Procedure for Periodic Manufacturer's Service

1.A PMS Checklist Babytherm 8004/8010 form, available from the Dräger Medical, Inc. Technical Service Department, shall be completed by the Technical Service

Representative each time a PMS is performed.

2.Caution: Service, Inspection and Calibration are to be performed by Dräger trained and authorized Service Personnel only!

3.Note: Test Equipment for this procedure is listed in the document "Test equipment for CCS devices". DO NOT USE any test equipment having an expired calibration date.

Test equipment for CCS devices

4.Warning: During Service and Inspection some systems may be disconnected. Ensure these systems are reconnected, test equipment is removed and all covers are replaced after testing!

5.Perform Periodic Manufacturer's Service according to the document "Test Certificate Babytherm 8004/8010" except for test items

1.4.18 through 1.4.18.1 1.4.25 through 1.4.36,

1.4.39 through 1.4.41,

2.2 through 2.2.1,

3 through 3.3 "Safety Checks" (Electrical safety test according to VDE 0751), and 4.7:

Test Certificate Babytherm 8004/8010

6.Perform electrical safety test according to the following Instructions "Babytherm

8004/8010, Electrical Safety Test in the USA and Canada":

With Biotek Model 501

With DNI Nevada Model µ-Test 2000

If necessary record the measured values additional to the PMS Checklist in the following form:

Electrical Safety Test Form Babytherm

7.Completely fill in the header information of the document PMS Checklist

Babytherm 8004/8010.

6132.300 Babytherm 8004/8010

PMS

July 2003

1/2