Navigator™ is a U.S. trademark of
Mindray DS USA, Inc.
Nellcor® is U.S. registered trademark of Nellcor
Puritan Bennett Inc.
NIV Line™ is a trademark of Oridion
Medical Ltd.
Oxiband® is a U.S. registered trademark of
Nellcor Puritan Bennett Inc.
OxiMax® is a US registered trademark of
Nellcor Puritan Bennett Inc.
Oxisensor® is a U.S. registered trademark of
Nellcor Puritan Bennett Inc.
Oxismart® is a U.S. registered trademark of
Nellcor Puritan Bennett Inc.
Panorama® is a U.S. trademark of
Mindray DS USA, Inc.
Passport 2® is a U.S. registered trademark of
Mindray DS USA, Inc.
®
FilterLine
is a U.S. registered trademark of
Oridion Medical Ltd.
®
LaserJet
is a U.S. registered trademark of
Hewlett Packard.
®
is a U.S. registered trademark of
LNOP
Masimo Corp.
®
Masimo SET
is a U.S. registered trademark
of Masimo Corp.
®
Max-Fast
is a U.S. registered trademark of
Nellcor Puritan Bennett Inc.
®
MediCO
is a registered trademark of
2
Oridion Medical Ltd.
®
miniMediCO
is a registered trademark of
2
Oridion Medical Ltd.
®
Microstream
is a U.S. registered trademark
of Oridion Medical Ltd.
PatientNet® is a U.S. registered trademark of GE
Medical Systems Information Technologies.
Spectrum® is a U.S. registered trademark of
Mindray DS USA, Inc.
Spectrum OR™ is a U.S. trademark of Mindray
DS USA, Inc.
Velcro® is a registered trademark of Velcro
Industries B.V.
View 12™ is a U.S. trademark of
Mindray DS USA, Inc.
Vigilance® is a US registered trademark of
Edwards Lifesciences Corporation.
Visa® is a U.S. registered trademark of Mindray
DS USA, Inc.
0070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Table of Contents
Foreword....................................................................................................................................................... v
Male 15 Pin D-Shell to Female 15 Pin D-Shell and Open Ended to Female 15 Pin D-Shell .................................1 - 8
Serial Port to Gas Module II/SE/SE with Spirometry Cable (P/N 0012-00-1276-XX) ..............................................1 - 9
Serial Port to Serial Port Cable (P/N 0012-00-1275-01) .....................................................................................1 - 10
26 pin Molex to Mini Phone Plug (DPD Sync Cable)............................................................................................1 - 11
ECG Shielded Lead Wires ...............................................................................................................................1 - 12
ECG Shielded Lead Wires ...............................................................................................................................1 - 13
Panorama Mobility Lead Wires ........................................................................................................................1 - 14
ECG Cable ESIS and Non ESIS ........................................................................................................................ 1 - 15
Panorama Mobility Cable (ESIS and Non ESIS) .................................................................................................. 1 - 16
™
View 12
12 Lead Wire Set ........................................................................................................................................... 1 - 17
Serial Port to RJ 45 Cable (VISA) ......................................................................................................................1 - 20
Installation Menu ............................................................................................................................................1 - 38
ECG Only ..............................................................................................................................................1 - 19
ECG/IBP (only for serial numbers MSXXXXX-K5 and higher) .........................................................................1 - 19
BIS Troubleshooting .................................................................................................................................1 - 34
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-02i
Table of Contents
System Information Menu................................................................................................................................. 1 - 42
Warning and Guidelines ................................................................................................................................. 4 - 1
Test Equipment and Special Tools Required........................................................................................................ 4 - 2
Pixel Test ................................................................................................................................................ 4 - 6
Color Test...............................................................................................................................................4 - 7
Temperature Verification........................................................................................................................... 4 - 22
Perform Verification and NIBP Calibration – Annually ..................................................................................5 - 1
Perform Verification and CO
User Preventative Maintenance Introduction .......................................................................................................5 - 2
Care and Cleaning of the Monitor .................................................................................................................... 5 - 2
Decontamination of the Monitor........................................................................................................................5 - 2
Care and Cleaning of SpO
Cleaning and Re-use of a Nellcor
Cleaning CO
Sensors, Adapters and Sampling Components ..............................................................................5 - 3
2
Sterilization and Cleaning of Reusable Cuffs ......................................................................................................5 - 3
Care and Cleaning of Gas Module................................................................................................................... 5 - 4
Care and Cleaning of 3 and 5-lead ECG Cables and Leadwires ..........................................................................5 - 5
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-02iii
Table of Contents
This page intentionally left blank.
iv0070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
ForewordIntroduction
Foreword
The Spectrum®/Spectrum OR™ Service Manual is intended as a guide for technically
qualified personnel during repair and calibration procedures.
This publication may have been updated to reflect product design changes and/or manual
improvements.
Note
Unauthorized servicing may void the remainder of the warranty. Check with the factory or
with a local authorized representative to determine the warranty status of a particular
instrument.
Warning
The Spectrum®/Spectrum OR™ operates on line voltages. Therefore, an electric shock
hazard may exist when the instrument covers are removed. Repair and calibration
procedures should only be performed by qualified personnel who proceed with care and
follow proper servicing techniques. Warnings are given in various Chapters, as well as in
other appropriate locations.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-02v
IntroductionWar nin g
This page intentionally left blank.
vi0070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
1.0
Repair Information
1.1Introduction
This chapter of the Service Manual provides the necessary technical information to perform
repairs to the instrument. The most important prerequisites for effective troubleshooting are
through understanding of the instrument functions as well as understanding the theory of
operation.
1.2Safety Precautions
In the event the instrument covers are removed, observe the following warnings and
guidelines.
1. Do not short component leads together.
2. The instrument covers must not be removed by other than qualified technical personnel
who have received supplementary instructions regarding maintenance of medical
equipment or has equivalent experience in this area.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 1
Troubleshooting GuidelinesRepair Information
1.3Troubleshooting Guidelines
1. Identify the problem - Due to the wide variety of potential symptoms certain
problems may be more subtle than others. One approach to troubleshooting is to set up
the instrument as described in Chapter 4.0. Following the guidelines of the tests will help
determine the problem if one exists.
2. Avoid shorting component leads - During repair procedures, it can become
tempting to make a series of quick measurements. Always turn the power off before
connecting and disconnecting the test leads and probes. The accidental shorting of
leads can easily stress the components and cause a second failure (aside from the safety
risk).
3. Use the Proper equipment - The equipment listed below is suggested to fulfill a
wide range of troubleshooting requirements. It is imperative to use the designated
equipment in order to ensure proper results of any and all test procedures.
4. Clean up the repair area - After any repair, especially after any soldering or
desoldering, clean off the repair area with alcohol and a stiff brush. This will remove
any residual solder flux, in turn allowing the instrument to return to its original
appearance.
1.4Exchange Programs
An exchange program for certain assemblies in the instrument is available. In many cases
replacement of the complete assembly will result in the most expedient repairs.
1.5Special Tools Required
•DVM
• Digital Mercury Manometer - 0 to 300 mmHg
• Safety Analyzer
• Patient Simulator
• Test Chamber / Dummy Cuff (P/N 0138-00-0001-01 (700 cc) or -03 (500 cc)
• BISx Sensor Simulator (P/N 0454-00-0060)
1 - 20070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationDisassembly Instructions
1.6Disassembly Instructions
Before disassembling the unit, perform the following:
• Power down the unit and remove the line cord
• Remove all cable assemblies from the left side, right side and rear of the unit
• Remove any batteries that were installed
• Perform all work on a properly grounded ESD workstation
Removal of the Front Housing
1. Place the unit face down on a protective surface.
2. Loosen the screw from the Comm-Port or filler Port. Remove the Comm-Port or filler port
from the rear of the unit. Remove the eight screws from the rear of the unit.
3. Turn the unit over and carefully remove the front housing assembly.
4. Disconnect the 80 pin ribbon cable from the J1 of the Display / Keypad board mount in
the front housing.
Removal of the TFT Panel
1. Remove Display driver Cable assembly from J5.
2. Remove the Inverter cable assembly from J8 and the display itself.
3. Remove the encoder cable assembly from J4.
4. Remove the Speaker cable assembly from J2.
5. Remove the five screws that secure the Panel board to the front housing and place to the
side. Lift the board up and out.
6. Remove the Keypad cable assembly from J10.
7. Remove the two screws that secure the Speaker holder assembly. Remove speaker
assembly.
8. Remove the two screws that secure the High voltage inverter board to the left rail
assembly.
9. Remove the high voltage cable assembly from J2 of the inverter board. Place board
assembly to the side.
10. Turn the front assembly over and remove the encoder knob from the front.
11. Remove the nut and washer that secures the encoder to the front housing. Push the
encoder out the back of the housing and place to the side.
12. Turn the front housing back over.
13. Remove the two standoffs below the display and place to the side.
14. Remove the four screws that secure the Display assembly to the front housing brackets.
15. Slide the Display to the bottom of the front housing and lift the right side past the bracket
to remove the display.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 3
Disassembly InstructionsRepair Information
Removal of the Module Interface Board
1. Remove the front housing as stated in “Removal of the Front Housing” on page 1-3.
2. Remove the four screws that secure the board to the main frame.
3. Lift board and remove.
Removal of the Main CPU board (Main Frame)
1. Remove the front housing assembly as stated in “Removal of the Front Housing” on
page 1-3.
2. Remove the NIBP Pump assembly and bracket as stated in “Removal of the NIBP Pump”
on page 1-4.
3. Remove the eight (8) screws that secures the metal shield to the back housing.
4. Remove the Power Supply assembly from the rear of the unit as stated in “Removal of the
Power Supply” on page 1-6. Once the Power Supply is removed carefully lift the Main
CPU assembly up (about one inch).
5. Disconnect the Recorder cable from J8.
6. Disconnect the connector J13 (power switch).
7. Disconnect the CO
8. Disconnect the SpO2 connector from the SpO2 board assembly (Masimo® or Nellcor®).
9. Disconnect the connector J203. (Panel board).
10. Carefully lift the CPU board assembly from the back housing.
11. Carefully angle and lift the CPU board assembly from the back housing.
12. Disconnect the connector from J202 and remove the SpO
13. Remove the ten screws that secure the CPU board to the metal frame.
connector from J 23 (CO2 module)
2
assembly.
2
Removal of the NIBP Pump
1. Remove the Front housing assembly as stated in “Removal of the Front Housing” on
page 1-3.
2. Disconnect the tubing from the inline pump filter.
3. Disconnect the connector from J8.
4. Remove Pump assembly from holding bracket.
Removal of the NIBP Module.
1. Remove the Front housing assembly as stated in “Removal of the Front Housing” on
page 1-3.
2. Remove the NIBP Pump as stated “Removal of the NIBP Pump” on page 1-4.
3. Disconnect the cable from J1.
4. Unfasten the NIBP fitting on the side of the back housing with a 3/8 inch nut driver.
5. Slide the NIBP module from the rear of the unit carefully and remove.
1 - 40070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationDisassembly Instructions
Removal of the Masimo® SpO2 Module
1. Remove the Front Housing assembly as stated “Removal of the Front Housing” on
page 1-3.
2. Remove the Main frame assembly as stated “Removal of the Main CPU board (Main
Frame)” on page 1-4.
3. Remove the three screws that secure the Masimo SpO
4. Remove the Cable assembly from J3 of the SpO
5. Lift the Masimo SpO2 Module up and remove.
module to the standoffs.
2
Module.
2
Removal of the CO2 Module
6.
Remove the front housing assembly as stated “Removal of the Front Housing” on page 1-3.
7. Remove the Main frame assembly as stated “Removal of the Main CPU board (Main
Frame)” on page 1-4.
8. Remove the Patient Connector Panel as stated in “Removal of the Patient Connector
Panel” on page 1-5.
9. Remove the four screws that secure the CO2 module to the back housing assembly.
10. Lift the Module up and out of the back housing.
Removal of the Patient Connector Panel
1. Remove the two screws that secure the Patient Connector Panel housing the back
housing.
2. Slide the Patient Connector housing toward the back.
3. Swing the housing open and remove the CO
connector retainer clip (optional).
4. Disconnect the Ribbon cable assembly from the module.
5. Remove the two screws that secure the SpO2 connector to the Panel assembly.
exhaust tubing, connector and input
2
Removal of the Recorder Assembly
1. Open the recorder door and unloosen the captive screws in the rear of the recorder.
2. Slide the recorder from the opening and remove.
Removal of Recorder Interface Board
1. Remove the Front Housing as stated in “Removal of the Front Housing” on page 1-3.
2. Remove the Main frame as stated in “Removal of the Main CPU board (Main Frame)” on
page 1-4.
3. Remove the recorder assembly as stated in “Removal of the Recorder Assembly” on
page 1-5.
4. Remove the Cable assembly from J3.
5. Remove the two screws and pull the board from the unit.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 5
Disassembly InstructionsRepair Information
Removal of the Power Supply
1. Insert a narrow flat blade into each of the four slots and release each tab.
2. Ensure not to damage each tab. Remove the plastic cover.
3. Remove the four screws from the corners of the metal housing.
4. For units with Li-ion batteries only, a control cable is connected to the power
supply as shown in Figure 3-13 on page 3 - 17. The power supply cannot be completely
removed without first disconnecting this cable. Slide the power supply out of the rear of
the monitor until the control cable connector is exposed. Disconnect the control cable.
5. Slide the power supply out of the opening and remove.
Removal of the Battery Holder Assembly
1. Be sure the batteries are removed from the battery holder assembly.
2.
Remove front housing assembly as stated in “Removal of the Front Housing” on page 1-3.
3. Remove the main frame assembly as stated in “Removal of the Main CPU board (Main
Frame)” on page 1-4.
4. Remove CO2 module and CO2 mounting brackets.
5. Remove the five screws that secure the housing to the back housing.
6. Lift and remove the battery holder assembly.
Removal of the 608 MHz Radio Assembly
1. Remove the 2 screws from the radio assembly base.
2. Pull the assembly back from the monitor and place to the side.
1 - 60070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationNurse Call Cable (3 Pin Circular to Unterminated)
NC
NO
Common
9 ft +/- 6 in.
1
2
3
1 in +/- .13 in.
6 ft +/- 1 in
E1
E2
E3
E4
E5
E6
E7
3
2
1
Connector
shield
Circuit Board
E8
BLK
RED
WHT
Cable shield
Cable
shield
BLK
RED
WHT
1.7Nurse Call Cable (3 Pin Circular to Unterminated)
1.7.1P/N 0012-00-1277-01/-02
FIGURE 1-1 Nurse Call Cable (3 Pin Circular to Unterminated)
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 7
VGA Extension CablesRepair Information
Length +/- 6"
Length +/- 6"
1.8VGA Extension Cables
1.8.1Male 15 Pin D-Shell to Female 15 Pin D-Shell and Open Ended to
Female 15 Pin D-Shell
FIGURE 1-2 VGA Extension Cables
DESCRIPTIONPART NUMBERLENGTH (FT.)
Terminated0012-00-0852-016
Open Ended Unterminated0012-00-0852-0225
Open Ended Unterminated0012-00-0852-0350
Open Ended Unterminated0012-00-0852-04100
1 - 80070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationSerial Port to Gas Module II/SE/SE with Spirometry Cable (P/N 0012-00-1276-XX)
2X +/-.131.00
L
1
5
6
9
SEE DETAIL A
DETAIL A
P1
P2
1
2
3
4
5
6-9
N.C.
N.C.
N.C.
N.C.
N.C.
N.C.
24-25
23
22
8-21
7
1-6
P2
P1
1.9Serial Port to Gas Module II/SE/SE with Spirometry
Cable (P/N 0012-00-1276-XX)
-01 12" 9 pin mini D serial to 25 pin D shell
-02 72" 9 pin mini D serial to 25 pin D shell
FIGURE 1-3 Serial Port to Gas Module II/SE/SE with Spirometry Cable
(P/N 0012-00-1276-XX)
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 9
Serial Port to Serial Port Cable (P/N 0012-00-1275-01)Repair Information
Connector Shield
Connector Shield
Cable Shield
N.C.
N.C.
N.C.
N.C.
N.C.
N.C.
1
2
3
4
5
6-9
1
2
3
4
5
6-9
P1P2
Female
Female
P1P2
10 ft +/-1 in.
1.10Serial Port to Serial Port Cable
(P/N 0012-00-1275-01)
FIGURE 1-4 Serial Port to Serial Port Cable (P/N 0012-00-1275-01)
1 - 100070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair Information26 pin Molex to Mini Phone Plug (DPD Sync Cable)
P1
P2
1
2625
2
L
P2
P1
7
6
5
4
3
2
1
8
9
10
11
12
13
14
15
16
17
18
19
20
TIP
RING
SLV
21
22
23
24
25
26
GND
RSV'd
SYNC
Connector shield
Cable shield
1.1126 pin Molex to Mini Phone Plug (DPD Sync Cable)
FIGURE 1-5 26 pin Molex to Mini Phone Plug (DPD Sync Cable)
PART NUMBERLENGTH
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 11
0012-00-1301-018 in. +/- 1 in.
0012-00-1301-0210 ft. +/- 6 in.
ECG Shielded Lead WiresRepair Information
1.12ECG Shielded Lead Wires
P/N 0012-00-1262-XX
DESCRIPTIONDASH #
18" pinch 5 lead set Domestic-01
24" pinch 5 lead set Domestic-02
40" pinch 5 lead set Domestic-03
18" pinch 5 lead set International-04
24" pinch 5 lead set International-05
40" pinch 5 lead set International-06
18" pinch 3 lead set Domestic-07
24" pinch 3 lead set Domestic-08
40" pinch 3 lead set Domestic-09
18" pinch 3 lead set International-10
24" pinch 3 lead set International-11
40" pinch 3 lead set International-12
3/40", 2/60" pinch 5 lead set Domestic-13
3/40", 2/60" pinch 5 lead set International-14
FIGURE 1-6 ECG Shielded Lead Wires
1 - 120070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationECG Shielded Lead Wires
GREEN
BLACK
RED
WHITE
BROWN
L" +/- 1.00"
GREEN
BLACK
RED
WHITE
BROWN
BLACK
YELLOW
GREEN
RED
WHITE
L" +/- 1.00"
BLACK
YELLOW
GREEN
RED
WHITE
BLACK
RED
WHITE
L" +/- 1.00"
TWO PINS
SHORTED
TWO PINS
SHORTED
KEYING BLOCKS
ARE TO BE THE SAME
COLOR AS LEAD WIRE
BLACK
RED
WHITE
YELLOW
GREEN
RED
L" +/- 1.00"
TWO PINS
SHORTED
TWO PINS
SHORTED
KEYING BLOCKS
ARE TO BE THE SAME
COLOR AS LEAD WIRE
YELLOW
GREEN
RED
LEAD
CONNECTOR
PATIENT END
TERMINATION
SHIELD
1.13ECG Shielded Lead Wires
P/N 0012-00-1261-XX
DESCRIPTIONDASH #
18" snap 5 lead set Domestic-01
24" snap 5 lead set Domestic-02
40" snap 5 lead set Domestic-03
18" snap 5 lead set International-04
24" snap 5 lead set International-05
40" snap 5 lead set International-06
18" snap 3 lead set Domestic-07
24" snap 3 lead set Domestic-08
40" snap 3 lead set Domestic-09
18" snap 3 lead set International-10
24" snap 3 lead set International-11
40" snap 3 lead set International-12
3/40", 2/60" snap 5 lead set Domestic-13
3/40", 2/60" snap 5 lead set International-14
FIGURE 1-7 ECG Shielded Lead Wires
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 13
Panorama Mobility Lead WiresRepair Information
1.14Panorama Mobility Lead Wires
P/N 0012-00-1503-XX
DESCRIPTIONDASH #
24", snap, 5 lead set, Domestic-02
24", snap, 3 lead set, Domestic-05
24", snap, 5 lead set, International-11
24", snap, 3 lead set, International-14
FIGURE 1-8 Panorama Mobility Lead Wires
1 - 140070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationECG Cable ESIS and Non ESIS
1.15ECG Cable ESIS and Non ESIS
P/N 0012-00-1255-XX
DESCRIPTIONDASH #
10’ Straight Non ESIS-01
20’ Straight Non ESIS-02
10’ Rt Angle Non ESIS-03
20’ Rt Angle Non ESIS-04
10’ Straight ESIS-05
20’ Straight ESIS-06
10’ Rt Angle ESIS-07
20’ Rt Angle ESIS-08
FIGURE 1-9 ECG Cable ESIS and Non ESIS
ANSI/AAMI EC53-1995IEC CONVENTIONAL STANDARD
LEADCOLORLEADCOLOR
VBrownChest (C)White
Right Arm (RA)WhiteRight Arm (R)Red
Left Leg (LL)RedLeft Leg (F)Green
Left Arm (LA)BlackLeft Arm (L)Yellow
Right Leg (RL)GreenRight Leg (N)Black
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 15
Panorama Mobility Cable (ESIS and Non ESIS)Repair Information
Non ESIS
ESIS
E (RL/N)
B (LA/L)
C (LL/F)
A (RA/R)
D (V/C)
F
E (RL/N)
B (LA/L)
C (LL/F)
A (RA/R)
D (V/C)
F
1.16Panorama Mobility Cable (ESIS and Non ESIS)
P/N 0012-00-1502-XX
DESCRIPTIONDASH #
Non ESIS, 10’, USA-01
Non ESIS, 20’, USA-02
ESIS, 10’, USA-03
ESIS, 20’, USA-04
Non ESIS, 10’, International-05
Non ESIS, 20’, International-06
ESIS, 10’, International-07
ESIS, 20’, International-08
FIGURE 1-10 Panorama Mobility Cable (ESIS and Non ESIS)
ANSI/AAMI EC53-1995IEC CONVENTIONAL STANDARD
LEADCOLORLEADCOLOR
VBrownChest (C)White
Right Arm (RA)WhiteRight Arm (R)Red
Left Leg (LL)RedLeft Leg (F)Green
Left Arm (LA)BlackLeft Arm (L)Yellow
Right Leg (RL)GreenRight Leg (N)Black
1 - 160070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
NIBP: Check CalibrationThe software has detected that the
Unable to obtain a BPPatient movementWait until patient is calm or
Reading too high or too lowIncorrect cuff sizeMeasure Patient limb, use
attempted but no reading was
possible and the retry limit has not
been reached.
cycle has been completed.
has been exceeded.
The hardware overpressure limit
has been exceeded.
overpressure transducer is out of
calibration.
Cuff or hose NOT attached /
leaking
HR irregular / arrhythmia presentCheck Patient and notify
Blood pressure is out
of range.
Improper cuff size / brandMeasure patient limb. Use
Patient movementWait until patient is calm or
Retry will be attempted.
Check for leaks and quality of
peripheral pulses. Decrease
patient movement. Switch cuff
to another limb.
Switch cuff to another limb.
Decrease patient movement.
Press START to retry. Be
prepared to auscultate BP
manually.
Contact Customer Support.
Power cycle unit.
If message reappears, contact
Customer Support.
Press STOP to clear the
hardware overpressure.
If message reappears, contact
Customer Support.
Have the unit calibrated.
If problem persists contact
Customer Support.
gently hold limb.
Check all connections.
Physician.
Check Patient and verify BP
with manual method.
only properly sized
accessories.
correct cuff.
gently hold limb.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 25
Troubleshooting MenusRepair Information
1.25.3SpO2 Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
SpO
: No SensorSensor is not plugged in to the
2
SpO
: Sensor Off
2
(Masimo SET
SpO
®
Only)
: InterferenceNoise detected on the pulse
2
Spectrum.
Sensor may not be connected to
the patient.
signal prevents pulse
discrimination.
SpO
: Pulse SearchHardware settings are being
2
adjusted in order to discriminate
a pulse waveform.
SpO
: No Pulse (Nellcor
2
Only)
SpO
: FailureThe system has detected an
2
SpO
: Low Perfusion
2
(Masimo SET Only)
SpO
: Too Much Light
2
(Masimo SET Only)
No detectable pulse is measuredCheck to patient connection and
unrecoverable failure of the
SpO
system.
2
Patient perfusion is low.Check to patient connection and
There is too much ambient room
light for the sensor to function
properly
SpO
: Unrecognized
2
Sensor (Masimo SET Only)
SpO
: Communication
2
Error
The sensor is not recognized by
the Monitor.
The monitor and the SpO2
modules are not communicating
properly.
SpO
: Board FaultMasimo SET board failed to
2
operate properly.
SpO
: Sensor FaultDefective Sensor.Replace Sensor.
2
: Motion (Nellcor
SpO
2
Only)
SpO
: Check Sensor
2
(Nellcor Only)
Unable to obtain SpO
reading
Motion is detectedDecrease patient motion, check
The SpO2 module has sensed a
poor connection or a bad sensor
Patient has poor perfusion.Switch limbs / Notify physician.
2
Sensor not on Patient.Reapply sensor.
Cables loose / not connected.Check connections, switch cable.
Ambient light.Switch limbs and cover sensor
No SpO
waveformWaveform not selected to
2
Display.
Cable or sensor not plugged inCheck cable and sensor
Low amplitude SpO
signal
2
sensor on same limb as
SpO
2
cuff.
Patient has poor perfusion.Switch limb / Notify physician.
Plug the sensor into the monitor.
Check patient connection.
Decrease patient motion, check
sensor.
Change to site where pulse is
stronger if patient is
vasoconstricted. Change or
readjust sensor if loose.
patient status.
Power cycle unit. If message
reappears, contact Customer
Support.
patient status.
Minimize the room light around
the patient. Check sensor.
Replace the sensor with a
recommended sensor.
Power cycle unit. If problem
persists, contact Customer
Support.
See Proper Service Menu:
Suggestion.
sensor.
Reconnect the same sensor. If
problem persists, replace sensor
with opaque material.
Go to the Display Setup Menu, choose to display Pleth
in the waveform area.
Check sensor placement, move
as necessary.
1 - 260070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationTroubleshooting Menus
1.25.4Temperature Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
Temperature Probes not
Worki ng
Temperature not displayedImproper display setupCheck display setup in Monitor
Poor contact from probes to bodyCheck the body surface contact
at the probe tip
Reposition or apply
thermoconductive gel
Setup Menu and change as
desired
Cable not plugged inCheck the cable
1.25.5Respiration and CO
MESSAGE/PROBLEMREASONSOLUTION
Resp. Waveform Too LargeScales set inappropriately.Change lead selection.
Resp. Waveform Too SmallPatient breathing is shallow or
False Apnea AlarmApnea delay may be
No Resp. Waveform or Rate
Displayed
CO
: FilterLine® DisconnectedThe FilterLine is not connected
2
CO
: Warming UpThe CO2 sensor has not
2
CO
: Auto-zero In ProgressThe CO2 sensor is. performing
2
CO
: Auto-zero RequestedAn Auto-zero was
2
CO
: FailureCO2 system failure.Contact Technical Support.
2
Troubleshooting
2
Change Respiration scale.
Change lead selection.
patient is turned on side.
Scale set inappropriately.Change respiration scale.
Choose an.other apnea delay
improperly set.
Patient may be having frequent
episodes of CVA.
Scale size may be too low.Change Respiration scale
Respiration turned Off.Turn respiration On (Off will be
Patient connected using ESIS
choke cable.
Cable not connected.Check cable.
to the monitor.
reached its operating
temperature.
(The monitor was just turned
on).
an auto-zero
automatically requested by the
system.
Reposition electrodes to better
detect respirations.
displayed in Resp. window).
Check that proper patient cable
is used. Use non ESIS patient
cable.
Connect the FilterLine.
Wait for the message to go
away.
It takes typically 30 seconds for
the sensor to warm up.
Wait for the auto-zero to
complete.
Wait for the auto-zero to
complete.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 27
Troubleshooting MenusRepair Information
MESSAGE/PROBLEMREASONSOLUTION
CO2: OcclusionSampling pump line is blocked
while the CO
pump is on.
sidestream
2
Check sampling line and filter
for blockage, clear sampling
line if possible.
Replace sampling line if
necessary.
Disconnect and reconnect the
FilterLine from the Spectrum in
order to clear this message.
CO
: PurgeThe system has detected a
2
blocked FilterLine
®
and has
Check FilterLine and replace if
necessary.
attempted to unblock it by
temporarily increasing the flow
rate.
CO
: Check Flow RateThe system has detected a high
2
or low flow rate.
Check FilterLine and replace if
necessary.
“CHK Lead” MessageIncreased impedance caused
by one of the following:
Chest hair under electrodes.Prep chest.
Dried electrode gel.Change electrodes.
Electrode off.Replace electrode.
Lead off.Replace lead.
Cracked lead wires. Replace lead wires.
Poor skin prep.Clean and abrade skin before
applying electrodes.
“CVA” MessageCan be caused by shallow
Check the patient.
breathing or an apnea event.
Patient HR and respiratory rate
identical.
Adjust scales or leads if
necessary.
Check the patient.
1.25.6Gas Module Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
GM: Warming UpAppears when the system has been
turned on, and the sensors have not
reached their stable operating
temperature.
GM: Agent Warming UpThis message appears after the GM:
Warming Up message disappears.
It indicates that the Agent ID Bench is
warming up and readings will not be
available.
GM: Exhaust BlockedAppears when the system detects a
blockage at the exhaust gas outlet, as
indicated by an increase in internal
pressure.
1 - 280070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Wait for the message to go
away. It takes up to five
minutes for the device to
warm up.
Wait for the message to go
away. It takes up to five
minutes from power-up for the
Agent ID Bench to warm up.
Remove waste gas
scavenging assembly, check
if message disappears.
Check exhaust line for
blockage and clear if
possible. If message persists
contact Customer Support.
Repair InformationTroubleshooting Menus
MESSAGE/PROBLEMREASONSOLUTION
GM: Mixed AgentsAppears when more than one
anesthetic agent is detected by the
system.
GM: Air LeakAppears when the system detects a
pneumatic leak.
Also may appear when the Gas
Module has been turned on without a
sample line attached.
Gas Module has been on for a long
period of time without the Spectrum
Monitor being on.
GM: Replace TrapIndicates residue build-up on the
Message will disappear
when a single agent is
detected again.
Turn Gas Module and
Spectrum Off.
Install/check sample lines,
filters, water trap and
electrical connections.
Turn off Gas Module.Turn on
Gas Module and Spectrum
Monitor.
Replace water trap reservoir.
water trap membrane that is
decreasing air flow.
GM: OcclusionAppears when the system detects an
obstruction in the sampling line or the
water trap bottle is full.
Empty and rinse water trap.
Change water trap if
necessary. Check sampling
line and filter for blockage,
clear sampling line if
possible. Replace sampling
line and/or filter if necessary.
If problem persists, contact
Customer Support.
GM: Zero In ProgressAppears when the system is zeroing
all of it’s channels. This appears
This is normal operation.
Wait for message to clear.
whether initiated by the user or is
automatic.
GM: CO
Zero ErrorAppears when the system has been
2
unable to successfully zero the CO
sensor.
Manually start zeroing the
system again. If problem
2
persists, contact Customer
Support.
GM: O
Zero ErrorAppears when the system has been
2
unable to successfully zero the O
sensor.
Manually start zeroing the
system again. If problem
2
persists, contact Customer
Support.
GM: N
O Zero ErrorAppears when the system has been
2
unable to successfully zero the N
sensor.
Manually start zeroing the
system again. If problem
O
2
persists, contact Customer
Support.
GM: Agent Zero ErrorAppears when the system has been
unable to successfully zero the
anesthetic agent sensor.
Manually start zeroing the
system again. If problem
persists, contact Customer
Support.
GM: Pump OffAppears when the system has turned
off the pump due to a pneumatic
error.
Restart the pump from the
Gas Menu. If problem
persists, contact Customer
Support.
GM: Agent Mismatch HAL
GM: Agent Mismatch ISO
Appears when the system detects
Halothane as the primary agent and
the manually selected agent is not
Halothane.
Appears when the system detects
Isoflurane as the primary agent and
the manually selected agent is not
Isoflurane.
Match the Agent
administered with the Agent
selected, or select Agent
Auto ID.
Match the Agent
administered with the Agent
selected, or select Agent
Auto ID.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 29
Troubleshooting MenusRepair Information
MESSAGE/PROBLEMREASONSOLUTION
GM: Agent Mismatch ENF
Appears when the system detects
Enflurane as the primary agent and
the manually selected agent is not
Enflurane.
GM: Agent Mismatch SEV
Appears when the system detects
Sevoflurane as the primary agent and
the manually selected agent is not
Sevoflurane.
GM: Agent Mismatch DES
Appears when the system detects
Desflurane as the primary agent and
the manually selected agent is not
Desflurane.
GM: Unknown AgentAppears when the system detects a
Match the Agent
administered with the Agent
selected, or select Agent
Auto ID
Match the Agent
administered with the Agent
selected, or select Agent
Auto ID
Match the Agent
administered with the Agent
selected, or select Agent
Auto ID
Use recognized agent
gas that does not match the
spectroscopic signatures of the five
known anesthetic agents
GM: Cannot Zero...
RETRYING
Appears when the Spectrum
requests Zeroing (either on the
automatic cycle or by a user request)
and the Gas Module is unable to
Allow system to retry without
intervention. If problem
persist, contact Customer
Support.
initialize the cycle
GM: CO
UncalibratedAppears after an unsuccessful
2
calibration attempt of the CO
The numeric data for CO
as - - -, and the CO
a flatline
will appear
2
waveform will be
2
sensor.
2
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Customer
Support.
GM: O
UncalibratedAppears after an unsuccessful
2
calibration attempt of the O
The numeric data for O
as - - -, and the O
will appear
2
waveform will be a
2
flatline
sensor.
2
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Customer
Support.
GM: N
O UncalibratedAppears after an unsuccessful
2
calibration attempt of the N
sensor. The numeric data for N
appear as - - -, and the N
waveform will be a flatline
2
O
2
O
2
O will
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Customer
Support.
GM: Agents UncalibratedAppears after an unsuccessful
calibration attempt of the agent
sensor. The numeric data for all
agents will appear as - - -, and the
agent waveform will be a flatline
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Customer
Support.
GM: FailedAppears when the Gas Module
Contact Customer Support.
detects an unrecoverable error in its
own operation
GM: DisconnectedAppears when the Spectrum cannot
detect signals being sent by the Gas
Module
Ensure Gas Module is turned
on and interface cable is
properly connected. If
problem persists, contact
Customer Support.
1 - 300070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationTroubleshooting Menus
MESSAGE/PROBLEMREASONSOLUTION
Sampling ErrorAppears when a sampling error
Not Ready For
Calibration
Calibration Error,
Sampling Error
Calibration Error, Zeroing
Error
occurs on one or more Gas Module
channels during calibration
Appears when the Gas Module is
unable to initialize calibration
Appears when a sampling error
occurs in all four Gas Module
channels during calibration
Appears when the Gas Module
cannot perform a Zeroing during
calibration
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 31
Troubleshooting MenusRepair Information
1.25.7IBP Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
Damped Invasive
Wav efo rm
IBP not Displayed / No
IBP Waveform
Abnormally High or Low
readings
Unable to ZeroStopcock not open to
Air bubbles in tubing.Eliminate air from tubing.
Kinked catheter.Change position of catheter,
check patient.
Catheter against wall of blood
vessel.
Blood in tubingPump pressure bag up to 300
Catheter partially occluded with
clot.
Improper Setup.Check display setup in monitor
Cable not plugged inCheck cable.
Transducer not connected.Check transducer connection.
Stopcock turned improperly.Check transducer.
Transducer not zeroed.Check and zero the transducer.
Transducer too HIGH or to LOW.Check patient adjust transducer
atmosphere.
Cable/Transducer not
plugged in.
Check for leaks at connector, flush
catheter.
mmHg.
Consult physician.
setup.
rezero.
Check transducer.
Check cable.
1.25.8PAWP Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
Unable to WedgeImproper catheter positionCheck PA catheter, notify
1 - 320070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationTroubleshooting Menus
1.25.9EPM Cardiac Output Troubleshooting
MESSAGE/
PROBLEMREASONSOLUTION
CO value higher /
lower than expected
No measurement/
Unable to measure
CO Signal Under
Range
CO Out of RangeAppears if the CO is out of the
Irregular curveImproper injection procedure.Check hospital policy, inject in
Delayed InjectionAppears if the time between the start of
Injectate Temp ErrorAppears when the temperature of the
Noisy BaselineCardiac Output waveform baseline is
MeasuringAppears once an injection is detected
Injectate Temp. Out
of Range
Computation constant incorrect for PA
catheter type, injectate temperature,
and injectate volume.
Catheter may be kinked or not in proper
position.
Unstable temperature.Check injectate temperature.
No temperature or temperature out of
range.
Time elapsed for measurement.Discard bolus fluid.
Appears if the CO curve is not sufficient
for a CO calculation or if a curve is not
detected within thirty (30) seconds
measurable range
(0.2 l/min to 20.0 l/min). Computation
constant incorrect for PA catheter type,
injectate temperature, and injectate
volume.
Catheter may be kinked or not in proper
position.
Patient movement during injection.Have patient lay still during
Measurement displayed if CO curve
has multiple peaks, failure to return to
baseline or irregularities in curve.
the CO measurement and the onset of
the temperature change is more than
fifteen (15) seconds
injectate is too warm (>27ºC) or the
difference between the injectate and the
blood temperature is <8ºC.
unstable
during the process of a CO run.
Appears when the temperature of the
injectate is too warm (>27ºC) or the
difference between the injectate and the
blood temperature is <8ºC.
Check computation constant
and enter correct data.
Notify physician.
Flush PA catheter.
Check patient.
Wait f o r Ready or Inject when Ready message to
appear. Rebolus when ready.
Rebolus if necessary
Check computation constant
and enter correct data. Rebolus
when ready.
a smooth and fluid bolus.
Notify physician.
bolus procedure.
Rebolus when ready
Rebolus when ready.
Ensure that the CO bolus is
initiated within 15 seconds
Check the temperature of the
injectate
Check injectate fluid, insure
fluid is not under warm lights,
near a warming blanket or
another warm source.
Rebolus when ready.
Check injectate fluid, insure
fluid is not under warm lights,
near a warming blanket or
over-chilled in the ice bath.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 33
Troubleshooting MenusRepair Information
MESSAGE/
PROBLEMREASONSOLUTION
Inject When ReadyAppears if Auto Start is enabled,
stable temperatures are detected
ReadyAppears if Auto Start is not enabled,
and stable temperatures are detected
Inject NowAppears once START has been
pressed, before bolus is initiated
Please WaitAppears after fluid bolus is initiated and
Cardiac Output is being calculated
Bolus when ready
Press START when ready
Bolus when ready
Wait until message disappears
1.25.10Vigilance Cardiac Output Troubleshooting
MESSAGE/
PROBLEMREASONSOLUTION
CO: Check Vigilance.This message is displayed when an
alert or alarm status has been sent
from the Vigilance to the Spectrum.
Refer to the Edwards Vigilance
Monitor Operator’s Manual or
contact Edwards Lifesciences
Corporation for assistance.
®
1.25.11BIS Troubleshooting
MESSAGE/
PROBLEMREASONSOLUTION
BIS: Check SensorIncorrect sensor application.Read Instructions on sensor
Poor sensor connections.Check sensor connections.
Sensor Check fails.Re-prep again or replace
Defective PIC.Replace the PIC.
Defective BISx.Replace the BISx.
Problem is detected relating to sensor
ground element. Sensor Overcurrent
Sensor is using too much current.
BIS: No SensorDisconnected sensor.Connect the sensor.
Poor or contaminated connection
between sensor and PIC.
Disconnected PIC.Connect the PIC.
Defective PIC.Replace the PIC.
Defective BISx.Replace the BISx.
BIS: ArtifactArtifact, such as those generated by
motion or eye blinks, is causing loss of
EEG recognition. The BIS value and other
trend variables that are adversely
affected by artifact are not displayed.
Within the USA: (800)-424-3278
Outside the USA: (949)-250-2500
package and re-prep sensor.
sensor. Verify Sensor Check
passes.
Disconnect and examine sensor
connection. Clean any
contamination present. Replace
sensor if necessary.
Connect/clean connection
between sensor and PIC.
Attempt to identify and
eliminate artifact source.
1 - 340070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationTroubleshooting Menus
MESSAGE/
PROBLEMREASONSOLUTION
BIS: Check SQI LevelEMG Bar indicates electrical activity that
may be interfering with EEG recognition.
Defective PIC.Verify Sensor Check passes. If
Defective BISx.Replace the BISx.
NOTE: This message is displayed when
the Signal Quality is less than half of the
level desirable for optimal monitoring
conditions or when the signal quality is
too low to accurately calculate a BIS
value. This may occur as the result of
artifact (non-EEG signal) such as that
generated from motion (patient
movement or eye blinks) or the presence
of electrocautery, warming blankets, or
other devices.
BIS: Invalid SensorPoor or contaminated connection
between sensor and PIC.
Defective sensor.Replace the sensor.
Defective PIC.Replace the PIC.
Defective BISx.Replace the BISx.
BIS: NoiseThe signal from the electrode goes
beyond the measurable range
If EMG bar is illuminated,
attempt to determine and
eliminate cause.
not, replace PIC.
Connect/clean connection
between sensor and PIC.
Re-prep the electrodes and
check all connections.
1.25.12Alarm Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
High or Low or No Alarm
Sound
No Arrhythmia
Alarm Sound
Alarms continue to Sound
despite pressing MUTE
No Alarm printout with
Alarm violation
Alarm limits not setGo to Alarm Setup and adjust
Alarm Mute All, On time has
not expired
Arrhythmia option not installedCall Sales Rep to purchase option
Arrhythmia Alarms offGo to Monitor Setup /
Monitor is in learning modeWait until learning is concluded
More than one alarm is activePress MUTE or MUTE ALL key to
Print on Alarm is set to OffGo to Print Menu and set Print
alarms
Press MUTE ALL to reactivate
alarms
Advanced Setup to activate
alarm
and monitor patient closely
silence
Check Patient
on Alarm to On
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 35
Troubleshooting MenusRepair Information
1.25.13Trends Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
No Trends displayedNo Trend triggers setGo to the Monitor Setup
Trend page is scrolledUse scroll button in Trend Menu
Menu and set NIBP Trend,
Trend Interval or Alarm
Trend as desired
to scroll to top of Trend Menu
1.25.14Printer/Recorder Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
Recorder Report Appears
Totally Blank
Local Printer Door OpenThe printer door is not closedClose the printer door
Local Printer Out Of PaperPrinter out of paperReplace with a new roll of paper
Printer BusyPrinter received multiple print
Local Printer Unable To
Print
Check Laser PrinterLaser printer is busy,
No print on AlarmAlarm printing not activeGo to Print Setup Menu and
Trends not printingPrint Trend not pressedPress PRINT TREND when trend
Thermal paper may be installed
incorrectly (up-side down)
requests at one time
The system has detected an
unrecoverable printer failure
disconnected, out of paper or has
a fault condition
No Trends displayedUse scroll feature to scroll to the
No paperCheck / Replace paper
Remove paper and re-install with
paper feeding off of the spool
from the bottom
Wait until the printer is not busy
Power cycle unit. If message
reappears, contact Customer
Support.
Check laser printer
set Print on Alarm to On
window is open
top of the trend then press
PRINT TREND
1 - 360070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationTroubleshooting Menus
1.25.15Monitor/Display Troubleshooting
MESSAGE/PROBLEMREASONSOLUTION
No trace for a desired
parameter
Trace Not MovingFREEZE key may have been
Display Appears to be OffMains power switch may not be
Disabled Alarm ToneMUTE key pressed.Check for alarm mute symbol
Cooling Fan FailureThe unit running on AC power
Patient Information did not
appear on display
Incorrect Date or TimeData not entered or entered
Improper attachment of
transducer or cable to monitor
Faulty transducer or cable.Try a new transducer or cable.
pressed.
on.
Unit may not be plugged into an
AC outlet.
If used as a portable, battery
pack may be drained.
12-lead card removed without
disabling 12-lead.
Beep volume low.Increase beep volume.
and the cooling fan is not
operational.
No data entered.Enter proper patient data.
Done was not selected from
keypad after entering data.
incorrectly.
Check transducer / cable
connection.
Press the FREEZE key to
unfreeze the trace.
Check mains power switch on
side panel.
Check power cord (Is it plugged
in?)
If battery pack is drained, plug
into an AC outlet to recharge the
battery
Always disable 12-lead card
prior to removal.
Power unit back on.
Contact Customer Support.
and message.
Contact Customer Support.
Go to the proper keypad enter
data, select Done when
finished.
Follow instructions from “How
to Set the Clock / Date and
Time”.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 37
Installation MenuRepair Information
1.26Installation Menu
1.26.1Installation Mode
FIGURE 1-19 Installation Menu
The following items are set through the Installation Mode menu: Language, Country, Date
Format, Time Format, NIBP Timeout, CO & PAWP Timeout, Enable EPM, Apnea Latch
(Spectrum OR only), Enable Mute All Permanent Selection (Spectrum OR only), Enable
Arrhythmia All Off Selection (Spectrum OR only), Temperature Units, Weight Units, Height
Units, CO
monitor defaults from card, Set up Serial Port 1, Set up Serial Port 2, Re-boot in demo mode,
Restore factory defaults, System Information, Change Password (Spectrum OR only), and
Options.
1. Enter Installation Mode by pressing and holding the DISCHARGE key (Spectrum
2. Set each item as necessary. The operation of the menu is the same as that of the normal
Units, SpO2 Tones (Spectrum OR only), Copy monitor defaults to card, Copy
2
®
the TRENDS key (Spectrum OR
operating mode. To save all of the selected settings, choose Save Current before
exiting this menu. To access the normal operation screen, power the unit OFF and ON
again.
™
) while powering ON the monitor.
) or
1 - 380070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationInstallation Menu
The following table describes the Installation Menu structure:
MENU TITLE ON
SCREENMENU CHOICESDEFAULT
ACTIONS/
COMMENTS
Save CurrentSelect to save current
settings as defaults
Select LanguageSet up at factorySelect to change
language
Select CountrySet up at factorySelect to change
country
Date FormatM/D/Y, D/Y/MPer countrySelect to change date
format
Time Format12, 24 hourPer countrySelect to change time
format
NIBP Timeout15, 30, 45, 60 mins15Selects time old and
data is removed from
screen
CO & PAWP Timeout15, 30, 45 mins, 1, 2,
4 hr
15 minSelect to change CO
and PAWP timeout.
Enable EPMYes, NoNoSelect to enable EPM
module
Apnea LatchOn, OffOnSelect to turn apnea
alarm latching on or
off.
Enable Mute All
Permanent Selection
Yes, NoYesSelect to enable or
disable the Permanent
Audio Off menu
selection.
Enable Arrhythmia All
Off Selection
Yes, NoNoSelect to enable or
disable the Arrhythmia
All Off menu selection.
Temperature Units° F, ° C° F - USA
° C - All others
Weight Unitslbs, kglbs - USA
kg - All others
Height Unitsft/ inches, cmft/ inches - USA
cm - All others
CO
UnitsmmHg, %, kPammHgSelect to change CO2
2
Select to change
temperature units
Select to ch ange weigh t
units
Select to change height
units
units
SpO
TonesStandard Tones,
2
Alternate Tones
Copy monitor defaults
to card.
Standard TonesSelect to change the
SpO
tones.
2
Select to copy the
monitor defaults and
settings to a data
transfer card inserted
into PCM2
Copy monitor defaults
from card.
Select to copy the
monitor defaults and
settings from a data
transfer card inserted
into PCM2
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 39
Installation MenuRepair Information
MENU TITLE ON
SCREENMENU CHOICESDEFAULT
Set up Serial Port 1None, VISA with
admit, DIAP, Accutorr,
Gas Module, Patient
Net (menu choices are
present only if options
are installed)
Set up Serial Port 2None, VISA with
admit, DIAP, Accutorr,
Gas Module, Patient
Net (menu choices are
present only if options
are installed)
Re-boot in demo modeYes, NoNoSet to YES to start the
Restore factory
defaults
System InformationSelect to set up users
Change PasswordSelect to change
Options Select to add/view
NoneSelect to set up a serial
NoneSelect to set up a serial
ACTIONS/
COMMENTS
output protocol port
output protocol port
monitor in
demonstration mode on
next power-up. Normal
monitoring will resume
after cycling power in
demonstration mode.
Select to restore factory
defaults
screen
password.
options
1.26.2Transferring Monitor Default Settings
When installing several monitors with identical display and alarm settings, a Transfer Card
(Spectrum
can be used to copy the settings from monitor to monitor.
1. Insert the Transfer Card into the PCM2 slot on the right side of the source monitor.
2. Access the Functions menu, and select Copy monitor defaults to card from the
3. Remove the card and insert it into the PCM2 slot of the receiving monitor.
4. Enter Installation Mode on the receiving monitor by pressing and holding the
5. Select Copy monitor defaults to card. A status message will report completion of
6. Select Save Current and power-cycle the receiving monitor to enter normal monitoring
menu. A status message will report completion of the transfer.
®
DISCHARGE key (Spectrum
) or the TRENDS key (Spectrum OR™) while
powering ON the monitor.
the transfer.
mode.
1 - 400070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationInstallation Menu
1.26.3Option Installation
1. Install option PCMCIA card (P/N 0996-00-0053-XXXXX) into slot PCM2.
2. Select “Install” from the menu selection. Press the Navigator™ knob to activate new
option.
3. Verify proper operation as stated in chapter 4.0.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 41
System Information MenuRepair Information
1.27System Information Menu
FIGURE 1-20 System Information Menu
This menu is accessed by selecting System Information from the Installation Menu.
Information is entered via a on-screen keyboard. Select each letter using the Navigator
™
Knob. When finished, rotate to Previous Menu and select using the Navigator Knob.
MENU TITLE ON
SCREEN
Previous MenuSelect to return to Previous Menu
Property OfSelect to set up Property of name
LocationSelect to set up Location
DepartmentSelect to set up Department
ContactSelect to set up Contact
PhoneSelect to set up Phone
Accutor Baud Rate
(Spectrum only)
DIAP Baud Rate9600, 19200Select to change the DIAP baud rate
Enable NetworkNot Used
IP Address:Select to set up IP Address
Subnet Mask ID:Select to set up Subnet Mask ID
Wireless IP Address
(Spectrum only)
MENU
CHOICESDEFAULTTEXT STRINGS
9600, 1200Select to change the Accutorr
protocol baud rate
Select to set up Wireless IP Address
1 - 420070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationSystem Information Menu
MENU TITLE ON
SCREEN
Wireless Subnet Mask ID
(Spectrum only)
Laser Printer IP Address
(Spectrum only)
Network Name:Select to set up Network name
Device ID
MENU
CHOICESDEFAULTTEXT STRINGS
Select to set up Wireless Subnet
Mask ID
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 43
Trend StorageRepair Information
1.28Trend Storage
1.28.1Installation and Use of the Extended Trend Feature
The Spectrum®/Spectrum OR™ monitor is capable of storing, in non-volatile memory, up
to 120 trend values for each parameter, and for Spectrum
CRG data. With the addition of the Extended Trend card (P/N 0996-00-0052-01), the
Spectrum
parameter, and for Spectrum
This feature is added to the Spectrum
card (P/N 0996-00-0052-01) into slot PCM1 on the right side of the monitor. The card
should be inserted before monitor power-up, and never removed during monitor operation.
In order to guard against accidental removal, the card slot is designed so that a tool is
required to eject the card after insertion.
The Extended Trend feature is automatically enabled when the unit is powered-up following
card insertion.
®
/Spectrum OR™ is capable of storing up to 500 trend values for each
®
only, up to 12 hours of OXY-CRG data.
®
/Spectrum OR™ by inserting the Extended Trend
®
only, up to 6 minutes of OXY-
NOTE:All prior trended information is lost upon installation of the
Extended Trend feature.
1.29Software Download
1.29.1Download Operation
A download flashcard must be installed in PCMCIA slot PCM2 prior to power up. When the
unit is powered on, the valid download card will be read and downloaded to the Host CPU
module only. While the images are being downloaded a beep will sound approximately
once every second until the download is complete. Once complete, the unit will then run
automatically, first displaying the Spectrum
normal runtime mode. A display box will be displayed if it is determined that the images
downloaded to the Host CPU module do not match the images in the attached CP and/or
Front-end modules and/or EPM modules. This display box, shown below, will indicate the
Host CPU software version, the current version on the CP, Front-end and/or EPM modules
and the software version being downloaded to the modules. The images will automatically
load and the status message Complete will appear, after the download is complete.
NOTE:To ensure that the EPM module download is complete,
reboot the unit.
ModuleModule SW
Ver s i o n
The above modules are being downloaded. When the download is complete, press the
normal screen button to continue, or reboot.
®
/Spectrum OR™ logo then proceeding to
Flash SW VersionStatusDownload
Percent
FIGURE 1-21 Software Download Status Screen
1 - 440070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Repair InformationSoftware Download
Status Messages
Pending:This is displayed for a module awaiting download.
Active:This is displayed once a pending module has turned active.
Erasing:This is displayed when an image is being erased.
Loading:This is displayed when an image is being downloaded.
Verify:This is displayed while the checksum is being verified.
Fault:This is displayed if there is a downloading error. If a fault occurs, the module
download screen will not be able to be removed. Contact customer service.
Complete: This is displayed for a module when download is complete.
NOTE:If a fault condition occurs, an error message will be
recorded in the error log.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-021 - 45
Software DownloadRepair Information
This page intentionally left blank.
1 - 460070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
2.0
Block Diagrams
2.1Introduction
The Block Diagrams indicate the internal organization of the instrument. The block diagrams
are used to gain both familiarity with the instrument and to locate malfunctioning PC boards
as readily as possible.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-022 - 1
Block DiagramBlock Diagrams
P1
SW1
SW3SW6
SW7
SW10SW13
SW16
SW17
SW14SW11SW8
SW19SW4
SW2
SW15
SW12
SW9
SW20
SW5
SW21
SW18
CPU Board AssemblyP/N 0670-00-0739-02or 0670-00-0782-03
33Standoff Hex Male/Female With Lock Washer0361-00-0164
34Label Part Number Serial NumberN/A
35Power Supply Assembly0014-00-0070
36Dashpot Assembly0103-00-0465
NSMB1 Connector Shroud0334-00-1668
N.A. Not Available
0217-00-0012
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-023 - 33
Comm-PortIsometric Drawings and Part List
This page intentionally left blank.
3 - 340070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
4.0
Calibration Procedure
4.1Introduction
The following procedures are provided to verify the proper operation of the Spectrum®/
Spectrum OR
problems within the Spectrum®/Spectrum OR™ hardware. A menu driven interface
similar to that of the Spectrum
tests.
™
Monitor. Service Diagnostics provide the capability of diagnosing
®
/Spectrum OR™ User Interface, is used to execute all
CAUTION:Calibration is not to be performed while monitoring a
patient.
4.2Warning and Guidelines
In the event that the instrument covers are removed, observe the following warnings and
general guidelines:
1. Do not short component leads together.
2. Perform all steps in the exact order given.
3. Use extreme care when reaching inside the opened instrument. Do not contact exposed
metal parts which may become electrically active.
4. Read and understand each step of the procedure prior to beginning the step.
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-024 - 1
Test Equipment and Special Tools RequiredCalibration Procedure
4.3Test Equipment and Special Tools Required
• Digital Mercury manometer with bulb and valve 0-500 mmHg - Netech Digimano Accuracy 0.25% Full Range
• Test Chamber/
Dummy Cuff
•DVM
• Patient Simulator
• Digital Flow Meter
• Safety AnalyzerDempsy Model or equivalent
• BISxBIS Sensor Simulator P/N 0454-00-0060
P/N 0138-00-0001-01 (700 cc) or -03 (500 cc)
4 - 20070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Calibration ProcedureDiagnostics
4.4Diagnostics
To enter the diagnostic mode:
1. Turn the power OFF.
2. Pressing and hold the FREEZE key (Spectrum®) or the SPIROMETRY key
(Spectrum OR
will appear on screen. Release the FREEZE key or the SPIROMETRY key.
3. Rotate the Navigator™ Knob to move the cursor within the Diagnostics Main
Menu. Pressing the Navigator knob will select the desired test and open the second
test menu.
™
) while powering ON the monitor. The Diagnostics Main Menu
FIGURE 4-1 Diagnostics Main Menu
FIGURE 4-2 Diagnostics Test Menu
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-024 - 3
DiagnosticsCalibration Procedure
Keypad / Control Knob Test
Press keys and the text will appear on screen. Rotate the control knob to move the cursor. Press and hold the Normal Screen Key to exit.
Key Pressed: None
Control Knob Detent:
4.4.1Keypad / Control Knob Test
When this menu is selected the unit will perform an echo test by displaying the name of the
key that was pressed.
1. A blank key name will appear on screen
2. When a key is pressed the name of that key will be displayed in the key name window.
3. Exercise each key to verify proper operation.
4. A second window with blank boxes will be displayed on screen.
5. When rotating the control knob the blank boxes will illuminate with each active detent.
6. Press the PRINT key to print the test result via the internal recorder.
7. Press the NORMAL key and hold to return to the Diagnostics Test Menu.
FIGURE 4-3 Keypad / Control Knob Test
4 - 40070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Calibration ProcedureDiagnostics
4.4.2Recorder Test
Select the Chart Grid ASCII Characters menu.
FIGURE 4-4 Recorder Test Menu
The printer will print the Recorder Test pattern as shown in the figure below.
FIGURE 4-5 Recorder Test Strip
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-024 - 5
DiagnosticsCalibration Procedure
4.4.3Display Tests
The display test offers the choice of a Pixel Test or a Color Test.
FIGURE 4-6 Display Tests Menu
4.4.4Pixel Test
The pixel test will verify the proper operation the display. On screen one half of the screen
will be illuminated, the second half will be black. Pressing the Navigator
illuminate the second half of the screen, the first half will be black. Pressing the Navigator
knob a third time will activate the Display Test Menu screen.
FIGURE 4-7 Pixel Test Menu
™
knob will
4 - 60070-10-0556-02Spectrum®/Spectrum OR™ Service Manual
Calibration ProcedureDiagnostics
White
Red
Blue
Green
4.4.5Color Test
The color test will verify the four basic colors of the display. Press the Navigator™ knob to
view the selected color screens in full illumination. The colors are Red, Blue, Green and
White.
FIGURE 4-8 Color Test
Spectrum®/Spectrum OR™ Service Manual0070-10-0556-024 - 7
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.