WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept
any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are discovered within the warranty period. The warranty does not cover components that are replaced during maintenance work, e.g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper handling or unauthorized opening of the instrument invalidates any warranty claim.
The compact precision handheld meter handylab pH 12 enables you to carry
out pH measurements rapidly and reliably.
The handylab pH 12 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven calibration procedures and the special AutoRead function support you in your work with the handylab pH 12.
1Keypad
2Sample display
3Jack field
ba75368e0401/2011
5
Overviewhandylab pH 12
CAL
RUN/ENTER
AR
STO
RCL
M
M
STO
CAL
RCL
AR
RUN/ENTER
1.1Keypad
Key functions
Select the measuring mode <M>:
–pH value
–ORP voltage
Save a measured value
<STO>
Switch measuring instrument on/off
<ON/OFF>
Calibration of the current measured variable
<CAL>
Display/transmit measured values
<RCL>
Activate/deactivate the AutoRead function
<AR>
Select the measuring mode
Increase values, scroll
<▲>
6
Select the measuring mode
Decrease values, scroll
<▼ >
Confirm entries, start AutoRead
<RUN/ENTER>
ba75368e0401/2011
handylab pH 12Overview
%
°F
S
TempError
mg/L
% ppm
mol/L
ISE
RCL
8
8.
8
88
8
8
mV/pH
°C
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
TP
1
UpH
Time
Baud
Day.Month No.
Ident
Year
Auto
Store
ConCal
Status display
Meas. value display
Function and
temperature display
Sensor symbol
R
e
f
5
123
4
1.2Display
1.3Jack field
The meter has either a DIN socket or a BNC socket.
The figure shows a meter with DIN socket (1).
ba75368e0401/2011
1pH electrode or single-rod measuring electrode
2Reference electrode
3Plug-in power supply (optional)
4Temperature probe
5Serial interface RS 232 / analog output (recorder)
Caution
Only connect electrodes to the measuring instrument that cannot return any
voltages or currents that are not allowed
(> SELV and > current circuit with current limiting).
Nearly all electrodes - especially SI Analytics electrodes - fulfill these conditions.
7
Overviewhandylab pH 12
1.4Technical Data
Dimensions
and weight
Mechanical
structure
Electrical safety
Test certificates
Ambient
conditions
Length [mm]172
Width [mm]80
Height [mm]37
Weight [kg]Approx. 0.3
Type of protectionIP 66
Protective classIII
cETLus, CE
Storage- 25 °C ... + 65 °C
Operation-10 °C ... + 55 °C
Relative atmospheric
< 90 % of annual average
humidity
pH/ORP
measuring ranges
Precision of
pH/ORP
(± 1 digit)
Measuring rangeResolution
pH- 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV]- 999.9 ... + 999.9
- 1999 ... + 1999
0.001
0.01
0.1
1
T [°C]- 5.0 ... + 105.00.1
pH
(± 2 pH units from the
± 0.005 at + 15 °C ... + 35 °C
± 0.01
calibration point)
U [mV]± 0.3 at + 15 °C ... + 35 °C
± 1
8
ba75368e0401/2011
handylab pH 12Overview
T [°C]NTC 30:
Accuracy ± 0.1
PT 1000:
AccuracyOperating temperature
± 0.5 0 °C ... 15 °C
± 0.1 15 °C ... 35 °C
± 1 35 °C ... 55 °C
Correction
function
Temperature input
Manually [°C]- 20 ... + 130
Analog outputAutomatic switchover if a recorder is connected.
Output signalCorresponds to the electrode voltage,
i.e. 50-62 mV/pH unit
Accuracy± 0.1 % of the measured value ± 4 mV
Internal resistance< 5 Ohm (current limited to max. 0.2 mA
output current)
Serial interfaceAutomatic switchover if a PC or printer is connected.
TypeRS232, data output
Baud rateCan be set to
1200, 2400, 4800, 9600 Baud
Data bits8
Stop bit2
ba75368e0401/2011
ParityNone
HandshakeRTS/CTS + Xon/Xoff
Cable length Max. 15 m
9
Overviewhandylab pH 12
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
Power supply
Applicable
guidelines and
norms
Batteries4 x 1.5 V alkali-manganese batteries, Type
AA
Operational lifeApprox. 5000 operating hours
Mains operationFRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1.5 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
EMCEC guideline 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Instrument safetyEC guideline 2006/95/EC
EN 61010-1
ANSI/UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Climatic classVDI/VDE 3540
Type of protectionEN 60529
10
ba75368e0401/2011
handylab pH 12Safety
2Safety
This operating manual contains basic instructions that you must follow during
the commissioning, operation and maintenance of the pH meter. Consequently, all responsible personnel must read this operating manual before
working with the measuring system. The operating manual must always be
available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in the lab-
oratory.
We assume that, as a result of their professional training and experience, the
operators will know the necessary safety precautions to take when handling
chemicals.
Safety
instructions
Further notes
The individual chapters of this operating manual use safety instructions such
as the label shown below to indicate various hazards or dangers:
Caution
indicates instructions that must be followed precisely in order to avoid the
possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of the
pH and ORP measurement in the field and laboratory.
The technical specifications as given in section 1.4 T
observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized.
Any other use is considered to be unauthorized.
ECHNICAL DATA must be
ba75368e0401/2011
11
Safetyhandylab pH 12
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines
and norms for electronic measuring instruments (see section 1.4 T
D
ATA).
It left the factory in a safe and secure technical condition.
ECHNICAL
Function and oper-
ating safety
The smooth functioning and operational safety of the measuring instrument
can only be guaranteed if the generally applicable safety measures and the
specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the measuring instrument
can only be guaranteed under the environmental conditions that are specified
in section 1.4 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the instrument reaches
room temperature before putting the instrument back into operation.
Safe operationIf safe operation is no longer possible, the instrument must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of time
z is visibly damaged
Obligations of the
purchaser
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following
laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
12
ba75368e0401/2011
handylab pH 12Commissioning
3Commissioning
3.1Scope of delivery
z Handheld meter, handylab pH 12
z Operating manual and short operating manual
z 4 batteries, 1.5 V Mignon type AA (in the instrument)
3.2Initial commissioning
Perform the following activities:
z Set the date and time
z Connect the plug-in power supply.
Setting the date
and time
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date ap-
pears on the display.
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
ba75368e0401/2011
12Set the current time with <▲> <▼>.
13Confirm with <RUN/ENTER>.
The instrument switches to the pH measuring mode.
13
Commissioninghandylab pH 12
Ref
2
1
3
Connecting the
plug-in power sup-
ply (option)
You can either operate the measuring instrument with batteries or with the
plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (12 V DC). This saves the batteries.
Caution
The line voltage at the operating site must lie within the input voltage range
of the original plug-in power supply (see
section 1.4 T
ECHNICAL DATA).
Caution
Use original plug-in power supplies only (see
section 1.4 T
ECHNICAL DATA).
1Plug the jack (1) into the socket (2) of the measuring instru-
ment.
2Connect the original SI Analytics plug-in power supply (3) to an
easily accessible mains socket.
Note
You can also measure without the plug-in power supply.
14
ba75368e0401/2011
handylab pH 12Operation
4Operation
4.1Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
Subsequently, the slope and asymmetry used appear for approx. one second one after the other. In addition, the calibration
procedure of the last calibration is shown (AutoCal TEC or AutoCal DIN or no display in the delivery state or following initialization).
The measuring instrument then automatically switches to the
measuring mode that was last selected.
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the measuring instrument off if no key has been pressed for an hour. The energy saving
feature is not active when the AutoStore function is active.
The energy saving feature is also not active
z if the power is supplied by the plug-in power supply,
z if the communication cable and a PC with a running communication pro-
gram are connected,
z if the recorder cable is connected,
z if the printer cable is connected (for external printers).
ba75368e0401/2011
15
Operationhandylab pH 12
4.2Measuring
4.2.1General information
You can measure the following variables:
z pH value
z ORP voltage
Preparatory
activities
Perform the following preparatory activities when you want to measure:
1Connect the electrode to the measuring instrument.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3In conjunction with the electrode, check or calibrate the mea-
suring instument. How to calibrate is described in section 4.3.
4Select the measuring mode with <M>.
Caution
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232 interface is not galvanically isolated.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values.
Calibrate regularly before measuring.
16
Temperature
sensor
You can measure with or without a temperature sensor. If a temperature sensor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the temperature sensor used. Therefore, you can connect temperature sensors of the
NTC30 or Pt1000 type.
ba75368e0401/2011
handylab pH 12Operation
mV/pH
mol/L
8
Auto
Store
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
29
9
6
°C
TP
pH
The temperature measurement is absolutely essential for a reproducible pH
measurement. If the measurement is made without a temperature sensor,
proceed as follows:
1Measure the current temperature of the test sample using a
thermometer.
2Enter the temperature on the instrument: Set the temperature
value of the test sample using <▲ > <▼>.
4.2.2Measuring the pH value
1Perform the preparatory activities according to section 4.2.1.
2Immerse the pH electrode in the test sample.
Setting the
resolution
3Press the <M> key until pH appears on the status display. The
pH value appears on the display.
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the <M> key
while pressing the <RUN/ENTER> key.
ba75368e0401/2011
17
Operationhandylab pH 12
AutoRead
(drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
ReproducibilityResponse time
Better than 0.02> 30 seconds
1Call up the pH measuring mode with <M> and <▲> <▼> .
2Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
18
ba75368e0401/2011
handylab pH 12Operation
/pH
ConCal
Auto
Store
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
8
pH
1
%
°F
S
Time
Baud
Day.MonthNo.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
3
1
2
2
0
mV
°C
U
TP
1
7
4.2.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring instrument can measure the ORP voltage (U) of a solution.
1Perform the preparatory activities according to section 4.2.1.
2Immerse the ORP electrode in the test sample.
3Press the <M> key until U appears on the status display.
The ORP voltage (mV) of the test sample appears on the display.
4Wait for a stable measured value.
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes
using a test solution.
ba75368e0401/2011
19
Operationhandylab pH 12
4.3Calibrating
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines the current values of the asymmetry and slope of the electrode
and stores them in the measuring instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z When the sensor symbol flashes:
– after the calibration interval has expired
– after a voltage interruption, e. g. after changing the batteries
Calibration intervalThe calibration interval (Int 3) is set to 7 days in the factory. You can select
the interval in the range of 1 ... 999 days (see section 4.6 C
You can select between 3 calibration procedures:
ONFIGURATION).
AutoCal TECis specially adapted to the SI Analytics technical buffer solutions as a fully au-
tomatic two-point or three-point calibration. The buffer solutions are auto-
matically recognized by the measuring instrument. Depending on the
instrument setting (see section 4.6 C
ONFIGURATION), the instrument displays
the relevant buffer nominal value or the current electrode voltage in mV. The
calibration can be terminated after the first buffer solution. This corresponds
to a single-point calibration. For this, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoCal DINis specially adapted to the permanently programmed buffer solutions in ac-
cordance with DIN19266 as a fully automatic two-point or three-point cali-bration. The buffer solutions are automatically recognized by the measuring
instrument. Depending on the instrument setting (see section
4.6 C
ONFIGURATION), the instrument displays the relevant buffer nominal val-
ue or the current electrode voltage in mV. The calibration can be terminated
after the first buffer solution. This corresponds to a single-point calibration.
For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
ConCalThis function is a conventional two-point calibration using two buffer solu-
tions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point calibra-
tion using any buffer solution and is used as a high-speed method.
AutoReadWhen calibrating with AutoCal TEC and AutoCal DIN, the AutoRead function
Displaying
calibration data
20
is automatically activated.
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
Each time the instrument is switched on, the calibration data are shown on
the display for a short time (see section 4.1 S
STRUMENT). In order to view the calibration data, switch the measuring instru-
The calibration protocol contains the calibration data of the current calibration.
You can transmit the calibration protocol to a printer via the serial interface
(see page 42).
Note
You can automatically print a calibration protocol after the calibration. To do
so, connect a printer to the interface according to section 4.5.3 before calibrating. After a valid calibration, the record is printed.
Sample printout:
ba75368e0401/2011
21
Operationhandylab pH 12
Calibration
evaluation
After the calibration, the measuring instrument automatically evaluates the
current status of the electrode. The asymmetry and slope are evaluated separately. The worst evaluation appears on the display.
DisplayAsymmetry
[mV]
Slope
[mV/pH]
-15 ... +15-60.5 ... -58
-20 ... +20-58 ... -57
-25 ... +25-61 ... -60.5
or
-57 ... -56
-30 ... +30-62 ... -61
or
-56 ... -50
Preparatory
activities
Clean the electrode according
to the electrode operating
manual
E3
Eliminate the error according to
chapter 6 W
HATTODOIF...
< -30 or
> 30
< -62 or
> -50
1Switch on the measuring instrument with <ON/OFF>.
2Connect the pH electrode to the measuring instrument.
3Keep the buffer solutions ready.
4Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without temperature sensor.
22
ba75368e0401/2011
handylab pH 12Operation
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
t
°C
TP
pH
AutoCal TEC
4.3.1AutoCal TEC
For this procedure, use any two or three technical buffer solutions in ascending or descending order
(pH values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the SI Analytics technical buffer
solution pH 10.01. Other buffer solutions can lead to an erroneous calibration. The correct buffer solutions are given in the SI Analytics catalog or in the
Internet.
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section 4.6 C
the instrument displays the relevant buffer nominal value or the current electrode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
ONFIGURATION),
1Press the <CAL> key repeatedly until the Ct1 display indicator
and the function display AutoCal TEC appears. The sensor
symbol displays the evaluation of the last calibration (or no
sensor symbol in the delivery state or after the measurement
parameter has been reset).
2If necessary, enter the temperature of the first buffer solution
with <▲> <▼>.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ba75368e0401/2011
23
Operationhandylab pH 12
cm
3
7
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
0
03
7
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
88
44
22
TP
TP
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2C
t
°C
TP
pH
AutoCal TEC
5When the measured value is stable, Ct2 appears.
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This
corresponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the
electrode.
6To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
7If necessary, enter the temperature of the second buffer solu-
tion with <▲> <▼>.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
24
ba75368e0401/2011
handylab pH 12Operation
1
TDS
S
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDS
1
S
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
88
44
08
14
4
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
1
22
TP
TP
O
L
S
RCL
Sal
cm
AutoCal DIN
Day.Month
S
Tref20
Cal
Ident
No.
Year
Time
Baud
1
TDS
Sal
Store
AR
Tref25
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
Lin
OU
nLF
Auto
LoBat
ARng
°
F
9
4
pH
5
mV/pH
AutoCal TEC
YA
S
0
RCL
Sal
cm
AutoCal DIN
Time
Day.Month
Sal
Store
AR
8
No.
Ident
Tref20
Tref25
Year
Baud
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
TDS
6
Lin
Cal
1
S
OU
nLF
Auto
LoBat
ARng
°
F
2
pH
mV
AutoCal TEC
10When the measured value is stable, AR disappears. The in-
strument displays the value of the slope (mV/pH) and the evaluation of the calibration in calibration range I.
ba75368e0401/2011
11Press the <RUN/ENTER> key. The instrument displays the
value of the asymmetry in calibration range I.
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This
corresponds to a two-point calibration.
The instrument uses the slope and asymmetry of the calibration
range I.
25
Operationhandylab pH 12
Three-point
calibration
12Press the <RUN/ENTER> key
to continue the three-point calibration.Ct3 appears on the display.
13Thoroughly rinse the electrode with distilled water.
14If necessary, enter the temperature of the third buffer solution
with <▲> <▼>.
15Immerse the pH electrode in the third buffer solution.
16Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display.
When the measured value is stable, AR disappears.
The instrument displays the value of the slope (mV/pH) and the
evaluation of the calibration in calibration range II.
17Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibration range II. The three-point calibration is completed.
18To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing <M>.
The values of the two-point calibration for slope and asymmetry will then remain stored.
26
ba75368e0401/2011
handylab pH 12Operation
AutoCalTEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
d
°C
TP
pH
AutoCal DIN
4.3.2AutoCal DIN
For this procedure, use two or three different standard buffer solutions according to DIN 19266 in ascending or descending order (type A, C, D or F
with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section 4.6 C
the instrument displays the relevant buffer nominal value or the current electrode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
1Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).
ONFIGURATION),
ba75368e0401/2011
2If necessary, enter the temperature of the buffer solution with
<▲> <▼>.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
27
Operationhandylab pH 12
AutoCalTEC
AutoCalTEC
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
Tref25
Auto
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
0
17
4
pHpH
C
°
C
°
ARAR
mV
88
44
22
TP
TP
7
1
AutoCal DIN
AutoCal DIN
AutoCalTEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2C
d
°C
TP
pH
AutoCal DIN
5When the measured value is stable, Cd2 appears.
Note
At this point, the AutoCal DIN calibration can be terminated with <M>. This
corresponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the
electrode.
28
6To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
7If necessary, enter the temperature of the second buffer solu-
tion with <▲> <▼>.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ba75368e0401/2011
handylab pH 12Operation
AutoCalTEC
AutoCalTEC
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
Tref25
Auto
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
1
89
9
pHpH
C
°
C
°
ARAR
mV
88
44
22
TP
TP
2
1
AutoCal DIN
AutoCal DIN
AutoCalTEC
YA
S
0
RCL
Sal
cm
Time
Day.Month
Sal
Store
AR
8
No.
Ident
Tref20
Tref25
Year
Baud
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
TDS
6
Lin
Cal
1
S
OU
nLF
Auto
LoBat
ARng
°
F
2
pH
mV
AutoCal DIN
10When the measured value is stable, AR disappears. The in-
strument displays the value of the slope (mV/pH) and the evaluation of the calibration in calibration range I.
ba75368e0401/2011
11Press the <RUN/ENTER> key. The instrument displays the
value of the asymmetry in calibration range I.
Note
At this point, the AutoCal DIN calibration can be terminated with <M>. This
corresponds to a two-point calibration.
The instrument uses the slope and asymmetry of the calibration
range I.
29
Operationhandylab pH 12
Three-point
calibration
12Press the <RUN/ENTER> key
to continue the three-point calibration.Cd3 appears on the display.
13Thoroughly rinse the electrode with distilled water.
14If necessary, enter the temperature of the third buffer solution
with <▲> <▼>.
15Immerse the pH electrode in the third buffer solution.
16Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display.
When the measured value is stable, AR disappears.
The instrument displays the value of the slope (mV/pH) and the
evaluation of the calibration in calibration range II.
17Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibration range II. The three-point calibration is completed.
18To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing <M>.
The values of the two-point calibration for slope and asymmetry will then remain stored.
30
ba75368e0401/2011
handylab pH 12Operation
4.3.3ConCal
Single-point
calibration
Two-point
calibration
Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the buffer
solution is to that of the test sample.
For this procedure, use two buffer solutions:
z pH 7.0 ± 0.5
z any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature sensor.
1Press the <CAL> key repeatedly until the display ASY and the
function display ConCal appears. The sensor symbol displays
the evaluation of the last calibration (or no sensor symbol in the
delivery state or after the measurement parameter has been
reset).
ba75368e0401/2011
2If necessary, enter the temperature of the first buffer solution
with <▲> <▼>.
3Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4Press the <RUN/ENTER> key.
The measured pH value appears on the display.
5Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
6Press the <RUN/ENTER> key. The value of the asymmetry
(mV) and the sensor symbol appear on the display.
7Press the <RUN/ENTER> key. SLO appears on the display.
Note
At this point, the ConCal calibration can be broken off with <M>. This corresponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the electrode.
31
Operationhandylab pH 12
AutoCal DIN
AutoCalTEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
OS
L
°C
TP
pH
ConCal
8To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
9If necessary, enter the temperature of the second buffer solu-
tion with <▲> <▼>.
10Immerse the pH electrode in the second buffer solution.
11Press the <RUN/ENTER> key. The second pH value appears
on the display.
12Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
13When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibration.
14Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
32
ba75368e0401/2011
handylab pH 12Operation
15To return to the measuring mode: Press the <M> key.
ba75368e0401/2011
33
Operationhandylab pH 12
Ident
°C
2
4
TP
mV/pH
mol/L
8
Auto
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
ConCal
Time
Baud
Day.Month
Year
1
U
%
°F
S
mg/L
% ppm
ISE
4
29
9
6
pH
Store
No.
4.4Saving
The measuring instrument has an internal data memory. It can store up to
800 data records.
A complete data record consists of:
z Storage location
z Date/time
z Measured value
z Temperature
z Temperature measuring procedure
z ID number
You can transmit measured values (data records) to the data storage in two
ways:
z Save manually
z Switch on AutoStore (Int 1).
4.4.1Saving manually
You can transmit a measured value to the data storage as follows:
1Press the <STO> key.
The current number (location number No.) of the next free storage location appears under the current measured value on the
display.
2Confirm with <RUN/ENTER>.
The display switches to entering the ID number.
34
ba75368e0401/2011
handylab pH 12Operation
No.
°C
2
4
TP
mV/pH
mol/L
8
Auto
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
ConCal
Time
Baud
Day.Month
Year
1
U
%
°F
S
mg/L
% ppm
ISE
2
29
9
6
pH
Store
Ident
3Using <▲> <▼>, enter the required ID number
(1 ... 999).
4Confirm with <RUN/ENTER>.
The instrument changes to the measuring mode.
Message StoFullThis message appears when all of the 800 storage locations are occupied.
You have the following options:
Saving the current measured value.
The oldest measured value (storage
location 1) will be overwritten by this
Returning to the measuring mode without
saving
Outputting the data storagesee
Clearing the memorysee
Press
<RUN/ENTER>.
press any key
section 4.4.3
section 4.4.4
ba75368e0401/2011
35
Operationhandylab pH 12
LoBat
Tref25nLF
/
K
%
RCL
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
S
8
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DINAutoCalTEC
Time
Baud
Day.MonthNo.
Ident
Year
F
F
O
t
1
n
1
Store
Auto
4.4.2Saving automatically
The save interval (Int 1) determines the chronological interval between automatic save processes.
After the fixed interval has expired, the current data record is transmitted to
the storage and to the interface.
Setting the save
interval
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min,
15 min, 30 min, 60 min):
1Press the <RUN/ENTER> key and hold it down.
2Press the <STO> key. Int 1 appears on the display.
3Set the required interval between the saving procedures with
<▲> <▼>.
36
4Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.
5As soon as all of the 800 storage locations are occupied,
AutoStore is terminated (Int 1 = OFF).
If there are not enough storage locations available for your
measurements:
– Output and backup the data storage (see page 38) and
– clear the memory (see page 43).
ba75368e0401/2011
handylab pH 12Operation
e
r
F
9
9
9
S
1
LoBat
Tref25nLF
/
K
%
RCL
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DINAutoCalTEC
Time
Baud
Day.MonthNo.
Year
1
Store
Auto
Ident
6Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
7Set the required ID number with <▲> <▼>.
8Confirm with <RUN/ENTER>.
The instrument switches to the pH measuring mode and starts
the measuring and saving process.
AutoStore flashes on the display.
Switching off
AutoStore
Note
The AutoStore function is interrupted if you start other functions, e.g. output
the data storage.
After the function is finished, the AutoStore function is continued. By this,
however, temporal gaps in the recording of the measured values will occur.
Switch AutoStore off by:
z setting the save interval (Int 1) to OFF, or
z switching the measuring instrument off and then on again.
ba75368e0401/2011
37
Operationhandylab pH 12
Store
Auto
Ident
S
1
LoBat
Tref25nLF
/
K
%
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DINAutoCalTEC
Time
Baud
Day.MonthNo.
Year
p
S
T
O
d
I
s
RCL
mV/pH
mol/L
8
Auto
Store
TempError
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
68
6
6
°C
TP
pH
RCL
4.4.3Outputting the data storage
You can output the contents of the data storage:
z Stored data on the display
z Calibration data on the display
z Stored data on the serial interface
z Calibration protocol to the interface
Outputting stored
data on the display
1Press the <RCL> key repeatedly until StO dISP appears on the
display.
2Press the <RUN/ENTER> key.
A measured value appears on the display.
The storage location of the data record is displayed for approx.
2 s, then the respective temperature appears.
38
You can perform the following activities:
Display further elements of the data record
(ID number, date, time, storage location)
Advance one data record (storage location) Press <▲>
Go back one data record (storage location)Press <▼>
Press
<RUN/ENTER>.
ba75368e0401/2011
handylab pH 12Operation
Store
Auto
Ident
S
1
LoBat
Tref25nLF
/
K
%
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DINAutoCalTEC
Time
Baud
Day.MonthNo.
Year
R
S
t
o
S
E
RCL
Note
If you want to search for a certain element (e.g. date), proceed as follows:
1Using <RUN/ENTER>, select the element (e.g. date).
2Press <▲> or <▼> repeatedly until the required date appears
on the display.
After approx. 2 s the temperature of the displayed measured
value appears.
Outputting stored
data to the
interface
1Press the <RCL> key repeatedly until Sto SEr appears on the
display.
2Press the <RUN/ENTER> key.
The complete storage content is transmitted to the interface;
during the data transmission the numbers of the currently
transmitted storage locations run through. After the data transmission, the measuring instrument automatically switches to
the pH measuring mode.
Note
You can cancel the transmission with <M> or <RUN/ENTER>.
ba75368e0401/2011
After the instrument number, the printout contains the complete storage contents in ascending order of the storage location numbers.
39
Operationhandylab pH 12
Device No.: 12345678
No. 1:
09.03.99 17:10
pH 10.01 25 °C
Tman AR
Ident : 1
No. 2:
09.03.99 17:11
pH 10.01 24.7 °C
Tauto AR
Ident : 1
No. 3:
09.03.99 17:12
305 mV
Tauto
Ident : 13
...
ARng
ConCal
Store
Auto
0
No.
LoBat
Day.Month
Year
Ident
UpH
AR
%
°F°C
TP
mV/pH
mol/L
TempError
AutoCal DIN
CalError
AutoCalTEC
Time
Baud
1
S
mg/L
% ppm
ISE
LC
A
pi
s
RCL
d
TP
°C
ConCal
AR
AutoCal DIN
mol/L
8
Auto
Store
LoBat
CalError
ARng
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
0
L
s
pH
45
9
AutoCal TEC
mV/pH
RCL
-
Sample printout:
Outputting the
calibration data
on the display
1Press the <RCL> key repeatedly until CAL dISP appears on
the display.
2Press the <RUN/ENTER> key.
The slope (mV/pH) in the calibration range I and the calibration
procedure of the last calibration appear on the display
40
ba75368e0401/2011
handylab pH 12Operation
3Press <RUN/ENTER> to display the value of the asymmetry
(mV) in the calibration range I.
4In two-point calibration:
Using <M> or <RUN/ENTER>, you can switch back to the pH
measuring mode.
In three-point calibration:
Press <RUN/ENTER> to display the slope [mV/pH] in the calibration range II (SLO2).
ba75368e0401/2011
5Press <RUN/ENTER> to display the value of the asymmetry
(mV) in the calibration range II (ASY2).
6Using <M> or <RUN/ENTER>, you can switch back to the pH
measuring mode.
41
Operationhandylab pH 12
d
ARng
ConCal
Store
Auto
0
No.
LoBat
Day.Month
Year
Ident
UpH
AR
%
°F°C
TP
mV/pH
mol/L
TempError
AutoCal DIN
CalError
AutoCalTEC
Time
Baud
1
S
mg/L
% ppm
ISE
LC
A
rs
e
RCL
Outputting the
calibration
protocol on the
interface
1Press the <RCL> key repeatedly until CAL appears on the dis-
play.
2Press the <RUN/ENTER> key.
The calibration protocol is transmitted to the interface.
After the data transmission, the measuring instrument automatically switches to the last active measuring mode.
Note
You will find a sample calibration protocol in P
CORD, page 21.
RINTINGTHECALIBRATIONRE-
42
ba75368e0401/2011
handylab pH 12Operation
RCL
d
Auto
Ident
S
1
LoBat
Tref25nLF
/
K
%
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DINAutoCalTEC
Time
Baud
Day.MonthNo.
Year
r
S
t
o
c
l
Store
4.4.4Clearing the memory
With this function, you can delete stored data records. 800 storage locations
will then be available again.
Note
The Clear memory function only appears when there are data records stored
in the memory. Otherwise, the measuring instrument automatically switches
to the measuring mode.
Proceed as follows to clear all data records:
1Switch off the measuring instrument.
2Press the <STO> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display.
Subsequently, Sto clr appears.
ba75368e0401/2011
4Confirm the clearing process with <RUN/ENTER>.
Pressing any other key prevents the clearing, the data records
will remain stored.
Note
The calibration data remain stored and can be called up.
43
Operationhandylab pH 12
4.5Transmitting data
You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 36), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured values
are periodically output on the interface (see below).
z With the Output data storage function (page 38), calibration data or saved
measured values are output on the interface.
z Via the analog recorder output (page 46), measured values are output as
voltage values.
z With a communication kit (accessory), data can be transmitted bidirection-
ally (page 47).
Note
If you connect a recorder (analog output), the output on the digital interface
is switched off.
4.5.1Data transmission interval (Int 2)
The interval to the data transmission (Int 2) determines the chronological interval between automatic data transmissions. After the selected interval expires, the current data record is transmitted to the interface.
Note
When the AutoStore function is active, the data transmission is performed according to the setting of the save interval (Int 1). Set the save interval (Int 1)
to OFF to activate the Data transmission interval (Int 2).
44
ba75368e0401/2011
handylab pH 12Operation
Store
Auto
LoBat
Tref25nLF
/
K
%
RCL
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
S
8
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
OUpH
mV/pH
AutoCal DINAutoCalTEC
Time
Baud
Day.MonthNo.
Ident
Year
F
F
O
t
2
n
1
Setting the
Data transmission
interval
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min,
5 min, 10 min, 15 min, 30 min, 60 min):
1Press the <RUN/ENTER> key and hold it down.
2Press the <RCL> key. Int 2 appears on the display.
3Set the required interval between the saving procedures with
<▲> <▼>.
4Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pHmeasuring mode.
ba75368e0401/2011
45
Operationhandylab pH 12
1 free
2 Plug coding
3 Ground
4 Analog output
4
(internal resistance < 5 Ohm)
4.5.2Recorder (analog output)
You can transmit data to a recorder via the analog output. Connect the analog output to the recorder via the Z 394 interface cable.
The data output automatically switches to Recorder output.
Socket
assignment
2
1
3
RS 232
REC
Note
The analog output is activated automatically in the cable by connecting 2 and
3.
At the analog output, there is a low-impedance electrode signal (corresponding to 50 ... 62 mV/pH).
46
ba75368e0401/2011
handylab pH 12Operation
1 CTS
2 RxD
3 Ground
4TxD
4
4.5.3PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an external
printer.
Use the Z 395 (PC) or Z 391 (ext. printer) cable to connect the interface to
the instruments.
The data output automatically switches to the RS232 interface.
Caution
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
The handheld meter can be remotely controlled from a PC. This requires a
communication kit (communication cable + communication program) which is
available upon request. The instrument is then controlled via commands that
simulate keystrokes and request the current display contents.
Note
A more detailed description is provided within the scope of delivery of the
communication kit.
ba75368e0401/2011
47
Operationhandylab pH 12
Day.MonthNo.
Time
1
2
M
m
%
mg/l
cm
S/
cm
S
mbar
mV/pH
t
n
Store
Auto
LoBat
Tref25nLF
/
K
%
RCL
Sal
°
AR
TP
C
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
OUpH
AutoCal DINAutoCalTEC
Ident
Year
0
0
8
4
Baud
4.6Configuration
You can adapt the measuring instrument to your individual requirements. To
do this, the following parameters can be changed (the status on delivery is
marked in bold):
Baud rate1200, 2400, 4800, 9600
Baud rate
Display during the
pH calibration
Calibration interval (Int 3)1 ... 7 ... 999 d
Date/timeAny
pH resolution 0.01 or 0.001
Note
You can leave the configuration menu at any time with <M> . The parameters
that have already been changed are stored.
1Switch off the measuring instrument.
2Press the <M> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
Buffer nominal value, or current
electrode voltage
48
4Select the required Baud rate with <▲> <▼>.
5Confirm with <RUN/ENTER>.
CAL diSP appears on the display.
ba75368e0401/2011
handylab pH 12Operation
2
d
°C
TP
mV/pH
mol/L
Auto
Store
RCL
LoBat
CalError
ARng
AR
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
3n
t
pH
Time
AutoCal DIN
AutoCal TEC
ConCal
I
d
S
1
M
m
%
mg/l
cm
S/
cm
mbar
mV/pH
D
i
s
Store
Auto
Ident
LoBat
Tref25nLF
/
K
%
Sal
°
AR
TP
C
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
OUpH
AutoCal DINAutoCalTEC
Time
Baud
No.
Year
4
0
9
0
Day.Month
Display during the
pH calibration
6Select the required display during the pH calibration with
<▲> <▼>.
mV: Display of the current electrode voltage
/pH: Display of the buffer nominal value.
7Confirm with <RUN/ENTER>. Int 3 appears on the display.
Calibration interval
Date and time
8Set the required interval in days with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The date flashes on the display.
ba75368e0401/2011
10Set the date of the current day with <▲> <▼>.
49
Operationhandylab pH 12
11Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
12Set the current month with <▲> <▼>.
13Confirm with <RUN/ENTER>.
The year appears on the display.
14Set the current year with <▲> <▼>.
15Confirm with <RUN/ENTER>.
The hours flash on the display.
16Set the current time with <▲> <▼>.
17Confirm with <RUN/ENTER>.
The minutes flash on the display.
18Set the current time with <▲> <▼>.
19Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH
measuring mode.
Setting the
resolution
1Press the <RUN/ENTER> key and hold it down.
2Press the <M> key.
The measured values are displayed with a high resolution, e.g.
pH = 4.012.
3Press the <RUN/ENTER> and <M> keys once more.
The measured values are displayed with a low resolution, e.g.
pH = 4.01.
50
ba75368e0401/2011
handylab pH 12Operation
4.7Reset
You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.
Measurement
parameters
Configuration
parameters
The following measured parameters (pH InI) are reset to the default condition:
Measuring modepH
Asymmetry0 mV
Slope -59.16 mV/pH
Calibration proceduresAutoCal TEC
Temperature, manual25 °C
Resolution (pH display)0.01 (low resolution)
Display during the
Buffer nominal value
pH calibration
Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.
The following configuration parameters (InI) are reset to the delivery status:
Baud rate4800
ba75368e0401/2011
Interval 1
(automatic save)OFF
Interval 2
(for data transmission)OFF
51
Operationhandylab pH 12
Tref25
Tref20
AutoCalTEC
RCL
LoBat
Cal
ARng
AR
Lin
nLF
Store
Sal
°
°
TP
cm
1/
F
C
%
/
K
8
8
1
S/
cm
m
M
TDS
Sal
S
Auto
mbar
%
mg/l
O
mV/pH
cm
AutoCal DIN
Time
Baud
Day.MonthNo.
Ident
Year
o
n
n
I
I
pH
OUpH
ARng
Y
Sal
Sal
/
K
%
TP
C
°
°
F
Store
cm
1/
1
S
M
m
%
mg/l
cm
S/
cm
mV/pH
mbar
Baud
4
Day.MonthNo.
Time
Auto
LoBat
Tref25nLF
RCL
AR
Cal
Lin
Tref20
TDS
AutoCal DIN
AutoCalTEC
Ident
Year
o
n
I
n
I
Resetting the
measuring
parameters
1Press the <RUN/ENTER> key and hold it down.
2Press the <CAL> key.
3Using <▲> <▼>, switch between no and YES.
YES: Resetting the measuring parameters
no: Retaining settings.
4Confirm with <RUN/ENTER>.
The measuring instrument switches to the configuration parameters.
Resetting the
configuration
parameters
5Using <▲> <▼>, switch between no and YES.
YES: Resetting the configuration parameters
no: Retaining settings.
6Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH
measuring mode.
52
ba75368e0401/2011
handylab pH 12Maintenance, cleaning, disposal
1
2
5Maintenance, cleaning, disposal
5.1Maintenance
The measuring instrument is almost maintenance-free.
The only maintenance task is replacing the batteries.
LoBat indicates that the batteries should be changed. The batteries are then
largely depleted.
Replacing the
batteries
1Open the housing after the instrument has been switched off:
– Undo the four screws on the underside of the instrument
– Pull down the lower cover (1).
ba75368e0401/2011
2If necessary, take the four depleted batteries (2) out of the bat-
tery compartment.
3Place four new batteries (type Mignon AA) in the battery com-
partment.
4Close the lower cover (1).
Caution
Make sure that the poles of the batteries are the right way round.
The
±signs on the batteries must correspond to the ± signs in the battery
compartment.
Only use leakproof alkaline manganese batteries.
53
Maintenance, cleaning, disposalhandylab pH 12
Note
For the maintenance of the electrodes, follow the corresponding operating
manual.
5.2Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lintfree cloth. Disinfect the housing with isopropanol as required.
Caution
The housing is made of a synthetic material (ABS). Thus, avoid contact with
acetone and similar detergents that contain solvents. Remove any splashes
immediately.
5.3Disposal
PackingThis measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
instrument against damage during transport.
BatteriesThis note refers to the battery regulation that applies in the Federal Republic
of Germany. We would ask end-consumers in other countries to follow their
local statutory provisions.
Note
This instrument contains batteries. Batteries that have been removed must
only be disposed of at the recycling facility set up for this purpose or via the
retail outlet.
It is illegal to dispose of them in household refuse.
Measuring
instrument
Dispose of the measuring instrument as electronic waste at an appropriate
collection point. It is illegal to dispose of the instrument in household refuse.
54
ba75368e0401/2011
handylab pH 12What to do if...
6What to do if...
Error message 0FL
CauseRemedy
pH electrode:
– Not connected– Connect electrode
Error message E3
– Air bubble in front of the
– Remove air bubble
diaphragm
– Air in the diaphragm– Extract air or
moisten diaphragm
– Cable broken– Replace electrode
– Gel electrolyte dried out– Replace electrode
CauseRemedy
Electrode
– Diaphragm contaminated– Clean diaphragm
– Membrane contaminated– Clean membrane
– Moisture in the plug– Dry plug
– Electrolyte out of date– Replenish electrolyte or
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low– Use suitable electrode
– Temperature too high– Use suitable electrode
– Organic liquids– Use suitable electrode
CauseRemedy
– Batteries almost empty– Replace batteries
(see section 5.1 MAINTENANCE)
56
ba75368e0401/2011
handylab pH 12What to do if...
Obviously incor-
rect
measured values
Instrument does
not react to key-
stroke
CauseRemedy
pH electrode:
– pH electrode unsuitable– Use suitable electrode
– Temperature difference
between buffer and test
– Adjust temperature of buffers
or test samples
sample too high
– Measurement procedure not
– Follow special procedure
suitable
CauseRemedy
– Operating condition
undefined or EMC load
unallowed
– Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Display to
Probe symbol
flashes
Message StoFull
CauseRemedy
– Time-out of the interface– Check the instrument that is
connected
CauseRemedy
– Calibration interval expired– Recalibrate the measuring
system
CauseRemedy
– All memory locations are full– Output data storage and clear
data storage
ba75368e0401/2011
57
What to do if...handylab pH 12
58
ba75368e0401/2011
handylab pH 12Lists
7Lists
This chapter provides additional information and orientation aids.
AbbreviationsThe list of abbreviations explains the indicators and the abbreviations that ap-
pear on the display and in the manual.
Specialist termsThe glossary briefly explains the meaning of the specialist terms. However,
terms that should already be familiar to the target group are not described
here.
IndexThe index will help you to find the topics that you are looking for.
ba75368e0401/2011
59
Listshandylab pH 12
Abbreviations
ARAutoRead (drift control)
ARngAutomatic range switching
Measuring instrument measures with highest resolution
ASYAsymmetry
AutoCal DINAutomatic pH calibration with buffer solutions prepared
according to DIN 19 266
AutoCal TECAutomatic pH calibration with SI Analytics technical buf-
fer solutions according to DIN 19267
°CTemperature unit, degrees Celsius
CalCalibration
Cd...Display indicator during calibration for pH measure-
ments. Indicates the selection of the buffer data record
for buffer solutions prepared according to DIN 19 266
ConCalConventional single-point or two-point calibration for pH
measurements
Ct...Display indicator during calibration for pH measure-
ments. Indicates the selection of the buffer data records
for SI Analytics technical buffer solutions
InIInitialization
Resets individual basic functions to the status they had
on delivery
LoBatBatteries almost empty(Low Battery)
mVVoltage unit
mV/pHUnit of the electrode slope (internat. mV)
OFLDisplay range exceeded (Overflow)
pHpH value
SSlope (internat. k)
SELVSafety Extra Low Voltage
SLOSlope setting on calibration
60
TPTemperature measurement active (Temperature Probe)
U
ASY
Asymmetry
ba75368e0401/2011
handylab pH 12Lists
Glossary
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains within the tolerance.
AsymmetryDesignation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is
immersed in a solution with the pH of the nominal electrode zero point
(electrodes: pH = 7).
AutoRangeName of the automatic selection of the measuring range.
AutoReadName for a function to check the stability of the measured value.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
DiaphragmThe junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between
two solutions and makes electrolyte exchange more difficult. The expression, junction, is also used for ground or junction-less transitions.
Electrode zero pointThe zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero.
Normally, this is at 25 °C.
Electromotive force of
an electrode
The electromotive force U of the electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the
galvanic voltages of the electrode. Its dependency on the pH results
in the electrode function which is characterized by the parameters,
slope and zero point.
Measured parameterThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured valueThe measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring systemThe measuring system comprises all the devices used for measuring,
e. g. measuring instrument and probe. In addition, there is the cable
and possibly an amplifier, terminal strip and armature.
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
MultiCal
ba75368e0401/2011
vent.
®
Name stating that a measuring instrument provides several calibration
procedures.
61
Listshandylab pH 12
Offset potentialThe measurable potential of a symmetrical electrode, the membrane
of which is immersed in a solution with the pH of the nominal electrode
zero point. The asymmetry is part of the offset potential.
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface
(e. g. a gold or platinum surface).
pH valueThe pH is a measure of the acidic or basic effect of an aqueous solu-
tion. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH
value is the value of a pH measurement.
PotentiometryName of a measuring technique. The signal (depending on the mea-
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
ResetRestoring the original condition of all settings of a measuring system.
ResolutionSmallest difference between two measured values that can be dis-
played by a measuring instrument.
SlopeThe slope of a linear calibration function.
Standard solutionThe standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Test sampleDesignation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
62
ba75368e0401/2011
handylab LF 12Index
Index
A
Analog output 46
Asymmetry 20
Authorized use 11
AutoCal DIN 20, 27
AutoCal TEC 20, 23
AutoRead 17
C
Calibration evaluation 22
Calibration procedures 20
Calibration protocol 21
ConCal 20, 31
Connecting a printer 47
Connecting the plug-in power
supply
14
D
Data record 34
Data transmission (interval)
45
Default condition 51
Display 7
Displaying calibration data
Operating safety 12
ORP electrode 19
ORP voltage 19
P
Printing
calibration protocol
21
R
Remote control 47
Replacing the batteries 53
Reset 51
RS232 interface 47
S
Safety 11
Safety precautions 11
Save interval 36
Scope of delivery 13
Setting the baud rate 48
Setting the date 13, 49
Setting the resolution 50
Setting the resolution (pH) 17
Setting the time 13, 49
Single-point calibration 20
Single-point calibration (pH)