Sirona Galaxis Operating Manual

MQKOMNM
d^i^ufp
kÉï=~ë=çÑW==
bеЦдблЬ
léÉê~íçêDë=j~åì~ä
61 23 488 D 3474
Sirona Dental Systems GmbH
Operator's Manual GALAXIS
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
3
Sirona Dental Systems GmbH
Operator's Manual GALAXIS Contents
bеЦдблЬ
Contents
1 Introduction.............................................................................................................. 6
1.1 Structure of the documents .......................................................................... 6
1.2 General information on the use of GALAXIS................................................ 7
1.3 Limited functionality of model version GAX5................................................ 8
2 Technical description............................................................................................... 9
2.1 RCU (Reconstruction Control Unit) .............................................................. 9
2.2 GALAXIS PC ................................................................................................ 9
2.3 Network ........................................................................................................ 9
3 Starting GALAXIS.................................................................................................... 10
4 Closing GALAXIS .................................................................................................... 11
5 Creating 3D X-ray images ....................................................................................... 12
6 Workspaces............................................................................................................. 13
6.1 Explanation................................................................................................... 13
6.2 Structure....................................................................................................... 13
6.2.1 Toolbar .............................................................................................. 13
6.2.2 Views................................................................................................. 15
6.2.3 Tab .................................................................................................... 15
6.2.4 What is the examination window?..................................................... 16
6.3 "Panoramic view" workspace ....................................................................... 16
6.4 "MPR/Radiology" workspace........................................................................ 17
6.5 "Ceph p.a./a.p." workspace .......................................................................... 18
6.6 "Ceph lateral" workspace ............................................................................. 19
6.7 "Detail" workspace ....................................................................................... 20
7 Views....................................................................................................................... 22
7.1 Panoramic view ............................................................................................ 22
7.1.1 Examination window ......................................................................... 22
7.2 3D view......................................................................................................... 23
7.2.1 3D view in the workspaces................................................................ 24
7.2.2 Interaction ......................................................................................... 24
7.3 Longitudinal view.......................................................................................... 24
61 23 488 D 3474
Sirona Dental Systems GmbH
Contents Operator's Manual GALAXIS
7.4 Transversal view ........................................................................................... 25
7.5 Axial plane..................................................................................................... 27
7.6 Coronal plane................................................................................................ 28
7.7 Sagittal plane ................................................................................................ 28
8 Functions.................................................................................................................. 29
8.1 2D snapshot .................................................................................................. 29
8.1.1 Complete examination ....................................................................... 29
8.1.2 Individual view.................................................................................... 29
8.2 Printing .......................................................................................................... 30
8.2.1 Complete examination ....................................................................... 30
8.2.2 Individual view.................................................................................... 30
8.3 Navigating through slices .............................................................................. 30
8.4 Standard mouse pointer................................................................................ 31
8.5 Mouse adjuster: Adjust brightness and contrast with mouse ........................ 31
8.6 Moving the view ............................................................................................ 32
8.7 Positioning the crosshair with a single click .................................................. 32
8.8 Resetting the displays ................................................................................... 32
8.9 Measurements .............................................................................................. 32
8.9.1 Measuring lengths.............................................................................. 33
8.9.2 Measure angles.................................................................................. 33
8.10 Reconstructing a panoramic view ................................................................. 34
8.11 Zoom function ............................................................................................... 37
8.12 Transfer function ........................................................................................... 37
8.13 Findings-based work ..................................................................................... 38
8.13.1 Explanation ........................................................................................ 38
8.13.2 Creating a new finding ....................................................................... 38
8.13.3 Selecting a finding.............................................................................. 39
8.13.4 Changing the position of a finding...................................................... 39
8.13.5 Editing a finding.................................................................................. 39
8.13.6 Deleting a finding ............................................................................... 39
8.13.7 Creating a report ................................................................................ 39
8.14 Reconstructing a "Ceph p.a./a.p." view......................................................... 40
8.15 Reconstructing a lateral Ceph view............................................................... 41
8.16 Creating a detail volume ............................................................................... 43
8.17 Showing and hiding orientation lines and findings ........................................ 44
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
5
Sirona Dental Systems GmbH
Operator's Manual GALAXIS Contents
bеЦдблЬ
8.18 Settings ........................................................................................................ 44
9 Create GALILEOS Viewer CD................................................................................. 46
61 23 488 D 3474
Introduction Sirona Dental Systems GmbH
Structure of the documents Operator's Manual GALAXIS
1 Introduction
Contents of this manual The present manual provides instructions and information on how to use the
GALAXIS software:
● Comprehensive instructions on use of the software.
● An introduction to the acquisition technique.
● Possibilities for visualizing and analyzing digital 3D X-ray images.
GALILEOS Viewer The manual also describes the functionality of the GALILEOS Viewer. Not all
GALAXIS functions are available in the GALILEOS Viewer.
1.1 Structure of the documents
empty
Structure of the documents
The symbols and character formats used in the present manual have the fol­lowing meaning:
WARNING
Identifies warnings where a medium risk of injury to persons exists if they are not observed.
CAUTION
Identifies safety information where the following hazards exist if th ey are no t observed: Slight risk of injury to persons, risk of property damage or damage to the product.
NOTICE
Assistance
Identifies additional information, hints and tips.
✔ Prerequisite ➢ Action
or
➢ 1., 2., …
Result
Requests you to do something.
See chapter "Structure of the documents [ ➙ 6]“
Identifies a reference to another text passa­ge.
●List Identifies a list.
“Text between quotation marks“
Identifies commands, menu items or quota­tions.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
7
Sirona Dental Systems GmbH Introduction
Operator's Manual GALAXIS General information on the use of GALAXIS
bеЦдблЬ
1.2 General information on the use of GALAXIS
Intended use GALAXIS is a software for dental practices and clinics that enables the pre-
paration of three-dimensional volume reconstructions of the maxillofacial re­gion using the GALILEOS dental volume tomograph.
GALAXIS is a component of the GALILEOS system.
The volume reconstructions can be used both for planning and for diagnostic purposes with the help of GALAXIS.
● For example for oral and maxillofacial surgery, plastic surgery and ENT medicine.
The GALILEOS system The GALILEOS system comprises:
● X-ray unit
The X-ray unit produces two-dimensional images of the head region
via a cone beam with a rotation sequence.
● RCU (Reconstruction Control Unit)
A reconstruction software converts two-dimensional images into a
three-dimensional volume reconstruction.
The RCU also manages the software licenses.
● GALAXIS workstation(s)
Enable(s) the creation of a new three-dimensional volume recon-
struction.
(Call to the X-ray unit to create a new image and to the RCU to con­vert this image into a three-dimensional volume reconstruction).
The three-dimensional volume reconstructions are displayed and
processed via the GALAXIS software.
Management of the three-dimensional volume reconstructions
(saving them to and retrieving them from the SIDEXIS database).
● SIDEXIS database
Storage of three-dimensional volume reconstructions.
IMPORTANT INFORMATION
Add-on modules The functionality of GALAXIS can be extended by installing add-on modules
(e.g. "GALILEOS Implant").
Such add-on modules are supplied with a separate user manual and there­fore are not further explained in this manual.
CAUTION
Compliance with national regulations
Make sure that the use of GALAXIS is compliant with all relevant national laws and regulations.
CAUTION
Restrictions imposed by U.S. American Federal Law
Based on US federal law, the sale of this unit to or by order of physicians, dentists or other trained medical personnel is subject to certain limitations.
61 23 488 D 3474
Introduction Sirona Dental Systems GmbH
Limited functionality of model version GAX5 Operator's Manual GALAXIS
1.3 Limited functionality of model version GAX5
Explanation GALAXIS has a limited functionality in connection with model version GAX5.
Version GAX5 is a component of the
"GALILEOS Compact"
package.
Identification
All functions not available for model version GAX5 are marked accordingly with a special logo in this document.
NOTICE
In case of problems with add-on modules, please contact the corresponding software provider.
5
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
9
Sirona Dental Systems GmbH Technical description
Operator's Manual GALAXIS RCU (Reconstruction Control Unit)
bеЦдблЬ
2 Technical description
2.1
RCU (Reconstruction Control Unit)
Minimum requirements for the RCU (included in the scope of supply)
2.2 GALAXIS PC
Minimum requirements for the PC used for the display GALAXIS
2.3 Network
Minimum requirements
Processor: at least CPU 2,0 GHz dual-core
RAM: at least 2 GB RAM
Hard disk: > 200 MB
Drive: DVD-ROM (dual-layer)
Operating system: Windows XP Professional Service Pack 2 or ne-
wer
Windows Vista Business
Windows Vista Ultimate
Windows 7 (32 / 64bit)
Graphics system: > 128 MB, minimum resolution 1024x768 pixels
Screen: Suitable for diagnostic purposes
Network connection: Network RJ45, 100 MBit/s (1 GBit/s recommen-
ded)
Processor: at least CPU 1,6 GHz dual-core
RAM: at least 2 GB RAM
Hard disk: at least 5 GB
Drives: DVD-ROM (dual-layer)
Operating systems: Windows XP Professional Service Pack 3 or ne-
wer
Windows Vista Business (Service Pack 1 or high­er)
Windows Vista Home Premium (Service Pack 1 or higher)
Windows 7 (32 / 64bit)
Graphics system: externally > 128 MB,
Minimum resolution 1024x768 pixels
16.7 mil. colors ("TrueColor")
Screen: Suitable for diagnostic purposes
Network connection: Network RJ45, 100 MBit/s (1 GBit/s recommen-
ded)
Infrastructure: Network RJ45, 100 MBit/s (1 GBit/s recommen-
ded)
61 23 488 D 3474
Starting GALAXIS Sirona Dental Systems GmbH
Operator's Manual GALAXIS
3 Starting GALAXIS
Explanation GALAXIS is started via SIDEXIS XG.
Prerequisites In order to start GALAXIS, a 3D X-ray image must be opened in the SIDEXIS
database first.
Simplified invocation 1. Start SIDEXIS XG.
ª The GALAXIS button is displayed as not active (faded) on the toolbar
of the SIDEXIS XG desktop.
2. Select a patient with an existing 3D X-ray image.
3. Select a 3D X-ray image.
ª The selected 3D X-ray image is displayed in SIDEXIS. The GALAXIS
button is displayed as active.
4. Press the GALAXIS button. ª GALAXIS and the selected 3D X-ray image are loaded.
Alternative start options The following alternative options exist for starting GALAXIS via the the SIDE-
XIS XG desktop instead of by pressing the GALAXIS button:
● Via context menu: Select the
"3D"
menu option.
●Via the
"3D"
button in the upper right-hand corner of the 3D image.
NOTICE
If no 3D X-ray images are available, you must create a 3D X-ray image first (see the chapter entitled "Creating a 3D image").
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
11
Sirona Dental Systems GmbH Closing GALAXIS
Operator's Manual GALAXIS
bеЦдблЬ
4 Closing GALAXIS
Close ➢ Go to the
"Workspace"
menu bar and select the
"Exit"
menu option.
ª GALAXIS is then closed and the SIDEXIS XG desktop appears.
61 23 488 D 3474
Creating 3D X-ray images Sirona Dental Systems GmbH
Operator's Manual GALAXIS
5 Creating 3D X-ray images
Explanation The 3D X-ray images are created via the SIDEXIS XG exposure readiness
function and managed in the SIDEXIS database parallel to other X-ray ima­ges (IO "Intraoral", XP "Panoramic", XC "Ceph", etc.).
At the same time, the 3D X-ray images are stored on the
"3D"
tab card.
Required documentation The operation of the GALILEOS X-ray device is explained in the
"GALILEOS
Operating Instructions"
.
The operation of SIDEXIS XG software is explained in the
"SIDEXIS XG Ope
-
rator's Manual"
.
Creation 1. Start SIDEXIS XG (see
"SIDEXIS XG Operator's Manual"
).
2. Select a patient.
3. Press the 3D Scan button.
4. Press the release button.
After a few minutes, the icon for the 3D X-ray image is displayed on the SI­DEXIS XG desktop.
The 3D X-ray image is now managed in the SIDEXIS database in the same way as any other X-ray image.
CAUTION
X-rays
Make sure that all locally applicable radiation protection regulations are ob­served! Also make sure that the application information provided in the
"GA
-
LILEOS Operating Instructions"
is observed (especially regarding patient
positioning and the start of exposure)!
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
13
Sirona Dental Systems GmbH Workspaces
Operator's Manual GALAXIS Explanation
bеЦдблЬ
6 Workspaces
Explanation Optimized workspaces are provided for different tasks.
A workspace shows several different views of various slices through the vo­lume.
These views are interconnected via the examination window.
6.1 Explanation
Optimized workspaces are provided for different tasks.
A workspace shows several different views of various slices through the vo­lume.
These views are interconnected via the examination window.
6.2 Structure
6.2.1 Toolbar
Explanation The toolbar consists of buttons which contain graphic symbols (icons).
A function can be initiated by clicking one of these icons.
A Title bar (display of the patient's name and date of birth and the
date of image acquisition)
B Menu bar
C Toolbar [ ➙ 13]
D Views [ ➙ 15] (Example:
"Panorama with 3D"
)
E Tab [ ➙ 15]
F Options bar (see the section "Showing and hiding orientation li-
nes and findings" [ ➙ 44])
61 23 488 D 3474
Workspaces Sirona Dental Systems GmbH
Structure Operator's Manual GALAXIS
Standard tools
Diagnostic tools
A Standard mouse pointer [ ➙ 31]
B Enlarging [ ➙ 37]
C Reducing [ ➙ 37]
D Hand (move view) [ ➙ 32]
E Navigating into the volume [ ➙ 30]
F Navigating out of volume [ ➙ 30]
G Open panoramic view [ ➙ 22]
H Length measurement [ ➙ 33]
I Print workspace [ ➙ 30]
J Create GALILEOS Viewer CD [ ➙ 46]
K Adjusting the brightness
L Create detail volume [ ➙ 43]
M Measure angles [ ➙ 33]
N 2D snapshot [ ➙ 29]
O Adjust contrast
5
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
15
Sirona Dental Systems GmbH Workspaces
Operator's Manual GALAXIS Structure
bеЦдблЬ
Evaluation tools
6.2.2 Views
Explanation Several different views are displayed in a single workspace.
● The currently active view, which can be recognized by its orange colored frame and letters, can be processed directly.
● The slice displayed in the active view is marked by a crosshair in all other views.
Selecting a view ➢ Click the desired view.
ª The color of the view's title bar then changes to indicate that the view is
active.
6.2.3 Tab
Explanation You can select different workspaces by clicking the corresponding tabs.
A Create new finding
(see section "Creating a new finding" [ ➙ 38])
B List of findings (list of created findings)
(see section "Selecting a finding" [ ➙ 39])
C Copy current view for finding
(see section "Changing the position of a finding" [➙ 39])
D Properties of finding
(see section "Editing a finding" [ ➙ 39])
E Delete finding
(see section "Deleting a finding" [ ➙ 39])
F Create report
(see section "Creating a report" [ ➙ 39])
NOTE: This function is only available with a current version of the REPORTER software.
61 23 488 D 3474
Workspaces Sirona Dental Systems GmbH
"Panoramic view" workspace Operator's Manual GALAXIS
6.2.4 What is the examination window?
Explanation The display in views is based on a single, three-dimensional vector point in
the reconstructed volume.
● Various different slices of the volume are displayed starting from this vec­tor point, depending on the view concerned.
● At the same time, the vector point is located in the crosshair of the re­spective slice axes.
For this reason, the crosshair is located at the same position within
the volume in all of the views.
Depending on the type of view concerned, the vector point is located
either in the center of the view or in the examination window of the view.
If you navigate through several slices in a view or in the examination
window, the slices in the other views will automatically be adjusted accordingly.
empty
Example:
If you, for example, move the examination window of the panoramic view, the transverse views, the longitudinal view and the curved line in the axial plane will all be altered accordingly.
6.3 "Panoramic view" workspace
Explanation After you start GALAXIS, the selected 3D X-ray image will be displayed in the
classical panoramic view.
In addition to the panoramic view workspace, there is an additional work­space with both a panoramic and a 3D view.
Brief overview In each view, the fine lines of the crosshair indicate the positions in the other
views to which this corresponds.
empty
Example:
● The horizontal line in the examination window indicates the position in the axial plane.
In this way, all views are allocated to each other and the viewer can
easily find his bearings.
● The axial slice also contains a curved line which indicates the position in the examination window.
NOTICE
Only one view can be active at any given time.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
17
Sirona Dental Systems GmbH Workspaces
Operator's Manual GALAXIS "MPR/Radiology" workspace
bеЦдблЬ
Views
6.4 "MPR/Radiology" workspace
Explanation The Radiology workspace contains the three classical (axial, coronal and la-
teral) planes of the 3D volume as well as the 3D view.
Views
A Panoramic view [ ➙ 22]
B 3D view [ ➙ 23] (Only in the "Panoramic view with 3D" work-
space)
C Longitudinal view [ ➙ 24]
D Transversal view [ ➙ 25]
E Axial plane [ ➙ 27]
61 23 488 D 3474
Workspaces Sirona Dental Systems GmbH
"Ceph p.a./a.p." workspace Operator's Manual GALAXIS
empty
Plane D
The view D (Default: Sagittal) can be freely adjusted. The name of the "sagit­tal" plane is changed to reflect the current position.
6.5 "Ceph p.a./a.p." workspace
Validity
This workspace is not available for GALAXIS in connection with model versi­on GAX5.
Explanation The Ceph view is calculated from the entire reconstructed volume.
Individual thin slices can also be viewed inside the Ceph view with the exami­nation window.
See the section "Reconstructing a p.a./a.p. Ceph view [ ➙ 40]".
A Axial plane
B 3D view [ ➙ 23]
C Coronal plane [ ➙ 28]
D Sagittal plane [ ➙ 28]
E Tilt
F Reset rotation and tilt
G Rotation
5
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
19
Sirona Dental Systems GmbH Workspaces
Operator's Manual GALAXIS "Ceph lateral" workspace
bеЦдблЬ
Views
6.6 "Ceph lateral" workspace
Validity
This workspace is not available for GALAXIS in connection with model versi­on GAX5.
Explanation The Ceph view is calculated from the entire reconstructed volume.
Individual thin slices can also be viewed inside the Ceph view with the exami­nation window.
See the section "Reconstructing a lateral Ceph view [ ➙ 41]".
A Ceph a.p.
B 3D view [ ➙ 23]
C Axial plane
D Sagittal plane [ ➙ 28]
5
61 23 488 D 3474
Workspaces Sirona Dental Systems GmbH
"Detail" workspace Operator's Manual GALAXIS
Views
6.7 "Detail" workspace
Scope
This workspace is not available for GALAXIS in connection with model versi­on GAX5.
Explanation The "Detail" workspace shows an enlarged view of the area inside the exami-
nation window.
The voxels are reconstructed and redisplayed with an edge length of 150 µm.
See the section "Creating a detail volume [ ➙ 43]".
A Ceph lateral
B 3D view [ ➙ 23]
C Axial plane
D Coronal plane
5
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
21
Sirona Dental Systems GmbH Workspaces
Operator's Manual GALAXIS "Detail" workspace
bеЦдблЬ
Views
A Axial plane
B 3D view [ ➙ 23]
C Coronal plane
D Sagittal plane [ ➙ 28]
61 23 488 D 3474
Views Sirona Dental Systems GmbH
Panoramic view Operator's Manual GALAXIS
7 Views
7.1
Panoramic view
Explanation The panoramic view is reconstructed based on a U-shaped region correspon-
ding to the mandibular arch.
● The examination window [ ➙ 22] is located inside the panoramic view.
● The U-shaped region is displayed in the axial plane of the corresponding workspace.
● The displayed panoramic view can be reconstructed by altering the U­shaped region (see chapter: Reconstructing a panoramic view [ ➙ 34].)
7.1.1 Examination window
Explanation The examination window is located inside the panoramic view.
The examination window displays a thin, vertical slice of the volume along the mandibular arch.
Configuration
Position The position of the examination window is shown by the curved line in the axi-
al slice of the corresponding workspace, as well as by the vertical line in the transversal views and in the longitudinal view.
NOTICE
The subjective image impression of the reconstructed panoramic view is not necessarily comparable to the conventional blurring technique (e.g. OR­THOPHOS) due to the dental volume tomography technique used.
The detail accuracy has been verified by numerous scientific publications.
A Title bar.
B Position of the transversal view (TSA) [ ➙ 25].
C Position of the axial plane [ ➙ 27].
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
23
Sirona Dental Systems GmbH Views
Operator's Manual GALAXIS 3D view
bеЦдблЬ
Move The examination window can be moved vertically and parallel to the mandi-
bular arch.
1. Place the mouse pointer above the title bar of the examination window.
2. Press and hold down the left mouse button.
3. Move the mouse as desired.
Go to position A double-click in any position in one of the views (apart from 3D) centers the
examination window to this position.
This method is very useful for navigating quickly between different points.
Navigating through slices You can navigate freely through the slices of the displayed region in the ex-
amination window In this way, the examination window provides access to the third dimension in the panoramic view.
1. Place the mouse pointer inside the examination window.
2. Press and hold down the left mouse button.
3. Move the mouse up and down as desired.
● Upward movements of the mouse display the slice positions located further lingual.
● Downward movements of the mouse display the slices located further buccal.
7.2 3D view
Explanation In the 3D view, the volume is displayed as a 3D object.
The 3D view simplifies orientation within the displayed volume.
There is an optimized display for each workspace with a 3D view.
Configuration The positions of the different slices are displayed as wireframes.
● The currently active slice is highlighted in orange.
NOTICE
Navigating through slices
In all views (apart from the 3D view), you can use the mouse to navigate fre­ely through the slices.
This is described in more detail in the following sections.
61 23 488 D 3474
Views Sirona Dental Systems GmbH
Longitudinal view Operator's Manual GALAXIS
7.2.1 3D view in the workspaces
Panoramic with 3D
MPR/Radiology
7.2.2 Interaction
Rotate 3D view 1. Place the mouse pointer inside the 3D view.
2. Press and hold down the left mouse button.
3. Move the mouse. ª You can rotate the 3D view depending on the movement of the mouse.
Set zoom level 1. Place the mouse pointer inside the 3D view.
2. Turn the scroll wheel on the mouse. ª You can change the zoom level of the 3D view depending on the move-
ment of the scroll wheel.
Move 3D view 1. Place the mouse pointer inside the 3D view.
2. Press and hold down the right mouse button.
3. Move the mouse. ª You can move the 3D view depending on the movement of the mouse.
7.3 Longitudinal view
Explanation The longitudinal view is a straight slice oriented longitudinally to the examina-
tion window.
This slice can be tilted about the axis of the axial plane.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
25
Sirona Dental Systems GmbH Views
Operator's Manual GALAXIS Transversal view
bеЦдблЬ
Configuration
Tilting ➢ Move the slider with the mouse pointer.
ª This tilts the longitudinal view.
● A tilted longitudinal view is displayed in the transversal views [ ➙ 25] tog­ether with an oblique line.
Tilt reset ➢ Press the reset button (D).
ª The longitudinal view returns to the untilted state.
Navigating through vertical slices You can freely navigate through the vertical slices of the displayed region in
the longitudinal view.
1. Place the mouse pointer inside the examination window.
2. Press and hold down the left mouse button.
3. Move the mouse up and down as desired.
● Upward movements of the mouse display the slice positions located further lingual.
● Downward movements of the mouse display the slices located further buccal.
7.4 Transversal view
Explanation The transversal views are slices that are oriented perpendicular to the man-
dibular arch.
A Longitudinal view
B Axial plane [ ➙ 27]
C Slider
"tilt"
D Reset button
"Reset tilt"
61 23 488 D 3474
Views Sirona Dental Systems GmbH
Transversal view Operator's Manual GALAXIS
Configuration GALAXIS shows a transversal view underneath the PAN view.
● The middle view corresponds to the transversal view of the vertical line running through the center of the examination window.
● A transversal view displaced by 5 millimeters from the center line is dis­played in the left and right views.
Navigating through slices You can navigate through the transversal views of the displayed region.
CAUTION
Risk of confusing slices
In GALAXIS the transversal views are always displayed in the direction of view toward the left temporormandibular joint.
A Oral/lingual side
B Labial/buccal side
C Axial position of the examination window [ ➙ 22] (axial plane [ ➙
27])
D Vertical position of the examination window [ ➙ 22]
E Slider
"tilt"
F Reset button
"Reset tilt"
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
27
Sirona Dental Systems GmbH Views
Operator's Manual GALAXIS Axial plane
bеЦдблЬ
The position of the examination window [ ➙ 22] then changes along the man­dibular arch.
1. Place the mouse pointer inside a transversal view.
2. Press and hold down the left mouse button.
3. Move the mouse up and down as desired.
● Upward movements of the mouse display the slices toward the right side of the head.
● Downward movements of the mouse display the slices toward the left side of the head.
● This method is used in all 2D slice views in GALAXIS.
7.5 Axial plane
Explanation The axial plane displays the horizontally positioned slices of the volume at the
same height as the horizontal lines shown in the other views.
It is part of the Radiology, Panoramic view with 3D and Panoramic view work­spaces.
Setup in the panoramic view The position of the mandibular arch and of the examination window is shown
in the displayed slice.
Navigating through axial slices You can freely navigate through the axial slices of the displayed region in the
axial plane view.
1. Place the mouse pointer inside the axial plane view.
2. Press and hold down the left mouse button.
3. Move the mouse up and down as desired.
● Upward movements of the mouse display the slice positions located further inferior.
● Downward movements of the mouse display the slices located further su­perior.
A Mandibular arch
B Position of the transversal view [ ➙ 25]
C Slice displayed in the examination window [ ➙ 22] (parallel to
mandibular arch)
D Longitudinal view [ ➙ 24]
61 23 488 D 3474
Views Sirona Dental Systems GmbH
Coronal plane Operator's Manual GALAXIS
Orientation It is possible to adjust the orientation of the axial view (see the section "Set-
tings" [ ➙ 44]).
7.6 Coronal plane
Explanation The coronal plane displays the frontal slices of the volume and is a compo-
nent of the Radiology workspace.
● Upward movements of the mouse display the slice positions located further posterior.
● Downward movements of the mouse display the slices located further su­perior.
7.7 Sagittal plane
Explanation The sagittal plane displays the frontal slices of the volume and is a component
of the Radiology workspace.
Operation The sagittal plane can be rotated in any direction with the slider.
A button for resetting to the untilted state is located at the bottom left end of the slider.
● Upward movements of the mouse display the slice positions located further to the left.
● Downward movements of the mouse display the slice positions located further to the right.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
29
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS 2D snapshot
bеЦдблЬ
8 Functions
8.1
2D snapshot
Explanation With the
"2D Screenshot"
function, GALAXIS offers the possibility of transfer­ring the current view or the complete current examination to SIDEXIS XG in the form of two-dimensional views.
The views thus transferred can either be added to an existing examination in SIDEXIS XG or to a newly created examination.
Use The
"2D Screenshot"
function can be used to document work in the SIDEXIS
XG database.
8.1.1 Complete examination
Creating a new exam
➢ Press the
"2D Screenshot"
button.
or
➢ Go to the
"Workspace"
menu bar and select the
"2D Screenshot"
option
> Select
"Save in new exam"
.
Adding to an existing exam ➢ Go to the
"Workspace"
menu bar and select the
"2D Screenshot"
option
> Select
"Save in current exam"
.
Hiding the examination window and auxiliary lines
empty
Explanation
With the option
"Use plain , hi-res grey value images"
you can determine wh­ether the existing examination windows and auxiliary lines are to be hidden when executing the
"2D Screenshot"
function.
Activation is indicated by a check mark in front of the
"Use plain , hi-res grey
value images"
option.
empty
Activation and deactivation
➢ Select the
"Use plain , hi-res grey value images"
option via the
"Work
-
space"
menu bar.
8.1.2 Individual view
Creating a new exam ➢ Go to the
"Image"
menu bar and select the menu option
"2D Screens
-
hot">"Save in new exam"
.
Adding to an existing exam ➢ Go to the
"Image"
menu bar and select the menu option
"2D Screens
-
hot">"Save in current exam"
.
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Printing Operator's Manual GALAXIS
Hiding the examination window and auxiliary lines
empty
Explanation
You can use the option
"Use plain , hi-res grey value images"
to determine whether existing examination windows and orientation lines are shown or hid­den when you execute the
"2D Screenshot"
function.
Activation is indicated by a check mark in front of the
"Use plain , hi-res grey
value images"
menu option.
empty
Activation and deactivation
➢ Use the
"Image"
menu bar to select the
"Use plain , hi-res grey value ima
-
ges"
menu item.
8.2 Printing
Explanation With the
"Print…"
function, GALAXIS offers the possibility of printing either the
current view or the complete current examination.
8.2.1 Complete examination
Printout
➢ Press the
"Print…"
button.
or
➢ Go to the
"Workspace"
menu bar and select the
"Print…"
menu option.
Preview ➢ Go to the
"Workspace"
menu bar and select the
"Print Preview…"
menu
option.
8.2.2 Individual view
Printout ➢ Go to the
"Image"
menu bar and select the
"Print…"
menu option.
Preview ➢ Go to the
"Image"
menu bar and select the
"Print Preview…"
menu opti-
on.
8.3 Navigating through slices
Explanation In addition to the possibility of navigating through slices directly with the left
mouse button, this can also be achieved using the
"Navigate into the volume
by one slice."
and
"Navigate out of the volume by one slice."
buttons.
Navigating into the volume
➢ Press the
"Navigate into the volume by one slice."
button.
ª The current view then moves one slice into the volume each time you
press the button.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
31
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Standard mouse pointer
bеЦдблЬ
Navigating out of volume
➢ Press the
"Navigate out of the volume by one slice."
button.
ª The current view then moves one slice out of the volume each time you
press the button.
8.4 Standard mouse pointer
Explanation The default mode for the mouse pointer is referred to as the
"Default Cursor"
.
In this mode, the mouse pointer is displayed as an arrow on the GALAXIS desktop.
The possibilities for changing over from another mode to
"Default Cursor"
are
described below.
Change-over Change-over to the
"Default Cursor"
mode can be performed in the following
ways:
➢ Click the
"Default Cursor"
button.
or
➢ Click on the
"Default Cursor"
menu option on the
"Analysis"
menu bar.
or
➢ Press the space bar on the PC keyboard.
8.5 Mouse adjuster: Adjust brightness and contrast
with mouse
Explanation The so-called mouse adjuster offers a simple and elegant way of adjusting
brightness and contrast. It enables you to adjust brightness and contrast si­multaneously.
Change-over
Change-over to the
"Mouse adjuster"
mode can be performed in the following
ways:
➢ Click on the
"Mouse adjuster"
menu option on the
"Analysis"
menu bar.
or
➢ Hotkey: Press the key combination
"Ctrl"+"Shift"+"M"
.
Display of the function
As long as this function is active, the mouse pointer in the exam workspace has another shape.
Operation 1. In the exam workspace, press and hold down the left mouse button.
2. Move the mouse as desired.
ª Move it horizontally to change the contrast.
ª Move it vertically to change the brightness.
➢ Release the left mouse button when the display settings correspond to
your wishes.
You can repeat the adjustment as often as you like from step 1 onwards as long as the mouse adjuster is active.
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Moving the view Operator's Manual GALAXIS
8.6 Moving the view
Explanation In some cases, the complete slice plane is longer displayed after the view is
enlarged.
Using the
"Hand"
mode, you can freely move the view as required.
Any region of the slice plane can thus be displayed.
Change-over Change-over to the
"Hand"
mode can be performed in the following ways:
➢ Click the
"Hand"
button.
or
➢ Select the
"Hand"
menu option on the
"Analysis"
menu bar.
8.7 Positioning the crosshair with a single click
Explanation In certain situations, it is practical to position the crosshair at a specific loca-
tion with a single click.
Operation
1. Select the
"Crosshair"
menu option on the
"Analysis"
menu bar.
2. Click the desired position in a view. ª All slices will then be updated to display the selected position according
to the slice orientation.
8.8 Resetting the displays
Explanation If you click the
"Back to default"
function, all views in all of the displays are
reset to their default values.
empty
Affected default values
● Navigating through slices
●Zoom
●Move
● Position of examination window
● Position of 3D view
● Contrast and brightness
Reset ➢ Use the
"Analysis"
menu bar to select the
"Back to default"
menu option.
or
➢ Enter [Ctrl] + [r] via the keyboard.
8.9 Measurements
Explanation GALAXIS offers the following measurement functions:
● Measure lengths [ ➙ 33]
● Measure angles [ ➙ 33]
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
33
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Measurements
bеЦдблЬ
8.9.1 Measuring lengths
Explanation This function is used to indicate the distance between two selected points of
a view in millimeters.
Measuring accuracy: ± 0.15mm
Validity This measuring function can be used for the following views:
● Longitudinal view
● transversal view (TSA)
● Axial plane
● Coronal plane
● Sagittal plane
Command
➢ Click the
"Measure Distance"
button.
or
➢ Click on the
"Measure Distance"
menu option on the
"Analysis"/"Measu
-
re"
menu bar.
or
➢ Press the
"[D]"
key on the PC keyboard.
ª The shape of the mouse pointer then changes to a small cross in the ap-
plicable views.
Measurement 1. Place the mouse pointer at the starting point of the measurement.
2. Confirm the starting point by clicking the left mouse button.
3. Drag the mouse pointer to the desired end point.
4. Confirm the measured length by clicking the left mouse button.
Deleting 1. Mark the measurement with the mouse.
2. Press the Delete key on the PC keyboard. ª The measurement is deleted.
8.9.2 Measure angles
Explanation This function is used to indicate the angular relationship between three selec-
ted points in a view.
The angle is indicated in degrees.
Validity This measuring function can be used for the following views:
● Longitudinal view
● transversal view (TSA)
● Axial plane
● Coronal plane
● Sagittal plane
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Reconstructing a panoramic view Operator's Manual GALAXIS
Command
➢ Click the
"Measure Angle"
button.
or
➢ Click on the
"Measure Angle"
menu option on the
"Analysis"/"Measure"
menu bar.
or
➢ Press the
"[A]"
key on the PC keyboard.
ª The shape of the mouse pointer then changes to a small cross in the ap-
plicable views.
Measurement 1. Move the mouse pointer to the spot where you want to place the apex of
the angle.
2. Confirm this position by clicking the left mouse button.
3. Draw the first leg of the angle to be measured using the mouse pointer.
4. Confirm this position by clicking the left mouse button.
5. Draw the second leg of the angle by dragging the mouse pointer.
6. Confirm the angle by clicking the left mouse button.
Deleting 1. Mark the measurement with the mouse.
2. Press the Delete key on the PC keyboard. ª The measurement is deleted.
8.10 Reconstructing a panoramic view
Explanation If the quality of the automatically generated views in the Panoramic view
workspace does not meet your diagnostic requirements, you can reconstruct the Panoramic view workspace.
This can be done via the
"New Panorama"
dialog box.
Command
➢ Press the
"Panorama…"
button.
or
➢ Use the
"Analysis"
menu bar to select the
"Panorama…"
menu option.
ª The
"New Panorama"
dialog box appears.
Configuration The central element in the
"New Panorama"
dialog box is the display of an
axial plane of the current volume.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
35
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Reconstructing a panoramic view
bеЦдблЬ
● A yellow, U-shaped region (A) which is used to reconstruct the panoramic view is displayed in this axial plane.
● Located inside of this region (A) is a blue line (B) which defines the man­dibular arch.
Goal of optimization The form, position and size of the U-shaped region (A) are to be optimized in
such a way that a higher quality panoramic view is obtained.
Preview A preview of the panoramic view with the current settings is displayed in the
lower part of the
"New Panorama"
dialog box.
The position of the axial slice is indicated in this preview by a yellow line.
Optimization points ● The region (A) should include all teeth and the jaw.
● The region (A) should be positioned so that the roots of the teeth are dis­played as centrally as possible. Ideally, the blue line of the mandibular arch (B) should run through the center of the roots of the teeth.
Tools
empty
Moving the U-shaped region
The U-shaped region can be selected and moved.
empty
Changing the shape of the dentition
The shape of the dentition (anterior teeth) can be changed via the radio but­tons in the
"Jaw Shape"
dialog box.
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Reconstructing a panoramic view Operator's Manual GALAXIS
empty
Changing the jaw size
The jaw size can be changed via the radio buttons in the
"Size"
dialog box.
empty
Rotating the U-shaped region
The U-shaped region can be rotated by altering the value entered in the
"Turn
left/right"
input field or by actuating the rotation buttons located next to it.
● This can be helpful if the patient was not optimally positioned during the acquisition of the X-ray image.
empty
Changing the thickness of the U-shaped region
The thickness of the U-shaped region can be defined by changing the value entered in the
"Thickness"
input field.
empty
Resetting the changes made
All of the changes made since the last time the dialog box was opened can be reset by clicking the
"Back to default"
button.
Aids The
"New Panorama"
dialog box features tools which simplify editing of the
U-shaped region.
empty
Detailed preview
A detailed preview of the current settings can be displayed by clicking the
"High quality preview"
button.
● This may take several seconds.
● You can thus display the exact results of your settings before closing the dialog box.
empty
Simultaneous display of all axial slices
The "Overlay" function simplifies alignment of the U-shaped region whenever one setting is to be used for both the upper and the lower jaw.
This enables you to obtain an overview of the position of the dentition and of the individual teeth.
➢ Press the
"Superimpose"
button.
ª The sum of all axial slices will then be displayed.
NOTICE
In most images, the maximum jaw size returns the best results.
NOTICE
A thickness of greater than 200% returns the best results in most images.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
37
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Zoom function
bеЦдблЬ
empty
Resetting the
"Superimpose"
function
➢ Press the
"Single"
button.
ª One axial slice will then be displayed.
empty
Navigating through axial slices
To navigate through the axial slices, proceed as follows:
➢ Press the
"Navigate into the volume by one slice."
button.
ª The current view then moves in one slice.
1. Press the
"Navigate out of the volume by one slice."
button.
ª The current view then moves one slice out of the volume each time
you press the button.
2. Confirm your setting with the
"OK"
button.
8.11 Zoom function
Explanation Using the
"Zoom in"
and
"Zoom out"
functions, you can increase or decrease
the size of any view.
Enlarging ➢ Roll the mouse scroll wheel forward (up).
or
➢ Click the
"Zoom in"
button.
or
➢ Click on the
"Zoom in"
menu option on the
"View"/"Zoom"
menu bar.
Reducing ➢ Roll the mouse scroll wheel back (down).
or
➢ Click the
"Zoom out"
button.
or
➢ Click on the
"Zoom out"
menu option on the
"View"/"Zoom"
menu bar.
8.12 Transfer function
Explanation The display of the medium shades of gray can be changed via a gamma cor-
rection in the transfer function
Command ➢ Select the
"Transfer function…"
menu option on the
"View"
menu bar.
ª The
"Transfer function"
dialog box then appears.
Operation 1. Move the
"Gamma"
slider as required.
2. Confirm your setting with the “
"OK"
” button.
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Findings-based work Operator's Manual GALAXIS
8.13 Findings-based work
8.13.1 Explanation
Explanation Findings act as a type of "bookmark" in the volume. When you select a fin-
ding, the views are reconstructed in exactly the same way as they appeared when the finding was set.
Layout ● Findings that you have created are displayed as a sphere in the volume,
or as a circle in 2D slices.
● If you select the finding in a view, the finding is then displayed in the work­space.
● If you select the entry for the finding from the list on the findings toolbar, the whole view is reconstructed.
This means that the workspace that was active when the finding was created is selected.
Forwarding findings It is possible to forward findings that you have created in GALAXIS to col-
leagues and supervisors:
● Created findings (including all view settings) are saved when a Viewer CD is created and can be displayed in the Viewer (see the section "Crea­ting a GALILEOSViewer CD" [ ➙ 46]).
The user of the Viewer CD then sees the views in exactly the same way as when the findings were created.
● Created findings can be used as a basis for automatically created reports (see the section "Creating a findings report" [ ➙ 39])
8.13.2 Creating a new finding
1. Use the center point of the examination window to navigate to the position at which you want to create a finding.
2. Press the
"Insert Finding"
button.
ª The
"Insert Finding"
dialog box appears.
3. Enter a name for the finding in the
"Title"
text box.
4. Enter a description of the region in the
"Regio"
text box.
5. Enter the finding in the
"Finding"
text box.
6. Press the
"OK"
button.
ª The finding is saved.
Options for findings Extended options can be set in the
"Insert Finding"
dialog box (or in the
"Fin
-
ding details"
dialog box):
"Save contrast and brightness with this finding"
NOTICE
Right-click in the
"Finding"
field to open a menu containing standard texts for findings. These texts can be changed in SIDEXIS. For more information, re­fer to the document "SIDEXIS XG / Operator’s Manual" (order number: 59 62 134).
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
39
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Findings-based work
bеЦдблЬ
"Save panning and zoom with this finding"
If these options are deactivated, when the finding is focused, the settings are no longer displayed in the same way as they were when the finding was created, but the general settings are retained instead.
8.13.3 Selecting a finding
➢ Select the required finding from the list of findings.
or
➢ Select the required finding in the views.
8.13.4 Changing the position of a finding
1. Select the appropriate finding.
2. Use the center point of the examination window to navigate to the re­quired position.
3. Press the
"Move finding to current position"
button.
8.13.5 Editing a finding
1. Select the appropriate finding.
2. Press the
"Properties"
button.
ª The
"Properties"
dialog box appears.
3. Enter a name for the finding in the
"Title"
text box.
4. Enter a description of the region in the
"Regio"
text box.
5. Enter the finding in the
"Finding"
text box.
6. Press the
"OK"
button.
ª The finding is saved.
8.13.6 Deleting a finding
1. Select the appropriate finding.
2. Press the
"Remove finding..."
button.
ª A confirmation prompt box opens.
3. Press the
"Yes"
button.
ª The finding is deleted.
8.13.7 Creating a report
Created findings can be used as a basis for automatically created reports.
As a prerequisite, the relevant current version of the REPORTER software (sold separately) must be installed.
NOTICE
Right-click in the
"Finding"
field to open a menu containing standard texts for findings. These texts can be changed in SIDEXIS. For more information, re­fer to the document "SIDEXIS XG / Operator’s Manual" (order number: 59 62 134).
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Reconstructing a "Ceph p.a./a.p." view Operator's Manual GALAXIS
For more information on the REPORTER software, refer to the document "REPORTER / Installation and Operator's Manual" (order no.: 62 59 316).
8.14 Reconstructing a "Ceph p.a./a.p." view
Validity
This function is not available for GALAXIS in connection with model version GAX5.
Explanation If a p.a. or an a.p. Ceph view is required, it can be reconstructed with the
"Ce
-
ph p.a./a.p. …"
function. The Ceph view is reconstructed from the entire vo-
lume.
The starting values for reconstructing the Ceph view can be corrected prior to reconstruction. This can be helpful if the patient was not properly positioned during image acquisition.
Command ➢ Click on the
"Ceph p.a./a.p. …"
menu option on the
"Analysis"
menu bar.
ª The
"New Ceph"
dialog box then appears.
Design The central element in the
"New Ceph"
dialog box is the display of an axial
plane of the current volume.
The initial settings are made via this axial plane.
Preview A coarse p.a. or a.p. preview of the current settings is displayed in the lower
part of the
"New Ceph"
dialog box.
The position of the axial slice is indicated in this preview by a yellow line.
Selection The Ceph type (a.p. or p.a.) can be selected via radio buttons in the
"Ceph
type"
dialog box.
Alignment
empty
Rotating the axial plane
The axial slice can be rotated by altering the value entered in the
"Turn left/
right"
input field or by actuating the rotation buttons located next to it.
empty
Tilting the axial plane forward or backward
The axial slice can be tilted horizontally by changing the value entered in the
"Nod up/down"
input field.
empty
Resetting the changes made
All of the changes made since the last time the dialog box was opened can be reset by clicking the
"Back to default"
button.
Aids The
"New Ceph"
dialog box features tools which simplify editing.
empty
Detailed preview
A detailed preview of the current settings can be displayed by clicking the
"High quality preview"
button.
● This may take several seconds.
● You can thus display the exact results of your settings before closing the dialog box.
5
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
41
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Reconstructing a lateral Ceph view
bеЦдблЬ
empty
Simultaneous display of all axial slices
This enables you to obtain an overview of the position of the dentition and of the individual teeth.
➢ Press the
"Superimpose"
button.
ª The sum of all axial slices will then be displayed.
empty
Reset
➢ Press the
"Single"
button.
ª One axial slice will then be displayed.
empty
Navigating through axial slices
To navigate through the axial slices, proceed as follows:
➢ Click on the
"Navigate into the volume by one slice."
button.
ª The current view then moves in one slice.
➢ Click on the
"Navigate out of the volume by one slice."
button.
ª The current view then moves in one slice.
Accepting settings ➢ Press the
"OK"
button.
ª The
"New Ceph"
dialog box is then closed and a Ceph view is recon-
structed.
8.15 Reconstructing a lateral Ceph view
Validity
This function is not available for GALAXIS in connection with model version GAX5.
Explanation If a lateral Ceph view is required, it can be reconstructed with the
"Ceph lateral
…"
function. The Ceph view is reconstructed from the entire volume.
The starting values for reconstructing the Ceph view can be corrected prior to reconstruction. This can be helpful if the patient was not properly positioned during the acquisition of the X-ray image.
Command ➢ Click on the
"Ceph lateral …"
menu option on the
"Analysis"
menu bar.
ª The
"New Ceph"
dialog box then appears.
Design The central element in the
"New Ceph"
dialog box is the display of an axial
plane of the current volume.
The initial settings are made via this axial plane.
Preview A coarse lateral preview of the current settings is displayed in the lower part
of the
"New Ceph"
dialog box.
5
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Reconstructing a lateral Ceph view Operator's Manual GALAXIS
The position of the axial slice is indicated in this preview by a yellow line.
Selection The Ceph type (LR or RL) can be selected via radio buttons in the
"Ceph type"
dialog box.
Alignment
empty
Rotating the axial plane
The axial slice can be rotated by altering the value entered in the
"Turn left/
right"
input field or by actuating the rotation buttons located next to it.
empty
Tilting the axial plane to the left or to the right
The axial slice can be tilted to the left or to the right by changing the value en­tered in the
"Tilt left/right"
input field.
empty
Resetting the changes made
All of the changes made since the last time the dialog box was opened can be reset by clicking the
"Back to default"
button.
Aids The
"New Ceph"
dialog box features tools which simplify editing.
empty
Detailed preview
A detailed preview of the current settings can be displayed by clicking the
"High quality preview"
button.
● This may take several seconds.
● You can thus display the exact results of your settings before closing the dialog box.
empty
Simultaneous display of all axial slices
This enables you to obtain an overview of the position of the dentition and of the individual teeth.
➢ Press the
"Superimpose"
button.
ª The sum of all axial slices will then be displayed.
empty
Resetting the
"Superimpose"
function
➢ Press the
"Single"
button.
ª One axial slice will then be displayed.
empty
Navigating through axial slices
To navigate through the axial slices, proceed as follows:
➢ Click on the
"Navigate into the volume by one slice."
button.
ª The current view then moves one slice into the volume each time you
press the button.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
43
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Creating a detail volume
bеЦдблЬ
➢ Click on the
"Navigate out of the volume by one slice."
button.
ª The current view then moves one slice out of the volume each time you
press the button.
Accepting settings ➢ Press the
"OK"
button.
ª The
"New Ceph"
dialog box is then closed and a Ceph view is recon-
structed.
8.16 Creating a detail volume
Scope
This function is not available for GALAXIS in connection with model version GAX5.
Explanation If a detail reconstruction is required, one can be produced with the
"Close-Up"
function.
The
"Close-Up"
function creates a detail reconstruction from the region dis-
played inside the examination window.
The region inside the examination window is reconstructed with a voxel size of 150 µm.
Command
➢ Click the
"Close-Up"
button.
or
➢ Use the
"Analysis"
menu bar to select the
"Close-Up"
menu option.
ª The
"Close-Up"
dialog box appears.
Configuration The examination window from the panoramic view is displayed in 3 planes
(axial, coronal and sagittal).
This is the region of the detail reconstruction to be created.
If the displayed region does not correspond to the desired region, you must close the dialog box and reposition the examination window.
Operation
empty
Assigning names
A name can be assigned to the detail reconstruction in the
"Name"
input field.
This name will then be displayed on the tab.
If no name is assigned, automatic incrementation will take place.
empty
Accepting settings
➢ Press the
"OK"
button.
ª The
"Close-Up"
dialog box is then closed and a detail reconstruction is
produced.
5
61 23 488 D 3474
Functions Sirona Dental Systems GmbH
Showing and hiding orientation lines and findings Operator's Manual GALAXIS
8.17 Showing and hiding orientation lines and findings
empty
Explanation
In the options bar (next to the "Sirona" logo), you can determine whether the orientation lines and findings are displayed or hidden.
To do this, choose the relevant button.
empty
Show/hide orientation lines
This function enables you to show or hide the yellow orientation lines (cros­shairs) in the 2D views.
empty
Show/hide findings
This function enables you to show/hide findings in the views. For more infor­mation, refer to the section "Findings-based work" [ ➙ 38].
8.18 Settings
Explanation In the
"Customize"
menu option, you can make settings that affect how GA-
LAXIS is closed and viewed.
Command Use the
"View"
menu bar to select the
"Customize…"
menu option.
Operation
empty
"Closing"
tab card
When GALAXIS is closed, SIDEXIS XG can transfer the slices displayed in the current workspace to the SIDEXIS database either as individual views or as part of an examination.
The following option fields are available:
● Option button
"Automatically make a screenshot and place it in a new ex
-
am."
● Option button
"Automatically make a screenshot and place it in the cur
-
rent exam."
● Option button
"Just close the scan; don't make a screenshot."
● Option button
"Prompt me each time I close a scan."
empty
"View"
tab card
"Image Quality/Performance"
area
If you select the
"This setting is not recommended for less powerful
computers, as it can decrease the application's interactivity."
check-
box, you can reduce the noise of the planes in the views.
Under certain circumstances, this measure may reduce interactivity due to the considerably higher computing effort it requires.
If you select the checkbox
"Update all slices continuously when na
-
vigating in a given slice."
, all slice images are continually updated
when navigating.
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
45
Sirona Dental Systems GmbH Functions
Operator's Manual GALAXIS Settings
bеЦдблЬ
Under certain circumstances, this measure may reduce interactivity due to the considerably higher computing effort it requires.
"Viewing Direction"
area
Here you can select the relevant option field to determine the direction of view of the sagittal and axial slice images.
empty
Accepting settings
➢ Press the
"OK"
button.
61 23 488 D 3474
Create GALILEOS Viewer CD Sirona Dental Systems GmbH
Operator's Manual GALAXIS
9 Create GALILEOS Viewer CD
Explanation You can burn the 3D data to CD together with the panoramic view and the 3D
viewer software GALILEOS Viewer.
● The recipient of this GALILEOS Viewer CD can view the image data in­teractively in 3D without having to install SIDEXIS or GALAXIS.
● GALILEOS Viewer
The GALILEOS Viewer provides all GALAXIS functions except for a few restrictions (creating detail volumes and saving are not possible).
● As an alternative to burning the data to CD, you can copy the data to a directory or to another medium (e.g. to a USB stick).
Open dialog box
➢ Press the
"Create GALILEOSViewer CD (Wrap&Go)"
button.
or
➢ Go to the
"Workspace"
me nu bar and select the
"Create GALILEOSView
-
er CD (Wrap&Go)"
menu option.
ª The dialog box appears.
Selecting additional files and functions
"Anonymize patient data"
check box
This function makes the patient data anonymous.
"Include findings"
check box
NOTICE
The GALILEOS Viewer can also be used to view planes and image data from the GALILEOS Implant software interactively in 3D. However, the data cannot be changed (see also GALILEOS Implant operating instructions).
61 23 488 D 3474 D 3474.208.01.06.02 04.2010
47
Sirona Dental Systems GmbH Create GALILEOS Viewer CD
Operator's Manual GALAXIS
bеЦдблЬ
This function integrates the list of findings.
"Attachments"
area
You can use the
"Add"
button to select additional files (for example, a re-
port on the current case in PDF format or a ZIP file with DICOM data).
The selected files are displayed in the text box.
You can use the
"Remove"
button to delete selected files.
Selecting a function ● Option button
"Write to a CD"
You want to create a GALILEOS Viewer CD.
● Option button
"Copy this directory"
You want to copy data.
Use the
"Select directory…"
button to select the required target directory.
Start burn or copy procedure ➢ Press the
"Burn CD"
button.
ª The GALILEOS Viewer CD is created or the data is copied to the selected
directory.
tЙ=кЙлЙкоЙ=нЬЙ=кбЦЬн=нз=г~вЙ=~еу=~днЙк~нбзел=пЬбЕЬ=г~у=ДЙ=кЙимбкЙЗ=ЗмЙ=нз=нЙЕЬебЕ~д=бгйкзоЙгЙенлK
«=pбкзе~=aЙен~д=pулнЙгл=dгДe=OMNM pйк~ЕЬЙW==ЙеЦдблЕЬ= mкбенЙЗ=бе=dЙкг~еу a=PQTQKOMUKMNKMSKMO MQKOMNM ûKJkêKW= NNO=RSV
pбкзе~=aЙен~д=pулнЙгл=dгДe
áå=íÜÉ=rp^W
c~Дкбвлнк~Й=PN SQSOR=_ЙелЬЙбг dЙкг~еу пппKлбкзе~KЕзг
pбкзе~=aЙен~д=pулнЙгл=ii` QUPR=pбкзе~=aкбоЙI=pмбнЙ=NMM `Ь~кдзннЙI=k`=OUOTP rp^
lêÇÉê=kç
SN=OP=QUU=a=PQTQ
Loading...