Getinge VS 60, VS 90, VS 130 Service manual

SERVICE MANUAL
VS 60/90/130
TABLE OF CONTENTS
INTRODUCTION CONTROL PANEL PERIODIC MAINTENANCE START TROUBLE SHOOTING COMPONENT CHECK ADVANCED TROUBLE SHO OTING COMPONENT PLACEMENT MENU NAVIGATION WIRING DIAGRAM INSTALLATION TECHNICAL DATA FEEDBACK SPARE PARTS LIST
Rev D TABLE OF CONTENTS
INTRODUCTION
WARNING!
As the sterilizer works with steam under pressure, some parts of it may become warm. You should therefore exercise the greatest caution whenever working on the sterilizer.
Hot steam may flow out of the sterilizer when the door is ope­ned.
When handling materials and loading the sterilizer, bear in mind that the chamber and the area close to the chamber en­trance is very warm.
Liquids must be sterilised in special programs intended for liquids. The sterilizer's liquid program is solely intended for the sterilisation of liquids in open vessels.
General
The VS-series consists of vertical sterilizers with pre­programmed sterilisation programs. The sterilizer works with steam produced in a chamber, fed with wa­ter from an external connection.
WARNING TEXTS IN THI S MANUAL
There are 4 di fferent admonitory levels in the manual:
WARNING! Factor, circumstance or state that can give rise to the risk of personal injury.
CAUTION! Factor, circumstance or state that can give rise to the risk of damage to machinery or the process.
IMPORTANT! Clarification that facilitates the understanding of that described in this manual.
TIP! Factor, circumstance or state that can facilitate execution or the process.
The following warning symbols is placed on the out­side of the sterilizer (symbol acc. to ISO 3864)
If national requirements do not accept this symbol, the symbol must be replaced by a suitable symbol valid in the country were the sterilizer is used.
The warning symbol indicates that the person, re­sponsible for the sterilizer must be fully conversant with all requirements mentioned in the documents for the sterilizer.
INTRODUCTION
Warranty
IMPORTANT!
The warranty on the delivered product is invalidated by faulty installation or an overlooked/ incorrect service interval/mainten­ance It must be possible to verify service/maintenance.
A 2 year warranty is provided. However, not on those components considered to be consumables, for ex­ample, door gaskets. For service work, reference should be made to agreements with the supplier.
The Product Liability Act only applies when the in­structions in this manual have been followed in their entirety.
Copyright
Copyright © 2005 Getinge Skärhamn AB. All rights re­served.
All technical documentation provided by Getinge Skärhamn AB is copyright protected and the property of Getinge Skärhamn AB.
This publication may be downloaded from Getinge Skärhamn's official website: www.skarhamn.getinge.com.
The technical documentation is supplied as is. The documentation can contain printing errors, technical errors or other errors. Getinge Skärhamn AB reserves the right to make alterations without prior notice.
No part of this publication may be copied without the expressed permission of Getinge Skärhamn AB.
Manufacturer
Getinge Skärhamn AB Industrivägen 5 SE-47131 Skärhamn Sweden
Disposal
This product consists of (percentage by weight) 76 % Stainless steel
4 % Plastic 9 % Aluminium 5 % Brass 2 % Copper 4 % Electronics
When disposing of the sterilizer, it must be sorted at source in accordance with the requirements set out in the WEEE directive (2002(96/EC). Contact your dealer when disposing of the sterilizer.
INTRODUCTION
CONTROL PANEL
2 3 4 5 6 7 8 9 10 11
Layout of the control panel
1. PROGRAM SELECTOR button – selection of the program
2. START button – start of the program
3. RESET button – reset a fault
4. STOP button – emergency stop
5. Yellow status lamp – program in progress
6. Green status lamp – program finished
7. Red status lamp – alarm
8. ENTER button – select items from the menu
9. Arrow keys – navigate through menus/lists
10. Pressure gauge – shows the chamber pressure
11. Display unit
1
Display unit
Shows the current status, program progres s, number of minutes after program start, number of completed cycles and error codes.
IN STANDBY MODE SHOW S:
Process name (program)
Sterile time/temperature
Menu
Press the DOWN arrow to CYCLE COUNTER (twice) and then press the ENTER button now press the DOWN arrow to read the pressure and load sen­sor temperature (when the sterilizer features a liquid program).
SHOWN DURING THE PRO CESS:
Process name
Process time
Home position: Press the LEFT arrow for at least 4 seconds
To specifically read the load sensor temperature (when the sterilizer features a liquid program), press the ENTER button at the process name.
Operation, check
as not been locked before the start button has been pressed. Read
INCORRECT OPERATION, CHECKLIST
The table below lists the most common operating faults that result in an alarm (error code) or that cer­tain symptoms are experienced.
Check that usage was correct.
ALARM (ERROR CODE) REASON
Close door Door h
the text below the table, Correct operation, checklist .
Sterile error The sterilizer may have been overloaded. Read the Technical data se c-
tion.
Door open The handle has been activated befo re the finished signal has been dis-
played. Read the text below the table, Correct operation, checklist . Process stopped The stop button has been pressed by the operator. Heating error Can be due to items caught in the door. No water Perhaps the water supply tap has not been opened
SYMPTOM REASON
Process abnormally long Can be due to:
no water on the incoming supply.
sterilizer overloaded. Read the Technical data section.
leakage.
WARNING!
Consider the risk of burns when loading the sterilizer.
CAUTION!
Only sterilise items intended for sterilisation with steam and that can withstand at least 121 de­grees. Only use loading equipment espe­cially designed for the VS series. Each standard basket may be loaded with max. 12 kg of goods. To ensure optimal ergonomics use a lifting device when processing heavy loads.
CORRECT OPERATION, CHECKLIST
1 Placing items in the sterilising chamber. Only use
loading equipment especially designed for the VS series. Carefully check that nothing is caught
between the door and chamber. Place moisture -sensitive items and low-mass
items at the top of the chamber.
Place heavy items at the bottom of the sterilis-
ing chamber. Place dish-shaped items dish side down. Place empty containers (bottles, test tubes,
etc.) with their opening downwards.
2 Close the door and pull the handle clockwise to
lock.
3 Press the START button, the cycle starts and will
be completed automatically.
OPERATION, CHECK
Recommended accessories
Basket, part no. 566687401 Solid, stainless steel container 48325031
4 When the display shows FINISH, the green status
lamp comes on and when the pressure gauge shows zero the process is completed.
5 The display never shows FINISH and the green
status lamp for process finished never comes on if the process is interrupted.
6 As soon as the door is opened FINISH is no
longer shown on the display, this provides infor­mation that if FINISH is not shown on the display, then the goods in the chamber may have been loaded by “someone else”, but not sterilised.
7 If the red status lamp for an alarm is on after a
sterilising process, the goods should be consid­ered as unsterilised and must then be sterilised again.
When the door is opened FINISH is no longer dis-
played. The display now shows:
INSTRUMENT 1340C (last selected program)
00.05.00_1340C
WARNING!
Remember the goods are hot after sterilising. Never attempt to open the door until the pressure gauge has re­turned completely to zero.
LIQUID STERILISATION
A special liquid program and a load sensor in the chamber are required to sterilise liquids. The liq­uid process is solely intended for liquid sterilisa­tion in open receptacles.
Bottles should be filled to max. 80 % of their vol- ume.
Place the load sensor in the largest container. When there is no space in the liquid for the load sensor, use a control receptacle where the load sensor can be placed. The volume of the liquid in the control receptacle should be as great or grea­ter than the liquid intended for sterilisation. Boiling point for the liquid in the control vessel must be the same as for the liquid to be sterilised.
Use a receptacle with the smallest possible vol- ume for the fastest processing.
OPERATION, CHECK
IMPORTANT!
If the process must be cancelled for one reason or another press the STOP button; the red status lamp will come on and an audio signal with sound. Now press the RESET button and wait for the pressure gauge to drop to zero. When the red status lamp indicat­ing an alarm has gone out, press the RESET button. Open the door.
IMPORTANT!
When a process is interrupted, there is a risk of steam streaming out of the chamber when the door is opened.
INTERRUPTED PROCESS
If the sterilizer stops the process: 1 Press the RESET button; the audio signal stops. 2 Wait until the pressure indication shows zero. 3 When the red status lamp has gone out: 4 Press the RESET button; the alarm text is no
longer displayed
If a process in progress has to be interrupted: 1 Press the STOP button; the text PROCESS
STOPPED is shown on the display. The audio sig-
nal sounds and the red status light comes on.
2 Press the RESET button; the audio signal stops. 3 Wait until the pressure indication shows zero. 4 When the red status lamp has gone out, press the
RESET button; the alarm text is no longer dis­played.
OPERATION, CHECK
PERIODIC MAINTENANCE
General
WARNING!
The outer casing may only be removed by technicians with do­cumented experience.
WARNING!
Bear in mind when cleaning that the chamber and the area close to the chamber entrance is very warm.
TIP!
The following Service kits are available:
Disks for solenoid valves
Grease for door gaskets
WARNING!
The electrical power supply must always be disconnected with all types of maintenance work.
The manufacturer recommends at full utilization of ca­pacity that at least the following maintenance opera­tions should be performed at the specified intervals with by service technicians. Monthly and weekly main­tenance is carried out by the operator as set out in the user manual.
Annual maintenance
ANNUALLY OR AT LEAST EVERY 400 CYCLES UNLESS OTHERWISE STATED.
Replace the filter where appropriate (accessory). Check that the filter is not wet.
Replace the plunger and O-ring on MV-1,MV-2, MV-6 and MV-8. Refer to the section Component check, 3. Solenoid valves.
Check the door and its flange. Lubricate the mov- ing parts by the door gasket. Use a grease suit­able for use in the food industry.
Replace the door gasket. At least every 3rd year or 1200 cycles. See the section Component check, 20. Door, Door gasket.
Measure the pressure using a calibrated absolute test pressure gauge, use the intended flange marked VT on the sterilizer. See the section
Component check, 12. Absolute pressure sen­sor.
Inspect and rectify any leakage from pipe cou- plings and instrument connections. See the sec­tion Advanced trouble shooting, Trouble shooting leakage .
Check and clean the water outlet in the chamber. At least every 800 cycles.
Check that the monthly and daily maintenance have been carried out correctly.
PERIODIC MAINTENANCE
Perform a functional check of the electrical com- ponents. See the section Component check. At least every 1200 cycles or every 3rd year.
Reset the service alarm from the Service mes- sages men u. (situated under SE-
TUP/SYSTEM/SERVICE in the menu tree)
SAFETY CHECKS
Check that the door seals against the steam pres- sure during the sterilising process. See the sec­tion Component check, 20. Door.
Check that the sterilizer can not be started when the door has not been closed. See the section Component check, 20. Door.
Check that the door's counter-hold spring works and prevents crush injuries. See the section
Component check, 20. Door, check the door spring
Inform the customer about the significance of the warning signs.
Inform the customer of the risks involved with the program for sterilising liquids, if installed. Particu­larly the information that it is only permitted to sterilise liquids inopen vessels.
PERIODIC MAINTENANCE
Monthly maintenance
MONTHLY, OR AS REQUI RED
Clean the bottom filter. See the section Compo­nent check, 1. Chamber.
Clean the chamber. See the section Component check, 1. Chamber.
Clean the door gasket and lubricate with the gre- ase recommended by Getinge.
Daily maintenance
Fig. 1
Check that the overheating protection sensor is fi t- ted on the heating element, see fig 1. See the section Component check, 6. Overheating pro- tection.
Check that the level guard in the bottom of the chamber is clean, fig. 2. See the section Component check, 1. Chamber.
Fig. 2
PERIODIC MAINTENANCE
START TROUBLE SHOOTING
4 Final
3 Verify the
1 Start
Trouble shooting orientation
START TROUBLE SHOOTING HERE
Choose the error code/codes displayed and the symp­tom experienced in the table below. Carry out the pr e-
verification
Choose the
error code
symptom
scribed actions in the stated order, until the fault has been identifie d and corrected.
CHECK PROCEDURE 1 START
Select the alarm (error code) or symptom in the tables below.
2 Action
action
IMPORTANT!
The actions stated in the tables are usually specific checks on the components. However, the actions can also represent advanced trouble shooting and indicate the specific component to be checked.
IMPORTANT!
After the action (for example, component replacement) the ac­tion must be verified (e.g. man­ual activation). In addition, it must be verified that the error code/customer's experi­enced symptom has been cor­rected. Several actions may be needed with subsequent action verifica­tion, remember to always end
with verification against the error code/symptom.
2 ACTION
Carry out the actions set out in the table. Carry out the actions in the stated order, continue until the fault has been corrected.
a. Check the components. In the event of a
component fault, repair or replace it. Instruc­tions are enclosed with the spare part or are set out after each check.
b. Should the check indicate that the component
is OK, return to the table in this section and continue with next action as per point 2.
3 VERIFY ACTION
After repair/replacement the action must be veri­fied. Instructions concerning verification are en­closed with the spare part or are set out after each check.
4 FINAL VERIFICATION
(ERROR CODE/SYMPTOM VERIFICATION)
After ac tion verification it must be verified that the displayed error code is not displayed again and the experienced symptom no longer remains. This is done by running a program and checking that no error codes are displayed or the symptoms are experienced.
START TROUBLE SHOOTING
Alarm (error codes visible on the display)
EXPLANATIONS
Explanations to references to sections stated in the tables below:
CC : chapter Component check Advanced chapter Advanced trouble
trouble shooting: shooting Other references are given in plain text.
ERROR CODE EXPLANATION ACTIONS IN ORDER OF PRIORITY
GETINGE PACS 300 GETINGE PACS 300 DIS­PLAY
CLOSE THE DOOR Door unlocked at point of start 1. Operation, check
DOOR OPEN Door open during process 1. Operation, check
POWER FAILURE Power failure during process 1. Check external fuses.
PT-100 ERROR Error on load sensor during process 1. CC, 15. Load temperature sensor
SERVICE Service indicator. 1. Contact the service technician
Process does not start, frozen proc­ess
1. CC, 19. Software PACS cold start
2. CC, 18. Display
3. CC, 17. Circuit board
2. CC, 20. Door, (micro-switch)
3. CC, 17. Circuit board
2. CC, 20. Door, (micro-switch)
3. CC, 17. Circuit board
2. CC, 6. Overheating protection
3. CC, 17. Circuit board (Fuse)
4. CC, 17. Circuit board
2. CC, 17. Circuit board AI
STERILE ERROR Incorrect pressure/steam tempera-
ture in the sterile phase
PROCESS STOPPED The process has been stopped with
the STOP button
PRESSURE SENSOR ERROR
UPDATE Battery replacement or circuit board
Pressure sensor incorrectly cali­brated
fault
1. Operation, check
2. CC, 15. Load temperature sensor
3. CC, 12. Absolute pressure sensor,
4. CC, 17. Circuit board
1. Operation, check
2. CC, 18. Display
1. Advanced trouble shooting, Checks with error code
1. CC, 17. Circuit board
START TROUBLE SHOOTING
ERROR CODE EXPLANATION ACTIONS IN ORDER OF PRIORITY
NO WATER Timeout level guard 1. Operation, check
2. CC, 10. Level guard
3. CC, 3. Solenoid valve MV-3
HEATING ERROR Timeout heater 1. CC, 6. Overheating protection
2. CC, 5. Relay
3. CC, 4. Heater
4. CC, 17. Circuit board
HEATING ERROR Leakage Trouble shoot according to the section
Trouble shooting leakage
MANUAL MODE Component manually ac tivated 1. Operation, check
2. An output on the circuit board has been set to manual mode. ( menu tree Diagnostic)
POST TREATMENT E R­ROR
POST TREATMENT E R­ROR
Outlet valve clogged CC, 3. Solenoid valve, MV-1 Screen in the chamber clogged CC, 3. Strainer
START TROUBLE SHOOTING
Symptom (experienced by user)
EXPLANATIONS
Explanations to references to sections stated in the tables below:
CC: chapter Component check Advanced chapter Advanced trouble
trouble shooting: shooting Other references are given in plain text.
SYMPTOM ACTIONS IN ORDER OF PRIORITY
No ready signal 1. CC, 12. Absolute pressure sensor
2. CC, 3. Solenoid valve MV-2
3. CC, 11. Air filter
Instruments rust in the sterilizer 1. Poor quality instruments
2. Water cleaning filter – type de -ionization, check (see documentation supplie d with the water fil­ter)
Process time abnormally long 1. See the section installation. Electricity, Water
requirement
2. Advanced trouble shooting, Leakage test, check
3. Operation, check
4. CC, 22. Strainer
Humming noise in standby mode 1. CC, 3. Solenoid valves
Door hard to open/cannot be opened 1. CC, 3. Solenoid valve MV-8
2. CC, 20. Door, (lock)
Pressure gauge shows pressure with the door open 1. CC, 14. Pressure gauge
Steam leaks from the door 1. CC, 20. Door, door gasket
2. CC, 3. Solenoid valve MV-5
3. CC, 21 Pressure relief valve door gasket
Black background on the display 1. CC, 23 Fan.
2. CC, 18 Display
Unstable measurement value and text on display 1. CC, 23 Fan.
2. CC, 18 Display
3. CC 17 Circuit board
4. Check the power supply (Fluctuations?)
START TROUBLE SHOOTING
COMPONENT CHECK
TABLE OF CONTENTS
1. General ............................................................ 2
2. Chamber .......................................................... 2
3. Pipe/pipe couplings.......................................... 2
4. Solenoid valves................................................ 3
5. Heater............................................................... 5
6. Relay................................................................ 6
7. Overheating protection, resetting.....................6
8. Non return valve BV-1...................................... 7
9. Non return valve BV-2...................................... 7
10. Non return valve BV-3...................................... 7
10. Non return valve BV-3...................................... 7
11. Level guard ...................................................... 8
12. Air filter (accessory) ......................................... 8
13. Absolute pressure sensor ................................ 8
14. Ejector.............................................................. 9
15. Pressure gauge................................................9
16. Load temperature sensor (accessory)........... 10
17. Pressure relief valve....................................... 10
18. Circuit board...................................................11
19. Display............................................................ 12
20. PACS software...............................................12
21. Door................................................................ 14
22. Pressure relief valve, door gasket.................. 16
23. Screen............................................................ 16
24. Fan................................................................. 16
COMPONENT CHECK
page 1
1. General
IMPORTANT!
Only use Teflon tape that is in­tended for sealing pipe couplings against steam pressure.
It is extremely important for safe functionality that all nipples are tightened so that air, which can jeopardize the sterilisation result, can not enter the chamber. If pipe couplings must be removed while performing checks, think about the following when assembling:
Use ample Teflon tape, so that the tape partly forms a seal and partly seals the threads.
2. Chamber
The chamber walls are cleaned using a mild lime dis­solving agent. Rinse well. Drain by closing the door, the bottom valve then opens automatically.
BOTTOM FILTER (ALSO SEE 23. SCREEN)
Lift out the filter. Clean
LEVEL GUARD (ALSO SEE 11. LEVEL GUARD)
Check that the rocker moves up and down easily.
TIP!
Check whether steam is let out with an overpressure.
Use soapy water on cold com­ponents
Use a mirror on hot compo­nents
3. Pipe/pipe couplings
Check: Start a 134 °C program. Trouble shoot when the steri­lizer is in “Sterilisation” mode.
Check whether steam is let out. Action: Check according to respective component checks. Also see 1. General.
Verification: Check all pipes/pipe couplings again.
COMPONENT CHECK
page 2
IMPORTANT!
All valves open in an energised state.
4. Solenoid valves
Pay attention to solenoid valves that hum! The noise may be a forewarning about the coil starting to over­heat due to it not working in a closed iron circuit. This can occur when the moving core is influenced by the magnetic forces from the winding and dirt prevents the coil from reaching the end position. An air gap is then formed in the magnetic circuit, which lowers the induc­tance in the coil resulting in increased current.
MV-1
Is fitted in the bottom of the sterilizer and has the task of evacuating any remaining water in the sterilizer at the end of the process.
Check: Check that both the plunger and the area it runs in are clean. Always change the O-ring that seals between the coil and housing after removing the coil. Activate MV-3 and fill the chamber. Activate MV-1 and MV4 (ejector). Check that the water runs out of the cham­ber again. Measure the voltage across the coil.
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Check that the water level in the chamber is main­tained for a long period (about 15 minutes).
MV-2
Is fitted at the rear, top edge of the sterilizer. Takes care of pressure equalization after the post-vacuum phase by releasing air into the chamber. The valve is open when the sterilizer is in standby mode if the door is locked. It becomes de-energised as soon as the door is opened.
Check: Check that both the plunger and the area it runs in are clean. Always change the O-ring that seals between the coil and housing after removing the coil. Activate MV-4 and MV-2 on the diagnostic menu. The pressure should not drop appreciably below atmospheric press­sure. Measure the voltage across the coil.
Check for leakage: Dismantle BV-2. Activate MV-3 and the heater. No steam should come out of MV-2.
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Run a sterilisation program. Check for leakage ac-
COMPONENT CHECK
page 3
cording to Check and that pressure equalization takes place at the end of the program.
MV-3
Is fitted directly on the inlet. Loads the chamber with water at the start of the process.
Check the opening function: Activate MV-3 in the diagnostic menu; water should then flow into the chamber.
Check the closing function: In standby mode, check that the water level in the chamber does not rise once the valve to the chamber has been closed.
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Run a sterilisation program. Check that no water en­ters the chamber after the process has ended and the door is open.
IMPORTANT!
Minor steam leakage can be due to a leaking gasket and has noth­ing to do with the MV-5 function
MV-4
Opens the water flow to the ejector to produce a vac­uum
Check: Check that water does not run out of the drain hose in standby mode
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Run a process and check again that water does not run out of the drain hose in standby mode.
MV-5
Pressurises the door gasket before a program. Check:
During the process, check that the door seals against steam.
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Check the door for steam leakage.
MV-6
Vents the chamber while the program is running. Opens at the given time interval. Also simplifies eva-
COMPONENT CHECK
page 4
evacuation of the chamber pressure after sterilisation is completed.
Check: Build up a pressure in the chamber. Open MV-6. The chamber pressure should drop. Check that the cham­ber pressure is maintained when MV-6 is closed. Pi­pes on the outlet side should be cooler than on the pressurised side.
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Run a program. Now verify according to Check, see above.
MV-8
Evacuates the door gasket pressure after the com­pleted process.
IMPORTANT!
A symptom of poor opening func­tion in the valve is difficult door opening.
TIP!
Ensure the heater is de-energised before you measure the resis­tance.
Check: Close and lock the door. Open MV-5 and open MV-8. Water should flow out of the drain.
Action: Clean and replace the plunger and O-ring. Replace the entire valve if necessary.
Verification: Run a process. The door should be easy to unlock af­ter the completed process.
5. Heater
Is fitted in the bottom of the chamber and has the task of heating the water dosed for the process.
Check: Activate the heater. Measure with a clamp-on amme­ter, this should show about 8 A/coil. The heater output is 2000 W x3 coils, which gives a resistance of about 27 Ω/coil. Check the resistance.
Action: Replace the heater with incorrect resistance.
If the clamp-on ammeter shows 0 A, and the relay and circuit board are OK. Replace the heater.
Verification: Verify by measuring the resistance.
COMPONENT CHECK
page 5
IMPORTANT!
Note that if overheating protection trips due to a temperature outside of the working range (0 - 300 °C), the sterilizer must cool for a pe­riod so that the resetting button shall remain pressed in. Wait until the sterilizer is within the working range again. Note that it may also require a great deal of force to press in the overheating protection.
6. Relay
Is fitted in the distribution box. Its task is to connect the heater to the power supply.
Check: Activate the heater, measure the coil on the relay us­ing a voltmeter. The coil is across A1-A2. Measure the voltage on the outgoing connection 2 - 4 - 6
7. Overheating protec­tion, resetting
Protects the heater against overheating, the protection trips at approx. 300 °C. Requires manual resetting. The sensor is positioned on the heater coil. The reset­ting button is positioned on the protection. A symptom that the overheating protection has tripped is the ti­meout alarm, heating error.
Possible causes:
That the sterilizer has been transported in tem-
peratures below freezing.
That a fault has occurred in the temperature regu-
lation of the heater.
That leakage results in the water being consumed
during the process.
The unit should always be de-energised when reset­ting.
CAUTION!
Check that the overheating protec­tion sensor is fitted correctly on the heater. Otherwise the sensor has no function.
Check: Check that the manual resetting button is pressed in. Measure the voltage on the incoming and outgoing pins on the protection. Voltage on the incoming but not on the outgoing, means that the protection has tripped or is faulty.
Action: Replace the overheating protection. Make sure that the sensor is correctly fitted on the heater. Also see daily maintenance
Verification: Check that the sterilizer display is lit.
COMPONENT CHECK
page 6
8. Non return valve
BV-1
Is fitted between MV-1 and the chamber and has the purpose of preventing return flow through MV-3
Check: Fill the chamber with an open door by opening MV-3. Shut the main tap and let MV-3 remain energised and open MV-4. Check whether water is running out from the drain or that the level in the chamber drops over a long period of time, at least 15 minutes.
Action: Replace BV-1.
Verification: As Check, see above.
9. Non return valve
BV-2
Is fitted between the air filter and MV-2. Its task is to prevent steam, with a fault on MV-2 being forced thro­ugh the filter.
Check: Close and lock the door. Remove the air filter. Build up a pressure; open MV5 and MV1, close MV1, acti­vate the heater, open MV2. The pressure should rise to 1 bar according to the front pressure gauge. Check that no steam comes out of the air filter holder.
Action: Replace BV-2. Check that the air filter is not wet.
Verification: As Check, see above.
10. Non return valve
BV-3
Is fitted after MV-1 and has the task of preventing wa­ter from the ejector returning up into the chamber
Check: Open MV-4 and MV-1. Check that no water enters the chamber.
Action: Replace BV-3
Verification: As Check, see above.
COMPONENT CHECK
page 7
11. Level guard
Is fitted in the lower part of the chamber above the heater. The task is to check the filling of the chamber.
Check: Move the rocker arm up and down so the digital input (Service, Diagnostic, Digital input) should switch be­tween 1 and 0.
12. Air filter (accessory)
Its task is to supply the sterilizer with sterile filtered air after the post vacuum phase.
Check: The air filter must be dry. Check that pressure equali­zation takes place after the post vacuum phase in the program.
Action: Replace the filter.
Verification: Run a program. Also see section BV-2, check.
CAUTION!
Connect the pressure gauge for the absolute pressure to the outlet on the lead-through at the rear of the chamber.
TIP!
Start a program and stop it by using the “Pause” switch in the vacuum and sterile phase. Then adjust the A-value (sterile phase) and the B-value (vacuum)
TIP!
Use the program CS1000 (for PC), for a basic calibration procedure.
13. Absolute pressure
sensor
Its task is to monitor and control the process. Check:
Carried out with an independently, calibrated pressure gauge for the absolute pressure, whose values are compared with the values shown on the display. Area: 0 - 4 bar absolute pressure corresponds to 4 – 20 mA, output signal with an open lid approx. 8 mA
Action: When the deviation is significant, replace the sensor and calibrate. With minor deviation, calibrate. Calibra­tion is carried out via the menu tree. Step using the UP arrow and the ENTER button through SETUP/ SYSTEM/ (password)/ CALIBRATION/ MANUAL CA­LIBRATION. Select the sensor and adjust index A or B up or down depending on the type of fault indica­tion. B with vacuum and A with a sterile pressure.
Verification: Run a sterilisation program. Perform with an inde­pendently, calibrated pressure gauge for the absolute pressure whose pressure values are compared with the values shown on the display.
COMPONENT CHECK
page 8
14. Ejector
The ejector is mounted on the pipe from the water in­let to the drain. The ejector's function is to create an underpressure in the chamber.
Check: Activate MV-1 and MV-4 through the diagnostic menu. The pressure shall, under the condition that BV-3 is problem free and the water pressure is above 1.5 Bar, reach a vacuum greater than 800 mBar within 15 min­utes.
Action: Clean the ejector. Any dirt or leakage in the ejector connections cause impaired vacuum capacity.
Verification: Check the depth of vacuum according to the check section above.
15. Pressure gauge
CAUTION!
The pressure gauge only gives an indication of the pressure and should not be used during calibra­tion.
Is placed on the front of the sterilizer. Shows the chamber pressure even if the power is switched off.
Check: When the door is open the pressure should show 0 ±0.1 bar. Check that the pointer indicates a pressure higher and lower than zero during a cycle.
Action: Calibrate the pressure gauge according to the section Pressure gauge adjustment. If it is not possible, re­place the pressure gauge.
Verification: Run a program. Compare the pressure gauge with the pressure indication on the display.
PRESSURE GAUGE ADJUSTMENT
Make sure the door is open.
Open the front.
Loosen the brass nipple at the rear of the pres-
sure gauge and lift out the pressure gauge.
Pry loose the pressure gauge glass by inserting a
screwdriver in the notch at the top of the pressure gauge glass.
Insert narrow tipped pliers under the pointer and
lift it off.
Adjust and press back the pointer in the right posi-
tion.
COMPONENT CHECK
page 9
16. Load temperature
sensor (accessory)
TEMPERATURE SENSOR
The sensor is placed in the lead-through to the cham­ber and is connected to contact piece X2 on the circuit board. It monitors the sterile process and provides the display with information about the temperature of the liquid that sterilises and controls that the liquid is un­der 95°C before the door can be opened.
Check: Loosen the connection on the circuit board and meas­ure the resistance on the sensor with an ohmmeter. The sensor's resistance should at 20 ° be about 120 ohm. The higher the temperature, the higher the resis­tance.
Action: With a resistance above 150 ohm or 0 ohm. Replace the sensor.
Check for leakage: Check that there is no steam leakage from the sen­sor's lead-through – silicone seal.
WARNING!
Adjustments or actions that alter the opening pressure on the pressure relief valve may only be performed by persons with docu­mented experience.
Action: With leakage – replace the silicone seal.
17. Pressure relief valve
The function of the pressure relief valve is to relieve the pressure in the chamber, should this rise uncon­trolledly.
Check: Start a process. When “sterilisation” is reached, close MV-1 and MV-6 as well as start the heater. Check us­ing an independently calibrated pressure gauge for absolute pressure that the opening pressure is 3.7 ±0.14 bar (absolute pressure). Check that the pres­sure relief valve does not leak water or steam at nor­mal sterilisation pressure – max 3.15 bar.
Action: Clean the valve. Replace the valve if necessary.
Verification: Run up the chamber pressure, activate the heater and close MV1 and MV6. Read when the pressure relief valve trips. Check that the valve does not leak during a program
COMPONENT CHECK
page 10
CAUTION!
The display unit and the circuit board unit must always be kept in shielded bags until the compo­nents are to be installed in the sterilizer. Make sure not to be electro statically charged when these parts are handled, by e.g. touching an earthed heater or by
WARNING!
Lithium battery. Explosion risk. Replace the battery with the same type as recommended by Getinge.
TIP!
Battery When battery replacement is nec­essary and no battery is available, the sterilizer can temporarily be run with jumper (see section 25 Software PACS, Cold start) in cold start mode. The calendar and process counter do not work in cold start mode.
18. Circuit board
Is located in the casing to the right. Its function is to bridge all input and output signals with the software PACS. The circuit board houses, among others, the flash memory that holds the PACS software and the battery.
Check the circuit board:
Incoming voltage on X23.
That PT100/PT1000 jumpers are located cor-
rectly: PT 100 position
Fuse 315 mA.
Voltage on the battery, 3.6 V.
Connect the cables and connectors.
Check the relays: To check the function of a relay, activate each output, connect a voltmeter between Line N (X23) and each output. The voltmeter should show approx. 230 VAC. On connection COM 1 ( X32) the board is equipped with a serial communication port (RS232) for connec­tion e.g. to a computer.
Action: Replacing the battery.
Make a backup of the software according to the sec­tion 25 Software PACS, if possible. Cut off the battery solder tags as close to the battery as possible. Do not solder on the circuit board! Solder in the new battery on the old solder tags. Check that the new battery's voltage is 3.7 V Perform a cold start according to 25 Software PACS,
Cold start. Action:
If the above does not correspond, correct or replace the board.
Verification: Run a prepacked program.
Replacing the circuit board
Move over the flashmemory circuit from the old board. Check the circuit board according to the section above. Perform a cold start according to 25 Software PACS, Cold start. Set the time and date and the language version Go to the menu TOOLS-PACS and select SAVE RAM TO FLASH.
COMPONENT CHECK
page 11
19. Display
Is placed on the front of the sterilizer. Its task is to show values and act as the interface to the PACS software. The display board is connected to X26 on the circuit board. The keypad in the decal is con­nected to contacts J1 and J3 on the display card.
Check: Check the cables including connection X26. Check jumper JP1. Should be fitted between pins 1 and 2.
Action: Replace the display card.
Verification: Run a program and make sure the display communi­cates without interference.
20. PACS software
Is installed in the flash memory on the circuit board. Its function is to monitor and govern the entire steri­lizer and the processes run on it. PACS also logs events and alarms.
Action:
MAKE A BACKUP.
Follow the following when the sterilizer's program ver­sion is to be saved on a PC: NOTE. The calibration settings will not be saved on this back up.
X24
Connect your PC to contact X32 on the sterilizer use the communications cable. Start the CS 100 program. State the service password Go to the menu TOOLS-PACS to RAM/ UPLOAD TO FILE. Now save the backup file on the hard drive in the PC.
LOAD A NEW BACKUP
Connect your PC to contact X32 on the sterilizer use the communications cable. Start the CS 100 program. State the password Go to the menu TOOLS-PACS and select SAVE RAM TO FLASH. Go to the same menu and select RAM/ DOWNLOAD FROM FILE. Set the time and date and the language version Go to the menu TOOLS-PACS and select SAVE RAM TO FLASH.
COMPONENT CHECK
page 12
Verification: connect an external pressure gauge intended for ab­solute pressure. Run a process and check the calibra­tion against the display.
BACK UP THROUGH A NEW FLASH MEMORY
Connect your PC to contact X32 on the sterilizer use the communications cable. Start the CS 100 program. State the password Go to the menu TOOLS-PACS and select SAVE RAM TO FLASH Switch off the power and change FLASH circuit Make a “cold start” see below Set the time and date and the language version Go to the menu TOOLS-PACS and select SAVE RAM
TO FLASH
COLD START
Switch off the power and place the jumper (marked with an arrow in picture) to position Cold start. Switch on the power again, wait a few seconds. Place the jumper in the Normal position
POST VACUUM ADJUSTMENT
How to navigate through the menus, see the section Main menus. Password. Step as set out in the sec­tion Menu navigation. The post vacuum time is ad­justable up to 30 minutes. A value lower than the fac­tory setting can not be made.
COMPONENT CHECK
page 13
21. Door
Door is locked mechanically by a bayonet mount and 10 flanges. The door gasket seals through the gasket groove being pressurised at the start of the process which then presses against the door.
Open and close the sterilizer door several times in a row and check that there is no play on the door and hinge. Also check that the door spring has not be­come too weak or has broken.
Check: Check that a process can not be started if the door is not closed correctly. Check the hinge joints, micro­switch and door lock.
Action:
Clean and lubricate the door's moving parts.
Check the wear to the hinge joint, change if nec-
essary.
IMPORTANT!
The door must be fully open (110°) to carry out the action.
Check the micro-switch by activating it manually
and read the display (switches between the pro­gram parameter and “
Check the door lock moves easily and does not
jam. Check especially that it reaches all the way in the deactivated position.
Verification: Open and close the door. The door's movement should not be more resistant than it can be closed us­ing normal force. In general as in Action.
Check the door spring: Check that the door spring has not become too weak or has broken by opening the door. The door spring shall prevent the door from shutting.
Action:
Remove the door's plastic cover
Open the door as far as possible
Loosen the nut (6)
START OK” on the display)
7 6 8
Loosen the index screw (7). Remove the door
Remove both clasps holding the spring axle. Se-
cured with six nuts under the axle
Remove the circlip (8) located at the end of the
spring axle.
Pull off the axle
COMPONENT CHECK
page 14
Change the spring. Press the spring leg in the
door holder.
Refit in the reverse order
Centre the door with the help of the screw (7)
Verification: Open the door and check that the door spring pre­vents the door from shutting.
DOOR GASKET
Door is locked mechanically by a bayonet mount and 10 flanges. The door gasket seals through the gasket groove being pressurised at the start of the process which then presses against the door.
Check: Check that the gasket is in one piece and that the contact surfaces are clean and lubricated with a food industry grease.
Action: The door gasket should be replaced after max 1200 cycles/once every 3rd year. Close the water connec­tion before replacing the door gasket. Open the door Remove the gasket using a blunt tool Check that the gasket groove is clean Clean the groove if necessary Grease the new gasket Press the new gasket in the groove fully
IMPORTANT!
Check that steam does not blow out around the door when the pressure is built up.
Verification: Close the door and run a program 134 °C check for leakage. Also see section Component check, 4. So- lenoid valves, MV-5 and MV-8.
COMPONENT CHECK
page 15
22. Pressure relief val-
ve, door gasket
Used to release the water pressure in the door gasket groove if the pressure exceeds 4.8 bar
Check: Check that the pipe after the pressure relief valve is not hot during “sterilisation”.
Action: Replace the pressure relief valve
Verification: Run a new process and check whether the pipe is hot
23. Screen
Is positioned in the bottom of the chamber to prevent a stoppage in the outlet.
Check: Remove the screen, by lifting it straight up from the chamber
Action: Clean the screen
Verification: Make a visual inspection that the screen is clean.
24. Fan
The fan is positioned behind the casing and is con­nected to contact piece X16 on the circuit board. Its task is to lower the temperature under the casing when the process is run.
Check: Loosen the casing and check that the fan rotates when a process is run. Check that the fan is voltage fed.
Action: Replace the fan
COMPONENT CHECK
page 16
ADVANCED TROUBLE SHOOTING
Flow chart
This section concerns advanced trouble shooting. The different sub -sections below follow on from the previ­ous sectionSTART TROUBLE SHOOTING. This me­ans that all trouble shooting starts there.
ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION
Process description
Steam is generated in the sterilizer through an exter-
5.
6. 7.
8.
3.
nal water feed that may be preceded by a softening fi l­ter.
Condensation is led off to the drain.
1. The door gasket is pressurized by MV-5 opening and releasing water into the gasket groove.
2. The sterilizer is filled with water until the level gu­ard cuts in.
3. The heater heats the water and the pressure ri­ses.
1. 2.
4.
9.
10.
4. The steam pressure is monitored by the pressure sensor which, via a PID -regulator, controls the heater.
5. The temperature of the steam is monitored by the pressure sensor (through a theoretical calcula­tion).
6. The solenoid valve MV-1 is used to empty the wa­ter from the chamber.
7. The steam pressure is released via MV1 and MV6
8. The water pressure powers an ejector, which cre­ates an underpressure in order to lowe r the chamber pressure below atmospheric pressure. MV4 releases the water pressure on the ejector.
9. MV-2 is used to release the air in the chamber.
10. MV-8 releases the door gasket's water pressure. The door can now be opened.
ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION
13.
11. 12.
14.
15.
16.
17.
18.
LIQUID LOAD
11. The door gasket is pressurised.
12. The sterilizer is filled with water until the level gu­ard cuts in.
13. The heater heats the water and the pressure ri­ses.
14. The steam pressure is monitored by the pressure sensor which, via a PID -regulator, controls the heater.
15. The load sensor monitors the temperature of the liquid to be sterilised. The load sensor temper a­ture determines when the sterilizer switches to sterilisation.
16. The solenoid valve MV-1 is used to empty the wa­ter from the chamber.
17. The steam pressure is released via MV1 and MV6 through ramping to prevent decoction.
18. When the liquid reaches a temperature below 95°C (controlled by the load sensor) MV-8 re­leases the gasket pressure and the door can be opened. The process is complete.
PRE-TREATMENT
The purpose of pre-treatment is to remove air from the chamber and from the goods. Air prevents the requi­site contact between the steam and micro -organisms to be killed.
Pre-treatment consists of steam injection and evacua­tion pulses in different processes depending on the program selected. In addition, the humidification es­sential to killing mainly takes place during the pre­treatment stage.
Pre-treatment process:
Blow through (1) MV-1 opens, the chamber is filled with water and the heater then heats the water. MV-6 opens dur­ing the preset time.
Pressure increases (2) MV-6 closes and the heater heats. The pressure rises to a preset level. The sterilisation pressure should be reached during the final increase in pressure. Sterilisation begins when the sterilisa­tion pressure is reached.
ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION
STERILISATION
Sterilisation takes place by maintaining the saturated steam's relation (boiling point at a specific pressure).
The sterilisation phase is in progress during a pre ­programmed number of minutes at the preselected temperature/pressure.
Sterilisation process:
MV-6 opens during the preset time intervals.
The pressure is controlled by means of the heater,
which heats as required.
POST TREATMENT
The purpose of post treatment is to reduce the mois­ture content in the goods. During post treatment the pressure drops to atmospheric pressure or lower – depending on the selected program. Post treatment is completed by air being drawn in until the chamber is at atmospheric pressure.
Post treatment process:
MV-1 opens and releases the remaining water.
MV-4 opens and the ejector starts.
MV-6 opens after an additional few of minutes.
When the right vacuum is reached (800 mBar)
time metering starts.
When the time elapses MV-1, MV-4 and MV-6 close. MV-2 opens and air is drawn in and the pressure rises towards atmos pheric pressure. MV-8 opens and the door gasket pressure is re­leased. The process is finished.
ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION
PROCESS DESCRIPTION - LIQUID LOAD
PRE-TREATMENT
The purpose of pre-treatment is to remove air from the chamber and from the goods. Air prevents the requi­site contact between the steam and micro-organisms to be killed.
Pre-treatment consists of steam injection and evacua­tion pulses in different processes depending on the program selected. In addition, the humidification es­sential to killing mainly takes place du ring the pre­treatment stage.
Pre-treatment process:
Blow through (1) MV-1 opens, the chamber is filled with water and the heater then heats the water. MV-6 opens dur­ing the preset time.
Pressure increases MV-6 closes and the heater heats. The pressure rises to a preset level just above the sterilisation pressure to accelerate the heating of the liquid. When the load sensor temperature approaches the sterilisation temperature, the steam pressure drops to the sterilisation pressure. Sterilisation starts.
ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION
STERILISATION
Sterilisation takes place by maintaining the saturated steam's relation (boiling point at a specific pressure).
The sterilisation phase is in progress during a pre ­programmed number of minutes at the preselected temperature/pressure. The load sensor monitors the temperature of the liquid.
Sterilisation process:
MV-6 opens during the preset time intervals.
The pressure is controlled by means of the heater,
which heats as required.
POST TREATMENT
The purpose of post treatment is to lower the tem­perature of the liquid below boiling point at atmos­pheric pressure.
Post treatment process:
MV-4 opens and the ejector starts.
MV-1 opens at the preset time intervals to ensure
a slow pressure decrease and prevent decoction of the liquid. The remainin g process water is re­leased in connection with ramping.
When atmospheric pressure is reached and the temperature of the liquid (controlled by the load sensor) is below 95°C, MV-1 and MV-4 close. MV-2 opens and the atmosphere in the chamber and the atmosph ere is the room are kept at the same level. MV-8 opens and the door gasket pressure is released. The process is finished.
ADVANCED TROUBLE SHOOTING, PROCESS DESCRIPTION
Trouble shooting leak­age
IMPORTANT!
Trouble shooting takes place:
At overpressure Follow the method prescribed in
Check component, 2. Pipe/pipe couplings.
TIP!
Check whether steam is let out with an overpressure.
On cold components: use soapy water
On hot components use a mirror
1 Check for door leakage as set out in the section
Check component, 20. Door.
2 Check for leakage from the pressure relief valve
as set out in the section Check component, 16. Pressure relief valve.
3 Check for leakage from the solenoi d valves as set
out in the section Check component, 3. Sole- noid valves.
4 Check for leakage from the pipe couplings as set
out in the section Check component 2. Pipe/pipe couplings.
5 Check for leakage on the temperature sensor's
lead-throughs as set out in the section Check component, 15. Load temperature sensor.
6 Check for leakage from the ejector as set out in
the section Check component, 13. Ejector.
Calibration instruction
1 Connect a calibrated test pressure gauge to the
connection labelled PT on the sterilizer. 2 Start a 134 °C program. 3 Let the program run to post treatment 4 When the vacuum has stabilized, calibrate the B
parameter for the pressure sensor. 5 Check that the external pressure gauge shows
the same pressure as on the sterilizer's display. 6 Allow the sterilizer to finish. 7 Start a new 134 °C program. Let the program run
to “sterilisation”. Calibrate the A parameter for the
load sensor. Check that the external pressure
gauge shows the same pressure as on the steri-
lizer's display. 8 The sterilizer will calculate the theoretical sterile
temperature, based on the calibrated pressure va-
lues. 9 Allow the program to finish. Check the vacuum
setting during the post vacuum phase.
ADVANCED TROUBLE SHOOTING, TROUBLE SHOOTING LEAKAGE
MENU NAVIGATION
Orientation
Navigation through the menus is done using the four arrow keys UP arrow, DOWN arrow, LEFT arrow, RIGHT arrow and the ENTER key.
Navigate to your choice using the UP arrow and DOWN arrow . The selection is indicated on the dis­play by a cursor below the value to be changed.
To return to the previous level, press the left arrow when the menu position is not selectable.
Press the ENTER key to activate your menu selec- tion.
When a menu selection has been activated, one of two actions can be performed.
When the selection is a new menu position, a fur-
ther selection can be made.
TIP!
You can return at anytime to the main menu by pressing the left key and holding it down for 4 sec­onds.
When the value can be edited, use the arrow keys
as set out above. The selection can either be shown on one or two
rows. When one row is shown, this means the row can be selected and activated using the ENTER key. When two rows are shown, the first row shows the se­lected name and the second row the value linked to the selection.
MENU NAVIGATION
Edit values
Press the ENTER key to set the display in edit mode, i.e. make the selected value editable. The cursor is positioned to the far left below the first character when the value is selectable.
EDIT PRESET VALUES
Use the UP arrow and DOWN arrow to change a va­lue. Different values are shown each time the arrow key is pressed.
Save the value by pressing the ENTER key.
EDITING NUMERICAL VALUES
TIP!
When adjusting the sterile or post vacuum times, the value can not be set below the preset value.
The value can consist of several characters. The cha­racter with the cursor below can be edited.
The highlighted character can be increased or de­creased by using the UP arrow or DOWN arrow.
Use the RIGHT arrow or LEFT arrow to select the next editable character to the right respective left.
When you have finished, save the values by pressing the ENTER key.
MENU NAVIGATION
THE MENU TREE
Main menus
Main selection: Program name, Program version, Phase, Cycle counter and Setup menu.
Password service 120416 Navigate to your choice using the UP arrow and
DOWN arrow. The selection is indicated on the dis-
play by a cursor below the value to be changed. To return to the previous level, press the LEFT arrow
when the menu position is not selectable. Press the ENTER key to activate your menu selec-
tion.
1. PROGRAM NAME Program name: Selected program Params 1 Params 2: Sterilisation time & temp (or
Heating up) Select program: Sub-menu to selected program.
Important, the sterilizer has a PROGRAM SELEC-
TOR button for this on the front. Change parameter: Possibility to change the sterile
time and post vacuum time. Password 558
2. PROGRAM VERSION
E.g.
050121
3. PHASE
E.g.
PRE-TREATMENT
4. CYCLE COUNTER
Possibility to read the sterilizer's counter, pressure and temperatures. E.g.
0210
Setup menu
5. SETUP MENU Print Last Cycle: Not used. System: To access the System menu. Password: State System menu: See the section System menu. About: Software version
MENU TREE, MAIN MENUS
System menu
Navigate to your choice using the UP arrow and DOWN arrow . The selection is indicated on the display by a cursor below the value to be changed.
To return to the previous level, press the LEFT arrow when the menu position is not selectable.
Press the ENTER key to activate your menu selection.
The sterilizer's menu for installation specific settings and unique machine settings. Main selection: Calendar, Calibration, configuration, Dipswitch, Service and Flash memory.
1. CALENDAR MENU
Time and date settings
2. CALIBRATION MENU
Only possible with manual calibration See separate section Calibration menu
3. CONFIGURATION MENU See separate section Configuration menu
4. DIPSWITCH MENU
Not used.
5. SERVICE MENU See separate section Service menu
6. FLASH MEMORY MENU
Save adjustments and changes to the flash memory. Should always be done once adjustments and calibrations have been made.
MENU TREE, SYSTEM MENU, ETC
Calibration menu
Main selection: Manual calibration, Automatic ca­libration and Compensation table.
Navigate to your choice using the UP arrow and DOWN arrow . The selection is indicated on the dis-
play by a cursor below the value to be changed. To return to the previous level, press the LEFT arrow
when the menu position is not selectable. Press the ENTER key to activate your menu selec-
tion.
1. MANUAL CALIBRATION
Always use MANUAL CALIBRATION. To calibrate, see the section Advanced trouble shooting, Cali-
bration 00 Not used 01 Not used 02 Not used 03 Pressure : Step the cursor to each value to be
changed. Verify with Enter
04 Not used 05 Not used 06 Not used 07 Load sensor: Step the cursor to each value to be
changed. Verify with Enter
2. AUTOMATIC CALIBRATION
Not used
MENU TREE, CALIBRATION MENU
Configuration menu
Main selection: Calendar, Calibration, configura­tion, Dipswitch, Service and Flash memory.
Navigate to your choice using the UP arrow and DOWN arrow . The selection is indicated on the dis-
play by a cursor below the value to be changed. To return to the previous level, press the LEFT arrow
when the menu position is not selectable. Press the ENTER key to activate your menu selec-
tion.
1. LANGUAGE DATE UNI T
Settings related to countries
Language setting: Display language Date format Pressure format: NOTE! Must not be changed! The
sterilizer is calibrated based on bar. Temperature format: NOTE! Must not be changed!
The sterilizer is calibrated base don degrees Celsius.
2. PRINTER Printing Mode: Not used. Logging interval low: Logging interval for the total
process excluding the sterilisation phase. Logging interval high: Logging interval for the sterili-
sation phase only.
3. ALARM CLOCK
Not used
4. PACS ADDRESS
Possibility to give the sterilizer an Id. When there is more than on sterilizer with the printer in the same lo­cation. 1-99
5. COMMUNICATION
Settings for the Com-ports (3) RS232 or RS485 Set for communication with the service software. Must not be changed!
MENU TREE, CONFIGURATION MENU
Service menu
Main selection: Error log, Service messages and Diagnostics.
Navigate to your choice using the UP arrow and DOWN arrow . The selection is indicated on the dis-
play by a cursor below the value to be changed. To return to the previous level, pres s the LEFT arrow
when the menu position is not selectable. Press the ENTER key to activate your menu selec-
tion.
1. ERROR LOG
20 most recent error codes The type of error code, in which phase of the program
and when.
2. SERVICE MESSAGES
Resetting of the service alarm (alarm every 400 proc­ess)
3. DIAGNOSTICS
Manual test of components. Set the value under MAN and control with S:1 or S:0
Analog input: Shows the water, pressure and tem­perature values
Analog output: Not used. Digital input: Shows the door switch and an external
level switch
Digital output: Check of all components Test User Flag: Not used Test System Flag: Not used Test Printer: Not used Test LED/buzzer: Functional check of the sterilizer's
audio signal and LEDs.
MENU TREE, SERVICE MENU
Component placement, internal
1 Circuit board 2 3 Ejector 4 Solenoid valve MV-3 5 Non return valve BV-1 6 Solenoid valve MV-4 7 Solenoid valve MV-5 8 Solenoid valve MV-6 9 Solenoid valve MV-8 10 Heater 11 Relay 12 Overheating protection, ÖV300 13 Chamber 14 Level guard 15 Solenoid valve MV-1 16 Absolute pressure sensor 17 Pressure relief valve (Chamber) 18 VT/PT-outlet 19 Pressure relief valve (Door gasket) 20 Solenoid valve MV-2 21 BV-2 22 BV-3
COMPONENT PLACEMENT, INTERNAL
23 24
COMPONENT PLACEMENT, INTERNAL
COMPONENT PLACEMENT
Component placement, outside
1. Water supply
2. Water connection
3. Water outlet
4. Handle
5. Control panel
6. Reset button, overheating protection
7. Printer port RS232 (accessory)
8. Mains supply
9. Door lock
COMPONENT PLACEMENT, OUTSIDE
CIRCUIT BOARD DIAGRAM
X1 Absolute pressure sensor X2 Temperature sensor X3 X9 X10:A1 Switch (Door) X10:A3 Level guard water X16:2 L1 (Phase) X16:3 Fan X16:4 MV-1 X16:5 MV-2 X17:3 MV-5
X17:4 MV-4 X17:5 MV-3 X18:3 MV-6
X18:4 Relay (Heater) X18:5 MV-8 X19:4 Recorder X20 X21 X23 X23:1 L1 (Phase) X23:2 Neutral X24 PC communication X25 X26 Display X27 X32
CIRCUIT BOARD DIAGRAM
WIRING DIAGRAM
KRETSKORTSCHEMA
INSTALLATION
CAUTION!
Installation must be carried by technicians with documented experience.
WARNING!
Make sure that the area where the sterilizer is to be located can withstand a load of at least 250 kg.
INTERNAL TRANSPORT
The easiest method to move the sterilizer is using a pallet truck or a barrow with a minimum lifting limit of 120 kg.
UNPACKING
Check that the packaging is intact. Check that the supplied goods correspond with the
order. Check that the sterilizer is free from defects. Transport damage must be reported immediately to the
carrier.
Setup
PREPARATIONS
Setup ought to be carried out by technicians with documented experience.
The sterilizer must be located indoors with a maximum ambient temperature of 35 °C and a relative, no condensing, humidity of maximum 95 %.
CAUTION!
Ambient temperature, max 35 °C. Ambient humidity, max 95 %.
PLACEMENT
The sterilizer is a floor model and must stand free, it must not be enclosed. It should be positioned so that loading and unloading can be performed without risk. It must be possible to open the door fully. See the recommended distances in the picture Follow local safety regulations.
Once the placement of the sterilizer has been established, anchor the sterilizer to the floor by each of the three feet. This prevents the sterilizer from tipping. A 13 mm hole has been drilled in each foot. Use a screw and plug intended for the surface in question. Make sure there is a slight backward incline, i.e. the side of the door should be slightly lower than the front, about 2 - 3 mm. This is accomplished by placing a washer under each of the front feet.
INSTALLATION
VS 90/60 1550mm VS130 1750mm
130mm
INSTALLATION
Electrical connection
IMPORTANT!
The electrical connection of the unit must be carried out by trained personnel
Prior to carrying out any connection work, ensure that the characteristics stated on the nameplate correspon d to the mains supply to which it is to be connected. In case of uncertainty as to the unit supply voltage,
VOLTAGE FLUCTUATION:
The electrical components tolerate a supply voltage fluctuation of between -10 and +10 %. For larger fluctuations, it is necessary to use a voltage stabilisation system placed between the circuit-breaker and the processor unit transformer. In this case, contact the sales office.
POWER-SURGE PROTECTION:
A power-surge protection device (over current protection) must be placed on the sterilizer’s AC power supply. This protection is the responsibility of the customer.
CIRCUIT-BREAKER:
A device enabling the sterilizer to be electrically isolated must be inserted on all the live conductors (including neutral). This switch or circuit-breaker must be part of the building’s electrical equipment, positioned near the unit, and must be easily accessible and marked as the power switch for the unit, while stating its function.
CAUTION!
The sterilizer must be declared to your electricity supplier in case of a 230 single phase connection.
If the unit is connected in polyphase. Its characteristics are as follows:
Breaking power 63 A breaking the 3 phases as well as neutral Earth terminal making it possible to connect the isolating device to the earth conductor without braking it. The on and off positions must be marked: I on and O off
Box IP 54 TYPE- Power (kW) Voltage (V) Current (A)
VS 60 - 130 6.1KW 230V Single phase 30A VS 60 - 130 6.1KW 230X3V 20A VS 60 - 130 6.1KW 400X3V + N 16A Connect the sterilizer to the earth and according to the
specification of the manual. The installation must be protected in accordance with the regulations in force. Failure to comply with this instruction may result in electrocution of the user and damage to the unit.
CONNECTION POWER SUPPLY
Open the front box.
INSTALLATION
WARNING!
Never work with energized equipment. The sterilizer must be earthed.
400V 3 phases connection 5-wire cable (3 phase + Neutral + earth) No jumper
230V 3 phases connection 4-wire for 230V (3 phase (3 phase + earth) Jumper between L2-N
230V single phase connection 3 wire (1phase + Neutral + earth). Jumper between L1-L2-L3
With appropriate section to 60227 and 60245. Pass the power cable through the rear housing provided
for this purpose. Connect the power cable earth connector to the terminal
marked with the earth symbol Connect the other end of the cable to the circuit-
breaker/cut-off device located on the outside. Change of voltage: This adaptation requires certain parts to be changed;
use the “Jumper kit” furnished with the machine to perform this change. For further information see the electrical diagram, or get in touch with the selling company.
Terminal connection:
Also see “Wiring diagram X1” 400 V 3 Phases Jumper between E-G-H 230 V 3 Phases Jumper between A-G, B-E, C-H 230 V Mono Phase Jumper between N-A-B-C
INSTALLATION
CAUTION!
The water supply must be fitted with a shutoff tap, as the sterilizer must not be left without supervision with the water pressurized.
WARNING!
Risk of hot water from the drain hose, exercise the greatest possible caution.
Water connection
The incoming water must have a pressure between 1.5 bar and 6 bar with a temperature of about 15 °C. Water quality <4 dH
The sterilizer is connected to a ½" water connection with a shutoff tap. The sterilizer must not be connected with hoses less than 12 mm. The hoses supplied with the sterilizer are fitted with Gaz ½" G connections. The hoses (two) are labelled “Inlet” and “Drain”.
The drain should be connected to a 6" cesspool, with a capacity of 10 litres/min that can withstand 100° C. Condensation can occur.
The drain hose must end at least 3 cm above the water surface in the well.
The drain hose must not go higher than to the lower edge of the chamber.
Diverse
Should the sterilizer be transported in temperatures below -5 °C, the overheating protection on the sterilizer may need to be reset. The error code Heating error when a process is started indicates a symptom for this.
Keep all instructions and description of operation close to the sterilizer.
Make sure the operator has read the User manual. All other documentation supplied with the sterilizer must
be looked after by the operator. The documentation shall be kept with the sterilizer until its service life comes to an end.
INSTALLATION
TECHNICAL DATA
Dimensions/weight
Width/Depth 570/700 mm Height VS60/VS90/VS130 1030/1030/1230 mm Load height VS60/VS90/VS130 940/940/1140 mm Weight VS60/VS90/VS130 100/106/115 kg Operating weight VS60/VS90/VS130 107/113/122 kg
Capacity
Chamber, diameter 444 mm Chamber, depth VS60/VS90/VS130 384/590/858 mm Chamber, volume VS60/VS90/VS130 60/90/130 L Instrument load VS60/VS90/VS130 12/24/36 kg
Electrical specifications
Electrical connection according to rating plate Control voltage according to rating plate Total output 6 kW (3 x 2 kW) Power consumption 5 A (3 x 400 V), 9 A (3 x 230 V) Heat emission while operational, closed door about 615 W Heat emission in standby, open door 0 W
Other data
Chamber, calculation pressure 3.0 bar Chamber, provisions AFS 1999:4 (ÅKN 71), Isbest PED 97/23 EEC) Chamber/door, material EN 1.4404, AISI 316L Water quality Max 4dH Water pressure 1.5 - 6 bar Water and drainage connections ½” Water consumption during steam phase 7 litres Water consumption per process 75 - 100 litres Sound level less than 53 dB
Not intended for the USA market
TECHNICAL DATA
Type designations, select
? VS60
? VS90
? VS130
Serial number:
_______________________________
Points of view, tips or criticism:
_______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________ _______________________________
FEEDBACK
Getinge Skärhamn works continuously to improve its products.
Would you like to help us by submitting your points of view, tips or criticism? Copy this page and send or fax it to us.
All suggestions that result in the improvement of our product will be rewarded. Getinge Skärhamn reserves the right to assess whether the suggestion is an im­provement
Points of view were submitted by:
Name .................................................
Telephone: .................................................
Address: .................................................
.................................................
.................................................
E-mail: .................................................
Send to:
Getinge Skärhamn AB Industrivägen 5 SE-47131 Skärhamn, Sweden Fax:+46 (0)304 600 229 E-mail: info@seska.getinge.com
Thanks in advance.
FEEDBACK
Loading...