Getinge VS 60, VS 90, VS 130 Service manual

4 (1)

SERVICE MANUAL

VS 60/90/130

TABLE OF CONTENTS

INTRODUCTION

CONTROL PANEL PERIODIC MAINTENANCE

START TROUBLE SHOOTING COMPONENT CHECK ADVANCED TROUBLE SHOOTING COMPONENT PLACEMENT

MENU NAVIGATION

WIRING DIAGRAM

INSTALLATION

TECHNICAL DATA

FEEDBACK

SPARE PARTS LIST

Rev D

TABLE OF CONTENTS

INTRODUCTION

WARNING!

As the sterilizer works with steam under pressure, some parts of it may become warm. You should therefore exercise the greatest caution whenever working on the sterilizer.

Hot steam may flow out of the sterilizer when the door is opened.

When handling materials and loading the sterilizer, bear in mind that the chamber and the area close to the chamber entrance is very warm.

Liquids must be sterilised in special programs intended for liquids. The sterilizer's liquid program is solely intended for the sterilisation of liquids in open vessels.

G e n e r a l

The VS-series consists of vertical sterilizers with preprogrammed sterilisation programs. The sterilizer works with steam produced in a chamber, fed with water from an external connection.

WARNING TEXTS IN THIS MANUAL

There are 4 different admonitory levels in the manual:

WARNING!

Factor, circumstance or state that can give rise to the risk of personal injury.

CAUTION!

Factor, circumstance or state that can give rise to the risk of damage to machinery or the process.

IMPORTANT!

Clarification that facilitates the understanding of that described in this manual.

TIP!

Factor, circumstance or state that can facilitate execution or the process.

The following warning symbols is placed on the outside of the sterilizer (symbol acc. to ISO 3864)

If national requirements do not accept this symbol, the symbol must be replaced by a suitable symbol valid in the country were the sterilizer is used.

The warning symbol indicates that the person, responsible for the sterilizer must be fully conversant with all requirements mentioned in the documents for the sterilizer.

INTRODUCTION

IMPORTANT!

The warranty on the delivered product is invalidated by faulty installation or an overlooked/ incorrect service interval/maintenance It must be possible to verify service/maintenance.

W a r r a n t y

A 2 year warranty is provided. However, not on those components considered to be consumables, for ex - ample, door gaskets. For service work, reference should be made to agreements with the supplier.

The Product Liability Act only applies when the instructions in this manual have been followed in their entirety.

C o p y r i g h t

Copyright © 2005 Getinge Skärhamn AB. All rights reserved.

All technical documentation provided by Getinge Skärhamn AB is copyright protected and the property of Getinge Skärhamn AB.

This publication may be downloaded from Getinge Skärhamn's official website: www.skarhamn.getinge.com.

The technical documentation is supplied as is. The documentation can contain printing errors, technical errors or other errors. Getinge Skärhamn AB reserves the right to make alterations without prior notice.

No part of this publication may be copied without the expressed permission of Getinge Skärhamn AB.

Manufacturer

Getinge Skärhamn AB Industrivägen 5

SE-47131 Skärhamn Sweden

Disposal

This product consists of (percentage by weight)

76 % Stainless steel

4 % Plastic

9 % Aluminium

5 % Brass

2 % Copper

4 % Electronics

When disposing of the sterilizer, it must be sorted at source in accordance with the requirements set out in the WEEE directive (2002(96/EC).

Contact your dealer when disposing of the sterilizer.

INTRODUCTION

CONTROL PANEL

Layout of the control panel

1.

PROGRAM SELECTOR button –

 

 

6.

Green status lamp – program finished

 

selection of the program

 

 

7.

Red status lamp – alarm

 

 

 

 

 

 

 

 

 

 

 

2.

START button – start of the program

 

 

8.

ENTER button – select items from the menu

 

 

 

 

 

 

 

3.

RESET button – reset a fault

 

 

9.

Arrow keys – navigate through menus/lists

 

 

 

 

 

 

 

4.

STOP button – emergency stop

 

 

10.

Pressure gauge – shows the chamber pressure

 

 

 

 

 

 

 

5.

Yellow status lamp – program in progress

11.

Display unit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

 

 

5

 

 

 

6

 

9

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

7

10

8

Display unit

Shows the current status, program progress, number of minutes after program start, number of completed cycles and error codes.

IN STANDBY MODE SHOWS:

Process name (program)

Sterile time/temperature

Menu

Press the DOWN arrow to CYCLE COUNTER (twice) and then press the ENTER button now press the DOWN arrow to read the pressure and load sensor temperature (when the sterilizer features a liquid program).

SHOWN DURING THE PROCESS:

Process name

Process time

Home position:

Press the LEFT arrow for at least 4 seconds

To specifically read the load sensor temperature (when the sterilizer features a liquid program), press the ENTER button at the process name.

 

 

Operation, check

 

 

 

INCORRECT OPERATION, CHECKLIST

 

 

The table below lists the most common operating

 

 

faults that result in an alarm (error code) or that cer-

 

 

tain symptoms are experienced.

 

 

Check that usage was correct.

 

 

ALARM (ERROR CODE)

REASON

 

 

Close door

Door has not been locked before the start button has been pressed. Read

 

the text below the table, Correct operation, checklist.

 

 

Sterile error

The sterilizer may have been overloaded. Read the Technical data se c-

 

tion.

 

 

Door open

The handle has been activated before the finished signal has been dis-

 

played. Read the text below the table, Correct operation, checklist.

 

 

Process stopped

The stop button has been pressed by the operator.

 

 

Heating error

Can be due to items caught in the door.

 

 

No water

Perhaps the water supply tap has not been opened

 

 

SYMPTOM

REASON

 

 

Process abnormally long

Can be due to:

 

no water on the incoming supply.

 

sterilizer overloaded. Read the Technical data section.

 

leakage.

 

 

 

 

WARNING!

Consider the risk of burns when loading the sterilizer.

CAUTION!

Only sterilise items intended for sterilisation with steam and that can withstand at least 121 degrees.

Only use loading equipment especially designed for the VS series. Each standard basket may be loaded with max. 12 kg of goods. To ensure optimal ergonomics use a lifting device when processing heavy loads.

CORRECT OPERATION, CHECKLIST

1Placing items in the sterilising chamber. Only use loading equipment especially designed for the VS series. Carefully check that nothing is caught between the door and chamber.

Place moisture-sensitive items and low-mass items at the top of the chamber.

Place heavy items at the bottom of the sterilising chamber.

Place dish-shaped items dish side down.

Place empty containers (bottles, test tubes, etc.) with their opening downwards.

2Close the door and pull the handle clockwise to lock.

3Press the START button, the cycle starts and will be completed automatically.

OPERATION, CHECK

Use a receptacle with the smallest possible volume for the fastest processing.
OPERATION, CHECK
Place the load sensor in the largest container. When there is no space in the liquid for the load sensor, use a control receptacle where the load sensor can be placed. The volume of the liquid in the control receptacle should be as great or greater than the liquid intended for sterilisation. Boiling point for the liquid in the control vessel must be the same as for the liquid to be sterilised.
Bottles should be filled to max. 80 % of their volume.
A special liquid program and a load sensor in the chamber are required to sterilise liquids. The liquid process is solely intended for liquid sterilisation in open receptacles.
00.05.00_1340C
LIQUID STERILISATION

Recommended accessories

Basket, part no.

566687401

Solid, stainless steel container 48325031

WARNING!

Remember the goods are hot after sterilising.

Never attempt to open the door until the pressure gauge has returned completely to zero.

4When the display shows F I N I S H, the green status lamp comes on and when the pressure gauge shows zero the process is completed.

5The display never shows F I N I S H and the green status lamp for process finished never comes on if the process is interrupted.

6As soon as the door is opened F I N I S H is no longer shown on the display, this provides information that if F I N I S H is not shown on the display, then the goods in the chamber may have been loaded by “someone else”, but not sterilised.

7If the red status lamp for an alarm is on after a sterilising process, the goods should be considered as unsterilised and must then be sterilised again.

When the door is opened F I N I S H is no longer displayed. The display now shows:

INSTRUMENT 1340C (last selected program)

IMPORTANT!

If the process must be cancelled for one reason or another press the STOP button; the red status lamp will come on and an audio signal with sound.

Now press the RESET button and wait for the pressure gauge to drop to zero.

When the red status lamp indicating an alarm has gone out, press the RESET button.

Open the door.

IMPORTANT!

When a process is interrupted, there is a risk of steam streaming out of the chamber when the door is opened.

INTERRUPTED PROCESS

If the sterilizer stops the process:

1Press the RESET button; the audio signal stops.

2Wait until the pressure indication shows zero.

3When the red status lamp has gone out:

4Press the RESET button; the alarm text is no longer displayed

If a process in progress has to be interrupted:

1Press the STOP button; the text PROCESS STOPPED is shown on the display. The audio signal sounds and the red status light comes on.

2Press the RESET button; the audio signal stops.

3Wait until the pressure indication shows zero.

4When the red status lamp has gone out, press the RESET button; the alarm text is no longer displayed.

OPERATION, CHECK

PERIODIC MAINTENANCE

G e n e r a l

WARNING!

The outer casing may only be removed by technicians with documented experience.

WARNING!

Bear in mind when cleaning that the chamber and the area close to the chamber entrance is very warm.

The manufacturer recommends at full utilization of capacity that at least the following maintenance operations should be performed at the specified intervals with by service technicians. Monthly and weekly maintenance is carried out by the operator as set out in the user manual.

TIP!

The following Service kits are available:

Disks for solenoid valves

Grease for door gaskets

WARNING!

The electrical power supply must always be disconnected with all types of maintenance work.

A n n u a l m a i n t e n a n c e

ANNUALLY OR AT LEAST EVERY 400 CYCLES UNLESS OTHERWISE STATED.

Replace the filter where appropriate (accessory). Check that the filter is not wet.

Replace the plunger and O-ring on MV-1,MV-2, MV-6 and MV-8. Refer to the section Component check, 3. Solenoid valves.

Check the door and its flange. Lubricate the moving parts by the door gasket. Use a grease suitable for use in the food industry.

Replace the door gasket. At least every 3rd year or 1200 cycles. See the section Component check, 20. Door, Door gasket.

Measure the pressure using a calibrated absolute test pressure gauge, use the intended flange marked VT on the sterilizer. See the section

Component check, 12. Absolute pressure se n- sor.

Inspect and rectify any leakage from pipe couplings and instrument connections. See the section Advanced trouble shooting, Trouble shooting leakage.

Check and clean the water outlet in the chamber. At least every 800 cycles.

Check that the monthly and daily maintenance have been carried out correctly.

PERIODIC MAINTENANCE

Perform a functional check of the electrical components. See the section Component check. At least every 1200 cycles or every 3rd year.

Reset the service alarm from the Service messages menu. (situated under SETUP/SYSTEM/SERVICE in the menu tree)

SAFETY CHECKS

Check that the door seals against the steam pressure during the sterilising process. See the section Component check, 20. Door.

Check that the sterilizer can not be started when the door has not been closed. See the section

Component check, 20. Door.

Check that the door's counter-hold spring works and prevents crush injuries. See the section

Component check, 20. Door, check the door spring

Inform the customer about the significance of the warning signs.

Inform the customer of the risks involved with the program for sterilising liquids, if installed. Particularly the information that it is only permitted to sterilise liquids inopen vessels.

PERIODIC MAINTENANCE

Getinge VS 60, VS 90, VS 130 Service manual

Fig. 1

Fig. 2

M o n t h l y m a i n t e n a n c e

MONTHLY, OR AS REQUIRED

Clean the bottom filter. See the section Component check, 1. Chamber.

Clean the chamber. See the section Component check, 1. Chamber.

Clean the door gasket and lubricate with the grease recommended by Getinge.

D a i l y m a i n t e n a n c e

Check that the overheating protection sensor is fitted on the heating element, see fig 1. See the section Component check, 6. Overheating protection.

Check that the level guard in the bottom of the chamber is clean, fig. 2.

See the section Component check, 1. Chamber.

PERIODIC MAINTENANCE

START TROUBLE SHOOTING

T r o u b l e s h o o t i n g o r i e n t a t i o n

4 Final

1 Start

verification

Choose the

 

error code

 

symptom

3 Verify the

2 Action

action

 

IMPORTANT!

The actions stated in the tables are usually specific checks on the components. However, the actions can also represent advanced trouble shooting and indicate the specific component to be checked.

IMPORTANT!

After the action (for example, component replacement) the action must be verified (e.g. manual activation).

In addition, it must be verified that the error code/customer's experienced symptom has been corrected.

Several actions may be needed with subsequent action verification, remember to always end with verification against the error code/symptom.

START TROUBLE SHOOTING HERE

Choose the error code/codes displayed and the symptom experienced in the table below. Carry out the prescribed actions in the stated order, until the fault has been identified and corrected.

CHECK PROCEDURE

1START

Select the alarm (error code) or symptom in the tables below.

2ACTION

Carry out the actions set out in the table. Carry out the actions in the stated order, continue until the fault has been corrected.

a.Check the components. In the event of a component fault, repair or replace it. Instructions are enclosed with the spare part or are set out after each check.

b.Should the check indicate that the component is OK, return to the table in this section and continue with next action as per point 2.

3VERIFY ACTION

After repair/replacement the action must be verified. Instructions concerning verification are enclosed with the spare part or are set out after each check.

4FINAL VERIFICATION

(ERROR CODE/SYMPTOM VERIFICATION)

After action verification it must be verified that the displayed error code is not displayed again and the experienced symptom no longer remains. This is done by running a program and checking that no error codes are displayed or the symptoms are experienced.

START TROUBLE SHOOTING

A l arm (error codes visible on the display)

EXPLANATIONS

Explanations to references to sections stated in the tables below:

CC :

chapter Component check

Advanced

chapter Advanced trouble

trouble shooting:

shooting

Other references are given in plain text.

ERROR CODE

EXPLANATION

ACTIONS IN ORDER OF PRIORITY

 

 

 

 

GETINGE PACS 300

Process does not start, frozen proc-

1.

CC, 19. Software PACS cold start

GETINGE PACS 300 DIS-

ess

2.

CC, 18. Display

PLAY

 

3.

CC, 17. Circuit board

 

 

 

 

CLOSE THE DOOR

Door unlocked at point of start

1.

Operation, check

 

 

2.

CC, 20. Door, (micro-switch)

 

 

3.

CC, 17. Circuit board

 

 

 

 

DOOR OPEN

Door open during process

1.

Operation, check

 

 

2.

CC, 20. Door, (micro-switch)

 

 

3.

CC, 17. Circuit board

 

 

 

 

POWER FAILURE

Power failure during process

1.

Check external fuses.

 

 

2.

CC, 6. Overheating protection

 

 

3.

CC, 17. Circuit board (Fuse)

 

 

4.

CC, 17. Circuit board

 

 

 

 

PT-100 ERROR

Error on load sensor during process

1.

CC, 15. Load temperature sensor

 

 

2.

CC, 17. Circuit board AI

 

 

 

 

SERVICE

Service indicator.

1.

Contact the service technician

 

 

 

 

STERILE ERROR

Incorrect pressure/steam tempera-

1.

Operation, check

 

ture in the sterile phase

2.

CC, 15. Load temperature sensor

 

 

3.

CC, 12. Absolute pressure sensor,

 

 

4.

CC, 17. Circuit board

 

 

 

 

PROCESS STOPPED

The process has been stopped with

1.

Operation, check

 

the STOP button

2.

CC, 18. Display

 

 

 

 

PRESSURE SENSOR

Pressure sensor incorrectly cali-

1.

Advanced trouble shooting, Checks

ERROR

brated

 

with error code

 

 

 

 

UPDATE

Battery replacement or circuit board

1.

CC, 17. Circuit board

 

fault

 

 

 

 

 

 

START TROUBLE SHOOTING

ERROR CODE

EXPLANATION

ACTIONS IN ORDER OF PRIORITY

 

 

 

 

NO WATER

Timeout level guard

1.

Operation, check

 

 

2.

CC, 10. Level guard

 

 

3.

CC, 3. Solenoid valve MV-3

 

 

 

 

HEATING ERROR

Timeout heater

1.

CC, 6. Overheating protection

 

 

2.

CC, 5. Relay

 

 

3.

CC, 4. Heater

 

 

4.

CC, 17. Circuit board

 

 

 

HEATING ERROR

Leakage

Trouble shoot according to the section

 

 

Trouble shooting leakage

 

 

 

 

MANUAL MODE

Component manually activated

1.

Operation, check

 

 

2.

An output on the circuit board has

 

 

 

been set to manual mode. ( menu

 

 

 

tree Diagnostic)

 

 

 

POST TREATMENT ER-

Outlet valve clogged

CC, 3. Solenoid valve, MV-1

ROR

 

 

 

POST TREATMENT ER-

Screen in the chamber clogged

CC, 3. Strainer

ROR

 

 

 

 

 

 

 

START TROUBLE SHOOTING

S y m p t o m ( e x p e r i e n c e d b y u s e r )

EXPLANATIONS

Explanations to references to sections stated in the tables below:

CC:

chapter Component check

Advanced

chapter Advanced trouble

trouble shooting:

shooting

Other references are given in plain text.

SYMPTOM

ACTIONS IN ORDER OF PRIORITY

 

 

 

No ready signal

1.

CC, 12. Absolute pressure sensor

 

2.

CC, 3. Solenoid valve MV-2

 

3.

CC, 11. Air filter

 

 

 

Instruments rust in the sterilizer

1.

Poor quality instruments

 

2.

Water cleaning filter – type de-ionization, check

 

 

(see documentation supplied with the water fil-

 

 

ter)

 

 

 

Process time abnormally long

1.

See the section installation. Electricity, Water

 

 

requirement

 

2.

Advanced trouble shooting, Leakage test, check

 

3.

Operation, check

 

4.

CC, 22. Strainer

 

 

 

Humming noise in standby mode

1.

CC, 3. Solenoid valves

 

 

 

Door hard to open/cannot be opened

1.

CC, 3. Solenoid valve MV-8

 

2.

CC, 20. Door, (lock)

 

 

 

Pressure gauge shows pressure with the door open

1.

CC, 14. Pressure gauge

 

 

 

Steam leaks from the door

1.

CC, 20. Door, door gasket

 

2.

CC, 3. Solenoid valve MV-5

 

3.

CC, 21 Pressure relief valve door gasket

 

 

 

Black background on the display

1.

CC, 23 Fan.

 

2.

CC, 18 Display

 

 

 

Unstable measurement value and text on display

1.

CC, 23 Fan.

 

2.

CC, 18 Display

 

3.

CC 17 Circuit board

 

4.

Check the power supply (Fluctuations?)

 

 

 

START TROUBLE SHOOTING

COMPONENT CHECK

TABLE OF CONTENTS

 

1.

General ............................................................

2

2.

Chamber ..........................................................

2

3.

Pipe/pipe couplings ..........................................

2

4.

Solenoid valves ................................................

3

5.

Heater...............................................................

5

6.

Relay ................................................................

6

7.

Overheating protection, resetting.....................

6

8.

Non return valve BV-1......................................

7

9.

Non return valve BV-2......................................

7

10.

Non return valve BV-3......................................

7

10.

Non return valve BV-3......................................

7

11.

Level guard ......................................................

8

12.

Air filter (accessory) .........................................

8

13.

Absolute pressure sensor ................................

8

14.

Ejector ..............................................................

9

15.

Pressure gauge................................................

9

16.

Load temperature sensor (accessory) ...........

10

17.

Pressure relief valve.......................................

10

18.

Circuit board...................................................

11

19.

Display............................................................

12

20.

PACS software...............................................

12

21.

Door................................................................

14

22.

Pressure relief valve, door gasket..................

16

23.

Screen............................................................

16

24.

Fan .................................................................

16

COMPONENT CHECK page 1

IMPORTANT!

Only use Teflon tape that is intended for sealing pipe couplings against steam pressure.

TIP!

Check whether steam is let out with an overpressure.

Use soapy water on cold components

Use a mirror on hot components

1. General

It is extremely important for safe functionality that all nipples are tightened so that air, which can jeopardize the sterilisation result, can not enter the chamber. If pipe couplings must be removed while performing checks, think about the following when assembling:

Use ample Teflon tape, so that the tape partly forms a seal and partly seals the threads.

2. Chamber

The chamber walls are cleaned using a mild lime dissolving agent. Rinse well. Drain by closing the door, the bottom valve then opens automatically.

BOTTOM FILTER (ALSO SEE 23. SCREEN)

Lift out the filter. Clean

LEVEL GUARD (ALSO SEE 11. LEVEL GUARD)

Check that the rocker moves up and down easily.

3. Pipe/pipe couplings

Check:

Start a 134 °C program. Trouble shoot when the sterilizer is in “Sterilisation” mode.

Check whether steam is let out.

Action:

Check according to respective component checks. Also see 1. General.

Verification:

Check all pipes/pipe couplings again.

COMPONENT CHECK page 2

IMPORTANT!

All valves open in an energised state.

4. Solenoid valves

Pay attention to solenoid valves that hum! The noise may be a forewarning about the coil starting to overheat due to it not working in a closed iron circuit. This can occur when the moving core is influenced by the magnetic forces from the winding and dirt prevents the coil from reaching the end position. An air gap is then formed in the magnetic circuit, which lowers the inductance in the coil resulting in increased current.

MV-1

Is fitted in the bottom of the sterilizer and has the task of evacuating any remaining water in the sterilizer at the end of the process.

Check:

Check that both the plunger and the area it runs in are clean. Always change the O-ring that seals between the coil and housing after removing the coil. Activate MV-3 and fill the chamber. Activate MV-1 and MV4 (ejector). Check that the water runs out of the chamber again. Measure the voltage across the coil.

Action:

Clean and replace the plunger and O-ring. Replace the entire valve if necessary.

Verification:

Check that the water level in the chamber is maintained for a long period (about 15 minutes).

MV-2

Is fitted at the rear, top edge of the sterilizer. Takes care of pressure equalization after the post-vacuum phase by releasing air into the chamber. The valve is open when the sterilizer is in standby mode if the door is locked. It becomes de-energised as soon as the door is opened.

Check:

Check that both the plunger and the area it runs in are clean. Always change the O-ring that seals between the coil and housing after removing the coil. Activate MV-4 and MV-2 on the diagnostic menu. The pressure should not drop appreciably below atmospheric presssure. Measure the voltage across the coil.

Check for leakage: Dismantle BV-2. Activate MV-3 and the heater. No steam should come out of MV-2.

Action:

Clean and replace the plunger and O-ring. Replace the entire valve if necessary.

Verification:

Run a sterilisation program. Check for leakage acCOMPONENT CHECK

page 3

Loading...
+ 39 hidden pages