These Operating Instructions in pdfformat are for information only.
They are not a replacement for the
Operating Instructions supplied with
the machine/device and options.
Hemodialysis System
Operating Instructions
Software version: 3.52
Edition: 5/09.06
Part no.: M38 816 1
0123
Fresenius
Medical
Care
Table of ContentsPage
1Index
2Important Information
2.1Important Information on the Operating Instructions............................................................ 2-1
2.2Important Information on the System...................................................................................... 2-3
2.1Important Information on the Operating Instructions
2.1.1How to Use the Operating Instructions
IdentificationThe document can be identified by the following information on the title
page and on the labels, if any:
– Software version of the system
– Edition of the technical document
– Part number of the technical document
Page identificationThe page identification 1-3, for example, refers to Chapter 1, page 3.
Editorial informationThe editorial information 1/01.05, for example, refers to: 1. edition,
January 2005.
ChangesDocument changes will be released as new editions or supplements. In
general, this manual is subject to change without notice.
Importance of the
instructions
Description of the optionsChapters 15 to 28 describe the operation of the options. For further
These Operating Instructions are part of the accompanying documents
and an integral part of the system. They contain information necessary
for the use of the system.
The Operating Instructions must be carefully studied before attempting
to operate the system.
Before the responsible organization may start operating the system, the
person responsible for the operation must have been instructed by the
manufacturer on how to use the system and must be thoroughly familiar
with the contents of the Operating Instructions.
The system may only be operated by persons certificated to have been
instructed on the proper operation and handling of the unit.
information please refer to the appropriate chapters. (e.g. The SN
Specifications are listed in chapter 11 System Description.)
Fresenius Medical Care 5008 OP 5/09.062-1
Chapter 2: Important Information
2.1.2Signification of the Safety Precautions
Explanation of the Caution and Note symbols used:
Caution
Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on
individuals.
Note
Informs the operator that in case of a failure to follow the steps as
described, a specific function will be executed incorrectly or will not be
executed at all, or will not produce the desired effect.
2.1.3Signification of the Highlight Symbol
Explanation on the following symbol:
Here you will find hints on easy handling.
2-2Fresenius Medical Care 5008 OP5/09.06
2.2Important Information on the System
2.2.1Brief Description
Dialysis treatments with the hemodialysis system 5008 can be
performed without any additional equipment. The hemodialysis system
controls and monitors the dialysate circuit and the extracorporeal blood
circuit.
The monitor comprises of four keys. All entries are made via a highresolution color monitor (touch screen). The current treatment data are
shown on the display.
In the dialysate circuit, product water is heated, degassed, mixed with
hemodialysis concentrate, and delivered to the dialyzer. Inflowing and
outflowing quantities are balanced volumetrically. The pressure at the
dialyzer is adjusted depending on the ultrafiltration rate selected and the
type of dialyzer used.
Chapter 2: Important Information
The blood in the extracorporeal blood circuit is transported through the
dialyzer. The blood can be continuously heparinized. An air bubble
detector prevents infusion of air. Any dangerous loss of blood is
prevented by a blood leak detector, a fluid detector and by monitoring
the venous return pressure. The arterial pressure monitoring unit
detects an aspiration of the needle in the vessel.
The hemodialysis system 5008 is designed for both acetate dialysis and
bicarbonate dialysis. The mixing ratio, the Na
bicarbonate concentration may be programmed within certain limits.
The hemodialysis system allows programming of Na and UF profiles.
ISO-UF (ultrafiltration without dialysate flow) may be performed.
The dialysate flow can be adjusted from 100 to 1000 ml/min, in
increments of 100 ml/min. The AutoFlow function automatically
regulates the dialysate flow, depending on the dialyzer type and blood
flow.
The 5008 hemodialysis system reflects the latest state of technology. It
is equipped with all safety systems required for its function and for
patient safety. It complies with the requirements of EN 60601-1 (IEC
601-1). The BPM (optional) complies with the EN 1060-1 standard for
non-invasive sphygmomanometers, Part 1 General Requirements.
+
concentration and the
The 5008 hemodialysis system is classified as Class II b (MDD)
equipment.
Fresenius Medical Care 5008 OP 5/09.062-3
Chapter 2: Important Information
2.2.2Intended Use
OFields of application
OSide effects
OContraindications
The 5008 hemodialysis system is designed for performing chronic and
acute hemodialysis. It can be used in home dialysis, hemodialysis and
limited care centers and clinical hemodialysis.
Hemodialysis therapies occasionally cause hypotension, nausea,
vomiting and cramps in some patients. In addition, the package inserts
enclosed with the consumables (e.g. hemodialysis concentrates,
dialyzers) must be observed.
– Hyperkalemia (only with potassium-containing hemodialysis
A different method of extracorporeal treatment may be indicated in
hemodynamically unstable patients.
ORestrictions
None
2.2.3Target Group
The system may only be installed, operated and used by persons with
the appropriate training, knowledge and experience.
2.2.4Duties of the Responsible Organization
The responsible organization assumes the following responsibilities:
– Compliance with the national or local installation, operation, use and
maintenance regulations
– Respect of the accident prevention regulations
– Correct and safe state of the system
– Permanent availability of the Operating Instructions
2-4Fresenius Medical Care 5008 OP5/09.06
2.2.5Disclaimer of Liability
2.2.6Guarantee / Warranty
Chapter 2: Important Information
The system has been approved for use with the consumables and
accessories listed in the Operating Instructions.
Should the responsible organization wish to use other consumables
and accessories than those listed in the Operating Instructions, the
responsibility to ensure the correct function of the system lies
exclusively with the responsible organization. The applicable legal
regulations must be complied with (e.g. in Germany the Medical Device
Directive, MDD and the MPBetreibV = German regulation for the
operation of medical products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the system resulting from the use of non-approved or unsuitable
consumables or accessories.
OGuarantee
OWarranty
2.2.7Safety Precautions
OBasic safety precautions
For guarantee refer to the respective sales contracts.
The customer's rights of warranty depend on the applicable legal
regulations.
Caution
When using a RO unit or CDS the following must be observed:
Operating Instructions of the RO unit or CDS used.
When cleaning the RO unit and its supply lines, the hemodialysis
system must be disconnected from the RO unit at the water supply.
During cleaning of the CDS distribution tubings, the hemodialysis
system must be separated from the CDS.
Fresenius Medical Care 5008 OP 5/09.062-5
Chapter 2: Important Information
OElectric hazards
Caution
The use of additional extension cables or multiway sockets / connectors
is prohibited.
2.2.8Additional Optional Equipment Supplied by Fresenius Medical Care
– DIASAFE®plus
– AquaUNO (single station ´reverse osmosis unit)
For connecting the AquaUNO to the 5008 hemodialysis system, the
two following cables must be used:
Control cable connection set: 3 meters (part no.: M37 525 1) or
Prior to the initial start-up thoroughly study the information given in
chapter 11.
The 5008 hemodialysis system must be in a perfect state. If the 5008
hemodialysis system shows signs of mechanical damage preventing
safe operation, stop using the machine. Applied parts that are damaged
must be replaced.
The following must be observed when entering parameters:
The parameters entered must be verified by the operator, i.e. the
operator must check that the values entered are correct. If the
verification reveals a deviation between the desired parameters and the
parameters displayed on the system, the setting must be corrected
before activating the function.
The actual values displayed must be compared with the desired values
specified.
The technical safety checks and technical measurement checks
required must be performed every 2 years.
2-6Fresenius Medical Care 5008 OP5/09.06
2.2.13 Repair
2.2.14 Technical Documentation
Chapter 2: Important Information
Assembly, extensions, adjustments, modifications or repairs may only
be carried out by the manufacturer or persons authorized by him.
Upon request the manufacturer will provide circuit diagrams,
descriptions, spare parts lists and other documents. These are intended
to support trained personnel in servicing and repairing the machine.
The following is also available on request:
– Test procedure by which the effectiveness of sterilization or
disinfection has been verified.
– Comments, concerning the expected recirculation of the blood flow
in the extracorporeal circuit in Single-Needle treatments, if the
recommended administration sets, dialyzers, fistula needles and
catheters are used.
Fresenius Medical Care 5008 OP 5/09.062-7
Chapter 2: Important Information
2.3Addresses
ManufacturerFresenius Medical Care AG & Co. KGaA
Please address any inquiries to:
D-61346 Bad Homburg
+49 (0)6172/609-0
www.fmc-ag.com
Service
Central Europe
International
Service
Local Service
Fresenius Medical Care
Deutschland GmbH
Geschäftsbereich Zentraleuropa
Kundendienst / Servicecenter
Steinmühlstraße 24
61352 Bad Homburg
Germany
Phone: +49 6172 609-7100
Fax: +49 6172 609-7102
E-mail: ServicecenterD@fmc-ag.com
Fresenius Medical Care
Deutschland GmbH
Service Support International
Hafenstrasse 9
D-97424 Schweinfurt
Germany
Phone: +49 9721 678-333 (hotline)
Fax: +49 9721 678-130
2-8Fresenius Medical Care 5008 OP5/09.06
3Design
3.1Front View
1
Chapter 3: Design
1Monitor
2Extracorporeal blood module
3Hydraulics
2
3
Fresenius Medical Care 5008 OP 5/09.063-1
Chapter 3: Design
3.2Rear View
1Monitor
2External connection options
3Push handle
4Fan filter (service door)
5Power connection (supply point)
1
6Line holder (for transport)
7Service door
8Hydraulic connectors
2
3
4
5
6
7
8
3-2Fresenius Medical Care 5008 OP5/09.06
3.3Lateral View, Left Side
Chapter 3: Design
1Cover, tray, cuff holder or shunt
interlock
2Concentrate rack (extractable)
3Brake
Remove the cover from the tray:
(a) Push the cover down and turn it.
(b) Pull the cover out.
a
b
To extract the concentrate rack:
1
Push with your foot from the front against the
rack.
To retract the concentrate rack:
Push with your foot from the front against the
rack until it clicks into place.
To apply or release the brake:
(a) Push the lever down to apply the brake.
(b) Push the lever down to release the brake.
a
2
3
b
Fresenius Medical Care 5008 OP 5/09.063-3
Chapter 3: Design
3.4Lateral View, Right Side
1IV pole
1
2Dialyzer holder
3Shunt door for dialysate lines
4Shunt interlock
5Dialysate return line
(dialyzer coupling blue)
6Dialysate supply line
(dialyzer coupling red)
7Concentrate rack (extractable)
8Brake
9Leakage sensor, filter chamber
2
3
4
5
6
14
13
12
11
10Particle filter, dialysate
11Filter 1 – DIASAFE
®
plus, right
12Filter 2 – ONLINEplus™, left
13Door, filter chamber
14Filter chamber
10
9
7
8
3-4Fresenius Medical Care 5008 OP5/09.06
Chapter 3: Design
To adjust the IV pole:
Push the knob (a) upwards and
b
a
simultaneously extract or retract the IV pole
(b).
Dialyzer holder:
b
Push the lever (a) to the left to insert the
c
dialyzer. The dialyzer can be moved to any
desired position (b). Press or pull the lever (c)
to swivel the dialyzer holder to the right.
a
(When the right-hand door is opened, the
dialyzer holder will automatically move to the
right.)
To open or close the shunt door:
b
Open the shunt door by flipping it to the top
(a). Close the shunt door by flipping it down
a
(b).
To remove the dialysate couplings:
Push the lever down and hold it, and remove
the dialysate coupling.
To move the hemodialysis system:
The hemodialysis system can be moved in all
directions.
Fresenius Medical Care 5008 OP 5/09.063-5
Chapter 3: Design
3.5Monitor Front
1
2
7
3
4
5
6
1Display failure sensor (hidden)
2On/Off LED/key (green)
(LED is illuminated – system in operation. LED is flashing – system
is connected to power supply, standby.)
3Blood system Stop LED/key (red)
4Blood system Start LED/key (green)
5Mute LED/key (red)
(LED is illuminated – audible alarm suppressed. LED is flashing –
audible alarm active.)
6Screen
7Operating mode indicator (green, yellow, red)
LED is green to indicate correct operation.
LED is yellow in case of a warning or an info.
LED is yellow and flashing in Emergency mode.
LED is red in case of an alarm.
LED is not illuminated during the cleaning programs.
3-6Fresenius Medical Care 5008 OP5/09.06
3.6Monitor Rear
1
2
3
Chapter 3: Design
4
1Card receptacle
(for PatientCard/UserCard/ServiceCard)
2Loudspeaker
3Recessed handle
4Monitor arm
To move the monitor:
To bring the monitor into the desired position,
it can be swiveled about three axes (a), (b),
(c).
a
b
c
(a) To move it, hold the monitor at the points
shown.
(b) Insert card.
a
b
Fresenius Medical Care 5008 OP 5/09.063-7
Chapter 3: Design
3.7Extracorporeal Blood Module
19
10
11
1
2
3
4
5
6
7
8
9
18
17
16
15
14
13
12
1Line holder
2Blood pump
3Heparin pump (if present)
4Arterial pressure measurement unit
5Substituate pump
6Arterial occlusion clamp
7Substituate catch/lock (blue)
8Substituate port, hidden by the substituate catch (blue)
9Rinse port, hidden by the rinse port catch (grey)
10Rinse port catch (grey)
11Groove
12Leakage sensor, extracorporeal blood module
13Venous occlusion clamp
14Venous monitoring function (optical detector, air bubble detector)
15Locator for venous bubble catcher
16Venous monitoring function (level detector)
3-8Fresenius Medical Care 5008 OP5/09.06
Loading...
+ 298 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.