WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot
accept any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are discovered within the warranty period. The warranty does not cover components that are replaced during maintenance work, e.g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper
handling or unauthorized opening of the instrument invalidates any
warranty claim.
To ascertain the warranty liability, return the instrument and proof of
purchase together with the date of purchase freight paid or prepaid.
This compact precision handheld meter enables you to carry out pH /
ORP measurements and concentration measurements using ion selective electrodes rapidly and reliably.
The handheld meter provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven MultiCal
®
calibration procedures and the special AutoRead
function support you in your work with the handheld meter.
6
M
C
AL
S
T
O
R
C
L
AR
pH
9
2
29
4
8
°C
TP
O
3
2
1
1Keypad
2Sample display
3Jack field
Note
If you need further information or application notes, you can obtain the
following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
5
OverviewpH/ION 3400i
1.2SETs of equipment
The measuring instrument is also available as part of individual SETs
of equipment.
You will find additional information on this and other accessories in the
WTW catalog or via the Internet.
6
pH/ION 3400iOverview
Set (sample configuration):
1pH/ION 3400i measuring instrument
2Stand
3Measuring beaker,50 ml
450 ml pH buffer solution, STP 4
550 ml pH buffer solution, STP 7
6Storing solution for pH electrodes
7pH combination electrode
8Plug-in power supply (optional)
7
OverviewpH/ION 3400i
1.3Keypad
STOM
ARRCLCAL
RUN/ENTER
Key functions
M
STO
CAL
RCL
AR
Select the measured variable <M>:
– pH value / ORP voltage
– Ion concentration
Save a measured value
<STO>
Switch measuring instrument on/off
<ON/OFF>
Calibrate the currently set measured variable
<CAL>
Display/transmit measured values
<RCL>
Activate/deactivate the AutoRead function
<AR>
Select the measuring mode, increase values,
scroll <▲ >
Select the measuring mode, decrease values,
scroll <▼ >
RUN/ENTER
Confirm entries, start AutoRead
<RUN/ENTER>
8
pH/ION 3400iOverview
1.4Display
Status display
ISE
UpH
S
8
1
Baud
Time
Day.Month No.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
1.5Jack field
8
8.
Auto
ConCal
8
Store
88
8
°F
°C
CalError
ARng
mol/L
mV/pH
% ppm
mg/l
%
Tem pEr ro r
AR
TP
RCL
Probe symbol
Meas. value display
Function and
temperature display
5
4
f
e
R
123
1pH electrode or ISE (combination) electrode
2Reference electrode
3Plug-in power supply (optional)
4pH temperature sensor
5Serial interface RS 232 / analog output (recorder)
9
OverviewpH/ION 3400i
1.6 Operating structure
The following overview diagram shows which keys you have to press
to select between the different measuring modes:
10
pH/ 3400iSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measuring instrument. Consequently, all responsible personnel must read this
operating manual before working with the measuring system. The operating manual must always be available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in
the laboratory.
We assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take
when handling chemicals.
Safety instructionsThe individual chapters of this operating manual use the following safe-
ty labels to indicate various types of danger:
Warning
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of
the pH and ORP measurement or the measurement of the ion concentration in the field and laboratory.
The technical specifications as given in chapter 7 T
must be observed. Only the operation and running of the measuring
instrument according to the instructions given in this operating manual
is authorized.
Any other use is considered to be unauthorized.
ECHNICALDATA
11
SafetypH/ 3400i
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter
7T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that
are specified in chapter 6 W
HATTODOIF....
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
Obligations of the pur-
chaser
12
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
pH/ 3400iCommissioning
3Commissioning
3.1Scope of delivery
z 3400i handheld meter
z Plug-in power supply (optional)
z Operating manual and short operating manual
z 4 batteries, 1.5 V Mignon type AA (in the instrument)
Setting the date and
time
For details of scope of delivery of SETs, see chapter 1.2 SET
EQUIPMENT and WTW catalog.
3.2Initial commissioning
Perform the following activities:
z Set the date and time
z Connect the plug-in power supply (optional)
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date ap-
pears on the display.
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
SOF
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
12Set the current time with <▲> <▼>.
13
CommissioningpH/ 3400i
13Confirm with <RUN/ENTER>.
The instrument switches to a measuring mode.
14Switch the instrument off using <ON/OFF>.
Connecting the original
plug-in power supply
unit
You can either operate the measuring instrument with batteries or with
the plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (12 VDC). This saves the batteries.
Warning
The line voltage at the operating site must lie within the input voltage
range of the original plug-in power supply (see chapter 7 T
TA).
ECHNICALDA-
Warning
Use original plug-in power supplies only (see
chapter 7 T
ECHNICALDATA).
2
f
e
R
1
3
14
1Plug the jack (1) into the socket (2) of the measuring instru-
ment.
2Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.
pH/ 3400iOperation
4Operation
4.1Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
After this, the measuring instrument automatically switches to
the measuring mode.
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour.
The energy saving feature is not active
z if the power is supplied by the plug-in power supply,
z if the AutoStore function is active,
z if the communication cable and a PC with a running communication
program are connected,
z if the printer cable is connected (for external printers).
15
OperationpH/ 3400i
4.2pH value / ORP voltage
4.2.1General information
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the measuring instrument with the elec-
trode.
4Using <▲ > <▼>, toggle between the pH or mV measuring
modes.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values. Calibrate regularly before measuring.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
Temperature sensorYou can measure with or without a temperature sensor. If a tempera-
ture sensor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the
temperature sensor used. Therefore, you can connect temperature
sensors of the NTC30 or Pt1000 type.
The temperature measurement is absolutely essential for a reproducible pH measurement. If the measurement is made without a temperature sensor, proceed as follows:
16
1Measure the current temperature of the test sample using a
thermometer.
2Enter the temperature on the instrument: While pressing the
<RUN/ENTER> key, enter the temperature value of the test
sample with <▲ > <▼>.
pH/ 3400iOperation
4.2.2Measuring the pH value
1Perform the preparatory activities according to section 4.2.1.
2Immerse the pH electrode in the test sample.
3If required, enter the temperature of the test sample with <▲>
<▼> while keeping the <RUN/ENTER> key depressed.
4Press the <▲> <▼> keys until pH appears in the status display.
The pH value appears on the display.
ISE
U
pH
29
8
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
6
1
Time
Day.MonthNo.
Year
AutoCal DIN
LoBat
AutoCalTEC
Baud
Ident
8
9
2
Auto
ConCal
4
Store
Setting the resolution
AutoRead AR
(drift control)
1Press the <RUN/ENTER> key and hold it down.
2Press the <M> key. The measured values are displayed with a
high resolution, e.g. pH = 4.012.
3Press the <RUN/ENTER> and <M> key once more.
The measured values are displayed with a low resolution, e.g.
pH = 4.01.
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values.
For identical measurement conditions, the following criteria apply:
ReproducibilityResponse time
Better than 0.02> 30 seconds
1Call up the pH measuring mode with <M> and <▲> <▼> .
2Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
17
OperationpH/ 3400i
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
18
pH/ 3400iOperation
4.2.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. SenTix ORP, the measuring
instrument can measure the ORP voltage (mV) of a solution.
1Perform the preparatory activities according to section 4.2.1.
2Submerse the ORP electrode in the sample.
3Press the <▲> <▼> keys until the U status display appears.
The ORP voltage (mV) of the test sample appears on the display.
4Wait for a stable measured value.
ISE
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
9
8
pH
U
2
Auto
ConCal
0
4
Store
5
8
CalError
ARng
mol/LmV/pH
mV
% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
19
OperationpH/ 3400i
4.3pH calibration
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instrument. Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z When the sensor symbol flashes
– (after the calibration interval has expired)
– after a voltage interruption, e. g. after changing the batteries
You can select one of 3 calibration procedures:
AutoCal TECis specially matched to the WTW technical buffer solutions as a fully au-
tomatic two-point calibration. The buffer solutions are automatically
recognized by the measuring instrument. Depending on the instrument
setting (see section 4.8 C
relevant buffer nominal value or the current electrode voltage in mV.
The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. When doing so, the instrument
uses the Nernst slope (-59,16 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
ONFIGURATION), the instrument displays the
AutoCal DINis specially adapted to the permanently programmed buffer solutions in
ConCalThis function is a conventional two-point calibration using two buffer
AutoReadWhen calibrating with AutoCal TEC and AutoCal DIN, the AutoRead
Displaying the calibra-
tion data
Printing the calibration
record
accordance with DIN 19266 as a fully automatic two-point calibration.
The operating sequence of the AutoCal DIN calibration corresponds to
that of the AutoCal TEC calibration. The calibration can only be terminated after the first buffer solution (single point calibration).
solutions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point calibration using any buffer solution, which is used as a high-speed
method.
function is automatically activated.
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
You can view the data of the last calibration on the display. The proceeding is described on page 48.
The calibration protocol contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see page 51).
20
pH/ 3400iOperation
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.7.2
before calibrating. After a valid calibration, the record is printed.
Calibration evaluationAfter the calibration, the measuring instrument automatically evaluates
the calibration. The asymmetry and slope are evaluated separately.
The worst evaluation appears on the display.
DisplayAsymmetry
[mV]
Slope
[mV/pH]
-15 ... + 15-60.5 ... -58
-20 ... +20-58 ... -57
-25 ... +25-61 ... -60.5
or
-57 ... -56
-30 ... +30-62 ... -61
or
-56 ... -50
Clean the electrode according to the
electrode operating manual
E3
Eliminate the error according to chapter 6 W
HATTODOIF...
< -30 or
> 30
... -62 or
... -50
21
OperationpH/ 3400i
Preparatory activities
1Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2Keep the buffer solutions ready.
3Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature sensor (the TP display indicator is missing from
the display).
4.3.1AutoCal TEC
For this procedure, use any two WTW technical buffer solutions (pH
values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the WTW technical buffer
solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead
to an erroneous calibration. The correct buffer solutions are given in the
WTW catalog or in the Internet.
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see
section 4.8 C
ONFIGURATION ), the instrument displays the relevant buff-
er nominal value or the current electrode voltage in mV.
Note
Skip the steps 2 and 7 if you use a pH electrode with a temperature
sensor.
1Press the <CAL> key repeatedly until the Ct1 display indicator
and the function display AutoCal TEC appears. The sensor
symbol displays the evaluation of the last calibration (or no
sensor symbol in the delivery state or after the measurement
parameters have been reset).
22
pH/ 3400iOperation
Lin
8
Tref20
1
8
pH
t
2
Cal
U
pH
2
Auto
ConCal
OU
1
4
AutoCal DIN
AutoCal TEC
3
4
Store
Sal
mV
%
m
mbar
8
Auto
ARng
3
°F
°C
8
CalError
ARng
/pH
mg/l
S/
cm
M
cm
1/
°
%
°
C
F
Store
AR
mol/LmV/pH
mV
% ppm
mg/L
%
TempError
AR
cm
Sal
K
/
RCL
TP
RCL
TP
S
7
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
ISE
0
8
U
pH
2
Auto
ConCal
0
4
Store
0
8
CalError
ARng
mol/LmV
mV/pH% ppm
mg/L
°F
%
°C
TempError
AR
TDS
S
C
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
2If required, enter the temperature of the first buffer solution with
<▲> <▼> while keeping the <RUN/ENTER> key depressed.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ISE
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
TP
RCL
5When the measured value is stable, Ct2 appears.
ISE
U
pH
2C
8
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
23
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. When doing so, the in-
8
t
2
Auto
ConCal
4
Store
OperationpH/ 3400i
strument uses the Nernst slope (-59,16 mV/pH at 25 °C) and determines the asymmetry of the electrode.
6Thoroughly rinse the electrode with deionized water.
7If required, enter the temperature of the second buffer solution
with <▲> <▼> while keeping the <RUN/ENTER> key de-
pressed.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ISE
U
pH
8
45
0
mV/pH
mV
% ppm
°F
°C
CalError
ARng
°F
°C
CalError
ARng
mol/L
mg/L
TP
%
TempError
RCL
AR
mol/L
mV/pH
% ppm
mg/L
%
TempError
AR
1
TP
RCL
S
80
14
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DINAutoCal DIN
LoBat
10When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The probe symbol
shows the evaluation of the current calibration.
ISE
8
2
Auto
ConCal
U
pH
4
4
Store
S
6
1
TimeDay.MonthNo.
Year
AutoCal DIN
LoBat
AutoCal TEC
Baud
Ident
8
9
s
Auto
ConCal
L
Store
ISE
S
Baud
Time
Day.MonthNo.
Year
Ident
LoBat
AutoCalTECAutoCalTEC
0
8
U
pH
2
Auto
ConCal
0
4
Store
8
8
CalError
ARng
mol/L
mV
mV/pH% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
24
11Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
pH/ 3400iOperation
ISE
U
pH
4
y
CalError
ARng
mol/LmV/pH
mV
% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
S
5
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
12Switch to the measuring mode with <M>.
8
9
a
Auto
ConCal
s
Store
25
OperationpH/ 3400i
4.3.2AutoCal DIN
For this procedure, use two different standard buffer solutions according to DIN 19266 (type A, C, D or F with pH values at 25 °C of: 1.679 /
4.006 / 6.865 / 9.180).
Note
Skip the steps 2 and 7 if you use a pH electrode with a temperature
sensor.
1Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).
ISE
U
pH
1C
8
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
8
d
2
Auto
ConCal
4
Store
2If required, enter the temperature of the first buffer solution with
<▲> <▼> while keeping the <RUN/ENTER> key depressed.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ISE
U
pH
8
mV/pH
mV
% ppm
°F
°C
CalError
ARng
mol/L
mg/L
%
TempError
RCL
AR
1
TP
S
77
14
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DINAutoCal DIN
LoBat
8
7
2
Auto
ConCal
4
Store
ISE
S
Baud
Time
Day.MonthNo.
Year
Ident
LoBat
AutoCalTECAutoCalTEC
0
8
U
pH
2
Auto
ConCal
0
4
Store
6
8
CalError
ARng
mol/L
mV
mV/pH% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
26
pH/ 3400iOperation
5When the measured value is stable, Cd2 appears.
ISE
U
pH
2C
8
mV/pH% ppm
°F
°C
CalError
ARng
mol/L
mg/L
%
TempError
RCL
AR
TP
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
8
AutoCal DIN
LoBat
AutoCalTEC
Note
At this point, the AutoCal DIN calibration can be terminated with <M>.
This corresponds to a single-point calibration. When doing so, the in-
strument uses the Nernst slope (-59,16 mV/pH at 25 °C) and determines the asymmetry of the electrode.
d
2
Auto
ConCal
4
Store
6Thoroughly rinse the electrode with deionized water.
7If required, enter the temperature of the second buffer solution
with <▲> <▼> while keeping the <RUN/ENTER> key depressed.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ISE
U
pH
8
mol/L
mV/pH
mV
% ppm
°F
°C
CalError
ARng
mg/L
%
TempError
RCL
AR
TP
S
23
19
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DINAutoCal DIN
LoBat
8
7
2
Auto
ConCal
4
Store
S
1
Baud
Time
Day.MonthNo.
Year
Ident
LoBat
AutoCalTECAutoCalTEC
ISE
1
8
U
pH
2
Auto
8
4
Store
ConCal
0
8
CalError
ARng
mol/L
mV
mV/pH% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
27
OperationpH/ 3400i
10When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The probe symbol
shows the evaluation of the current calibration.
ISE
U
pH
45
0
°F
°C
CalError
ARng
mol/L
mV/pH
% ppm
mg/L
TP
%
TempError
RCL
AR
S
-
1
Baud
TimeDay.MonthNo.
Year
Ident
8
AutoCal DIN
LoBat
AutoCalTEC
11Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
9
s
Auto
ConCal
L
Store
ISE
U
pH
4
y
CalError
ARng
mol/LmV/pH
mV
% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
S
5
6
1
Baud
TimeDay.MonthNo.
Year
Ident
8
AutoCal DIN
LoBat
AutoCalTEC
12Switch to the measuring mode with <M>.
9
a
Auto
ConCal
s
Store
28
pH/ 3400iOperation
4.3.3ConCal
Two-point calibrationFor this procedure, use two buffer solutions:
z pH 7.0 ± 0.5
z any other buffer solution
Note
Skip the steps 2 and 8 if you use a pH electrode with a temperature
sensor.
1Press the <CAL> key repeatedly until the ASY display and the
ConCal function display appears. The sensor symbol displays
the evaluation of the last calibration (or no sensor symbol in the
delivery state or after the measurement parameter has been
reset).
ISE
U
pH
YA
8
mV/pH% ppm
°F
°C
CalError
ARng
mol/L
mg/L
%
TempError
RCL
AR
TP
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
8
S
2
Auto
ConCal
4
Store
2If required, enter the temperature of the first buffer solution with
<▲> <▼> while keeping the <RUN/ENTER> key depressed.
3Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4Press the <RUN/ENTER> key.
The measured pH value appears on the display.
5Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
6When the measured value is stable, press the <RUN/ENTER>
key. The value of the asymmetry appears.
7Press the <RUN/ENTER> key. SLO appears.
Note
At this point, the ConCal calibration can be broken off with <M>. This
corresponds to a single-point calibration. When doing so, the instru-
ment uses the Nernst slope (-59,16 mV/pH at 25 °C) and the fixed
asymmetry of the electrode.
29
OperationpH/ 3400i
ISE
U
pH
OS
8
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
8Thoroughly rinse the electrode with deionized water.
9If required, enter the temperature of the second buffer solution
with <▲> <▼> while keeping the <RUN/ENTER> key de-
pressed.
8
L
2
Auto
ConCal
4
Store
10Immerse the pH electrode in the second buffer solution.
11Press the <RUN/ENTER> key.
12Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
13When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibration.
ISE
U
pH
25
0
°F°C
CalError
ARng
mol/L
mV/pH
% ppm
mg/L
TP
%
TempError
RCL
AR
S
-
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
8
9
s
Auto
ConCal
L
Store
30
14Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
pH/ 3400iOperation
ISE
U
pH
3
y
CalError
ARng
mol/L
mV/pH
mV
% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
S
5
6
1
Baud
Time
Day.MonthNo.
Year
Ident
8
AutoCal DIN
LoBat
AutoCalTEC
15Switch to the measuring mode with <M>.
9
a
Auto
ConCal
s
Store
31
OperationpH/ 3400i
4.4Ion concentration
4.4.1General information
Preparatory activitiesPerform the following preparatory activities when you want to measure
the concentration using an ion selective electrode:
1Connect the ion selective electrode and the reference elec-
trode to the measuring instrument.
2If necessary, connect a temperature sensor (sockets 2 and 4,
see section 1.5 J
3If necessary, press the <M> key repeatedly until the status dis-
play ISE appears.
4Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without temperature sensor.
ACKFIELD on page 9.
Note
Incorrect calibration of ion selective electrodes leads to incorrect measured values. Calibrate regularly before measuring.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
Temperature sensorYou can measure with or without a temperature sensor. If a tempera-
ture sensor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the
temperature sensor used. Therefore, you can connect temperature
sensors of the NTC30 or Pt1000 type.
32
pH/ 3400iOperation
The temperature measurement is absolutely essential for a reproducible ISE measurement. If the measurement is made without a temperature sensor, proceed as follows:
1Measure the current temperature using a thermometer.
2Bring the test sample and calibration standards to the same
temperature (± 2 °C).
Note
For ISE measurements and ISE calibration, it is not possible to enter
temperature values manually.
4.4.2Measuring the ion concentration
To perform ion concentration measurements, proceed as follows:
AutoRead AR
(Drift control)
1Perform the preparatory activities according to section 4.4.1.
2Immerse the ISE electrode and, if necessary, the temperature
sensor in the test sample.
ISE
UpH
71
8
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
%
TempError
RCL
AR
TP
S
0
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable effect on the reproducibility of the measured value.
For identical measurement conditions, the following criteria apply:
ReproducibilityResponse time
8
5
2
Auto
ConCal
0
Store
better ± 0.05 mV> 30 seconds
33
OperationpH/ 3400i
1Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
2Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
3If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
4To terminate the AutoRead function: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
4.4.3Calibration for ion selective measurements
Why calibrate?Ion sensitive electrodes age and are temperature dependent. This
changes the slope. As a result, an inexact measured value is displayed.
Calibration determines the current value of the slope of the electrode
and stores it in the measuring instrument.
Thus, you should calibrate before each measurement (if possible), and
at regular intervals.
Note
For calibration, use tempered standard solutions with a temperature
that differs by maximum
± 2 °C from the temperature of the test sam-
ple.
When to calibrate?z Before each ISE measurement if possible
z After connecting another ion selective electrode or reference elec-
trode
z When the sensor symbol flashes after a voltage interruption, e.g.
battery change
ConCalThis is the conventional two-point or three-point calibration procedure
that uses 2 or 3 freely selectable standard solutions. The concentration
expected for the measurement determines the concentration of the calibration standards.
34
AutoReadThe calibration procedure automatically activates the AutoRead func-
tion. To accept the current measured value without waiting for the AutoRead function to finish, just press <RUN/ENTER>.
pH/ 3400iOperation
TP displayIf you use a temperature sensor, a flashing TP display indicates that the
difference in temperature of the standard solutions is too high (more
than 2 °C).
Displaying the
calibration data
Printing the
calibration protocol
You can view the data of the last calibration on the display. The proceeding is described on page 48.
The calibration protocol contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see page 51).
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.7.2
before calibrating. After a valid calibration, the record is printed.
> Std 1, Std 3 must be > Std 2
< Std 1, Std 3 must be < Std 2
With a two-point calibration, the calculated slope is used for the entire
measuring range of the electrode. With a three-point calibration, the entire measuring range is divided into measuring range I and II. Different
calibration values apply for the two measuring ranges. The measured
value of standard solution 2 separates the two measuring ranges from
each other.
Note
Among other things, the measuring accuracy depends on the selected
standards solutions. Therefore, the standard solutions selected should
cover the expected value range of the following concentration measurement.
36
pH/ 3400iOperation
Two-point calibration
1Press the <CAL> key. The previously set up concentration of
the first standard solution and ConCal appear on the display.
ISE
UpH
00
1
mol/LmV/pH% ppm
°F°C
CalError
ARng
mg/l
%
TempError
AR
TP
RCL
S
1
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
2Enter the current concentration of the first standard solution
with <▲> <▼>.
3Immerse the ISE electrode and, if necessary, the temperature
sensor in the first standard solution.
4Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.
S
0
t
Auto
ConCal
d
Store
ISE
UpH
5
4
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
°C
TempError
AR
/pH
%
TP
RCL
S
0
1
1
Baud
TimeDay.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
5As soon as a stable value is achieved, AR stops flashing. The
previously set up concentration of the second standard solution
appears on the display.
2
0
2
Auto
ConCal
3
Store
37
OperationpH/ 3400i
ISE
UpH
00
2
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/l
TP
%
TempError
RCL
AR
S
2
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
6Enter the current concentration of the second standard solution
with <▲> <▼>.
7Thoroughly rinse the ion selective electrode and, if necessary,
the temperature sensor with deionized water.
8Immerse the ion selective electrode and, if necessary, the tem-
perature sensor in the second standard solution.
S
0
t
Auto
ConCal
d
Store
9Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.
ISE
UpH
4
4
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
°C
AR
/pH
TP
%
TempError
RCL
S
0
1
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
10As soon as a stable value is reached, the AR display indicator
stops flashing. The slope (mV) of the electrode after the twopoint calibration appears on the display.
The probe symbol shows the evaluation of the two-point calibration for the measuring range I.
2
8
2
Auto
ConCal
3
Store
38
pH/ 3400iOperation
ISE
U
pH
4
4
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
%
°C
TempError
AR
/pH
TP
RCL
S
2
1
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
11To return to the measuring mode: Press the <M> key or go on
to the three-point calibration by pressing the <RUN/ENTER>
key.
2
6
2
Auto
ConCal
3
Store
Three-point
calibration
12The previously set up concentration of the third standard solu-
tion appears on the display.
ISE
UpH
00
3
°F°C
CalError
ARng
mol/LmV/pH% ppm
mg/l
TP
%
TempError
RCL
AR
S
5
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
13Enter the current concentration of the third standard solution
with <▲> <▼>.
14Thoroughly rinse the ion selective electrode and, if necessary,
the temperature sensor with deionized water.
15Immerse the ion selective electrode and, if necessary, the tem-
perature sensor in the third standard solution.
16Press the <RUN/ENTER> key.
The electrode voltage appears on the display, the AR display
indicator flashes.
S
0
t
Auto
ConCal
d
Store
39
OperationpH/ 3400i
ISE
U
pH
2
6
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
°C
TempError
AR
/pH
%
TP
RCL
S
2
1
1
Baud
TimeDay.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
17As soon as a stable value is reached, the AR display indicator
stops flashing. The slope (mV) of the electrode for the measuring range II is displayed.
The probe symbol shows the evaluation of the three-point calibration for the measuring range II.
2
1
2
Auto
ConCal
2
Store
ISE
U
pH
1
4
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
°C
TempError
AR
/pH
%
TP
RCL
S
3
1
1
Baud
TimeDay.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
18To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing
<M>. The values of the two-point calibration will then remain stored.
2
2
2
Auto
ConCal
3
Store
40
pH/ 3400iOperation
4.5Calibration interval (Int 3)
The flashing probe symbol reminds you to calibrate regularly (pH/
ORP). When the fixed calibration interval (Int 3) has expired, the probe
symbol flashes. It is still possible to measure. The calibration interval is
set to 7 days in the factory.
Note
To ensure the high measuring accuracy of the measuring system, cal-
ibrate after the calibration interval has expired.
Setting the calibration
interval
You can change the interval set in the factory (1 ... 999 days):
1Switch off the measuring instrument.
2Press the <M> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configuration.
4Press the <RUN/ENTER> key repeatedly until Int 3 appears.
Example:
ISE
U
pH
3n
7
mol/LmV/pH% ppm
°F
°C
CalError
ARng
mg/L
%
TempError
RCL
AR
TP
S
I
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
d
t
2
Auto
ConCal
d
Store
5Set the required interval (in days) until the next calibration with
<▲> <▼>.
6Confirm with <RUN/ENTER>.
7Switch to the measuring mode with <M>.
41
OperationpH/ 3400i
4.6Saving
The measuring instrument has an internal data memory. It can store up
to 500 data records.
A complete data record consists of:
z Number of the storage location
z Date/time
z Measured value
z Temperature
z Temperature measuring procedure
z ID number
You can transmit measured values (data records) to the data storage
in two ways:
z Save manually
z Switch on AutoStore (Int 1), (see page 44).
Note
If you connect a recorder (analog output), the output on the digital interface is switched off.
4.6.1Saving manually
You can transmit a measured value to the data storage as follows:
1Press the <STO> key.
The current number (location number No.) of the next free storage location appears under the current measured value on the
display.
ISE
U
pH
29
4
°F°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
1
Time
Day.Month
Year
LoBat
6
Baud
No.
Ident
8
AutoCal DIN
AutoCalTEC
9
2
Auto
ConCal
4
Store
42
2Confirm with <RUN/ENTER>.
The display switches to entering the ID number.
pH/ 3400iOperation
ISE
U
pH
29
2
°F°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
6
1
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
3Using <▲> <▼>, set the required ID number
(1 ... 999).
4Confirm with <RUN/ENTER>.
The measured values are stored. The instrument changes to
the measuring mode.
8
9
2
Auto
ConCal
4
Store
Message StoFull
StoFullThis message appears when all of the 500 storage locations are occu-
StoFullStoFull
pied.
You have the following options:
Saving the current measured value.
The oldest measured value (storage location 1)
will be overwritten by this
Returning to the measuring mode without savingpress any key
Outputting the data storagesee
Clearing the memorysee
Press
<RUN/ENTER>
section 4.6.3
section 4.6.4
43
OperationpH/ 3400i
4.6.2Switching on AutoStore (Int 1)
The save interval (Int 1) determines the chronological interval between
automatic save processes. After the fixed interval has expired, the current data record is transmitted to the internal storage and to the interface.
Setting the
save interval
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):
1Press the <RUN/ENTER> key and hold it down.
2Press the <STO> key. Int 1 appears on the display.
ISE
UpH
1n
f
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
I
1
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
3Set the required interval between the saving procedures with
The number of free memory locations appears on the display.
ISE
U
pH
94
E
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
I
1
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
5Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
F
9
r
Auto
ConCal
E
Store
pH/ 3400iOperation
ISE
U
pH
9
1
CalError
ARng
mol/LmV/pH% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
S
4
I
1
Baud
Time
No.Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
6Set the required ID number with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode and starts the measuring and saving procedure. AutoStore flashes on the display.
9
r
Auto
ConCal
EF
Store
As soon as all of the 500 storage locations are occupied, AutoStore is
terminated (Int 1 = OFF). If there are not enough storage locations
available for your measurements:
z Output and backup the data storage (see page 46) and
z Clear the memory (see page 52).
Note
The AutoStore function is interrupted if you start other functions, e.g.
output the data storage. After the function is finished, the AutoStore
function is continued. By this, however, temporal gaps in the recording
of the measured values will occur.
Switching off AutoStoreSwitch AutoStore off by:
z setting the save interval (Int 1) to OFF, or
z switching the measuring instrument off and then on again.
45
OperationpH/ 3400i
4.6.3Outputting the data storage
You can output the contents of the data storage:
z Stored data on the display
z Calibration data on the display
z Stored data on the serial interface
z Calibration protocol to the interface
Outputting stored data
on the display
1Press the <RCL> key repeatedly until StO dISP appears on the
display.
TDS
S
T
S
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
2Press the <RUN/ENTER> key.
A measured value appears on the display.
The storage location of the data record is displayed for approx.
2 s, then the respective temperature appears.
ISE
Lin
d
Tref20
Cal
I
AutoCal DINAutoCalTEC
U
pH
S
6
1
Baud
Time
Day.MonthNo.
Year
LoBat
Ident
AutoCal DIN
AutoCalTEC
8
2
Auto
ConCal
OUpH
O
s
6
4
Store
Sal
mV/pH
%
m
mbar
1/
°
F
p
Auto
ARng
68
°F
°C
8
CalError
ARng
mg/l
S/
cm
cm
M
Sal
cm
°
K
%
/
C
Store
AR
RCL
mol/LmV/pH% ppm
mg/L
%
TempError
AR
TP
TP
RCL
46
You can perform the following activities:
Display further data of the data record
(ID number, date, time, storage location)
Advance one data record
(storage location)
Go back one data record (storage location)Press <▼ >
Press <RUN/ENTER>
Press <▲>
pH/ 3400iOperation
Note
If you want to search for a certain element of the data record (e.g. date),
proceed as follows:
1Using <RUN/ENTER>, select the element (e.g. date).
2Press <▲> or <▼> repeatedly until the required element ap-
pears on the display.
After approx. 2 s the temperature of the displayed measured
value appears.
Outputting stored data
to the interface
1Press the <RCL> key repeatedly until Sto SEr appears on the
display.
TDS
S
t
S
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
2Press the <RUN/ENTER> key.
The complete contents of the storage are transmitted to the interface. During the data transmission the instrument increments the storage numbers. After the data transmission, the
instrument automatically switches to the last active measurement mode.
Note
You can cancel the transmission with <M> or <RUN/ENTER>.
Lin
Tref20
Cal
S
AutoCal DINAutoCalTEC
OUpH
o
E
Sal
mV/pH
%
m
mbar
1/
°
F
R
Auto
ARng
mg/l
S/
cm
M
cm
°
%
C
Store
AR
cm
Sal
K
/
RCL
TP
The transmitted data contains the entire contents of the storage in incrementing order of the location numbers.
47
OperationpH/ 3400i
Sample printout:
No. 1:
09.03.99 17:10
pH 10.013 25 °C
Tman AR
Ident : 1
No. 2:
09.03.99 17:11
pH 10.104 24.7 °C
Tauto AR
Ident : 1
No. 3:
09.03.99 17:12
305.2 mV
Tauto
Ident : 13
...
Outputting the
calibration data
on the display
1Press the <RCL> key repeatedly until CAL disp appears on the
display.
ISE
UpH
S
mol/LmV/pH% ppm
mg/L
0
1
Baud
Time
No.
Day.Month
Year
Ident
d
AutoCal DIN
LoBat
AutoCalTEC
2Press the <RUN/ENTER> key.
The data of the last calibration of all measured variables appears in the following sequence:
z pH: Slope SLO and asymmetry ASY
z ISE: Slope of the electrode for calibration range I (SLO1)
Information concerning the calibration procedure is output as
well.
Auto
ConCal
and calibration range II (SLO2)
A
s
Store
LC
pi
°F
°C
CalError
ARng
%
TempError
AR
TP
RCL
48
pH/ 3400iOperation
ISE
U
pH
45
0
CalError
ARng
4
y
CalError
ARng
mol/L
mV/pH
% ppm
mg/L
°F
%
°C
TempError
AR
mol/LmV/pH
mV
% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
TP
RCL
S
-
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
3Press <RUN/ENTER> to display the value of the asymmetry
(mV).
ISE
8
s
Auto
ConCal
U
pH
9
L
Store
S
5
6
1
Baud
Time
Day.MonthNo.
Year
Ident
AutoCal DIN
LoBat
AutoCal TEC
8
9
a
Auto
ConCal
s
Store
4Press <RUN/ENTER> to display the slope of the ion selective
electrode in the calibration range I.
ISE
U
pH
4
1
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
%
°C
TempError
AR
/pH
TP
RCL
S
1
2
s
3
L
Auto
ConCal
O
Store
1
Baud
Time
Day.Month
No.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
5Press <RUN/ENTER> to display the slope of the ion selective
electrode in the calibration range II.
49
OperationpH/ 3400i
ISE
U
pH
4
2
CalError
ARng
mol/L
mV
% ppm
mg/L
°F
°C
TempError
AR
/pH
%
TP
RCL
S
1
2
s
3
L
Auto
ConCal
O
Store
1
Baud
Time
Day.Month
No.
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
6With <M> you can switch back to the last active measuring
mode.
50
pH/ 3400iOperation
Outputting the calibra-
tion protocol on the
interface
1Press the <RCL> key repeatedly until CAL SEr appears on the
display.
ISE
UpH
S
mol/LmV/pH% ppm
mg/L
0
1
Baud
Time
No.
Day.Month
Year
Ident
d
AutoCal DIN
LoBat
AutoCalTEC
2Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmitted to the interface. After the data transmission, the instrument
automatically switches to the last active measurement mode.
With this function, you can delete stored data records. 500 storage locations will then be available again.
Note
The Clear memory function only appears when there are data records
stored in the memory. Otherwise, the measuring instrument automatically switches to the last active measuring mode.
Proceed as follows to clear all data records:
1Switch off the measuring instrument.
2Press the <STO> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display.
ISE
UpH
oS
rc
°F°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
0
1
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
4Confirm the clearing process with <RUN/ENTER>.
Pressing any other key prevents the clearing, the data records
will remain stored.
Note
The calibration data remain stored and can be called up.
d
t
l
Auto
ConCal
Store
52
pH/ 3400iOperation
4.7Transmitting data
You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 44), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured
values are periodically output on the interface (see below).
z With the Output data storage function (page 46), calibration data or
saved measured values are output on the interface.
4.7.1Data transmission interval (Int 2)
The interval to the data transmission (Int 2) determines the chronological interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
Setting the
Data transmission
interval
Note
The setting of the interval (Int 2) is only effective if the save interval
(AutoStore function) is switched off!
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min, 5
min, 10 min, 15 min, 30 min, 60 min):
1Press the <RUN/ENTER> key and hold it down.
2Press the <RCL> key. Int 2 appears on the display.
ISE
UpH
2n
fO
°F°C
CalError
ARng
mol/L
mV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
1
Time
Day.Month
Year
LoBat
I
Baud
No.
Ident
8
AutoCal DIN
AutoCalTEC
t
f
Auto
ConCal
Store
3Set the required interval between the saving procedures with
<▲> <▼>.
4Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode.
53
OperationpH/ 3400i
S
Note
When the AutoStore function is active at the same time, the data transmission is performed according to the setting of the save interval (Int 1).
Set the save interval (Int 1) to OFF to activate the Data transmission in-terval (Int 2).
4.7.2PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an external printer. Use the AK340/B (PC) or AK325/S (ext. printer) cable to
connect the interface to the instruments. The data output automatically
switches to the RS232 interface.
Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
Socket assignment
Set up the following transmission data on the PC/printer:
The measuring instrument can be remotely controlled from a PC. This
requires the KOM pilot communication kit. It is available as an accessory. The instrument is then controlled via commands that simulate
keystrokes and request the current display contents.
Note
A more detailed description is provided within the scope of delivery of
the communication kit.
55
OperationpH/ 3400i
4.8Configuration
You can adapt the measuring instrument to your individual requirements. To do this, the following parameters can be changed (the status
on delivery is marked in bold):
Baud rate1200, 2400, 4800, 9600
Display during the
pH calibration
Intervals
Calibration (Int 3)
in days (d)
Date/timeAny
Note
You can leave the configuration menu at any time with <M>. The parameters that have already been changed are stored.
1Switch off the measuring instrument.
2Press the <M> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
Buffer nominal value, or current
electrode voltage
– pH: 1 ... 7 ... 999 dInt 3
– ISE: no interval
56
Baudrate
TDS
S
t
n
1
Baud
TimeDay.MonthNo.
Year
Ident
Tref25
LoBat
nLF
4Select the required Baud rate with <▲> <▼>.
5Confirm with <RUN/ENTER>. CAL dISP appears on the dis-
play.
Lin
4
Tref20
Cal
8
AutoCal DIN
AutoCalTEC
OUpH
2
0
Sal
mV/pH
%
m
mbar
1/
°
F
0
Auto
ARng
mg/l
S/
cm
M
cm
°
%
C
Store
AR
cm
Sal
K
/
RCL
TP
pH/ 3400iOperation
ISE
UpH
LC
pi
°F°C
CalError
mol/L
mV/pH
/pH
% ppm
mg/L
TP
%
TempError
RCLARng
AR
S
0
1
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
6Select the required display during the pH calibration with
<▲> <▼>.
mV: Display of the current electrode voltage
/pH: Display of the buffer nominal value.
7Confirm with <RUN/ENTER>. On the display, Int 3 appears.
A
sd
Auto
ConCal
Store
Calibration
intervals
Date and time
ISE
U
pH
3n
8
CalError
ARng
4
4
CalError
ARng
mol/LmV/pH% ppm
mg/L
°F
%
°C
TempError
AR
mol/L
/pH
mV% ppm
mg/L
°F
%
°C
TempError
AR
TP
RCL
TP
RCL
S
I
1
Baud
Time
Day.MonthNo.
Year
Ident
d
AutoCal DIN
LoBat
AutoCal TEC
8Set the required interval (in days) with <▲> <▼>.
9Confirm with <RUN/ENTER>. The date flashes on the display.
ISE
U
2
Auto
pH
t
d
Store
ConCal
S
1
2
0
3
9
Auto
ConCal
0
Store
1
Time
Year
LoBat
Baud
No.Day.Month
Ident
AutoCal DIN
AutoCalTEC
10Set the date of the current day with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The date (month) flashes on the display.
12Set the current month with <▲> <▼>.
57
OperationpH/ 3400i
13Confirm with <RUN/ENTER>.
The year appears on the display.
14Set the current year with <▲> <▼>.
15Confirm with <RUN/ENTER>.
The hours flash on the display.
16Set the current time with <▲> <▼>.
17Confirm with <RUN/ENTER>.
The minutes flash on the display.
18Set the current time with <▲> <▼>.
19Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode.
58
pH/ 3400iOperation
4.9Reset
You can reset (initialize) the measurement parameters and the configuration parameters separately from one another.
Measurement
parameters
pH
Configuration
parameters
The following measuring parameters are reset to the delivery status:
Measuring modepH
Slope -59.16 mV/pH
Asymmetry0 mV
Manual temperature input25 °C
Resolution of pH display0.001 (high resolution)
Note
The calibration data gets lost when the measuring parameters are reset. Recalibrate after performing a reset.
The following configuration parameters (InI) are reset to the delivery
status:
Baud rate4800
Display during the
Buffer nominal value
pH calibration
Interval 1
(automatic save)OFF
Interval 2
(for data transmission)OFF
59
OperationpH/ 3400i
Resetting the
measuring parameters
1Press the <RUN/ENTER> key and hold it down.
2Press the <CAL> key. The setting to reset the pH measuring
parameters appears on the display.
ISE
U
pH
1I
o
°F
°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
S
I
1
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
3Using <▲> <▼>, toggle between no and YES.
YES: Resetting the pH measuring parameters
no: Retaining settings.
4Confirm with <RUN/ENTER>.
The measuring instrument switches to the configuration parameters.
8
n
O
Auto
ConCal
n
Store
60
pH/ 3400iOperation
Resetting the
configuration parame-
ters
1
ISE
S
I
Baud
Time
No.
Day.Month
Year
Ident
AutoCal DIN
LoBat
AutoCalTEC
5Using <▲> <▼>, toggle between no and YES.
YES: Resetting the configuration parameters
no: Retaining settings.
6Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the last
active measuring mode.
8
pH
U
O
Auto
ConCal
n
n
Store
1I
o
°F°C
CalError
ARng
mol/LmV/pH% ppm
mg/L
TP
%
TempError
RCL
AR
61
OperationpH/ 3400i
62
pH/ 3400iMaintenance, cleaning, disposal
5Maintenance, cleaning, disposal
5.1Maintenance
The measuring instrument is almost maintenance-free.
The only maintenance task is replacing the batteries.
LoBat on the display indicates that the batteries should be changed.
The batteries are then largely depleted.
Replacing the batteries
2
1
1Open the housing after the instrument has been switched off:
– Undo the four screws on the underside of the instrument
– Pull down the lower cover (1).
2If necessary, take the four depleted batteries (2) out of the bat-
tery compartment.
3Place four new batteries (type Mignon AA) in the battery com-
partment.
4Close the lower cover (1).
Warning
Make sure that the poles of the batteries are the right way round.
The
±signs on the batteries must correspond to the ± signs in the bat-
tery compartment. Only use leakproof alkaline manganese batteries.
63
Maintenance, cleaning, disposalpH/ 3400i
Note
For maintenance of electrodes and ion-selective electrodes, follow the
corresponding operating manual.
5.2Cleaning
Occasionally wipe the outside of the measuring instrument with a
damp, lint-free cloth. Disinfect the housing with isopropanol as required.
Warning
The housing is made of a synthetic material (ABS). Thus, avoid contact
with acetone and similar detergents that contain solvents. Remove any
splashes immediately.
5.3Disposal
PackingThis measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the instrument against damage during transport.
BatteriesThis note refers to the battery regulation that applies in the Federal Re-
public of Germany. We would ask end-consumers in other countries to
follow their local statutory provisions.
Note
This instrument contains batteries. Batteries that have been removed
must only be disposed of at the recycling facility set up for this purpose
or via the retail outlet.
It is illegal to dispose of them in household refuse.
Measuring instrumentDispose of the measuring instrument as electronic waste at an appro-
priate collection point. It is illegal to dispose of the instrument in household refuse.
64
pH/ 3400iWhat to do if...
6What to do if...
6.1pH system messages
Error message 0FL
0FL
0FL0FL
CauseRemedy
pH electrode:
– Not connected– Connect electrode
Error message E3
E3
E3E3
– Air bubble in front of the
– Remove air bubble
diaphragm
– Air in the diaphragm– Extract air or
moisten diaphragm
– Cable broken– Replace electrode
– Gel electrolyte dried out– Replace electrode
CauseRemedy
Electrode
– Diaphragm contaminated– Clean diaphragm
– Membrane contaminated– Clean membrane
– Moisture in the plug– Dry plug
– Electrolyte out of date– Replenish electrolyte or
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low– Use suitable electrode
– Temperature too high– Use suitable electrode
– Organic liquids– Use suitable electrode
CauseRemedy
pH electrode:
– pH electrode unsuitable– Use suitable electrode
– Temperature difference
between buffer and test
– Adjust temperature of buffers
or test samples
sample too high
66
– Measurement procedure not
suitable
– Follow special procedure
pH/ 3400iWhat to do if...
6.2ISE system messages
Error message 0FL
0FL
0FL0FL
CauseRemedy
Ion-selective electrode:
– Not connected– Connect the electrode
– Cable broken– Replace the electrode
Error message E3
E3
E3E3
CauseRemedy
Ion-selective electrode:
Concentration
display flashes
– Slope not in the range
50 ... 70 mV or
– Replace the electrode
– recalibrate
23 ... 35 mV
– Moisture in the plug– Dry plug
– Electrode obsolete– Replace the electrode
– Electrode broken– Replace the electrode
Measuring instrument:
– Calibration procedures:
– Select the correct order
Incorrect order of the
standards with 3-point
calibration
– Socket damp– Dry socket
CauseRemedy
In measurement:
– The measured value is outside
the range determined through
calibration
– Select the calibration
standards so that the
measured value is in between
them
In calibration:
U2U1–5mV≤
U3U2–5mV≤
– Calibration standards lie too
closely together. Select other
calibration standards
67
What to do if...pH/ 3400i
TP and Temp Error flash
Calibration Std 2 flashes
Calibration Std 3 flashes
CauseRemedy
In calibration:
– Calibration standards do not
have the correct temperature
(max. temperature difference
– Temper the calibration
standards
±
2 °C)
In measurement:
– Test sample does not have the
– Temper the test sample
correct temperature (max.
temperature difference
± 2 °C)
CauseRemedy
– Standard 2 = Standard 1– Use different calibration
standards
CauseRemedy
Display LoBat
Instrument does not
react
to keystroke
Requirement not met:
– Use different standard 3
– Standard 3 > Standard 2, if Std
2 > Std 1 or
– Standard 3 < Standard 2, if Std
2 < Std 1
6.3General errors
CauseRemedy
– Batteries almost empty– Exchange the batteries (see
section 5.1 M
CauseRemedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Switch the instrument on
while pressing the <CAL> key
AINTENANCE)
68
pH/ 3400iWhat to do if...
Display to
to
toto
CauseRemedy
– Time-out of the interface– Check the instrument that is
connected
Probe symbol flashes
CauseRemedy
– Calibration interval expired– Recalibrate the measuring
system
Message StoFull
StoFull
StoFullStoFull
CauseRemedy
– All memory locations are full– Output data storage and clear
data storage
69
What to do if...pH/ 3400i
70
pH/ 3400iTechnical data
7Technical data
Dimensions and weight
Length [mm]172
Width [mm]80
Height [mm]37
Weight [kg]Approx. 0.3
Mechanical structure
Electrical safety
Ambient
conditions
pH/ORP measuring
ranges
Type of protectionIP 66
Protective classIII
Storage- 25 °C ... + 65 °C
Operation-10 °C ... + 55 °C
Climatic class2
Measuring rangeResolution
pH- 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV]- 999.9 ... + 999.9
- 1999 ... + 1999
0.001
0.01
0.1
1
T [°C]- 5.0 ... + 105.00.1
Precision of pH/ORP
(± 1 digit)
pH (± 2 pH units from
± 0.003 at + 15 °C ... + 35 °C
the calibration point)
U [mV]± 0.2 at + 15 °C ... + 35 °C
± 0.1
T [°C]NTC 30:
± 0.1
PT 1000:
± 0.5 at 0 °C ... 15 °C
± 0.1 at 15 °C ... 35 °C
± 1 at 35 °C ... 55 °C
71
Technical datapH/ 3400i
pH temperature input:
ISE measuring ranges
- 20 °C ... + 130 °C
Measuring range 1
Resolution
Measuring range 2
Resolution
Measuring range 3
Resolution
Measuring range 4
Resolution
Usable
ISE electrodes
ISE calibration
procedure
Combination and double electrodes with connector according to DIN
19262
Two point calibration or three point calibration with standards suiting
the sample
ISE slope ranges± 25 to 35 mV and ±50 to 70 mV
0.000 ... 9.999 mg/l
0.001 mg/l
0.00 ... 99.9 mg/l
0.01 mg/l
0.0 ... 999.9 mg/l
0.1 mg/l
0 ... 1999 mg/l
1mg/l
Analog outputAutomatic switchover when the recorder is connected by the cable,
AK 323.
Output signalsCorresponds to the electrode voltage,
i.e. 50-62 mV/pH unit
Accuracy± 0.1 % of the measured value ± 4 mV
Internal resistance< 5 Ohm (current limited to max. 0.2 mA
output current)
72
pH/ 3400iTechnical data
Serial
interface
Power supply
Automatic switchover when a PC or a printer is connected via the cable,
AK 340/B or AK 325/S.
TypeRS232, data output
Baud rateSelectable 1200, 2400, 4800, 9600 baud
Data bits8
Stop bit2
ParityNone
HandshakeRTS/CTS + Xon/Xoff
Cable length Max. 15m
Batteries4 x 1.5 V alkali-manganese batteries, type
AA
Operational lifeApprox. 3000 operating hours
Mains (optional)The following specifications apply to all
plug-in power supplies:
Max. overvoltage connection category II
Plug-in power supply unit
(Euro, US , UK, Australian plug)
FRIWO FW7555M/09, 15.1432
Friwo Part. No. 1822089
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1,5 A
Plug-in power supply with Euro plug:
FRIWO FW1199, 11.7864
Friwo Part. No. 1762613
Input: 230 V ~ / 50 Hz / 5.6 VA
Output: 12 V = / 130 mA / 1.56 VA
Plug-in power supply with US plug:
FRIWO FW1199, 11.7880
Friwo Part. No. 1794043
Input: 120 V ~ / 60 Hz / 6 VA
Output: 12 V = / 150 mA
Plug-in power supply with UK plug:
FRIWO FW1199, 11.7872
Friwo Part No. 1816491
Input: 230V ~ / 50 Hz / 5.6 VA
Output: 12 V = / 130 mA / 1.56 VA
73
Technical datapH/ 3400i
Guidelines
and norms used
EMCE.C. guideline 89/336/EEC
EN 61326-1:1997
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A
Instrument safetyE.C. guideline 73/23/EEC
EN 61010-1 A2:1995
Climatic classVDI/VDE 3540
Type of protectionEN 60529:1991
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
74
Test certificates
cETLus, CE
pH/ 3400iLists
8Lists
This chapter provides additional information and orientation aids.
AbbreviationsThe list of abbreviations explains the indicators and the abbreviations
that appear on the display and in the manual.
Specialist termsThe glossary briefly explains the meaning of the specialist terms. How-
ever, terms that should already be familiar to the target group are not
described here.
IndexThe index will help you to find the topics that you are looking for.
75
ListspH/ 3400i
Abbreviations
ARAutoRead (drift control)
ARngAutomatic range switching
Measuring instrument measures with highest resolution
ASYAsymmetry
AutoCal DINAutomatic pH calibration with buffer solutions pre-
pared according to DIN 19 266
AutoCal TECAutomatic pH calibration with WTW technical buff-
er solutions according to DIN 19267
°CTemperature unit, degrees Celsius
CalCalibration
Cd...Display indicator during calibration for pH mea-
surements. Indicates the selection of the buffer
data record for buffer solutions prepared according
to DIN 19 266
Cm...Display indicator during calibration for pH mea-
surements. Indicates the selection of buffer data
records for buffer solutions of the Merck company
ConCalConventional single-point or two-point calibration
for pH measurements
Ct...Display indicator during calibration for pH mea-
surements. Indicates the selection of the buffer
data records for WTW technical buffer solutions
E3Error message
see chapter 6 W
HATTODOIF...
InIInitialization
Resets individual basic functions to the status they
had on delivery
ISEIon sensitive electrode
LoBatBatteries almost empty(Low Battery)
mVVoltage unit
mV/pHUnit of the electrode slope (internat. mV)
76
OFLDisplay range exceeded (Overflow)
pHpH value
pH/ 3400iLists
SSlope (internat. k)
SELVSafety Extra Low Voltage
SLOSlope setting on calibration
TPTemperature measurement active (Temperature
Probe)
U
ASY
Asymmetry
77
ListspH/ 3400i
Glossary
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains within the tolerance.
AsymmetryDesignation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is
immersed in a solution with the pH of the nominal electrode zero point
(WTW electrodes: pH = 7).
AutoRangeName of the automatic selection of the measuring range.
AutoReadWTW name for a function to check the stability of the measured value.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
DiaphragmThe junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between
two solutions and makes electrolyte exchange more difficult. The expression, junction, is also used for ground or junction-less transitions.
Electrode zero pointThe zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero.
Normally, this is at 25 °C.
Electromotive force of
an electrode
The electromotive force U of the electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the
galvanic voltages of the electrode. Its dependency on the pH results
in the electrode function which is characterized by the parameters,
slope and zero point.
Measured parameterThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured valueThe measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring systemThe measuring system comprises all the devices used for measuring,
e. g. measuring instrument and probe. In addition, there is the cable
and possibly an amplifier, terminal strip and armature.
78
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
MultiCal
®
WTW name stating that a measuring instrument provides several calibration procedures.
pH/ 3400iLists
Offset potentialThe measurable potential of a symmetrical electrode, the membrane
of which is immersed in a solution with the pH of the nominal electrode
zero point. The asymmetry is part of the offset potential.
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface
(e. g. a gold or platinum surface).
pH valueThe pH is a measure of the acidic or basic effect of an aqueous solu-
tion. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH
value is the value of a pH measurement.
PotentiometryName of a measuring technique. The signal (depending on the mea-
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
Reference
temperature
Fixed temperature value to compare temperature-dependent measured values. For conductivity measurements, the measured value is
converted to a conductivity value at a reference temperature of 20 °C
or 25 °C.
ResetRestoring the original condition of all settings of a measuring system.
ResolutionSmallest difference between two measured values that can be dis-
played by a measuring instrument.
SlopeThe slope of a linear calibration function.
Standard solutionThe standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Temperature
Value of the slope of a linear temperature function.
coefficient
Temperature functionName of a mathematical function expressing the temperature behav-
ior of a test sample, a probe or part of a probe.
Test sampleDesignation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
79
ListspH/ 3400i
80
pH/ION 3400iIndex
A
Asymmetry 20
Auslieferzustand 59
Authorized use 11
AutoCal DIN 20, 26
AutoCal TEC 20, 22
AutoRead 17
Remote control 55
Replacing the batteries 63
Reset 59
Resetting the measuring parameters
RS232 interface 54
60
S
Safety 11
safety precautions 11
Save interval 44
Scope of delivery 13
Setting the baud rate 56
Setting the date 13, 57
Setting the resolution (pH) 17
Setting the time 13, 57
single-point calibration 20
Single-point calibration (pH)