input mode is ended and the sealer is ready to seal.
The sealer now cools down or heats up.
Important
to confirm.
↵↵
The display looks like this.
To store the temperatures
If the set temperature does not correspond to the target temperature, transport does not start. If this happens, wait until the correct temperature is reached.
Contamination of the sterilized goods “Danger of patient infection”
See section 2.5, page 4.
(see section 6.4, page 23.)
UM-GS57-V1.03-GB 15
Page 16
Pressure:
65 N
5.11
Pressure measurement is optional. If pressure measurement is not used, this key does nothing.
Set press.: 65 N ÷
Pressure measurement is set at the factory to 65 N. There is usually no need to change this setting. The pressure settings may only be changed by service personnel.
5.12
The sealer is equipped with one printer (single-line) or two printers (two-line). The sealer can print different data at the same time during the sealing process: This is how to set the print information:
Print: Select >
Pressing the individual symbols activates the following data:
Pressure measurement
Press the key.
÷ 9N
÷ ÷
Printer setting
Press the key.
Press
Press
The sealer is in input mode.
To delete a symbol, press it again.
to confirm.
↵↵
to confirm.
↵↵
Date of manufacture (= Fab)
Date of expiry (= Exp)
Personal ID number (= Pnr)
Batch number (= Lot)
Text selection (= Txt)
The display looks like this.
The display looks like this.
UM-GS57-V1.03-GB 16
Page 17
With one printer, all the data are printed on one line. With two printers, the batch number and the text are printed on the second line. This reduces the line length.
Press ↵↵↵↵ to confirm.
After confirmation, the character width and line length appear automatically on the display.
Symbol width: 1.6mm Print-length: 68mm
Three printing widths are programmed at the factory.
The correct printing width for the packaging material in use can be selected by pressing the <> symbol on the alphanumeric keypad.
Important
Behind every symbol selected for printing there are data that were
The printing width must be matched to the size of the packaging material. The packaging material must be slightly wider than the printing width. The line width is only indicated when at least one printed is switched on.
previously entered and confirmed.
Only the information shown on the display is printed.
When the packaging material is fed in, the paper side must be down.
It’s also possible to print the Barcode on the material. The barcode can be switched on by connecting the sealer to a computer on the serial port. (see programmers manual for further instructions)
If the printer is not used for a long time, we recommend removing the print cassette (see section 5.12.2, page 19) to prevent the ink drying out.
The display looks like this
UM-GS57-V1.03-GB 17
Page 18
5.12.1
You can use the GS 46 sealer to print out the process data for documentation purposes.
This is how to print the process data:
Print Process-report: 03-18-07 / 180°C 65N
Printing process data
Insert a piece of packaging material.
The sealer prints the following process data:
This information is printed on the bag. This can be detached for documentation purposes. When the process data have been printed, sealing can continue.
Press the key twice.
Date of manufacture Temperature (Pressure measurement) Speed Total counter reading
The display looks like this.
UM-GS57-V1.03-GB 18
Page 19
5.12.2
When the print quality deteriorates, the printing cassette must be replaced. A short­term improvement in the print quality can be achieved as follows:
Changing the printing cassette
Hot surface
See section 2.6, page 5.
1. Short-term improvement in print quality
Switch off the sealer.
Open the cover (1).
Move the blue lever (4) to the midway position (4), or if necessary to the upper position (6). The print quality will be improved for a short time.
2. Changing the printing cassette
Switch off the sealer.
Open the cover (1).
Move the blue lever (4) to the lowest position.
Tilts back the printer (2).
Pull out the printing cassette (3) in the direction of the arrow.
Insert a new printing cassette (7) at an angle into the carrier on the ink ribbon holder and move the carrier up until it latches into place.
Leave the blue lever (4) in the lowest position.
Close cover (1). The GS-57 sealer is ready for use.
UM-GS57-V1.03-GB 19
Page 20
5.13
This is what to do if material has become trapped:
5.14
With the label printer, which is available as a accessory, you can create barcode labels. For this, the label printer is connected to the socket marked LABEL on the back of the sealer, on the right, using the cable supplied. Now install the label printer and load the labels and the transfer band, as described in the printer instructions. The printer can print a label for every bag that passes through, or only on request. To print a label for each bag, switch on the label printer, as described in section, switching components on and off (see section 7.1, page 24.). If you have more than one sealer with label printers, you can select the label printers with numerals between 1 and 9. The barcode begins with the numeral with which the label printer was switched on.
In order to generate a single label, press the printer symbol three times and then press Enter to confirm that you want an extra label. This function works with the printer switched on or off. With the printer switched off, the barcode begins with the numeral last used.
Reverse button
Press the button.
The transport mechanism runs in the opposite direction and the material can easily be removed.
“Damage to the sealer”
See section 2.4, page 3.
Label printing
UM-GS57-V1.03-GB 20
Page 21
Setting the default values
6
These are the default settings of the GS-57 sealer:
Date and time to Central European time
Date sequence day - month – year
User password (1 – 2 – 3 – 4)
Standby setting at 50 °C and 5 minutes
This is how to change the default values:
M
Menu
1. Password 3. Heating
2. Language 4. Settings
Press the key on the alphanumeric keypad.
Enter the password (the factory default password is 1234).
The display looks like this.
6.1
Menu
Password:
Changing the password
The functions for making changes are protected by a password. If text 0 is selected the text can be freely edited.
M
Press the “1” key on the alphanumeric keypad.
The display value that can be set blinks. The sealer is in input mode.
The alphanumeric keypad is used for data input.
You can enter any combination of four digits.
Enter a new password. If you enter 0, the password option is disable. Press ↵↵↵ to confirm.
Input mode is ended and the sealer is ready to seal.
Press the key.
The display looks like this.
Press to confirm.
UM-GS57-V1.03-GB 21
Page 22
Menu
Menu
6.2
Changing the language
M
Press the key.
Press the “2” key on the alphanumeric keypad.
Language: English Date: mm – dd - yy
To select one of the languages stored, press the delete key, then press ↵↵↵ to confirm. The date code also will change, it is possible to choose three different date codes with every language.
Input mode is ended and the sealer is ready to seal.
6.3
The sealer has a standby mode. After a set time of five minutes, for instance, the sealing temperature of the sealer drops automatically. The sealer is in standby mode. Pressing any key returns the temperature to the set sealing temperature. When the sealer is switched on the conveyor belt stats every 10 minutes for a few seconds.
Changing heating standby parameters
M
Press the “3” key on the alphanumeric keypad.
Press the key.
Standby temp: 50°C Standby timer: 5 min
Press
The set values appear blinking on the display in sequence. The sealer is in input mode.
The alphanumeric keypad is used for data input.
Enter new standby temperature, press confirm.
Input mode is ended and the sealer is ready to seal.
to confirm.
↵↵
The display looks like this.
The display looks like this.
to confirm. Enter time, press
↵↵
↵↵
to
Important
UM-GS57-V1.03-GB 22
To disable the standby function set the time to 0.
The standby timer can be set between 1 and 20 minutes.
Page 23
6.4
To make it easier to set the sealing temperature, four different temperatures can be stored.
This is how to change a preset temperature:
Menu
Changing a preset temperature
M
Press the “4“ key on the alphanumeric keypad.
Press the key. If necessary, enter a password.
Set 1: 120°C Set 2: 150°C Set 3: 180°C Set 4: 200°C
Press ↵↵↵↵ to confirm.
The first temperature that can be set blinks. The sealer is in input mode.
The alphanumeric keypad is used for data input.
The temperatures can be changed. Set 1 and press ↵↵↵ to confirm. Set 2 and press ↵↵↵ to confirm. Set 3 ↵↵↵ and press
to confirm.
↵↵ Set 4 ↵↵↵ and press ↵↵↵ to confirm.
Input mode is ended and the sealer is ready to seal.
The display looks like this.
Important
The maximum temperature is 200°C, the minimum temperature is 50°C. The stored temperatures can be selected. (see section 5.10, page 14.)
UM-GS57-V1.03-GB 23
Page 24
Service functions
7
A service technician can use the service functions of the sealer as described in this paragraph. The service password should be known by the technician.
7.1
This is how to switch a component on or off:
Menu
Printer 1: 1 Press 1 Printer 2: 0 Label 0
7.2
Switching components on and off
M
Press the “5” key on the alphanumeric keypad.
Press ↵↵↵↵ to confirm.
The first temperature that can be set blinks.
Input mode is ended and the sealer is ready to seal.
Press the key on the keypad.
Enter the service password.
The sealer is in input mode. Switch components on or off one after another (1 = ON / 0= OFF).
Showing the total counter reading
The display looks like this.
M
Menu
Enter the service password.
Press the “6” key on the alphanumeric keypad.
Total: 00000565
The reset function is barred.
For example, 565 packages have been sealed.
Press the key on the keypad.
The display looks like this.
UM-GS57-V1.03-GB 24
Page 25
Menu
Temp :
180°C
Sens1:
0
7.3
Sens2: 0
The transport sensor (sensor 1) is on the left-hand side of the sealer. The counter sensor (sensor 2) is on the right-hand side of the sealer. If the sensors no longer work properly you can order spare parts for the sealer or send the sealer to GETINGE for repair.
Testing sensors
M
Enter the service password.
Press the “7” key on the alphanumeric keypad.
Press the key on the keypad.
The display looks like this.
l
UM-GS57-V1.03-GB 25
Page 26
Troubleshooting guide
8
State of device Fault Corrective action
Display doesn’t light up
Pressure error on display ”pressure to low” ”Pressure to high” Temperature error on display Various problems Contact an authorized service
Bad seal result
Crumpled paper Paper crumpled up in device Press reverse and remove
Error warning on display Every possible warning Needs service by an
No connection between sealer and computer
label printer doesn’t print
In all other cases the device needs service or repair by an authorized technician or a specialized dealer.
Device not connected Connect device Power switch not pushed Push power switch Defect fuse Replace fuse Defect power cable Replace power cable Various problems Contact an authorized service
technician
technician
Seal-seam not closed Temperature to cold; set-
temperature according to specification from the seal­bag manufacturer
Seal-seam rippled or burned Temperature to warm; set-
temperature according to specification from the seal­bag manufacturer
Seal-bag fed in upside down Feed in seal-bag on a proper
way
Paper rests in device Press reverse and remove
paper rests
paper
authorized technician or a
specialized dealer Wrong printer height Adjust printer height Bad printing result Defect ink ribbon Change printing cassette RS232 cable or USB cable not connected Defect RS232 cable or USB cable RS232 cable connected to the wrong serial port Wrong settings Change settings RS232 cable not connected Connect RS232 cable Defect RS 232 cable Replace RS232 cable RS232 cable connected to the wrong serial port
Connect RS232 cable or USB
cable
Replace RS232 cable or USB
cable
Connect RS232 cable in the
right serial port
Connect RS232 cable in the
right serial port
UM-GS57-V1.03-GB 26
Page 27
Technical data
9
Sealer GS-57
Temperature settings 50 - 200°C
Dimensions w x d x h 665 x 280 x 215 mm
Dimensions with feeder plate w x d x h 755 x 505 x 215 mm
Weight 21 kg approx
Transport speed 10 m/min approx
Seal seam dimensions 4 x 2.5 mm approx
Seal-edge settings 15-35 mm
Max. power consumption 850W
Power supply 100-240V / 50-60Hz
Mains fuse 4A (115V : 8A)
Printing features Date of manufacture, expiry date
alphanumeric personal code and batch code Text Symbols to DIN 980 Sealing parameters Bar code with external printer
Options Second printer
Force measurement
Every GS-57 sealer is set up to produce a seal seam that complies with DIN 58953-7. Every sealer has a certificate of initial validation.
Always process and dispose faulty devices or parts in an environment-friendly way or send them back to the manufacturer.
UM-GS57-V1.03-GB 27
Page 28
Wear and replacement parts
10
Fax order to: _________________
Quantity
Description
Ordering no.
From: ___________________________________________________
___________________________________________________
Series no. ___________________________________________________
Signature: ______________ Date: ______________
Other parts are available from GETINGE on request.
PTFE strip for heating element (70 cm) 4427032
PTFE strip for transport rails (130cm) 4427030
Pressure roller, upper, complete 4419016
Pressure roller, lower 4425723
Transport belt (one) 4416005
Toothed belts for pressure roller 4416016
Bearing bush (Ø10 x Ø12) 4415015
Ball bearing 4415000
Ink ribbon
4421001
UM-GS57-V1.03-GB 28
Page 29
Nijverheidsstraat 6, 5961PJ, Horst, NL
Manufactured by: Getinge Infection Control Kleeven Medical BV. PO Box 69, SE-310 44 Getinge, Sweden Phone: +46 35 15 55 00 Phone: +31 (0)77 398 43 06 Fax: +46 35 549 52 Fax: +31 (0)77 398 71 34 info@getinge.com www.getinge.com info@kleeven.nl www.kleeven.com