These Operating Instructions in pdfformat are for information only.
They are not a replacement for the
Operating Instructions supplied with
the machine/device and options.
Hemodialysis System
Operating Instructions
Software version: 3.52
Edition: 5/09.06
Part no.: M38 816 1
0123
Fresenius
Medical
Care
Table of ContentsPage
1Index
2Important Information
2.1Important Information on the Operating Instructions............................................................ 2-1
2.2Important Information on the System...................................................................................... 2-3
2.1Important Information on the Operating Instructions
2.1.1How to Use the Operating Instructions
IdentificationThe document can be identified by the following information on the title
page and on the labels, if any:
– Software version of the system
– Edition of the technical document
– Part number of the technical document
Page identificationThe page identification 1-3, for example, refers to Chapter 1, page 3.
Editorial informationThe editorial information 1/01.05, for example, refers to: 1. edition,
January 2005.
ChangesDocument changes will be released as new editions or supplements. In
general, this manual is subject to change without notice.
Importance of the
instructions
Description of the optionsChapters 15 to 28 describe the operation of the options. For further
These Operating Instructions are part of the accompanying documents
and an integral part of the system. They contain information necessary
for the use of the system.
The Operating Instructions must be carefully studied before attempting
to operate the system.
Before the responsible organization may start operating the system, the
person responsible for the operation must have been instructed by the
manufacturer on how to use the system and must be thoroughly familiar
with the contents of the Operating Instructions.
The system may only be operated by persons certificated to have been
instructed on the proper operation and handling of the unit.
information please refer to the appropriate chapters. (e.g. The SN
Specifications are listed in chapter 11 System Description.)
Fresenius Medical Care 5008 OP 5/09.062-1
Chapter 2: Important Information
2.1.2Signification of the Safety Precautions
Explanation of the Caution and Note symbols used:
Caution
Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on
individuals.
Note
Informs the operator that in case of a failure to follow the steps as
described, a specific function will be executed incorrectly or will not be
executed at all, or will not produce the desired effect.
2.1.3Signification of the Highlight Symbol
Explanation on the following symbol:
Here you will find hints on easy handling.
2-2Fresenius Medical Care 5008 OP5/09.06
2.2Important Information on the System
2.2.1Brief Description
Dialysis treatments with the hemodialysis system 5008 can be
performed without any additional equipment. The hemodialysis system
controls and monitors the dialysate circuit and the extracorporeal blood
circuit.
The monitor comprises of four keys. All entries are made via a highresolution color monitor (touch screen). The current treatment data are
shown on the display.
In the dialysate circuit, product water is heated, degassed, mixed with
hemodialysis concentrate, and delivered to the dialyzer. Inflowing and
outflowing quantities are balanced volumetrically. The pressure at the
dialyzer is adjusted depending on the ultrafiltration rate selected and the
type of dialyzer used.
Chapter 2: Important Information
The blood in the extracorporeal blood circuit is transported through the
dialyzer. The blood can be continuously heparinized. An air bubble
detector prevents infusion of air. Any dangerous loss of blood is
prevented by a blood leak detector, a fluid detector and by monitoring
the venous return pressure. The arterial pressure monitoring unit
detects an aspiration of the needle in the vessel.
The hemodialysis system 5008 is designed for both acetate dialysis and
bicarbonate dialysis. The mixing ratio, the Na
bicarbonate concentration may be programmed within certain limits.
The hemodialysis system allows programming of Na and UF profiles.
ISO-UF (ultrafiltration without dialysate flow) may be performed.
The dialysate flow can be adjusted from 100 to 1000 ml/min, in
increments of 100 ml/min. The AutoFlow function automatically
regulates the dialysate flow, depending on the dialyzer type and blood
flow.
The 5008 hemodialysis system reflects the latest state of technology. It
is equipped with all safety systems required for its function and for
patient safety. It complies with the requirements of EN 60601-1 (IEC
601-1). The BPM (optional) complies with the EN 1060-1 standard for
non-invasive sphygmomanometers, Part 1 General Requirements.
+
concentration and the
The 5008 hemodialysis system is classified as Class II b (MDD)
equipment.
Fresenius Medical Care 5008 OP 5/09.062-3
Chapter 2: Important Information
2.2.2Intended Use
OFields of application
OSide effects
OContraindications
The 5008 hemodialysis system is designed for performing chronic and
acute hemodialysis. It can be used in home dialysis, hemodialysis and
limited care centers and clinical hemodialysis.
Hemodialysis therapies occasionally cause hypotension, nausea,
vomiting and cramps in some patients. In addition, the package inserts
enclosed with the consumables (e.g. hemodialysis concentrates,
dialyzers) must be observed.
– Hyperkalemia (only with potassium-containing hemodialysis
A different method of extracorporeal treatment may be indicated in
hemodynamically unstable patients.
ORestrictions
None
2.2.3Target Group
The system may only be installed, operated and used by persons with
the appropriate training, knowledge and experience.
2.2.4Duties of the Responsible Organization
The responsible organization assumes the following responsibilities:
– Compliance with the national or local installation, operation, use and
maintenance regulations
– Respect of the accident prevention regulations
– Correct and safe state of the system
– Permanent availability of the Operating Instructions
2-4Fresenius Medical Care 5008 OP5/09.06
2.2.5Disclaimer of Liability
2.2.6Guarantee / Warranty
Chapter 2: Important Information
The system has been approved for use with the consumables and
accessories listed in the Operating Instructions.
Should the responsible organization wish to use other consumables
and accessories than those listed in the Operating Instructions, the
responsibility to ensure the correct function of the system lies
exclusively with the responsible organization. The applicable legal
regulations must be complied with (e.g. in Germany the Medical Device
Directive, MDD and the MPBetreibV = German regulation for the
operation of medical products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the system resulting from the use of non-approved or unsuitable
consumables or accessories.
OGuarantee
OWarranty
2.2.7Safety Precautions
OBasic safety precautions
For guarantee refer to the respective sales contracts.
The customer's rights of warranty depend on the applicable legal
regulations.
Caution
When using a RO unit or CDS the following must be observed:
Operating Instructions of the RO unit or CDS used.
When cleaning the RO unit and its supply lines, the hemodialysis
system must be disconnected from the RO unit at the water supply.
During cleaning of the CDS distribution tubings, the hemodialysis
system must be separated from the CDS.
Fresenius Medical Care 5008 OP 5/09.062-5
Chapter 2: Important Information
OElectric hazards
Caution
The use of additional extension cables or multiway sockets / connectors
is prohibited.
2.2.8Additional Optional Equipment Supplied by Fresenius Medical Care
– DIASAFE®plus
– AquaUNO (single station ´reverse osmosis unit)
For connecting the AquaUNO to the 5008 hemodialysis system, the
two following cables must be used:
Control cable connection set: 3 meters (part no.: M37 525 1) or
Prior to the initial start-up thoroughly study the information given in
chapter 11.
The 5008 hemodialysis system must be in a perfect state. If the 5008
hemodialysis system shows signs of mechanical damage preventing
safe operation, stop using the machine. Applied parts that are damaged
must be replaced.
The following must be observed when entering parameters:
The parameters entered must be verified by the operator, i.e. the
operator must check that the values entered are correct. If the
verification reveals a deviation between the desired parameters and the
parameters displayed on the system, the setting must be corrected
before activating the function.
The actual values displayed must be compared with the desired values
specified.
The technical safety checks and technical measurement checks
required must be performed every 2 years.
2-6Fresenius Medical Care 5008 OP5/09.06
2.2.13 Repair
2.2.14 Technical Documentation
Chapter 2: Important Information
Assembly, extensions, adjustments, modifications or repairs may only
be carried out by the manufacturer or persons authorized by him.
Upon request the manufacturer will provide circuit diagrams,
descriptions, spare parts lists and other documents. These are intended
to support trained personnel in servicing and repairing the machine.
The following is also available on request:
– Test procedure by which the effectiveness of sterilization or
disinfection has been verified.
– Comments, concerning the expected recirculation of the blood flow
in the extracorporeal circuit in Single-Needle treatments, if the
recommended administration sets, dialyzers, fistula needles and
catheters are used.
Fresenius Medical Care 5008 OP 5/09.062-7
Chapter 2: Important Information
2.3Addresses
ManufacturerFresenius Medical Care AG & Co. KGaA
Please address any inquiries to:
D-61346 Bad Homburg
+49 (0)6172/609-0
www.fmc-ag.com
Service
Central Europe
International
Service
Local Service
Fresenius Medical Care
Deutschland GmbH
Geschäftsbereich Zentraleuropa
Kundendienst / Servicecenter
Steinmühlstraße 24
61352 Bad Homburg
Germany
Phone: +49 6172 609-7100
Fax: +49 6172 609-7102
E-mail: ServicecenterD@fmc-ag.com
Fresenius Medical Care
Deutschland GmbH
Service Support International
Hafenstrasse 9
D-97424 Schweinfurt
Germany
Phone: +49 9721 678-333 (hotline)
Fax: +49 9721 678-130
2-8Fresenius Medical Care 5008 OP5/09.06
3Design
3.1Front View
1
Chapter 3: Design
1Monitor
2Extracorporeal blood module
3Hydraulics
2
3
Fresenius Medical Care 5008 OP 5/09.063-1
Chapter 3: Design
3.2Rear View
1Monitor
2External connection options
3Push handle
4Fan filter (service door)
5Power connection (supply point)
1
6Line holder (for transport)
7Service door
8Hydraulic connectors
2
3
4
5
6
7
8
3-2Fresenius Medical Care 5008 OP5/09.06
3.3Lateral View, Left Side
Chapter 3: Design
1Cover, tray, cuff holder or shunt
interlock
2Concentrate rack (extractable)
3Brake
Remove the cover from the tray:
(a) Push the cover down and turn it.
(b) Pull the cover out.
a
b
To extract the concentrate rack:
1
Push with your foot from the front against the
rack.
To retract the concentrate rack:
Push with your foot from the front against the
rack until it clicks into place.
To apply or release the brake:
(a) Push the lever down to apply the brake.
(b) Push the lever down to release the brake.
a
2
3
b
Fresenius Medical Care 5008 OP 5/09.063-3
Chapter 3: Design
3.4Lateral View, Right Side
1IV pole
1
2Dialyzer holder
3Shunt door for dialysate lines
4Shunt interlock
5Dialysate return line
(dialyzer coupling blue)
6Dialysate supply line
(dialyzer coupling red)
7Concentrate rack (extractable)
8Brake
9Leakage sensor, filter chamber
2
3
4
5
6
14
13
12
11
10Particle filter, dialysate
11Filter 1 – DIASAFE
®
plus, right
12Filter 2 – ONLINEplus™, left
13Door, filter chamber
14Filter chamber
10
9
7
8
3-4Fresenius Medical Care 5008 OP5/09.06
Chapter 3: Design
To adjust the IV pole:
Push the knob (a) upwards and
b
a
simultaneously extract or retract the IV pole
(b).
Dialyzer holder:
b
Push the lever (a) to the left to insert the
c
dialyzer. The dialyzer can be moved to any
desired position (b). Press or pull the lever (c)
to swivel the dialyzer holder to the right.
a
(When the right-hand door is opened, the
dialyzer holder will automatically move to the
right.)
To open or close the shunt door:
b
Open the shunt door by flipping it to the top
(a). Close the shunt door by flipping it down
a
(b).
To remove the dialysate couplings:
Push the lever down and hold it, and remove
the dialysate coupling.
To move the hemodialysis system:
The hemodialysis system can be moved in all
directions.
Fresenius Medical Care 5008 OP 5/09.063-5
Chapter 3: Design
3.5Monitor Front
1
2
7
3
4
5
6
1Display failure sensor (hidden)
2On/Off LED/key (green)
(LED is illuminated – system in operation. LED is flashing – system
is connected to power supply, standby.)
3Blood system Stop LED/key (red)
4Blood system Start LED/key (green)
5Mute LED/key (red)
(LED is illuminated – audible alarm suppressed. LED is flashing –
audible alarm active.)
6Screen
7Operating mode indicator (green, yellow, red)
LED is green to indicate correct operation.
LED is yellow in case of a warning or an info.
LED is yellow and flashing in Emergency mode.
LED is red in case of an alarm.
LED is not illuminated during the cleaning programs.
3-6Fresenius Medical Care 5008 OP5/09.06
3.6Monitor Rear
1
2
3
Chapter 3: Design
4
1Card receptacle
(for PatientCard/UserCard/ServiceCard)
2Loudspeaker
3Recessed handle
4Monitor arm
To move the monitor:
To bring the monitor into the desired position,
it can be swiveled about three axes (a), (b),
(c).
a
b
c
(a) To move it, hold the monitor at the points
shown.
(b) Insert card.
a
b
Fresenius Medical Care 5008 OP 5/09.063-7
Chapter 3: Design
3.7Extracorporeal Blood Module
19
10
11
1
2
3
4
5
6
7
8
9
18
17
16
15
14
13
12
1Line holder
2Blood pump
3Heparin pump (if present)
4Arterial pressure measurement unit
5Substituate pump
6Arterial occlusion clamp
7Substituate catch/lock (blue)
8Substituate port, hidden by the substituate catch (blue)
9Rinse port, hidden by the rinse port catch (grey)
10Rinse port catch (grey)
11Groove
12Leakage sensor, extracorporeal blood module
13Venous occlusion clamp
14Venous monitoring function (optical detector, air bubble detector)
15Locator for venous bubble catcher
16Venous monitoring function (level detector)
3-8Fresenius Medical Care 5008 OP5/09.06
17Line holder for SafeLine™
18Line holder
19Venous pressure port
Open or close the doors on the upper side as
shown in the illustration.
Chapter 3: Design
Fresenius Medical Care 5008 OP 5/09.063-9
Chapter 3: Design
OBlood pump
1Holder (shape-coded) for line guide
1
OHeparin pump
5
2Rotor
3Handle for an emergency operation
4
3
2
4Key/ejector (for inserting and removing
the line segment)
5Line pulleys
1Barrel holder with syringe detector
2Bracket
3Fixation for the plunger
4Grip handle
5Clamping brackets
1
2
3
4
5
3-10Fresenius Medical Care 5008 OP 5/09.06
OVenous fill level and air monitoring function
1
Chapter 3: Design
1Tension lever with level detector (for the
venous bubble catcher)
2Locator for venous bubble catcher
3Optical detector
4Air bubble detector (ABD)
5Line housing
2
3
4
5
Fresenius Medical Care 5008 OP 5/09.063-11
Chapter 3: Design
3.8Extracorporeal Blood Module with Additional Functions
5
6
1
2
3
4
7
8
9
10
BPM (option)1Blood pressure cuff
2Cuff holder
3Pressure port (BPM)
4Pressure tubing
SN (option)5Single-Needle pressure port
6Single-Needle pump
7Holder for SN chamber (with mark)
11Connector for CDS 1, red (Central Delivery System) acid 1
12Connector for CDS 2, red (Central Delivery System) acid 2
(option)
13Connector for BIC, blue (central bicarbonate supply) (option)
3-14Fresenius Medical Care 5008 OP 5/09.06
Chapter 3: Design
3.11External Connection Options/Connection to Power Supply
Caution
Before connecting any optional equipment, observe the notes under
Specifications.
1
4
5
1LAN (local area network) network connection
2Service/diagnostics, RS232, 24 V
Connector for AquaUNO (single station reverse osmosis unit)
3Alarm output (staff call)
4Power connection (supply point)
23
5Power switch
Fresenius Medical Care 5008 OP 5/09.063-15
Chapter 3: Design
3-16Fresenius Medical Care 5008 OP 5/09.06
4Graphical User Interface
4.1After Turning Power on to the System
START-UP SCREEN
The display shows the machine type, the
current software version and the clinical data
(on request) for approx. 15 seconds.
Chapter 4: Graphical User Interface
SELECTION SCREEN
The following selections are possible:
– Treatment
– Cleaning program (e.g. Rinse)
Touch the desired button to make your
selection.
Fresenius Medical Care 5008 OP 5/09.064-1
Chapter 4: Graphical User Interface
4.2Overview (Screen)
Operating
mode
12345678
16
Pressure
displays
15
Menu
section
Status
indicator
Dialysate flow
StatusBlood flowPatient IDStatus
Message
button
Info
indicator
Heparin
9
UF Timer I/O
10
Emergency
Button
11
Blood
pressure
12
Options
menus
13
HEPARIN
14
Menu buttons
1Operating mode
Displays the operating mode of the system (e.g. Dialysis).
In addition, a progress bar is displayed, depending on the
operating mode, e.g. in the Rinse mode.
2Dialysate flow status indicator
– Flow turned on – waves green (grey bar is moving.)
– Bypass – waves green (grey bar is not moving.)
– Flow turned off – waves grey
3Status
Displays data on the system condition. (Software, error memory,
cleaning status, system info)
4Message button
Allows retrieval of information, warnings and alarms (3 maximum)
4-2Fresenius Medical Care 5008 OP5/09.06
Chapter 4: Graphical User Interface
5Info
Displays information on the current procedure.
6Heparin status indicator
– Pump switched on – drop green
(Grey bar is moving.)
– Pump switched off – drop grey
7Patient ID (patient identification)
Treatment data sheet will be displayed.
Combined with the use of the patient card, it is possible to retrieve
current treatment data. Storage of 3 previous treatments.
8Blood flow
Displays the effective blood flow.
Rocker switch for increasing + / reducing – the effective blood flow.
9UF Timer I/O
Button for starting/stopping ultrafiltration and the timer function.
10Emergency button
11Blood pressure
(Displayed only, if BPM option is available.)
12Options menus
Via the OPTIONS menu button, it is possible to program up to
four option menus with direct access.
13HEPARIN
(displayed only, if selected in the Operator Setup)
14Menu buttons
Corresponding menu opens automatically during operation
OR
touch button for opening the respective menu.
15Menu section
In the center of the screen, the appropriate data for each menu is
displayed.
Indicators/buttons/diagrams/graphics are displayed depending on
the Setup settings.
16Pressure displays
ART (arterial pressure)
VEN (venous pressure)
The actual value is displayed as a numerical value and as a bar.
The alarm window is displayed in block representation, according
to the window size.
Touch the ART or VEN field for setting the alarm limits.
4.3General Operation Philosophy
It is possible to control all treatment sections via the screen menu.
Fresenius Medical Care 5008 OP 5/09.064-3
Chapter 4: Graphical User Interface
OScreen colors
The fields in the header bar are:
Grey in the normal operating mode
Orange during the functional test (T1 test)
Orange during rinse procedure of the extracorporeal blood circuit, until the minimum
rinse volume has been reached.
Yellow during the cleaning programs
BLUE
Selection possible
GREEN
Active
GREY
Not active
Example
Examples
UF goal value field
UF MENU button
Examples
UF Timer I/O indicator
TREATMENT button
Emergency menu I/O indicator
Selection not possible
Example
CLEANING button
4-4Fresenius Medical Care 5008 OP5/09.06
OCourse of the treatment
Chapter 4: Graphical User Interface
9 menu buttons in 3-D-design are placed at the bottom screen bar,
representing the chronological course of operation. The change to the
corresponding menus is performed automatically when the respective
conditions have been fulfilled. (Exception: DIALYSATE MENU, UF MENU, OPTIONS and SYSTEM)
Fresenius Medical Care 5008 OP 5/09.064-5
Chapter 4: Graphical User Interface
ODesign of the menu structure
Treatment data may be
changed directly on the main
screen.
Touching the OK button
accepts changed data.
To enter data for more
parameters, touch this OK
level button to accept the
changed data and to open
the respective menu.
4-6Fresenius Medical Care 5008 OP5/09.06
Chapter 4: Graphical User Interface
4.4Examples for Data Entry (Treatment Data)
OSetting via the numeric keypad (for example setting the prescribed Na)
1
2
4
3
1. Touch the Prescr. Na field.
2. Enter the desired prescribed Na via the keypad.
Check the entered value (prescribed value).
(Grey keys prevent implausible entries.)
3. Touch the OK button to accept the entered value.
Visually check the accepted value.
4. Touch the C button for making corrections.
Fresenius Medical Care 5008 OP 5/09.064-7
Chapter 4: Graphical User Interface
OSetting via the rocker switch (for example changing the venous alarm limits)
132a2b45
1. Touch the VEN field.
2. a, Adjustment of window width – left
b, Adjustment of window position – right
3. Adjust the desired alarm window via the
+/– rocker switch.
Check the entered alarm window value in the venous pressure
display (prescribed value).
4. Touch the OK button to accept the selected alarm window.
Visually check the accepted alarm window.
5. Touch the C button for making corrections.
4-8Fresenius Medical Care 5008 OP5/09.06
4.5Screen Saver
Chapter 4: Graphical User Interface
SCREEN SAVER
Displays the following data:
– the arterial and the venous pressures
– the UF parameters goal, rate and volume
– the effective blood flow
– the remaining treatment time in the center
– the last measured blood pressure
(Only if the BPM system option exists.)
– the BVM rate and - under UF goal - the maximum UF goal = “+“ as
well as the minimum UF goal = “-“
(Only if the BVM system option exists.)
It is only displayed during the treatment, following a certain timed delay
after the last screen action. (Timed delay adjustable in the Operator
setup.)
The SCREEN SAVER disappears when any part of the screen is
touched.
The SCREEN SAVER disappears immediately:
– when a message is given (info, warning or alarm),
– when the BPM (optional) starts a measurement.
Fresenius Medical Care 5008 OP 5/09.064-9
Chapter 4: Graphical User Interface
4-10Fresenius Medical Care 5008 OP 5/09.06
Fold-Out Sheet
5Preparation
Fresenius Medical Care 5008 OP 5/09.06
Connecting the concentrate container (e.g. acid)PREPARATION SCREEN
Connecting the bag (e.g. bibag®)
BLOOD SYSTEM SCREEN
DIALYSATE SCREEN
UF SCREEN
PREPARATION SCREEN
HEPARIN SCREEN
5Preparation
5.1Preparation using ONLINEplus™
Irrespective of the treatment mode, all 5008 ONLINEplus™
hemodialysis systems can be operated without rinse or infusion
solutions provided in NaCl bags. The fluid volumes required for
preparation, bolus administration or during reinfusion will then be
produced ONLINE by the 5008 hemodialysis system according to the
actual requirements, thus saving both cost and time.
5.1.1Turning the Hemodialysis System On
Caution
Chapter 5: Preparation
The stability of the 5008 hemodialysis system must be ensured.
Establish the water and power supply.
Press the On/Off key. (Turn the hemodialysis system on!)
The On/Off LED is illuminated.
START-UP SCREEN
The display shows the machine type, the current software version and
the clinical data (on request) for approx. 15 seconds.
Caution
After a downtime of more than 72 hours, a cleaning program must be
performed completely before starting the treatment.
If necessary, check the hemodialysis system for presence of residual
disinfectant. (see chapter 8 Cleaning).
Note
If the message: Defective battery is acknowledged by pressing the Skip
key, it might be that the audible alarm will not be generated, if a power
failure occurs.
Fresenius Medical Care 5008 OP 5/09.065-1
Chapter 5: Preparation
5.1.2The Following Must be Observed when Using Consumables
Caution
The system has been approved for use with the consumables and
accessories listed in the Operating Instructions.
Should the responsible organization wish to use other consumables
and accessories than those listed in the Operating Instructions, the
responsibility to ensure the correct function of the system lies
exclusively with the responsible organization. The applicable legal
regulations must be complied with (e.g. in Germany the Medical Device
Directive, MDD and the MPBetreibV = German regulation for the
operation of medical products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the system resulting from the use of non-approved or unsuitable
consumables or accessories.
Caution
The symbols printed on the packaging of the consumables have to be
observed. The symbols are described in the chapter System
Description (consumables symbols).
When using consumables, it is important to take note of the following
symbols:
Do not reuse
2
Use by
Caution
The consumables may only be used if the packaging and the respective
consumable including the protective caps used are not damaged. The
protective caps must not have fallen off.
The plastics used for the consumables may not be compatible with
components of drugs or disinfectants. If they are planned to be used,
the compatibility of the consumables' components must be ensured
before the treatment. If connectors made of polycarbonate are for
example exposed to aqueous solutions with the pH value > 10 or to
aliphatic solutions this will cause tension cracks.
5-2Fresenius Medical Care 5008 OP5/09.06
5.1.3Selecting the Concentrate Supply
OConnecting the concentrates
Caution
Concentrate:
The concentrate displayed on the screen must comply with the
specifications mentioned on the acid or the acetate container or on the
bag. This also applies to the concentrate composition in CDS operation.
Concentrate packages:
– Assure that the packages used contain sufficient concentrate to
complete the treatment.
– Use only the dedicated coded containers or the bibag
bicarbonate dialysis.
Bicarbonate dry concentrate bibag
Only the bibag
The bibag
Only use the bibag
concentrate according to the prescribed dilution. Other mixing ratios
may lead to a hazard for the patient.
Chapter 5: Preparation
®
®
®
manufactured by Fresenius Medical Care may be used.
®
must only be used for one treatment.
®
in combination with acid bicarbonate hemodialysis
:
for
Acid and basic bicarbonate hemodialysis concentrate have to be diluted
immediately prior to application only. The bag's content must be used
up within 12 hours after dilution. Discard residual volumes. The powder
is non-pyrogenic.
Conductivity limits:
The alarm limits are automatically set around the expected value.
The actual value of the conductivity display must have attained the
expected desired value after a maximum of 10 minutes.
Should this not be the case, the actual value must first be checked in
the laboratory. Change or check the concentrate, if necessary, or call
service.
Note
The bicarbonate suction tube must be inserted into the rinse chamber
during the bibag
®
treatment.
Extract the concentrate rack.
Bicarbonate dialysisTo connect the (acid) concentrate container:
Push the latch (1) upwards. Open the concentrate flap. Place the red
concentrate suction tube (2) into the acid container. Close the
concentrate flap (3) until it clicks into place.
Fresenius Medical Care 5008 OP 5/09.065-3
Chapter 5: Preparation
CDS, Central Delivery
System (option)
Acetate dialysisConnect the concentrate container.
To connect the bibag®:
Push the latch (1) upwards. Open the bicarbonate flap. Remove the
®
bibag
bibag
from its packaging. Remove the foil from the bibag®. Attach the
®
(2). Close the bicarbonate flap (3) until it clicks into place.
OR
To connect the bicarbonate container:
Insert the bicarbonate suction tube (blue) into the bicarbonate
container.
Close the bicarbonate flap.
Note
The responsible organization is responsible for the proper installation
and function of the CDS.
Insert the concentrate suction tube (red) into the acetate container.
The bicarbonate suction tube (blue) remains in the rinse chamber.
Selecting a treatmentSELECTION SCREEN
Touch the Treatment field.
If there is no tubing system inserted, the system automatically moves to
the BLOOD SYSTEM screen.
BLOOD SYSTEM SCREEN
The T1 test is now running in parallel with the preparation of the
hemodialysis system. The color of the header bar is orange for the
duration of the T1 test.
The operating mode display shows the progress of the T1 test.
Message: T1 test completed is displayed for a moment after successful
completion of the T1 test.
5-4Fresenius Medical Care 5008 OP5/09.06
Chapter 5: Preparation
5.1.4Important Items to be Considered Before and During the Treatment
Caution
Aseptic technique:
Use aseptic technique for all bloodside connections and all connections
in the area where sterile solutions are to be used.
Caution
Preventing contamination:
Use tubing systems with hydrophobic filters at the pressure lines to
prevent cross-contamination.
Connect the hydrophobic filters so that an ingress or loss of air is not
possible and that any wetting by fluid is reliably avoided, also in case of
pressure fluctuations.
If a hydrophobic filter has become wet, the tubing system must be
replaced.
On tubing systems with additional connection sites, a replacement
pressure line may be connected (accessory available from Fresenius
Medical Care).
The blood in the pressure line must not be forced back by means of a
syringe. This could damage the hydrophobic membrane and thus lead
to a contamination.
If fluid may have passed the hydrophobic filter, the system must be
checked for contamination after completion of the treatment. If the
system is contaminated, it has to be taken out of service. All affected
components have to be disinfected or replaced in accordance with the
manufacturer's specifications before the system is put into operation
again.
Fresenius Medical Care 5008 OP 5/09.065-5
Chapter 5: Preparation
Caution
When inserting the tubing systems, the following precautions
must be respected:
– The tubing systems have to be free of kinks, tension and twists and
must not be jammed (risk of hemolysis). Use the line holders
provided.
– Ensure the correct position of the screwed connections, especially
of the connection sites to the patient, the dialyzer and the system
and check or correct them during the treatment if necessary. Take
the appropriate measures if required (e.g. retightening of the Luer
Lock connection or replacement of the tubing system).
– Check the protective caps for tight fit and tighten them if necessary.
– The lines for the supply of infusions should always be clamped,
except if they are needed.
– During long-term operation, the blood lines must be changed after
24 hours at the latest.
– Do not use cannulas with a diameter of > 20 gauge to pierce the
septum of the injection sites. Insert the cannula vertically and in the
center of the septum. Disinfect the injection sites with 70% alcohol
before use.
– The blood pump must be set to the diameter of the pump segment,
refer also to the product label of the blood lines. If a wrong line
diameter is set, this may cause significant deviations in the blood
flow and thus in the dialysis dose.
– Materials which come directly or indirectly into contact with blood
polycarbonate, latex-free rubber, ABS.
– The minimum temperature of the tubing systems during use is 18 °C.
Caution
Delivery operation of the pump(s) with open doors
(blood pump, substituate pump, optional Single-Needle pump):
When the doors are open and the rotor of the pump(s) is running, make
sure that no objects, such as fingers, hair or ball point pens, come into
contact with the rotor (risk of injury).
Arterial pressure measurement unit:
Prevent foreign objects from coming into contact with the arterial
pressure measurement unit.
Heparin pump:
If heparin syringes of third party suppliers are used, the operator is
responsible and has to ensure that the syringe data displayed match the
actual data.
Heparin syringes without Luer lock are not recommended as the
connection between the heparin syringe and the blood lines may come
loose. If heparin syringes without Luer lock are used it is the operator’s
responsibility to ensure that the connection between the heparin syringe
and the blood lines does not loosen inadvertently.
IV pole:
Securely fix bags or other objects to be hung from the IV pole.
5-6Fresenius Medical Care 5008 OP5/09.06
Chapter 5: Preparation
Caution
Before the treatment, check:
– The safe connection of all connection sites of the tubing system.
– The tightness of the tubing system during and after priming.
– Retighten the connections and replace the tubing system, if
necessary.
– The absence of air in the tubing system and the correct position of
all fluid levels.
To be observed when working on the tubing system during the
treatment:
If the position of the tubing system or of one of its components is
changed, the correct position of the entire tubing system must be
restored afterwards, above all the correct position of the line guides.
During the treatment check at appropriate intervals:
– The condition of the patient.
– The function of the hemodialysis system and the extracorporeal
blood circuit. Pay particular attention to the venous insertion site, as
a possible dislocation of the venous cannula may not always be
detected by the pressure monitoring system.
– The tubing system for leakages or possible loosening of connections
as well as entry of air.
– The fluid level in the venous bubble catcher. Correct it, if required
(desired level: approx. 1 cm below the upper edge of the cover)
Caution
Venous alarm limit:
The lower venous alarm limit must be set as close as possible to the
actual venous pressure value.
Note
The dialyzer holder is not suitable for rectangular plate dialyzers.
Note
For hygienic reasons, the blood lines should be inserted immediately
prior to the treatment only.
If the blood lines were inserted more than 8 hours before the treatment,
malfunctions may occur. Correcting these malfunctions may require
removing the present blood lines and inserting new blood lines.
Fresenius Medical Care 5008 OP 5/09.065-7
Chapter 5: Preparation
5.1.5Preparing the Extracorporeal Blood Circuit with ONLINEplus™
OInserting the arterial and the venous tubing system
When inserting the tubing system, follow the description of the menu
displayed.
Wings of the
syringe cylinder
Thumb rest of the
syringe plunger
16
17
12
2
6
1
4
13
5
10
14
11
15
3
7
8
9
Open the doors of the Extracorporeal Blood Module.
1 Insert the line guide into the blood pump until a signal is sounded.
The arterial pressure measurement unit is opened.
(After closing the doors the pump segment will be automatically
inserted into the blood pump.)
2 Insert the arterial blood line into the line holder.
3 Connect the arterial blood line to the lower port of the dialyzer.
4 Insert the arterial pressure dome into the arterial pressure
measurement unit.
5 Insert the arterial blood line into the arterial occlusion clamp.
5-8Fresenius Medical Care 5008 OP5/09.06
Chapter 5: Preparation
Caution
If no heparin syringe is used, retighten the cap of the heparin line and
close the clamp.
6 Connect the heparin syringe to the arterial tubing system.
To place the heparin syringe in the holder:
Press on the clamping brackets to move the grip handle to its lower
position.
Place the heparin syringe between the barrel holders.
The syringe wings must be positioned between the barrel holders
and the bracket.
The thumb rest of the syringe plunger now must be positioned
between the clamps of the grip handle.
Press on the clamping brackets to move the grip handle to its
starting position.
Caution
When inserting the heparin syringe, the following precautions must be
respected:
– Only use heparin syringes with a volume of up and equal to 30 ml.
– Ensure that the heparin syringe is correctly inserted into the heparin
pump and locked. Follow the description and the illustration.
7 Insert the venous bubble catcher into the level detector. Mind the
locator for the venous bubble catcher.
8 Insert the venous line into the optical detector/air bubble detector.
The line must be positioned completely inside the line housing.
Caution
The following must be observed for the air bubble detector:
– No ultrasound-conducting objects and agents may be used.
– The line housing must be clean and dry.
9 Insert the venous blood line into the venous occlusion clamp.
10 Insert the venous blood line into the line holder.
11 Connect the venous blood line to the upper port of the dialyzer.
12 Connect the venous pressure line to the venous pressure connector.
Caution
Tightly connect the hydrophobic filter of the venous pressure line to the
venous pressure connector in order to avoid an ingress or loss of air.
Inserting the lines may be continued only after successful completion of
the T1 test.
13 Insert the SafeLine™ line guide into the substituate pump until a
signal is sounded.
(After closing the doors the SafeLine™ pump segment will be
automatically inserted into the substituate pump.)
14 Insert the SafeLine™ into the line holder.
Fresenius Medical Care 5008 OP 5/09.065-9
Chapter 5: Preparation
15 Connect the arterial patient access line to the SafeLine™.
b
a
c
16 To connect the substituate connector to the substituate port:
The catch for the substituate port (blue) is in initial position (a).
Pull and turn the substituate catch (blue) counterclockwise into
position (b).
Push the substituate connector firmly into the substituate port.
Pull and turn the substituate catch (blue) clockwise until it clicks into
place at position (c). If necessary to make sure of a tight closure
push in the substituate catch (blue).
a
b
c
17 To insert the rinse connector (connected onto the venous patient
line) into the rinse port:
The catch for the rinse port (grey) is in initial position (a).
Pull and turn the rinse port catch (grey) counterclockwise into
position (b).
Push the rinse connector firmly into the rinse port.
Pull and turn the rinse port catch (grey) clockwise until it clicks into
place at position (c). If necessary to make sure of a tight closure
push in the rinse port catch (grey).
Close the doors.
(The line segment(s) are automatically inserted, the arterial pressure
measurement unit closes.)
The system automatically switches to the PREPARATION SCREEN.
Message: Connect dialyzer couplings!
5-10Fresenius Medical Care 5008 OP 5/09.06
5.1.6Connecting the Dialysate Lines
Note
The dialysate lines may only be connected after the T1 test has been
completely terminated.
Message: Connect dialyzer couplings!
Open the shunt door.
Connect the dialysate supply line (dialyzer coupling red) to the dialyzer
(on the side of the venous blood outlet port).
Connect the dialysate return line (dialyzer coupling blue) to the dialyzer
(on the side of the arterial blood inlet port).
Close the shunt door.
5.1.7Priming the Extracorporeal Blood Circuit with ONLINEplus™
Chapter 5: Preparation
OStarting the Rinse procedure
PREPARATION SCREEN
Check/set the rinse volume.
Check/set the delivery rate of the blood pump.
The rinse volume and the delivery rate are automatically set to the value
preselected in the Operator Setup. Change the rinse volume and the
delivery rate if necessary.
Message: Do not connect patient! Minimum rinse volume not reached.
– Rinse Continue
If not yet done, the parameters for dialysate, UF and the
heparin pump now may be checked/set.
Touch the Continue button to continue rinsing. (Blood pump I/O
indicator green.)
Fresenius Medical Care 5008 OP 5/09.065-11
Chapter 5: Preparation
ORinse procedure completed
Endless rinse will start when the Online rinse volume and the Online UF
rinse volume have been reached.
The delivery rate of the blood pump is automatically reduced to
50 ml/min, and the dialysate flow is reduced to EcoFlow (100 ml/min).
In case of ONLINEplus™, the total flow is composed of EcoFlow
(100 ml/min) and the respective substituate rate.
5.1.8Checking/Setting the Dialysate Parameters
Note
Calcium carbonate precipitations may occur in the bicarbonate dialysis,
depending on the use and the dose of the concentrates and the duration
of the treatment. Detailed information will be provided by the
manufacturer on request.
The Prescr. Bic can be reduced to 25.0 mmol/l during preparation
(adjustable in the Technician's Setup). If the Bic reduction is selected,
25.0 mmol/l appears below the value Prescr. Bic. The Bic reduction is
de-activated when the treatment starts.
After completion of the T1 test, the EcoFlow (100 ml/min) is
automatically selected. The flow can be changed as desired.
On the PREPARATION SCREEN you can directly check,
select and change the Prescr. Na and Prescr. Bic
parameters.
In the DIALYSATE MENU
Check the dialysate parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.1.9Checking/Setting the UF Parameters
On the PREPARATION SCREEN you can directly check,
select and change the UF goal, UF time and UF rate
parameters.
5-12Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
Note
If only the UF rate is entered for ultrafiltration (instead of volume and
time), check the UF rate displayed in the Dialysis menu for plausibility
after saving the data.
In the UF MENU
Possible setting variants:
– UF goal/UF time (UF rate is calculated)
– UF goal/UF rate (UF time is calculated)
– UF rate/UF time
–UF rate
–Time
– UF profiles
– ISO goal/ISO time (ISO rate is calculated)
– ISO goal/ISO rate (ISO time is calculated)
The following must be observed in case of ISO-UF:
The ISO-UF treatment type can be started at any time and can be
repeated as often as necessary.
The parameters entered at the beginning of the treatment (UF goal and
UF time) must be taken into consideration.
In case of a combination with UF and Na profiles, first enter the ISO UF
parameters. Then set the respective profiles.
The total volume to be removed (UF goal), the total treatment time (UF
time) or the UF goal and the UF rate must always be programmed. The
ISO data goal and time cannot be higher than the UF goal/time.
The UF goal/UF time or UF goal/UF rate parameters must first be
entered.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.1.10 Checking/Setting the Sodium and UF Profiles
Caution
When using Na profiles, the following precautions must be observed:
The balancing neutrality of the profiles was computed for a dialysis dose
of KT/V = 1.2. In case of higher deviations (KT/V > 1.4; KT/V < 1.0) the
balancing neutrality may not always be achieved.
Basic requirements for setting the profiles:
the UF parameters must have been set.
Fresenius Medical Care 5008 OP 5/09.065-13
Chapter 5: Preparation
ONa profiles
In the DIALYSATE MENU
Check the dialysate parameters.
Minimum value that can be set for Start Na: 3 mmol higher than the
prescribed Na
Set the desired profile.
Check/set Start Na (maximum value).
(The minimum Na value is automatically adjusted.)
The following must be observed:
– The treatment may also be started with the Na profile only.
– It is only possible to select matching profile groups.
(E.g. if UF profile 1 has been selected, only Na profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
Concentrate, Prescr. Na and Prescr. Bic parameters.
– The minimum UF time must be set:
Profiles 1, 2: UF time 2:00 hrs
Profile 3: UF time 3:30 hrs
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
OUF profiles
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
In the UF MENU
Check the UF parameters.
Minimum UF parameters that can be set:
Profiles 1, 2: UF goal 200 ml, UF time 2:00 hrs, UF rate 10 ml
Profile 3: UF goal 200 ml, UF time 3:30 hrs, UF rate 10 ml
Set the desired profile.
Check/set the start rate.
(The minimum profile rate is automatically adjusted.)
The following must be observed:
– The treatment may also be started with the UF profile only.
– It is only possible to select matching profile groups.
(E.g. if Na profile 1 has been selected, only UF profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
UF time and UF rate parameters.
– ISO UF parameters and Na parameters may be selected only before
starting or after ending a profile.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5-14Fresenius Medical Care 5008 OP 5/09.06
5.1.11 Checking/Setting the Heparin Pump Parameters
Caution
Administer the heparin dose according to the physician's instructions.
In the HEPARIN menu
Check the heparin pump parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
Chapter 5: Preparation
Fresenius Medical Care 5008 OP 5/09.065-15
Chapter 5: Preparation
5-16Fresenius Medical Care 5008 OP 5/09.06
5.2Preparation with Rinse Solution Bag
If the 5008 ONLINEplus™ hemodialysis system cannot be operated
without rinse solution bags, rinse or infusion solutions provided in NaCl
bags may be used instead.
5.2.1Turning the Hemodialysis System On
Caution
The stability of the 5008 hemodialysis system must be ensured.
Establish the water and power supply.
Press the On/Off key. (Turn the hemodialysis system on!)
The On/Off LED is illuminated.
START-UP SCREEN
Chapter 5: Preparation
The display shows the machine type, the current software version and
the clinical data (on request) for approx. 15 seconds.
Caution
After a downtime of more than 72 hours, a cleaning program must be
performed completely before starting the treatment.
If necessary, check the hemodialysis system for presence of residual
disinfectant. (see chapter 8 Cleaning).
Note
If the message: Defective battery is acknowledged by pressing the Skip
key, it might be that the audible alarm will not be generated, if a power
failure occurs.
Fresenius Medical Care 5008 OP 5/09.065-17
Chapter 5: Preparation
5.2.2The Following Must be Observed when Using Consumables
Caution
The system has been approved for use with the consumables and
accessories listed in the Operating Instructions.
Should the responsible organization wish to use other consumables
and accessories than those listed in the Operating Instructions, the
responsibility to ensure the correct function of the system lies
exclusively with the responsible organization. The applicable legal
regulations must be complied with (e.g. in Germany the Medical Device
Directive, MDD and the MPBetreibV = German regulation for the
operation of medical products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the system resulting from the use of non-approved or unsuitable
consumables or accessories.
Caution
The symbols printed on the packaging of the consumables have to be
observed. The symbols are described in the chapter System
Description (consumables symbols).
When using consumables, it is important to take note of the following
symbols:
Do not reuse
2
Use by
Caution
The consumables may only be used if the packaging and the respective
consumable including the protective caps used are not damaged. The
protective caps must not have fallen off.
The plastics used for the consumables may not be compatible with
components of drugs or disinfectants. If they are planned to be used,
the compatibility of the consumables' components must be ensured
before the treatment. If connectors made of polycarbonate are for
example exposed to aqueous solutions with the pH value > 10 or to
aliphatic solutions this will cause tension cracks.
5-18Fresenius Medical Care 5008 OP 5/09.06
5.2.3Selecting the Concentrate Supply
OConnecting the concentrates
Caution
Concentrate:
The concentrate displayed on the screen must comply with the
specifications mentioned on the acid or the acetate container or on the
bag. This also applies to the concentrate composition in CDS operation.
Concentrate packages:
– Assure that the packages used contain sufficient concentrate to
complete the treatment.
– Use only the dedicated coded containers or the bibag
bicarbonate dialysis.
Bicarbonate dry concentrate bibag
Only the bibag
The bibag
Only use the bibag
concentrate according to the prescribed dilution. Other mixing ratios
may lead to a hazard for the patient.
Chapter 5: Preparation
®
®
®
manufactured by Fresenius Medical Care may be used.
®
must only be used for one treatment.
®
in combination with acid bicarbonate hemodialysis
:
for
Acid and basic bicarbonate hemodialysis concentrate have to be diluted
immediately prior to application only. The bag's content must be used
up within 12 hours after dilution. Discard residual volumes. The powder
is non-pyrogenic.
Conductivity limits:
The alarm limits are automatically set around the expected value.
The actual value of the conductivity display must have attained the
expected desired value after a maximum of 10 minutes.
Should this not be the case, the actual value must first be checked in
the laboratory. Change or check the concentrate, if necessary, or call
service.
Note
The bicarbonate suction tube must be inserted into the rinse chamber
during the bibag
®
treatment.
Extract the concentrate rack.
Bicarbonate dialysisTo connect the (acid) concentrate container:
Push the latch (1) upwards. Open the concentrate flap. Place the red
concentrate suction tube (2) into the acid container. Close the
concentrate flap (3) until it clicks into place.
Fresenius Medical Care 5008 OP 5/09.065-19
Chapter 5: Preparation
CDS, Central Delivery
System (option)
Acetate dialysisConnect the concentrate container.
To connect the bibag®:
Push the latch (1) upwards. Open the bicarbonate flap. Remove the
®
bibag
bibag
from its packaging. Remove the foil from the bibag®. Attach the
®
(2). Close the bicarbonate flap (3) until it clicks into place.
OR
To connect the bicarbonate container:
Insert the bicarbonate suction tube (blue) into the bicarbonate
container.
Close the bicarbonate flap.
Note
The responsible organization is responsible for the proper installation
and function of the CDS.
Insert the concentrate suction tube (red) into the acetate container.
The bicarbonate suction tube (blue) remains in the rinse chamber.
Selecting a treatmentSELECTION SCREEN
Touch the Treatment field.
If there is no tubing system inserted, the system automatically moves to
the BLOOD SYSTEM screen.
BLOOD SYSTEM SCREEN
The T1 test is now running in parallel with the preparation of the
hemodialysis system. The color of the header bar is orange for the
duration of the T1 test.
The operating mode display shows the progress of the T1 test.
Message: T1 test completed is displayed for a moment after successful
completion of the T1 test.
5-20Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
5.2.4Important Items to be Considered Before and During the Treatment
Caution
Aseptic technique:
Use aseptic technique for all bloodside connections and all connections
in the area where sterile solutions are to be used.
Caution
Preventing contamination:
Use tubing systems with hydrophobic filters at the pressure lines to
prevent cross-contamination.
Connect the hydrophobic filters so that an ingress or loss of air is not
possible and that any wetting by fluid is reliably avoided, also in case of
pressure fluctuations.
If a hydrophobic filter has become wet, the tubing system must be
replaced.
On tubing systems with additional connection sites, a replacement
pressure line may be connected (accessory available from Fresenius
Medical Care).
The blood in the pressure line must not be forced back by means of a
syringe. This could damage the hydrophobic membrane and thus lead
to a contamination.
If fluid may have passed the hydrophobic filter, the system must be
checked for contamination after completion of the treatment. If the
system is contaminated, it has to be taken out of service. All affected
components have to be disinfected or replaced in accordance with the
manufacturer's specifications before the system is put into operation
again.
Fresenius Medical Care 5008 OP 5/09.065-21
Chapter 5: Preparation
Caution
When inserting the tubing systems, the following precautions
must be respected:
– The tubing systems have to be free of kinks, tension and twists and
must not be jammed (risk of hemolysis). Use the line holders
provided.
– Ensure the correct position of the screwed connections, especially
of the connection sites to the patient, the dialyzer and the system
and check or correct them during the treatment if necessary. Take
the appropriate measures if required (e.g. retightening of the Luer
Lock connection or replacement of the tubing system).
– Check the protective caps for tight fit and tighten them if necessary.
– The lines for the supply of infusions should always be clamped,
except if they are needed.
– During long-term operation, the blood lines must be changed after
24 hours at the latest.
– Do not use cannulas with a diameter of > 20 gauge to pierce the
septum of the injection sites. Insert the cannula vertically and in the
center of the septum. Disinfect the injection sites with 70% alcohol
before use.
– The blood pump must be set to the diameter of the pump segment,
refer also to the product label of the blood lines. If a wrong line
diameter is set, this may cause significant deviations in the blood
flow and thus in the dialysis dose.
– Materials which come directly or indirectly into contact with blood
polycarbonate, latex-free rubber, ABS.
– The minimum temperature of the tubing systems during use is 18 °C.
Caution
Delivery operation of the pump(s) with open doors
(blood pump, substituate pump, optional Single-Needle pump):
When the doors are open and the rotor of the pump(s) is running, make
sure that no objects, such as fingers, hair or ball point pens, come into
contact with the rotor (risk of injury).
Arterial pressure measurement unit:
Prevent foreign objects from coming into contact with the arterial
pressure measurement unit.
Heparin pump:
If heparin syringes of third party suppliers are used, the operator is
responsible and has to ensure that the syringe data displayed match the
actual data.
Heparin syringes without Luer lock are not recommended as the
connection between the heparin syringe and the blood lines may come
loose. If heparin syringes without Luer lock are used it is the operator’s
responsibility to ensure that the connection between the heparin syringe
and the blood lines does not loosen inadvertently.
IV pole:
Securely fix bags or other objects to be hung from the IV pole.
5-22Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
Caution
Before the treatment, check:
– The safe connection of all connection sites of the tubing system.
– The tightness of the tubing system during and after priming.
– Retighten the connections and replace the tubing system, if
necessary.
– The absence of air in the tubing system and the correct position of
all fluid levels.
To be observed when working on the tubing system during the
treatment:
If the position of the tubing system or of one of its components is
changed, the correct position of the entire tubing system must be
restored afterwards, above all the correct position of the line guides.
During the treatment check at appropriate intervals:
– The condition of the patient.
– The function of the hemodialysis system and the extracorporeal
blood circuit. Pay particular attention to the venous insertion site, as
a possible dislocation of the venous cannula may not always be
detected by the pressure monitoring system.
– The tubing system for leakages or possible loosening of connections
as well as entry of air.
– The fluid level in the venous bubble catcher. Correct it, if required
(desired level: approx. 1 cm below the upper edge of the cover)
Caution
Venous alarm limit:
The lower venous alarm limit must be set as close as possible to the
actual venous pressure value.
Note
The dialyzer holder is not suitable for rectangular plate dialyzers.
Note
For hygienic reasons, the blood lines should be inserted immediately
prior to the treatment only.
If the blood lines were inserted more than 8 hours before the treatment,
malfunctions may occur. Correcting these malfunctions may require
removing the present blood lines and inserting new blood lines.
Fresenius Medical Care 5008 OP 5/09.065-23
Chapter 5: Preparation
5.2.5Preparing the Extracorporeal Blood Circuit with Rinse Solution Bag
OInserting the arterial and the venous tubing system
When inserting the tubing system, follow the description of the menu
displayed.
13
12
2
11
Wings of the
syringe cylinder
Thumb rest of the
syringe plunger
7
1
4
5
3
8
9
10
14
6
Open the doors of the Extracorporeal Blood Module.
1 Insert the line guide into the blood pump until a signal is sounded.
The arterial pressure measurement unit is opened.
(After closing the doors the pump segment will be automatically
inserted into the blood pump.)
2 Insert the arterial blood line into the line holder.
3 Connect the arterial blood line to the lower port of the dialyzer.
4 Insert the arterial pressure dome into the arterial pressure
measurement unit.
5 Insert the arterial blood line into the arterial occlusion clamp.
6 Connect the arterial patient connection of the tubing system to the
rinse solution bag.
5-24Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
Caution
If no heparin syringe is used, retighten the cap of the heparin line and
close the clamp.
7 Connect the heparin syringe to the arterial tubing system.
To place the heparin syringe in the holder:
Press on the clamping brackets to move the grip handle to its lower
position.
Place the heparin syringe between the barrel holders.
The syringe wings must be positioned between the barrel holders
and the bracket.
The thumb rest of the syringe plunger now must be positioned
between the clamps of the grip handle.
Press on the clamping brackets to move the grip handle to its
starting position.
Caution
When inserting the heparin syringe, the following precautions must be
respected:
– Only use heparin syringes with a volume of up and equal to 30 ml.
– Ensure that the heparin syringe is correctly inserted into the heparin
pump and locked. Follow the description and the illustration.
8 Insert the venous bubble catcher into the level detector. Mind the
locator for the venous bubble catcher.
9 Insert the venous line into the optical detector/air bubble detector.
The line must be positioned completely inside the line housing.
Caution
The following must be observed for the air bubble detector:
– No ultrasound-conducting objects and agents may be used.
– The line housing must be clean and dry.
10 Insert the venous blood line into the venous occlusion clamp.
11 Insert the venous blood line into the line holder.
12 Connect the venous blood line to the upper port of the dialyzer.
13 Connect the venous pressure line to the venous pressure connector.
Caution
Tightly connect the hydrophobic filter of the venous pressure line to the
venous pressure connector in order to avoid an ingress or loss of air.
14 Connect the venous patient connection of the tubing system to the
prime collection bag. (Prime collection bag not included in scope of
delivery of the tubing system.)
Close the doors.
(The line segment(s) are automatically inserted, the arterial pressure
measurement unit closes.)
Break the cone on the rinse solution bag.
Fresenius Medical Care 5008 OP 5/09.065-25
Chapter 5: Preparation
The system automatically switches to the PREPARATION SCREEN.
Message: Connect dialyzer couplings!
5.2.6Connecting the Dialysate Lines
Note
The dialysate lines may only be connected after the T1 test has been
completely terminated.
Message: Connect dialyzer couplings!
Open the shunt door.
Connect the dialysate supply line (dialyzer coupling red) to the dialyzer
(on the side of the venous blood outlet port).
Connect the dialysate return line (dialyzer coupling blue) to the dialyzer
(on the side of the arterial blood inlet port).
Close the shunt door.
5.2.7Priming the Extracorporeal Blood Circuit with Rinse Solution Bag
OStarting the Rinse procedure
PREPARATION SCREEN
Check/set the rinse volume.
Check/set the delivery rate of the blood pump.
The rinse volume and the delivery rate are automatically set to the value
preselected in the Operator Setup. Change the rinse volume and the
delivery rate if necessary.
Message: Do not connect patient! Minimum rinse volume not reached.
– Rinse Continue
Do not connect a patient if rinsing is interrupted during the T1 test or
during a cleaning program.
5-26Fresenius Medical Care 5008 OP 5/09.06
Touch the Blood pump I/O button to continue the Rinse procedure.
(Blood pump I/O indicator green.)
ORinse procedure completed
Message: Rinse volume reached. – Rinse Continue – Circulation StartMute LED is flashing. Audible signal
Connect the venous tubing system to the rinse solution bag.
Touch the Start button to launch precirculation. (Blood pump I/O
indicator green.)
If T1 test active:
Message: CAUTION! T1 test still in progress! Patient cannot be
connected!
Chapter 5: Preparation
Touch the Blood pump I/O button to continue the circulation
procedure. (Blood pump I/O indicator green.)
5.2.8Checking/Setting the Dialysate Parameters
Note
Calcium carbonate precipitations may occur in the bicarbonate dialysis,
depending on the use and the dose of the concentrates and the duration
of the treatment. Detailed information will be provided by the
manufacturer on request.
The Prescr. Bic can be reduced to 25.0 mmol/l during preparation
(adjustable in the Technician's Setup). If the Bic reduction is selected,
25.0 mmol/l appears below the value Prescr. Bic. The Bic reduction is
de-activated when the treatment starts.
After completion of the T1 test, the EcoFlow (100 ml/min) is
automatically selected. The flow can be changed as desired.
On the PREPARATION SCREEN you can directly check,
select and change the Prescr. Na and Prescr. Bic
parameters.
In the DIALYSATE MENU
Check the dialysate parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Fresenius Medical Care 5008 OP 5/09.065-27
Chapter 5: Preparation
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.2.9Checking/Setting the UF Parameters
On the PREPARATION SCREEN you can directly check,
select and change the UF goal, UF time and UF rate
parameters.
Note
If only the UF rate is entered for ultrafiltration (instead of volume and
time), check the UF rate displayed in the Dialysis menu for plausibility
after saving the data.
In the UF MENU
Possible setting variants:
– UF goal/UF time (UF rate is calculated)
– UF goal/UF rate (UF time is calculated)
– UF rate/UF time
–UF rate
–Time
– UF profiles
– ISO goal/ISO time (ISO rate is calculated)
– ISO goal/ISO rate (ISO time is calculated)
The following must be observed in case of ISO-UF:
The ISO-UF treatment type can be started at any time and can be
repeated as often as necessary.
The parameters entered at the beginning of the treatment (UF goal and
UF time) must be taken into consideration.
In case of a combination with UF and Na profiles, first enter the ISO UF
parameters. Then set the respective profiles.
The total volume to be removed (UF goal), the total treatment time (UF
time) or the UF goal and the UF rate must always be programmed. The
ISO data goal and time cannot be higher than the UF goal/time.
The UF goal/UF time or UF goal/UF rate parameters must first be
entered.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5-28Fresenius Medical Care 5008 OP 5/09.06
5.2.10 Checking/Setting the Sodium and UF Profiles
Caution
When using Na profiles, the following precautions must be observed:
The balancing neutrality of the profiles was computed for a dialysis dose
of KT/V = 1.2. In case of higher deviations (KT/V > 1.4; KT/V < 1.0) the
balancing neutrality may not always be achieved.
Basic requirements for setting the profiles:
the UF parameters must have been set.
ONa profiles
In the DIALYSATE MENU
Check the dialysate parameters.
Minimum value that can be set for Start Na: 3 mmol higher than the
prescribed Na
Chapter 5: Preparation
OUF profiles
Set the desired profile.
Check/set Start Na (maximum value).
(The minimum Na value is automatically adjusted.)
The following must be observed:
– The treatment may also be started with the Na profile only.
– It is only possible to select matching profile groups.
(E.g. if UF profile 1 has been selected, only Na profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
Concentrate, Prescr. Na and Prescr. Bic parameters.
– The minimum UF time must be set:
Profiles 1, 2: UF time 2:00 hrs
Profile 3: UF time 3:30 hrs
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
In the UF MENU
Check the UF parameters.
Minimum UF parameters that can be set:
Profiles 1, 2: UF goal 200 ml, UF time 2:00 hrs, UF rate 10 ml
Profile 3: UF goal 200 ml, UF time 3:30 hrs, UF rate 10 ml
Set the desired profile.
Check/set the start rate.
(The minimum profile rate is automatically adjusted.)
Fresenius Medical Care 5008 OP 5/09.065-29
Chapter 5: Preparation
The following must be observed:
– The treatment may also be started with the UF profile only.
– It is only possible to select matching profile groups.
(E.g. if Na profile 1 has been selected, only UF profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
UF time and UF rate parameters.
– ISO UF parameters and Na parameters may be selected only before
starting or after ending a profile.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.2.11 Checking/Setting the Heparin Pump Parameters
Caution
Administer the heparin dose according to the physician's instructions.
In the HEPARIN menu
Check the heparin pump parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5-30Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
5.3Single-Needle (Option) Preparation Using ONLINEplus™
Irrespective of the treatment mode, all 5008 ONLINEplus™
hemodialysis systems can be operated without rinse or infusion
solutions provided in NaCl bags. The fluid volumes required for
preparation, bolus administration or during reinfusion will then be
produced ONLINE by the 5008 hemodialysis system according to the
actual requirements, thus saving both cost and time.
5.3.1Turning the Hemodialysis System On
Caution
The stability of the 5008 hemodialysis system must be ensured.
Establish the water and power supply.
Press the On/Off key. (Turn the hemodialysis system on!)
The On/Off LED is illuminated.
START-UP SCREEN
The display shows the machine type, the current software version and
the clinical data (on request) for approx. 15 seconds.
Caution
After a downtime of more than 72 hours, a cleaning program must be
performed completely before starting the treatment.
If necessary, check the hemodialysis system for presence of residual
disinfectant. (see chapter 8 Cleaning).
Note
If the message: Defective battery is acknowledged by pressing the Skip
key, it might be that the audible alarm will not be generated, if a power
failure occurs.
Fresenius Medical Care 5008 OP 5/09.065-31
Chapter 5: Preparation
5.3.2The Following Must be Observed when Using Consumables
Caution
The system has been approved for use with the consumables and
accessories listed in the Operating Instructions.
Should the responsible organization wish to use other consumables
and accessories than those listed in the Operating Instructions, the
responsibility to ensure the correct function of the system lies
exclusively with the responsible organization. The applicable legal
regulations must be complied with (e.g. in Germany the Medical Device
Directive, MDD and the MPBetreibV = German regulation for the
operation of medical products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the system resulting from the use of non-approved or unsuitable
consumables or accessories.
Caution
The symbols printed on the packaging of the consumables have to be
observed. The symbols are described in the chapter System
Description (consumables symbols).
When using consumables, it is important to take note of the following
symbols:
Do not reuse
2
Use by
Caution
The consumables may only be used if the packaging and the respective
consumable including the protective caps used are not damaged. The
protective caps must not have fallen off.
The plastics used for the consumables may not be compatible with
components of drugs or disinfectants. If they are planned to be used,
the compatibility of the consumables' components must be ensured
before the treatment. If connectors made of polycarbonate are for
example exposed to aqueous solutions with the pH value > 10 or to
aliphatic solutions this will cause tension cracks.
5-32Fresenius Medical Care 5008 OP 5/09.06
5.3.3Selecting the Concentrate Supply
OConnecting the concentrates
Caution
Concentrate:
The concentrate displayed on the screen must comply with the
specifications mentioned on the acid or the acetate container or on the
bag. This also applies to the concentrate composition in CDS operation.
Concentrate packages:
– Assure that the packages used contain sufficient concentrate to
complete the treatment.
– Use only the dedicated coded containers or the bibag
bicarbonate dialysis.
Bicarbonate dry concentrate bibag
Only the bibag
The bibag
Only use the bibag
concentrate according to the prescribed dilution. Other mixing ratios
may lead to a hazard for the patient.
Chapter 5: Preparation
®
®
®
manufactured by Fresenius Medical Care may be used.
®
must only be used for one treatment.
®
in combination with acid bicarbonate hemodialysis
:
for
Acid and basic bicarbonate hemodialysis concentrate have to be diluted
immediately prior to application only. The bag's content must be used
up within 12 hours after dilution. Discard residual volumes. The powder
is non-pyrogenic.
Conductivity limits:
The alarm limits are automatically set around the expected value.
The actual value of the conductivity display must have attained the
expected desired value after a maximum of 10 minutes.
Should this not be the case, the actual value must first be checked in
the laboratory. Change or check the concentrate, if necessary, or call
service.
Note
The bicarbonate suction tube must be inserted into the rinse chamber
during the bibag
®
treatment.
Extract the concentrate rack.
Bicarbonate dialysisTo connect the (acid) concentrate container:
Push the latch (1) upwards. Open the concentrate flap. Place the red
concentrate suction tube (2) into the acid container. Close the
concentrate flap (3) until it clicks into place.
Fresenius Medical Care 5008 OP 5/09.065-33
Chapter 5: Preparation
CDS, Central Delivery
System (option)
Acetate dialysisConnect the concentrate container.
To connect the bibag®:
Push the latch (1) upwards. Open the bicarbonate flap. Remove the
®
bibag
bibag
from its packaging. Remove the foil from the bibag®. Attach the
®
(2). Close the bicarbonate flap (3) until it clicks into place.
OR
To connect the bicarbonate container:
Insert the bicarbonate suction tube (blue) into the bicarbonate
container.
Close the bicarbonate flap.
Note
The responsible organization is responsible for the proper installation
and function of the CDS.
Insert the concentrate suction tube (red) into the acetate container.
The bicarbonate suction tube (blue) remains in the rinse chamber.
Selecting a treatmentSELECTION SCREEN
Touch the Treatment field.
If there is no tubing system inserted, the system automatically moves to
the BLOOD SYSTEM screen.
BLOOD SYSTEM SCREEN
The T1 test is now running in parallel with the preparation of the
hemodialysis system. The color of the header bar is orange for the
duration of the T1 test.
The operating mode display shows the progress of the T1 test.
Message: T1 test completed is displayed for a moment after successful
completion of the T1 test.
5-34Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
5.3.4Important Items to be Considered Before and During the Treatment
Caution
Aseptic technique:
Use aseptic technique for all bloodside connections and all connections
in the area where sterile solutions are to be used.
Caution
Preventing contamination:
Use tubing systems with hydrophobic filters at the pressure lines to
prevent cross-contamination.
Connect the hydrophobic filters so that an ingress or loss of air is not
possible and that any wetting by fluid is reliably avoided, also in case of
pressure fluctuations.
If a hydrophobic filter has become wet, the tubing system must be
replaced.
On tubing systems with additional connection sites, a replacement
pressure line may be connected (accessory available from Fresenius
Medical Care).
The blood in the pressure line must not be forced back by means of a
syringe. This could damage the hydrophobic membrane and thus lead
to a contamination.
If fluid may have passed the hydrophobic filter, the system must be
checked for contamination after completion of the treatment. If the
system is contaminated, it has to be taken out of service. All affected
components have to be disinfected or replaced in accordance with the
manufacturer's specifications before the system is put into operation
again.
Fresenius Medical Care 5008 OP 5/09.065-35
Chapter 5: Preparation
Caution
When inserting the tubing systems, the following precautions
must be respected:
– The tubing systems have to be free of kinks, tension and twists and
must not be jammed (risk of hemolysis). Use the line holders
provided.
– Ensure the correct position of the screwed connections, especially
of the connection sites to the patient, the dialyzer and the system
and check or correct them during the treatment if necessary. Take
the appropriate measures if required (e.g. retightening of the Luer
Lock connection or replacement of the tubing system).
– Check the protective caps for tight fit and tighten them if necessary.
– The lines for the supply of infusions should always be clamped,
except if they are needed.
– During long-term operation, the blood lines must be changed after
24 hours at the latest.
– Do not use cannulas with a diameter of > 20 gauge to pierce the
septum of the injection sites. Insert the cannula vertically and in the
center of the septum. Disinfect the injection sites with 70% alcohol
before use.
– The blood pump must be set to the diameter of the pump segment,
refer also to the product label of the blood lines. If a wrong line
diameter is set, this may cause significant deviations in the blood
flow and thus in the dialysis dose.
– Materials which come directly or indirectly into contact with blood
polycarbonate, latex-free rubber, ABS.
– The minimum temperature of the tubing systems during use is 18 °C.
Caution
Delivery operation of the pump(s) with open doors
(blood pump, substituate pump, optional Single-Needle pump):
When the doors are open and the rotor of the pump(s) is running, make
sure that no objects, such as fingers, hair or ball point pens, come into
contact with the rotor (risk of injury).
Arterial pressure measurement unit:
Prevent foreign objects from coming into contact with the arterial
pressure measurement unit.
Heparin pump:
If heparin syringes of third party suppliers are used, the operator is
responsible and has to ensure that the syringe data displayed match the
actual data.
Heparin syringes without Luer lock are not recommended as the
connection between the heparin syringe and the blood lines may come
loose. If heparin syringes without Luer lock are used it is the operator’s
responsibility to ensure that the connection between the heparin syringe
and the blood lines does not loosen inadvertently.
IV pole:
Securely fix bags or other objects to be hung from the IV pole.
5-36Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
Caution
Before the treatment, check:
– The safe connection of all connection sites of the tubing system.
– The tightness of the tubing system during and after priming.
– Retighten the connections and replace the tubing system, if
necessary.
– The absence of air in the tubing system and the correct position of
all fluid levels.
To be observed when working on the tubing system during the
treatment:
If the position of the tubing system or of one of its components is
changed, the correct position of the entire tubing system must be
restored afterwards, above all the correct position of the line guides.
During the treatment check at appropriate intervals:
– The condition of the patient.
– The function of the hemodialysis system and the extracorporeal
blood circuit. Pay particular attention to the venous insertion site, as
a possible dislocation of the venous cannula may not always be
detected by the pressure monitoring system.
– The tubing system for leakages or possible loosening of connections
as well as entry of air.
– The fluid level in the venous bubble catcher. Correct it, if required
(desired level: approx. 1 cm below the upper edge of the cover)
Caution
Venous alarm limit:
The lower venous alarm limit must be set as close as possible to the
actual venous pressure value.
Note
The dialyzer holder is not suitable for rectangular plate dialyzers.
Note
For hygienic reasons, the blood lines should be inserted immediately
prior to the treatment only.
If the blood lines were inserted more than 8 hours before the treatment,
malfunctions may occur. Correcting these malfunctions may require
removing the present blood lines and inserting new blood lines.
Fresenius Medical Care 5008 OP 5/09.065-37
Chapter 5: Preparation
5.3.5Preparing the Extracorporeal Blood Circuit with ONLINEplus™
OInserting the arterial and the venous tubing system
When inserting the tubing system, follow the description of the menu
displayed.
Wings of the
syringe cylinder
Thumb rest of the
syringe plunger
18
19
4
2
3
8
7
1
6
15
14
16
13
12
17
5
9
10
11
Open the doors of the Extracorporeal Blood Module.
1 Insert the line guide into the blood pump until a signal is sounded.
The arterial pressure measurement unit is opened.
(After closing the doors the pump segment will be automatically
inserted into the blood pump.)
2 Connect the SN pressure line to the Single-Needle pressure port.
3 Insert the SN chamber in its holder.
4 Insert the SN line guide into the Single-Needle pump until a signal is
sounded.
(After closing the doors the SN pump segment is automatically
inserted in the Single-Needle pump, if the level detector detected
fluid.)
5 Connect the arterial blood line to the lower port of the dialyzer.
6 Insert the arterial pressure dome into the arterial pressure
measurement unit.
7 Insert the arterial blood line into the arterial occlusion clamp.
5-38Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
Caution
If no heparin syringe is used, retighten the cap of the heparin line and
close the clamp.
8 Connect the heparin syringe to the arterial tubing system.
To place the heparin syringe in the holder:
Press on the clamping brackets to move the grip handle to its lower
position.
Place the heparin syringe between the barrel holders.
The syringe wings must be positioned between the barrel holders
and the bracket.
The thumb rest of the syringe plunger now must be positioned
between the clamps of the grip handle.
Press on the clamping brackets to move the grip handle to its
starting position.
Caution
When inserting the heparin syringe, the following precautions must be
respected:
– Only use heparin syringes with a volume of up and equal to 30 ml.
– Ensure that the heparin syringe is correctly inserted into the heparin
pump and locked. Follow the description and the illustration.
9 Insert the venous bubble catcher into the level detector. Mind the
locator for the venous bubble catcher.
10 Insert the venous line into the optical detector/air bubble detector.
The line must be positioned completely inside the line housing.
Caution
The following must be observed for the air bubble detector:
– No ultrasound-conducting objects and agents may be used.
– The line housing must be clean and dry.
11 Insert the venous blood line into the venous occlusion clamp.
12 Insert the venous blood line into the line holder.
13 Connect the venous blood line to the upper port of the dialyzer.
14 Connect the venous pressure line to the venous pressure connector.
Caution
Tightly connect the hydrophobic filter of the venous pressure line to the
venous pressure connector in order to avoid an ingress or loss of air.
Inserting the lines may be continued only after successful completion of
the T1 test.
15 Insert the SafeLine™ line guide into the substituate pump until a
signal is sounded.
(After closing the doors the SafeLine™ pump segment will be
automatically inserted into the substituate pump.)
16 Insert the SafeLine™ into the line holder.
Fresenius Medical Care 5008 OP 5/09.065-39
Chapter 5: Preparation
17 Connect the arterial patient access line to the SafeLine™.
b
a
c
18 To connect the substituate connector to the substituate port:
The catch for the substituate port (blue) is in initial position (a).
Pull and turn the substituate catch (blue) counterclockwise into
position (b).
Push the substituate connector firmly into the substituate port.
Pull and turn the substituate catch (blue) clockwise until it clicks into
place at position (c). If necessary to make sure of a tight closure
push in the substituate catch (blue).
a
b
c
19 To insert the rinse connector (connected onto the venous patient
line) into the rinse port:
The catch for the rinse port (grey) is in initial position (a).
Pull and turn the rinse port catch (grey) counterclockwise into
position (b).
Push the rinse connector firmly into the rinse port.
Pull and turn the rinse port catch (grey) clockwise until it clicks into
place at position (c). If necessary to make sure of a tight closure
push in the rinse port catch (grey).
Close the doors.
(The line segment(s) are automatically inserted, the arterial pressure
measurement unit closes.)
The system automatically switches to the PREPARATION SCREEN.
Message: Connect dialyzer couplings!
5-40Fresenius Medical Care 5008 OP 5/09.06
5.3.6Connecting the Dialysate Lines
Note
The dialysate lines may only be connected after the T1 test has been
completely terminated.
Message: Connect dialyzer couplings!
Open the shunt door.
Connect the dialysate supply line (dialyzer coupling red) to the dialyzer
(on the side of the venous blood outlet port).
Connect the dialysate return line (dialyzer coupling blue) to the dialyzer
(on the side of the arterial blood inlet port).
Close the shunt door.
5.3.7Priming the Extracorporeal Blood Circuit with ONLINEplus™
Chapter 5: Preparation
OStarting the Rinse procedure
PREPARATION SCREEN
Check/set the rinse volume.
Check/set the delivery rate of the blood pump.
The rinse volume and the delivery rate are automatically set to the value
preselected in the Operator Setup. Change the rinse volume and the
delivery rate if necessary.
Message: Do not connect patient! Minimum rinse volume not reached.
– Rinse Continue
If not yet done, the parameters for dialysate, UF and the
heparin pump now may be checked/set.
Touch the Continue button to continue rinsing. (Blood pump I/O
indicator green.)
Fresenius Medical Care 5008 OP 5/09.065-41
Chapter 5: Preparation
ORinse procedure completed
Endless rinse will start when the Online rinse volume and the Online UF
rinse volume have been reached.
The delivery rate of the blood pump is automatically reduced to
50 ml/min, and the dialysate flow is reduced to EcoFlow (100 ml/min).
In case of ONLINEplus™, the total flow is composed of EcoFlow
(100 ml/min) and the respective substituate rate.
5.3.8Checking/Setting the Dialysate Parameters
Note
Calcium carbonate precipitations may occur in the bicarbonate dialysis,
depending on the use and the dose of the concentrates and the duration
of the treatment. Detailed information will be provided by the
manufacturer on request.
The Prescr. Bic can be reduced to 25.0 mmol/l during preparation
(adjustable in the Technician's Setup). If the Bic reduction is selected,
25.0 mmol/l appears below the value Prescr. Bic. The Bic reduction is
de-activated when the treatment starts.
After completion of the T1 test, the EcoFlow (100 ml/min) is
automatically selected. The flow can be changed as desired.
On the PREPARATION SCREEN you can directly check,
select and change the Prescr. Na and Prescr. Bic
parameters.
In the DIALYSATE MENU
Check the dialysate parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.3.9Checking/Setting the UF Parameters
On the PREPARATION SCREEN you can directly check,
select and change the UF goal, UF time and UF rate
parameters.
5-42Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
Note
If only the UF rate is entered for ultrafiltration (instead of volume and
time), check the UF rate displayed in the Dialysis menu for plausibility
after saving the data.
In the UF MENU
Possible setting variants:
– UF goal/UF time (UF rate is calculated)
– UF goal/UF rate (UF time is calculated)
– UF rate/UF time
–UF rate
–Time
– UF profiles
– ISO goal/ISO time (ISO rate is calculated)
– ISO goal/ISO rate (ISO time is calculated)
The following must be observed in case of ISO-UF:
The ISO-UF treatment type can be started at any time and can be
repeated as often as necessary.
The parameters entered at the beginning of the treatment (UF goal and
UF time) must be taken into consideration.
In case of a combination with UF and Na profiles, first enter the ISO UF
parameters. Then set the respective profiles.
The total volume to be removed (UF goal), the total treatment time (UF
time) or the UF goal and the UF rate must always be programmed. The
ISO data goal and time cannot be higher than the UF goal/time.
The UF goal/UF time or UF goal/UF rate parameters must first be
entered.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.3.10 Checking/Setting the Sodium and UF Profiles
Caution
When using Na profiles, the following precautions must be observed:
The balancing neutrality of the profiles was computed for a dialysis dose
of KT/V = 1.2. In case of higher deviations (KT/V > 1.4; KT/V < 1.0) the
balancing neutrality may not always be achieved.
Basic requirements for setting the profiles:
the UF parameters must have been set.
Fresenius Medical Care 5008 OP 5/09.065-43
Chapter 5: Preparation
ONa profiles
In the DIALYSATE MENU
Check the dialysate parameters.
Minimum value that can be set for Start Na: 3 mmol higher than the
prescribed Na
Set the desired profile.
Check/set Start Na (maximum value).
(The minimum Na value is automatically adjusted.)
The following must be observed:
– The treatment may also be started with the Na profile only.
– It is only possible to select matching profile groups.
(E.g. if UF profile 1 has been selected, only Na profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
Concentrate, Prescr. Na and Prescr. Bic parameters.
– The minimum UF time must be set:
Profiles 1, 2: UF time 2:00 hrs
Profile 3: UF time 3:30 hrs
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
OUF profiles
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
In the UF MENU
Check the UF parameters.
Minimum UF parameters that can be set:
Profiles 1, 2: UF goal 200 ml, UF time 2:00 hrs, UF rate 10 ml
Profile 3: UF goal 200 ml, UF time 3:30 hrs, UF rate 10 ml
Set the desired profile.
Check/set the start rate.
(The minimum profile rate is automatically adjusted.)
The following must be observed:
– The treatment may also be started with the UF profile only.
– It is only possible to select matching profile groups.
(E.g. if Na profile 1 has been selected, only UF profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
UF time and UF rate parameters.
– ISO UF parameters and Na parameters may be selected only before
starting or after ending a profile.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5-44Fresenius Medical Care 5008 OP 5/09.06
5.3.11 Checking/Setting the Heparin Pump Parameters
Caution
Administer the heparin dose according to the physician's instructions.
In the HEPARIN menu
Check the heparin pump parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
5.3.12 Checking/Setting the Single-Needle Parameters
In the SINGLE-NEEDLE menu
Chapter 5: Preparation
Check the Single-Needle parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
Fresenius Medical Care 5008 OP 5/09.065-45
Chapter 5: Preparation
5-46Fresenius Medical Care 5008 OP 5/09.06
Chapter 5: Preparation
5.4Single-Needle (Option) Preparation with Rinse Solution Bag
If the 5008 ONLINEplus™ hemodialysis system cannot be operated
without rinse solution bags, rinse or infusion solutions provided in NaCl
bags may be used instead.
5.4.1Turning the Hemodialysis System On
Caution
The stability of the 5008 hemodialysis system must be ensured.
Establish the water and power supply.
Press the On/Off key. (Turn the hemodialysis system on!)
The On/Off LED is illuminated.
START-UP SCREEN
The display shows the machine type, the current software version and
the clinical data (on request) for approx. 15 seconds.
Caution
After a downtime of more than 72 hours, a cleaning program must be
performed completely before starting the treatment.
If necessary, check the hemodialysis system for presence of residual
disinfectant. (see chapter 8 Cleaning).
Note
If the message: Defective battery is acknowledged by pressing the Skip
key, it might be that the audible alarm will not be generated, if a power
failure occurs.
Fresenius Medical Care 5008 OP 5/09.065-47
Chapter 5: Preparation
5.4.2The Following Must be Observed when Using Consumables
Caution
The system has been approved for use with the consumables and
accessories listed in the Operating Instructions.
Should the responsible organization wish to use other consumables
and accessories than those listed in the Operating Instructions, the
responsibility to ensure the correct function of the system lies
exclusively with the responsible organization. The applicable legal
regulations must be complied with (e.g. in Germany the Medical Device
Directive, MDD and the MPBetreibV = German regulation for the
operation of medical products).
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the system resulting from the use of non-approved or unsuitable
consumables or accessories.
Caution
The symbols printed on the packaging of the consumables have to be
observed. The symbols are described in the chapter System
Description (consumables symbols).
When using consumables, it is important to take note of the following
symbols:
Do not reuse
2
Use by
Caution
The consumables may only be used if the packaging and the respective
consumable including the protective caps used are not damaged. The
protective caps must not have fallen off.
The plastics used for the consumables may not be compatible with
components of drugs or disinfectants. If they are planned to be used,
the compatibility of the consumables' components must be ensured
before the treatment. If connectors made of polycarbonate are for
example exposed to aqueous solutions with the pH value > 10 or to
aliphatic solutions this will cause tension cracks.
5-48Fresenius Medical Care 5008 OP 5/09.06
5.4.3Selecting the Concentrate Supply
OConnecting the concentrates
Caution
Concentrate:
The concentrate displayed on the screen must comply with the
specifications mentioned on the acid or the acetate container or on the
bag. This also applies to the concentrate composition in CDS operation.
Concentrate packages:
– Assure that the packages used contain sufficient concentrate to
complete the treatment.
– Use only the dedicated coded containers or the bibag
bicarbonate dialysis.
Bicarbonate dry concentrate bibag
Only the bibag
The bibag
Only use the bibag
concentrate according to the prescribed dilution. Other mixing ratios
may lead to a hazard for the patient.
Chapter 5: Preparation
®
®
®
manufactured by Fresenius Medical Care may be used.
®
must only be used for one treatment.
®
in combination with acid bicarbonate hemodialysis
:
for
Acid and basic bicarbonate hemodialysis concentrate have to be diluted
immediately prior to application only. The bag's content must be used
up within 12 hours after dilution. Discard residual volumes. The powder
is non-pyrogenic.
Conductivity limits:
The alarm limits are automatically set around the expected value.
The actual value of the conductivity display must have attained the
expected desired value after a maximum of 10 minutes.
Should this not be the case, the actual value must first be checked in
the laboratory. Change or check the concentrate, if necessary, or call
service.
Note
The bicarbonate suction tube must be inserted into the rinse chamber
during the bibag
®
treatment.
Extract the concentrate rack.
Bicarbonate dialysisTo connect the (acid) concentrate container:
Push the latch (1) upwards. Open the concentrate flap. Place the red
concentrate suction tube (2) into the acid container. Close the
concentrate flap (3) until it clicks into place.
Fresenius Medical Care 5008 OP 5/09.065-49
Chapter 5: Preparation
CDS, Central Delivery
System (option)
Acetate dialysisConnect the concentrate container.
To connect the bibag®:
Push the latch (1) upwards. Open the bicarbonate flap. Remove the
®
bibag
bibag
from its packaging. Remove the foil from the bibag®. Attach the
®
(2). Close the bicarbonate flap (3) until it clicks into place.
OR
To connect the bicarbonate container:
Insert the bicarbonate suction tube (blue) into the bicarbonate
container.
Close the bicarbonate flap.
Note
The responsible organization is responsible for the proper installation
and function of the CDS.
Insert the concentrate suction tube (red) into the acetate container.
The bicarbonate suction tube (blue) remains in the rinse chamber.
Selecting a treatmentSELECTION SCREEN
Touch the Treatment field.
If there is no tubing system inserted, the system automatically moves to
the BLOOD SYSTEM screen.
BLOOD SYSTEM SCREEN
The T1 test is now running in parallel with the preparation of the
hemodialysis system. The color of the header bar is orange for the
duration of the T1 test.
The operating mode display shows the progress of the T1 test.
Message: T1 test completed is displayed for a moment after successful
completion of the T1 test.
5-50Fresenius Medical Care 5008 OP 5/09.06
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.