WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept
any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are discovered within the warranty period. The warranty does not cover components that are replaced during maintenance work, e.g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper handling or unauthorized opening of the instrument invalidates any warranty claim.
The compact precision handheld meter handylab pH 12 enables you to carry
out pH measurements rapidly and reliably.
The handylab pH 12 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven calibration procedures and the special AutoRead function support you in your work with the handylab pH 12.
1Keypad
2Sample display
3Jack field
ba75368e0401/2011
5
Overviewhandylab pH 12
CAL
RUN/ENTER
AR
STO
RCL
M
M
STO
CAL
RCL
AR
RUN/ENTER
1.1Keypad
Key functions
Select the measuring mode <M>:
–pH value
–ORP voltage
Save a measured value
<STO>
Switch measuring instrument on/off
<ON/OFF>
Calibration of the current measured variable
<CAL>
Display/transmit measured values
<RCL>
Activate/deactivate the AutoRead function
<AR>
Select the measuring mode
Increase values, scroll
<▲>
6
Select the measuring mode
Decrease values, scroll
<▼ >
Confirm entries, start AutoRead
<RUN/ENTER>
ba75368e0401/2011
handylab pH 12Overview
%
°F
S
TempError
mg/L
% ppm
mol/L
ISE
RCL
8
8.
8
88
8
8
mV/pH
°C
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
TP
1
UpH
Time
Baud
Day.Month No.
Ident
Year
Auto
Store
ConCal
Status display
Meas. value display
Function and
temperature display
Sensor symbol
R
e
f
5
123
4
1.2Display
1.3Jack field
The meter has either a DIN socket or a BNC socket.
The figure shows a meter with DIN socket (1).
ba75368e0401/2011
1pH electrode or single-rod measuring electrode
2Reference electrode
3Plug-in power supply (optional)
4Temperature probe
5Serial interface RS 232 / analog output (recorder)
Caution
Only connect electrodes to the measuring instrument that cannot return any
voltages or currents that are not allowed
(> SELV and > current circuit with current limiting).
Nearly all electrodes - especially SI Analytics electrodes - fulfill these conditions.
7
Overviewhandylab pH 12
1.4Technical Data
Dimensions
and weight
Mechanical
structure
Electrical safety
Test certificates
Ambient
conditions
Length [mm]172
Width [mm]80
Height [mm]37
Weight [kg]Approx. 0.3
Type of protectionIP 66
Protective classIII
cETLus, CE
Storage- 25 °C ... + 65 °C
Operation-10 °C ... + 55 °C
Relative atmospheric
< 90 % of annual average
humidity
pH/ORP
measuring ranges
Precision of
pH/ORP
(± 1 digit)
Measuring rangeResolution
pH- 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV]- 999.9 ... + 999.9
- 1999 ... + 1999
0.001
0.01
0.1
1
T [°C]- 5.0 ... + 105.00.1
pH
(± 2 pH units from the
± 0.005 at + 15 °C ... + 35 °C
± 0.01
calibration point)
U [mV]± 0.3 at + 15 °C ... + 35 °C
± 1
8
ba75368e0401/2011
handylab pH 12Overview
T [°C]NTC 30:
Accuracy ± 0.1
PT 1000:
AccuracyOperating temperature
± 0.5 0 °C ... 15 °C
± 0.1 15 °C ... 35 °C
± 1 35 °C ... 55 °C
Correction
function
Temperature input
Manually [°C]- 20 ... + 130
Analog outputAutomatic switchover if a recorder is connected.
Output signalCorresponds to the electrode voltage,
i.e. 50-62 mV/pH unit
Accuracy± 0.1 % of the measured value ± 4 mV
Internal resistance< 5 Ohm (current limited to max. 0.2 mA
output current)
Serial interfaceAutomatic switchover if a PC or printer is connected.
TypeRS232, data output
Baud rateCan be set to
1200, 2400, 4800, 9600 Baud
Data bits8
Stop bit2
ba75368e0401/2011
ParityNone
HandshakeRTS/CTS + Xon/Xoff
Cable length Max. 15 m
9
Overviewhandylab pH 12
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
Power supply
Applicable
guidelines and
norms
Batteries4 x 1.5 V alkali-manganese batteries, Type
AA
Operational lifeApprox. 5000 operating hours
Mains operationFRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1.5 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
EMCEC guideline 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Instrument safetyEC guideline 2006/95/EC
EN 61010-1
ANSI/UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Climatic classVDI/VDE 3540
Type of protectionEN 60529
10
ba75368e0401/2011
handylab pH 12Safety
2Safety
This operating manual contains basic instructions that you must follow during
the commissioning, operation and maintenance of the pH meter. Consequently, all responsible personnel must read this operating manual before
working with the measuring system. The operating manual must always be
available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in the lab-
oratory.
We assume that, as a result of their professional training and experience, the
operators will know the necessary safety precautions to take when handling
chemicals.
Safety
instructions
Further notes
The individual chapters of this operating manual use safety instructions such
as the label shown below to indicate various hazards or dangers:
Caution
indicates instructions that must be followed precisely in order to avoid the
possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of the
pH and ORP measurement in the field and laboratory.
The technical specifications as given in section 1.4 T
observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized.
Any other use is considered to be unauthorized.
ECHNICAL DATA must be
ba75368e0401/2011
11
Safetyhandylab pH 12
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines
and norms for electronic measuring instruments (see section 1.4 T
D
ATA).
It left the factory in a safe and secure technical condition.
ECHNICAL
Function and oper-
ating safety
The smooth functioning and operational safety of the measuring instrument
can only be guaranteed if the generally applicable safety measures and the
specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the measuring instrument
can only be guaranteed under the environmental conditions that are specified
in section 1.4 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the instrument reaches
room temperature before putting the instrument back into operation.
Safe operationIf safe operation is no longer possible, the instrument must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of time
z is visibly damaged
Obligations of the
purchaser
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following
laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
12
ba75368e0401/2011
handylab pH 12Commissioning
3Commissioning
3.1Scope of delivery
z Handheld meter, handylab pH 12
z Operating manual and short operating manual
z 4 batteries, 1.5 V Mignon type AA (in the instrument)
3.2Initial commissioning
Perform the following activities:
z Set the date and time
z Connect the plug-in power supply.
Setting the date
and time
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date ap-
pears on the display.
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
ba75368e0401/2011
12Set the current time with <▲> <▼>.
13Confirm with <RUN/ENTER>.
The instrument switches to the pH measuring mode.
13
Commissioninghandylab pH 12
Ref
2
1
3
Connecting the
plug-in power sup-
ply (option)
You can either operate the measuring instrument with batteries or with the
plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (12 V DC). This saves the batteries.
Caution
The line voltage at the operating site must lie within the input voltage range
of the original plug-in power supply (see
section 1.4 T
ECHNICAL DATA).
Caution
Use original plug-in power supplies only (see
section 1.4 T
ECHNICAL DATA).
1Plug the jack (1) into the socket (2) of the measuring instru-
ment.
2Connect the original SI Analytics plug-in power supply (3) to an
easily accessible mains socket.
Note
You can also measure without the plug-in power supply.
14
ba75368e0401/2011
handylab pH 12Operation
4Operation
4.1Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
Subsequently, the slope and asymmetry used appear for approx. one second one after the other. In addition, the calibration
procedure of the last calibration is shown (AutoCal TEC or AutoCal DIN or no display in the delivery state or following initialization).
The measuring instrument then automatically switches to the
measuring mode that was last selected.
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the measuring instrument off if no key has been pressed for an hour. The energy saving
feature is not active when the AutoStore function is active.
The energy saving feature is also not active
z if the power is supplied by the plug-in power supply,
z if the communication cable and a PC with a running communication pro-
gram are connected,
z if the recorder cable is connected,
z if the printer cable is connected (for external printers).
ba75368e0401/2011
15
Operationhandylab pH 12
4.2Measuring
4.2.1General information
You can measure the following variables:
z pH value
z ORP voltage
Preparatory
activities
Perform the following preparatory activities when you want to measure:
1Connect the electrode to the measuring instrument.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3In conjunction with the electrode, check or calibrate the mea-
suring instument. How to calibrate is described in section 4.3.
4Select the measuring mode with <M>.
Caution
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232 interface is not galvanically isolated.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values.
Calibrate regularly before measuring.
16
Temperature
sensor
You can measure with or without a temperature sensor. If a temperature sensor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the temperature sensor used. Therefore, you can connect temperature sensors of the
NTC30 or Pt1000 type.
ba75368e0401/2011
handylab pH 12Operation
mV/pH
mol/L
8
Auto
Store
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
29
9
6
°C
TP
pH
The temperature measurement is absolutely essential for a reproducible pH
measurement. If the measurement is made without a temperature sensor,
proceed as follows:
1Measure the current temperature of the test sample using a
thermometer.
2Enter the temperature on the instrument: Set the temperature
value of the test sample using <▲ > <▼>.
4.2.2Measuring the pH value
1Perform the preparatory activities according to section 4.2.1.
2Immerse the pH electrode in the test sample.
Setting the
resolution
3Press the <M> key until pH appears on the status display. The
pH value appears on the display.
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the <M> key
while pressing the <RUN/ENTER> key.
ba75368e0401/2011
17
Operationhandylab pH 12
AutoRead
(drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
ReproducibilityResponse time
Better than 0.02> 30 seconds
1Call up the pH measuring mode with <M> and <▲> <▼> .
2Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
18
ba75368e0401/2011
handylab pH 12Operation
/pH
ConCal
Auto
Store
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
8
pH
1
%
°F
S
Time
Baud
Day.MonthNo.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
3
1
2
2
0
mV
°C
U
TP
1
7
4.2.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring instrument can measure the ORP voltage (U) of a solution.
1Perform the preparatory activities according to section 4.2.1.
2Immerse the ORP electrode in the test sample.
3Press the <M> key until U appears on the status display.
The ORP voltage (mV) of the test sample appears on the display.
4Wait for a stable measured value.
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes
using a test solution.
ba75368e0401/2011
19
Operationhandylab pH 12
4.3Calibrating
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines the current values of the asymmetry and slope of the electrode
and stores them in the measuring instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z When the sensor symbol flashes:
– after the calibration interval has expired
– after a voltage interruption, e. g. after changing the batteries
Calibration intervalThe calibration interval (Int 3) is set to 7 days in the factory. You can select
the interval in the range of 1 ... 999 days (see section 4.6 C
You can select between 3 calibration procedures:
ONFIGURATION).
AutoCal TECis specially adapted to the SI Analytics technical buffer solutions as a fully au-
tomatic two-point or three-point calibration. The buffer solutions are auto-
matically recognized by the measuring instrument. Depending on the
instrument setting (see section 4.6 C
ONFIGURATION), the instrument displays
the relevant buffer nominal value or the current electrode voltage in mV. The
calibration can be terminated after the first buffer solution. This corresponds
to a single-point calibration. For this, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoCal DINis specially adapted to the permanently programmed buffer solutions in ac-
cordance with DIN19266 as a fully automatic two-point or three-point cali-bration. The buffer solutions are automatically recognized by the measuring
instrument. Depending on the instrument setting (see section
4.6 C
ONFIGURATION), the instrument displays the relevant buffer nominal val-
ue or the current electrode voltage in mV. The calibration can be terminated
after the first buffer solution. This corresponds to a single-point calibration.
For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
ConCalThis function is a conventional two-point calibration using two buffer solu-
tions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point calibra-
tion using any buffer solution and is used as a high-speed method.
AutoReadWhen calibrating with AutoCal TEC and AutoCal DIN, the AutoRead function
Displaying
calibration data
20
is automatically activated.
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
Each time the instrument is switched on, the calibration data are shown on
the display for a short time (see section 4.1 S
STRUMENT). In order to view the calibration data, switch the measuring instru-
The calibration protocol contains the calibration data of the current calibration.
You can transmit the calibration protocol to a printer via the serial interface
(see page 42).
Note
You can automatically print a calibration protocol after the calibration. To do
so, connect a printer to the interface according to section 4.5.3 before calibrating. After a valid calibration, the record is printed.
Sample printout:
ba75368e0401/2011
21
Loading...
+ 47 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.