Printed in Germany
We reserve the right to alter the design or any data given in this handbook. This is
due to the fact that we are permanently developing ou r pr od u cts .
The illustrations are not binding. If there are divergences between your product and
the Technical Handbook, please contact us.
Reprint, translation and duplication need to be a pproved by INFICON GmbH. Please
contact your sales people for an updated version of the Technical Handbook.
•Please read this manual before using the T-Guard™ Leak Detection Sensor.
•Keep the manual in a way that you can use it at any time.
The INFICON T-Guard™ Leak Detection Sensor has been designed for safe and
efficient operation when used properly and in ac cord an ce with this Tec hn ica l
Handbook. It is on the user's own responsibility carefully to read and strictly to
observe all safety precautions described in this chapter and throughout this
Technical Handbook. The T-Guard™ Leak Detection Sensor must only be operated
under the conditions described in this Technical Handbook. It must be operated and
maintained by trained personnel only. Consult local, state, and national authorities
regarding specific requirements and regulations. In case you have any further
queries on safety, operation and/or maintenance, please contact our office nearest
to you.
1.2Warning and danger symbols
Indicates procedures that must be strictly observed to prevent haza rd s to per sons.
Indicates procedures that must strictly be observed to prevent damage to, or
destruction of T-Guard™ Leak Detection Sensor.
4Operating instructions
jina85e 1.fmTechnical Handbook(1112)
1.3Graphic conventions
NoticePoints to very useful pieces of information.
1 Points to an operation that you have to perform.
⇒ Points to the result of an operation you accomplished.
→ Points to the button you have to press.
•Shows a listing.
1.4Glossary
Sensitivity
Characterises the ratio of the helium signal to the measured signal.
I·Stick
Optional memory extension of the T-Gu ard™ Leak Detection Sensor.
Calibration
Measuring time
Parameter set
RS232
Purge
T-Guard
Equalisation of measured signal and leakage rate.
Time from the beginning of the measurement to the output of the results.
Group of parameters, e.g. measuring time, chamber volume, carrier gas, etc.
Interface standard for data communication between electronic devices.
Removal of excessive helium from the air.
Tracer (Gas) Guard
jina85e 1.fmTechnical Handbook(1112)
Operating instructions5
Carrier Gas
Trigger
Background
Zero
Cycle time
Large gas flow which transports a small gas flow.
Limit value of the rejection leakage rate for th e detection of le aks of the test sa mple .
The air contains 5 ppm (0.0005%) of helium by nature. For goo d measurements with
the T-Guard, the helium content in the air must be less than 10 ppm.
"Reset to Zero" is an automatic function of the T-Guard™ Leak Detection Sensor.
Time from the beginning of the measurement to the next possible measurement.
6Operating instructions
jina85e 1.fmTechnical Handbook(1112)
2Safety
Skilled personnel
Warning
Warning
2.1Intended use
The T-Guard™ Leak Detection Sensor is a helium leak dete ction sensor which does
not need any vacuum chamber. It can find leaks in test samples and quantify them
when they are filled with helium under pressure.
Using the T-Guard™ Leak Detection Sensor, the chamber with the test sa mple does
not have to be evacuated.
Measured at atmospheric pressure, the minimum detectable leak rate is 1×10
l/s. This depends on the chamber volume and the measuring time, the rejection l eak
rate will mostly be between 1·10
Do not let water penetrate into the body or hoses of the T-Guard™ Leak Detection
Sensor. Do not forget to use the external filters.
Only use accessories and spare parts from INFICON.
2.2Requirements to be fulfilled by the user
-4
and 1·10-2 mbar l/s.
-6
mbar
The T-Guard™ Leak Detection Sensor must only be started and operated by
properly trained staff.
Get used to the functioning of the device. Yo u may only use the device after reading
and understanding the manual.
2.3Restrictions of use
Disregarding the following precautions could result in serious injury:
For all contacts of the I/O port, a maximum voltage of 60 V DC or 25 V AC must not
be exceeded or reached based on the protective cond uctor or the earthing devices.
Please refer to the information given in the following chapters.
For maintenance work, disconnect the T-Guard™ Leak Detection Se nsor from the
power supply.
jina85e 2.fmTechnical Handbook(1112)
Safety7
Warning
Before replacing the fuses, the T-Guard™ Leak Detection Sensor must be
Caution
Caution
Caution
disconnected from the power supply.
Disregarding the following precautions could result in damage to the equipment.
The T-Guard™ Leak Detection Sensor must not be operated while standing in
water or when exposed to drip water. The same applies to all other kinds of liquid.
The T-Guard™ Leak Detection Sensor should only be used inside rooms.
Avoid the T-Guard getting in contact with bases, acids, and solvents, as well as
being exposed to extreme climatic conditions.
Do not suck up liquids into the T-Guard™.
8Safety
jina85e 2.fmTechnical Handbook(1112)
3Description
This chapter describes the exterior design of the T-Guar d™ Leak Detection Sensor,
its accessories and how to use it.
3.1Housing
The housing of the unit consists of an oval cylinder. On top, there are several
electrical connections and the ventilation opening. On the rear side, ther e are th ree
hose connectors.
Fig. 4-1 Front view of the T-Guard™ Leak Detection Sensor
Fig. 4-2 Back view of the T-Guard™ Leak Detection Sensor
jina85e 3.fmTechnical Handbook(1112)
Description9
3.2Interfaces
The T-Guard™ Leak Detection Sensor has 5 interfaces for control and information
exchange.
Fig. 4-3
Power Connector
Only two of the four pins are used for the power supply. The plus pole is marked with
„1+“, the minus pole is marked with „2-“.
Fuse
The fuse reacts to reverse polarity and a current exceeding 7 A.
Item DescriptionItemDescription
1Power supply connector5PLC-interface
2Fuse 6RS232 PC-interface
3Interface for the control unit7Analog Recorder Output
4I•Stick -connector8Status LEDs
jina85e 3.fmTechnical Handbook(1112)
10Description
Interface for the control unit
Here, the INFICON control unit (Cat. No. 551-100 or 551-101) is connected via its
connecting cable (Cat. No. 551-102 or 551-103) to control the T-Guard™ Leak
Detection Sensor manually.
NoticeThe connection cable must only be plugged in or unplugged when the
supply voltage of the T Guard™ is turned off or disconnected.
I•Stick -connector
You can store 25 parameter sets on the INFICON-I-Stick and also download them
from there.
PLC-interface
The T-Guard™ Leak Detection Sensor has 6 configurable digital PLC-Inputs and
8 configurable digital PLC-outputs.
(Refer to 4.6.1Selection of functions assigned to the PLC inputs. and 4.6.3Selection
of functions assigned to the PLC outputs..)
RS232 PC-interface
Connect any RS232 capable device here with an uncrossed RS232 cable in order to
control the T-Guard™ Leak Detection Sensor or to read out information. The latter
takes less than 100 ms.
Analog Recorder Output
The T-Guard™ Leak Detection Sensor has two separately configurable analog
recorder outputs (0 - 10 V).
They have a resolution of 16 bit and a hardware limited refresh rate of about 2 Hz.
For a higher refresh rate, please use the RS232 interface.
For detailed Information on how to use the analog output refer to chapter 5.6.
jina85e 3.fmTechnical Handbook(1112)
Description11
Status LEDs
Certain operating conditions can be indicated by 5 status LEDs:
The first LED:lights up during operation
The four remaining LEDsindicate the status of the configurable digital outputs 1-4.
LED No. 2 is red and in standard setting, it will light up to
indicate an error or a warning. The desired output function
can be set (refer to4.6.3).
Fig. 4-4 Status LEDs
12Description
jina85e 3.fmTechnical Handbook(1112)
3.3Vacuum diagram
On top of the vacuum diagram, there are the measurement inlet "IN" and the
reference inlet "Ref". Both inlets are equipped with an external filter to protect TGuard™ Leak Detection Sensor from sucking in dust or fluids. Three valves are
integrated in the T Guard™ Leak Detection Sensor (see Fig. 4-5).
Valve V1 opens and closes the measurement inlet. Valve V3 opens and closes the
reference inlet. Valve V2 opens and closes the connection between V1 and V3.
These three valves are connected to two flow-restricted capillaries which lead to the
Wise Technology™ Sensor.
The upper inlet of the Wise Technology™ Sensor is the fine inlet, the lower one is
the gross inlet, which is less sensitive to helium than the fine inlet.
The Wise Technology™ Sensor detects the heliu m in the gas stream, which passes
by. The total pressure, which is applied to the Wise Technology™ Sensor is detected
by the Pressure Gauge p1.
The fore pump, which provides the gas stream inside T-Guard™ Leak Detection
Sensor is also throttled to produce constant pressure and flow for the Wise Technology™ Sensor.
Fig. 4-5 Vacuum diagram of the T-Guard™ Leak Detection Sensor
Item DescriptionItem Description
1Inlet “IN”8Pressure gauge p1
2Reference “Ref”9Pump
3Filter 210Pump connector “OUT”
4Valve V
5Valve V
6Throttle 213Valve V
7Wise Technology™ Sensor14Filter 1
jina85e 3.fmTechnical Handbook(1112)
3
2
11Throttle 3
12Throttle 1
1
Description13
When the T-Guard™ Leak Detection Sensor is not measuring at the moment or is
READY , V2 and V3 are open, V1 is closed (refer to Fig. 4-6). Only gas from the
reference inlet gets to the sensor.
Fig. 4-6 Valve position in READY mode.
Item DescriptionItemDescription
1Inlet “IN”8Pressure gauge p1
2Reference “Ref”9Pump
3Filter 210Pump connector “OUT”
4Valve V
5Valve V
3
2
6Throttle 213Valve V
7Wise Technology™ Sensor14Filter 1
11Throttle 3
12Throttle 1
1
14Description
jina85e 3.fmTechnical Handbook(1112)
In Standby mode, all valves except V2 are closed and the pressure before the Wise
Technology™ Sensor is equal to the base pressure of the fore pump.
(refer to Fig. 4-7)
Fig. 4-7 Valve configuration in S
TANDBY mode.
Item DescriptionItem Description
1Inlet “IN”8Pressure gauge p1
2Reference “Ref”9Pump
3Filter 210Pump connector “OUT”
4Valve V
5Valve V
3
2
6Throttle 213Valve V
11Throttle 3
12Throttle 1
1
7Wise Technology™ Sensor14Filter 1
jina85e 3.fmTechnical Handbook(1112)
Description15
In the rare event that the sensor is contaminated with helium, only the valve V3 to the
reference inlet "Ref" is open, to purge the sensor. In this case, highly pure nitrogen
or external air has to be led to the reference inlet to decontaminate the sensor as fast
as possible.
Fig. 4-8 Valve configuration during decontamination
Item DescriptionItemDescription
1Inlet “IN”8Pressure gauge p1
2Reference “Ref”9Pump
3Filter 210Pump connector “OUT”
4Valve V
5Valve V
6Throttle 213Valve V
7Wise Technology™ Sensor14Filter 1
3
2
11Throttle 3
12Throttle 1
1
jina85e 3.fmTechnical Handbook(1112)
16Description
In FINE measurement mode, V1 and V2 are open, V3 is closed (see Fig. 4-9). The
sensor receives only gas from the measurement inlet and shows the highest sensitivity to helium from the measurement line.
Fig. 4-9 Valve configuration in FINE mode
Item DescriptionItem Description
1Inlet “IN”8Pressure gauge p1
2Reference “Ref”9Pump
3Filter 210Pump connector “OUT”
4Valve V
5Valve V
3
2
6Throttle 213Valve V
7Wise Technology™ Sensor14Filter 1
11Throttle 3
12Throttle 1
1
jina85e 3.fmTechnical Handbook(1112)
Description17
In GROSS measurement mode, V1 and V3 are open, V2 is closed (see Fig. 4-10).
This way, the Wise Technology™ Sensor only receives little helium from the measurement line and normally even less helium from the reference line.
Fig. 4-10 Valve configuration in GROSS mode
Item DescriptionItemDescription
1Inlet “IN”8Pressure gauge p1
2Reference “Ref”9Pump
3Filter 210Pump connector “OUT”
4Valve V
5Valve V
6Throttle 213Valve V
7Wise Technology™ Sensor14Filter 1
3
2
11Throttle 3
12Throttle 1
1
Other valve configurations than described above are not used by the software and
cannot be set by the user.
18Description
jina85e 3.fmTechnical Handbook(1112)
3.4Wise Technology™ Sensor
The helium detector (Wise Technology™ Sensor) consists of a closed glass
container with a measurement device for the precise determination of the pressure
inside the glass housing and a membrane chip with a large number of small quartz
windows.
The membrane is permeable only for helium; the membrane avoids all other co mponents of air from entering the glass housing.
The quartz membrane is heated so that the permeation for helium is sufficie ntly high
and fast.
Inside of the glass housing, the total pressure is measur ed precisely. As only helium
can enter the glass housing, the total pressu re is equa l to the pa rtial pressure of
helium.
The determined total pressure inside the housing is p roportional to the helium partial
pressure outside the sensor.
jina85e 3.fmTechnical Handbook(1112)
Description19
3.5Accessories
1
9
12
11
2
3
4
5
14
6
7
8
10
13
In addition to the main unit, two different types of the control panel and the corresponding connection cables are available.
3.5.1Control unit for desktop operation
Install the control panel on a flat workbench to avoid slipping. It is used to control all
functions and read all values of the T-Guard™ Leak Detection Sensor.
The control panel contains an LC Display, START, S
and also 8 soft keys to navigate in the menu on the display and to enter values.
NoticeAll 3 LEDs in the buttons are inactive with T-Guard™ Leak Detec tio n
ERO button with LED
Sensor.
jina85e 3.fmTechnical Handbook(1112)
Press the START button to start a measurement using the selected method. The
measurement will then finish automatically.
STOP button
Press the STOP button to cancel the measurement immediately.
MENU button
In order to change the settings or to read information about T-Gua rd™ Leak Detection Sensor, open the menu by pressing the MENU button or the Soft Key (No.4) with
the MENU symbol ().
You will go back to the position in the menu you left before.
By pressing the M
ENU button again you will get back to the m easurement menu. By
pressing the MENU button for two seconds you will get into the main menu.
Fig. 4-12 Main Menu appearance
Soft Keys
The functions of the 8 soft keys depend on the currently opened menu.
Soft key No. 1Description
1 and 8very often have the functions B
4often has the function to completely leave the menu ( )
ACK / ESC and OK, in this order,
MENU buttonEnter/Leave the menu
Entering values and units
The numerical values are entered in the the MENU page as follows:
Soft key No. 1Description
SC-button)
1 (E
2( ↑ )increments the numerical value.
3( ↓ )decrements the numerical value.
6 and 7( ↑ and ↓ )changes the unit
8 (OK)saves the entered value.
jina85e 3.fmTechnical Handbook(1112)
leave this menu page and return to the previous menu page. The
value will not be changed.
Description21
Fig. 4-13 You can change the value and the unit on this menu page
Entering the Menu-PIN
1 Press the ESC-button (soft key No.1) to leave this menu page and to return to the
previous menu page. The Menu-PIN will then neither be changed nor be entered.
2 Pressing a number button opens a submenu where you can select one of these two
numbers. After that, the cursor moves to the next digit to the right.
3 You can move the cursor to the left or to the right using the soft keys No. 4 and 8.
( ← and → ).
4 OK appears for soft key No. 8 when the last digit has been reached.
5 Pressing OK causes the PIN to be stored.
Status Line in the Display
22Description
Fig. 4-14 Entering the Menu-PIN
The lowest line of the display is the status line. It shows general information, for
instance, the actual measurement mode.
jina85e 3.fmTechnical Handbook(1112)
3.5.219" control unit
You can insert this control unit in a 19’’ rack or in a panel cut-out as shown in Section
8.5.
This control panel is used in the same way as it is for the desktop control unit, only
the START, STOP and ZERO buttons are at other positions.
Fig. 4-15 Control unit for the installation in a 19"-rack
On its front side, the 19"-control unit refers to protection class IP40.
3.5.3Connection Cable for Control Unit, 5 m
This cable connects the control unit and the T-Guard™ Leak Detection Sensor.
Fig. 4-16 Connection cable for control unit, cat. No. 551-102
jina85e 3.fmTechnical Handbook(1112)
Description23
3.5.4Connector set
The connector set includes all Phoenix connectors for connecting the electrica l interfaces.
Fig. 4-17 Connector set, cat. No. 551-110
3.5.5Fore Pump
24Description
Fig. 4-18 Fore pump, cat No. 200 002 929
This fore pump has two stages, is brushless and runs with a supply voltage of
24 V DC.
jina85e 3.fmTechnical Handbook(1112)
Loading...
+ 59 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.