6.4.2 Certificate of Conformity and System Validation..............................................170
6.5 Scope of delivery and ordering designations.......................................................... 172
Index ........................................................................................................................ 175
785 DMP Titrino
Overview
Herisau Switzerland
f = 50 - 60 Hz
S = 15 W
Pol
Ref
Ind I
Ind II
Key-
board
1
RS 232
1 Overview
Front view of instrument: Rear view of instrument:
7
8
9
10
1
2
Remote
1.785
Nr.
A
E
13
12
B
2
A
11
100 - 240 V
Pt 100/1000
C
Made by Metrohm
3
4
6
785 DMP Titrino
5
1
Overview
Front view of instrument:
1Exchange Unit
2Display
3Control keys and indicator lamps on the Titrino
Key < >Power switch
Key < 4 > Switching stirrer ON/OFF
Key <DOS>Dosing key. Dispensing is performed as long as
<DOS> is being pressed. Used e.g. to prepare
the Exchange Unit. The dispensing rate can be
set with potentiometer (5).
Key <STOP/FILL>- Stops procedures, e.g. titrations, conditioning.
- Filling after manual dosing with <DOS>.
Key <START>Starts procedures, e.g. titrations, conditioning.
Identical with key <START> of the separate
keypad.
Indicator lamps:
"COND."Lamp flashes when conditioning is performed
and the titration vessel is still wet. It is on if con-
ditioning is ok.
"STATISTICS"Lamp is on when the "statistics" function
(calculation of mean and standard deviation) is
on.
"SILO"Lamp is on when silo memory (for sample data)
is on.
4Setting of display contrast
5Controls the dosing rate during manual dosing with <DOS> and
subsequent filling
6Opening for the memory card
2
785 DMP Titrino
Rear view of instrument:
7RS232 interfaces
2 separate interfaces for the connection of printer, balance, and computer
8Remote lines (input/output)
for the connection of the Remote Box, Sample Changers, robots etc.
9Connection of electrodes and temperature sensor
• 2 high-impedance measuring inputs for pH and U measurements.
They can either be used separately or for differential potentiometry,
see page 158.
Important: If you work with both measuring inputs in the same vessel, the same reference electrode must be used.
• 1 measuring input for polarized electrodes, e.g. KF electrode
• 1 measuring input for PT100 or Pt1000 temperature sensor
Overview
10Connection for separate keypad
11Connection for power cable
With power supplies where the voltage is subject to severe HF disturbances, the Titrino should be operated via an additional power filter,
e.g. Metrohm 615 model.
12Connection for stirrer
728 Magnetic Stirrer, 722 Rod Stirrer, 703 or 727 Ti Stand
Supply voltage: 10 VDC (I ≤ 200 mA)
13Rating plate
with fabrication, series and instrument number
785 DMP Titrino
3
2.1. Keypad
SILO
–9STATISTICS
MEAS/HOLD
+8USER
→↑←
↓
STOP
START
CARD
/6CAL.DATA
*
USER METH
)3DEF
(2C-FMLA
MODE
REPORTS
PRINT
CURVE
2 Manual operation
2.1 Keypad
PARAMCONFIG
EP
SMPL
DATA
7
RS
4
C
5
1
.
QUIT
ABC
–
6.2132.070
MN
0
CLEARENTER
CONFIGConfiguration.
PARAMParameters.
SMPL DATA Sample data.
STATISTICS ON/OFF switching of statistics calcu-
lations of consecutive determination,
see page 41.
MEAS/HOLDON/OFF switching of measurements
between titrations and hold during titrations.
SILOON/OFF switching of silo memory for
sample data, see page 60.
CAL.DATACalibration data, see page 58.
CARDManagement of memory card, see
;
page 55.
C-FMLACalculation values, see page 40.
DEFFormulas, data output, sequence for
TIP, see page 38ff.
USER METH Management of internal method mem-
ory, see page 53.
PRINTPrinting of reports, see page 45.
REPORTSResult output.
MODEMode selection, see page 13.
USERUser name, see page 48.
CURVESwitching result/curve display.
←,→Selection of special values (dialog
marked with ":")
↑,↓Cursor key for navigation.
CLEARClears values, set special values.
ENTERStores values.
STOPStops methods.
QUITQuits inquiries, waiting times, printing.
STARTStarts methods.
The third functions (inscriptions in the triangle) on the keys of the keypad are used for
formula entry, see page 38.
4
785 DMP Titrino
2.2 Principle of data input
remote box: OFF
• If you press a key you will find a group of inquiries in
the display.
In the first line you see where you are: You pressed
key <CONFIG> and you are now in the inquiries
"configuration".
• The cursor is inverted. In our example the cursor is
on the inquiry ">monitoring". You can move the cursor up and down with keys <↑> and <↓>.
• If a dialog text is marked with ">", it contains a
group of inquiries itself. You go to this group pressing <ENTER>.
Example inquiries of "peripheral units":
The first two lines indicate again where you are.
Then you find the inquiries.
If a dialog text of an inquiry is marked with ":", you
can select a value with keys <←> and <→>
(forward/backward).
• A value is stored with <ENTER> and the cursor
moves to the next inquiry.
• With key <QUIT> you move one level up, in our example you go back to ">peripheral units".
If you press <QUIT> once more you quit the in-
quiries in "configuration" altogether.
• If you can scroll, ↓ or ↑ appear in the right lower or
upper corner of the display.
• Press key <USER METH>.
Place the cursor to ">store method" and press
<ENTER>.
The name of the method which is currently in the
working memory is displayed.
• Delete this name with <CLEAR>.
• Open the "text writing mode" with key <ABC>.
You can now select the desired character by means
of the cursor keys, then confirm this character. Select the next character...
When you confirmed the last character, i.e. your
name is complete, you quit the text writing mode
with <QUIT>.
Confirm now the name with <ENTER>.
• During text input you can correct typing errors with
<CLEAR>:
<CLEAR> deletes the characters one by one.
• If you wish to modify an existing name (e.g. if you
have names like Text 1, Text 2, Text 3), do not delete the existing name before you start the text input
mode. Proceed then as follows:
1. Press <USER METH>, place the cursor to
">store method" and press <ENTER>.
2. Open the text writing mode directly: Press key
<ABC>.
3. <CLEAR> now deletes the characters one by
one or you can add additional characters.
4. If your text is complete, leave the text writing
mode with <QUIT> and confirm the text with
<ENTER>.
6
785 DMP Titrino
2.4 Configuration, key <CONFIG>
Key <CONFIG> serves to enter device specific data.
Monitoring of instrument validation, pH calibration,
service interval and printout of diagnostic report.
peripheral units:
Selection of printer, balance, stirrer control and COM
for manual report output.
auxiliaries:
e.g. setting of dialog language, date time, type of result
display.
RS232 settings COM1 and 2:
RS parameters for the COM's.
common variables:
Values of common variables.
The display texts of the Titrino are shown to the left.
The values are the default values.
2.4 Configuration, key <CONFIG>
>monitoring
validation: OFF
time interval 365 d
time counter 0 d
calibration: OFF
Monitoring functions
Monitoring the validation interval (ON, OFF)
Monitoring is carried out at the end of the titrations and
when the Titrino is switched on. If the monitoring
responds the message "validate instrument" appears.
The message vanishes with <CLEAR>. At the same
time the counter is reset to zero.
If "on" has been set:
Time interval for validation (1...9999 d)
see also page 168.
Time counter (0...9999 d)
Counts the number of days since the last time the
counter was reset.
Monitoring the pH calibration interval (ON, OFF)
Monitoring is carried out at the end of the titrations and
when the Titrino is switched on if the measuring input
has a current calibration date. If the monitoring
responds the message "calibrate electrode" appears.
The counter will be reset to zero on the next pH
calibration of this input.
meas.input: 1
time interval 7 d
time counter 0 d
785 DMP Titrino
If "on" has been set:
Measuring input (1, 2, diff.)
Time interval for validation (1...9999 d)
Time counter (0...9999 d)
Counts the number of days since the last time the
counter was reset.
7
2.4. Configuration, key <CONFIG>
service: OFF
next service YYYY-MM-DD
system test report: OFF
>peripheral units
send to COM1: IBM
send to COM2: IBM
Monitoring the service interval (ON, OFF)
Monitoring is carried out after the Titrino has been
switched on. If the monitoring responds the message
"Service is due" appears. The message vanishes with
<CLEAR>.
If "on" has been set:
Date of next service (YYYY-MM-DD)
System test report printout (ON, OFF)
With "on" the report of the system test is printed out
after the Titrino has been switched on, see also page
168.
Settings for peripheral units
Selection of printer (Epson, Seiko, Citizen, HP, IBM) at
the Titrino COM1
"Epson", for Epson
"Seiko", e.g. for DPU-414
"Citizen", e.g. for iDP 562 RS
"HP" e.g. for Desk Jet types. Place curves always at the
beginning of a page as you cannot have them over 2
pages.
"IBM" for all printers with IBM character set Table 437
and IBM graphics, as well as for the data transmission
to a computer or a data system.
man.reports to COM: 1
balance: Sartorius
stirrer control: OFF
COM of Titrino for the output of manually triggered
reports (1, 2, 1&2)
Manually triggered reports e.g. with <PRINT> .... .
Exception <PRINT><REPORTS>: These reports are
outputted on the COM as defined in the method.
Selection of balance (Sartorius, Mettler, Mettler AT,
AND, Precisa)
Sartorius:Models MP8, MC1
Mettler:Models AM, PM and balances with 011,
012, and 016 interfaces
Mettler AT: Model AT
AND:Models ER-60, 120, 180, 182, FR-200, 300
and FX-200, 300, 320
Precisa:Models with RS232C interface
Automatic switching ON/OFF of the stirrer in the titration
sequence (ON, OFF)
If stirrer control is ON, the stirrer will be switched on at
the beginning and switched OFF at the end of a determination. For SET with conditioning the stirrer will be
switched off in the inactive state.
In the modes MEAS, CAL, and TIP the stirrer is not
switched automatically.
For stirrer control the red switch on the stirrer unit must
be ON.
8
785 DMP Titrino
2.4 Configuration, key <CONFIG>
remote box: OFF
keyboard: US
barcode: input
>auxiliaries
Connection of a remote box (on ,OFF)
To the remote socket for PC keyboard and barcode
reader, see page 155.
If "on" has been set:
Type of PC keyboard (US, German, French, Spanish,
Swiss.)
The barcode reader is used as an input aid, see page
155.
Input:The barcode string goes to the entry field in
which the cursor is currently located.
Method:The barcode string goes to the entry field
"Methods" in the silo memory.
Id1:The barcode string goes to the entry field
"Id1". (Similar for Id2 and Id3.)
Smpl size:The barcode string goes to the entry
field "smpl size".
General settings
dialog: english
date 1998-04-23
time 08:13
run number 0
auto start OFF
start delay 0 s
result display: bold
Selection of dialog language (english, deutsch,
francais, español, italiano, portugese, svenska)
Current date (YYYY-MM-DD)
Format: Year-month-day, entry with leading zeros.
Current time (HH-MM)
Format: Hours-minutes, entry with leading zeros.
Current run number for result output (0...9999)
The sample number is set to 0 when the instrument is
switched on and incremented on every determination.
Automatic starts of titrations. (1...9999, OFF)
Number of automatic starts ("number of samples").
Used for instrument interconnections in which the
external instrument does not initiate a start. Not
advisable in connections with Sample Changers.
Start delay (0...999 999 s)
Delay time after start of methods. Abort start delay time
with <QUIT>.
Type of result display at the end of the determination
(bold, standard)
bold: The calculated results are displayed in bold
characters.
standard: Displays the whole information, e.g. results,
endpoints, messages etc.
785 DMP Titrino
9
2.4. Configuration, key <CONFIG>
dev.label.
program 785.0010
>RS232 settings COM1
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
>common variables
Individual identification of devices (up to 8 ASCII
characters)
Will be printed in the result report, see page 45.
The values of all common variables are displayed. For
creating of common variables see page 43.
10
785 DMP Titrino
2.4 Configuration, key <CONFIG>
Settings with key <CONFIG> and power ON
Proceed as follows:
1. Switch the Titrino OFF.
2. Press <CONFIG> and keep it pressed during switching the Titrino ON.
The display shows the following:
Setup
>lock
>curve
>lock
<configuration>: OFF
<parameters>: OFF
<smpl data>: OFF
recall method: OFF
store method: OFF
delete method: OFF
lock:
Locking keys <CONFIG>, <PARAM> and <SMPL
DATA>, and the functions "recall method", "store
method" and "delete method" of the internal method
memory in the Titrino.
curve:
Changes the appearance of the curve printout.
lock
"ON" means that the corresponding function is not
accessible anymore.
The corresponding key is locked.
The corresponding function in the internal method
memory of the Titrino is locked.
785 DMP Titrino
11
2.4. Configuration, key <CONFIG>
>curve
grid: ON
frame: ON
scaling: Full
width 0.75
length 0.1
curve
The settings are valid for COM1 and COM2.
If you change the printer type, the following settings are
initialized according to the printer.
Grid drawing (ON, OFF)
Frame drawing (ON, OFF)
Type of scaling (Full, Auto)
Full: The scaling goes from the greatest to the smallest
value.
auto: The scaling from tick to tick, e.g. the
smallest/greatest values lie in between the first/last tick.
Width (0.2...1.00)
1 is greatest width. If you set 1 you may loose the label
at the right margin.
Length (0.01...1.00)
Volume axis
Curve length per cylinder volume
0.01100 cm
0.110 cm
0.52 cm
11 cm
Time axis:
Curve length
0.01100 cm
0.110 cm
0.5 2 cm
11 cm
12
785 DMP Titrino
2.5 Selection of the mode, key <MODE>
MODE
2.5 Selection of the mode, key <MODE>
;
–
mode
mode: DET
measured quantity: pH
Press key <MODE> until the desired mode is
displayed and confirm with <ENTER>.
Select the measured quantity pH, U, Ipol, Upol, (T) with
<←> or <→> and confirm it also with <enter>.
The following modes can be selected:
• DET: Dynamic Equivalence-point Titration
• MET: Monotonic Equivalence-point Titration
• SET: Set Endpoint Titration.
• CAL: pH Calibration.
• MEAS: Measuring.
• TIP: Titration Procedure. Linking of various
commands and methods to a titration procedure.
These standard modes are equipped with a set of
standard parameters. They only need few settings in
order to be ready to work.
TIP is an empty "shell". The TIP sequence has to be
defined with <DEF>, see page 49.
You will find a survey of the modes in the short
Instructions for use, page 10.
785 DMP Titrino
13
2.6. Parameters, key <PARAM>
2.6 Parameters, key <PARAM>
The key <PARAM> is used for the entry of values that
PARAM
2.6.1 Parameters for DET and MET
determine the modes. Values marked with "cond." are
accessible during the conditioning in the SET mode.
"**titr." means that these values can be changed during the titration. They influence the ongoing determination. Other values can only be changed in the inactive
state.
The display texts of the Titrino are shown to the left.
The values are the default values.
determine the course of the titration and measured
value acquisition.
stop conditions:
Parameters for the automatic termination of the titration.
statistics:
Calculation of mean values and standard deviation,
see page 41.
evaluation:
Parameters for the evaluation of EP's, fix EP's, and
pK/HNP.
preselections:
ON/OFF of various auxiliary functions such as automatic requests after the start and activate pulse.
General titration parameters
Measuring point density (0...9)
0 means highest density, 9 lowest. Selection of the
meas.pt.density, see page 20.
14
min.incr. 10.0 ul
DET
V step 0.10 ml
MET
Minimum increment (0...999.9 uL)
The increment is dispensed at the beginning of the
titration and in the region of the equivalence point.
Size of volume increment (0...9.999 mL)
Dosing step. Small volume increments are needed to
determine blank values or to assure accuracy with
highly unsymmetrical curves. Selection of size of the
increment, see page 20.
785 DMP Titrino
2.6 Parameters, key <PARAM>
dos.rate max. ml/min
**titr.
signal drift 50 mV/min
**titr.
equilibr.time 26 s
**titr.
Dosing rate for volume increments (0.01...150 mL/min,
max.)
<CLEAR> sets "max.".
The maximum rate depends on the Exchange Unit:
Exchange Unitmax.
Drift criterion for measured value acquisition. (input
range depends on the measured quantity:
pH, U, Ipol: 0.5...999 mV/min, OFF
Upol:0.05...99.9 uA/min, OFF)
<CLEAR> sets "OFF".
This type of measured value acquisition is often called
an equilibrium titration.
"OFF" means that the measured value is acquired after
an equilibration time. This may be useful for slow
titration reactions or when the response of the
electrode assembly is slow.
Waiting time for measured value acquisition. (0...9999 s,
OFF)
<CLEAR> sets "OFF".
If no new equilibration time has been entered, the
Titrino calculates an equilibration time appropriate to
the drift, see page 19. The measured value is acquired
as soon as the first criterion (drift or time) has been
met.
start V: OFF
start V 0.0 ml
factor 0
dos.rate max. ml/min
**titr.
Type of start volume (OFF, abs., rel.)
"OFF": start volume switched off
"abs":absolute start volume in mL
"rel.":relative start volume to sample size.
If "abs." is set:
Absolute start volume (0...999.99 mL)
If "rel." is set:
Factor for relative start volume (0...±999 999).
Calculated as: start V in mL = factor * sample size
Dosing rate for start volume (0.01...150 mL/min, max.)
<CLEAR> sets "max.".
Maximum rate depends on the Exchange Unit:
Exchange Unitmax.
Waiting time, e.g. for equilibration of the electrode after
the start or reaction time after dosing of start volume.
The pause can be aborted with <QUIT>.
Measuring input for pH and U (1, 2, diff.)
Request for measuring input for pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 158.
With polarized electrodes, instead of the measuring
input the
polarization current (-127...127 uA)
or the
polarization voltage (-1270...1270 mV, in steps of 10
mV)
is inquired.
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby state to a
measurement. "OFF" means that the test is not
performed.
Titration temperature (-170.0...500.0 °C)
The temperature is continuously measured if a T
sensor is connected. This parameter is used for
temperature compensation in pH titrations.
>stop conditions
stop V: abs.
**titr.
stop V 99.99 ml
**titr.
factor 999999
**titr.
stop pH OFF
**titr.
Stop conditions for the titration
If several stop conditions have been set, the criterion
which is met first applies.
Type of stop volume (abs., rel., OFF)
"abs.": absolute stop volume in mL.
"rel.":relative stop volume to sample size.
"OFF": stop volume switched off. Stop volume is not
monitored.
If "abs." is set:
Absolute stop volume (0...9999.99 mL)
If "rel." is set:
Factor for relative stop volume (0...±999 999)
Calculated as: Stop V in mL = factor * sample size
Stop at measured value (input range dependent on
measured variable:
pH:0... ± 20.00, OFF
U, Ipol: 0... ± 2000 mV, OFF
Upol:0... ± 200.0 uA, OFF)
<CLEAR> sets "OFF".
"OFF" means that the criterion is not monitored.
16
785 DMP Titrino
2.6 Parameters, key <PARAM>
stop EP 9
**titr.
filling rate max. ml/min
**titr.
>evaluation
EPC 5
Stop after a number of EP's have been found (1...9,
OFF)
<CLEAR> sets "OFF".
"OFF" means that the criterion is not monitored.
Filling rate (0.01...150 mL/min, max.)
<CLEAR> sets "max.".
The maximum rate depends on the Exchange Unit:
Exchange Unitmax.
Equivalence point criterion (input range depends on
mode.
For DET:
0...200
For MET:
pH:0.10...9.99
U, Ipol:1...999 mV
Upol:0.1...99.9 uA)
Threshold for the size of the jump, see page 22.
EP recognition: all
low lim.1 pH -20.00
up lim.1 pH 20.00
Recognition of EP's which fulfill the EP criterion. (all,
greatest, last, window, OFF)
Selection of equivalence point recognition:
"all":All equivalence points are recognized.
"greatest": Only the greatest (steepest) equivalence
point is recognized.
"last":Only the last equivalence point is
recognized.
"window": Only EP's in specified windows are
recognized.
"OFF":Evaluation switched off.
If "window" is selected, lower and upper limits of
windows are inquired.
Lower limit of window 1
and
upper limit of window 1
(Input ranges for both inquiries depend on the
measured quantity):
pH:0...±20.00, OFF
U, Ipol:0...±2000 mV, OFF
Upol:0...±200.0 uA, OFF)
<CLEAR> sets "OFF".
Only equivalence points are recognized which lie within
the set lower and upper limits. The equivalence point
numbering is defined with the windows, see page 22.
Window inquiries continue until the lower limit is set to
"OFF". Up to 9 possible windows.
Always set both limits to ≠ OFF for a valid window.
785 DMP Titrino
17
2.6. Parameters, key <PARAM>
fix EP1 at pH OFF
pK/HNP: OFF
>preselections
req.ident: OFF
Interpolation of volumes at fixed times (input range
depends on the measured quantity:
pH:0...±20.00, OFF
U, Ipol:0...±2000 mV, OFF
Upol:0...±200.0 uA, OFF)
<CLEAR> sets "OFF".
If a fix end point has been set, the volume value for the
input measured value is interpolated from the curve,
see also page 23. The volume values are available as
C5X.
Fix EP's are inquired until "OFF" is set. Up to 9 fix EP's.
Evaluation of pK or HNP (ON, OFF)
pK evaluation in case of pH titrations and half
neutralization potential for U, see page 23.
Preselections for the sequence
Request of identifications after start of titration. (id1,
id1&2, all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
req.smpl size: OFF
limit smpl size: OFF
low lim. 0.0
up lim. 999999
activate pulse: OFF
Request of sample size after start of titration (value, unit,
all, OFF)
"all" requests the value, then the unit.
Limiting value check for sample size (ON, OFF)
With "on" the error message "sample size out." appears
if the entry is outside the set limits. The limiting values
are shown in the display window.
The absolute value of the limit is checked during sample size input and during the calculation of the results.
If "on" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
Pulse output on I/O line "activate" (L6, pin 1) of the
remote socket (ON, OFF)
After the start, the activate pulse is outputted and the stirrer switched on.
The start delay time is waited off.
The sample identifications and the sample size are re-
quested.
The start volume is dispensed (no meas.value acquisition)
and the pause is waited off.
During the titration the volume increments are dispensed
and after each increment a measured value is acquired.
Meas.values are either acquired drift controlled
("equilibrium titration") or a after a fixed waiting time. If you
have not intentionally changed the equilibration time, it is
calculated according to the formula:
150
equilibr.time in s = —————— + 5
ûDrift + 0.01
The criterion (drift or equilibration time) which is first met
applies. This avoids "infinite" titrations. If the parameter
"signal drift" is set "OFF", the measured values are acquired after a fixed equilibration time.
The titration is terminated according to the first criterion
which is met.
The stirrer is switched off.
Evaluations and calculations are carried out.
Data are outputted.
785 DMP Titrino
19
2.6. Parameters, key <PARAM>
Reagent feeding and EP evaluation of DET
The reagent feeding of DET is similar to the controlling, a human being would apply in
manually controlled titrations: Great volume increments are dosed far away from the EP,
small increments in the region of the equivalence point.
The size of the volume increments dosed by the Titrino is determined by the following
parameters:
meas.pt.densityThe measuring point density is entered as a relative
value from 0...9. Input of a low number means small
volume increments, i.e. a large measuring point density
on the curve. A curve results which reproduces all fine
details. "Fine details", however, also include signal noise,
which can easily lead to unwanted equivalence points. A
high number, i.e. low measuring density, on the other
hand, allows a more rapid titration. The standard value of
4 is suitable for most cases.
If you work with small cylinder volumes (1 or 2 ml), a
small measuring point density may be advisable. In
these cases you should also lower the drift for
meas.value acquisition and set a higher EPC.
min.incr.Defines the minimum possible increment. This minimum
increment is dosed at the beginning of the titration and
in the region of the equivalence point (for steep curves).
Use low minimum increments only, if small volumes of
titrant consumption are expected, e.g. in micro titrations;
otherwise unwanted equivalence points may arise. The
standard value of 10.0 uL is suitable for most cases.
EPC
The EP's are evaluated according to a special
METROHM procedure which is so sensitive that even
weak equivalence points are determined correctly.
Equivalence Point Criteria.
The preset EPC is compared to the found ERC
(Equivalence point Recognition Criteria) for each
evaluated equivalence point. The size of the ERC is
displayed in the titration curve, see page 47. The ERC is
the first derivative of the titration curve overlaid with a
mathematical function so that small maxima become
higher and great maxima smaller. EP's whose ERC is
below the preset EPC will not be recognized. For most
cases the standard value of 5 is suitable. The evaluation
can be repeated at any time after the titration in a "dry
run" with changed evaluation criteria. The old titration
data are not deleted until a new titration is started.
20
785 DMP Titrino
2.6 Parameters, key <PARAM>
Reagent feeding and EP evaluation of MET
In monotonic titrations, the volume increment is constant over the whole titration curve.
V stepVolume increment.
A prerequisite for good accuracy is the correct size of
the volume increments. A good value is given by V step
= 1/20 VEP (VEP = volume of the EP). In any case, the
increment size should always be between 1/10 VEP and
1/100 VEP; with steep jumps preferably in the region of
1/100 and with flat jumps preferably in the region of
1/10. The accuracy of the evaluation can not be
increased by dispensing small increments as the
changes in the measured value can then be of the same
order of magnitude as the noise. This can produce
"ghost EP's"!
The EP's are localized with an algorithm which is based
on Fortuin and has been adapted by METROHM for
numeric procedures. (METROHM Bulletin 2, No. 10,
1971). Here, the greatest change in the measured value
is sought (∆n). The exact equivalence point is
determined with an interpolation factor, which depends
on the delta values before and after ∆n:
Equivalence Point Criteria.
The preset EPC is compared to the found ERC
(Equivalence point Recognition Criteria) for each
evaluated equivalence point. The size of the ERC is
displayed in the titration curve, see page 47. ERC is the
sum of the measured value changes before and after the
break:
∆
+∆
n-2
+∆n+∆
n-1
n+1
+∆
n+2
(In certain cases there are only 3 or 1 summand).
EP's whose ERC is below the preset EPC will not be
recognized. For most titrations the standard value is
suitable.
The evaluation can be repeated at any time after the
titration in a "dry run" with changed evaluation criteria.
The old titration data are not deleted until a new titration
is started.
785 DMP Titrino
21
2.6. Parameters, key <PARAM>
U/mV
V/ml
U/mV
V/ml
EP recognition criteria for DET and MET
The parameter "EP recognition" offers you a range of possibilities to ensure selection of
the EP you are interested in: If the desired jump is very large, you can select the "greatest"
jump (with DET the steepest jump will be evaluated). Thus you always obtain just one EP
per titration (EP1).
If you wish to determine the sum of different components (e.g. acid or base numbers), the
"last" jump can be the correct one.
And finally you can set a "window" for each expected EP.
EP windows
EP windows are used
• to suppress disturbing influences and EP's which are not needed.
• to increase the liability for the calculation of the results. The EP windows make an
unequivocal assignment of the EP's possible: per window one EP is recognized; the
numbering of EP's is defined by the windows so that even if EP's are missing, the
calculations are still performed with the correctly assigned EP volumes.
An EP window defines the range in which an EP is expected. EP's outside these ranges
are not recognized. Windows are defined on the measured value axis.
EP window 2
EP window 1
EP 1
upper lim. 1
lower lim. 1
EP 1
+EP window 1
EP 2
2 EP's are recognized. Their
numbering is defined by the windows:
Window 1 ⇒ EP1
Window 2 ⇒ EP2
If more than 1 EP is expected, a
window must be set for each EP.
Windows must not overlap. They may
only touch each another.
Rule: If there are more than 1 EP in a
window, the first jump is recognized as
EP1, the second is not recognized.
EP1 is marked as EP1 + to indicate
that more than one EP has been found
in the window
22
785 DMP Titrino
2.6 Parameters, key <PARAM>
Fix EP's
Fix EP's allow determination of the associated volume value for every inputted measured
value on the titration curve. This function is useful for performing standard methods such
as TAN/TBN determinations. For the evaluation of fix EP's, the pH calibration is advisable.
The volume values of the fix EP's are available for the calculation as C5X:
Fix EP1 ⇒ C51
:
Fix EP9 ⇒ C59Maximum 9 fix EP's are possible.
pK and HNP evaluation
The following relation (Henderson-Hasselbach), derived from the law of mass action,
exists between the activities of a conjugate acid-base pair in aqueous solutions:
pH = pK + log (aB/aA)
When the activities aB = aA, then pH = pK is valid. This value corresponds to the pH at
the half neutralization point and can be taken from the titration curve. For pK determinations, a careful pH calibration is necessary.
The pK value determined is an approximate value owing to the fact that the ionic strength
of the solution has not been taken into account. For more exact values, titrations must be
performed with decreasing ionic strength and the results extrapolated to zero ionic
strength.
The evaluation of pK's in aqueous solutions is limited to
pK>3.5due to the leveling effect of strong acids in aqueous solutions
pK<10.5because for weaker acids no inflection points can be found aqueous
solutions.
pK evaluation for polybasic acids and for acid mixtures is also possible.
In non-aqueous solutions, the half neutralization potential (HNP) is often used instead of
pK. The HNP is evaluated accordingly.
A start volume must be smaller than half of the equivalence point volume of the first EP.
The pK/HNP values are available for calculation as C6X.
Minimum/maximum evaluation
Volumes at minimum or maximum voltage values are extrapolated. These volumes are
accessible as variables C49 and C48 resp. and can be used in formulas.
Parameters for the termination of the titration.
statistics:
Calculation of mean values and standard deviation,
see page 41.
preselections:
ON/OFF of various auxiliary functions such as automatic requests after the start and activate pulse.
Control parameters for EP1 or EP2, resp.
Preset EP1 (input range depends on the measured
quantity:
pH:0...±20.00, OFF
U, Ipol:0...±2000 mV, OFF
Upol:0...±200.0 uA, OFF)
<CLEAR> sets "OFF".
If EP1 is "OFF", no further inquiries under SET1 appear.
dynamics OFF
**titr.
max.rate 10.0 ml/min
**titr.
Distance from EP where constant dosing should stop
and controlling begins. (control range, input range
depends on the measured quantity:
pH:0.01...20.00, OFF
U, Ipol:1...2000 mV, OFF
Upol:0.1...200.0 uA, OFF)
<CLEAR> sets "OFF".
"OFF" means largest control range, i.e. low titration.
Outside the control range, dispensing is performed
continuously, see also page 30.
Maximum dosing rate (0.01...150 mL/min, max.)
<CLEAR> sets "max.".
This parameter determines primarily the addition rate
outside the control range, see also page 30.
The maximum rate depends on the Exchange Unit:
Exchange Unitmax.
This parameter determines the addition rate right at the
start and the end of the titration, see also page 30.
This parameter influences the titration speed and
therefore its accuracy very strongly: A smaller min.rate
results in a slower titration.
Type of stop criteria (drift, time)
Titration stops if EP is and stop drift is reached. (1...999
uL/min)
Titration stops if there is no dosing during t(delay).
(0...999 s, INF)
<CLEAR> sets "INF"
Switch off when the end point is reached and the set
time after the last dispensing has elapsed.
If "INF" is set, an inquiry regarding the stop time
appears.
If t(delay) is "INF"
Stop after a time (0...999 999 s, OFF)
<CLEAR> sets "OFF".
Stop after the set time after the start of the titration.
"OFF" means no stop, i.e. titration for an "infinitely" long
time.
>titration parameters
titr.direction: auto
pause 1 0 s
**titr.
start V: OFF
cond.
start V 0.0 ml
cond.
General titration parameters
Direction is set automatically (+, -, auto)
auto: The direction is set automatically by the Titrino
(sign [U
- EP]).
first
+:Direction of higher pH, higher voltage (more
"positive"), larger currents.
- : Direction of lower pH, lower voltage, smaller
currents.
The titration direction is fixed if two EP's are set. In this
case, an input for titration direction has no meaning.
Pause 1 (0...999 999 s)
Waiting time before start volume, e.g. for equilibration
of the electrode after the start. The waiting time can be
aborted with <QUIT>.
Type of start volume (OFF, abs., rel.)
"OFF":start volume switched off
"abs.":absolute start volume in mL
"rel.":relative start volume to sample size.
If "abs." is set:
Absolute start volume (0...999.99 mL)
factor 0
cond.
785 DMP Titrino
If "rel." is set:
Factor for relative start volume (0...±999 999).
Calculated as: start V in mL = factor * sample size
25
2.6. Parameters, key <PARAM>
dos.rate max. ml/min
**titr.
pause 2 0 s
**titr.
extr.time 0 s
**titr.
meas.input: 1
Dosing rate for start volume
(0.01...150 mL/min, max.)
<CLEAR> sets "max.".
The maximum rate depends on the Exchange Unit:
Exchange Unitmax.
Waiting time after start volume, e.g. reaction time after
dosing of a start volume. The waiting time can be
aborted with <QUIT>.
Extraction time (0...999 999 s)
During this time the titration is already running. It will be
not stopped (also if the EP is already reached) before
the extraction time is terminated. The extraction time
can be aborted with <QUIT>.
Measuring input (1, 2, diff.)
Inquiry only with measured quantities pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 158.
I(pol) 1 uA
U(pol) 400 mV
electrode test: OFF
temperature 25.0 °C
cond.
time interval 2 s
cond.
With polarized electrodes, instead of the measuring
input the
polarization current (-127...127 uA)
or the
polarization potential (-1270...1270 mV, in steps of 10
mV)
is inquired.
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive state to a measurement.
"OFF" means that the test is not performed.
Titration temperature (-170.0...500.0°C).
Temperature is measured at the start of the titration if a
T sensor is connected. The value is used for
temperature compensation in pH titrations.
Time interval (1...999 999 s)
Time interval for acquisition of a measured value into
the measuring point list.
26
785 DMP Titrino
2.6 Parameters, key <PARAM>
>stop conditions
stop V: abs.
**titr.
stop V 99.99 ml
**titr.
factor 999999
**titr.
filling rate max. ml/min
Stop conditions for titration
If this is not "normal", i.e. after reaching the EP.
Type of stop volume (abs., rel., OFF)
"abs.":absolute stop volume in mL.
"rel.":relative stop volume to sample size.
"OFF":stop volume switched off. Stop volume
is not monitored.
If "abs." is set:
Absolute stop volume (0...9999.99 mL)
If "rel." is set:
Factor for relative stop volume (0...±999 999)
Calculated as:
Stop V in mL = factor * sample size
Filling rate after the titration
(0.01...150 mL/min, max.)
<CLEAR> sets "max.".
The maximum rate depends on the Exchange Unit:
Exchange Unitmax.
Automatic conditioning of titration vessel. (ON, OFF)
If conditioning is "on", between the titrations the titration
solution is constantly maintained at the (1st) end point.
When conditioning is performed, the volume drift can
be displayed during the conditioning:
Display of drift during conditioning (ON, OFF).
Volume drift.
Type of drift correction (auto, man. OFF)
auto: The value of the drift will be stored automatically
at the start of the titration.
Value for manual drift correction (0...99.9 ul/min)
Request of identifications after start of titration (id1,
id1&2, all, OFF)
After start, sample identifications can be requested
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
785 DMP Titrino
27
2.6. Parameters, key <PARAM>
req.smpl size: OFF
cond.
limit smpl size: OFF
cond.
low lim. 0.0
cond.
up lim. 999999
cond.
activate pulse: OFF
cond.
Request of sample size after start of titration (value, unit,
all, OFF)
"all" the value and the unit will be requested.
Limiting value check for sample size (ON, OFF)
With "on" the error message "sample size out." appears
if the entry is outside the set limits. The limiting values
are shown in the display window.
The absolute value of the limit is checked during sample size input and during the calculation of the results.
If "on" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
Pulse output on I/O line 6 (L6, pin 1) of the remote
socket (first, all, cond., OFF)
see page 167.
28
785 DMP Titrino
Titration sequence of SET
<START>
(Start delay)
(Start conditions)
Calculations
Data output
(Reconditioning)
(Stirrer OFF)
2.6 Parameters, key <PARAM>
(Activate pulse)
(Stirrer ON)
(Preconditioning)
(<START>
(Activate pulse)
(Start delay)
(Request ident.)
(Request smpl size)
(Extraction time)
Titration with test of
stop criterion
After the start, the activate pulse is outputted and the stirrer switched on.
The start delay time is waited off.
If conditioning is on, the sample solution is titrated until
the (first) EP is reached. The display shows then
drift OK 2.3 ul/min
or
SET pH conditioning
The vessel is now conditioned. The titration can be started
with <START>.
The sample identifications and the sample size are requested.
The temperature is measured if a T sensor is connected.
Pause 1 is waited off, the start volume is dispensed and
pause 2 waited off.
The titration to EP1 then to EP2 is executed. If the extrac-
tion time is not yet over when EP1 has been reached, the
titration will only be terminated after the extraction time is
over.
The stirrer is switched off when there is no conditioning.
Calculations are carried out.
Data are outputted.
Conditioning is carried out.
785 DMP Titrino
29
2.6. Parameters, key <PARAM>
U/mV
V/ml
U/mV
V/ml
Control parameters
The control parameters can be set separately for each end point. Optimize your control
parameters for routine analyses for samples with a rather low content.
During the titration, reagent dosing occurs in 3 phases:
U/mV
EP
1. Initial dosing:
Here the dosing rate
increases constantly. The rate
starts with "min.rate" and
goes up to "max.rate".
2. Continuous dosing:
V/ml
t/s
Dosing is performed at the
maximum rate "max.rate" until
the control range (dynamics)
is reached.
3. Control range (dynamics):
In this range, dosing is
performed in single steps.
The last dosing steps are
controlled by the parameter
"min.rate".
Initial
dosing
Continuous
dosing
V/ml
Control
range
Trial settings for the size of the control range
Set a large control range for steep curves. Relatively flat curves, on the other hand, need
a smaller control range. You can get an initial, good approximation for the start of the
control range from the intersection point of the tangents:
30
Control
range
"steep curves""flat curves"
EP
Control
range
EP
785 DMP Titrino
2.6 Parameters, key <PARAM>
Relation between the stop criteria "time" and "drift"
The stop criterion "time", t(delay), means that the end point must be exceeded for a
certain period of time. In other words, after the last dosed increment, time t is allowed to
elapse before the titration is stopped. The size of this last increment depends on the
volume of the Exchange Unit used. With a 20 mL Exchange Unit, the smallest possible
increment is 2 ul. With a t(delay) = 5 s, the last 2 ul reagent must thus suffice for 5 s or
longer. This results in a drift of ≤ 2 uL/5 s = 24 uL/min (the drift can be less than 24
uL/min as it is not known whether the last increment would also suffice for 10 s). If you
have been working up to this point with a 20 mL Exchange Unit and a t(delay) = 5 s, you
can set a value of ≤ 24 uL/min as stop drift. The following table shows several values for
the maximum stop drift.
t(delay)
5 s10 s20 s
min.incr. (Exchange Unit)
0.5 uL (5 ml)6 uL/min3 uL/min1.5 uL/min
1 uL (10 ml)12 uL/min6 uL/min3 uL/min
2 uL (20 ml)24 uL/min12 uL/min6 uL/min
5 uL (50 ml)60 uL/min30 uL/min15 uL/min
Same t(delay) with a range of extremely small volume increments means different switchoff points. In case the stop criterion "drift" is used, the stopping point remains the same.
785 DMP Titrino
31
2.6. Parameters, key <PARAM>
If you have entered the endpoint and the control range (dynamics), the default values for
the other control parameters should suffice for the first titration. If you encounter
difficulties in optimizing your titration, the following table will be of use.
How to proceed if ...
ProblemPossible causes and corrective measures
Dosing at the end too long and
with too small increments.
"Never ends!"
• Increase "min.rate". Perform an experiment with a
much higher min.rate.
• Change switch-off criterion. Attempt, e.g. to increase
the stop drift or use a shorter t(delay) as stop
criterion.
• Possibly pass an inert gas through the titration
vessel.
"Overshoots". Titration is not
controlled, i.e. at the end single
pulses are not dosed.
• Lower "max.rate".
• Set larger control range.
• Set "min.rate" much lower.
• Optimize arrangement of electrode and buret tip and
improve stirring, see page 160. This is particularly
important with very fast titration reactions and with
steep curves.
Titration time is too long.• Set higher "min.rate".
determine the measurement.
statistics:
Calculation of mean values and standard deviation,
see page 41.
preselections:
ON/OFF of various auxiliary functions such as automatic requests after the start and activate pulse.
Measuring parameters
Drift criterion for measured value acquisition (input
range depends on the measured quantity:
pH, U, Ipol:0.5...999 mV/min, OFF
Upol:0.05...99.9 uA/min, OFF
T:0.5...999 °C/min, OFF)
<CLEAR> sets "OFF".
"OFF" means that the measured value is acquired after
a fixed equilibration time.
Waiting time for measured value acquisition (0...9999 s,
OFF)
<CLEAR> sets "OFF".
If no new equilibration time has been entered, the
Titrino calculates an equilibration time appropriate to
the drift with the formula
150
equilibr.time (in s) = ДДДДДДДДД + 5
ûDrift + 0.01
The measured value is acquired when the first criterion
(drift or time) is met. With drift and time "OFF", you will
have an "infinite" measurement.
meas.input: 1
I(pol) 1 uA
U(pol) 400 mV
electrode test: OFF
785 DMP Titrino
Measuring input for pH and U. (1, 2, diff.)
Inquiry only with measured quantities pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 158.
With polarized electrodes, instead of the measuring
input the
polarization current (-127...127 uA)
or the
polarization potential (-1270...1270 mV, in steps of 10
mV)
is inquired.
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby mode to a
measurement. "OFF" means that the test is not
performed.
33
2.6. Parameters, key <PARAM>
temperature 25.0 °C
time interval 2 s
>preselections
req.ident: OFF
req.smpl size: OFF
limit smpl size: OFF
low lim. 0.0
up lim. 999999
Temperature (-170.0...500.0 °C)
Temperature is measured at the start if a T sensor is
connected.
This parameter is used for temperature compensation
in pH measurements.
Time interval (1...999 999 s)
Time interval for acquisition of measured values.
Preselections for the sequence
Request of sample identifications after start of titration
(id1, id1&2, all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after start of titration (value, unit,
all, OFF)
Limiting value check for sample size (ON, OFF)
With "on" the error message "sample size out." appears
if the entry is outside the set limits. The limiting values
are shown in the display window.
The absolute value of the limit is checked during sample size input and during the calculation of the results.
If "on" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
activate pulse: OFF
Pulse output on line "activate" (L6, pin 1) of the remote
socket (ON, OFF)
see page 166.
34
785 DMP Titrino
2.6.4 Parameters for CAL
The calibration interval may be monitored, see page 7.
2.6 Parameters, key <PARAM>
parameters
>calibration parameters
>statistics
>calibration parameters
meas.input: 1
cal.temp. 25.0 °C
buffer #1 pH 7.00
buffer #2 pH 4.00
buffer #3 pH OFF
calibration parameters
determine the calibration procedure.
statistics:
Calculation of mean values and standard deviation,
see page 41.
Calibration parameters
Measuring input (1, 2, diff.)
Measuring input 1 or 2 or differential amplifier;
Connection of electrodes, see page 158.
Calibration temperature (-20.0 ... 120.0 °C)
If a T sensor is connected, the temperature will be
measured.
The calibration temperature can also be input during
the calibration sequence.
pH value of first buffer (0...±20.00)
The pH value of the buffers can be put in during the
calibration sequence.
pH value of second and the following buffers
(0...±20.00, OFF)
<CLEAR> sets "OFF".
Buffers are requested until "OFF" is set. This gives an
n-point calibration. Up to 9 buffers.
A regression line will be calculated in calibrations with
more than 2 buffers.
signal drift 2 mV/min
equilibr.time 110 s
electr.id
sample changer cal: OFF
785 DMP Titrino
Drift for measured value acquisition (0.5...999 mV/min,
OFF)
<CLEAR> sets "OFF".
"OFF" means that the measured value is acquired after
an equilibration time.
Equilibration time (0...9999 s, OFF)
<CLEAR> sets "OFF".
If a new equilibration time has not been entered, the
Titrino calculates an equilibration time appropriate to
the drift, see page 33. The measured value is acquired
as soon as the first criterion (drift or time) has been
met.
If drift and time are both set to "OFF", the measured
value acquisition is immediate.
Electrode identification (up to 8 characters).
Calibration with sample changer (ON, OFF)
In calibrations with a sample changer, there are no
35
2.6. Parameters, key <PARAM>
<START>
(Start delay)
Buffer 1 pH
Buffer 1 pH
etc.
Measuring buffer 1
Measuring buffer 2
Data output
(Activate pulse)
hold points in the calibration sequence for inputs.
Calibration temperature and pH values of the buffers
(which are temperature dependent) must therefore be
entered in advance. The inputs in key <PARAM> are
valid.
activate pulse: OFF
Calibration sequence
Measuring cal.temp.
or
entry
Pulse output on the line "activate" (L6, pin 1) of the
remote socket (all, first, OFF)
See page 167.
After the start, the activate pulse is output.
The start delay time is waited off.
Then, the calibration temperature is measured.
It no T sensor is connected, you enter the temperature
manually. Store the value with <ENTER> or continue with
<START> (T is not stored).
Enter the nominal value of the first buffer.
Store the value with <ENTER> or continue with
<START> (the value is not stored).
The first buffer is measured.
Enter the nominal value of the second buffer.
Store the value with <ENTER> or continue with
<START> (the value is not stored).
Leave the calibration with <STOP> ⇒ 1 point calibration.
The second buffer is measured.
As many buffers appear as have been specified in the
<PARAM> key (up to 9).
You may leave the calibration any time with <STOP>.
Data are output.
The calibration data are available for calculation:
C46: pHas
C47: Electrode slope
Calibration data can be viewed at any time with the
<CAL.DATA> key and the calibration report printed out
using the key sequence
<PRINT><CAL.DATA><ENTER>.
36
785 DMP Titrino
2.6 Parameters, key <PARAM>
2.6.5 Parameters for TIP
In TIP, several commands and methods can be linked to make a titration procedure. The
TIP sequence can be defined with <DEF>, see page 49.
parameters
>sequence
>statistics
>preselections
>preselections
req.ident.: OFF
req.smpl size: OFF
limit smpl size: OFF
low lim. 0.0
up lim. 999999
sequence
Parameters for the TIP sequence see page 50.
statistics:
Calculation of mean values and standard deviation,
see page 41.
preselections:
ON/OFF of various auxiliary functions such as automatic requests after the start and activate pulse.
Preselections for the sequence
Request of sample identifications after start (id1, id1&2,
all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after the start (value, unit, all,
OFF)
Limiting value check for sample size (ON, OFF)
With "on" the error message "sample size out." appears
if the entry is outside the set limits. The limiting values
are shown in the display window.
The absolute value of the limit is checked during sample size input and during the calculation of the results.
If "on" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
meas.mode: OFF
meas.input: 1
I(pol) 1 uA
U(pol) 400 mV
electrode test: OFF
temperature 25.0 °C
785 DMP Titrino
Measured quantity (pH, U, Ipol, Upol, T, OFF)
Quantity for measurements with key <MEAS/HOLD>.
Measuring input (1, 2, diff.)
Inquiry only with measured quantities pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 158.
With polarized electrodes, inquiry of
polarization current (-127...127 uA) or
polarization potential (-1270...1270 mV, steps of 10 mV)
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby state to a
measurement. "OFF" means that the test is not
performed.
Temperature (-170.0...500.0 °C)
Temperature for pH compensation. Its value has to be
entered manually even if a T sensor is connected.
Key <DEF> contains various inquiries for result
calculations and data output. The data of this key are
method specific and they are stored in the method
memory together with the method.
formula:
Formulas for result calculations.
The display texts of the Titrino are shown to the left.
The values are the default values.
Input of formulas
Enter formula number (1...9)
You can calculate up to 9 results per method.
Enter a number 1...9.
Input of formula
Example:
RS1=EP1*C01/C00
Enter formula by means of 3rd functions of keyboard.
Here you will find operands, mathematical operations
and parentheses. Operands require a number as an
identification. You can use the following operands:
EPX: EP's. X = 1...9
RSX: Results which have already been calculated with
previous formulas. X = 1...9.
CXX: Calculation constants. XX = 00...79.
38
Rules:
• Calculation operations are performed in the
algebraic hierarchy: * and / before + and -.
• Store formula with <ENTER>.
• Calculation quantities and operands can be deleted
with <CLEAR> one by one.
• To delete a complete formula press <CLEAR>
repeatedly until only RSX remains in the display.
Confirm with <ENTER>.
If a formula is stored with <ENTER>, result text,
number of decimals and result unit will be requested:
785 DMP Titrino
2.7 Result calculations
RS1 text RS1
Text for result output (up to 8 characters)
Text input see page 6.
RS1 decimal places 2
RS2 unit: %
Number of decimal places for result (0...5)
Selection of result unit (%, ppm, g/l, mg/ml, mol/l,
mmol/l, g, mg, ml, mg/pc, s, ml/min, no unit or up to 6
characters).
RS1 limit control: OFF
Limit control for the result (on, off)
The limits are checked each time a result is calculated.
If "on" has been set:
Sets line L13 of the remote socket (OFF, active, pulse)
if the result lies outside the limits.
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00Sample size, see page 59.
C01...C19Method specific operands, see page 40. They are stored with the
method in the method memory.
C21...C23Sample specific operands, see page 59ff.
C26, 27 Mean values from silo calculations.
C30...C39Common variables.
C40Initial measured value of the sample, last measured value in MEAS.
C41End volume.
C42Determination time.
C43Volume drift for SET with conditioning.
C44Temperature.
C45Dispensed start volume.
C46Asymmetry-pH (calibration).
C47Electrode slope (calibration).
C48Volume value at maximum voltage in the curve (no evaluation in CAL and
TIP).
C49Volume value at minimum voltage in the curve (no evaluation in CAL and
TIP).
C51...C59Fix EP for DET and MET.
C61...C69pK/HNP values for DET and MET.
C70...C79Temporary variables for calculations in TIP.
785 DMP Titrino
39
2.7. Result calculations
C-FMLA
Input method specific operands C01...C19, key <C-FMLA>
C
1
With <C-FMLA> the operands C01...C19 can be put
in. For the calculation, the operands are used, which
were introduced in the formula.
The inputs method specific and are store in method
memory.
The calculation report can be printed with the key sequence
<PRINT><←/→> (press keys repeatedly until "calc" appears in the display) <ENTER>
40
785 DMP Titrino
2.8 Statistics calculation
DEF
2.8 Statistics calculation
Mean values, absolute and relative standard deviations are calculated.
The <DEF> key is used to allocate results for statistics
calculation.
The entries are specific to the method and are stored in
the method memory.
mean:
Assigns values for statistics calculations.
The display texts of the Titrino are shown to the left.
The values are the default values.
Allocations for statistics calculations
Number n of single values for statistics calculation.
(1...9)
You can perform statistics calculation using up to 9
results (RSX), endpoints (EPX) or variables (CXX). For
MN1, the default value RS1 is entered.
Delete allocation with <CLEAR> + <ENTER>
Each mode has an inquiry group ">statistics" in key
<PARAM>
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
785 DMP Titrino
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n":Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.
41
2.8. Statistics calculation
How do you obtain statistics calculations?
1) Enter the allocations for the statistics calculation, see page 41.
2) Switch on the statistics calculations: Either with <STATISTICS> or set the status under
<PARAM>, "> statistics" to "ON". The "STATISTICS" LED is on. Storing a method in
the method memory, the status of the statistics calculation is retained.
3) Change the number of the individual values n under "mean n", if necessary.
4) Perform at least 2 titrations. The statistics calculation are constantly updated and
printed. The values are printed in the short and full result report.
5) The statistics report can be printed with <PRINT><STATISTICS><ENTER>.
Rules:
• Recalculated results are incorporated in the statistics calculation.
• If a result of a particular titration can not be calculated, no results for this determination
are incorporated in the statistics calculation. However, the sample counter is still
operative, i.e. the statistics calculation start afresh when the number of required
individual determinations has been performed.
• If the statistics are switched off ("statistics" LED no longer on), results are no longer
entered in the statistics table. But the table remains unchanged. When the statistics
are switched on again, you can immediately continue working.
• If you delete results, all results of the determination with index n are removed from the
statistics evaluation.
• On method change, the old statistics table is cleared and the statistics instructions of
the new method followed.
• Old results in the statistics table can be deleted with "delete all" (<PARAM>,
">statistics", "res.tab:").
42
785 DMP Titrino
2.9 Common variables
2.9 Common variables
Common variables are used for:
• Determination of a titer with a method. This titer is stored permanently as C3X. The
operand C3X can be used in various other methods like any other operand.
• Determination of a blank values with a method . Using this blank value in various other
methods.
• Determination of a result with method. Reconciliation of this result in various other
methods.
You may view the values of the common variables with <CONFIG>.
With <DEF>, results (RSX), endpoints (EPX), variables
(CXX) or mean values (MNX) can be allocated as
common variables.
The entries are specific to the method and are stored in
the method memory.
common variables:
Assigns values as common variables.
The display texts of the Titrino are shown to the left.
The values are the default values.
Allocation for common variables
Common variable C30...C39 (RSX, EPX, CXX, MNX)
Results (RSX), endpoints (EPX), variables (CXX), and
means (MNX) can be assigned.
The values of the common variables remain in force for
all methods until they are overwritten or deleted. They
can be viewed under the <CONFIG> key.
Delete allocation with <CLEAR>+<ENTER>.
785 DMP Titrino
43
2.10. Data output
DEF
2.10 Data output
2.10.1 Reports for the output at the end of a determination
With <DEF>, the report sequence at the end of the
determination is defined.
The entries are specific to the method and are stored in
the method memory.
report:
Definition of report blocks to be printed automatically at
the end of the determination.
The display texts of the Titrino are shown to the left.
The values are the default values.
Report sequence for COM1 (input range depends on
the mode:
DET: full, short, mplist, curve, derive, comb, scalc full,
Select a block. If you require more than one report
block, set a ";" as a delimiter between the blocks.
report COM1:full;curve
Identical for COM2.
Meaning of the report blocks:
fullFull result report with raw results, calculations and statistics.
shortShort result report with calculations and statistics.
mplistMeasuring point list.
curveTitration curve (with DET and MET) or volume vs. time (with SET) or
measured value vs. time (with MEAS)
derive1st derivative of the titration curve (with DET)
combCombined titration curve and 1st derivative (with DET)
scalc fullFull report of silo calculations.
scalc srtShort report of silo calculations.
paramParameter report.
calcReport with formulas and operands.
calibCalibration data.
ffForm feed on printer.
Original reports which are put out automatically at the end of the titration can be printed
with recalculated values at any time. Key sequence:
<PRINT><REPORTS><ENTER>.
44
785 DMP Titrino
2.10 Data output
Original reports have double dashes ==== at the end, whereas recalculations are
marked by single dashes ----.
Report outputs can be stopped with <QUIT>.
Example of reports:
Full result report
'fr
785 DMP Titrino 0P1/101 785.0010
date 1999-06-27 time 08:54 14
card label: Appl.785
pH(init) 2.88 DET pH ********
smpl size 0.372 g
EP1 2.083 ml 7.64
RS1 3.80 g/l
device label Titr 1 sign:
============
'cu
785 DMP Titrino 0P1/101 785.0010
date 1999-06-27 time 08:54 14
start V 0.000 ml DET pH ********
1.0 ml/div dpH=1.0/div
Card label if a card was inserted, see page
56.
Device label (if there is a designation, see
page 10) and manual signature
Titration curve
Scaling of volume and pH axis
============
785 DMP Titrino
45
2.10. Data output
Additional possibilities for report outputs
In addition to the reports which are printed at the end of the titration, various other reports
can be put out. There are 2 possibilities to select the reports:
1) <PRINT><←/→><ENTER>Cursor is pressed repeatedly until the desired report
appears in the display.
2) <PRINT><keyX><ENTER>key X is the key under which the appropriate data
are entered.
List of the "keys X":
Report<Key X>
Configuration reportCONFIG
Parameter reportPARAM
Current sample dataSMPL DATA
Statistics report with the individual resultsSTATISTICS
All sample data from the silo memorySILO
Calibration dataCAL.DATA
Content of the current card directoryCARD
Operands C01...C19C-FMLA
Contents of the <DEF> keyDEF
Contents of the method memory with details of the memory
USER METH
requirements of the individual methods and the remaining bytes
Complete report sequence of the last determination, as defined
REPORTS
under the <DEF> key in the method
Result display without printer
If you work without printer, we recommend to work with the standard character set for
result display (settings in key <CONFIG>, ">auxiliaries", see page 10). You will get the
complete information on your determinations: Calculated results, endpoints, messages
etc.
46
785 DMP Titrino
2.10.2 Display of the titration curve
After the titration, the curve can be viewed.
Switch between "curve" and "result display" with key <CURVE>.
You can trace the curve with keys <↑> and <↓>. In
the text field to the left of the curve the index of the
current measured value is displayed in the first line. In
the subsequent lines, the corresponding measured
values are shown.
If you place the cursor in DET and MET curves onto the
EP, the data of this EP are also shown.
• The EP number and its ERC (Endpoint Recognition
Criterion).
• The EP volume.
2.10 Data output
785 DMP Titrino
47
2.11. User name, key <USER>
USER
2.11 User name, key <USER>
+
←
user
name: Boss
>delete
The key <USER> manages the user names.
User names can be entered directly or selected with
the keys <←> and <→>.
Name:
Selection or input of user name.
Delete:
Delete user name.
The display texts of the Titrino are shown below at the
left.
name:User name (up to 10 ASCII characters)
User names can be entered directly or selected with
the keys <←> and <→>.
The operator name is printed out in the report.
The operator name remains in the instrument until it is
deleted (or until the RAM is initialized).
If no operator name is to be printed out the operator
"blank" can be selected.
>delete
name:
Delete user name
Enter the name directly or select it with the keys <←>
and <→>. <ENTER> will delete the name from the
list of user names.
48
785 DMP Titrino
2.12 TIP, Titration procedure
2.12 TIP, Titration procedure
TIP (Titration Procedure) is used to link several commands in a sequence.
TIP is selected with <MODE> and <ENTER>. TIP is an "empty shell" in which the
sequence of the determination must be defined.
of TIP.
The display texts of the Titrino are shown to the left.
The values are the default values.
Sequence
Select a step with keys <←> and <→>:
method:Method from the user memory or from
the card.
pause:Waiting time
L4, L6 output: Set an output.
info:Hold sequence and write a message
into the display.
stirrerON/OFF
Confirm the step with <ENTER> and enter the
parameter for the selected step, see below.
The request for the second step follows etc. Up to 30
steps can be selected.
On completion of the sequence definition, exit the
inquiry with <QUIT>.
785 DMP Titrino
49
2.12. TIP, Titration procedure
Information for the commands (steps):
CommandMeaningInput range
methodMethod from the user memory or from the card. This
method runs as a submethod.
pauseWaiting time. The waiting time can be aborted with
<QUIT>.
<CLEAR> sets "inf" (= infinitely long pause time).
L4, L6 outputSet L4 output (pin 3) resp. L6 output (pin 1) of the
remote socket.
active = 0 V, inactive = 5 V, pulse > 100 ms, off =
output is not used.
Cable Titrino (L6) - Dosimat: 6.2139.000.
Important: A pulse (e.g. a pulse from monitoring or an
activate pulse in a submethod) can set an active output
to inactive! At the end of the TIP method, the outputs
are set to "inactive".
infoMessage in the display .
The TIP sequence is held and the message displayed.
Continue the sequence with <START>, <QUIT> or
<ENTER>.
stirrerSwitching stirrer ON/OFF.
In TIP the stirrer is not switched automatically in the
submethods. At the end of TIP the stirrer is switched off
(if stirrer control is ON).
Name
0...999 999 s,
inf.
active,
inactive,
pulse, off
up to 16
characters
ON, OFF
The parameters of the sequence can be viewed and changed at any time under the
<PARAM> key.
50
785 DMP Titrino
2.12 TIP, Titration procedure
<START>
(Start delay)
Switching ON stirrer
Set output
TIP calculations
Pause
Info
C7X
TIP data output
Sequence of TIP
As there is no preset sequence of TIP, in what follows the procedure is illustrated by a
sequence that contains all available commands.
The start delay time is waited off.
(Request ident.)
(Request smpl size)
Submethod
Calculations
Data output
The sample identifications and the sample size
are requested.
Stirrer is switched on. The stirrer is not operated
automatically in submethods of TIP. At the end of
TIP, however, the stirrer is switched off (with stirrer control ON).
Submethods are processed according to their
parameters. They run to completion, including
calculations and data output (e.g. curves). The
determination data of submethods are overwritten in the next sequence step of TIP. Those values which are needed for higher-level calculations must therefore be assigned to temporary
variables C7X.
The pause is waited off.
Outputs on the socket "Remote" can be set.
A message can be written in the display. The se-
quence is held until it is continued manually (with
<START>, <QUIT> or <ENTER>).
785 DMP Titrino
TIP higher-level calculations are carried out.
Data output in TIP (TIP contains no more deter-
mination data, i.e. curves must be put out within
the submethods).
51
2.12. TIP, Titration procedure
DEF
Preparation of submethods for use in TIP
All titration data, i.e. curves and lists of measured points must be put out in the
submethod as they are overwritten on return to TIP.
Individual values from the submethod, e.g. endpoints or calculated results must be stored
as temporary variables C7X. This allows them to be used in TIP for further calculations.
Reevaluations of data of a submethod are not possible in TIP. The submethods should
thus be thoroughly wet tested before they are used in TIP.
Assignment of temporary variables in the submethod:
With key <DEF> temporary variables can be
assigned.
The entries are specific to the method and are stored in
the method memory.
temporary variables
for higher-level calculations.
The display texts of the Titrino are shown to the left.
The values are the default values.
Temporary variables
Assignment of result, endpoints or variables (RSX, EPX,
CXX)
Values of the submethod to be used in TIP
calculations.
In TIP calculations variables C7X from different submethods can be used, formula entry
see page 38.
Note:
We recommend to execute the calculations in TIP, as they can only be recalculated "dryly"
in TIP itself, e.g. with a different sample size.
52
785 DMP Titrino
2.13 Method memory, keys <USER METH> and <CARD>
USER METH
2.13 Method memory, keys <USER METH> and
<CARD>
2.13.1 Key <USER METH>
user methods
>recall method
>store method
>delete method
>recall method
method name:
)
3
Management of the internal method memory with key
<USER METH>.
Select method name with keys <←> and <→>.
recall method:
Loads a method from the internal method memory into
the working memory.
store method:
Stores the method which is in the working memory in
the internal method memory.
delete method:
Deletes a method from the internal method memory.
The display texts of the Titrino are shown to the left.
The values are the default values.
Recall method
Recall method from the internal method memory to the
working memory (input of method name, which is
included in the memory).
If a method identification is entered which is not found
in the method memory, the selected value blinks.
>store method
method name:
>delete method
method name:
785 DMP Titrino
Store method
Store method from the working memory to the internal
method memory (up to 8 ASCII characters).
If a method with an identical name is already stored,
you are requested if you wish to overwrite the old
method. With <ENTER> it is overwritten, with
<QUIT> you return to the entry.
Delete method
Delete method from the internal method memory (input
of method name, which is included in the memory).
For safety, you are again asked if you really wish to
delete the method. With <ENTER> it is deleted, with
<QUIT> you return to the working memory.
If a method name is entered which is not found in the
method memory, the selected value blinks.
53
2.13. Method memory, keys <USER METH> and <CARD>
The contents of the method memory can be printed with the key sequence
<PRINT><USER METH><ENTER>
Document your methods (e.g. parameter report, def. report and C-fmla report)!
With a PC and the 6.6008.XXX Vesuv program, you should carry out a complete method
backup from time to time.
Making the backup on the card, pay attention of the exchange date of the battery!
Management of the method memory on the card with
key <CARD >.
Select names with keys <←> and <→>.
On the first line you find the name of the current
directory (here user meth.).
recall method:
Loads a method from the current directory of the card
into the working memory.
store method:
Stores the method, which is in the working memory, in
the current directory of the card.
delete method:
Deletes a method from the current directory of the
card.
change directory:
Changes the current directory of the card.
create directory:
Creates a new directory on the card.
delete directory:
Deletes a directory on the card.
backup:
Backup of the internal method memory on the card.
reload:
Reloads a backup from the card into the internal
method memory.
format:
Formats the card.
change battery:
Date for changing the battery.
>recall method
>store method
>delete method
>change directory
>create directory
>delete directory
dir.name:
785 DMP Titrino
The display texts of the Titrino are shown to the left.
The values are the default values.
Recall, store, delete method
Identical with the corresponding function of the internal
method memory. The functions operate on the current
directory of the card.
Change, create, delete directory
If a directory is deleted, it is deleted together with all its
methods.
(directory name with up to 10 characters)
55
2.13. Method memory, keys <USER METH> and <CARD>
>backup
dir.name:
>reload
dir.name:
>format
card label
Backup of the internal method memory on the
card
(directory name with up to 10 characters)
Existing methods in this directory will be deleted, and
all methods from the internal method memory are
stored in the directory.
Reload methods from the card to the internal
method memory
(directory name with up to 10 characters)
Primary all methods from the internal method memory
are deleted. Afterwards all methods from the directory
of the card are copied into the internal method
memory.
Format the card
Card label (with up to 8 characters)
The card label will be printed in the card report and in
the full result report (if the card is inserted, see page
45).
format: no
Confirmation of formatting the card (yes, no)
Upon formatting the card, all data on the card are
deleted.
After formatting a new card, the date for battery change
must be entered.
>change battery
date
Battery change date
Date for changing the battery (YYYY-MM-DD)
see page 57.
Possibilities of the card
Methods can be stored on the card.
• Simple exchange of methods between different users, laboratories, factories. Every
user can use his card with his own methods. The user specific card label will be
printed in the full result report, if the card remains inserted.
• Methods can be stored in different directories, e.g. in directories depending on the
sample type or the user.
• The card can be used as an extended memory.
• You will find many application methods on the application card 6.6036.000.
56
785 DMP Titrino
2.13 Method memory, keys <USER METH> and <CARD>
Internal method memory and methods on the card
If methods are called from TIP or from the silo memory, the Titrino will search these
methods primary in the internal method memory, and afterwards in the current directory
of the card.
Important: We do not recommend not to store identical methods in the current directory
of the card as well as in the internal method memory at the same time. Otherwise, you
have to update both methods!
Card battery
The card is supplied by a battery. This battery must be changed periodically. Note the
lifetime of the battery which is indicated in the manual of the card.
If the card is inserted in the Titrino you will receive a warning if the battery is low on
switching on the Titrino or if the battery change date is expired.
The battery is in a case of the card. Read the note enclosed with the card before
changing the battery. Leave the card in the Titrino when you change the battery to ensure
sufficient power supply.
Important: The battery lifetime refers to a storing temperature of 25 °C. With higher
temperatures, the lifetime is shorter.
Therefore do not
• carry the card on the body
• store the card near radiators
• have the card in the sun.
Write protection
The write protection prevents all functions, which write on the card (store methods, delete
methods, change directory - the current directory is written on the card, create directory,
delete directory, backup, format the card, change date of the card battery). Reading
functions are possible.
Write protection is on, if the lash in front of the card is on the right.
Reports
• Methods of the current directory, key sequence <PRINT><CARD><ENTER>.
• Methods of the whole card: key sequence <PRINT><←/→><ENTER>; press the
cursor keys repeatedly until "card" appears in the display.
Ordering designations
Memory card with 128 Kbytes memory space............................................ 6.2245.010
Application card with application file...........................................................6.6036.000
785 DMP Titrino
57
2.14. Calibration data, key <CAL.DATA>
CAL.DATA
2.14 Calibration data, key <CAL.DATA>
*
5
cal.data
>input 1
>input 2
>input diff.
>input 1
pH(as) 7.00
slope 1.000
temp 25.0 °C
With <CAL.DATA>, the current pH calibration data of
all measuring inputs can be seen. Calibration data are
entered here automatically on completion of a
calibration.
input 1:
Calibration data for measuring input 1.
Identical for input 2 and diff.
The display texts of the Titrino are shown to the left.
The values are the default values.
pH calibration data for measuring input 1
Asymmetry pH (0...±20.00)
Entered automatically after a calibration with
measuring input 1.
Slope (0...±9.999)
Entered automatically after a calibration with
measuring input 1.
Calibration temperature (-20.0...120.0 °C)
Will be printed automatically after calibration with
measuring input 1.
cal.date
Date of last calibration (no entry possible)
If the calibration data "pH(as)" and/or "slope" are
changed by a manual entry, the date entry is deleted.
This date is the reference date for calibration
monitoring, see page 7.
electr.id
Electrode identification of calibrated electrodes (no
entry possible)
If an electrode identification has been entered in the
CAL mode, it is automatically entered after the
calibration.
The calibration report with the current measuring input data can be printed at any time
with the key sequence
<PRINT><CAL.DATA><ENTER>.
58
785 DMP Titrino
2.15 Current sample data, key <SMPL DATA>
SMPL
2.15 Current sample data, key <SMPL DATA>
The key <SMPL DATA> can be used to enter the
DATA
smpl data
id#1 or C21
id#2 or C22
id#3 or C23
smpl size 1.0 g
smpl unit: g
current sample data. The contents of this key change
when the silo memory is switched on, see page 61.
Instead of entering the current sample data with
<SMPL DATA>, you can request these data
automatically after start of determinations.
Configuration: <PARAM>, ">preselections".
Current sample data can be entered live. For working
with the silo memory see page 60.
id#1...3 or C21...C23, sample identifications:
The sample identifications can also be used as sample
specific calculation variables C21...C23.
smpl size:
Sample size.
The sample size can be monitored, see e.g. page 18.
The limits are then displayed in this window.
smpl unit:
Unit of the sample size.
smpl data
id#1 or C21
id#2 or C22
id#3 or C23
smpl size 1.0 g
smpl unit: g
The display texts of the Titrino are shown to the left.
The values are the default values.
Sample data
Sample identification 1...3 or sample specific operand
C21...C23 (up to 8 characters).
Sample identifications or sample specific operands
can be entered using the keypad, via a balance with a
special input device or via barcode reader.
Sample size (6-digit number: ±X.XXXXX)
Entry using keypad, via balance or via barcode reader.
Unit of sample size (g, mg, ml, ul, pc, no unit or up to 5
characters)
Select unit with <←/→>.
785 DMP Titrino
59
2.16. Silo memory for sample data
2.16 Silo memory for sample data
In the silo memory or pushup storage, sample data (method, identifications and smpl
size) can be stored. This is useful, e.g. when you work with Sample Changers and other
automatic sample addition systems or if you wish an overview of your determination
results, see page 64.
SILO
9
-
Press the key <SILO> for working with the silo
memory. The status LED "silo" is on when the silo
memory is switched on. The silo memory works by the
FIFO principle (First In, First Out).
If the silo memory is switched on, sample data are routed to the last free line of the silo
memory. If no new value is put in, the value from the last line is automatically copied. In
this manner, data can be simply taken over when they remain unchanged.
When the instrument is started, the sample data are fetched from the next silo line.
Organization of the silo memory
255
1
2
255
1
2
6
35
Silo memory contains 35 lines.
Next free line is 36
6 of the 35 lines have been
processed. Free lines from 36 to
255 and from 1 to 6.
1 silo line needs between 18 and 120 bytes memory capacity.
Filling the silo memory with a connected balance
If the silo memory is filled from the balance, you must ensure that there is sufficient space
in the silo memory for the required number of silo lines! The number of free bytes is given
in the user memory report.
When the sample data are entered from a balance, the transfer of the sample size is
taken as the end of the silo line. You should not send data from the balance and edit the
silo memory at the same time.
For mixed operation, manual input of identifications and sample sizes from a balance, the
values from the balance are sent into the line in which editing just takes place. Confirmed
the data with <ENTER> at the Titrino.
60
35
785 DMP Titrino
2.16 Silo memory for sample data
SMPL
Key <SMPL DATA> with the silo memory switched on
Sample data can be entered into the silo memory with
key <SMPL DATA>.
DATA
edit silo lines:
smpl data
>edit silo lines
>delete silo lines
>delete all silo lines
cycle lines: OFF
save lines: OFF
Entering sample data into the silo memory.
delete silo lines:
Deletes single silo lines.
delete all silo lines:
Deletes the whole silo memory.
The display texts of the Titrino are shown to the left.
The values are the default values.
>edit silo lines
silo line 1
method:
id#1 or C21
id#2 or C22
id#3 or C23
smpl size 1.0 g
smpl unit: g
Input for silo memory
Silo line (1...255)
The next free line is displayed automatically. Lines
already occupied can be corrected.
Method with which the sample is processed (method
name from the method memory)
If no method name has been entered, the sample is
processed with the method in the working memory.
Selection of the method with <←/→>.
Sample identification 1..3 or sample specific calculation
variables C21...C23 (up to 8 characters)
Sample size (6-digit number: ±X.XXXXX)
The method specific limits are tested on result
calculation.
Unit of sample size (g, mg, ml, ul, pc, no unit or up to 5
characters)
Select unit with <←/→>.
>delete silo lines
delete line n OFF
785 DMP Titrino
Delete individual silo lines
Line number of the line to be deleted(1...255, OFF)
<CLEAR> sets "OFF".
Deleted lines remain in the silo memory. Access is
blocked during the processing. To show that a line has
been deleted, they are marked with "*". The symbol *
indicates that the line has been deleted.
Deleted lines can be reactivated if the appropriate line
is re-edited.
61
2.16. Silo memory for sample data
>delete all silo lines
delete all: no
cycle lines: OFF
save lines: OFF
Delete all silo lines
Confirmation (yes, no)
When all silo lines are deleted, the silo is completely
empty: The line numbering starts again with 1.
With "ON", worked off silo lines will be copied to the
highest line of the silo memory (ON, OFF)
Data cycling "on" is useful if you constantly have to
process the same sample data. In such a case, the
processed silo line is not deleted, but copied to the
next free line, see below. If you work in this mode, you
should not enter any new silo lines during the
determinations.
Store results in the silo memory (ON, OFF)
Determination results will be stored as C24 or C25 in
the silo memory according to the allocations in the
methods, see page 64.
"save lines" can only be set to "OFF" if the silo is
completely empty.
Silo memory with data cycling "on"
1
255
2
35
Silo memory contains 35 lines.
Next free line is 36.
1
255
2
6
35
41
6 of 35 lines have been processed.
The processed lines have been
copied to the end of the silo
memory: your silo is filled up to line
41.
62
785 DMP Titrino
2.17 Storing determination results and silo calculations
DEF
2.17 Storing determination results and silo calculations
2.17.1 Storing determination results
If the sample-specific data of the silo memory should be kept after the determination and
supplemented by results, the following entries are necessary:
1. In the method under <DEF>
Assignment of the determination results to C24 and/or C25:
2. In the silo memory, <SMPL DATA> (when the silo memory is switched in):
"save lines: on"
Assignment of determination results
(
The determination results are assigned in key <DEF>.
2
The display texts of the Titrino are shown to the left.
Important:
Ensure that there is still sufficient space for storing the results C24 and C25. (In the report
<PRINT><USER METH><ENTER> the number of free bytes is shown.) Result name,
value and unit are stored. The memory requirements can be estimated as follows:
Result with text (8 characters) and unit (5 characters):32 bytes
Measured value C40, value without unit:22 bytes
The values are the default values.
Silo calculations
Assignment to C24 (RSX, EPX, CXX)
Calculated results (RSX), endpoints (EPX) or variables
CXX can be stored as C24.
Same procedure for C25.
785 DMP Titrino
63
2.17. Storing determination results and silo calculations
After several samples have been processed, the silo memory report can have the
following appearance (printout with <PRINT><SILO><ENTER>):
'si
785 DMP Titrino 0P1/101 785.0010
date 1999-06-27 time 08:54 14
>silo
cycle lines: OFF
save lines: ON
sl method id 1/C21 id 2/C22 id 3/C23 C00 C24 C25
+ 1 11-2 A/12 94-09-12 0.233g 0.142ml/min 98.53%
+ 2 11-2 A/13 94-09-12 0.286g 0.138ml/min 95.75%
/ 3 11-2 A/14 94-09-12 0.197g 0.145ml/min 100.61%
4 11-2 A/15 94-09-12 0.288g NV NV
5 11-2 A/16 94-09-12 0.263g NV NV
processed silo
lines with
saved results
The silo lines can be marked as follows (at very left of report):
+Silo line has been processed. It cannot be edited anymore.
*A silo line not yet processed has been deleted.
-A processed silo line has been deleted and hence removed from the silo
calculations.
/The last processed silo line. Recalculation will be considered e.g., if the sample data
of this line are changed.
No marking: The silo line is awaiting processing.
For silo lines ≥100, the first digit will be overwritten by the marking.
2.17.2 Silo calculations
Mean value and standard deviation of the results available in the silo memory can
subsequently be calculated over the entire series.
The following details can be entered in the method under <DEF>:
>silo calculations
C24=
C25=
match id: OFF
Silo calculations
Assignment to C24 and C25
Calculated results (RSX), endpoints (EPX) and
variables (CXX) can be stored as C24.
Identical for C25.
Which sample identifications must match in order to
combine of the results (id1, id1&2, all, OFF)
"OFF" means no matching ids, all samples which have
been processed with the same method are combined,
see examples below.
64
785 DMP Titrino
2.17 Storing determination results and silo calculations
Starting from the following silo report:
'si
785 DMP Titrino 0P1/101 785.0010
date 1999-06-27 time 08:54 14
>silo
cycle lines: OFF
save lines: ON
sl method id 1/C21 id 2/C22 id 3/C23 C00 C24 C25
+ 1 11-2 A/12 94-09-12 0.233g 0.142ml/min 98.53%
+ 2 0-15 A/13 94-09-12 0.286g 0.9976 NV
+ 3 0-15 A/13 94-09-12 0.197g 0.9947 NV
+ 4 11-2 A/12 94-09-12 0.288g 0.138ml/min 95.75%
/ 5 11-2 A/15 94-09-12 0.263g 0.145ml/min 100.61%
*
Assignment
for C24 only
*
*
with "match id: off" the following silo calculation report (scalc full) is obtained:
:
method id 1/C21 id 2/C22 id 3/C23 mean +/-s n
11-2 * * * Rate 0.142ml/min 0.0035 3
Content 98.30% 2.438 3
0-15 * * * Titer 0.9962 0.00105 2
All samples
which have been
processed with
the same method are combined
With "match id: id1" the following silo calculation report (scalc full) is obtained:
:
method id 1/C21 id 2/C22 id 3/C23 mean +/-s n
11-2 A/12 * * Rate 0.140ml/min 0.0028 2
Content 97.14% 1.966 2
0-15 A/13 * * Titer 0.9962 0.00105 2
11-2 A/15 * * Rate 0.145ml/min 0.000 1
Content 100.61% 0.000 1
Sample processed with the
same method
and having the
same id1 are
combined
The short silo calculation report contains only calculations for the current sample.
:
method id 1/C21 id 2/C22 id 3/C23 mean +/-s n
11-2 A/15 * * Rate 0.145ml/min 0.000 1
Content 100.61% 0.000 1
The mean values of the silo calculations are available for further result calculations as C26
and C27 and can be used in the Titrino in formulas.
Mean value of C24 ⇒ C26
Mean value of C25 ⇒ C27
Important:
• If work is performed with silo calculations, the method name must be entered in the
silo memory.
• Results will be overwritten in the silo recalculation, as long as the silo line is marked
with "/". If you do not wish such an input, e.g. because you work off an urgent sample
between a series, disconnect the silo.
785 DMP Titrino
65
2.17. Storing determination results and silo calculations
• Calculations and assignments are carried out in the following order:
1. Calculation of the results RSX
2. Assignment of temporary variables C7X for TIP
3. Calculation of means MNX
4. Assignment of silo results C24 and C25
5. Silo calculations
6. Assignment of means C26 and C27 from silo calculations
7. Assignment of common variables C3X
66
785 DMP Titrino
3 Operation via RS232 Interface
3.1 General rules
The Titrino has an extensive remote control facility that allows full control of
the Titrino via the RS 232 interface, i.e. the Titrino can receive data from an
external controller or send data to an external controller. CR and LF are used
as terminators for the data transfer. The Titrino sends 2xCR and LF as termination of a data block, to differentiate between a data line which has CR and
LF as terminators. The controller terminates its commands with CR and LF. If
more than one command per line is sent by the controller, “;” is used as a
separator between the individual commands.
The data are grouped logically and easy to understand. Thus e.g., for the selection of the dialog language, the following must be sent
&Config.Aux.Language "english"
3.1 General rules
whereby it is sufficient to only transmit the boldface characters, thus:
&C.A.L "english"
The quantities of the commands above are:
Configconfiguration data
Auxauxiliaries, various data
Languagesetting the dialog language
The data are hierarchically structured (tree form). The quantities that occur in
this tree are called objects in the following. The dialog language is an object
which can be called up with the
&Config.Aux.Language
command.
If one is in the desired location in the tree, the value of the object can be queried.
&Config.Aux.Language $Q Q means Query
The query command $Q initiates the issuing of the value on the instrument
and the value emission is triggered. Entries which start with $, trigger something. They are thus called triggers.
785 DMP Titrino
Values of objects can not only be queried, they can also be modified. Values
are always entered in quotes, for example:
&Config.Aux.Language "english"
67
3.1. General rules
3.1.1 Call up of objects
An excerpt from the object tree is represented below:
3rd node
2nd node
1st node
0st node
LanguageProg
Aux
Config
RSSet
&
Root
RulesExample
The root of the tree is designated by &.
The branches (levels) of a tree are marked with
a dot (.) when calling up an object.
When calling up an object, it is sufficient to give
only as many letters as necessary to uniquely
assign the object. If the call is not unequivocal,
Calling up the dialog language
&Config.Aux.Language or &C.A.L
the first object in the series will be recognized.
Mode
Upper- or lowercase letters may be used.&C.A.L or &c.a.l
An object can be assigned a value. Values are
signified at the beginning and end by quotes (").
Entering the dialog language:
&C.A.L"english"
They may contain up to 24 ASCII characters.
Numerical values can contain up to 6 digits, a
negative sign, and a decimal point. Numbers
correct entry of numbers:
"0.1"
with more than 6 characters are not accepted;
more than 4 decimal places are rounded off. For
numbers <1, it is necessary to enter leading
incorrect entry of numbers
"1,5" or "+3" or ".1"
zeros.
The current object remains until a new object is
called.
New objects can be addressed relative to the
old object:
entry of another dialog language:
"deutsch"
From the root to node 'Aux': &C.A
Forward from node 'Aux' to 'Prog': .P
A preceding dot leads forwards to the next level
in the tree.
More than one preceding dot leads one level
backwards in the tree. n node backwards re-
Jump from node 'Prog' to node 'Aux' and select
a new object 'Language' at this level: ..L
quire n+1 preceding dots.
If you must jump back to the root, enter a preceding &.
Change from node 'Language' via the root to
node 'Mode': &M
68
785 DMP Titrino
3.1 General rules
3.1.2 Triggers
Triggers initiate an action on the Titrino, for example, starting a process or
sending data. Triggers are marked by the introductory symbol $.
The following triggers are possible:
$GGoStarts processes, for ex. starting the mode run or setting the RS 232 interface
parameters
$SStopStops processes
$HHoldHolds processes
$CContinueContinues processes after Hold
$QQueryQueries all information from the current node in the tree forward up to and
including the values
$Q.PPathQueries the path from the root of the tree up to the current node
$Q.HHighestQueries the number of son nodes of the current node
Index
$Q.N"i"NameQueries the name of the son node with index i, i = 1 - n
$DDetail-InfoQueries the detailed status information
$UqUitAborts the data flow of the instrument, for example, after $Q
The triggers $G and $S are linked to particular objects, see the summary table
page 77ff.
All other triggers can be used at any time and at all locations on the object
tree.
Examples:
Querying the value of the baud rate: &Config.RSSet.Baud $Q
Querying all values of the node "RSSet": &Config.RSSet $Q
Querying the path of the node "RSSet": &Config.RSSet $Q.P
Start mode: &Mode $G
Querying the detailed status: $D
785 DMP Titrino
69
3.1. General rules
3.1.3 Status messages
In order to have an efficient control by an external control device, it must also
be possible to query status conditions; they provide information on the status
of the Titrino. The trigger $D initiates output of the status. Status messages
consist of the global status, the detailed status and eventual error messages,
e.g. $S.Mode.SET;E26. The global status informs on the activity of the process, while the detailed status conditions show the exact activity within the
process.
The following global status conditions are possible:
$GGo:The Titrino is executing the last command.
$HHold:The Titrino has been held ($H, key <meas/hold> or by an error which
effects the hold status)
$CContinue:The Titrino has been restarted actively after hold
$RReady:The Titrino has executed the last command and is ready
$SStop:A process has been aborted in an "unnatural manner". e.g. stopped or aborted
because there was an error.
Detailed status conditions
Status conditions of the global $G:
$G .Mode.DET .Inac:Instrument at the beginning or at the end of a titration.
.Req .Id1:Instrument in the DET mode, requesting Id1 after titration start.
.Id2:Instrument in the DET mode, requesting Id2 after titration start.
.Id3:Instrument in the DET mode, requesting Id3 after titration start.
.Smpl: Instrument in the DET mode, requesting sample size after titration start.
.Unit: Instrument in the DET mode, requesting unit of sample size after titration
start.
.Start:Instrument in the DET mode, processing the start conditions.
.Titr:Instrument in the DET mode, titrating.
$G .Mode.MET...As DET.
$G .Mode.SET .Inac:Instrument at the beginning or at the end of a titration.
.Req .Id1:Instrument in the SET mode, requesting Id1 after start.
.Id2:Instrument in the SET mode, requesting Id2 after start.
.Id3:Instrument in the SET mode, requesting Id3 after start.
.Smpl: Instrument in the SET mode, requesting sample size after start.
.Unit: Instrument in the SET mode, requesting unit of sample size after start.
.Start:Instrument in the SET mode, processing the start conditions.
.SET1:Instrument in the SET mode, titrating to the first endpoint.
.SET2:Instrument in the SET mode, titrating to the second endpoint.
.Cond.Ok:Instrument in the SET, conditioning, endpoint reached (after the first
startup from the standby mode).
.Cond.Prog: Instrument in the SET mode, conditioning, endpoint not reached
(Conditioning progressing).
$G .Mode.MEAS .Inac:Instrument at the beginning or at the end of a titration.
.Req .Id1:Instrument in the MEAS mode, requesting Id1 after start.
.Id2:Instrument in the MEAS mode, requesting Id2 after start.
.Id3:Instrument in the MEAS mode, requesting Id3 after start.
.Smpl: Instrument in the MEAS mode, requesting sample size after start.
.Unit: Instrument in the MEAS mode, requesting unit of sample size after start.
.Meas:Instrument in the MEAS mode, measuring.
$G .Mode.CAL .Inac:Instrument at the beginning or at the end of a calibration
.Req.Temp:Instrument in the CAL mode, requesting calibration temperature.
.Meas.Temp: Instrument in the CAL mode, measuring calibration temperature.
.Req.Buf1:Instrument in the CAL mode, requesting pH of buffer 1.
70
785 DMP Titrino
.Meas.Buf1: Instrument in the CAL mode, measures buffer 1.
.Req.Buf2:Instrument in the CAL mode, requesting pH of buffer 2.
.Meas.Buf2: Instrument in the CAL mode, measures buffer 2.
etc.
$G .Assembly.Bur .Fill: Buret in filling process
.ModeDis: Buret in DIS mode
In TIP, its global status as well as the step number (X) is available.
$G .TIP.X.Inac:Instrument at the beginning or at the end of a TIP.
.Req .Id1:Instrument in the TIP mode, requesting Id1 after start.
.Id2:Instrument in the TIP mode, requesting Id2 after start.
.Id3:Instrument in the TIP mode, requesting Id3 after start.
.Smpl: Instrument in the TIP mode, requesting sample size after start.
.Unit: Instrument in the TIP mode, requesting unit of sample size after start.
.Pause:Instrument in the TIP mode, in pause.
.Info:Instrument in the TIP mode, in info.
.Mode...:Instrument in the TIP mode, working off a submethod. The detailed
status messages of the submethod appear, see above.
Status conditions of the global $H:
The status message of the action which has been held appears.
If the process is held because a monitored limit has been violated, its status
message is $H.Mode.XXX.Titr.
3.1 General rules
Status conditions of the global $C:
The status conditions of the global $C are identical with the ones of the global
status $G. They appear when the process has been restarted actively from
the status "Hold" ($C, key <meas/hold> or automatically after elimination of
an error).
785 DMP Titrino
71
3.1. General rules
Status conditions of the global $R:
$R .Mode.XXXX.QuickMeas: Quick manual measurement from the initial status in mode XXXX.
$R .Mode.DET .Inac:Instrument in the DET mode, inactive.
$R .Mode.MET .Inac:Instrument in the MET mode, inactive.
$R .Mode.SET .Inac:Instrument in the SET mode, inactive.
.Cond.Ok:Instrument in the SET mode, conditioning, endpoint reached.
.Cond.Prog: Instrument in the SET mode, conditioning, endpoint not reached.
$R .Mode.MEAS .Inac:Instrument in the MEAS mode, inactive.
$R .Mode.CAL .Inac:Instrument in the CAL mode, inactive.
$R .Assembly.Bur.ModeDis: Buret in the DIS mode, inactive.
$R .TIP.Inac:Instrument in TIP, inactive.
Status conditions of the global $S:
$S .Mode.XXXX.QuickMeas: Quick manual measurement from the initial status in mode XXXX.
The instrument gives the status from which it has been stopped. The detailed
status information is therefore identical to for the global status $G.
Violation of monitored limits with action "end" give the status message
$S.Mode.XXX.Inac;EYYY.
72
785 DMP Titrino
3.1 General rules
3.1.4 Error messages
Error messages are added to the status messages and separated from them
by the sign ";".
E8Card read/write error.
Exit: Send new command.
E9Wrong card, a card has been removed/inserted during the inquiry.
Exit: Send new command.
E10The card has lost data.
Exit: Send new command.
E18Card battery low (it is between 2.37...2.64 V).
Exit: Send new command.
E20Check exchange unit.
Exit: Mount Exchange Unit (properly) or &m $S.
E21Check electrode, short circuit.
Exit: Rectify fault or &m $S.
E22Check electrode, break.
Exit: Rectify fault or &m $S.
E23Division by zero.
Exit: The error message disappears on next startup or on recalculation.
E24Check drive unit.
Exit: Connect drive unit (correctly) or &m $S.
E26Manual stop.
Exit: The error message disappears on next startup.
E27Stop V reached in SET.
Exit: The error message disappears on next startup.
E28Wrong object call up
Exit: Send correct path for object. Start path at root.
E29Wrong value or no value allowed.
Exit: Send correct value or call up new object.
E30Wrong trigger, this trigger is not allowed or carrying-out of action not
possible.
Exit: Send correct trigger (exception: $D) or call up new object.
E31Command is not possible in active status. Repeat command in inactive
status.
Exit: Send new command.
E32Command is not possible during titration. Repeat command during the
conditioning phase or in inactive status.
Exit: Send new command.
E33Value has been corrected automatically.
Exit: Send new command.
E34Instrument at the end of the titration and sample data is edited; the instrument
at rest or editing during filling.
Exit: &m $S.
785 DMP Titrino
73
3.1. General rules
RS receive errors:
E36Parity
Exit: <QUIT> and ensure settings of appropriate parameters at
both devices are the same.
E37Framing error
Exit: <QUIT> and ensure settings of appropriate parameters at
both devices are the same.
E38Overrun error. At least 1 character could not be read.
Exit: <QUIT>
E39The internal working-off buffer of the Titrino is full (>82
characters).
Exit: <QUIT>
RS send errors:
E42CTS=OFF No proper handshake for more than 1 s.
Exit: <QUIT> Is the receiver switched on and ready to receive?
E43The transmission of the Titrino has been interrupted with XOFF for
at least 6 s.
Exit: Send XON or <QUIT>
E45The receive buffer of the Titrino contains an incomplete command
(LF missing). Sending from the Titrino is therefore blocked.
Exit: Send LF or <QUIT>.
E120Overrange of the primary measured value (pH, U, Ipol, Upol or T
with MEAS T). The secondary measured value (temperature) may
be instable as well.
Exit: Correct error or &m $S.
E121Measuring point list overflow (more than 500 measuring points).
Exit: The error message disappears on next startup.
E122EP overflow.
Exit: The error message disappears on next startup or on
recalculation.
E123Missing EP for calculation.
Exit: The error message disappears on next startup or on
recalculation.
E124Number of EP does not correspond with the set windows.
Exit: The error message disappears on next startup or on
recalculation.
E125Missing fix EP for calculation, has not been defined.
Exit: The error message disappears on next startup or on
recalculation.
E126Fix-EP outside of measuring point list.
Exit: The error message disappears on next startup or on
recalculation.
E128No new mean.
Exit: The error message disappears on next startup or on
recalculation.
E129No new common variable, old value remains.
Exit: The error message disappears on next startup or on
recalculation.
74
785 DMP Titrino
3.1 General rules
E130Wrong sample. For SET with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
E131No EP set for SET.
Exit: The error message disappears on next startup.
E132Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
E133Silo full.
Exit: Send new command.
E134No method. A method, which is required from the silo memory or in TIP, does
not exist.
Exit: The error message disappears on next startup.
E135Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
E136Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
E137XXX Bytes are missing so that the method, the silo line could not be stored or
not enough RAM for running TIP.
Exit: Send new command.
E155No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
E157No sequence defined in TIP.
Exit. The error message disappears on next start.
E158A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
E160No new temporary variable.
Exit: The error message disappears on next start.
E161Measurement range of the secondary measured value (temperature)
exceeded. The primary measured value (pH, U, Ipol, Upol) can also be
unstable.
Exit: Rectify error or &m $S.
E166Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
E172In TIP, a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
E177Accessing to the memory card, the card was not (properly) inserted.
Exit. The error message disappears on next start.
E178The date of changing the battery of the card is expired.
Exit. The error message disappears on next start.
E180Memory card write-protected.
Exit: Send new command.
E181Memory card not formatted.
Exit: Send new command.
E182Memory card not accessible.
Exit: Send new command.
E183A directory with the same name exists already on the memory card.
Exit: Send new command.
785 DMP Titrino
75
3.1. General rules
E196Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
E197Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
E198Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
E199Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.
E205Calibration interval is expired.
Exit: The error message disappears on next calibration or if you delete the
calibration.
E212Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
E213Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
E214Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.
E270Overload in dosing element.
Exit: <STOP> then <CLEAR>. The dosing element will be initialized. If the
error reappears, check the wet part.
76
785 DMP Titrino
3.2 Remote control commands
3.2 Remote control commands
3.2.1 Overview
The internal object tree can be divided into the following branches:
& Root
à ModeMethod parameters
à UserMethAdministration of the internal user-memory for methods
à MemoryCardAdministration of the memory card
à ConfigInstrument configuration
à SmplDataSample specific data
à HotkeyKeys with direct access
à InfoCurrent Data
à AssemblyComponent data
à SetupSetting the operating mode
À DiagnosisDiagnostics program
à .SelectMode selectionDET,MET,SET,
³MEAS,CAL,TIP3.2.2.3.
à .DETQuantityMeasured quantity for DETpH, U, Ipol, Upolditto
à .METQuantityMeasured quantity for METpH, U, Ipol, Upolditto
à .SETQuantityMeasured quantity for SETpH, U, Ipol, Upolditto
à .MEASQuantityMeasured quantity for MEASpH, U, Ipol, Upol, Tditto
à .NameName of current methodread only/read+write3.2.2.4.
à .Parameter*Parameter of current mode, see below
³
à .DefDefinitions for data output
³ Ã .FormulasCalculation formulas
³ ³ Ã .1for result 1
³ ³ ³ Ã .FormulaCalculation formulaspecial3.2.2.5.
³ ³ ³ Ã .TextRSText for result outputup to 8 ASCII charditto
³ ³ ³ Ã .DecimalNumber of decimal places0...2...5ditto
³ ³ ³ Ã .UnitUnit for result outputup to 6 ASCII charditto
³ ³ ³ Ã .LimitsLimits for resultON, OFFditto
³ ³ ³ Ã .LoLimLower limit0...±999 999ditto
³ ³ ³ Ã .UpLimUpper limit0...±999 999ditto
³ ³ ³ Ã .OutputOutput on L13active, pulse, OFFditto
³ ³ ³ :up to 9 results
³ Ã .SiloCalcSilo calculations
³ ³ Ã .AssignAssignment
³ ³ ³ Ã .C24Store as variable C24RSX,EPX,CXX3.2.2.6.
³ ³ ³ Ã .C25Store as variable C25RSX,EPX,CXX
³ ³ Ã .MatchIdMatching of Id'sid1, id1&2, all, OFF
³ Ã .ComVarAssignment of common variables
³ ³ Ã .C30for C30RSX,EPX,CXX,MNX3.2.2.7.
³ ³ Ã up to C39
³ Ã .ReportReports at the end of determination
³ ³ Ã .Assign1Output to COM1special3.2.2.8.
³ ³ Ã .Assign2Output to COM 2as COM1
³ Ã .MeanAssignment for mean calculation
³ ³ Ã .1MN1
³ ³ ³ Ã .AssignInput of variableRSX, EPX, CXX3.2.2.9.
³ ³ ³ :
³ Ã .TempVarAssignment of temporary variables
³ ³ Ã .C70for C70RSX,EPX,CXX3.2.2.10.
³ ³ Ã up to C79
à .CFmlaCalculation constants
³ Ã .1Calculation constant C01
³ ³ Ã .ValueInput of value0...±999 9993.2.2.11.
³ Ã up to C19
78
785 DMP Titrino
3.2 Remote control commands
*ParameterTree part "Parameters for DET"
à .TitrParaTitration parameters
³ Ã .MptDensityMeasuring point density0...4...93.2.2.12.
³ Ã .MinIncrMinimum increment0...10.0...999.9ditto
³ Ã .DosRateDispensing rate for increments0.01...150.0, max.3.2.2.13.
³ Ã .SignalDriftDrift for meas. value acquisitiondepends on meas.quant. 3.2.2.14.
³ Ã .UnitSigDriftUnit of measured value driftread onlyditto
³ Ã .EquTimeEquilibrium time0...26...9999, OFFditto
³ Ã .StartVStart volume
³ ³ Ã .TypeType of start volumeabs., rel., OFF3.2.2.15.
³ ³ Ã .VVolume for absolute start volume0...999.99ditto
³ ³ Ã .FactorFactor for relative start volume0...±999 999ditto
³ ³ Ã .RateDispensing rate for start volume0.01...150.0, max.ditto
³ Ã .PauseWaiting time0...999 9993.2.2.16.
³ Ã .MeasInputMeasuring input1, 2, diff.3.2.2.17.
³ Ã .IpolPolarization current0...1...±127ditto
³ Ã .UpolPolarization voltage0...400...±1270ditto
³ Ã .PolElectrTestTest for polarized electrodesON, OFFditto
³ Ã .TempTitration temperature-170.0...25.0...500.03.2.2.18.
³
à .StopCondStop conditions
³ Ã .VStopStop volume
³ ³ Ã .TypeType of stop volumeabs., rel., OFF3.2.2.19.
³ ³ Ã .VVolume for absolute stop volume0...99.99...9999.99ditto
³ ³ Ã .FactorFactor for relative stop volume0...±999 999ditto
³ Ã .MeasStopStop measured value pH, U, Idepends on meas.quant. 3.2.2.20.
³ Ã .UnitMStopUnit of stop measured valueread onlyditto
³ Ã .EPStopStop after a number of EP's1...9, OFF3.2.2.21.
³ Ã .FillRateFilling rate0.01...150.0, max.3.2.2.22.
³
à .StatisticsStatistics
³ Ã .StatusStatus of statistics calculationON, OFF3.2.2.23.
³ Ã .MeanNNo. of individual determinations2...20ditto
³ Ã .ResTabResult table
³ ³ Ã .Selectoriginal,delete n,delete allditto
³ ³ Ã .DelNDeletion of individual results1...20ditto
³
à .EvaluationEvaluation
³ Ã .EPCEP criterion0...5...2003.2.2.24.
³ Ã .RecognitionEP recognition
³ ³ Ã .SelectType of EP recognitionall,greatest,last,window,OFFditto
³ ³ Ã .WindowWindow
³ ³ ³ Ã .1up to 9 windows
³ ³ ³ ³ Ã.LowLimLower limit window 1depends on meas.quant. ditto
³ ³ ³ ³ Ã.UpLimUpper limit window 1depends on meas.quant. ditto
³ Ã .FixEPFix endpoints
³ ³ Ã .1up to 9 fix EP's
³ ³ ³ Ã.ValueMeasured value for fix EP1depends on meas.quant. ditto
³ Ã .pKpK or HNP evaluationON, OFFditto
³
à .PreselPreselections
³ Ã .IReqRequest of Id's after startid1, id1&2, all, OFF3.2.2.25.
³ Ã .SreqRequest of smpl size after startvalue, unit, all, OFFditto
³ Ã .LimSmplSizeLimits for sample size3.2.2.26.
³ ³ Ã .StatusStatus of limit controlON, OFFditto
³ ³ Ã .LoLimLower limit0.0...999 999ditto
³ ³ Ã .UpLimUpper limit0.0...999 999ditto
³ Ã .ActPulseOutput of a pulseON, OFF3.2.2.27.
785 DMP Titrino
79
3.2. Remote control commands
*ParameterTree part "Parameters for MET"
à .TitrParaTitration parameters
³ Ã .VStepVolume increment0...0.10...999.93.2.2.12.
³ Ã .DosRateDispensing rate for increments0.01...150.0, max.3.2.2.13.
³ Ã .SignalDriftDrift for meas. value acquisitiondepends on meas.quant. 3.2.2.14.
³ Ã .UnitSigDriftUnit of measured value driftread onlyditto
³ Ã .EquTimeEquilibrium time0...26...9999, OFFditto
³ Ã .StartVStart volume
³ ³ Ã .TypeType of start volumeabs., rel., OFF3.2.2.15.
³ ³ Ã .VVolume for absolute start volume0...999.99ditto
³ ³ Ã .FactorFactor for relative start volume0...±999 999ditto
³ ³ Ã .RateDispensing rate for start volume0.01...150.0, max.ditto
³ Ã .PauseWaiting time0...999 9993.2.2.1.
³ Ã .MeasInputMeasuring input1, 2, diff.3.2.2.18.
³ Ã .IpolPolarization current0...1...±127ditto
³ Ã .UpolPolarization voltage0...400...±1270ditto
³ Ã .PolElectrTestTest for polarized electrodesON, OFFditto
³ Ã .TempTitration temperature-170.0...25.0...500.03.2.2.19.
³
à .StopCondStop conditions
³ Ã .VStopStop volume
³ ³ Ã .TypeType of stop volumeabs., rel., OFF3.2.2.20.
³ ³ Ã .VVolume for absolute stop volume0...99.99...9999.99ditto
³ ³ Ã .FactorFactor for relative stop volume0...±999 999ditto
³ Ã .MeasStopStop measured value pH, U, Idepends on meas.quant. 3.2.2.21.
³ Ã .UnitMStopUnit of stop measured valueread onlyditto
³ Ã .EPStopStop after a number of EP's1...9, OFF3.2.2.22.
³ Ã .FillRateFilling rate0.01...150.0, max.3.2.2.23.
³
à .StatisticsStatistics
³ Ã .StatusStatus of statistics calculationON, OFF3.2.2.24.
³ Ã .MeanNNo. of individual determinations2...20ditto
³ Ã .ResTabResult table
³ ³ Ã .Selectoriginal,delete n,delete allditto
³ ³ Ã .DelNDeletion of individual results1...20ditto
³
à .EvaluationEvaluation
³ Ã .EPCEP criteriondepends on meas.quant. 3.2.2.25.
³ Ã .RecognitionEP recognition
³ ³ Ã .SelectType of EP recognitionall,greatest,last,window,OFFditto
³ ³ Ã .WindowWindow
³ ³ ³ Ã .1up to 9 windows
³ ³ ³ ³ Ã.LowLimLower limit window 1depends on meas.quant. ditto
³ ³ ³ ³ Ã.UpLimUpper limit window 1depends on meas.quant. ditto
³ Ã .FixEPFix endpoints
³ ³ Ã .1up to 9 fix EP's
³ ³ ³ Ã.ValueMeasured value for fix EP1depends on meas.quant. ditto
³ Ã .pKpK or HNP evaluationON, OFFditto
³
à .PreselPreselections
³ Ã .IReqRequest of Id's after startid1, id1&2, all, OFF3.2.2.26.
³ Ã .SReqRequest of sample size after startvalue, unit, all, OFFditto
³ Ã .LimSmplSizeLimits for sample size3.2.2.27
³ ³ Ã .StatusStatus of limit controlON, OFFditto
³ ³ Ã .LoLimLower limit0.0...999 999ditto
³ ³ Ã .UpLimUpper limit0.0...999 999ditto
³ Ã .ActPulseOutput of a pulseON, OFF3.2.2.28.
80
785 DMP Titrino
3.2 Remote control commands
*ParameterTree part "Parameters for SET"
à .SET1Control parameters for EP1
³ Ã .EPEndpoint 1depends on meas.quant. 3.2.2.28.
³ Ã .UnitEpUnit of endpointread onlyditto
³ Ã .DynDynamicsdepends on meas.quant. 3.2.2.29.
³ Ã .UnitDynUnit of dynamicsread onlyditto
³ Ã .MaxRateMaximum dosing rate0.01...10...150, max.ditto
³ Ã .MinRateMinimum dosing rate0.01...25.0...9999ditto
³ Ã .StopTitration stop
³ ³ Ã .TypeType of stop criteriondrift, time3.2.2.30.
³ ³ Ã .DriftStop drift1...20...999ditto
³ ³ Ã .TimeSwitch-off delay time0...10...999, infditto
³ ³ Ã .StopTStop time0...999 999, OFFditto
à .SET2Control parameters for EP2, as for EP1
à .TitrParaTitration parameters
³ Ã .DirectionTitration direction+, -, auto3.2.2.31.
³ Ã .XPauseWaiting time before start volume0...999 9993.2.2.32.
³ Ã .Start VStart volume
³ ³ Ã .TypeType of start volumeabs., rel., OFF3.2.2.15.
³ ³ Ã .VVolume for absolute start volume0...999.99ditto
³ ³ Ã .FactorFactor for relative start volume0...±999 999ditto
³ ³ Ã .RateDispensing rate for start volume0.01...150.0, max.ditto
³ Ã .PauseWaiting time after start volume0...999 9993.2.2.16.
³ Ã .ExtrTExtraction time0...999 9993.2.2.33.
³ Ã .MeasInputMeasuring input1, 2, diff.3.2.2.18.
³ Ã .IpolPolarization current0...1...±127ditto
³ Ã .UpolPolarization voltage0...400...±1270ditto
³ Ã .PolElectrTestTest for polarized electrodesON, OFFditto
³ Ã .TempTitration temperature-170.0...25.0...500.03.2.2.19.
³ Ã .TDeltaTime interv. for meas.acquisition1...2...999 9993.2.2.34.
à HotKeyKeys with direct access
:Ã .UserUser name3.2.2.91.
³ Ã .NameInput of user nameup to 10 ASCII charditto
³ Ã .DeleteDelete user$Gditto
³ ³ Ã .NameInput of user nameup to 10 ASCII charditto
³ Ã .DelAllDelete all users$Gditto
³ Ã .ListList of users
³ ³ Ã .1User 1
³ ³ ³ Ã .NameName of userread onlyditto
³ ³ Ã up to 99
785 DMP Titrino
89
3.2. Remote control commands
&Info
ObjectDescriptionInput rangeReference
& Root
:
à InfoCurrent data
:Ã .ReportTransmission of formatted reports $G3.2.2.92.
³ Ã .SelectReport typeconfiguration, parameters,
³smpl data, statistics, silo, calib
³C-fmla, def, user method, full,
³short, mplist, curve, derive, comb,
³adj para,scalc full, scalc srt, calc,
³act dir, mem card, all, ffditto
³
à .CalibrationDatapH calibration data$G3.2.2.93.
³ Ã .Inp1For measuring input 1
³ ³ Ã .pHasAsymmetry pH0...7.00...±20.00ditto
³ ³ Ã .SlopeSlope of electrode0...1.000...±9.999ditto
³ ³ Ã .TempCalibration temperature-170.0...25.0...500.0ditto
³ ³ Ã .DateDate of calibrationread onlyditto
³ ³ Ã .ElectrodeIdId of the calibrated electroderead onlyditto
³ Ã .Inp2For measuring input 2, as for input 1
³ Ã .DiffFor differential input, as for input 1
³
à .ChecksumsChecksums$G3.2.2.94.
³ Ã .MPListChecksum of meas.point listread only ditto
³ Ã .ActualMethodChecksum of current methodread only ditto
³
à .DetermDataDetermination data$G3.2.2.95.
³ Ã .WriteRead/write for several nodsON, OFF
³ Ã .ExVVolume of Exchange/Dosing unitread only/read+writeditto
³ Ã .MPListMeasuring point list
³ ³ Ã .1Measuring point 1
³ ³ ³ Ã .AttributeAttributeread only/read+writeditto
³ ³ ³ Ã .XX coordinateread only/read+writeditto
³ ³ ³ Ã .YY coordinateread only/read+writeditto
³ ³ ³ Ã .Z1Z1 coordinateread only/read+writeditto
³ ³ ³ Ã .Z2Z2 coordinateread only/read+writeditto
³ ³ ³for each measuring point
³
à .TitrResultsTitration results
³ Ã .RSCalculated results
³ ³ Ã .11
st
result
³ ³ ³ Ã .ValueValueread only3.2.2.96.
³ ³ Ã up to 9 results
"Info", continuation
³ Ã .EPEndpoint
³ ³ Ã .11
st
result
³ ³ ³ Ã .VValueread only ditto
³ ³ ³ Ã .MeasMeasured valueread only
³ ³ ³ Ã .MarkMark if more than 1 EP per window read only
³ ³ Ã up to 9 EP's
³ Ã .VarVariables C4X
³ ³ Ã .C40Start measured valueread only/read+write ditto
³ ³ Ã .C41Titration end volumeread only/read+write
³ ³ Ã .C42Titration timeread only/read+write
90
785 DMP Titrino
3.2 Remote control commands
³ ³ Ã .C43Volume drift in SETread only/read+write
³ ³ Ã .C44Titration temperatureread only/read+write
³ ³ Ã .C45Start volumeread only/read+write
³ ³ Ã .C46Asymmetry pHread only
³ ³ Ã .C47Slope of electroderead only
³ ³ Ã .C48Volume at maximum voltageread only/read+write
³ ³ Ã .C49Volume at minimum voltageread only/read+write
³ ³ Ã .DTimeTime for drift corr.or dosing timeread only/read+write
³ Ã .FixEPFix EP
³ ³ Ã .51C51
³ ³ ³ Ã .ValueValueread only3.2.2.97.
³ ³ ³up to 59
³ Ã .pKpK/HNP
³ ³ Ã .61C61
³ ³ ³ Ã .ValueValueread only ditto
³ ³ ³up to 69
³ Ã .TempVarTemporary variables C7X
³ ³ Ã .C70up to C79read only/read+write ditto
³
à .StatisticsValStatistics values
³ Ã .ActNNumber of results in chartread only3.2.2.98.
³ Ã .11
st
mean
³ ³ Ã .MeanMeanread only ditto
³ ³ Ã .StdAbsolute standard deviationread only ditto
³ ³ Ã .RelStdRelative standard deviationread only ditto
³ Ã up to 9 mean values
³
à .SiloCalcValues of silo calculations
³ Ã .C24Values of variable C24
³ ³ Ã .NameNameread only3.2.2.99.
³ ³ Ã .ValueValueread only ditto
³ ³ Ã .UnitUnitread only ditto
³ Ã .C25as for C24
³ Ã .C26Values of variable C26
³ ³ Ã .ActNNumber of single valuesread onlyditto
³ ³ Ã .MeanMean valueread only ditto
³ ³ Ã .StdAbsolute standard deviationread only ditto
³ ³ Ã .RelStdRelative standard deviationread only ditto
³ Ã .C27as for C26
³"Info", continuation
à .ActualInfoCurrent data
³ Ã .InputsI/O Inputs
³ ³ Ã .StatusLine statusread only3.2.2.100.
³ ³ Ã .ChangeChange of line statusread only ditto
³ ³ Ã .ClearClear change$G ditto
³ Ã .Outputsas for I/O Inputs
³ Ã .AssemblyFrom Assembly
³ ³ Ã .CyclNoCycle numberread only3.2.2.101.
³ ³ Ã .CounterAssembly counterread only3.2.2.102.
³ ³ ³ Ã .VVolume counterread onlyditto
³ ³ ³ Ã .ClearClears counter$Gditto
³ ³ Ã .MeasMeasured valueread only3.2.2.103.
³ Ã .TitratorFrom Titrator
³ ³ Ã .CyclNoCycle numberread only3.2.2.104.
³ ³ Ã .VVolumeread only ditto
³ ³ Ã .MeasMeasured indicator voltageread only ditto
³ ³ Ã .dVdtVolume drift dV/dtread only ditto
³ ³ Ã .dMeasdtMeasured value driftread only ditto
785 DMP Titrino
91
3.2. Remote control commands
³ ³ Ã .dMeasdV1st deviation of titration curveread only ditto
³ ³ Ã .ERCERC from DETread only ditto
³ ³ Ã .TTemp.as secondary meas.read only ditto
³ Ã .MeasPtEntry in measuring point list
³ ³ Ã .IndexIndex of entryread only3.2.2.105.
³ ³ Ã .XX coordinateread only ditto
³ ³ Ã .YY coordinateread only ditto
³ ³ Ã .Z1Z1 coordinateread only ditto
³ ³ Ã .Z2Z2 coordinateread only ditto
³ Ã .EPEP entry
³ ³ Ã .IndexIndex of entryread only ditto
³ ³ Ã .XX coordinateread only ditto
³ ³ Ã .YY coordinateread only ditto
³ Ã .DisplayDisplay
³ ³ Ã .L1Text line 1up to 32 ASCII char3.2.2.106.
³ ³ Ã up to line 8
³ ³ Ã .DelAllDelete display$G ditto
³ Ã .ComportComport
³ ³ Ã .NumberCOM where PC is connectedread only3.2.2.107.
³
à .AssemblyAssembly
³ Ã CycleTimeCycle timeread only3.2.2.108.
³ Ã ExVVolume of Exchange unitread only ditto
92
785 DMP Titrino
3.2 Remote control commands
&Assembly
ObjectDescriptionInput rangeReference
& Root
:
à AssemblyAssembly control
à . BurBuret
³ Ã .RatesRates
³ ³ Ã .ForwardForward rate
³ ³ ³ Ã .SelectType of rate controldigital, analog3.2.2.109.
³ ³ ³ Ã .DigitalDigital rate0...150, max. ditto
³ ³ Ã .Reverseas for forward rate
³ ³ ³ Ã .SelectType of rate controldigital, analog ditto
³ ³ ³ Ã .DigitalDigital rate0...150, max. ditto
³ Ã .FillFill$G,$H,$C3.2.2.110.
³ Ã .ModeDisDispensing$G,$S,$H,$C3.2.2.111.
³ ³ Ã .SelectType of dispensing controlvolume, time ditto
³ ³ Ã .VVolume to be dispensed0.0001...0.1...9999 ditto
³ ³ Ã .TimeTime to dispense0.25...1...86 400 ditto
³ ³ Ã .VStopLimit volume0.0001...9999, OFF ditto
³ ³ Ã .AutoFillFilling after each incrementON, OFF ditto
³
à .MeasMeasuring
³ Ã .StatusMeasuring ON/OFFON, OFF3.2.2.112.
³ Ã .MeasInputSelection of measuring input1, 2, diff., Ipol, Upol, Temp ditto
³ Ã .IpolPolarization current0...1...±127 ditto
³ Ã .UpolPolarization voltage0...400...±1270 ditto
³
à .OutputsI/O outputs
³ Ã .AutoEODAutomatic output of EODON, OFF3.2.2.113.
³ Ã .SetLinesSet I/O lines$G ditto
³ ³ Ã .L0Signal on L0active,inactive,pulse,OFF ditto
³ ³ Ã up to L13
³ Ã .ResetLinesReset I/O lines$G ditto
³
à .StirrerStirrerON, OFF3.2.2.114.
³
785 DMP Titrino
93
3.2. Remote control commands
&Setup
ObjectDescriptionInput rangeReference
& Root
:
à SetupSettings for the operating mode
à .ComportOutput of automatic info1,2,1&23.2.2.115.
à .KeycodeSend key codeON, OFF3.2.2.116.
à .TreeSending format of path info
³ Ã .ShortShort format of pathON, OFF3.2.2.117.
³ Ã .ChangedOnlyPaths of modified nodes only ON, OFFditto
³
à .TraceMessage on changed valuesON, OFF3.2.2.118.
³
à .LockLock key functions
³ Ã .KeyboardLock all keyboard keysON, OFF3.2.2.119.
³ Ã .ConfigLock <CONFIG> key ON, OFF ditto
³ Ã .ParameterLock <PARAM> keyON, OFF ditto
³ Ã .SmplDataLock <SMPL DATA> keyON, OFF ditto
³ Ã .UserMethLock functions
³ ³ Ã .RecallLock “loading”ON, OFF ditto
³ ³ Ã .StoreLock “saving”ON, OFF ditto
³ ³ Ã .DeleteLock “deletion”ON, OFF ditto
³ À .DisplayLock display functionON, OFF ditto
³
à .ModeSetting waiting intervals
³ Ã .StartWaitWaiting time after startON, OFF3.2.2.120.
³ Ã .FinWaitWaiting time after runON, OFF ditto
³
à .SendMeasAutomatic sending of measured values
³ Ã .SendStatusConnect/disconnect sendingON, OFF3.2.2.121.
³ Ã .IntervalTime interval0.08...4...16200, ditto
³MPList
³ Ã .SelectSelectionAssembly, Titrator3.2.2.122.
³ Ã .AssemblyFrom assembly
³ ³ Ã .CyclNoCycle numberON, OFF3.2.2.123.
³ ³ Ã .VVolumeON, OFF ditto
³ ³ Ã .MeasMeasured indicator voltageON, OFF ditto
³ Ã .TitratorFrom Titrator
³ ³ Ã .CyclNoCycle numberON, OFF3.2.2.124.
³ ³ Ã .VVolumeON, OFF ditto
³ ³ Ã .MeasMeasured indicator voltageON, OFF ditto
³ ³ Ã .dVdtVolume drift dV/dtON, OFF ditto
³ ³ Ã .dMeasdtMeasured value driftON, OFF ditto
³ ³ Ã .dMeasdV1st deviation of titration curveON, OFF ditto
³ ³ Ã .ERCERC from DETON, OFF ditto
³ ³ Ã .TTemp.as secondary meas.ON, OFF ditto
³
94
785 DMP Titrino
3.2 Remote control commands
"Setup", continuation
à .AutoInfoAutomatic message for changes3.2.2.125.
³ Ã .StatusSwitch AutoInfo on/off ON, OFF ditto
³ Ã .PWhen mains is switched onON, OFF ditto
³ Ã .TTitrator infos
³ ³ Ã .RWhen “ready”ON, OFF ditto
³ ³ Ã .GWhen method startedON, OFF ditto
³ ³ Ã .GCWhen start is initiatedON, OFF ditto
³ ³ Ã .SWhen stoppedON, OFF ditto
³ ³ Ã .BBegin of methodON, OFF ditto
³ ³ Ã .FEnd of processON, OFF ditto
³ ³ Ã .EErrorON, OFF ditto
³ ³ Ã .HWhen “hold"ON, OFF ditto
³ ³ Ã .CContinue after "hold"ON, OFF ditto
³ ³ Ã .OConditioning OKON, OFF ditto
³ ³ Ã .NConditioning not OKON, OFF ditto
³ ³ Ã .ReRequest after startON, OFF ditto
³ ³ Ã .SiSilo emptyON, OFF ditto
³ ³ Ã .MEntry in measuring point listON, OFF ditto
³ ³ Ã .EPEntry in EP listON, OFF ditto
³ ³ Ã .RCRecalculation of results doneON, OFF ditto
³ Ã .CComport infos
³ ³ Ã .B1When COM1 sends a reportON, OFF ditto
³ ³ Ã .R1When COM1 is ready againON, OFF ditto
³ ³ Ã .B2When COM2 sends a reportON, OFF ditto
³ ³ Ã .R2When COM2 is ready againON, OFF ditto
³ Ã .IChanging an I/O inputON, OFF ditto
³ Ã .OChanging an I/O outputON, OFF ditto
³
à .GraphicsChanging the curve output
³ Ã .GridGrid on curveON, OFF3.2.2.126.
³ Ã .FrameFrame on curveON, OFF ditto
³ Ã .ScaleType of depending axisFull, Auto ditto
³ Ã .RecorderLength of axes
³ ³ Ã .RightLength of meas value axis0.2...0.5...1.00 ditto
³ ³ Ã .FeedLength of paper drive axis0.01...0.05...1.00 ditto
³
à .PowerOnRESET (power on)$G3.2.2.127.
à .InitialiseSet default values$G3.2.2.128.
³ Ã .SelectSelection of branchActMeth,Config,Silo,Calib
³Assembly,Setup,All ditto
à .RamInitInitialization of working mem.$G3.2.2.129.
à .InstrNoDevice Identification$G3.2.2.130.
à .ValueInput of device identification8 ASCII characters ditto
785 DMP Titrino
95
3.2. Remote control commands
&Diagnose
ObjectDescriptionInput rangeReference
& Root
:
à DiagnoseDiagnose
à .ReportOutput of adjustment parameters$G3.2.2.131.
96
785 DMP Titrino
3.2 Remote control commands
3.2.2 Description of the remote control commands
3.2.2.1.Mode$G, $S, $H, $C
Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size
after the start (see 3.2.2.25) as well as after inquiries of calibration temperature and pH values of buffers (see 3.2.2.41 and 3.2.2.42).
3.2.2.2.Mode.QuickMeas$G, $S
Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the <meas/hold> key. In TIP, the measured quantity is selected with
&Mode.Parameter.Presel, see 3.2.2.49.
With an ongoing measurement, the current mode can be started. This stops
the measurement automatically.
785 DMP Titrino
3.2.2.3.Mode.SelectDET, MET, SET, MEAS, CAL, TIP
Mode.DETQuantitypH, U, Ipol, Upol
Mode.METQuantitypH, U, Ipol, Upol
Mode.SETQuantitypH, U, Ipol, Upol
Mode.MEASQuantitypH, U, Ipol, Upol, T
Selection of the standard mode. Mode and the measured quantity belong to
the complete selection.
If a method is selected from the method memory, the nodes &Mode.Select
and &Mode.XXXQuantity are overwritten with mode and measured quantity of
the corresponding user method.
3.2.2.4.Mode.Nameread only
Name of the current method in the working memory. $Q sends 8 ASCII characters. Standard methods carry the name ********. The node can be set
read + write, see 3.2.2.95.
3.2.2.5.Mode.Def.Formulas.1.FormulaEPX, CXX, RSX, +, -, *, /, (, )
Mode.Def.Formulas.1.TextRSup to 8 ASCII characters
Mode.Def.Formulas.1.Decimal0...2...5
Mode.Def.Formulas.1.Unitup to 6 ASCII characters
Mode.Def.Formulas.1.LimitsON, OFF
Mode.Def.Formulas.1.Outputactive, pulse, OFF
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 38.
Example: "(EP2-EP1)*C01/C00"
In addition to the formula, a text for result output, the number of decimal
places and a unit for the result output can be selected. "No unit" is selected
with the blank string.
97
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.