Sirona Galileos Maintenance manual

d^ifiblp
fелйЙЕнбзе=~еЗ=г~бенЙе~еЕЙ
pбкзе~=aЙен~д=pулнЙгл=dгДe
Dear customer
You would like to have and will have many years of satisfaction with your páêçå~ X-ray unit.
Your dental dealership offers you service by specially trained engineers for this purpose. The maintenance should ensure that your product is permanently safe and operational. All components subject to normal wear and tear are checked and, if necessary, replaced. Maintenance work may be performed by the operator only if this is described on the part of Sirona Dental Systems GmbH, otherwise only authorized service engineers of Sirona Dental Systems GmbH or its authorized dealers may be entrusted with the work. In case you have not concluded a maintenance contract, please contact the customer service department of your dental dealership. The performed maintenance must be documented in this document, which must be kept with the unit.
We wish you much success and pleasure with your quality product from pfolk^=aЙен~д=pулнЙглK
61 25 533 D 3437 D 3437.105.01.03.02
Sirona Dental Systems GmbH
List of Contents
1 Yearly inspection by the system
owner....................................................5
1.1 Technical documents ............................... 5
1.2 System accessories ................................. 6
1.3 Surfaces of the unit .................................. 7
1.4 X-ray unit .................................................. 8
1.5 System safety .......................................... 8
1.6 Exposure safety ....................................... 9
1.7 Key functionality ..................................... 10
1.8 Light localizer functionality ..................... 10
1.9 Swivel arm/bite block/forehead
support functionality ............................... 11
1.10 Image quality .......................................... 11
1.11 Height adjustment functionality .............. 12
1.12 Documenting your yearly inspection ...... 12
ЗЙмнлЕЬЙеЦдблЬ
2 Yearly calibration by the service
engineer (for the USA refer to the
Maintenance Instructions)................13
2.1 Customer data ....................................... 13
2.2 Calibrating the X-ray unit ....................... 14
2.3 Documenting the yearly calibration ........ 15
3 Maintenance by the service
engineer .............................................16
3.1 Customer data ....................................... 16
3.2 Checking the records ............................. 17
3.3 Labels on the system ............................. 17
3.4 Serial numbers of the system ................ 18
3.5 Functional test ........................................ 19
3.6 Checking the tube data .......................... 20
3.7 Checking the cables for damage ........... 20
3.8 Safety checks ......................................... 21
3.9 Final work ............................................... 21
3.10 Documenting maintenance work ............ 22
61 25 533 D 3437 D 3437.105.01.03.02
3
Sirona Dental Systems GmbH
61 25 533 D 3437
4 D 3437.105.01.03.02

Sirona Dental Systems GmbH 1 Yearly inspection by the system owner

1 Yearly inspection by the system
owner
Inspection and maintenance must be performed at predetermined intervals to guarantee the health and safety of all patients, users and third parties.
In order to guarantee the operational safety and reliability of your product, you as the system owner should check your unit through at least once a year or commission your dental depot to do so.
In addition, your dental depot offers you maintenance of the system by specially trained engineers; see two Chapter in this document.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!
If one or more checks to be performed do not lead to satisfactory results, please contact your dental depot.

1.1 Technical documents

Technical documents Answer questions with yes (9) or no (−)
Date
(please enter month/year)
Operating Instructions available?
Installation Instructions available?
Installation Requirements availa­ble?
Installation Report / Warranty Passport available?
Declaration of Conformity by sys­tem integrator available?
X-ray System Logbook available (for Germany only)?
If no Ö Order missing documents from your dental depot.
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
5
1 Yearly inspection by the system owner Sirona Dental Systems GmbH

1.2 System accessories

System accessories (see Operating Instructions) Answer questions with yes (9) or no (−)
Not all accessories listed here may be included in the scope of supply, cross out what does not apply.
Date (please enter month/year)
Test phantom avail­able?
Forehead support available?
Rigid bite block available?
Head fastening strap available?
Chin rest, complete available?
Mandibular bite plate holder availa­ble?
Maxillary bite plate holder available?
If no Ö Order missing accessories from your dental depot.
61 25 533 D 3437
6 D 3437.105.01.03.02
Sirona Dental Systems GmbH 1 Yearly inspection by the system owner
Fig. 1:

1.3 Surfaces of the unit

Cover parts Answer questions with yes (9) or no (−)
Date
(please enter month/year)
All cover parts attached at X-ray unit (Fig. 1)?
All screws on unit housing availa­ble?
Cover parts and mirrors free of damage (no surface damages, e.g. scratches)?
All cover parts in clean condition?
Clean X-ray unit with the recom­mended cleaning agents (see Operating Instructions).
If no Ö Inform your dental depot.
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
7
1 Yearly inspection by the system owner Sirona Dental Systems GmbH
Fig. 1:
Tube assembly housing
Fig. 1: Fig. 2:
Power cable
Control cable
Anti-kink sleeve
2nd ground wire

1.4 X-ray unit

X-ray tube assembly and stand condition Answer questions with yes (9) or no (−)
Date (please enter month/year)
Tube assembly housing palpably/ visibly oil-free (Fig. 1)?
If no Ö Inform your dental depot.

1.5 System safety

Safety Answer questions with yes (9) or no (−)
Date
(please enter month/year)
Coiled cable with strain relief and anti-kink sleeve OK (Fig. 1)?
Power cable and 2nd ground wire free of damage (Fig. 1)?
Control cable free of damage (Fig. 1)?
All screws that fasten the X-ray unit to the wall and the support bracket tightly secured and/or unit securely fastened to the stand (Fig. 2)?
For Japan only: Is the unit imme­diately switched off when the emergency shutdown switch is activated?
If no Ö Immediately order repair by a service engineer.
8 D 3437.105.01.03.02
61 25 533 D 3437
Sirona Dental Systems GmbH 1 Yearly inspection by the system owner
Error message
Radiation indication
kV/mAs lights
Fig. 1:
Radiation indicationno longer lights up. Acoustic signal no longer sounding, when radiation button is released prematurely
Return key
Light localizer
GALILEOS Easypad
T
R
Prog.
kV
mAs
E1 10 07
Radiation indication
kV/mAs lights
Radiation indicationno longer lights up. Acoustic signal no longer sounding, when radiation button is released prematurely
Error message
Fig. 2:
GALILEOS Multipad
Return key
Light localizer

1.6 Exposure safety

Exposure safety Answer questions with yes (9) or no (−)
Date
(please enter month/year)
No repeated occurrence of a cer­tain error message (Fig. 1/2)?
Optical and acoustic signal present when radiation is released (Fig. 1/2)?
Exposure is immediately inter­rupted when radiation button is released prematurely (Fig. 1/2)?
Does rotational movement occur without any atypical running noises?
If no Ö Inform your dental depot.
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
9
1 Yearly inspection by the system owner Sirona Dental Systems GmbH
Fig. 1:
GALILEOS Easypad
T
R
Prog.
kV
mAs
E1 10 07
GALILEOS Multipad
Fig. 1:
Vertical light beam

1.7 Key functionality

Key functionality Answer questions with yes (9) or no (−)
Date
(please enter month/year)
Membrane foil of the Easypad/ Mutipad keys without cracks or holes (Fig. 1)?
All numbers and letters on the Easypad/Multipad show up com­pletely?
Can the Easypad/Mutipad be positioned reliably and does it remain in the desired position?
If no Ö Inform your dental depot.

1.8 Light localizer functionality

Light localizer Answer questions with yes (9) or no (−)
Date (please enter month/year)
Vertical and horizontal light beam visible (Fig. 1)?
If no Ö Inform your dental depot.
10 D 3437.105.01.03.02
61 25 533 D 3437
Sirona Dental Systems GmbH 1 Yearly inspection by the system owner
Fig. 1:

1.9 Swivel arm/bite block/forehead support functionality

Swivel arm/bite block/forehead support Answer questions with yes (9) or no (−)
Date
(please enter month/year)
Can the swivel arm be easily swiveled, is the locking mecha­nism OK, does the swivel arm lock into position (Fig. 1)?
When swiveled out, is return movement not possible?
Does the clamping device for the bit block work correctly?
Does the clamping device for the forehead support work correctly?
If no Ö Inform your dental depot.

1.10 Image quality

Image quality Answer questions with yes (9) or no (−)
Date (please enter month/year)
Typical exposure parameters e.g mAs are not adjusted in order to achieve a constant image quality?
No recurring deterioration of the image sharpness or contrast?
If no Ö Inform your dental depot.
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
11
1 Yearly inspection by the system owner Sirona Dental Systems GmbH

1.11Height adjustment functionality

Height adjustment Answer questions with yes (9) or no (−)
Date
(please enter month/year)
Acoustic signal sounds when height is adjusted?
Height adjustment works without jolting, without slumping and with­out atypical running noises (e.g. clacking, scuffing)?
Does the height setting remain the same even after the system has been switched off for a longer period of time?
If no Ö Inform your dental depot.

1.12Documenting your yearly inspection

The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has informed the competent dental depot in case of any defects.
Year Inspection date: Name: Signature:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
61 25 533 D 3437
12 D 3437.105.01.03.02

Sirona Dental Systems GmbH 2 Yearly calibration by the service engineer (for the USA refer to the Maintenance Instructions)

i
2 Yearly calibration by the service
engineer (for the USA refer to the Maintenance Instructions)
General information for the service engineer
The GALILEOS Service Manual (Order No. 61 25 665) is indispensable for performing the calibration. All of the measures required for the calibration are described there in the Chapter “Calibrating the unit.” Furthermore, the Operating Instructions which have been supplied together with the unit are required.
In addition to the annual check by the system owner or other authorized persons, a yearly calibration must also be performed by the service engineer.
Should it be necessary to replace defective components along with parts subject to wear, this must be agreed previously with the customer or dental practice staff.
If components bearing a serial number are replaced, the new serial number must be recorded in the table in Chapter 3.4‚ “Serial numbers of the system”.
Discontinuation of spare part deliveries: For each system, the spare part deliveries are discontinued after a certain period of time. The system owner will be responsible for safety-relevant failures of systems which continue in operation after that time and can no longer be serviced due to missing spare parts.

2.1 Customer data

Customer
Last name:
First name:
Street:
Post code / city:
Phone:
Space for practice stamp
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
13
2 Yearly calibration by the service engineer (for the USA refer to the Maintenance Instructions) Sirona Dental Systems GmbH

2.2 Calibrating the X-ray unit

Assessing X-ray exposures Answer questions with yes (9) or no (−)
Date
(enter month/year)
Mechanical unit adjustment OK? See Service Manual
Diaphram exposure OK? See Service Manual
Beam field OK? See Service Manual
Dosimetry OK? See Service Manual
Sensor calibration OK? See Service Manual
Iris calibration OK? See Service Manual
Shading calibration OK? See Service Manual
Distorsion calibration OK? See Service Manual
Geometric calibration OK? See Service Manual
Calibration of touchscreen OK? See Service Manual
If no Ö Correct fault.
61 25 533 D 3437
14 D 3437.105.01.03.02
Sirona Dental Systems GmbH 2 Yearly calibration by the service engineer (for the USA refer to the Maintenance Instructions)

2.3 Documenting the yearly calibration

The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has handed over the unit in fully functional condition..
Year Date of
check:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Biometric phantom Serial No.:
Distorsion phantom Serial No.:
Name: Dealer: Signature:
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
15

3 Maintenance by the service engineer Sirona Dental Systems GmbH

i
3 Maintenance by the service engineer
General information for the service engineer
For performing maintenance work, the Service Manual GALILEOS, (Order No. 61 25 665) is indispensable. All measures which are required for maintenance are described there in Chapter “Maintenance”. Furthermore, the Operating Instructions which have been supplied together with the unit are required.
In addition to the yearly inspection by the system owner or persons authorized by him/her, preventive maintenance has to be performed after 4, 7, 10 years and afterwards every two years.
Should it be necessary to replace defective components along with parts subject to wear, this must be agreed previously with the customer or dental practice staff.
If components bearing a serial number are replaced, the new serial number must be recorded in the table in Chapter 3.4‚ “Serial numbers of the system”.
Discontinuation of spare part deliveries: For each system, the spare part deliveries are discontinued after a certain period of time. The system owner will be responsible for safety-relevant failures of systems which continue in operation after that time and can no longer be serviced due to missing spare parts.

3.1 Customer data

Customer
Last name:
First name:
Street:
Post code / city:
Phone:
Space for practice stamp
61 25 533 D 3437
16 D 3437.105.01.03.02
Sirona Dental Systems GmbH 3 Maintenance by the service engineer
Fig. 1:
fелн~дд~нбзе=oЙйзкн=L=t~кк~ену=m~ллйзкн
Customer Dealer
Address, tel.: Address, tel.:
Customer No.:Order No:
Serial numbers Pos. Serial-No.
n o p q r s t u v
n
o
s
r
q
p
Dongle
u
t
11
v
Software Hardware
GALILEOS V _ _ . _ _ _ _ _
GALILEOS Implant
V _ _ . _ _
GALAXIS V _ _ . _ _ SIDEXIS V _ _ . _ _
RCU PC _ _ _
Information on installation:
Dispatch mode License file:
E-MAIL: .............................................................@..................................................... CD (dispatch by mail)
Visual inspection of packaging for possible damage. Remarks:
Shockwatch OK? On the unit On the X-ray detector
Yes
Yes
No
No
Confirmation of completeness of consignment (see installation instructions for scope of supply).
Dongle value explained to customer.
Installation of GALILEOS performed error-free, including RCU server and software modules.
•Patient management software: Manufacturer: Type: Version:
• Implantplaning software: Manufacturer: Type: Version:
•Installed products:
(X-ray/Treatment center)
SIRONA Competitors
Commissioning performed error-free according to the pre-installation instructions and installation instructions as well as country-specific safety controls where required. The unit, including accessories and all documents in accordance with the Operating Instructions, was turned over to the the customer in proper and functioning condition. USA only: The customer received his copy of the Federal Report of Assembly FD 2579. The person signing confirms having the required technical knowledge for performing commisioning and training and fulfilling and complying with the requirements outline above.
Name
Customer: Trained person: Sales rep.: Technician:
Signature
Customer: Trained person: Sales rep.r: Technician:
Date: Date: Date: Date:
d^ifiblp
11
Maintenance overview: To ensure proper functioning, annual maintenanceand calibration must be per­formed by a specialized dental dealer. This must be documented in the "Inspection and Maintenance" report.
Sirona Dental Systems GmbH, Fabrikstraße 31, 64625 Bensheim, Germany Ä.-Nr: 108 924 61 47 941 D3437.077.01.03.02 09.2007
FORTHE CUSTOMER

3.2 Checking the records

Checking the records of the Answer questions with yes (9) or no (−) yearly inspection by the system owner
Maintenance interval (after years)
Date (enter month/year)
Yearly inspection by the system owner has been performed (see Chapter 1)?
Yearly calibration has been per­formed by the service engineer (see Chapter 2 or Maintenance Instructions)?
Records are complete?
All questions answered with yes?
All work which has led to a nega­tive result during the checks have been performed?
4 7 10 12 14
ЙеЦдблЬ
If no Ö Perform yearly inspection prior to maintenance.

3.3 Labels on the system

Comparing labels on the system with Installation Report / Warranty Passport Answer questions with yes (9) or no (−)
Maintenance interval (after years)
Date (enter month/year)
All labels according to Installation Report / Warranty Passport affixed (Fig. 1)?
All labels legible?
4 7 10 12 14
If no Ö Replace missing and/or illegible labels.
61 25 533 D 3437 D 3437.105.01.03.02
17
3 Maintenance by the service engineer Sirona Dental Systems GmbH

3.4 Serial numbers of the system

Please enter new Serial No. here when replacing a component:
X-ray unit Date Signature
Serial No. (at initial start-up) Serial No. Serial No.
X-ray tube assembly Date Signature
Serial No. (at initial start-up) Serial No. Serial No. Serial No. Serial No.
Tube Date Signature
Serial No. (at initial start-up) Serial No. Serial No. Serial No. Serial No.
Remote Date Signature
Serial No. (at initial start-up) Serial No. Serial No. Serial No.
X-ray detector Date Signature
Serial No. (at initial start-up) Serial No. Serial No.
Test phantom Date Signature
Serial No. (at initial start-up) Serial No. Serial No.
61 25 533 D 3437
18 D 3437.105.01.03.02
Sirona Dental Systems GmbH 3 Maintenance by the service engineer

3.5 Functional test

Prior to starting with maintenance work, a functional test must always be performed.
In doing so, it is helpful to read out the error messages, if present. Refer to Chapter 5.2‚ “Selecting Service routines”.
Inform the customer or dental practice staff about any defects you have detected.
Checking the system functions (observe Operating Instructions) Answer questions with yes (9) or no (−)
Maintenance interval (after years)
Date (enter month/year)
Is there no visible abrasion on the HA motor and threaded rod? See Service Manual
No atypical running noises audi­ble during height adjustment?
Jolt-free fine positioning possible during height adjustment?
Does the height setting remain the same even after the system has been switched off for a longer period of time?
Is the limit switch for height adjust­ment functional?
Acoustic signals audible during height adjustment?
Clamping device for forehead sup­port works correctly?
Clamping device for bite block works correctly?
Can the swivel arm be easily swiveled, is the locking mecha­nism OK, does the swivel arm lock into position?
4 7 10 12 14
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
If no Ö Correct fault.
19
3 Maintenance by the service engineer Sirona Dental Systems GmbH

3.6 Checking the tube data

Checking the tube data Answer questions with yes (9) or no (−)
Maintenance interval (after years)
Date (enter month/year)
Actual kV/mAs values OK? Fan working?
See Service Manual. Does the temperature sensor in
the tube assembly indicate a value between ca. 18° and 65°C? See Service Manual.
4 7 10 12 14
If no Ö Correct fault.

3.7 Checking the cables for damage

Checking the cables Answer questions with yes (9) or no (−)
Maintenance interval (after years)
4 7 10 12 14
Date (enter month/year)
Power cable OK? See Service Manual.
Protective ground wire OK?
Control cable OK?
Data transfer cable OK?
If no Ö Correct fault.
61 25 533 D 3437
20 D 3437.105.01.03.02
Sirona Dental Systems GmbH 3 Maintenance by the service engineer

3.8 Safety checks

Move unit to working height. Switch power OFF at the main switch for the building installation or disconnect the power plug. Remove covers from the unit.
Performing safety checks Answer questions with yes (9) or no (−)
Maintenance interval (after years)
Date (enter month/year)
Idling roll free of damage? See Service Manual.
Do grounding straps and cable shields have complete and firm contact?
Cable shielding OK?
Protective ground wire resistance complies with specifications? See Service Manual.
Protective ground wire connec­tions properly attached?
Unit’s leakage current complies with specifications? See Service Manual.

3.9 Final work

4 7 10 12 14
ЙеЦдблЬ
If no Ö Correct fault.
Final work (observe Operating Instructions) Answer questions with yes (9) or no (−)
Maintenance interval (after years)
Date (enter month/year)
System cleaned?
61 25 533 D 3437 D 3437.105.01.03.02
4 7 10 12 14
If no Ö Clean system.
21
3 Maintenance by the service engineer Sirona Dental Systems GmbH

3.10Documenting maintenance work

The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has handed over the unit in fully functional condition.
Date of mainte­nance
Name of engineer Name of dealer Signature of engineer
61 25 533 D 3437
22 D 3437.105.01.03.02
Sirona Dental Systems GmbH 3 Maintenance by the service engineer
Space for remarks
ЙеЦдблЬ
61 25 533 D 3437 D 3437.105.01.03.02
23
3 Maintenance by the service engineer Sirona Dental Systems GmbH
61 25 533 D 3437
24 D 3437105.01.03.02
tЙ=кЙлЙкоЙ=нЬЙ=кбЦЬн=нз=г~вЙ=~еу=~днЙк~нбзел=пЬбЕЬ=г~у=ДЙ=кЙимбкЙЗ=ЗмЙ=нз=нЙЕЬебЕ~д=бгйкзоЙгЙенлK
«=pбкзе~=aЙен~д=pулнЙгл=dгДe=OMMS pйк~ЕЬЙW=ЙеЦдблЕЬ= mкбенЙЗ=бе=dЙкг~еу a=PQPTKNMRKMNKMPKMO===MSKOMMU ûKJkêKW= NMV=VSV fгйкбг¨=Йе=^ддЙг~ЦеЙ
pбкзе~=aЙен~д=pулнЙгл=dгДe=
áå=íÜÉ=rp^W
c~Дкбвлнк~É=PN pбкзе~=aЙен~д=pулнЙгл=ii` aJSQSOR=_ЙелЬЙбг QUPR=páêçå~=aêáîÉI=pìáíÉ=NMM lêÇÉê=kç dÉêã~åó `Ь~кдзннЙI=k`=OUOTP пппKлбкзе~KЗЙ rp^
SN=OR=RPP=a=PQPT