iina74e1-h (1408) Translation of the original handbook
Catalog-No.
550-500A,
550-501A
UL5000
Helium Leak Detector
from software version
V 5.14
Table of contents
1 General Information 6
1.1Notes on the Use of this Handbook 6
1.1.1Safety Symbols 6
1.1.2Indications 7
1.1.3Symbols of Vacuum Technology 7
1.1.4Definition of Terms 7
1.2Support from INFICON Service 9
1.2.1Service Centers 10
1.3Introduction 12
1.3.1Purpose 12
1.3.2Technical Data 13
1.3.2.1Physical Data 13
1.3.2.2Electrical Data 14
1.3.2.3Other Data 14
1.3.2.4Ambient Conditions 14
1.4Unpacking 14
1.4.1Supplied Equipment 15
1.4.2Accessories and Options 16
1.4.2.1Sniffer line SL200 16
1.4.2.2Toolbox 16
1.4.2.3Helium Bottle Holder 16
1.4.2.4ESD Mat 16
1.4.2.5RC1000 Remote control 17
2 Installation 18
2.1Transportation 18
2.2Working Location 20
2.3Electrical Connections 21
2.3.1Mains Power 21
2.3.2Connections for the Data Acquisition Systems 23
2.3.2.1 Accessories 24
2.3.2.2Digital Out 24
2.3.2.3Digital In 25
2.3.2.4Recorder 26
2.3.2.5RS232 26
2.3.2.6Remote Control RC1000 / Wireless Transmitter 26
2.4Vacuum Connections 27
2.4.1 Inlet Port 27
2.4.2Exhaust 27
2.4.3Vent 27
2.4.4Purge 27
2.5Default parameters 28
3 First Operation Check 29
3.1Needed Equipment 29
3.2Description of the Initial Operation 29
iina74e 0.fmtechnical handbook(1408)
2Table of contents
3.2.1Startup and Measure 29
3.2.2Internal Calibration 32
3.2.3Verification 32
4 Description and Working Principle 33
4.1Introduction 33
4.2Components of the UL5000 33
4.2.1Vacuum System 33
4.2.2Control Panel 35
4.2.2.1LC Display 35
4.2.2.2START Button 35
4.2.2.3STOP Button 36
4.2.2.4ZERO Button 36
4.2.2.5MENU Button 37
4.2.2.6Soft Keys 37
4.2.2.7Numerical Entries 38
4.3Working Modes 39
4.3.1Vacuum Mode 39
4.3.2Sniffer Mode 40
5 Operation of the UL5000 41
5.1Display 41
5.2The Screen in Run-Up Mode 41
5.3Display in stand-by mode 41
5.3.1Purging 42
5.4The Screen in Measurement Mode 42
5.4.1Call for Calibration 42
5.4.2Speaker Volume 43
5.4.3Status Line in the Display 43
5.4.4Numerical Display Mode 44
5.4.5Trend Mode 44
5.4.6HYDRO•S on / off 44
6 Description of the Menu 45
6.1Main Menu 45
6.2View 47
6.2.1Scale linear/logarithmic 47
6.2.2Display-range auto/manual 48
6.2.3Time axis 48
6.2.4Contrast 49
6.2.5Background in Stand-by 49
6.2.6Decimal places 50
6.2.7Lower display limit 50
6.3Mode 51
6.4Trigger & Alarms 52
6.4.1Trigger Level 1 52
6.4.2Trigger Level 2 53
iina74e 0.fmtechnical handbook(1408)
6.4.3Volume 53
6.4.4Units 54
Table of contents3
6.4.5Alarm delay 54
6.4.6Audio alarm type 55
6.4.6.1Pinpoint 55
6.4.6.2Leak rate prop. 55
6.4.6.3Setpoint 56
6.4.6.4Trigger alarm 56
6.5Calibration 56
6.6Settings 57
6.6.1Vacuum settings 58
6.6.1.1Automatic purge 58
6.6.1.2Vent delay 59
6.6.1.3Vacuum ranges 59
6.6.1.4HYDRO•S 60
6.6.1.5Leak rate internal test leak 60
6.6.1.6Machine factor 61
6.6.1.7Booster TMP mode 61
6.6.2Zero & Background 61
6.6.2.1Background Suppression 62
6.6.2.2Calculate Inlet Area Background 62
6.6.2.3Zero 63
6.6.3Mass 63
6.6.4Interfaces 64
6.6.4.1Control Location 64
6.6.4.2Recorder output 65
6.6.4.3RS232 Protocol 66
6.6.4.4Scaling Recorder Output 67
6.6.5Miscellaneous 68
6.6.5.1Time&Date 68
6.6.5.2Language 68
6.6.5.3Leak rate filter 69
6.6.5.4Mains Frequency 69
6.6.5.5 Service interval exhaust filter 69
6.6.5.6Service message exhaust filter 70
6.6.6Parameter save / load 70
6.6.6.1Save parameter set 71
6.6.6.2Load parameter set 71
6.6.7Monitoring functions 71
6.7Information 74
6.7.1Service 74
6.8Access Control 75
6.8.1Access to CAL function 75
6.8.2Access to Trigger&Alarme menu 76
6.8.3Change Device PIN 76
6.8.4Change Menu-PIN 76
4Table of contents
7 Calibration 77
7.1Introduction 77
7.2The calibration routines 77
7.2.1Internal Calibration 78
7.2.1.1Automatic Internal Calibration 78
iina74e 0.fmtechnical handbook(1408)
7.2.1.2Manual Internal Calibration 78
7.2.2External Calibration 78
7.3Factor of Calibration - Range of Values 81
8 Error And Warning Messages 82
8.1Hints 82
8.2List of Errors & Warnings 83
9 Maintenance works 88
9.1Hints 88
9.2INFICON Service 89
9.3Key for Maintenance Schedule 89
9.4Maintenance Plan 90
9.5Maintenance groups 91
9.5.11500 hours maintenance 91
9.5.24000 hours maintenance 92
9.5.38000 hours maintenance 93
9.5.416000 hours maintenance 94
9.6Description of the maintenance work 95
9.6.1Opening the UL5000 95
9.7Check/Replace the filter cell 96
9.8Replacing the Exhaust Silencer 98
9.9Turbomolecular pump TMH 071 99
9.10Scroll Pump 99
Appendix 100
Declaration of Conformity 102
iina74e 0.fmtechnical handbook(1408)
Table of contents5
1General Information
Caution
Warning
STO P
Danger
Skilled personnel
Notice: We recommend that you carefully read this technical handbook to ensure
optimum operating conditions right from the start.
This technical handbook contains important informations on the functions, insta llation, start-up and operation of the UL5000.
General
We reserve the right to modify the design and the specified data. The illustrations are
not binding.
1.1Notes on the Use of this Handbook
1.1.1Safety Symbols
Important remarks concerning operational safety and protection are emphasised as
follows:
Information on correct handling or use. Disregard can lead to malfunctions or minor
equipment damage.
Information on preventing extensive equipment and environmental damage.
Information on preventing any kind of physic al injury.
Indicates procedures that must be performed by skilled personnel only.
iina74e 01.fmtechnical handbook(1408)
6General Information
1.1.2Indications
TippInformation on helpful procedures.
Notice: Information on special technical requirements that the user must comply
with.
The references to diagrams consists of the Chapter number, figure number and the
item number in this order. For example: Fig. 2-4/7 refers to item 7 in the figure 4 of
Chapter 2.
1.1.3Symbols of Vacuum Technology
Given in the following are some important vacuum symbols which are used in this
manual.
Vacuum pump in general
Turbomolecular pump
Pressure gauge
1.1.4Definition of Terms
Autoranging
The range of the preamplifier and the vacuum ranges are selected automatically.
The autoranging feature of the UL5000 covers the entire range or leak rates
depending on the selected operating mode. Not only the leak rate signal, but also the
pressure in the test sample (inlet pressure P1) and the fore vacuum pressure (P2)
are used for control purposes. Range switching between the main ranges is
performed via valves. Fine range switching within the main ranges is implemented
by switching over the gain factor of the preamplifier.
Autotune Mass alignment
This function automatically aligns the mass spectrometer so that a maximum leak
rate is displayed. The control processor changes the voltage which erates the ions
in the selected mass range until a maximum ion current is detected by the ion
detector. During each calibration the mass alignment is run automatically.
Valve
iina74e 01.fmtechnical handbook(1408)
General Information7
Auto zero
Determination and automatic adaptation of the helium background.
Through this function, the internal zero level of the instrument is determined which is
then subtracted from the current leak rate signal. This function is run during the calibration process or when operating the start pushbutton, provided the UL5000 has
been running previously for at least 20 seconds in the standby or vent mode. If the
helium background previously suppressed shoul d drop so that for the du ration of the
zero time only the display limit will be displayed, the zero level is adapted automatically.
Menu
The menu allows the user to program the UL5000 according to his requirements. The
menu has a tree architecture.
Default
Status of the UL5000 when supplied by the factory.
GROSS
GROSS is a measurement mode which allows high inlet pressure (1 to 15 mbar).
The smallest detectable leak rate is 1×10
-6
mbar l/s.
FINE
FINE is the medium measurement mode with inlet pressure between 2 and 0 ,4 mbar.
Detection limit is 1×10
-10
mbar l/s.
ULTRA
ULTRA is the most sensitive measuring range with inlet pressures below 0,4 mbar.
The minimum detectable leak rate is < 5×10
-12
mbar l/s.
Foreline pressure
Pressure in the foreline between Turbo pump and scroll pump.
Minimum detectable leak rate
The smallest leak rate the UL5000 is able to detect ( 5E-12 mbar l/s).
Internal helium background
The existing helium partial pressure in the measurement system. The level of the
internal helium background is measured in the Stand-by mode and subtracted fr om
the measured signal.
Measure Measurement mode
The UL5000 measures the leak rate of the test sample.
I/we hereby declare that the information on this form is complete and accurate and that I/we will assume any further costs that may
arise. The contaminated product will be dispatched in accordance with the applicable regulations.
Organization/company
Address Post code, place
Phone Fax
Email
Name
Date and legally binding signature Company stamp
1) or not containing any amount
of hazardous residues that
exceed the permissible ex posure limits
Process related contamination of product:
toxic no 1) yes
caustic no 1) yes
biological hazard no yes 2)
explosive no yes 2)
radioactive no yes 2)
other harmful substances no1) yes
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has
been submitted. Non-completion will result in delay.
This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Copies:
Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which the product may have come into contact with:
Trade/product name
Chemical name
(or symbol)
Precautions associated
with substance
Action if human contact
Description of product
Type
Article Number
Serial Number
Reason for return
Operating fluid(s) used (Must be drained before shipping.)
The product is free of any substances which are damaging to
health yes
This form can be downloaded
from our website.
2) Products thus contam i nated will not be ac cepted without written
evidence of decontami nation!
If an instrument is returned to INFICON or an authorised representative of INFICON,
please indicate whether the instrument is free of substances damaging to health or
whether it is contaminated. If it is contaminated also indicate the nature of the hazard.
INFICON must return any appliances without a Declaration of Contamination to the
sender’s address. A form for stating details as to the type of contamination is reproduced in Fig. 1-1.
A maintenance and service contract is recommended.
iina74e 01.fmtechnical handbook(1408)
www.inficon.com leakdetection.service@inficon.co
Fig. 1-1: Declaration of Contamina tion form
General Information9
1.2.1Service Centers
In case you urgently need assistance please get in touc h with the loca l IN FICON
Service in your country or the service hotline in Cologne, Germany:
Algeriajhj@agramkow.dkFinlandjhj@agramkow.dk
Agramkow
Sonderborg
Belarusleakdetection.service@inficon.com FranceChristophe.Zaffanella@oerlikon.com
INFICON GmbH
Cologne
Belgiumleakdetection.service@inficon.com Germanyleakdetection.service@inficon.com
INFICON GmbH
Cologne
Brazilfernandoz@prestvacuo.com.brHungaryadam.lovics@kon-trade.hu
PV Pest Vácuo Ltda.
Santa de Parnaíba
Bulgarialeakdetection.service@inficon.com Indiaasdash@hotmail.com
INFICON GmbH
Cologne
Canadareachus@vpcinc.caIrelandreach.unitedkingdom@inficon.com
Vacuum Products Canada Ltd.
Ontario
Central Americainfoqro@meisa.comItalydavide.giovanetti@inficon.com
MEISA S.a. de C.V.
Querètaro
Chinareach.china@inficon.comIsraelurimark@mark-tec.co.il
INFICON LTD
Hong Kong
INFICON LTD
Beijing
INFICON LTD
Guangzhou
INFICON LTD
Shanghai
Czech Republicfilip.lisec@inficon.comKoreareach.korea@inficon.com
INFICON GmbH
Pilsen
Denmarkjhj@agramkow.dkINFICON Ltd.
Fax: +49 221 56788-9112
Sloveniamedivak@siol.netUSAservice.usa@inficon.com
Medivac
Ljubljani
South America except Brazilinfoqro@meisa.comInficon Inc.
MEISA S.a. de C.V.
Querètaro
Phone: +386 15 63 91 50
Fax: +386 17 22 04 51
Phone: +52 44 22 12 36 15
Fax: +52 44 22 12 19 40
Inficon Inc.
East Syracuse, NY
San Jose, CA
Inficon Inc.
Austin, TX
Phone: +1.315.434.1167
Fax: +1.315.434.2551
Phone: +1.408.361.1200
Fax: +1.408.362.1556
Phone: +1.512.448.0488
Fax: +1.512.448.0398
iina74e 01.fmtechnical handbook(1408)
General Information11
1.3Introduction
STO P
Danger
STO P
Danger
Caution
1.3.1Purpose
The UL5000 is a helium leak detector. This instrument may be used to detect the
location and the size of leaks on objects under test in two different ways:
•when the test sample has been evacuated first and is spraye d with helium on the
or
•when a helium overpressure is provided in the test sample and the test sample is
Caution: Danger of explosion:
Hydrogen forms a highly explosive gas mixture with air.
Great caution is necessary when using hydrogen! No smoking, no naked flames,
avoid sparks.
outside. It is required that a vacuum connection is provided between the UL5000
and the test sample (vacuum mode).
searched from the outside with a sniffer probe which is attached to the inlet port
(sniffer mode).
Dangerous gases pollute the machine.
So you must not use the machine for detecting toxical, acidly, microbiological,
explosive, radioactive or other noxious matters.
If you plan to detect noxious matters please contact the manufacturer. Rules for
decontamination will be developed then. If the leak detector already has been in
contact with dangerous gases please fill the declaration of contamination and sent it
back with the leak detector to INFICON.
When there are parts that have to be cleaned, please fill the declaration of decontamination, too, and send it to INFICON before you send the parts.
The UL5000 is to be used for leak detection only. It must not be used as a pumping
system (esp. pumping aggressive or humid gases.)
•in sniffer mode< 5×10
Maximum Helium leak rate which can be displayed0.1 mbar l/s
Measurement range12 decades
Time constant of the leak rate signal (blanked off,
63% of the final value)
Roughing capability25 m
Pumping speed (Helium) at the inlet (EN 1518)30m³/h, 21.1 cfm
•in vacuum mode
Detectable masses2, 3 and 4
Mass spectrometer180°
Ion source2 filaments;
Inlet portDN 40 KF
Run-up time (after starting) 4min
To get down to the minimum detected leak rate range some conditions must be
fulfilled:
•UL5000 has fully warmed up
•Ambient conditions must be stable (temperature, no vibration/
accelerations.)
•The part under test has been evacuated long enough (background
is no longer decreasing)
•ZERO must be active
iina74e 01.fmtechnical handbook(1408)
General Information13
1.3.2.2Electrical Data
1.3.2.3Other Data
Power supply, single phase (model dependent)
Part no. 550 - 500A230 V 50 Hz
Part no. 550 - 501A115 V 60 Hz
Power consumption1200 VA
Type of protectionIP20
Power cords (EU, USA, UK)3 m
Valvessolenoid
Dimensions (L × W × H) incl. handle in mm1068 × 525 × 1083
Dimensions (L × W × H) incl. handle in inches42 × 21 × 43
Weight in kg140
Weight in lbs308
Noise level dB (A)< 70
Noise level dB (A) 0.5 m distance55.9
Audio alarm dB (A)90
Contamination level (to IEC 60664-1)2
Overvoltage category (to IEC 60664-1)II
1.3.2.4Ambient Conditions
For use within buildings
Permissible ambient temperatur e (d ur in g op er at ion)+10 °C … +40 °C
Permissible storage temperature0 °C … +60 °C
Max. rel. humidity80% non cond en sin g
Max. permissible height above sea level (during operation) 2000 m
1.4Unpacking
Unpack the UL5000 immediately after delivery, even if it will be installed later on.
Examine the shipping container for any external damage. Completely remove the
packaging materials.
Check the UL5000 is complete (1.4.1Supplied Equipment) and carefully examine
the UL5000 visually.
If any damage is discovered, report it immediately to the forwarding agent and
insurer. If the damaged part has to be replaced, please contact the orders
department.
TippRetain the packaging materials in the event of complaints about damage.
TippFor unpacking please use the wedge which is part of the packaging.
14General Information
iina74e 01.fmtechnical handbook(1408)
1.4.1Supplied Equipment
•Helium Leak Detector UL5000
•Exhaust filter (mounted)
•Exhaust hose adapter with clamps (see arrow 1)
•power cord fixture (see arrow 2)
•Set of fuses (see arrow 3)
•Set of tools (see arrow 4)
•Bellow Clips (2 + 2) (see arrow 5)
•Folder with documents
– Technical Handbook UL5000
– Spare Parts List UL5000
•hooks to wrap power cord (with screws) (see arrow 6)
•Tool to open the UL5000 (see arrow 7)
•O-Ring with filter (for use at applications with particles)
•Tool box (detachable)
•ESD mat
iina74e 01.fmtechnical handbook(1408)
Fig. 1-2 Accessories
General Information15
1.4.2Accessories and Options
The following parts can be ordered additionally:
•Sniffer Line SL20014005
•Adapter 40 / 25KF (required to connect SL200)211-283
•Tool Box (detachable)551-000
•Helium Bottle Holder551-001
•ESD Mat551-002
•Remote control in two versions:
•Spray gun with hose16555
•Set of plugs20099024
•Leak Ware (software package)14090
1.4.2.1Sniffer line SL200
By use of the sniffer line the UL5000 can easily be converted to a sniffer leak
detector. The length of the sniffer line is 4m (i.e. 12 feet).
1.4.2.2Toolbox
The toolbox is a detachable compartment with a lockable lid. Fittings and small
fixtures can be stored plus the hand set (See Chapter 1.4.2.5). The storage volume
is approximately 5 l.
The toolbox is placed on the working surface and jammed by the handle.
– RC1000C wired
– RC1000WL wireless
Extension Cable, 10 m
551-010
551-015
14022
1.4.2.3Helium Bottle Holder
The helium bottle holder allows you to carry a helium re servoir and a sp ray gun with
the UL5000. Only small and midsize bottles (max 10 l, 200 bar) will fit without influencing the stability of the UL5000.
1.4.2.4ESD Mat
This mat is put on the working surface of the UL5000 and is cla mped a nd grou nded
by the inlet port ring. It avoids electrical discharges between the working surface and
sensitive test parts.
16General Information
iina74e 01.fmtechnical handbook(1408)
1.4.2.5RC1000 Remote control
The RC1000 is a wireless remote control that allows to operate the UL5000 from
distance up to 100 m. It provides the functions START, STOP/VENT, ZERO and
speaker volume, and displays leak rate in bargaraph or in chart mode. (see also
Technical Handbook RC1000.)
Measured values can be stored in an internal memory for up to 24 hours of recording
time. The data can easily be downloaded to a USB stick to save it.
An internal trigger can be set to provide a warning if the limit leak rates are exceeded.
An optical warning is shown on the display and an acoustic warning signal is
sounded on the integrated loudspeaker or the connected headphones.
The RC1000 remote control is housed in a robust housing to enable ergonomic
working. Magnets on the underside of the unit enable it to be attached to horizontal
or vertical metal surfaces.
The RC1000 also enables remote operation of the le ak test device in question using
a connection cable of up to 28 metres in length.
Fig. 1-3 RC1000WL wireless remote control
iina74e 01.fmtechnical handbook(1408)
General Information17
2Installation
Caution
Caution
Caution
Caution
Caution
2.1Transportation
The UL5000 is not equipped with any crane eyes and must therefore not be
transported using lifting equipment.
The UL5000 must only be pushed along u sing the handle provided for this purpose.
Don’t use the handle to lift.
Your feet can be pinched.
Keep your feet away from the rollers. .
Your feet can be run over.
Do not pull the unit, push it.
When transporting over longer distances the or iginal packaging must be u sed. The
castors must not be fixed when the UL5000 is shipped in a crate.
UL5000 with Triscroll TS 620
For transportation the chassis plate where the pump is mounted on has to be
secured by a transportation fixing.
This transportation fixing consists of 2 screws at ch assis of the UL5000 (one on each
side).
To get access to these screws remove the side covers of the UL5000.
There are orange labels on the bottom part pointing to the screws:
iina74e 02.fmtechnical handbook(1408)
18Installation
Fig. 2-1
For transportation fixing the screws are tightened to the chassis plate. For operation
of the UL5000 the screws should be loosened.
To loosen the screws first loosen the counter nut that is accessible from underneath:
Fig. 2-2
Loosen the screws approximately 10 mm above the chass i s plat e an f tigh te n then
iina74e 02.fmtechnical handbook(1408)
the counter nuts again:
Installation19
Fig. 2-3
Warning
Caution
Caution
Caution
For transportation tighten the screws again and fix them by the counter nuts.
2.2Working Location
Move the UL5000 to the desired position and arrest the castors.
The UL5000 is designed for indoor use only.
Ensure a sufficient air cooling. The air inlet as well as the air discharge openings
must never be obstructed.
20Installation
The UL5000 can be locked by arresting the castors of the front wheels to avoid
movements on skewness.
Make sure that you can always reach the mains plug.
iina74e 02.fmtechnical handbook(1408)
STO P
Danger
Caution: Exhaust gases and fumes:
STO P
Danger
Warning
Exhaust gases and fumes from oil-sealed pumps may be harmful to health.
For operation in poorly ventilated rooms, an exhaust pipe should be connected to
exhaust connection 5 depending on the application and gases used.
It is recommended that you check all major helium sources in the vicinity of the
UL5000 within a radius of about 10 m for the presence of any big leaks. You may use
the sniffer probe for this.
2.3Electrical Connections
2.3.1Mains Power
Notice: Generally the local regulations for electrical connections m ust be observed.
Caution: mains voltage:
Improperly grounded or fused products can lead to fatal injuries.
The product must be connected in accordance with local regulations and properly
grounded. Interruption of the ground conducto r within or outside of the device is not
permitted!
Only fuses of the specified type and rating are to be used as replacements.
Before connecting the UL5000 to the mains you must make sure that the mains
voltage rating of the UL5000 coincides with the locally available mains voltage. The
instrument must exclusively be connected to a single phase supply with fuses for
installation (Circuit breaker 16A max. according to IEC/EN 60898 with tripping
characteristic B).
The mains voltage rating for the UL5000 can be read off from the name plate beneath
the mains socket Fig. 2-6/7 at the back side. This voltage is fixed and can not be
changed.
A separate fuse for each of the mains conductors has been integrated into the main s
switch.
The mains voltage is applied to the instrument via the detachable mains cable which
is supplied with the instrument. A mains socket Fig. 2-6/7 is available for th is purpose
at the back side of the instrument.
iina74e 02.fmtechnical handbook(1408)
Installation21
STO P
Danger
Only 3-core mains cables having a protection ground conductor must be used.
Operation of the UL5000 where the ground conductor has been left unconnected is
not permissible.
TippThe power cord can be secured against coming out by using the power
cord fixture as shown:
Fig. 2-4. secure fixture power cord
22Installation
Fig. 2-5. storing power cord
iina74e 02.fmtechnical handbook(1408)
2.3.2Connections for the Data Acquisition Systems
Caution
Caution
1. Accessories
2. Digital Out
3. Digital In
4. Recorder
5. RS232
6. Remote Control RC1000 / Wireless
Transmitter
7. Mains Socket
8. Mains Switch
9. Hole to mount cable hooks.
10. Speaker
11. Vent
12. Purge
Fig. 2-6 Connections
TippThe sockets: Accessories, Digital Out, Digital In and Recorder have pin 1
on top. The pin numbers are counted downwards. The socket 2 and 3 are
coded mechanically to avoid a confusion with the counter plug. For the
connection with the counter plug (set of plugs 20099024) remove the
plastic pins at the plug, accordingly the plug fits the socket.
Notice: The connections for external devices are safely separated from the mains
and safe low voltage.
The electronic of the device can be destroyed. So just connect de vices to the le ak
detector that are separated from the mains.
Only connect devices that don’t exceed 25V AC/1A
iina74e 02.fmtechnical handbook(1408)
Installation23
2.3.2.1 Accessories
2.3.2.2Digital Out
The following accessories may be connected to this port Fig. 2-6/1:
•external venting valve
•sniffer probe
Contact pins 1 and 3 are fused with a 0.8 A slow-blow fuse. The amount of power
which can be drawn is limited to 10 W. The contacts are numbered from top to
bottom.
PinAssignment
1
2GND24
3, 6Input
4, 5, 7, 8Output
+24 V, constantly applied, power supply for
the INFICON partial flow valve or sniffer line.
The following relay outputs are available for further signal processing. The maximum
rating for the relay contacts is 25V AC/1A.
PinAssignment
1+24V, bridged with pin 1
of socket „Digital In“
2GND_24V10Error
3Trigger 111Warning
4Trigger 212Purge
5Reserved13Measure
6Zero active14Recorder Strobe
7Ready15Common dig. out
8CAL active16Reserved
PinAssignment
9CAL request
Description of the operation mode of the Digital Out.
Trigger 1:
Is open in case Trigger Level 1 is exceeded or the machine is not in condition of
measuring.
Trigger 2:
Is open in case Trigger Level 2 is exceeded or the machine is not in condition of
measuring.
Zero active:
Is closed in case Zero function is running.
Ready:
Is closed in case machine is ready for measurement (Emission on, no error).
CAL active
Closed when machine is in calibrating routine.
CAL Request
Is opened in case of calibration request. During external calibration an open output
indicates that the external calibrated leak has to be closed.
Error
Open when an error is shown.
24Installation
Warning
Open when a warning is shown.
iina74e 02.fmtechnical handbook(1408)
2.3.2.3Digital In
Purge
Closed when purge is active.
Measure
Closed in case a machine is in measure mode.
Recorder Strobe
Closed in case recorder output is invalid. Only used when record output is set on
„leak rate“.
These inputs can be used to control the UL5000 with a programmable logic control
(PLC).
Description of operation mode of the Digital In.
Zero:
Change from low to high: activate zero
Change from high to low: deactivate zero
Start:
Change from low to high: activate START
Stop:
Change from low to high: activate STOP
Additionally the inlet is vented, when this input is longer high than the specified vent
delay.
Purge:
Change from low to high: activate purge
Change from high to low: deactivate purge
Clear:
Change from low to high: confirm error message
CAL:
Change from low to high:
When machine is in stand-by mode: start internal calibration. In case machine is
measurement mode: start external calibration. (Premise: external calibration test
leak has to be open and leak rate signal is stable)
Change from high to low:
External calibration: approve that external test leak is closed and leak rate signal is
stable.
High means: U > 13 V( ap pro xim at ely 7m A)
Low means: U < 7 V
The level of the logic signals must not exceed 35V.
iina74e 02.fmtechnical handbook(1408)
Notice: Signals at these inputs are only accepted if the location of control is set to
„PLC“ or „Local and PLC“. 6.6.4.2
Installation25
2.3.2.4Recorder
The recorder output Fig. 2-6/4 may be used to log the leak rate, the inlet pressure
and the forevacuum pressure.
The measured values are provided by way of an analogue signal in the range of 0 V
… 10 V. The resolution is limited to 10 mV. The instrument which is connected to the
recorder output (e. g. X(t) chart recorder) should have an input re sistance of no less
than 2.5 k. The measured values are available through pins 1 and 4. The reference
potential (GND) is available at pins 2 and 3. The contacts are numbered from top to
bottom.
TippA diagramm showing pressures and leakrate versus voltage is attached in
the appendix.
Notice: The chart recorder outputs are electrically isolated from other plugs. If, in
spite of this, hum interference is apparent it is recommended to operate the
UL5000 and the chart recorder from the same mains phase. If this is not
possible, you must make sure that the frame ground of both instruments
is kept at the same potential.
PinAssignment
1Analog 1
2GND
3GND
4Analog 2
2.3.2.5RS232
This RS232 C interface Fig. 2-6/5 is wired as data communication equipment (DCE)
and permits the connection of a personal computer (PC) for monitoring and data
logging. The connection is made through a 9 pin sub-D socket.
PinAssignment
2RXD
3TXD
5GND
7RTS
8CTS
2.3.2.6Remote Control RC1000 / Wireless Transmitter
This Remote Control interface Fig. 2-6/6 is a serial interface to control the UL5000
by the RC1000. The RC1000 Remote Control can be connected via the wireless
transmitter or via an extension cable with a RJ45 plug. Refer to the RC1000
Technical Handbook for more information.
PinAssignment
2+24V (fuse 0.8 A time lag)
30V
4RXD (intern. RS232)
5TXD (intern. RS232)
26Installation
iina74e 02.fmtechnical handbook(1408)
2.4Vacuum Connections
Warning
Warning
2.4.1 Inlet Port
The inlet port is located on the top of the UL5000. The size of the flange is DN 25 KF.
Risk of injury due to sucking connection flange (inlet port).
If the Vacuum-Mode of the UL5000 is activated, the connection flange may suck
bodily parts around the connection flange.
Keep bodily parts off the connection flange.
A test object or a test chamber has to be connected to the inlet port if the vacuum
mode is chosen (Refer to Chapter 6.3).
The inlet port is also used for the connection of the sniffer line.
2.4.2Exhaust
The exhaust Fig. 2-6/12 flange is located underneath the UL5000 at the back side.
The size of the flange is DN 16 KF.
When shipped only the exhaust filter body is preassembled. The filter cartridge is
supplied together with the leak detector and can be installed at the exhaust.
Instead of this an exhaust line can be connected to the exhaust by the exhaust
adapter.
Depending on the chamber the UL5000 is attached to and the gas inside the
chamber lethal gases can be spoiled into the air through the exhaust.
2.4.3Vent
Usually the parts under test are vented with ambient air when the test is finished. If
it is required the parts can be vented with a different gas (i. e. fresh air, dry air,
nitrogen, …) at max. 1050 mbar pressure. In this case a vent hose has to be
connected to the hose coupling Fig. 2-6/10.
2.4.4Purge
For purge modes it is recommended to use Helium-free gases at atmospheric
pressure. Ambient air can be contaminated with Helium due to
In this case a gas supply line (i. e. nitrogen, fresh air, …) should be connected to the
hose coupling Fig. 2-6/11. The pressure of these gas line must not exceed
1050 mbar.
iina74e 02.fmtechnical handbook(1408)
The connector 10 and 11 in Fig. 2-6 are quick connectors for hose diameters of 8/
6mm.
spraying o r charging.
Installation27
2.5Default parameters
The following parameters are set like shown when in the menu of the UL 5000 under
Settings Parameters saveload, „load default values“ is chosen.
Auto-scaling:On
Scalinglogarithmic
Display range:4 decades
Time axis:32 seconds
LCD invers:OFF
Background in stand by mode:OFF
Automatic calibration request:OFF
Mass:4 (helium)
Recorder Output:leak rate
Volume:2
Zero time:5 seconds
Leak rate unit:mbar l/s
Mode:Vacuum with
Trigger level 1:1×10
Trigger level 2:1×10
Leak rate external test leak (Vacuum):1×10
Leak rate external test leak (Sniffer):1×10
Vent delay:2 seconds
Automatic purge:OFF
Pressure unit:mbar
Minimum volume:0
Beep:ON
Maximum evacuation time:30 m i nu te s
Audio Alarm Type:Trigger Alarm
Maximum inlet pressure when sniffing:1 mbar
Minimum Inlet pressure when sniffing:0,1 mbar
Number of decimal place at leak rate displayed:1
HYDRO•S
-9
mbar l/s
-8
mbar l/s
-7
mbar l/s
-5
mbar l/s
28Installation
Scroll display:On
Particle protection:Off
Direct access to calibration:On
Contamination protection:Off
Switch off limit for contamination protection:1E-3 mbar l/s
Control location:Local
Alarm delay:30 seconds
Leak rate filter:I•Cal
Zero:enable
HYDRO•S automaticenabled
iina74e 02.fmtechnical handbook(1408)
3First Operation Check
Warning
Caution
The steps for an initial operation are descr ibed in this chapter. It is expla ined how to
switch on the UL5000, how to measure and how to carry out an internal calibration.
Notice: If anything unexpected happens during the initial operation or the leak
detector acts in a strange way the UL5000 can be switched off by the mains
switch at any time.
3.1Needed Equipment
The following parts will be needed:
•A blind flange 40 KF (if not preassembled at the inlet port).
•A helium test leak with a DN 40 KF adapter (optional).
3.2Description of the Initial Operation
Please proceed the following description step by step to sta rt th e initial op er at ion .
See Chapter 5 for a more detailed description.
3.2.1Startup and Measure
1 Unpack the UL5000 and inspect it for any external damage (See Chapter 1.4
Unpacking).
2 Connect the instrument to the mains power (See Chapter 2.3.1Mains Power).
3 Switch on the leak detector by using the mains switch Fig. 2-6/8.
Caution: Abrupt movements.
Abrupt movements can damage the running turbo pump.
Avoid abrupt movement and vibration of the instrument (e.g. running over cables,
door sills) during operation and up to 4 minutes after switching off since the turbo
pump can be damaged.
Don’t switch machine on when ambient temperature is below 10°C.
iina74e 03.fmtechnical handbook(1408)
After power on a welcoming picture appears on the screen of the control panel Fig.
3-1/1, then status information on the speed of the turbo pump, the foreline pressure,
the emission and the active filament are given.
First Operation Check29
The start up procedure takes about 4 minutes and the end is indicated by a beep.
1
2
The UL5000 is in Stand-by mode now.
Fig. 3-1: Top view of the UL5000
Pos. DescriptionPos. Description
1Control Panel2Inlet Port
4 Check if the inlet port Fig. 3-1/2 is blanked off. If not, please mount a blind flange with
O-Ring on the inlet port.
5 Press the START ButtonFig. 3-2/6. The inlet will be evacuated and the measured
leak rate will be displayed a moment later.
This is the measurement mode. If a test part was connected you would start spraying
Helium to identify leaks.
30First Operation Check
iina74e 03.fmtechnical handbook(1408)
Loading...
+ 78 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.