Drager Medical Air Compressor Operation manual

0 (0)

Drager

“MEDICAL AIR COMPRESSOR

OPERATION AND SERVICE MANUAL

SPECIFICATIONS:

Minimum Continuous Output: 60 Ipm © 37 psi

Outlet

Pressure:

37-49

psi

 

 

 

Dew

Point Depression:

C

below

room ambient

Power

Requirement:

115V,

60Hz, 10

Amps

Current

Leakage:

 

 

normal

polarity open ground

 

 

 

 

 

 

less

than 85

microamps

Circuit

Breaker:

15

Amps

 

 

 

Weight:

126

Ibs.

 

 

 

 

 

 

Noise

Level:

52

dBa

+

2 dBa

 

 

Compressor

Safety

Valve:

85

psig t

5 psig

*PRECAUTIONS

Before

operating

or

servicing

the

air

compressor,

read

and

understand

all

the accompanying

manuals.

Be

familiar

with

all

the precautions, notes

and

warnings. Drager is

not

responsible

for

damages

caused

by

misuse.

Do

not

use

the

compressor

if repair is needed. Replace broken

or

worn parts

immediately.

Follow

the

instructions

in this

service

manual to

perform scheduled

maintenance

and

repairs.

 

 

 

 

 

 

 

WARNING:

To avoid shock hazard, refer all servicing to qualified

 

 

 

-

personnel.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WARNING:

Do

not use the compressor in the

presence

of flammable

 

 

-

anesthetics.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WARNING:

The

compressor

is

grounded.

DO

NOT

plug

it into an

ungrounded

 

 

outlet

or

an adaptor.

 

 

 

 

 

 

 

 

 

 

 

 

WARNING:

Never

put oxygen

 

into the

compressor.

The

electrical

components

 

-

are

not

approved

for oxygen

use.

 

 

 

 

 

 

 

 

 

CAUTION:

It is

recommended

that

you

install

a

transparent

water

trap

at

 

 

the

inlet

of all blenders,

respirators,

etc.

to

 

protect them from

 

 

moisture in case the drying system

 

of the

compressed

air

source

 

 

(portable

or piped)

fails.

The

water

 

trap

should

be checked

and

 

|

emptied

daily.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAUTION:

The

compressor

is

of the oil-less type and

MUST

NOT

be

 

 

 

lubricated.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0-18

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ОКАСЕК

ОЛ

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

JUNE

199!

| TABLE OF CONTENTS

ROUTINE MAINTENANCE

WEEKLY QUARTERLY YEARLY

BETWEEN 8000-9000 HOURS TROUBLE SHOOTING

A. PRESSURE GAUGE NEEDLE STICKS OR VIBRATES

B.OVERHEATING

C.NOISE/EXCESSIVE VIBRATION

D.WATER PROBLEMS

E.HIGH PRESSURE

F.LOW PRESSURE

 

If

the

compressor

is

not running

-

If

the

compressor

is

running

SERVICE

A. TO CHECK FOR A FAULTY SOLENOID

B. ADJUSTING THE PRESSURE SWITCH

EXTERNAL AIR RESERVOIR KIT

A.AIR OUTLET MODIFICATIONS

B. INSTALLING OPTIONAL EXTERNAL AIR RESERVOIR KIT

REPLACEMENT PARTS LIST

WIRING SCHEMATIC

PNEUMATIC SCHEMATIC

ROUTINE MAINTENANCE

O © © © © © 00 “Y Y Y“ Y“ OO Mn NN

Note: After performing any maintenance procedure, verify the performance of the unit to the specifications on page 1, use an RT-200 or equivalent measuring device.

WEEKLY

Remove the

foam inlet filter

on

the front

of the compressor. Wash it in warm soapy water, rinse well, allow to dry

and reinstall.

Replacement

part

#13989.

 

QUARTERLY

Wash/replace the air pump intake filter elements:

 

“FILTER,

COMPLETE INTAKE

1.

_Unthread the 8 screws from the rear door and remove

968-402

 

door.

 

2.Locate the muffler in the left side of the compartment.

Unscrew intake filter assy. from the muffler.

3.Snap the cap off of the intake filter and pull the filter

 

elements out. Wash them in warm soapy

water.

Rinse

 

 

 

well, allow to dry and

reinstall.

While

the

cap is

off, wipe it

 

 

 

and the inside

of the

filter

with

a

clean

dry

cloth.

 

FILTER, FELT

4.

Reassemble by

reversing

steps

1

through

3.

 

 

 

901-425

 

 

 

 

 

 

 

 

 

 

— FILTER,

FOAM

 

 

 

 

 

 

 

 

 

FIGURE i

901-426

 

 

 

 

 

 

 

 

 

 

 

YEARLY

A

Preventive Maintenance Kit should

be installed, part #14003.

It

includes:

|.

Inlet Filters

 

(6)

#13989

..

Water

Jar

Filter

#924-401

3.

Water Jar O-ring

#13530

4.

Regulator

Diaphragm

#923-402

5.

Relief

Valve

Diaphragm

#923-401

6.

Intake

Filter

Element

#901-425

7.

Intake

Filter

Element

#901-426

A. To install the intake filter element see 'QUARTERLY' page 2.

B.To install the water trap jar o-ring and water trap filter:

‘Remove the inlet filter from the front of the unit.

=No

FIGURE 2

2.

‘Loosen

the 2

screws

that

hold

the

inlet

grille

in place and remove

 

grille.

 

 

 

 

 

 

 

 

3

Pull the

tubing

off of

the

bottom

of

the

water

jar.

À,

4. 'Unthread the bowl and remove.

5 ‘Unthread the retainer cap and remove the filter.

6 Install the new filter and reinstall the retainer cap.

7Carefully position the new O-ring on top of the water jar and rethread. !t is important to make this seal tight.

L = mues que)

© “To install the relief valve and regulator diaphragms:

1

‘Unthread

the

8

screws

from the

rear

door and

remove

the

door.

2.

Relief valve

has

a white

locking

ring,

regulator

has a

red

locking

|

ring.

 

 

 

 

 

 

 

 

 

3.

Unthread and remove the bonnet portion of the relief valve.

4.

Remove the worm diaphragm, install the replacement and

 

.

reassemble

relief valve.

 

 

 

 

 

 

5.

Unthread

the

bonnet from the regulator.

 

 

 

FIGURE 3

REGULATOR O A Y DIAPHRAGM

-

RELIEF

 

VALVE

/

DIAPHRAGM

 

FIGURE 4

Drager Medical Air Compressor Operation manual
Loading...
+ 7 hidden pages