Xylem Oxix Dissolved User Manual

Dissolved Oxygen Transmitter
Dissolved Oxygen Transmitter
Installation and User Manual
1GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
2GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
CE CERTIFICATE OF CONFORMITY
This product complies with the requirements concerning electromagnetic compatibility (EMC) stipulated in Council directive no. 89/336/EEC of 3rd may 1989, alteret at directive no. 92/31/ EEC, on the approximation of the laws of the member states relating to electromagnetic compat­ibility.
CE approvals
We declare that the Oxix
®
transmitter complies with the values stipulated in EN-61010-1 and EN
61326-1.
Ex approval
The Oxix
®
transmitter is approved for mounting in explosive amospheres: Ex II G3.
3
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

Table of Contents

1. Introduction 7
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Operating Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2. Safety, Repair and Product Identification 13
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Physical Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Repair ................................................................13
Product Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3. Specifications and Order Numbers 15
Specifications ..........................................................15
Order Numbers .........................................................17
4. Sensor Installation 19
Physical Sensor Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Preparation ...........................................................20
Positioning the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
5. Sensor Calibration 21
Zero-point Calibration Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
2-point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
6. Converter 23
Electrical Mounting .....................................................23
Power Supply .........................................................24
Display Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Changing the Power Supply Voltage 230/115 V AC . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Analogue Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Digital Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Converter Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Sensor and Mounting Kit / Connection Board Terminals .........................30
7. System Configuration Examples 31
Magflux® with Blind Lid and Oxix® with Display Unit ...........................32
Wiring Schematic #1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
®
SuSix
with Blind Lid, Magflux® with Blind Lid and a Remote Display .............34
Wiring Schematic #2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
4GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
8. Startup 37
Initial Checks before Power-up ............................................37
Initial Power-up ........................................................37
Initial Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Language Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Display Read-out, one connected unit ......................................38
Display Read-out, several connected units ...................................39
Display Keys ..........................................................39
9. Oxix® Menus 41
Converter Overview Menu ................................................41
Specify Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Datalogger ...........................................................44
Graph Display .........................................................46
Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Sensor Name .........................................................49
Converter Setup ........................................................51
Averaging ............................................................52
Units ...............................................................52
mA Output ...........................................................53
Cleaning .............................................................54
High alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Low alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Sensor error ..........................................................58
Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
Service Menu - Digest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Service Menu - Detailed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Display Setup ..........................................................72
Language ............................................................73
Set clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Modbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Display SW version .....................................................77
10. Mechanical Dimensions 79
Sensor ..............................................................79
Converter and Display Unit ...............................................80
Immersion Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
11. Maintenance 83
Cleaning the sensor .....................................................83
Cleaning the converter display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Functional Tests ........................................................83
Service Contract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
5
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Appendix A. Pop-up and Error Messages 85
Appendix B. MJK-Field Link Software 89
Important Notes! ......................................................89
A. Connect a PC to the Oxix
B. Save Log Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
C. Upgrade Display Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
D. Upgrade Converter Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
E. Install/Add Languages ................................................94
®
............................................89
Appendix C. Front Panel Cut-out Drawing 95
Appendix D. Log Files 97
Appendix E. Input/Output Connections 101
Main Menu Overview 102
Converter Setup Menu Overview 106
Service Menu Overview 114
Display Setup Menu Overview 120
Index Index-1
6GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
7
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

1. Introduction

Thank you for choosing the MJK Oxix® Dissolved Oxygen Transmitter. We have done our utmost to design and manufacture a quality transmitter that satisfies your requirements:
High performance optical sensor•
No calibration•
No spareparts needed in sensor's 10 year lifetime•
Air or water jet cleaning system is standard•
Measurements not affected by sunlight or dry environment•
Large, easy-to-read display with trend curve capability•
Multiple languages•
Built-in datalogger•
No moving parts, oxygen consumption or risk of poisoning•
Mounting accessories • (optional)
The Oxix ever, read this manual first to get familiar with the Oxix® transmitter and all its features.
The equipment must be handled and operated as instructed by the manufac­turer, MJK Automation A/S to ensure stable operation and accurate measure­ments.
You can always contact your local representative or the MJK hot lines for advice and guidance:
®
transmitter is easy to install, calibrate and put into service. How-
Europe• Tel.: +45 45 56 06 56 E-mail: mjk@mjk.com
Denmark• Tel.: +45 45 56 06 56 E-mail: mjk@mjk.dk
Norway• Tel.: +47 69 20 60 70 E-mail: mjk@mjk.no
Sweden• Tel.: +46 53 31 77 50 E-mail: kontoret@mjk.se
Holland• Tel.: +31 251 672171 E-mail: mjknl@mjk.com
USA• Tel.: +1 847 482 8655 E-mail: mjkusa@mjk.com
Australia• Tel.: +61 3 9758 8533 E-mail: ns@mjk.com
Visit our web sites at www.oxix.dk and www.mjk.com to learn more about MJK Automation, our other products and the people behind them.
®
Oxix
is a registered trademark of MJK Automation A/S, Denmark.
8GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter

About this Manual

This manual is divided into one table of contents, eleven chapters, five appen­dices, four continuous menu overview flow charts and one index.
1. Introduction
Contains a presentation of the Oxix
structure of this manual, the operating principles and the top features.
2. Safety, Repair and Product Identification
Provide answers to issues regarding safety, mounting, repair,
restrictions and product identification.
3. Specifications and Order Numbers
The system specs and order numbers are listed for quick access.
4. Transmitter Installation
Contains installation guidelines for positioning of the transmitter in liquids,
distances to walls and other solid faces, and installation in tubes.
5. Sensor Calibration
The Oxix
®
sensor is factory calibrated and requires no re-calibration.
6. Converter
Describes the physical specifications and installation rules for the convert-
er such as power supply, in- and outputs, sensor/converter/ display configurations, etc.
7. System Configuration Examples
Illustrate how Oxix
®
Dissolved Oxygen Transmitter can be interconnected
with for example MJK MagFlux® flow meters in real-life scenarios.
8. Startup
Describes important checks and basic settings to get started including the
display and keyboard user interface.
9. Oxix
®
Menus
Contains a comprehensive description of the Oxix
utilities.
10. Mechanical Dimensions
Lists the Oxix
®
Dissolved Oxygen Transmitter dimensions.
11. Maintenance
Contains procedures for cleaning the sensor and the display.
®
Dissolved Oxygen Transmitter, the
®
menus, options and
9
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Appendix A. Pop-up Events / Error Messages
Lists possible pop-up messages, explains their meaning and offer
solutions to error conditions.
Appendix B. MJK-Field Link Software Upgrade
Describes in detail how to utilise the unique and intuitive MJK-Field Link
software program to upload new display and converter firmware versions.
Appendix C. Front Panel Cut-out Drawing
A 1:1 scale drawing of the front panel outline and cut-out area for installa-
tion and mounting purposes.
Appendix D. Log Files
An example log file illustrates the format and explains the entries.
Appendix E. Digital Input/Output Connections
Illustrates how the digital in- and outputs can be interconnected.
* * * * *
Main Menu Overview
A continuous presentation of the top level menu structure.
Converter Setup Menu Overview
A continuous presentation of the converter setup menu structure.
Service Menu Overview
A continuous presentation of the service menu structure.
Display Setup Menu Overview
A continuous presentation of the display configuration menu structure.
10GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter

Operating Principles

The MJK Oxix® Dissolved Oxygen Transmitter is designed for measuring the oxygen contents in open tanks exposed to sunlight, wells and closed contain­ers.
The Oxix® sensor function is based on the fluorescence principle and does not consume oxygen like membrane-type sensors. It is rugged and designed to handle tough applications.
Light pulses with a wavelength of 475 nm are transmitted to a ruthenium compound immobilized in a sol-gel matrix. The ruthenium absorbs the light energy by changing the outer electron’s ener­gy level. The unstable electron then collapses and returns back to its original energy state and at the same time emits the surplus energy as a photon with a wavelength of 600 nm. This process is also refered to as fluorescing.
When the intensity of the transmitted light pulses is tightly controlled, the amount of fluorescing is predictable and repeatable. If free oxygen molecules are present, the photon energy is absorbed by the oxygen molecules without any light emission, and the amount of fluorescing is consequently reduced (also refered to as fluorescence quenching). It is therefore possible to determine the amount of dissolved oxygen by meas­uring the amount of fluorescence quenching (light intensity).
11
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
ll
RS­485
RS­485
RS­485
Max. 1000 m
1 pc. Oxix Converter with display
(
206505
)
3 pcs. Oxix without display (206508)
4 pcs. Oxix sensors (206510)
ll ll
300 m
Max.

Features

Flexible Installation
MJK’s modular design allows up to 300 meters (950ft.) between the sensor and the converter, and the Display Unit can be mounted up to 1000 m (3000 ft.) from the signal converter with ordinary twisted wires.
One Display Unit can control up to four MagFlux® flow meters, SuSix® trans­mitters and/or Oxix® transmitters for greater economy, space savings and an improved overview of the multiple measurement values.
Simple to Operate
The Oxix® Display Unit has a mobile phone-like menu structure and can dis­play text in several selectable languages.
PC Connection
®
Oxix
allows downloading setup configurations, uploading new instrument software updates and display customization. The onboard data logger cap­tures 20.000 readings which are displayed as an electronic graph and can be retrieved as a CSV file on a PC. This connectivity is achieved with a common USB port and the free MJK Field Link software.
12GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Flexible In- and Outputs
The Oxix® Transmitter has one 4-20 mA analogue output, two digital outputs for alarms or control, and one digital input for resetting alarms, batches, etc.
User-definable Text
Up to five lines of text and readings can be configured by the user. The graphic display is automatically adjusted to show the largest characters pos­sible. Alarms can be displayed as pop-up alarms until they are reset.
Modbus® Communication
The Display Unit uses the Modbus® RTU communication protocol to con­nect to the Oxix® Transmitter. The Transmitter can in turn connect to SCADA systems with its Modbus® communication protocol.
13
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
2. Safety, Repair and Product Identification

Safety Instructions

Read this manual carefully.•
Pay attention to the environment on the installation site.•
Wear necessary protective equipment and follow all current safety regula-• tions.
®
Oxix•
can invoke a start signal for dangerous machinery. Always ensure that connected machinery and other equipment are effectively put out of service (that is to remove the main fuses and lock main and security switches in off-position) before commencing configuration, fault finding, servicing, maintenance work, etc.
• WARNING: There is a risk of lethal, electrical shock from the "Mains sup-
ply" terminals "N" and "L". Be careful not to touch these terminals when the Oxix
WARNING:• Incorrect installation or use can lead to body injury and/or material damage!

Physical Mounting

The Oxix® converter/display unit must not be mounted in explosion hazard­ous areas!
Oxix® may only be installed and commissioned by authorized MJK personnel, and all national and local regulations must be complied with.
®
front cover is removed.

Repair

Repair must only be made by authorized MJK personnel or by a service rep­resentative approved by MJK.
Product Identification
A delivery will usually consist of an Oxix® sensor and an Oxix® converter with or without a display unit.
Check that the item(s) delivered corresponds to the ordered item(s). The part number is printed on an identification label stuck on the converter shipping box and on a label on the converter itself.
14GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
15
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
3. Specifications and Order Numbers
Specifications
Oxix® Transmitter
Accuracy +/- 0,1% of reading
Measuring input RS 485
Analog output One active 4 - 20 mA, galvanically isolated (max. load 800 W)
Digital outputs
Digital inputs
Communication MODBUS
Interface RS 485 for connection to Display Unit or PLC
Power supply 10 - 30 V DC, or 24 V AC, 50 / 60 Hz ± 10 % or
Power consumption Max. 10 W
Cabinet material Polycarbonate, glass reinforced
Enclosure rating IP 67, NEMA 6
Temperature range - 20 … 60 °C
Weight 1,1 kg
One voltage-free electromechanical relay (max. 50 V DC / 1 A) One optically isolated MOSFET relay (max. 50 VAC / V DC / 120 mA)
One, max. 30 V DC, < 5 V DC = 0(low), > 10 V DC = 1(high), pulse length > 100 ms
®
RTU-mode, 9600 baud, 2-wire RS 485, slave-mode
115 V AC, 50 / 60 Hz ± 10 % or 230 V AC, 50 / 60 Hz ± 10 %
Display Unit
Enclosure rating Dust and waterproof IP 67, NEMA 6 (when mounted on converter)
Housing material Polycarbonate, glass reinforced
Protection lid Transparent polycarbonate
Display White backlit LCD-display (64 x 128 pixels) with softkeys
Indication Indication measurement, configuration and graph
Clock Real-time clock with built-in battery backup
Communication MODBUS
Interface RS 485
Memory 256 Kb Flash memory, 20.000 entries with date, time and value
Interface USB 1,1 type mini B, Female
Temperature range - 20 … 60 °C
®
RTU-mode, 9600 baud, 2-wire RS 485, master-mode
16GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Oxix® Sensor
Dissolved oxygen saturated
Measurement principle Optical, near infra-red fluorescence (l
Sensor check Automatic self-diagnostics
Pressure 6 bar (100 psi)
Materials Boby and Head: PU, PVC, silicone, epoxy and stainless steel 316 SS
Cleaning system Built-in air or water jet
Cable 4 x 0,326 mm
Cable length 10 m (33 ft.), can be extended up to 600 mm
Response time t
Flow velocity No flow required
Power supply 12 V DC (10 -16 V DC), approx. 10 mA
Output RS 485, 9600 baud, 2-wire
Accuracy Better than 1% of actual reading (0,02 mg/l)
Resolution Better than 0,02 mg/l
Temperature range 0 - 60 ºC
Temperature compensation 0 - 50 ºC
Enclosure rating IP 68 to IEC 529 (10 m), NEMA 6
Dimensions 50 x 130 mm (diameter x length)
Mounting
Weight 0,77 kg (1.7 lbs)
0 - 30.0 mg/l or ppm (parts per million) 0 - 120 %
= 475 nm, I
in
Optical lens: Sapphire O-ring: Viton Cable: PUR
2
, shielded, outer diameter Ø 5.0 mm (22 AWG)
less than 1 second, t95 less than 60 second
90
1,5 inch NPT thread 0,25 inch NPT thread for jet spray cleaning tube
= 600 nm)
out
17
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

Order Numbers

Oxix® Converters
®
206304 Oxix
206305 Oxix
206306 Oxix
206307 Oxix
206308 Oxix
206309 Oxix
Oxix® Sensor
206510 Oxix
Converter w/ display, 10 - 30 V DC
®
Converter w/ display, 230 / 115 V AC
®
Converter w/ display, 24 V AC
®
Converter without display, 10 - 30 V DC
®
Converter without display, 230 / 115 V AC
®
Converter without display, 24 V AC
®
Sensor
Accessories for Oxix
155230 Oxix
155236 Oxix
®
Sensors
®
Insertion adapter, 1½" NPT inner thread -> 2" RG outer thread
®
Immersion fitting
691120 Cable for Oxix
- 1½" NPT to 1¼" RG Adapter
®
sensor (PUR)
18GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
19
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

4. Sensor Installation

WARNING!
The Oxix® transmitter must not be powered until the transmitter is correctly mounted and connected!
WARNING!
Connections to the Oxix® transmitter must not be disconnected while the transmitter is powered.
Important! To ensure safe and optimal functionality of the Oxix® transmitter, the instructions below must be followed closely.
The sensor must be mounted in a location where it is possible to perform • a representative dissolved oxygen sat. measurement.
Fittings and mountings must be capable of carrying the sensor's • weight (0,77 kg / 1.7 lbs), and vibrations from for example pumps and other equipment must be avoided.
The system should be installed to facilitate access to the Oxix• operator panel and sensor without risking personal injury during normal operation, maintenance, cleaning of sensor, etc.
The Oxix•
®
transmitter must be connected to a mains outlet that complies
with national and local regulations.
All cables must be installed, connected and mounted to minimize me-• chanical impact and electrical influence from other installations.
The sensor cable’s maximum tensile load is adjusted to the sensor weight.•
The sensor is a electro-mechanical instrument. Be careful not to damage • the sensor lens. The accuracy will deteriorate if foreign material accumu­lates on it, or it is subject to abrasive cleaning procedures. Only use a soft cloth for cleaning. NEVER EVER a steel brush!
All • fuses are located on the Oxix
®
converter’s motherboard.
®
transmitter
20GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter

Physical Sensor Installation

Preparation

Select a position that can produce a valid and representative measuring value and provide sufficient space for physical cleaning of the sensor on a regular basis.
Positioning the Sensor
The positioning of the sensor is very application dependant. There are, how­ever, some "do"s and "don't"s that can be used as guidelines.
Do
In • aeration tanks or basins with significant fluctuation in depth the sensor should be mounted on a foating device that keeps the sensor submerged to a fixed depth from the surface.
In tanks or basins with no or little fluctuation in depth the sensor may • be mounted in a fixed position approximately 1/4 below the surface.
The optimal position at installations in flow-through aeration basins with • one measuring point is near the tank inlet.
The optimal positions at installations in flow-through aeration basins • with two measuring points are near the tank inlet and the tank outlet.
Don't
Do not position the sensor at or very close to the surface. Strong wind • aeration leads to increased dissolved oxygen concentration in the sur­face layer which may not be representative.
Do not use a fixed sensor position in aeration tanks/basins with signifi-• cant depth fluctuations. The dissolved oxygen saturation varies with the depth.
21
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

5. Sensor Calibration

The Oxix® sensor is factory calibrated. No further calibration is required.

Zero-point Calibration Check

A zero-point calibration check may be carried out as described below.
Required materials
1 bucket•
1 measuring cup•
1 tablespoon•
Sodium sulfite (Na•
Procedure
1. Dissolve 3 tablespoons of sodium sulfite ( Na container) containing 4 liters of clean tap water (1 US gallon, 0.9 UK gallon).
2. Stir for about 1 minute.
3. Lower the sensor into the solution and let it rest at least half an hour.
4. Make sure that the sensor's measuring element is facing down, and no air bubbles stick to the lense.
5. Check that the the Oxix
)
2SO3
®
readout is zero.
) in a bucket (or open
2SO3

2-point Calibration

2-point calibration is normally not required and is not recommended.
For very special applications it is possible to perform a 2-point calibration of the Oxix tive for further details.
®
transmitter. If required, contact an MJK sales or service representa-
22GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
23
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Mains supply
-mA- -D02--D01--DI--Display-
+ +A B C2 N02C NO N01 PE N L
- -
C1
+DCGND A B
Red
Black
White
Green
Shield
521275
Line
2 wire
shielded cable.
4 wire
shielded cable.
display unit
display unit
Display unitConverter #4Converter #1
Max. four MJK converters with Modbus can be connected to one display unit:
521274
2,5
Main supply
Supply Fuse
10-30 V DC 1 A T
24 V AC 500 mA T
115 V AC 125 mA T
230 V AC 63 mA T
Sensor supply
Fuse 200 mA T

6. Converter

Electrical Mounting

Warning!
The Oxix® converter/display unit must not be mounted in explosion hazardous areas!
Loosen the four screws (positions indicated by arrows) and remove the dis­play unit to gain access to the terminals.
24GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter

Power Supply

The flow converter must be supplied from a properly fused mains outlet, a 24 volt AC outlet, or a 10 - 30 V DC power supply/battery.
Power Supply 230 V AC, 115 V AC or 24 V AC
Terminal Designation
PE Protective ground
N 230 / 115 / 24 V AC neutral
L 230 / 115 / 24 V AC live
Power Supply 10 - 30 V DC
Terminal Designation
PE Protective ground
DC neutral
+ DC live
The internal fuse ratings are:
Internal Fuse Ratings
Voltage Rating Order no. Dimension
230 V AC 0.063 mA T 550030 5 x 20 mm
115 V AC 0.125 mA T 550035 5 x 20 mm
24 V AC 0.5 mAT 550049 5 x 20 mm
10 - 30 V DC 1.0 AT 550051 5 x 20 mm
The technical specifications for a 10 - 30 V DC power supply/battery are:
Technical Specifications for 10 - 30 V DC Power Supply
Power consumption without display < 5 W
Power consumption with display < 8 W
Peak start current @12 V DC,1 second Approx. 1,5 A
Peak start current @24 V DC,1 second Approx. 1 A
Display Connector
The display wires (and the remote display extension cable) are colour coded as follows:
Display Connector
Wire colour Designation
Red + supply
White RS-485 A
Grey RS-485 B
Black - supply
25
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Changing the Power Supply Voltage 230/115 V AC
To change the input mains voltage from 230 V AC to 115 V AC (or vice versa) proceed as follows:
1. Loosen the four screws on the front and lift out the display (see also page
23).
2. Note down the color and position of the wires in the terminal blocks, and then loosen the terminal screws.
3. Unscrew the two screws that hold the metal cover and then remove it.
4. Unscrew the four screws that hold the mother PCB (the printed circuit board with all the electrical components).
5. Remove the mother PCB and look at the back side. This is what you should expect to see for a 230 V AC configuration:
230 V AC single jumper position
6. Unsolder and remove the jumper indicated by the arrow.
26GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
7. See on the following picture where two jumpers must be inserted for a 115 V AC configuration:
115 V AC dual jumper positions
8. Insert and solder 2 jumpers (wires) in the positions indicated by the black arrows.
9. Turn around the PCB and replace the 63 mAT fuse on the right side with a 125 mAT fuse ( or vice versa going from 115 VAC to 230 VAC).
10. Re-insert the PCB and re-assemble the converter (see also steps 4, 3, 2 and 1).
27
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Analogue Output
The analogue output is an active output with a maximum load of 800Ω.
Analogue Output
Terminal Designation
AO + 4-20 mA AO
-
4-20 mA
The analogue output can be programmed for indication of primary units.
See details on page 53.
Digital Output
Oxix® has two digital outputs - DO1 with an opto (light triggered) relay and DO2 with an electro-mechanical relay.
They can both be programmed for the following functions:
- Sensor error
- High and low alarms
Opto Relay
Terminal Designation
DO1 Com Max. 50 V DC /120 mA DO1 NO
Electro-mechanical relay
Terminal Designation
DO2 Com Max. 50 V DC /1 A DO2 NO
DO1 shares the common terminal (Com) with DI.
See also application examples in Appendix E. Digital Input/Output Connec­tions on page 101.
28GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Digital Input
Oxix® has one digital input which can be activated with a voltage higher than 10 V DC and de-activated with a voltage lower than 5 V DC.
Digital Input
Terminal Designation
DI Com DI Max. 30 V DC
The digital input (DI) can be programmed for the following functions:
- start and pause cleaning function
- Solid curve
See also application examples in Appendix E. Digital Input/Output Connec­tions on page 101.
29
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Mains supply
-mA- -D02--D01--DI--Display-
+ +A B C2 N02C NO N01 PE N L
- -
C1
+DCGND A B
Red
Black
White
Green
Shield
521275
Line
2 wire
shielded cable.
4 wire
shielded cable.
display unit
display unit
Display unitConverter #4Converter #1
Max. four MJK converters with Modbus can be connected to one display unit:
521274
2,5
Main supply
Supply Fuse
10-30 V DC 1 A T
24 V AC 500 mA T
115 V AC 125 mA T
230 V AC 63 mA T
Sensor supply
Fuse 200 mA T
Converter Terminals
Oxix® converter terminals
Flip the green label with "SuSix® / Oxix®" printed on it (see arrow) to the right to gain access to the sensor connection terminals (see next page).
30GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Mains supply
-mA- -D02--D01--DI--Display-
+ +A B C2 N02C NO N01 PE N L
- -
C1
Line
2 wire
shielded cable.
4 wire
shielded cable.
display unit
display unit
Display unitConverter #4Converter #1
Max. four MJK converters with Modbus can be connected to one display unit:
521274
2,5
Main supply
Supply Fuse
10-30 V DC 1 A T
24 V AC 500 mA T
115 V AC 125 mA T
230 V AC 63 mA T
Sensor supply
Fuse 200 mA T
+DCGND A B
Red
Black
White
Green
Shield
521275
www.mjk.com
DK: +45 45 56 06 56 NO: +47 69 20 60 70 SE: +46 53 31 7 7 50 NL: +31 251 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Sensor and Mounting Kit / Connection Board Terminals
Important!: Make sure the right wiring indication label is fitted below the terminal block (label indicated by arrow). The label must clearly be marked Oxix®.
Wrong wiring may cause sensor destruction !
Connect the sensor wires to the Oxix® converter with the wires coming from the sensor as shown below (see arrow).
Terminals
Wires from sensor Converter terminals
Oxix® converter terminals
- not used Black GND
White A
Green B
Shield - (shield GND)
Red +DC
31
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
7. System Configuration Examples
MJK Oxix® transmitters share communication protocols with other MJK prod­ucts like the MJK Magflux® flow meter and the MJK Connect data transmitter, and can also share display units. This section illustrates two configuration examples and includes cable connections and part numbers.
Magflux® with blind lid and Oxix® with display unit
Oxix® with blind lid, Magflux® with blind lid and a remote display unit
32GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Magflux® with Blind Lid and Oxix® with Display Unit
Configuration: Magflux® flow sensor/converter with a blind lid connected to a Oxix
Part numbers for this configuration:
2x7xxx MagFlux® Flow Sensor
2x7791x MagFlux® Connverter with blind lid
579035 MagFlux® Gel Potting Kit IP68
2-wire, shielded, twisted cable (Magflux® <--> Oxix®)
206510 Oxix® Sensor
206304-6 SuSix® /Oxix® Converter with Display Unit
®
transmitter with display unit
33
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Mains
supply
-mA- -D02--D01--DI--Display-
+ +A B C2 N02C NO N01 PE N L
- -
C1
Line
2 wire
shielded cable.
4 wire
shielded cable.
display unit
display unit
Display unitConverter #4Converter #1
Max. four MJK converters with Modbus can be connected to one display unit:
521274
2,5
Main supply
Supply Fuse
10-30 V DC 1 A T
24 V AC 500 mA T
115 V AC 125 mA T
230 V AC 63 mA T
Sensor supply
Fuse 200 mA T
+DCGND A B
Red
Black
White
Green
Shield
521275
www.mjk.com
DK: +45 45 56 06 56 NO: + 47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 25 1 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533

Wiring Schematic #1

for " Magflux® with Blind Lid and Oxix® with Display Unit" from the opposite page (page 32).
Oxix® with Display Unit
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot "Display -"
12Shield
Magflux® with Blind Lid
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot "Display -"
12Shield
34GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
SuSix® with Blind Lid, Magflux® with Blind Lid and a Re­mote Display
Configuration: An Oxix® transmitter with a blind lid connected to a Mag­flux® flow sensor with a blind lid and common remote Display Unit.
Part numbers for this configuration:
206510 Oxix® Sensor
206304-6 SuSix® /Oxix® Converter
2x7xxx MagFlux® Flow Sensor
2x791x MagFlux® Converter with blind lid
579035 MagFlux® Gel Potting Kit IP68
207930 Wall Mounting Kit
2-wire, twisted shield cable (Oxix® <--> Magflux®)
4-wire, twisted shield cable (Magflux® <--> Display Unit)
35
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Mains
supply
-mA- -D02--D01--DI--Display-
+ +A B C2 N02C NO N01 PE N L
- -
C1
Line
2 wire
shielded cable.
4 wire
shielded cable.
display unit
display unit
Display unitConverter #4Converter #1
Max. four MJK converters with Modbus can be connected to one display unit:
521274
2,5
Main supply
Supply Fuse
10-30 V DC 1 A T
24 V AC 500 mA T
115 V AC 125 mA T
230 V AC 63 mA T
Sensor supply
Fuse 200 mA T
+DCGND A B
Red
Black
White
Green
Shield
521275
www.mjk.com
DK: +45 45 56 06 56 NO: + 47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 25 1 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Wiring Schematic #2
for "Oxix® with Blind Lid and a Magflux® with Blind Lid connected to a com­mon wall mounted Display Unit" on the opposite page (page 34).
Oxix® with Blind Lid
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot "Display -"
®
Magflux
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot "Display -".
4-wire, shielded twisted cable
1. Connect lead "1" to slot "Display +"
2. Connect lead "2" to slot "Display A"
3. Connect lead "3" to slot "Display B"
4. Connect the twisted shield to slot "Display -".
with Blind Lid
12Shield
12Shield
1
2
3
Wall Mounted Display Unit
4-wire, shielded twisted cable
1. Connect lead "1" to slot "Display +"
2. Connect lead "2" to slot "Display A"
3. Connect lead "3" to slot "Display B"
4. Connect the twisted shield to slot "Display -".
4 and shield
1
2
3
36GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
37
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

8. Startup

Initial Checks before Power-up

Before switching on power the following steps must be checked.
1. The local mains power supply voltage corresponds to the voltage printed on the identification label of the transmitter.
2. All electrical connections are made in accordance with the electrical con­nection diagram shown with special emphasis on the sensor connection described on page 30.
3. All terminal screws are tightened.
4. All cable glands are tightly secured.
5. All grounding connections are made in accordance with the instructions in this manual (see page 21).

Initial Power-up

1. Turn on power.
2. Within 60 seconds the Oxix sor detection program. During this process you might briefly see "SuSix1" and/or "MagFlux1" on the display. This is normal and to be expected.
®
sensor will be found by the automatic sen-
Initial Measurement
1. Make sure that the sensor is completely covered by liquid.
2. Turn on power to the transmitter and wait 15 minutes before taking the initial measurement.
Language Selection
1. The default display language is English. If another language is required, proceed with step 2.
2. Select "Display Setup" from the Main Menu.
3. Select "Language" from the Display Setup menu and chose the required language (see page 73).
38GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Oxix1
Oxygen sat.
%
Select Set up
Converter Overview Menu
Display Read-out, one connected unit
All Oxix® display read-outs are illustrated and described in this manual.
Chapter 6. Oxix® Menus (on page 41) gives a detailed description of the 5-line LCD screen displays shown during setup, configuration and normal opera­tion.
Display and keyboard for reading and programming
The display shown above will in the remaining part of this manual be stylized to appear like this:
39
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Oxix1
Oxygen sat.
%
Converter Overview Menu
Oxix1 91.9 % Susix1 578 mg/L
Back Select Display
Oxix1
Oxygen sat.
%
Select Set up
Converter Overview Menu
Specify Main Screen Data Logger Password Set Sensor Name Converter Setup Display Setup
Back Select OK
Select line Linie 1 - Sensor name Linie 2 - Meas. Linie 3 - Linie 4 - Linie 5
Back Select OK
Specify Main Screen
Select line content
SuSix1 [X]
Meas.
FTU
Select Setup
Main Menu
Converter Top Display
SuSix1 [X]
Meas.
FTU
Select Setup
Converter Top Display
Specify Main Screen Data Logger Password Set Sensor Name Converter Setup Display Setup
Back Select OK
Main Menu
SuSix1 [X]
Prim.
mg/l
Second 0.1%
Select Setup
Display Read-out, several connected units
When several MJK units are interconnected, for example an Oxix® dissolved oxygen transmitter and a SuSix® turbidity and suspended solids transmitter with different names and Modbus ID numbers, a "Display Overview Menu" is available at top level (press "Back" repeatedly):
All connected units are displayed and sorted by their ID numbers, and conse­quently each unit can be selected and set up as required:
Important: More connected units can only by managed as described above, if each unit have been assigned a unique name and Modbus ID number. See page 75 for details.
Display Keys
The keys and the soft keys (variable key functions determined by the display firmware) are used for initial programming and normal operation of the flow meter.
The function(s) of the four available keys is displayed at the bottom of the dis­play. The symbols and actual functions are described in detail in the sections that describe the individual menus.
or
40GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Contrast Adjustment
Adjust the display contrast by pressing the two outmost keys simultaneously (indicated by the keys) and press the up/down keys as required.
Save the new setting by pressing the two outmost keys simultaneously.
System Reset
You can reset and refresh all system displays and key combinations by press­ing all four keys simultaneously. It is a so-called coldboot, where all displays are refreshed and all settings/configurations are kept with only one exception: the language setting.
This feature is especially useful during for example a service call, where the display language (Dutch, Danish, Swedish etc.) cannot be understood by the local serviceperson. A system reset immediately sets the display language to (GB) English.
The system reset must not be confused with the ultimate "Recover factory settings" (see page 60) which deletes all local configurations/settings and replaces them with default factory settings.
41
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Oxix1
Oxygen sat.
%
Select Set up
Converter Overview Menu
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Login
Set log interval
Datalogger
Select line
Specify main Screen Converter setup
Password
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu
9. Oxix® Menus
All the Oxix® menus and sub menus are shown and described in the following sections.
Continous overviews of the menu and sub-menu structures are presented at the end of this manual:
Converter Overview Menu• (page 102)
Main Menu Overview• (page 102)
Converter Setup Overview• (page 106)
Service Menu Overview• (page 114)
Display Setup Menu Overview• (page 120)

Converter Overview Menu

1. Press the "Setup" key on the Oxix® display (see below) to enter the Oxix®
Converter Overview Menu.
2. Press the up/down keys to highlight the required menu line in the Main Menu (here: Specify main screen) and then press OK to select it.
Main Menu
42GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Login Activate/deactivate Set new password
Back Select OK
Set log interval
0020 sec
Back Select >
Datalogger
Sensor name Oxix 1
Back Select >
Sensor name
Select line Linie 1 - Sensor name Linie 2 - Oxygen sat. Linie 3 ­Linie 4 ­Linie 5 Back Select OK
Specify main Screen Converter setup
Select line content Not in use Sensor name Measurement Temperature Clock Back Select OK
Password
Type password
0000
Back Select >
Set new password
Back Select >
Activate Deactivate
Back Select OK
Login
Activate/
Deactivate
Set new password

Specify Main Screen

The "Specify Main Screen" menu allows you to customize the Oxix® display to suit your requirements. You can add and remove five available display lines and name/configure them individually.
1. Press the up/down keys to highlight the required menu line and then press OK to select it.
2. Press the up/down keys to highlight the required option and press OK.
The available options are:
Not in use
The line not used. The freed space will be used by the other lines.
Sensor name
TBD
Measurement
Measurement displays the measured values in time in the predefined unit:
%, mg/l or ppm.
Note: The units are defined later on in the "Converter setup/Units" menu on
page 52.
43
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Temperature
The temperature may be displayed in degrees Celcius (ºC) or in degrees Fahrenheit (ºF). See also page 52.
Clock
Time and date
The size of the display lines will automatically increase or decrease as the number of display lines is removed or added to maximize the field of view for the measured values
44GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Login Activate/deactivate
Set log interval
0020 sec
Datalogger
Select line
Linie 1 - Sensor name
Specify main Screen Converter setup
Password
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu
Login Activate/deactivate Set new password
Back Select OK
Set log interval
0020 sec
Back Select >
Datalogger
Sensor name Oxix 1
Back Select >
Sensor name
Averaging Units mA output Cleaning High Alarm
Password
Datalogger
Oxix® provides a data logger with a capacity of approx. 20,000 entry points. See Appendix D on page 97 for examples and descriptions of log files.
The data logger operates after the FIFO principle (First In, First Out). If the data logger is full and new data are coming in, the oldest data are dis­carded.
1. Press the up/down keys to highlight the required menu line
(here: Datalogger) and then press OK to select it.
2. Press the up/down keys to change the highlighted digit, and then press
the right-key (>) to proceed to the next digit.
3. Continue until all digits have been set and press OK.
The log interval can be set in intervals from 10 seconds up to 9999 seconds.
45
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
The data log contains:
Date•
Time•
Dissolved oxygen values •
In case of a power failure, the data logger continues when power returns.
If more converters are connected to one Display Unit, each converter has its own individual log interval and can be sorted.
All converters share the same memory of 20.000 entry points.
The log data can be shown on the Display Unit or be stored in an external CSV file. MJK-Field Link software is needed for transfer of data into CSV file format via the USB port in the display unit. The format can be read in for ex­ample Microsoft
®
Excel® (see Appendix D on page 97 for details).
An example of the log capacity of one sensor versus the time interval is shown in the following table.
Log interval Log capacity
Seconds Minutes Hours Days
10 - 55,56 2,31
30 - 166,67 6,94
- 1 333 14
- 5 1667 69
- 10 3333 139
- 30 10000 417
- 60 20000 833
46GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Oxix1
Oxygen
98 %
Select Graph
100 %
2008-09-21
11:25:43 84 %
<< >>< >
Oxix1
Oxygen
98 %
Select Graph
Graph Display
The content of the Data Logger can displayed on the Display Unit by pressing the up/down keys, highlighting "Oxygen sat." and selecting "Graph"
The Display Unit then shows the Graph screen. To return to the Main Menu display screen, select "esc" by pressing the up/down keys simultaneously.
The Y axis is automatically scaled according to the maximum value of the mA output.
Double-arrow keys jump forward and backward one screen frame at a time. Single-arrow keys move the cursor forward and backward on the screen.
The actual values at the cursor position is shown at the bottom of the screen.
47
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Login
Set log interval
Datalogger
Select line
Specify main Screen Converter setup
Password
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu
Login Activate/deactivate Set new password
Back Select OK
Set log interval
0020 sec
Back Select >
Datalogger
Sensor name Oxix 1
Back Select >
Sensor name
Averaging Units mA output Cleaning High Alarm Low Alarm Sensor Error Status Factory settings Service menu
Back Select OK

Password

Type password
0000
Back Select >
Set new password
0000
Back Select >
Activate Deactivate
Back Select OK
Login
Activate/
Deactivate
Set new password
Password
A password provides (and prevents) access to all the settings in the Display and Converter unit. The code consists of a numeric 5-digit code between 0 and 65535. If your current password is lost or forgotten, contact MJK for instructions.
1. Press the up/down keys to highlight the required menu line (here: Password) and then press OK to select it.
2. Press the up/down keys to highlight the required option and then press OK to select it.
The available options are:
Login
Use the up/down keys to set the digits one by one. Continue with > until
all digits have been set and then press OK.
48GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Activate/deactivate
Write protection. Selecting "Activate" means that a password must be en-
tered to change vital settings. "Deactivate" disables password protection. If your current password is lost or forgotten, the password protection can be overruled with the code "01750".
Change password
The present 5-digit password can be changed as required.
49
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Login
Set log interval
Datalogger
Select line
Specify main Screen Converter setup
Password
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu
Sensor Name
A unique name and/or number, a function or a location can be assigned for a sensor (here: "Oxix1"). It is consequently shown on the main display with up to 4 display lines.
1. Press the up/down keys to highlight the required menu line (here: Sensor Name) and then press OK to select it.
2. To change the default sensor name "Oxix1" to for example "Aeration Tank Inlet", press the left/right keys to highlight the wanted character.
50GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Set log interval
0020 sec
Back Select >
Datalogger
Select line Linie 1 - Sensor name Linie 2 - Oxygen sat. Linie 3 - Linie 4 - Linie 5 Back Select OK
Specify main Screen Converter setup
Select line content Not in use Sensor name
Login
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu
Oxix1 91.9 % Susix1 578 mg/L
Back Select Display
Display setup
Sensor name Aeration Tank Inlet
Back Select >
Sensor name
3. Press the up/down keys to change the highlighted character and then press > to proceed to the next character.
4. Continue with > until all numeric and alphabetical characters have been set.
5. Press OK.
The available characters depend on the chosen language. English, for exam­ple, provides the following character set: a b c d e f g h i j k l m n o p q r s t u v w x y z A B C D E F G H I J K L M N O P Q R S T U V W X Y Z <space> 1 2 3 4 5 6 7 8 9 and 0.
51
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Login
Set log interval
Datalogger
Select line
Specify main Screen Converter setup
Password
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu

Converter setup

Language Set Clock Modbus Factory settings Display SW version
Back Select OK
Display setup
Averaging Units mA output Cleaning High Alarm Low Alarm Sensor Error Status Factory settings Service menu
Back Select OK
Converter Setup
"Converter setup" provides configuration options for calibration, alarms, units, etc. See detailed descriptions in the following sections, and a continuous overview on page 106 at the end of this manual.
1. Press the up/down keys to highlight the required menu line (here: Converter setup) and then press OK to select it.
2. A number of options appears beginning with "Averaging":
52GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Averaging
Units
mA output
D.O. Units % mg/l ppm
Enable mA output? Yes No
Back Select OK
Averaging
0001 sec
(from 1 sec to 120 min)
Back Select >
This menu is for display only.
Cleaning
Electro-mechanical relay Opto relay Not in use
Back Select OK
Units
mA output
D.O. Units % mg/l ppm
Back Select OK
4mA output
4 mA = 00.00 mg/l
Back Select OK
Yes
Enable mA output? Yes No
Back Select OK
Temperature unit °C °F
Back Select OK
Temperature
This menu is for display only.
Cleaning
Interval Use DI to clean Clean now Maintenance Back Select OK
Electro-mechanical relay Opto relay Not in use
Back Select OK
Averaging
To avoid unwanted fluctuations in the measurement reading, the "Averaging" options smoothes the measurements over a predefined period of time.
Units
The measuring and temperature units can be set from this menu.
53
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
mA output
20 mA output
20 mA = 25.00 mg/l
Back Select OK
4mA output
4 mA = 00.00 mg/l
Back Select OK
Yes
Enable mA output? Yes No
Back Select OK
Clean now Yes No
Clean now
Cleaning
Interval Use DI to clean Clean now Maintenance Back Select OK
Interval Off On
Use DI to cleanInterval Maintenance
Use DI to clean No Yes
Electro-mechanical relay Opto relay Not in use
Back Select OK
mA Output
When a Oxix® is connected to a power supply for the first time, the mA out­put is automatically set to provide 4 mA at zero and 20 mA at the theoretical maximum value.
Changes in the mA setting will not affect the relay output settings.
Note: Both values can be set in the range 10 % to 120 %, making it
possible to increase or decrease the mA signal.
The mA output is an active output, and the maximum load is 800 Ohm.
The upper limit for the mA output is 20,5 mA 3,75 mA indicates the mA output is not in use 3,5 mA indicates values under 4 mA level 20,5 mA indicates value higher than 20 mA level.
54GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Clean now Yes No
Back Select OK
Reset Maintenance Yes No
Back Select OK
Maintenance Runtime 1000 hour Count 500
Back Select OK
Clean now
Select cleaning interval
060 Min
(20 - 255 Min)
Back Select >
Cleaning
Interval Use DI to clean Clean now Maintenance Back Select OK
Interval Off On
Back Select OK
Use DI to clean
Interval Maintenance
DI function NO NC
Back Select OK
Use DI to clean No Yes
Back Select OK
Delay
00100 ms
Back Select >
Electro-mechanical relay Opto relay Not in use
Back Select OK
Cleaning time
005 sec
Back Select OK
Cleaning time
005 sec
Back Select OK
Measurement delay
10 sec
Back Select OK
Cleaning
Air or water jet cleaning can be defined based on local conditions.
55
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Clean now Yes No
Back Select OK
Reset Maintenance Yes No
Back Select OK
Maintenance Runtime 1000 hour Count 500
Back Select OK
Clean now
Use DI to cleanInterval Maintenance
DI function NO NC
Back Select OK
Use DI to clean No Yes
Back Select OK
Delay
00100 ms
Back Select >
Cleaning time
005 sec
Back Select OK
A typical cleaning interval for aeration basins would be 2 hours (120 minutes). In colder climates, condensation may form and then freeze in the tubing. This can be prevented by setting the interval to 5 - 20 minutes.
The cleaning time would typically be 30 seconds.
The measurement delay (recovery time) may also be used to coordinate with other processes. The delay merely freezes the last reading before cleaning begins. A typical value would be 60 seconds.
56GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
On Setpoint
5,50 %
Back Select >
Relay function NO NC
Back Select >
Off Setpoint
5,75 %
Back Select >
Delay
00100 ms
Back Select >
Electro-mechanical relay
On Setpoint
7,00 %
Back Select >
High Alarm
Relay function NO NC
Back Select >
Off Setpoint
6,75 %
Back Select >
Electro-mechanical relay Opto relay Display only Not in use
Back Select OK
Delay
00100 ms
Back Select >
Electro-mechanical relay
Opto relay Display only
On Setpoint
7,00 %
Back Select >
Relay function NO NC
Back Select >
Off Setpoint
6,75 %
Back Select >
Delay
00100 ms
Back Select >
On Setpoint
7,00 %
Back Select >
Off at flow
6,75 %
Back Select >
Delay
00100 ms
Back Select >
High alarm
A set point can be set to trigger an alarm (here: a high alarm) and to activate one of two relays or simply to display the alarm on the display.
57
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Electro-mechanical relay
Delay
00100 ms
Back Select >
mA indication High Low Off Back Select OK
Relay function NO NC
Back Select OK
Low Alarm
Electro-mechanical relay Opto relay Display only Not in use
Back Select OK
On Setpoint
5,50 %
Back Select >
Relay function NO NC
Back Select >
Off Setpoint
5,75 %
Back Select >
Delay
00100 ms
Back Select >
Opto relay
Display only
On Setpoint
5,50 %
Back Select >
Relay function NO NC
Back Select >
Off Setpoint
5,75 %
Back Select >
Delay
00100 ms
Back Select >
On Setpoint
5,50 %
Back Select >
Off Setpoint
5,75 %
Back Select >
Delay
00100 ms
Back Select >
Electro-mechanical relay
Low alarm
A set point can be set to trigger an alarm (here: a low alarm) and to activate one of two relays or simply to display the alarm on the display.
58GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Sensor error
Electro-mechanical relay Opto relay Display only Not in use
Back Select OK
Electro-mechanical relay
Opto relay
Display only
Delay
00100 ms
Back Select >
Delay
00100 ms
Back Select >
Delay
00100 ms
Back Select >
Relay function NO NC
Back Select OK
mA indication High Low Off Back Select OK
mA indication High Low Off Back Select OK
mA indication High Low Off Back Select OK
Relay function NO NC
Back Select OK
Sensor error
Oxix® has two digital outputs - DO1 with an opto (light triggered) relay and DO2 with an electro-mechanical relay. They can both be programmed for a sensor error, whenever the 4 - 20 mA range has been exceeded.
59
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Type password
00000
Back Select >
Service menu
Pending errors xxx xxx
Back Select OK
Factory setting
Recover factory settings Cancel Default Force to SuSix Force to Oxix Back Select OK
Status
DO 1 [ x ] Selected function DO 1 [ \ ] Selected function DI [ ] Selected function mA output 10,234 mA
Back Select OK
Pending warnings xxx xxx
Back Select OK
Pending information xxx xxx
Back Select OK
Pending alarms xxx xxx
Back Select OK
Status
The "Status" option provides an overview of the in- and output status.
The check boxes can contain four different characters:
X A cross (X) in a check box indicates an active state
/ A blinking forward slant (/) in a check box indicates that the input/
output is in the process of being activated. Eventually it turns into a steady X.
60GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Type password
00000
Back Select >
Service menu
Factory setting
Recover factory settings Cancel Default Force to SuSix Force to Oxix Back Select OK
\ A blinking backward slant (\) in a check box indicates that the
input/output is in the process of being de-activated. Eventually the check box is cleared.
An empty check box indicates a de-activated state.
DO 1 [ X ] Selected function
The selected function linked to digital output D01 is active (an "X" in the
check box).
DO 2 [ X ] Selected function
The selected function linked to digital output D02 is in the process of be-
ing de-activated (an "X" in the check box).Digital output D02 for Totalizer Forward is activated (a ticked check box).
DI [ ] Selected function
The selected function linked to digital input DI de-activated (an empty
check box). A "selected function" could for example be "Cleaning".
mA output 10.234 mA
The present analog output current is 10.234 mA. Pending errors, alarms, warnings and infos are displayed for the selected in- or output.
Factory settings
The "Factory settings" menu provides resetting to the original factory settings (except for the language setting).
61
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Sensor Data Sensor Calibration Converter SW. Version. Product data Reset Counter Time Internal meas. & cal. Calibrate mA Test Primary value Read eventlog Comunication log Back On Stock Autodetect sensor type Main supply frequency Back Select OK
Sensor data
Sensor serial number
30487
Sensor software version
2
Back OK
Converter 846001-002
Bootload 840020-002
Back
Set actual measurement
10,20 mg/l
Back Select OK
Sensor calibration
Sensor calibration 1 point calibration 2 point calibration Cancel Back Select OK
OK
Clear calibration No Yes
Back Select OK
Confirm zero oxygen No Yes
Back Select OK
1 point
2 point
Type password
00000
Back Select >
Service menu
Sensor Data Sensor Calibration Converter SW. Version. Product data Reset Counter Time Internal meas. & cal. Calibrate mA Test Primary value Read eventlog Comunication log Back On Stock Autodetect sensor type Main supply frequency Back Select OK
Sensor data
Sensor serial number
30487
Sensor software version
2
Back OK
Converter 846001-002
Bootload 840020-002
Back
Sensor calibration
Sensor calibration 1 point calibration 2 point calibration Cancel Back Select OK
OK
Service Menu - Digest
The "Service Menu" provides options intended for service personnel during installation, calibration, operation, monitoring and maintenance.
All parameters can be read without any restrictions, but certain parameters can only by changed after a password has been typed.
62GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Averaging
Units
mA output
Converter setup
D.O. Units % mg/l ppm
Back Select OK
Averaging Units mA output Cleaning High alarm Low alarm Sensor error Status Factory settings Service menu
Back Select OK
Yes
Enable mA output? Yes No
Back Select OK
Averaging
0001 sec
(from 1 sec to 120 min)
Back Select >
This menu is for display only.
Sensor Data Sensor Calibration Converter SW. Version. Product data Reset Counter Time Internal meas. & cal. Calibrate mA Test Primary value Read eventlog Comunication log Back On Stock Autodetect sensor type Main supply frequency Back Select OK
Sensor data
Sensor serial number
30487
Sensor software version
2
Back OK
Converter 846001-002
Bootload 840020-002
Back
Set actual measurement
10,20 mg/l
Back Select OK
Sensor calibration
Sensor calibration 1 point calibration 2 point calibration Cancel Back Select OK
OK
Clear calibration No Yes
Back Select OK
Confirm zero oxygen No Yes
Back Select OK
1 point
2 point
Type password
00000
Back Select >
Service menu
Sensor Data Sensor Calibration Converter SW. Version. Product data Reset Counter Time Internal meas. & cal. Calibrate mA Test Primary value Read eventlog Comunication log Back On Stock Autodetect sensor type Main supply frequency Back Select OK
Sensor data
Sensor serial number
30487
Sensor software version
2
Back OK
Converter 846001-002
Bootload 840020-002
Back
Sensor calibration
Sensor calibration 1 point calibration 2 point calibration Cancel Back Select OK
OK

Service Menu - Detailed

The "Service Menu", a submenu in "Converter setup", is intended for MJK and other authorized service personnel during installation, calibration, commissioning, operation, monitoring and maintenance.
All parameters can be read without any restrictions, but certain parameters can only by changed after a password has been typed.
63
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Sensor data
Sensor serial number
30487
Sensor software version
2
Back OK
Converter SW version
Converter 846001-002
195
Bootload 840020-002
684
Back
Product ID
Back Select >
Sensor calibration
Sensor calibration 1 point calibration 2 point calibration Cancel Back Select OK
Converter SW version
Converter 846001-002
195
Bootload 840020-002
684
Back
Product data
Product ID
000000000000
Back Select >
HW type no
0510305004
Back Select >
Track no
00000
Back Select >
Product type no
00000206305
Back Select >
Set actual measurement
10,20 mg/l
Back Select OK
Sensor calibration
Sensor calibration 1 point calibration 2 point calibration Cancel Back Select OK
OK
Clear calibration No Yes
Back Select OK
Confirm zero oxygen No Yes
Back Select OK
1 point
2 point
Reset Counter 1 System Runtime 2304 Since Reset 2304 Runtime (days) 0
Back Select OK
Set actual measurement
10,20 mg/l
Back Select OK
Sensor data
"Sensor data" displays the factory-assigned sensor serial number and
the present sensor software version.
Sensor calibration
Sensor calibration is read-only. Only MJK servicepersons can access
the option using the proper password.
64GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Converter SW version
Converter 846001-002
195
Bootload 840020-002
684
Back
Product data
Product ID
000000000000
Back Select >
Calibrate Vcc Calibrate V_ref
Back Select OK
Internal meas. & cal.
Reset counter time
Reset Counter 1 System Runtime 2304 Since Reset 2304 Runtime (days) 0
Converter SW ver.
The converter software version and the bootload version are displayed.
From that you can determine whether a software (firmware) upgrade is required or not.
65
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Product data
Product ID
000000000000
Back Select >
HW type no
0510305004
Back Select >
Track no
00000
Back Select >
Product type no
00000206305
Back Select >
Calibrate Vcc Calibrate V_ref
Back Select OK
Internal meas. & cal.
View Vcc
- - - -
Back Select OK
Calibrate Vcc
- - - -
Back Select >
View V_ref
- - - -
Back Select OK
Calibrate Vref
- - - -
Back Select >
Reset counter time
Reset Counter 1 System Runtime 2304 Since Reset 2304 Runtime (days) 0
Back Select OK
Calibrate Vcc Calibrate V_ref
Back Select OK
Internal meas. & cal.
Reset counter time
Reset Counter 1 System Runtime 2304 Since Reset 2304 Runtime (days) 0
Back Select OK
Product Data
Key product data like ID, hardware type no., track no. and HW type no.
are displayed.
Reset Counter Time
The counter can be reset for fault finding, error correction and similar
service procedures, and the actual runtime is displayed.
66GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Calibrate Vcc Calibrate V_ref
Back Select OK
Internal meas. & cal.
View Vcc
- - - -
Back Select OK
Calibrate Vcc
- - - -
Back Select >
View V_ref
- - - -
Back Select OK
Calibrate Vref
- - - -
Back Select >
Internal Meas. & Cal.
Internal measurements and calibrations are for service personnel use only!
The supply voltage (Vcc) and the reference voltage (Vref) can be viewed and adjusted.
67
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Set primary value
- - - - mg/l
Back Select >
Test primary value
Enable test Yes No
Back Select OK
Current Set fixed current Calib. current DAC zero current DAC top current
Back Select OK
Calib. current
- - - - uA
Back Select OK
Calibrate mA
Current
- - - - uA
Back Select OK
Set fixed current Release mA Minimum mA 4 mA 12 mA 20 mA Maximum mA Back Select OK
DAC zero current
- - - - uA
Back Select OK
DAC top current
- - - -
Back Select OK
No Hour Event Extension 1 154 0 0 2 154 0 0 3 154 0 0 4 154 0 0 Back Select >
Read eventlog
No Hour Event Extension 5 154 0 0 6 154 0 0 7 154 0 0 8 154 0 0 Back Select >
No Hour Event Extension 13 154 0 0 14 154 0 0 15 154 0 0 16 154 0 0 Back Select >
No Hour Event Extension 9 154 0 0 10 154 0 0 11 154 0 0 12 154 0 0 Back Select >
No Hour Event Extension 17 154 0 0 18 154 0 0 19 154 0 0 20 154 0 0 Back Select >
Calibrate mA
From this menu the mA reading can be calibrated against a multimeter.
68GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Set primary value
- - - - mg/l
Back Select >
Test primary value
Enable test Yes No
Back Select OK
No Hour Event Extension 1 154 0 0 2 154 0 0 3 154 0 0 4 154 0 0 Back Select >
Read eventlog
No Hour Event Extension 5 154 0 0 6 154 0 0 7 154 0 0 8 154 0 0 Back Select >
Recover Factory Settings Cancel Back on Stock Sensor Factory Reset
Back Select OK
Back on stock
Sensor Factory Reset No Yes
Back Select Ok
Modbus Rx ok 4950 Rx CRC error 4564 Rx paritet error 4545 Rx adr. error 4566 Rx wrong id 4950 Tx count 3454 Back Select OK
Communication log
Sensor Rx ok 4950
Test primary value
"Test primary value" is for MJK service personnel use only. It can be set to
for example 100% to test a state or condition for a given input value.
69
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
No Hour Event Extension 1 154 0 0 2 154 0 0 3 154 0 0 4 154 0 0 Back Select >
Read eventlog
No Hour Event Extension 5 154 0 0 6 154 0 0 7 154 0 0 8 154 0 0 Back Select >
No Hour Event Extension 13 154 0 0 14 154 0 0 15 154 0 0 16 154 0 0 Back Select >
No Hour Event Extension 9 154 0 0 10 154 0 0 11 154 0 0 12 154 0 0 Back Select >
No Hour Event Extension 17 154 0 0 18 154 0 0 19 154 0 0 20 154 0 0 Back Select >
Recover Factory Settings Cancel Back on Stock Sensor Factory Reset
Back Select OK
Back on stock
Sensor Factory Reset No Yes
Back Select Ok
Modbus Rx ok 4950 Rx CRC error 4564 Rx paritet error 4545 Rx adr. error 4566 Rx wrong id 4950 Tx count 3454 Back Select OK
Communication log
Sensor Rx ok 4950 Rx CRC error 4564 Rx paritet error 4545 Rx count 4950 Tx count 3454
Back Select OK
Autodetect sensor type Use Not in use
Back Select OK
Autodetect sensor type Mains supply frequency
Read Eventlog
View the events within a certain time frame.
70GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Recover Factory Settings Cancel Back on Stock Sensor Factory Reset
Back Select OK
Back on stock
Sensor Factory Reset No Yes
Back Select Ok
Modbus Rx ok 4950 Rx CRC error 4564 Rx paritet error 4545 Rx adr. error 4566 Rx wrong id 4950 Tx count 3454 Back Select OK
Communication log
Sensor Rx ok 4950 Rx CRC error 4564 Rx paritet error 4545 Rx count 4950 Tx count 3454
Back Select OK
Autodetect sensor type Use Not in use
Back Select OK
Mains supply frequency 50 Hz 60 Hz
Back Select OK
Autodetect sensor type Mains supply frequency
Recover Factory Settings Cancel Back on Stock Sensor Factory Reset
Back Select OK
Back on stock
Sensor Factory Reset No Yes
Back Select Ok
Autodetect sensor type Use Not in use
Back Select OK
Mains supply frequency 50 Hz 60 Hz
Back Select OK
Autodetect sensor type Mains supply frequency
Communication log
View the Modbus and sensor communication logs. Primarily intended for
fault detection.
Back on stock
The factory settings can be re-etablished with the "Back on stock" option.
71
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Autodetect sensor type Use Not in use
Back Select OK
Mains supply frequency 50 Hz 60 Hz
Back Select OK
Autodetect sensor type Mains supply frequency
Mains supply frequency 50 Hz 60 Hz
Back Select OK
Mains supply frequency
Autodetect sensor type
The option can be toggled as required. "Use" is the factory default setting.
Mains supply frequency
Select the national/local mains supply frequency: 50 Hz for Europe, Africa, Asia, Australia and South America, and 60 Hz for USA, Canada, Mexico, Latin America and the Carabian.
72GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Specify main Screen Datalogger Password Sensor name Converter setup Display setup
Back Select OK
Login Activate/deactivate
Set log interval
0020 sec
Datalogger
Select line
Linie 1 - Sensor name
Specify main Screen Converter setup
Password
Oxix1
Oxygen sat.
%
Select Set up
Main Menu
Converter Overview Menu
Set Clock
Language
Language Set Clock Modbus Factory settings Display SW version
Back Select OK

Display setup

Display Setup
The "Display Setup" menu provides options for language, clock and factory settings, Modbus parameters and software version display.
1. Press the up/down keys to highlight the required menu line (here: Display setup) and then press OK to select it.
2. Press the up/down keys to highlight the required optione and then press OK.
The available options are described in detail on the following pages.
73
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Time Format 24 Hour AM/PM
Back Select OK
Language (GB) English (US) English Dansk Deutsch Español Français Italiano Nederlands Norsk Suomalainen Svenska Back Select OK
Change device addresse
Scanner
Set Clock
Language
Add device
Change device addresse
Set Clock
2006-01-08 10:53:12
Back Select >
Language
The Oxix® is installed with a language package, and (GB) English is the default language.
In the "Language" menu several national languages can be selected.
Additional languages can be added using the MJK-Field Link software pro­gram (see also page 94).
1. Chose the required language using the up/down keys up and then press OK.
74GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Time Format 24 Hour AM/PM
Back Select OK
Change device addresse
Add device Change Device addresse Remove Device Change Display ID No.
Back Select OK
Scanner
Set Clock
Modbus
Add device
Change device addresse
Remove device
Remove device
Set Clock
2006-01-08 10:53:12
Back Select >
Set clock
The "Set clock" menu provides setting of the built-in clock and the time for­mat.
1. Select the date and time digits with the left/right keys and change the set­ting with the up/down keys.
2. Press OK and repeat step 1 for all remaining digits.
3. Press OK to save the time/clock settings.
75
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Cancel Delete Log Restore default Language Restore factory setting Back Select OK
Change device addresse
Device 1
Back Select OK
Add device Change Device addresse Remove Device Change Display ID No.
Back Select OK
Scanner
Back
Modbus
Factory settings
Add device
Devices found
Device 1
Back Select OK
Change device addresse
Change device 1
Back Select OK
Remove device
Device 1
Back Select OK
Remove device
Change Display ID No.
1
Back Select OK
Change display ID no.
Number of Displays
1
Back Select OK
Modbus
The Oxix® is delivered with a standard Modbus RTU protocol, and up to four Oxix® ,Connect® and/or MagFlux® units can be connected for concurrent operation.
When several converters are to be added, you must change each converter's device address to a unique Modbus ID address before connecting the con­verter to the Display Unit.
Add device
Four Oxix® Converters can be connected to one Display Unit using the RS485 interface connection. Note that only one Display Unit can be used with several converters.
The Display Unit can easily be moved to another converters for readout and configuration.
1. Select "Add Device", click OK and wait for the scan process to finish.
2. The found converters are finally displayed.
76GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Change device address
To change each converter's device adress it is required that you connect the Display Unit to each converter one by one to change each converter's device address. If neglected, a conflict between the units will arise when the RS 485
serial loop is established.
Change for example "Device 1" to "Device 2" as follows:
1. Press OK and use the up/down keys to change "1" to "2".
2. Press OK.
The Oxix
®
converter ID is now changed to "2", and more converters can be
added to the Display Unit with the "Add Device" option.
Remove Device
If a device is to be removed from the Display Unit, proceed as follows:
1. Select the required device with the up/down keys.
2. Click OK to remove it.
The device is now removed from the Display unit and display returns to the "Modbus" menu.
Change controller ID (display)
The display controller ID must be changed when several displays operate with the same converter(s)
77
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Cancel Delete Log Restore default Language Restore factory setting Back Select OK
Display 841013-009 SW Aug 15 2008 15:14:20 Language 841513-009
Back
Factory settings
Display SW version
Display 841013-009 SW Aug 15 2008 15:14:20 Language 841513-009
Back
Display SW version
Factory settings
The "Factory settings" menu provides log deletion and resetting to factory settings of the display unit.
1. Press the up/down keys to highlight the required option and press OK.
Important:
The data logger will be reset, and all devices will be deleted from the dis-
play Unit. This is indicated by the MJK start-up screen which is displayed within a few seconds.
Display SW version
The "Display SW version" menu provides a read-out of the display software (actually firmware) version, the build date and last-edited date of the the multi­ple languages file.
78GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
79
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
1/4” tube fitting
1 - 1/2” NPT
112 mm (4.2”)
Measuring element
1/4” tube
Cable to Oxix Converter
130 mm (5”)
57 mm (1.9”)

10. Mechanical Dimensions

Sensor

80GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter

Converter and Display Unit

162
156
144
150
186
89
83
M20
183
85
79
81
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Immersion Fitting
®
Immersion fitting
Oxix order no. 155235
Wall bracket for extension
Extension for universal bracket
Universal bracket
1000 mm
Oxix® sensor
82GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
83
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

11. Maintenance

The Oxix® sensor requires no maintenance as such. It may, however, be cleaned on a regular basis to maintain its maximum accuracy.
In very slimy environments the built-in water or air jet function may be used with advantage.
Clean the transmitter for example once a month to begin with, until an optimal cleaning interval can be determined.

Cleaning the sensor

1. Remove the Oxix® sensor from the liquid.
2. Clean the sensor with a soft, moist cloth and clean water. A mild detergent or dishwashing liquid may be used.
Warning: The accuracy will deteriorate, or be completely ruined at the worst, if the measuring optic element is scratched by for example the use of a steel brush during cleaning.
3. Dry the sensor with a soft cloth.
4. Immerse the Oxix least 15 minutes to stabilize the reading.
5. The cleaning intervals may need to be shortened or extended, all depend­ing on the usage, the liquid composition and the location.
®
sensor into the liquid to be measured and leave it for at

Cleaning the converter display

When seen necessary the Oxix® converter display may be cleaned with a soft, moist cloth.

Functional Tests

The Oxix® transmitter calibration can, if wanted, be checked from the serv­ice menu option "Sensor calibration". This option is read-only. In the unlikely event that the unit fall out of calibration, an MJK serviceperson must be called to fix or re-calibrate the transmitter.

Service Contract

A service contract maintains the equipment reliability and accuracy. Get in contact with MJK’s national sales and service representatives.
84GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
85
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

Appendix A. Pop-up and Error Messages

Alarms
Current overrange Alarm - The mA connection has exceeded
the 4 - 20 mA range. For example 3,5 or 20,5 mA.
Factory reset occurred Alarm - A factory reset has occured. Some
settings have changed. Check all used set­tings.
High alarm Alarm - The high set point has been de-
tected.
Low alarm Alarm - The low set point has been de-
tected.
mA connection Alarm - The mA connection is not connect-
ed. If the current-out is used, deactivate mA Current out to stop this alarm.
Errors
Calibration Error - 1. Perform calibration (see page 21)
and/or check the units and ranges (see page
52).
Eeprom write error Error - Eeprom write error. If error keeps
popping up, contact MJK.
Measuring start failed Error - Start failed. Try again.
Measuring timeout Error - Timeout. Try again.
Missing adjustment Error - The measured value has not been
entered.
Not in solid mode! Error - Calibration can only be performed
when in solid mode. Change mode and try again.
86GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Sensor Error - General sensor error that may also
be transferred to a relay output. Contact MJK.
Sensor communication Error - Modbus communication errrors nor-
mally imply that a device has been moved, added, removed or has changed status in an illegal manner. Rearranged cables or wires can also invoke this error message.
Reset the system to cancel the error condi­tion.
Sensor hardware Error - Change the sensor hardware.
Sensor setting Error - Some settings are wrong or have
been changed.
Sensor supply Error - The sensor supply is too low or too
high. Check the sensor cable connection.
System Error - Internal VCC error. See the event log
for details.
System error Error - The system software is inoperable in-
cluding faulty alarm and sensor bits. Contact MJK.
D.O. offset Error - A zero-point sensor problem. The
sensor needs a zero-point calibration. Contact MJK.
Warnings
Sensor range exceeded Warning - The measured value is out of
range. Check the measuring range and the measuring units.
Info
Measurement hold Info - The measurement is set on hold dur-
ing the cleaning cycle or by an error.
Sensor needs cleaning Info - The sensor needs cleaning.
System reset occurred Info - The converter has been subject to a
software or hardware reset.
87
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
USB
Flash programming please
Programming is in progress but not finished.
wait
Flash programming done Programming is finished.
Flash crc error Cyclic redundancy check error.
Unknown command flash
Unvalid command issued.
user text
Flash address overrun A write error hos occured.
File phase error Internal USB error.
File write failure A write error hos occured.
USB protocol overrun A communication error has occured.
88GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
89
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

Appendix B. MJK-Field Link Software

The MJK-Field Link software package provides several utilities that will be described in detail on the following pages:
PC <-> Oxix•
Saving log data•
Display firmware upgrade•
Converter firmware upgrade•
Language installation•

Important Notes!

1. Current Oxix® log data must be saved in a file before transferring and installing new firmware. See "2. Save Log Data" overleaf.
2. Note down all display and converter settings before upgrading new firmware.
3. The upgrade procedures below must be carried out in strict alphabetical order. First procedure A, then B, C, D and finally E.

A. Connect a PC to the Oxix®

1. Insert the Oxix® MJK-Field Link CD-ROM into the PC's disk drive. The MJK-Field Link opening menu will appear after a short while (see be­low). If the auto-run function does not execute, locate the "MJK-Field-Link 840110-007.exe" file (or a similar file) and click "Run".
®
transmitter interconnection
90GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
2. Unscrew the four screws that hold the Display Unit.
3. Lift out the Display Unit and connect a USB mini A/B cable to the mini USB female B connector on the rear of the front panel.
4. Connect the other end of the USB cable (max. 4.7 metres long) to the PC.
5. Select "File" in the menu bar and click "Mount USB". If the connection is successful, a "Device Clock" window for PC and de­vice time synchronisation is displayed (if PC and device clock were out of sync.) along with a link status display (see below).
If the connection fails, first remove the cable and then re-connect it.
91
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
B. Save Log Data
1. Select "LogData" in the menu bar and click "Save Log to File".
2. Assign a file name, find the destination you want to save the file to, and click "Save". Consequently the file is saved, and the contents is displayed.
C. Upgrade Display Firmware
Important: Always upgrade the display firmware before you upgrade the converter software.
1. Select "Firmware" in the menu bar and click "Display".
2. Locate the "MJK Field FW" directory on the CD-ROM, select the 841xxx­xxx.hex file (or similar 841*.hex) and click "Open".
3. Click OK when the "Display Firmware" dialogue box appears and prompts for programming the display Flash memory.
The display unit firmware is then transferred, and the display unit is conse­quently programmed. Expect this process to take a few minutes.
92GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
4. A "Searching for MJK-Field" dialogue box then appears and prompts for disconnecting and re-connecting the USB cable.
Do as suggested and click OK.
5. Meanwhile the Oxix
®
Display Unit restarts and displays the MJK logo, the new link status is displayed, and the Display Unit informs that the PC is once again connected. Click OK on the PC and the Oxix® Display Unit to return to the Oxix®.
6. Adjust the Oxix
®
Display Unit screen contrast by pressing the two outmost keys simultaneously (indicated by the keys) and press the up/down keys as required.
7. Save the new setting by again pressing the two outmost keys simultane­ously.
93
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
D. Upgrade Converter Firmware
1. Select "Firmware" in the menu bar and click "Converter".
2. Locate the "MJK Field FW" directory on the CD-ROM, select the "842xxx­xxx.hex" file (or similar) and click "Open".
3. Select the required Oxix "Converter Firmware" dialogue box appears and prompts for selecting the converter to program.
The converter firmware is then transferred, and the converter Flash pro-
gramming is initiated. Expect this process to take a few minutes.
4. When the Flash programming is done, click OK on the PC and the Oxix Display Unit to return to the Oxix® display.
5. If required you may adjust the Oxix pressing the two outmost keys simultaneously, press the up/down keys as required, and save the new setting by again pressing the two outmost keys simultaneously.
®
(option 1, 2, 3 or 4) and click OK when the
®
Display Unit screen contrast by
®
94GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Cut-out area
Frontpanel outline
Up
E. Install/Add Languages
All languages are removed during a display firmware upgrade. Consequently the required languages must be re-installed at this stage.
1. Select "File" in the menu bar and click "Languages".
2. A dialogue appears showing the currently installed languages (none follow­ing a firmware upgrade). You can now either add or delete the required languages.
3. Click "Add", select the required language text file (or language pack) in the "MJK Field Display Unit text files" directory on the CD-ROM (for example 841513-008.txt) and click "Open".
4. Select for example "841500 J Europe N+W+S+US" in the "Add Language" dialogue box (N+W+S+US = North, West, South and USA) and click OK.
5. The languages are then transferred and loaded. If required, you can remove one or more languages by selecting them from the list, clicking "Delete" and "Close".
95
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Cut-out area
Frontpanel outline
Up

Appendix C. Front Panel Cut-out Drawing

The dotted line indicates the front panel outline and measures 155 x 145 mm.
96GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
This page intentionally left blank.
97
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009

Appendix D. Log Files

The example log file shown below is the result of a CSV file from the converter (see "Save Log Data" procedure on page 91) having been opened in with Microsoft® Excel® spread sheet utility.
The example below illustrates a Oxix® converter being logged every 10 sec­onds.
The entries are described on the following page.
98GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Index Converter ID (here four converters: 32, 33, 34 and 35)
MB Address Modbus address (here: 1, 2, 3 and 4)
Unit SI unit according to the MJK unit codes (here: 246=NTU)
MJK Unit Codes
Code no. Unit
16 GPM gallons per minute
17 l/min litres per minute
19 m
23 MGD million gallons per day
24 l/sec litres per second
28 m
57 % percent
95 mg/l milligrams per litre
96 kg/l kilograms per litre
97 g/l grams per litre
130 ft
138 l/h litres per hour
139 ppm parts per million
246 NTU Nephelometric Turbidity Units
248 FNU Formazin Nephelometric Units
249 EBC European Brewing Convention
250 FTU Formazine Turbidity Units
3
/h cubic meters per hour
3
/sec cubic meters per second
3
/h cubic feet per hour
Value ID Modbus address for flow (here: 600)
Device Type The device type (here: 9 = SuSix®)
MJK Device Type Codes
Code no. Unit
1 Display
2 MagFlux
3 Connect
4 702 Comtroller
5 Shuttle
6 pH-transmitter
7 Redox
8 Pressure transmitter
9 SuSix
10 Field log
11 Oxix
Frequency Log interval in seconds (here: 10 seconds)
Range Max Maximum value for graph
99
Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Range Min Minimum value for graph
Precision The precision of the SI value
Column J The sensor name as entered by the operator.
Time Date and time in Central European Standard Time (CEST)
UTime The time in UNIX format. Number of seconds since
January 1, 1970.
SI Value The value in SI units (Système International d'Unités)
100GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
To plot the log file's data, proceed as follows:
1. Start Excel and open the csv log file.
2. Sort by "Index", click the "Diagram Guide" tool in the toolbar and select the required curve or graph (see example on the next page).
Loading...