Xylem OxiTop Control OC 100 Operating Manual

OPERATING MANUAL
ON/OFF
GLP
RUN/ENTER
OC 1x0
®
ba31114e07 08/2014

OxiTop® Control OC 100

®
OxiTop

BOD/OXYGEN CONSUMPTION: BOD, RESPIRATION, BIOGAS DETERMINATION

C or OxiTop® C/B
Copyright © 2016 Xylem Analytics Germany GmbH
Printed in Germany.
OxiTop® Control OC 100 Contents
OxiTop® Control OC 100 - Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 The system OxiTop® Control. . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 The measurement principle. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Data management. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.4 Literature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.1 Safety information in the operating manual . . . . . . . . . . . 9
2.1.2 Safety signs on the meter. . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 Requirements for safe operation . . . . . . . . . . . . . . . . . . . 9
2.2.3 Unauthorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.1 Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.2 Battery status signals . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.3 Energy saving feature (automatic switch-off) . . . . . . . . . 12
3.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.1 Inserting the supply battery holder - controller . . . . . . . . 12
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 General operating principles . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1.1 Keys - controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1.2 Socket field - controller. . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 Switching on the controller . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Switching on the measuring heads. . . . . . . . . . . . . . . . . 16
4.3 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3.2 Menus, functions and dialogs. . . . . . . . . . . . . . . . . . . . . 18
4.3.3 The course of the measurement. . . . . . . . . . . . . . . . . . . 19
5 Operating mode: BOD Standard / BOD Routine . . . . . . . . . 21
5.1 Sample preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2 Starting the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Calling up all data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.4 SAMPLE STATUS TYPE list (sample management) . . . . . . . 26
5.4.1 Showing a sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.4.2 Erasing data of finished samples . . . . . . . . . . . . . . . . . . 28
5.4.3 Showing the measuring head list . . . . . . . . . . . . . . . . . . 29
ba31114e07 08/2014 3
Contents OxiTop® Control OC 100
5.4.4 Calling up data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.4.5 Calling up data - Stop. . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.5 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.5.1 Sample statistics (BOD Standard operating mode) . . . . 34
5.5.2 Excluding a curve (BOD Standard operating mode) . . . 35
5.5.3 Cursor interrogation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.4 Curves display for cold samples. . . . . . . . . . . . . . . . . . . 37
5.5.5 Measured values outside the measuring range . . . . . . . 37
6 Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7 GLP/TOOLS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.1 GLP/TOOLS main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.2 Show free measuring heads . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.3 Show settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.4 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.4.1 Operating mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.4.2 Operational life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.4.3 Date/time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.4.4 GLP (Operating mode BOD Standard ) . . . . . . . . . . . . . 45
7.4.5 GLP - calibration interval (- Calinterval) . . . . . . . . . . . . . 46
7.4.6 Data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.4.7 AutoTemp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.4.8 Switch-off interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4.9 Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.5 Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.5.1 Showing the measuring heads. . . . . . . . . . . . . . . . . . . . 51
7.5.2 Measuring head information. . . . . . . . . . . . . . . . . . . . . . 51
7.5.3 Controller information. . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.5.4 cal-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.5.5 Pneumatic test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.6.1 Erasing finished samples . . . . . . . . . . . . . . . . . . . . . . . . 57
7.6.2 Reset/release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
7.6.3 Reading out data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
8 Transmitting data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
8.1 Achat OC software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
9 Maintenance, cleaning, disposal. . . . . . . . . . . . . . . . . . . . . . 63
9.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
9.1.1 General maintenance activities . . . . . . . . . . . . . . . . . . . 63
9.1.2 Replacing the supply batteries (controller). . . . . . . . . . . 63
9.1.3 Replacing the data backup battery (controller). . . . . . . . 65
9.1.4 Changing the supply battery (measuring head) . . . . . . . 67
9.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
9.2.1 Cleaning of the sample bottles. . . . . . . . . . . . . . . . . . . . 68
9.2.2 Cleaning of the controller and measuring heads . . . . . . 68
9.3 Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4 ba31114e07 08/2014
OxiTop® Control OC 100 Contents
10 What to do if.... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.1 Display messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.2 Display messages on the charging status of the batteries . . . 74
10.2.1 Batteries in the controller . . . . . . . . . . . . . . . . . . . . . . . . 74
10.2.2 Battery in the measuring head . . . . . . . . . . . . . . . . . . . . 74
10.3 General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
11 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
11.1 Controller OxiTop® OC 100. . . . . . . . . . . . . . . . . . . . . . . . . . . 79
11.2 Measuring head OxiTop
®
-C or OxiTop®-C/B . . . . . . . . . . . . . 80
ba31114e07 08/2014 5
Contents OxiTop® Control OC 100
6 ba31114e07 08/2014
OxiTop® Control OC 100 Overview


2
0
m
l
lt
m
2
Op
T
T
V
VV
TR
OM
BOD
 
 
M(O2) Molecular weight (32000 mg/mol) R Gas constant (83.144 lmbar/molK) T
0
Reference temperature (273.15 K)
T
m
Measuring temperature
V
t
Bottle volume (nominal volume in ml)
V
l
Sample volume in ml a Bunsen absorption coefficient (0.03103) p(O
2
) Difference of the D. O. partial pressure (mbar)

1 Overview

1.1 The system OxiTop
®
Control
The classical application field for the OxiTop® measuring heads and controller is the BODx determination (BODx = Biochemical Oxygen Demand for the time x). The evaluation of biological degradability (e.g. test according to OECD 301F) is also part of this field.
With different sample vessels, the OxiTop
®
measuring heads and the OxiTop®
Controller 110 can also be used for other applications such as: Evaluation of respiration and toxicity in earth, sludge, waste and sediment
(e.g. extraction of contaminated earth according to recovery concepts)
Evaluation of the respiration rate of cell culturesMicrobiological growth and stress examinationsMeasurement of anaerobic degradation processes (e.g. biogas determina-
tion with the measuring head OxiTop
®
-C/B)
Accessories: see WTW catalog and price list

1.2 The measurement principle

The respirometric measurement is a pressure measurement. If oxygen is consumed in a closed vessel at a constant temperature, a negative pressure develops. If a gas is released, an overpressure develops. The OxiTop measuring head measures and stores this pressure for the whole duration of a measurement once started.
®
The OxiTop
OC 100 controller collects the pressure values from the
measuring heads and processes them. The formula shown below is the basis for all calculations for the BOD using the
®
values from the OxiTop
measuring head.
®
ba31114e07 08/2014 7
Overview OxiTop® Control OC 100
The interpretation of the pressure differences in the temporal course depends on the measured material and its preparation and on the sample manipulation (e. g. intermediate aerations) during the measuring period.

1.3 Data management

The measuring head records the measured values and stores the measuring data. Using the controller, the collected and stored data are read out from the measuring head and stored in the “sample management“. Starting at the OxiTop
®
sample management, it is possible to make the graphical and statis-
tical evaluation of the sample data. The data can be transferred from the controller to a computer via the RS232
interface. The read-out software Achat OC facilitates to process the stored data further.
Additionally, the data can be transferred to an IR printer via the IR interface.

1.4 Literature

Further information and application reports on the subject of measuring with the system OxiTop
CD-ROM Applications for OxiTop
®
Control can be obtained at no cost from WTW:
®
Control
CD-ROM Principles of measuring technique
8 ba31114e07 08/2014
OxiTop® Control OC 100 Safety

2 Safety

2.1 Safety information

2.1.1 Safety information in the operating manual

This operating manual provides important information on the safe oper­ation of the instrument. Read this operating manual thoroughly and make yourself familiar with the instrument before putting it into opera­tion or working with it. The operating manual must b e kept in the vicinity of the instrument so you can always find the information you need.
Important safety instructions are highlighted in this operating manual. They are indicated by the warning symbol (triangle) in the left column. The signal word (e.g. "CAUTION") indicates the danger level:
WARNING
indicates a possibly dangerous situation that can lead to seri­ous (irreversible) injury or death if the safety instruction is not followed.
CAUTION
indicates a possibly dangerous situation that can lead to slight (reversible) injury if the safety instruction is not followed.
NOTE
indicates a possibly dangerous situation where goods might be damaged if the actions mentioned are not taken.

2.1.2 Safety signs on the meter

Note all labels, information signs and safety symbols on the instrument. A warning symbol (triangle) without text refers to safety information in this operating manual.

2.2 Safe operation

2.2.1 Authorized use

Only the operation and running of the instrument according to the instructions and technical specifications given in this operating manual is authorized (see section 11 T
ECHNICAL DATA, page 79).
Any other use is considered unauthorized.

2.2.2 Requirements for safe operation

Note the following points for safe operation:
The instrument may only be operated according to the authorized
ba31114e07 08/2014 9
Safety OxiTop® Control OC 100
use specified above.
The instrument may only be supplied with power by the energy sources mentioned in this operating manual.
The instrument may only be operated under the environmental conditions mentioned in this operating manual.
The instrument may only be opened if this is explicitly described in this operating manual (example: Changing the batteries).

2.2.3 Unauthorized use

The instrument must not be put into operation if:
it is visibly damaged (e.g. after being transported)
it was stored under adverse conditions for a lengthy period of time (storing conditions, see section 11 T
ECHNICAL DATA, page 79).
10 ba31114e07 08/2014
OxiTop® Control OC 100 Commissioning

3 Commissioning

3.1 Scope of delivery

Controller OxiTop® OC 100  Battery holder with batteries (3 x Mignon, AA, AM3, LR6)Operating manualAdditional scope of delivery: depending on set composition

3.2 Power supply

The controller OxiTop® OC 100 and measuring head OxiTop®-C or OxiTop®­C/B are both battery-powered.
Additionally, the controller has a data backup battery. This battery ensures that your measuring data and instrument settings are stored in the controller OxiTop
®
OC 100.

3.2.1 Batteries

Controller Power supply 3 alkaline batteries (alkali manganese),
Size: Mignon, AA, AM3, LR6 Runtime: > 100 h (approx. 1000 start-ups in normal use)
Data backup 1 x lithium battery CR2430 (accessory).
Runtime: typically 4 years
Measuring head
Power supply Batteries: 2 x lithium battery CR2430 (accessory).
Runtime: typically 2 years

3.2.2 Battery status signals

®
The OxiTop
-OC 100 controller and the OxiTop® measuring heads are pow­ered by batteries. A low battery status of the controller or measuring head is automatically indicated by a display message and a signal tone after the instru­ment is switched on (see section 10.2 D
STATUS OF THE BATTERIES, page 74).
ISPLAY MESSAGES ON THE CHARGING
1st level (warning level) Battery LoBat ! 2nd level (error level) Battery empty !
The current battery status can also be queried in the menu:
 Controller: Menu GLP/TOOLS - Check - Controller Info  Measuring head: Menu GLP/TOOLS -Check - info
ba31114e07 08/2014 11
Commissioning OxiTop® Control OC 100
LISTED 8F93 LaboratoryEquipment E163694
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: (1)This devise may not cause harmful interference, and (2) this devise must accept any interference received, including interference that may cause undesired operation.
Ser.-Nr.:05150013
OxiTop OC 1x0
WTW 82362 Weilheim
Made in Germany
1x CR 2430 Lithium
3xAlkaline, AA size, 1,5V
1
1
1
1

3.2.3 Energy saving feature (automatic switch-off)

The instrument switches off automatically following the last key actuation after expiry of the specified switch-off interval.
To specify the switch-off interval: see section 7.4.8 S
WITCH-OFF INTERVAL, page
49.

3.3 Initial commissioning

Perform the following activities: Insert the supply battery holder in the controller (see section 3 .3.1 I
THE SUPPLY BATTERY HOLDER - CONTROLLER, page 12)
Switch on the instrumentSet the date and time (see section 7.4 S
ETTINGS, page 43)

3.3.1 Inserting the supply battery holder - controller

Loosen the 4 screws underneath the housing using a Phillips screwdriver.
NSERTING
1Screws
Place the controller on the lower case.Remove the upper case and put it down to the right next to the lower case
with the display downwards.
Connect the supply battery holder with the power supply contacts of the
upper case.
Insert the supply battery holder in the lower case with the open side down.
12 ba31114e07 08/2014
OxiTop® Control OC 100 Commissioning
1
2
1 Supply batteries in the battery holder (lower case) 2 Data backup battery (upper case)
Set the upper case on top of the lower case, turn the controller and tighten
the housing screws using the screwdriver.
Switch on the controller. The battery insertion was successful if no error
message appears concerning the supply batteries.
ba31114e07 08/2014 13
Commissioning OxiTop® Control OC 100
14 ba31114e07 08/2014
OxiTop® Control OC 100 Operation
ON/OFF
GLP
RUN/ENTER
1
2
5
34
7
8
6
The function keys are used to start a function or to change to another function. Confirmed data and settings are preserved.
1 <On/Off> Switching on/off
2 <PRINT> Printing of measurement data and set-
tings via IR interface
3 <RUN/
ENTER>
Confirmation of entries
4<><> Select and change settings
Function keys: 5 <ORG> Sample management: List of samples,
reading/calling up/deleting the data of indi­vidual measuring heads or samples, shows measuring heads or samples
6 <COMM> Communication with measuring heads:
Start measuring, change or show settings, call up data
7 <GRAPH> Evaluation: Graphic and numerical dis-
play of measuring data
8 <GLP> GLP/Tools: Change or display settings,
display free measuring heads, display or change settings, perform checks or main­tenance
ON/OFF
RUN/ENTER
GLP

4Operation

4.1 General operating principles

4.1.1 Keys - controller

ba31114e07 08/2014 15
Operation OxiTop® Control OC 100
1

4.1.2 Socket field - controller

1 RS232 interface
The RS232 interface is used for communication with the PC. Usage of the RS232 interface is exclusively intended with the WTW software "Achat OC“ (see section 8.1 A
CHAT OC SOFTWARE, page 62).

4.2 Switching on

4.2.1 Switching on the controller

The current date and time appear for approx. 2 seconds (important for the allocation of sample numbers). If the date or time is wrong, correct them under GLP/TOOLS.
The instrument is in the sample management mode (operating mode BOD Routine , in the default condition).

4.2.2 Switching on the measuring heads

The measuring heads are immediately ready for operation. The controller switches the measuring heads on and off during communication.
To avoid malfunctions:
If you use two or more controllers simultaneously make sure that the distance between the controllers is at least 3 meters!
16 ba31114e07 08/2014
OxiTop® Control OC 100 Operation

4.3 Navigation

4.3.1 Operating modes

BOD Routine:
(preselected)
Measurement of single samples for a BOD
(x = 0.5h ... 99d) at 20 °C
x
The measuring range and filling volume can be selected out of seven per-
manent ranges and the corresponding filling volumes.
The function is permanently switched on
BOD Standard:  Measurement of up to 12 parallel sample processes per overall sample for
a BOD
(x = 0.5h ... 99d) at 20 °C
x
The measuring range and filling volume can be selected out of seven per-
manent ranges and the corresponding filling volumes.
Automatic sample statistics with averagingThe AutoTemp function and the GLP mode can be switched on
ba31114e07 08/2014 17
Operation OxiTop® Control OC 100

4.3.2 Menus, functions and dialogs Example: Pressing the <COMM> button causes the controller to change to the

SELECT ACTION menu (communication with the measuring heads).
Signal tone
Action: Press button. The instrument displays:
This symbol in the operating manual means: A signal tone sounds.
Description of what the action caused and possible further request:
SELECT ACTION menu (communication with the mea­suring heads): Preselected: - Start sample.
Select a menu item with <><>. The selected function or line is displayed in inverse video. Confirm the menu item with <RUN/ENTER>. The submenu is opened or the function is carried out.
Measuring head Each measuring head that receives a command from the controller indicates
this by a short flashing signal.
®
The OxiTop
measuring head is indicated by the symbol
18 ba31114e07 08/2014
OxiTop® Control OC 100 Operation
GLP
Time
End of measurement
Start of measurement
AutoTemp
Measurement
Data transfer
Starting the measuring heads
Sample management
Evaluation
Print
Print

4.3.3 The course of the measurement

Preparation before
starting the mea-
surement
Measuring
Select operating modeSelect settings
Operating mode and settings cannot be changed for a measure­ment once it is started!
ba31114e07 08/2014 19
Operation OxiTop® Control OC 100
20 ba31114e07 08/2014

OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine

RUN/ENTER
5 Operating mode: BOD Standard / BOD Rou-
tine
The BOD Standard mode (parallel sample process for up to 12 measuring heads per overall sample) is preset in the delivery condition of the controller.
To change to the BOD Routine operating mode (single samples): see section
7.4.1 O The handling of controller and measuring heads is basically the same in both
operating modes. Where there are differences for the BOD Routine operating mode they are described.

5.1 Sample preparation

See application reports.
Screw the OxiTop close them tightly.
PERATING MODE, page 43.
®
measuring heads onto the measuring vessels and
NOTE
Never use joint grease or other lubricants for your OxiTop
®
measuring heads. Some of these products contain solvents that can cause severe damage to the plastic housing of the measuring head. The sealing of the measuring vessels is perfectly adequate without grease. However, you should always wipe off heavy contamination and
®
particles on the sealing surfaces of the rubber sleeves and OxiTop measuring heads. WTW accepts no liability for damage due to the use of joint grease.

5.2 Starting the measurement

SELECT ACTION menu (com­munication with the measuring heads): Preselected: - Start sample.
(In BOD Routine mode appears: START).
Select the measuring range with <><>. The filling volume required is given in the right hand column.
The controller stores the setting (memory function: the last selected measurement range is set).
ba31114e07 08/2014 21
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
Confirm the selected measure­ment range for the sample. The automatically assigned sample number (YY/MM/DD and sequential number) is given in the header line.
Additional information: Type of measurement, run duration, measurement range,
final date, ID number.
Temp display => the AutoTemp function (see section 7.1
GLP/TOOLS
MAIN MENU, page 41) is switched on.
GLP display => the GLP function (see section 7 GLP/
TOOLS, page 41) is switched on.
Change the ID number for the additional identification of the sample (e.g. sampling location) as follows:
Select the - I.D. number menu item with <><>Confirm with <RUN/ENTER>.  Set the required Id number (setting range 001 ... 255) with
<><>.
Confirm with <RUN/ENTER>. Use <PRINT>
(in BOD Routine mode only), see section 6 P
to print out the entire sample information
RINTING, page
39). Confirm the start of measure-
ment. Contact selection:
The controller repeatedly sends the start information in the scanner mode until feedback is successfully received from the OxiTop
®
measuring head. After the successful start message
from the measuring head:
Displays ! started ! From this
point in time, the sample exists within the SAMPLE STATUS TYPE list (sample manage- ment).
Subsequently, the request to start the next measuring head is made automatically.
In BOD Routine operating mode there is no 1... identifier in the display, and the controller automatically returns to the entry menu after starting the measuring head.
22 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
Start the next measuring head (2). Contact selection:
The number of the measuring head is automatically incre­mented.
The controller continues to work with the contact selection, i.e. you can now start a sequence of measuring heads (parallel sample process) without having to press any further keys by holding the controller to the next measuring head to be started.
When all the measuring heads provided for this sample have been started: Confirm Stop start with <RUN/ENTER>. The controller returns to the entry menu.
If no measuring head is started (e.g. because the controller was not held to - or not close enough to - a measuring head):
The start was stopped. Use <RUN/ENTER> - Continue
start to confirm “Continue start” and hold the controller to a measuring head (see above).
Return to entry level menu with
-Stop  start
If an attempt is made to start a measuring head that was already started:
Displays already used! Use <RUN/ENTER> to confirm -
New  and hold the controller to a free measuring head (see above).
Return to entry level menu with
-Stop  start
In the delivery state, the controller automatically makes space when Start is selected if the measured value memory is full. To do this, it deletes the oldest finished sample (if a finished sample is available). You can change this setting to “manual erase” (see section 7.4.6 D
ATA STORAGE, page 47).
ba31114e07 08/2014 23
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
40 cm to 1 m
RUN/ENTER

5.3 Calling up all data

This function is used to call up the data of all measuring heads, no matter in which operating mode they were started.
To call up the data of individual measuring heads: see section 5.4 SAMPLE STATUS TYPE
LIST (SAMPLE MANAGEMENT), page 26.
SELECT ACTION menu (com­munication with the measuring heads): Preselected: - Start sample.
Select the - Call up all data menu item with <><>
Point the controller at the measuring heads:
Queries the data of all active measuring heads in the scanner mode.
The controller stores the data obtained and updates the
SAMPLE STATUS TYPE list (sample management).
Duration of a pass: approx. 3 seconds for 12 measuring heads in a stirring system.
If the measuring heads do not all respond in the 1st pass of the data query, the controller searches for the missing measuring heads in the scanner mode for approximately a further 7 seconds. Without having to press another key, you can continue the data queries at other locations (e.g. other stirring plat­forms or other thermal cabinets). Approximately 7 seconds after the last call up was performed, the scanner mode breaks off automatically and the following display appears:
24 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
40 cm to 1 m
RUN/ENTER
Preselected:
- Continue transfer: A new pass is made only for measuring heads that have not yet responded.
Return to the entry menu with -
Stop
For information on searching for missing measuring heads, see section 10 W
HAT TO DO IF..., page
69
Point the controller at the measuring heads:
The call up of further measuring heads is performed in the scan­ner mode.
Message that all the data of all the measuring heads has been called up. The instrument then returns to the entry menu.
Immediately after the complete measurement data records of a fin­ished measurement have been called up, the corresponding mea­suring head is given the free status. The measuring head can be used for a new measurement. The relevant sample in the SAMPLE STATUS TYPE list (sample management) is marked as finished (see section 5.4 SAMPLE STATUS TYPE
MANAGEMENT), page 26), as soon as the measurement time has
LIST (SAMPLE
expired and the data of all measuring heads of this sample have been called up.
ba31114e07 08/2014 25
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
3
2
1

5.4 SAMPLE STATUS TYPE list (sample management)

Entry into the sample management. A list of samples appears in the display (if samples are avail­able):
1 Sample number:
Date (JJ/MM/TT) and consecutive number
2 Temporal process of the sample:
– Status bar partly filled:
The sample is not yet ready.
– Status bar filled:
The ready and complete data set of the sample can be called up from the measuring head.
– Status bar filled plus checkmark:
The sample is ready. The complete data set is given in the con­troller for evaluation.
3 BOD type and run duration.
– B20h: BOD with 20 hours run duration – BOD5: BOD with 5 days run duration
The samples started in the BOD Routine operating mode also appear in the SAMPLE STATUS TYPE lists (sample management) of all other modes. Data of samples that have been started in other operating modes do not appear in the sample management of the BOD Routine operating mode.
Reporting order:
At the upper end of the list: finished samples (if available) Under this: current samples
Sorting of the samples: according to date and sequential number 01 ... 99, from the oldest to the newest sample.
26 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
RUN/ENTER
40 cm to 1 m
RUN/ENTER
Select a sample. According to the sample selected, one of the two following menus appears.
Finished sample Current sample
The header line contains the following data of the selected sample:
Sample numberID number (e.g. I001) BOD type and run duration.

5.4.1 Showing a sample

This function is used to find the measuring heads or samples. Sample labeling is not necessary.
Point the controller at the measuring heads:
SAMPLE STATUS TYPE list (sample management)
Use <><> to select a sam­ple.
Select the -Show sample menu item with <><>
The controller transmits the call up of the selected sample. The allocated measuring heads flash for approx. 5 seconds.
After the message, the controller automatically returns to the previous menu.
ba31114e07 08/2014 27
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
RUN/ENTER

5.4.2 Erasing data of finished samples

This function erases the data of finished samples from the SAMPLE STATUS TYPE list ( sample management) of the controller. You can only erase a sample if it is finished; in non-finished samples, the menu item -Erase sample does not appear.
SAMPLE STATUS TYPE list (sample management)
Use <><> to select a fin­ished sample.
Select the - Erase sample menu item with <><>.
Safety prompt with possibility to return.
The sample was erased.
After the message, the controller automatically returns to the SAMPLE STATUS TYPE list (sample management).
28 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
RUN/ENTER
40 cm to 1 m

5.4.3 Showing the measuring head list

This function is used to allocate individual measuring heads to the rele­vant sample in the SAMPLE STATUS TYPE list (sample management) and to find individual measuring heads.
Select the - list menu item with <><>
The list of measuring heads of the selected sample is displayed together with the serial numbers of the measuring heads. (In the BOD Routine operating mode, this list always contains only one measuring head.)
Point the controller at the measuring heads:
Use <><> to select a measuring head and confirm with <RUN/ENTER>. The controller repeatedly transmits the call up of the selected measuring head. The measuring head flashes for approx. 5 seconds.
Use <ORG> to return to the SAMPLE STATUS TYPE list (sample management).

5.4.4 Calling up data

This function is used to call up the data of individual samples. To call up the data of all samples, see section 5.3 C
ALLING UP ALL DATA,
page 24.
ba31114e07 08/2014 29
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
RUN/ENTER
40 cm to 1 m
RUN/ENTER
SAMPLE STATUS TYPE list (sample management)
Use <><> to select a sam- ple whose run duration is not yet com-
pleted (bar not yet full)
data after complete mea-
surement that have not yet been called up (bar full, no checkmark)
Jumps to submenu of the selected sample. Preselected: - Call up data
Point the controller at the measuring heads:
Sequential call up of the mea­suring heads allocated to the sample (in this example: 7) in the scanner mode:
The controller
stores the data obtained updates the
SAMPLE STATUS TYPE list (sample management)
Duration of a pass: approx. 3 seconds for 12 mea­suring heads in a stirring sys­tem.
The controller confirms the call up performed.
The controller then returns automatically to the submenu.
30 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
RUN/ENTER
If the measuring heads do not all respond, the following dis­play appears:
Renewed start of the data call up.
The controller continues to search for missing measuring heads in the scanner mode. Without having to press another key, you can continue the data queries at other locations (e.g. other stirring platforms or other thermal cabinets). Approximately 7 seconds after the last call up was per­formed, the scanner mode breaks off and the following dis­play appears:
Use <RUN/ENTER> to restart the data transfer (see above).
Return to the
SAMPLE STATUS TYPE list (sample management).
Transferred and completed samples appear with a check­mark next to them. (The complete measurement data records of all finished measuring heads of the sam­ple have been procured.)
Immediately after the successful data transfer of a sample, the allo­cated measuring heads are given the free status. The measuring heads can be used for a new measurement.
The user administration function is not active in the delivery condi­tion.

5.4.5 Calling up data - Stop

If a measuring head is missing or defective and the controller cannot call up the sample as a result, you can stop the data transfer with the -Stop menu item.
ba31114e07 08/2014 31
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
RUN/ENTER
After confirmation, the display message shown here appears.
Three displays are then possible:
Case 1 The sample is still running and
individual measuring heads of the sample are not attainable.
Case 2 The run duration of the sample
has ended and the finished, all data of the attainable measuring heads have been transferred.
-Erase: Removes missing measuring
then
heads from the data stock. The completed sample is given the “finished” status.
-Back: Jumps back to the previous menu.
Abort using a function key: The stopped sample is not declared as “finished”.
Case 3: No measuring head of a sample
is attainable. (In the BOD Routine operating mode: The relevant measuring head is not attainable).
-Erase: Removes a sample from the data stock.
then
-Back: Jumps back to the previous menu.
- Show
:
As in - Show sample, but selectively for missing measuring heads (identification option if only the sender of the measuring head is defective. Otherwise, see the section 10 W
IF..., page 69)
HAT TO DO
32 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
1
2
- Missing
:

5.5 Evaluation

Only appears when the run duration of the sample has expired. Here you can remove the data of a missing or defec­tive measuring head from the SAMPLE STATUS TYPE list (sample management) of the controller. In the BOD Standard operating mode, you can set a sample to the "finished" status (checkmark ) if all the other measuring heads are already “fin­ished”: Erases the missing measuring heads from the
SAMPLE STATUS TYPE list (sample management).
SAMPLE STATUS TYPE list
(sample management) Use <><> to select a sam-
ple. Use <PRINT> to print out the
results with curves. Evaluation of the sample:
Display of all the curves together with mean value data. Check the display for outliers.
Use <PRINT> to print the results with curves (according to print­ing format - see
1 End of measuring range
the section 6 P
RINTING, page 39)
2 Mean value
Example: Overall sample with outlier.
In the BOD Routine operating mode, the evaluation shows the selected sample as a single curve with the measured value data. Use <RUN/ENTER> to switch between the cursor interroga­tion (see section 5.5.3 C
URSOR INTERROGATION, page 36) dis-
play of the curve with measured value data.
Scroll through the parallel sam­ples: Display of individual curves (cyclical pass) with data of the BOD final value.
Use <PRINT> to print out a sin­gle result with a single curve.
ba31114e07 08/2014 33
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER

5.5.1 Sample statistics (BOD Standard operating mode)

Starting point: Display of a single curve.
Changes to sample statistics. The menu selection shown here
only appears if the number of measuring heads (n) is at least
2. If n = 1 pressing of <RUN/
ENTER>
changes directly to the
cursor interrogation.
When jumping to the selection menu from “Show all curves“, the menu item - Exclude curve is not displayed.
Results of the finished sam-
Data of the current sample:
ple:
Mean value SD: standard deviation
(from n = 3)
n: number of measuring
end datecurrent mean value n: number of measuring
heads
heads
<PRINT>
Printout of the results with curves
<PRINT>
Printout of the data (current mean value, n, sample infor­mation) with curve paths to date
34 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine
RUN/ENTER
RUN/ENTER
RUN/ENTER

5.5.2 Excluding a curve (BOD Standard operating mode) This function is used to temporarily exclude a single curve (e.g. an

outlier) from the evaluation and averaging of an overall sample.
The curve is only excluded temporarily! The data stock of the SAMPLE STATUS TYPE list (sample management) does not change. The excluded curve is present again when the call is repeated.
The function "List of measuring heads“ (see section 5.4 SAMPLE STATUS TYPE
LIST (SAMPLE MANAGEMENT), page 26) is used
to find leaky or defective single samples.
Display of single curves: Select curve.
Use <><> to select -
Exclude curve.
- Exclude curve is preselected.
-Back: Returns to the previous
menu.
Message Curve excluded!.
Updated display (curve excluded, mean value recalcu­lated).
Use <PRINT> to print the updated results with curves (without the excluded curve).
ba31114e07 08/2014 35
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
1
2
3
RUN/ENTER

5.5.3 Cursor interrogation

Starting point: Display of a single curve with measured value data.
Use <><> to select -Cursor query (this menu is not dis- played in the BOD Routine oper­ating mode).
Use <><> to run through the curve values.
Use <PRINT> to print out the current display.
1 The scaling of the Y-axis
is automatically adapted to the measured value
2 Measuring time at the
cursor position
3 Measured value at the
cursor position
Or from the display of all the single curves:
Use <><> to run through the curve mean values.
Use <PRINT> to print out the current display.
Return to the previous menu using <RUN/ENTER>. Return to the display of all the single curves using <GRAPH>.
36 ba31114e07 08/2014
OxiTop® Control OC 100 Operating mode: BOD Standard / BOD Routine

5.5.4 Curves display for cold samples

Display of a single curve.
Display of all curves.
When samples are inserted that are too cold the maximum AutoTemp time is not sufficient for the sample temperature to reach the incubation temperature: Warming the sample can cause excess pressure.
Display: Negative values of the curve will be truncated and the curve originates from the time axis but not the origin.

5.5.5 Measured values outside the measuring range

In the following cases, undef. (undefined) is displayed instead of the measured value or mean value during evaluation:
A measured curve exceeding the measuring range at any point of its
path.
A measured curve undercutting the measuring range at its end
point.
Sample displays:
A measured value exceeds the measuring range (Overflow).
A measured value undercuts the measuring range (Underflow).
ba31114e07 08/2014 37
Operating mode: BOD Standard / BOD Routine OxiTop® Control OC 100
A measured value temporarily leaves the mea­suring range.
After the exclusion of the defective curve (see section 5.5.2 EXCLUDING
A CURVE (BOD STANDARD OPERATING MODE), page 35), the controller
displays the mean value again.
38 ba31114e07 08/2014
OxiTop® Control OC 100 Printing

6Printing

Switch on the IR printer. Point the controller at the printer:
When printing from the SAMPLE STATUS TYPE list (sam­ple management) or from the evaluation:
The controller transfers the report data of the selected sam­ple to the IR interface.
Print date and print timeSample numberMeasured value: mean value, SD, n (with parallel samples)
or end value (with single samples)
Type and run duration, measurement rangeStart date and start time, end date, ID number If GLP is switched on: Note, GLP OnIf the AutoTemp function is switched on:
Note, AutoTemp On
Curve(s) + BOD mean value or end valueWith GLP On: List of the serial numbers of the measuring
heads allocated to the sample (only for samples started in the BOD Standard measuring modes)
ba31114e07 08/2014 39
Printing OxiTop® Control OC 100
Printing from the evaluation:
If the controller is
in the display of all curves with cursor displayin the display of the single curves with or with-
out cursor interrogation
the printout is a copy of the display indication.
40 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
GLP

7 GLP/TOOLS

7.1 GLP/TOOLS main menu

GLP/TOOLS menu:
Select a menu item with <><>. Confirm with <RUN/ENTER>.
The selected menu opens up or the function is carried out.
GLP/TOOLS menu:
- Show free

- Show Settings The current settings are displayed here.
- Settings Here you can undertake or change the following settings:
- Check - Show  (finished/all)
- Maintenance - Erase sample
This function is used to identify free measuring heads. Thus an additional sample labeling is not required. Free measuring heads can be used to start new samples.
- Operation mode
- Measuring time
- date/time
-GLP; calibration interval (not in the BOD Routine operating mode)
- Memory erase!
- AutoTemp (BOD Standard
- Switch-off interv.(not in the BOD Routine operating mode)
- Temperature- Height- Air pressure- Limit pressure- Language
- info (with report printout)
- Controller Info (with report printout)
- Cal Test
- Pneumatic test
(in the controller - SAMPLE STATUS TYPE list (sample management))
- Reset/release
- Restore data
ba31114e07 08/2014 41
GLP/TOOLS OxiTop® Control OC 100
GLP
40 cm to 1 m
GLP
RUN/ENTER

7.2 Show free measuring heads

This function is used to identify free measuring heads. Free measuring heads can be used to start new samples.
GLP/TOOLS menu: Preselected: - Show free .
Press <RUN/ENTER>.
Point the controller at the measuring heads:
All free measuring heads flash for approx. 5 seconds.

7.3 Show settings

The current settings are displayed here.
GLP/TOOLS menu: Select the - Show Settings menu
item with <><>
The following example illustrates the presettings in the
BOD Standard operating mode:
SHOW SETTINGS menu:
A list containing the current set­tings is displayed.
Use <><> to scroll. (Automatic stop at the beginning and end of the list)
In the "Show settings“ menu: Print out the whole list of current settings (as in the display).
42 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
GLP
RUN/ENTER

7.4 Settings

With this function you can undertake or change settings. The table below shows the settings in the delivery state.
GLP/TOOLS menu. Select the - Settings menu item with
<><>
Default setting (bold)
Setting
Operating mode BOD Routine BOD Standard Operational life 5 days
Date current current
and setting range
(0.5 h to 99 days)
5 days (0.5 h to 99 days)
Time current current GLP Off
(permanently set)
On
Off
Calibration interval --- 12 months
to 36 months
Clearing the memory automatic
or manual
AutoTemp On
(permanently set)
Switch-off interval 5 minutes
(permanently set)
automatic
or manual
On
Off
5 minutes to 15 minutes
Language German German

7.4.1 Operating mode

You can select the following operating modes:
BOD StandardBOD Routine
SETTINGS menu: Select the - Operation mode
menu item with <><>.
ba31114e07 08/2014 43
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
Confirm with <RUN/ENTER>. Select the required operating mode <><>. Confirm with <RUN/ENTER>.
-Back:: Return to the main menu SETTINGS:

7.4.2 Operational life

Here you determine the run duration of the measurement.
SETTINGS menu: Use <><> to select the -
Measuring time menu item.
Set the days (1 to 99) or hours (0.5 to 23) with <><>. Confirm with <RUN/ENTER>.
Display: Adjusted meas. time. Delivery condition: 5 days.
44 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER

7.4.3 Date/time

Here you set the date and time for the controller (important for the assignment of sample numbers).
SETTINGS menu: Use <><> to select the -
date/time menu item.
Use <><> to set each num­ber block on a black back­ground. Confirm with <RUN/ENTER>.
Set the day, month, year, min­utes, seconds one after the other.
The run duration of samples already started is not affected by changing the date or time.

7.4.4 GLP (Operating mode BOD Standard )

The GLP (Good Laboratory Practice) set of rules requires a detailed description of all equippings and procedures in a laboratory.
These descriptions include the room equipment of the laboratory, the instruments used, and the measuring procedures laid down, and require the complete documentation of the samples tested. The type and frequency of the use of test resources is to be described and docu­mented as well.
To make the documentation of the test resource monitoring easier an d simpler the OxiTop
®
Control system has a GLP mode.
When this mode is switched on, the following data are documented:
the calibration intervals that have been setthe calibrations that have been performed
The reports contain the series numbers of the measuring head and controller and the date of the calibration.
The GLP mode monitors the calibration intervals and blocks the start of a new measurement if a calibration will be due within the measuring time.
ba31114e07 08/2014 45
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
Switching the GLP func-
tion on or off
SETTINGS menu: Use <><> to select the -
GLP menu item.
GLP menu: Preselected: -GLP
Confirm with <RUN/ENTER>. Use <><> to select either GLP On or Off.
Confirm with <RUN/ENTER>.

7.4.5 GLP - calibration interval (- Calinterval)

Set the time period (1 to 36 months) here. When it expires, the instru­ment registers that the next test resource monitoring of the measuring heads is due. After the calibrating interval expires, measuring is blocked until the calibration is performed or the Off setting is selected.
If the end of the calibrating interval set up lies within the measuring time of a measurement to be started, this measurement cannot be started.
GLP menu: Use <><> to select the -
Calinterval menu item.
Confirm with <RUN/ENTER>.
Use <><> to set the time period for the calibrating interval (1 - 36 months).
Default setting: 12 months. Confirm with <RUN/ENTER>.
46 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER

7.4.6 Data storage

Here, you can set whether the controller should automatically erase the oldest finished samples if the memory is full to create space for new measurement data (setting auto).
With the manual setting, the message "Memory lack! Erase finished sample/s! is displayed if the memory is full (see section 10 W
IF..., page 69).
HAT TO DO
Back up your data regularly!
SETTINGS menu: Use <><> to select the -
Memory menu item.
Preselected: MEMORY -Erase
auto To change it:
Confirm with <RUN/ENTER>. Use <><> to select either auto or manual. Confirm with <RUN/ENTER>.
-Back (return to the SETTINGS
menu)

7.4.7 AutoTemp

The AutoTemp function controls the automatic start of the measure­ment after checking the temperature adaptation.
The pretemperature regulation to the precise incubator temperature is recommended but not essential. Recommendation: e.g. regulate the temperature of the sample for BOD5 measurements from 15 °C up to 20 °C.
You can tightly close the sample bottle with the measuring head imme­diately and start the measurement. The AutoTemp function then auto­matically triggers the start of the actual measuring after checking the temperature adaptation. The measuring time of the AutoTemp phase (adaptation phase plus the test phase) is added to the sample measuring time selected in the settings.
Pretemperature regula-
tion of the sample
With the AutoTemp function switched on and, adhering to the recom­mendation according to the table, the error quota that results from the temperature adaptation of the sample to the incubator temperature,
ba31114e07 08/2014 47
GLP/TOOLS OxiTop® Control OC 100
T
Incubator
, is smaller than 1% of the selected measurement range final
value.
Measuring time of the measure­ment
1 day 2 days T 3 days T 4 days T 5 ... 99 days T
Recommended sample temper­ature at the start of the mea­surement
T
Incubator-0.5K ... TIncubator Incubator-1K ... TIncubator Incubator-2K ... TIncubator Incubator-3K ... TIncubator Incubator-5K ... TIncubator
BOD5 15°C ... 20°C
The AutoTemp function The AutoTemp function is made up of the adaptation phase and the
test phase. Adaptation phase
The phase without evaluation of the pressure process. The duration of the adaptation phase is defined for the various measuring times according to the table. The adaptation of the microbiology to the char­acteristics of the sample is made in this phase and small temperature deviations, too high and too low temperatures of the sample, can be compensated for. The measuring system is zeroed after the adaptation phase has expired.
Sequence of the test
phase:
Test phase
The phase in which the further pressure process direction in the sample bottle is checked. The test phase is defined for the various measuring times according to the table. In this phase, the continuing temperature deviation can be compen­sated if the temperature of the sample is too low.
With a further drop in pressure (consumption) or constant pressure after the adaptation phase, the pressure value at the end of the adap­tation phase is the starting point of the measurement.
On a further increase of pressure following the adaptation phase (the sample is still too cold), the turning point of the pressure process at which the pressure increase changes into a pressure drop, is the starting point of the measurement. If the sample is too warm the test phase is dropped (temper the sample according to the table on the last page).
If no starting point is found (according to the procedure given in points 1 and 2) after the termination of the AutoTemp phase (time limit exceeded), the last measuring point of the AutoTemp phase forms the starting point.
This means that the BOD curve in the graphical evaluation does not emerge from the coordinate origin at the zero time point. A sample that was too cold was started.
48 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER
BOD measuring time
Duration of the adapta­tion phase
Duration of the test phase
0.5 to 23 hours Since the measurement times are very short, the sys­tem always suppresses the AutoTemp phase here even if the AutoTemp function is switched on in the set­tings.
1, 2, 3, 4, 5 days 14, 28, 42, 56, 70 minutes Max. 28, 56, 84, 112, 140
minutes
6 to 99 days
70 minutes Max. 140 minutes
in a 1 day pattern
Switch the AutoTemp function on or off (only in operating modes
BOD Standard):
SETTINGS menu Use <><> to select the -
AutoTemp menu item.
Preselected: - AutoTemp On To change it:
Confirm with <RUN/ENTER>. Use <><> to select either -
AutoTemp On or Off.
Confirm with <RUN/ENTER>.
-Back (to the AutoTemp menu)

7.4.8 Switch-off interval

The time interval after the last time a key is pressed can be set here. After this time interval expires, the controller switches itself of f to save energy. Possible settings: 5 to 15 minutes. Default setting: 5 minutes.
In BOD Routine mode the switch-off interval is permanently set to 5 min and cannot be changed.
SETTINGS menu: Use <><> to select the -
Switch-off interv. menu item.
ba31114e07 08/2014 49
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
Preset:
-Interval: 5 minutes To change it:
Confirm with <RUN/ENTER>. Set the required number of min­utes with <><>. Confirm with <RUN/ENTER>.
-Back (to the SETTINGS menu)

7.4.9 Language

®
Select the language here in which the displays of the OxiTop
control­ler appear. The controller has the following 5 languages stored in it (default German):
German - English - French - Italian - Spanish.
SETTINGS menu: Use <><> to select the -
Language menu item.
Preset:
-Mode German To change it:
Confirm with <RUN/ENTER>. Set the required language with <><>. Confirm with <RUN/ENTER>. From now on the displays appear in the selected language.
-Back (return to the SETTINGS
menu)
SETTINGS menu.
-Back (to the GLP/TOOLS
menu ) or with <GLP>.
50 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER
RUN/ENTER
40 cm to 1 m
RUN/ENTER

7.5 Check

Main menu GLP/TOOLS. Use <><> to select the -
Check menu item.

7.5.1 Showing the measuring heads

CHECK menu: Preselected:
-Show 
Point the controller at the measuring heads:
All the measuring heads that are addressed flash for 5 sec­onds.

7.5.2 Measuring head information

SHOW  menu: Preselected:
- Show free 
Use <><> to select either
- Free  or
-All 
.
CHECK menu: Select the - info menu item
with <><>.
ba31114e07 08/2014 51
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
The measuring head flashes. A display appears on the con-
troller giving the following infor­mation: Series number of the measur-
ing head,
Battery status
(OK/LOBAT/EMPTY!!),
the next calibration date (only
in operating modeBOD Standard)
Status of the measuring head
(free/used/defective).
If the measuring head is "used": Use <><> to scroll
through the display of further information:
– Start date – Final date – Sample number – Type and – Measuring range of the
measurement
Use <RUN/ENTER> to return
to the CHECK menu.
Repeat the procedure for each measuring head.
If the measuring head does not respond, the following display appears after approximately 7 seconds:
Query stopped!
You can select between
- Continue query (see above) and
-Stop (return to the CHECK
menu)
52 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER

7.5.3 Controller information

CHECK menu. Use <><> to select the -
Controller Info menu item.
The display shows the following information: Number of measuring heads
for which there is still memory capacity
Status of the supply batteries Status of the data backup
batteries (OK/LOBAT).
Use <><>
to scroll to the
specification of the
Software version Series number
(can differ from the example shown here)
Return to the CHECK menu: With <RUN/ENTER>.

7.5.4 cal-test

The cal-test function is used to test the sealing of the system measur ing head - rubber sleeve - sample bottle and the operability of the system
®
OxiTop
Control.
In the BOD Standard mode with the GLP On setting, the con troller indi­cates when the next cal test is due, depending on the set calibration interval (see section 7.4.4 GLP (O
PERATING MODE BOD STANDARD ),
page 45).
®
To perform the test, you need the WTW test resource, OxiTop
PM,
order number 209 333. Sample preparation: see operating manual of the OxiTop
®
PM test
resource.
CHECK menu. Use <><> to select the -Cal
Test menu item.
ba31114e07 08/2014 53
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
The instrument automatically allocates the sample number (in the header line). The filling volume (164 ml) and the type together with the run time of 5 days are preset.
For information on the further handling of the sample up to the “finished” status: See section 5.2 S
TARTING THE MEASUREMENT,
page 21. The Cal sample appears together with the other samples in the
SAMPLE STATUS TYPE list (sample management). The BOD type is CAL:
SAMPLE STATUS TYPE list (sample management)
Evaluation of the Cal
test
Use <><> to select the test sample.
Entry into the evaluation. Display of the selected sample as a curve with measurement value data.
Compare measurement value with lot test value (according to operating manual of OxiTop
®
PM) . Now you can set a new calibra-
tion date for the measuring head with the controller.
- Show  (to find the relevant measuring head): Function and messages as described (see section 5.4 SAMPLE STATUS TYPE
SAMPLE MANAGEMENT), page 26.
(
54 ba31114e07 08/2014
LIST
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER
40 cm to 1 m
- Set Cal date
This menu item only appears if the controller is in the
BOD Standard operating mode, GLP is switched on and the test
is finished.
-Stop: Return to the curve display. The Cal test is considered as not performed.
The controller sets a new test date in the measuring head. The controller calculates the new test date from the current date + the specified check interval (see section 7.4.5 GLP -
INTERVAL (- CALINTERVAL), page
CALIBRATION
46). The controller displays the set-
ting of the calibration date.
If the date was not set successfully (e.g. because the control­ler was not held to - or not held close enough to - the measur­ing head):
Repeat the procedure using
<RUN/ENTER>.
Then continue as described above.

7.5.5 Pneumatic test

The pneumatic test tests the measurement precision of the measuring head. It says nothing about the long-term impermeability of the system.
To perform the pneumatic test, you require the test resource, OxiTop
®
PT (WTW order number 209 334).
ba31114e07 08/2014 55
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
RUN/ENTER
The user interface of the controller guides you through the test:
CHECK menu. Use <><> to select the -
Pneumatic test  menu item.
Set the plunger of the syringe on
®
the OxiTop
PT test resource to 5 scale parts. Tightly screw the measuring head to be tested onto the Oxi-
®
Top
PT test resource.
(Exceeding the time causes a return to the CHECK menu).
(Exceeding the time causes a return to the CHECK menu).
The controller display shows the result of the pneumatic test.
Use <RUN/ENTER> to return to the CHECK menu.

7.6 Maintenance

GLP/TOOLS menu. Use <><> to select the -
Maintenance menu item.
56 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER
RUN/ENTER
RUN/ENTER
RUN/ENTER
RUN/ENTER
RUN/ENTER

7.6.1 Erasing finished samples

Here you can erase the data of finished samples that are already eval­uated or no longer required in order to free memory in the controller.
MAINTENANCE menu:
-Erase sample preselected
ERASE FINISHED SAMP.
menu:
Preselected: - From sample no.
-all
-Back (to the MAINTENANCE
submenu) The controller displays the list of
the finished samples. The oldest finished sample is marked.
With <>, you can mark further samples. With <>, you can remove the marking again.
The controller asks if you really want to erase the marked sam­ples from the memory.
After confirmation, the display message shown here appears for 2 seconds and then the con­troller returns to the menu ERASE FINISHED SAMP..
In the selection of erase “All“ samples, the following display appears:
The controller asks if you really want to erase the samples from the memory.
Continue: see above.
ba31114e07 08/2014 57
GLP/TOOLS OxiTop® Control OC 100
If no finished samples are available in the memory, the follow­ing display appears:
58 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
RUN/ENTER
RUN/ENTER

7.6.2 Reset/release

This function can be used to release measuring heads again that were unintentionally started.
After carrying out the - Reset/release  function the data of the measuring head are erased!
MAINTENANCE menu: Use <><> to select the -
Reset/release menu item.
The serial number of the mea­suring head and the sample number appear on the display.
- Reset/release
Confirm with <RUN/ENTER>. The release for the measuring head has been performed..
-Back: (measuring head not released)
Display message:
Reset performed!
Repeat the procedure for each measuring head to be released.
Display message when the last measuring head of a sample has been released:
Use <RUN/ENTER> to return to the RESET/RELEASE menu.
ba31114e07 08/2014 59
GLP/TOOLS OxiTop® Control OC 100
RUN/ENTER
RUN/ENTER
RUN/ENTER

7.6.3 Reading out data

In case you lost your controller or your controller is defective, the "Restore data" function facilitates to restore the data of running measurements using a new or other controller. This means the data are not lost!
To perform this function the memory of the controller mu st be abso­lutely empty! The data of each single measuring head are restored one after the other in a sequence. It is not possible to leave this function and to continue it afterwards because then the memory would no longer be empty.
Select the - Restore data menu item.
Select the -Restore data menu item.
Confirm the restoration for one measuring head and hold the controller to the measuring head. Contact selection:
The controller displays the serial number of the selected measuring head.
After confirming again the data of the measuring head are read out and stored in the controller.
The Id number of the sample is not restored, it is always I999.
60 ba31114e07 08/2014
OxiTop® Control OC 100 GLP/TOOLS
The controller returns to the RESTORE DATA menu. Perform this procedure for every measuring head with a mea­surement running, without leav­ing the RESTORE DATA function.
If the data of a measuring head were already restored the following is dis­played for approx. 3 seconds:
Then the controller returns to the RESTORE DATA menu. Perform the restoration with another measuring head.
If the memory of the controller is not empty the following is displayed for approx. 3 seconds:
After this, the RESTORE DATA menu is displayed again. It is not possible to restore the data with the controller. Contact the WTW service department.
It is possible to restore the data of all running measurements if the controller is in the BOD Routine operating mode. In the SAMPLE STATUS TYPE list (sample management), however, only the samples that were started in the BOD Routine operat ing mode are displayed in this case. To have all samples shown, switch over to another operating mode (see section 7 GLP/TOOLS, page 41).
ba31114e07 08/2014 61
Transmitting data OxiTop® Control OC 100

8 Transmitting data

The RS232 interface is used to transmit data to a PC (see section 4.1.2 S
OCKET FIELD - CONTROLLER, page 16).
For details, see operating manual of the Achat OC software.

8.1 Achat OC software

With the Achat OC software, you can Execute the sample management of the controller on the PC (back
up the data regularly!)
Comfortably display the sample management together with addi-
tional information on the screen
Select samples in the PC and transfer the measurement data of the
selected samples from the controller to the PC
Export the measurement data for processing with spreadsheet pro-
grams
For details, see operating manual of the Achat OC software.
62 ba31114e07 08/2014
OxiTop® Control OC 100 Maintenance, cleaning, disposal
LISTED 8F93 LaboratoryEquipment E163694
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: (1)This devise may not cause harmful interference, and (2) this devise must accept any interference received, including interference that may cause undesired operation.
Ser.-Nr.:05150013
OxiTop OC 1x0
WTW 82362 Weilheim
Made in Germany
1x CR 2430 Lithium
3xAlkaline, AA size, 1,5V
1
1
1
1
1
2

9 Maintenance, cleaning, disposal

9.1 Maintenance

9.1.1 General maintenance activities

The only maintenance activity required is replacing the batteries.

9.1.2 Replacing the supply batteries (controller)

®
Switch off the controller OxiTopLoosen the 4 screws underneath the housing using a Phillips screw-
driver.
-OC 100.
1Screws
Place the controller on the lower case.Remove the upper case and put it down to the right next to the lower
case with the display downwards.
1 Supply batteries in the battery holder (lower case) 2 Data backup battery (upper case)
ba31114e07 08/2014 63
Maintenance, cleaning, disposal OxiTop® Control OC 100
Remove the supply battery holder from the fixing in the lower case
and turn it around.
Remove the empty supply batteries.Insert the new supply batteries - 3 pieces, alkaline (alkali manga-
nese), Size: Mignon, AA, AM3, LR6).
NOTE
Ensure that they are inserted in the correct position! (The poles are marked in the supply battery holder.) Always replace the complete set of batteries.
Turn the supply battery holder around again and place it in the fixing
in the lower case.
Set the upper case on top of the lower case, turn the controller and
tighten the housing screws using the screwdriver.
Switch on the controller. The battery change was successful if no
error message appears concerning the supply batteries.
It is recommended to use only brand name batteries of the type
specified as “Alkaline“. Using other types of battery can adversely affect the reliable functioning.
The supply batteries have no influence on data integrity.
Dispose of used batteries according to the local regulations of your country.
End users within the European Union are obligated to return used batteries (even ecologically compatible ones) to a collection point set up for recycling purposes.
Batteries are marked with the crossed-out waste container symbol. Therefore, they may not be disposed with the domestic waste.
64 ba31114e07 08/2014
OxiTop® Control OC 100 Maintenance, cleaning, disposal
LISTED 8F93 LaboratoryEquipment E163694
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: (1)This devise may not cause harmful interference, and (2) this devise must accept any interference received, including interference that may cause undesired operation.
Ser.-Nr.:05150013
OxiTop OC 1x0
WTW 82362 Weilheim
Made in Germany
1x CR 2430 Lithium
3xAlkaline, AA size, 1,5V
1
1
1
1
1
2

9.1.3 Replacing the data backup battery (controller)

Evaluate and back up all the measurement data.
Options available: – print out the results (see section 6 P
RINTING, page 39)
– save the data in the PC using the software “Achat OC“, see sec-
tion 8.1 A
CHAT OC SOFTWARE, page 62
– handwritten documentation of the results.
Documentation of your instrument settings:
printout of your instrument settings, see section 7.3 S
HOW SETTINGS,
page 42.
Switch off the controller OxiTop
®
-OC 100.
Loosen the 4 screws underneath the housing using a Phillips screw-
driver.
1Screws
Place the controller on the lower case.Remove the upper case and put it down to the right next to the lower
case with the display downwards.
1 Supply batteries in the battery holder (lower case) 2 Data backup battery (upper case)
ba31114e07 08/2014 65
Maintenance, cleaning, disposal OxiTop® Control OC 100
Remove the empty data backup battery.Insert the new data backup battery (1 piece, lithium battery,
CR2430, e.g. WTW type Batt/ OxiTop
®
).
NOTE
Ensure that they are inserted in the correct position! The positive pole must point upwards.
Set the upper case on top of the lower case, turn the controller and
tighten the housing screws using the screwdriver.
Switch on the controller. The message Please Set system clock!
appears.
This message is a reminder and should be acknowledged by press­ing <RUN/ENTER>.
The battery change was successful if no error message appears
concerning the data backup battery.
Enter the current date/time and your settings.
Removing the data backup battery and the supply battery simulta­neously will erase all data (instrument reset)!
Dispose of used batteries according to the local regulations of your country.
End users within the European Union are obligated to return used batteries (even ecologically compatible ones) to a collection point set up for recycling purposes.
Batteries are marked with the crossed-out waste container symbol. Therefore, they may not be disposed with the domestic waste.
66 ba31114e07 08/2014
OxiTop® Control OC 100 Maintenance, cleaning, disposal

9.1.4 Changing the supply battery (measuring head)

Change the battery after the measurement has been terminated.Press in the snap-action holder and open the upper case.
Remove the batteries.
Insert new batteries (2 x lithium batteries, CR2430, e.g. WTW type
Batt/OxiTop
®
). When doing so make sure the batteries are inserted
in the correct position!
Insert the upper case with the lug in the locking pin (lower case).
Caution! Do not crush the cable connection!
Close the upper case (let the hook snap into place). Check if the battery change was successful:
In the GLP/TOOLS -Check - info menu, Batteries: OK must appear. The batteries do not affect the data integrity of the OxiTop
®
measuring
head.
Dispose of used batteries according to the local regulations of your country.
End users within the European Union are obligated to return used batteries (even ecologically compatible ones) to a collection point set up for recycling purposes.
Batteries are marked with the crossed-out waste container symbol. Therefore, they may not be disposed with the domestic waste.
ba31114e07 08/2014 67
Maintenance, cleaning, disposal OxiTop® Control OC 100

9.2 Cleaning

9.2.1 Cleaning of the sample bottles

See application reports.

9.2.2 Cleaning of the controller and measuring heads

Occasionally wipe the outside of the measuring instrument with a damp, lint-free cloth.
NOTEThe housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain solvents. Remove any splashes immediately.

9.3 Disposal

At the end of its operational lifetime, the instrument must be returned to the disposal or return system statutory in your country. If you have any questions, please contact your supplier.
68 ba31114e07 08/2014
OxiTop® Control OC 100 What to do if...

10 What to do if...

10.1 Display messages

Display message Cause Remedy
already used! The selected measuring head has
already been started for a measure­ment and can only be used for a new measurement after the current mea­surement is finished and the mea­surement data completely retrieved.
Select a new and free measuring
head or
-Stop start orRelease the measuring head (sec-
tion 7.6.2 R
ESET/RELEASE, page 59)
Please Set system clock! Data backup battery was changed. Set the clock (see section 7.4.3
D
ATE/TIME, page 45).
defective! The selected measuring head is
defective. Please return it to WTW.
No samples available!
There are no finished samples stored in the sample management.
Select a new and free measuring
head or
-Stop startTransfer the unfinished samples of
the sample management to “fin­ished” status. To do this, the measurement data of all measuring heads must be called up and evaluation and docu­mentation of the measurement data of the finished samples must be performed.
If the measurement data of the
measuring heads are not yet fin­ished, no instant remedy is possi­ble! The memory capacity limit of the instrument has been reached.
ba31114e07 08/2014 69
What to do if... OxiTop® Control OC 100
Display message Cause Remedy
No active !  No measurement has been
started.
The measuring heads are placed
at another location (e.g. other lev­els in the incubator, other incuba­tors, further storage locations).
The measuring heads have no
optical contact with the controller: – Distance too great, angular
position imprecise
– IR window of the measuring
head averted from the control­ler
– Shading by other object.
Controller defective
cal test due on
18.07.98! (sample date)
GLP is switched onThe test resource monitoring is
due within the measuring time of measurement for the selected measuring head.
Modification of the distances or
angle
Search at other storage locationsCheck the controllerCheck the measuring heads (see
section 10.3 G
ENERAL INFORMATION,
page 75)
Use another free measuring head
that has been checked
Start the due measuring head with
the Cal test (see section 7.5.4
TEST, page 53)
CAL-
Lack of memory possi­ble! Memory for 11 free ! Continue
Date/time is not correctly set in the
controller.
This message is used as information in the operating mode. There are a lot of sample data stored in the sample management. The memory is almost full. It is still possible to start a sample as a parallel sample start with 11 mea­suring heads (the number 11 used here is an example, possible num­bers: 1...11 ).
Switch off GLP and start the mea-
suring head (if your laboratory prac­tice permits this).
Set the date/time in the controller. Problem solution, if necessary:
Erase finished samples
section 7.6.1 E
SAMPLES, page 57
RASING FINISHED
Alternative: Changing the instru-
ment setting to automatic erasing of finished samples: section 7.4.6 D
ATA STORAGE, page
47
70 ba31114e07 08/2014
OxiTop® Control OC 100 What to do if...
Display message Cause Remedy
Memory lack! No ready samples autom. erasable!
Memory lack! Erase finished sample/s!
The instrument works with the set­ting, Erase memory auto (automati­cally). The memory is full and only unfinished samples are stored in the sample management. The instrument can only erase fin­ished samples automatically.
The instrument works with the set­ting, Erase memory "manually". The memory is full.
Transfer the unfinished samples of
the sample management to “fin­ished” status. To do this, the measurement data of all measuring heads must be called up and evaluation and docu­mentation of the measurement data of the finished samples must be performed.
If the measurement data of the
measuring heads are not yet fin­ished, no instant remedy is possi­ble! The memory capacity limit of the instrument has been reached.
Transfer the unfinished samples of
the sample management to “fin­ished” status:
Call up the measurement data of all
measuring heads
Perform the evaluation and docu-
mentation for the measurement data of the finished samples
Erase finished samples from the
sample management: section 7.6.1 E
SAMPLES, page 57
RASING FINISHED
Note: Only the finished samples in the sample management can be erased.
ba31114e07 08/2014 71
What to do if... OxiTop® Control OC 100
Display message Cause Remedy
undef.
(Display in the curve pre­sentation or sample sta­tistics)
10 of 12
called up
The selected measurement range
has been exceeded.
The sample filled was too warm.The AutoTemp function is
switched off.
The sample filled was very cold
and has a low consumption behavior (small BOD value).
The system is not sealed (bottle
internal pressure = atmospheric pressure).
Two measuring heads have not
been found by the controller
Select the correct measurement
range (see application reports on the CD-ROM, Principles of measur­ing technique).
Pre-temper the sample more pre-
cisely.
Switch on the AutoTemp function
(only effective with run durations of more than 1 day).
Check the system for leaks by
means of a visual check: Are the sealing surfaces of the bot­tle and measuring head clean and fault-free? Are there cracks in the sleeve, bottle or measuring head?
Perform a cal-test (see section
7.5.4
CAL-TEST, page 53).
Check the incubator temperature.Modification of the distances or
angle.
(sample message, no. is variable)
The measuring heads are placed
at another location (e.g. other lev­els in the incubator, other incuba­tors, further storage locations)
The measuring heads have no
optical contact with the controller:
- Distance too great, angular posi­tion imprecise
- IR window of the measuring head averted from the controller
- Shading by other objects
The missing measuring heads are
defective.
Search at other storage locations.Search for and check measuring
heads (see section 10.3 G
INFORMATION, page 75)
ENERAL
72 ba31114e07 08/2014
OxiTop® Control OC 100 What to do if...
Display message Cause Remedy
0 of 19
called up
(sample message, no. is variable)
No measuring head was found by
the controller.
The measuring heads are placed
at another location (e.g. other lev­els in the incubator, other incuba­tors, further storage locations)
The measuring heads have no
optical contact with the controller. – Distance too great, angular
position imprecise
– IR window of the measuring
head averted from the control­ler
– Optical shading by other
objects
The controller is defective.
Modification of the distances or
angle
Search at other storage locationsCheck the controller (see section
10.3 G
ENERAL INFORMATION, page
75)
ba31114e07 08/2014 73
What to do if... OxiTop® Control OC 100
10.2 Display messages on the charging status of the bat­teries

10.2.1 Batteries in the controller

Display message Notes/ explanation
The warning appears for approx. 3 seconds. The instrument then continues to run normally. The instrument can still be safely operated within the specifications. When the message first appears, there is still a running reserve available. Please obtain new batteries and replace the old ones!
The message appears for approx. 3 seconds. The instrument then switches itself off. The instrument can no longer be used. The supply batteries must be replaced by new ones.

10.2.2 Battery in the measuring head

Display message Notes/ explanation
The instrument can still be safely operated within the specifications. The running reserve of the instrument is still suffi­cient for the measurement to be started. Please obtain new batteries and replace the old ones!
The instrument can no longer be used. The supply batteries must be replaced by new ones. It is possible to start a new measuring head or to complete a sample start.
74 ba31114e07 08/2014
OxiTop® Control OC 100 What to do if...

10.3 General information

Cause Remedy
No or missing sample entries in the sample management
At the beginning of the measurement, no mea­surement curves are dis­played Measurement curves do not emerge from the ori­gin
No measurement curves are displayed although the measurement has already been running for an extended period of time
Search for free measur­ing heads for a new mea­surement
The BOD Routine operating mode
is switched on. Samples were started in the other operating modes.
The sample filled and started was
too cold.
The AutoTemp function is switched
off.
No data were retrieved from the
measuring heads.
The AutoTemp phase is still run-
ning (see section 7.4.7 A
UTOTEMP,
page 47).
Switch on the BOD Standard oper-
ating mode section 7.4.1 O
PERATING MODE,
page 43
Temper the sample more precisely.Switch on the AutoTemp function
(only effective with run durations of more than 1 day).
Call up the data of the measuring
heads. See section 5.3 C
DATA, page 24.
ALLING UP ALL
Wait for the end of the AutoTemp
phase then start the evaluation.
Perform the - Show free  function (see section 7 GLP/TOOLS, page
41). The controller causes the free measuring positions to flash for 5 seconds.
Measuring head uninten­tionally started for mea­surement
Measuring head started with incorrect settings
Measuring head is required for another sam­ple
Search for a defective measuring head
A measuring head is missing or is defective.
Requirement: To determine the corre­sponding sample
The measuring head can be released again with the command Reset/release: see section 7.6.2 R
ESET/RELEASE, page 59.
The controller guides you further through the operation. If only one measuring head is stored under the relevant sample number (in the
BOD Routine operating modes, this is always the case), the controller auto­matically erases the sample in the sample management.
Perform a measuring head reaction test (see above). A measuring head that repetitively shows no reaction to the test is defective. If the defective measuring head cannot be established in this way, perform the measuring head check (see “Single check”). In doing this, each individual measuring head must be checked until the defective measuring head has been found.
Call up the function “Call up data” from the sample management for each indi­vidual unfinished sample. The corresponding sample can be determined by this. For subsequent error handling, see section 5.4.4 C
ALLING UP DATA, page 29.
ba31114e07 08/2014 75
What to do if... OxiTop® Control OC 100
Cause Remedy
Which measuring head belongs to which mea­surement curve?
The precision and sealing of the system sample bot­tles plus measuring head should be checked
Measuring head check (single check)
(To which running sam­ple does the selected measuring head belong?)
Prepare the relation separate curve – number of measuring head; see sec-
tion 5.5 E
VALUATION, page 33.
Read number of measuring head out of headlineHave measuring head shown; see section 5.4.1 S
HOWING A SAMPLE, page
27
see section 7.5.4
CAL-TEST, page 53
Single check method: Menu GLP/TOOLS -Check - info
The controller guides you further through the operation.All single information on the state of the measuring heads is listed, i.e. the
measuring head responds.
The check can be undertaken at any time without affecting the measure-
ment that is running.
If no reaction at all can be determined, the batteries should be changed and
the measuring head check repeated.
The measuring precision cannot be checked with this!
The measuring precision of the measuring head should be checked.
Perform measuring head reaction test
See section 7.5.4
CAL-TEST, page 53 or section 7.5.5 PNEUMATIC TEST, page 55
Menu GLP/TOOLS - Check -All :All optically attainable measuring heads must flash for approx. 5 seconds
independent of their status.
The check can be undertaken at any time without affecting the measure-
ment that is running.
This test only checks the reaction to commands.If no measuring head reacts, the controller should be checked.
76 ba31114e07 08/2014
OxiTop® Control OC 100 What to do if...
Cause Remedy
Check of the controller Controller information
See section 7.5.3 C
ONTROLLER INFORMATION, page 53
All single information on the state of the controller is listed. Th e check can be undertaken at any time without affecting the measurements that are run­ning.
Perform the command: -All . See GLP/TOOLS - Check - Show .
(The check is used in this case to test the IR interface.) All working measur­ing heads must flash for 5 seconds.
The controller does not react / is defective – the data of the running mea­surements should be restored
Incorrect time display on being switched on
Perform a measuring head check (see section 7.5.2 M
INFORMATION, page 51). (The check is used in this case to test the IR inter-
EASURING HEAD
face.) Here, a working measuring head should supply its status data.
Keyboard, display and signal tone should show the required reaction.Check the function of the clock (prerequisite: Clock is set).
Switch off the instrument and switch it on again. Time and date each appear showing the current values.
Check RS232 interface (only required when using the WTW software,
ACHAT OC): – Connect the controller to your PC by means of the interface cable
AK540/B.
– Switch on the controller and start the PC program, ACHAT OC. In doing
so, the PC program checks the RS232 interface.
Send the controller to WTW. Using a repaired or new controller (memory must be empty!) perform the “Restore data” procedure (see section 7.6 M
NANCE, page 56).
Data backup battery was changed.Summer/winter time change has
Set the clock (see section 7.4.3 D
ATE/TIME, page 45).
AINTE-
taken place.
ba31114e07 08/2014 77
What to do if... OxiTop® Control OC 100
Cause Remedy
The IR printer does not react
The printer is not switched onThe printer has no optical contact
with the controller: – Distance too great, angular posi-
tion imprecise
– IR window of the printer averted
from the controller
– Shading by other objects.
Printer batteries are empty. No paper or the wrong paper is
loaded.
Printing is not possible in the oper-
ating state of the controller selected.
The printer or the controller is
defective.
Switch on the printerEstablish optical contactCheck or change the printer batter-
ies. Please read the operating instruc­tions of the printer.
Check in the operating manual of
the controller whether printing is possible in the state selected. Note: During printing, the message Printing active ... always appears on the display
78 ba31114e07 08/2014
OxiTop® Control OC 100 Technical data

11 Technical data

11.1 Controller OxiTop® OC 100

Measuring ranges 0 ... 40 / 80 / 200 / 400 / 800 / 2000 / 4000 mg/l BOD Run durations and data
records Per measurement 0.5 hours 180
Display Memory capacity Data of 120 measuring heads Power supply Supply batteries: 3 pieces, alkaline (alkali manganese), size:
Measuring time of the measure­ment Datasets
1.5 hours 270
2.5 hours 300
3.5 hours 315
4.5 hours 324
5.5 hours 330 All whole hours
(1, 2, 3 to 23 hours) All whole days
(1, 2, 3 to 99 days) LCD graphic display 64 x 128 pixels
Mignon, AA, AM3, LR6 Data backup battery: 1 piece lithium battery, CR2430, WTW order no. 209 012
360
360
Battery runtime
Supply battery Data backup battery
Protective class 3, instrument safety: N 61010 -1 Type of protection IP 41 EMC EN 61326-1, FCC Class A Ambient temperature Storage: -25°C ... +65°C
Dimensions 45 x 100 x 200 mm (H x W x D) Weight Approx. 390 g Test certificates CE
> 1000 start-ups in case of normal use (the instrument has an automatic shut off) typically 4 years
Operation: +5°C ... +40°C
ba31114e07 08/2014 79
Technical data OxiTop® Control OC 100

11.2 Measuring head OxiTop®-C or OxiTop®-C/B

Measuring principle Piezoresistive pressure sensor Operation pressure
range Maximum allowed pres-
sure Accuracy ± 1% of the measured value ± 1 hPa Resolution 1 hPa Display LED pilot lamp Power supply 2 x lithium battery CR2430 Battery runtime Typically 2 years, in case of normal use Protective class 3, instrument safety: EN 61010-1 Type of protection IP 54 EMC EN 61326-1, FCC Class A Ambient temperature Storage: -25°C ... +65°C
Dimensions H: 70 mm, Weight 95 g
500 ... 1350 hPa (mbar).
2000 hPa (mbar)
Operation: +5°C ... +50°C
Ø 70 mm
Test certificates CE
80 ba31114e07 08/2014
OxiTop® Control OC 100
A
Application reports . . . . . . . . . . . . . . . . . . . . 8
AutoTemp . . . . . . . . . . . . . . . . . . . . . . . . .47
B
Batteries
Exchange . . . . . . . . . . . . . . . . . . . . . . . 63
insert (controller supply batteries) . . . . 12
BOD
Routine . . . . . . . . . . . . . . . . . . . . . . . . . 21
Standard . . . . . . . . . . . . . . . . . . . . . . . 21
C
Cal-Test . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Course of the measurement . . . . . . . . . . . 19
G
GLP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
GLP - calibration interval . . . . . . . . . . . . . . 46
GLP/TOOLS
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Scope of delivery . . . . . . . . . . . . . . . . . . . . 11
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Show settings . . . . . . . . . . . . . . . . . . . . . . 42
Socket field - controller . . . . . . . . . . . . . . . 16
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Switch-off interval . . . . . . . . . . . . . . . . . . . 49
T
Technical data . . . . . . . . . . . . . . . . . . . . . . 79
Transmitting data . . . . . . . . . . . . . . . . . . . . 62
I
Initial commissioning . . . . . . . . . . . . . . . . . 12
K
Keys - controller . . . . . . . . . . . . . . . . . . . . . 15
L
Language . . . . . . . . . . . . . . . . . . . . . . . . . . 50
M
Menus and dialogs . . . . . . . . . . . . . . . . . . . 18
O
Operating modes . . . . . . . . . . . . . . . . . . . .17
BOD Routine . . . . . . . . . . . . . . . . . . . . 21
BOD Standard . . . . . . . . . . . . . . . . . . . 21
P
Pneumatic test . . . . . . . . . . . . . . . . . . . . . . 55
Power supply . . . . . . . . . . . . . . . . . . . . . . .11
Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
R
Reading out data . . . . . . . . . . . . . . . . . . . . 60
S
Safe operation . . . . . . . . . . . . . . . . . . . . . . . 9
ba31114d07 08/2014 81
OxiTop® Control OC 100
82 ba31114d07 08/2014
What can Xylem do for you?
We're a global team unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. We move, treat, analyze, and return water to the environment, and we help people use water efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to xyleminc.com.
®
Service address:
Xylem Analytics Germany Sales GmbH & Co. KG WTW Dr.-Karl-Slevogt-Str. 1 82362 Weilheim Germany
Tel.: Fax: E-Mail
Internet:
+49 881 183-325 +49 881 183-414 wtw.rma@xyleminc.com
www.WTW.com
Xylem Analytics Germany GmbH Dr.-Karl-Slevogt-Str. 1 82362 Weilheim Germany
Loading...