Xylem Magflux User Manual

Electromagnetic Flow Meter
Electromagnetic Flow Meter
Installation and User Manual
1EN 3.05 MagFlux Manual 1212
FW: 842018/842009
Electromagnetic Flow Meter
Declaration of Conformity
Konformitetserklæring
Vi, MJK Automation A/S, DK-2850 Nærum, påtager os det fulde ansvar for at produktet
Declaration of Conformity
We, MJK Automation A/S, DK-2850 Nærum, declare under our sole responsibility that the product
som denne erklæring angår, er i overensstemmelse med følgende standard(er) eller andre normdokument(er).
EN61000-6-3/-4:2001, EN61000-6-1/-2:1999
efter bestemmelserne i direktiv
89/336/EEC; 92/31/EEC
Declaration de conformite
Nous, MJK Automation A/S, DK-2850 Nærum, déclarons sous notre seule responsabilité que le produit
to which this declaration relates is in conformity with the following standard(s) or other normative document(s).
EN61000-6-3/-4:2001, EN61000-6-1/-2:1999
following the provisions of Directive
89/336/EEC; 92/31/EEC
Dichiarazione di conformità
Noi, MJK Automation A/S, DK-2850 Nærum, dichiariamo sotto la nostra esclusiva respon­sabilità che l’apparecchio
Konformitätserklärung
Wir, MJK Automation A/S, DK-2850 Nærum, erklären in alleiniger Verantwortung, dass das Produkt
auf das sich diese Erklärung bezieht mit der/den folgenden Nor me(en) oder normativen Dokument(en) übereinstimmt.
EN61000-6-3/-4:2001, EN61000-6-1/-2:1999
Gemäss den Bestimmungen der Richtlinie
89/336/EEC; 92/31/EEC
Declaración de Conformidad
Nosotros, MJK Automation A/S, DK-2850 Nærum, declaramos bajo nuestra única responsabili­dad que el producto
auquel se réfère cette déclara­tion est conforme á la (aux) norme(s) ou autre(s) document(s) normatif(s)
EN61000-6-3/-4:2001, EN61000-6-1/-2:1999
conformément aux dispositions de Directive
89/336/EEC; 92/31/EEC
al quale questa dichiarazione si riferisce, è conforme alla seg­uente normativa(e) standard o ad altri documenti di normativa(e)
EN61000-6-3/-4:2001, EN61000-6-1/-2:1999
conformemente alla disposizioni della Direzione
89/336/EEC; 92/31/EEC
21.11.2006
Jens Kruse
2EN 3.05 MagFlux Manual 1212
al cual se refiere esta de­claración, está en conformidad con la(s) siguente(e) norma(s) u otros documentos normativos
EN61000-6-3/-4:2001, EN61000-6-1/-2:1999
según las disposiciones de la(s) directiva(s)
89/336/EEC; 92/31/EEC
Electromagnetic Flow Meter
Table of Contents
1. Introduction 7
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Operating Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
The MagFlux
®
Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2. Safety, Repair and Product Identification 13
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Physical Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Product Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3. Flow Sensor 15
Minimum conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Liner selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Electrode selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Sizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Sizing Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Pipe System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Cabinet Mounted on the Flow Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Potential Equalization and Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Type 7100/7200/7600 in Conductive Pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Type 7100/7200/7600 in Non-conductive Pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Type 7300/7400 in Conductive Pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Type 7300/7400 in Non-conductive Pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
4. Flow Converter 25
Electrical Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Changing the Power Supply Voltage 230/115 V AC . . . . . . . . . . . . . . . . . . . . . . . . .27
Analogue Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Digital Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Local (compact) Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Remote Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Converter Connection Board - Local . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Converter Connection Board - Remote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
5. System Configuration Examples 33
Compact Converter and Display Unit on Flow Sensor. . . . . . . . . . . . . . . . . . . . . . . .33
Remote Converter with Connection Box on Flow Sensor . . . . . . . . . . . . . . . . . . . . . .34
Wiring Schematic #1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
3
FW: 842018/842009
Electromagnetic Flow Meter
Remote Display and Multiple Converters Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Wiring Schematic #2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
6. Startup 39
Initial Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Initial Flow Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Flow Direction Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Language Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Display Read-out, one connected unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Display Read-out, several connected units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Display Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
7. MagFlux Menus 47
Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Specify Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Factory Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Graph Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Set Sensor Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Converter Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Minimum Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
mA Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Flow Forward . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
Flow Reverse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Forward & Rev. (12mA=0) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Forward & rev. (4mA=0) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Not in use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Settings and Limits for Resettable Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Resettable Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Resettable Forward Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Resettable Reverse Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Resettable Sum Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Non-Resettable Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
NR Forward Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
NR Reverse Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
NR Sum Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Batch Counters 1 & 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Settings and Limits for Batch Counters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
High Flow/Low Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Flow Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Empty Pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
4EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
24h Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
Display Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Set Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
Modbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Factory Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
Display SW Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
8. Mechanical Dimensions 101
Standard Flanges - EN-1092-1:2001 - Europe . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Standard Flanges - ANSI B 16.5 - U.S.A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Standard Flanges - AS-4087-2004 - Australia . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Standard Flanges - AS-4087-2004 (cont'd) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Standard Flanges - AS-4087-2004 (cont'd) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Appendix A. Pop-up and Error Messages 107
Appendix B. MJK-Field Link Software 111
Appendix C. FAQs - Frequently Asked Questions 115
Appendix D. Front Panel Cut-out Drawing 118
Appendix E. Test Certificate 119
Appendix F. Log Files 120
Appendix G. Digital Input/Output Connections 123
Appendix H. Remote Slave Display Unit 124
Connection Board - with a remote "slave" display unit . . . . . . . . . . . . . . . . . . . . . .124
Converter Connection Board - with a "master" display unit . . . . . . . . . . . . . . . . . . . 125
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Configuration for the master unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Configuration for the slave unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Main Menu Overview 126
Converter Setup Menu Overview 128
Service Menu Overview 138
Display Setup Menu Overview 142
5
FW: 842018/842009
Electromagnetic Flow Meter
6EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
1. Introduction
Thank you for choosing the MJK MagFlux® electromagnetic flow meter. We have done our utmost to design and manufacture a flow meter that satisfies your requirements.
The MagFlux® is suitable for flow measurement in all kinds of conduc­tive fluids, and it is especially suited for flow measurement of water, waste water, sludge and other fluids containing particles.
The flow meter is easy to install and put into service. However, read this manual first to learn about the MagFlux® electromagnetic flow meter and all its features.
You can always contact your local representative or the MJK hot lines for advice and guidance:
Europe Tel.: +45 45 56 06 56 E-mail: mjk@mjk.com
Denmark Tel.: +45 45 56 06 56 E-mail: mjk@mjk.dk
Norway Tel.: +47 69 20 60 70 E-mail: mjk@mjk.no
Sweden Tel.: +46 53 31 77 50 E-mail: kontoret@mjk.se
Holland Tel.: +31 251 672171 E-mail: mjknl@mjk.com
USA Tel.: +1 847 482 8655 E-mail: mjkusa@mjk.com
Australia Tel.: +61 3 9758 8533 E-mail: mjkaus@mjk.com
Visit our web sites at www.magflux.dk and www.mjk.com to learn more about MJK Automation, our other products and the people behind them.
MagFlux
®
is a registered trademark of MJK Automation A/S, Denmark.
Trademarks & acknowledgements
Mµ Connect, Connect, MagFlux, Chatter, SuSix, Oxix, pHix and Shuttle are ® registered trademarks of MJK Automation A/S, Denmark.
Expert and Instrument Link are ™ trademarks of MJK Automation A/S, Denmark.
VLT® is a registered trademark of Danfoss A/S. All other trademarks are the property of their respective owners.
MJK Automation A/S is a Xylem Brand.
7
FW: 842018/842009
Electromagnetic Flow Meter
About this Manual
The manual is divided into a table of contents, eight chapters, seven ap­pendices and an index.
1. Introduction
Contains a presentation of the MJK MagFlux
meter, the structure of this manual and the operating principles.
2. Safety, Repair and Product Identification
Provide answers to issues regarding safety, mounting, repair,
restrictions and product identification.
3. Flow Sensor
Describes the physical specifications and installation guidelines for the
flow sensor such as liner and electrode selection, example pipe systems, sizing chart, etc.
4. Flow Converter
Describes the physical specifications and installation rules for the flow
converter such as power supply, in- and outputs, sensor/converter/ display configurations, etc.
5. System Configuration Examples
Illustrates how MagFlux
®
sensors, converters and display units can be
interconnected in real-life scenarios.
®
electromagnetic flow
6. Startup
Describes important checks and basic settings to get started including
the display and keyboard user interface.
7. MagFlux Menus
Contains a comprehensive description of the MagFlux
®
menus,
options and utilities.
8. Mechanical Dimensions
Lists the flange and sensor dimensions and specifications.
* * * * *
8EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Appendix A. Pop-up Events / Error Messages
Lists possible pop-up messages, explains their meaning and offer
solutions to error conditions.
Appendix B. MJK-Field Link Software Upgrade
Describes in detail how to utilise the unique and intuitive MJK-Field Link
software program to upload new display and converter firmware versions.
Appendix C. FAQs - Frequently Asked Questions
Contains answers to questions that are often asked.
Appendix D. Front Panel Cut-out Drawing
A 1:1 scale drawing of the front panel outline and cut-out area for
installation and mounting purposes.
Appendix E. Test Certificate
Shows an example test certificate which is part of MagFlux
shipments.
Appendix F. Log Files
An example log file illustrates the format and explains the entries.
Appendix G. Digital Input/Output Connections
Illustrates how the digital in- and outputs can be interconnected.
®
* * * * *
Main Menu Overview
Is a continuous presentation of the main menu structure. The size of
this manual does not provide sufficient space for showing the complete menu structure in one image.
Converter Setup Menu Overview
Is a continuous presentation of the converter setup menu structure.
Service Menu Overview
Is a continuous presentation of the service menu structure.
Display Setup Menu Overview
Is a continuous presentation of the display setup menu structure.
9
FW: 842018/842009
Electromagnetic Flow Meter
Operating Principles
The electromagnetic flow meter is an instrument for measuring the flow of conductive fluid using Faraday's electromagnetic induction law, and consequently the fluid must be electrically conductive.
As illustrated below a magnetic field with density B - perpendicular to the direction of flow - stretches across a fluid flowing within an electrically isolated pipe.
The magnetic flux will induce a voltage difference (E) that can be measured between two electrodes arranged perpendicular to the direction of flow and the magnetic field. The voltage is proportional to the velocity (V) of the fluid.
E = B x D x V x k [Volt] where E is the voltage that is induced between the two electrodes B is the magnetic flux density D is the distance between the two electrodes V is the fluid velocity k is a dimensionless constant The flow of the fluid Q (m3/s) is given by the following formula:
Q = π x D2 x V / 4 => V = 4 / (π x D2) where
π : is the constant pi (= 22/7) D: is the internal diameter V: is the fluid velocity The combination of above formulas and gives the following formula:
E = K x B x D x Q x 4 / (π x D x D) = 4 K x B x Q / (π x D)
Evidently the voltage E is proportional to the actual flow.
10EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
The MagFlux® Flow Meter
Measurement
Electromagnetic flow meters may have either direct current (DC) or alter­nating current (AC) excitation.
When systems are AC excited, electrostatic and/or electromagnetic noise may influence the measurements. A DC excited system however, is less sensitive to noise produced by electromagnetic induction, simply be­cause the voltage induced over the electrodes will be a DC voltage. The drawbacks are measurement errors due to electro-chemical polarization between the electrodes and the fluid.
The MagFlux® electromagnetic flow meters are excited by a 2.5 Hz square wave and thereby eliminate the drawbacks of both DC and AC systems.
A microprocessor receives and measures the amplitude of the pulse with a 16-bit resolution, and converts and displays the result on the display unit.
Electrode Cleaning
Electrode cleaning is always in service to maintain clean and accurate electrodes. The operating principle is a 55 Hz AC voltage superimposed on the normal 2.5 Hz square wave. This step effectively reduces the risk of a dirt layer build-up on the electrodes that eventually will deteriorate and finally isolate electrical contact to the flow media.
11
FW: 842018/842009
Electromagnetic Flow Meter
This page intentionally left blank.
12EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
2. Safety, Repair and Product Identification
Safety instructions
1. Read this manual carefully.
2. Pay attention to the environment on the installation site.
3. Wear necessary protective equipment and follow all current safety regulations.
4. The MagFlux Always ensure that connected machinery and other equipment are effectively put out of service (that is to remove the main fuses and lock main and security switches in off-position) before commencing configu­ration, fault finding, servicing, maintenance work, etc.
5. WARNING: There is a risk of lethal, electrical shock from "Mains sup­ply" terminals N and L. Be careful not to touch these terminals while the MagFlux
Physical Mounting
The MagFlux® flow converter/flow meter must not be mounted in explosion hazardous areas!
Repair
Repair must only be made by MJK or by a service representative ap­proved by MJK.
®
can invoke a start signal for dangerous machinery.
®
is being serviced.
13
FW: 842018/842009
Electromagnetic Flow Meter
Product Identification
A delivery will usually consist of a MagFlux® converter and a MagFlux® flow sensor. Check that the item(s) delivered corresponds to the ordered item(s). The part number and the calibration code are printed on an iden­tification label sticked onto the flow converter shipping box and on a label on the flow converter itself. The part number, the serial number, the cali­bration code and the electrode/lining data for the flow sensor are printed on a green name plate (see example below).
This calibration code (Cal. code) is unique and provides the MagFlux® converter with information about flow sensor number, nominal diameter of the flow sensor and calibration data for the flow sensor.
The current converter firmware requires an 8-character input, but it also accepts 6 characters plus two "Ok"s. The example above (dw5uq4) would require that you enter: d w 5 u q 4 "Ok" "Ok", where "Ok" indicates that you press the "Ok" key without selecting any character. Only small letters can be entered from the MagFlux® keypad.
If changes are needed to the unique sensor calibration code following initial setup, this is the calibration code that the "Sensor Calibration Code" menu must contain (see pages 44 and 86).
14EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
3. Flow Sensor
The following conditions must be satisfied to get the full benefit of the
MagFlux® flow sensor.
Minimum conductivity
• The conductivity of the media must be greater than 5 µS/cm.
Liner selection
• Use Teflon® lining for chemicals and food industries
• Use hard rubber lining for drinking water and waste water
• Use soft rubber lining for water with abrasive particles
Electrode selection
Steel AISI 316TI-1.4571 for general purpose, sewage, water and district heating systems
Hastelloy for sea water
Titanium and platinum for chlorine and other aggressive chemicals
Mounting location
• To obtain a stable and accurate flow measurement, it is very important that the flow sensor is mounted correctly in the pipe system
• There must be no flow fluctuations
• Avoid locations where vacuum can occur; especially for flow sensors with Teflon™ linings
• Avoid locations with vibrations from for example pumps
• Avoid locations with extensive temperature changes
• Avoid corrosive environments and locations with a great risk of con­densation, or consult factory for special builds for these locations
• Take care that condensate and water cannot enter the connector box on the flow sensor.
• There must be sufficient free space around the flow sensor.
IMPORTANT: Observe that the correct flow direction is set in the
MagFlux
flow direction is "A" (flow direction towards left).
®
"Converter Setup" menu and in the "Service Menu". Default
15
FW: 842018/842009
Electromagnetic Flow Meter
The flow directions A or B are clearly shown on the name plate. Default flow direction is "A".
Pressure loss
The pressure loss can easily be determined, if the nominal pipe diameter is greater than the MagFlux® flow sensor. See the diagram below:
Pressure loss graph
0.003
The diagram illustrates that decreasing the internal diameter from 100 mm (DN) to 80 mm (DO) will cause a pressure loss of 0.003 Bar @3 m/s.
16EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
10 m/s
Accuracy
According to the type and size of the flow sensor, the measuring accuracy will be better than 0.25%, provided that the flow sensor has the correct dimension.
[%]
5
4
3
2
1
0,20,10 0,5 21 5
Sizing
The flow sensor should be selected so the flow velocity through the sensor will be between 0.2 - 10 m/s.
MJK recommends that flow velocities in tubes are kept between 1 - 3 m/s for reliable and safe operation. See also the dimensioning table below and the dimensioning chart on the following page.
Min and max flow
DN
3 5.09 254 20 226 11304 65 2,39 119 6 20.4 1018 25 353 17676 80 3,62 181
8 36.2 1810 32 579 28944 100 5,65 283 10 56.5 2827 40 905 45360 125 8,84 442 15 127 6362 50 1414 70560 150 12,7 636
Min and max flow
DN
200 22.6 1131 450 115 5726 900 458 22902 250 35.3 1767 500 141 7069 1000 565 28274 300 50.9 2545 600 204 10179 1200 814 40715 350 69.3 3464 700 277 13854 The Qmax. is equal to 20 mA 400 90.5 4524 800 362 18095
Qmin Qmax
0.2 m/s 10 m/s 0.2 m/s 10 m/s 0.2 m/s 10 m/s
[l/h] [l/h] [l/h] [l/h] [m
Qmin Qmax
0.2 m/s 10 m/s 0.2 m/s 10 m/s 0.2 m/s 10 m/s
3
/h] [m3/h] [m3/h] [m3/h] [m3/h] [m3/h]
[m
DN
DN
Qmin Qmax
Qmin Qmax
17
Qmin Qmax
DN
3
Qmin Qmax
DN
factory preset
FW: 842018/842009
/h] [m3/h]
Electromagnetic Flow Meter
DN 1000 - 40"
DN 900 - 36"
DN 800 - 32"
DN 700 - 28"
DN 600 - 24"
DN 500 - 20"
DN 450 - 18"
DN 400 - 16"
DN 350 - 14"
DN 300 - 12"
DN 250 - 10"
DN 200 - 8"
DN 150 - 6"
DN 125 - 5"
DN 100 - 4"
DN 80 - 3"
DN 65 - 2 1/2"
DN 50 - 2"
DN 40 - 1 1/2"
DN 32 - 1 1/4"
DN 25 - 1"
DN 20 - 3/4"
DN 15 - 1/2"
DN 10 - 3/8"
DN 8 - 5/16"
DN 6 - 1/4"
DN 3 - 1/8"
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
2
3
4
5
6
7
8
9
10
0, 2
0, 3
0, 4
0, 5 0, 6 0, 7
0, 8 0, 9
1
2
3
4
5 6 7
8 9
10
20
30
40
50 60 70
80 90
0, 02
0, 01
0, 03
0, 04
0, 05 0, 06 0, 07
0, 08 0, 09
0, 1
20 000
30 000
10 0
20 0
30 0
40 0
50 0
60 0 70 0 80 0 90 0
10 00
20 00
30 00
40 00
50 00
60 00 70 00 80 00
90 00
10 000
1,5
Sizing Chart
Example: A MagFlux with an internal diameter of 100 mm can measure flow rates between approx. 290m3/h and 5.6 m3/h, and the fluid velocity is
1.5 m/s at a flow rate of approx. 42 m3/h.
Fluid velocity [m/s]
Size
3
/h]
Flow [m
18EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
3
2
3
2
3
2
Pipe System
1. The flow sensor must be mounted in a location which is free from inter­fering elements like valves, Ts, bends, pumps, etc. to ensure a laminar flow without turbulence upstream of the flow sensor. For that reason the flow sensor must be mounted in a straight pipe at a distance from interfering elements of minimum 3 x DN upstream and minimum 2 x DN downstream.
Flow direction
Flow direction
Flow direction
Important: Valves should always be mounted on the downstream side
of the flow sensor!
19
FW: 842018/842009
Electromagnetic Flow Meter
2. If it becomes necessary to use reducers, the inner angle must not
exceed 7.5°.
The minimum length to keep the angle below 7.5° can be checked by
means of the formula below:
L = (D - d) x 7.63
where "D" is the large diameter and "d" the small diameter of the re-
ducer.
Example: If a flow sensor in dimension DN 80 is mounted downstream
of a 100 mm pipe, the reducer must then have a length of minimum 152,6 mm in order to keep the inner angle below 7,5 °.
3. Flange connections must be assembled concentrically on both the up-
stream and the downstream side. Measuring accuracy will be affected by turbulence in the liquid from poorly made connections.
Important: Gaskets and grounding rings must also be mounted con-
centrically!
4. The flow sensor should always be filled with liquid. For that reason the
flow sensor must not be mounted at the highest point of the pipe sys­tem or in free outlets, where gravity could empty or partially empty the pipe.
20EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
5. The flow sensor can be mounted vertically or horizontally.
If the flow sensor is mounted vertically, the flow direction should always
be upwards. In that way the effect from possible bubbles in the liquid will be significantly reduced, just as it will ensure that the flow sensor is always filled with liquid.
In case the liquid is carrying particles, for example when measuring
sludge, sewage, etc., the flow sensor must be mounted vertically.
6. When mounting horizontally in pipes with free downstream outlet, the flow sensor should be mounted such that it will always be filled with liq­uid, for example in a bend situated lower than the height of the outlet.
In case the liquid is carrying particles, e.g. when measuring sludge,
sewage etc. the flow sensor must be mounted vertically.
21
FW: 842018/842009
Electromagnetic Flow Meter
7. When mounting horizontally the flow sensor can be rotated max. +/- 45°
seen from the connection end. If the flow sensor is rotated more than 45°, one of the electrodes may not be in full contact with the liquid.
Cabinet Mounted on the Flow Sensor
The cabinet is mounted on the flow sensor with four pcs. 6x12 hex cap screws.
22EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Potential Equalization and Grounding
Type 7100/7200/7600 in Conductive Pipes
Conductive pipes
Equalization connection
Soft gasket
Ground connection.
NB! The flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm
Type 7100/7200/7600 in Non-conductive Pipes
2
.
Non-conductive pipes
Grounding ring (Min. thickness = 5 mm.)
Potential equalization connection
Soft gasket
Ground connection.
NB! The flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm
23
2
FW: 842018/842009
.
Electromagnetic Flow Meter
Type 7300/7400 in Conductive Pipes
Potential equalization connection
Conductive pipe
Ground connection.
NB! The flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm
Type 7300/7400 in Non-conductive Pipes
Grounding ring (Min. thickness = 5 mm.)
Potential equalization connection
Soft gasket
Soft gasket
Conductive pipe
2
.
Soft gasket
Non-conductive pipe
Ground connection.
NB! The flow sensor must be connected to an effective ground connection, and the wire dimension must be at least 1.5 mm
24EN 3.05 MagFlux Manual 1212
Non-conductive pipe
2
.
Electromagnetic Flow Meter
521255
Ground from
flow pipe
Red
White
White
Black
Green
White
L2E2E1
L1GND
Mains
supply
-mA--RE2--RE1--DI1--Display-
+ +A B DI COM CNO NO PE N L
- -
FUSE
www.mjk.com
Fuse rating: 100 mAT @ 115 V AC
63 mAT @ 230 V AC
521254
Connection for top
mounted operator panel
Live
Neutral
Protective earth
PE N L
Terminals
for remote
operator panel
+ A B
-
Up to four MagFlux® flow sensors can be connected to one operator panel:
4 wire cable.2 wire cable.
A, B
+, A, B,
-
4-20 mA (Active.)
Max. 500
Electronic relay
Max. 50 V 120 mA resistive load
Mechanical relay
Max. 50 V 1A resistive load
< 5 VDC = O ; > 10 VDC = I
Max. 30 VDC
+DI COM COMNO NO
-
DO1DI DO2 AO
FUSE
MAINS SUPPLY
Sensor #1
Sensor #4
Operator
panel
4. Flow Converter
Electrical Mounting
Warning: The MagFlux® flow converter / flow sensor must not be mounted in explosion hazardous areas!
Loosen the four screws (position indicated by arrows) and remove the display unit to gain access to the terminals.
25
FW: 842018/842009
Electromagnetic Flow Meter
Power Supply
The flow converter must be supplied from a properly fused mains outlet, a 24 volt AC outlet, or a 10 - 30 V DC power supply/battery.
Power Supply 230 V AC, 115 V AC or 24 V AC
Terminal Designation
PE Protective ground
N 230 / 115 / 24 V AC neutral
L 230 / 115 / 24 V AC live
Power Supply 10 - 30 V DC
Terminal Designation
PE Protective ground
DC neutral
+ DC live
The internal fuse ratings are:
Internal Fuse Ratings
Voltage Rating Order no. Dimension
230 V AC 0.063 mA T 550030 5 x 20 mm
115 V AC 0.125 mA T 550035 5 x 20 mm
24 V AC 0.5 mAT 550049 5 x 20 mm
10 - 30 V DC 1.0 AT 550051 5 x 20 mm
The technical specifications/requirements for a 10 - 30 V DC power supply/battery are:
Technical Specifications for 10 - 30 V DC Power Supply
Power consumption without display < 5 W
Power consumption with display < 8 W
Peak start current @12 V DC,1 second Approx. 1,5 A
Peak start current @24 V DC,1 second Approx. 1 A
26EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Changing the Power Supply Voltage 230/115 V AC
To change the input mains voltage from 230 VAC to 115 VAC (or vice versa) proceed as follows:
1. Loosen the four screws on the front and lift out the display (see also page 25).
2. Note down the color and position of the wires in the terminal blocks, and then loosen the terminal screws.
3. Unscrew the two screws that hold the metal cover and then remove it.
4. Unscrew the four screws that hold the mother PCB (the printed circuit board with all the electrical components).
5. Remove the mother PCB and look at the back side. This is what you should see (230 V AC configuration):
6. Unsolder and remove the jumper indicated by the arrow.
27
FW: 842018/842009
Electromagnetic Flow Meter
7. Look at the following picture to see where two new jumpers must be
inserted (115 V AC configuration):
8. Insert and solder 2 jumpers (wires) in the positions indicated by the ar-
rows.
9. Turn around the PCB and replace the 63 mAT fuse on the right side
with a 125 mAT fuse ( or vice versa going from 115 VAC to 230 VAC).
10. Re-insert the PCB and re-assemble the flow meter (see also steps 4, 3,
2 and 1).
28EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Analogue Output
The analogue output is an active output with a max. load of 800Ω.
Analogue Output
Terminal Designation
AO + 4-20 mA AO
-
4-20 mA
The analogue output can be programmed for indication of:
- direct flow
- inverse flow
- bidirectional flow
- absolute flow
See details on pages 63 - 66 incl.
Digital Outputs
MagFlux® has two digital outputs - DO 1 with an opto (light triggered)
relay and DO 2 with a mechanical relay. They can both be programmed for the following functions:
- high and low flow alarms
- pulse output for counters R Totalizer Forward and Reverse
- batch 1 and batch 2 counters
- system error alarm
- empty pipe alarm
- flow direction indication
Opto Relay (DO1)
Terminal Designation
DO 1 Com Max. 50 V DC /120 mA DO 1 NO
Mechanical Relay (DO2)
Terminal Designation
DO 2 Com Max. 50 V DC /1 A DO 2 NO
DO 1 shares the common terminal (Com) with DI.
29
FW: 842018/842009
Electromagnetic Flow Meter
See also application examples in Appendix G. Digital Input/Output Con­nections on page 123.
Digital Input
MagFlux® has one digital input which is activated with a voltage higher
than 10 V DC and de-activated with a voltage lower than 5 V DC.
Digital Input
Terminal Designation
DI Com DI Max. 30 V DC
The digital input (DI) can be programmed for the following functions:
- start and pause batch 1 and batch 2 counters
- reset counters R Totalizer Forward and Reverse
- reset counter R Totalizer Sum DI shares the common terminal (Com) with DO 1. See also application examples in Appendix G. Digital Input/Output Con-
nections on page 123.
Local (compact) Flow Sensor
Connect the compact (local) flow sensor to the flow converter with the wires coming from the flow sensor as shown below.
Compact Flow Sensor
Terminal Signal Designation Colour from sensor
1
2 E1 Electrode Red
3 E2 Electrode White
4 GND Ground Black
5 L1 Coil Blue
6 L2 Coil Brown
7, 8 and 9 not used
Liquid GND
Built-in liquid ground electrode
30EN 3.05 MagFlux Manual 1212
Black/shield
Electromagnetic Flow Meter
Remote Flow Sensor
Connect the remote flow sensor to the flow converter with the included cable (part number 691080).
Important: Do not use other cable types! Important: Terminals 1 and 4 must be shunted!
The extension cable length between converter and sensor must under normal circumstances not exceed 50 meters. Power lines running in paral­lel and noisy environments may reduce the max. length to less than 50 meters. Consult MJK personnel for advice and guidance.
Remote Flow Sensor 691080
Terminal Signal Designation Colour
1
Liquid GND Shield
before March 2011
- -
2 E1 Wire pair no. 1 Red Red
3 E2 Wire pair no. 1 White White
4 GND Wire pair no. 3 Black Black
5 L1 Wire pair no. 2 Green Blue
6 L2 Wire pair no. 2 White Brown
7, 8 and 9 not used
Colour
from March 2011
Converter Connection Board - Local
Red
White
Black
Blue
Brown
Electronic relay
max. 50 V DC 120 mA
Electromechanical relay
max. 50 V AC / 1 A
Conn ection for
Flow Senso r
For attached
Display Unit
For remote
Display Unit
DI 1
Digital output 1 DO 1
Digital input
10-30 V DC
4-20 mA max. 800
DO 2Digital output 2
4-20 mA output
31
63 mAT @ 230 V AC
Fuse rating: 125 mAT @ 115 V AC
Power supply
FW: 842018/842009
Electromagnetic Flow Meter
Converter Connection Board - Remote
691080 cable before March 2011
Connection for local sensor
A
1 2 3 4 5 6 7 8 9
Connection for remote converter
B
1 2 3 4 5 6 7 8 9
Connection for local sensor
A
1 2 3 4 5 6 7 8 9
Plug for local display unit
Connection for remote display unit
Plug for local display unit
691080 cable from March 2011
Connection for remote converter
B
1 2 3 4 5 6 7 8 9
32EN 3.05 MagFlux Manual 1212
Connection for remote display unit
Electromagnetic Flow Meter
5. System Configuration Examples
Compact Converter and Display Unit on Flow Sensor
Configuration: The MagFlux® Converter and Display Unit is mounted directly on the MagFlux® Flow Sensor.
Order numbers for this configuration: 207xxx MagFlux® Flow Sensor 207920 MagFlux® Converter with Display Unit for sensor mounting
207920 consists of: 207940 Display Unit 807000 Converter PCB 800070 Field Cabinet
bottom part for sensor mounting
33
FW: 842018/842009
Electromagnetic Flow Meter
Remote Converter with Connection Box on Flow Sensor
Configuration: The MagFlux® Converter and Display Unit is remote mounted for example when the sensor is to be buried or submerged.
Order numbers for this configuration: 207xxx MagFlux® Flow Sensor 207925 MagFlux® Converter with Display Unit for wall mounting 691080 MagFlux® Sensor cable 579035 MagFlux® Gel Potting Kit IP68
207920
A
1 2 3 4 5 6
207925 consists of: 207940 Display Unit
207020 MagFlux connection PCB (see picture below)
820050 Blind Lid 800070 Cabinet bottom part
for sensor mounting 800075 Cabinet bottom part
for wall mounting
B
1 2 3 4 5 6
Converter connection PCB
34EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
A
207920
1 2 3 4 5 6 7 8 9
Wiring Schematic #1
for "Remote Converter with Connection Box on Flow Sensor" on the op­posite page (page 34).
Flow Sensor
Connect one end of the MJK cable to the flow sensor as de­scribed in the table below. All 3 shields are twisted and connected in slot 1. The white lead in pair no. 3 is not used.
B Connection for remote converter
Terminal Signal Wire Colour before
1
GND Shield
2 E1 Pair no. 1 Red Red
3 E2 Pair no. 1 White White
4 GND Pair no. 3 Black Black
- - Pair no. 3 White -
5 L1 Pair no. 2 Green Blue
6 L2 Pair no. 2 White Brown
7, 8, 9 not used
March 2001
- -
Colour from March 2011
B
1 2 3 4 5 6 7 8 9
Blue
Brown
Remote Converter (and Display Unit)
Connect the other end of the MJK cable to the remote converter as shown to the right. The "1 Shield" lead goes into "Ground from flow pipe", the "2 Red" lead goes into "E1 Red", the "3 White" lead into "E2 White", and so on.
Blue
Brown
35
FW: 842018/842009
Electromagnetic Flow Meter
Remote Display and Multiple Converters Wiring
Configuration: Two locally mounted MagFlux® Flow Sensors and con­verters with blind lid and one remote mounted Display Unit.
The communication between the sensors and the Display Unit is executed on shielded twisted 2- and 4-wire cables using the Modbus tion protocol on RS-485.
Order numbers for this configuration: 207xxx MagFlux® Flow Sensor (2 pcs.) 207925 MagFlux® Converter with
Display Unit for wall mounting
®
communica-
207925 consists of: 207940 Display Unit 207020 MagFlux connection
PCB 820050 Blind lid 800070 Cabinet bottom part
for sensor mounting 800075 Cabinet bottom part
for wall mounting
Note: The total maximum cable length for the 2- and 4-wire shielded twisted cables is 1000 meters.
36EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Wiring Schematic #2
for the "Remote Display and Multiple Converters Wiring" example on the opposite page (page 36).
Local flow sensor 1
2-wire twisted shield cable:
Connect lead "1" to slot "A", lead "2" to slot "B" and the twisted shield to slot "-".
1 2
Shield
Local flow sensor 2
2-wire twisted shield cable:
Connect lead "1" to slot "A", lead "2" to slot "B" and twisted shield to slot "-".
4-wire twisted shield cable:
Connect lead "1" to slot "A", lead "2" to slot "B", lead "3" to slot "+" and lead "4" and twisted shield to slot "-".
213
4 and shield
Remote display unit
4-wire twisted shield cable:
Connect lead "1" to slot "A", lead "2" to slot "B", lead "3" to slot "+" and lead "4" and twisted shield to slot "-".
37
FW: 842018/842009
Electromagnetic Flow Meter
This page intentionally left blank.
38EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
6. Startup
Initial Checks
Before switching on power the following steps must be checked.
1. The local mains power supply voltage corresponds to the voltage printed on the identification label of the flow converter.
2. All electrical connections are made in accordance with the electrical connection diagram shown on page 32.
3. All terminal screws are tightened.
4. All cable glands are tightly secured.
5. All grounding connections are made in accordance with the instruc­tions in this manual (see page 23).
Initial Flow Measurement
1. Make sure that the flow sensor is completely filled with liquid.
2. Turn on power to the flow converter and wait one minute.
3. Verify or enter the calibration code of the flow sensor in the "Converter Setup/Service Menu/Sensor Calibration Code" menu (see also page 86).
If the entered calibration code is incorrect, and for example indicates a
wrong internal diameter, read the correct calibration code on the con­verter as explained on page 14.
Flow Direction Check
1. Allow free flow through the sensor.
2. Check on the display unit screen that the flow value increases. If the value goes negative, check that the "Flow direction" option is set cor­rectly and/or check the electrical connections on the flow sensor.
Language Selection
1. The default display language is English. If another language is required, proceed with step 2.
2. Select "Display Setup" from the Main Menu.
3. Select "Language" from the Display Setup menu and chose the re­quired language (see page 96).
39
FW: 842018/842009
Electromagnetic Flow Meter
Display Read-out, one connected unit
All MagFlux® display read-outs are illustrated and described in this manual. Chapter 6. MagFlux Menus (on page 47) gives a detailed description of the
displays shown during setup, configuration and normal operation. A 5-line LCD screen displays symbols and the actual status during setup,
configuration, servicing and normal operation (see also page 48).
Keyboard for reading and programming the MagFlux
The display shown above will in the remaining part of this manual be stylized to appear like this:
MagFlux
Flow FTot 18.77 m
Select
4.1 m
40EN 3.05 MagFlux Manual 1212
3
/h
3
Setup
Electromagnetic Flow Meter
Oxix1
Oxygen sat.
91.9 %
Select
Setup
Display Read-out, several connected units
When several MJK units are interconnected, for example an Oxix® dis­solved oxygen transmitter and MagFlux® flow meter with different names and Modbus ID numbers, a "Display Overview Menu" is available at top level (press "Back" repeatedly):
4.1 m3/h
MagFlux Oxix1 91.9 %
All connected units are displayed and sorted by their ID numbers, and consequently each unit can be selected and set up as required:
Important: More connected units can only by managed as described above, if each unit have been assigned a unique name and Modbus ID num­ber. See page 98 for details.
Display Keys
The keys and the soft keys (variable key functions determined by the display firmware) are used for initial programming and normal operation of the flow meter.
The function(s) of the four available keys is displayed at the bottom of the display. The symbols and actual functions are described in detail in the sections that describe the individual menus.
MagFlux
Flow FTot 18.77 m
Select
4.1 m
Back Select
3
/h
3
Setup
or
Setup
Oxix1
Oxygen sat.
Select
91.9 %
Setup
41
FW: 842018/842009
Electromagnetic Flow Meter
Contrast Adjustment
Adjust the display contrast by pressing the two outmost keys simultane­ously (indicated by the keys) and press the up/down keys as required.
Save the new setting by pressing the two outmost keys simultaneously.
System Reset
You can reset and refresh all system displays and key combinations by pressing all four keys simultaneously.
This feature is especially useful during for example a service call, where the display language (Dutch, Danish, etc.) cannot be understood by the service person. A system reset immediately sets the display language to (GB) English.
The system reset must not be confused with the ultimate "Recover factory settings" (see page 100) which deletes all local configurations/settings and replaces them with default factory settings.
42EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Main Menu
Initial Setup
Initial setup is normally performed by MJK personnel and is required to get started with the MagFlux® flow meter.
A unique sensor calibration code provides the MagFlux® flow converter with information about flow sensor number, nominal diameter and calibra­tion data.
If changes to the unique sensor calibration code are needed following physical installation and initial setup, the "Sensor Calibration Code" menu must be addressed. A password is not required to enter or change the sensor calibration code. Toggle through the displayed digits with the right-arrow key when prompt­ed for a password and press OK.
1. Press the "Setup" key on the LCD display (see below) to enter the
MagFlux
®
Main Menu.
MagFlux
Flow FTot 18.77 m
Select
4.1 m
Setup
3
/h
3
2. Press the up/down keys to highlight the required menu line (here: "Converter Setup") and then press OK to select it.
Specify Main Screen Factory Settings Data Logger Password Set Sensor Name
Converter Setup
Converter Setup Display Setup
Back Select OK
43
FW: 842018/842009
Electromagnetic Flow Meter
Sensor Calibration Code |
Back Select >
S/N not entered
Set Flow Direction Follow Arrow A Follow Arrow B
Back Select OK
MagFlux1
Flow 0.00 m
3
/h
FTot 00.000 m
3
Select Setup
A B
3. Then select "Service Menu" and "Sensor Calibration Code".
Converter Setup
Service Menu
Converter Setup Minimum Flow Averaging Units mA Output R Totalizer Forward R Totalizer Reverse R Totalizer Sum NR Totalizer Forward NR Totalizer Reverse NR Totalizer Sum Batch Counter 1 Batch Counter 2 High Flow Low Flow Flow Direction Empty Pipe 24h Flow Status
Service Menu
Service Menu Back Select OK
Sensor Calibration Code
Sensor Calibration Code
Converter SW Ver. Product Data Reset Counter Time Internal Meas. & Cal. Minimum Velocity Calibrate Flow now Calibrate Volume Calibrate mA Test Velocity Freeze Coil Read Event Log Back on Stock Back Select OK
Sensor Calibration Code
Sensor Calibration Code dw5uq4
Back Select OK
Set Flow Direction Follow Arrow A Follow Arrow B
A B
Back Select OK
4. Enter the calibration code read from the converter label (or pass through or change the displayed calibration code) and then press OK. See also details regarding the calibration code on page 14.
5. Select the correct flow direction in the "Set Flow Direction" menu and press OK.
44EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Type Password
0000
Back Select >
Service Menu
Sensor Calibration Code Converter SW Ver. Product Data Reset Counter Time Internal Meas. & Cal. Minimum Velocity Calibrate Flow now Calibrate Volume Calibrate mA Test Velocity Freeze Coil Read Event Log Back On Stock Back Select OK
OK
Converter SW Ver.
Product Data
Back
Product ID
56136
Back Select >
Product type No.
0000207910
Back Select >
Track No.
00000
Back Select >
HW type No.
00000510250
OK
OK
OK
Reset Counter Time
Reset Counter 11 Systen Runtime 1632838 Since reset 426838 Runtime (days) 18 Back Select OK
Converter 842009-005 Build 7659
Build 719
Bootload 840020-002
6. Pass through the "DN" menu and press OK.
Sensor Calibration Code
Sensor Calibration Code 7V3RI0
The MagFlux® is now configured with default settings and is ready for use.
Note: If the sensor has not been pre-configured from the factory with a sensor calibration code, the calibration code procedure is somewhat different form the just mentioned.
Back Select >
Set Flow Direction Follow Arrow A Follow Arrow B
<A ----------------------------- B>
Back Select OK
DN
00025
Back Select >
OK
OK
See the steps on the following page to configure a non-configured sensor with a calibration code.
45
FW: 842018/842009
Electromagnetic Flow Meter
1. Apply power to the sensor.
2. When the "Sensor Calibration Code" menu appears (see below), enter the calibration code read from the converter label and press OK.
S/N not entered
Sensor Calibration Code |
Back Select >
Set Flow Direction Follow Arrow A Follow Arrow B
A B
Back Select OK
MagFlux1
Flow 0.00 m FTot 00.000 m
Select Setup
3
/h
3
3. Select the correct flow direction from the "Set Flow Direction" menu and press OK.
4. Sensor calibration code configuration is now finished, and you are returned to the Main Menu.
46EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Main Menu
7. MagFlux Menus
All the MagFlux® menus and sub menus are shown and described in the following sections.
Continous overviews of the menu and sub-menu structures are presented on the gatefolds at the end of this manual:
Main Menu Overview (page Gf-1)
Converter Setup Menu Overview (page Gf-2)
Service Menu Overview (page Gf-3)
Display Setup Menu Overview (page Gf-4)
Main Menu
1. Press the "Setup" key on the MagFlux® display (see below) to enter the
MagFlux® Main Menu.
MagFlux
Flow FTot 18.77 m
Select
4.1 m
3
/h
3
Setup
The MagFlux® Main Menu contains a number of sub menus (see figure
below).
Specify Main Screen
Specify Main Screen Factory Settings Data Logger Password Set Sensor Name Converter Setup Display Setup
Back Select OK
2. Press the up/down keys to highlight the required menu line (here: Specify Main Menu) and then press OK to select it.
47
FW: 842018/842009
Electromagnetic Flow Meter
Select Line Line 1 - Sensor Name Line 2 - Flow Line 3 - FTot Line 4 Line 5 Back Select OK
Specify Main Screen
Select Line Content Not in use Flow RFTot RRTot RTTot FTot RTot Cnt Flow Direction Batch 1 Batch 2 Batch 1 Dn Batch 2 Dn Batch 1 Cnt Batch 2 Cnt Batch 1 Vol Batch 2 Vol Clock
Back Select OK
Specify Main Screen
The "Specify Main Screen" menu allows you to customize the MagFlux® display to suit your requirements. You can add and remove the five avail­able display lines and configure them individually.
1. Press the up/down keys to highlight the required menu line and then press OK to select it.
2. Press the up/down keys to highlight the required option and press OK.
48EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
The available options are:
Not in use
The line will not be used. The set free space will be used by the other
lines.
Sensor Name
The actual sensor name like a number, a location, a name or a function
will be displayed Note: The actual sensor name is defined later on in the "Set Sensor
Name" menu on page 59. Flow Actual flow rate (in units chosen in the menu for primary units).
RFTot
Resettable Forward Totalizer counter 32bit (unit L) 4 294 967 295 L
RRTot
Resettable Reverse Totalizer counter 32bit (unit L) 4 294 967 295 L
RTTot
Resettable Total Totalizer counter 32bit (unit L) 2 147 483 647 L (+/- half of Tot)
FTot
Forward Totalizer counter 64bit (unit mm
3
) 18 446 744 073 709 551 615 mm RTot Reverse Totalizer counter 64bit (unit mm
3
) 18 446 744 073 709 551 615 mm
Cnt
Totalizer sum Counter 64bit (unit mm
3
) 9 223 372 036 854 775 807 mm3 (+/- half of Tot)
Flow Direction
Flow direction "A" or "B" will be displayed.
3
3
49
FW: 842018/842009
Electromagnetic Flow Meter
Batch 1
Timer for batch volume 1
Batch 2
Timer for batch volume 2
Batch 1 DN Displays the missing amount of Batch volume 1 (DN = down)
Batch 2 DN Displays the missing amount of Batch volume 2 (DN = down)
Batch 1Cnt Batch 1 Counts (number of batches)
Batch 2 Cnt Batch 2 Counts (number of batches)
Batch 1 Vol Batch 1 Volume
Batch 2 Vol Batch 2 Volume
Clock
Time and date
The size of the display lines will automatically increase or decrease as the number of display lines is removed or added to maximize the field of view for the measured values
50EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Main Menu
Factory Settings
The "Factory Settings" menu resets the display to default settings, to met­ric settings or to US settings.
Note: The converter settings are not changed from this menu. Converter settings are covered in "Converter Setup" on page 61.
1. Press the up/down keys to highlight the required menu line (here: Fac-
tory Settings) and then press OK to select it.
Specify Main Screen Factory Settings
Factory Settings
Data Logger Password Set Sensor Name Converter Setup Display Setup
Back Select OK
2. Press the up/down keys to highlight the required option and then press OK to select it.
Factory Settings
Recover Factory Settings Cancel Default Set to Metric Set to US
Back Select OK
The available options are: Cancel Exit the menu without changes.
51
FW: 842018/842009
Electromagnetic Flow Meter
Default
When "Default" is selected, the following settings are not affected and remain as chosen by the user:
• Sensor Name
• Device ID
• Flow Direction
• DN, size
• Calibration
• Calibration Code
The in- and outputs are not activated from the factory. Activate the in- and outputs by configuring the functions with for example the default values shown in the following tables:
• MagFlux Default DI/DO Settings
• MagFlux Default 20 mA Settings
MagFlux Default DI/DO Settings
Digital Output 1 (opto) - DO1 Digital Output 2 (mech.) - DO2 Digital Input - DI
Counter R Totalizer Forward Counter R Totalizer Forward RESET R Totalizer Forward
Counter R Totalizer Reverse Counter R Totalizer Reverse RESET R Totalizer Reverse
Batch Counter 1 Batch Counter 1 RESET R Totalizer Sum
Batch Counter 2 Batch Counter 2 Start / Pause Batch counter 1
Low Flow Low Flow Start / Pause Batch counter 2
High Flow High Flow
Flow Direction Flow Direction
Empty Pipe Empty Pipe
Note: The same function can not be on assigned to both relays.
52EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
MagFlux Default 20 mA Settings
Metric US
DN [mm] Flow [m3/h] Size Flow [GPM]
3 0,25
6 1 1/4¼" 10
8 2
10 5 3/8" 10
15 5 ½" 30
20 10 3/4" 50
25 20 1" 75
32 30 1¼" 100
40 50 1½" 200
50 75 2" 300
65 100 2½" 500
80 200 3" 800
100 300 4" 1000
125 400 5" 2000
150 600 6" 3000
200 1000 8" 5000
250 2000 10" 7700
300 2500 12" 10000
350 3000 14" 10000
400 4500 16" 20000
450 6000 18” 25000
500 7000 20” 30000
600 10000 24” 40000
700 15000 28” 60000
800 20000 32” 80000
900 25000 36” 100000
1000 30000 40” 100000
1200 40000 48” 200000
53
FW: 842018/842009
Electromagnetic Flow Meter
Set to metric All converter-related units can be set to the following units (default
values in bold italics)
:
Flow Unit
m3/sec
l/sec
l/min
l/h
m3/h
RFTot RRTot RSTot FTot RTot Cnt
l l l
hl hl hl hl hl hl
kl kl kl kl kl kl
3
m
m
Totalizers
l l l
3
3
m
3
m
3
m
Set to US All converter-related units can be set to the following units (default
values in bold italics)
Totalizers
3
3
ft
3
ft
3
ft
gal gal gal
Flow Unit
ft3/h
MGD
GPM
RFTot RRTot RSTot FTot RTot Cnt
3
ft
ft
gal gal gal
MG MG MG MG MG MG
3
m
3
ft
54EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Main Menu
Data Logger
Magflux® provides a data logger with a capacity of approx. 20,000 entry
points. See Appendix E on page 120 for examples and descriptions of log files.
The data logger operates after the FIFO principle (First In, First Out). If the data logger is full and new data are coming in, the oldest data are dis­carded.
1. Press the up/down keys to highlight the required menu line (here: Data Logger) and then press OK to select it.
Specify Main Screen Factory Settings Data Logger
Data Logger
Password Set Sensor Name Converter Setup Display Setup
Back Select OK
2. Press the up/down keys to change the highlighted digit, and then press the right-key (>) to proceed to the next digit.
Data Logger
Set Log Interval
Back Select >
0010 sec
3. Continue until all digits have been set and press OK. The log interval can be set in intervals from 10 sec. up to 9999 sec.
55
FW: 842018/842009
Electromagnetic Flow Meter
The data log contains:
• Date
• Time
• Flow values
In case of a power failure, the data logger continues when power returns. If more converters are connected to one Display Unit, each converter has
its own individual log interval and can be sorted. All converters share the same memory of 160.000 entry points. The log data can be shown on the Display Unit or be stored in an external
CSV file. MJK-Field Link software is needed for transfer of data into CSV file format via the USB port in the display unit. The format can be read in for example Microsoft
An example of the log capacity of one sensor versus the time interval is shown in the following table.
Seconds Minutes Hours Days
10 - 444,45 18,51
30 - 1333,34 55,55
- 1 2667 111
- 5 13333 555
- 10 26667 1111
- 30 80000 3333
- 60 160000 6667
®
Excel® (see Appendix B on page 120 for details).
Log interval Log capacity
56EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
10.000M3/H
2006-08-21
16:09:42 3.039M3/H
<< >>< >
Innlet flow
flow
32,5
m3/h
cnt 1 312535 m3 Select Graph
Graph Display
The content of the Data Logger can displayed on the Display Unit by pressing the up/down keys simultaneously (esc), highlighting "Flow" and selecting "Graph"
Inlet flow
flow
32,5 m3/h
cnt 1 312535 m3 Select Graph
The Display Unit then shows the Graph screen. To return to the Main Menu display screen, select "esc" by pressing the up/down keys simulta­neously.
The Y axis is automatically scaled according to the Qmax of the mA output. Double-arrow keys jump forward and backward one screen frame at a
time. Single-arrow keys move the cursor forward and backward on the screen.
The actual values at the cursor position is shown at the bottom of the screen.
57
FW: 842018/842009
Electromagnetic Flow Meter
Main Menu
Password
A password provides (and prevents) access to all the settings in the Display and Converter unit. The code consists of a numeric 5-digit code between 0 and 65535. If your current password is lost or forgotten, the password protection can be overruled with the code "01750".
1. Press the up/down keys to highlight the required menu line (here: Pass-
word) and then press OK to select it.
Specify Main Screen Factory Settings Data Logger
Password
Password Set Sensor Name Converter Setup Display Setup
Back Select OK
2. Press the up/down keys to highlight the required option and then press
OK to select it.
Password
Type Password 00000
Back Select >
Login
Login Activate/Deactivate Change Password
Back Select OK
Activate/
Deactivate
Activate Deactivate
Back Select OK
58EN 3.05 MagFlux Manual 1212
Change Password
Set New Password
00000
Back Select >
Electromagnetic Flow Meter
Main Menu
The available options are:
Login
Use the up/down keys to set the digits one by one. Continue with >
until all digits have been set and then press OK.
Activate/Deactivate
Write protection. Selecting "Activate" means that a password must
be entered to change vital settings. "Deactivate" disables password protection.If your current password is lost or forgotten, the password protection can be overruled with the code "01750".
Change Password
The present 5-digit password can be changed as required.
Set Sensor Name
A unique name and/or number, a function or a location can be assigned for a sensor (here: "MagFlux1"). It is consequently shown on the main display with up to 4 display lines.
MagFlux1
Flow
32,5
Cnt 1 312535 m3 Back Select Setup
m3/h
1. Press the up/down keys to highlight the required menu line (here: Set Sensor Name) and then press OK to select it.
Specify Main Screen Factory Settings Data Logger Password
Set Sensor Name
Set Sensor Name Converter Setup Display Setup
Back Select OK
59
FW: 842018/842009
Electromagnetic Flow Meter
Set Sensor Name
MagFlux1
Back Select >
Set Sensor Name
2. To change the default sensor name (MagFlux1), press the left/right keys
to highlight the wanted character.
3. Press the up/down keys to change the highlighted character and then
press > to proceed to the next character.
4. Continue with > until all numeric and alphabetical characters have been
set to for example "Inlet flow" (see below).
5. Press OK.
The available characters depend on the chosen language. English, for example, provides the following character set: a b c d e f g h i j k l m n o p q r s t u v w x y z A B C D E F G H I J K L M N O P Q R S T U V W X Y Z <space> 1 2 3 4 5 6 7 8 9 and 0.
Inlet flow
Flow
32,5
Cnt 1 312535 m3 Back Select Setup
m3/h
60EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Main Menu
Converter Setup
Converter Setup Minimum Flow Averaging Units mA Output R Totalizer Forward R Totalizer Reverse R Totalizer Sum NR Totalizer Forward NR Totalizer Reverse NR Totalizer Sum Batch Counter 1 Batch Counter 2 High Flow Low Flow Flow Direction Empty Pipe
Status Service Menu
Minimum Flow
24h Flow
Converter Setup
"Converter Setup" provides configuration options for volumes, batches, units, etc. See detailed descriptions in the following sections and in the overview pages at the end of this manual.
1. Press the up/down keys to highlight the required menu line (here: Con­verter Setup) and then press OK to select it.
Specify Main Screen Factory Settings Data Logger Password Set Sensor Name
Converter Setup
Converter Setup Display Setup
Back Select OK
2. Press the up/down keys to highlight the required option and then press OK to select it.
The available options are described in detail in the following sections.
61
FW: 842018/842009
Electromagnetic Flow Meter
Averaging
Units
Averaging
0010 sec
Back Select >
Flow Forward Flow Reverse Forward & rev. (12mA=0) Forward & rev. (4 mA=0) Not in use Back Select OK
Flow
m
3
/sec l/sec l/min l/h m
3
/h
3
Units
mA Output
Flow Forward Flow Reverse Forward & rev. (12mA=0) Forward & rev. (4 mA=0) Not in use Back Select OK
Flow
m
3
/sec l/sec l/min l/h m
3
/h
3
Minimum Flow
The "Minimum Flow" option sets the minimum flow rate.
Magflux® flow meters are default set to the values shown on page 17.
The unit for the flow rate can be set from the "Units" option (see page 64).
Minimum Flow
Minimum Flow
0.177 m
Back Select >
3
/h
1. Select the digits one by one with the left/right keys and set the value with the up/down keys.
2. Press OK to save the final setting.
Averaging
The "Averaging" option sets the time period within which the measure­ments are smoothed and averaged.
Averaging
Averaging
Back Select >
1. Select the digits one by one with the left/right keys and set the value with the up/down keys.
2. Press OK to save the final setting.
0010 sec
62EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
mA Output
Flow Forward Flow Reverse Forward & rev. (12mA=0) Forward & rev. (4 mA=0) Not in use Back Select OK
(4mA = 0 Flow)
(12mA = 0 Flow, +4mA = - 20mA)
Units
The "Units" option sets the unit for the flow rate. The available units are shown below.
Units
3
m
/sec l/sec l/min l/h
3
m
/h
3
ft
/h MGD GPM Back Select OK
Flow
1. Press the up/down keys to highlight the required unit and then press OK to select it.
mA Output
When a MagFlux® is connected to a power supply for the first time, the mA output is automatically set to provide 4 mA at zero flow and 20 mA at a flow corresponding to the theoretical Qmax of the flow sensor.
Changes in the mA setting will not affect the relay output settings. Note: Both values can be set in the range 10 % to 120 %, making it
possible to increase or decrease the mA signal.
The "mA Output" parameters can be configured in four different ways. Each option is described in the following sections.
mA Output
Flow Forward Flow Reverse Forward & rev. (12mA=0) Forward & rev. (4 mA=0) Not in use Back Select OK
63
FW: 842018/842009
Electromagnetic Flow Meter
mA Output
Flow Forward Flow Reverse Forward & rev. (12mA=0) Forward & rev. (4 mA=0) Not in use Back Select OK
Flow Forward
mA Output
20 mA = 1.000 m
3
/h
Back Select >
(4mA = 0 Flow)
Flow Reverse
mA Output
4 mA = 1.000 m
3
/h
20 mA = -1.000 m
3
/h
Back Select >
(12mA = 0 Flow, +4mA = - 20mA)
Forward & rev. (12mA=0)
mA Output
4 mA = 0,00 m
20 mA = 1.000 m
Back Select >
(4mA = 0 Flow)
0
4
8
12
16
20
-100% -800 -600 -400 -200 0 200 400 600 800 100%
Flow
mA
Using the factory setup MagFlux® returns to default mA settings corre­sponding to the chosen MagFlux® flow sensors Qmin and Qmax.
The mA output is an active output, and the maximum load is 800 Ohm. The upper limit for the mA output is 20,5 mA
3,75 mA indicates the mA output is not in use 3,5 mA indicates flow under 4 mA level 20,5 mA indicates flow higher than 20 mA level.
The 20 mA maximum output is rounded up to the nearest round figure.
Example: The maximum flow of 17.676 m3/h for a DN25 is rounded up to 20.000 m3/h.
Flow Forward
The mA signal provides 4 mA at zero flow and 20 mA at Qmax in the for­ward direction.
mA Output
20 mA = 1.000 m
Back Select >
3
/h
mA
-100% -80 -60 -40 -20 0 20 40 60 80 100%
20
16
12
8
4
0
Flow
64EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
mA Output
Flow Forward Flow Reverse Forward & rev. (12mA=0) Forward & rev. (4 mA=0) Not in use Back Select OK
Flow Forward
mA Output
20 mA = 1.000 m
3
/h
Back Select >
(4mA = 0 Flow)
Flow Reverse
mA Output
4 mA = 1.000 m
3
/h
20 mA = -1.000 m
3
/h
Back Select >
(12mA = 0 Flow, +4mA = - 20mA)
Forward & rev. (12mA=0)
mA Output
4 mA = 0,00 m 20 mA = 1.000 m
Back Select >
(4mA = 0 Flow)
0
4
8
12
16
20
-100% -80 -60 -40 -20 0 20 40 60 80 100%
Flow
mA
0
4
8
12
16
20
-100% -800 -600 -400 -200 0 200 400 600 800 100%
Flow
mA
0
4
8
12
16
20
-100% -800 -600 -400 -200 0 200 400 600 800 100%
Flow
mA
(4mA = 0 Flow)
Forward & rev.
(4 mA=0)
tbd
Not in use
Flow Reverse
The mA signal provides 4 mA at zero flow and 20 mA at Qmax in the re­verse direction.
mA Output
20 mA = 1.000 m
Back Select >
mA
-100% -800 -600 -400 -200 0 200 400 600 800 100%
Forward & Rev. (12mA=0)
The mA signal provides 4 to 20 mA reverse to forward direction with 12 mA at zero flow.
mA Output
mA
-100% -800 -600 -400 -200 0 200 400 600 800 100%
4 mA = 0,00 m 20 mA = 1.000 m
Back Select >
20
16
12
8
4
0
Flow
20
16
12
8
4
0
Flow
65
3
/h
3
/h
3
/h
FW: 842018/842009
Electromagnetic Flow Meter
(4mA = 0 Flow)
Forward & rev.
(4 mA=0)
tbd
Not in use
0
4
8
12
16
20
-100% -800 -600 -400 -200 0 200 400 600 800 100%
Flow
mA
Forward & rev. (4mA=0)
The mA signal provides 4 to 20 mA for forward and reverse flow with 4 mA at zero flow.
mA Output
4 mA = 0,00 m 20 mA = 1.000 m
Back Select >
20
16
3
/h
3
/h
mA
-100% -800 -600 -400 -200 0 200 400 600 800 100%
12
8
4
0
Flow
Not in use
The "mA Output" option is not being used.
66EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Totalizers
The MagFlux® provides six totalizers, each with two or three output op­tions. Up to two totalizers can be assigned to one digital ouput.
The totalizers are:
RFTot Resettable Forward Totalizer RRTot Resettable Reverse Totalizer RSTot Resettable Sum Totalizer
and
NR FTot Non-Resettable Forward Totalizer NR RTot Non-Resettable Reverse Totalizer NR Cnt Non-Resettable (Counter) Sum Totalizer
The available output options for the resettable totalizers are:
Mechanical Relay (DO2) Opto Relay (DO1) (light-triggered electronic relay) Display only (on the Display Unit)
The available output option for the non-resettable totalizers is:
Display only (on the Display Unit)
If a chosen output is not available (dedicated to another function), a "Device Exception" pop-up message will inform the operator.
Settings and Limits for Resettable Totalizers
Totalizer Forward Unit
Available units are l, hl, kl, m
Reset with Digital Input
Reset the totalizer via the digital input (DI)
On-time
The pulse length of the opto digital output (DO 1) is adjustable from 1 ms to 10 secs.
The pulse length of the mechanical digital relay (DO 2) is adjustable from 100 ms to 10 secs.
3
, ft3, gal and MG.
67
FW: 842018/842009
Electromagnetic Flow Meter
Volume between Pulses
Select a volume between pulses in steps of: 0,001, 0,01, 0,1, 1, 10, 100 and 1000.
MagFlux Totalizer Output Setting
Metric
Optical DO 1 @100 mSec.
DN [mm] min. max. min. max. Unit
3 0,0001 0,1 0,01 0,1 m3
6 0,0001 1 0,01 1 m3
8 0,001 1 0,1 1 m3
10 0,001 1 0,1 1 m3
15 0,001 1 0,1 1 m3
20 0,001 1 0,1 1 m3
25 0,01 10 1 10 m3
32 0,01 10 1 10 m3
40 0,01 10 1 10 m3
50 0,01 10 1 10 m3
65 0,01 100 1 100 m3
80 0,1 100 10 100 m3
100 0,1 100 10 100 m3
125 0,1 100 10 100 m3
150 0,1 100 10 100 m3
200 0,1 1000 10 1000 m3
250 1 1000 100 1000 m3
300 1 1000 100 1000 m3
350 1 1000 100 1000 m3
400 1 1000 100 1000 m3
450 1 1000 100 1000 m3
500 1 1000 100 1000 m3
600 1 10000 100 10000 m3
700 1 10000 1000 10000 m3
800 10 10000 1000 10000 m3
900 10 10000 1000 10000 m3
1000 10 10000 1000 10000 m3
1200 10 10000 1000 10000 m3
Mechanical DO 2
@100 mSec.
68EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
If a required volume betwen pulses is not contained in the above steps, for example for a pre-set to a sampler, use the batch counter option.
The relays are limited to a minimum and a maximum of pulses pr. hour depending on the 20 mA scale and the on-time of the relay in milli­seconds.
DO 1 (Opto Relay)
The maximum limit is 500 pulses pr sec at 1 ms pulse length.
The formula for the minimum flow unit pr pulses is: (flow at 20mA) x (pulse delay in mSec) / 1800000
Example: Min. flow unit/pulses for DN 100 max. flow 300 m3/h ?
3
Minimum flow unit pr pulses = 300 m 0,016 m3 which is rounded to 0,1 m3 by the MagFlux.
The formula for the maximum flow unit pr pulses is: (flow at 20mA)
Example: DN 100 max. flow 300 m
Maximum flow unit pr pulses = 300 m
x 100 mSec / 1800000 =
3
/h ?
3
DO 2 (Mechanical Relay)
The maximum limit is 120 pulses pr hour.
The formula for minimum flow unit pr pulses is: (flow at 20mA) / 120
Example: DN 100 max. flow 300 m3/h ?
Minimum flow unit pr pulses = 300 / 120 = 2,5 m3 which is rounded to 10 m3 by the MagFlux.
The formula for maximum flow unit pr pulses is: (flow at 20mA)
Example: DN 100 max. flow 300 m3/h ?
69
3
FW: 842018/842009
Maximum flow unit pr pulses = 300 m
Electromagnetic Flow Meter
Mechanical Relay (DO2) Opto Relay (DO1) Display only
R Totalizer
Reverse
Mechanical Relay (DO2) Opto Relay (DO1) Display only
Back Select OK
On-time
00100 mSec
Back Select >
Totalizer Forward Unit l hl kl m
3
ft
3
gal MG Back Select OK
Volume between Pulses
01.00000 m
3
Back Select >
Reset with Digital Input Yes No
Back Select OK
Resettable Totalizers
Totalizers can be reset from the Main display (see page 71) and via the dig­ital input DI (see page 30).
R Totalizer
Forward
Mechanical Relay (DO2) Opto Relay (DO1) Display only
Back Select OK
Mechanical Relay (DO2)
Totalizer Forward Unit l hl kl
3
m
3
ft gal MG Back Select OK
Reset with Digital Input Yes No
Back Select OK
On-time
00100 mSec
Back Select >
Volume between Pulses
01.00000 m
Back Select >
Totalizer Forward Unit l hl kl
3
m
3
ft gal MG Back Select OK
Reset with Digital Input Yes No
Back Select OK
On-time
Back Select >
Volume between Pulses
3
Back Select >
Opto Relay (DO1)
00100 mSec
01.00000 m
Totalizer Forward Unit l hl kl
3
m
3
ft gal MG Back Select OK
Reset with Digital Input Yes No
Back Select OK
3
Display only
70EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Resettable Forward Totalizer
RFTot
The Totalizer counts the forward flow volume according to the primary flow direction selected at startup.
The Totalizer counts in selected units and is resettable. The counter can be connected to the relays or to the display only. To reset the counter, the totalizer must be present in the Main display
and be selected by the up/down keys followed by a click on Reset. Alternatively it can be reset via the digital input DI.
Resettable Reverse Totalizer
RRTot
The Totalizer counts the reverse flow volume according to the primary flow direction selected at startup.
The Totalizer counts in selected units and is resettable. The counter can be connected to the relays or to the display only. To reset the counter, the totalizer must be present in the Main display
and be selected by the up/down keys followed by a click on Reset. Alternatively it can be reset via the digital input DI.
Resettable Sum Totalizer
RSTot
The Totalizer counts the sum of forward flow volume minus the reverse flow volume according to the primary flow direction selected at startup.
The Totalizer counts in selected units and is resettable. The counter can be connected to the relays or to the display only. To reset the counter, the totalizer must be present in the Main display
and be selected by the up/down keys followed by a click on Reset. Alternatively it can be reset via the digital input DI.
71
FW: 842018/842009
Electromagnetic Flow Meter
Batch Counter 1
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Converter setup
Path: Main Menu - > Converter setup -> NR Totalizer Forward / NR Totalizer Reverse / NR Totalizer Sum / Batch Counter 1 / Batch Counter 2
Non-Resettable Totalizers
NR Totalizer
Forward
Totalizer Forward Unit l hl kl
3
m
3
ft gal MG Back Select OK
NR Totalizer
Reverse
Totalizer Reverse Unit l hl kl
3
m
3
ft gal MG Back Select OK
NR Totalizer
Sum
Totalizer total Unit l hl kl
3
m
3
ft gal MG Back Select OK
NR Forward Totalizer
NR FTot
The Totalizer counts the forward flow volume according to the primary flow direction selected at start up.
The Totalizer counts in selected units and cannot be reset. The counter can only be selected from the Main display.
NR Reverse Totalizer
NR RTot
The Totalizer counts the reverse flow volume according to the primary flow direction selected at start up.
The Totalizer counts in selected units and cannot be reset. The counter can only be selected from the Main display.
NR Sum Totalizer
NR Cnt
The Totalizer counts the sum of forward flow volume minus the reverse flow volume according to the primary flow direction selected at startup.
The Totalizer counts in selected units and cannot be reset. The counter can only be selected from the Main display.
72EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Batch Counter 1
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Opto Relay (DO1)
Number of preSelected (Units)
00.002778 m
3
Back Select >
Not in use
Converter setup
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Adaptive
Automatic
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
On-time
00100 mSec
Back Select >
Number of preSelected (Units)
00.002778 m
3
Back Select >
Manually
Batch Counter 2
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Automatic
On-time
00100 mSec
Back Select >
Batch Counters 1 & 2
The batch counters can operate in three different ways: automatically, manually or adaptively, and they deduct a reverse flow from the forward flow.
73
FW: 842018/842009
Electromagnetic Flow Meter
Batch Counter 1
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Opto Relay (DO1)
Number of preSelected (Units)
00.002778 m
3
Back Select >
Not in use
Converter setup
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Adaptive
Maximum Correction
20%
Back Select >
Reset Batch Error Yes No
Back Select OK
Automatic
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
On-time
00100 mSec
Back Select >
Number of preselected (Units)
00.002778 m
3
Back Select >
Number of preSelected (Units)
00.002778 m
3
Back Select >
Manually
Batch Counter 2
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Automatic
On-time
00100 mSec
Back Select >
Number of preselected (Units)
00.02778m
3
Back Select >
Batch Counter Option Chart (cont'd)
In the following descriptions of the batch counters, the function or digital status is illustrated as follows:
Function or DO activated
Function or DO deactivated
74EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Batch Counter 1
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Opto Relay (DO1)
Number of preSelected (Units)
00.002778 m
3
Back Select >
Not in use
Converter setup
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Adaptive
Maximum Correction
20%
Back Select >
Automatic
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
On-time
00100 mSec
Back Select >
Number of preselected (Units)
00.002778 m
3
Back Select >
Number of preSelected (Units)
00.002778 m
3
Back Select >
Manually
Batch Counter 2
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Automatic
On-time
00100 mSec
Back Select >
Number of preselected (Units)
00.02778m
3
Back Select >
Function or DO activated
Function or DO deactivated
Automatic Batch Counter
The automatic batch counter issues a signal, when a preset volume is reached.
On-time
The period of time in which the relay is activated (in on-state)
Number of preselected (units)
The preset batch volume.
V (volume)
Automatic
Preset Volume
DO 1 or 2
Time
75
FW: 842018/842009
Electromagnetic Flow Meter
Opto Relay (DO1)
Number of preSelected (Units)
00.002778 m
3
Back Select >
Not in use
Converter setup
Adaptive
Batch Counter 1 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Reseting the Batch Counter 1 Automatic Manually Adaptive
Back Select OK
Number of preSelected (Units)
00.002778 m
3
Back Select >
Manually
Batch Counter 2
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Opto Relay (DO1)
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Automatic
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
On-time
00100 mSec
Back Select >
Number of preSelected (Units)
00.002778 m
3
Back Select >
Manually
Automatic
V (volume)
Time
DO 1 or 2
Preset Volume
Time
Function or DO activated
Function or DO deactivated
Manual Batch Counter
The counter issues signals as determined by manual start, stop and
pause commands. A manual stop is equivalent to a reset command.
Number of preSelected (Units)
Manual
Preset Volume
The preset batch volume.
V (volume)
Start
Stop
Pause
DO 1 or 2
76EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Number of preSelected (Units)
00.002778 m
3
Back Select >
Not in use
Converter setup
Adaptive
Maximum Correction
20%
Back Select >
Reset Batch Error Yes No
Back Select OK
Batch Counter 2
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
Mechanical Relay (DO2) Opto Relay (DO1) Not in use
Back Select OK
Mechanical Relay (DO2)
Opto Relay (DO1)
Number of preselected (Units)
00.002778 m
3
Back Select >
Not in use
Converter setup
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Adaptive
Maximum Correction
20%
Back Select >
Reset Batch Error Yes No
Back Select OK
Automatic
Batch Counter 2 Unit l hl kl m
3
ft
3
gal MG Back Select OK
Pause with Digital Input Yes No
Back Select OK
Reseting the Batch Counter 2 Automatic Manually Adaptive
Back Select OK
On-time
00100 mSec
Back Select >
Number of preselected (Units)
00.02778m
3
Back Select >
Number of preSelected (Units)
00.002778 m
3
Back Select >
Manually
Automatic
Manual
V (volume)
Time
DO 1 or 2
Preset Volume
V (volume)
Start
Preset Volume
Stop
Pause
DO 1 or 2
Function or DO activated
Function or DO deactivated
Adaptive Batch Counter
The counter adapts compensates for overrun and insufficient flow.
Number of preSelected (Units)
Maximum Correction
Reset Batch Error
The preset batch volume.
The correction in percent that triggers an error message or an alarm.
Reset the batch correction percentage to 0%.
V (volume)
Adaptive
Preset Volume
overflow %
Dynamic stop of Batch
Start
DO 1 or 2
Stop
77
FW: 842018/842009
Electromagnetic Flow Meter
Settings and Limits for Batch Counters
Batch counter units
Available units are l, hl, kl, m3, ft3, gal and MG.
Start, stop or pause with Digital Input
The manual and the adaptive batch counters can be started, stopped or paused via the digital input (DI)
On-time
The pulse length of the opto digital output (DO 1) is adjustable from 1 ms to 10 secs.
The pulse length of the mechanical digital relay (DO 2) is adjustable from 100 ms to 10 secs.
Volume between pulses
Select a volume between pulses as listed in the table overleaf. The relays are limited to a minimum and a maximum of pulses pr. hour
depending on the 20 mA scale and the on-time of the relay in milli­seconds.
78EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
MagFlux Batch Counter Output Settings
Metric
Optical DO 1 @100 mSec. Mechanical DO 2 @100 mSec.
DN [mm] min. max. min. max. Unit
Automatic Manually/
adaptive
3 0,0000139 0,0006944 0,25 0,0020833 0,0006944 0,25 m
6 0,0000556 0,0027778 1 0,0083333 0,0027778 1 m
8 0,0001111 0,0055556 2 0,0166667 0,0055556 2 m
10 0,0002778 0,0138889 5 0,0416667 0,0138889 5 m
15 0,0002778 0,0138889 5 0,0416667 0,0138889 5 m
20 0,0005556 0,0277778 10 0,0833333 0,0277778 10 m
25 0,001111 0,055556 20 0,166667 0,055556 20 m
32 0,001667 0,083333 30 0,250000 0,083333 30 m
40 0,002778 0,138889 50 0,416667 0,138889 50 m
50 0,004167 0,208333 75 0,625000 0,208333 75 m
65 0,005556 0,277778 100 0,833333 0,277778 100 m
80 0,011111 0,555556 200 1,666667 0,555556 200 m
100 0,016667 0,833333 300 2,500000 0,833333 300 m
125 0,022222 1,111111 400 3,333333 1,111111 400 m
150 0,033333 1,666667 600 5,000000 1,666667 600 m
200 0,055556 2,777778 1000 8,333333 2,777778 1000 m
250 0,111111 5,555556 2000 16,66667 5,555556 2000 m
300 0,138889 6,944444 2500 20,83333 6,944444 2500 m
350 0,166667 8,333333 3000 25,00000 8,333333 3000 m
400 0,250000 12,500000 4500 37,50000 12,500000 4500 m
450 0,333333 16,666667 6000 50,00000 16,666667 6000 m
500 0,388889 19,444444 7000 58,33333 19,444444 7000 m
600 0,555556 27,777778 10000 83,33333 27,777778 10000 m
700 0,833333 41,666667 15000 125,0000 41,666667 15000 m
800 1,111111 55,555556 20000 166,6667 55,555556 20000 m
900 1,388889 69,444444 25000 208,3333 69,444444 25000 m
1000 1,666667 83,333333 30000 250,0000 83,333333 30000 m
1200 2,222222 111,111111 40000 333,3333 111,111111 40000 m
Automatic Manually/
adaptive
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
79
FW: 842018/842009
Electromagnetic Flow Meter
Limitation of the automatic batch counter on DO 1 (Opto Relay)
The maximum limit is 500 pulses pr sec at 1 ms pulse length.
The formula for the minimum flow unit pr pulses is: (flow at 20mA) x (pulse delay in mSec) / 1800000
Example: Min. flow unit/pulses for DN 100 max. flow 300 m3/h ?
Minimum flow unit pr pulses = 300 m 0,016 m3 which is rounded to 0,1 m3 by the MagFlux.
The formula for the maximum flow unit pr pulses is: (flow at 20mA)
Example: DN 100 max. flow 300 m3/h ?
Maximum flow unit pr pulses = 300 m
Limitation of the automatic batch counter on DO 2 (mech. relay)
The maximum limit is 120 pulses pr hour.
The formula for minimum flow unit pr pulses is: (flow at 20mA) / 120
Example: DN 100 max. flow 300 m3/h ?
Minimum flow unit pr pulses = 300 / 120 = 2,5 m3
3
x 100 mSec / 1800000 =
3
The formula for maximum flow unit pr pulses is: (flow at 20mA)
Example: DN 100 max. flow 300 m3/h ?
Maximum flow unit pr pulses = 300 m
3
Limitation of the manual and adaptive batch counter on DO 1 or DO 2
The formula for minimum flow unit pr pulses is: (flow at 20mA) / 360
Example: DN 100 max. flow 300 m3/h ?
Minimum flow unit pr pulses = 300 / 360 = 0,833 m3
80EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Low Flow
Mechanical Relay (DO2) Opto Relay (DO1) Display only Not in use
Back Select OK
ON at Flow
176,4 m3/h
Back Select >
Relay Function NO NC
Back Select >
Off at Flow
170,0 m3/h
Back Select >
Delay
00010 Sec
Back Select >
Mechanical Relay (DO2)
Path: Main Menu -> Converter setup -> High Flow / Low Flow
The formula for maximum flow unit pr pulses is: (flow at 20mA)
Example: DN 100 max. flow 300 m3/h ?
Maximum flow unit pr pulses = 300 m
3
High Flow/Low Flow
The high and low flow options provide alarms and trigger relays when certain volume limits are exceeded or not reached.
High Flow
Mechanical Relay (DO2) Opto Relay (DO1) Display only Not in use
Back Select OK
Mechanical Relay (DO2)
ON at Flow
10.000 m
Back Select >
OFF at Flow
9.800 m
Back Select >
Delay
00010 Sec
Back Select OK
Relay Function NO NC
Back Select >
3
/h
3
/h
Opto Relay (DO1)
ON at Flow
10.000 m3/h
Back Select >
OFF at Flow
9.800 m3/h
Back Select >
Delay
00010 Sec
Back Select OK
Relay Function NO NC
Back Select >
ON at Flow
Back Select >
OFF at Flow
Back Select >
Delay
Back Select OK
Display only
10.000 m3/h
9.800 m3/h
00010 Sec
81
FW: 842018/842009
Electromagnetic Flow Meter
The available output options for High Flow and Low Flow are:
Mechanical Relay (DO2) Opto Relay (DO1) (light-triggered, electronic relay) Display only (signals to the Display Unit only) Not in use (the High/Low Flow option is not being used)
On at Flow
Issues an alarm and/or sends a signal when for example a high flow
limit is exceeded.
OFF at Flow Cancels an alarm and/or sends a signal when recovering from for
example an overflow situation.
Delay
A delay in seconds can be set to compensate for positive and negative
spikes in the flow. The lenght is the time between an incident and set­ting an alarm.
Relay Function
NO (Normally Open) determines the relay state under normal
conditions.
NC (Normally Closed) determines the relay state under normal
conditions.
82EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Empty Pipe
Mechanical Relay (DO2) Opto Relay (DO1) Display only Not in use
Back Select OK
Mechanical Relay (DO2)
Relay Function NO NC
Back Select OK
Delay
00010 Sec
Back Select >
Flow Direction
The "Flow Direction" option indicates the flow direction of the flow and determines the open/closed state of the relay.
The default Flow Direction is set to "Direction A". If flow "Direction A" and the "NO" function are chosen, the relay is on at a negative flow ("Direction B").
Flow Direction
Mechanical Relay (DO2) Opto Relay (DO1) Display only Not in use
Back Select OK
Mechanical Relay (DO2)
Delay
00010 Sec
Back Select >
Relay Function NO NC
Back Select OK
Opto Relay (DO1)
Delay
00010 Sec
Back Select >
Relay Function NO NC
Back Select >
Display only
Delay
00010 Sec
Back Select >
The available output options for "Flow Direction" are:
Mechanical Relay (DO2) Opto Relay (DO1) (light-triggered electronic relay) Display only (signals to the Display Unit only) Not in use (the Flow Direction option is not being used)
Delay
A delay in seconds can be set to compensate for positive and negative
spikes in the flow. The lenght is the time between an incident and set­ting an alarm.
Relay Function
NO (Normally Open) determines the relay state under normal conditions. NC (Normally Closed) determines the relay state under normal conditions.
83
FW: 842018/842009
Electromagnetic Flow Meter
Empty Pipe
Status
DO1 Not in use DO2 R Totalizer Forward DI R Totalizer Forward mA output 6.250 mA
Back Select OK
Empty Pipe
A signal and/or an alarm can be set, when the sensor becomes practically empty (see also pages 19 - 22 incl.), or if the conductivity drops below 5 µS/ cm.
Mechanical Relay (DO2) Opto Relay (DO1) Display only Not in use
Back Select OK
Mechanical Relay (DO2)
Delay
00010 Sec
Back Select >
Relay Function NO NC
Back Select OK
Opto Relay (DO1)
Delay
00010 Sec
Back Select >
Relay Function NO NC
Back Select >
Display only
Delay
00010 Sec
Back Select >
The available output options for "Empty Pipe" are:
Mechanical Relay (DO2) Opto Relay (DO1) (light-triggered electronic relay) Display only (signals to the Display Unit only) Not in use (the Empty Pipe option is not being used)
Delay
A delay in seconds can be set to compensate for positive and negative
spikes in the flow. The lenght is the time between an incident and set­ting an alarm.
Relay Function
NO (Normally Open) determines the relay state under normal conditions. NC (Normally Closed) determines the relay state under normal conditions.
84EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Type Password
0000
Back Select >
Service Menu
24h Flow
The "24h Flow" option measures the total flow (FTot) for 1 day (24 hours) and logs at midnight (12 PM). Contact MJK for detailed information.
Valid settings are "No" and "Yes" (default is "No").
Status
The "Status" option provides an overview of the in- and output status. The check boxes can contain four different characters to indicate the current status at that very moment:
X A cross (X) in a check box indicates an active state / A blinking forward slant (/) in a check box indicates that the input/
output is in the process of being activated. Eventually it turns into a steady X.
\ A blinking backward slant (\) in a check box indicates that the
input/output is in the process of being de-activated. Eventually the check box is cleared.
An empty check box indicates a de-activated state.
Status
DO1 Not in use DO2 R Totalizer Forward DI R Totalizer Forward mA output 6.250 mA
Back Select OK
DO1 Not in use
Digital output D01 is active (an empty check box).
DO2 X Totalizer Forward
Digital output D02 for Totalizer Forward is activated (a ticked check
box).
DI Totalizer Forward
Digital input DI for Totalizer Forward is active (an empty check box)
mA output 6.250 mA
The present analog output current is 6.250 mA.
85
FW: 842018/842009
Electromagnetic Flow Meter
Service Menu
The "Service Menu" provides options intended for service personnel dur­ing installation, calibration, operation, monitoring and maintenance.
All parameters can be read without any restrictions, but certain parameters can only by changed after a password has been typed.
Service Menu
Type Password
Back Select >
Sensor Calibration Code Converter SW Ver. Product Data Reset Counter Time Internal Meas. & Cal. Minimum Velocity Calibrate Flow now Calibrate Volume Calibrate mA Test Velocity Freeze Coil Read Event Log Back on Stock Back Select OK
0000
OK
Sensor Calibration Code
Each and every MagFlux
®
has a unique calibration code that contains calibration data for the specific unit. A calibration code could for exam­ple be dw5uq4 or 7v3ri0.
The current converter firmware requires an 8-character input, but it also accepts 6 characters plus two "Ok"s. If the code is dw5uq4, enter:
d w 5 u q 4 "Ok" "Ok"
where "Ok" indicates that you press the "Ok" key without selecting any character. Only small letters can be entered from the MagFlux® key­pad.
86EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Type Password
0000
Back Select >
Service Menu
Sensor Calibration Code Converter SW Ver. Product Data Reset Counter Time Internal Meas. & Cal. Minimum Velocity Calibrate Flow now Calibrate Volume Calibrate mA Test Velocity Freeze Coil Read Event Log Back On Stock Back Select OK
OK
Converter SW Ver.
Product Data
Back
Product ID
56136
Back Select >
Product type No.
0000207910
Back Select >
Track No.
00000
Back Select >
HW type No.
OK
OK
OK
Reset Counter Time
Reset Counter 11 Systen Runtime 1632838 Since reset 426838 Runtime (days) 18 Back Select OK
Internal Meas. & Cal.
View Vgate Coil Current View ADC Factor Calibrate Vground Calibrate Vref View Vwater View Raw ADC View ADC Current Back Select >
View Vgate
Back Select >
Only for terminal use
Must not be surrended
A
B
C
D
E
Date Modification
Scale:
Converter 842009-005 Build 7659
Build 719
Bootload 840020-002
Sensor Direction Test
IMPORTANT: All letters are small letters (e.g. “a” and not “A”), and numbers are big characters (e.g. “0”, zero and not “o”, small o).
A potential source of error is incorrect reading of numbers and letters, where a small “o” is confused with a zero (“0”), or a small “l” is con­fused with the number one (“1”).
Note also that the default flow direction from factory is “A”. See page 14 regarding the calibration code. A password is not required to add or change the sensor calibration
code.
Sensor Calibration Code
Sensor Calibration Code 7V3RI0
Back Select >
Set Flow Direction Follow Arrow A Follow Arrow B
<A ----------------------------- B>
Back Select OK
DN
Back Select >
OK
OK
00025
87
FW: 842018/842009
Electromagnetic Flow Meter
Converter SW Ver.
Product Data
Product ID
56136
Back Select >
OK
Reset Counter Time
Reset Counter 11 Systen Runtime 1632838 Since reset 426838 Runtime (days) 18 Back Select OK
Internal Meas. & Cal.
View Vgate Coil Current View ADC Factor Calibrate Vground Calibrate Vref View Vwater
Minimum Velocity
0.100 m/ sec
Back Select >
Only for terminal use
Must not be surrended
A
B
C
D
E
Date Modification
Scale:
Skala for tegning
Sensor Direction Test
Reset Counter Time
Reset Counter 11 Systen Runtime 1632838 Since reset 426838 Runtime (days) 18 Back Select OK
Internal Meas. & Cal.
View Vgate Coil Current View ADC Factor Calibrate Vground Calibrate Vref View Vwater View Raw ADC View ADC Current Back Select >
Minimum Velocity
Minimum Velocity
0.100 m/ sec
Back Select >
View Vgate Coil Current
View Vgate
3624
Back Select >
Coil Current
148
Back Select >
View ADC Current View Raw ADC
View ADC Current
09593127.00000
Back Select >
Input noise 0.18% Raw ADC 10791 Gain 4.000 Velocity 5.21656 m/sec Flow 0.0055786500 m
3
/sec
Back Select >
Sensor Direction Test
Converter SW Ver.
The converter software version and bootload version are displayed,
so that you can determine whether a software (firmware) upgrade is required or not.
Converter 842009-005 Build 7659 Bootload 840020-002 Build 719
Back
Product Data
Key product data like ID, hardware type no., track no., etc. are dis-
played.
Product Data
Product ID
Back Select >
Product type No.
Back Select >
Track No.
Back Select >
HW type No.
Back Select >
56136
OK
0000207910
OK
00000
OK
00000510250
88EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Internal Meas. & Cal.
View Vgate Coil Current View ADC Factor Calibrate Vground Calibrate Vref
Minimum Velocity
Minimum Velocity
0.100 m/ sec
Back Select >
Only for terminal use
Must not be surrended
A
B
C
D
E
Date Modification
Scale:
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Skala for tegning
Menu system for MagFlux Converter Converter Setup
Service Menu
Calibrate Flow now
View flow now
0.0019816387 m
3
/sec
Back Select >
Sensor Direction Test
Minimum Velocity
Minimum Velocity
0.100 m/ sec
Back Select >
A
B
C
D
E
Date Modification
Scale:
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Date:
Constructor:
Part no.:
Software version
Skala for tegning
Menu system for MagFlux Converter Converter Setup
Service Menu
2006 - 09 - 20
841010
LSP
Calibrate Flow now
View flow now
0.0019816387 m
3
/sec
Back Select >
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Date:
Constructor:
Part no.:
Software version
Menu system for MagFlux Converter Converter Setup
Service Menu
2006 - 09 - 20
841010
LSP
Calibrate Flow now
View flow now
0.0019816387 m
3
/sec
Back Select >
Reset Counter Time
The counter reset and runtime can be reset for fault finding, error cor-
rection and similar service procedures.
Internal Meas. & Cal.
Internal measurements and calculations are for service technician use
only, and are used for adjustments and calibrations.
Minimum Velocity
The incoming data is below the measuring accuracy. The option forces
the minimum velocity to zero.
Reset Counter Time
Reset Counter 11 Systen Runtime 1632838 Since reset 426838 Runtime (days) 18 Back Select OK
Internal Meas. & Cal.
Sensor Direction Test View Vgate Coil Current View ADC Factor Calibrate Vground Calibrate Vref View Vwater View Raw ADC View ADC Current Back Select >
Minimum Velocity
0.100 m/ sec
Back Select >
Minimum Velocity
89
FW: 842018/842009
Electromagnetic Flow Meter
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Date:
Constructor:
Part no.:
Software version
2006 - 09 - 20
841010
LSP
Calibrate mA
Current Set fixed current Calib. current DAC zero current DAC top current Back Select OK
Test Velocity
Set velocity
00.000 m/sec
Back Select >
Velocity
4.039745 m/sec
Flow
7.136209 m
3
/h
Back Select OK
OK
Enable test Yes No
Back Select OK
OK
Freeze Coil
Freeze coil Positive Negative
Back Select OK
Freeze coil Disable Enable
Back Select OK
OK
OK
Current
20000 uA
Back Select >
OK
Set fixed current Release mA Minimum mA 4mA = 12 mA 20 mA Maximum mA
Back Select OK
OK
Calib. current 20500 uA
Calibrate Flow Now
First "View Flow Now" is displayed, and then afterwards you can cali-
brate the flow.
Calibrate Flow now
View flow now
0.0019816387 m
Back Select >
Calibration volume
The volume counter can be reset and a new value can be entered.
3
/sec
Calibration Volume
Reset counter Yes No
Back Select OK
OK
Reset counter Yes No
Back Select OK
View counter
1163.1950000000 m
Back Select OK
Set volume
1163.1950000000 m
Back Select OK
OK
3
OK
3
90EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Test Velocity
Set velocity
00.000 m/sec
Back Select >
Velocity
4.039745 m/sec
Flow
7.136209 m
3
/h
Back Select OK
OK
Enable test Yes No
Back Select OK
OK
Freeze Coil
Freeze coil Positive Negative
Back Select OK
Freeze coil Disable Enable
Back Select OK
OK
Read Event Log
No Hour Event Extension 1 1956 0 0 2 1956 3 0 3 1956 205 0 Back Next
OK
No Hour Event Extension 18 306 205 0 19 306 208 0 20 163 206 0 Back OK
Calibrate mA
From this menu the mA reading can be calibrated against a multimeter.
Calibrate mA
Current Set fixed current Calib. current DAC zero current DAC top current Back Select OK
Current
Back Select >
Set fixed current Release mA Minimum mA 4mA = 12 mA 20 mA Maximum mA
Back Select OK
Calib. current 20500 uA
OK
20000 uA
OK
OK
Back Select >
OK
DAC zero current 3955 uA 776
Back Select >
OK
DAC top current 19997 uA 3814
Back Select >
91
FW: 842018/842009
Only for terminal use
Must not be surrended
A
B
C
Date Modification
Scale:
Skala for tegning
Menu system for MagFlux Converter Converter Setup
Service Menu
Freeze Coil
Freeze coil Positive Negative
Back Select OK
Freeze coil Disable Enable
Back Select OK
OK
Read Event Log
No Hour Event Extension 1 1956 0 0 2 1956 3 0 3 1956 205 0 Back Next
OK
No Hour Event Extension 18 306 205 0 19 306 208 0 20 163 206 0 Back OK
Read sensor data
DN 00050
OK
Back
OK
Serial no 00006
OK
Back
OK
Sensor Cal 19600
OK
Back
Electromagnetic Flow Meter
Test Velocity
Test velocity is for are for service technician use only and can be used
to compare the actual velocity against the flow/velocity/diameter chart.
Test Velocity
Set velocity
Back Select >
Velocity
Flow
Back Select OK
Enable test Yes No
Back Select OK
00.000 m/sec
OK
4.039745 m/sec
3
7.136209 m
OK
/h
92EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Only for terminal use
Must not be surrended
A
B
C
D
E
Date Modification
Scale:
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Date:
Constructor:
Part no.:
Software version
Skala for tegning
Menu system for MagFlux Converter Converter Setup
Service Menu
2006 - 11 - 22
841010
LSP
Read Event Log
No Hour Event Extension 1 1956 0 0 2 1956 3 0 3 1956 205 0 Back Next
OK
No Hour Event Extension 18 306 205 0 19 306 208 0 20 163 206 0 Back OK
Back On Stock
Recover Factory Settings Cancel Back On Stock
Back Select OK
Read sensor data
DN 00050
OK
Back
OK
Serial no 00006
OK
Back
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Date:
Constructor:
Part no.:
Software version
Menu system for MagFlux Converter Converter Setup
Service Menu
2006 - 11 - 22
841010
LSP
Back On Stock
Recover Factory Settings Cancel Back On Stock
Back Select OK
Read sensor data
DN 00050
OK
Back
OK
Serial no 00006
OK
Back
Freeze Coil
For service technician use only. The sensor coil current and voltage val-
ues can be "frozen" to enable read-out using an standard multimeter.
Freeze coil Positive Negative
Back Select OK
Freeze coil Disable Enable
Back Select OK
Read Eventlog
See the events within a certain time frame.
Freeze Coil
OK
Read Event Log
No Hour Event Extension 1 1956 0 0 2 1956 3 0 3 1956 205 0 Back Next
No Hour Event Extension 18 306 205 0 19 306 208 0 20 163 206 0 Back OK
OK
93
FW: 842018/842009
Electromagnetic Flow Meter
Back On Stock
Recover Factory Settings Cancel Back On Stock
Back Select OK
MJK AUTOMATION A/S Byageren 7 DK-2800 Nñrum Fax: +45 45 56 06 46 Tel.: +45 45 56 06 56
Date:
Constructor:
Part no.:
Software version
2006 - 11 - 22
841010
LSP
Read sensor data
Access the sensor's key data.
Read sensor data
DN 00050
OK
OK
OK
OK
OK
OK
OK
3
/h
Back
Serial no 00006
Back
Sensor Cal 19600
Back
Sensor zero
0.000000 m
Back
Back on Stock
The factory settings can be re-etablished. with the "Back On Stock"
option.
Back On Stock
Recover Factory Settings Cancel Back On Stock
Back Select OK
94EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Main Menu
Display Setup
The "Display Setup" menu provides configuration options for language, clock and factory settings, Modbus parameters and software version display.
1. Press the up/down keys to highlight the required menu line (here: Dis­play Setup) and then press OK to select it.
Specify Main Screen Factory Settings Data Logger Password Set Sensor Name Converter Setup Display Setup
Display Setup
Back Select OK
2. Press the up/down keys to highlight the required optione and then press OK.
Display Setup
Language Set Clock Modbus Factory Settings Display SW Version
Communication
Back Select OK
The available options are described in detail on the following pages. Note: The option "Communication" is only visible (and selectable), if a
communication module is installed and connected to the MagFlux display. If the communication module is removed, the option disappears automati­cally. A description of available communication settings can be found in the manual for communication modules "Communication Moduler for Dis­plays".
95
FW: 842018/842009
Electromagnetic Flow Meter
Language
Language
The MagFlux® is installed with a language package, and GB English is the default language.
In the "Language" menu several national languages can be selected.
(GB) English (US) English Dansk Deutsch Español Français Korean Italiano Nederlands Norsk Soumalainen Svenska
Language
Back Select OK
Additional languages can be added using the MJK-Field Link software program (see also Appendix B on page 115).
1. Chose the required language using the up/down keys up and then
press OK.
96EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Set Clock
The "Set Clock" menu provides setting of the built-in clock and the time format.
Set Clock
Time Format 24 hour AM/PM
Back Select OK
Set Clock
2006-09-19 12:43:11
Back Select >
1. Select the date and time digits with the left/right keys and change the setting with the up/down keys.
2. Press OK and repeat step 1 for all remaining digits.
3. Press OK to save the time/clock settings.
97
FW: 842018/842009
Electromagnetic Flow Meter
Modbus Display SW Version
Change Display ID No.
1
Back Select OK
Change Display ID No.
Number of Displays
1
Back Select OK
Modbus
The MagFlux® is delivered with a standard Modbus RTU protocol, and up to four MagFlux® converters can be connected for concurrent operation.
Add Device Change Device Address Remove Device Change Display ID No. Set No. of Retries
Back Select OK
Add Device
Scanning
Back
Devices Found
Device 1
Back OK
Change Device Addresse
Change Device Address
Device 1
Back Select OK
Change Device 1
Back Select OK
Remove Device
Remove Device
Device 1
Back Select OK
1
When several converters are to be added, it is necessary to change each converter device address to a unique ID address before connecting the converter to the Display Unit.
Add Device
Four MagFlux® Converters can be connected to one Display Unit using the RS485 interface connection. Note that only one Display Unit can be used with several converters.
The Display Unit can easily be moved to another converters for readout and configuration.
1. Select "Add Device", click OK and wait for the scan process to finish.
2. The found converters are finally displayed.
Change Device Address
To change each converter's device adress it is required that you connect
98EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
Display 841xxx-xxx Build Sep 15 2006 11:18:24 Language 2006 09 21
Back
Recover Factory Settings Cancel Delete Log Restore Default Language Restore Factory Settings
Back Select OK
Factory Settings
the Display Unit to each converter one by one. If neglected, a conflict be­tween the units will arise when the RS 485 serial loop is established.
Change for example "Device 1" to "Device 2" as follows:
1. Press OK and use the up/down keys to change "1" to "2".
2. Press OK. The MagFlux
can be added to the Display Unit with the "Add Device" option.
Remove Device
If a device is to be removed from the Display Unit, proceed as follows:
1. Select the required device with the up/down keys.
2. Click OK to remove it. The device is now removed from the Display unit and display returns to the
"Modbus" menu.
®
converter ID is now changed to "2", and more converters
Change Display ID No.
Change Display ID No.
Back Select OK
Number of Displays
Back Select OK
1
1
Change Display ID No.
The display ID number and the number of displays may have to be
Set No. of Retries
Set No. of Retries
This will reset the display.
Back Select >
005
changed to avoid conflicts on the Modbus.
Set No. of Retries
If the data communication lines are subject to noise or other disturbances, the number of retries can be raised to increase the chances of a succesful change of settings.
99
FW: 842018/842009
Electromagnetic Flow Meter
Factory Settings
The "Factory Settings" menu provides log deletion and resetting to factory settings of the display unit.
Factory Settings
Recover Factory Settings Cancel Delete Log Restore Default Language Restore Factory Settings Back Select OK
1. Press the up/down keys to highlight the required option and press OK.
Important:
The Data Logger will be reset, and all devices will be deleted from the
display Unit. This is indicated by the MJK start-up screen which is displayed within a few seconds.
Display SW Version
The "Display SW Version" menu provides a read-out of the display soft­ware (actually firmware) version, the build date and last-edited date of the the multiple languages file.
Display SW Version
Display 841xxx-xxx Build Sep 15 2006 11:18:24 Language 2006 09 21
Back
100EN 3.05 MagFlux Manual 1212
Loading...