Metrohm 893 User Manual

893 Professional Biodiesel Rancimat
Manual
8.893.8001EN / 2014-05-09
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
893 Professional Biodiesel Rancimat
8.893.8001EN / 2014-05-09
Manual
fpe
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://documents.metrohm.com.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 Introduction 1
1.1 Instrument description ......................................................... 1
1.2 About the documentation ................................................... 2
1.2.1 Symbols and conventions ........................................................ 2
1.3 Safety instructions ................................................................ 3
1.3.1 General notes on safety ........................................................... 3
1.3.2 Electrical safety ........................................................................ 3
1.3.3 Flammable solvents and chemicals ........................................... 5
1.3.4 Recycling and disposal ............................................................. 5
2 Overview of the instrument 6
2.1 Front of the instrument ........................................................ 6
2.2 Rear of the instrument ......................................................... 7
Table of contents
2.3 Instrument display ................................................................ 8
3 Installation 10
3.1 Setting up the instrument .................................................. 10
3.1.1 Packaging .............................................................................. 10
3.1.2 Checks .................................................................................. 10
3.1.3 Location ................................................................................ 10
3.2 Mounting accessories ......................................................... 11
3.2.1 Mounting the internal air supply ............................................ 11
3.2.2 Mounting the external air supply ........................................... 13
3.2.3 Assembling the reaction and measuring vessels ..................... 14
3.2.4 Inserting vessels / Establishing tubing connections ................. 17
3.2.5 Mounting the exhaust air collection tube ............................... 18
3.3 Power socket ...................................................................... 19
3.3.1 Replacing fuses ...................................................................... 19
3.3.2 Power supply cable and power socket ................................... 20
3.3.3 Switching the instrument on and off ...................................... 20
3.4 Connecting a computer ...................................................... 22
3.4.1 Connecting the 893 Professional Biodiesel Rancimat and the
computer .............................................................................. 22
4 Operation 23
4.1 Rancimat method ............................................................... 23
4.2 Calibration functions .......................................................... 24
4.2.1 Determining the cell constant ................................................ 24
4.2.2 Determining the temperature correction ................................ 25
893 Professional Biodiesel Rancimat
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
4.3 Determinations ................................................................... 28
4.3.1 Preparing the instrument and the accessories ......................... 28
4.3.2 Preparing the determination .................................................. 29
4.3.3 Cleaning the accessories ........................................................ 31
4.3.4 Cleaning the oil separator ...................................................... 32
5 Operation and maintenance 35
5.1 General notes ...................................................................... 35
5.1.1 Care ...................................................................................... 35
5.1.2 Maintenance by Metrohm Service .......................................... 35
5.2 Replacing the dust filter .................................................... 36
5.3 Regenerating or replacing the molecular sieve ............... 36
5.4 Quality management and qualification with Metrohm .. 37
6 Troubleshooting 38
6.1 Problems ............................................................................. 38
7 Technical specifications 43
7.1 General data ....................................................................... 43
7.2 Temperature control and measurement .......................... 43
7.3 External temperature sensor ............................................. 44
7.4 Conductivity measurement ................................................ 44
7.5 Gas flow regulation ............................................................ 45
7.6 USB interface ...................................................................... 45
7.7 Power socket ...................................................................... 45
7.8 Safety specifications ........................................................... 45
7.9 Electromagnetic compatibility (EMC) ................................ 46
7.10 Ambient temperature ......................................................... 46
7.11 Dimensions/Material .......................................................... 46
8 Conformity and warranty 47
8.1 Quality Management Principles ........................................ 47
8.2 Warranty (guarantee) ......................................................... 48
9 Accessories 50
■■■■■■■■
IV
Index 52
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 Front 893 Professional Biodiesel Rancimat ......................................... 6
Figure 2 Rear 893 Professional Biodiesel Rancimat .......................................... 7
Figure 3 Instrument display ............................................................................. 8
Figure 4 Mounting accessories for the air supply ........................................... 11
Figure 5 Drying flask - Individual parts ........................................................... 12
Figure 6 Drying flask mounted in place ......................................................... 13
Figure 7 Equipping the reaction and measuring vessels ................................. 15
Figure 8 Mounting the air tube: correct - incorrect ........................................ 17
Figure 9 Mounting the foam barrier .............................................................. 17
Figure 10 Instrument display with instrument name ........................................ 20
Figure 11 Instrument display with "no connection" symbol ............................. 21
Figure 12 Instrument display without instrument name and symbol ................ 21
Figure 13 Instrument display with serial number ............................................. 21
Figure 14 Measuring arrangement (schematic representation) ......................... 24
Figure 15 Assembling the reaction vessel for determining the temperature correc-
tion ................................................................................................. 26
Figure 16 Oil separator (6.2753.200), disassembled ........................................ 33
Figure 17 Dust filter - conditions ..................................................................... 36
Table of figures
893 Professional Biodiesel Rancimat
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction

1.1 Instrument description

The 893 Professional Biodiesel Rancimat is a computer-controlled measur­ing device for determining the oxidation stability of biodiesel and biodiesel blends (mixture of biodiesel and conventional diesel fuel) according to the standards EN 14112 and EN 15751.
It is equipped with two heating blocks each with 4 measuring positions. Each block can be heated individually, i.e., 4 samples can each be mea­sured at two different temperatures or 8 samples at the same tempera­ture. The measurements at the individual measuring positions can be started individually for this.
The 893 Professional Biodiesel Rancimat is controlled by means of the StabNet computer software and a computer that is connected via the USB interface. Up to 4 instruments can be controlled and monitored by each computer, hence allowing a maximum of 32 samples to be analyzed at the same time. The evaluation algorithm of the computer software determines the break point of the Rancimat curve and thus the induction time fully automatically. Besides the induction time, the so-called stabil- ity time, i.e., the time until a defined conductivity change has been achieved, can also be determined. In the case of conductivity changes (stages) which do not have anything to do with the autoxidation, the eval­uation can be suspended for definite time intervals. The results deter­mined can be further processed mathematically.

1 Introduction

Each Rancimat curve can also be evaluated manually. A computer-sup­ported tangential method is available for this, in which you can position the tangents anywhere on your curves. This makes evaluations possible in extreme cases as well.
The results of the determinations are saved in a database together with all method and determination data. Determinations can be searched for, sorted, filtered, exported and printed in the Database program part. Apart from graphically displaying individual and multiple curves, the soft­ware is also capable of conducting recalculations with changed parame­ters and extrapolating the results to a certain temperature.
893 Professional Biodiesel Rancimat
■■■■■■■■
1

1.2 About the documentation

1.2 About the documentation
CAUTION
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe opera­tion of the instrument.

1.2.1 Symbols and conventions

The following symbols and formatting may appear in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
■■■■■■■■■■■■■■■■■■■■■■
Instruction step
Carry out these steps in the sequence shown.
Method Dialog text, parameter in the software
File New Menu or menu item
[Next] Button or key
WARNING
This symbol draws attention to a possible life-threat­ening hazard or risk of injury.
WARNING
This symbol draws attention to a possible hazard due to electrical current.
WARNING
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
WARNING
This symbol draws attention to a possible biological hazard.
■■■■■■■■
2
CAUTION
This symbol draws attention to possible damage to instruments or instrument parts.
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■

1.3 Safety instructions

1.3.1 General notes on safety

WARNING
This instrument may only be operated in accordance with the specifica­tions in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.
Hot reaction vessels
1 Introduction
NOTE
This symbol highlights additional information and tips.
The reaction vessels can become very hot.
Avoid any contact with the hot reaction vessels. Place these in the ves­sel holders provided for cooling down.
Flammable substances
The heating block of the 893 Professional Biodiesel Rancimat can be heated to 159.9 °C.
Flammable substances may ignite at such temperatures.
Adjust the oven's maximum heating temperature to the sample that is to be analyzed.

1.3.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
WARNING
WARNING
893 Professional Biodiesel Rancimat
■■■■■■■■
3
1.3 Safety instructions
■■■■■■■■■■■■■■■■■■■■■■
WARNING
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
WARNING
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Mains voltage
WARNING
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
Protection against electrostatic charges
WARNING
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.
■■■■■■■■
4
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■

1.3.3 Flammable solvents and chemicals

WARNING
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. fume cup-
board).
Keep all sources of flame far from the workplace. Clean up spilled liquids and solids immediately. Follow the safety instructions of the chemical manufacturer.

1.3.4 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
1 Introduction
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
893 Professional Biodiesel Rancimat
■■■■■■■■
5

2.1 Front of the instrument

1
2
3
4
5
6
7
8

2 Overview of the instrument

2.1 Front of the instrument
■■■■■■■■■■■■■■■■■■■■■■
Figure 1 Front 893 Professional Biodiesel Rancimat
Iso-Versinic® tubing (6.1839.000)
1
For connecting the reaction vessel to the measuring vessel. A fluoroelastomer.
Glass measuring vessel (6.1428.030)
3
FEP tubing 250 mm (6.1805.080)
5
For supplying air into the reaction vessel.
Reaction vessels (6.1429.050)
7
Measuring vessel cover (6.0913.130)
2
Contains an integrated conductivity measur­ing cell.
Instrument display
4
Displays the status of the instrument and the individual measuring positions.
Reaction vessel cover (6.2753.107)
6
Start buttons
8
■■■■■■■■
6
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20

2.2 Rear of the instrument

2 Overview of the instrument
Figure 2 Rear 893 Professional Biodiesel Rancimat
Electrode connector
1
For connecting the conductivity measuring cell integrated in the measuring vessel cover.
Power switch
3
For switching the instrument on and off.
I = ON / 0 = OFF.
Power socket
5
For important information on the power connection, see Chapter 3.3.
Drying flask cap (6.1602.145)
7
Cap for the drying flask.
Drying flask (6.1608.050)
9
Drying flask filled with molecular sieve (6.2811.000).
FEP tubing 130 mm (6.1805.010)
11
For connecting the Air out connector to the Air/N2 in connector during normal opera-
tion with the internal air pump.
Air supply connector
2
For connecting the FEP tubing 250 mm.
Fuse holder
4
For replacing fuses (see Chapter 3.3.1, page
19).
Collection tube holder
6
For fastening the optional exhaust air collec­tion tube (6.2757.000).
Bottle holder
8
For fastening the drying flask.
"Air out" connector
10
"Air/N2 in" connector
12
893 Professional Biodiesel Rancimat
■■■■■■■■
7

2.3 Instrument display

893_1
1 2 3 4
9 8 7 6
10
5
■■■■■■■■■■■■■■■■■■■■■■
Type plate
13
Contains specifications concerning supply voltage and serial number.
Dust filter (6.2724.010)
15
"To flask" connector
17
"From flask" connector
19
2.3 Instrument display
Pt100 connector
14
For connecting an external temperature sen­sor.
USB connector
16
For connecting the computer.
FEP tubing 250 mm (6.1805.080)
18
For supplying the air from the internal pump to the drying flask.
FEP tubing 250 mm (6.1805.080)
20
For supplying the air from the drying flask to the reaction vessels.
1
3
■■■■■■■■
8
Figure 3
Measuring vessel cover red
This measuring position is not available for a determination (determination is running or multiple determination has not yet been completed).
Instrument name
The instrument name display corresponds to the configuration in StabNet.
Instrument display
Measuring vessel cover green
2
A determination can be started on this mea­suring position.
Measuring vessel cover gray
4
This measuring position is not available for starting a determination (instrument not connected to computer or no method loa­ded).
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
2 Overview of the instrument
Conductivity display block B
5
Shows the measured conductivity.
Dash (-) is displayed = Conductivity cannot be displayed (no sensor connected or no valid measuring signal).
Temperature display block B
7
Shows the temperature measured on block B (gray background: heater switched off; red background: temperature not stable; green background: temperature stable).
Gas flow display block A
9
Shows the gas flow measured on block A (gray background: gas flow switched off; white background: gas flow switched on).
Gas flow display block B
6
Shows the gas flow measured on block B (gray background: gas flow switched off; white background: gas flow switched on).
Temperature display block A
8
Shows the temperature measured on block A (gray background: heater switched off; red background: temperature not stable; green background: temperature stable).
Conductivity display block A
10
Shows the measured conductivity.
Dash (-) is displayed = Conductivity cannot be displayed (no sensor connected or no valid measuring signal).
893 Professional Biodiesel Rancimat
■■■■■■■■
9

3.1 Setting up the instrument

3 Installation

3.1 Setting up the instrument

3.1.1 Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

3.1.2 Checks

Immediately after receipt, check whether the shipment has arrived com­plete and without damage by comparing it with the delivery note.

3.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.
■■■■■■■■■■■■■■■■■■■■■■
Place the instrument in a location of the laboratory which is suitable for operation and free of vibrations and which provides protection against corrosive atmosphere and contamination by chemicals to the extent possi­ble.
The instrument should be protected against excessive temperature fluctu­ations and direct sunlight.
NOTE
In order to improve accessibility of the measuring positions, the instru­ment can also be placed on the optionally available turning ring (6.2059.000).
■■■■■■■■
10
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
1 2 3 4 5 6 7 8
9
10
11
12

3.2 Mounting accessories

3.2.1 Mounting the internal air supply

The gas in the 893 Professional Biodiesel Rancimat is normally supplied using the internal air pump, which aspirates laboratory air. For air supply and air purification, the following accessories must be mounted on the rear of the 893 Professional Biodiesel Rancimat:
3 Installation
Figure 4 Mounting accessories for the air supply
FEP tubing 130 mm (6.1805.010)
1
"Air/N2 in" connector
3
"From flask" connector
5
"To flask" connector
7
Drying flask cap (6.1602.145)
9
Cap for the drying flask.
Bottle holder
11
For fastening the drying flask.
Mount the accessories for the air supply as follows:
"Air out" connector
2
Dust filter (6.2724.010)
4
FEP tubing 250 mm (6.1805.080)
6
For supplying the air from the drying flask to the reaction vessel.
FEP tubing 250 mm (6.1805.080)
8
For supplying the air from the internal pump to the drying flask.
Drying flask (6.1608.050)
10
Drying flask filled with molecular sieve (6.2811.000).
Filter tube (6.1821.040)
12
1
Mounting the dust filter
Insert the dust filter on the connector marked with Filter on the
rear of the 893 Professional Biodiesel Rancimat.
If the laboratory air is heavily contaminated, a tubing for supply-
ing fresh air can be connected to the dust filter.
893 Professional Biodiesel Rancimat
■■■■■■■■
11
3.2 Mounting accessories
■■■■■■■■■■■■■■■■■■■■■■
NOTE
The dust filter serves for filtering the air aspirated through the air pump and must be replaced at periodic intervals (see Chapter 5.2, page 36).
2
Mounting the drying flask
Drying flask cap (6.1602.145)
1
Bottle (6.1608.050)
3
Figure 5 Drying flask - Individual parts
Filter tube (6.1821.040)
2
Fill the molecular sieve into the drying flask. Screw the filter tube on the bottom side of the drying flask cap
into the opening marked with a dot (on the upper side).
Screw the drying flask cap with mounted filter tube onto the dry-
ing flask and insert it in the flask holder on the rear of the 893 Professional Biodiesel Rancimat.
Screw one end of the FEP tubing 250 mm to the opening on the
drying flask cap marked with a dot.
Screw the other end of the FEP tubing to the From flask connec-
tor on the rear of the 893 Professional Biodiesel Rancimat.
Screw one end of the second FEP tubing 250 mm to the second
opening on the drying flask cap.
Screw the other end of the second FEP tubing to the To flask
connector.
■■■■■■■■
12
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
Figure 6 Drying flask mounted in place
NOTE
The molecular sieve serves to adsorb interfering oxidizing gases as well as water from the aspirated air.
You can regenerate it in the drying oven at +140 - +180 °C for 24 to 48 h (see Chapter 5.3, page 36).
3
Mount the FEP tubing for the air supply
Screw one end of the FEP tubing 130 mm to the Air out connec-
tor.
Screw the other end of the FEP tubing to the Air/N
tor.

3.2.2 Mounting the external air supply

If the laboratory air is heavily contaminated, an external gas supply with synthetic air can be provided.
For this, the corresponding accessories must be mounted on the rear of the 893 Professional Biodiesel Rancimat.
NOTE
If air is supplied externally, the gas flow cannot be regulated in the computer program.
in connec-
2
893 Professional Biodiesel Rancimat
The gas flow must be set manually using the reducing valve and the gas flow display.
■■■■■■■■
13
3.2 Mounting accessories
Mount the accessories for the external air supply as follows:
1
Mounting the FEP tubing
Screw one end of the FEP tubing 130 mm to the Air/N
nector (2-12) on the rear of the 893 Professional Biodiesel Ranci­mat.
Screw the tubing adapter M6/olive (6.1808.020) onto the other
end of the FEP tubing.
2
Connecting the gas supply
Mount the compressed air bottle on the M6/olive tubing adapter
(6.1808.020).
Adjust the air flow by means of the reducing valve at the com-
pressed air bottle.

3.2.3 Assembling the reaction and measuring vessels

The following figure shows in detail how the accessory parts for measur­ing the oxidation stability have to be mounted and connected to one another.
■■■■■■■■■■■■■■■■■■■■■■
in con-
2
■■■■■■■■
14
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
1
2
3
4
5
6 7 8
9
10
11
12
13
14
15
16
17
18
19
20
3 Installation
Figure 7 Equipping the reaction and measuring vessels
FEP tubing 250 mm (6.1805.080)
1
For supplying air into the reaction vessel.
Thread adapter M8/M6 (6.1808.090)
3
Air tube (6.2418.130)
5
Thread adapter M8/M6 connector
7
Foam barrier (6.1451.010)
9
Optional foam barrier.
Tubing adapter M8/olive (6.1808.050)
11
For connecting the Iso-Versinic® tubing to the In opening.
Opening "In"
13
15
893 Professional Biodiesel Rancimat
For supplying the air to the measuring ves­sel.
Connector plug
Iso-Versinic® tubing (6.1839.000)
2
For connecting the reaction vessel to the measuring vessel. A fluoroelastomer.
O-ring (6.1454.040)
4
Tubing connector
6
For connecting the Iso-Versinic® tubing.
Reaction vessel cover (6.2753.107)
8
Reaction vessel (6.1429.050)
10
Opening "Out"
12
For removing the air from the measuring vessel.
Label field
14
For attaching labels (e.g. 6.2250.000 lami­nated labels).
Measuring vessel cover (6.0913.130)
16
Contains an integrated conductivity measur­ing cell.
■■■■■■■■
15
3.2 Mounting accessories
■■■■■■■■■■■■■■■■■■■■■■
PTFE cannula (6.1819.080)
17
For supplying the air to the measuring solu­tion.
Protection ring
19
Proceed as follows to mount the measuring vessel and the reaction vessel:
1
Mounting the measuring vessel cover
Screw the PTFE cannula from above into the In opening of the
Screw the M8/olive tubing adapter into the In opening of the
Pour 50 - 80 mL of deionized water into the measuring vessel (the
Place the measuring vessel cover onto the measuring vessel.
2
Mounting the reaction vessel cover
Place the O-ring (6.1454.040) over the upper end of the air tube. Feed the air tube (6.2418.xx0) into the connector of the reaction
Gently screw the M8/M6 thread adapter into the connector while
Optional: If determinations are being carried out with highly
Blow out the reaction vessel with nitrogen to free it from foreign
Before putting the cover in place, briefly rotate the upper part of
Place the reaction vessel cover on the reaction vessel.
Electrode
18
Glass measuring vessel (6.1428.030)
20
measuring vessel cover.
measuring vessel cover.
exact quantity depends on the application).
vessel cover (6.2753.107) from above.
pressing the air tube against the thread adapter from below. Then fix the air tube onto the reaction vessel cover by firmly tightening the thread adapter.
foaming samples, clamp the foam barrier (6.1451.010) onto the air tube.
substances (e.g. dust or cardboard shreds).
the reaction vessel in your hand. This slightly lubricates the glass and facilitates removal of the covers after the measurement.
■■■■■■■■
16
NOTE
When mounted without foam barrier, the air tube must be in a vertical position in the reaction vessel.
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
Figure 8 Mounting the air tube: correct - incorrect
WARNING
The foam barrier can melt if it projects too deeply into the heating block.
Ensure that the foam barrier is at least 7 cm above the base of the reaction vessel.
Figure 9
Mounting the foam barrier

3.2.4 Inserting vessels / Establishing tubing connections

After you have assembled the reaction and measuring vessels, insert them in the 893 Professional Biodiesel Rancimat and establish the tubing con­nections (see Chapter 2.1, page 6) as follows:
1
Inserting the measuring vessel
Fill distilled water into the measuring vessel. Place the measuring vessel cover onto the measuring vessel. Insert the measuring vessel into the openings provided on the 893
Professional Biodiesel Rancimat. While doing so, carefully guide the connector plug into the electrode connector.
Connect the black Iso-Versinic
adapter of the measuring vessel cover.
893 Professional Biodiesel Rancimat
®
tubing to the M8/olive tubing
■■■■■■■■
17
3.2 Mounting accessories
■■■■■■■■■■■■■■■■■■■■■■
2
Mounting the tubing for the air supply
Screw the FEP tubing 250 mm to the air supply connections of
the 893 Professional Biodiesel Rancimat.
3
Inserting the reaction vessel
Fill the reaction vessel with the sample. After the required reaction temperature has been reached, insert
the reaction vessel with the mounted reaction vessel cover in the openings provided on the 893 Professional Biodiesel Rancimat.
4
Establishing the tubing connections
Connect the Iso-Versinic
®
tubing that is connected to the measur­ing vessel cover to the tubing connector of the reaction vessel cover.
Screw the FEP tubing 250 mm which is connected to the M8/olive
tubing adapter of the 893 Professional Biodiesel Rancimat to the M8/M6 thread adapter of the reaction vessel cover.

3.2.5 Mounting the exhaust air collection tube

The optional exhaust air collection tube (6.2757.000) [Link target not
found in publication context!] can be mounted on the 893 Professional
Biodiesel Rancimat for targeted removal of the exhaust air.
NOTE
A total of 8 pieces of Iso-Versinic® tubing (6.1839.000) are required in addition to the exhaust air collection tube.
Proceed as follows to mount the collection tube:
1
Mounting the exhaust air collection tube
Insert the exhaust air collection tube with both support pins into
the collection tube holders on the 893 Professional Biodiesel Ran­cimat in such a way that the connector to the exhaust air removal is located to the rear.
2
Connecting the measuring vessels
Screw the M8/olive tubing adapter into the Out opening of the
measuring vessel cover.
Connect one end of the Iso-Versinic
ing adapter.
®
tubing to the M8/olive tub-
■■■■■■■■
18
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
Insert the other end of the Iso-Versinic
Seal the unused openings on the collection tube with the
3
Connecting the exhaust air collection tube
Connect a suitable tubing to the connector of the exhaust air col-

3.3 Power socket

There is a risk of fire if the instrument is operated with an incorrect mains fuse!
Follow the regulations below for the mains connection.
®
tubing into the corre-
sponding opening on the collection tube.
enclosed stoppers.
lection tube and connect it to an active suction device (e.g. water jet pump).
WARNING

3.3.1 Replacing fuses

The fuse holder of the 893 Professional Biodiesel Rancimat contains two fuses:
Two slow-acting 4 A fuses
Ensure that the instrument is never put into operation with fuses of another type, otherwise there is a risk of fire!
Proceed as follows to replace defective fuses:
1
Pulling out the power supply cable
Pull the power supply cable out of the power socket of the 893
2
Removing the fuse holder
Release the fuse holder located on the rear of the instrument
WARNING
Professional Biodiesel Rancimat.
above the power socket by pressing the catch spring and pull out the holder completely.
893 Professional Biodiesel Rancimat
■■■■■■■■
19
3.3 Power socket
3
Replacing the fuses
Carefully remove the defective fuses from the fuse holder and
replace them with two new slow-acting 4 A fuses:
4
Inserting the fuse holder
Push the fuse holder back into the instrument until it latches into
place.

3.3.2 Power supply cable and power socket

Power supply cable
The instrument is supplied with a three-core power supply cable and a plug with protective grounding. If another plug has to be mounted, then the yellow/green conductor (IEC standard) must be connected to the pro­tective ground (protection class 1).
WARNING
■■■■■■■■■■■■■■■■■■■■■■
Any interruption to the grounding within or outside the instrument can make it dangerous!
Power socket
Plug the power supply cable into the power socket (2-5) of the 893 Pro­fessional Biodiesel Rancimat.

3.3.3 Switching the instrument on and off

The 893 Professional Biodiesel Rancimat is switched on and off using the power switch (2-3). The instrument display is switched on when the instrument is switched on.
Figure 10
The instrument 893 Professional Biodiesel Rancimat has been recognized by the StabNet computer program and the instrument name entered has been transmitted.
Instrument display with instrument name
■■■■■■■■
20
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
Figure 11 Instrument display with "no connection" symbol
The USB connection between the instrument 893 Professional Biodiesel Rancimat and the computer has been interrupted and the corresponding symbol is displayed.
Figure 12
Instrument display without instrument name and symbol
The instrument 893 Professional Biodiesel Rancimat is connected to the computer but the StabNet computer program has been closed.
Figure 13
Instrument display with serial number
The instrument 893 Professional Biodiesel Rancimat has been started, but the StabNet computer program was not previously started.
893 Professional Biodiesel Rancimat
■■■■■■■■
21

3.4 Connecting a computer

■■■■■■■■■■■■■■■■■■■■■■
3.4 Connecting a computer
3.4.1 Connecting the 893 Professional Biodiesel Rancimat and the com­puter
NOTE
The StabNet computer program has to be installed before you can connect the instrument to the computer.
The StabNet computer program allows you to control up to 4 instru­ments.
Connect and set up the 893 Professional Biodiesel Rancimat as follows:
Establish a connection between the USB interface (2-16) of the 893
1
Professional Biodiesel Rancimat and the required USB interface on the computer using the USB cable (6.2151.130).
Start the StabNet computer program.
2
Switch on the 893 Professional Biodiesel Rancimat using the power
3
switch.
Wait for the instrument 893 Professional Biodiesel Rancimat to be
4
detected and installed.
Enter the instrument information into the dialog fields of the 893
5
Professional Biodiesel Rancimat configuration.
■■■■■■■■
22
NOTE
You will find detailed information regarding the StabNet com­puter program in the Tutorial.
The instrument name entered in the configuration must appear on the instrument display (3-3).
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
4 Operation

4.1 Rancimat method

Biodiesel, also referred to as RME (rapeseed methyl ester) or FAME (fatty acid methyl esters), is usually obtained from oil seeds and used in the automotive industry either in pure form or as a mixture with conventional diesel fuel. As is the case with the natural oils, the fatty acid methyl esters obtained from them also exhibit a relatively low level of storage stability, as they are oxidized by atmospheric oxygen. These oxidation processes, which progress slowly at ambient temperatures, are referred to as autoxi- dation. They start with radical reactions on unsaturated fatty acids and undergo a process involving multiple stages resulting in diverse decompo­sition products, in particular peroxides as primary oxidation products and alcohols, aldehydes and carboxylic acids as secondary oxidation products.
In the Rancimat method, the sample is exposed to an air flow at a con­stant temperature of 110 °C. Highly volatile, secondary oxidation products (for the most part formic acid and acetic acid) are transferred into the measuring vessel with the air flow, where they are absorbed in the mea­suring solution (distilled water). Here the conductivity is continuously reg­istered. The organic acids can thus be detected by an increase in the con­ductivity. The time until these secondary reaction products occur is refer­red to as the induction time or induction period, which is a good charac­teristic for the oxidation stability.

4 Operation

The Rancimat method has been developed as an automated version of the extremely complex AOM (active oxygen method) for determining the induction time of fats and oils. This method has established itself over the course of time and has also been applied for derivatives of oils and fats. It has become an accepted method for biodiesel and biodiesel blends in various national and international standards, e.g. EN 15751, EN 14112 and ASTM 6751.
893 Professional Biodiesel Rancimat
■■■■■■■■
23

4.2 Calibration functions

■■■■■■■■■■■■■■■■■■■■■■
Figure 14 Measuring arrangement (schematic representation)
Reaction vessel
1
Heating block
3
Conductivity measuring cell
5
4.2 Calibration functions

4.2.1 Determining the cell constant

The cell constant of the conductivity sensor (6.0913.130) is normally at
1.1 ± 0.1 cm-1. This accuracy is sufficient for an induction time determi­nation since only the shape of the curve is evaluated. For the determina­tion of the stability time, however, the conductivity change is measured in absolute terms. As a rule, the time until an increase in conductivity of 50 µS/cm is determined. The cell constant for the conductivity electrode being used has to be calibrated in order to be able to measure the con­ductivity correctly.
Sample
2
Measuring vessel
4
Measuring solution
6
■■■■■■■■
24
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
The cell constants can be either entered manually or determined automat­ically by means of a defined standard solution, e.g. the conductivity stand­ard 100 µS/cm (6.2324.010).
NOTE
You will find detailed information regarding the StabNet computer program in the Tutorial.

4.2.2 Determining the temperature correction

Temperature correction indicates the deviation of the current sample temperature from the temperature of the heating block and forms part of the method as a parameter.
It can be determined automatically with the calibrated, external tempera­ture sensor.
4 Operation
893 Professional Biodiesel Rancimat
■■■■■■■■
25
4.2 Calibration functions
■■■■■■■■■■■■■■■■■■■■■■
■■■■■■■■
26
Figure 15 Assembling the reaction vessel for determining the tempera-
FEP tubing 250 mm (6.1805.080)
1
For supplying air into the reaction vessel.
Thread adapter M8/M6 (6.1808.090)
3
Temperature sensor opening
5
For inserting the temperature sensor.
Reaction vessel cover (6.2753.107)
7
Spacer (6.2042.040)
9
ture correction
Pt100 temperature sensor
2
(6.1111.020)
O-ring (6.1454.040)
4
Connector
6
For connecting the thread adapter M8/M6.
Air tube (6.2418.130)
8
Reaction vessel (6.1429.050)
10
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
Preparing the determination of the temperature correction
The figure shows in detail how the accessory parts are to be assembled for the determination of the temperature correction. Proceed as follows:
1
Preparing the reaction vessel cover
Mount the air tube on the reaction vessel cover. Clamp the first spacer at a distance of approx. 12 cm from the
lower end onto the air tube.
Clamp the second spacer at a distance of approx. 5 cm from the
lower end onto the air tube.
Insert the temperature sensor from above into the temperature
sensor opening and fasten it in the corresponding openings of the spacers.
2
Preparing the reaction vessel
Fill the reaction vessel with 6 g silicone oil (6.2326.000). Place the reaction vessel cover with the temperature sensor on
the reaction vessel.
Push the temperature sensor all the way down (the sensor must
touch the vessel base).
3
Inserting and connecting the reaction vessel
Insert the reaction vessel with the mounted reaction vessel cover
in the measuring position 2 or 3 of the required heating block.
Screw one end of the FEP tubing 250 mm to the M8/M6 thread
adapter of the reaction vessel cover.
Screw the other end of the FEP tubing to the corresponding con-
nector of the 893 Professional Biodiesel Rancimat.
Connect the temperature sensor to the Pt100 connector (2-14)
on the rear of the 893 Professional Biodiesel Rancimat.
893 Professional Biodiesel Rancimat
NOTE
You will find detailed information regarding the StabNet computer program in the Tutorial.
■■■■■■■■
27

4.3 Determinations

4.3 Determinations

4.3.1 Preparing the instrument and the accessories

The cleanliness of instrument and accessory parts is an indispensable pre­requisite for reliable, reproducible and correct analysis results. Even the slightest contamination could catalytically accelerate the oxidative decomposition and lead to completely incorrect results. Therefore, always observe the instructions for use of measuring and reaction vessels in this chapter.
Check and prepare the instruments and vessels as follows:
1
Checking the positions for reaction vessels
Check whether the positions in the heating block are clean and
empty. Blow out contamination and dust in the positions with nitrogen. If the instrument is not used, always seal the corresponding posi­tions with the stoppers.
2
Filling the measuring vessels
■■■■■■■■■■■■■■■■■■■■■■
NOTE
Only use measuring vessels and accessories which are abso­lutely clean and in a flawless condition.
Fill the cleaned measuring vessels with 60 mL distilled water
each. For analysis times of more than 24 h approx. 7 mL more distilled water must be added per day to compensate for the evaporation loss, so that the electrodes remain safely immersed.
3
Inserting the measuring vessels
Place the clean measuring vessel covers equipped with a PTFE
cannula on the measuring vessels.
Insert the measuring vessels with the measuring vessel covers into
the openings provided for this on the 893 Professional Biodiesel Rancimat while carefully guiding the connector plugs of the cover into the electrode connectors.
■■■■■■■■
28
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
4
Weighing in samples
NOTE
Use new reaction vessels and air tubes for each measure­ment. Blow out the reaction vessels with nitrogen before use so that any adhering particles are removed.
Weigh in 7.5 g of the samples in each reaction vessel.
5
Mounting the accessories
Hold the upper edge of the reaction vessel in your hand (e.g. in
the recess between your thumb and index finger) and rotate the glass once. This serves to cover the degreased glasses with a light fat film so that the vessel covers can be removed more easily after the deter­mination.
Insert an air tube into the connector of the reaction vessel cover,
fix it with the O-ring and fasten it by screwing in the M8/M6 thread adapter.
Place the reaction vessel cover on the reaction vessel. Turn the
cover in such a way that the air tube is as close as possible to the vessel wall.
Connect the Iso-Versinic
®
tubing to the tubing connector of the
reaction vessel cover.
Place the prepared reaction vessel in the vessel holder.
4 Operation

4.3.2 Preparing the determination

1
Selecting the method (StabNet)
In the Workplace program part, click on the symbol within the
block A area and select a Method for block A.
If required, also select a method for block B.
893 Professional Biodiesel Rancimat
■■■■■■■■
29
4.3 Determinations
■■■■■■■■■■■■■■■■■■■■■■
NOTE
Different methods with different temperatures can be selected for block A and block B.
The gas flow for the two blocks A and B cannot be switched on and off individually for each block but only collectively. If methods with different gas flows are loaded on the two blocks, then the value from the block in which the gas flow is switched on is used.
2
Starting the heater (StabNet)
In the Workplace program part, click on the [Start] button for
Heater within the block A area.
If required, also switch on the heater for Block B.
The color of the temperature display on the 893 Professional Biodie­sel Rancimat changes to red during the heating phase.
The color of the temperature display on the 893 Professional Biodie­sel Rancimat changes to green once the setpoint temperature has been reached.
Heating period to 120 °C: approx. 45 min.
NOTE
If you wish to switch off the heating, click on the [Stop] button.
3
Enter sample identification (StabNet)
Enter the sample identifications Ident and Info 1 - 3 for all sam-
ple positions used.
The entries for Ident and Info 1 - 3 can be selected from the Text
templates.
NOTE
The Info 2 and Info 3 info fields can be activated in the sub­window Properties - Sample data in the Method program part.
■■■■■■■■
30
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
4
Connecting and inserting reaction vessels
NOTE
The temperature defined in the method has to be reached before you insert the reaction vessels, i.e., the temperature display has to be green.
Seal the unused positions with stoppers or empty reaction vessels
for protection against impurities.
Connect the Iso-Versinic
®
tubing which is connected to the reac­tion vessel covers to the M8 tubing adapter/olive of the measur­ing vessel covers.
Screw the pieces of FEP tubing 250 mm to the M8/M6 thread
adapters of the reaction vessel covers and the air supply connec­tors of the 893 Professional Biodiesel Rancimat.
Insert the prepared reaction vessels into the recesses of the heat-
ing block.
Press the start button to start the data recording immediately
after inserting each single reaction vessel.
4 Operation

4.3.3 Cleaning the accessories

1
Cleaning the measuring vessels and accessories
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Clean used measuring vessels after pouring off the measuring
solution with acetone or 2-propanol or in the dishwasher. Pre-clean with dishwashing detergent in the case of severe con­tamination.
Thoroughly rinse with distilled water. Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and thoroughly rinse with distilled water. Pre-clean with dishwashing detergent in the case of severe con­tamination. Remove the protective ring to facilitate cleaning of the electrodes.
Clean used measuring vessels after pouring off the measuring
solution with acetone or 2-propanol or in the dishwasher. Pre-clean with dishwashing detergent in the case of severe con­tamination.
Thoroughly rinse with distilled water.
893 Professional Biodiesel Rancimat
■■■■■■■■
31
4.3 Determinations
■■■■■■■■■■■■■■■■■■■■■■
Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and thoroughly rinse with distilled water. Pre-clean with dishwashing detergent in the case of severe con­tamination. Remove the protective ring to facilitate cleaning of the electrodes.
2
Cleaning the reaction vessels and accessories
Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.
Clean the reaction vessel covers with acetone or 2-propanol or
in the dishwasher and rinse with distilled water. Pre-clean with dishwashing detergent in the case of severe con­tamination.
Then heat the reaction vessel covers in the drying oven for 2
hours at 80 °C.
NOTE
Replace the reaction vessel covers if they no longer sit tightly on the reaction vessel or the material is brittle.
3
Cleaning the tubing
Clean the Iso-Versinic
the dishwasher and rinse with distilled water. Pre-clean with dishwashing detergent in the case of severe con­tamination.
Then heat the Iso-Versinic
hours at 80 °C.

4.3.4 Cleaning the oil separator

The oil separator is available as an optional accessory.
®
tubing with acetone or 2-propanol or in
®
tubing in the drying oven for 2
■■■■■■■■
32
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
4 Operation
Figure 16 Oil separator (6.2753.200), disassembled
Body
1
Body with the IN and OUT connectors.
1
2
3
Base
2
Base with O-ring (6.1454.050) used for seal­ing the oil separator.
Disassembling the oil separator
Remove the base.
NOTE
Remove the base carefully.
Do not use sharp or pointed tools.
Cleaning the oil separator bodies
Clean the oil separator bodies with acetone or 2-propanol or in
the dishwasher and rinse with distilled water. Pre-clean with dishwashing detergent in the case of severe con­tamination.
Then heat the oil separator bodies in the drying oven for 2
hours at 80 °C.
Cleaning the oil separator bases
Clean the oil separator bases with 2-propanol or in the dish-
washer and rinse with distilled water. Pre-clean with dishwashing detergent in the case of severe con­tamination.
Then heat the oil separator bases in the drying oven for 2
hours at 80 °C.
Mount new O-rings where necessary.
893 Professional Biodiesel Rancimat
■■■■■■■■
33
4.3 Determinations
■■■■■■■■■■■■■■■■■■■■■■
NOTE
Do not allow the oil separator bases to come into contact with
acetone for extended periods of time.
NOTE
Replace the O-rings if they no longer sit tightly on the body or if the material is brittle.
4
Assembling the oil separator
Press the base onto the body by hand until it sits flush.
■■■■■■■■
34
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance

5.1 General notes

5.1.1 Care

The 893 Professional Biodiesel Rancimat requires appropriate care. Excess contamination of the instrument may result in functional disruptions and a reduction in the service life of the otherwise sturdy mechanics and elec­tronics.
Spilled chemicals and solvents should be removed immediately. Above all, the plug connections on the rear of the instrument (in particular the power socket) should be protected from contamination.
CAUTION

5 Operation and maintenance

Although this is largely prevented by design measures, the power plug should be unplugged immediately if aggressive media have found their way into the interior of the instrument to prevent serious damage to the instrument electronics. In such cases, Metrohm Service must be informed.

5.1.2 Maintenance by Metrohm Service

Maintenance of the 893 Professional Biodiesel Rancimat is best carried out as part of annual service, which is performed by specialist personnel from Metrohm. If you are frequently working with caustic and corrosive chemi­cals, we recommend a shorter maintenance interval.
Metrohm Service offers every form of technical advice for maintenance and service of all Metrohm instruments.
893 Professional Biodiesel Rancimat
■■■■■■■■
35

5.2 Replacing the dust filter

5.2 Replacing the dust filter
The dust filter (2-15) is mounted on the opening marked with Filter on the rear of the instrument and serves for filtration of the air aspirated through the air pump. It must be checked at periodic intervals and replaced in the case of more intense contamination (6.2724.010).
Figure 17 Dust filter - conditions
■■■■■■■■■■■■■■■■■■■■■■
Dust filter new
1
New dust filters are white on the intake side.
Dust filter full
3
Full dust filters have a dark to black color on the intake side and must be exchanged.
Dust filter used
2
Used dust filters are discolored on the intake side.

5.3 Regenerating or replacing the molecular sieve

NOTE
Regenerate the molecular sieve regularly.
The intervals at which you have to regenerate the molecular sieve depends on the humidity in the laboratory and also on the frequency of use of the instrument.
The molecular sieve filled in the drying flask (2-9) serves to adsorb disrup­tive oxidizing gases as well as of the water from the aspirated air.
■■■■■■■■
36
You can regenerate the molecular sieve in the drying oven at approx. +140 - +180 °C for 24 to 48 h. You can order a new molecular sieve under the order number 6.2811.000.
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
CAUTION
Do not fill the hot molecular sieve directly into the drying flask after regeneration, as otherwise the plastic filter on the filter tube will melt.
Wait until the molecular sieve has cooled down before filling.
5 Operation and maintenance

5.4 Quality management and qualification with Metrohm

Quality management
Metrohm offers you comprehensive support in implementing quality man­agement measures for instruments and software.
Qualification
Please contact your local Metrohm representative for support in qualifica­tion of instruments and software. The Installation Qualification (IQ) and Operational Qualification (OQ) are offered by Metrohm represen­tatives as a service. They are carried out by trained employees using stan­dardized qualification documents and in accordance with the currently applicable requirements of the regulated industry.
Maintenance
The electronic and mechanical functional groups of Metrohm instruments can and should be checked by specialist personnel from Metrohm as part of a regular preventive maintenance schedule. Please ask your local Metrohm representative regarding the precise terms and conditions involved in concluding a corresponding maintenance agreement.
For detailed information on this topic, please visit www.metrohm.com.
893 Professional Biodiesel Rancimat
■■■■■■■■
37

6.1 Problems

6 Troubleshooting

6.1 Problems
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
The pump is louder than normal.
No air flow can be detected in the reac­tion vessel (it does not bubble), although the pump is running.
The air flow is blocked somewhere before or after the pump.
Check dust filter and replace it if necessary. Check filter tube on the drying flask cap for
blockages and, if necessary, gently tap on it to remove them.
Examine the FEP tubing with regard to
blockages and replace it if necessary.
Extra air is aspirated from elsewhere other than the dust filter. There is a leak somewhere in the system
Check the FEP tubing for cracks, kinks, etc.
and fasten it tightly. Replace it if necessary.
Place the drying flask cap correctly on the
drying flask and screw tight.
before the pump.
The air supply is blocked. Remove the FEP tubing from the thread
adapter. Here, a slight air flow must be perceptible. If this is not the case, please contact Metrohm Service.
Check FEP tubing for blockages. If neces-
sary, clean or replace.
Check the thread adapter and the air tube
on the reaction vessel cover for blockages. If necessary, clean or replace.
■■■■■■■■
38
The FEP tubing for the air supply is defective.
The FEP tubing for the air supply is not connected correctly.
The air tube does not immerse in the sample.
Check the FEP tubing for cracks, kinks etc. If necessary, replace.
Tighten the FEP tubing on both sides.
Press the reaction vessel cover all the way
down.
Use more sample.
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
Problem Cause Remedy
6 Troubleshooting
No air flow can be detected in the mea­suring vessel (it does not bubble), although an air flow can be discerned in the reaction vessel.
The connection is blocked. Check the tubing connector on the reac-
tion vessel cover for blockage and, if neces­sary, clean.
Check the Iso-Versinic
®
tubing for blockage
and, if necessary, clean it.
Check the tubing adapter and the PTFE
cannula on the measuring vessel cover for blockage and, if necessary, clean.
The connection is leaking. Check the Iso-Versinic® tubing for leakages
and, if necessary, replace.
The reaction vessel cover does not sit correctly or tightly enough.
If the reaction vessel cover is oblique or not
completely mounted, press it all the way down.
If the reaction vessel cover is loose on the
reaction vessel despite correct assembly, the cover has to be replaced.
The connection is wrongly connected.
Make sure that the PTFE cannula for air
supply is connected to the In opening of the measuring vessel cover.
Make sure that the Iso-Versinic
®
tubing is connected to the tubing adapter that is mounted on the In opening.
Make sure that the reaction vessel is con-
nected to the measuring vessel that belongs to the corresponding measuring position.
The induction times are not reproducible
The reaction vessels used are not clean.
for multiple determi­nations.
The reaction vessels used are scratched on the inside.
The reaction vessel cover does not sit correctly or tightly enough.
893 Professional Biodiesel Rancimat
Clean the reaction vessels of particles (dust,
cardboard, etc.) with nitrogen before weighing in the sample.
Only use new, unused reaction vessels.
Only use new, unused reaction vessels.
If the reaction vessel cover is oblique or not
completely mounted, press it all the way down.
If the reaction vessel cover is loose on the
reaction vessel despite correct assembly, the cover has to be replaced.
■■■■■■■■
39
6.1 Problems
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
The stability times are not reproducible for multiple determi­nations.
The connection to the measuring vessel is not mounted correctly.
The temperature in differ­ent positions of a heating block differs, as sample has burnt on at one or more places in the recess of the heating block.
The sample is not homoge­nous.
The cell constant was not determined or does not correspond to the value entered.
The conductivity measuring cell is contaminated.
Ensure that no air can escape through leaks when transferring from the reaction vessel to the measuring vessel.
If necessary, carefully remove contamination from the cold heating block.
Homogenize the sample.
Determine the cell constant. Make sure that the assignment of the con-
ductivity sensors is correct, so that the determined cell constant actually corre­sponds to the measuring cell used.
Make sure that the measuring cell is not
contaminated. Clean if necessary.
Check the measuring cell and, if necessary, clean.
The induction time is longer/shorter than expected.
The stability time is longer/shorter than expected.
See also: The induction times are not reproducible for multiple determina­tions.
The temperature is not selected correctly.
See also: The induction times are not reproducible for multiple determina­tions.
The conductivity change is not correctly defined.
See also: The stability times are not reproducible for multiple determinations.
Make sure that the correct method for the
determination has been selected.
Check whether the Sample temperature
and the Temperature correction are indicated correctly in the method.
Make sure that the value defined for the con­ductivity change in the method is correct.
■■■■■■■■
40
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
Problem Cause Remedy
See also: The induction time is longer/shorter than expected.
6 Troubleshooting
The measurement curves are extremely noisy.
The air supply for the mea­suring solution is directed to the conductivity measur­ing cell.
Gas bubbles adhere to the conductivity measuring cell during measurement.
During the measurement, sample evaporates in the reaction vessel and con­denses in the measuring vessel. This results in con­tamination of the conduc­tivity measuring cell, which in turn promotes adher­ence of gas bubbles.
Loosen the tubing adapter on the measuring vessel cover and turn the PTFE cannula such that the air is no longer directed to the elec­trode and fix it in this position.
Make sure that the measuring cell is clean
and free of fat. If necessary, clean thor­oughly.
In some cases, ultrapure water contains a
large proportion of dissolved air. In this case, degas the ultrapure water before the measurement for 5 to 10 min in a vacuum.
Keep the measuring time as short as possi-
ble, by about 4 to 6 h. The induction time can be reduced by about half by increasing the temperature by 10 °C.
Reduce the temperature to such an extent
that sample evaporation is minimized or eliminated. However, this can substantially extend the measuring time. The induction time approximately doubles by reducing the temperature by 10 °C.
Use an oil separator (6.2753.200).
The curve shows a step which means that the induction time is no longer determined cor-
Side reactions occur at the start or during the mea­surement. These reactions cause the conductivity in the measuring cell to rise.
rectly.
893 Professional Biodiesel Rancimat
Use the Evaluation suppression in the
method.
In addition to Endpoint(s) activate also
Conductivity (e.g. 200 µS/cm) as stop cri- terion in the method and select the Stop once all the criteria have been fulfilled
option. The evaluation parameters can be optimized on the basis of this curve, or the curve can be evaluated manually.
Increase the Evaluation sensitivity
method parameter.
■■■■■■■■
41
6.1 Problems
Problem Cause Remedy
■■■■■■■■■■■■■■■■■■■■■■
The curve shows a step at the start of the measurement, which has not occur­red in previous measurements.
The induction time is not evaluated automatically, although a signifi­cant break point can be noticed in the curve.
The reaction vessel cover and/or the tubing still con­tain residues from previous measurements. These resi­dues are then transferred to the measuring vessel with the flow of warm air during a new measure­ment.
The Evaluate induction time option is deactivated in the method.
The Evaluation suppres- sion option preventing the evaluation of the curve in the corresponding time period is defined in the method.
Automatic detection of the induction time is not yet possible.
Thoroughly clean the reaction vessel cover
and the Iso-Versinic® tubing.
Replace the reaction vessel cover and the
Iso-Versinic® tubing from time to time.
If an oil separator has been used, thor-
oughly clean it.
Activate the evaluation of the induction time in the method.
Deactivate the corresponding option in the method.
Keep the determination running until the induction time is automatically found.
The measurement aborts without an endpoint being found.
The curve progression is too flat, with the result that automatic detection of the induction time is not possible.
An inappropriately high value was used in the method for the Evalua- tion sensitivity. This value makes a curve evaluation impossible.
The measurement has been stopped manually.
A time or conductivity defined as a stop criterion in the method is reached before the endpoint.
Reduce the Evaluation sensitivity
method parameter.
Evaluate the curves manually using tan-
gents.
Enter a lower value (e.g. 1.0) for the Evalua- tion sensitivity option in the method.
Keep the measurement running until the end­point is automatically found.
Activate the Stop once all the criteria
have been fulfilled option in the method.
Increase the value for time or conductivity. Deactivate time or conductivity as stop cri-
terion.
■■■■■■■■
42
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications

7.1 General data

7 Technical specifications

Number of instru­ments that can be connected
Number of sam­ples per instru­ment
Sample amount Several grams / several milliliters
1 - 4
1 - 8 (4 each per heating block)

7.2 Temperature control and measurement

Sample tempera­ture setting range
Temperature cor­rection
Temperature mea­suring range
Resolution 0.1 °C
50 - 150 °C
±0 - 9.9 °C (Can be entered manually or determined automatically using the exter­nal temperature sensor.)
0 - 300 °C
Maximum devia­tion of the heating block temperature from the set value (50 - 220 °C)
Reproducibility of the set tempera­ture
Max. temperature difference between different measuring posi­tions per block
Temperature fluc­tuations
±0.3 °C
typ. ±0.2 °C
typ. < 0.3 °C
typ. < 0.1 °C (With operating temperature attained, reaction vessels inserted and identically filled and 20 L/h air throughput.)
893 Professional Biodiesel Rancimat
■■■■■■■■
43

7.3 External temperature sensor

■■■■■■■■■■■■■■■■■■■■■■
Shutdown temper­ature
Heating time for the instrument
External tempera­ture of the instru­ment
180 ± 7 °C (Resetting and troubleshooting carried out by Metrohm Service.)
approx. 45 min (from 20 °C to 120 °C)
< 50 °C (at operating temperature 150 °C)
7.3 External temperature sensor
Sensor
4-pin for sensor Pt100 (6.1111.020)

7.4 Conductivity measurement

Sensor
Electrode Conductivity measuring cell with 2 stainless steel electrodes
Measuring princi­ple
Conductivity measuring cell, integrated in the measuring vessel cover (6.0913.130)
AC current measurement with 1 kHz frequency and approx. 1.0 V amplitude (peak to peak)
Cell constant 1.0 - 1.2
(The exact value can be entered manually or determined automatically.)
Measuring range 0 - 400 µS/cm (at c = 1)
Resolution 0.1 µS/cm
Display range 0 - 999 µS/cm
Maximum devia­tion from mea­sured value
±(0.5 µS/cm + 1% of the measured value)
■■■■■■■■
44
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■

7.5 Gas flow regulation

Pump Membrane pump (brushless motor)
7 Technical specifications
Volumetric flow range
Maximum devia­tion from the set range
1 - 25 L/h at 25 °C and 1,013 mbar
±(0.25 L/h + 5% of the measured value)

7.6 USB interface

USB connector
USB plug type B

7.7 Power socket

Nominal voltage range
Frequency 50 and 60 Hz (autosensing)
Power consump­tion
Fuse Diameter 5 mm, length 20 mm
100 - 120 V and 220 - 240 V ± 10% (autosensing)
450 VA
4.0 ATH (slow-acting)
max

7.8 Safety specifications

This instrument fulfills the following electrical safety requirements:
CE marking in accordance with the EU directives:
2006/95/EC (Low Voltage Directive, LVD) 2004/108/EC (EMC Directive, EMC)
Design and testing According to EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1 protection
class I, EN/IEC 61010-2-010, EN/IEC 60529, degree of protection IP20.
Safety instructions This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.
893 Professional Biodiesel Rancimat
■■■■■■■■
45

7.9 Electromagnetic compatibility (EMC)

7.9 Electromagnetic compatibility (EMC)
Emission Standards fulfilled:
EN/IEC 61326-1 EN/IEC 61000-6-3 EN 55011 / CISPR 11 EN/IEC 61000-3-2 EN/IEC 61000-3-3
Immunity Standards fulfilled:
EN/IEC 61326-1 EN/IEC 61000-6-2 EN/IEC 61000-4-2 EN/IEC 61000-4-3 EN/IEC 61000-4-4 EN/IEC 61000-4-5 EN/IEC 61000-4-6 EN/IEC 61000-4-11 EN/IEC 61000-4-14 EN/IEC 61000-4-28
■■■■■■■■■■■■■■■■■■■■■■

7.10 Ambient temperature

Nominal function range
Storage –20 - +70 °C
Transport –40 - +70 °C
+5 - +45 °C (at 20 - 80% humidity)

7.11 Dimensions/Material

Width
Height 277 mm (without accessories)
Depth 462 mm
Weight 16.1 kg (without accessories)
Cover material Baydur®110 FR-6 with flame retardation for fire class UL94VO, CFC-
Base material Steel sheet, coated
383 mm
free
■■■■■■■■
46
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
8 Conformity and warranty

8.1 Quality Management Principles

Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number 10872-02, issued by SQS (Swiss Association for Quality and Management Systems). Internal and external audits are carried out periodically to assure that the standards defined by Metrohm’s QM Manual are maintained.
The steps involved in the design, manufacture and servicing of instruments are fully documented and the resulting reports are archived for ten years. The development of software for PCs and instruments is also duly docu­mented and the documents and source codes are archived. Both remain the possession of Metrohm. A non-disclosure agreement may be asked to be provided by those requiring access to them.
The implementation of the ISO 9001:2000 quality management system is described in Metrohm’s QM Manual, which comprises detailed instruc­tions on the following fields of activity:

8 Conformity and warranty

Instrument development
The organization of the instrument design, its planning and the intermedi­ate controls are fully documented and traceable. Laboratory testing accompanies all phases of instrument development.
Software development
Software development occurs in terms of the software life cycle. Tests are performed to detect programming errors and to assess the program’s functionality in a laboratory environment.
Components
All components used in the Metrohm instruments have to satisfy the qual­ity standards that are defined and implemented for our products. Suppli­ers of components are audited by Metrohm as the need arises.
Manufacture
The measures put into practice in the production of our instruments guar­antee a constant quality standard. Production planning and manufacturing procedures, maintenance of production means and testing of compo­nents, intermediate and finished products are prescribed.
Customer support and service
Customer support involves all phases of instrument acquisition and use by the customer, i.e. consulting to define the adequate equipment for the analytical problem at hand, delivery of the equipment, user manuals, train-
893 Professional Biodiesel Rancimat
■■■■■■■■
47

8.2 Warranty (guarantee)

ing, after-sales service and processing of customer complaints. The Metrohm service organization is equipped to support customers in imple­menting standards such as GLP, GMP, ISO 900X, in performing Opera­tional Qualification and Performance Verification of the system compo­nents or in carrying out the System Validation for the quantitative determi­nation of a substance in a given matrix.
8.2 Warranty (guarantee)
Metrohm guarantees that the deliveries and services it provides are free of defects in materials, design or manufacturing.
The general warranty period is 36 months (exclusions below) from the date of delivery, or 18 months in the event of continuous operation. The warranty remains valid on the condition that the servicing is provided by a service organization authorized by Metrohm at defined intervals and with a defined scope.
The warranty period for anion suppressors of the type "MSM" is 120 months from the date of delivery or 60 months in the case of continuous operation.
■■■■■■■■■■■■■■■■■■■■■■
The warranty period for IC separation columns is 90 days after start-up.
For third-party components that are recognizable as such, the manufac­turer's warranty regulations apply.
For instruments sold under the Metrohm NIRSystems brand, a full 16­month warranty is applicable. In the event of continuous operation, the warranty period is reduced by half.
Consumables and materials with limited storage life and glass breakage in the case of electrodes or other glass parts are excluded from the warranty.
Warranty claims cannot be asserted if the ordering party has failed to meet its payment obligations according to schedule.
During the warranty period, Metrohm undertakes either to replace free of charge or to credit the purchaser for any modules or components that can be shown to be faulty. Any transport or customs fees that may apply are the responsibility of the ordering party.
The precondition for this is that the ordering party has to specify the arti­cle number, the article designation, an adequate error description, the delivery date and (if applicable) the serial number or chip data in the Sup­port Tracker. Metrohm then decides whether a replacement or a credit note is to be issued or whether the faulty part has to be returned using the Return Material Authorization (RMA). If a replacement or credit note is issued, the ordering party undertakes to store the faulty part for at least 24 months in accordance with the current storage directives (in compli­ance with ESD guidelines) and to hold it in readiness for onsite inspection
■■■■■■■■
48
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
8 Conformity and warranty
or for return shipment to Metrohm. Metrohm reserves the right to invoice the ordering party for these articles, including retroactively, in the event of noncompliance with these preconditions.
The same warranty periods that are specified for a corresponding new part apply to parts that are replaced or repaired within the above-men­tioned warranty periods. However, replacement or repair of a part does not extend the warranty period of the entire system.
Deficiencies arising from circumstances that are not the responsibility of Metrohm, such as improper storage or improper use, etc., are expressly excluded from the warranty.
Metrohm also offers a 120-month spare parts availability guarantee and a 60-month PC software support warranty, calculated from the date on which the product is withdrawn from the market. The content of this war­ranty is the ability of the customer to obtain functioning spare parts or appropriate software support at market prices during the time of the war­ranty period. This does not apply for software products sold under the Metrohm NIRSystems brand.
If Metrohm AG is unable to meet this obligation due to circumstances beyond the control of Metrohm AG, then the ordering party shall be offered alternative solutions at preferential conditions.
893 Professional Biodiesel Rancimat
■■■■■■■■
49

9 Accessories

Up-to-date information on the scope of delivery and optional accessories for your instrument can be found on the Internet.
When you receive your new instrument, we recommend downloading the accessories list from the Internet, printing it out and keeping it together with the manual for reference purposes.
Instruments currently sold
If you do not know the article number of your instrument, proceed as fol­lows:
Downloading the accessories list
■■■■■■■■■■■■■■■■■■■■■■
NOTE
Go to the Metrohm website http://www.metrohm.com/com.
1
2
Click on .
The Search webpage will be displayed.
Enter a search term relating to the instrument into the search field
3
and click on Find.
The search results will be displayed.
In the search results, select the Devices tab (if it is not already
4
selected) and then click on the Metrohm article number of the required instrument (e.g. 2.852.0050).
The page with information pertaining to the searched article is dis­played.
Select the Parts tab.
5
The complete list of accessories with the scope of delivery and the optional accessories will be displayed.
6
Click on .
■■■■■■■■
50
893 Professional Biodiesel Rancimat
■■■■■■■■■■■■■■■■■■■■■■
9 Accessories
The Partslists webpage will be displayed.
Select the desired output language.
7
With the article number entered, click on the command Generate
8
PDF.
The PDF file with the accessories data will be created in the language selected.
Direct access for all instruments
If you are unable to find your instrument using the search as described above, this may be due to the instrument not being sold anymore. Using the article number, you can download accessories lists for all instruments as follows:
Downloading the accessories list
Type http://partslists.metrohm.com into your Internet browser.
1
The Partslists webpage will be displayed.
Select the desired output language.
2
Enter the article number and click on the Generate PDF command.
3
The PDF file with the accessories data will be created in the language selected.
893 Professional Biodiesel Rancimat
■■■■■■■■
51

Index

Index
■■■■■■■■■■■■■■■■■■■■■■
A
Accessories
Mount ................................ 11
Air purification
Mount accessories .............. 11
Assemble
Oil separator ....................... 34
Autoxidation ............................ 23
C
Cell constant
Determine .......................... 24
Clean
Accessories ......................... 31
Measuring vessel ................ 31
Oil separator ....................... 33
Reaction vessel ................... 32
Tubing ................................ 32
Computer
Connect ............................. 22
Conductivity measurement
Measuring range ................ 44
Resolution .......................... 44
D
Determination
Prepare ............................... 29
Disassemble
Accessories ......................... 33
Oil separator ....................... 33
Drying flask
Mount ................................ 12
Dust filter
Mount ................................ 11
Replace .............................. 36
E
Electrostatic charge .................... 4
External air supply
Mount accessories .............. 13
External temperature ................ 44
F
Fuse
Replace .............................. 19
G
Gas flow
Range ................................. 45
Guarantee ................................ 48
H
Heater
Range ................................. 43
Start manually .................... 30
Switch off ........................... 30
Heating period ......................... 30
Heating time ............................ 44
I
Induction time .......................... 23
Installation qualification ............ 37
Internal air supply
Mount accessories .............. 11
M
Mains voltage ............................. 4
Maintenance agreement .......... 37
Measuring vessel
Assemble ............................ 14
Molecular sieve
Regenerate ................... 13, 36
O
Operational qualification .......... 37
Q
Quality management ................ 37
R
Reaction vessel
Assemble ............................ 14
S
Safety instructions ...................... 3
Sample identification
Enter .................................. 30
Service ....................................... 4
Shutdown temperature ............ 44
Stability time .............................. 1
Switch on ................................. 20
T
Temperature correction
Determine .......................... 25
Temperature measurement
Measuring range ................ 43
Reproducibility .................... 43
Resolution .......................... 43
W
Warranty .................................. 48
■■■■■■■■
52
893 Professional Biodiesel Rancimat
Loading...